Professional Documents
Culture Documents
Service Manual For The CARESTREAM X-Ray Generator: Publication No. 7K3254 Rev E 2023-06-19
Service Manual For The CARESTREAM X-Ray Generator: Publication No. 7K3254 Rev E 2023-06-19
7K3254 Rev E
2023-06-19
7K3254 – 2023-06-19 1
Service Manual
This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.
Table of Contents
Description Page
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Safety Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Electrical/Mechanical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Usability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Flammability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Manual Conventions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
General Use Cautions and Special Messages. . . . . . . . . . . . . . . . . . . . . . . . . . 6
Radiation and Magnetic Field Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Electrical and Flammable Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Safety Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Dataplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Modification Record Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Essential Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Product Disclosure Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Conforming Standards - Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Acoustic Noise Emission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
EMC Compliance Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
EMC Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Specification and Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Electrical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
AC Mains Cable Requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Mechanical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Main Component Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Compatibility. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Abbreviations and Definitions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
AC Main Cable Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Auxiliary Transformer Line Voltage Connection . . . . . . . . . . . . . . . . . . . . . . . . . 28
Stator and Tube Thermal Switch Connections. . . . . . . . . . . . . . . . . . . . . . . . . . 29
High Tension Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
X-ray Tube Housing Ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Emergency-Off Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Room Interface Board Inputs & Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Oil Level Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
DIP Switch Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Dual Speed Starter Tube Select Table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Jumper Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Switch On Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Service Tool Overview (DirectView). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
File Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Tools Menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Setting Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
Help Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
2 2023-06-19 – 7K3254
Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Functional Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
System ON (9H4401) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
DC Bus & Power Distribution (9H4400) . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Room Interface (9H4402) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
X-ray Exposure (9H4403) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
kV Control and Feedback (9H4404) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Filament Drive and mA Control (9H4405). . . . . . . . . . . . . . . . . . . . . . . . . . 54
Dual Speed Starter (9H4407) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Temp Sensor / Programming Jumpers (7K1362) . . . . . . . . . . . . . . . . . . . . 55
Serial Communications (9H4409) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Automatic Exposure Control (9H4408) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Interconnect Diagram (7K1361) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
General Remarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Strategy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Start-up Sequence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Switch On Problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Error Code List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Generator Error Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Generator Warning Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Service Tool Warning message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Error Code Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Error Code: Console Error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Service Tool Error: Console Error. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
LED Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
LED Indicators on MCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
LED Indicators on AEC board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
LED Indicators on RCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
LED Indicators on FCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Test Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Adjustment and Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
MCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
FCB. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
RCB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Inverter-Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
AEC Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
Room Interface Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108
Tank . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
Other Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Generator Firmware Upgrade Procedure (DirectView). . . . . . . . . . . . . . . . . . . . 114
Generator Firmware Upgrade Procedure (ImageView) . . . . . . . . . . . . . . . . . . . 117
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 119
Equipment Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 120
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Illustrated Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Figure 1 Generator Right View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Figure 2 Inverter Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
Figure 3 Generator Left View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124
7K3254 – 2023-06-19 3
Service Manual
4 2023-06-19 – 7K3254
Section 1: Safety Information
The information contained herein is based on the experience and knowledge relating to the subject matter gained by
Carestream Inc. prior to publication.
No patent license is granted by this information.
Carestream Health, Inc. reserves the right to change this information without notice, and makes no warranty, express
or implied, with respect to this information. Carestream shall not be liable for any loss or damage, including
consequential or special damages, resulting from any use of this information, even if loss or damage is caused by
Carestream's negligence or other fault.
Overview
The CARESTREAM X-ray Generator is a High Voltage X-ray Generator which is intended for use in
film-based stationary radiographic X-ray systems and also provides a digital interface for digital
radiography (DR) equipment. The intended use for the generator is to provide high voltage to X-ray tube
between anode and cathode, filament current to tube cathode filament, voltage to drive tube anode to
rotate, and interface connection to room equipment. The CARESTREAM X-ray generator does not
include a console panel or console PC to control the generator, but does provide a RS-232 port for a
console to communicate with the generator. This port also includes a remote power on/off connection.
Personnel operating and maintaining the CARESTREAM X-ray Generator should receive training and
be familiar with all aspects of operation and maintenance. To ensure safety, read the Safety Information
section carefully before using the system and observe all Cautions, Importants, and Notes located
throughout this manual and other manuals supplied with the equipment.
Safety Notice
• Do not modify this equipment without authorization of the manufacturer.
• If this equipment is modified, appropriate inspection and testing must be conducted to ensure
continued safe use of equipment.
• Allow only authorized, properly trained personnel to operate this equipment.
• Be certain that all individuals authorized to use the equipment are aware of the danger of excessive
exposure to X-ray radiation.
• This equipment is sold with the understanding that the manufacturer, its agents, and
representatives do not accept any responsibility for over-exposure of patients or personnel to X-ray
radiation.
• The manufacturer does not accept any responsibility for over-exposure of patients or personnel to
X-ray radiation generated by this equipment as a result of poor operating techniques or procedures.
• The manufacturer accepts no responsibility for any generator not maintained or serviced according
to this service and installation manual, or for any generator that has been modified in any way.
Electrical/Mechanical
Caution
Only a qualified authorized Service Provider should replace electrical and mechanical components.
The following are mechanical safety precautions:
• Do not operate the equipment with covers or access panels removed.
• Route cables properly to eliminate hazards from tripping.
Usability
The design and development of a diagnostic X-ray incorporated a usability engineering process in
accordance with IEC 60601-1-6: Medical Electrical Equipment, Part 1-6: General Requirements for
Safety - Collateral Standard: Usability.
7K3254 – 2023-06-19 5
Service Manual
However, it is not possible or practical to resolve every potential usability issue without affecting the
intended use of the system.
Therefore, some precautions must be observed. These precautions appear throughout the manual.
Training
This equipment is intended for use by appropriately educated and skilled radiological healthcare
professionals who have received specific training on the operation and use of this equipment.
Caution
Only qualified personnel may operate the X-ray generator.
For training in the operation of this equipment, contact Carestream Inc.
Flammability
Not suitable for use in the presence of a flammable anesthetic mixture with oxygen or nitrous oxide.
Manual Conventions
This manual uses three types of messages to emphasize information or potential risks to personnel or
equipment: Note, Important and Caution.
Note
Notes provide additional information, such as expanded explanations, hints, or reminders.
Important
Important highlights critical policy information that affects how you use this manual and this product.
Caution
Cautions point out procedures that you must follow precisely to avoid injury to yourself, others, damage
to the system or any of its components, loss of data, or corruption of files in software applications.
Disregarding the caution statement may lead to abnormal use.
6 2023-06-19 – 7K3254
Radiation
Use of controls or adjustments or performance of procedures other than those specified herein may
result in hazardous radiation exposure.
Caution
Do not operate this equipment outside of its operating environment limits. Doing this may cause the
equipment to malfunction. The operating environment limits are as follows.
Caution
This product contains lead. Disposal of components that contain these materials may be regulated due
to environmental conditions. For disposal or recycling information, contact your local authorities or visit
the Electronics Industry Alliance web-site at http://www.eiae.org.
Caution
The generator includes no user-serviceable parts. For service assistance contact Carestream Inc.
