Thermo Forma 3940 - Rev 19 - 29 Cu. Ft. Capacity Environmental Chamber Models 3940 and 3949 - Operati-2

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 38

Models 3940 and 3949

29 cu. ft. capacity


Environmental Chamber

Operating and Maintenance Manual


Manual No: 7033940 Rev. 19
Models 3940 and 3949 ___________________________________________________________________________

Read This Instruction Manual.

Failure to read, understand and follow the instructions in


this manual may result in damage to the unit, injury to operat-
ing personnel, and poor equipment performance.

CAUTION! All internal adjustments and maintenance must


be performed by qualified service personnel.

Refer to the serial tag on the back of this manual.

The material in this manual is for information purposes


only. The contents and the product it describes are subject to
change without notice. Thermo Forma makes no representations
or warranties with respect to this manual. In no event shall
Thermo Forma be held liable for any damages, direct or inciden-
tal, arising out of or related to the use of this manual.

MANUAL NUMBER 7033940


-- 21172/IN-3067 1/30/03 Refer to description below ccs
19 21228/IN-3065 12/31/02 Reduced noise on optional CO2 control (electrical schematics) ccs
18 21173/IN-3057 11/15/02 Added ground shield to temperature and humidity probe ccs
17 20479/IN-3013 10/7/02 Added 1900140 IR CO2 option as an accessory aks
16 20474/IN-3013 3/6/02 Added optional heatless dryer aks
15 20342/PIP-012 10/3/01 Revised electrical schematics aks
14 19942 6/8/01 Replaced timer 275030 with 410048 aks
13 19935 5/8/01 Updated 3940-71 drawing aks
12 IN-2858 3/13/01 Revised control panel access and revised electrical schematics aks
11 19617 12/21/00 New door (105063) - updated electrical schematics ccs
REV ECR/ECN DATE DESCRIPTION By

i
Models 3940 and 3949 ______________________________________________________________________________Safety

Important operating and/or maintenance instructions. Read the accompanying text carefully.
Ce symbole attire l'attention de l'utilisateur sur des instructions importantes de fonctionnement et/ou d'entretien. Il
peut être utilisé seul ou avec d'autres symboles de sécurité. Lire attentivement le texte d'accompagnement.
Wichtige Betriebs- und/oder Wartungshinweise. Lesen Sie den nachfolgenden Text sorgfältig.
Importante instruccions de operacion y/o mantenimiento. Lea el texto acompanante cuidadosamente.

Potential electrical hazards. Only qualified persons should perform procedures associated with this symbol.
Ce symbole attire l'attention de l'utilisateur sur des risques électriques potentiels. Seules des personnes qualifiées
doivent appliquer les instructions et les procédures associées à ce symbole.
Gefahr von Stromschlägen. Nur qualifizierte Personen sollten die Tätigkeiten ausführen, die mit diesem Symbol beze-
ichnet sind.
Potencial de riesgos electricos. Solo personas das capacitadadas deben ejecutar los procedimientos asociadas con este
simbulo.

Equipment being maintained or serviced must be turned off and locked off to prevent possible injury.
Risques potentiels liés à l'énergie. L'équipement en entretien ou en maintenance doit être éteint et mis sous clé pour
éviter des blessures possibles.
Geräte, an denen Wartungs- oder Servicearbeiten durchgeführt werden, müssen abgeschaltet und abgeschlossen wer-
den, um Verletzungen zu vermeiden.
El equipo recibiendo servicio o mantenimiento debe ser apagado y segurado para prevenir danos.

Hot surface(s) present which may cause burns to unprotected skin, or to materials which may be damaged by elevated
temperatures.
Présence de surface(s) chaude(s) pouvant causer des brûlures sur la peau non protégée, ou sur des matières pouvant
être endommagées par des températures élevées.
Heiße Oberfläche(n) können ungeschützter Haut Verbrennungen zufügen oder Schäden an Materialien verursachen,
die nicht hitzebeständig sind.
Superficias calientes que pueden causar quemaduras a piel sin proteccion o a materiales que pueden estar danados
por elevadas temperaturas.

√ Always use the proper protective equipment (clothing, gloves, goggles, etc.)
√ Always dissipate extreme cold or heat and wear protective clothing.
√ Always follow good hygiene practices.
√ Each individual is responsible for his or her own safety.

ii
Models 3940 and 3949 _________________________________________________________________________Service

iii
Models 3940 and 3949 ___________________________________________________________________Table of Contents

Table of Contents
a. Checking the Steam Generator Safety Thermostat
Section 1 - Receiving . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 3
1.1 Preliminary Inspection . . . . . . . . . . . . . . . . . . . . . . .1 - 1
1.2 Visible Loss or Damage . . . . . . . . . . . . . . . . . . . . . .1 - 1 Section 6 - Specifications . . . . . . . . . . . . . . . . . . . . . . . . .6 - 1
1.3 Responsibility for Shipping Damage . . . . . . . . . . . .1 - 1
Section 7 - Spare Parts . . . . . . . . . . . . . . . . . . . . . . . . . . .7 - 1
Section 2 - Installation and Set-Up . . . . . . . . . . . . . . . . .2 - 1
2.1 Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1 Section 8 - Refrigeration Schematics . . . . . . . . . . . . . . . .8 - 1
2.2 Water (Humidity System) and Drain Connections . .2 - 1
a. Connecting Water Inlet for Humidity System . . .2 - 1 Section 9 - Electrical Schematics . . . . . . . . . . . . . . . . . . .9 - 1
b. Alternate Water Supply for Humidity System . . .2 - 1
c. Connecting the Drain Line . . . . . . . . . . . . . . . . . .2 - 1
2.3 Remote Data Output . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 2
a. RS-232 Data Output . . . . . . . . . . . . . . . . . . . . . . .2 - 2
b. Remote Alarm Contacts . . . . . . . . . . . . . . . . . . . .2 - 2
2.4 Power Connection . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 2
2.5 Start-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 2
2.6 Setting the Overtemp Safety Thermostat . . . . . . . . .2 - 2
2.7 Setting the Undertemp Safety Thermostat . . . . . . . .2 - 2
2.8 Preparing the (Optional) CoBex Recorder . . . . . . . .2 - 3
a. Installing the Battery . . . . . . . . . . . . . . . . . . . . . . .2 - 3
b. Changing the Chart Paper . . . . . . . . . . . . . . . . . . .2 - 3
c. Changing the Pen . . . . . . . . . . . . . . . . . . . . . . . . .2 - 3
d. Calibrating the Chart Recorder . . . . . . . . . . . . . . .2 - 3
2.9 Offset Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 4
2.10 Honeywell Recorder (optional) . . . . . . . . . . . . . . . .2 - 4
2.11 Controller Configuration . . . . . . . . . . . . . . . . . . . . .2 - 4
2.12 1900140 IR CO2 Option . . . . . . . . . . . . . . . . . . . . . .2 - 5

Section 3 - Start Up and Operation . . . . . . . . . . . . . . . . .3 - 1


3.1 The Control Panel . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
3.2 Setting Incubator Operating Temperature . . . . . . . . . .3 - 3
3.3 Setting Incubator Operating Humidity . . . . . . . . . . . .3 - 3
3.4 Programming the controllers . . . . . . . . . . . . . . . . . . .3 - 3
a. Removing the Mechanical Lockout . . . . . . . . . . .3 - 3
b. Removing the Software Lockout . . . . . . . . . . . . .3 - 3
3.5 Air Exchange Ventilator Caps . . . . . . . . . . . . . . . . . .4 - 1
3.6 Heatless Dryer (optional) . . . . . . . . . . . . . . . . . . . . . . .4 - 1

