1. Technical Specifications
Items Description
Items Description
Items Description
Reproducibility <0.5%
Items Description
Items Description
Position 1 Rhodium
Position 2 Silver
This tube is an anode-grounded rotating anode tube with a focal track made of respectively Tungsten(W).
It has two focal spots, with nominal focal spot values of 0.3 and 0.1 according to IEC60336 (2005).
The target angle(s) with respect to the specified reference axis are:
Target angle with respect to the specified Small focal spot 0.1 ~ 10°
reference axis
Large focal spot 0.3 ~ 16°
3
Strictly Confidential: Do not reproduce or distribute before prior consent from DRTECH Corp.
Source to Image Distance (SID)
For the large focal spot the reference axis intercepts the image reception area in the middle of a row of the
detector parallel to the chest wall, 93 mm from the first active row
For the small focal spot the reference axis intercepts the image reception area in the middle of a row of
the detector parallel to the chest wall, 93 mm from the first active row
When performing examinations in contact view, the breast support contains a reciprocating anti-scatter
grid (Potter-Bucky device).
This grid is positioned with the septa parallel to the chest wall side for compatibility with angulated
operation (such as tomosynthesis). For this reason, the movement of the grid has very small amplitude (<
2 mm) to preserve the capability to image the region close to the chestwall of the patient.
A special speed-law has been designed to perform a satisfactory erasure of the grid lines in spite of this
short amplitude.
Items Description
4
Strictly Confidential: Do not reproduce or distribute before prior consent from DRTECH Corp.
Items Description
Items Description
Source-to-image distance
660 mm ± 6.6 mm
(SID)
1.3.7 Compression
Items Description
5
Strictly Confidential: Do not reproduce or distribute before prior consent from DRTECH Corp.
Items Description
▪ Minimum display: 3 daN
▪ Maximum display: 30 daN
▪ 24X30 cm screening paddle
▪ 18x24 cm screening paddle
▪ Sliding perforated 18x24 cm paddle
Paddle
▪ Small breast and implant 10x24 cm paddle
▪ Spot Paddle
▪ Magnification Paddle
Items Description
Light field to X-ray difference Chest wall, right, left and nipple side < SID 2%
Items Description
1.3.10 Others
Items Description
6
Strictly Confidential: Do not reproduce or distribute before prior consent from DRTECH Corp.
Items Description
▪ C-ARM angle
Display information on gantry
▪ Compression force
base
▪ Compression thickness
Weight ≤ 300 kg
Items Description
7
Strictly Confidential: Do not reproduce or distribute before prior consent from DRTECH Corp.
1.5 Dimension
8
Strictly Confidential: Do not reproduce or distribute before prior consent from DRTECH Corp.
1.5.2 Dimensions of the Acquisition Workstation
9