CRP IP

AA-Pro-ICRP-3 (06/2020) Ref : ICRP-XXXX

APPLICATION Selectra Pro Serie


Performance data, TouchPro application : Consult Instructions for Use

General Sample types


Name CRP IP v3 Normal
Abbreviated name CRP3 Allowed Yes
Code - Predilution No

Mode Two Point Reference values


Test parameters 1

Wavelength 340 nm Male 0.0 - 10.0 mg/L


Units mg/L Female 0.0 - 10.0 mg/L
Decimals 1 decimals Pediatric XXX - XXX mg/L
Test repeats 1 Panic XXX - XXX mg/L
Controls
Control 1 XXX
Concentration limits 0.0 - 190.0 mg/L Control 2 XXX
Calibrator name CRP IP CAL SET Control 3 XXX
Correlation
Prozone check No Correlation factor 1.000
Correlation offset 0.0 mg/L

General Measurement
Name CRP IP v3 Slope Blank No
Predilution No Point one, two -3 236 sec
Sample
Sample blank No
Test parameters 2

normal volume rerun volume Factor 1


16.0 µL 16.0 µl Absorbance limits
Reagents Absorbance -0.100 - 1.000 Abs
Reagent blank No R. Absorbance -0.100 - 0.300 Abs
normal volume rerun volume
Reagent addition at - 2.25 minutes (R1) Substrate depletion 3.000 Abs
250 µL 250 µl
Reagent addition at 2.90 minutes (R2)
0 µL 0 µl
Reagent addition at 4.70 minutes (R3)
25 µL 25 µl

Calibrator Predilution
Name CRP IP CAL SET Auto predilution No
Number of standards 6
Calibration

Calibration accepted -
Repeats 1
Interval 21 days
Curvefit Calibration limits
Algorithm Smoothed cubic spline Low High
1.0 % Absorbance 0.000 % 0.000 %
Auto accept calibration No Concentration 0.000 % 0.000 %

N° Used Concentration
#1 Standard #1 Enter conc.
#2 Standard #2 Enter conc.
#3 Standard #3 Enter conc.
#4 Standard #4 Enter conc.
#5 Standard #5 Enter conc.
#6 Standard #6 Enter conc.