Download as pdf or txt
Download as pdf or txt
You are on page 1of 14

Ultrasound in Med. & Biol., Vol. 27, No. 4, pp.

551–564, 2001
Copyright © 2001 World Federation for Ultrasound in Medicine & Biology
Printed in the USA. All rights reserved
0301-5629/01/$–see front matter

PII: S0301-5629(01)00357-X

● Original Contribution

EFFECTS OF TRANSDUCER, VELOCITY, DOPPLER ANGLE, AND


INSTRUMENT SETTINGS ON THE ACCURACY OF COLOR DOPPLER
ULTRASOUND

SANDY F. C. STEWART
Hydrodynamics and Acoustics Branch, Center for Devices and Radiological Health, Food and Drug Administration,
Rockville, MD, USA

(Received 7 August 2000; in final form 22 January 2001)

Abstract—The accuracy of a commercial color Doppler ultrasound (US) system was assessed in vitro using a
rotating torus phantom. The phantom consisted of a thin rubber tube filled with a blood-mimicking fluid, joined
at the ends to form a torus. The torus was mounted on a disk suspended in water, and rotated at constant speeds
by a motor. The torus fluid was shown in a previous study to rotate as a solid body, so that the actual fluid velocity
was dependent only on the motor speed and sample volume radius. The fluid velocity could, thus, be easily
compared to the color Doppler-derived velocity. The effects of instrument settings, velocity and the Doppler angle
was assessed in four transducers: a 2.0-MHz phased-array transducer designed for cardiac use, a 4.0-MHz
curved-array transducer designed for general thoracic use, and two linear transducers designed for vascular use
(one 4.0 MHz and one 6.0 MHz). The color Doppler accuracy was found to be significantly dependent on the
transducer used, the pulse-repetition frequency and wall-filter frequency, the actual fluid velocity and the
Doppler angle (p < 0.001 by analysis of variance). In particular, the phased array and curved array were
observed to be significantly more accurate than the two linear arrays. The torus phantom was found to provide
a sensitive measure of color Doppler accuracy. Clinicians need to be aware of these effects when performing color
Doppler US exams. (E-mail: sxs@cdrh.fda.gov) © 2001 World Federation for Ultrasound in Medicine &
Biology.

Key Words: Color Doppler accuracy, Doppler ultrasound phantom, Color Doppler ultrasound.

INTRODUCTION assume that color Doppler systems are accurate, or may


interpret color images in a quantitative or semiquantita-
The assessment of color Doppler accuracy of clinical
tive fashion, either consciously or subconsciously, pos-
ultrasound (US) systems is an important, but neglected,
topic. Given the enormous number of exams performed sibly leading to diagnostic errors. This latter effect may
annually worldwide, color Doppler accuracy should be be more likely to show up in serial exams, where data
given more attention than it is. One of the impediments from one examination are compared against that from
to the assessment of Doppler accuracy is the lack of another. Although reproducibility is often more impor-
generally accepted Doppler accuracy phantoms. Color tant than absolute accuracy in serial exams, differences
Doppler, in particular, suffers from the lack of a suitable in color images due to variations with instrument settings
technique for assessing accuracy. Because color Doppler or transducer used may, nevertheless, be interpreted as
is generally assumed to be used qualitatively, a common having substantive diagnostic meaning.
belief may be that it doesn’t need to be assessed for Numerous methods have been previously described
accuracy. However, there are at least two arguments in for assessing the accuracy of Doppler in commercial US
favor of the development and use of color Doppler ac- systems. String phantoms have been used extensively for
curacy standards: 1. numerous quantitative uses for color calibrating pulsed Doppler (Walker et al. 1982; Phillips
Doppler US have been developed, 2. clinicians may et al. 1990). A phantom using a moving belt of foam was
developed for assessing color Doppler (Rickey et al.
Address correspondence to: Sandy F. C. Stewart, Ph.D., Hydro- 1992). A rotating rubber disk has also been used to assess
dynamics and Acoustics Branch, Food and Drug Administration, 9200
Corporate Blvd. (HFZ-132), Rockville, MD 20850 USA. E-mail:
Doppler accuracy (McDicken et al. 1983; Fleming et al.
sxs@cdrh.fda.gov 1994; Schwarz et al. 1995). However, string phantoms

