Technical Data
DANEO 400
2 OMICRON
DANEO 400 technical data
OMICRON 3
DANEO 400
Maximum error
Sampling frequency 10 kHz / 40 kHz 10 kHz / 40 kHz 10 kHz 40 kHz
Frequency range DC 10 Hz ... 1 kHz 1 kHz ... 4 kHz 1 kHz ... 10 kHz
10 mV 0.08 + 0.50 0.20 + 0.30 0.20 + 0.30 0.20 + 0.30
100 mV 0.08 + 0.07 0.08 + 0.05 0.16 + 0.04 0.16 + 0.04
1V 0.08 + 0.02 0.08 + 0.02 0.16 + 0.04 0.16 + 0.04
10 V 0.08 + 0.02 0.08 + 0.02 0.16 + 0.04 0.16 + 0.04
100 V 0.08 + 0.02 0.08 + 0.02 0.16 + 0.04 0.16 + 0.04
600 V 0.08 + 0.02 0.08 + 0.02 0.16 + 0.04 0.16 + 0.04
4 OMICRON
DANEO 400 technical data
1.4 Harmonics
The first order harmonic (order 1, designated as f1) is the fundamental component. Higher harmonic
orders are 2 to 25.
There are two THD figures available, THDf and THDr, which are relative to the fundamental component
and to the RMS value, respectively. The THDf calculation conforms to the definitions in IEEE P1495,
IEEE 519, and IEC 61000. The THDf may exceed 100 %. Vi and Ii are the magnitudes (RMS values) of
the individual spectral components.
N N
2 2
Vi or
Ii
i=2 i=2 -
THDf = ---------------------
- ⋅ 100% THDf = ------------------- ⋅ 100%
V1 I1
The THDr does not exceed 100 %. Vrms and Irms are the “fast RMS” values of the harmonic calculation.
or
···
N
2
Vi
N N
2 2
V rms = Vi
and
I rms = Ii
i=1 i=1
The maximum order (N) used in the calculations depends on several parameters such as nominal
frequency and is always higher than the maximum order i=25 offered for acquisition and analysis.
The following table enumerates the harmonics accuracy where the error is derived from the harmonic
calculation only. The measurement error for processed quantities such as voltages and currents,
chapter 1.3 "Analog measurement inputs" on page 3, is added to the harmonic accuracy error.
Harmonics accuracy
Calculated quantity Error Conditions
Magnitudes 0.1 % f1 = fnom
(Order 1 ... 25) (of fundamental) 15 Hz ≤ fnom ≤ 70 Hz
THD 0.2 % THDf < 100 %
OMICRON 5
DANEO 400
6 OMICRON
DANEO 400 technical data
Input signal
Input signal
deglitched
Tdeglitch Tdeglitch
Input signal
Input signal
debounced
OMICRON 7
DANEO 400
The diagram on Figure 1-3 shows the load limit curve for DC voltages. For AC voltages, a maximum
power of 2000 VA is achieved.
Figure 1-3: Load limit curve for relays on the binary outputs with DC voltages
8 OMICRON
DANEO 400 technical data
1.8 USB
OMICRON 9
DANEO 400
1.11 Cleaning
To clean DANEO 400, use a cloth dampened with isopropanol alcohol. Prior to cleaning, always unplug
the power cord from power supply and unplug all connectors so that all hazardous life parts are
disconnected and the device is switched off.
10 OMICRON
DANEO 400 technical data
OMICRON 11