Section 4 - Routine Maintenance . . . . . . . . . . . . . . . . . . .4 - 1


4.1 Cleaning the Incubator . . . . . . . . . . . . . . . . . . . . . . . .4 - 1

Section 5 - Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
5.1 Accessing the Electrical Components . . . . . . . . . . . .5 - 1
5.2 Replacing Over/Undertemp Probe & Thermostat . . .5 - 1
5.3 Replacing the Humidity/Temperature Sensor . . . . . .5 - 1
5.4 Replacing the Optional Recorder and Probe(s) . . . .5 - 1
5.5 Removing the Top Section . . . . . . . . . . . . . . . . . . . . .5 - 2
5.6 Setting the Door Heater Control . . . . . . . . . . . . . . . . .5 - 2
5.7 Cleaning and Adjusting Humidity Steam Generator . .5 - 3

iv
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up

Section 1 - Receiving

1.1 Preliminary Inspection 1.3 Responsibility for Shipping Damage

This incubator was thoroughly inspected and carefully For products shipped F.O.B. Marietta, Ohio, the responsi-
packed prior to shipment and all necessary precautions were bility of Thermo Forma ends when the merchandise is loaded
taken to ensure its safe arrival. Immediately upon receipt, onto the carrier’s vehicle.
before the incubator is moved from the receiving area, carefully On F.O.B. Destination shipments, Thermo Forma and the
examine the shipment for loss or damage. Unpack the incubator carrier’s responsibility ends when your Receiving Department
and inspect its interior and exterior for any in-transit damage. personnel sign a free and clear delivery receipt.
Whenever possible, Thermo Forma will assist in settling
claims for loss or in-transit damage.
1.2 Visible Loss or Damage

If any loss or damage is discovered, note any discrepancies


on the delivery receipt. Failure to adequately describe such evi-
dence of loss or damage may result in the carrier refusing to
honor a damage claim. Immediately call the delivering carrier
and request that their representative perform an inspection. Do
not discard any of the packing material and under no circum-
stances move the shipment from the receiving area.

1-1
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up

Section 2 - Installation and Set-Up The water inlet is the 1/8” FPT connection located on the
rear top left corner of the incubator. Water inlet pressure must
not exceed 40 PSI. a manual shut-off valve should be installed
between the main water supply and the incubator.
2.1 Location

Locate the unit on a firm, level surface in an area of mini- To prevent mineral buildup on heater coils and
mum ambient temperature fluctuation. A minimum of twelve humidity generator walls, it may be necessary to
(12) inches of clearance is required at the top of the incubator clean the humidity generator and immersion
and a minimum of three (3) inches of clearance is required at heaters with a non-metallic abrasive pad and
the rear of the incubator. This space is necessary to allow ade- flush thoroughly every two to three months. Refer
quate air flow around the refrigeration system. to Section 5.7, Cleaning and Adjusting the
Humidity Steam Generator.

2.2 Water (Humidity System) and Drain Connections


b. Alternate Water Supply for Humidity System
a. Connecting Water Inlet for Humidity System
If an in-house water supply of the required purity range
(50K to 1M Ohm) is not available, an alternate water supply
Do not attempt to operate the humidity system method can be used. A large vented carboy (5 gal. minimum) of
without filling it with water. The humidification water in the required purity range can be placed on top of the
heater will overheat, and the overtemp safety ther- unit. Connect it to the 1/8” FPT water inlet fitting, located on
mostat will shut the humidification system down. the rear top left corner of the incubator.

The humidity reservoir will require approximately one IMPORTANT: The maximum water consumption of this incu-
quart (0.946 liter) of water on the initial filling. For best opera- bator could be as high as 2 gallons per day. When using an
tion of the incubator, sterilized distilled, demineralized or de- alternate water supply method, it is imperative that the humidity
ionized water should be used in the humidity reservoir. Water reservoir not be allowed to run dry, as permanent damage to the
purity should be in the resistance range of 50K Ohm/cm to 1M steam generator will result.
Ohm/cm, or a conductivity range of 20.0 uS/cm to 1.0 uS/cm.
Refer to ASTM Standard D5391-93 or D4195-88 for measuring c. Connecting the Drain Line
water purity.
Distillation systems, as well as some types of reverse The cabinet’s 3/8” MPT drain line connection is located on
osmosis water purity systems, can produce water in the quality the rear (lower left side) of the cabinet. A P-trap (Figure 2-1) is
range specified. Tap water is not recommended as it may con- included with the unit and must be installed on the connection.
tain chlorine, which can deteriorate the stainless steel. Tap To install the drain connection:
water may also have a high mineral content, which would pro-
1. Using Teflon pipe thread tape, tape the threads on the
duce a build-up of scale in the reservoir. High purity, ultra pure
cabinet drain connection.
or milli-q water is not recommended as it is an extremely
aggressive solvent and will deteriorate the stainless steel. High 2. Using an open end adjustable wrench, install the P-trap
purity water has a resistance of above 1M Ohm to 18M Ohm. onto the connection. Make sure that the trap is posi-
Even high purity water can contain bacteria and organic con- tioned down.
taminants. Water should always be sterilized or treated with a 3. Push a piece of 3/8” ID tubing onto the trap and direct
decontaminant, safe for use with stainless steel as well as safe the tubing to a convenient drain. Install a hose clamp on
for the product, prior to being introduced into the humidity the tubing, if desired.
reservoir.

Distilled or de-ionized water used in the humidity


reservoir must be within a water quality resistance
range of 50K Ohm/cm to 1M Ohm/cm to protect Figure 2-1
and prolong the life of the stainless steel. Use of
water outside the specified range will decrease the Cabinet drain P-trap
operating life of the unit and may void warranty.

2-1
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up

2.3 Remote Data Output 2.5 Start-Up

a. RS-232 Data Output When the humidification system is operational, the incuba-
tor may be started. Preset the controls as follows:
The Model 3940/3949 incubator is equipped with an RS- Overtemp Safety Thermostat . . . . . . . . . . . . . Fully Clockwise
232 Serial Communication Interface for the remote transmis- Undertemp Safety Thermostat . . . . . . Fully Counterclockwise
sion of temperature and humidity data. A terminal strip is locat- Main Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ON
ed on the back of the incubator for convenience. Refer to Humidity Controller . . . . . . . . . . . . . . . . . . . .Desired Setpoint
Figure 2-2 for terminal pin identification. Temperature Controller . . . . . . . . . . . . . . . . . .Desired Setpoint
Door Heater . . . . . . . . . . . . . . . . . . . . . . . . . .40% (factory set)

For best overall performance of the incubator, the refrigera-


tion switch should be turned on for most applications. When
running Low or No humidity at high temperatures, the refriger-
ation switch may be turned off.

The defrost switch must be turned on when the


temperature setpoint is 10°C, or below.

