Presents:
Electrosurgery
Introduction, application & Hazards
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Electricity can be created by forcing electrons to flow from atom to
atom.
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
DC
(Direct Current)
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
DC
(Direct Current)
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
AC
(Alternative Current)
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
AC
(Alternative Current)
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Voltage : Volts – V
Force pushing current through the resistance,
measured in Volts
Current : Amperes – I
Flow of electrons during a period of time,
measured in amperes
Resistance : Ohm – Ω
Obstacle to the flow of current, measured in ohms
(impedance = resistance)
Power : Watts – P P=I x V
Frequency : Hertz – HZ
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
1 2
5
1
2
4
3
Complete Circuit
3
4 5 Current
2-22
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Resistance in different Tissues
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Resistance in different Tissues
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Resistance in different Tissues
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Resistance in different Tissues
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Resistance in different Tissues
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Flows to ground
Follow path of least resistance .
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Direct Current - Electrocautery
+
Voltage
Reference
-
Time
Reference
-
Time
2-24
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Developed in 1926 by
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
- Electrolyte
- Faradic
- Thermal
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
- DC current –
ionizing
- - AC current -
Vibration
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
- Faradic
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
High Voltage – High Leakage
Return current may exit in ways other than
the patient – may cause a burn .
Safety Standard is :
► 100 KHZ – 150000 Microamperes
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Electrosurgery : 2 Types : Bipolar and
Monopolar
Monopolar -Cut & Coagulate
Bipolar – Coagulate
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Active electrode at
surgical site
Return electrode at
another site
Current flows through
the body between the
electrodes
Voltage ranges:
• Cut: 1350 - 4,000 V
• Blend: 2975 – 4500 V
• Coag: 3000 - 9,000 V
2-25
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Footswitching
Pencil
Rocker Switch
Pencil
Two-Button
Pencil
Three-Button
Pencil
2-26
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Return
Active
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
• Active and return electrodes within the
instrument;
no patient return electrode (pad) required
• Current flow is confined to tissue between
electrodes
• Relatively low voltage (300V – 1200V)
• Used for delicate tissue and confined surgical
sites:
• Laparoscopy
• Spinal
• Neurosurgery
• Ophthalmology
Current between
the tines
2-27
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Tines
2-28
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Pad Site burns are caused by adverse conditions
at the pad- patient interface which result in
increased current density .
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Brûlure sous la plaque
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
ALARME
EP-19
Covidien | | Confidential
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
COURANT HF
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
E7509 HRA5
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Generator
Patient Jack
Active Jack
Circuit
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Interrogation
Current
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
REM
(Return Electrode
Monitoring)
insuffisant contact
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
No Plate Burns
Generator More Efficient
100,000,000 cases without any incidence
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Standard range of safe resistance
70 Ω
5Ω 135 Ω
Decreased resistance 60 Ω 84 Ω
changes the baseline
baseline, so the
Baseline 40% increase
40% upper limit changes as Resistance in resistance
well.
50 Ω 70 Ω
Baseline 40% increase
Resistance in resistance
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
RE Site Burning:
1. Single Plate
2. Quality of Pad
3. NO REM
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Alternate Site Burning:
1. Small Area
2. Sharp Surrounding
4. Happens Immediately
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Other Sores :
1. Bed Sore
2. Chemical Sore
3. Pressure Sore
4. Humidity Sore
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Pure
PureCut
Cut Blend Cut Desiccate Fulgurate
100 % ON 60 % ON – 40 % OFF
Very High Voltage to
Less Volts Higher Voltage push thru Air
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Cut
COAGULATION
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Cells Explosion
ES-07
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Cut
Energy
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Cut
Energy
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Lower current density
causes dehydration and
shriveling of the cells.
Energy Intracellular fluid evaporates
BLADE/ BALL/
FORCEPS
Dehydration thru
heating Coagulated Cells
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Ball Electrode Cells dehydrate
through slower
heating effet
ES-19
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
ES-23
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
With blend, we are raising the ___________
voltage
?
but we are decreasing the ______
duty
? _______
cycle
? .
What is the difference between the Pure Cut and Pure Coag
waveforms?
2-30
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
ES-28
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Fulguration (lightning) is used
to control bleeding. Low (6%)
duty cycle, but high voltage.
The tip of the electrode should
be held slightly over the tissue.
ForceTriad™ energy
platform
Coag Duty Cycles:
Fulgurate – 6.5%
Spray – 4.6%
2-33
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Force FX™-C
electrosurgical generator
Force Argon™
electrosurgical unit
Argon gas
tanks
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
◦ Creates uniform conductive plasma bridge between
electrode and tissue
• Less smoke, odor
• Non contact in coagulation
• Decreases blood loss, re-bleeding
• Less tissue damage
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Gas Coag & Extend/Retract
on/off Cut Shroud
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Cut Blend Coag
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
• Heat is generated locally by a high frequency current that flows from the
electrodes.
• A probe is inserted into the center of the tumor for about 10-15 minutes.
• The local heat melts (coagulates) the tissue that is adjacent to the probe.
