91CwNxQvmjL (3) Book

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 27

Manual No.: J-300S-M-E-09-001(REV.

01)
2009. 03

Forced Convection
Drying Oven
Operation Manual

Model : J-300S

Please read this Operation Manual carefully before use.


Operation Manual

Table of Contents

Use of Operation Manual ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 3


To SUPERVISOR in charge of this equipment ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 4
SAFETY SYMBOLS ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 5
Safety Precautions ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 6

1. Intended use of the Apparatus ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 8


2. Feature of product ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 8
3. Appearance ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 9
4. Installation ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 10
5. Name & functions of the controller ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 11
6. Maintenance and service ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 19
7. Trouble shooting ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 20
8. After-sales service ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 20
9. Specification
9.1 Technical Specification ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 21
9.2 Component list ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 22
9.3 Assembly Drawing ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 23
9.4 Circuit Diagram ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 23
9.5 CE certificate ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 25
9.6 ISO certificate ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 26
Warranty certification ‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥‥ 27

J-300S-M-E-09-001(REV.01) Page 2
Operation Manual

Use of Operation Manual

z Please read through and understand this Operation Manual before operating the product. After
reading, always keep the manual nearby so that you may refer to it as needed. When moving the
product to another location, be sure to bring the manual as well.

z If you find any incorrectly arranged or missing pages in this manual, they will be replaced. If the
manual it gets lost or soiled, a new Operation Manual can be purchased. In either case, please
contact your JISICO agent, and provide the "Model No." given on cover.

z This manual has been prepared with the utmost care; however, if you have any questions, or note
any errors or omissions, please contact your JISICO distributor or agent.

Reproduction and reprinting of this operation manual, whole or partially, without our permission is prohibited.
Both unit specifications and manual contents are subject to change without notice.

Copyright ⓒ 2004-2009 JISICO CO., LTD.


Printed in Republic of Korea.

J-300S-M-E-09-001(REV.01) Page 3
Operation Manual

To SUPERVISOR in charge of this equipment.

z If the operator does not read the language used in this manual,
translate the manual into appropriate language.

z Help the operator in understanding this manual before use in


operation.

z Keep this manual near the tester for easy access by the operator.

J-300S-M-E-09-001(REV.01) Page 4
Operation Manual

SAFETY SYMBOLS

This operation manual and this product use the following safety symbols.
Please keep in mind the meaning of each of the symbols to ensure safe use of the product and to
prevent any damages, accidents, or problems resulting from negligence of these safety
symbols. (As using symbols may depend on the product, not all of the symbols may be used.)

SAFETY SYMBOLS
z Indicates the possibility of personal injury or death.
Never fail to follow the operating procedure.

z Do not proceed beyond a WARNING sign until the noted


conditions are fully met to satisfactory and understood.

z Indicates the existence of possibility of damage to the product


or to the operator in operation.

z Never fail to follow the operating procedure.

z Indicates the existence of possibility of damage to the product


or to the operator in operation when cautions are not observed
and followed properly.

z Indicates additional information such as general warning,


caution, risk of danger, and operating procedure.

z When this mark is indicated on the product, refer the relevant


section of the Operation Manual.

J-300S-M-E-09-001(REV.01) Page 5
Operation Manual

Safety Precautions

The following safety precautions must be observed and followed to avoid fire hazard, electrical
shock, accidents, and other failures. Keep them in mind and make sure that all the precautions
are observed and followed properly. JISICO CO., LTD. assumes no liability nor any responsibility
against any damages, accidents, or problems resulting from negligence of the precautions.

Users
z This product must be used only by qualified personnel who understand the
contents of this operation manual.
z If it is handled by disqualified personnel, personal injury may result.
Be sure to handle it under supervision of qualified personnel.

Purpose of use
z If the product is to be used for purposes not described in this manual,
contact your JISICO agent in advance.

Input Power
z Use the product with the specified input power voltage.
z For applying power, use the AC power cable provided. The shape of the
plug differs according to the power voltage and areas. Use the cable
which is suitable for the line voltage used.

Fuse
z With products with a fuse holder on the exterior surface, the fuse can be
replaced with a new one.
z When replacing a fuse, use the one which has appropriate shape, ratings,
and specifications.