Figure 2-2

b. Remote Alarm Contacts 2.6 Setting the Overtemp Safety Thermostat

Remote alarm connections are also included on the termi-


Allow the chamber temperature and humidity to stabilize
nal strip (Figure 2-2) providing Normally Open (N.O.) and
then set the overtemp safety thermostat as follows:
Normally Closed (N.C.) contacts. C is the Common terminal.
The remote alarm will activate when either the incubator’s tem- 1. Turn the overtemp control knob slowly counterclock-
perature or humidity go out of the set alarm limits. wise until the audible alarm sounds and the overtemp
indicator lights.
2. Turn the overtemp control knob clockwise approximate-
ly two units on the scale. The alarm should be silenced
2.4 Power Connection
and the overtemp indicator light should go out. The
overtemp safety thermostat is now set a few degrees
See the serial tag on the side of the unit for electrical spec-
above the control temperature setpoint. When the cham-
ifications, or refer to the electrical schematics at the end of this
ber temperature rises to the overtemp control point, the
manual.
alarm system will activate, power to the heaters will shut
off, and the chamber temperature will be maintained at
Connect the incubator to a grounded, dedicated the overtemp control point.
circuit. The power cord connector is the mains When an overtemp condition occurs, the cause must be
disconnect device for the incubator. Position the determined and corrected before normal operation under the
incubator so the unit can be easily disconnected. main temperature controller can be resumed.
Note: When the chamber temperature control setpoint is
For Model 3940, plug the provided 10 ft. power cord with changed, the overtemp safety thermostat must be reset to
the NEMA 14-20 plug into the power outlet connection on the accommodate the change.
back of the cabinet, then into the grounded dedicated electrical
circuit.
For Model 3949, the electrical junction box is located on
the rear top section of the incubator. With the junction box
switch OFF, connect the incubator to an adequate power source.

2-2
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up

If the battery is weak or not connected, the green LED will


2.7 Setting the Undertemp Safety Thermostat flash. If power is lost to the cabinet, the LED will also flash.
When replacing the 9-volt battery, use only an alkaline style
Allow the chamber temperature and humidity to stabilize, battery. Dispose of the old battery following established envi-
then set the undertemp safety thermostat as follows: ronmental practices.
1. Turn the undertemp control knob slowly clockwise until
the audible alarm sounds and the undertemp indicator b. Changing the Chart Paper
lights.
2. Turn the undertemp control knob counterclockwise 1. Press the Change Chart button (#3) and hold it for 1 sec-
approximately two units on the scale. The alarm should ond until the pen begins to move to the left of the chart.
be silenced and the undertemp indicator light should go See Figure 2-3.
out. 2. Remove the existing chart by unscrewing the center
The undertemp safety thermostat is now set a few degrees knob securing it.
below the control temperature setpoint. When the chamber tem- 3. Install the new chart, positioning it so that the correct
perature drops to the undertemp control point, the alarm system time line coincides with the time line groove on the
will activate, power to the compressor will shut off, and the chart plate.
chamber temperature will be maintained at the undertemp con- 4. Replace the center knob and screw it tightly against the
trol point. chart.

When an undertemp condition occurs, the cause must be c. Changing the Pen
determined and corrected before normal operation under the
main temperature controller can be resumed. 1. Using a small flat blade screwdriver, loosen the 2 screws
Note: When the chamber temperature control setpoint is holding the pen arm and remove the pen and arm as an
changed, the undertemp safety thermostat must be reset to assembly.
accommodate the change. 2. Unsnap the plastic hinge securing the pen. Remove and
discard the old pen.
3. Install the new pen by snapping the hinge securely
2.8 Preparing the (Optional) CoBex Recorder
around the pen arm.
a. Installing the Battery 4. Re-install the pen assembly by sliding the pen arm under
the screws, positioning the pen tip in the time line
The seven-day circular chart recorder is located on the groove. Tighten the screws.
front of the incubator cabinet and is protected by a glass door. 5. Push the Chart Change button and hold it for 1 second
until the pen begins to move back onto the chart.
Note: Make sure that the pen is marking on the chart. It may
be necessary to gently lift the pen onto the chart paper.

d. Calibrating the Chart Recorder

Place an accurate thermometer(s)


in the chamber next to the recorder’s
probe(s). After about three minutes,
compare the thermometer with the chart
recorder. For 2 pen operations, also
compare the second thermometer.
Note: For 2 pen operations, first select
the pen you wish to calibrate. Hold
Figure 2-4
Figure 2-3 down the #1 arrow for the red (#1) pen
or the #2 arrow for the blue (#2) pen, until the light goes out.
To prepare the recorder for operation, open the glass door Refer to Figure 2-4. Then adjust as necessary.
and snap the connector onto the 9-volt battery as shown in If an adjustment is necessary, press either the #1 or #2 but-
Figure 2-3. If the unit is operating, the green LED lights steady. ton to move the pen left or right. The button must be held
If the unit is not turned on, the LED blinks. about five seconds before the pen begins to move. Release the
2-3 button when the pen matches the thermometer.
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up

