CSSD Equip

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 18

Specifications

CSSD

Index Item Description Code Total


1 HEAT SEALING MACHINE ST12040 2
2 SET, STEAM GUN ST18010 2
3 SINK, 2-BAY, COUNTER CLEANUP ST19020 1
4 STERILIZER, PLAZMA ST16035 1
5 STERILIZER, STEAM, DOUBLE DOOR ST16010 1
6 STERILIZER, STEAM, SINGLE DOOR ST16004 1
7 TABLE, WORK LARGE, 2000 x 600 mm ST16040 2
8 TABLE, WORK SMALL, 1200 x 600 mm ST16041 1
9 ULTRASONIC CLEANER ST20227 1
10 WASHER / DECONTAMINATOR, 185 LITER, DD ST17041 1
Masharif Hospital
Item Description HEAT SEALING MACHINE
Category CSSD
Item Code ST12040
Item Qty 2

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1 Compact and lightweight table top model.

2 Digitally controlled sealing temperature.

3
Protected against overheating
4
LCD display of sealing status
5
The package heat sealer shall have a spring-loaded hinge to open and
close the jaws
6
Heating time approximately 10 seconds or less
7
Counter records total number of seals performed
8
Accepts a wide range of bag sizes, Auto pack, and dust cover pouches; up
to 18 inches with a 1/2 inch seal.
9
Visual alarm to notify the operator when the bag has been sufficiently
sealed and the heating element has been cooled
10
Sealing width of heating element approximately 15 mm
11
The temperature range from 100 to 200 C
12
Feeder speed not less than 8 m/minute.
13
Accessories: Roll Holder and cutter
14
Optional: Barcode Reader attachment
Masharif Hospital
Item Description SET, STEAM GUN
Category CSSD
Item Code ST18010
Item Qty 2

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1 A stationary, wall-mounted cleaning unit for manual washes of trolleys or
other movable equipment

2 The unit includes container for chemical disinfectant to be injected into


the hospital-supplied water

3 The right amount of disinfectant is automatically injected into the water


regardless of water pressure

4 Capability to switch to rinse with pure water

5 Impact-resistant rubber-covered spray nozzle

6 Approximately 2 meter long, reinforced rubber hose

7 Non-return valve unit and a main shut-off valve are included in the supply
Hose frame to be included

8 Accessories: Comes with set of stainless steel brushes and gloves


Masharif Hospital
Item Description SINK, 2-BAY, COUNTER CLEANUP
Category CSSD
Item Code ST19020
Item Qty 1

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1 Bench top of stainless steel, floor stand of stainless steel including
stainless steel under shelf.

2 Double bowl of 500 x 400 x 200 mm fitted in middle of table, complete


with water mixer, Brass bottom valve etc.

3 Overall dimensions around: 2000 x 650 x 900mm


Masharif Hospital
Item Description STERILIZER, PLAZMA
Category CSSD
Item Code ST16035
Item Qty 1

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1 Low temp sterilizer that provides an efficient sterilization solution for
heat & moisture sensitive equipment in CSSD.

2 Floor standing castor mounted for convenient positioning

3 Construction: Made from high quality Medical Stainless Steel 316L

4 Chamber: Usable Volume 100L approximately.

5 Shelves: Removable & Fixed

6 3 Preprogrammed cycles: Normal, Advanced & Endoscopic

7 Ability to sterilize Double & Triple channel Lumen

8 Sterilize Flexible endoscopes single and Dual channel

9 LCD display of cycle parameters & error status (preferably full message)

10 Built-in printer with Printout of cycle processes

11 Alarms (Audible & Visual) for power failure, wet cycle and leakage

12 Safety against opening door while chamber temperature or pressure is


high
13 Cassette agent delivery system

14 Chemical Indicators: tapes, strips & peel pouches

15 Self-contained Biological Indicator with 48-hour readout


Masharif Hospital
Item Description STERILIZER, STEAM, DOUBLE DOOR
Category CSSD
Item Code ST16010
Item Qty 1

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1 Double door sterilizer for the sterilization of wrapped and unwrapped
porous and non-porous instruments and supplies in the CSSD