Cover
z There are parts inside the product which may cause physical hazards. Do
not remove the external cover. If the cover must be removed, contact your
JISICO agent in advance.

J-300S-M-E-09-001(REV.01) Page 6
Operation Manual

Installation
z When installing products be sure to observe Installation Precaution
described in this manual.
z To avoid electrical shock, connect the protective ground terminal to
electrical ground (safety ground).
z When applying power to the products from a switchboard, be sure work
is performed by a qualified and licensed electrician or is conducted
under the direction of such a person.
z Be sure to use the AC power cable provided.

Relocation
z Turn off the power switch and then disconnect all cables when
relocating the product.
z Use two or more persons when relocating the product which weights
more than 20 kg. The weight of the products can be found on the rear
panel of the product and/or in this operation manual.
z Use extra precautions such as using more people when relocating into
or out of present locations including inclines or steps. Also handle
carefully when relocating tall products as they can fall over easily.
z Be sure the operation manual be included when the product is relocate

Maintenance
z To avoid electrical shock, be absolutely sure to unplug the AC power
cable or stop applying power before performing maintenance or
checking.
z Do not remove the cover when performing maintenance or checking. If
the cover must be removed, contact your JISICO agent in advance.
z To maintain performance and safe operation of the product, it is
recommended that periodic maintenance, checking, cleaning, and
calibration be performed.

SERVICE
z Internal service is to be done by JISICO service engineers.
z If the product must be adjusted or repaired, contact your JISICO agent.

J-300S-M-E-09-001(REV.01) Page 7
Operation Manual

1. Intended Use of the Equipment

Forced convection drying oven is used when you need quicker dry for glass instrument or when you
experiment degeneration and hardening of reagent.

This device is also used to experiment preheating prior to heating test, drying for architecture materials
and electronic parts and heat-resistant test.

This device provides excellent experiment environment with surveying self life of food in hard
condition, eliminating moisture, hardening or softening experiment of food and chemicals by heating,
dry sterilization of glass device for microorganism and eliminating moisture contained in food.

2. Feature of Product

z Temperature is regulated by PID controller microprocessor in membrane touch type.

z Overheating is prevented by automatic shut-off circuit.

z One touch type door catcher allows you to open/close the door without any vibration so as not to
affect the contents in the chamber.

z Yon can select two kind timer modes ; normal mode starting the machine as soon as you switch
on, and the other timer mode starting to operate at 1℃ from the pre-set temperature

z This product is multipurpose forced convection oven used in various fields such as life engineering,
pharmacy, medicine science, biology and the like, and is realized to be considered greatly
convenient use and safety.

z Adoption for thermophilic heat-resistant material and use for thermophilic silicone door packing at
the outside of chamber and door provides supper effect of resistant- heat and prevents loss of
heating.

z The structure of chamber is optimized for uniform temperature distribution in chamber.

z Safety circuit is adopted to prevent the device from over current and overheating of heater.

J-300S-M-E-09-001(REV.01) Page 8
Operation Manual

3. Appearance

Circulation fan &


Temperature sensor (installed inside)

Ventilation hole

Packing

Rack support

Shelf

Observation
window

Door catch

Door

Heater (installed inside)

Rubber foot

Control panel

Main power switch


= E.L.B. : Electrical Leakage Breaker

[ Fig.1 : Front View ]

ON position

Emergency
Break Button
OFF position

E.L.B
Electrical Leakage Breaker
( = Main Power Switch )

[ Fig.2 : Main Power Switch ]

J-300S-M-E-09-001(REV.01) Page 9
Operation Manual

4. Installation

4.1 Precautions for Installation

z This equipment is designed and manufactured to operate properly only in use of rated voltage.
Refer to specification section of this manual for rated voltage before installation.
z Use AC power cable that comes with (or installed to) the product.
z Do not touch the Power Cord wet handed.
z Avoid locations where the equipment is exposed to direct sunshine.
z Do not place the equipment in a corrosive atmosphere.
z Do not use the equipment in a flammable atmosphere.
z Do not use the equipment where ventilation is poor.
z Do not locate the equipment in a dusty location.
z Do not use in an unstable place that are subject to vibrations.
z Avoid locations of high humidity.
z Avoid locations where the equipment is exposed to high temperature.
z Recommended temperature range : 20℃ to 40 ℃
z Use the equipment indoors only.
z Avoid locations of altitude higher than 2000m.
z Secure adequate space around the power plug.

