Ca

Calcium (OCPC)
BiossaysTM Clinical Chemistry

Intended Use
The Snibe BiossaysTM Ca reagent is used for quantitative determination of Calcium in human serum.

Assay Principle
In this method 8-Hydroxyquinoline is added to remove interference by magnesium and iron. O-Cresolphtha-
lein-complexone reacts with calcium in an alkaline solution to form a Ca-OCPC complex. The absorbance of
the Ca-OCPC complex is measured chromatically at 570 nm. The resulting increase in absorbance of the
reaction mixture is directly proportional to the calcium concentration in the sample.

Assay Characteristics
1~19 years: 2.26 mmol/L~2.66 mmol/L (2.5th-97.5th percentiles)
Expected Values 20~79 years: 2.11 mmol/L~2.52 mmol/L (Verified)

Limit of Detection (LoD) The LoD for Ca is 0.2 mmol/L

Measuring Range 0.2-3.75 mmol/L

Precision 1.94%-2.56% (1.56-3.10)

Package
REF Contents Supporting Analyzer
R1: 3×60 mL, R2: 3×20 mL,
130509001LC Calibrator: 1×2 mL Biossays BC1200
130509001SC R1: 1×60 mL, R2: 1×20 mL, Biossays BC2200
Calibrator: 1×2 mL Biolumi 8000

R1: 2×30 mL, R2: 2×10 mL, Biossays 240


130509001AC Biossays 240 Plus
Calibrator: 1×2 mL

BiossaysTM is a trademark of Snibe.