Radiation
The generator connects with X-ray tube which produces ionizing radiation. Operators must meet all
international, national, state, and local requirements and regulations.
Radiation
This X-ray unit may be dangerous to patients and operators unless safe exposure factors and operating
instructions are observed.
Caution
Only allow trained X-ray personnel to use the operator console.
Caution
The generator is not suitable for operation in the presence of a flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
Safety Symbols
The following safety-related symbols are found on the equipment.
7K3254 – 2023-06-19 7
Service Manual
5 Heavy caution
8 2023-06-19 – 7K3254
Safety Symbol Location
7K3254 – 2023-06-19 9
Service Manual
10 2023-06-19 – 7K3254
3
7K3254 – 2023-06-19 11
Service Manual
Dataplate
Figure 1 Dataplate Template
Note
The CE Mark is a mandatory conformity marking for EU Market access.
The ETL Mark is proof of product safety compliance to North American safety standards.
For the specification, refer to Section 2.
12 2023-06-19 – 7K3254
Modification Record Label
Classification
Type of protection against electric shock: Class I Equipment
Permanently installed ordinary equipment
The degree of protection against ingress of water: IPX0
According to the degree of safety of application in the presence of a flammable anaesthetic mixture with
air or with oxygen or nitrous oxide: Equipment not suitable for use in the presence of a flammable
anaesthetic mixture with air or with oxygen or nitrous oxide.
The mode of operation: CONTINUOUS OPERATION WITH INTERMITTENT LOADING.
Essential Performance
• Accuracy of loading factors (IEC 60601-2-54)
• Reproducibility of radiation output (IEC 60601-2-54)
• Linearity and constancy of radiation output (IEC 60601-2-54)
Radiation
Loss of essential performance above can lead to excessive radiation and cause harm to the patient.
Follow all instructions in this manual to avoid such happening.
7K3254 – 2023-06-19 13
Service Manual
14 2023-06-19 – 7K3254
For out of China market
Carestream Health, Inc., hereby declares under its sole responsibility that the product(s) listed are
made in conformity with ANNEX I, Essential Requirements and ANNEX VII, EC Declaration of
Conformity of the European Economic Community Medical Device Directive, [Directive 93/42/EEC];
and Article 4 of the Directive on the restriction of the use of certain hazardous substances in electrical
and electronic equipment [Directive 2011/65/EU].
Canada:
CAN/CSA-C22.2 No. 601.1-M90 (R2005) - Medical Electrical Equipment
CAN/CSA-C22.2 No. 601.1S1-94 (R1999) - Supplement No. 1-94 to CAN/CSA-C22.2 No. 601.1-M90
CAN/CSA-C22.2 No. 601.1B-90 (R2006) - Amendment 2 to CAN/CSA-C22.2 No. 601.1-M90
CAN/CSA-C22.2 No. 60601-1-1-02 (R2006) - Medical Electrical Systems
CAN/CSA-C22.2 No. 60601-1-14 - Medical Electrical Equipment - Part 1: General requirements for
safety and essential performance
Europe
EN 60601-1:1990, + Amendment 1:1993, + Amendment 2:1995 - Medical Electrical Equipment
EN 60601-1-4:1996, + Amendment 1:1999 - Programmable Electrical Medical Systems
EN 60601-1-6:2004 - Usability
EN 60601-2-7:1998 - Safety of High-Voltage Generators of Diagnostic X-ray Generators, 2nd Edition
EN 60601-1:2006+A1:2013 - Medical Electrical Equipment - Part 1: General requirements for safety
and essential performance
EN 60601-1-6:2010 - Medical Electrical Equipment - Part 1: General requirements for safety and
essential performance - Collateral Standard: Usability
EN 60601-2-54:2009 - Medical Electrical Equipment - Part 2-54: Particular requirements for the basic
safety and essential performance of X-ray equipment for radiography and radioscopy
International
IEC 60601-1:1988, + Amendment 1:1991, + Amendment 2:1995 - Medical Electrical Equipment
IEC 60601-1-4:1996, + Amendment 1:1999 - Programmable Electrical Medical Systems
IEC 60601-1-6:2004 - Usability
IEC 60601-2-7:1998 - Safety of High-Voltage Generators of Diagnostic X-ray Generators, 2nd Edition
IEC 60601-1:2005+A1:2012- Medical Electrical Equipment - Part 1: General requirements for safety
and essential performance
IEC 60601-1-6:2010 - Medical Electrical Equipment - Part 1: General requirements for safety and
essential performance - Collateral Standard: Usability - 3rd Edition
IEC 60601-2-54:2009 - Medical Electrical Equipment - Part 2-54: Particular requirements for the basic
safety and essential performance of X-ray equipment for radiography and radioscopy
7K3254 – 2023-06-19 15
Service Manual
16 2023-06-19 – 7K3254
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Carestream X-ray Generator is intended for use in the following electromagnetic environment.
The customer or the user of the Carestream X-ray Generator should ensure that it is used in such
an environment.
Immunity Test IEC 60601 Compliance Electromagnetic Environ-
Test Level Level ment - Guidance
Electrostatic ±8 kV contact ±8 kV contact Floors should be wood, con-
Discharge (ESD) ±15 kV air ±15 kV air crete or ceramic tile. If floors
IEC 61000-4-2 are covered with synthetic
material, the relative humidity
should be at least 30 %.
Electrical Fast ±2 kV for power sup- ±2 kV for power supply Mains power quality should be
Transient/Burst ply lines lines that of a typical commercial or
IEC 61000-4-4 ±1 kV for input/out- ±1 kV for input/output hospital environment.
put lines lines
Surge ±1 kV line (s) to line ±1 kV line (s) to line (s) Mains power quality should be
IEC 61000-4-5 (s) ±2 kV line (s) to earth that of a typical commercial or
±2 kV line (s) to earth hospital environment.
Voltage Dips, Short 0 % UT; 1 cycle 0 % UT; 1 cycle Mains power quality should be
Interruptions and and 70 % UT; 25/30a) and 70 % UT; 25/30a) that of a typical commercial or
Voltage Variations cycles cycles hospital environment. If the
on Power Supply 0% UT; 250/300a) 0% UT; 250/300a) cycle user of the Carestream X-ray
Input Lines IEC cycle Generator requires continued
61000-4-11 operation during power mains
interruptions, it is recom-
mended that the Carestream
X-ray Generator be powered
from an uninterruptible power
supply or battery.
Power Frequency 30 A/m 30 A/m Power frequency magnetic
(50/60 Hz) Mag- fields should be at levels char-
netic Field IEC acteristic of a typical location
61000-4-8 in a typical commercial or hos-
pital environment.
a) For example, 10/12 means 10 periods at 50 Hz or 12 periods at 60 Hz.
