Key signal testing points: TP81 Reset signal of the monitoring CPU TP84 U7 stepping signal

controlled by the monitoring CPU TP88 U7 enabling signal controlled by the monitoring CPU
TP87 U7 direction signal controlled by the monitoring CPU TP85 U7 current size signal
controlled by the monitoring CPU TP25, TP26 Motor coupler detection signal TP24 Bubble
detection output signal TP29 Ultrasonic emission enabling signal TP36 Holze switch door
detection output signal TP98 Pressure detection signal TP38 Drip enabling signal TP23 Drip
detection signal 2-9 2.3.2 Key Board Serial Keys communication Battery indicator light AC
indicator light Red-yellow alarm light Wifi module (optional) LCD LCD backlight control circuit
From control board Hall sensor Motor detecting board interface Ultrasonic transceiver circuit
Gold capacitor circuit Key Board Chart Key testing points TP50 VDD 3.3V TP51 VCC 5.0V TP32
LCD backlight power TP53 Signal of battery indicator light TP52 Signal of AC indicator light TP24
Serial output signal of Wifi module TP25 Serial input signal of Wifi module TP56 Gold capacitor
power supply voltage TP54, TP55 Red-yellow alarm light drive signal TP1 Holze switch door
detection signal TP34, TP35 Double coupling motor testing signal TP4, TP5 Ultrasonic emission
signal TP8 Ultrasonic receiving signal 2.3.3 Motor Detection Board Motor speed detection
adopts double coupling circuit, which detects direction and speed; a gap on the detection
encoder is used for detecting the camshaft of pump body. TP1, TP2 Motor coupler output
signa