Atlan A300 With E-Charting

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

Atlan Family

A350 XL

Atlan A300 with e-charting

#available in Trolley or ceiling version ( compact size )

#available in compact or XL version www.draeger.com


Internal

why Dräger?
Dräger Technology

1. Automatic system test that allow user to do other preparation


while Atlan self test
2. Intuitive touchscreen allows user to displays user need
3. Econometer & Low flow wizard would serve user to apply minimal
& low flow anesthesia for “Patient safety”
4. Electronic charting to optimize low flow by automatically record
patient data

Ventilation Technology

1. Piston technology, bring ICU level ventilation technology with no


medical gas needed “more economic”
2. Fresh gas decoupling, no intervention to Paw & tidal volume at O2
Flush for “Patient safety”
3. Pause mode, stop ventilation if any leakage on ETT for “User
Safety” cause by no anaesthetic gas while repotitioning ETT
4. With AutoFlow, the set tidal volume is applied for all mandatory
volume-controlled breaths with the lowest required pressure for
“Patient Safety”

www.draeger.com
Operating characteristics (Trolley variant)
Weight of the compact variant Approx. 135 kg (298 lbs), basic setup
Dimensions of the large variant (may deviate with hardware Approx (W x H x D) 74.5 cm x 140.3 cm x 69.2 cm
options) (29.3 in x 55.2 in x 27.2 in)
Dimensions of the work surface on the large variant Width approx. 47 cm (18.5 in), depth approx. 38 cm (15.0 in)
Storage space and work surface 1 lockable drawer, large version with 2 additional drawers
2 x serial ports (RS232) (MEDIBUS.X protocol), 1 x USB port,
Data interfaces 1 x LAN
Ambient conditions
Intended patient population Adults, paediatric patients, and neonates
Gas supply
3-gas version (O2/AIR/N2O), electronic measurement and monitoring of supply pressure of all connected gases (for gas cylinders with optional Dräger pressure reducer)
Central gas supply, supply pressure for O2, AIR, N2O 2.7 to 6.9 kPa x 100 (39 to 100 psi)
Gas supply with gas cylinders (O2, AIR, N2O) 1 or 2 standing gas cylinders (option)
Holder for 1 additional standing gas cylinder (option)
Fresh-gas delivery
Electronically driven piston ventilator (E-Vent plus), fresh-gas decoupled, ventilation without drive gas, i.e. no medical gases are
consumed in operating the ventilator (regardless of gas supply). All patient-gas leading components are autoclavable.
Manual /Spontaneous (Man/Spon)
Volume-controlled:
time controlled (VC-
CMV) Pressure-
Ventilation modes controlled: time
controlled (PC-
CMV) AutoFlow
time controlled (VC-
CMV/AF) Volume-
controlled,
synchronised (VC-
SIMV) Pressure-
controlled,
synchronised (PC-
SIMV) AutoFlow,
synchronised (VC-
SIMV/AF)
CPAP / PSV
External fresh-gas outlet
Vaporizer Sevoflurene and Isoflurene included
Anaesthetic gas scavenging system (AGS)
Available as active or passive anaesthetic gas scavenging system for operation with and without adequate scavanging system infrastructure; detection of excessive suction flows,
with connector for sample gas disposal when using third-party patient gas measurement modules.
For connection to an anaesthetic gas scavenging system
Active AGS With a control valve (option) or an ejector (option)
For connection to a disposal system with low or no suction flow
Passive AGS Maximum inlet flow 0.5 L/min With overpressure valve and underpressure valve
Measuring systems and displays
15.3" (38.9 cm) touchscreen, configurable screen contents, smart
Screen alarm management with extensive support system
Depending on the machine configuration simultaneous display
of 2, 3 or 4 real-time curves for:
concentration of CO2, O2, and anaesthetic
agents, airway pressure, inspiratory and
Screen configuration expiratory flow; display of virtual flow tubes
for O2, AIR, N2O, of tabular trends, quick
access to 3 screen layouts.
Front Panel with LC display of airway pressure, supply status of
Device status display battery and gases (CGS + cylinders)
Display of graphical trends or mini-trends simultaneously with
real-time curves and volume-pressure loop;
Advanced trend display additional data export functions via USB
storage device
Minute volume (MV) and tidal volume (VT and ΔVT); respiratory
rate (frequency); peak inspiratory pressure
(PIP), plateau pressure (Pplat), mean airway
pressure (Pmean), PEEP; dynamic compliance
(Cdyn), resistance (R), elastance (E), external
Ventilation monitoring pressure gauge for indicating the pressure in
the internal breathing system
Bar diagram display of volume and tidal volume, simultaneous
display of 2 loops: Volume-pressure and
Advanced ventilation monitoring (Option) flow-volume, reference loop
Gas monitoring
The device can monitor inspiratory O2 concentration or use the integrated patient-gas measurement module for O2, N2O and anaesthetic agents
Inspiratory and expiratory gas concentration of O2, N2O, CO2
and anaesthetic agents, automatic
Version with integrated patient-gas measurement module identification of isoflurane, sevoflurane,
desflurane, halothane, enflurane, detection
of anaesthetic gas mixtures, age-corrected
xMAC display; sample gas returned to the
breathing circuit
Econometer for displaying fresh-gas efficiency (optionally
including trend and/or in the form of
Advanced gas monitoring (Option by PGM module) low-flow wizard), determination of
consumption and uptake
(determination of
uptake only for anaesthetics), fresh-gas and
anaesthetics per case and since last zeroing
Safety functions
– The integrated device checklist and illustrated step-by-step instructions for daily machine preparation help to comply with national guidelines, such as DGAI (Deutschland),
ASA/APSF (USA), AAGBI (UK)
– Man/Spon ventilation with dosing of O2, AIR (N2O option) and anaesthetic agents possible even when switched off (emergency start-up)
– Backup manual mode allows the direct change to manual ventilation while maintaining gas and ventilation monitoring; O2, AIR (N2O option) and anaesthetic agents from
the vaporisers can be continuously delivered
– Mechanical ventilation with ambient air in case of complete failure of the gas supply, the change to intravenous anaestheticagents is required
– Real-gas O2 test integrated in automatic machine self-test (option with patient-gas measurement module)
Comfort functions and other features
– Automatic start-up and self-test of machine including calibration of all sensors and testing of all control valves; normally no user action necessary after start of test
– Autoset function for adjusting all alarm limits, CBM mode (cardiac bypass mode)
– Breathing bag as an indicator of fresh-gas deficiency and leaks
– Pause mode for short-term interruptions of ventilation
– Data storage on USB storage device (alarm history, self-test results, screen shots, trends and machine configurations)
– Time-saving transfer of device default settings and configurations to other Atlan devices (export and import of configurationdata via USB storage device)
– Integrated, dimmable illumination of working and documentation surfaces
– Central brake, smooth running castors with optional cable deflectors

