Professional Documents
Culture Documents
Pentra XL 80
Pentra XL 80
REF: 1300018562
Int. Ref. Doc: RAA022GEN
Contents
1 - Introduction
1. Revision ..................................................................................................................3
7. Specifications ......................................................................................................23
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 0-1
Contents
Technical Manual
0-2 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Contents
1. Boards ....................................................................................................................3
3. Connectors ..........................................................................................................10
4. Fan ........................................................................................................................13
5. Synoptics .............................................................................................................13
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 0-3
Contents
Technical Manual
0-4 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Contents
5 - Software Versions
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 0-5
Contents
Technical Manual
0-6 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Contents
07 - Maintenance
1. Maintenance ..........................................................................................................2
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 0-7
Contents
8 - Exploded Views
1. List ......................................................................................................................... 2
2. Valves ..................................................................................................................... 9
Technical Manual
0-8 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Contents
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 0-9
Contents
Technical Manual
0 - 10 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Introduction
1 - Introduction
1. Revision .......................................................................................................................... 3
7. Specifications .............................................................................................................. 19
7.1. Technical specifications ............................................................................................................... 19
7.2. Physical specifications ................................................................................................................. 20
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1-1
Introduction
Technical Manual
1-2 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Revision
1. Revision
Reference Internal Reference Software Version Date
RAA022AA RAA022AA V1.0 March 2002
P80: V1.3
RAA022BEN RAA022BEN April 2004
PXL80: V1.1
P80: V1.7
RAA022CEN RAA022CEN March 2008
PXL80: V1.6
P80/XL80: V1.11
RAA022DEN RAA022DEN March 2013
PXLR: V2.0
RAA022EEN RAA022EEN V2.0 December 2013
1300012448 RAA022FEN V2.3 February 2015
1300018562 RAA022GEN V2.4 June 2016
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1-3
Introduction
Legal Information
2. Legal Information
This instrument complies with the Standards and Directives named in the Declaration of Conformity. The latest
version of the CE Declaration of Conformity for this instrument is available on www.horiba.com
The Information in this manual is distributed on an "As Is" basis, without warranty. While every precaution has
been taken in the preparation of this manual, HORIBA Medical will not accept any liability to any persons or
entities with respect to loss or damage, caused or alleged to be caused directly or indirectly by not following
the instructions contained in this manual, or by using the computer software and hardware products described
herein in a manner inconsistent with our product labeling.
To alert the operator of potentially hazardous conditions, one of the headings which are described below is
provided wherever necessary throughout this text.
Flags a procedure that if not followed properly, can prove to be extremely hazardous to either the
operator or the environment or both.
Emphasizes an operating procedure that must be followed to avoid possible damage to the
instrument or erroneous test results.
Emphasizes important information especially helpful to the operator before, during or after a specific
operational function.
Technical Manual
1-4 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Legal Information
2.4. Graphics
All graphics, including screens, printouts and photographs, are for illustration purposes only and are not
contractual.
2.5. Trademarks
Other product names mentioned within this publication may be trademarks or registered trademarks of their
respective owners.
All rights reserved. No part of this book may be reproduced or transmitted in any form or by any means,
electronic, mechanical, photocopying, recording, or otherwise, without the prior written permission of HORIBA
Medical.
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1-5
Introduction
Operational Conditions
3. Operational Conditions
3.1. Environment
■ The operation of the ABX Pentra 80 / ABX Pentra XL80 / Pentra XLR should be restricted to indoor location
use only. The instrument is operational at an altitude of 3000 m (9800 ft) maximum.
■ The instrument is designed for safety from voltage surges according to INSTALLATION CATEGORY II and
POLLUTION DEGREE 2 (IEC 61010-1).
■ Please contact your local representative for more information regarding operation locations, when it does not
comply with the recommended specifications.
3.2. Location
Keep in mind that the instrument and printer weigh approximately 55 kg (121 lbs) [ABX Pentra 80/
ABX Pentra XL80] or 57 kg (126 lbs) [Pentra XLR]. To move the instrument, two persons are required
and must use the lifting handles provided in the installation kit.
The power switch and power supply connection should always be accessible. When positioning the
system for operational use, leave the required amount of space for easy accessibility to these items.
3.3. Grounding
Proper grounding is required when installing the system. Check the wall outlet ground (earth) for proper
grounding to the facilities electrical ground. If you are unsure of the outlet grounding, contact your facilities
engineer to verify the proper outlet ground.
Technical Manual
1-6 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Operational Conditions
Humidity Conditions:
Relative humidity of 80% maximum for temperatures up to 31°C (88°F), decreasing linearly to 50% at 40°C
(104°F).
■ The instrument has been designed to produce less than the accepted level of electromagnetic interference in
order to operate in conformity with its destination, allowing the correct operation of other instruments also in
conformity with their destination.
■ In case of suspected electromagnetic noise, check that the instrument has not been placed in the proximity
of electromagnetic fields or short wave emissions, e.g. radar, X-rays, scanners, cell phones, etc.
If a power failure or sudden extinction occurs during routine series, eject the racks still in the
instrument. Check the integrity of results and patient identification in accordance with the history of
the samples for all the tubes present in the machine during the incident. It is recommended to install
the system on UPS (Uninterruptible Power Supply).
■ Grounding is required. Check that the earth wall-plug is correctly connected to the laboratory grounding
system. If there is no such system, a ground stake should be used.
■ Use only the main supply cable delivered with the instrument.
■ Main supply voltage fluctuations must not exceed +/- 10% of the nominal voltage.
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1-7
Introduction
Operational Conditions
Instrument disposal
This product should be disposed of and recycled at the end of the useful life in accordance with
European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) and/or
European Directive 2006/66/EC on batteries and accumulators.
If any doubt, please contact your local HORIBA Medical representative service department.
Condition for storage and transportation: Temperature from -20°C to +50°C (-4°F to 122°F).
Before the instrument removal from use, transportation or disposal, perform a general cleaning procedure and
a draining procedure on the instrument.
3.9. Installation
An HORIBA Medical representative will install your instrument, software and printer.
Verify that all of the parts from the packages are present:
Package content:
■ Analyzer
Technical Manual
1-8 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Operational Conditions
■ Documentation CD-ROM
■ Daily Guide
■ Safety Information booklet
■ Reagents, Controls and Calibrators CD-ROM
■ Installation kit
■ External Barcode reader
■ Printer
■ Lifting handles
■ Racks
■ Waste tank
■ External tank sensor
■ Diluent container opening tool
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1-9
Introduction
Labels and connections
The power switch and power supply connection should always be accessible. When positioning the
system for operational use, leave the required amount of space for easy accessibility to these items.
Technical Manual
1 - 10 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Labels and connections
The laser used in this instrument is a Class 3B laser (complies with IEC 60825 - 1:2007). Its beam
presents, by definition, a high safety hazard.
2 - Waste output 1
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 11
Introduction
Labels and connections
4 - 4 USB connections
3 4
Technical Manual
1 - 12 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Labels and connections
3 - 4 USB connections
2 3
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 13
Introduction
Labels and connections
1 2 3 4 5
6 7
8 9 10
1 - VGA 06 - Printer
2 - Mouse 07 - COM3: Unused port
3 - Keyboard 08 - COM2: Labo Link
4 - LAN 09 - COM1: To mother board
5 - USB ports 10 - Power
4 5 4
1
3 2
1 - Power
2 - COM2 (laboratory link)
3 - COM1 (to motherboard)
4 - 4 USB ports
5 - LAN
1 - USB port
Technical Manual
1 - 14 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Labels and connections
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 15
Introduction
Labels and connections
Technical Manual
1 - 16 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Labels and connections
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 17
Introduction
Instrument general cleaning
The external surfaces of the instrument must be decontaminated considering the biological environment.
■ Fungicidal,
Please also refer to the WHO (World Health Organization) guidelines: "Laboratory Biosafety Manual,
3rd edition", for further information.
Technical Manual
1 - 18 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Instrument general cleaning
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 19
Introduction
Warning and precautions
The instrument must be operated as instructed in the technical manual. Any other use might compromise the
system integrity and might be hazardous for the operator.
This instrument complies with the Standards and Directives named in the Declaration of Conformity. The latest
version of the CE Declaration of Conformity for this instrument is available on www.horiba-abx.com.
■ The reagents and accessories stipulated by HORIBA Medical have been validated in accordance
with the European Directive for in vitro medical devices (98/79/CE).
■ The use of any other reagents and accessories may place at risk the performance of the
instrument, engaging the user’s reponsability. In this case, HORIBA Medical takes no responsability
for the device nor for the results rendered.
■ Disposable gloves, eye protections and lab coats must be worn by the operator.
■ Local or national regulations must be applied in all the operations.
■ Mobile phones should not be used in proximity of the instrument.
■ All peripheral devices should be IEC compatible.
The duration of warranty is stipulated in the Sales conditions associated with the purchase of this instrument.
To validate the warranty, ensure the following is adhered to:
■ The system is operated under the instructions of this manual.
■ Only software or hardware specified by HORIBA Medical is installed on the instrument. This software must
be the original copyright version.
■ Services and reparations are performed by a HORIBA Medical authorized technician, using only HORIBA
Medical approved spare parts.
■ The electrical supply of the laboratory adheres to national or international regulations.
■ The system is operated according to HORIBA Medical’s recommendations.
■ Specimens are collected and stored in normal conditions.
■ Reagents used are those specified in the user manual.
■ Proper tools are used when maintenance or troubleshooting operations are performed (Refer to the
Maintenance chapter).
If this instrument has been supplied to you by anyone other than HORIBA Medical or an authorized
representative, HORIBA Medical cannot guarantee this product in terms of specifications, latest
revisions and latest documentation. Further information may be obtained from your authorized
representative.
Technical Manual
1 - 20 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Warning and precautions
LASER: it is strictly forbidden to disable laser cover sensors as it may cause operator (and persons
around) injuries. Laser protection covers must never be handled nor opened. Class 3B laser product
(Standard series IEC 60825 - 1:2007).
Operator injury may occur from an electric shock. Electronic components can shock and injure the
user. Do not tamper with the instrument and do not remove any components (covers, doors, panels,
etc.) unless otherwise instructed within this document.
Danger of explosion if the battery is not replaced correctly. When replacing the battery, always use
the same and/or equivalent type recommended by the manufacturer. Dispose of used batteries
according to the manufacturer's specific instructions.
Moving parts: It is strictly forbidden to disable sensors as it may cause operator (and persons
around) injuries. Protection covers must not be opened during instrument operations.
Opening the doors and covers during instrument operations triggers the instrument emergency
stop.
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 21
Introduction
Warning and precautions
Consider all specimens, reagents, calibrators, controls, etc. that contain human blood or serum as
potentially infectious. Use established, good laboratory working practices when handling
specimens. Wear protective gear, gloves, lab coats, safety glasses and/or face shields, and follow
other biosafety practices as specified in the OSHA Blood borne Pathogens Rule (29 CFR part 1910.
1030) or equivalent biosafety procedures.
HORIBA Medical uses disinfectant product for instrument decontamination (including the touch screen) and
highly recommends it to decontaminate your instrument.
Technical Manual
1 - 22 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Specifications
7. Specifications
7.1.1. Parameters
* PDW, PCT, ALY%, ALY#, LIC%, LIC# and RHCc have not been established as indications for use
in the United States for this instrument. Their use should be restricted to research or investigational
use only (RUO parameters).
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 23
Introduction
Specifications
7.1.4. Reagents
■ ABX Diluent (10 Liters or 20 Liters).
■ ABX Cleaner (1 Liter, Integrated).
■ ABX Basolyse II (1 Liter, Integrated).
■ ABX Eosinofix (1 Liter, Integrated).
■ ABX Lysebio (0.4 Liter or 1 Liter, Integrated or 1 Liter, Integrated [ABX Pentra 80 and ABX Pentra XL80]).
■ ABX Fluocyte (0.5 Liter, Integrated) [Pentra XLR only].
■ ABX Minoclair (0.5 Liter, non-integrated).
Power requirements
From 100 Vac to 240 Vac (+/-10%)
Power supply:
50 Hz to 60 Hz
Maximum power consumption: 230 W
Technical Manual
1 - 24 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Specifications
Power requirements
Maximum heat output: 670 kJ/h (635 BTU/h)
Printer: Refer to the printer manual
Computer Characteristics
Color LCD touch screen: 12 in.
Windows XP or Windows NT 4.0 for ABX Pentra 80
Operating System:
Windows XP or Windows 7 for ABX Pentra XL80 and Pentra XLR
Celeron 433 MHz, or
Processor:
Dual Core Intel Atom 1.46 GHz
128 MB for ABX Pentra 80
RAM (Random Access Memory):
4 Go for ABX Pentra XL80 and Pentra XLR
CD drive or external DVD/CD drive connected to the instrument USB port
Hard drive: 10 GB minimum
RS232, Ethernet, 4 USB connections
Capacity: 10000 results + graphics
Keyboard
Mouse
Laser specifications
Laser Melles Griot, Class IIIb Diode Laser
Wavelength: 488 nm
Emission power: 50 mW
Power consumption: 9W
Heat output: Negligible
Dimensions: 120 x 55 x 73 mm (Width x Depth x Height)
Dilution ratios
RBC/PLT: 1/10000 HGB: 1/250
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 25
Introduction
Specifications
Dilution ratios
WBC/BAS: 1/200 RET (Pentra XLR only): 1/15000
LMNE: 1/80
HGB measurement
Modified Drabkin method: Cyanide-free determination for ABX Lysebio
Light source: Electroluminescent diode
Wavelength: 550 nm
Bandwidth: 10 nm
Technical Manual
1 - 26 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Instrument description & technology
1
2
3 4
5 6 7
Left side:
5
3
4
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 27
Introduction
Instrument description & technology
Right side:
1
4
2
3
Technical Manual
1 - 28 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Introduction
Reagent specifications
9. Reagent specifications
The HORIBA Medical reagents specified for this instrument has been approved in accordance with
the European Directive 95/79/CE (Annexe III) for in vitro medical devices.
The CD-ROM RAX055 delivered with your instrument provides Reagents, Controls and Calibrators
leaflets/MSDS. The latest versions of these documents are available on:
www.horiba-abx.com/documentation.
When disposing of waste, protective clothing must be worn (lab coats, gloves, eye protections, etc.).
Follow your local and /or national guidelines for biohazard waste disposal.
■ If required, waste can be neutralized before being discarded. Follow your laboratory protocol when
neutralizing and disposing of waste.
■ The disposal of the waste container must be done according to the local or national regulatory requirements.
HORIBA Medical manufactures and markets reagents, calibrators and control bloods specially
designed for use with this analyzer. The use of products not recommended by HORIBA Medical may
give erroneous results or instrument operation problems. Contact your HORIBA Medical authorized
representative for all information regarding the recommended products.
10. Printer
Use the printer supplied or approved by HORIBA Medical.
Contact your local HORIBA Medical representative for more information about printer compatibility
and consumable part numbers.
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 1 - 29
Introduction
Printer
Technical Manual
1 - 30 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2-1
Hydraulic & Pneumatic Principles
Hydropneumatic connections
1. Hydropneumatic connections
Technical Manual
2-2 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2-3
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
2-4 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2-5
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
2-6 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2-7
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
2-8 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2-9
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
2 - 10 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2 - 11
Hydraulic & Pneumatic Principles
Hydropneumatic connections
Technical Manual
2 - 12 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Valves functions
2. Valves functions
Valve number Subject Functions
1 Differencial diluent Selects flowcell sheath 2 or LMNE chamber
2 Differencial diluent
3 Differencial diluent Selects flowcell sheath 1 or sheath 2
4 Flowcell sample supply Opens the pathway from the LMNE chamber to the flowcell
5 Flowcell sample injector Opens the waste path for sample LMNE syringe
6 Rinse output control Selects rinse for needle rinse block or WBC/Baso chamber
7 ABX Lysebio Controls ABX Lysebio entry in/exit from the reagent syringe
8 ABX Cleaner Controls ABX Cleaner entry in/exit from the reagent syringe
9 ABX Eosinofix Controls ABX Eosinofix entry in/exit from the reagent syringe
10 ABX Diluent Controls ABX Diluent entry in/exit from the reagent syringe
11 Diluent ouput control Routes ABX Diluent to the needle rinse block or the heating coil
12 ABX Basolyse II Controls ABX Basolyse II entry in/exit from the reagent syringe
Routes ABX Diluent from the flowcell and the LMNE syringe to
13 Flowcell rinse
the counting syringe
14 Counting syringe Sets counting syringe to atmosphere
15 WBC and RBC/Plt count Opens vacuum count line for WBC and RBC chambers
16 Counting syringe Drains counting syringe
Sets draining syringe #2 to atmosphere (through the Rinse
17 Draining syringe #2
chamber)
18 Draining syringe #2 Drains draining syringe #2
Fills the ABX Diluent tank or sets the ABX Diluent tank to
19 Diluent reservoir
atmosphere
20 Needle Needle cleaning
21 Needle Flushing out of the needle rinsing block
22 Needle Directs ABX Diluent into interior or exterior of needle
23 Needle Needle cleaning
Sets draining syringe #1 to atmosphere (through the Rinse
24 Draining syringe #1
chamber)
WBC/Baso Counting Routes ABX Diluent to the heating coil or sweeps flow for
25
head counting head
Routes ABX Diluent (via the heating coil) to the HGB or RBC
26 Diluent chamber select
chamber
27 Diluent circuit Fills WBC\BAS chamber with ABX Diluent
Opens the path fromthe flowcell output to the LMNE chamber to
28 Flowcell drain
drain it
29 ABX Diluent circuit Entry of ABX Diluent
30 Counting syringe Flushing\Draining of the counting syringe
31 Chambers Drains the rinse chamber
32 Chambers Drains the first dilution chamber
33 Chambers Drains theLMNE chamber
34 Chambers Drains the RBC chamber
35 Chambers Drains the WBC\BAS chamber
Selects and transfers a sample from the first dilution for RBC
36* RBC dilution
counting
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2 - 13
Hydraulic & Pneumatic Principles
Valves functions
Technical Manual
2 - 14 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydraulic circuit
3. Hydraulic circuit
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2 - 15
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
2 - 16 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2 - 17
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
2 - 18 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2 - 19
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
2 - 20 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2 - 21
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
2 - 22 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2 - 23
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
2 - 24 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Hydraulic & Pneumatic Principles
Hydraulic circuit
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 2 - 25
Hydraulic & Pneumatic Principles
Pneumatic diagrams
4. Pneumatic diagrams
See pneumatic diagrams on next page.