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
RF Ablation System
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
MW Ablation System
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Electrosurgery Hazards
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Hazards In Laparotomy surgeries:
1) Direct Coupling
2) Insulation Failure
3) Capacitive Coupling
4) Residual Heat
5) Electromagnetic Interface
6) Electrosurgical Smoke
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
1) The Leakage Current
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
2) The Place of Plate
❖Muscular
❖No Hair
❖No Scar
❖Optimum 50 cm
❖Direction
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
3) Open Circuit Activation
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
4) Using Spray Mode in pick ups
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
5) Electrosurgical Smoke
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Designed to:
• Capture particulates and absorb
smoke and odors
• Improve visibility
• Reduces potential health hazards
Used in both open and laparoscopic
procedures
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
ست الپاروسکوپ
Medtronic (Covidien)
مانیتور مدیکال
کمرا هد و پروسسور
منبع نور
اینسافلیتور
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
1) Direct Coupling
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
2) Insulation Failure
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
3) Capacitive Coupling
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
4) Residual Heat
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
5) Electromagnetic Interference
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
6) Electrosurgical Smoke
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
6) Electrosurgical Smoke
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
6) Electrosurgical Smoke
ULTRAVISION VENTING
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
6) Electrosurgical Smoke
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
New Technologies
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Valleylab History of Innovation
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Force 2 in 1986
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Old Technologies
No Instant Response
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Old Technologies
No Instant Response
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Old Technologies
No Instant Response
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Old Technologies
No Instant Response
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Cut set at 40 Watts
Ohms
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology
(Tissue Response)
Founded in 1995
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology
Presets
Output
VARIATIONS
200 times
PER SEC
REGULATIONS
MEASUREMENT
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology
CONSTANT
Power POWER
IMPEDANCE
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Cut set at 40 Watts
Ohms
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology
Valleylab FX8
434000/s
Force FX-8C
200/s
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
LigaSure Vessel Sealing System
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
LigaSure Vessel Sealing System
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
LigaSure :System Operation
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Formerly Tyco Healthcare Covidien | | Confidential
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Formerly Tyco Healthcare Covidien | | Confidential
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Formerly Tyco Healthcare Covidien | | Confidential
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
LigaSure Vessel Sealing System
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Bipolar Coagulated Vessel Characteristics
Open Loop System – No feedback
Relies on tissue shrinking and proximal Thrombus for Ligation
Lumen Still Apparent
Tissue Damage determined by Power Setting , activation time and
shape of Power Curve
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Instant Response Technology TissueFect Technology
20000X / SEC
200 X / SEC
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
New LigaSure
LS10
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
مشخصات و قابلیتهای
سیستم انسداد عروق کمپانی Covidien
مدل Valleylab LS10
دارای ضریب شکست
دارای کانکتور صفر تا 3برابر فشار
انسداد عروق تا
هوشمند سیستولیک (تا 400
قطر 7میلی
( Smart میلیمتر جیوه)
متر (دارای
Connector
) جهت تاییدیه FDA
شناسایی و آمریکا)
دارای ابعاد اندازه گیری و
تشخیص کوچک و وزن ارزیابی
هندپیسهای 5/5کیلوگرم امپدانس
مختلف جهت حمل و بافت 20000
نقل آسان. قابلیت انسداد عروق بار در ثانیه
در 2تا 4ثانیه
دارای کنترل بصری (باالترین سرعت
و خودآموز و عملکرد کامال هوشمند بنحویکه انسداد عروق نسبت
نمایش اطالعات هیچگونه قابلیت تنطیم وات برای به تکنیکهای دیگر)
ساده برای کاربر روی دستگاه وجود ندارد قابلیت کارکرد
یادگیری ،درک و با هندپیسهای
عیب یابی دستگاه. مدرن کمپانی و
دارای تکنولوژی Tissue Effect Plusکه باعث هوشمندی در هندپیسهای
شناسایی و ارزیابی بافت هدف و تنظبم انرژی بطور خودکار در دستگاه قدیمی
بوده بطوریکه میزان پخش حرارتی ( )Thermal Spreadبه
بافتهای جانبی حداکثر 1میلی متر برای هر مدل هندپیس میباشد.
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
FT10
TissueFect™ Reads tissue: 434,000 times
per second1
Software Upgradeable Ethernet connection (has WiFi
capabilities that are not yet
enabled)
Bipolar Cable Compensation Reads cable length and width
for consistent electrosurgical
output
Touch Screen Single, simplified touch screen
New ES Settings Soft coag
Dimensions • Weight (kg): 10.1
• Width (cm): 35.8
• Height (cm): 17.0
• Depth (cm): 46.2
Performance • Lower max jaw temperature3
• Faster sealing times (1-4
seconds vs. 3-6 seconds with
ForceTriad™)4
• Automatic power settings
require minimal setup and
mimize need for futher
handling during surgery
• Simple, intuitive controls
Confidential. For internal use only. Notand information
for distribution displays
© copyright Valleylab 5
2006, a division of Tyco Healthcare Group LP All Rights Reserved
Valleylab Mode Key Message
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
3 Button
Power Control from Sterile Field
Teflon Coated Blade
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Concepts: You control it!
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Sonicision ™
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
Ultrasonic Dissection System
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
R0024175
Indicator
Reusable Generator LED Torque Wrench
• Low temperature hydrogen peroxide
gas
plasma sterilization (i.e. STERRAD™*)
Audio
Indicator
Ultrasonic Dissector
• Single patient use
Reusable Battery
• Low temperature hydrogen peroxide
gas
plasma sterilization (i.e. STERRAD™*)
Confidential. For internal use only. Not for distribution © copyright Valleylab 2006, a division of Tyco Healthcare Group LP All Rights Reserved
R0024175
Torque Knob