MAGLUMI anti-TPO(CLIA)

1. INTENDED USE
The kit has been designed for the quantitative determination of anti-TPO in
human serum.
The method can be used for samples over the range of 0-100 IU/ml.
130203011M 100 The test has to be performed on the Maglumi fully auto analyzer (Including
Maglumi 1000, Maglumi 2000, Maglumi 2000 plus).

2. SUMMARY AND EXPLANATION OF THE TEST


Thyroid peroxidase or thyroperoxidase (TPO) is an enzyme expressed
mainly in the thyroid that liberates iodine for addition onto tyrosine residues
on thyroglobulin for the production of thyroxine (T4) or triiodothyronine (T3),
Shenzhen New Industries Lotus Global Co., Ltd thyroid hormones.[1] In humans, thyroperoxidase is encoded by the TPO
Biomedical Engineering Co., Ltd 15 Alexandra Road gene.
4F,Wearnes Tech Bldg,
London UK Inorganic iodine enters the body primarily as iodide, I-. After entering the
Science&Industry Park,
NW8 0DP thyroid follicle (or thyroid follicular cell) via a Na+/I- symporter (NIS) on the
Nanshan,Shenzhen, China 518057
Tel. + 44-20-75868010 basolateral side, iodide is shuttled across the apical membrane into the
Tel. + 86 755 26508518
Fax.+ 44-20-79006187 colloid via pendrin, after which thyroid peroxidase oxidizes iodide to atomic
Fax.+ 86 755 26508339 iodine (I) or iodinium (I+). The "organification of iodine," the incorporation of
iodine into thyroglobulin for the production of thyroid hormone, is nonspecific;
that is, there is no TPO-bound intermediate, but iodination occurs via reactive
iodine species released from TPO. The chemical reactions catalyzed by
thyroid peroxidase occur on the outer apical membrane surface and are
FOR PROFESSIONAL USE ONLY mediated by hydrogen peroxide.
Store at 2...8 °C TPO is stimulated by TSH, which upregulates gene expression. It is inhibited
by the thioamide drugs, such as propylthiouracil and methimazole.
It is a frequent epitope of autoantibodies in autoimmune thyroid disease
COMPLETELY READ THE INSTRUCTIONS (most common is Hashimoto's thyroiditis) so the antibody titer can be used to
BEFORE PROCEEDING assess disease activity in patients that have developed such antibodies.

3. PRINCIPLE OF THE TEST


Sandwich immunoluminometric assay:
en
Mouse anti-human IgG is used to label nano magnetic microbeads, and use
purified TPO antigen to coat ABEI. Sample, Calibrators, or Control, with
buffer and nano magnetic microbeads coated with TPO antigen are mixed
SYMBOLS USED ON LABELS
thoroughly and incubated at 37℃, after sediment in a magnetic field, decant
Authorized Representative in Europe the supernatant, then cycle washing for 1 time. Then add ABEI Label,
incubation and forming a sandwich, washing for the 2nd time. Subsequently,
the starter reagents are added and a flash chemiluminescent reaction is
Manufacturer initiated. The light signal is measured by a photomultiplier as RLU within 3
seconds and is proportional to the concentration of anti-TPO present in
controls or samples.
Attention. See Instructions For Use
4. KIT COMPONENTS
4.1 Material supplies

Reagent Integral for 100 determinations


Contents of kit
Nano magnetic microbeads: TRIS buffer, 1.2%(W/V),
2.5ml
0.2%NaN3, coated with TPO antigen
In vitro diagnostic medical device Calibrator, low 2.5ml
(In vitro diagnostic use) Calibrator, high 2.5ml
Buffer 12.5ml
Lot number ABEI Label: purified TPO antigen labeled ABEI
12.5ml
contains BSA, 0.2%NaN3
Diluent 25ml
Catalogue Code
All reagents are provided ready-to-use.
*Please prepare 0.9% sodium chloride solution in case of insufficient diluents.
Expiry date (Use by…)
Accessories required but not provided
Maglumi Reaction module

Temperature limitation Maglumi Starter kit 1+2


( store at 2...8 °C) Maglumi Light check
Maglumi Wash /System Liquid

Number of tests 4.2 Preparation of the Reagent Integral


Before the sealing is removed, gentle and careful horizontal shaking of the
Reagent Integral is essential (avoid foam formation!) Remove the sealing and
Keep away from sunlight turn the small wheel of the magnetic microbeads compartment to and fro
untill the colour of the suspension has changed into brown. Place the Integral
into the reagent area and let it stand there for 30 mins. During this time, the
Biological risks magnetic microbeads are automatically agitated and completely
resuspended.
Do not interchange Nano Magnetic Microbeads from different reagents!