4.2 How to install

z Use with a minimum distance all around of 20cm from walls of other items
z Locate in a flat place.
z Connect the power cord to an outlet with earthling point.
z Use POWER CORD with specification higher than 125V, 30A.
z In case of connecting to an outlet without earthling point, connect lead wire to the earth after
connecting grounding adapter.

4.3 How to ground Earth wire

z Earth wire must be grounded to prevent any electrical accidents.


z Improper or no grounding may cause electrical shock.
z Grounding terminal is included in the power plug. Therefore, there is no need to earth if power cord
is plugged to a power outlet with grounding terminal.
z If power outlet with grounding terminal is not available, install adapter to plug to ground earth wire
from green terminal or install earth wire to the grounding terminal of the equipment.

J-300S-M-E-09-001(REV.01) Page 10
Operation Manual

5. Name and functions Controller.

① ②


④ ⑤

[PID Controller]

① : PV display window : Indicating current temperature

② : SV / Time display window : Indicating temperature setup and time.

③ : Heating Status Display : Lamp is turned on when heater is ON.

④ : ALARM Status Display : Lamp blinks when it is alarmed.

when the unit is abnormal condition, the alarm is working.


: TIMER Function Status Display : Lamp is turned ON when timer function is in use.

t_A : Lamp blinks below setting temperature and Lamp is turned on


⑤ when temperature is reach the setting temperature.
t_B : Lamp is turned on when press the RUN key.
⑥ : Auto Tuning Status Display : Lamp blinks when automatic tuning is in process.

: MODE Select Key : Used for setting up temp. time correction and other related parameters.

: Shift Key : Used to move decimal point of setting value.

: Increase Value Key : Used to change setting value.

: Auto Tuning and Timer Key : AT key function is starts if you press AT key for more than 4

seconds.

: Heater RUN/STOP Key

: POWER On/Off Key

J-300S-M-E-09-001(REV.01) Page 11
Operation Manual

z Key switch function feature of Temperature controller.

- Power ON / OFF Key : Turns Power ON or OFF. (It does NOT operate in setting mode.)
- RUN/STOP Key : RUN/STOP Heater. (It does NOT operate in setting mode.)
- MODE Key :
․ Press MODE Key once in Normal Display Mode and changes to Temperature, Time Setting
Mode.
․ Press MODE Key for over 6 seconds and changes to individual Parameter Setting Mode.
․ Press Mode key once again and the parameter changes to another parameter,
and pressing the MODE Key for more than 2 seconds ends Parameter Setting Mode, and
returns to Normal status.
- Shift Key :
․ Moves the decimal point of setting data in each MODE.
․ Press Shift Key for over 6 seconds and changes to Factory Setting Mode.
- Increase setting value Key : Increase set-value of each decimal point
- AT (Auto tuning key) :
․ Displays present time, then temperature in turn when At key is pressed in Normal Display
Mode.
․ AT key function is starts if you press AT key for more than 4 seconds in Normal Display
Mode.
․ By pressing AT key for more than 4 seconds during Auto Tuning Process ends Auto Tuning
Process instantly.
․ Displays present temperature when AT key is pressed once during Auto Tuning Process.

J-300S-M-E-09-001(REV.01) Page 12
Operation Manual

5.2 Configuring controller.

5.2.1 Setting Temperature and Time MODE.

CPU with software which is capable of digital PID-Auto tuning is equipped and contained regulation
device of supper safety class such as the regulation function for temp. correction for Pt100, heating and
others.

