ELECTROLYTE ANALYSER

Series

QUICK START GUIDE

Revision: 03-2021, UM/58-S/21/B www.erbamannheim.com


QUICK START GUIDE EC 90 ELECTROLYTE
ANALYSER
Series

The purpose of this guide is to provide essential information required for using the EC 90 series.
Nevertheless, operators are instructed to carefully read the user manual before starting operation. All
precautions, safety and operating instructions must be read, understood and followed.

Instrument and EC cartridge overview

Maintenance door:
access to pump,
Touch Screen valves and tubing

Front door:
access to
EC cartridge Sample cover:
compartment access to sample
aspiration unit

Biosensor
Fluidic connector

EC cartridge
connector

Lever arm

Aspiration needle

Proximity sensor
EC cartridge
compartment

FOR DETAILED INFORMATION please refer to the EC 90 User Manual PAGE 3 www.erbamannheim.com
QUICK START GUIDE EC 90 ELECTROLYTE
ANALYSER
Series

Start-up

Open maintenance door, remove Remove left side panel to access Confirm pump tube clamped and
1 foam block & clamp pump tube 2 valve V5
7 tube clips removed 8 Open front door and lift lever arm up

Remove tube clips from valves Close maintenance door and left side Remove cap & cap sealing from EC Insert new EC cartridge & close
3 V1, V2, V3 & V5 4 panel and connect to main power 9 cartridge and peel off sensor film 10 lever and front door

Confirm insertion & wait for


5 Log in with dummy account 6 Set system settings 11 startup completion 12 Insert printer paper (optional)

PAGE 4 FOR DETAILED INFORMATION please refer to the EC 90 User Manual PAGE 5 www.erbamannheim.com
QUICK START GUIDE EC 90 ELECTROLYTE
ANALYSER
Series

Performing measurements Performing on-demand calibration and QC

Select/deselect parameters in Select sample type, volume, measuring Initiate smart sensor calibration 1P or 2P calibration is performed
1 main screen & start measurement 2 mode & enter further information 1 directly in main screen 2 according to system requirements
e.g. EC cartridge e.g. EC cartridge plus iCa

Option 2

Option 1

Lift sample cover & pass sample Trigger aspiration via proximity To register a new lot, access QC Provide all information
3 to aspiration needle 4 sensor or status button 1 material screen and press add button 2 to complete registration

Close sample cover & wait until Test results are automatically Start measurement in QC Select QC level and lot number &
5 completion of measurement 6 displayed in results menu 3 results screen 4 proceed with measurement

PAGE 6 FOR DETAILED INFORMATION please refer to the EC 90 User Manual www.erbamannheim.com PAGE 7 www.erbamannheim.com
Erba Lachema s.r.o.
Karásek 2219/1d, 621 00 Brno, Czech Republic
Phone: +420 517 077 111
E-mail: diagnostics@erbamannheim.com
www.erbamannheim.com

Quality system certified ISO 13485