PERFORMING AN ANALYSIS CYCLE


- Switch on the instrument
- Click START

Wait for the instrument to complete the initialization phase

Gently invert the tube 10-12 times for complete


inversion of the tube so that the air bubble that forms run through it
from one end to the other, immediately before starting the analysis.

From this moment on, it is possible to introduce the sample for testing.

To start a cycle, insert the tubes orienting them as shown in the


photo in one of the free positions. If a barcode reader (optional) is
present, read the sample identification code before inserting the
tube. After insertion, the instrument scans to determine the blood
level at zero sedimentation time.
During the first scan a “CHECK TUBE” pop-up window appears which determines the
display of the tube type (normal or pediatric) and the blood level, after which the
countdown will be activated.

Procedure for map TUBES (BD MICROTAINER®, MICROVETTE® APT


(SARSTEDT))

To analyze a MAP tube sample, touch the screen in one of the


empty positions to open the NEW SAMPLE window.

Select “Microtube MAP” YES (Y) and insert the tube in one of the
free positions.
The tube will be represented as a pediatric tube, with the letter
M on the cap, indicating that they are MAP tubes.

If most of the tubes are BD Microtainer® MAP, in the settings menu, MICROTUBE MAP
parameter, select ON.
July 2021

EXECUTION OF ANALYSIS CYCLE - RANDOM ACCESS


- During a cycle it will be possible at any time to insert a new sample in each of the free
positions present.
- Mix the sample thoroughly before inserting it into the instrument as described above

RESULTS
At the end of the 20 minutes (14 min for Pediatric) the instrument displays the result.
Green color indicates ESR within the normal range.
Red color indicates an ESR that has exceeded the normal level.
The results are expressed in mm / h (Westergren 1st hour).

The results can be printed via Bluetooth printer (Optional)


You will need to connect it to the instrument first from the menu.
Printing can take place automatically at the end of the analysis, or

- Press the ARCHIVE key


- Select the samples for consultation / printing of the results

QUALITY CONTROL

Entering the settings menu, under QC, you can enter the
characteristics of each control, just read the bar code of the QC
Diesse. Then, every time the control is inserted in any position of
the instrument, it will be recognized as QC, and displayed on the
screen, as shown in the image.
The tube with the white cap indicates the Normal control, the
one with the black cap the Pathological control.

In case you are using another trade control, it is recommended that you
create dedicated barcode labels and manually enter the data.

REFILL
- Press the SETTINGS button
- Select REFILL on the second page
- Insert the TEST DEVICE NEXT in the appropriate compartment
(as shown in the figure)
- Press the CHECK key
- Press OK to confirm the test number to reload
- Automatic update of the number of tests displayed on the
instrument screen (top left)
July 2021

TROUBLESHOOTING
ERROR MESSAGES CAUSE REMEDY
A cycle was started with some Remove the tubes and restart the run.
Remove all test tubes tubes already inserted. Turn off the instrument, turn it over and try to
Otherwise, there may be a clean it internally. If the problem persists,
foreign body in front of the contact technical assistance.
sensors.
Low (Lo) Blood level inside the tube too If possible, add blood up to established levels.
low
High (HI) - Blood level inside the tube too - If possible, remove blood up to the
high established levels
-Interference caused by labels - Try to reposition the tube as described
previously
ERR Generic error, pressing on the - Check that the tube used is compatible with
error message a number will the instrument or the possible presence of a
appear that differentiates the number of labels greater than 2 in case remove
cause. the excess labels. Check for any lumps or clots.
See complete troubleshooting - Try to repeat the test
in the User Manual
Printer error, retry? Printer not found. Make sure the printer is turned on. Otherwise,
deselect the Bluetooth mode (OFF)

Exam not available. Need refill The executable test counter is Follow the instructions to recharge the
equal to zero instrument with a new transponder described
in chapter 4.
USB pendrive not found The instrument does not Remove and reinsert the USB stick or change
recognize the USB stick. it.

Warning! Remain n tests The number of tests remaining It is recommended to recharge the instrument
is less than 50 as soon as possible with a new transponder

CHECK DEVICE already used The transponder used is empty. Use a new transponder

CHECK DEVICE not found The instrument does not Remove and reinsert the transponder. Use
recognize the transponder used another transponder