Professional Documents
Culture Documents
91CwNxQvmjL (3) Book
91CwNxQvmjL (3) Book
91CwNxQvmjL (3) Book
01)
2009. 03
Forced Convection
Drying Oven
Operation Manual
Model : J-300S
Table of Contents
J-300S-M-E-09-001(REV.01) Page 2
Operation Manual
z Please read through and understand this Operation Manual before operating the product. After
reading, always keep the manual nearby so that you may refer to it as needed. When moving the
product to another location, be sure to bring the manual as well.
z If you find any incorrectly arranged or missing pages in this manual, they will be replaced. If the
manual it gets lost or soiled, a new Operation Manual can be purchased. In either case, please
contact your JISICO agent, and provide the "Model No." given on cover.
z This manual has been prepared with the utmost care; however, if you have any questions, or note
any errors or omissions, please contact your JISICO distributor or agent.
Reproduction and reprinting of this operation manual, whole or partially, without our permission is prohibited.
Both unit specifications and manual contents are subject to change without notice.
J-300S-M-E-09-001(REV.01) Page 3
Operation Manual
z If the operator does not read the language used in this manual,
translate the manual into appropriate language.
z Keep this manual near the tester for easy access by the operator.
J-300S-M-E-09-001(REV.01) Page 4
Operation Manual
SAFETY SYMBOLS
This operation manual and this product use the following safety symbols.
Please keep in mind the meaning of each of the symbols to ensure safe use of the product and to
prevent any damages, accidents, or problems resulting from negligence of these safety
symbols. (As using symbols may depend on the product, not all of the symbols may be used.)
SAFETY SYMBOLS
z Indicates the possibility of personal injury or death.
Never fail to follow the operating procedure.
J-300S-M-E-09-001(REV.01) Page 5
Operation Manual
Safety Precautions
The following safety precautions must be observed and followed to avoid fire hazard, electrical
shock, accidents, and other failures. Keep them in mind and make sure that all the precautions
are observed and followed properly. JISICO CO., LTD. assumes no liability nor any responsibility
against any damages, accidents, or problems resulting from negligence of the precautions.
Users
z This product must be used only by qualified personnel who understand the
contents of this operation manual.
z If it is handled by disqualified personnel, personal injury may result.
Be sure to handle it under supervision of qualified personnel.
Purpose of use
z If the product is to be used for purposes not described in this manual,
contact your JISICO agent in advance.
Input Power
z Use the product with the specified input power voltage.
z For applying power, use the AC power cable provided. The shape of the
plug differs according to the power voltage and areas. Use the cable
which is suitable for the line voltage used.
Fuse
z With products with a fuse holder on the exterior surface, the fuse can be
replaced with a new one.
z When replacing a fuse, use the one which has appropriate shape, ratings,
and specifications.
Cover
z There are parts inside the product which may cause physical hazards. Do
not remove the external cover. If the cover must be removed, contact your
JISICO agent in advance.
J-300S-M-E-09-001(REV.01) Page 6
Operation Manual
Installation
z When installing products be sure to observe Installation Precaution
described in this manual.
z To avoid electrical shock, connect the protective ground terminal to
electrical ground (safety ground).
z When applying power to the products from a switchboard, be sure work
is performed by a qualified and licensed electrician or is conducted
under the direction of such a person.
z Be sure to use the AC power cable provided.
Relocation
z Turn off the power switch and then disconnect all cables when
relocating the product.
z Use two or more persons when relocating the product which weights
more than 20 kg. The weight of the products can be found on the rear
panel of the product and/or in this operation manual.
z Use extra precautions such as using more people when relocating into
or out of present locations including inclines or steps. Also handle
carefully when relocating tall products as they can fall over easily.
z Be sure the operation manual be included when the product is relocate
Maintenance
z To avoid electrical shock, be absolutely sure to unplug the AC power
cable or stop applying power before performing maintenance or
checking.
z Do not remove the cover when performing maintenance or checking. If
the cover must be removed, contact your JISICO agent in advance.
z To maintain performance and safe operation of the product, it is
recommended that periodic maintenance, checking, cleaning, and
calibration be performed.
