6-4 USB 생산 이력 저장 에러
6.4.1 생산이력 저장 설정 0
6.4.2 USB 저장 0
When “AC Connection Error” message is displayed in the touch screen, and the zero value is
under 0(zero) or over 90000 in LCD in the main board, the zero value must be adjusted by use of
tmip-port.
*Direction of turning
1
6.1.2 LOADCELL TEST
This is for the test by use of tester whether the loadcell is normal or not
**If the load cell is used by other brand’s one, SEN+,- must be short like as above photo.
2
6.1.3 Communication Error
<Confirm LED>
(Main board)
There are 2 LED at the place of left below on the main board.
*Upper LED: When the touch screen is normally connected with the main board,
the red color of LED is a light on.
*Lower LED: When the touch screen is normally connected with the main board,
the yellow color of LED is a light on.
So, you can confirm the communication status between the touch screen and main board
by LED light color.
If the cable has a wrong connection between the touch screen and main board,
“AD connection Error” message can be displayed on the touch screen.
3
6.2 Rejector Error
Default Value
PASS 101
UNDER 135
OVER 125
Double Entry 125
Foreign material 145
detect
4
6.2.3 When the double entry comes out (In case of Pusher typr rejector)
The double entry items cannot be rejected in case of Pusher type rejector.
So, there comes out the rejecting error.
This condition must be confirmed first when the rejector is used by the pusher type.
5
6.3 Conveyor driving Error
Emergency Button
In addition,
if there happens the conveyor driving error by emergency stop button, “SA” message is automatically
displayed on Inverter by self-protecting function in inverter.
또한, 비상정지 상태로 인해 컨베이어 구동 에러가 발생하였을 때 인버터에서는 인버터의 보호기능이
동작해서 표시부가 아래와 같은 “SA”라는 에러 표시가 자동적으로 나타납니다.
(Please refer to the guide list of each error message for the Inverter)
If the motor is out of order, or has a wrong wiring connection, error message will be displayed on the
Inverter.
Please refer to the below table.
6
<Table. Inverter Error>
7
6.3.3 When the speed data does not input in the part number setting,
When the speed data does not input in the part number setting, this causes the conveyor driving
error. Please check whether the speed is set correctly.
In order to check this, enter into “Setup -- P/N Set - Speed Set”
(If the speed data does not be input, please set it like as below photos.))
O.K.
No-Good
8
6.4 Save (back-up) error by USB
Step 1
Insert the USB memory, then enter into
the accumulation screen..
Step 2
If the blue light on the USB icon comes On,
the back-up is a normal condition..
9
Step 3
The message of “Do you want to delete inside saved
data?” in Pop-up window will be displayed when you
press the icon.
Step 4
If you touch “Yes”, the data will save to USB
memory after inside data in USB is deleted.
.
10
6.1.6 Password Error
If you forget the password, please re-set it by the following method.
Initial setting password in factory is 1234.
Step 1
If you touch the point of upper right corner on the
setup screen, the pop-up window is displayed like
as the photo
.
.
Step 2
After you input the master pass “56”,
input again two digits of setting date in
program.
The password is composed of total 4 digits.
Step 3
You can enter into the User setup screen.
And, the new password can be set..
11
Method 2)
If you do not know the saved date in the program, you can use this method 2.
Step 1
If you touch the point of upper right corner on the setup screen, the pop-up window is displayed like as the
photo.
Step 2
If you input the password “5600”, then enter, the password error message is displayed.
After clicking OK, the master password pop-up window will appear again..
Step 3
Enter 5688 in the master password that pops up secondly and press Enter to return to the user setting screen.
You will be entered with the password set disabled and you can reset your password.