551
552 Ultrasound in Medicine and Biology Volume 27, Number 4, 2001

are not particularly suitable for 2-D color imaging, be-


cause the moving string forms a stationary, narrow color
image that cannot be easily interpreted. Furthermore,
these devices use moving solids that may not produce a
realistic backscatter signal.
Phantoms that use parabolic fluid flow in a tube
(McDicken 1986) or between parallel plates (Boote and
Zagzebski 1988) have some advantages. These phantoms
use a blood-mimicking fluid instead of a solid and pro-
vide a more physiologic model of arterial flow. These
have been used to calibrate volume flow measurements
(Picot et al. 1995; Forsberg et al. 1995; Holland et al.
1996) and to derive the velocity profile (Hein and
O’Brien 1992). The associated velocity gradients are
high, however, so that the fluid velocity is a strong
function of position. The calibration of Doppler US Fig. 1. Torus phantom for assessing color Doppler accuracy.
requires precise positioning of the US sample volume,
combined with an accurate evaluation of the velocity
profile. Errors in measuring the volume flow rate, corre- effects of instrument settings, Doppler angle and velocity
lating the volume flow to the velocity profile, and ne- magnitude were also investigated.
glecting entrance length effects limit the usefulness of
tube phantoms. In addition, turbulence alters the velocity METHODS
profile, even at moderate flow rates; thus, invalidating
Torus phantom
the known relationship between velocity and radius and
The torus phantom used was described previously
limiting the useful velocity range of the phantom.
(Stewart 1999). Briefly, the phantom was made from a
An accuracy phantom designed for color Doppler
bicycle inner tube measuring 0.064 cm in thickness and
was recently described that overcomes some of the lim-
3.0 cm in diameter. The tube was filled with a blood-
itations of previously available phantoms (Stewart 1999).
mimicking fluid optimized for US studies that had an
The torus phantom moves a blood-mimicking fluid at a
acoustic velocity equal to 1548 ⫾ 5 m/s (R.G. Shelley
known rate with a low radial velocity gradient, making it
Ltd., North York, Ontario, Canada) (Rickey et al. 1995).
ideal for color Doppler tests. At a velocity of 115.2 cm/s The ends were clamped onto a short cylinder to form a
(equivalent to 40 revolutions/min), the velocity gradient torus, which was then glued onto the rim of a 52-cm
perpendicular to the plane of the color Doppler image is diameter disk (Fig. 1). The torus had a minor radius of
approximately 4.2 s⫺1 (Stewart 1999). This is substan- 3.0 cm and a major radius of approximately 28 cm. The
tially less than the velocity gradient at the wall of a 1-cm torus was submerged in degassed water with the axis
diameter tube phantom at a peak velocity of 115.2 cm/s, aligned vertically, and rotated by a gearmotor controlled
which is calculated to be 460.8 s⫺1. Spectral broadening by a digital velocity feedback system. Numerical simu-
of the Doppler signal due to the velocity gradient is, thus, lations and flow visualization studies have shown that the
minimized. The torus phantom also benefits from a high fluid within such a rotating torus spins as a solid body at
maximum velocity, around 300 cm/s in the current de- the same angular velocity as the torus, after about 5 to
sign. The phantom was shown to provide a sensitive 10 s acceleration time (Stewart 1999). The circumferen-
measure of color Doppler accuracy in tests of a commer- tial velocity of the fluid (calculated as the sample volume
cial US system, using a single transducer designed for radius times the torus angular velocity) was the standard
vascular use. Significant effects of some of the system’s against which the Doppler-derived velocities were com-
instrument settings on Doppler accuracy were reported pared. In this paper, “accuracy” is defined as the ratio of
(Stewart 1999). the Doppler-derived velocity to the actual fluid velocity.
Manufacturers of US systems typically provide a
variety of transducers optimized for different functions, Doppler ultrasound system
such as vascular, cardiac and obstetrical examinations. An ATL Ultramark 9 HDI US system was used for
The current investigation was designed to assess the all laboratory experiments. The four different transducers
effect of the choice of transducer on the color Doppler tested are listed in Table 1. The transducer under test was
accuracy. Four transducers were examined, including mounted in a four-axis positioning system (three linear
two linear arrays, a curved array, and a phased array. The and one rotational, the rotational axis being parallel to
Color Doppler accuracy ● S. F. C. STEWART 553

Table 1. Transducers used in study. varied considerably with PRF. However, an earlier study
showed that color Doppler accuracy did not vary sub-
Transmit frequency
Transducer Scanhead type (MHz) stantially with the power output, beyond a threshold level
(Stewart 1999).
P3-2 Phased array 2.0 In the second series of experiments, the study was
C5-40R Curved array 4.0
L5 Linear array 4.0 repeated with the Doppler angle set to 70°. PRF and
L10-5 Linear array 6.0 WFF settings had to be adjusted to take into account the
lower perceived velocity at this higher Doppler angle. In
the third series of experiments, the Doppler angle was set
to 55°. All the velocities were used with the L10-5
the torus major radius). The active face was suspended transducer at 55°. Only one fluid velocity was used for
approximately 1.0 to 2.0 cm above the torus, with the US the P3-2, C5-40R, and L5 transducers, corresponding to
beam focused on the center of torus fluid. The degassed an intermediate torus speed of 40 revolutions/min.
water bath in which the torus was immersed provided a The color box was centered within the torus tube
path for the US beam between the transducer and the (visible in the 2-D imaging B-mode) and positioned 4 ⫾
torus. The circumferential velocity of the moving fluid 2 cm from the transducer face. A single focal zone was
was measured by Doppler and compared to the velocity always used, and was positioned as close as possible to
calculated by the motor speed multiplied by the radial the middle of the color Doppler box. In all cases, the
position of the sample volume. The radial position was color baseline was maximally shifted downward, which
calibrated by focusing the transducer on the gap between effectively doubled the Nyquist limit and allowed a
two sound-absorbing rectangles glued to the torus higher maximum velocity to be measured.
mounting disk (Stewart 1999), which was a known dis-
tance from the center of the torus. By using the position- Color Doppler image processing and data analysis
ing system, the radial position of the sample volume Images from the US system’s color video output
within the torus could be accurately set. were captured using a Flashpoint 128 video frame grab-
ber (Integral Technologies, Indianapolis, IN) mounted in
Color Doppler experiments a personal computer running Windows 95. Images cap-
Three series of experiments were performed for tured with a resolution of 640 ⫻ 480 pixels and color
each of the four transducers. In the first series, the effects depth of 24 bits (i.e., 8 bits per color red, green and blue)
of the pulse-repetition frequency (PRF), wall-filter fre- were saved onto hard disk in Windows bitmap (BMP)
quency (WFF), and fluid velocity on Doppler accuracy image file format. A custom color/velocity map on the
were investigated at a Doppler angle of 40°. Tests were US system was used to encode velocities in the red
run at fluid velocities equal to 28.8, 57.6, 115.2, 172.8, image plane, using a linear velocity-to-color translation
230.4 and 288.0 cm/s, corresponding to torus rotational scheme. Red byte values varied from 0 to 255, with 0
speeds of 10, 20, 40, 60, 80 and 100 revolutions/min. The corresponding to zero velocity and 255 corresponding to
other color instrument settings, including color gain, the maximum velocity. Green was added to make the
output power, color sensitivity, color persistence and color maps more visible, with pixel values of 0 and 197
dynamic motion differentiation, were fixed between ex- corresponding to half the maximum velocity and the
periments (Table 2). The spatial peak temporal average maximum velocity, respectively. The green pixel values
intensity (derated), or ISPTA.3, averaged 20 mW/cm2, but were ignored when calibrating the velocity/color map or
making measurements.
After an image was acquired, the velocity-color
Table 2. ATL Ultramark 9 HDI nominal instrument settings calibration was done. A simple linear regression was
used in color Doppler accuracy tests. performed between the color and velocity, using the
pixels within the color bar. Two assumptions were made:
Instrument parameter Setting used 1. the velocity was linear with position down the color
HDI/high frequency on/on bar, and 2. the minimum velocity corresponded to black
Variance off and maximum velocity corresponded to the maximum
Output power ISPTA.3 ⫽ 20 mW/cm2, MI ⫽ 0.5 color in the color bar (so that the y-intercept was near
Color gain 94%
Color sensitivity Maximum ⫽ 16 zero). Each image was separately calibrated.
Color persistence Minimum ⫽ 0 Image files stored on disk were analyzed by an
Dynamic motion differentiationTM Off analysis program written in Borland Delphi version 3.0
ISPTA.3 ⫽ spatial peak temporal average intensity, derated; MI ⫽ (Inprise Corp., Scotts Valley, CA). After an image was
mechanical index. read into this program, the color bar was calibrated
554 Ultrasound in Medicine and Biology Volume 27, Number 4, 2001