7K3254 – 2023-06-19 17
Service Manual
a.The ISM (industrial, scientific and medical) bans between 150 k Hz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0
MHz, 5.3 MHz to 5.4 MHz, 7 MHz to 7.3 MHz, 10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07
MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz, 24.89 MHz to 24.99 MHz, 28.0 MHz to 29.7 MHz and
50.0 MHz to 54.0 MHz.
b. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radios, AM and FM radio broadcasts, and TV broadcasts
cannot be predicted theoretically with accuracy. To assess the electromagnetic envionment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the [ME EQUIPMENT or ME SYSTEM] is used exceeds the
applicable RF compliance level above, the [ME EQUIPMENT or ME SYSTEM] should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as re-orienting or relocating the Carestream X-ray Generator.
c. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
18 2023-06-19 – 7K3254
Recommended Separation Distances Between
Portable and Mobile RF Communications Equipment and the Carestream X-ray
Generator
The Carestream X-ray Generator is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the
user of the Carestream X-ray Generator can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Carestream X-ray Generator as
recommended in the following table, according to the maximum output power of the
communications equipment.
Rated Maximum Separation Distance According to Frequency of Transmitter
Output Power of (m)
Transmitter
150 kHz to 80 MHz 80 MHz to 800 800 MHz to 2.7
w)b
MHz GHz
b) For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in metres (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
7K3254 – 2023-06-19 19
Service Manual
EMC Information
In accordance with the intended use, some models of this series of X-ray generators comply with the
European Council Directive concerning Medical Devices. The CE marking affixed to compliant products
signifies this. One of the harmonized standards of this Directive defines the permitted levels of
electromagnetic emission from this equipment and its required immunity from the electromagnetic
emissions of other devices.
It is not possible, however, to exclude with absolute certainty the possibility that other high frequency
electronic equipment, which is fully compliant to the EMC regulations, will not adversely affect the
operation of this generator. If the other equipment has a comparatively high level of transmission power
and is in close proximity to the generator, these EMC concerns (the risk of interference) may be more
pronounced. It is therefore recommended that the operation of equipment of this type such as mobile
telephones, cordless microphones and other similar mobile radio equipment be restricted from the
vicinity of this X-ray generator.
The Carestream X-ray Generators are intended to be used in the hospital environment except for near
active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM for magnetic
resonance imaging, where the intensity of EM DISTURBANCES is high.
Caution
Use of this equipment adjacent to or stacked with other equipment should be avoided because it could
result in improper operation. If such use is necessary, this equipment and the other equipment should
be observed to verify that they are operating normally.
Caution
Only connect cables specified below to the Carestream X-ray Generators.
Cable Manufacturer Name Model Name Length (m) Shielded (Y/N)
RS-232 Cable Generic Generic ≤ 26 m Y
Use of accessories, transducers and cables other than those specified in this manual or provided by
Carestream could result in increased electromagnetic emissions or decreased electromagnetic
immunity of this equipment and result in improper operation , and can influence basic safety and cause
excessive radiation.
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the Carestream X-ray
Generators, including cables specified in this manual. Otherwise, degradation of the performance of
this equipment could result.
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals
(CISPR 11 class A). If it is used in a residential environment (for which CISPR 11 class B is normally
required) this equipment might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or re-orienting the
equipment.
20 2023-06-19 – 7K3254
Section 2: Specification and Introduction
Electrical Specifications
Power Requirement
CGF-65 CGF-80
(65 kW) (80 kW)
Nominal input voltage 380 V (ac) ±10 %,3 phase, no neutral
400 V (ac) ± 10 %,3 phase, no neutral
480 V (ac) ± 10 %,3 phase, no neutral
Line Frequency 50/60 Hz
Momentary Current
380 V 140 A 166 A
400 V 135 A 158 A
480 V 115 A 143 A
Nominal Current ≤5A
Momentary Power 100 kVA 125 kVA
Consumption
Output Specifications
CGF-65 CGF-80
(65 kW) (80 kW)
Inverter Frequency 240 kHz
kVp range 40 - 150 kV
kVp step variable in 1 kV step
kV accuracy ± (5 % +1 kV)
Rise time (10 - 90 %) < 1.5 ms
Time range 1.0 to 6300 ms
Time accuracy ±(10 % +1) ms
mAs range 0.1 to 800 mAs 0.1 to 1000 mAs
mA range ±(10 % +0.2) mAs
mA range 10 to 800 mA 10 to 1000 mA
Coefficient of linearity 0.05 (station to station) mAs
Coefficient of reproducibility ≤ 0.05
Service Manual
Minimum
Recommended
Mains Minimum Minimum
Line Momentary Minimum Disconnect to Recommended Recommended Maximum
Generator voltage Line Recommended Generator Transformer Ground wire main
Model (VAC) Current Service Rating (5 m max) Rating size resistance
CGF-65 380 140 A 80 A #6 13.3 mm 85 kVA #6 13.3 mm 0.14 ohm
(65 kW) 400 135 A 80 A #6 13.3 mm 85 kVA #6 13.3 mm 0.15 ohm
480 115 A 80 A #6 13.3 mm 85 kVA #6 13.3 mm 0.21 ohm
CGF-80 380 166 A 100 A #4 21 mm 105 kVA #4 21 mm 0.11 ohm
(80 kW) 400 158 A 100 A #4 21 mm 105 kVA #4 21 mm 0.13 ohm
480 143 A 80 A #6 13.3 mm 105 kVA #6 13.3 mm 0.17 ohm
Note
1. All wiring and grounding should comply with the local electrical code.
2. All wiring must be copper.
3. The disconnect switch must be located within reach of the operator.
Environmental Specifications
Transport and Storage Environments (with package)
Temperature: -25 to 70 °C (-13 to 158 F)
Humidity: 5 to 95 % (non-condensing) Atmospheric Pressure: 54 kPa -110 kPa
Operating Environments
Temperature: 10 to 40 °C (50 to 104 F)
Humidity: 20 to 80 % (non-condensing)
Atmospheric Pressure: 70 kPa -110 kPa, -70 to 3000 meters (-230 to 9843 feet)
22 2023-06-19 – 7K3254
Mechanical Specifications
7K3254 – 2023-06-19 23
Service Manual
24 2023-06-19 – 7K3254
10 11
9
8 12
13
14
15
16
17 18
1. HV tank
2. AEC (Automatic Exposure Control)
3. MCB (Main Control Board)
4. FCB (Filament Control Board)
5. Dual Speed RCB (Dual Speed Rotor Control Board)
6. Inv B (Inverter Assembly B)
7. Inv A (Inverter Assembly A)
8. Mains Input Cable Access
9. Main Circuit Breaker
10. Cable Access
11. Filter
12. RIB (Room Interface Board)
13. LS (Low Speed) Phase-shift Capacitor
14. HS (High Speed) Phase-shift Capacitor
15. Rectifier
16. Contact
17. Fuse
18. Auxiliary Transformer
7K3254 – 2023-06-19 25
Service Manual
Compatibility
The generator is compatible with the following tube types:
Manufacturing Model Stator
Toshiba E7252X XS-AL
XRR4631G XS-BM
Varian RAD92 Q
RAD60 Q
Kailong H1086 R
Options
The generator has the following options:
1. 9 pin connector AEC board. 15 pin connector AEC board is standard delivery.
2. Varian fixed detector interface board.
26 2023-06-19 – 7K3254
Section 3: Installation
Unpacking
1. Inspect the shipping pack(s) for evidence of shipping damage. If there is evidence of shipping
damage, note this in the event that a damage claim is justified. Taking pictures of the damage is
also recommended.
2. Remove the cardboard outer pack from the buttom of the cardboard box. See the cautionary note
below before removing the pack.
