Professional Documents
Culture Documents
RPSRPSG 160 Spec
RPSRPSG 160 Spec
User’s Manual
+ +
EN 60601-1
IEC 60601-1
■ Features ■ Applications
5"× 3" compact size Oral irrigator
Medical safety approved (2 x MOPP) according to Hemodialysis machine
ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1 Medical monitors
Suitable for BF application with appropriate system consideration Sleep apnea devices
110W convention, 160W force air Pumps machine
EMI class B for class Ⅰ configuration
No load power consumption under 0.5W by PS-ON control (G model)
5Vdc standby output, Power Good, Power Fail ;
Remote sense for 5~15V ■ GTIN CODE
Protections: Short circuit / Overload / Over voltage / Over temperature MW Search: https://www.meanwell.com/serviceGTIN.aspx
■ Model Encoding
RPS G - 160 - 5
Output voltage
Rated wattage
Type
Series name
Block Diagram
fosc :100KHz(5V)
70KHz(12~48V)
+5VSB
DETECTION
CIRCUIT
POWER SWITCHING RECTIFIERS
& +5Vsb
PWM CONTROL FILTER +S
EMI RECTIFIERS POWER RECTIFIERS +V
I/P & &
FILTER PFC SWITCHING FILTER -V
-S
PG PG
FG DETECTION
CIRCUIT
O.L.P. CONTROL
O.V.P.
PS/ON PS/ON
CONTROL
100
100
With FAN
90
80
Without FAN 80
60 70
LOAD (%)
LOAD (%)
60
40
50
20
40
-20 0 10 20 30 40 50 60 70 (HORIZONTAL) 90 95 100 115 120 140 160 180 200 220 240 264
FAN
5cm
5 20.5CFM min.
SVR1
M1 CN3
HS1 1
2
3
4
1
2
76.2
66.2
HS2 3
FS1
4
3 5
CN1 2 6
1 HS3
7
FS2 8
CN2 CN901
M2 LED
CN901
4 -ψ
1234
3.5
5 117
127
PS/ON
5mA(max.)
34.6
GND
3 max.
AC Input Connector (CN1) : JST B3P-VH or equivalent Power Good Connector(CN3):JST B4B-XH or equivalent
Pin No. Assignment Mating Housing Terminal Pin No. Status Mating Housing Terminal
1 AC/N 1 PG
2 No Pin JST VHR JST SVH-21T-P1.1 2 GND
or equivalent or equivalent JST XHP JST SXH-001T-P0.6
3 AC/L 3 -S or equivalent or equivalent
4 +S
Installation Manual