INSTALLATION OPTIMUS RAD

7.2. Installation procedure

S Provide the service PC with the hardware key and switch it ON.
The hardware key provides access to special program settings and to menu ”Faultfind”.
Standard programming is possible without a hardware key.

S Connect the PC to X5 on EZ139 CENTRAL UNIT CU via a serial data cable:


(A 5m long data cable can be ordered via 12NC: 4512 130 56931)

PC: COM1 Generator: EZ139X5


(9-pole, female) (25-pole, female)

1 20
6
*2 2*
*3 3*
4 6
8
*5 7*
7 5
8 4

* = sufficient for communication RAD / RF

S Insert the floppy disks containing the self-unpacking exe-files of the firmware in the disk drive of the PC:
-- OMA : 4512 116 022xx
-- OMB : 4512 116 023xx
For unpacking on the harddisk of the PC about 5MB are needed.

S Generate a directory e.g. [C:\OPT_R_35] on the PC by typing <md C:\OPT_R_35> or use WIN commands.

S Copy the firmware from both floppy disks to the PC into the same directory C:\OPT_R_35 by typing
<copy A:\*.* C:\OPT_R_35> or use WIN commands.

S Start unpacking the programs by typing <OMAxxxxx.exe> and <OMBxxxxx.exe> or doubleclick on


<*.exe> --files.
The programs unpack all files needed for the update of the firmware and the newest service tools.

S After unpacking [OMAxxxxx.exe] and [OMBxxxxx.exe] can be deleted on the harddisk by typing
<del OMAxxxxx.exe> and <del OMBxxxxx.exe>.

S For the current contents of [OMA/OMBxxxxx.exe] read [OMA/OMBxxxxx.txt] on the floppy disks.

S For a new installation of the generator firmware see section 4 REPLACEMENT, chapter 3
”Exchange of firmware ...”.

S Call the installation program by typing <xrgscope> or <xrgscope lcd> for PCs with LCD screen.

2--28 (c/02.1) OPTIMUS RAD


E 1998 Philips Medical Systems OPTIMUS_RAD_2_c021

ALL RIGHTS RESERVED