Technical Manual
2 - 26 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
1 3
2 1 2.3
2
2
2.3
3 1
70
1 2
2
2.5
3 2.3
3 1
1 2 1 3
2.5 1.6
3
2
3
3
2 1
1 3
2.3
2
2 1
Electric & Electronic Principles
1. Boards ............................................................................................................................ 3
1.1. APD Preamplifier board for Pentra XLR: CBR035A (1201761035) ................................................. 3
1.2. Carriage board: XAA428 (1209103428) .......................................................................................... 3
1.3. LEDs board: XAA429AS (1209109429) .......................................................................................... 3
1.4. Main board for ABX Pentra 80: XAA456 (1209102456) .................................................................. 4
1.5. LMNE amplifier board for ABX Pentra 80 / ABX Pentra XL80: XAA458 (1209109458) ................. 5
1.6. Motor board: XAA459BS (1209109459) ......................................................................................... 6
1.7. Solenoid board: XAA478AS (1209109478) ..................................................................................... 6
1.8. Main board for PXL80 / PXLR: XAA694 (1209102694/1209101694) .............................................. 7
1.9. CIS OD Amplifier board for Pentra XLR: XAA695A (1209101695) .................................................. 8
1.10. Hub board for Pentra XLR or Pentra XL80: XAA705A (1209101705) ........................................... 8
3. Connectors .................................................................................................................. 10
4. Fan ................................................................................................................................ 12
5. Synoptics ..................................................................................................................... 12
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 3-1
Electric & Electronic Principles
Technical Manual
3-2 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Electric & Electronic Principles
Boards
1. Boards
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 3-3
Electric & Electronic Principles
Boards
TP45
R248
Optical Gain Gain TP9 TP7 TP5
RBC/PLT WBC/BASO
60V_LMNE
60V_WBC
60V_RBC
TP1
LMNE
Resistive Gain
Resistive Threshold
TP4 TP19
TP2 TP13 TP12
R159 TP15
TP30
Threshold
R160
R136
R158
TP55
Optical
J44
bench fan
Vertical
TP66
carriage
motor current R114
U48
External reset E1
TP65 U49 U50
Carriage
Board J1 R126 J6
J46
U53
Motors J3
Board
J38
LEDS
Board U55 U56
J42
Emergency
J19
posit° assy
U59
U63 U60
U64
Valves
#1 - 6
J16 J17
Valves
J30
Draining #1
R177 R171 R78 R151
Valves Draining #2
DS6 TP48
#24 - 30 detection
U70
J18 J14 Counting
J41
U74 DS8
Valves
TP68 TP46 detection
DS9
#31 - 35
J28
Transfer
J37
DS10 detection
TP67
Valves J4
#13 - 19 TP42 TP40 TP41 TP43
J27
TP50
TP51 J5
J33
RS232#3
Internal
Barcode
*: If equipped with Reagent Diluent level Room Heating room Waste level Power Power supply of barcode reader on J5
Takasago valves Heater detection Heater door detection detection supply Jumper on E2= VCC on J5-8 (by default)
Technical Manual
3-4 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Electric & Electronic Principles
Boards
1.5. LMNE amplifier board for ABX Pentra 80 / ABX Pentra XL80:
XAA458 (1209109458)
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 3-5
Electric & Electronic Principles
Boards
J1 J2
DS1
+24V
Motor Reagent
J3
J4
U1 syringe
Home
Reagent syr. motor current
U2 R16
TP1 Motor
Injection
J5
J6
U3
Home syringe
Injection syr. motor current
TP10 R23
TP2
Motor Counting
J7
J8
U4
Home syringe
Counting syr. motor current
U5 R38
TP3 Motor Draining #2
J9
U6 J10 Home syringe
Draining#2 syr. motor current
R50
TP4
Home
J11
Sample horizontal
J12
U8 carriage
Carriage motor current Motor
U7 R59
TP5
Motor Draining #1
J13
J14
U9 syringe
Home
Draining#1 syr. motor current
R73
J15
TP6 Home
Rack transfer
J16
U11 Motor
U10
Transfer motor current
Rack transfer & R92
J17
TP7
loading switches
J18
U15
U14
TP9 R122
Mixer
J20 J21 J22 J23 motor current
Technical Manual
3-6 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Electric & Electronic Principles
Boards
Laser
safety switch
Laser
system
5V power
supply harness
Optical
bench fan
Sampling
syringe
motor current
Vertical
carriage
motor current
Carriage
Board
Motors
Board
LEDs
Board
Emergency
posit° assy
Valves
#1 - 6
Valves
#7 - 12 Draining #1
detection
Draining #2
Valves detection
#24 - 30
Counting
Valves detection
#31 - 35
Transfer
detection
Valves
#13 - 19
+24V RS232#2
Valves PC panel
#36 - 37 +5V +15V -15V
Shutter
solenoïd
+ Valves RS232#3
Internal
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 3-7
Electric & Electronic Principles
Boards
1.10. Hub board for Pentra XLR or Pentra XL80: XAA705A (1209101705)
Technical Manual
3-8 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Electric & Electronic Principles
Flat cables
2. Flat cables
List
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 3-9
Electric & Electronic Principles
Connectors
3. Connectors
List
Reference From to
CBC008A
J5 on Main board (xaa456b) Tube barcode reader
(1201681008)
XBA322D External reagent waste rear
Waste level detection sensor
(1209114322) connector
XBA342A Extension cable connector
Horizontal carriage motor home
(1209111342) (xba469a)
XBA356A
Frame ground Ground shielding (Pentra XLR)
(1209111356)
XBA387A
J31 on main board (xaa456b) Heater/sensor for reagent heating coil
(1209112387)
XBA389B
J26 on main board (xaa456b) DIL1/HGB chamber
(1209112389)
XBA393A
J44 on main board (xaa456b) Upper fan
(1209111393)
XBA396A
J1 on carriage board (xaa428c) Vertical carriage home infrared cell
(1209111396)
XBA398C
J23 on main board (xaa456b) RBC chamber
(1209113398)
XBA398C
J24 on main board (xaa456b) WBC chamber
(1209113398)
XBA399C
J22 on LMNE amplifier board LMNE flowcell
(1209113399)
Tube piercing needle high switch
XBA427A Tube piercing motor (B3)
J19 on main board (xaa456b)
(1209111427) Tube piercing needle low switch
Tube recognition switches
Rack loading home switch
Rack counting #1 switch
Rack counting #2 switch
Rack output tray nearly full switch
XBA428A Transfer cover open switch
(1209111428) Transfer end check switch
J17 on motor board (xaa459b) Transfer check #1 switch
(schematic not Transfer check #2 switch
available) Tube presence check switch
Tube height position switch
Emergency door status switches
Emergency door solenoïd harness (XBA471A: Solenoïd
molex 230 mm)
XBA431A J37 on main board (xaa456b) LMNE transfer infrared cell
XBA431A J41 on main board (xaa456b) Counting detection infrared cell
XBA431A J29 on main board (xaa456b) Draining #2 infrared cell
XBA431A J30 on main board (xaa456b) Draining #1 infrared cell
XBA432A
J15 on motor board (xaa459b) Rack transfer motor home
(1209111432)
XBA432A
J22 on motor board (xaa459b) Tube mixer µstep motor home
(1209111432)
Technical Manual
3 - 10 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Electric & Electronic Principles
Connectors
Reference From to
XBA432A
J23 on motor board (xaa459b) Grabber status infrared cell
(1209111432)
XBA458A
J2 on carriage board (xaa428c) Sample syringe home switch
(1209111458)
XBA459A
J8 on motor board (xaa459b) Counting syringe home switch
(1209111459)
XBA459A
J10 on motor board (xaa459b) Drain syringe #2 Home switch
(1209111459)
XBA460A
J4 on motor board (xaa459b) Reagent syringe home switch
(1209111460)
XBA460A
J6 on motor board (xaa459b) LMNE syringe home switch
(1209111460)
XBA461A
J27 on main board (xaa456b) Chamber heating door security switch
(1209111461)
XBA461A
J14 on motor board (xaa459b) Drain syringe #1 home switch
(1209111461)
XBA462A
J3 on carriage board (xaa428c) Vertical carriage motor (B1)
(1209111462)
XBA462A
J4 on carriage board (xaa428c) Sample syringe motor (B2)
(1209111462)
XBA463A
J18 on motor board (xaa459b) Rack loading motor (B4)
(1209111463)
XBA463A
J16 on motor board (xaa459b) Rack transfer motor (B5)
(1209111463)
XBA464A
J7 on motor board (xaa459b) Counting syringe motor (B9)
(1209111464)
XBA464A
J9 on motor board (xaa459b) Draining syringe #2 motor (B8)
(1209111464)
XBA465A
J19 on motor board (xaa459b) Tube mixer µstep motor (B6)
(1209111465)
XBA465A
J13 on motor board (xaa459b) Drain syringe #1 motor (B12)
(1209111465)
XBA465A
J3 on motor board (xaa459b) Reagent syringe motor (B11)
(1209111465)
XBA465A
J5 on motor board (xaa459b) LMNE syringe motor (B10)
(1209111465)
XBA469A Horizontal carriage motor home harness (XBA432A:
J11 on motor board (xaa459b)
(1209111469) Infrared photocell 600 mm (2 tabs))
XBA470A
J21 on motor board (xaa459b) Mixer head locking solenoïd
(1209111470)
XBA471A
Rack switches harness (xba428a) Emergency door solenoïd
(1209111471)
XBA478A
J12 on motor board (xaa459b) Horizontal carriage motor (B7)
(1209112478)
XBA486A
J38 on main board (xaa456b) LEDs extension cable connector (XBA487A)
(1209111486)
XBA487A
XBA486A J1 on LEDs board (xaa429a)
(1209111487)
XBA488A
J47 on main board (xaa456b) External reagent waste rear connector
(1209111488)
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 3 - 11
Electric & Electronic Principles
Connectors
Reference From to
XBA572A
J2 on Main board (xaa694a) Temperature sensor (Pentra XLR)
(1209111572)
XBA645A
DAD136A XBA646A
(1209111645)
LV# 36
XBA646A XBA646A
LV# 37
XBA838A
J4 on CIS OD Amplifier board J100 on ADP Preamplifier board (Pentra XLR)
(1209111838)
XBA839A
J55 on Main board (xaa694a) J5 on CIS OD Amplifier board (Pentra XLR)
(1209111839)
XBA840A
J8 on Main board (xaa694a) Shutter solenoïd, EV39 and EV39 (Pentra XLR)
(1209111840)
XBA843A
J2 on CIS OD Amplifier board Flowcell anode (Pentra XLR)
(1209111843)
XBA844A
Ground Flowcell cathode (Pentra XLR)
(1209111844)
XBA845A
Ground Tube grounding (Pentra XLR)
(1209111845)
XDA605A
J28 on main board (xaa456b) Diluent reservoir
(1209131605)
XDA813B
J33 on main board (xaa456b) Heater/fan/sensor for chamber heating
(1209132813)
Optical bench lamp
DBN006C J45 on the main board
Power supply module
(1202163008) J3 on the PC panel supply board
J2 on the motor board
Technical Manual
3 - 12 Ref: 1300018562- Int. Ref. Doc.: RAA022GEN
Electric & Electronic Principles
Fan
4. Fan
By default, the fan is engaged when the temperature is over 23°, or when the laser interlock sensor is triggered.
The reason it has been set that way is that the ABX Pentra 80 and the ABX Pentra XL80 require that the fan is
always active. Since the ABX Pentra 80 and the ABX Pentra XL80 do not have a laser bench, there is no laser
interlock sensor, and the fan is therefore always engaged. For the Pentra XLR, the fan is engaged only when
the temperature is over 23° or if the laser bench cover is open.
5. Synoptics
See the synoptic diagrams on the next page.
Technical Manual
Ref: 1300018562- Int. Ref. Doc.: RAA022GEN 3 - 13
PC
RS232 RS232
POWER
HOST INTERNAL
SUPPLY
PRINTER COM 3
VGA USB
Analysis Cycle Technology
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4-1
Analysis Cycle Technology
Analysis cycle description (principles)
Technical Manual
4-2 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Analysis cycle description (principles)
■ Translation
of the sampling carriage over the
WBC/BASO counting chamber.
■ Positioningthe needle point into a tangential
flow of 2 mL of ABX Basolyse II and
synchronised distribution of 10 µL of blood.
■ Bubbling.
Dilution
Blood Volume: 10 µL
ABX Basolyse II Volume: 2000 µL
Dilution Rate: 1/200
■ Translation
of the sampling carriage over the
LMNE/RET mixing chamber.
■ Positioning the needle point into a tangential
flow of 1 mL of ABX Eosinofix and synchronised
distribution of 25 µL of blood.
■ Bubbling.
Dilution
Blood Volume: 25 µL
ABX Eosinofix Volume: 1000 µL
ABX Diluent Volume: 1000 µL
Final Dilution Rate: 1/80
■ Translation
of the sampling carriage over the
Rinse chamber.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4-3
Analysis Cycle Technology
Analysis cycle description (principles)
HGB:
■ Distribution of 0.4 mL of ABX Diluent, then
distribution of 0.36 mL of ABX Lysebio into the
DIL1 chamber during bubbling.
■ RBC/PLT counting and HGB measurement.
■ WBC/BASO chamber rinsing: Draining and
filling with 1 mL of ABX Cleaner and 1.5 mL of
ABX Diluent.
Technical Manual
4-4 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Analysis cycle description (principles)
■ Translation
of the sampling carriage over the
DIL1 chamber (first dilution).
■ Distribution
of 0.88 mL of ABX Diluent during
the descent of the needle at the bottom of the
chamber.
■ Distribution of 10 µL of blood when the needle
is at the bottom of the chamber.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4-5
Analysis Cycle Technology
Analysis cycle description (principles)
■ Translation
of the sampling carriage over the
WBC/BASO counting chamber.
■ Positioning the needle point into a tangential
flow of 2 mL of ABX Basolyse II and
synchronised distribution of 10 µL of blood.
■ Bubbling.
Dilution
Blood Volume: 10 µL
ABX Basolyse II Volume: 2000 µL
Dilution Rate: 1/200
■ Translation
of the sampling carriage over the
LMNE/RET mixing chamber.
■ Positioning the needle point into a tangential
flow of 1 mL of ABX Eosinofix and synchronised
distribution of 25 µL of blood.
■ Bubbling.
Dilution
Blood Volume: 25 µL
ABX Eosinofix Volume: 1000 µL
ABX Diluent Volume: 1000 µL
Final Dilution Rate: 1/80
Technical Manual
4-6 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Analysis cycle description (principles)
Dilution
Initial Blood Volume: 10 µL (1/200)
Reagent Volume (RBC chamber): 2500 µL
Final Dilution Rate: 1/10000
HGB:
■ Distribution of 0.4 mL of ABX Diluent, then
distribution of 0.36 mL of ABX Lysebio into the
DIL1 chamber during bubbling.
■ RBC/PLT counting and HGB measurement.
■ WBC/BASO chamber rinsing: Draining and
filling with 1 mL of ABX Cleaner and 1.5 mL of
ABX Diluent.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4-7
Analysis Cycle Technology
Analysis cycle description (principles)
Technical Manual
4-8 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Analysis cycle description (principles)
■ Distribution
of 2.5 mL of ABX Fluocyte in the
LMNE/RET chamber.
Dilution
Initial Blood Volume: 10 µL (1/200)
Reagent Volume (RBC chamber): 2500 µL
Final Dilution Rate: 1/10000
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4-9
Analysis Cycle Technology
Analysis cycle description (principles)
■ RBC counting.
■ Translation
of the sampling carriage over the
DIL1 chamber.
■ Sampling of 28 µL of the first dilution.
■ Translation
of the sampling carriage over the
LMNE/RET chamber.
■ Distribution of the 28 µL of first dilution.
■ Bubbling.
Dilution
Initial Blood Volume from Dil1: 28 µL (1/200)
ABX Fluocyte Volume: 2500 µL
Dilution rate: 1/89
Temperature: 35°C
Incubation time: 50 seconds
Final Dilution Rate: 1/15000
■ Translation
of the sampling carriage over the
Rinse chamber.
Technical Manual
4 - 10 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Analysis cycle description (principles)
Dilution
■ Final Dilution Rate: 1/15000
■ Dilution
transfer trough the LMNE/RET flowcell
and counting during 18 seconds.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4 - 11
Analysis Cycle Technology
Measuring principles
2. Measuring principles
3 µL discarded
3 µL discarded
Technical Manual
4 - 12 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Measuring principles
Air
22 µL discarded
3 µL discarded
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4 - 13
Analysis Cycle Technology
Measuring principles
2.2.1. RBC/PLT
Measurement of impedance variation generated
by the passage of cells through a calibrated
micro aperture.
Results
Number of cells counted per volume unit x
calibration coefficient
Histograms
RBC: Distribution curves on 256 counting
channels from 30 fL to 300 fL.
Technical Manual
4 - 14 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Measuring principles
RBC/PLT Dilutions
**: Two successive dilutions are carried out: Primary Dilution for RBC and PLT:
Blood: 10 µL
Vol. ABX Diluent: 1700 µL Dilution: 1/170
Results
Final HGB result represents: absorbance value obtained x coefficient of calibration.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4 - 15
Analysis Cycle Technology
Measuring principles
Technical Manual
4 - 16 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Measuring principles
Results
WBC: The number of cells counted per volume during a specified period of time X the WBC coefficient of
calibration.
BAS: The proportion of BAS from total WBC (Basophils and WBC nuclei) counted per volume during a specified
period of time X the WBC calibration coefficient.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4 - 17
Analysis Cycle Technology
Measuring principles
Technical characteristics of the WBC counts during the acquisition of the matrix
Impedance with
Initial blood volume: 25 µL Method:
hydrofocus
Vol. ABX Eosinofix: 1000 µL Aperture diameter: 60 µm
Vol. ABX Diluent: 1000 µL Flow diameter: 42 µm
Final dilution rate: 1/80 Injection duration 12 seconds
Temperature of
35°C Volume injected 72 µL
reaction:
Incubation duration: 12 seconds
Technical Manual
4 - 18 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Measuring principles
Poorly-stained (agranular)
Low absorbance
cell in the flowcell
Results
From these measurements, a matrix is drawn up with volumes on the X-axis and optical transmission on the Y-
axis. The study of the matrix image permits the clear differentiation of 4 out of 5 leukocyte populations. As a
matter of fact, the basophil population is very small compared to the others.
■ LYMPHOCYTES: The lymphocytes being small with regular shapes, are positioned in the lower part of both
the optical axis and volume axis. Normal lymphocyte populations are generally observed with a good volume
homogeneity. The far left side of the lymphocyte zone should normally be empty, however if small
lymphocytes are present, small populations may exist in this area. The presence of platelet aggregates is
detected by a distribution pattern which moves from the origin of the matrix (background zone) into the
lymphocyte zone. The NRBCs with their cytoplasmic membranes lysed like the erythrocytes, will have their
nuclei situated to the far left side of the lymphocyte zone.
■ MONOCYTES: The monocytes, being cells with large kidney shaped nuclei and a large non-granular
cytoplasm, will neither be scattered nor absorb a large amount of light. They will therefore be positioned in the
lower part of the optical axis but clearly to the right of the volume axis.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4 - 19
Analysis Cycle Technology
Measuring principles
■ NEUTROPHILS: The neutrophils, with their cytoplasmic granules and their generally segmented nuclei, will
scatter light depending on their morphological complexity. A hypersegmented neutrophil will give an
increased optical response with respect to a young neutrophil population which will be in the upper position
of the optical axis depending on the presence of segmentation and/or granules.
■ EOSINOPHILS: With reagent action on cytoplasmic membranes, the leukocytes keep their native size and
only eosinophils are colored for optical separation. Eosinophils will be situated in the upper part of the optical
Y-axis due to their strong absorbance qualities and their size, which is nearly equivalent to large neutrophils.
■ Additional parameters: LIC (Large Immature Cells) and ALY (Atypical Lymphocytes) complete the panel
available on the matrix.
The immature granulocytic cells are detected by their larger volumes and by the presence of granules which
increase the intensity of the scattered light. Therefore, cells such as metamyelocytes will be found clearly to
the right of the neutrophils and nearly at the same level. Myelocytes and promyelocytes will be found in
saturation position on the far right of the matrix. These last three populations will be counted as LIC (Large
Immature Cells) and their given results are included in the neutrophil value. The blast cells will be found
generally to the right of the monocytes, and, as such, will increase the LIC count. Small blasts will be found
between the normal lymphocytes and monocytes. Platelets and debris from erythrocyte lysis represent the
background noise population located in the lower left area of the matrix. Most of the population partition
thresholds are fixed and give the limits of the morphological normality of leukocytes. Changes in the
morphology of a population will be expressed on the matrix by a shifting of the corresponding population.
A Blast alarm is generated from increased counts within the LIC area. This is correlated with Blast detection
on the Basophil curve.
Large lymphocytes are detected in the ALY (Atypical Lymphocytes) zone where reactive lymphoid forms,
stimulated lymphocytes and plasmocytes are also to be found.
Technical Manual
4 - 20 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Measuring principles
2.4.1. Technology
■ The instrument samples 35 µL of blood and injects 10 µL of it in the DIL1 chamber. It then mixes it with 1.7 mL
of ABX Diluent.
■ Then the instrument samples 28 µL of diluted blood and mixes it with 2.5 mL of ABX Fluocyte.
This reagent contains a fluorescent stain which is specific to nucleic acids: thiazol orange (Thiazol orange is
a patented product from Becton Dickinson San Jose, CA, USA).
■ The dilution is warmed at 35°C for 50 seconds. The stain molecules enter through the cell membrane and fix
the ribonucleic acid molecules.
■ Thisbinding increases the fluorescence (enhancement: ~3000x). After 50 seconds, the solution is transferred
to the laser optical bench to be measured.
■ The laser optical bench simultaneously measures the fluorescence of the cells passing through the measuring
point into the flowcell, and volume by impedance.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4 - 21
Analysis Cycle Technology
Measuring principles
Technical Manual
4 - 22 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Measuring principles
BIBLIOGRAPHY
1- Koepke JF, Koepke JA. Reticulocytes. Clinical lab. Haemat. (1986) 8:169-179.
2- Davis BH. Immature Reticulocyte Fraction (IRF): By any name, a useful clinical parameter of erythropoietic
activity. Laboratory Hematology (1996) 2:2-8
3- Davis BH, Bigelow NC. Flow cytometric reticulocyte quantification using thazole orange provides clinically
useful reticulocyte maturity index. Arch. Pathol. Lab. Med (1989) 113:684-689
4- Davis BH, Bigelow NC, Ball ED, Mills L, Cornwell G. Utility of flow cytometric reticulocytes quantification as
a predictor of engraftment in autologous bone marrow transplantation. Am J Hematol (1989) 32:81-87.
5- B.H. DAVIS, M. DICORATO, N.C. BIGELOW, M.H. LANGWEILER; Proposal for standardization of flow
cytometric reticulocytes maturity index measurements. Cytometry 14:318 1993
6- B.H. DAVIS ET AL. Flow Cytometric Reticulocyte Analysis Multiinstitutional Interlaboratory Correlation Study.
Am. J. Clin. Pathol. oct. 1994, vol. 102: 468-477
7- G.F. RIEDLER, R. ZINGG; Tabulae Haematologicae, Rocom, 23-30.
8- Clin. Lab. Haem. 1999, 21, 3-10 JX CORBERAND, C SEGONDS, A M FONTANILLES, JP CAMBUS, G
FILLOLA, P LAHARRAGUE. Evaluation of the PENTRA 120 Haemaotogy Analyser in a university hospital setting
9- Eur J Clin Chem Clin Biochem 1997; 35(2): 105-111 Clinical Performance of Leukocyte Differential on the
HORIBA ABX Pentra 120 Haematological Analyzer Giuseppe Lippi, Marco Nicoli, Nereo Modena and
Giancesare Guidi.
10- Hematol Cell Ther (1999) 41: 47-50 Evaluation of the monocyte counting by the Pentra 120. Comparison
with the manual method and fluoro-flow cytometry. M. Morillon, J. Maslin, J.J. De Pina, F.J. Louis and G. Martet
11- Clin Chem Lab Med 2000; 38(3):245-249 Determination of reticulocytes: Three methods compared,
Rüdiger Siekmeier, Alexa Bierlich and Werner Jaroß
12- Journal of hematology ISSN 0390-6078 Assessment of hematologic progenitor engraftment by complete
reticulocyte maturation parameters after autologus and allogeneic hematopoietic stem cell transplantation.
Antonio Torres, Joaquin Sánchez, Didier Lakomsky, Josefina Serrano, Miguel A. Alvarez, Carmen Martin,
Carmen Valls, Lourdes Nevado, Antonia Rodriguez, Javier Casaño, Francisco Martinez, Pedro Gómez.
13- Laboratory Hematology 7:75-80 Comparison of the Automated Reticulocyte Counts and Immature
Reticulocyte Fraction Measurements Obtained With the HORIBA ABX PENTRA 120 Retic Blood Analyzer and
the Sysmex XE-2100 Automated Hematology Analyzer. CAROL BRIGGS, DONA GRANT, SAMUEL J. MACHIN
14- American Journal of clinical pathology. Automated Reticulocyte Counting and Immature Reticulocyte
Fraction Measurement. Comparison of HORIBA ABX PENTRA 120 Retic, Sysmex R-2000, Flow Cytometry, and
Manual counts. Francis Lacombe, MD, PhD, Laurent Lacoste, Jean Philippe Vial, MD, Alex Briais, Josy Reiffers,
MD, Michel R. Boisseau, MD, and Philippe Bernard, MD.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4 - 23
Analysis Cycle Technology
Raw counts
3. Raw counts
The raw counts are printed at the bottom of the Report printout.
■ WBC1 / WBC2: Number of WBCs counted in WBC/BASO chamber during the first and second count.
■ RBC1 / RBC2: Number of RBCs counted in the RBC/PLT chamber during the first and second count.
■ HCT1 / HCT2: First and second HCT measurements.
■ PLT1 / PLT2: Number of PLT counted in RBC/PLT chamber during the first and second count.
■ HGBB: Measurement of the HGB blank.
■ HGBM: HGB measurement.
■ LYM / MON / NEU / EOS: Number of cells located in their respective population areas.
■ NO / LL / RM / LN / FLN / FMN / FNE / RN: Number of cells located in their respective flag areas.
■ ALY / C127: Number of cells located in their respective areas.
■ LIC: Number of cell located on RN + RM + C127
C127
EOS
NO
FNE
RN
LN NEU
N
FM
FLN
MON RM
LL LYM
ALY
■ WBC: WBCs used for for the LMNE and BASO balance adjustments.
Technical Manual
4 - 24 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Analysis Cycle Technology
Raw counts
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 4 - 25
Analysis Cycle Technology
Raw counts
Technical Manual
4 - 26 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Software Versions
5 - Software Versions
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 5-1
Software Versions
Service software overview
Gain Adjustment
LMNE Adjustment
Aperture Current
Pulse Adjustment
Laser Adjustment
RET Adjustment
Autoclean Technical Measurement
Thermic adjustment
M.D.S.S. adjustment
Liquid sensor adjustment
Vacuum control
Sampler adjust.
Bubbling
Gains
LMNE Calibration
Technician Menu Cycles
Calibration Coefficients
Blank Values
System Tools
Burn in cycle
Barcode Setup
Others
Initialization
Service Miniclean Check Motors
Check Valves
Check Sensors
Sampler Test
Holder Adjustment
Rack Adjustment
Mechanical
Drain Chambers
Prime Cycles
Super User Menu Unprime Cycle
Clean Cycles
Hydraulic
Cycles Counter
Run Park Syringes Position
Run Maintenance Carriage Pos.
Calibration Coefficients
Concentrated Cleaning Others
Technical Manual
5-2 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Software Versions
Windows Explorer Access
3. Software releases
First release of technical manual: V1.0.1
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 5-3
Software Versions
Software releases
Technical Manual
5-4 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Alarms & Errors
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 6-1
Alarms & Errors
Analyzer error messages
Technical Manual
6-2 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Alarms & Errors
Transfer error messages
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 6-3
Alarms & Errors
STAT mode error messages
Technical Manual
6-4 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Alarms & Errors
User error messages
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 6-5
Alarms & Errors
Analyzer internal error types messages
Technical Manual
6-6 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Maintenance
07 - Maintenance
1. Maintenance .................................................................................................................. 2
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 7-1
Maintenance
Maintenance
1. Maintenance
1.1. Introduction
No special adjustment or maintenance has to be done on your equipment if the recommended startup
and shutdown procedures are explicitly respected.
Refer to the user manual for more details on the daily rinsing and cleaning of the system.
Technical Manual
7-2 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Maintenance
Prevention against the electrostatic discharges
■ An antistatic ground mat is one of a number of antistatic devices designed to help eliminate static
electricity. It has a conductive material embedded within the mat which collects the static. The mat
would need to be grounded (earthed). This is usually accomplished by plugging it into the grounded
line in an electrical outlet. It is important to discharge at a slow rate, therefore a resistor should be used
in earthing the mat. This ground mat allows you to connect an antistatic wrist strap.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 7-3
Maintenance
Prevention against the electrostatic discharges
- A wrist strap:
- A crocodile clip:
Technical Manual
7-4 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Maintenance
Prevention against the electrostatic discharges
■ Install
the wrist strap on the first ground
cord and connect the first ground cord to
the second snap of the antistatic mat.