4.3 Storage of the Reagents Integral


 Sealed:Stored at 2-8℃ until the expiry date.
130203011M-v1.0-EN 1/2
 Opened:Stable for 4 weeks. After this period, it is still possible to
keep on using the Reagent Integral provided that the controls are 11 Results
found within the expected ranges. 11.1 Calculation of Results
 Keep upright for storage.  The analyzer automatically calculates the anti-TPO concentration in
 Keep away from direct sunlight.. each sample by means of a calibration curve which is generated by a
2-point calibration master curve procedure. The results are expressed
5. Origin of Calibrators in IU/ml. For further information please refer to the Maglumi fully auto
Calibrators in the Reagent Kit are from Biodesign. Operator’s Manual.
Biological root: distilled from human serum, processed by SDS PAGE  Test results need NOT to multiply dilution rate!
purification, with a purity ≥95%. No HBsAg, anti-HCV, and anti-HIV is found.
11.2 Interpretation of Results
6. Calibration  Reference values:< 10 IU/ml.
6.1 2 point recalibration
 Results may differ between laboratories due to variations in
Via the measurement of calibrators, the predifined master curve is adjusted population and test method. Each laboratory should establish its own
(recalibrated) to a new, instrument-specific measurement level with each reference range.
calibration.
12. Limitations of the procedure
6.2 Frequency of Recalibration 12.1
 After each exchange of lot (Reagent Integral or Starter Reagents). Use anti-TPO value as a kind of auxiliary material for other testing data when
 After 4 weeks and/or each time a new Integral is used in diagnosis. Assay results should be utilized in conjunction with other clinical
(recommendation). and laboratory data to assist the clinician in making individual patient
 after each servicing of the Maglumi Fully Auto analyzer. management decisions.
 If controls are beyond the expected range.
A skillful technique and strict adherence to the instructions are necessary to
 Room temperature changes exceed 5℃(recommendation).
obtain reliable results. Bacterial contamination of samples or repeated
7. Sample Collection, Material and Storage freeze-thaw cycles may affect the test results. Assay results should be
 Collect samples using standard procedures. utilized in conjunction with other clinical and laboratory data to assist the
 Sample material: serum. clinician in making individual patient management decisions.
 Store at 2-8℃: 24 hours.
 For longer storage periods: freeze to below - 20℃. 13. Performance Characteristics
 Avoid repeated freezing and thawing cycles. 13.1 Accuracy
 Stored samples should be thoroughly mixed prior to use (Vortex Consider Calibrator high of known concentration as a sample, dilute it by 1:2
mixer). ratio with diluent, and measure its diluted concentration for 10 times. Then
 *Vacuum tubes calcaluate the recovery of measured concentration and expected
(a) Blank tubes are recommended type for collecting samples. concentration. The recovery should be within 90% -110%.
(b) If plasma sample is needed, EDTA tube is conformed has no
effect on the results RLUs. 13.2 Precision
(c) Liquaemin Sodium tube is found to increase the sample RLU and Intra-assay coefficient of variation was evaluated on Calibrator High
cause test results deviation. repeatedly measured 10 times in the same assay, calculating their coefficient
(d) Please ask SNIBE for advice if special additive must be used in of variation, the results should ≤10%.
the sample blood. Inter-assay coefficient of variation was evaluated on three batches of kit,
repeatedly measured 10 times of Calibrator High, calculating three batches of
8. WARNING AND PRECAUTIONS FOR USERS kit for Calibrator High between the measured values of the coefficients of
 For use in IN-VITRO diagnostic procedures only. variation, the results should ≤15%.
 Do not interchange reagents from different lots. Do not use kit
