floor area and structure should fully consider the convenience of medical
work. Specific requirements are as follows:
a) X-ray shielding requirements for the room: X-ray shielding must comply
with the regulations of the health and epidemic prevention department.
b) Room size requirements: Room area ≥ 5.5m × 4.5m, room height ≥ 3m,
minimum door size ≥ 1.6m × 2.2m (width × height).
d) Requirements for the机房 floor: The foundation should be solid, flat, and
clean. The room should not be cluttered with items unrelated to diagnostic
work, and it should be kept dry and well-ventilated.
j) Cable trench requirements: Design cable trenches for wiring within the
room. The direction and width of the cable trench can be excavated based
on the position of the equipment. The cable trench cover plate should have
sufficient strength and be removable for easy maintenance in the future.
It is recommended to arrange the cable lines of the unit underground in a
hidden line layout. Between each component, pre-drilled wiring channels
should be prepared, especially for concrete floors. Pay attention to this
issue. Pay attention to moisture prevention and prevention of rodent
damage to cables in the wiring channel. It is recommended to place a metal
-made trough-shaped wiring channel in a concrete trench. Figure 2 Cross-
section of the trench.
Figure 2 Cross-sectional view of the trench.
c) Monitor the temperature change rate within the computer room and ensure it
does not exceed 10 degrees per hour to prevent condensation water from
forming inside the control cabinet and damaging equipment.
Based on the site, the position of equipment placement and the installation
site requirements for each component are as follows:
generator
chest stand
lead glass
flimming bed
Console
1) When the floor load capacity in the computer room does not meet the
requirements, it will cause the unit to be unable to be installed.
2) When the net height in the机房 is less than 2800mm, it will cause the unit
to be unable to operate.