SERVICE
z Internal service is to be done by JISICO service engineers.
z If the product must be adjusted or repaired, contact your JISICO agent.
J-300S-M-E-09-001(REV.01) Page 7
Operation Manual
Forced convection drying oven is used when you need quicker dry for glass instrument or when you
experiment degeneration and hardening of reagent.
This device is also used to experiment preheating prior to heating test, drying for architecture materials
and electronic parts and heat-resistant test.
This device provides excellent experiment environment with surveying self life of food in hard
condition, eliminating moisture, hardening or softening experiment of food and chemicals by heating,
dry sterilization of glass device for microorganism and eliminating moisture contained in food.
2. Feature of Product
z One touch type door catcher allows you to open/close the door without any vibration so as not to
affect the contents in the chamber.
z Yon can select two kind timer modes ; normal mode starting the machine as soon as you switch
on, and the other timer mode starting to operate at 1℃ from the pre-set temperature
z This product is multipurpose forced convection oven used in various fields such as life engineering,
pharmacy, medicine science, biology and the like, and is realized to be considered greatly
convenient use and safety.
z Adoption for thermophilic heat-resistant material and use for thermophilic silicone door packing at
the outside of chamber and door provides supper effect of resistant- heat and prevents loss of
heating.
z Safety circuit is adopted to prevent the device from over current and overheating of heater.
J-300S-M-E-09-001(REV.01) Page 8
Operation Manual
3. Appearance
Ventilation hole
Packing
Rack support
Shelf
Observation
window
Door catch
Door
Rubber foot
Control panel
ON position
Emergency
Break Button
OFF position
E.L.B
Electrical Leakage Breaker
( = Main Power Switch )
J-300S-M-E-09-001(REV.01) Page 9
Operation Manual
4. Installation
z This equipment is designed and manufactured to operate properly only in use of rated voltage.
Refer to specification section of this manual for rated voltage before installation.
z Use AC power cable that comes with (or installed to) the product.
z Do not touch the Power Cord wet handed.
z Avoid locations where the equipment is exposed to direct sunshine.
z Do not place the equipment in a corrosive atmosphere.
z Do not use the equipment in a flammable atmosphere.
z Do not use the equipment where ventilation is poor.
z Do not locate the equipment in a dusty location.
z Do not use in an unstable place that are subject to vibrations.
z Avoid locations of high humidity.
z Avoid locations where the equipment is exposed to high temperature.
z Recommended temperature range : 20℃ to 40 ℃
z Use the equipment indoors only.
z Avoid locations of altitude higher than 2000m.
z Secure adequate space around the power plug.
z Use with a minimum distance all around of 20cm from walls of other items
z Locate in a flat place.
z Connect the power cord to an outlet with earthling point.
z Use POWER CORD with specification higher than 125V, 30A.
z In case of connecting to an outlet without earthling point, connect lead wire to the earth after
connecting grounding adapter.
J-300S-M-E-09-001(REV.01) Page 10
Operation Manual
① ②
⑥
③
④ ⑤
[PID Controller]
: MODE Select Key : Used for setting up temp. time correction and other related parameters.
: Auto Tuning and Timer Key : AT key function is starts if you press AT key for more than 4
seconds.
J-300S-M-E-09-001(REV.01) Page 11
Operation Manual
- Power ON / OFF Key : Turns Power ON or OFF. (It does NOT operate in setting mode.)
- RUN/STOP Key : RUN/STOP Heater. (It does NOT operate in setting mode.)
- MODE Key :
․ Press MODE Key once in Normal Display Mode and changes to Temperature, Time Setting
Mode.
․ Press MODE Key for over 6 seconds and changes to individual Parameter Setting Mode.
․ Press Mode key once again and the parameter changes to another parameter,
and pressing the MODE Key for more than 2 seconds ends Parameter Setting Mode, and
returns to Normal status.
- Shift Key :
․ Moves the decimal point of setting data in each MODE.
․ Press Shift Key for over 6 seconds and changes to Factory Setting Mode.
- Increase setting value Key : Increase set-value of each decimal point
- AT (Auto tuning key) :
․ Displays present time, then temperature in turn when At key is pressed in Normal Display
Mode.