2.9 Offset Calibration


To turn the hardware lockout back on:
It may be necessary to calibrate the temperature or humidi- 1. Press the Up and Down Arrow keys simultaneously for
ty controllers to match an independent temperature or humidity 3 seconds. The word “InPt” will appear in the upper dis-
sensor. To do so, follow the next few steps. play, and “set” will appear in the lower display.
1. Suspend an independent, calibrated sensor(s) in the cen- 2. Press the Down Arrow once, then continue to press the
ter of the interior chamber. Mode key until “LOC” appears in the lower display. The
2. Allow approximately 30 minutes for the incubator to upper display will show 0. Press the Up Arrow until 2
stabilize. appears.
3. Turn off the main power switch. 3. Press the Display key once.
4. Wear a grounding wrist strap to avoid damaging any of 4. Turn off the main power switch.
the electrical components. 5. Wear a grounding wrist strap to avoid damaging any of
the electrical components.
6. Remove the 982 controller module(s) by pressing in the
four retaining tabs, two on the right side, two on the left
side. Pull the controller module out by gently rocking it
Figure 2-5 from side to side.
7. Looking at the top and right side of the controller mod-
5. Remove the 982 controller module(s) by pressing in the ule, locate the red DIP switches. Use your fingernail or
four retaining tabs, two on the right side, two on the left a small screwdriver, to turn on SW 2 by moving the
side. Refer to Figure 2-5. Pull the controller module out white toggle towards the back of the module.
by gently rocking it from side to side.
8. Return the controller into its frame and press firmly on
6. Looking at the top and right side of the controller mod- the top and bottom of the bezel until all four locking
ule, locate the red DIP switches indicated in Figure 2-6. tabs “click” into place.
Use your fingernail
or a small screw- 9. Turn on the main power switch.
driver, to turn off
SW 2 by moving 2.10 Honeywell Recorder (optional)
the white toggle
towards the front of The Honeywell, DR 4500 Recorder is a one to four-chan-
the module as nel microprocessor-based, circular chart recorder.
shown in the illus- The recorder is capable of recording both temperature and
tration. humidity and printing alphanumeric chart data on blank heat-
7. Return the con- sensitive chart. Refer to the Honeywell Configuration at the
troller into its frame Figure 2-6 end of this section and the supplemental Honeywell Recorder
and firmly press on Manual.
the top and bottom of the bezel until all four locking
tabs “click” into place. 2.11 Controller Configuration
8. Turn on the main power switch.
9. Press the Up and Down Arrow keys simultaneously for The Watlow 982 Temperature and Humidity
3 seconds. The word “InPt” will appear in the upper dis- Controllers have been configured at the factory. Copies of the
play, and “set” will appear in the lower display. Watlow Configuration records are included on the following
10. Press the Down Arrow once, then continue to press the pages.
Mode key until “LOC” appears in the lower display. The
upper display will show 2. Press the Down Arrow until 0
The controller should not be re-configured without first con-
appears.
sulting Thermo Forma Service department at 1-888-213-1790.
11. Press the Mode key once, then the Up Arrow once.
“InPt” will appear in the upper display, and “set” will be Refer to Section 3 of this manual for operating instructions
in the lower display. Press the Mode key until “CAL 1” for the Watlow 982 controllers. The Watlow factory manuals
appears in the lower display. Press the Up or Down
are included with this manual.
Arrow key to either add or subtract an offset value. This
would be the difference between the actual value shown
on the controller, and the reference sensor value.
12. Press the Display key once. 2-4
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up
2.12 1900140 IR CO2 Option c. CO2 Controller Calibration
This section applies to units with the 1900140 IR CO2 If it should become necessary to calibrate the CO2 con-
option only. troller, perform the following procedure. From the standard
operating display (setpoint in bottom display, actual CO2 reading
a. Connecting the CO2 Source (Refer to Figure 2-3) in the upper display):
Enter the Setup Menu by pressing the t up-arrow and
For the most economical use, the liquid CO2 supply tanks
6down-arrow keys
should be without siphon simultaneously on the
tubes, so that only CO2 gas Watlow 93 controller for
enters the incubator injection three seconds. The
system. Two tanks may be lower display shows the
joined together with a mani- Lock parameter, and the
fold to ensure a continuous upper display shows its
CO2 supply. current level. Press
Install a two-stage pres- the6 down-arrow until
sure regulator, with indicating 0 is displayed. Then,
gauges, at the supply cylinder use the Advance key to
outlet. The high-pressure Figure 2-8
move through the menus
gauge should have an indicat- until the "CAL" prompt appears.
ing range of 0 to 2000 psig to
monitor tank pressure. The Once at the "CAL" prompt, press thet up-arrow or 6
low-pressure gauge should down-arrow to add or subtract the percentage of CO2 to make
have an indicating range of 0 Figure 2-7 the display accurate. Press the Advance key until LOC appears.
to 30 psig to monitor input Press the t up-arrow key to place "2" in the upper display.
pressure to the incubator injection system. A suitable two-stage Continue to press the Advance key until the setpoint display
pressure regulator is available. appears.
The CO2 source must be regulated at a pressure level
between 5 and 10 psig. Higher pressure levels may damage the d. CO2 Control and Indicators
CO2 control system. The user should determine the most eco-
nomical pressure level, between 5 and 10 psig appropriate for Sample Port (Refer to Figure 2-9)
the desired CO2 percentage in the chamber. Use only sufficient The sample port is used for checking CO2 percentage in the
pressure to maintain recovery time after door openings. incubator chamber by an independent test instrument (such as
with a Fyrite, or similar CO2 test instrument).
To connect the CO2 supply:
· Insert the copper tubing provided with the unit as far as it To prevent CO2 loss, the sample port must be capped when it is
will go into the nut of the CO2 connection. not in use.
· Turn the nut until it is finger tight.
CO2 Alarm (Refer to Figure 2-9)
· For reference, scribe the nut at the 6:00 position.
The CO2 alarm is factory set to activate when the
· While holding the fitting body steady with a wrench, tighten chamber CO2 content
the nut 1-1/4 turns until the mark is at the 9:00 position. deviates ± 5%. When a
· Securely attach the CO2 line to the open end of the copper CO2 alarm occurs, the
tubing. CO2 Alarm indicator on
· Check the tubing connection for leaks. the control panel lights
and the audible alarm
b. Setting the CO2 Content sounds.
The Watlow 93 CO2 controller's upper display shows The CO2
the actual CO2 content inside the chamber. The lower display alarm high and low set-
shows the CO2 setpoint. points are established
through the Watlow 93
Before setting the CO2 Content, allow the chamber tempera-
CO2 controller (ALO,
ture and humidity to stabilize. Do not open door during the sta-
and AHI). Refer to the
bilization period.
configuration record
To set the CO2 content, press the up or down triangles included at the end of
on the Watlow 93 Controller. Section 2.
Figure 2-9
2-5
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up

3940 (WATLOW 982) CONFIGURATION RECORD


CUSTOMER:
JOB NUMBER:
UNT SERIAL NUMBER:
CONTROL TYPE: Temperature
PREPARED BY: DATE
COMPLETED BY: DATE

Switch Configuration:
Main Boards Input 1 Board Output 3 Option board
SW1 ON SW1 OFF Jumper switch setting:
SW2 ON SW2 OFF Form B
SW3 ON

Software Configuration:
Operations Menus
SYS: Ei1S NA Ent4 NA A3LO -73.3 A4Hi NA
Ei2S NA A2LO NA A3Hi 34.0 Aut OFF
Ent3 NA A2Hi NA A4LO NA

PID: Pb1 2.0 dE1 NA It2 NA dB 0.0


rE1 0.50 Ct1 5 rA2 0.25
It1 NA Pb2 1.0 dE2 NA
rA1 0.19 rE2 0.25 Ct2 30

PROG:See Programming Sheets if required.

Setup Menus
InPt: In1 RT.D CAL1 ** rL2 NA CAL2 NA
dEC1 NA rtd1 DIN rH2 NA Hunt NA
rL1 -20.0 Ftr1 2 LrnL NA
rHi 80.0 In2 NA LrnH NA

OtPt: Ot1 Ht AL2 NA LAt3 nLA SIL4 NA


PrC1 NA LAt2 NA SIL3 OFF Aout NA
HyS1 0.1 SIL2 NA Ot4 NA PrC3 NA
Ot2 CL Ot3 AL3 AL4 NA ArL NA
PrC2 NA AL3 Pr1 HyS4 NA ArH NA
HyS2 0.1 HyS3 0.1 LAt4 NA ACAL NA

gLbL: C_F C Anun OFF PtyP ti PStr StPt


Err nLA LoP -100 gSd 0 LOC 2 *
Ei1 NO HiP 100 POut Cont
Ei2 NA AtSP 90 IdSP NA
* LOC should be set at 0 until factory testing and calibration is complete.
** This is a calibration factor and will vary from unit to unit.
COM: bAUd 9600 dAtA 70 PrOt FULL Addr 0
IntF NA

2-6
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up

3940 (WATLOW 982) CONFIGURATION RECORD


CUSTOMER:
JOB NUMBER:
UNT SERIAL NUMBER:
CONTROL TYPE: Humidity
PERPARED BY: DATE:
COMPLETED BY: DATE:

Switch Configuration:
Main Boards Input 1 Board Output 3 Option board
SW1 ON SW1 ON Jumper switch setting:
SW2 ON SW2 ON Form B
SW3 ON

Software Configuration:
Operations Menus
SYS: Ei1S NA Ent4 NA A3LO 0* A4Hi NA
Ei2S NA A2LO NA A3Hi 100* Aut OFF
Ent3 NA A2Hi NA A4LO NA
* These values are customer setable to there alarm points.
PID: Pb1 8 dE1 NA It2 NA dB 0
rE1 0.05 Ct1 5.0 rA2 0.35
It1 NA Pb2 7 dE2 NA
rA1 0.35 rE2 0.03 Ct2 30

PROG:See Programming Sheets if required.