3. Set aside the cardboard pack.
6. Keep the shipping packs. In case of shipping damage, place the unit(s) back in their shipping
pack(s) and notify the carrier and the customer support department.
7. Remove and set aside the screws and washers securing the cover to the generator chassis.
8. Carefully lift the cover off the chassis.
7K3254 – 2023-06-19 27
Service Manual
Procedures:
1. Pass the AC mains cable through the cable clamp at the upper rear of the generator cabinet.
2. Connect the ground wire to the chassis ground connector.
3. Connect the mains power wires to the terminals on the main circuit breaker 1L1, 3L3, and 5L5.
4. Tighten the clamps to secure the cable.
5. Verify the main circuit breaker is set to the ON position and not tripped.
Note
The AC mains cable conductor size for circuit breaker should be in the range 18 - 4 AWG.
Chassis ground
AC mains connection
The auxiliary transformer line voltage connection should be changed to the appropriate position on the
transformer primary according to the nominal line voltage.
1. Verify the mains voltage used and that it meets the generator power requirements. Loosen the
clamping screws for the current line-voltage tap.
2. Connect input power supply line to the proper connector, according to input line voltage.
3. Retighten both of the clamping screws.
28 2023-06-19 – 7K3254
Note
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with the
protective earth.
The AC mains cable is routed into the generator via the cable clamp on the rear of the generator.
All the other cables should be routed into the generator main cabinet through the cable access slots at
the upper rear of the generator. The cables should be clamped with the cable strain relief.
For connections that must be made to the rotor control board, the AEC board, and the main control
board, route the cables and secure them with tie-wraps by the way of the tie wrap slot. All cables should
be kept away from high voltage areas in the cabinet, and wrapped neatly in place.
3. Route the X-ray tube stator cable through the cable access slot and connect the cable to terminal
blocks on the Room Interface Board as below. Secure the cable with tie wraps.
7K3254 – 2023-06-19 29
Service Manual
Designator Connect to
SHIFT P8-3 SHIFT
MAIN P8-2 MAIN
COMMON P8-1 COM
GROUND GND
THERMAL SWITCH (TERM1) J4-1
THERMAL SWITCH (TERM1) J4-3
The stator cable must be shielded and the shielded layer must be properly grounded at both tube side
and generator side.
Connect the generator end of the shielded layer to the ground terminal on the back plate.
Caution
Obtain and familiarize with Safety Data Sheet or Article Information sheet before using chemicals
1. Verify that the HT cable connectors are clean and in good condition. Use alcohol to completely
clean the connectors on the cables and the connectors at the HT Tank and X-ray tube.
2. Remove the dust caps that cover the high voltage terminals on the HT tank and tube.
3. Insulate the high tension cable connectors.
Note
For Carestream systems, we use silicone washer and insulating oil to insulate the high tension cable
connectors. This is available in kit, part number TL5515.
a. Use two vials of oil in each connector of the HT Tank. Make sure the oil level is still sufficient
by fully seating and then removing the candlestick to see the resulting oil level on the side of
the candlestick. The oil should reach at least 1/3 the length of the candlestick.
b. Place one silicon washer in each connector of the x-ray tube. If desired, a drop or two of
insulating oil can be placed on the washer before installing the washer. Placing a thin film of oil
on the washer before installing it into the x-ray tube, will help prevent the washer from sticking
and make it easier to remove at a later time. Do not place excess oil in the x-ray tube
connectors.
4. Connect the anode and cathode cables to the HT tank and tube. Ensure that the cables are plugged
into the proper connectors.
5. Make the HT cable connectors tight between the connector insulator and the ring nut.
30 2023-06-19 – 7K3254
To prevent tube stator cable from pulling out, it’s recommended to follow two steps below to fix the tube
stator cable with high tension cable.
1. Bind the stator cable together with the high tension cable by using tie wrap.
2. Make the combined cables above go through the cable access hole by using tie wrap to better
fixing. See the illustration below.
Emergency-Off Circuit
If the external emergency-off circuit is connected to the generator, remove the short jump cable from
P54- 1 to P54-3 on main control board. Then connect the emergency-off switch to P54-1 and P54-3.
If it’s not necessary to connect the external emergency-off circuit, link P54 pin1-3.
See 9H4401 "system on" in "Schematics“.
7K3254 – 2023-06-19 31
Service Manual
Radiation
Do not disable the function detecting room door interlock, unless the corresponding room equipment is
not present. Consult the applicable regulations before disabling any exposure interlocks. Do not violate
any regulations for x-ray safety. Never bypass the x-ray tube thermal switch interlock. Observe ionizing
radiation personal protection at all times.
Table 3-1 Generator Inputs
Pin Definition
J1-7 BUCKY1_READY BUCKY1 ready. A contact closure from pin7 to
J1-8 GND pin8 indicates BUCKY1 READY. Low level active.
J2-7 BUCKY2_READY BUCKY2 ready. A contact closure from pin7 to
J2-8 GND pin8 indicates BUCKY2 READY. Low level active.
J3-1 DOOR_INTERLOCK Door interlock. A contact closure from pin1 to
J3-3 GND pin3 indicates the room door is closed. Low level
active.
J4-1 TUBE_THERMAL_SWITCH Tube overheat switch. A thermal switch closure
J4-3 GND from pin1 to pin3 indicates that the tube is not
over heated.
Note
These inputs are meant for dry contacts only. Do not apply any voltage source to these inputs.
Outputs
Table 3-2 shows the Bucky and auxiliary power outputs from the generator. Refer to Figure 1 for correct
connection.
32 2023-06-19 – 7K3254
Table 3-2 Generator Outputs
Note
The generator is factory configured not to supply power to BUCKY_START OR BUCKY_OUT. The
Bucky outputs can be configured to supply 24 V (dc),110 V (ac), or 220 V (ac) by short jump cable if
required. Please confirm the compatibility with the Bucky used in this installation before operating.
In some installations, BUCKY_START relay, K1or K2 on the room interface board may drive the inputs
of optocouplers or other low current devices on the Bucky. The leakage current through the R snubber
is connected across the relay contacts (R6 for K1, R5 for K3) and may be sufficient to energize the
Bucky inputs when the relays are open. If this is experienced, the snubber should be removed.
Removing the parallel resistor will resolve this problem.
7K3254 – 2023-06-19 33
Service Manual
5220,17(5)$&(%2$5' )
- 3
9$&287 `
IRUFROOLPDWRU - 3
9$& )URP$X[LOLDU\
3
( FRQWURO
( 9$& 3 WUDQVIRUPHU
( - ( (
%XFN\UHWXUQ %8&.<3:5 3
-
) ( 9$&
%XFN\RXW (
- . 9
%XFN\6WDUW 9 ) ),/$0(17&21752
3 3
- ( / %2$5'
%XFN\5'< 9
9 3 3
- (
%XFN\*1'
8
5
73
3 3 ,2
8
,2
5 3 3
9 4
' 9 9 '
%XFN\UHWXUQ - 8
%XFN\RXW -
. 73 5
- .