Fig. 2. Average of 520 color Doppler images using P3-2 Fig. 4. Color Doppler image using C5-40R curved array. Green
phased-array transducer at a fluid velocity of 115.2 cm/s and line is an arc of the circle fitted to curved face of transducer.
Doppler angle of 70°. The upper and lower red lines have been The center of the circle defines the radii along which the US is
fitted to upper and lower walls of the torus tube. The middle red radiated. Straight yellow line is direction of fluid flow (equiv-
line (arrow) is the average of the two and is the assumed alent to middle red line in Fig. 2). Green sector outline delimits
direction of fluid flow. pixels to be processed. Only those pixels pi tinted in purple
have a Doppler angle ␾i ⫽ ␾e ⫾ 1°, where ␾e ⫽ 70° in this
case. Fluid velocity ⫽ 115.2 cm/s.

against the maximum velocity corresponding to the PRF.


The velocities were then corrected for the Doppler angle
given transducer, Doppler angle and velocity were aver-
␾ between the fluid velocity and the transducer beam, by
aged, and a line was fitted to the grey-scale images of the
dividing the velocity by cos(␾).
averaged top tube wall (Fig. 2) as follows. A narrow
In most cases, the transducers were set at an oblique
rectangle (yellow box in Fig. 2) was drawn manually
angle to the torus tube (as in Fig. 2), so that the direction
around the image of the walls to tell the program which
of the fluid velocity had to be determined from the
pixels to include. Then linear regression was performed
B-mode image of the tube walls. All the images for a
between the x and y positions of each pixel within the
yellow rectangle, with a weighting factor equal to the
intensity of the pixel at that point. This gave a straight
line that was the best fit to the tube wall. The procedure

Fig. 3. Color Doppler image using P3-2 phased-array trans-


ducer. Yellow cross labeled “O” at the top indicates center
defined by radii of ultrasound beam. Straight yellow line is
direction of fluid flow (equivalent to middle red line in Fig. 2).
Green sector outline delimits pixels to be processed. Only those Fig. 5. Color Doppler image using L5 linear array. Direction of
pixels pi tinted in purple have a Doppler angle ␾i ⫽ ␾e ⫾ 1°, fluid flow is horizontal. Doppler angle ⫽ 70°. Fluid velocity ⫽
where ␾e ⫽ 70° in this case. Fluid velocity ⫽ 115.2 cm/s. 115.2 cm/s.
Color Doppler accuracy ● S. F. C. STEWART 555