■ Put the bracelet around your wrist.
To ensure a good contact, ensure that the wrist strap is correctly tightened around your wrist.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 7-5
Maintenance
Maintenance tables and kits
The instrument maintenance is divided into three different maintenance kits. The maintenance schedule depends on the
number of cycles per day. Follow the table to know the part number of the kits that need to be used.
Technical Manual
7-6 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Maintenance
Maintenance kits
4. Maintenance kits
XEC236AS (1219179236)
XEA486FS (1209159486)
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 7-7
Maintenance
Maintenance kits
Technical Manual
7-8 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Maintenance
Maintenance procedures
5. Maintenance procedures
Maintenance and adjustments required for this instrument are divided into procedures
according to the concerned assemblies. This should make any update easier as all
interventions can be done with the corresponding procedure on its own.
Each procedure has to be read entirely before starting intervention.
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 7-9
Maintenance
Maintenance procedures
Technical Manual
7 - 10 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Maintenance
Maintenance procedures
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 7 - 11
Maintenance
Maintenance procedures
Technical Manual
7 - 12 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
RAS324G
Installation
RAS324G: Installation
■ Concerns
■ Working conditions
■ Instrument installation
■ First start of the instrument
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
1. Instrument Installation
1.1. Unpacking
Package list
Technical Manual
RAS324G - 2 RAA022
Installation
■ Install
the instrument lifting handles in the
appropriate location.
Technical Manual
RAA022 RAS324G - 3
Installation
■ Lift the instrument using the four handles up on to the installation bench.
■ Open the thermic panel (righthand side of the instrument).
■ Remove the left panel door then the upper cover to access the optical bench.
Technical Manual
RAS324G - 4 RAA022
Installation
■ Check that the power supply module voltage of the instrument corresponds to the nominal voltage supply of
the laboratory and the country.
■ Connect the power supply cable.
■ Connect the printer.
Connections until the CCC039A (1201841039)
computer reference
Technical Manual
RAA022 RAS324G - 5
Installation
Power requirements:
- Analyzer: 250 W
- Laser printer: 350 W
If the printer is connected with a cable that was not supplied by HORIBA Medical, make sure this
cable is shielded.
Technical Manual
RAS324G - 6 RAA022
Installation
Keyboard
Mouse
5 - LAN
1- USB port
Technical Manual
RAA022 RAS324G - 7
Installation
A ferrit has been installed in factory on the network wire (2 turns) close to the connector on the
ABX Pentra XL80 and the Pentra XLR.
■ Use the "Luer" connectors, straws and stoppers supplied within the installation kit to connect integrated
reagents, diluent and wastes.
Technical Manual
RAS324G - 8 RAA022
Installation
1 2
1 - ABX Cleaner
2 - ABX Basolyse II
3 4 5
3 - ABX Eosinofix
4 - ABX Lysebio
6 - Waste container
7 - ABX Diluent
6 7
Risk of erroneous results if the diluent container is installed more than 80 cm (31.5 in.) below the
instrument.
Diluent input tubing: Cristal 3.2 x 6.4 / 2 meters (80 in.) maximum
Waste output tubing: Cristal 4x6 / 2 meters (80 in.) maximum
Technical Manual
RAA022 RAS324G - 9
Installation
2.1. Racks
The racks are identified by means of barcode labels. Prepare two or three racks to use during the installation.
These labels have to be placed on the racks as follows:
■ Front side:
■ Rear side:
■ Check the type of the keyboard used (Qwerty or Azerty). From "Settings > System > Local Settings", press
"Edit" and then press "Change Input Locales" (ABX Pentra 80) or "Keyboard Settings" (ABX Pentra XL80
and Pentra XLR).
Technical Manual
RAS324G - 10 RAA022
Installation
■ Check and adjust (if necessary) the date and time. From "Settings > System > Local Settings", press "Edit"
and then press "Change Date / Time".
■ Adjust Date & Time then press Validate to save changes.
■ Prime reagents using the barcode reader to enter the lot number and to update the expiry date automatically.
Go to the "Status" menu to do so.
■ Run a priming cycle of all reagents to be sure that they all are correctly primed by going to "Service >
Super User Menu > Hydraulic System > Prime cycles" and pressing "ALL REAGENTS".
■ Make sure that all the panels are fitted on the instrument.
■ Go to "Service > Super User Menu > Mechanical System > Check Sensors" and make sure that all the
sensors are green.
■ Check that the racks are properly transfered to the reception tray from "Service > Technician Menu > Gains
> Sampler adjust", and then press "Check Rack Transfer Moving".
Technical Manual
RAA022 RAS324G - 11
Installation
2.7. Startup
■ Perform the "Check up after intervention" procedure (Refer to RAS340D: Check up after intervention).
■ After
the installation is completed, save (either on the floppy disk provided with the system or on a USB key)
and print the instrument configuration by following procedure RAS357E: Save and restore settings.
This step has to be performed on ABX Pentra XL80 and Pentra XLR from software version V2.2.0.
■ Go to Services > Technician Adjustments > Others > Blank Values / Type.
■ Verify
that the online help is installed on the instrument in the appropriate language. If not, install it using the
"Documentation CD-ROM" supplied with the instrument.
Technical Manual
RAS324G - 12 RAA022
RAS325F
6 month maintenance
■ Concerns
■ Rinsing block
■ Reagent syringe
■ Draining syringe 1
■ Draining syringe 2
■ Counting syringe
■ Instrument & Filter cleaning
■ Axis lubricating
■ Required tools
■ Hexagonal keys
■ Dynamometric screwdrivers A302, A301, A300
■ Cutting pliers
■ Flat screwdriver
■ Cruciform screwdrivers
■ Torx keys
■ Required products
■ ABX Minoclair
■ Ethanolor essence C
■ XEA821A (1209151821): Vactra oil 100 mL
■ Intervention time
■2 hours
■ Frequency
Technical Manual
RAS325F - 2 RAA022
6 month maintenance
1. Rinsing block
O-ring replacement
1 x FAA013A (1202401013), 1 x FAA055A (1202401055), 1 x FAA057A (1202401057), 1 x GBG296A
(1203201296), 1 x GBG297A (1203201297), 1 x GBG292A (1203201292)
GBG296A
FAA055A (1203201296)
(1202401055)
GBG297A
(1203201297)
GBG292A
(1203201292)
FAA057A
(1202401057)
Technical Manual
RAA022 RAS325F - 3
6 month maintenance
2. Reagent syringes
O-ring replacement
1 x FAA073A (1202401073) & 4 x XDA622A (1209131622)
1
■ On a piece of absorbant paper open the
reagents syringe (9 x CHC M3x12 screws, 2 x
FX M3x12 screws).
2
■ Replaceall O-rings and washers with the new
ones supplied in the 6 month maintenance kit.
■ Use one drop of silicon grease between two
fingers to lubricate the O-rings.
■ Reassemble in reverse order:
8
- 1 - Reagent syringe body GBG033A
(1203201033) 3
- 2 - XDA622A (1209131622) x4
- 3 - Piston GBC030A (1202801030) x4
- 4 - FX M3x12 screws x2
- 5 - CHC M3x12 screws x9
- 6 - Reagent syringe bottom plate 7
- 7 - Hgb lyse reagent piston GBC031A
(1202801031) 4
- 8 - O-ring FAA073A (1202401073) 6
Technical Manual
RAS325F - 4 RAA022
6 month maintenance
Technical Manual
RAA022 RAS325F - 5
6 month maintenance
On Pentra XLR, the Draining syringe 2 and the counting syringe are located in a lower position, as
shown on the pictures below:
■ Switch the instrument off, disconnect the power supply cable then remove the covers (Refer to RAS342D:
Front panel & covers dismantling).
■ Drain the syringe by pushing the piston up and down several times.
■ Open the syringe:
- 4 x CHC M4x16 screws on the old model
- Rotation of the syringe body on the new model.
Technical Manual
RAS325F - 6 RAA022
6 month maintenance
Open
Close
■ Replace the O-ring with the new one supplied in 60 Vthe 6 month maintenance kit.
■ Use one drop of silicon grease between 2 fingers to lubricate the O-ring and the piston.
1-Motor
assembly
2-O-ring
■ Make sure that the washer (4) is present on the (FAA017A) -
old model, then reassemble in reverse order.
3-Anti-extrusion
ring (GBG219A)
4-Washer
(GBG148A)
5-Syringe body
6-CHC M4x16
screws (x4)
The chamber body of the new syringe is directly screwed to the motor block. When the body is
reinstalled on the motor block, it has to be rotated until the mark on the body is located between the
two marks of the motor block.
Technical Manual
RAA022 RAS325F - 7
6 month maintenance
■ Switch the instrument off, disconnect the power supply cable and remove the covers (Refer to RAS342D:
Front panel & covers dismantling).
■ Follow the same procedure as for draining syringe 1 to replace the O-ring.
■ Switch the instrument off, disconnect the power supply cable and remove the covers (Refer to RAS342D:
Front panel & covers dismantling).
■ Follow the same procedure as for draining syringe 1 to replace the O-ring.
Technical Manual
RAS325F - 8 RAA022
6 month maintenance
4. Axes lubrication
Technical Manual
RAA022 RAS325F - 9
6 month maintenance
5. Instrument cleaning.
■ Thermostated compartment
■ Outer surfaces of the instrument (perpex, covers, reagent locations, etc.)
■ Waste connector plug
■ Liquid valve push button
■ Assemblies close to the needle
■ Tube holder assembly
■ Overflow trays
5.2. Procedure
■ Dilute the 9° Cl bleach to 1 part of bleach for 4 parts of deionized water (1/5).
■ Instrument environment must be cleaned.
■ No sponge nor cloth must be used. Only use absorbant paper, and throw it away after use in contamination
bins. For small or sensitive assemblies, use delicate absorbant paper.
■ All assemblies suspected to have been in contact with biohazardous material must be disinfected with diluted
bleach (the stainless steel must be bleached below 30°C).
■ Blood stains or salt marks must be cleaned with spray detergent first.
5.3. Filter
■ Reinstall all the assemblies and set the instrument back to its initial configuration.
Technical Manual
RAS325F - 10 RAA022
6 month maintenance
Technical Manual
RAA022 RAS325F - 11
6 month maintenance
Technical Manual
RAS325F - 12 RAA022
RAS326E
Yearly maintenance
■ Concerns
■ Required tools
■ Hexagonal keys
■ Dynamometric screwdriver A302, A301, A300
■ Cutting pliers
■ Flat screwdriver
■ Cruciform screwdriver
■ Tork keys
■ Required products
■ ABX Minoclair
■ RBC/Plt Latex
■ Acetone
■ Intervention time
■2 hours
■ Frequency
Technical Manual
RAS326E - 2 RAA022
Yearly maintenance
Technical Manual
RAA022 RAS326E - 3
Yearly maintenance
Technical Manual
RAS326E - 4 RAA022
Yearly maintenance
■ Close the tube holder when the message "Please sample LATEX" appears.
Technical Manual
RAA022 RAS326E - 5
Yearly maintenance
Hold the top part of the reservoir so that both tubes are above the diluent reservoir.
Top screws tightening torque: 120 m.Nm (17 Ozf.in).
■ Tighten the 4 diluent reservoir screws according to the tightening torque above.
■ If
the tube under the reservoir body has been disconnected, connect it before installing the diluent reservoir
back to its proper location.
■ Install the diluent reservoir back to its proper location and block the screws.
Technical Manual
RAS326E - 6 RAA022
Yearly maintenance
Technical Manual
RAA022 RAS326E - 7
Yearly maintenance
■ Reassemble the electrode then repeat the same operation for the other electrode.
Technical Manual
RAS326E - 8 RAA022
Yearly maintenance
■ Carefully
dismantle the counting head and
plunge the aperture in distilled water.
■ Clean the chamber and the counting head with liquid soap.
Do not use any sharp objects inside the chamber and/or the aperture as it can cause damage.
Do not manipulate the aperture by using hard instruments. Clean the aperture with a soft paper or,
preferably, between 2 fingers.
It is recommended to reconnect the tubes on the counting head before reassembling the electrode
and the chamber in order to avoid applying constraint on the chamber.
Do not apply too much pressure on the electrode fixation screws as it can break the aperture.
Tightening torque: 400 m.Nm (56.8 Ozf.in).
■ Reassemble the electrode and repeat the same operation with the other counting head.
■ Reassemble in reverse order. Gently push down the chamber assembly when tightening the screws in order
to keep the correct position.
Technical Manual
RAA022 RAS326E - 9
Yearly maintenance
310 mm
320 mm
Loading tray
220 mm
Ejection tray
Technical Manual
RAS326E - 10 RAA022
RAS327F
2 year maintenance
■ Concerns
■ Required tools
■ Hexagonal Keys
■ Dynamometric screwdrivers A302, A301, A300
■ Cutting pliers
■ Flat screwdriver
■ Cruciform screwdrivers
■ Torx keys
■ Required products
■ ABX Minoclair
■ Intervention time
■2 hour
■ Frequency
■ The2 year maintenance kit for ABX Pentra 80/ABX Pentra XL80 XEC282AS (1209179282)
contains pistons, piercing needle and LMNE ABX Pentra 80/ABX Pentra XL80 coaxial.
■ The2 year maintenance kit for Pentra XLR XEC196BS (1209178196) contains piston, piercing
needle and LMNE Pentra XLR coaxial
Technical Manual
RAS327F - 2 RAA022
2 year maintenance
2. Maintenance
GBL0651
(1203500651)
Use new O-rings when you perform the 6 month maintenance at the same time.
Change the sampling needle XDA655ES (1209139655) when you perform the yearly maintenance at
the same time.
Use a dynamometric screwdriver to tighten the syringe. Tightenning torques: 120 mN.m \ 17 Ozf.in.
Technical Manual
RAA022 RAS327F - 3
2 year maintenance
1
■ On a piece of absorbant paper open the
reagents syringe (9 x CHC M3x12 screws, 2 x
FX M3x12 screws).
2
■ Replace the pistons.
■ Replace all O-rings and washers with the new
ones from the 6 month maintenance kit.
■ Use one drop of silicon grease between 2
fingers to lubricate the O-rings.
8
■ Reassemble in reverse order:
3
- 1 - Reagent syringe body
GBG033A (1203201033).
- 2 - XDA622A (1209131622) x4.
- 3 - Piston GBC030A (1202801030) x4.
- 4 - FX M3x12 screws x2. 7
- 5 - CHC M3x12 screws x9.
- 6 - Reagent syringe bottom plate.
- 7 - Hgb lyse reagent piston 4
GBC031A (1202801031). 6
- 8 - O-ring FAA073A (1202401073).
Technical Manual
RAS327F - 4 RAA022
2 year maintenance
■ Switch the instrument off and disconnect the power supply cable.
■ Remove the cover and the right-hand side door to access the draining syringe 1 (Refer to RAS342D: Front
panel & covers dismantling for further details).
■ Put a piece of absorbant paper under the syringe and disconnect all the tubes from the syringe.
■ Remove the syringe.
■ Drain the syringe by pushing the piston up and down several times.
■ Open the syringe
- 4 x CHC M4x16 screws on the old model
- Rotation of the syringe body on the new model.
Technical Manual
RAA022 RAS327F - 5
2 year maintenance
Open
Close
Use the new O-ring when you perform the 6 month maintenance at the same time.
O-ring FAA017A (1202401017) + Washer GBG148A (1203201148) = XDA621A (1209131621)
Technical Manual
RAS327F - 6 RAA022
2 year maintenance
1-Motor
assembly
2-O-ring
■ The chamber body of the new syringe is FAA017A
directly screwed to the motor block. When the (1202401017)
body is reinstalled on the motor block, it has to
be rotated until the mark on the body is located 3-Anti-extrusion
between the two marks of the motor block. ring GBG219A
(1203201219)
4-Washer
(GBG148A)
5-Syringe body
6-CHC M4x16
screws (x4)
■ Follow the same procedure as for the draining syringe 1 to replace the piston.
Technical Manual
RAA022 RAS327F - 7
2 year maintenance
■ Follow the same procedure as for the draining syringe 1 to replace the piston.
XBA399C (1209113399)
Technical Manual
RAS327F - 8 RAA022
2 year maintenance
XBA843A (1209111843)
Technical Manual
RAA022 RAS327F - 9
2 year maintenance
3. Instrument cleaning
■ Concerned assemblies:
- Thermostated compartment
- Outer surfaces of the instrument (perpex, covers, reagent locations, etc.)
- Waste connector plug
- Liquid valve push button
- Assemblies close to the needle
- Tube holder assembly
- Overflow trays
Cleaning:
■ Dilute the 9° Cl bleach to 1 part of bleach for 4 of deionized water (1/5).
■ Instrument environment must be cleaned.
■ No sponge nor cloth must be used. Only use absorbant paper and throw it away after use in contamination
bins. For small or sensitive assemblies, use delicate absorbant paper.
■ All assemblies suspected to have been in contact with biohazardous material must be disinfected with diluted
bleach (the stainless steel must be bleached below 30°C).
■ Blood stains or salt marks must be cleaned with spray detergent first.
■ Reinstall all the assemblies and set the instrument back to its initial configuration.
Technical Manual
RAS327F - 10 RAA022
RAS328D
■ Concerns
■ Required tools
■ Hexagonal keys
■ Clamps
■ Flat screwdriver
■ Torx keys
■ Required products
■ Fungicidal, bacterial, virus killing detergent spray,
non corrosive for metals, non plastic altering.
■ ABX Minoclair
■ 9° bleach
■ Deionized water
■ Distilled water
■ Absorbant paper
■ Intervention time
■1 hour 30 minutes
■ Frequency
■ On request
1. Preparation
■ Switch the instrument on.
■ Open the thermal door (right-hand side of the instrument).
■ Perform a "Concentrated Cleaning" from "Service > Super User Menu > Hydraulic System > Clean
Cycles".
■ Confirm the concentrated cleaning by pressing "OK".
■ When the cycle is finished, switch the instrument off and remove the power supply cable.
■ Remove the instrument covers (Refer to RAS342D: Front panel & covers dismantling).
■ Spray the bactericidal cleaner on all biohazardous areas (assemblies in contact with the biohazardous
material such as instrument covers, tube holder, keyboard, etc.) and wait 10 minutes.
2. Manual decontamination
■ Dilute the 9° Cl bleach to 1 part of bleach for 4 of deionized water (1/5).
■ Instrument environment must be cleaned and decontaminated.
■ No sponge nor cloth must be used. Only use absorbant paper and throw it away after use in contamination
bins. For small or sensitive assemblies use delicate absorbant paper.
■ All assemblies suspected to have been in contact with biohazardous material must be disinfected with diluted
bleach (stainless steel must be bleached below 30°C).
■ Blood stains or salt marks must be cleaned with spray detergent first.
■ Concerned assemblies:
- Outer surfaces of the instrument (perpex, covers, reagent locations, etc.)
- Waste connector plug
- Liquid valve push button
- Assemblies close to the needle
- Overflow tray
■ Put back all the assemblies and set the instrument back to its initial configuration.
Technical Manual
RAS328D - 2 RAA022
Decontamination & rinse
4. Instrument Rinse
Run this procedure before transporting the instrument, after a demonstration or before a long period without
functioning.
■ Run an Instrument general cleaning.
■ Remove the straws from the reagent bottles and put them in an empty bottle.
■ Remove the straw from the diluent cubitainer and put it in the empty bottle.
Technical Manual
RAA022 RAS328D - 3
Decontamination & rinse
■ Go to "Service > Super User Menu > Hydraulic System > Unprime Cycle" and then click "All".
Technical Manual
RAS328D - 4 RAA022
RAS329D
Chambers adjustment
■ Concerns
■ Required tools
■ Hexagonal Keys
■ Required products
■ None
■ Intervention time
■ 15 minutes
■ Frequency
■ On request
If the teflon part of the needle comes into contact with the gauge, the needle will be damaged.
■ Without lifting the needle, gently push the carriage over the WBC/BASO chamber.
Technical Manual
RAS329D - 2 RAA022
Chambers adjustment
■ If
you need to get a better position, free the
chamber support by loosening the 5 screws.
Technical Manual
RAA022 RAS329D - 3
Chambers adjustment
Technical Manual
RAS329D - 4 RAA022
RAS330E
Needle adjustment
■ Concerns
■ Required tools
■ GBG282A key for cover
■ Required products
■ None
■ Intervention time
■ 15 minutes
■ Frequency
■ On request
Make sure that the chamber assy is parallel to the carriage motion before checking or making an
adjustment of the needle position.
Follow RAS329D: Chambers adjustment.
The first dilution is performed in two steps to noticeably improve the dilution accuracy:
■ After blood aspiration and needle rinsing, 0.88 mL of ABX Diluent is distributed in the DIL1/HGB chamber
during the descent of the needle at the bottom of the chamber.
■ Then, 10 µL of blood is distributed when the needle is at the bottom of the chamber.
■ Finally, the needle moves up to the ABX Diluent distribution inlet and distributes 0.80 mL of ABX Diluent.
The needle adjustments can be performed from "Service > Technician Menu > Gains >
M.D.S.S. Adjustment".
Technical Manual
RAS330E - 2 RAA022
Needle adjustment
2. Home adjustment
■ Remove the covers (Refer to RAS342D: Front panel & covers dismantling).
■ Go to "Service > Technician Menu > Gains > M.D.S.S. Adjustment".
■ Press "Check Needle Home".
1 mm
Sampling needle
Notch
■ Use the notch in GBG282A tool (key for cover)
as a gauge to measure the gap.
Technical Manual
RAA022 RAS330E - 3
Needle adjustment
■ Remove the covers (Refer to RAS342D: Front panel & covers dismantling).
■ Go to "Service > Technician Menu > Gains > M.D.S.S. Adjustment".
■ Press "Check Carriage Home".
The carriage home position is factory adjusted and must not be changed. It represents a gap of
2 mm between the tooth wheel and the front of the carriage.
This carriage position is also the sampling position. If you need to readjust the Sampling position,
move the tube holder instead of changing the "Carriage Home" value.
2 mm
■ Check the gap between the tooth wheel and the front of the carriage and press "OK".
■ If the carriage home position needs to be adjusted, change the "Carriage Home" value and then press
"Check Carriage Home" again. Use the MAJ004A gauge to check the gap between the tooth wheel and the
front of the carriage.
By changing the carriage home value, the carriage positions will change. If this position is modified,
increase or decrease the "Dil 1 carriage position" with the same number of steps, not to loose the
other adjustments:
- Increase "Dil 1 carriage position" if you increase "Carriage Home".
- Decrease "Dil 1 carriage position" if you decrease "Carriage Home".
The sampling position in the tube holder will change too. Re-adjust the tube holder position to center
the needle in the tube holder’s sampling hole (Refer to RAS338).
Technical Manual
RAS330E - 4 RAA022
Needle adjustment
3. MDSS adjustment
3.1. Adjustment
■ Disconnect
all tubes connected to the inlets of each chamber (Dil1/HGB-Inlet1; LMNE-Inlet3; RBC-Inlet1
and WBC-Inlet2).
C: Needle position
■ Check the needle depth if necessary:
If the sampling needle is too low, decrease the number of steps of the needle.
If the sampling needle is too high, increase the number of steps of the needle.
■ If any modification has to be done, run again the "Check Carriage Position" from the begining to control
the correct position of the sampling needle.
■ If the adjustment is correct, put the tubing back to the inlet.
■ Press "OK" to continue.
Technical Manual
RAA022 RAS330E - 5
Needle adjustment
.
os
ep
i ag
c ar r po
s.