components beyond their labelled expiry date. 13.3 Sensitivity
 All samples, biological reagents and materials used in the assay must The sensitivity is defined as the concentration of anti-TPO equivalent to the
be considered potentially able to transmit infectious agents. They mean RLU of 20 replicates of the zero standard plus two standard deviations
should therefore be disposed of in accordance with the prevailing corresponding to the concentration from the standard curve. The sensitivity is
regulations and guidelines of the agencies holding jurisdiction over typically less than 0.38 IU/ml.
the laboratory, and the regulations of each country. Disposable
materials must be incinerated; liquid waste must be decontaminated 13.4 Specificity
with sodium hypochlorite at a final concentration of 5% for at least half The specificity of the anti-TPO assay system was assessed by measuring the
an hour. Any materials to be reused must be autoclaved using an apparent response of the assay to various potentially cross reactive analytes.
overkill approach(USP 24,2000,p.2143). A minimum of one hour at
121℃ is usually considered adequate, though the users must check 13.5 Linearity
the effectiveness of their decontamination cycle by initially validating it Conduct a logarithmic transform to the RLU value and concentration value of
and routinely using biological indicators. 6 standards. After a double logarithmic fitting, the absolute value of its
 The calibrators in this kit are prepared from bovine serum products. linearity should exceed 0.9800.
However, because no test method can offer complete assurance that
HIV, Hepatitis B Virus or other infectious agents are absent, these 14. References
reagents should be considered a potential biohazard and handled 1. Bomet H,Madec AM,Rodlen P,Latter R,Haond P,Allannlc H, Orglazzl
with the same precautions as applied to any serum or plasma J,Evaluation of Anti-TPO antlbody determlnation in Varlous cllnlcal
situations.In P.Carayon (Ed):Thyroperoxldase and Thyrold
specimen
Autolmmunlty 1990; 207: 315-320
2. Bussen SS,Steck T.Thyrold antibodles and their relation to antithrombin
9. Test Procedure antibodles,anticardlollpin antibodles and lupus anticoagulant in women
To ensure proper test performance, strictly adhere to the operating with recurrent spontaneous abortions (antithyrold,anticardlollpin and
instructions of the Maglumi Fully Auto analyzer. Each test parameter is antithrombin autoantbodles and lupus anticoagulant in habltual
indentified via a RFID tag on the Reagent Integral. For further information aborters).Eur J Obster Gynecol Reproduct Biol 1997; 74: 139-143
please refer to the Maglumi Fully Auto Operator’s Manual. 3. Doullay F,Ruf J,Carayon P,Codaccionl JL. Autoantibodles to
Thyroperoxldase in varlous thyrold and autolmmune dlseases In: P.
40μl Sample, calibrator or controls
Carayon (Ed):Thyroperoxldase and Thyrold Autolmmu--nity 1990; 207:
+100μl Buffer 285-29
+20μl Nano magnetic microbeads 4. Feldt-Rasmussen U.Analytical and cllnical performance goals for
10 min Incubation testing autoantibodles to thyroperoxldase, thyroglobulln, and
400μl each time Cycle washing thyrotropin receptor.Clin Chem 1996; 42 (1): 160-16
5. Plannenstiel P. Hotze LA, Stller B. Schlld drosen-krankheten. Dlagnose
+100μl ABEI label
and Theraple.3〃 completely revlsed edltion;Benll--ner Med. Verl. Anst.
10 min Incubation Berlin 1999
400μl each time Cycle washing 6. Rotl E. Gardinl E, Mlneiil R, Blanconl L, Braverman LE. Prev--alence of
3s Measurement anti-thyrold Peroxldase Antibodles in Serum in the Elderly:Comparison
with other Tests for Antithyrold Antibodles.Clin Chem 1992; 38 (1): 88-9
10. Quality Control 7. Scherbaum WA, Paschke R. Bedeutung der
Schilddrtisenant--senantikorper for Dlagnostlkund Verlaufsbeurteiiung
 Observe quality control guidelines for medical laboratories. von Sc- hilddrtisenerkrankungen. Internlst 1995; 36: 303-309
 Use suitable controls for in-house quality control.

130203011M-v1.0-EN 2/2