was then repeated for the bottom wall. The top and
bottom red lines in Fig. 2 are the fitted lines, and the
middle red line (arrow) has the average slope and inter-
cept of the top and bottom line, and is assumed to be the
direction of the fluid velocity.
The curved-array and phased-array transducers
emitted the US beam along radii of a circle, so that the
Doppler angle ␾ was a function of position (an effect that
is readily visible in Fig. 3). For the phased-array trans-
ducer, the center C defined by the US radii was marked
manually by a yellow cross (labeled “O” at the top of
Fig. 3). For the curved array, the center C was found by
fitting a circle to the arc indicating the transducer surface
(Fig. 4). The fitted circle is marked in green in Fig. 4; in
this case the calculated center C is beyond the visible part
of the image. In Figs. 3 and 4, the yellow line between Fig. 6. Pulsed Doppler spectrum for P3-2 transducer at a
cross marks is the center of the tube (and estimated Doppler angle of 40°. Fluid velocity ⫽ 115.2 cm/s. ROI is
outlined by rectangular box.
direction of the fluid velocity) found by the curve fitting
procedure described above (Fig. 2). For the curved- and
phased-array transducers, the beam steering angle was
always set within the range of 5 to 20. was drawn within the color box, and the color-encoded
After the center of radii was found in the phased- velocities within the parallelogram were divided by
array and curved-array transducers, a sector of a circle cos(␾) and averaged.
was drawn within the color box (defined by two radii and The Doppler-derived velocity was also dependent
two arcs, seen as green lines in Figs. 3 and 4). The on the variable angle between the plane of the US beam
direction of the US beam for each pixel pi within the and the torus itself, due to the mild curve of the torus
green box was determined by first computing the line major radius. However, this effect was assumed to be
between pi and the center C. The Doppler angle ␾i for negligible (Gill 1985; Stewart 1999).
pixel pi was then calculated from the angle between the
line defining the US beam for that pixel and the line
defining the fluid velocity. Only those pixels pi with Pulsed Doppler comparisons
angles ␾i ⫽ ␾e ⫾ 1 were analyzed further, where ␾e ⫽ Measurements were also taken using the P3-2 and
angle for a particular experiment (␾e ⫽ 40°, 55° or 70°). L10-5 transducers in pulsed Doppler mode, to provide a
The velocity at pi was divided by cos(␾i), and the cor- comparison to the color Doppler accuracy. The Doppler
rected velocities in the selected range of angles were then angle was set to 40° or 70°, using the same transducer tilt
averaged. In Figs. 3 and 4, the locus of points with 69.0° and beam steering angles as were used in the color
ⱕ ␾i ⱕ 71.0° are indicated by the purple tint. Doppler experiments. The ATL’s internal angle correc-
For the linear arrays, the Doppler angle ␾ was tion was used. The pulsed Doppler sample volume was
constant over the entire color image, because the US set to 2.5 mm, and the velocity, PRF and WFF were
beam was assumed to be emitted along parallel lines varied as in the color Doppler studies. The output and
(Fig. 5). The direction of the US beam was determined gain were set to avoid saturating the image. Spectral
from the angled side of the color box. When testing the images were downloaded to the personal computer and
linear arrays at 70°, the transducers were placed perpen- the mean velocity was calculated off-line. The velocity
dicular to the torus wall, and the US beam was steered 20 was first calibrated to the vertical direction. Then, at each
from vertical (as in Fig. 5). The direction of the fluid point in time ti in the spectrum, a mean velocity vi was
velocity, in this case, was assumed to be horizontal. In calculated, using the pixel intensity as a weight factor.
the case of the other two Doppler angles (55° and 40°), Finally an overall mean (⫾ SD) velocity Vm was calcu-
the beam was steered 20 from vertical, and the probe was lated by averaging the vi over time. A typical spectrum
then tilted an extra 15° or 30° with respect to the torus for the P3-2 transducer at a Doppler angle of 40° and
tube. In these latter two cases, the direction of fluid flow velocity of 115.2 cm/s is shown in Fig. 6. The region-
was calculated by fitting lines to the walls of the torus of-interest (ROI) is outlined by the rectangular box. The
tube in an averaged image (as in Fig. 2). The Doppler spikes are reflections from the short cylinder onto which
angle ␾ between the fluid velocity and the direction of the ends of the torus tube are clamped, and were ex-
the US beam was calculated. A graphic parallelogram cluded from the velocity measurements.
556 Ultrasound in Medicine and Biology Volume 27, Number 4, 2001

Fig. 7. Color Doppler-derived velocity as a function of transducer at a Doppler angle of 40°. (a) P3-2 phased array; (b)
L5 linear array; (c) C5-40R curved array; (d) L10-5 linear array. Fluid velocity ⫽ 115.2 cm/s. Effects of PRF and WFF
can also be seen.

Statistical analysis A sample size of 10 images were analyzed for each


The color Doppler accuracy was evaluated statisti- combination of transducer, velocity, Doppler angle, WFF
cally by analysis of variance, using a general linear and PRF. Statistical analysis was performed on a per-
model approach that automatically accounted for the sonal computer using Minitab release 12 (Minitab, State
unbalanced experimental design. The dependent variable College, PA).
was the accuracy (i.e., the ratio of the color Doppler-
derived velocity to the actual fluid velocity). The inde-
RESULTS
pendent variables were the transducer, Doppler angle,
PRF, WFF and actual fluid velocity. All independent Effects of transducer
variables had fixed levels, so no covariates were used. The color Doppler-derived velocity as a function of
Interaction terms among certain of the independent vari- transducer at a fluid velocity of 115.2 cm/s and a Doppler
ables were included in the model: (transducer ⫻ Doppler angle of 40° is shown in Fig. 7. The P3-2 phased-array
angle), (transducer ⫻ WFF), (transducer ⫻ actual fluid and C5-40R curved-array transducers had average accu-
velocity), (Doppler angle ⫻ WFF), and (Doppler an- racies (0.911 ⫾ 0.066 and 0.905 ⫾ 0.044, respectively;
gle ⫻ actual fluid velocity). Higher interaction terms and mean ⫾ SD) that were higher than that of the two linear
interaction terms involving PRF could not be used, be- arrays (0.842 ⫾ 0.046 and 0.828 ⫾ 0.065 for the L5 and
cause of disconnects within the data set (e.g., not all L10-5 transducer, respectively) as indicated in Table 3.
levels of PRF could be used with all levels of the other Differences in the range of PRF used reflect the different
variables). Bonferroni pair-wise comparisons were made transmission frequencies among the four transducers
among the levels of the five independent variables to used (Table 1).
determine the significance of differences among levels. The color Doppler-derived velocity as a function of
Color Doppler accuracy ● S. F. C. STEWART 557

transducer at the same velocity, but at 70° is shown in

0.915 ⫾ 0.136

0.907 ⫾ 0.079

0.808 ⫾ 0.077

0.826 ⫾ 0.092
All velocities
and angles
Fig. 8. Again, the P3-2 phased-array and C5-40R curved-
array transducers had higher average accuracies
(0.911 ⫾ 0.116 and 0.905 ⫾ 0.093, respectively) than the
two linear arrays (0.771 ⫾ 0.068 and 0.817 ⫾ 0.078,
respectively, Table 3).
The effect of transducer was statistically significant,

0.915 ⫾ 0.086
0.940 ⫾ 0.059
0.912 ⫾ 0.169
0.904 ⫾ 0.058
0.860 ⫾ 0.091
0.918 ⫾ 0.085
0.849 ⫾ 0.062
0.812 ⫾ 0.064
0.763 ⫾ 0.055
0.829 ⫾ 0.072
0.838 ⫾ 0.073
0.815 ⫾ 0.111
All velocities
at p ⬍ 0.001. The Bonferroni pair-wise comparisons
showed that each transducer was significantly different
from all the others, with an adjusted p ⬍ 0.0001.