O e
AS i ag
/B ar r
W BC E c
(2) L MN
(3)
.
os
g ep
a rria
1c
IL
(1 )D
s .
po
ia ge
arr
RB
C
c 2
(4)
■ If
you want to modify only one of the positions,
make sure you increase or decrease the next
4
motions with the same number of steps, so that
you do not loose the other adjustments. 3
Technical Manual
RAS330E - 6 RAA022
Needle adjustment
■ Make sure that the needle does not touch the inner side and bottom of the chamber.
■ Press "OK" until the window closes.
Technical Manual
RAA022 RAS330E - 7
Needle adjustment
1 B
■ If
any modification has been carried out, check
the MDSS adjustment. Refer to MDSS
adjustment.
Technical Manual
RAS330E - 8 RAA022
RAS331E
■ Concerns
■ Required tools
■ Hexagonal keys
■ Voltmeter
■ Required products
■ RBC / PLT latex: 1300019538
■ Intervention time
■ 30 minutes
■ Frequency
■ On request
PCB PCB
P80 PXL80 PXLR
P/N Movex P/N SAP
XAA456A X
XAA456B 1209102456 X X
XAA694A 1209101694 X
XAA694B 1209102694 X X
Technical Manual
RAS331E - 2 RAA022
Mother Board Adjustment
TP45
R248
Optical Gain Gain TP9 TP7 TP5
RBC/PLT WBC/BASO
60 V_LMNE
60 V_WBC
60 V_RBC
TP1
LMNE
Resistive Gain
Resistive Threshold
TP4 TP19
TP2 TP13 TP12
R159 TP15
TP30
Threshold
R160
R136
R158
TP24 TP17 TP33 TP16
TP14
R157
TP21 TP58 TP57
TP31
TP18 TP36 TP54 R133 TP27 Threshold
R135
TP55
Optical
J44
bench fan
Vertical
TP66
carriage
motor current R114
U48
External reset E1
TP65 U49 U50
Carriage
Board J1 R126 J6 BD32
J46
U53
Motors J3 RS232#1
Board
J38
LEDS
Board U55 U56
J42 JTAG
Emergency
J19
posit° assy
U59
U63 U60
U64
Valves
#1 - 6
J16 J17
Valves
J30
Draining #
R151
Valves Draining #
R177 R171 R78
DS6 TP48
#24 - 30 detection
U70
J18 J14 Counting
J41
U74 DS8
Valves
TP68 TP46 detection
DS9
#31 - 35
J28
Transfer
J37
TP50
TP51 J5
J33
RS232#3
Internal
Barcode
*: If equipped with Reagent Diluent level Room Heating room Waste level Power Power supply of barcode reader on J5
Takasago valves Heater detection Heater door detection detection supply Jumper on E2= VCC on J5-8 (by default)
Technical Manual
RAA022 RAS331E - 3
Mother Board Adjustment
Laser
safety switch
Laser
system
Optical
bench fan
Sampling
syringe
motor current
Vertical
carriage
motor current
Carriage
Board
Motors
Board
LEDs
Board
Emergency
posit° assy
Valves
#1 - 6
Valves
#7 - 12 Draining #1
detection
Draining #2
Valves detection
#24 - 30
Counting
Valves detection
#31 - 35
Transfer
detection
Valves
#13 - 19
+24V RS232#2
Valves PC panel
#36 - 37 +5V +15V -15V
Shutter
solenoïd
+ Valves RS232#3
#38 - 39 Internal
Barcode
Reagent Diluent level Heating room Room Laser ambiant Waste level Power
Heater detection door detection Heater T° sensor detection supply
in red: Pentra XLR Only
Technical Manual
RAS331E - 4 RAA022
Mother Board Adjustment
■ The power supply test points are located at the bottom of the main board and are not adjustable.
J16 J17
J30
R151
DS5 TP52
TP47 DS7
J29
R177 R171 R78
DS6 TP48
U70
J18 J14
J41
U74 DS8
TP68 TP46
DS9
J28
J37
DS10
TP67
J4
TP42 TP40 TP41 TP43
J27
TP50
TP51 J5
J33
DS1
DS15 DS16 J45
J31
Technical Manual
RAA022 RAS331E - 5
Mother Board Adjustment
The Aperture voltage test points are located in the right upper part of the main board and are not adjustable.
J26
HGB
R248
60 V_WBC
60 V_RBC
TP1
Aperture voltage
supply
Technical Manual
RAS331E - 6 RAA022
Mother Board Adjustment
■ The threshold test points are located at the top of the main board.
R248
LMNE OD: LMNELMNE CIS: RBC/PLT WBC/BASO
TP4 - R161 TP3 - R160
RBC:
Resistive Threshold
TP13 - R158
PLT R134
Otpical Threshold
Threshold
TP4 TP19
TP2 TP13
R159
RBC
R160
R136
TP3
R161
R158
Threshold
TP14
R157
Threshold
R133
PLT:
R135
TP2 - R159
■ Check and adjust the threshold voltages, if necessary, according to the following table.
■ The threshold test points are located at the top of the main board.
Technical Manual
RAA022 RAS331E - 7
Mother Board Adjustment
Technical Manual
RAS331E - 8 RAA022
Mother Board Adjustment
J26
HGB
R248
TP9 TP7 TP5
Gain
TP1
■ Check between Ground and TP1 that the voltage is in accordance with the value displayed on the screen.
A special cycle allows the dilution of the RBC/PLT latex. A special count program carries out the calculation of
the mean volume in the predefined zones and displays them every 700 ms.
The duration of the measurement cycle is 21 seconds.
■ The adjustments are made with
R133 potentiometer for RBC and
R135 for PLT.
■ Go to "Service > Technician Menu > Measurement\Gains" and press "RBC / PLT Gain".
TP45
R248
60 V_WBC
60 V_RBC
TP1
LMNE
Resistive Gain
WBC/BASO
Resistive Threshold
TP39 Gain
R134 High voltage supply
Otpical Threshold
TP4 TP19
TP2 TP13 TP12
R159 TP15
TP30
Threshold
R160
R136
R158
PLT
R135
TP59
TP44 TP23 TP22
Gain
RBC
Gain
Technical Manual
RAA022 RAS331E - 9
Mother Board Adjustment
■A special cycle allows the dilution of the RBC/PLT latex. A special count program carries out the calculation
of the mean volume in the predefined zones and displays them every 700 ms. The duration of the
measurement cycle is 21 seconds.
■ The adjustment is made using the R134 potentiometer.
■ Go to "Service > Technician Menu > Measurement > Gains" and press "Wbc Gain".
■ Target values (for 1300019538 latex lot): refer to the latest version of the RAN278 technical note.
■ The liquid sensor test points are located at the bottom of the main board.
■ From "Service > Technician Menu > Measurement > Gains > Liquid sensor adjustment", adjust all the
different voltage values according to the following table:
Technical Manual
RAS331E - 10 RAA022
Mother Board Adjustment
Draining #2
R177 R171 R78
U74 DS8
TP68 TP46 detection
DS9
J28
Transfer
J37
DS10 detection
TP67
J4
TP42 TP40 TP41 TP43
J27
TP50
TP51 J5
J33
DS1
DS15 DS16 J45
J31
Draining #1
detection
Draining #2
detection
Counting
detection
Transfer
detection
Technical Manual
RAA022 RAS331E - 11
Mother Board Adjustment
■ The motor current test points are located in the middle of the main board, on the left.
■ Check and adjust the motor current, if necessary, according to the following table.
U45 U46
Sampling TP66
syringe R114
motor current
U48
Sampling
syringe
motor current
Vertical
carriage
motor current
Technical Manual
RAS331E - 12 RAA022
Mother Board Adjustment
■ The laser power setup test point is located on the top of the main board, on the left.
■ Check and adjust if necessary the motor current according to the following table.
Technical Manual
RAA022 RAS331E - 13
Mother Board Adjustment
Technical Manual
RAS331E - 14 RAA022
RAS332D
■ Concerns
■ Required tools
■ Hexagonal keys
■ Voltmeter
■ Required products
■ None
■ Intervention time
■ 15 minutes
■ Frequency
■ On request
For more details about the dismantling of the instrument covers, refer to the procedure RAS342D: Front panel
& covers dismantling.
*: This motor current new value is only applicable on the new motorisation assemblies: in case a new
piston is installed on an old syringe assembly (motor + syringe), the motor current remains 4 V
+/- 0.05 V.
Technical Manual
RAS332D - 2 RAA022
Motor board adjustment
J1 J2
DS1
+24V
Motor Reagent
J3
U1
J4
Home syringe
Reagent syr. motor current
U2 R16
TP1 Motor
Injection
J5
J6
U3
Home syringe
Injection syr. motor current
TP10 R23
TP2
Motor
Counting
J7
J8
U4
Home syringe
Counting syr. motor current
U5 R38
TP3 Motor
Draining #2
J9
J10
U6 syringe
Home
Draining#2 syr. motor current
R50
TP4
Home
J11
Sample horizontal
J12
U8 carriage
Carriage motor current Motor
U7 R59
TP5
Motor Draining #1
J13
J14
U9 syringe
Home
Draining#1 syr. motor current
R73
J15
TP6 Home
Rack transfer
J16
U11 Motor
U10
Transfer motor current
Rack transfer & R92
J17
TP7
loading switches
J18
U15
U14
TP9 R122
Mixer
J20 J21 J22 J23 motor current
Technical Manual
RAA022 RAS332D - 3
Motor board adjustment
Technical Manual
RAS332D - 4 RAA022
RAS333E
Temperature adjustment
■ Concerns
■ Required tools
■ Thermometer
■ Required products
■ None
■ Intervention time
■ 30 minutes
■ Frequency
■ On request
■ Reagent Heating:
- Reference CTN value: Input the CTN value in this field.
- Temperature recorded: Allows to modify the temperature instruction.
- Run Cycle Number: Number of cycles to check temperature adjustment.
This control must be made with the thermal door closed and under stabilised thermic conditions (at
least 5 minutes after startup).
Technical Manual
RAS333E - 2 RAA022
Temperature adjustment
Make sure the thermometer sensor is pushed as deep as possible into the liquid.
■ After the thermometer probe has been installed, close the thermal door.
■ Enter 5 in the "Run Cycles Number" field (in the "Reagents Heating" area of the window) to run 5 cycles.
■ Press "Check Thermic" (in the "Reagents Heating" area of the window) to start controling the temperature.
■ Every time your hear a beeping sound coming from the machine, check that the temperature is around
35°C +/-1°C.
The check is validated if you obtain the right temperature at least 4 times during the control.
If an adjustment is necessary, when the heating block has been replaced for example, perform the
following steps of this procedure.
Technical Manual
RAA022 RAS333E - 3
Temperature adjustment
The heating coil CTN value is located on the CTN wire, close to the main board, connected on J31.
Be sure to read the correct value as the sticker may be up or down.
■ Adjust, as described for the control, the temperature by the mean of the "Temperature recorded" value in
the "Thermic adjustment" window.
■ The "Temperature recorded" value (in the "Reagents Heating" area of the window) must be rounded (e.g.
34, 35, 36, etc.):
- If "Temperature recorded" is increased, reagents pre-heating temperature increases.
- If "Temperature recorded" is decreased, reagents pre-heating temperature decreases.
■ Perform a reagent heating coil check after the adjustment as described at the beginning of the procedure.
■ If required, change the "Temperature recorded" value (in the "Reagents Heating" area of the window) and
check the temperature again until you obtain an acceptable temperature.
Technical Manual
RAS333E - 4 RAA022
Temperature adjustment
■ After the thermometer sensor has been installed, close the thermal door.
■ Wait during 20 minutes.
■ Check that the temperature is around 35°C +/-1°C. If not, perform an adjustment.
Technical Manual
RAA022 RAS333E - 5
Temperature adjustment
The thermostatic compartment CTN value is located on the CTN wire, close to the main board,
connected on J33. Be sure to read the correct value as the sticker may be up or down.
■ Adjust, as described for the control, the temperature by the mean of the "Temperature recorded" value in
the "Thermic Adjustment" window.
■ The "Temperature recorded" value (in the "Thermostatic Compartment" area of the window) must be
rounded (e.g. 34, 35, 36, etc.):
- If "Temperature recorded" is increased, reagents pre-heating temperature increases.
- If "Temperature recorded" is decreased, reagents pre-heating temperature decreases.
■ Press "Check Thermic" in the "Thermostatic Compartment" area.
■ Perform a thermostatic compartment temperature check after the adjustment as described at the beginning
of this procedure.
■ If required, change the "Temperature recorded" value and check the temperature again until you obtain an
acceptable temperature.
Technical Manual
RAS333E - 6 RAA022
Temperature adjustment
This control must be made with the laser bench cover closed and under stabilised thermic
conditions (at least 5 minutes after startup).
■ Remove the plate located at the back of the instrument to access the CTN sensor.
■ Press "Check Thermic" (in the "Optical Bench" area of the window) to control the temperature.
Technical Manual
RAA022 RAS333E - 7
Temperature adjustment
Technical Manual
RAS333E - 8 RAA022
RAS334E
Vacuum adjustment
■ Concerns
■ Required tools
■ Hexagonal keys
■ Barflex
■ Flat screwdrivers
■ Required products
■ None
■ Intervention time
■ 15 minutes
■ Frequency
■ On request
-50
-100
Vacuum (mbar)
-150 Draining
-200 Syringe 1
-250
-300
-350
Altitude (m)
1000 2000 3000
0
-50
-100
Vacuum (mbar)
-150 Draining
-200 Syringe 2
-250
-300
-350
Altitude (m)
■ The number of steps to adjust the vacuum to -200 mbar +/-5 mbar on the counting syringe will be different,
as shown on the following curve:
1000 2000 3000
0
Number of steps for -200 mbar +/-5
250
500
750 Counting
1000 Syringe
1250
1500
1750
Altitude (m)
Technical Manual
RAS334E - 2 RAA022
Vacuum adjustment
When you reach an altitude of 1000 m or more, the cubitainer of ABX Diluent should be installed at
the same level as the instrument.
■ Remove the covers (Refer to RAS342D: Front panel & covers dismantling).
Technical Manual
RAA022 RAS334E - 3
Vacuum adjustment
■ Press "OK".
■ Disconnect the barflex and connect the tube back to the draining syringe 1.
If the vacuum is not stable or if it is too low, check the vacuum circuit (tubings, valves, syringe or
O-ring) and repair if necessary.
■ Disconnect the tube on the draining syringe 2 (inlet 1) and replace it by the barflex tube.
Technical Manual
RAS334E - 4 RAA022
Vacuum adjustment
■ Remove the covers (Refer to RAS342D: Front panel & covers dismantling).
■ Press "OK".
■ Disconnect the barflex and connect the tube back to the counting syringe.
Technical Manual
RAA022 RAS334E - 5
Vacuum adjustment
Technical Manual
RAS334E - 6 RAA022
RAS335D
Bubbling adjustment
■ Concerns
■ Bubbling adjustment.
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■ 15 minutes
■ Frequency
■ The bubbling is factory adjusted and normally does not require any further adjustment.
■ If an adjustment is necessary, follow this procedure.
■ Go to "Service > Technician Menu > Gains > Bubbling".
■ Enter the new step value for the bubbling of each chamber, and press "Accept Values".
The RBC bubbling value can be modified for all instruments but it will not have any effect on
instruments equipped with Takasago valves.
Technical Manual
RAS335D - 2 RAA022
RAS336F
■ Concerns
■ Required tools
■ Hexagonal keys
■ Flat screwdriver
■ Required products
■ Latex
■ Fresh blood sample
■ Calibrator
■ Intervention time
■1 hour
■ Frequency
■ On request
This adjustment procedure must be carried out after the LMNE flowcell has been moved or replaced, or when
the LMNE tube has been changed.
This adjustment must be checked in case of quality problems on the LMNE matrix.
For Pentra XLR, refer to RAS735D: RET optical bench procedure.
The lamp voltage corresponds to the voltage on the LMNE amplifier board, between TP55 and TP53.
This voltage is defined by the light received (from the lamp) on the cell of the LMNE amplifier board.
Target value: 6.7 V +/- 0.8 V, adjustable with R414 on the LMNE amplifier board.
This volumetric adjustment defines the size of the cells counted through the micro aperture of the LMNE
flowcell.
Target value: 53 +/-2, adjustable with R136, located on the mother board.
Technical Manual
RAS336F - 2 RAA022
LMNE flowcell adjustment
LMNE ABS
LMNE amplifier
board cell
Duration between the resistive detection of the cell in the micro aperture (1) and the absorbance detection of
the cell in the optical flow (2). The duration is defined by the height of the optical flowcell, adjustable with
gauges.
Gauges
2. Requirements
Before any adjustment, ensure that:
- The pneumatic circuit functions correctly.
- The transfer time is correct.
- The RA# on normal human fresh blood is correct (Refer to Section 4, 3. Raw counts, page 4-24). A correct
RA# is around 90% of the number of WBC counted in the WBC chamber.
Technical Manual
RAA022 RAS336F - 3
LMNE flowcell adjustment
Never dismantle the emission gun or even unlock it. The adjustment is factory made and cannot be
performed on the field.
Technical Manual
RAS336F - 4 RAA022
LMNE flowcell adjustment
■ Check that the displayed values are within the range in the following table.
■ Ifthe values are out of range or if you want a better adjustment, follow the LMNE flowcell position adjustment
procedure.
Technical Manual
RAA022 RAS336F - 5
LMNE flowcell adjustment
Always perform this adjustment in a counterclockwise motion. If you have not reached a correct
adjustment, turn the screw clockwise and start turning counterclockwise again to obtain the best
adjustment.
3 mm
■ If
necessary, adjust this position by turning the
axial toothed wheel (B). B
Always perform this adjustment in a counterclockwise motion. If you have not reached a correct
adjustment, turn the screw clockwise and start turning counterclockwise again to obtain the best
adjustment.
Technical Manual
RAS336F - 6 RAA022
LMNE flowcell adjustment
J34 J23
R148
TP45
Optical Gain
RBC/PLT
LMNE
■ If
necessary, adjust the Resistive gain (LMNE
Resistive Gain
RES) to 53 +/-2 by turning R136 located on the PLT RBC
Resistive Threshold
TP39
Otpical Threshold
mother board. TP19
TP4 TP2
R159 TP13 TP12
Threshold
R160
R136
R161
TP3 TP56 Threshold
R158
TP24 TP17 TP33 TP16
TP31
TP18 TP36 TP54 R133 TP27
R135
TP44 TP23 TP22 Gain Gain
A
■ First, turn the lateral screw (A) to obtain a stable
transfer on the LMNE adjustment screen.
■ Turn
the axial toothed wheel (B) to obtain the
maximal absorbance. B
■ Adjust the lateral screw (A) again, then the axial
toothed wheel (B) to get a better absorbance.
■ If
necessary, press "Run LMNE Adjustment"
again.
■ Run several cycles on normal human fresh blood and check if the CO% (Refer to Section 4, 3. Raw counts,
page 4-24) can be increased by slightly turning the axial toothed wheel (B).
Technical Manual
RAA022 RAS336F - 7
LMNE flowcell adjustment
5. Final adjustment
Check and adjust the matrix position as follows:
■ Select the raw counts:
- ABX Pentra 80: Go to "Settings > System > Printer" and select "raw".
- ABX Pentra XL80: Go to "Settings > System > Printer" and select "raw", and then press "Print/Send" and
select "Print the run and raws in full page for selected rows".
■ Run 5 different normal human fresh blood samples and print the results.
■ Check the raw count on the printout (Refer to Section 03, 3.Raw counts):
- RA #: Number of cells counted. The RA# allows you to check the integrity of the pneumatical circuit from
the sampling needle to the LMNE flowcell. If the RA# is lower than the expected value (around 90% of the
WBC number), check the pneumatic circuit.
- CO%: Percentage of cells crossing the flowcell in the correct time (transfer time) between the aperture and
the optical measurement. A correct CO% should be > 95%.
- DIFF#: Number of impacts in LYM + MON + NEU + EOS population areas
■ Check the position of the populations on the matrix: the populations must be correctly placed in
their respective zone (box), without flags.
■ Check that the ALY and LIC values are as low as possible (average maximum 1.3%).
■ Check that no alarms are triggered.
Optical axis
Resistivity axis
1- Resistive gain adjustment:
- The resistive gain is set by R136 on the mother board.
- According to the resistive gain adjustment, populations of cells will move along the resistive axis in different
proportions according to the population:
- If the resistive gain is increased, cell populations will move to the right of the matrix.
- If the resistive gain is decreased, cell populations will move to the left of the matrix.
A good adjustment is obtained when the Lymphocyte population is correctly centered in the LYM box.
Technical Manual
RAS336F - 8 RAA022
LMNE flowcell adjustment
A good adjustment is obtained when the Lymphocyte population is close to the NL threshold.
■ Run the 5 normal human blood samples again to check the correct adjustment of the matrix.
Before adjusting the height of the flowcell, make sure the other parameters are correct (lamp voltage,
resistive gain, LMNE Abs. gain).
Technical Manual
RAA022 RAS336F - 9
LMNE flowcell adjustment
Technical Manual
RAS336F - 10 RAA022
RAS337D
■ Concerns
■ Required tools
■ None
■ Required products
■ Fresh and normal blood samples (5 different
samples at least)
■ Intervention time
■ 15 minutes
■ Frequency
■ On request
■ Prepare a minimum of five different human blood samples without WBC alarm.
■ When the analysis cycle ends, the first results are displayed on the result chart table.
■ When the tube holder opens, run another specimen. Repeat this action until you obtain a minimum of 5
different results.
Technical Manual
RAS337D - 2 RAA022
LMNE balance adjustment
■ Selectresults that you want to include in the statistical calculation by selecting the check boxes in the "Sel"
column.
■ To discard a result from the statistical calculation, deselect the check box.
Technical Manual
RAA022 RAS337D - 3
LMNE balance adjustment
Technical Manual
RAS337D - 4 RAA022
RAS338D
■ Concerns
■ Required tools
■ Hexagonal keys
■ Flat screwdriver
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
■ If necessary, adjust the position of the tube holder assembly by mean of the 2 screws to get a correct
adjustment.
Technical Manual
RAS338D - 2 RAA022
Tube holder adjustment
Technical Manual
RAA022 RAS338D - 3
Tube holder adjustment
■ If
necessary, adjust the tube holder locker to
obtain a correct play.
Without this play, the tube holder may not open correctly because of the tube cap.
Technical Manual
RAS338D - 4 RAA022
Tube holder adjustment
3. Needle adjustment
There is a high risk of damaging the needle if the teflon part of the needle comes into contact with
the tube holder.
When the adjustment of the tube holder is completed, adjust the "Tube Holder Position" to obtain a correct
needle sampling position.
■ Go to "Services > Technician Menu > Gains > M.D.S.S. Adjustment > Holder Adjustment" and press
"Control".
The needle moves down to the level of a tube position.
■ Gently press and turn the tube holder.
■ Check that the needle is as close as possible to the tube holder surface, without any actual contact.
■ Turn the tube holder to its initial position.
- If the adjustment is correct, press "OK". The needle moves up.
- If not, press "OK". The needle moves up and you can modify the number of steps in "Tube Holder Position".
Increase the number of steps if the needle is above the tube holder, decrease the number of steps if it is too
low inside the tube holder.
■ Press "Accept Values" and press "Control" again.
■ If
not, adjust the spring’s tightness using the
CHC screw to obtain the correct value, then
repeat the same steps as before.
Technical Manual
RAA022 RAS338D - 5
Tube holder adjustment
5. Adjustment check
■ Once all the adjustments have been carried out, make sure all the screws have been tightened properly.
■ Run a few samples in STAT mode to check that the tube holder operates correctly.