․ AT key function is starts if you press AT key for more than 4 seconds in Normal Display
Mode.
․ By pressing AT key for more than 4 seconds during Auto Tuning Process ends Auto Tuning
Process instantly.
․ Displays present temperature when AT key is pressed once during Auto Tuning Process.
J-300S-M-E-09-001(REV.01) Page 12
Operation Manual
CPU with software which is capable of digital PID-Auto tuning is equipped and contained regulation
device of supper safety class such as the regulation function for temp. correction for Pt100, heating and
others.
Normal Status
J-300S-M-E-09-001(REV.01) Page 13
Operation Manual
0 No Alarm
AL
(LOW ALARM) PV
* Can be
High Low Alarm ON 0FF ON
1 AL AH used as
(Absolute value)
* Adjust with ◀ Key, ▲ Key refrigeration
High Alarm PV
0FF ON
AH 2
(Absolute value) AH
ON 0FF
3 used as
MODE Key (Absolute value) AL
refrigeration
* Can be
Art High Low Alarm ON
PV
0FF ON
4 used as
(ALARM TYPE) (Relative value) SV-AL SV+AH
refrigeration
High Alarm PV
OFF ON
MODE Key 5
(Relative value)
* Can be
H.SAF Low Alarm ON OFF
PV
6 used as
(HI SAFETY ALARM) (Relative value)
refrigeration
MODE Key
4) H.ASF (HI SAFETY ALARM) : 0~99 (PT)
If present temperature is higher than "set value +
CooL
H.ASF value", then timer stops and changes to
(COOLING TEMPERATURE) STOP MODE.
MODE Key 5) COOL (COOLER SET TEMPERATURE)
: set within temperature range
HyS - displays only when cooler is in activation.
(HYSTERESIS) - If Set temperature > COOL set value,
COOL RELAY ==> OFF
MODE Key
- If Set temperature > COOL set value,
J-300S-M-E-09-001(REV.01) Page 14
Operation Manual
9) I (INTEGRAL FACTOR)
I
: 0~999.9
(INTEGRAL FACTOR)
- Set automatically after Auto Tuning process.
MODE Key
10) D (DIFFERENTIAL FACTOR)
: 0~999.9
D
- Set automatically after Auto Tuning process.
(DIFFERENTIAL FACTOR)
11) AR (ANTI RESET FACTOR)
MODE Key : 0~999.9
- Set automatically after Auto Tuning process.
Ar
12) DRAN (TEMPERATURE FIXED DISPLAY RANGE)
(ANTI RESET FACTOR)
: Displays present temperature in the range as
MODE Key
SET TEMPERATURE.
Ex) If set temperature is 50℃ and DRAN is 0.3℃
drAn
it displays present temperature within range of
(TEMP. FIXED RANGE)
49.7~50.3℃ as 50℃.
MODE Key
13) ACTP (TIMER STARTING TEMPERATURE VALUE)
LoC
(KEY LOCK)
MODE Key
z DO NOT CHANGE the SET VALUE, which is FACTORY SET DEFAULT VALUE !
z Changing the SET VALUE may result in MAL-FUNCTIONING !
J-300S-M-E-09-001(REV.01) Page 15
Operation Manual
Normal Status
1) SEN (SENSOR TYPE)
* Press SHIFT Key for
* PT TYPE : Pt display ==> DPt (Default)
over 6 seconds
* Beep sound JPt display ==> JPt
2) PO_m (POWER FAILURE MANAGEMENT)
SEn
* ON ==> START DRAN, PO_m ON (Default)
(SENSOR TYPE)
OFF ==> PO_m OFF
MODE Key
* Adjust with ▲ Key 3) A_TY (RELAY CONTACT TYPE)
* Displays when "ALARM RELAY" is selected.
PO_m
( Refer to (5) A.C.R Mode )
(POWER FAILURE
MANAGEMENT) 4) SR_A (START RELAY OFF in HIGH ALARM)
J-300S-M-E-09-001(REV.01) Page 16
Operation Manual
ON
z When temperature difference occurs during operation, press AT key for 4 seconds, and
automatic tuning process will begin, and AT lamp(Auto tuning lamp) will start to blink until
automatic tuning process is completed.
z When automatic tuning is in process, NEVER change setting values nor OPEN/CLOSE the door
until automatic tuning process is completed.
z Press AT key for 4 seconds during Auto Tuning Process , and Auto Tuning Process will
be stopped instantly.