Normal Status

* Press MODE Key


1) Temperature setting range
tEmP - PT TYPE : -99.9℃ ~ +299.9℃
(Temperature Setting) - Max. temperature is adjustable.
2) Time Setting Range
* Press MODE Key
- Day, Time MODE (TIME SCALE : d.h)
* Adjust with ◀ Key, ▲ Key : 0~99 days, 23 hours
- Hour, Minute MODE (TIME SCALE : h.m)
timE : 0~99 hours, 59 minutes
(Time Setting)
- Minute, Second MODE (TIME SCALE : m.s)
MODE Key : 0~99 minutes, 59 seconds
* If TIME set-value is 00.00 , it will operate
InS continuously.
(Input Shift Value) 3) InS (Input Shift Value)
= Input Deviation Value Correction
MODE Key
: -19.9~+19.9PT

*Press MODE Key for more than


2 seconds, it will return to
NORMAL Status

J-300S-M-E-09-001(REV.01) Page 13
Operation Manual

5.2.2 Controller Parameter Configuration Mode

Normal Status 1) AL : LOW ALARM


2) AH : HIGH ALARM
* Press MODE Key for over
3) ART : ALARM TYPE
6 seconds and changes to
individual Parameter Setting ALARM
ART ALARM OUTPUT REMARKS
Mode. MODE

0 No Alarm
AL
(LOW ALARM) PV
* Can be
High Low Alarm ON 0FF ON
1 AL AH used as
(Absolute value)
* Adjust with ◀ Key, ▲ Key refrigeration
High Alarm PV
0FF ON
AH 2
(Absolute value) AH

(HIGH ALARM) * Can be


Low Alarm
PV

ON 0FF
3 used as
MODE Key (Absolute value) AL

refrigeration
* Can be
Art High Low Alarm ON
PV
0FF ON
4 used as
(ALARM TYPE) (Relative value) SV-AL SV+AH

refrigeration
High Alarm PV
OFF ON
MODE Key 5
(Relative value)
* Can be
H.SAF Low Alarm ON OFF
PV

6 used as
(HI SAFETY ALARM) (Relative value)
refrigeration
MODE Key
4) H.ASF (HI SAFETY ALARM) : 0~99 (PT)
If present temperature is higher than "set value +
CooL
H.ASF value", then timer stops and changes to
(COOLING TEMPERATURE) STOP MODE.
MODE Key 5) COOL (COOLER SET TEMPERATURE)
: set within temperature range
HyS - displays only when cooler is in activation.
(HYSTERESIS) - If Set temperature > COOL set value,
COOL RELAY ==> OFF
MODE Key
- If Set temperature > COOL set value,

bEEP COOL RELAY ==> OFF


6) HYS (HYSTERESIS TEMPERATURE) : for AH, AL
(BUZZER ON TIME)
- 0~29.9 (PT)
MODE Key ex) If AH=50, High Alarm activates at 50℃, and
HYS=0.5 deactivates at 49.5℃.
P Same applies for AL.
(PROPORTIONAL FACTOR) 7) BEEP (BUZZER TIME)
MODE Key : 0~9999sec ( 0sec = continuous )
8) P (PROPORTIONAL FACTOR)
: 0~999.9
- Set automatically after Auto Tuning process.

J-300S-M-E-09-001(REV.01) Page 14
Operation Manual

9) I (INTEGRAL FACTOR)
I
: 0~999.9
(INTEGRAL FACTOR)
- Set automatically after Auto Tuning process.
MODE Key
10) D (DIFFERENTIAL FACTOR)
: 0~999.9
D
- Set automatically after Auto Tuning process.
(DIFFERENTIAL FACTOR)
11) AR (ANTI RESET FACTOR)
MODE Key : 0~999.9
- Set automatically after Auto Tuning process.
Ar
12) DRAN (TEMPERATURE FIXED DISPLAY RANGE)
(ANTI RESET FACTOR)
: Displays present temperature in the range as
MODE Key
SET TEMPERATURE.
Ex) If set temperature is 50℃ and DRAN is 0.3℃
drAn
it displays present temperature within range of
(TEMP. FIXED RANGE)
49.7~50.3℃ as 50℃.
MODE Key
13) ACTP (TIMER STARTING TEMPERATURE VALUE)

ACtP : Relative value of set temperature


14) LOC
(TIMER STARTING
: KEY LOCK
TEMPERATURE VALUE)
- Press Key Lock : YES ==> LOCK
MODE Key NO ==> UNLOCK

LoC
(KEY LOCK)

MODE Key

* Press MODE Key for more than


2 seconds, it will return to
NORMAL Status

z DO NOT CHANGE the SET VALUE, which is FACTORY SET DEFAULT VALUE !
z Changing the SET VALUE may result in MAL-FUNCTIONING !