Technical Manual
RAS338D - 6 RAA022
RAS339D
■ Concerns
■ Required tools
■ Hexagonal keys
■ Flat screwdriver
■ Required products
■ None
■ Intervention time
■ 30 minutes
■ Frequency
■ On request
Technical Manual
RAS339D - 2 RAA022
Power supply replacement
J16 J17
J30
R151
DS5 TP52
TP47 DS7
J29
R177 R171 R78
DS6 TP48
U70
J18 J14
J41
U74 DS8
TP68 TP46
DS9
J28
J37
DS10
TP67
J4
TP42 TP40 TP41 TP43
J27
TP50
TP51 J5
J33
DS1
DS15 DS16 J45
J31
Technical Manual
RAA022 RAS339D - 3
Power supply replacement
Technical Manual
RAS339D - 4 RAA022
RAS340D
■ Concerns
■ Required tools
■ None
■ Required products
■ Freshand normal blood samples
■ Calibrationblood samples
■ Control blood
■ Intervention time
■ 45 minutes
■ Frequency
■ On request
1. Repeatability
The measurement of repeatability is based on a set of 10 results obtained from the consecutive analyses of the
same blood sample.
This procedure can be performed either on CBC or DIFF analyses, with a limitation of 35 results per test.
Beyond the 35th result, data generated from the latest analysis will be disregarded.
Note that some results may be automatically rejected. In that case, a dialog box informs the user.
■ From "Quality Assurance", press "Within Run" and run several samples in DIFF mode. Check that the
obtained coefficients of variation are within the following limits.
2. Calibration
Calibration is used to determine the values of the calibration coefficients in order to calibrate the instrument
with known result samples.
■ From "Quality Assurance" press "Calibration".
■ Select the calibrator lot number you are going to use for calibration in the left area of the window.
■ If the calibrator is not listed, create it.
■ Press "Target" and check that the target values are correct. If not, press "Edit", scan your current calibrator
barcode with the barcode reader (or manually enter the target values supplied within the calibrator leaflet),
then confirm by pressing "OK". Exit the "Target Values" by pressing "Return".
■ Prepare the calibrator according to the specific instructions detailed in the calibrator package insert
(temperature, mixing, etc.).
■ Press "STAT Mode", open the calibrator vial and place it in the tube holder.
■ Close the tube holder door.
■ When the tube holder opens, remove the vial and recap it.
Risk of erroneous results if the calibrator is not continuously mixed between each analysis.
In order to obtain a correct calibration, it is recommended to sample the calibrator at least 5 times.
Technical Manual
RAS340D - 2 RAA022
Check up after intervention
■ When the analysis cycle ends, the results are displayed in the result grid. Run the calibrator at least 4 more
times.
In the event that the results show an analysis fault or a reject, the result is not stored. An error
message advising to reject the result is diplayed.
■ To discard a result from the statistical calculation, deselect the check box in the "Sel" column to remove the
result from the statistical calculation.
■ If the coefficient of variation is within the limits, and if the percentage difference between the target and the
mean value is less than 20, the instrument allows an automatic calibration. Press "Automatic Calibration" to
calculate the new coefficients, and then press "OK" to confirm and apply the new coefficients of calibration.
You may need to force the calibration sometimes in order to enter the new coefficients when
the instrument is very far from its normal range. A warning message will prompt you to
confirm. You must then validate and perform a control of the instrument.
Technical Manual
RAA022 RAS340D - 3
Check up after intervention
3. Control
■ Run several controls to check and validate the calibration. When using ABX Difftrol or ABX Minotrol Retic, use
the floppy disk or a USB key to update the lot parameters.
■ Select the "Reserved" check box to link the lot number to the QC results for further analyses.
It may take a few moments for the USB key to be recognized by the computer.
Technical Manual
RAS340D - 4 RAA022
RAS341E
■ Concerns
Optical bench:
■ Dismantling
■ Replacement
■ Control
■ Required tools
■ Hexagonal keys
■ Dynamometric screwdriver A302, A301, A300
■ Flat screwdriver
■ Torx keys
■ Required products
■ None
■ Intervention time
■ 30 minutes
■ Frequency
■ On request
Technical Manual
RAS341E - 2 RAA022
Optical bench dismantling & replacement
As the electronic devices are very sensitive to electrostatic discharges (ESD), the use of an antistatic
kit is mandatory during boards handling/dismantling.
■ Locate the new optical bench and tighten the silent blocks (about 4 turns).
Do not tighten the silent blocks too much, and make sure that the optical bench cannot be lifted.
Technical Manual
RAA022 RAS341E - 3
Optical bench dismantling & replacement
Be carefull not to damage the flowcell nor the emission gun with the hexagonal key.
Technical Manual
RAS341E - 4 RAA022
Optical bench dismantling & replacement
Technical Manual
RAA022 RAS341E - 5
Optical bench dismantling & replacement
3. Lamp replacement
■ Switch the instrument off and wait for the lamp to cool down for at least 10 minutes.
Do not touch the bulb with your fingers. This would reduce significantly the shelf life of the lamp. In
case of finger contact, clean the bulb with a solution of 90% alcohol and a soft paper.
Holder
Technical Manual
RAS341E - 6 RAA022
Optical bench dismantling & replacement
Before starting the adjustment, make sure there are no air bubbles in the flowcell. If there are, go to
"Service > Super User Menu > Hydraulic System > Clean Cycles" and press "Rinse Cytometer" to clean
the flowcell.
■ Disconnect
the coaxial cable and remove the
LMNE amplifier board.
■ Remove the 2 screws.
Technical Manual
RAA022 RAS341E - 7
Optical bench dismantling & replacement
■ If the adjustment is not correct, perform the Lamp height adjustment procedure.
■ If
the image is too low, decrease the distance
between the bench and the lamp holder.
■ If
the image is too high, increase the distance
between the bench and the lamp holder.
To do so:
■ Check the distance between the bench and the lamp holder.
■ Use the measured distance and your observation to estimate the thickness of the gauges required.
Technical Manual
RAS341E - 8 RAA022
Optical bench dismantling & replacement
■ Perform the Lamp orientation adjustment procedure when you have found the correct gauges.
Technical Manual
RAA022 RAS341E - 9
Optical bench dismantling & replacement
Technical Manual
RAS341E - 10 RAA022
RAS342D
■ Concerns
■ Required tools
■ Hexagonal keys
■ Flat screwdriver
■ Required products
■ None
■ Intervention time
■ 30 minutes
■ Frequency
■ On request
Technical Manual
RAS342D - 2 RAA022
Front panel & covers dismantling
■ Open the reagent cover (lift up, then pull it slowly down).
Technical Manual
RAA022 RAS342D - 3
Front panel & covers dismantling
■ Remove
the left panel by unscrewing the 2
CHC M4X6 at the front and loosening the 2
CHC M4X6 at the rear.
To remove the loading tray, open the 2 slides, then lift the tray from the rear to the front. Take
care not to damage the switches.
■ Remove the 4 reagent bottles (disconnect the tubes from the stoppers).
■ Remove the bottle recuperation tray.
■ Remove
the reagent plate by loosening the 4
CHC M4X6 screws.
Technical Manual
RAS342D - 4 RAA022
Front panel & covers dismantling
■ Remove
the bottle support by loosening the 4
CHC M4X6 screws.
Technical Manual
RAA022 RAS342D - 5
Front panel & covers dismantling
■ Remove the upper cover by unscrewing the 6 CHC M4X6 screws: 2 on the right side, 2 on the left side and 3
at the rear of the instrument.
Technical Manual
RAS342D - 6 RAA022
RAS343D
■ Concerns
■ Required tools
■ Hexagonal keys
■ Flat screwdriver
■ Required products
■ None
■ Intervention time
■ 15 minutes
■ Frequency
■ On request
Tube 2 reading zone Rack label reading zone Tube 1 reading zone
Bargraph zone
The tube reading zone depends on the label width and condition. If there is reflection, the reading
zone may be in several parts.
Technical Manual
RAS343D - 2 RAA022
Internal barcode reader adjustment
To reach the screws, there are two holes in the stripper plate.
The angle between the barcode reader and the label to be read must always be the same. The
barcode reader support plate purpose is only to move to the left or to the right.
Striper plate
Technical Manual
RAA022 RAS343D - 3
Internal barcode reader adjustment
3. Barcode configuration
Several standards of barcode labels with different encoding are available. According to the model of the
barcode used by the laboratory, it is necessary to define the settings of the read characters.
Following are the barcode standards that are the most largely used in laboratories:
- Code 2 of 5 interleaved
- Code 39
- Code 128
- Codabar
Among those typologies, some can be used with or without the check digit. The check digit is an added
encoding digit, located on the right part of the code, and corresponds to an algorithm returned value that
checks the integrity of the barcode reading.
The check digit is optional and is defined by the administrator of the barcode label edition software.
Before configuring the barcode, it is necessary to know the barcode models used in the laboratory and to note
the typology and the check digit exploitation.
Example:
1- Note the label models used in the laboratory as shown:
t
Laboratory XXX
Typology Check digit
Code 2 of 5 interleaved No
Code 128 Yes
Codabar Yes
Technical Manual
RAS343D - 4 RAA022
Internal barcode reader adjustment
- Open tab 2.
- Select "Code type = Code 128".
- Select "Check Digit = Not transmitted".
- Confirm by pressing "Accept values".
- Open tab 3.
- Select "Code type = Codabar".
- Select "Check Digit = Not transmitted".
- Confirm by pressing "Accept values".
■ Code Type:
This parameter is set according to the barcode typology used in the laboratory. If several barcode labels are
used, each one must be defined in one of the tabs available onscreen.
■ Code length:
The recommended value is "01-16".
■ Reading direction:
Cannot be configured. It means that the label is readable whatever its position on the tube (right-to-left or left-
to-right).
■ Check Digit:
This parameter is set according to the barcode typology used in the laboratory.
■ None:
Must be checked if the barcode model is "without Check Digit".
■ Transmit:
Must not be used.
■ Not transmitted:
Must be checked if the barcode model is "with Check Digit".
Technical Manual
RAA022 RAS343D - 5
Internal barcode reader adjustment
■ Transmit Start/Stop:
Must be set on "Disabled".
■ Codabar Start/Stop Char.:
Disabled when "Transmit Start/Stop" is disabled.
3.4. Troubleshooting
2- Labels are read, but an additional digit appears on the right side:
- Label includes "Check Digit", but the option has been defined as "None".
Technical Manual
RAS343D - 6 RAA022
RAS344G
■ Concerns
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■ 15 minutes
■ Frequency
■ On request
■ Check that the barcode reader is working properly by means of the reading test.
1224488 121314151617
reading12244881 reading 121314151617
with C.D.
reading 12345ABCDE
without C.D.
reading 12345ABCDEW
12345abcde QWERTY
12345ABCDE azerty
37859 123456/$
reading 37859 reading 123456/$
Technical Manual
RAS344G - 2 RAA022
External barcode reader configuration
■A beeping sound occurs after each reading. Read all the labels from top to bottom and from left to right.
■ When the last label is read, check the test labels once more.
For the code I2 of 5, it is mandatory to hold the reader in order to read the entire code and avoid a
bad interpretation.
■ If the reading test has failed for the DATALOGIC reader, proceed to the barcode setup as follows:
Enter or Exit Configuration
KBD-AT-ALT
I 2 of 5 Set Length
Technical Manual
RAA022 RAS344G - 3
External barcode reader configuration
Technical Manual
RAS344G - 4 RAA022
External barcode reader configuration
■ Never type data with the external keyboard when reading barcode label with the external barcode
reader. Risk of erroneous data entries!
■ Never use a PS2 keyboard at the same time as the external barcode reader (connected on the USB
port). Risk of erroneous data entries!
■ Never connect external barcode reader on PS2 port.
■ Check that the barcode reader is working properly by means of the reading test.
■ If the reading test has failed for the DATALOGIC reader, proceed to the barcode setup as follows:
USB-KBD-ALT
I 2 of 5 Set Length
Technical Manual
RAA022 RAS344G - 5
External barcode reader configuration
Technical Manual
RAS344G - 6 RAA022
RAS345D
■ Concerns
■ Required tools
■ Hexagonal keys
■ Flat screwdriver
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
■ Refer to RAS516B: Takasago valve assembly dismantling for more information on how to access the chamber.
Technical Manual
RAS345D - 2 RAA022
Heater assy replacement
2. Heater replacement
Install the new heater in the same way, in reverse order.
■ Use sleevings when necessary (for the tubes connected to the heater, for example).
■ Make sure that all the tubes are connected.
■ Enter the new CTN value and adjust the heating coil temperature after replacement.
■ When the new heater is installed, start the instrument and check that there are no leaks.
■ Enter the new CTN value (Follow the RAS333E: Temperature adjustment procedure).
■ Check and adjust the MDSS adjustment (Follow the RAS329D: Chambers adjustment procedure).
■ Complete a rinse cycle then prime all reagents.
■ Perform the RAS340D: Check up after intervention procedure in order to ensure that the instrument works
properly.
Technical Manual
RAA022 RAS345D - 3
Heater assy replacement
Technical Manual
RAS345D - 4 RAA022
RAS346D
Mixer replacement
■ Concerns
■ Required tools
■ Hexagonal keys
■ Flat screwdriver
■ Required products
■ Threadlocking: LAX003A
■ Intervention time
■1 hour
■ Frequency
■ On request
■ Before dismantling, mark with a pen the location of the rail on its support.
Technical Manual
RAS346D - 2 RAA022
Mixer replacement
■ Follow the RAS332D: Motor board adjustment procedure to access the motor board.
■ On the motor board, disconnect the following connectors:
- J19 (µstep motor)
- J22 (tube mixer motor home)
- J23 (grabber status)
- J21 (head locking solenoid)
Technical Manual
RAA022 RAS346D - 3
Mixer replacement
■ If
the instrument serial number is lower than P800154, find the mixer assembly location: mark the contour of
the mixer with a pen.
If the instrument serial number is higher than P800154, 2 squares are fixed on the support plate to keep the
location.
Never unscrew the squares. They are factory adjusted and that is the only way to keep the mixer
assembly position.
■ Unscrew the 3 CHC screws and carefully remove the mixer assembly.
Do not remove the wedge. The thickness of this wedge is factory calculated and associated with
this instrument to keep the correct distance between the mixer and the rack rails, apart from the
mixer assembly.
Technical Manual
RAS346D - 4 RAA022
Mixer replacement
2. Adjustments
If the instrument serial number is higher than P800154, no adjustments are necessary, except
entering the new movement values.
■ Plug the power supply cable then switch the instrument on.
■ Go to "Main Menu > Service > Technician Menu > Gains > Sampler adjustment".
■ In the "Mixer Mechanism Adjustment" window, enter the movement values (Move Forward 1, Open/Close
Grabbers / Move Back Home / Move Forward 2) noticed on the new mixer and previously written down.
The mixer adjustments are factory made and parameters are written on a label on the mixer body.
If the serial number is lower than P800154, check the following grabber positions procedure.
■ Check the transfer mechanism home (Refer to RAS347D: Automatic sampler adjustment - 3.1.Transfer
Home).
■ Check the "First transfer" movement (Refer to RAS347D: Automatic sampler adjustment - 3.4.First Transfer).
Technical Manual
RAA022 RAS346D - 5
Mixer replacement
Grabber Rail
Rack
A
Side view
Upper view
Grabber
9 8 7 6
Technical Manual
RAS346D - 6 RAA022
Mixer replacement
In case of grabber assy replacement, the screw (E) should be tightened using threadlocking
fluid (low strength LAX003A) to prevent mixer dysfunction.
The threadlocking fluid LAX003A is available for order (50 mL).
Procedure
■ Unscrew the 3 CHC screws (A)
■ Remove the plate (B).
■ Loosen the HC screw (E) and remove the grabber assy from the mixer axis.
■ Unscrew the 3 CHC screws (C).
■ Remove the white upper part (D).
■ Unscrew and remove the HC screw (E).
Technical Manual
RAA022 RAS346D - 7
Mixer replacement
■ Put threadlocking fluid on this screw (E) and reinstall it without tightening.
■ Reinstall the white upper part (D) using the 3 CHC screws (C).
■ Reinstall the grabber assy on the mixer axis.
■ Make sure that the grabbers are closed and that the axis is situated in the front of the grabber assy.
Grabber closed
Technical Manual
RAS346D - 8 RAA022
RAS347D
■ Concerns
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
The following adjustments must be checked and performed in the following order.
Make sure not to damage the different switches by manually moving the racks.
The Max Rack Loading value represents the maximum number of steps run by the slides to load a
rack. This number of steps must be superior to the number of steps necessary to load a rack,
without reaching a mechanical stop (noise) if there is no rack (to ensure the correct position of the
rack against the front plate).
This motion is never completed when there is a rack because it is stopped by the two switches.
Technical Manual
RAS347D - 2 RAA022
Automatic sampler adjustment
■ The gap between the slides and the front plate (1) must be smaller than the thickness of the rack (2).
SLIDE
SLIDE SLIDE
SLIDE
RACK
RACK
22 1 1
front plate
■ Check the gap, then press "Validate".
- If the gap (1) is too small (mechanical stop, noise), decrease the "Max Rack Loading" value.
- If the gap (1) is too big, increase the "Max Rack Loading" value.
■ Repeat the procedure until you obtain the correct adjustment.
The Stop Rack Loading value represents the number of steps after the detection of the switches,
to ensure the correct rack position.
■ Put a rack on the loading tray, then press "Check Stop Rack".
■ When the motion has stopped, make sure that there is a bit of play between the slides, the rack and the front
plate.
■ Press the rack on the front plate, then check the play (1) between the rack and the slides. It must be around
0.5 mm.
Technical Manual
RAA022 RAS347D - 3
Automatic sampler adjustment
PRESS
SLIDE
1 SLIDE
RACK
front plate
Make sure that the motor does not loose steps (noise) when the slides touch the rack, because it
can return a little bit if the motion is too big.
- If the play (1) is too big: increase the Stop Rack Loading value.
- If the play (1) is too small: decrease the Stop Rack Loading value.
The Transfer home value represents the offset of the Transfer carriage. Follow the procedure
in this order, because the First Transfer and the Rack Ejection depend on the Transfer Home.
Technical Manual
RAS347D - 4 RAA022
Automatic sampler adjustment
Technical Manual
RAA022 RAS347D - 5
Automatic sampler adjustment
Rack Hole
Carriage
Finger
At the end of the transfer carriage motion, the carriage motor is no longer supplied. Make sure it does
not move when you push the finger up.
■ Adjust the transfer home value to center the carriage finger into the hole of the rack:
- If the finger is on the right, increase the "Transfer Home" value.
- If the finger is on the left, decrease the "Transfer Home" value.
■ Press "Check Transfer Home" to check the position.
■ Repeat the procedure until you obtain the correct adjustment.
The Stop Transfer value represents the number of steps after the Stop Transfer switch detection.
After the transfer motion, you have only four seconds to check the position before the rack ejection.
Technical Manual
RAS347D - 6 RAA022
Automatic sampler adjustment
Stop Transfer
RACK Switch
Ejection Tray
Play(~0.5 mm)
Make sure that the motor does not loose steps (noise) once the transfer motion has stopped.
- If the rack is too much on the right, decrease the "Stop Transfer" value.
- If the rack is too much on the left, increase the "Stop Transfer" value.
The Rack Ejection value is the number of steps run by the transfer carriage to move the
ejection assembly.
Technical Manual
RAA022 RAS347D - 7
Automatic sampler adjustment
■ Put one rack on the loading tray and nine racks in the ejection tray against the ejection stops to make sure
that it can push ten racks.
■ Press "Check Rack Ejection".
■ Once the cycle has ended, the rack must be out of the Ejection stops but the ejection assembly must not be
at the mechanical stop.
EJECTION STOPS
- If the rack is not out of the Ejection stops, increase the "Rack Ejection Value".
- If the rack is too much out of the Ejection stops, or at the mechanichal stop, decrease the "Rack Ejection
Value".
■ Repeat the procedure until you obtain the correct adjustment.
The "First Transfer" value represents the number of steps of the First Carriage motion. Other motions
are always the same, representing the space between two positions of tube on the rack.
Adjust this value to center the rack position under the two tracers (followers).
Technical Manual
RAS347D - 8 RAA022
Automatic sampler adjustment
A B A
Left tracer
Rack
Rack motion
Technical Manual
RAA022 RAS347D - 9
Automatic sampler adjustment
Technical Manual
RAS347D - 10 RAA022
RAS348D
Internal PC replacement
■ Concerns
■ Required tools
■ Hexagonal Keys
■ Required products
■ None
■ Intervention time
■ 10 minutes
■ Frequency
■ On request
1. PC dismantling
The PC hard disk may contains some user specific data (Worklist, Archives, etc.). If the PC is
replaced, this information will not be accessible.
Read the technical note included in the kit before replacing the PC assembly.
■ Switch the instrument off and disconnect the power supply cable.
■ Remove the side and top covers of the instrument (Refer to RAS342D: Front panel & covers dismantling).
Technical Manual
RAS348D - 2 RAA022
Internal PC replacement
2. PC installation
■ Install the PC and tighten the fixation screws.
■ Connect the PC accessories.
■ Connect the LED connector.
■ Connect the power supply cable and start the instrument.
■ Log on and check that the instrument works properly.
Technical Manual
RAA022 RAS348D - 3
Internal PC replacement
Technical Manual
RAS348D - 4 RAA022
RAS349D
■ Concerns
■ Required tools
■ Hexagonal Keys
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
If possible, save and/or print the configuration before following this procedure (Refer to RAS357).
Read the technical note included in the kit before replacing the hard disk.
Technical Manual
RAS349D - 2 RAA022
PC Hard Disk replacement
Technical Manual
RAA022 RAS349D - 3
PC Hard Disk replacement
■ Return
the drive support and unscrew the 4
CHC M3x4 screws to free the hard disk.
Technical Manual
RAS349D - 4 RAA022
PC Hard Disk replacement
■ The hard disk is formated and updated with the latest version of the software.
■ Ifthe instrument mother board EEPROM version is not compatible with the Hard disk software
version, it is necessary to load the new version by using the CD-ROM provided with the Hard disk
(please refer to the corresponding technical note to load the version).
Make sure that the blue band on the FPC cable (Floppy disk reader connector) is in the correct
position.
■ Restart the system and restore the configuration (Refer to RAS357E: Save and restore settings).
Technical Manual
RAA022 RAS349D - 5
PC Hard Disk replacement
Do not remove the Left Angle Plate (A) in order to keep the PC/Cover adjustment.
Technical Manual
RAS349D - 6 RAA022
PC Hard Disk replacement
■ Disconnect the flat cable from the Hard Disk, and then from the CD-ROM Drive.
Technical Manual
RAA022 RAS349D - 7
PC Hard Disk replacement
■ Disconnect the FPC cable from the Floppy disk Reader, and remove the Drives Support with the Hard Disk,
Floppy Disk Reader and CD-ROM Reader.
Blue band of
FPC cable in
this position
■ Turn
the Drives Support over and unscrew the
4 CHC M3x4 screws to free the Hard Disk.
Technical Manual
RAS349D - 8 RAA022
PC Hard Disk replacement
■ the Hard Disk is formated and updated with the latest version of the software.
■ Ifthe instrument mother board EEPROM version is not compatible with the Hard disk software
version, it is necessary to load the new version by using the CD-ROM provided with the Hard disk
(please refer to the corresponding technical note to load the version).
Make sure that the flat cable connector is correctly plugged into the Hard Disk in such a way that 4
pins are free, and that the blue band on the FPC cable (Floppy disk reader connector) is in the
correct position.
If you install a DVD drive (P/n: CBT021A) on CCC005L instead of the CD-ROM drive, make
sure you twist the flat cable so that the flat cable connector is inverted on the hard disk side.
Failing to do so will prevent the DVD drive from being functional.
■ Restart the system and restore the configuration (Refer to RAS357E: Save and restore settings).
Technical Manual
RAA022 RAS349D - 9
PC Hard Disk replacement
Technical Manual
RAS349D - 10 RAA022
RAS351D
■ Concerns
■ Required tools
■ Hexagonal Keys
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
Technical Manual
RAS351D - 2 RAA022
PC floppy disk reader replacement
■ Remove the drives support with the hard disk, the floppy disk reader and the CD-ROM reader.