%XFN\6WDUW
8
- 3 3 ,2
%XFN\5'< 8 '5,9(5
- 5
%XFN\*1' 73
3 3 ,2
9
8PD[ 9 8
5220/,*+7 ,PD[ $
. 9 9
-
-
& 8
73 5
&
.
- 5
'225 8
- 9 3 3 ,2
,17(5/2&. 73 '5,9(5
8 ,2
3 3
9
9 9
78%( - 8
7+(50$/
- 5 5
6:,7&+ 73 5
.
9 8
7202 - 73 3 3 ,2
% 8 '5,9(5 %
(;32685(
- 3 3 ,2
,1387
3 3 ,2
9
3 3
- 8
9'& 3 3
` 5
2XW - 0$,1&21752/%2$5'
&38
Note
1. Refer to Table 3-1 for required input logic levels.
2. The bucky outputs can be configured for +24 V (dc), 110 V (ac) , or 220 V (ac) .
34 2023-06-19 – 7K3254
Oil Level Check
Caution
Obtain and familiarize with Safety Data Sheet or Article Information sheet before using chemicals.
The insulation oil is filled to the correct level in the factory before the tank is sealed. However, the oil
volume and air pressure in the tank will change because of temperature change or oil leakage. It's
necessary to balance the air pressure inside and outside of the tank during the system installation,
simply by loosening the oil fill port screw, waiting a second, and tightening the screw.
The insulation oil in the tank does not require periodic check in normal condition. However, if there is
possible oil loss, follow the procedure below to check the oil level.
1. Loosen and remove the oil fill port screw on the tank lid.
2. With oil fill port screw removed, make sure the rubber washer is removed.
3. Use a clean ruler or other material to measure the oil level. Measure the distance from the top of
the oil level to the top of oil port. Make sure the distance is between 18-25 mm. If the distance is
greater than 25 mm, replace the tank. The tank is not a field repairable part.
7K3254 – 2023-06-19 35
Service Manual
Jumper Setting
Verify JW101 on MCB is jumped on 1-2 position, which means normal operation mode.
36 2023-06-19 – 7K3254
Switch On Generator
Number Explanation
1 Service Tool software version.
2 Communication port number and generator firmware version. Presence of this
information indicates communication established between service tool and generator.
3 Anode heat content percentage (HU) and preparing/exposure indicator.
4 Exposure feedback information.
5 AEC related parameter. Only effective when AEC exposure mode is enabled.
6 Exposure station selection.
Service Tool is a software program used to make exposures, configure the generator, perform specific
calibrations, and view generator status. Menu bar is available on the upper left corner of the interface
screen.
7K3254 – 2023-06-19 37
Service Manual
Note
* means login is required via File->Login to perform any change.
File Menu
Serial Port
Follow these steps to set serial port.
1. From the File menu, select Serial Port.
2. Select Com Port from drop-down list.
3. Press OK.
The Serial Port window is shown below.
Reconnect
From the File menu, select Reconnect to reconnect the generator.
Login
From the File menu, select Login, and input service password to unlock specific functions in Service
Tool (mark with * in Table 3-3).
38 2023-06-19 – 7K3254
The Login window is shown below.
Upgrade Firmware
From the File menu, select Upgrade Firmware to open upgrade firmware window. Refer to Section 6
Replacement Adjustment for upgrade procedure.
Exit
From the File menu, select Exit to exit the software.
OR
Press the close icon on the upper right corner to exit the software.
Tools Menu
Tube Calibration
Before beginning tube auto calibration, the tube must be properly configured, and the generator
limitation should be programmed. It is recommended that the tube should be conditioned (seasoned),
particularly if the tube has not been used for a long period of time.
7K3254 – 2023-06-19 39
Service Manual
Radiation
The following procedures produce X-rays. Take all safety precautions to protect personnel from X-
radiation.
Caution
Always verify the manufacturer of the tube insert. If the X-ray tube has been rebuilt, the tube insert and
tube housing may be from different manufacturers.
2. Select the focus. Toggle the button to select the desired focal spot to calibrate (Small or Large).
Start with large focus.
3. Press and hold the X-ray button (or use the optional hand switch) to begin the calibration
procedure.
4. When finished calibrating the large focus, repeat the calibration on the small focal spot.
5. When the auto-calibration is completed, the Calibration Status displays Complete. Press Close
to return to the Service Tool.
AEC Calibration
This menu is to configure four AEC relevant settings: kV Compensation, Density Setup, Time
Compensation and Field Compensation. Generator supports two channels and 7 film screens, and
these four setting need to be configured on each different combination of film screen and channel.
Note
The demonstrated values are only examples. Configure the values according to system AEC
calibration procedure.
40 2023-06-19 – 7K3254
kV Compensation
Film screen response to kV is not linear. Therefore, the compensation must be provided in order to
maintain the constant film density as kV is changed for different anatomical studies. By selecting and
calibrating various kV breakpoints, the overall system response will be compensated so as to yield a
constant film density.
Density Setup
The Density Setup menu is as below.
7K3254 – 2023-06-19 41
Service Manual
Time Compensation
These values are configured to compensate exposure value according to different exposure time, the
default value will be '0' if no any special illustration.
Field Compensation
If field balance calibration failed by potentiometers on Ion chamber, adjust the Left Field, Central Field
and Right Field accordingly, then click Apply.
Tube Statistics
This utility shows the tube exposure count and generator exposure count. This also allows resetting of
the tube exposure counter. Generator exposure counter cannot be reset.
The Tube Statistics menu is as below.
42 2023-06-19 – 7K3254
To view Tube Statistics, follow these steps.
1. From the Tools menu, select Tube Statistics.
2. The Tube Statistics menu displays tube exposure counter and generator exposure counter, and
allows the tube exposure counter to be reset.
3. Press Close to return to the main page.
View Calibration Data
To view calibration data, follow these steps.
1. From the Tools menu, select View Calibration Data.
2. Toggle the button to select the desired focal spot to view (Small Focus or Large Focus).
3. Click close icon on the top right corner of the window to return to the main page.
View DC Bus Voltage
To view DC bus voltage, follow these steps.
1. From the Tools menu, select View DC Bus Voltage.
2. The pop-up window displays the DC bus voltage.
7K3254 – 2023-06-19 43
Service Manual
Note
Closely follow the calibration procedure to input the filament current. Making exposure with excessive
filament current may damage the x-ray tube.
AEC Field
This utility is designed for R&D test only.
Dual Energy Calibration
Refer to Appendix A: Tomo and Dual Energy Exposure Set Up Procedure for detail.
Setting Menu
Generator Limits
The Generator Limits function allows setting of the generator output limits defined below.
The Generator Limits menu is shown below.
44 2023-06-19 – 7K3254
Function Description
Max Generator Power (kW) Sets the maximum generator kW limit.
Max Generator Current( mA) Sets the maximum generator mA limit.
Min Generator Current (mA) Sets the minimum generator mA limit.
Max Generator Current - Time (mAs) Sets the maximum generator mAs limit.
Note
Before making any changes in this section, consult the x-ray tube data sheets to ensure that the
proposed changes do not exceed the manufacturers’ recommended limits. The original generator limits
should be recorded in a copy before continuing.
Follow these steps to set the generator limits. Refer to the definitions in the previous table.