Setup Menus
InPt: In1 0-5 CAL1 ** rL2 NA CAL2 NA
dEC1 0 rtd1 NA rH2 NA Hunt NA
rL1 0 Ftr1 2 LrnL NA
rHi 100 In2 NA LrnH NA

OtPt: Ot1 Ht AL2 NA LAt3 nLA SIL4 NA


PrC1 NA LAt2 NA SIL3 ON Aout NA
HyS1 1 SIL2 NA Ot4 NA PrC3 NA
Ot2 CL Ot3 AL3 AL4 NA ArL NA
PrC2 NA AL3 Pr1 HyS4 NA ArH NA
HyS2 1 HyS3 1 LAt4 NA ACAL NA

gLbL: C_F NA Anun ON PtyP ti PStr StPt


Err nLA LoP -100 gSd 0 LOC 2 *
Ei1 NO HiP 100 POut Cont
Ei2 NA AtSP 90 IdSP NA
* LOC should be set at 0 until factory testing and calibration is complete.
** This is a calibration factor and will vary from unit to unit.
COM: bAUd 9600 dAtA 70 PrOt FULL Addr 0
IntF NA

2-7
Models 3940 and 3949 ________________________________________________________________Installation and Set-Up

WATLOW 93 CONFIGURATION RECORD

CUSTOMER:
JOB NUMBER:
UNT SERIAL NUMBER:
CONTROL TYPE: CO2
PERPARED BY: DATE
COMPLETED BY: DATE:

Operational Parameters:

Pb1 2.5 It2


rE1 0.05 rA2
It1 dE2
rA1 0 Ct2
dE1 ALO -5.0
Ct1 10 AHI 5.0
Pb2 CAL *
rE2 AUt 0

Setup Parameters:

LOC 2 HSA 0.1


In 4-20 LAt NLA
dEC 0.0 SIL OFF
C_F rtd
rL 0.0 rP off
rH 20.0 rT
Ot1 ht PL 100
HSC 0.1 dSP nor
Ot2 de

Notes: * calibration offset may vary from unit to unit

Important: Input switches 1&2 must be set to "on"

2-8
Models 3940 and 3949_________________________________________________________________Start-Up and Operation

Section 3 - Start Up and Operation Overtemp Safety Control, Indicator Light, and Audible
Alarm (Figure 3-2)
The overtemp safety thermostat should be set slightly
above the operating temperature of the incubator. In the event
of an overtemp condition, the overtemp safety thermostat will:
• Activate the audible alarm and the overtemp indicator
light.
• Interrupt power to the heaters and maintain the incubator’s
cabinet temperature at the overtemp safety control point.
Note: The overtemp control is not directly calibrated. The
Figure 3-1, Reach-In Incubator Control Panel numbers (0 to 10) on the dial are for reference only and do not
shown with optional CoBex Recorder correspond to any specific temperature.
If an overtemp condition occurs, the alarm can only be
silenced by raising the overtemp safety thermostat setting.
3.1 The Control Panel However, the cause of the problem must be determined and cor-
rected before normal operation under the main temperature con-
Main Power Switch and Indicator Light (Figure 3-2) troller is resumed.

A The The main power switch controls power to the incubator.


main power indicator lights when the power switch is
on and the unit is receiving power.

Refrigeration Switch and Indicator Light (Figure 3-2)

B ationThe refrigeration switch controls power to the refriger-


system. The refrigeration indicator lights when the
refrigeration switch is on and the compressor is receiving
power.

Defrost Switch and Indicator Light (Figure 3-2)

C tem.The defrost switch controls power to the defrost sys-


The defrost timer is factory set to provide two 15-
minute defrost cycles during a twenty-four hour period.
The defrost indicator lights when the defrost switch is on
and the incubator is in a defrost cycle.

The defrost switch must be turned on when the


temperature setpoint is 10°C, or below.

Figure 3-2 3 -1
Models 3940 and 3949_________________________________________________________________Start-Up and Operation

Undertemp Safety Control, Indicator Light and Audible Audible Humidity Alarm and Indicator Display (Figure 3-3)
Alarm (Figure 3-2) The humidity alarm is a function of the humidity con-
The undertemp safety thermostat should be set I troller, Figure 3-4. When the cabinet humidity goes outside
E F slightly lower than the operating temperature of the
incubator. In the event of an undertemp condition, the
the set parameters of the controller, L3 indicator on the
controller lights, the audible alarm sounds, and the humidity
undertemp safety thermostat will: alarm indicator (Figure 3-3, Item I) on the control panel lights.
• Activate the audible alarm and the undertemp indicator
light. Hold/Run Key
• Interrupt power to the refrigeration system and maintain Pressing the Hold/Run key (Figure 3-4) on the con-
the incubator’s cabinet temperature at the undertemp safety
control point.
J troller silences the audible alarm and extinguishes the
humidity alarm indicator and the L3 indicator on the con-
Note: As with the overtemp safety control, the undertemp con- troller. A3HI/A3LO flashes on the lower display until the cabi-
trol is not directly calibrated. The numbers (0 to 10) on the dial net humidity returns to the system setpoint.
are for reference only and do not correspond to any specific Note: The humidity controller’s high
temperature. and low limits are set by the factory at
If an undertemp condition occurs, the alarm can only be 100% and 0%. Therefore, the system
silenced by lowering the undertemp safety thermostat setting. will go into the alarm state when the
However, the cause of the problem must be determined and cor- humidity exceeds these percentages by
rected before normal operation under the main temperature con- one percent. When operating the incu-
troller is resumed. bator near these high or low humidity
levels, frequent alarms may occur.
This will require that the controller’s
Humidify Switch and Indicator (Figure 3-3) high or low limit be reset to three or
four percent over the high limit or three
G powerTheto humidify switch controls the
the humidification system cir- or four percent under the low limit.
cuit. The humidity indicator light will Refer to the Watlow User’s guide pro-
cycle as the controller toggles between humid- vided. Refer also to the factory configura- Figure 3-4
ify and dehumidify. tion records located at the end of Section
2 of this manual.

Dehumidify Switch and Indicator (Figure


3-3)
The dehumidify switch will enable
H the refrigeration system if the refrigera-
tion switch is currently off. When con-
trolling humidity, the dehumidification switch
should be in the ON position for most appli- Figure 3-3
cations. The dehumidification light will cycle
on and off as the humidity controller toggles between humidify
and dehumidify. The dehumidify light will be de-energized
when the system is in Defrost mode.
Humidity is removed from the cabinet by the refrigeration
evaporator coil. Moisture accumulates on the coil and is collect-
ed into the evaporator pan. The accumulated moisture is then
plumbed to the back of the cabinet, which can be emptied to a
floor drain, or an evaporative device.