Technical Manual
RAA022 RAS351D - 3
PC floppy disk reader replacement
■ Change the floppy disk reader (Reference: CBT014A) and reassemble in reverse order.
Make sure that the FPC cable is correctly plugged, with the blue band in the correct position.
No floppy drives are installed on the Pentra XLR. They are no longer installed on the new Pentra 80
range instruments either.
Technical Manual
RAS351D - 4 RAA022
RAS352D
■ Concerns
■ Required tools
■ Hexagonal Keys
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
Technical Manual
RAS352D - 2 RAA022
PC CD-ROM drive replacement
Technical Manual
RAA022 RAS352D - 3
PC CD-ROM drive replacement
■ Unscrew the 2 CHC M2x3 screws with 2.5 AZ washers (A) and remove the upper drives bracket.
■ Unscrew the 2 CHC M3x4 with M3 washers (B) and remove the first lower drives bracket.
■ Unscrew the 2 CHC M2x3 screws with 2.5 AZ washers (C) and remove the second lower drive bracket.
Technical Manual
RAS352D - 4 RAA022
PC CD-ROM drive replacement
■ Change the CD-ROM drive (Reference CBT013A or CBT018A) and reassemble in reverse order.
Make sure that the FPC cable is correctly plugged, with the blue band in the correct position.
Do not tighten the screws too much in order not to damage the CD-ROM drive.
If you install a DVD drive (P/n: CBT021A) on CCC005L instead of the CD-ROM drive, make
sure you twist the flat cable so that the flat cable connector is inverted on the hard disk side.
Failing to do so will prevent the DVD drive from being functional.
Technical Manual
RAA022 RAS352D - 5
PC CD-ROM drive replacement
Technical Manual
RAS352D - 6 RAA022
RAS353D
■ Concerns
■ Required tools
■ Hexagonal Keys
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
■ Disconnect the flat cable from the touch screen interface board.
Technical Manual
RAS353D - 2 RAA022
PC touch screen replacement
On new instruments, a square located at the bottom of the screen allows you to keep the vertical
location of the touch screen.
■ Adjust the touch screen position to have its window centered with regards to the window of the screen.
■ Mount the touch screen using the tape included in the kit (cut 4 equal pieces).
The tape allows you to keep the touch screen in position before re-installing in the PC Cover.
■ Use a flat screwdriver to connect the flat cable on the interface touch screen board.
■ Put the plate back then fasten it using the 6 CHC screws.
■ Install the PC in its cover using the 6 CHC screws. Take care not to move the touch screen.
■ Install the PC on the instrument.
Technical Manual
RAA022 RAS353D - 3
PC touch screen replacement
3. Adjustment
■ Switch the instrument on.
■ Go to Menu > Service > Technician Menu > Others > System tools.
■A white window appears, with a red cross in the left upper corner.
■ Click on this cross, it will move to the next position.
■ Click the red cross for each position then press "Enter". The touch screen is calibrated.
Technical Manual
RAS353D - 4 RAA022
RAS354F
■ Concerns
■ Required tools
■ Hexagonal Keys
■ Voltmeter
■ Required products
■ None
■ Intervention time
■ 30 min
■ Frequency
■ On request
Before changing a mother board, make sure that the analyzer settings are saved on a floppy disk,
or USB media or on the hard disk. If not, refer to RAS357E: Save and restore settings to save and/
or print the analyzer settings.
Read the technical note included in the kit before replacing the mother board.
As the electronic devices are very sensitive to electrostatic discharges (ESD), the use of an antistatic
kit is mandatory during boards handling/dismantling.
Technical Manual
RAS354F - 2 RAA022
Instrument mother board replacement
■ Install the new mother board and tighten the 9 CHC M4x 6 screws.
■ Refer to RAS331E: Mother Board Adjustment to connect the connectors and adjust the mother board.
■ Restore the settings (Refer to RAS357E: Save and restore settings and follow the technical note included in
the kit).
It is very important to restore the settings after replacing the mother board. If you fail to do,
the system will not operate properly.
■ For Pentra XLR only, make sure you also change the LMNE Delay value (set it to "175") and
the LMNE Window value (set it to "150") from Service > Technician Menu > Technical
Measurement > Pulse Adjustment
Technical Manual
RAA022 RAS354F - 3
Instrument mother board replacement
Technical Manual
RAS354F - 4 RAA022
RAS355D
■ Concerns
■ Required tools
■ Hexagonal keys
■ Voltmeter
■ Required products
■ None
■ Intervention time
■ 30 min
■ Frequency
■ On request
As the electronic devices are very sensitive to electrostatic discharges (ESD), the use of an antistatic
kit is mandatory during boards handling/dismantling.
Technical Manual
RAS355D - 2 RAA022
Motor board replacement
Technical Manual
RAA022 RAS355D - 3
Motor board replacement
Technical Manual
RAS355D - 4 RAA022
RAS356F
■ Concerns
■ Required tools
■ Hexagonal keys
■ Setting rack GBL0339
■ Piercing switch adjustment tool: GBL0368
■ 0.6 mm gauge: MAJ004A
■ Required products
■ None
■ Intervention time
■2 hours
■ Frequency
■ On request
1. Sensor check
■ Go to Service > Super User Menu > Mechanical > Check Sensors.
■ Remove the left front cover (See RAS342D: Front panel & covers dismantling) to acces the switches (front
cover status becomes red on previous screen).
Technical Manual
RAS356F - 2 RAA022
Sensors check & adjustment
PUSH
Tube Sensor
■ Push on the sensor lever and check that the "Tube Sensor" status becomes red.
PUSH
Technical Manual
RAA022 RAS356F - 3
Sensors check & adjustment
RACK
RACK
Ejection tray
Technical Manual
RAS356F - 4 RAA022
Sensors check & adjustment
1.4. Waste
Waste Control
■ Remove the waste detector from the waste container, then move the detector from the bottom up and check
that the "Waste Control" status becomes red.
1.5. Covers
Front cover
■ Open the right front cover and check that the "Front Cover" status becomes red then close it and repeat this
operation with the left front cover.
Technical Manual
RAA022 RAS356F - 5
Sensors check & adjustment
Thermic Panel
■ Open the thermic panel and ckeck that the "Thermic Panel" status becomes red.
Laser Interlock
■ Remove the laser cover and check that the "Laser Interlock" status becomes red.
2. Sensors Adjustment
A special tool has been created to allow the adjustment of the different switches.
On this tool, different areas are available.
Technical Manual
RAS356F - 6 RAA022
Sensors check & adjustment
The adjustment of these two sensors is the same. These sensors are delivered on a bracket and are
pre-stressed. This adjustment is factory made. Never remove the switch from its bracket.
■ Loosen the CHC M4x12 screws to adjust the switch assy (switch + bracket + CHC M2x10 screws + flat nut).
■ Use the setting rack (GBL0339) to adjust the switch position with regards to the front plate:
- Place the setting rack in order to have the wheel of the switch in the area "1".
- Move the switch assy to have the wheel of the switch just in touch with the area "1" of the setting rack.
- Tighten the CHC M4x12 screws.
Technical Manual
RAA022 RAS356F - 7
Sensors check & adjustment
Adjust
Check
Check
- Place the Setting rack in order to have the wheel of the switch in the area "3".
- Check that the "Stop rack loading sensor" (left or right, depending of the switch) becomes red.
- Place the Setting rack in order to have the wheel of the switch in the area "2".
- Check that the "Stop rack loading sensor" (left or right, depending of the switch) becomes green.
■ Follow the same procedure for the second "Stop rack loading sensor".
Technical Manual
RAS356F - 8 RAA022
Sensors check & adjustment
Before any adjustment, move the setting rack under the stripper plate and check that it can move
under this plate.
The upper part of the setting rack is divided into 3 zones (A, B and C) witch have different heights.
The purpose of this procedure is to have detection of the "Tube height control" switch on the middle (B) and
higher zones (A), without detection on the lower zone (C) of the setting rack.
■ Move the setting rack in order to have the middle zone (B) under the left wheel.
Left wheel
Stop screw
A
C B
Setting rack
Technical Manual
RAA022 RAS356F - 9
Sensors check & adjustment
During this adjustment, make sure the lever of the left wheel does not touch the stop screw.
■ Move the setting rack to have the lower zone (C) under the left wheel, and check that the "Tube height control
status" stay green. If it is not the case, start the adjustment again.
■ To correctly check this adjustment, remove the setting rack, then move it again under the left wheel, starting
by the lower zone (C) (the status must stay green), then the middle zone (B), where the status must be red, as
for the higher zone (A).
A
C B
Stop screw
■ To adjust the stop screw, remove the setting rack and install a 0.6 mm gauge (MAJ004A) between the lever
and the switch box.
■ Adjust the stop screw until it touches the lever, then block it with the nut.
Lever
0.6 mm gauge
Tube height
Nut
control switch
Stop screw
The object of this adjustment is only to ensure that the lever does not touch the box of the switch.
Technical Manual
RAS356F - 10 RAA022
Sensors check & adjustment
PLAY
A B A
LEFT TRACER
Technical Manual
RAA022 RAS356F - 11
Sensors check & adjustment
The non-detection zone (B) width depends on the tracer and on the switch up and down position.
■ Loosen the two CHC M2x10 screws and adjust the switch position:
- If the non detection zone (B) is too small, pull the switch down.
- If the non detection zone (B) is too large, lift the switch up.
■ Install the setting rack against the ejection tray mechanical stop.
■ Loosen the two CHC M2x10 switch screws and adjust the "Rack End Transfer" switch to find the
commutation point. The "Rack end transfer" status must change from green to red.
Technical Manual
RAS356F - 12 RAA022
Sensors check & adjustment
■ Check that the switch commutes really frankly when removing the rack (The "Rack end transfer" status must
change from red to green).
This adjustment is the same as for the "Stop Rack Loading Left and Right" Sensors. The switch is
prestressed on the bracket. Never remove the switch from its bracket.
Those switches are connected in series. During the adjustment of one cover switch, always keep the
opposite cover closed.
2.5. Waste
Waste Control
■ No adjustment available.
Technical Manual
RAA022 RAS356F - 13
Sensors check & adjustment
2.6. Others
Left switch
Right cam
1 mm
Slide
Mechanical stop
Technical Manual
RAS356F - 14 RAA022
Sensors check & adjustment
Technical Manual
RAA022 RAS356F - 15
Sensors check & adjustment
Technical Manual
RAS356F - 16 RAA022
RAS357E
■ Concerns
■ Required tools
■ Floppy disk or USB drive
■ Required products
■ None
■ Intervention time
■ 30 min
■ Frequency
■ On request
Make sure you save and print the configuration after any installation.
1. Analyzer settings
■ Go to "Menu > Settings > Save and Restore > Configuration".
Technical Manual
RAS357E - 2 RAA022
Save and restore settings
Technical Manual
RAA022 RAS357E - 3
Save and restore settings
■ Press "Validate".
2. Workstation settings
■ Follow the same procedure as for analyzer settings using the buttons from the "Workstation Setting" area
instead.
Technical Manual
RAS357E - 4 RAA022
Save and restore settings
3. Print
■ Go to "Main Menu > Settings".
■ Press "Print/Send", the following popup window will be displayed.
Technical Manual
RAA022 RAS357E - 5
Save and restore settings
Technical Manual
RAS357E - 6 RAA022
RAS514B
■ Concerns
■ Required tools
■ Clamp
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
2. Springs replacement
■ Pull
the white part to close the grabber and to
move it against the front plate.
Technical Manual
RAS514B - 2 RAA022
Mixer assembly spring replacement
■ Reinstall the Mixer assembly by following the "Mixer assembly dismantling procedure" in reverse order.
Technical Manual
RAA022 RAS514B - 3
Mixer assembly spring replacement
Technical Manual
RAS514B - 4 RAA022
RAS516B
■ Concerns
■ Required tools
■ Hexagonal keys
■ Cruciform screwdriver
■ Required products
■ None
■ Intervention time
■ 15 minutes
■ Frequency
■ On request
1. Purpose
On instruments equiped with Takasago valves, it could be more difficult to access the chambers or the heater
assembly.
Follow the first part (Chamber access) of this procedure to remove the Takasago valves for maintenance on
those assemblies.
Follow the second part of this procedure (Valve assembly replacement) only to replace the Takasago valve
assembly or to dismantle a Takasago valve.
The second part of the procedure has to be used only in case of failure of the Takasago valve
assembly.
2. Chamber access
2.1. Dismantling
■ Drain the chambers: Main Menu > Super User Menu > Hydraulic > Drain Chambers.
■ Switch the instrument off and disconnect the power supply cable.
■ Remove the right door and the right side cover (Refer to RAS342D: Front panel & covers dismantling).
Technical Manual
RAS516B - 2 RAA022
Takasago valve assembly dismantling
■ In this configuration, it is possible to put the valve assembly on the ejection tray in order to work on the
chambers.
2.2. Reassembling
■ Connect the short tubing (from the DIL1/HGB 4 inlet) on the manifold.
■ Install the valve plate on the crosspieces using the 2 CHC screws.
Technical Manual
RAA022 RAS516B - 3
Takasago valve assembly dismantling
■ Install
the loop of the BASO3/RBC3 tubing
(yellow on the picture) in the loop of the BASO4
tubing (green on the picture) as shown on the
following picture.
Technical Manual
RAS516B - 4 RAA022
Takasago valve assembly dismantling
■ Reassemble in reverse order. Be careful about the tubing position (See “2.2.Reassembling, page 3”).
Technical Manual
RAA022 RAS516B - 5
Takasago valve assembly dismantling
4. Valve dismantling
■ Cut the tyraps maintening the valve wirings together.
■ Unscrew the 2 CHC screws (A) to remove the valves + manifold from the plate.
■ Unscrew the 2 cruciform screws (B) to remove the valve from the manifold. Be careful to the seal (C).
Technical Manual
RAS516B - 6 RAA022
RAS519B
■ Concerns
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■ 10 minutes
■ Frequency
■ On request
1. Purpose
The purpose of this procedure is the adjustment of the needle heigt in a tube located in a rack.
2. Adjustment
■ Enter: Service > Super User Menu > Mechanical System > Rack adjustment.
■ Select the type of rack you use (A or B, C is reserved for SARSTEDT rack).
■ The position of the needle in the tube can be adjusted by changing the "Depth" value.
- If the needle is too low, decrease the "Depth" value.
- If the needle is too high, increase the "Depth" value.
Technical Manual
RAS519B - 2 RAA022
RAS735D
■ Concerns
■ Required tools
■ Gauge set: GBD880A
■ Flowcell adj. knob: XDA555AS
■ Hexagonal keys
■ 10 mm flat key
■ Small flat screwdriver
■ Flat screwdriver
■ Voltmeter
■ Required products
■ RBC/PLT latex: 1300019538
■ Retic latex: LAD012AS
■ ABX Minotrol Retic levels 1-2-3
■ Fresh blood sample
■ Calibrator
■ Intervention time
■ Optical bench replacement: 1 hour
■ Flowcell replacement: 45 minutes
■ Frequency
■ On request
1. Purpose
2. Maintenance Plan
Technical Manual
RAS735D - 2 RAA022
RET optical bench
3. Disassembly / Reassembly
3.1.1. Disassembly
■ Switch the instrument off.
■ Disconnect the power supply cable.
■ Remove the left-hand side panel and the top cover to access the optical bench (Refer to RAS342D: Front
panel & covers dismantling for more information on how to remove covers).
As the Melles Griot laser is very sensitive to electrostatic discharges, install the protection provided
with the laser bench onto the connector when you disconnect it.
Technical Manual
RAA022 RAS735D - 3
RET optical bench
D
B
■ Disconnect the lamp connector (A).
C
■ Disconnect the connector on J2 (B).
■ Disconnect the flat cable on J3 (C). E
■ Disconnect the switch (D).
■ Disconnect the solenoid shutter (E).
Technical Manual
RAS735D - 4 RAA022
RET optical bench
Technical Manual
RAA022 RAS735D - 5
RET optical bench
3.1.2. Reassembly
■ Reassemble in reverse order.
■ Put the power supply cable back and switch the instrument on.
■ Go to "Service> Super User Menu > Hydraulic System > Prime Cycles" and press "ABX DILUENT" to run
an ABX Diluent prime cycle.
■ Go to "Service > Super User Menu > Hydraulic System > Clean Cycles" and press "Rinse Cytometer" to
clean the flowcell.
■ Perform the LMNE flowcell adjustment check procedure.
■ Perform the RET adjustment check procedure.
3.2.1. Disassembly
As the electronic devices are very sensitive to electrostatic discharges (ESD), the use of an antistatic
kit is mandatory during boards handling/dismantling.
Technical Manual
RAS735D - 6 RAA022
RET optical bench
■ Remove the left-hand side panel and the top cover to access the optical bench (Refer to RAS342D: Front
panel & covers dismantling for more information on how to remove covers).
JP101
3.2.2. Reassembly
■ Reassemble in reverse order.
■ Check the adjustment of the Reception gun.
■ Adjust the lamp voltage on the APD board on VR100.
3.3.1. Disassembly
As the electronic devices are very sensitive to electrostatic discharges (ESD), the use of an antistatic
kit is mandatory during boards handling/dismantling.
Technical Manual
RAA022 RAS735D - 7
RET optical bench
J1
3.3.2. Reassembly
■ Reassemble in reverse order.
3.4. Flowcell
3.4.1. Disassembly
■ Switch the instrument off.
■ Disconnect the power supply cable.
■ Remove the left-hand side panel and the top cover to access the optical bench (Refer to RAS342D: Front
panel & covers dismantling for more information on how to remove covers).
Technical Manual
RAS735D - 8 RAA022
RET optical bench
Be careful not to damage the lens of the optical bench with the screwdriver.
■ Place
some absorbent paper below the "Y"
connector (T1).
■ Disconnect the tubing of the external sheath
from the "Y" connector (T1_1).
Technical Manual
RAA022 RAS735D - 9
RET optical bench
Technical Manual
RAS735D - 10 RAA022
RET optical bench
3.4.2. Reassembly
Technical Manual
RAA022 RAS735D - 11
RET optical bench
Technical Manual
RAS735D - 12 RAA022
RET optical bench
Technical Manual
RAA022 RAS735D - 13
RET optical bench
If you lose the screws, make sure you replace them with screws that are not too long. Using screws
that are too long could damage the flowcell.
■ Put the power supply cable back and switch the instrument on.
■ Go to "Service> Super User Menu > Hydraulic System > Prime Cycles" and press "ABX DILUENT" to run
an ABX Diluent prime cycle.
Technical Manual
RAS735D - 14 RAA022
RET optical bench
■ Go to "Service > Super User Menu > Hydraulic System > Clean Cycles" and press "Rinse Cytometer" to
clean the flowcell.
■ Connect a voltmeter to TP1 (ground) and TP2 on the CIS OD amplifier board.
■ Gently move the flowcell until you get the maximum voltage.
■ Tighten the flowcell CHC M4X16 screws to keep the flowcell in the optimal position.
■ Perform the LMNE adjustment procedure.
The replacement of the flowcell requires that you re-set the displayed value of the laser power. Refer
to the Laser adjustment procedure.
4. Lamp replacement
■ Switch the instrument off and wait for the lamp to cool down for at least 10 minutes.
Do not touch the bulb with your fingers. This would reduce the shelf life of the lamp significantly. In
case of finger contact, clean the bulb with a solution of 90% alcohol and a soft paper.
Holder
Technical Manual
RAA022 RAS735D - 15
RET optical bench
5. Lamp adjustment
Before starting the adjustment, make sure there are no air bubbles in the flowcell. If there are, go to
"Service > Super User Menu > Hydraulic System > Clean Cycles" and press "Rinse Cytometer" to clean
the flowcell.
■ If the adjustment is not correct, perform the Lamp height adjustment procedure.
■ If
the image is too low, decrease the distance
between the bench and the lamp holder.
Technical Manual
RAS735D - 16 RAA022
RET optical bench
■ If
the image is too high, increase the distance
between the bench and the lamp holder.
Technical Manual
RAA022 RAS735D - 17
RET optical bench
To do so:
■ Perform the Lamp orientation adjustment procedure when you have found the correct gauges.
Technical Manual
RAS735D - 18 RAA022
RET optical bench
6. LMNE adjustment
■ The blocking screws (with red varnish on the threading) are factory adjusted and must never be
turned.
■ Only the axes screws (F) and (G) can be adjusted.
■ Connect a voltmeter to TP1 (ground) and TP2 on the CIS OD amplifier board.
Technical Manual
RAA022 RAS735D - 19
RET optical bench
Note that there is a rather big play for each of the reception gun screws. Make sure you take
it into account when you adjust each axis. You can tell you are out of the play when you start
to feel resistance when you turn the adjustment knob.
■ Turn
the lateral screw (G) using the adjustment
knob until the maximum voltage is reached. G
■ Repeat the 2 previous steps once more to make sure you have the maximum voltage on both axes.
Once you are certain you have reached the maximum value on each axis, start turning in the
opposite direction, just enough to go in the play so that you do not damage the blocking
system.
Technical Manual
RAS735D - 20 RAA022
RET optical bench
6.2. Flowcell
Never dismantle the emission gun or even unlock it. The adjustment is factory made and cannot be
performed on the field.
Technical Manual
RAA022 RAS735D - 21
RET optical bench
■ Check that the displayed values are within the range in the following table.
■ If
the values are out of range or if you want a better adjustment, follow the LMNE flowcell position adjustment
procedure.
Technical Manual
RAS735D - 22 RAA022
RET optical bench
■ If
necessary, adjust the Resistive gain (LMNE
RES) to 53 +/-2 by turning R136, located on
the mother board.
Technical Manual
RAA022 RAS735D - 23
RET optical bench
■ Gently
turn the lateral screw (J) using a flat
screwdriver until you obtain a stable transfer on
the LMNE adjustment screen.
■ Gently turn the axial (focus) toothed wheel (K)
until you obtain the maximal absorbance.
■ Adjust the lateral screw (J) again, then the axial
(focus) toothed wheel (K) until you obtain the
best absorbance.
You may need to press "Run LMNE
Adjustment" several times to obtain the right
adjustment. K
J
Always perform this adjustment in a counterclockwise motion. If you have not reached a correct
adjustment, turn the screw clockwise and start turning counterclockwise again to obtain the best
adjustment.
■ If the transfer time remains out of range, perform the Transfer time adjustment procedure.
■ Run this procedure in case the transfer time is out of range (target value after stabilization: between 140 µs
and 160 µs).
■ In case of the replacement of the optical flowcell, a gauge set may be necessary to adjust the transfer time.
This gauge set (GBD880A) is now available as service parts.
Before adjusting the height of the flowcell, make sure the other parameters are correct (lamp voltage,
resistive gain, LMNE Abs. gain).
Technical Manual
RAS735D - 24 RAA022
RET optical bench
■ Check the position of the populations on the matrix: the populations must be correctly placed in
their respective zone (box), without flags.
■ Check that the ALY and LIC values are as low as possible (average maximum 1.3%).
■ Check that no alarm is triggered.
Technical Manual
RAA022 RAS735D - 25
RET optical bench
Optical axis
■ If necessary, adjust the matrix position:
Resistivity axis
A good adjustment is obtained when the Lymphocyte population is correctly centered in the LYM box.
Technical Manual
RAS735D - 26 RAA022
RET optical bench
It is recommended to print the results to check the matrix because it is easier to see on paper if the
matrix is properly adjusted.
K
J
Always perform this adjustment in a counterclockwise motion. If you have not reached a correct
adjustment, turn the screw clockwise and start turning counterclockwise again to obtain the best
adjustment.
Technical Manual
RAA022 RAS735D - 27
RET optical bench
Technical Manual
RAS735D - 28 RAA022
RET optical bench
It is recommended to print the results to check the matrix because it is easier to see on paper if the
matrix is properly adjusted.