1. From the Setting menu, select Generator Limits.
2. Set the maximum kW via the Max. Generator Power (kW) dialog box.
3. Set the maximum mA via the Max. Generator Current (mA) dialog box.
4. Set the minimum mA via the Min. Generator Current (mA) dialog box.
5. Set the maximum mAs via the Max. Generator Current-Time (mAs) dialog box.
6. Press Apply to save the setting values.
7. Press Close to return to the main page.
Generator Date
The window below is just for viewing console time.Console time is synchronized with console PC.
Acutally generator time in below window is meaningless .Both generator and console time can't be set
by this menu.
The Generator Date menu is shown below.
7K3254 – 2023-06-19 45
Service Manual
Use these steps to set up the receptor parameters. Refer to the definitions in the previous table.
1. From the Setting menu, select Station Setting.
2. Select the first station to be programmed.
3. Check Enable Station box.
4. Select Rad Hangover(s). Use the arrow buttons to select the desired rad hangover time.
5. Check Auto Focus box to enable/disable auto focus switching function.
6. Select Interface. This function is reserved for future use.
7. Select AEC Channel. Use the arrow buttons to assign the desired AEC channel to the selected
station, or to disable AEC operation on that station.
8. Select AEC backup mAs. Use the arrow buttons to select the maximum backup mAs.
46 2023-06-19 – 7K3254
9. Select AEC back-up time (ms). Use the arrow buttons to select the maximum backup ms.
10. Check Door Interlock in the Standby, Prepare, Ready, or Exposure phase.
11. Check BUCKY1 Ready to check BUCKY1 enabled at Exposure phase.
12. Check BUCKY2 Ready to check BUCKY2 enabled at Exposure phase.
13. Check the Room Light in the Standby, Prepare, Ready, or Exposure phase to activate the room
light in corresponding phase.
14. Enable/disable BUCKY1 Start at the Exposure phase.
15. Enable/disable BUCKY2 Start at the Exposure phase.
16. Press Save to save the programming for the current station.
17. Press Edit Station name to change the station name.
18. Repeat steps 2 to 17 for the remaining stations.
19. Press Close to return to the main page.
Tube Setting
The Tube Setting function allows the desired tube type to be selected, and allows setting of the default
limits for that tube.
Filament Parameter
The Filament Parameter window is shown below.
Note
All filament parameters are fixed to prevent potentially creating a risky scenario for the X-ray tube. The
filament parameters will be automatically set based on the tube selection from Tube Files.
Function Description
Small Focus: Standby current The small focus standby filament current from the X-ray tube data
(mA) sheets.
Small Focus: Max current (mA) The small focus maximum filament current from the X-ray tube
data sheets.
Large Focus: Standby current The large focus standby filament current from the X-ray tube data
(mA) sheets.
Large Focus: Max current (mA) The large focus maximum filament current from the X-ray tube
data sheets.
Boost Time (ms) The filament rapid boost duration in order to quickly raise the
filament temperature.
Preheat Time (ms) The time that the filament is held at the required emission level
before an exposure is permitted.
7K3254 – 2023-06-19 47
Service Manual
Anode Setting
The Anode Setting window is shown below.
Function Description
Anode Warning (%) Sets the limit at which the anode heat-warning message is
displayed.
Adode Limitation (%) Sets the limit at which exposures will be inhibited. If the
anode heating is currently under the limit, the next
exposure will be inhibited if the generator calculates that
the exposure will exceed the anode HU limit.
Follow these steps to set up Anode Setting. Refer to the definitions in the previous table.
1. From the Setting menu, select Tube Setting.
2. Select Anode Setting tab.
3. Set the desired anode HU warning % via the Anode Warning (%) .
4. Set the desired anode HU limit % via the Anode Limitation (%).
5. Press Apply.
Tube Setting
The Tube Setting window is shown below.
Function Description
Max Tube Voltage (kV) Sets the maximum kV allowed.
Max Small Focus Current (mA) Sets the maximum mA in small focus.
Additional Rotator Boost time Sets the additional time (ms) for tube rotator to boost.
(ms)
48 2023-06-19 – 7K3254
Use these steps to set up Tube Setting. Refer to the definitions in the previous table.
1. From the Setting menu, select Tube Setting.
2. Select Tube Setting tab.
3. Set the maximum allowed kV via the Max. Tube Voltage (kV).
4. Set the maximum small focus mA via the Max. Small Focus Current (mA).
5. Set the additional time(ms) for tube rotator to boost via Additional Rotator Boost time (ms).
6. Press Apply.
Tube Files
The Tube Files window is shown below.
Select the tube type in the displayed list, then press Download Tube File.
Note
The selected tube type should be identical to the setting of DIP switch S102.
Follow these steps to set the error log to a different level. The system will capture the log by the log
level.
1. From the Setting menu, select Set Log Level.
2. Select Log Level at Error, Warning, Information and Debug. The default log level is Information.
3. Press OK.
Note
If Debug level log is being used for debugging, do not forget to change back to Information level.
Otherwise, large quantity of log will be created in hard disk.
7K3254 – 2023-06-19 49
Service Manual
Tomo Setting
Refer to Appendix A: Tomo and Dual Energy Exposure Set Up Procedure for detail.
Help Menu
About
From the Help menu, select About to view the about information, including service tool version.
50 2023-06-19 – 7K3254
Section 4: Theory of Operation
Introduction
This section contains the theory of operation for the CARESTREAM X-ray Generator. The theory of
operation is organized by functional blocks as per the functional drawings in Section 9.
System ON (9H4401)
The main control board on the CARESTREAM X-ray Generator includes power ON and power OFF
switches S501 and S502 that may be used to switch the generator on and off locally while working on
the equipment. These switches are connected in parallel with the generator ON and OFF switches that
are located on the interface box.
At the beginning, the Q503 is off and Q508 is on. Pressing any of the power ON switches turns on Q503
on the main control board. This turns on Q504, holding the collector of Q504 low. This latches Q503 in
the "on" state by holding the base of Q503 low when the ON switch is released.
With Q504 turned on, the base of Q505 is held low, turning Q505 off. The collector of Q505 is connected
to the ON / OFF control pins on the +12 V switching regulator on the generator main control board
(9H4400). With Q505 turned off, the regulator control pins will be pulled to a logic high state, enabling
the low-voltage supplies on the generator main control board (9H4400).
A supervisory circuit (not shown) monitors the +5 V (dc) supplies on the main control board. The
supervisory circuit releases the CPU reset line when these low voltage DC supplies are detected at
start-up. This causes the CPU to enter its startup sequence. The CPU will initialize itself, perform its
startup diagnostics, and then output a command via U202 on the main control board to energize pre-
charge contactor K3 on the rotor control board. The CPU will then wait for approximately 10 seconds
for DC bus voltage to reach certain value, detected by CPU through ADC, indicating that the DC bus
capacitors are charged. If CPU detects DC bus capacitor charging within the allotted time, the CPU will
output a command to U202 and the driver K5, etc. to energize the main power contactor K1 via the rotor
control board. Pre-charge contactor K3 will be de-energized after a maximum of 10 seconds, as the
pre-charge circuit is not needed once the DC bus capacitors are charged.
If CPU detects that DC bus voltage does not reach certain value within the allowed time limit, an error
message will be presented and the turn-on sequence will be aborted.