J-300S-M-E-09-001(REV.01) Page 15
Operation Manual

5.2.3 Controller Factory Configuration Mode

Normal Status
1) SEN (SENSOR TYPE)
* Press SHIFT Key for
* PT TYPE : Pt display ==> DPt (Default)
over 6 seconds
* Beep sound JPt display ==> JPt
2) PO_m (POWER FAILURE MANAGEMENT)
SEn
* ON ==> START DRAN, PO_m ON (Default)
(SENSOR TYPE)
OFF ==> PO_m OFF
MODE Key
* Adjust with ▲ Key 3) A_TY (RELAY CONTACT TYPE)
* Displays when "ALARM RELAY" is selected.
PO_m
( Refer to (5) A.C.R Mode )
(POWER FAILURE
MANAGEMENT) 4) SR_A (START RELAY OFF in HIGH ALARM)

MODE Key * NON => Does not function. (Default)


* OFF => START RELAY OFF.
A_ty
5) A.C_R (ALARM or COOLER relay SELECT)
(RELAY CONTACT TYPE) * ALr : ALARM RELAY function (Default)
MODE Key
* Cor : COOLER RELAY function
Sr_A 6) OF.AR (ALARM RELAY OUTPUT when OFF or STOP)
(START RELAY OFF in AH) * ON => ALARM RELAY ON (Default)
MODE Key * OFF => ALARM RELAY OFF
7) DP.ON (DECIMAL POINT DISPLAY)
A.C_r
* Displays when selected for Pt sensor.
(ALARM / COOLER SELECT
* ON => Displays decimal point for temperature
MODE Key
(Default)
oF.Ar * OFF => No decimal point for temperature
(OFF => ALARM OUTPUT) 8) TM.TY (TIME MODE SETTING)
MODE Key * t_A, : Timer is operating after setting temperature.
* t_B : Timer is operating press RUN key.
dP.on (DECIMAL POINT)
9) T_SC (TIME UNIT SETTING)
MODE Key
* Day, Time MODE (d.H)
tm.ty : 0~99 days, 23 hours

(TIME MODE SETTING) * Hour, Minute MODE (H.m ) : DEFAULT SETTING


: 0~99 hours, 59 minutes
MODE Key
* Minute, Second MODE ( m.s)
t_SC (TIME UNIT SETTING) : 0~99 minutes, 59 seconds
MODE Key * If TIME set-value is 00.00 , it will operate
continuously.
r n t
- Set automatically after Auto Tuning process.
(MAXIMUM IMPUT VALUE 10) r n t ( MAXIMUM INPUT VALUE)
MODE Key

* Press MODE Key for more than 2 seconds,


it will return to NORMAL Status
z DO NOT CHANGE the SET VALUE, which is FACTORY SET DEFAULT VALUE !
z Changing the SET VALUE may result in MAL-FUNCTIONING !

J-300S-M-E-09-001(REV.01) Page 16
Operation Manual

5.2.4 Output type

z Main output : * Output by PID control


* On during Heating (Heater LED is ON)
z Start Relay
When used as START RELAY
(when FACTORY CONFIGURATION MODE => SR_A => NON is selected)

ON

* OFF when SENSOR


OFF OFF
is disconnected

POWER ON Set time end

5.2.5 How to set automatic tuning for temperature.

z When temperature difference occurs during operation, press AT key for 4 seconds, and
automatic tuning process will begin, and AT lamp(Auto tuning lamp) will start to blink until
automatic tuning process is completed.
z When automatic tuning is in process, NEVER change setting values nor OPEN/CLOSE the door
until automatic tuning process is completed.

z Press AT key for 4 seconds during Auto Tuning Process , and Auto Tuning Process will
be stopped instantly.

5.3 How to set temperature.

1) Turn on power switch

2) Press MODE key at full length for 1 second.


Then tEmP is shown on display window of PV side.