3 -2
Models 3940 and 3949_________________________________________________________________Start-Up and Operation

3.2 Setting the Incubator’s Operating Temperature a. Removing the mechanical lockout
(Figure 3-5)
Make sure the unit is turned
KL The Watlow temperature
controller’s upper numerical
off.
Wearing a grounding wrist
display shows the actual temperature strap or maintaining constant con-
inside the incubator chamber. The tact with the metal cabinet, press
lower display shows the temperature in the four locking tabs on the
setpoint. frame of the Temperature con-
troller. There are two tabs on
Changing the Setpoint either side as shown in the front
and side views in Figure 3-7.
M point,Topress
raise or lower the set-
the up or down
When all tabs are unlocked, pull
the controller module out of its frame. Figure 3-7
arrow. Temperatures are set in .1°C
increments. Figure 3-5 Looking at the top of the module, locate the red DIP switch
indicated in Figure 3-8. With a fingernail or small screwdriver,
turn off SW #2 by moving the
white toggle down or towards
3.3 Setting the Incubator’s Operating Humidity
the front of the module. (SW#2
(Figure 3-6)
is the top switch when looking
at the module from the right
N O controller’s
The Watlow humidity
upper numerical side.)
display shows the actual humidity inside Very carefully, replace the
the incubator. The lower display shows module into its frame and press
the humidity setpoint. firmly on the top and bottom of
the bezel until all four locking
P To raise the setpoint, press the
up or down arrows. Humidity is set tabs “click” into place.
in one percent increments. Figure 3-8
b. Removing the Software Lockout
Figure 3-6 Press the up arrow and down arrow keys at the same time
3.4 Programming the controllers
and hold them for about three seconds. The words “inpt”
The Watlow temperature and humidity controllers have (input) and “set” will appear in the top and bottom displays.
been set at the factory to operate the incubator within the speci- (Figure 3-9) If numbers in the bottom display begin to scroll up
fications listed in Section 6 of this manual. Reference copies of or down, the keys have not been pressed simultaneously. Try
the Watlow configuration records are included at the end of again.
Section 2 of this manual. Press the up arrow until “gLbL” (global) appears in the
To prevent tampering, mechanical and software lockouts upper display. The word “set” will remain in the lower display.
are employed in the system. These lockouts must only be (Refer to Figure 3-10)
removed by persons skilled in configuring controller software.

Making program changes to either the tempera-


ture or humidity controllers will seriously alter the
performance of the incubator and therefore must
be made only by qualified persons. The con-
trollers should not be re-configured without first
consulting Thermo Forma Service Department, at
1-888-213-1790.

Figure 3-9 Figure 3-10


3 -3
Models 3940 and 3949_________________________________________________________________Start-Up and Operation

Press the Mode key to scroll through the menus until the Section 4 - Routine Maintenance
upper and lower displays show “2 Loc” (Figure 3-11). Press the
down arrow key until the 2 becomes 0 (zero) as shown in
Figure 3-12. 4.1 Cleaning the Incubator

De-energize all potential sources of energy to this


unit and lockout/tagout their controls. (O.S.H.A.
Regulation, Section 1910-147.)

The continued cleanliness of the stainless steel used in


Forma products has a direct effect on the appearance and opera-
tion of the unit. Use the mildest cleaning procedure that will do
the job effectively. Clean the outside of the incubator with soap
and water or with any non-abrasive commercial spray cleaner.
Clean the inside of the chamber with alcohol and/or soap and
water. Disinfect the interior panels with a general use laboratory
disinfectant, diluted according to the manufacturer’s instruc-
tions. rinse the surface thoroughly after each cleaning and wipe
Figure 3-11 Figure 3-12
the surfaces dry. always rub in the direction of the finish polish
lines.
Press the Display key to return to showing the setpoint and
actual temperatures. All safeguards are now removed and the Do not use chlorinated solvents on stainless steel
controller may be programmed. Refer to the Watlow publica- as they can cause rusting and pitting.
tions included with this operating manual.

Reprogramming the temperature or humidity Do not use volatile or aromatic solvents for
controllers will change the factory default cleaning inside the cabinet as their residue can
settings and may seriously alter the performance contaminate the cabinet environment.
of the incubator and void the warranty. Contact
the Thermo Forma Service Department, at 1-888-
213-1790.
The Thermopane glass door may be cleaned with commer-
cial glass cleaner or with a solution of ammonia and water.
3.5 Air Exchange Ventilator Caps

Air exchange for the incubator is regulated through the


manually adjustable intake and exhaust ventilator caps located
on the top of the cabinet.
When viewed from the front of the incubator, the intake
cap is on the left and the exhaust cap is on the right. The venti-
lator caps may be opened by turning counterclockwise, and
closed by turning clockwise.
For optimum performance of the unit, the vent caps should
be closed at all times.

3.6 Heatless Dryer P/N 1900139 (optional)

The optional heatless dryer provides dehumidification for


the incubator chamber. The dehumidify switch must be turned
on for the dryer to operate. The dryer is controlled with the
humidity controller.

4-1
Models 3940 and 3949_________________________________________________________________________________________________________Preventive Maintenance

PREVENTIVE MAINTENANCE
Incubators
Your equipment has been thoroughly tested and calibrated before shipment. Regular preventive maintenance is important to keep your unit function-
ing properly. The operator should perform routine cleaning and maintenance on a regular basis. For maximum performance and efficiency, it is rec-
ommended that the unit be checked and calibrated periodically by a qualified service technician.

The following is a condensed list of preventive maintenance requirements. See the specified section of the instruction manual for further details.

Thermo Forma has qualified service technicians, using NIST traceable instruments, available in many areas. For more information on Preventive
Maintenance or Extended Warranties, please contact us at the number below.

Cleaning and calibration adjustment intervals are dependent upon use, environmental conditions and accuracy required.

Tips for all incubators:


• Do NOT use bleach or any disinfectant that has high chloros.
• Use sterile, distilled or demineralized wqater.
• Avoid spraying cleaner on the CO2 sensor.
• Do not use powdered gloves for tissue cultures.

•Millcreek Road, Box 649 •Marietta, Ohio 45750 USA •740-373-4763


•USA and Canada 888-213-1790 •Telefax: 740-373-4189 •http://www.service@thermoforma.com

4-2
Models 3940 and 3949_______________________________________________________________________________________________________Preventive Maintenance

Preventive Maintenance for Model 3940/3949


Refer to Manual Section Action Daily Weekly Yearly

-- Inspect door latch, hinges and door gasket seal ;


3.5 Check air exchange ventilator caps for adjustment; open or close as required ;
4.1 Perform a complete decontamination procedure. Wipe down interior, shelves, Between Experiments
side panels with disinfectant. Rinse everything well with sterile distilled water. More frequent decontamination may be
required, depending on use and
environmental conditions.
2.9 * Verify and document all calibrations, at the minimum. ;
5.7 * Inspect and clean the humidity generator, at the minimum. ;
-- Clean drip pan and drain lines ;
-- Clean refrigeration system condenser ;
-- Verify defrost cycle for below 10°C operation ;
-- Change filters(under normal conditions) ;
* Qualified service technicians only
** Regular monitoring routines of the various levels in your unit is encouraged.

4-3
Models 3940 and 3949 _____________________________________________________________________________Service

Section 5 - Service 11. Pull the entire assembly out of the panel, and remove the
unit.
12. Replace the thermostat and probe.
Note: Reseal the probe access hole with Permagum and retie
the overtemp cable to the existing wires after replacing the
probe.