Effects of Doppler angle

0.976 ⫾ 0.120

0.760 ⫾ 0.094

0.961 ⫾ 0.068
0.892 ⫾ 0.039

0.768 ⫾ 0.053

0.871 ⫾ 0.029
0.845 ⫾ 0.052
The effect of the Doppler angle is shown for the

288.0
P3-2 phased-array and the L10-5 linear-array transducer



at 115.2 cm/s in Fig. 9. For the phased-array transducer,
the average accuracies at 40°, 55° and 70° were equal to
0.911 ⫾ 0.066, 0.940 ⫾ 0.059, and 0.911 ⫾ 0.116,
respectively. For the L10-5 linear array, the average

0.972 ⫾ 0.054

0.864 ⫾ 0.105

0.938 ⫾ 0.065
0.852 ⫾ 0.030

0.773 ⫾ 0.060
0.873 ⫾ 0.041
0.822 ⫾ 0.092
0.835 ⫾ 0.092
Table 3. Summary of color Doppler accuracy tests.
accuracies at 40°, 55° and 70° were equal to 0.828 ⫾

230.4
0.065, 0.842 ⫾ 0.074, and 0.817 ⫾ 0.078, respectively.



The other two transducers showed effects of the Doppler
angle that were similar in magnitude (Table 3).
Note that, at 40°, the angled configuration of the
transducer occasionally did not allow the sample volume

0.922 ⫾ 0.061

0.906 ⫾ 0.111
0.918 ⫾ 0.043

0.897 ⫾ 0.069
0.828 ⫾ 0.060

0.782 ⫾ 0.059
0.856 ⫾ 0.070
0.847 ⫾ 0.052
0.840 ⫾ 0.062
depth to be short enough for a PRF high enough to
172.8
measure the higher velocities without aliasing (Table 3).


Fluid velocity (cm/s)

Therefore, some of the higher velocities were missing in


the C5-40R and L10-5 transducer experiments at 40°.
The effect of Doppler angle on the overall accuracy
was statistically significant, at p ⬍ 0.001, although there
0.911 ⫾ 0.066
0.940 ⫾ 0.059
0.911 ⫾ 0.116
0.905 ⫾ 0.044
0.860 ⫾ 0.091
0.905 ⫾ 0.093
0.842 ⫾ 0.046
0.812 ⫾ 0.064
0.771 ⫾ 0.068
0.828 ⫾ 0.065
0.842 ⫾ 0.074
0.817 ⫾ 0.078

* Color Doppler-derived velocity divided by the actual fluid velocity (mean ⫾ SD).
were no clear trends. An increase in the effect of PRF
115.2

and WFF with higher angles is evident, however (Fig. 9).


The pair-wise comparisons showed that each Doppler

PRF could not be set high enough to measure velocity without aliasing.
angle was significantly different from all the others, with
an adjusted p ⬍ 0.0001.
0.910 ⫾ 0.060

0.960 ⫾ 0.157
0.901 ⫾ 0.068

0.933 ⫾ 0.072
0.852 ⫾ 0.074

0.796 ⫾ 0.078
0.814 ⫾ 0.065
0.835 ⫾ 0.069
0.792 ⫾ 0.101

Effects of PRF and WFF


57.6

The color Doppler-derived velocity as a function of


velocity, PRF and WFF for the P3-2 phased-array trans-
ducer at a Doppler angle of 40° is shown in Fig. 10. At
low PRFs, there was a tendency for the Doppler-derived
velocity to decrease with increase in WFF and, at high
0.858 ⫾ 0.103*

0.857 ⫾ 0.067
1.005 ⫾ 0.271
0.899 ⫾ 0.060

0.896 ⫾ 0.107

0.763 ⫾ 0.110
0.831 ⫾ 0.081
0.832 ⫾ 0.082
0.801 ⫾ 0.177

PRFs, the tendency was reversed. At an angle of 70°


28.8

(Fig. 11), the increase in Doppler-derived velocity with


WFF was enhanced at high PRF, but the decrease in
Doppler-derived velocity with WFF was suppressed at
low PRF. Overall, the Doppler accuracy was found to be
0.915 ⫾ 0.086 and 0.912 ⫾ 0.169 at Doppler angles of
Doppler
angle, °

40
55
70
40
55
70
40
55
70
40
55
70

40° and 70°, respectively (Table 3). Combined effects of


PRF and WFF were observed in all transducers, and have
been reported previously (Stewart 1999).
Transducer

The effect of WFF was statistically significant (p ⬍


C5-40R

L10-5

0.001). Pair-wise comparisons showed that each level of


P3-2

L-5

WFF was statistically significantly different from all of


558 Ultrasound in Medicine and Biology Volume 27, Number 4, 2001

Fig. 8. Color Doppler-derived velocity as a function of transducer at a Doppler angle of 70°. (a) P3-2 phased array; (b)
L5 linear array; (c) C5-40R curved array; (d) L10-5 linear array. Fluid velocity ⫽ 115.2 cm/s. Effects of PRF and WFF
can also be seen.