Workstation
TECHNICAL DATA
Display Specifications Touch Screen Display
Display screen 12.” color TFT
Resolution: 800 × 600
Waveforms Maximum of 11 waveforms
Ukuran Dimensi 344 mm (Weight) × 266 mm (Height) × 145 mm (Depth)
Berat 5 kg
Power supply 100 V to 240 V~, 50 Hz/60 Hz
Battery life 350 min (At 25±2 C, with (a) new fully charged
Battery charge time ≤ 390 min, 100% charge
SUPPORTED PARAMETER
ECG
Lead mode 3-lead : I, II, III
Heart rate calculation
Range ADU: 15 to 300 bpm
PED/NEO: 15 to 350 bpm
PVC
Range ADU: 0 to 300 PVCs/min
PED/NEO: 0 to 350 PVCs/min
ST value
Range –2.0 to 2.0 mV
RESPIRATION
Method Impedance between RA-LL, RA-LA
Measurement lead Default lead II
Measurement Range
Adult 0 to 120 rpm
Neo/Ped 0 to 150 rpm
Sweep 6.25 mm/s, 12.5 mm/s, 25mm/s, 50mm/s
Apnea Alarm time setup 10s, 15s, 20s, 25s, 30s, 35s, 40s, default is 20s
NIBP
Method Oscillometric
Mode Manual, Auto, Continuous
Measuring interval in auto mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 240, and 480 min
Continuous 5 min, interval is 5 s
Measuring type Systolic Pressure, Diastolic Pressure, Mean Pressure
Alarm type SYS, DIA, MAP
Pulse Rate
Measuring range 40 to 240 bpm
Accuracy ±3 or 3.5 %, whichever is larger
SpO2
Measuring range 0 to 100 %
Resolution 1%
TEMPERATURE
Technique thermal resistance
Position Skin
IBP dual invasive pressure
Technique Direct Invasive measurement
Measuring range
Art 0 to 300 mmHg
PA -6 to 120 mmHg
CVP/RAP/LAP/ICP -10 to 40 mmHg
P1/P2 -50 to 300 mmHg