K
J
Always perform this adjustment in a counterclockwise motion. If you have not reached a correct
adjustment, turn the screw clockwise and start turning counterclockwise again to obtain the best
adjustment.
Technical Manual
RAA022 RAS735D - 29
RET optical bench
Technical Manual
RAS735D - 30 RAA022
RET optical bench
Technical Manual
RAA022 RAS735D - 31
RET optical bench
7. RET adjustment
1 2
3
4
5
8
6 9
10
Number Description
On the field, you must systematically use the "Holder" option to launch a RET
1
adjustment cycle. The "Chamber" option is used for the factory adjustment.
2 "Ret Adjust" allows you to launch the RET adjustment cycle.
Selecting or deselecting "Cumulative matrix" determines the aspect of the
matrix:
- If you do not select this option, the matrix is updated every second and you
only see the current position of the latex balls on the matrix. The numerical
3 values on the left-hand side of the screen are also updated every second.
- If you select this option, the dots that show the location of the latex balls are
added on top of each other in order to form a cloud of dots. All the dots are
taken into account to compile the numerical values on the left-hand side of the
screen.
Changing the "Optical Gain" value causes the cloud of dots to move upward
(by increasing the value) or downward (by decreasing the value).
4
Note that the Optical Gain can be adjusted from this screen and from
Service > Super User Menu > Others. If you change its value in either screen,
the value in the other screen will be affected.
5 The "Optical CV" should be below 0.4 if the RET adjustment is correct.
The "Correlation" only takes into account the dots that are above the threshold
(channel 25) on the matrix. To get a proper adjustment, the cloud of dots must
be located above the threshold. It may be necessary to temporarily increase the
6 optical gain to get the dots located in the upper part of the matrix.
Technical Manual
RAS735D - 32 RAA022
RET optical bench
Number Description
The dots on the matrix need to be centered on the vertical axis and must be
8 above the threshold. Dots below the threshold will not be taken into account in
the correlation.
If you change the "Threshold" value on this screen, the electronic threshold will
not change. Only the position of the threshold on the matrix and the values in
"Upper" and "Lower" change.
9
The numerical values on the left-hand side of the screen will not be affected.
"Upper" and "Lower" display the average of dots located above and below the
threshold. These fields are only used for the factory adjustment.
The transfer time cannot be adjusted. On latex, it should be approximately
10
10 µs longer than the LMNE transfer time.
Before starting the RET adjustment procedure, you need to adjust the laser.
Technical Manual
RAA022 RAS735D - 33
RET optical bench
The goal of this procedure is to adjust the displayed value of the laser power only. This has no effect
on the actual laser power.
Technical Manual
RAS735D - 34 RAA022
RET optical bench
Make sure the LMNE adjustment is correct before attempting to perform the RET adjustment.
Make sure the laser cover is in place before attempting to perform the RET adjustment check
procedure. If the laser cover is not in place, the laser will not be able to function.
■A good adjustment:
Technical Manual
RAA022 RAS735D - 35
RET optical bench
Adjust the RET resistive gain by mean of the R257 potentiometer in order to get the latex cloud
centered on channel #100 (vertical line).
■ The blocking screws (with red varnish on the threading) are factory adjusted and must never be
turned.
■ The laser axes screws (U) and (V) are factory adjusted and must never be turned.
■ Only the lateral axis (L) can be adjusted.
Technical Manual
RAS735D - 36 RAA022
RET optical bench
Make sure the laser cover is in place before attempting to perform the RET adjustment procedure.
If the laser cover is not in place, the laser will not be able to function.
Technical Manual
RAA022 RAS735D - 37
RET optical bench
Once you are certain you have the best adjustment possible, start turning in the opposite direction,
just enough to go in the play so that you do not damage the blocking system.
■ Go back to the RET adjustment check procedure to make sure the adjustment is correct.
■ To adjust the RET matrix with control blood, proceed as described below:
OK OK
RET MFI RET% Run the 3
Control Control
Control levels of RET
Level 3 Level 1
control
NOK NOK
Main Menu / Service / Main Menu / Service /
Super User Menu / Others NOK
Super User Menu / Others
Adjust Optical Gain Adjust Sigma Factor F1
by steps of 5 by steps of 0.2
The RET matrix adjustment procedure is performed from Main Menu > Service > Super User Menu > Others.
Technical Manual
RAS735D - 38 RAA022
RET optical bench
Make sure the PIC value is between 4 and 16 at the risk of triggering a FIT alarm.
Technical Manual
RAA022 RAS735D - 39
RET optical bench
Technical Manual
RAS735D - 40 RAA022
RAS736C
■ Concerns
■ Required tools
■ Hexagonal keys
■ Required products
■ None
■ Intervention time
■ 30 minutes
■ Frequency
■ On request
As the electronic devices are very sensitive to electrostatic discharges (ESD), the use of an antistatic
kit is mandatory during boards handling/dismantling.
J6 J5
Technical Manual
RAS736C - 2 RAA022
Hub board replacement
Technical Manual
RAA022 RAS736C - 3
Hub board replacement
Technical Manual
RAS736C - 4 RAA022
RAS738E
■ Objectives
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■1 hour
■ Frequency
■ On request
Before setting up a new version, it is recommended to save analyzer settings, workstation settings
and database folder "ASP0X" on a USB flash drive from "Settings > Save and restore >
Configuration", and to print them by pressing "Print/Send". Refer to RAS357 procedure.
1 2
Technical Manual
RAS738E - 2 RAA022
Version installation and update
If you are using a version prior to the 2.3.0 version, you have to perform MDSS adjustment
after software installation.
Windows XP Windows 7
2- After the operating system boot, the 2 - After the system boot, the login screen appears.
screen becomes black, then blue. Press Ctrl + Alt + Suppr and then press Log off.
When the screen is blue, press the Shift key
for several seconds. The Logon
Information system window appears.
In case of incompatible laboratory apparatus message while running the BinLoader process,
please refer to the Software installation troubleshooting chapter at the end of this
document.
Technical Manual
RAA022 RAS738E - 3
Version installation and update
10- Click Next to continue the installation. The following window is displayed:
Technical Manual
RAS738E - 4 RAA022
Version installation and update
12- The installation is now completed. Click Finish to restart the computer.
Do not interrupt the software installation at this step.
13- After computer restart, press Validate if the following message appears: "Analyzer Serial Number
is missing. Only technician user can log in application".
14- Log in to the application using the initial password.
15- When prompted, enter a new password and press Validate.
Note that the default technician password will not be accepted anymore. Please create a new different
password.
16- After application start, press Validate if the following message appears: "Analyzer serial number
was retrieved".
Technical Manual
RAA022 RAS738E - 5
Version installation and update
17- Select the configuration of the instrument (with or without Takasago valves) and validate:
This screen never appears with a Pentra XLR since they are all equipped with Takasago valves.
18- Retart the instrument.
After a version update, the technician must go to Service > Technician Menu > Others > Barcode
Setup and press “Accept Values” to regenerate the barcode reader settings.
Windows XP Windows 7
2- After the operating system boot, the 2 - After the system boot, the login screen
screen becomes black, then blue. appears.
When the screen is blue, press the Shift key Press Ctrl + Alt + Suppr and then press Log off.
for several seconds. The Logon
Information system window appears.
Windows XP Windows 7
6- Go to Start > Settings > Control Panel 6 -Go to Start > Control Panel > Programs and
> Add/Remove Programs. Features.
Technical Manual
RAS738E - 6 RAA022
Version installation and update
Windows XP Windows 7
8- Click Add/Remove. 8- Click Uninstall.
Technical Manual
RAA022 RAS738E - 7
Version installation and update
To properly roll back to previous version, you must have previously saved:
■ The complete ASP03 folder located in C:\Program Files\ABX
■ The analyzer settings (see RAS357 more details).
Windows XP Windows 7
2- After the operating system boot, the 2 - After the system boot, the login screen
screen becomes black, then blue. appears.
When the screen is blue, press the Shift key Press Ctrl + Alt + Suppr and then press Log off.
for several seconds. The Logon
Information system window appears.
Windows XP Windows 7
5- Go to Start > Settings > Control Panel 5- Go to Start > Control Panel > Programs and
> Add/Remove Programs. Features.
Windows XP Windows 7
7- Click Add/Remove. 7- Click Uninstall.
Windows XP Windows 7
14- After the operating system boot, the 14- After the system boot, the login screen
screen becomes black, then blue. appears.
When the screen is blue, press the Shift key Press Ctrl + Alt + Suppr and then press Log off.
for several seconds. The Logon
Information system window appears.
Technical Manual
RAS738E - 8 RAA022
Version installation and update
Technical Manual
RAA022 RAS738E - 9
Version installation and update
2. Post-requisites
2.1. To update the counting window and the transfer time (PXLR only)
After installing the software version, it is very important to update the counting window and
the transfer time for Pentra XLR, as it is different from the other Pentra 80 range instruments.
1- Go to Service > Technician Menu > Technical Measurement > Pulse Adjustment.
2- Set the LMNE Delay to "175" and the LMNE Window to "150".
Even though the LMNE Delay is set to "175", the physical transfer time adjustment is still at
150 µs +/-10.
Technical Manual
RAS738E - 10 RAA022
Version installation and update
■ Go to Main menu > Settings > System > Analyzer Serial Number and check the serial number field.
■ Ifthe serial number is missing, enter it manually strictly in the same format as indicated in the label at the rear
of the instrument.
The serial number field is mandatory to generate an unlocking key for the system in case of
password loss.
Technical Manual
RAA022 RAS738E - 11
Version installation and update
Ensure the language of the application is English before starting this procedure (Settings >
System > Local Settings).
In order to switch the interface language to Japanese, Chinese, Russian* or Polish*, you first need to switch to
the appropriate language of Windows. To do so, please proceed as follow:
(*: not available on ABX Pentra 80)
1- Switch the instrument on.
Windows XP Windows 7
2- After the operating system boot, the 2 - After the system boot, the login screen appears.
screen becomes black, then blue. Press Ctrl + Alt + Suppr and then press Log off.
When the screen is blue, press the Shift key
for several seconds. The Logon
Information system window appears.
Technical Manual
RAS738E - 12 RAA022
Version installation and update
4. Help update
If necessary, the online help can be updated by performing these next steps:
1- Insert the CD-ROM RAXxxxx in the internal drive or the external drive.
2- Go to Settings > Save and Restore Settings.
3- Press "Update help" to access the User Manual screen.
4- Choose the language.
5- Press "Install" to update the online help on the instrument.
6- Remove the CD-ROM from the drive.
Technical Manual
RAA022 RAS738E - 13
Version installation and update
Windows XP Windows 7
5- After the operating system boot, the 5- After the system boot, the login screen appears.
screen becomes black, then blue. Press Ctrl + Alt + Suppr and then press Log off.
When the screen is blue, press the Shift key
for several seconds. The Logon
Information system window appears.
Windows XP Windows 7
5- After the operating system boot, the 5- After the system boot, the login screen appears.
screen becomes black, then blue. Press Ctrl + Alt + Suppr and then press Log off.
When the screen is blue, press the Shift key
for several seconds. The Logon
Information system window appears.
Technical Manual
RAS738E - 14 RAA022
Version installation and update
9- When all blinking at the same time, it means that the embedded software is crashed.
In that case, you have to reset the embedded firmware, depending on the motherboard type:
■ Press the reset button near the J6 connector during one second
■ Plug a jumper as described in the picture below on E1 strap near the J6 connector during one
second and remove it. Do not use metallic part.
Technical Manual
RAA022 RAS738E - 15
Version installation and update
Technical Manual
RAS738E - 16 RAA022
RAS816A
Password Recovery
■ Objectives
■ Required tools
■ None
■ Required products
■ None
■ Intervention time
■ N/A
■ Periodicity
■ On request
1. Subject
To strengthen cybersecurity level and comply with FDA requirements, technician passwords are not hardcoded
anymore in the software application.
It is now mandatory to change the password the first time you log in to the application. The technician must
keep this new password.
The new password policy does not allow to re-use the common default password for default profiles (for
example 1224 for Technician profile).
However, the 2.4.0 application still allows to create any users whatever the profile type is. For more information,
refer to the user manual.
In any case of password loss, the technician has the possibibility to generate another one, as described in the
two procedures bellow:
■ One allowing you to unlock the system in case of password loss
■ One giving a temporary access to end users at a technician level
The password recovery only works with ABX Pentra XL80 and Pentra XLR that have been updated in V2.4.0
and upper versions.
To generate a security code or a temporary code, the instrument serial number must already be entered in the
following software menu: Settings > System > Analyzer serial number.
Serial number must be entered in the same format as written on the label at the rear of the instrument.
Once inputted through the software application, the serial number is available:
■ When pressing the unlock button in the login screen
■ At the rear of the instrument
■ On the results printouts footer
■ In the following software screen: Settings > System > Analyzer serial number.
Technical Manual
RAS816A - 2 RAA022
Password recovery
Technical Manual
RAA022 RAS816A - 3
Password recovery
Technical Manual
RAS816A - 4 RAA022
Password recovery
The technician mode disables reagent and technical alarms. At the end of the intervention, log in as user or
super user when technician mode is no more necessary.
Technical Manual
RAA022 RAS816A - 5
Password recovery
Technical Manual
RAS816A - 6 RAA022
Exploded Views
8 - Exploded Views
1. List ................................................................................................................................... 2
2. Valves .............................................................................................................................. 9
3. Carriage assy ................................................................................................................ 10
4. Rinse block & needles ................................................................................................... 11
5. Reagent syringe ............................................................................................................ 12
6. DIFF syringe .................................................................................................................. 13
7. Sampling syringe ........................................................................................................... 14
8. Waste syringe (1 & 2) .................................................................................................... 15
9. Counting syringe assy ................................................................................................... 16
10. 2.5 mL dispenser assy (PXLR) .................................................................................... 17
11. Diluent tank assy ......................................................................................................... 17
12. Optical bench for P80 / PXL80 ................................................................................... 18
13. Optical bench for Pentra XLR ..................................................................................... 19
14. Sampling percutor ....................................................................................................... 20
15. Chambers .................................................................................................................... 21
16. Chamber parts ............................................................................................................ 22
17. Thermostated room .................................................................................................... 23
18. Tube holder assy ......................................................................................................... 24
19. Mixer assy ................................................................................................................... 25
20. Barcode readers ......................................................................................................... 26
21. Loader assy ................................................................................................................. 27
22. Transfer assy ............................................................................................................... 28
23. Ejection assy ............................................................................................................... 29
24. Front covers magnets & switches ............................................................................... 30
25. Main fan ...................................................................................................................... 30
26. Electronic boards ........................................................................................................ 31
27. Cups ............................................................................................................................ 32
28. Internal computer ........................................................................................................ 33
29. Reagent straws and tubings ....................................................................................... 34
30. Stickers ....................................................................................................................... 35
31. Covers ......................................................................................................................... 36
32. References .................................................................................................................. 37
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8-1
Exploded Views
List
1. List
N° in
Drawing Reference Designation
drawing
“2.Valves, page 9” 1 XDB045A VALVE, LIQUID 12 VALVES ASSY (1-12)
“2.Valves, page 9” 2 XDB041A VALVE, LIQ. 2 WAYS/NC W/O COIL
“2.Valves, page 9” 3 XDB042A VALVE, LIQ. 3 WAYS W/O COIL
“2.Valves, page 9” 4 XDB046A VALVE, LIQUID 7 VALVES ASSY (13-19)
“2.Valves, page 9” 5 XDB047A VALVE, LIQUID 4 VALVES ASSY (20-23)
“2.Valves, page 9” 6 XDB048A VALVE, LIQUID 7 VALVES ASSY (24-30)
“2.Valves, page 9” 7 XDB049A VALVE, LIQUID 5 VALVES ASSY (31-35)
“2.Valves, page 9” 8 XBA685A TAKASAGO EQUIPED VALVE ASSEMBLY
TAKASAGO WIRING KIT (DAD136A +
“2.Valves, page 9” 9 XEA959AS
XBA645A)
“2.Valves, page 9” 10 EZZ003A TAKASAGO VALVE MANIFOLD
TAKASAGO SEALS AND SCREWS
“2.Valves, page 9” 11 XEA960AS
ASSEMBLY
“3.Carriage assy, page 10” 1 XDA995CS KIT, CARRIAGE ASSY P80
“3.Carriage assy, page 10” 2 FBR011A BELT, NEEDLE BELT L=364 P60/P80
“3.Carriage assy, page 10” 3 FBR020A BELT, CARRIAGE L=660 P80
“3.Carriage assy, page 10” 4 XBA396A SENSOR, SAMP NEEDLE MOVE P60/80
“3.Carriage assy, page 10” 5 XBA474A MOTOR, RACK TRF / NEEDLE P80
“3.Carriage assy, page 10” 6 GBG082A NEEDLE, BLOCK SHAFT
“3.Carriage assy, page 10” 7 XDA930A CARRIAGE, PIERCING ASSY
“3.Carriage assy, page 10” 8 GBG172A CARRIAGE, SLIDING PLATE
“3.Carriage assy, page 10” 9 GBG176AS CARRIAGE, NEEDLE STOP P60C+/P80
“3.Carriage assy, page 10” 10 DBE014A CABLE, BUSHING D=3.2 BLACK
“3.Carriage assy, page 10” 11 GBG093A MOTOR, PULLEY P60
“3.Carriage assy, page 10” 12 XBA478B MOTOR, CARRIAGE P80
“3.Carriage assy, page 10” 13 XBA342A SENSOR, CARRIAGE MOVE P80
“3.Carriage assy, page 10” 14 XCA191A CHAMBER, ISOLATOR VENT. P80
“4.Rinse block & needles, page 11” 1 XEA810AS KIT, NEEDLE RINSE BLOCK PC+/P80
“4.Rinse block & needles, page 11” 2 FAA055A O-RING, SAMPL. NEEDLE
“4.Rinse block & needles, page 11” 3 FAA013A O-RING, NEEDLE RINSE BLOCK C+, P80
“4.Rinse block & needles, page 11” 4 FAA057A O-RING, PIERCING NEEDLE P120/C+/P80
“4.Rinse block & needles, page 11” 5 GBL0397 NEEDLE, PIERCING P80
“4.Rinse block & needles, page 11” 6 XDA655ES NEEDLE, SAMPLING NEEDLE
“5.Reagent syringe, page 12” 1 XDB001BS SYRINGE, REAGENT ASSY P60/P80
“5.Reagent syringe, page 12” 2 GBG033A SYRINGE, REAG BLOCK BODY P60/P80