Press key and key to set required temperature.

3) After setting the temperature, press MODE key once


Then timE is shown on display window of PV side.

Press key and key to set required time for timer function.

4) When setting on all units is finished, pressing MODE key at full length for 2 seconds,
and all setup values will be reset as set values and return to normal mode.

J-300S-M-E-09-001(REV.01) Page 17
Operation Manual

5.4 How to operate

z Check rated voltage of the equipment as well as check the specification. (AC110V, 50/60Hz).
-In case of using 110V, use a boosting transformer with enough capacity.
-Be sure to use AVR (Automatic Voltage Regulator) in the place with unstable voltage

1) Connect the power plug to the power outlet.


2) Turn ON the Main Power Switch located on the right-bottom side of the main body.
3) Turn ON the Power by pressing POWER On/Off Key once located on the controller
4) Set and configure temperature settings by pressing the keys on the Temperature Controller.
(Refer to Chapter 5 of this manual, page 11~16 for details.
5) Open the door, put the sample/specimen in the chamber, and close the door.
6) After the operation is completed, Turn OFF the Power by pressing POWER On/Off Key once
located on the controller.
Turn OFF the Main Power Switch located on the right side of the main body.
7) Ensure that the equipment has cooled down to a safe temperature to touch. Open chamber door and
take out the sample/specimen from the chamber, and close chamber door.
8) Disconnect the power plug from power outlet

z Internal wall or board of the chamber might a little soot and colored by the fume
generated inside the chamber.

J-300S-M-E-09-001(REV.01) Page 18
Operation Manual

6. Maintenance and service.

6.1 In case of cleaning inside of the chamber

z Ensure that the apparatus has cooled down to a safe temperature to touch.
- Disconnect the power plug from power outlet.
- Clean the inside part of the equipment with sponge or soft clothes getting wet by neutral cleaner.
- Clean the inside part of the equipment again with sponge or soft clothes getting wet by water.
- Clean with dry clothes.
- Do not use organic solvent.
- Do not use flammable or volatile chemicals such as benzene. alcohol to clean.

6.2 In case of cleaning exterior of the equipment

- Disconnect the power plug from power outlet.


- Ensure that the apparatus has cooled down to a safe temperature to touch.
- Clean the exterior of the equipment with sponge or soft clothes getting wet by neutral cleaner.
- Clean with dry clothes.
- Do not use organic solvent.
- Do not use flammable or volatile chemicals such as benzene. alcohol to clean.

z Do not use flammable or volatile chemicals such as benzene. alcohol to clean the equipment.
z This equipment is not designed nor manufactured Explosion-proof, so never place highly flammable
material or explosive material inside the equipment which may cause an explosion or a fire.

6.3 Cleaning electric parts.

-Clean with dry clothes.

6.3 When the equipment is not used for long time.

- Remove power plug from power outlet.


- Clean with soft clothes.
- Pack the equipment in a appropriate way and store in a safe place for storage.

J-300S-M-E-09-001(REV.01) Page 19
Operation Manual

7.Trouble shooting.

① In case apparatus fails to work at all :-

z Control unit switched off :-


=> Ensure unit is on.
z Power supply shortage :-
=> Refer to specification and check whether sufficient power is being supplied.
=> Is the breaker on the side of outer part open?
z Check if lamp is turned down.
=> Check unit in a power outlet that is known to work.
If still does not work, contact your local JISICO agent/distributor & have it repaired.

② In the machine inclined?


install the machine horizontally all the time.

③ In case general failure :-


=> Check unit in a power outlet that is known to work.
If still does not work, contact your local JISICO agent/distributor & have it repaired.

8. After-sales service.

8.1 Free repair

Faults which is responsible for manufacture in normal condition can be repaired with free for 1year from
purchase date (only, pressure vessel is zero year), and it is desirous to check following items when
requesting A/S.
. Part and condition generated fault
( It is necessary to explain in detail within limit of possibility )
. Model name
. Serial Number
. Purchase date (year, month, date)

8.2 Onerous repair

In following case, it is required to repair with compensation in spite of guarantee period.