Pressing any of the power-OFF switches turns off Q508 on the main control board, or pressing the
emergency stop button turns off Q504 on the main control board. Q503 will be turned off in either
circumstance mentioned above. Then, Q505 will turn on, taking the ON / OFF control lines on the +12
V AA and +12 V regulators low. This disables the switching regulators, shutting down the generator
control functions. The main power contactor will open, removing the AC input to the mains rectifier
assembly.
On the filament control board, LED DS1 confirms the presence of the +24 V (dc) supply, and LED DS2
indicates -35 V (dc) supply and LED DS3 indicates +35 V (dc) supply.
The BUCKY DRIVE output may be configured for +24 V (dc), 110 V (ac), or 220 V (ac) by setting room
interface board. Connecting E12 toE10 and E5 to E3 supplies +24 V (dc), wiring E12 to E6 and E1 to
E3 supplies 110 V (ac), and jumping E12 to E8 and E1 to E3 supplies 220 V (ac) (refer to 9H4402).
The cooling fan will be energized by turning on the MOSFET on the room interface board when inverter
is over-temperature.
As described in System ON (9H4401), the +12 V AA and +12 V switching regulator become operational
when the generator is switched on. The -12 V supply is derived from the +24 V. The +5 V supply is
derived from the +12 V AA supplies. The +1.8 V and +3.3 V supplies are derived from the +5 V supply,
respectively.
52 2023-06-19 – 7K3254
--The tank must be connected.
With the base of Q314 held low, Q314 will be off. The line connected to TP303 on the main control
board will then be pulled high via R356, pulling the control inputs on NAND gates U314A and U314B
on the main control board high, as described in kV Control and Feedback (9H4404).
The base of Q312 on the main control board must be held low to allow an exposure. This requires no
filament fault.
The base of Q315 on the main control board must be held low to allow an exposure. This requires no
beam fault.
The latch circuit of PT STOP on the main control board must be held high to allow an exposure. This
requires no PT STOP from AEC board.
Lastly, with the output of CPU on the main control board low, the collector of Q313 will be high. This
turns Q405, Q406 and Q407 on, isolating from the kV demand circuit, part of kV Control and Feedback
(9H4404).
7K3254 – 2023-06-19 53
Service Manual
U507, which will send a fault pulse to the CPU via U509 if a kV over-voltage condition is detected.
The main control board also contains circuits that guard against an inverter "shoot-through" fault. If a
shoot through fault is detected, T501 and / or T502 on the main control board will supply a current pulse
that is detected by comparator circuits U501 and U502. The relevant comparator will then generate a
fault pulse that is passed to the CPU via U509, causing the gate drive to be inhibited immediately, thus
preventing inverter damage due to this fault.
Refer to 9H4404, sheet 2:
The inverter boards produce the high power drive for the high voltage transformers. The output of the
inverter boards drive the primaries of the high voltage transformers via the resonant / sharing inductors
and the resonant capacitor and a current transformer that samples the tank primary current.
The tank has similar anode and cathode sections. The cathode and anode sections each have their
own high voltage transformer and high voltage multiplier board. The anode section generates the
anode voltage, 0 to 75 kV. The cathode section generates the cathode voltage, 0 to -75 kV. The anode
and cathode sections contain voltage dividers that supply kV anode and cathode feedback voltages.
The kV feedback from the tank is brought to P42 on the main control board as described earlier in this
section.
54 2023-06-19 – 7K3254
clamp the voltage across the series resistors during high voltage arcs.
The grounded mid-secondary circuit in the tank is connected to the mA test jacks E1 / E2 on the tank
lid board. Transient protectors on the tank lid board clamp the voltage across the mA measuring device
during high voltage arcs, and prevent the voltage at E1 and E2 from rising above approximately 15 V
if the mA test jack is removed and no measuring device is connected.
The anode mA feedback that appears at P42-1 and P42-2 on the main control board is scaled
approximately 0.4 volts = 100 mA of anode current, and is applied to the input of differential amplifier
U406D.
The output of U406D is taken to comparator U505, which detects high anode mA. U505 will send a fault
pulse to the CPU if an anode over-current condition is detected, causing the kV output to be inhibited
immediately, thus preventing generator damage due to the over-current fault.
The output of U406D is also fed to the input of U406A, which provides a scaled mA feedback voltage.
The mA is calibrated by R488 such that 1 volt = 100 mA at the output of U406A. The output of U406A
is buffered by U404D. The output of U404D is taken to CPU via A/D converter. The output of U406D is
also to voltage amplifier U406C, which provides a better quality (higher gain) feedback signal at low
anode currents. The mA is calibrated by R497 such that 1 volt = 10 mA at the output of U406C. The
output of U406C is buffered by U406B. The output of U406B is taken to CPU via A/D converter. The
mA feedback information is used by the CPU to regulate the X-ray tube mA and to perform mA
monitoring functions during exposures.
The cathode mA feedback at P42-4 and P42-3 on the main control board is used for cathode over-
current detection only. The cathode mA feedback is fed to the input of differential amplifier U405D. The
output of U405D is taken to comparator U506, which detects high cathode mA. U506 will send a fault
pulse to the CPU if a cathode over-current condition is detected, inhibiting the X-ray exposure, thus
preventing generator damage due to the over-current fault.
console PC does not receive a response from the generator, a communication error message will be
displayed.
DS102 on the main control board will flash when data is being received by the main control board from
the console. DS103 on the main control board will flash to indicate that the main control board is
sending data to the console.
U105 on the main control board is a RS-232 transceiver, and converts the +3.3 V TTL levels from the
CPU via the isolate IC U112 to RS-232 levels for the serial port.
The main control board includes two serial communications ports. These include P11 and J10. J10 is
for normal operation and P11 is for debug.
J10 and P11 can't be used at the same time.
DS101 on the main control board will flash at a consistent 1 Hz rate if the related CPU is operational.
56 2023-06-19 – 7K3254
Section 5: Troubleshooting
Introduction
General Remarks
Warning
Please wear the appropriate antistatic tool at all times when handling the PCBs.
To investigate the possible cause of generator failure, the generator offers several diagnostic methods:
• Power on self test
• Generator log (error code)
• Service tool view function (View DC bus voltage, view tube calibration data,etc.)
• Test point on PCBs
• LED indicators on PCBs
The generator contains self-diagnostic routines which greatly facilitate troubleshooting. Self-diagnostic
functions require that DSP function correctly. The LEDs DS101 on MCB that indicate DSP correct
operation. When DSP is OK, the LED will be flashing at 1 Hz frequency.
When error occurred, an error message will pop up on the console, then the generator will change to
the error phase, and no exposure is available. In this case, you need to close the pop-up window to turn
generator to be in normal status. If the error continually pops up, problem needs to be fixed first before
using.
As a general rule, the first step in any troubleshooting procedure is to verify correct power supply
voltages and perform a visual inspection of all boards and cable connections.
Log access:
• Generator will send the log to console PC istead of storing any log on PCB. The log will be stored
in hard disk. A txt file will be created for every 50 M.
