Quick Start Guide

QSFLUXUS_FG721V1-1EN
FLUXUS *721

Attention! Observe the "Safety Instructions for the Use in Explosive Atmosphere" (see document SIFLUXUS).

Transmitter
Sensor module

A B NEXT
QON
DISP
The supplied sensor mod-
display keyboard
ule (SENSPROM) has to
SNAP
Q- Q+

MUX DISP

QOFF

transducers
FAST MODE

be connected to the corre-


Ethernet interface equipotential sponding terminals of the
channel B LAN USB

USB interface bonding terminal


transmitter. The serial
KL12
KL11

transducers
KL7

KL9
AGN

BGN
ARS

BRS
BVS
AVS

T1A
T1B

T2A
T2B
T3A
T3B

T4A
T4B

P1+
P2+
P3+
P4+
P5a
P6a
P7a
AR

BR
BV

S1

S3
AV

A+
B-

power supply numbers of the sensor


KL4
KL10
KL8

L(+)
N(-)
PE

X_AV X_AR X_BV X_BR

channel A X2 X3
P5b
P6b
P7b
T1a
T1b

T2a
T2b
T3a
T3b

T4a
T4b

P1-
P2-
P3-
P4-
S1

S3

S
S

module and the transducer


must be identical.

outputs
sensor sensor
module module inputs
channel A channel B

Connection of Transducers
connection system TS

Attention!
For good high frequency shielding, it is important to
junction box JB01 junction box JB02, JB03
ensure good electrical contact between the external
shield and the cap nut (and the housing).

Connection to transmitter
transmitter

transmitter
terminal extension cable
AV white or marked cable (core)
AVS white or marked cable (internal shield)
ARS brown cable (internal shield)
AR brown cable (core)
20 mm
100 mm

terminal transducer cable with stripped cable ends


75 mm

transducer cable with stripped cable ends

AV transducer (brown cable, marked white)

transducer cable with SMB connectors


12 mm

AVS transducer (red cable)


ARS transducer (red cable)
AR transducer (brown cable)

terminal transducer cable with SMB connectors


X_AV SMB connector (brown cable, marked white)
extension cable

extension cable

X_AR SMB connector (brown cable, marked black)


70 mm

Connection to junction box

terminal extension cable


TV white or marked cable (core)
TVS white or marked cable (internal shield)
TRS brown cable (internal shield)
transmitter transmitter
TR brown cable (core)

transducer, junction box and transmitter have the same electrical potential
terminal transducer cable with stripped cable ends
V transducer (brown cable, marked white)
VS transducer (red cable)
RS transducer (red cable) Connection of Power Supply
R transducer (brown cable)
terminal connection AC connection DC
PE earth earth
terminal transducer cable with SMB connectors
N(-) neutral -DC
XV SMB connector (brown cable, marked white) L(+) phase +DC
XR SMB connector (brown cable, marked black)
The voltage is indicated on the plate below the terminal strip.

QSFLUXUS_FG721V1-1EN, 2016-07-15
Quick Start Guide FLUXUS *721

Parameter input Start measurement Navigation

• vertical selection ( ): key & and

Parameter Measurement • horizontal selection ( ): key " and $

• return to main menu: key BRK


• delete: key C

Channel

x : channel is activated
Parameter Channel >A< B Y Z - : channel is deactivated
Channel A: Measurement x - - - : no parameters
Select the Select the
channel. channel.
• select the channel: key " and $
Transducer A: Meas. point no.
• activate the channel: key & and
Connected Transd A
Select the Enter the
transducer. measuring point number.
Input of sound path
Outer diameter A: Sound path a
62.00 mm 2 NUM • sound path (even number):
The transducers are mounted on
Enter the Confirm
the same side of the pipe.
outer pipe diameter. the value.
• sound path (uneven number): a
Pipe material Transd. distance The transducers are mounted on
Carbon Steel 54 mm opposite sides of the pipe.
Select the
Use coupling foil or apply
pipe material.
a bead of acoustic
Wall thickness a - transducer distance
coupling compound to
3.00 mm the contact surface
of the transducers.
Enter the equipotential
pipe wall thickness. Mount the transducers on the bonding ter-
Lining pipe at the recommended minal
transducer distance a.
>NO< YES
Is pipe A: Measurement
lining present?

Roughness
Auto
Enter
the roughness. Press or until bar
graph Q is displayed in the lower
Fluid
line.
Water Change transducer position slightly,
Select until bar graph Q has max. length.
the fluid.

Fluid temp. Notes on transducer mounting Note!


22.0 C • Observe the recommended distance between When the transducers are mounted
the measuring point and the disturbance point. correctly, the engravings on the trans-
Enter the ducers form an arrow.
fluid temperature. • Clean the pipe.
• Use coupling foil or apply a bead of acoustic
Extension cable coupling compound to the contact surface
0.0 m of the transducers.

Enter the length • Mount the transducers laterally to the pipe,


if possible.
of the extension cable.

Transd. distance(50.00)
54 mm
Measure and enter trans-
ducer distance a. Measurement
Press .
Measurement
A:Volum. flow rate Stop measurement
54.5 m3/h Show parameters
Show measurement
Measurement

QSFLUXUS_FG721V1-1EN, 2016-07-15