QSFLUXUS_FG721V1-1EN
FLUXUS *721
Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).
Transmitter
Sensor module
A B NEXT
QON
DISP
The supplied sensor mod-
display keyboard
ule (SENSPROM) has to
SNAP
Q- Q+
MUX DISP
QOFF
transducers
FAST MODE
transducers
KL7
KL9
AGN
BGN
ARS
BRS
BVS
AVS
T1A
T1B
T2A
T2B
T3A
T3B
T4A
T4B
P1+
P2+
P3+
P4+
P5a
P6a
P7a
AR
BR
BV
S1
S3
AV
A+
B-
L(+)
N(-)
PE
channel A X2 X3
P5b
P6b
P7b
T1a
T1b
T2a
T2b
T3a
T3b
T4a
T4b
P1-
P2-
P3-
P4-
S1
S3
S
S
outputs
sensor sensor
module module inputs
channel A channel B
Connection of Transducers
connection system TS
Attention!
For good high frequency shielding, it is important to
junction box JB01 junction box JB02, JB03
ensure good electrical contact between the external
shield and the cap nut (and the housing).
Connection to transmitter
transmitter
transmitter
terminal extension cable
AV white or marked cable (core)
AVS white or marked cable (internal shield)
ARS brown cable (internal shield)
AR brown cable (core)
20 mm
100 mm
extension cable
transducer, junction box and transmitter have the same electrical potential
terminal transducer cable with stripped cable ends
V transducer (brown cable, marked white)
VS transducer (red cable)
RS transducer (red cable) Connection of Power Supply
R transducer (brown cable)
terminal connection AC connection DC
PE earth earth
terminal transducer cable with SMB connectors
N(-) neutral -DC
XV SMB connector (brown cable, marked white) L(+) phase +DC
XR SMB connector (brown cable, marked black)
The voltage is indicated on the plate below the terminal strip.
QSFLUXUS_FG721V1-1EN, 2016-07-15
Quick Start Guide FLUXUS *721
Channel
x : channel is activated
Parameter Channel >A< B Y Z - : channel is deactivated
Channel A: Measurement x - - - : no parameters
Select the Select the
channel. channel.
• select the channel: key " and $
Transducer A: Meas. point no.
• activate the channel: key & and
Connected Transd A
Select the Enter the
transducer. measuring point number.
Input of sound path
Outer diameter A: Sound path a
62.00 mm 2 NUM • sound path (even number):
The transducers are mounted on
Enter the Confirm
the same side of the pipe.
outer pipe diameter. the value.
• sound path (uneven number): a
Pipe material Transd. distance The transducers are mounted on
Carbon Steel 54 mm opposite sides of the pipe.
Select the
Use coupling foil or apply
pipe material.
a bead of acoustic
Wall thickness a - transducer distance
coupling compound to
3.00 mm the contact surface
of the transducers.
Enter the equipotential
pipe wall thickness. Mount the transducers on the bonding ter-
Lining pipe at the recommended minal
transducer distance a.
>NO< YES
Is pipe A: Measurement
lining present?
Roughness
Auto
Enter
the roughness. Press or until bar
graph Q is displayed in the lower
Fluid
line.
Water Change transducer position slightly,
Select until bar graph Q has max. length.
the fluid.
Transd. distance(50.00)
54 mm
Measure and enter trans-
ducer distance a. Measurement
Press .
Measurement
A:Volum. flow rate Stop measurement
54.5 m3/h Show parameters
Show measurement
Measurement
QSFLUXUS_FG721V1-1EN, 2016-07-15