2
Sterilizer vessel:

2.1 Chamber, jacket and door are constructed form solid, high quality type
316L or 316 Ti stainless steel

2.2 Required total chamber usable space approximately 650 Lt or equivalent

2.3 Chamber is fully isolated and fully jacketed

2.4 The sterilizer cabinet's external construction shall be stainless steel

2.5 Minimum chamber thickness shall be 5 mm

2.6 Rapid cooling

2.7 PLC or Microprocessor control of all parameters

2.8 Controls mounted far from heated surfaces

2.9 Control panel: LCD display with FULL messaging of status of cycles, date,
operator, alarms etc. the largest the best
3
Cycles:

3.1 Minimum number of cycles required is 6

3.2 Supplier/contractor should list the maximum number of cycles the


sterilizer is designed to operate

3.3 Automatic leak test

3.4 Audio-visual alarms for incomplete cycles or wet cycle

3.5 Bowie Dick test

3.6 Tamper-proof access code that prevents changes in programmed cycles

3.7 Power System Failure backup until the end of the Cycle

3.8 Alarms of vessel flood, under temperature, over temperature, drain,


steam

3.9 Modifications of operating cycles to suite particular loads

4
Door:

4.1 Automatic operated

4.2 Manual Loading and unloading

4.3 Vertical sliding with safety switches

4.4 The door must stop automatically if an obstruction is encountered

4.5 Door interlock System to prevent opening door at high pressure.

4.5 Sealed by silicone-rubber gasket

5
Air Filter:

5.1 The supplier/contractor shall specify the filter separation efficiency


5.2 The supplier shall specify the frequency of replacement based on number
of cycles

5.3
Thermal printer for cycle documentation

5.4
Pipes, valves
5.5
It is preferable if the material of the pipes & valves in the sterilizer to
receive building supplied steam are constructed from copper
6
Steam Source:
6.1 The primary steam supply is from the central steam supply, however, in
case of the central steam supply is out, then the proposed unit shall
operate on back up built-in steam generator
7 Accessories required: 2 original loading trolleys, 2 original carriage
trolleys, Chemical and biological indicators, Sterilization baskets…
8
Internal battery backup for all cycle memory
9
The supplier/contractor shall submit the list of standards (BS or EN) to
which the sterilizer is designed and constructed and a statement of
compliance.
10 The unit shall include all accessories for full functionality

11 The unit shall be CE marked and/or FDA approved


Masharif Hospital
Item Description STERILIZER, STEAM, SINGLE DOOR
Category CSSD
Item Code ST16004
Item Qty 1

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1 Single door sterilizer for the sterilization of wrapped and unwrapped
porous and non-porous instruments and supplies in the CSSD