- Fault by user mistake, unsuitable repair or modification
- Fault by negligence for carrying and handling during use
- Fault due to natural disasters such as fire, flood disaster and abnormal voltage
- Fault by using against direction for use.

J-300S-M-E-09-001(REV.01) Page 20
Operation Manual

9. Specification.

9.1 Technical Specification

Model J-300S

Type Forced convection system

Overall size(WxDxH) mm 580 × 600 × 820

Chamber size(WxDxH) mm 450 x 450 x 450

Capacity ℓ 91

Range Ambient +10℃ ~ 260℃

Using range 40℃ ~ 250℃

Accuracy ±0.5℃ at 100℃


Temperature
Uniformity ±1.0℃ at 100℃

Controller PID control

Regulator TRIAC

Thermocouple K type (CA)

Heater(incoloy sheath) W 1,200

Circulation fan & motor Ø150 x 60mm, 24W

Insulation Glass wool

Observation window Reinforced glass 5T

Self diagnosis function(Memory error, Temp. overheating,


Safety device
SSR short circuit, Sensor short circuit), ELB

Shelves 3 ea

Interior Stainless steel plate


Material
Exterior Steel plate with powder coated finish

Power source AC 110V, 50/60Hz, 1 phase

Net weight kg 51

Certificate CE

J-300S-M-E-09-001(REV.01) Page 21
Operation Manual

9.2 Component list

Description Standard Q'ty

Safety Glass 5.0t 2.00

Rubber LEG Bolt type 4.00

Packing Rubber (2m) 1.00

Cord 125V, 15A 1.00

Main board NTHC-1 1.00

Display board JSFP-2 1.00

Power relay HR705-2P 1.00

Heater 230V, 1phase, 600W 2.00

Insulation Glass wool 1.00

Main Switch ELB, 220V, 1phase, 15A 1.00

TRIAC TG25C600 1.00

Shelves Stainless steel 3.00

Thermocouple K type 1.00

Circulation
230V, 1phase, 24W 1.00
fan motor

Circulation fan Ø150 x 60H 1.00

J-300S-M-E-09-001(REV.01) Page 22
Operation Manual

9.3 Assembly Drawing

J-300S-M-E-09-001(REV.01) Page 23
Operation Manual

9.4 Circuit Diagram

J-300S-M-E-09-001(REV.01) Page 24
Operation Manual

9.5 CE certificate

J-300S-M-E-09-001(REV.01) Page 25
Operation Manual

9.6 ISO certificate

J-300S-M-E-09-001(REV.01) Page 26
Operation Manual

Warranty Certification.

Thank you for purchasing a JISICO product. Please take a moment to fill in the following and return it
to JISICO CO., LTD. for product registration.

z JISICO products are warranted against defect in material and workmanship for a period of one(1) year
after delivery to the first purchaser for use.
z This warranty does not apply to the damage resulting from misuse, alteration, transportation, abuse,
accidents, tampering, servicing attempted or performed by unauthorized service agency.
z Before use, read the manual thoroughly. JISICO can not be held responsible for the malfunctioning
resulting from the use of the product other than specified in operation manual.
z Call or fax Customer Service for a Return Material Authorization (RMA) number before returning the
product or any specific material.
z Put the RMA number on the shipping label, and a written detailed description of the problem.

Product Name
Model No. Serial No.
Delivery Date / / ( dd/mm/yy) Warranty One(1) year
JISICO Company
Agent/ Distributor Phone No.
Company Contact
Custom Address
Phone No. Fax. No.

JISICO only warrants the safety and proper operation of product under the condition that :
- The product is properly and in accordance with the instructions stated in the operation manual.
- No modifications are made.
- Only original spare parts are used. The use of other parts carries unknown risks and will void unit's
warranty.
- The back of the unit has not been opened.
Signed by

Chung Hee Kang / President


JISICO CO., LTD.
JISICO BLDG. 301-1, SEONGSU-DONG 2-GA, SEONGDONG-GU, SEOUL, 133-835, KOREA
Tel : +82 2 461 7071~3 +82 2 467 7071~3
Fax : +82 2 464 1962
http://www.jisico.co.kr
E-mail : s67@jisico.co.kr

J-300S-M-E-09-001(REV.01) Page 27