Servicing must be performed by qualified service


personnel only! 5.3 Replacing the Humidity/Temperature Sensor

1. Locate the probe mounting plate in the center of the


right side of the incubator interior.
De-energize all potential 2. Open the mounting plate by removing the screws that
sources of energy to this hold it in place.
unit and lockout/tagout
3. Locate the humidity sensor mounted on the inside of the
their controls.
panel in a black housing. Note the angle of the probe.
4. Grasp and unplug the probe from the probe cable.
5.1 Accessing the Electrical Components 5. When replacing the humidity sensor, be sure to mount
the probe at the same angle as it was originally mounted.
To gain access to the electrical components, remove the
two screws located on the left side of the control panel with a
phillips screwdriver. The control panel is hinged and will swing 5.4 Replacing the Optional Recorder and Probe(s)
open.
1. Open the incubator door, and locate the probe mounting
5.2 Replacing the Overtemp/Undertemp Probe and plate attached to the center of the right interior wall.
Thermostat Remove the mounting plate.
2. The recorder probe is attached to the lower end of the
1. Remove the incubator ceiling by remove the screws back of the mounting plate. Remove the probe by care-
holding it in place. fully sliding it out of the housing.
2. Remove the top three screws from the top of the right 3. Remove the screws securing the ceiling of the incubator
duct cover. and remove the ceiling.
3. Lean the duct sheet out, and remove the Permagum seal 4. Remove the top three screws on both edges of the right
from around the probe access hole. duct sheet.
4. Remove the 15” copper capillary overtemp probe by 5. Lean the duct sheet out in order to remove the
extracting the two plastic clips that hold the probe in Permagum seal from around the probe access hole.
place.
6. Remove the two screws located on the left side of the
5. Open the control panel by removing the two screws control panel and open the control panel door. Remove
located on the left side of the control panel. any Permagum from around the access hole.
6. Pull the overtemp probe up through the access hole and 7. Pull the probe(s) carefully up through the hole.
into the control panel.
8. Follow the probe cable(s) to the back of the recorder,
7. Follow the wires from the probe to the thermostat and carefully clip any plastic ties holding the cable(s) to
mounted on the control panel. Clip the plastic ties hold- other wiring.
ing the overtemp cable to the existing wiring.
9. Remove the three screws securing the recorder and pull
8. Pull the overtemp knob on the control panel off. it carefully out from the front of the control panel.
9. Remove the two screws that hold the overtemp assembly 10. Replace the recorder with the correct Forma part.
to the control panel.
Note: When replacing the recorder and probe(s), retie the probe
10. Disconnect the two wires from the back of the thermo- cable(s) to the existing wires.
stat assembly.

5-1
Models 3940 and 3949 _____________________________________________________________________________Service

Note: When raising the top section, notice the gasket seal at the
5.5 Removing the Top Section top opening of the incubator chamber. The gasket must be cor-
rectly positioned when reinstalling the top on the chamber.
If it becomes necessary to remove the top of the incubator 15. Slowly lift the entire top assembly up and off the lower
to move it through a low doorway, do the following: chamber section while carefully guiding the capillaries
1. Read the instructions completely before starting. and sensing bulbs out of the chamber area. Place the top
2. Provide adequate space and sufficient lighting to per- assembly onto the carpenter’s horses or other support
form the work. arrangement.
3. Disconnect the inlet water supply and the drain. 16. Reinstall the top assembly by reversing the above proce-
dure. Exercise care particularly when:
4. Before lifting the top section from the incubator, provide
carpenter’s horses or other suitable support when the top • Placing and aligning the sealer gasket on the 1/2”
is suspended. flange on top of the chamber when lowering the top in
place.
5. Remove all test material from the incubator chamber.
• Routing the temperature and/or humidity sensors and
6. All temperature and/or humidity sensing probes are
capillaries to prevent severely bending them.
mounted on a probe mounting bracket on the right side
of the incubator chamber. Remove the nuts securing the • Mounting the temperature and/or humidity sensor
bracket and then remove the probes from the bracket. bulbs on the mounting brackets.
Note the location and configuration of the probes. • Tightening the top mounting bolts alternately to ensure
7. The sensing bulb capillaries are routed behind the duct a balanced pressure on the gasket.
sheet up to the access port at the top of the unit behind
the control panel.
Note: Route all the probes between the duct sheet and the out-
side wall of the chamber so that when the top is raised, all 5.6 Setting the Door Heater Control
probes will move freely up and out. do not kink or bend the
capillaries.
8. The humidification system drain line (3/8” clear tygon) High voltage is present behind con-
is routed down through the access port and is secured to trol panel. Servicing must be per-
the condensate drain tube on the center left side of the formed only by qualified electrical
incubator (as viewed from the rear). Free the drain line, service personnel.
by pulling it away from the float tank (behind the con-
trol panel) and routing it down through the access port. The infinite heater control is
9. Open the control panel by removing the two screws located in the right side of the incu-
located on the left side of the control panel. bator top compartment behind the
10. Loosen the top gasket around each of the air exhaust control panel door. The heater
vents by turning the screw. varies the amount of door heat from
no heat (zero) to full heat (100) as
11. Remove the eight screws from the top cover of the incu- indicated by the control dial face. If
bator. the knob is turned past zero, a
12. Remove the four screws from each of the vent caps on “click” will indicate that all power
the top cover of the incubator. The top cover can now be to the door is shut off. If turned past Figure 5-1
removed. 100, a similar “click” will indicate that the heat is set at the
13. Remove the nine 5/16” x 4” hex head bolts, lock wash- maximum.
ers, and two flatwashers which secure the top assembly Initially the units leave the factory with the dial set at 40.
to the cabinet. Note the washer arrangement on the If desired, the amount of heat can later be reduced until mois-
bolts. ture appears on the door, then the heat advanced. However, in
14. Remove the black trim gasket located at the junction of fluctuating ambient conditions, it is recommended that a mini-
the top assembly and the main incubator section. The mum of 40% door heat be used.
ends of the gasket have been joined together at the rear
of the incubator.

5-2
Models 3940 and 3949 _____________________________________________________________________________Service