Press key and key to set required time for timer function.
4) When setting on all units is finished, pressing MODE key at full length for 2 seconds,
and all setup values will be reset as set values and return to normal mode.
J-300S-M-E-09-001(REV.01) Page 17
Operation Manual
z Check rated voltage of the equipment as well as check the specification. (AC110V, 50/60Hz).
-In case of using 110V, use a boosting transformer with enough capacity.
-Be sure to use AVR (Automatic Voltage Regulator) in the place with unstable voltage
z Internal wall or board of the chamber might a little soot and colored by the fume
generated inside the chamber.
J-300S-M-E-09-001(REV.01) Page 18
Operation Manual
z Ensure that the apparatus has cooled down to a safe temperature to touch.
- Disconnect the power plug from power outlet.
- Clean the inside part of the equipment with sponge or soft clothes getting wet by neutral cleaner.
- Clean the inside part of the equipment again with sponge or soft clothes getting wet by water.
- Clean with dry clothes.
- Do not use organic solvent.
- Do not use flammable or volatile chemicals such as benzene. alcohol to clean.
z Do not use flammable or volatile chemicals such as benzene. alcohol to clean the equipment.
z This equipment is not designed nor manufactured Explosion-proof, so never place highly flammable
material or explosive material inside the equipment which may cause an explosion or a fire.
J-300S-M-E-09-001(REV.01) Page 19
Operation Manual
7.Trouble shooting.
8. After-sales service.
Faults which is responsible for manufacture in normal condition can be repaired with free for 1year from
purchase date (only, pressure vessel is zero year), and it is desirous to check following items when
requesting A/S.
. Part and condition generated fault
( It is necessary to explain in detail within limit of possibility )
. Model name
. Serial Number
. Purchase date (year, month, date)
J-300S-M-E-09-001(REV.01) Page 20
Operation Manual
9. Specification.
Model J-300S
Capacity ℓ 91
Regulator TRIAC
Shelves 3 ea
Net weight kg 51
Certificate CE
J-300S-M-E-09-001(REV.01) Page 21
Operation Manual
Circulation
230V, 1phase, 24W 1.00
fan motor
J-300S-M-E-09-001(REV.01) Page 22
Operation Manual
J-300S-M-E-09-001(REV.01) Page 23
Operation Manual
J-300S-M-E-09-001(REV.01) Page 24
Operation Manual
9.5 CE certificate
J-300S-M-E-09-001(REV.01) Page 25
Operation Manual
J-300S-M-E-09-001(REV.01) Page 26
Operation Manual
Warranty Certification.
Thank you for purchasing a JISICO product. Please take a moment to fill in the following and return it
to JISICO CO., LTD. for product registration.
z JISICO products are warranted against defect in material and workmanship for a period of one(1) year
after delivery to the first purchaser for use.
z This warranty does not apply to the damage resulting from misuse, alteration, transportation, abuse,
accidents, tampering, servicing attempted or performed by unauthorized service agency.
z Before use, read the manual thoroughly. JISICO can not be held responsible for the malfunctioning
resulting from the use of the product other than specified in operation manual.
z Call or fax Customer Service for a Return Material Authorization (RMA) number before returning the
product or any specific material.
z Put the RMA number on the shipping label, and a written detailed description of the problem.
Product Name
Model No. Serial No.
Delivery Date / / ( dd/mm/yy) Warranty One(1) year
JISICO Company
Agent/ Distributor Phone No.
Company Contact
Custom Address
Phone No. Fax. No.
JISICO only warrants the safety and proper operation of product under the condition that :
- The product is properly and in accordance with the instructions stated in the operation manual.
- No modifications are made.
- Only original spare parts are used. The use of other parts carries unknown risks and will void unit's
warranty.
- The back of the unit has not been opened.
Signed by
J-300S-M-E-09-001(REV.01) Page 27