“5.Reagent syringe, page 12” 3 XDA622A O-RING, REAGENT SYRINGE + WASHER
“5.Reagent syringe, page 12” 4 GBC030A SYRINGE, REAGENT PISTON P60/P80
“5.Reagent syringe, page 12” 5 GBC031A SYRINGE, LYSE PISTON MIC/P60/P80
“5.Reagent syringe, page 12” 6 FAA073A O-RING, REAGENT SYRINGE D=6.3+
“5.Reagent syringe, page 12” 7 XDB003D SYRINGE, REAGENT+MOTOR ASSY P80
“5.Reagent syringe, page 12” 8 XDB002BS MOTOR, BLOCK FOR REAG. SYR. P80
“5.Reagent syringe, page 12” 9 XBA485A MOTOR, SYRINGE EQUIPPED P80
Technical Manual
8-2 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
List
N° in
Drawing Reference Designation
drawing
“5.Reagent syringe, page 12” 10 GBG034A SYRINGE, 5DIFF BLOCK BOTTOM P60
“5.Reagent syringe, page 12” 11 GBG035B SYRINGE, REAGENT HOLDER PLATE
“6.DIFF syringe, page 13” 1 GBG037A SYRINGE, 5DIFF BLOC BODY P60/80
“6.DIFF syringe, page 13” 2 FAA040A O'RING, 5DIFF SYRINGE D=12.1
“6.DIFF syringe, page 13” 3 FAA067A O'RING, 5DIFF SYRINGE D=2.4
“6.DIFF syringe, page 13” 4 GBG042A SYRINGE, CROSSPIECE P60/SPS/P80
“6.DIFF syringe, page 13” 5 GBG040A SYRINGE, 5DIFF PISTON P60/P80
“6.DIFF syringe, page 13” 6 XDA616AS NEEDLE, 190 µL SYR. P60/SPS/P80
“6.DIFF syringe, page 13” 7 XDA591AS SYRINGE, 5 DIFF SYRINGE ASSY P60/P80
“6.DIFF syringe, page 13” 8 XDB033AS SYRINGE, 5DIFF BLOCK+MOTOR ASSY
“6.DIFF syringe, page 13” 9 XDA826AS MOTOR, BLOCK 5DIFF SYR. P60/P80
“6.DIFF syringe, page 13” 10 XBA548AS MOTOR, FOR 5 DIFF SYRINGE
“6.DIFF syringe, page 13” 11 GBG038A SYRINGE, 5DIFF BOTTOM PLATE
“6.DIFF syringe, page 13” 12 XBA460A CABLE, SYR. SWITCH L=500 P60/P80
“7.Sampling syringe, page 14” 1 GBG044A SYRINGE, SAMP. SYR. BODY P60/P80
“7.Sampling syringe, page 14” 2 FAA064A O-RING, SAMPLING SYR. P60/P80
“7.Sampling syringe, page 14” 3 GBG048A SYRINGE, SAMPLING CROSSPIECE
“7.Sampling syringe, page 14” 4 XDA617AS NEEDLE, 100 µL SAMPL. SYR. P60/P80
“7.Sampling syringe, page 14” 5 XDA593AS SYRINGE, SAMPLING SYRINGE ASSY
“7.Sampling syringe, page 14” 6 XDA834AS SYRINGE, SAMPL BLOCK+MOTOR ASSY
“7.Sampling syringe, page 14” 7 XDA824AS MOTOR, BLOCK SAMPL. SYR. P60/P80
“7.Sampling syringe, page 14” 8 XBA547AS MOTOR, FOR SAMPLING SYRINGE
“7.Sampling syringe, page 14” 9 GBG045A SYRINGE, SAMPLING BOTTOM PLATE
“8.Waste syringe (1 & 2), page 15” 1 XDA827BS SYRINGE, WASTE CPTE P60/C+/P80
“8.Waste syringe (1 & 2), page 15” 2 CAE010A SWITCH, MICROSWITCH XCC5-81
“8.Waste syringe (1 & 2), page 15” 3 FAA017A O'RING, TANK MIN/AG+WASTE MIC
“8.Waste syringe (1 & 2), page 15” 4 GBG219A O'RING, ANTI EXTRUSION RING
“8.Waste syringe (1 & 2), page 15” 5 GBG272A SYRINGE, VACCUM PUMPBODY P60/80
“8.Waste syringe (1 & 2), page 15” 6 FAL009A SILENT BLOCK, FOR SUB ASSEMBLIES
“8.Waste syringe (1 & 2), page 15” 7 XBA458A CABLE, SYR. SWITCH L=300 P60/P80
“8.Waste syringe (1 & 2), page 15” 8 XBA462A CABLE, MOTORS L=300 P60/P80
“8.Waste syringe (1 & 2), page 15” 9 XDA621A O-RING, VAC/WASTE PUMP + WASHER
“9.Counting syringe assy, page 16” 1 XDA828BS SYRINGE, VACUUM CPTE P60/C+/P80
“9.Counting syringe assy, page 16” 2 CAE010A SWITCH, MICROSWITCH XCC5-81
“9.Counting syringe assy, page 16” 3 FAA017A O'RING, TANK MIN/AG+WASTE MIC
“9.Counting syringe assy, page 16” 4 GBG219A O'RING, ANTI EXTRUSION RING
“9.Counting syringe assy, page 16” 5 GBG273A SYRINGE, VACCUM PUMPBODY P60/80
“9.Counting syringe assy, page 16” 6 FAL009A SILENT BLOCK, FOR SUB ASSEMBLIES
“9.Counting syringe assy, page 16” 7 XBA458A CABLE, SYR. SWITCH L=300 P60/P80
“9.Counting syringe assy, page 16” 8 XBA462A CABLE, MOTORS L=300 P60/P80
“9.Counting syringe assy, page 16” 9 XDA621A O-RING, VAC/WASTE PUMP + WASHER
“10.2.5 mL dispenser assy (PXLR), page 17” 2 XDA553B 2.5 ML DISPENSER
“11.Diluent tank assy, page 17” 1 XDA626AS CHAMBER, DIL. TANK COVER P60/P80
“11.Diluent tank assy, page 17” 2 GAL094A CHAMBER, WAST. P120/DIL TANK P60
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8-3
Exploded Views
List
N° in
Drawing Reference Designation
drawing
“11.Diluent tank assy, page 17” 3 XEA286AS KIT, O'RING + WASHER P60/P80/P120
“11.Diluent tank assy, page 17” 4 XDA605A CHAMBER, DIL. TANK ASSY P60/P80
“12.Optical bench for P80 / PXL80, page 18” 1 DAJ007A OPTICAL, LAMP FOR BENCH P60/P80
“12.Optical bench for P80 / PXL80, page 18” 2 FAL006A SILENT BLOC, COMPR. VEP
“12.Optical bench for P80 / PXL80, page 18” 3 1300012654 CHAMBER, LMNE FLOWCELL P60/P80
“12.Optical bench for P80 / PXL80, page 18” 4 XBA399C FITTING, GROUND FLOWCELL P60/80
“12.Optical bench for P80 / PXL80, page 18” 5 XBA490A CABLE, LMNE FLOWCEL CATHOD WIRE
“12.Optical bench for P80 / PXL80, page 18” 6 XBA403A TUBING, SHIELD REPARTITOR P60
“12.Optical bench for P80 / PXL80, page 18” 7 XEA616BS KIT, FLOWCELL INJECTOR TUBE
“12.Optical bench for P80 / PXL80, page 18” 8 XBA193A CABLE, COAX BASO-O.D LMNE-LMNE
“12.Optical bench for P80 / PXL80, page 18” 9 XEA663A CABLE, COAX LMNE FLOWCEL P60/80
“12.Optical bench for P80 / PXL80, page 18” 10 XDA729BS OPTICAL, BENCH COMPLETE P80
“13.Optical bench for Pentra XLR, page 19” 1 XDB287AS RETIC OPTICAl BENCH XLR
“13.Optical bench for Pentra XLR, page 19” 2 1300012656 RETIC FLOWCELL XLR
“13.Optical bench for Pentra XLR, page 19” 3 DAJ007A OPTICAL, LAMP FOR BENCH
“13.Optical bench for Pentra XLR, page 19” 4 XAA695A CIS/OD PREAMPLIFIER BOARD XLR
“13.Optical bench for Pentra XLR, page 19” 5 CBR035A APD PREAMPLIFIER BOARD XLR
“13.Optical bench for Pentra XLR, page 19” 6 XBA844A CABLE, LMNE FLOWCEL CATHOD WIRE
“13.Optical bench for Pentra XLR, page 19” 7 XDB298AS SHUTTER ASSY
“13.Optical bench for Pentra XLR, page 19” 8 FAL016A SILENT BLOCK L17 M4x10 MM 55Sh
“13.Optical bench for Pentra XLR, page 19” 9 CAE034A SWITCH XCG5-J1Z1 (SAIA)
“13.Optical bench for Pentra XLR, page 19” 10 GBD880A CHAMBER, PEELABLE SPACER
“13.Optical bench for Pentra XLR, page 19” 11 GBL0555 LASER PROTECTION PLATE
“13.Optical bench for Pentra XLR, page 19” 12 KZZ054A SKIFFY KNURLED SCREW 140 0410
“13.Optical bench for Pentra XLR, page 19” 13 DBK034A WIRING HOLDER
“13.Optical bench for Pentra XLR, page 19” 14 DAC045A COAX. MMCX RG316 400MM
“13.Optical bench for Pentra XLR, page 19” 15 XBA845A OPTICAL FLOWCELL TUBE SHIELDING
“13.Optical bench for Pentra XLR, page 19” 16 XBA843A CIS OPTICAL COAX FLOWCELL XLR
“14.Sampling percutor, page 20” 1 XDB006AS SAMPLING, NEW PIERCING+CAM
“14.Sampling percutor, page 20” 2 XDB008AS SAMPLING, PIERCING OLD CAM
“14.Sampling percutor, page 20” 3 CAE006A SWITCH, MICROSWITCH XC5-81-S2
“14.Sampling percutor, page 20” 4 DAL016C MOTOR, REDUCTOR 24 V RAP. 40+GEAR
“15.Chambers, page 21” 1 XDB035C 5 CHAMBERS BLOCK ASSY EQUIPED
5 CHAMBERS TAKASAGO BLOCK ASSY
“15.Chambers, page 21” 1* XDB021E
EQUIPED
“15.Chambers, page 21” 2 XDB036BS 4 CHAMBERS BLOCK ASSY EQUIPED
4 CHAMBERS TAKASAGO BLOCK ASSY
“15.Chambers, page 21” 2* XDB017ES
EQUIPED
“15.Chambers, page 21” 3 XDB019BS WBC/BASO CHAMBER EQUIPED
“15.Chambers, page 21” 4 XDB250AS 4 CHAMBERS TAKASAGO BLOCK ASSY
“16.Chamber parts, page 22” 1 XDB035C 5 CHAMBERS BLOCK ASSY
“16.Chamber parts, page 22” 2 GBG325A 4 CHAMBERS BLOCK
“16.Chamber parts, page 22” 3 GBG326A WBC/BASO CHAMBER NAKED
“16.Chamber parts, page 22” 4 GBG275A FLAT JOINT, COUNTING HEAD
“16.Chamber parts, page 22” 5 FAA066A O’ RING, DRAINING CHAMBER
Technical Manual
8-4 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
List
N° in
Drawing Reference Designation
drawing
“16.Chamber parts, page 22” 6 FAK001A APERTURE 50 µ
“16.Chamber parts, page 22” 7 GBG003A DRAINING, 1/4 TURN
“16.Chamber parts, page 22” 8 GBG007A CHAMBER, DRAINING CHICANE
“16.Chamber parts, page 22” 9 GBG157A CHAMBER, COUNTING HEAD
“16.Chamber parts, page 22” 10 XBA398B COAXIAL RBC & WBC PENTRA 60 CE
“16.Chamber parts, page 22” 11 FAK003A APERTURE 80 µ
HGB PHOTOMETER (HGB PHOTOCELL
“16.Chamber parts, page 22” 12 + 12* XBA389B
ALONE: CAX005A)
“16.Chamber parts, page 22” 13 XBA621A SENSOR, CHAMBERS DRAINING P80
“17.Thermostated room, page 23” 1 XCA167A CHAMBER, ISOLATOR
“17.Thermostated room, page 23” 2 GBL0089 CUP, OVERFLOW T° ROOM P80
“17.Thermostated room, page 23” 3 XDB035C CHAMBERS, 5 CHAMBERS BLOCK
“17.Thermostated room, page 23” 4 GBC015A CLIP, MIX CH. HOLDER MIC/P120
“17.Thermostated room, page 23” 5 XDB031CS HEATER, BLOCK COMPLETE P60/P80
“17.Thermostated room, page 23” 6 XCA166A CHAMBER, ISOLATOR (LONG)
“17.Thermostated room, page 23” 7 XDA813B KIT, T° ROOM FAN + HARNESS P80
“17.Thermostated room, page 23” 8 EBB059AS FILTER, RUBBER CAP DEBRIS 25 MM
“18.Tube holder assy, page 24” 1 GBL0183S SAMPLING, STD TUBE HOLDER P80
“18.Tube holder assy, page 24” 1* GBL0372S SAMPLING, OPT TUBE HOLDER P80
“18.Tube holder assy, page 24” 2 XDA728A SAMPLING, PIERCING BLOCK P80
“18.Tube holder assy, page 24” 3 FCB001A SAMPLING, BRAKING GEAR MIC/SPS
“18.Tube holder assy, page 24” 4 DAM009A VALVE, SOLENOID PIERC. BLOCK
“18.Tube holder assy, page 24” 5 CAE006A SWITCH, MICROSWITCH XC5-81-S2
“18.Tube holder assy, page 24” 6 GBL0122 SAMPLING, LOCK PLATE TUBE HOLDER
“18.Tube holder assy, page 24” 7 CAE019A SWITCH, MICROSWITCH XCC5-81-S4
SAMPLING, STD TUBE HOLDER P80
“18.Tube holder assy, page 24” A XDB064A
(RAN273A)
SAMPLING, OPT TUBE HOLDER P80
“18.Tube holder assy, page 24” A* 1300015600
(RAN273A)
“18.Tube holder assy, page 24” B XDB063A SWITCH, EMERGENCY SWITCH ASSEMBLY
“18.Tube holder assy, page 24” C GBL0324 SPRING BLADE
“18.Tube holder assy, page 24” D CAE010A SWITCH, MICROSWITCH XCC5-81
“19.Mixer assy, page 25” 1 XDA972B MOTOR, MIXING COMPLETE P80
“19.Mixer assy, page 25” 2 XDA973AS SAMPLING, GRABBER BLOC ASSY P80
“19.Mixer assy, page 25” 3 DAM009A VALVE, SOLENOID PIERC. BLOCK
“19.Mixer assy, page 25” 4 XBA432A SENSOR, HOME RACK TRF/MOVE GRAB
“20.Barcode readers, page 26” 1 CBC008A PCB, BARCODE INTERNAL P80
“20.Barcode readers, page 26” 2 XBA725A PCB, EXT BARCODE USB PXLR
“20.Barcode readers, page 26” 3 XBA453B PCB, BARCODE EXTERNAL P80/XL80
“21.Loader assy, page 27” 1 XDA792AS SAMPLING, LEFT RACK LOADER P80
“21.Loader assy, page 27” 2 GBL0385 COVER, RACK LOADING TRAY P80
“21.Loader assy, page 27” 3 CAE019A SWITCH, MICROSWITCH XCC5-81-S4
“21.Loader assy, page 27” 4 XDA791AS SAMPLING, RIGHT RACK LOADER P80
“21.Loader assy, page 27” 5 XBA476A MOTOR, RACK LOADING AXIS P80
“21.Loader assy, page 27” 6 XBA475A MOTOR, RACK LOADING P80
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8-5
Exploded Views
List
N° in
Drawing Reference Designation
drawing
“21.Loader assy, page 27” 7 XDA796AS SAMPLING, RACK LOADING LOCK P80
“21.Loader assy, page 27” 8 XDA964A KIT, LOADING FRONT PLATE EQUIPED
“21.Loader assy, page 27” 9 GBL0352 COVER, NEW FRONT PLATE LOADING
XDA974AS SWITCH, LEFT LOADING EQUIP P80 (Switch
“21.Loader assy, page 27” 10
alone: CAE019A)
“21.Loader assy, page 27” 11 XDA793AS BELT, RACK LOADING
“22.Transfer assy, page 28” 1 XDA794A SAMPLING, TUBE DETECTION P80
“22.Transfer assy, page 28” 2 CAE011A SWITCH, MICROSWITCH XCH5-S2
“22.Transfer assy, page 28” 3 CAE010A SWITCH, MICROSWITCH XCC5-81
“22.Transfer assy, page 28” 4 XDA795A SAMPLING, RACK TRACER ASSY P80
“22.Transfer assy, page 28” 5 XDA732A CARRIAGE, SAMPLING ASSY P80
“22.Transfer assy, page 28” 6 XBA432A SENSOR, HOME RACK TRF/MOVE GRAB
“22.Transfer assy, page 28” 7 XBA474A MOTOR, RACK TRF / NEEDLE P80
“22.Transfer assy, page 28” 8 XAA478B PCB, SOLENOID RACK TRANSFER
“23.Ejection assy, page 29” 1 XDA971BS KIT, EJECTION PLATE EQUIPPED
“23.Ejection assy, page 29” 2 CAE011A SWITCH, MICROSWITCH XCH5-S2
“23.Ejection assy, page 29” 3 GBL0062 CARRIAGE, RACK EJEC. PUSHROD P80
XDA975A SWITCH, RIGHT LOADING EQUIP P80
“23.Ejection assy, page 29” 4
(Switch alone: CAE019A)
“23.Ejection assy, page 29” 5 GBL0068 COVER, RACK EJECTION TRAY
“24.Front covers magnets & switches,
1 FAJ008A COVER, CLOSING MAGNET P80
page 30”
“24.Front covers magnets & switches,
2 CAE020A SWITCH, MICROSWITCH XGG2A
page 30”
“25.Main fan, page 30” 1 XBA393A FAN, MAIN FAN 24 V P60/P80
XAA544* PCB, MAIN PCB P80 + VERSION (*: index
according to the version number)
XAA545* PCB, MAIN PCB PXL80 + VERSION (*: index
“26.Electronic boards, page 31” 1
according to the version number)
XAA694* PCB, MAIN PCB XLR + VERSION (*: index
according to the version number)
“26.Electronic boards, page 31” 2 XAA459BS PCB, MOTOR BOARD P80
“26.Electronic boards, page 31” 3 XAA429AS PCB, LED BOARD FOR COVER P60/80
“26.Electronic boards, page 31” 4 XAA458AS PCB, PREAMPLI BOARD for P80/PXL80
“26.Electronic boards, page 31” 5 DZZ018 CABLE, FERRITE FOR COAXIALS
“26.Electronic boards, page 31” 6 XAA695A PCB, CIS OD PREMPLI BOARD for XLR
“26.Electronic boards, page 31” 7 CBR035A PCB, APD PREAMPLIFIER BOARD XLR
“26.Electronic boards, page 31” 8 XAA705A PCB, HUB BOARD XLR
“27.Cups, page 32” 1 GBL0164 CUP, OVERFLOW REAG. BOTTLE P80
“27.Cups, page 32” 2 GBL0215 CUP, OVERFLOW TOP MOTOR PCB P80
“27.Cups, page 32” 3 GBL0332 CUP, OVERFLOW MOTOR BOARD P80
“27.Cups, page 32” 4 GBL0193 CUP, OVERFLOW LIQ. SYRINGES P80
“27.Cups, page 32” 5 GBL0194 DRAIN SYRINGE RECUP. TRAY
“27.Cups, page 32” 6 GBL0213 CUP, OVERFLOW MAIN BOARD
“27.Cups, page 32” 7 FBM003A COVER, REAGENT COVER HOLDER P80
“27.Cups, page 32” 8 GBL0149 CUP, FOR GRABBER
Technical Manual
8-6 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
List
N° in
Drawing Reference Designation
drawing
“27.Cups, page 32” 9 GBL0089 CUP, OVERFLOW T° ROOM P80
“28.Internal computer, page 33” 1 XAA511BS PCB, TOUCH PANEL FUTJI P80/P400
“28.Internal computer, page 33” 2 CBT014A PCB, FLOPPY DISK READER P80
CBT013A or
PCB, CD-ROM READER
“28.Internal computer, page 33” 3 CBT018A
PCB, DVDRW
or CBT021A
“28.Internal computer, page 33” 4 Refer to RAN326 Technical Note
“28.Internal computer, page 33” 5 XBA507AS FAN, FOR COMPUTER P80
“28.Internal computer, page 33” 6 CBR019A PCB, TACTILE BOARD FUTJI PC P80
“28.Internal computer, page 33” 7 ABC010A PCB, BACKLIGHT NEC COMPUTER P80
“28.Internal computer, page 33” 8 Refer to RAN326 Technical Note
“28.Internal computer, page 33” 9 DAK026A FAN, FAN FOR SOCKET 370 PC P80
CCC029A PCB, SODIM 256 MB P80
“28.Internal computer, page 33” 10
CCC033A PCP, DDR2 RAM
“29.Reagent straws and tubings, page 34” 1 GBG145A REAGENT, STRAW D=20 P60/P80
“29.Reagent straws and tubings, page 34” 2 GBG155A REAGENT, BOTTLE STOPPER D=25
“29.Reagent straws and tubings, page 34” 3 GBG245A REAGENT, STRAW D=28 P60/P80/ACT
“29.Reagent straws and tubings, page 34” 4 GAK302A REAGENT, BOTTLE STOPPER D=40
“29.Reagent straws and tubings, page 34” 5 XEA018A REAGENT, STRAW DILUENT L=360
“29.Reagent straws and tubings, page 34” 6 GBC284A REAGENT, STRAW D=28 XLR
EAE057AS
“29.Reagent straws and tubings, page 34” 7 TUBING, PTFE 1,59X0,8 LG=3M
(1202249057)
HAX0246 STICKER, RACK LABEL TYPE "A" 1-20
HAX0247 STICKER, RACK LABEL TYPE "B" 1-20
HAX0248 STICKER, RACK LABEL TYPE "A" 21-40
HAX0249 STICKER, RACK LABEL TYPE "B" 21-40
“30.Stickers, page 35” 1 HAX0250 STICKER, RACK LABEL TYPE "A" 41-99
HAX0251 STICKER, RACK LABEL TYPE "B" 41-99
HAX0252 STICKER, RACK LABEL TYPE "C" 1-20
HAX0253 STICKER, RACK LABEL TYPE "C" 21-40
HAX0254 STICKER, RACK LABEL TYPE "C" 41-99
HAE026B STICKER, VALVES JOUCO (1-10)
HAE027B STICKER, VALVES JOUCO (11-20)
“30.Stickers, page 35” 2
HAE028B STICKER, VALVES JOUCO (21-30)
HAE029B STICKER, VALVES JOUCO (31-40)
“30.Stickers, page 35” 3 HAX034A STICKER, PC CONNECTIONS P80
“30.Stickers, page 35” 4 HAU108A STICKER, LASER
“30.Stickers, page 35” 5 HAX0356 STICKER, PC USB CONNECTION
HAX0236 STICKER, P80/XL80 CHAMBER equipped
“30.Stickers, page 35” 6 with Takasago valves
HAX0357 STICKER, XLR CHAMBER
“30.Stickers, page 35” 7 HAX0355 STICKER, OPTICAL BENCH BLOCKING
XDA803AS COVER, FRONT LEFT COVER P80
“31.Covers, page 36” 1
XDA896AS COVER, FRONT LEFT COVER XL/XLR
XDA804AS COVER, FRONT RIGHT COVER P80
“31.Covers, page 36” 2
XDA897BS COVER, FRONT RIGHT COVER XL/XLR
“31.Covers, page 36” 3 XDA912BS COVER, COMPUTER COVER
“31.Covers, page 36” 4 XDA895AS COVER, RIGHT DOOR EQUIPED 80 / XL
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8-7
Exploded Views
List
N° in
Drawing Reference Designation
drawing
GBL0227 DOOR, MANUAL SAMPLING P80
“31.Covers, page 36” 5
GBL0313 DOOR, MANUAL SAMPLING XL / XLR
HAX0309 STICKER, COVER LABEL "PENTRA XL80
“31.Covers, page 36” 6
HAX0348 STICKER, COVER LABEL "PENTRA XLR"
“31.Covers, page 36” 7 HAX0307 STICKER, COVER LABEL "HORIBA Medical"
“31.Covers, page 36” 8 GBL0316 COVER, REAGENT COVER P80 / XL
“31.Covers, page 36” 9 GBL0318 GLUEING BOARD + RIGHT SIDE P80XL
“31.Covers, page 36” 10 GBL0315 LEFT SIDE COVER P80 XL
“31.Covers, page 36” 11 GBL0320 COVER, LEFT DOOR XLR
“31.Covers, page 36” 12 GBL0317 COVER, UPPER COVER
GBL0659
“31.Covers, page 36” 13 COVER, 1L LYSE REAG TRAY KIT
(1209179289)
Technical Manual
8-8 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Valves
2. Valves
1
4
3
7
5
6
10
11
9
8
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8-9
Exploded Views
Carriage assy
3. Carriage assy
1
8 11
4
2
12
10
14
6
3
7 13
Technical Manual
8 - 10 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Rinse block & needles
2 3 6
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 11
Exploded Views
Reagent syringe
5. Reagent syringe
9 3
x4
7
6
4
x4
8 5
1
10
11
Technical Manual
8 - 12 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
DIFF syringe
6. DIFF syringe
9
7
1
2
8
3 x4
4 x2
12
10
5
11
6 x2
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 13
Exploded Views
Sampling syringe
7. Sampling syringe
2 x2
5
8
4
Technical Manual
8 - 14 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Waste syringe (1 & 2)
Please refer to RAS325C: 6 month maintenance, Draining & counting syringes and to
RAH964 technical note for more information.
2 9
8 1
3
4
7
5
6 x4
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 15
Exploded Views
Counting syringe assy
Please refer to RAS325C: 6 month maintenance, Draining & counting syringes and to
RAH964 technical note for more information.
2 9
3
4
6 x4
Technical Manual
8 - 16 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
2.5 mL dispenser assy (PXLR)
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 17
Exploded Views
Optical bench for P80 / PXL80
7 10
1
6
2 x4
4
8
Technical Manual
8 - 18 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Optical bench for Pentra XLR
12
11
9 2
3
13
10
4
7 6
16
14
15
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 19
Exploded Views
Sampling percutor
2
4
3
1
Technical Manual
8 - 20 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Chambers
15. Chambers
1 1*
2*
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 21
Exploded Views
Chamber parts
10 x2
3
4 x4
12
9 x2
11
2
6
5
x4
8
x4
1 7 x4 12*
13
Technical Manual
8 - 22 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Thermostated room
6 1 x2
4 x3
2
8
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 23
Exploded Views
Tube holder assy
1
1* B
A
C
A*
D x4
7 x4
4
6
5 x2
2
Technical Manual
8 - 24 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Mixer assy
4 1 2
3
Index P/n Designation
1 XDA972B MOTOR, MIXING COMPLETE P80
2 XDA973AS SAMPLING, GRABBER BLOC ASSY P80
3 DAM009A VALVE, SOLENOID PIERC. BLOCK
4 XBA432A SENSOR, HOME RACK TRF/MOVE GRAB
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 25
Exploded Views
Barcode readers
Technical Manual
8 - 26 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Loader assy
8
9
2
1
10 3
11
7
6
5
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 27
Exploded Views
Transfer assy
2 4 x2
3
1
5
8
Technical Manual
8 - 28 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Ejection assy
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 29
Exploded Views
Front covers magnets & switches
1 x4
2 x2
Technical Manual
8 - 30 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Electronic boards
5 7
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 31
Exploded Views
Cups
27. Cups
7
6
5
4
8 9
Technical Manual
8 - 32 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Internal computer
5 1 7
3
4
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 33
Exploded Views
Reagent straws and tubings
6 1
3
4
Technical Manual
8 - 34 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
Stickers
30. Stickers
4
3
5
1 2
7
6
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 35
Exploded Views
Covers
31. Covers
8 11 12 4
9
13
10
or
7 3
2
5
1
Technical Manual
8 - 36 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
32. References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 37
Exploded Views
References
Technical Manual
8 - 38 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 39
Exploded Views
References
Technical Manual
8 - 40 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 41
Exploded Views
References
Technical Manual
8 - 42 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 43
Exploded Views
References
Technical Manual
8 - 44 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 45
Exploded Views
References
Technical Manual
8 - 46 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 47
Exploded Views
References
Technical Manual
8 - 48 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 49
Exploded Views
References
Technical Manual
8 - 50 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 51
Exploded Views
References
Technical Manual
8 - 52 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 53
Exploded Views
References
Technical Manual
8 - 54 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 55
Exploded Views
References
Technical Manual
8 - 56 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN
Exploded Views
References
Technical Manual
Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN 8 - 57
Exploded Views
References
Technical Manual
8 - 58 Ref: 1300018562 - Int. Ref. Doc.: RAA022GEN