Trima Accel® System Preventive Maintenance Checklist

Customer:
Technician:
Work Order: Serial Number: Date:

This checklist is to be used with PM Procedure 777098-829.


Checkmark or “X” indicates step was successfully completed.
1. Introduction
Test Equipment Verification Control Number Calibration Due Date
DVM or DMM:
Leakage Current box or meter
Pressure Monitor/Meter
Torque Screwdriver or Torque Wrench
2. General Mechanical Inspection and Cleaning
1) IV Pole
3) Display Console
4) Display Arm
5) Front Panel Valves
6) Pump Rotor Test
7) Door Alignment
8) Door Screws Test
9) Inner Cabinet
10) General Inspection
11) Brake System
12) General Cleaning
13) Filler Latch Lubrication
3. Electrical Safety and Power Tests
1) Grounding Straps Test
2) Ground Wire Resistance Test ≤ 0.3 Ω Ω
3) External Ethernet Isolation Test
4) Line Voltage Test Record Value VAC
5) Power Up Test
6) Normal Condition with Neutral Closed
 AC Power Voltages ≤ 165 VAC or ≤ 300 µA
 AC Power Voltages > 165 VAC ≤ 500 µA
Closed Neutral – Forward Record Value µA
Closed Neutral – Reverse Record Value µA
7) Single Fault Condition with Neutral Open
 AC Power Voltages ≤ 165 VAC or 35 to 600 µA
 AC Power Voltages > 165 VAC 35 to 1000 µA
Open Neutral – Forward Record Value µA
Open Neutral – Reverse Record Value µA
4. 5V Test +4.95 to +5.05 VDC or NA VDC or NA
5. OVP, OTW and Reset Test
6. Fan Direction and Failure Tests
1) Air Flow Direction
2) Fan Failure Detection
7. Machine and Software Information
2) Machine On Time Record Value
3) Number of Runs Record Value
4) Trima Accel Software Version Record Version
5) Service Software Version Record Version

1st Page: Customer 2nd Page: Administration 3rd Page: Service Engineer P/N 777100-430 PAGE 1 OF 2
Trima System Preventive Maintenance Checklist

Work Order: Serial Number: Date:


8. Trima Calibrations
1) Cassette Plate Down Test
2) Trima Accel Calibration
3) Install Calibration Files
9. Verifications
2) Alarm LED Test
3) AGC Upper Level Sensor Air Test +0.75 to +0.95 VDC VDC
4) AGC Lower Level Sensor Air Test +0.75 to +0.95 VDC VDC
5) Touch Screen Test
6) Access Pressure Sensor Test ………
0 mmHg Test –20 to +20 mmHg mmHg
+330 mmHg Test +290 to +370 mmHg mmHg
–250 mmHg Test –220 to –280 mmHg mmHg
+550 mmHg Test +500 to +600 mmHg mmHg
7) Centrifuge Pressure Sensor Test……
+500 mmHg Test +450 to +550 mmHg mmHg
+1300 mmHg Test +1170 to +1430 mmHg mmHg
10. Trima Accel Autotest
11. Pump Occlusion Tests
2) AC, Inlet, & Return Pump Occlusion
3) Plasma & Platelet Pump Occlusion
4) Machine Set-up
12. Fluid Run
1) Fluid Run Set-Up
2) AC Sensor Test
3) Stop Button Function Test
4) Draw Pressure Too Low Function Test
5) Return Pressure Too High Function Test
6) Centrifuge Pressure High Function Test
7) AGC Upper Level Sensor FLUID Test +0.15 to +0.55 VDC VDC
8) AGC Lower Level Sensor FLUID Test +0.15 to +0.55 VDC VDC
9) Power Interrupt Function Test
10) End of Fluid Run
11) Next Procedure Test
13. End of Preventive Maintenance
1) Logbook
2) Verify all Entries

Notes:

As applicable, leave the procedure and/or a copy of this checklist with the customer.

1st Page: Customer 2nd Page: Administration 3rd Page: Service Engineer P/N 777100-430 PAGE 2 OF 2