5.7 Cleaning and Adjusting the Humidity Steam 7. Carefully pry off the brass top of the float control reser-
Generator, P/N 505087 voir. Clean and disinfect it and the connecting tube
between the reservoir and the steam generator being
Depending upon the quality of water used in the humidifi- careful to not damage the float assembly.
cation system, it may be necessary to clean the humidity gener-
ator and immersion heaters every 2 to 3 months. 8. Clean the two heater elements and the part of the copper
capillary tube attached to them.
Materials Required: 6-foot stepladder
9. When all surfaces are clean, rinse with sterile distilled
Flat and Phillips screwdrivers water until all soap and disinfectant are removed.
Clamp-on style Ammeter 10. Open the distilled water valve and allow the reservoir to
Laboratory disinfectant fill. Measure the water level. The limit is three inches,
Siphon, sponge, & cleaning materials plus or minus one-quarter of an inch. The water level is
raised and lowered by carefully bending the float arm.
1. Remove all contents from the incubator, shut it off, and When set, replace the cover.
disconnect the power.
11. Replace the top of the steam generator and alternately
2. Turn off the valve supplying the sterile distilled water. tighten the four thumbscrews to evenly compress the
3. From the stepladder, remove the eight Phillips screws rubber gasket.
securing the top of the incubator cabinet.
4. When the steam generator has cooled, remove the four
thumbscrews.
a. Checking the Steam Generator Safety
Thermostat Calibration
The internal temperature of the
steam generator is hot enough to boil 1. Turn the slotted rheostat shaft of the temperature control
water. Make certain sufficient time is on the back of the steam generator fully clockwise.
given to allow the unit to completely 2. Remove all tools and cleaning materials from the top of
cool before removing the top. the incubator. Reconnect power to the unit and turn it
on.
5. Lift the top off of the steam generator and lay it to the 3. Set the RH level to 100% and open the incubator door.
side taking care to not crimp the copper capillary tube. This will cause the heaters in the steam generator to be
6. Siphon all of the water out of the steam generator and continuously energized.
clean it with a good quality laboratory detergent and dis-
infectant. Do not use any type of chloride cleaner. A
The next step requires working in the
bristle brush may be needed for stubborn rust and scale.
top section of the incubator where
Also clean the inside of the standpipe using a test tube
high voltages are present. This should
brush. Siphon and sponge the wash water out of the
only be done by qualified persons.
boiler. Repeat cleaning with soap and water as neces-
sary.

capillary
tube

standpipe

float

float
arm
rubber
float
gasket
Figure 5-2 valve

Figure 5-3
5-3
Models 3940 and 3949 _____________________________________________________________________________Service

4. Using a clamp-on ammeter, establish that current is Section 6 - Specifications


flowing in the blue wire going to the heater connectors.
Refer to Figure 5-2. Temperature
5. Proper current is about 3.6 amps. If the current is not Control ±0.1°C @ +37°C (98.6°F)
flowing and the humidity indicator is on, rotate the small Range 0°C (32°F) to +60°C (140°F)
slotted screw in the shaft of the safety thermostat coun- Sensor RTD
terclockwise, using a jeweler’s screwdriver, until the Controller Digital electronic proportional
current flows to the heaters. Setpoint Digital
Display Digital LED
6. Allow the steam generator to operate for about 30 min-
Readability 0.1°C
utes continuously. If the heaters cycle off due to the
Setability 0.1°C
safety thermostat, rotate the small slotted screw further
Uniformity ±0.3°C at 25°C to 37°C with six
counterclockwise to maintain the heater’s operation.
shelves installed*
7. After 30 minutes, turn the small slotted screw clockwise
very slowly until the heaters shut off. Shelves
8. Turn the slotted screwdriver 1-1/2 turns counterclock- Standard 6
wise. The thermostat is now properly adjusted. Maximum 19
9. Replace the cabinet top and return the incubator to serv- Dimensions 30.62”W x 25.81”F-B
ice. (77.78 cm x 65.56 cm)
Construction Solid stainless steel reinforced
Surface Area 5.4 sq. ft. (.51 sq. m) per shelf
Max. Per Chamber 104.3 sq. ft. (9.69 sq. m)
Clearance Adjustable on 3” (7.62 cm) centers
Loading 35 lbs. (16 Kg) (slide in and out)
50 lbs. (23 Kg) (stationary)

Construction
Volume 29 cu. ft. (823 liters)
Interior 304 2B stainless steel
Exterior Cold rolled steel
Insulation 2” (5.1cm) Foamed urethane
Outer Door Gasket Four sided vinyl compression
Finish Powder coated. Salt spray tests
exceed 1000 hrs. per ASTM
Standard B117-85

Weight
Net 800 lbs. (363 Kg)

Shipping
Motor 850 lbs. (386 Kg)
Air 940 lbs. (476 Kg)
Ocean 985 lbs. (447 Kg)

Temperature Alarm
Sensor Thermostat
Controller Thermostat
Setpoint Analog reference dial
Alarm Audible/visual

6-1
Models 3940 and 3949 _________________________________________________________________________Spare Parts

Section 7 - Spare Parts


Humidity
Control ±5% RH Part No. Description Qty
Range Above ambient to 95% 290163 RTD Temperature Sensor 1
Sensor Bulk polymer
230066* Fuse, Ceramic 10A 350V 1
Controller Electronic, direct set in % RH
Setpoint Digital 400051 Power Supply 1
Display Digital LED 290162 RH Sensor 1
Readability 1.0% 231186 Controller, Watlow 982 Auto Tune 2
Setability 1.0%
Steam Gen. Initial fill approx. 1 quart (.95 liters) 320277 Triac, 25A 2
285615* Fuse, Fusetron 25A 2
Fittings 620430* Humidifier Heater 1
Fill Port 1/8” FPT
270010 Thermostat, B-10 0-220F 1
Drain Port 3/8” (1.9 cm) FPT
131300 Blower Wheel 1
Unit Heat Load 410048 Defrost Timer 1
115V 5500 BTUH (1600W)
630090 Heater 1
220V 6000 BTUH (1750W)
205062 Condensing Unit, 1/4 HP 115V R-134A 1
Refrigeration 991334 Solenoid, 120V .101” Port 1
Compressor 1/4 Horsepower, air-cooled 285599 Thermal Fuse, 250V 17A 1
Refrigerants Non-CFC
830044 Blower Motor, 1/4 HP, 115V 1
Electrical
3940/4940 208-220V, 1 PH, 60Hz, 3P, 4W, *Immediate Stock
14.0 FLA
3949/4949 220-240V, 1 PH, 50/60Hz, 2P, 3W,
14 FLA
Line Cord None (lockable disconnect provided)

Dimensions
Exterior 38.00”W x 87.50”H x 32.00”F-B
(96.52 cm x 222.25 cm x 81.28 cm)
Interior 31.00”W x 60.00”H x 27.00”F-B
(78.74 cm x 152.40 cm x 68.58 cm)

Continuing research and improvements may result in speci-


fication changes at any time. Performance plus or minus
the least significant digit unless otherwise specified.

* Better than ±0.5°C uniformity at all other temperature


parameters.

7-1
Models 3940 and 3949 _________________________________________________________________________Spare Parts

7-2
Models 3940 and 3949 _________________________________________________________________________Spare Parts

BACK VIEW

7-3
Models 3940 and 3949 ______________________________________________________________Refrigeration Schematics
8-1
Models 3940 and 3949 __________________________________________________________________Electrical Schematics

9-1
Models 3940 and 3949 __________________________________________________________________Electrical Schematics

9-2
Models 3940 and 3949 __________________________________________________________________Electrical Schematics

9-3
Models 3940 and 3949 __________________________________________________________________Electrical Schematics

9-4
Models 3940 and 3949 __________________________________________________________________Electrical Schematics

9-5
Models 3940 and 3949 __________________________________________________________________Electrical Schematics

9-6
Models 3940 and 3949 _______________________________________________________________________________________________________Electrical Schematics

9-7
Models 3940 and 3949 _______________________________________________________________________________________________________Electrical Schematics

9-8
Models 3940 and 3949 _______________________________________________________________________________________________________Electrical Schematics

9-9
Thermo Forma
Millcreek Road, P.O. Box 649
Marietta, Ohio 45750
U.S.A.

Telephone (740) 373-4763


Telefax (740) 373-4189

You might also like