the others with an adjusted p ⬍ 0.0001, except for the 50 wise comparisons showed that each level of velocity was
Hz vs. 100 Hz and 50 Hz vs. 200 Hz comparisons, which significantly different from all of the others, with an
were not significant. adjusted p ⬍ 0.0001.
The effect of PRF was also statistically significant
(p ⬍ 0.001). The pair-wise comparisons showed that Effects of interactions
most pairs of PRF were significantly different at p ⬍ The analysis of variance showed that not only were
0.0001, but some of the close pairs at the extremes of the the effects of all the independent variables significant at
range (e.g., 0.4 kHz vs. 0.6 kHz at the low end, and 14 p ⬍ 0.001, but the five tested interaction terms were also
kHz vs. 16 kHz at the high end) were not significantly significant at this same p level. Thus, the effect of trans-
different from each other. The only close pairs in the ducer was itself significantly dependent on the levels of
middle of the range that were not significantly different Doppler angle, WFF and actual fluid velocity. The effect
were the 4.0 kHz vs. 4.5 kHz and 7.0 kHz vs. 8.0 kHz of Doppler angle was itself significantly dependent on
comparisons. the levels of WFF and actual fluid velocity. More com-
plex interactions could not be investigated due to limi-
Effects of velocity tations in the data set connectivity.
For the P3-2 transducer, the accuracy could be seen
to increase with increasing velocity at 40°, and decrease Pulsed Doppler results
with increasing velocity at 70° (Table 3). This effect of The pulsed Doppler-derived velocity as a function
velocity was not a clear trend because it was not repeated of velocity, PRF and WFF for the P3-2 transducer at a
for the other transducers. Nevertheless, the effect of Doppler angle of 40° is shown in Fig. 12. The accuracy
velocity was statistically significant, at p ⬍ 0.001. Pair- for pulsed Doppler at this Doppler angle averaged
Color Doppler accuracy ● S. F. C. STEWART 559

Fig. 9. Color Doppler-derived velocity as a function of Doppler angle: (a) P3-2 phased array, 40°; (b) P3-2 phased array,
55°; (c) P3-2 phased array, 70°; (d) L10-5 linear array, 40°; (e) L10-5 linear array, 55°; (f) L10-5 linear array, 70°. Fluid
velocity ⫽ 115.2 cm/s.

1.023 ⫾ 0.032, which was closer to 1.0 than the corre- trends emerge. First, some transducers are markedly
sponding accuracy (0.911 ⫾ 0.066) for color Doppler more accurate than others. Although the effects of the
(Fig. 10). At a Doppler angle of 70°, the pulsed Doppler other variables mask the differences somewhat, the pair-
accuracies averaged 1.012 ⫾ 0.116 for the P3-2 trans- wise comparisons clearly showed the differences among
ducer. For the L10-5 transducer, the accuracies averaged pairs of transducers to be significant. Second, the effect
0.952 ⫾ 0.089 and 0.920 ⫾ 0.057 for Doppler angles of of WFF and PRF formed a distinctive pattern. In general,
40° and 70°, respectively. In all cases, the effects of PRF as the PRF increased, the estimated velocity first de-
and WFF were lower in the pulsed Doppler than in the creased, then increased with WFF. Third, the variations
color Doppler case. in accuracy with velocity and Doppler angle were also
significant, but with much less of an observable trend.
DISCUSSION The general linear models approach used in this
The measured color Doppler accuracy has been study can accommodate an unbalanced experimental de-
shown to be dependent on all of the parameters varied in sign. This approach was necessary because not all values
this study: the transducer, Doppler angle, WFF, PRF and of PRF and WFF were available at all velocities, Doppler
velocity. In every case, the main effects were statistically angle and transducer. In some cases, aliasing prevented
significant, at a confidence level of p ⬍ 0.001. Three measurements from being taken at higher velocities. This
560 Ultrasound in Medicine and Biology Volume 27, Number 4, 2001

Fig. 10. Color Doppler-derived velocity as a function of velocity, PRF and WFF for the P3-2 phased-array transducer
at a Doppler angle of 40°: (a) 28.8 cm/s; (b) 57.6 cm/s; (c) 115.2 cm/s; (d) 172.8 cm/s; (e) 230.4 cm/s; (f) 288.0 cm/s.

flexibility was exploited to reduce the number of mea- timator, due to difficulty in filtering a Doppler signal
surements required by about 25%. Examination of the composed of only a few data points (Rickey and
L10-5 transducer data (taken at all three Doppler angles) Fenster 1996; Willemetz et al. 1989). The underesti-
suggested that the 55° data could safely be omitted for mation has been shown to be less when using pulsed
most of the velocities in the other three transducers. Doppler (Fig. 12), so is probably not due to errors of
Table 3 shows that, for the other transducers, experi- the torus phantom mechanism itself. An assessment of
ments were performed over the full velocity range at 70° the torus phantom error demonstrated it to be accurate
and 40°, but only at a single intermediate velocity at 55°. to within ⫾ 1.7% (Stewart 1999).
The missing velocities were, thus, bracketed on either The two linear-array transducers were found to be
side by velocities at 40° and 70°. Any missing data were significantly less accurate than the phased-array and
accounted for by the model, and the resulting p values curved-array transducers. This may be due to the
were automatically adjusted. Statistically significant ef- higher focusing in the linear arrays, which would
fects were demonstrated despite the unbalanced design. increase the Doppler bandwidth and make them more
susceptible to aliasing; thus, reducing the mean
Effects of transducer estimated frequency and lowering the measured veloc-
In color Doppler mode, all four transducers sig- ity.
nificantly underestimated the actual fluid velocity. Variations in the beam spread may be due to dif-
This may result from bias in the mean-frequency es- ferences of ⫾ 2 cm in the range of interest (position of
Color Doppler accuracy ● S. F. C. STEWART 561