2
Sterilizer vessel:

2.1 Chamber, jacket and door are constructed form solid, high quality type
316L or 316 Ti stainless steel

2.2 Required total chamber usable space approximately 250 Lt or equivalent

2.3 Chamber is fully isolated and fully jacketed

2.4 The sterilizer cabinet's external construction shall be stainless steel

2.5 Minimum chamber thickness shall be 5 mm

2.6 Rapid cooling

2.7 PLC or Microprocessor control of all parameters

2.8 Controls mounted far from heated surfaces

2.9 Control panel: LCD display with FULL messaging of status of cycles, date,
operator, alarms etc. the largest the best

3
Cycles:

3.1 Minimum number of cycles required is 6

3.2 Supplier/contractor should list the maximum number of cycles the


sterilizer is designed to operate

3.3 Automatic leak test

3.4 Audio-visual alarms for incomplete cycles or wet cycle

3.5 Bowie Dick test

3.6 Tamper-proof access code that prevents changes in programmed cycles

3.7 Power System Failure backup until the end of the Cycle

3.8 Alarms of vessel flood, under temperature, over temperature, drain,


steam

3.9 Modifications of operating cycles to suite particular loads

4
Door:

4.1 Automatic operated

4.2 Manual Loading and unloading

4.3 Vertical sliding with safety switches

4.4 The door must stop automatically if an obstruction is encountered

4.5 Door interlock System to prevent opening door at high pressure.

4.5 Sealed by silicone-rubber gasket

5
Air Filter:
5.1 The supplier/contractor shall specify the filter separation efficiency

5.2 The supplier shall specify the frequency of replacement based on number
of cycles

5.3
Thermal printer for cycle documentation

5.4
Pipes, valves
5.5
It is preferable if the material of the pipes & valves in the sterilizer to
receive building supplied steam are constructed from copper
6
Steam Source:
6.1 The primary steam supply is from the central steam supply, however, in
case of the central steam supply is out, then the proposed unit shall
operate on back up built-in steam generator
7 Accessories required: 1 original loading trolleys, 2 original carriage
trolleys, Chemical and biological indicators, Sterilization baskets…
8
Internal battery backup for all cycle memory
9
The supplier/contractor shall submit the list of standards (BS or EN) to
which the sterilizer is designed and constructed and a statement of
compliance.
10 The unit shall include all accessories for full functionality

11 The unit shall be CE marked and/or FDA approved


Masharif Hospital
Item Description TABLE, WORK LARGE, 2000 x 600 mm
Category CSSD
Item Code ST16040
Item Qty 2

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1
Used for sorting, wrapping and preparation of items for sterilizations in
the CSSD department
2
Constructed of stainless steel 316L or 304 stainless
3 Chamber, jacket and door are constructed form solid, high quality type
316L or 316 Ti stainless steel

4
Adjustable legs
5
Shelf to wrap supplies
6
Dimensions (HxWxL) mm: approximately 900x2000x600
Masharif Hospital
Item Description TABLE, WORK SMALL, 1200 x 600 mm
Category CSSD
Item Code ST16041
Item Qty 1

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1
Used for sorting, wrapping and preparation of items for sterilizations in
the CSSD department
2
Constructed of stainless steel 316L or 304 stainless
3 Chamber, jacket and door are constructed form solid, high quality type
316L or 316 Ti stainless steel

4
Adjustable legs
5
Shelf to wrap supplies
6
Dimensions (HxWxL) mm: approximately 900x1200x600
Masharif Hospital
Item Description ULTRASONIC CLEANER
Category CSSD
Item Code ST20227
Item Qty 1

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1 Floor- Unsurpassed ultrasonic energy, Plus detergent cleaning for
instruments and other hard goods.
2
Hands –free, toe switch operation via a powered load-tray elevator
3 The toe switch opens and closes the lid and begins the cleaning process
4
Automated water fill, Programmable controls, detergent injection,
solution heating and cycle timing
5
Push button for drain
6
Liquid level sensor and overflow protection
7
Automatic detergent dosing and low detergent indicator
8
Digital display for time, real temperature, set temp
9
Audible & visual alarms
10
Capacity: approximately 60 Liter
11 The unit shall include all accessories for full functionality
12 The unit shall be CE marked and/or FDA approved
Masharif Hospital
Item Description WASHER / DECONTAMINATOR, 185 LITER, DD
Category CSSD
Item Code ST17041
Item Qty 1

General Requirements
Manfacturer Supplier
Model No. Origin
Delivery Period Warranty
Utility Requirements
Environmental Req.
IT_Integration requirement
Unit Price Total Price

Specifications

Bidders
No Specifications corresponding Bidder
technical Comments
specification
1 Washer/decontamination units shall be used in the CSSD department to
clean surgical instruments, rigid tubular instruments, anesthesia materials
& laboratory glassware

2 Microprocessor controlled

3 Glass door for easy inspection

4 Capacity: approximately 150 Lt

5 Phases of each cycle: pre-wash, wash, rinse and thermal rinse, all as
standard

6 Drying capabilities

7 Rinsing capabilities post main any wash cycle

8 Adjustable cycles (instruments, gentle, utensils, glassware, plastic goods


and rubber goods)

9 Descaling cycle for removal of scale and other hard water deposits from
chamber and re-circulating piping

10 At least 20 pre-programmed cycles

Capability to program at least 20 additional program

Built-in storage for different types of detergents

Empty chemical alarm to prevent cycles running without detergent

Capability to be programmed to operate on a wash-only cycle


Internal battery to back up all cycle memory

Temperature device to sense the temperature inside the chamber

A temperature indicator/recorder should be provided to monitor and


record the chamber temperature throughout the cycle

Supplied with Instruments rack (with at least 2 levels, 3 baskets) and


Anesthesia rack

11 The unit shall include all accessories for full functionality


12 The unit shall be CE marked and/or FDA approved