Fig. 11. Color Doppler-derived velocity as a function of velocity, PRF and WFF for the P3-2 phased-array transducer
at a Doppler angle of 70°: (a) 28.8 cm/s; (b) 57.6 cm/s; (c) 115.2 cm/s; (d) 172.8 cm/s; (e) 230.4 cm/s; (f) 288.0 cm/s.

the color box) or focal zone, resulting from the necessity Doppler spectrum is near zero (e.g., right end of Fig. 10a,
of preventing steric interference between the transducer high PRF setting), the wall filter removes some of the
and torus tube. Some of the observed variation in accu- lower frequencies, thus biasing the mean frequency es-
racy may be due to this effect. timate upward. Increasing the WFF removes more of the
low frequencies, so the bias upward increases. This
Effects of PRF and WFF mechanism explains most of the interactions between the
The effects of the PRF and WFF can be explained PRF and WFF (Rickey and Fenster 1996; Willemetz et
by a linear signal analysis of the Doppler spectrum. If the al. 1989; Tysoe and Evans 1995). The magnitude of this
color baseline is centered, the wall filter acts as a notch effect was observed to be higher for color Doppler than
filter centered at zero. In this study, however, the color for pulsed Doppler (Fig. 10 vs. Fig. 12), most likely
baseline was maximally shifted, so that wall filter acts because of greater spectral broadening in the former.
like a broad bandpass filter dropping off at 0 kHz and at Velocity gradients cause spectral broadening of the
the PRF. When the mean of the Doppler spectrum is near Doppler signal, which makes the mean velocity estimate
the PRF (e.g., left end of Fig. 10a, low PRF setting), the more sensitive to the wall filter. This effect is also
filter removes some of the higher frequencies, thus bias- dependent on the beam width, and may be a contributing
ing the mean frequency estimate downward. Increasing factor in the differences seen among transducers. How-
the WFF removes more of the higher frequencies, thus ever, the velocity gradient in the torus phantom is small,
increasing the bias downward. When the mean of the on the order of 4.2 s⫺1 for a velocity of 115.2 cm/s, so
562 Ultrasound in Medicine and Biology Volume 27, Number 4, 2001

Fig. 12. Pulsed Doppler-derived velocity as a function of velocity, PRF and WFF for the P3-2 phased array transducer
at a Doppler angle of 40°: (a) 28.8 cm/s; (b) 57.6 cm/s; (c) 115.2 cm/s; (d) 172.8 cm/s; (e) 230.4 cm/s; (f) 288.0 cm/s.

that effects of the velocity gradient are considered to be effect of WFF at the highest PRF is much less in the 40°
minimal. Some of the variation in the effect of the wall case than at an equivalent PRF in the 70° case. This may be
filter on the accuracy may also be due to differences in because the Doppler spectrum (when shifted by the cosine
the amount of beam spread with differences in the range of the Doppler angle) is broader at a high angle than at a low
of interest and focal position. angle.
In the linear arrays, the beam was steered to a
Effects of velocity and Doppler angle constant 70°, regardless of Doppler angle, so there were
It is speculated that the effects of velocity and Doppler no differences in accuracy merely due to differences in
angle are for the most part secondary effects of the PRF, beam asymmetry with steering angle. In the phased and
because the usable range of PRF was highly dependent on curved arrays, the amount of steering varied somewhat
these two variables. Thus, a high velocity at a low Doppler (between 5° and 20°), so that a small amount of the
angle would be expected to have a similar effect as a lower observed variation in accuracy may be attributed to dif-
velocity at a higher Doppler angle; as far as the US system ferences in the beam asymmetry.
is concerned, the two are equivalent. However, the velocity
distribution would not necessarily be the same at the two Comparison with pulsed Doppler results
Doppler angles, even if the perceived mean velocity is the Pulsed Doppler was found to be substantially more
same. This effect can be seen in Figs. 10a and 11a. The accurate than color Doppler for the transducers tested.
Color Doppler accuracy ● S. F. C. STEWART 563

Most likely this is due to color Doppler’s mean fre- 3. Color Doppler accuracy was also found to be statis-
quency estimator having to work with fewer points (Wil- tically significantly dependent on the pulse-repetition
lemetz et al. 1989). The subsequent broader spectrum is frequency and wall-filter frequency (p ⬍ 0.001).
more susceptible to aliasing of frequency components, These two settings were observed to depend heavily
which contribute to a lower mean frequency estimate. on each other. In an earlier study (Stewart 1999),
This biases the velocity estimate downward to a greater other instrument settings, such as output power, color
extent than is seen in the pulsed Doppler case. This effect gain and box width, were observed to have much less
also explains the decreased effect of the wall filter in effect on color Doppler accuracy.
pulsed Doppler. The relatively narrow spectrum of 4. The effects of the Doppler angle and the magnitude
pulsed Doppler is less susceptible to components being of the actual fluid velocity were also found to be
filtered out than that of color Doppler, so that the effect statistically significant at p ⬍ 0.001. However, this
of the wall filter on the mean frequency is less. Never- may be an effect of changes in PRF and WFF,
theless, the L10-5 linear-array transducer was still less required to keep the velocity in the proper range
accurate than the P3-2 phased-array transducer in pulsed when either the actual velocity was changed, or the
Doppler mode, possibly due to its higher degree of perceived velocity was changed due to changes in
focusing and concomitant broader bandwidth. the Doppler angle.
5. The interactions among the independent variables
Clinical relevance tested complicated the effects considerably.
The clinical relevance of these effects is difficult to 6. Clinicians need be aware of these effects whether
assess in detail, but general conclusions can be made. interpreting color Doppler quantitatively or intu-
Quantitative use of color Doppler lags behind that of itively. These effects are especially important when
other US modes such as pulsed- and continuous-wave evaluating serial exams for changes in the character of
Doppler. However, it is possible that even when color blood flow.
Doppler is used qualitatively, the clinician may make a
conscious or unconscious correlation between color and Acknowledgements— The expert machining assistance of James Duff
velocity. If so, errors may creep in. and Bruce Fleharty in building the phantom and making many excellent
suggestions is gratefully acknowledged. The assistance of Dr. Gerald
The overall underestimation of the true velocity by R. Harris with the optical motor controller is also appreciated. The
the various transducers may not be as much of a problem assistance of Dr. Hector Lopez with the theoretical interpretations is
as the variations in accuracy from one exam to the next. also acknowledged. The mention of commercial products, their source,
or their use in connection with material reported herein is not to be
The P3-2 phased-array transducer will be preferred for construed as either an actual or implied endorsement of such products
cardiac exams, so that errors due to variations with by the Department of Health and Human Services.
transducer choice will be minimized. However, in eval-
uating vascular disease, if the L10-5 is used for one
REFERENCES
exam, and the L5 used for the next, errors in interpreta-
tion may result. In any case, minor changes in the wall Boote EJ, Zagzebski JA. Performance tests of Doppler ultrasound
equipment with a tissue and blood-mimicking phantom. J Ultra-
filter or PRF between serial exams may mimic changes sound Med 1988;7:137–147.
in the measured velocity where none exist or, conversely, Fleming AD, McDicken WN, Sutherland GR, Hoskins PR. Assessment
mask a significant clinical change. The extent to which of colour Doppler tissue imaging using test-phantoms. Ultrasound
Med Biol 1994;20:937–951.
this occurs on a day-to-day basis, and the significance to Forsberg F, Liu JB, Russell KM, Guthrie SL, Goldberg BB. Volume
clinical diagnosis, are unknown. flow estimation using time domain correlation and ultrasonic flow-
metry. Ultrasound Med Biol 1995;21:1037–1045.
SUMMARY Gill RW. Measurement of blood flow by ultrasound: Accuracy and
sources of error. Ultrasound Med Biol 1985;11:625– 641.
The conclusions of this study can be summarized as Hein IA, O’Brien WD. A flexible blood flow phantom capable of
independently producing constant and pulsatile flow with a predict-
follows: able spatial flow profile for ultrasound flow measurement valida-
1. The torus phantom was found to be a useful tool for tions. IEEE Trans Biomed Engr 1992;39:1111–1122.
detailed assessments of color Doppler accuracy over a Holland CK, Clancy MJ, Taylor KJW, Alderman JL, Purushothaman
K, McCauley TR. Volumetric flow estimation in vivo and in vitro
wide range of velocity, Doppler angle and instrument using pulsed-Doppler ultrasound. Ultrasound Med Biol 1996;22:
settings. 591– 603.
2. The four transducers tested gave estimated velocities McDicken WN. A versatile test-object for the calibration of ultrasonic
Doppler flow instruments. Ultrasound Med Biol 1986;12:245–249.
averaging 90% or less of the real velocity. The dif- McDicken WN, Morrison DC, Smith DSA. A moving tissue-equivalent
ferences among the transducers was statistically sig- phantom for ultrasound real-time scanning and Doppler techniques.
nificant at p ⬍ 0.001, with the accuracy of the phased Ultrasound Med Biol 1983;9:L455–L459.
Phillips DJ, Hossack J, Beach KW, Strandness DE Jr. Testing ultra-
array and curved array significantly greater than that sonic pulsed Doppler instruments with a physiologic string phan-
of the two linear array transducers. tom. J Ultrasound Med 1990;9:429 – 436.
564 Ultrasound in Medicine and Biology Volume 27, Number 4, 2001

Picot PA, Fruitman M, Rankin RN, Fenster A. Rapid volume flow rate of integrated backscatter as a measure of scattered ultrasound
estimation using transverse colour Doppler imaging. Ultrasound intensity? Ultrasound Med Biol 1995;21:231–242.
Med Biol 1995;21:1199 –1209. Stewart SFC. A rotating torus phantom for assessing color Doppler
Rickey DW, Fenster A. A Doppler ultrasound clutter phantom. Ultra- accuracy. Ultrasound Med Biol 1999;25:1251–1264.
sound Med Biol 1996;22:747–766. Tysoe C, Evans DH. Bias in mean frequency estimation of Doppler
Rickey DW, Picot PA, Christopher DA, Fenster A. A wall-less vessel signals due to wall clutter filters. Ultrasound Med Biol 1995;21:
phantom for Doppler ultrasound studies. Ultrasound Med Biol 671– 677.
1995;21:1163–1176. Walker AR, Phillips DJ, Powers JE. Evaluating Doppler devices using
Rickey DW, Rankin R, Fenster A. A velocity evaluation phantom for a moving string test target. J Clin Ultrasound 1982;10:25–30.
colour and pulsed Doppler instruments. Ultrasound Med Biol 1992; Willemetz JC, Nowicki A, Meister JJ, De Palma F, Pante G. Bias and
18:479 – 494. variance in the estimate of the Doppler frequency induced by a wall
Schwarz KQ, Bezante GP, Chen X. When can Doppler be used in place motion filter. Ultrason Imag 1989;11:215–225.

You might also like