Tools
• Standard service engineers tool kit
• Digital multimeter with capacitor mode and diode mode
• Digital oscilloscope
• Antistatic kit
Strategy
There are two categories of problems:
1. The generator cannot be switched on at all, no error code is reported. Refer to 'Switch on problem'.
2. Error messages are displayed on console or log file. Refer to 'Error code analysis'.
7K3254 – 2023-06-19 57
Service Manual
Start-up Sequence
When S1 is closed, green LED DS1,DS2, and DS3 on FCB stay ON.
1. Power ON switch is pressed, power on circuit on MCB sends out 'ON' command, power regulator
circuits start to work and generates 12 V, 5 V, 3.3 V, and 1.8 V low DC voltages, which will supply
all PCBs.
2. MCB begins to work. DS301 (red) on MCB is ON, (filament fault indicator), DS511, DS512 (yellow)
on MCB are ON, (± 12 V indicator), DS101 (green) on MCB is flashing, (DSP heart beat indicator),
DS202, DS203 (green) on MCB are ON, indicate default small filament and high speed rotor is
selected.
3. MCB tests following signals: hand switch signal, tube thermal switch signal, rotor status, and K1
status.
Test OK.
4. Energy RCB K3, pre-charge DC bus, DS205 (green) on MCB is ON indicating k3 is energized. The
red LED DS1 on RCB becomes bright gradually and stays lit when at the maximum brightness.
5. Generator monitors the DC bus voltage.
DC bus voltage reaches preset range in 10 seconds.
6. Energize main contact K1 and turn on DS206 (green) on MCB, which indicates that main contactor
driver signal is present. Then MCB will verify the status of main contactor K1. K1 will be regarded
to be energized after DS603 being turned ON.
7. Verify tube file.
Test OK.
8. Apply voltage on filament. DS301 on MCB turn OFF, indicate filament current present, generator
start up finished and internally ready for exposure.
Switch On Problem
1. If failed to switch on generator by console, try to switch on by S501 on MCB.
2. If switched on failed by S501, run the following troubleshooting tree to identify the root cause of
failure.
58 2023-06-19 – 7K3254
3. If switched on successfully by S501, check the cable connection from J10 of MCB to console.
7K3254 – 2023-06-19 59
Service Manual
a. Tripped state
C D E
A B
60 2023-06-19 – 7K3254
DANGER: DANGEROUS VOLTAGE
After the generator has been switched OFF, hazardous voltages are still applied to the DC bus.
Measure the DC voltage on rectifier bridge DC side, make sure the voltage is safe.
i> Disconnect the cable connected to terminal A and terminal B.
ii>Check whether the short circuit happens between two input cable of the inverter (in red
circle). If short circuit detected, the inverter is damaged.
c. The procedure for checking the circuit breaker:
i>Confirm the circuit breaker can be switched to the OFF position. If cannot the break circuit is
damaged.
ii>Check the input and output of circuit breaker is short connected when ON, and is
disconnected when OFF. If not, the circuit breaker is damaged.
7K3254 – 2023-06-19 61
Service Manual
ER015 Beam_Tank_NC
62 2023-06-19 – 7K3254
ER036 Communication Error Console
7K3254 – 2023-06-19 63
Service Manual
ER108 Large filament set parameter is more than max. filament current
ER109 Small filament set parameter is more than max. filament current
ER110 Prep-parameter_kV_Error
ER111 Prep-parameter_mA_Error
ER112 Prep-parameter_ms_Error
ER113 Prep-parameter_RS_Error
ER114 Prep-parameter_Fo_Error
64 2023-06-19 – 7K3254
The data restored in EEPROM corrupted.
The EEPROM breaks down.
Action:
1. Erase the EEPROM.
a. Turn off the generator, then set the 8th bit of DIP switch S101 on MCB to ON.
b. Turn on the generator, and wait for one minute.
c. Turn off the generator, and set the 8th bit of DIP switch S101 on MCB to OFF.
2. Restart the generator. If ER003 still occurs, replace the MCB.
3. If ER003 does not occur, download the tube file and restore generator backup file.
Error Code: ER004
Description: Generator CPU Real Time Clock error
Appears when: Generator power on
Information/Symptom: The DSP cannot read data from RTC.
Possible cause:
The RTC device does not work.
Action:
Replace the MCB.
Error Code: ER005
Description: Main Contactor error
Error type: Latch error
Appears when: Generator power on
Information/Symptom: Main contactor is not contacting.
Possible cause:
Main contactor breaks down.
Main contactor control relay breaks down.
Action:
1. Cut off the power supply, and check whether the contactor button on main contactor K1 is
protuberant.
2. If the contactor button on main contactor K1 is not protuberant, replace the main contactor.
3. Power on the generator. Check the LED DS206 on the MCB.
4. If LED DS206 is off, replace the MCB.
7K3254 – 2023-06-19 65
Service Manual
66 2023-06-19 – 7K3254
Action:
7K3254 – 2023-06-19 67
Service Manual
b. Oscilloscope setting: voltage scale to 5 V/div, time base 8 ms/div. The normal waveform of TP109
and TP110 is shown as below, and the normal waveform of TP113 and TP114 is also the same.
The normal waveform of TP111 and TP112 is shown as below, and the normal waveform of TP115
and TP116 is also the same.
68 2023-06-19 – 7K3254
FCB failure
MCB failure
Tube failure
Action:
7K3254 – 2023-06-19 69
Service Manual
Note
No exposure is required. The filament is glowing during system standby.
> Remove the collimator.
> Check whether the filament is lighting when either small focus or large focus is selected. See the
picture below.
> Otherwise, the tube is damaged.
b. Oscilloscope setting: voltage scale to 2 V/div, time base 20 us/div. The normal waveform of TP2
is shown as below, and the normal waveform of TP4 is also the same.
70 2023-06-19 – 7K3254
Action:
1. Confirm the tube “seasoning” procedure is complete.
2. If not, the tube needs “seasoning,” then reset the error.
3. Else, check the arcing trace at the high voltage plug.
4. If yes, clean the plug, coat the silicon grease and connect the plug reliably.
5. If no, replace the tube if the tube arcing is suspected, or replace the tank if the tank arcing is
suspected, or replace the high voltage cable if the high voltage cable arcing is suspected.
Error Code: ER010
Description: Beam_Anode Fault
Appears when: During exposure
Information/Symptom: Excessive current in anode is detected.
Possible cause:
Tank arcing
Tube arcing
High voltage cable arcing
mA feedback circuit failure
Action:
1. Confirm the tube "seasoning" procedure is complete.
2. Otherwise, the tube needs “seasoning,” then reset the error.
3. Else, check the arcing trace at the high voltage plug.
4. If yes, clean the plug, coat the silicon grease and connect the plug reliably.
5. Otherwise, replace the tube if the tube arcing is suspected, or replace the tank if the tank arcing is
suspected, or replace the high voltage cable if the high voltage cable arcing is suspected.
Error Code: ER011
Description: Beam_INVA Fault
Appears when: After the exposure
Information/Symptom: Excessive current in the inverter module A is detected.
Possible cause:
The MOSFET in Inverter module A has boken down.
Action:
Replace the inverter.
Error Code: ER012
Description: Beam_INVB Fault
Appears when: During exposure
Information/Symptom: Excessive current in the inverter module B is detected
Possible cause:
The MOSFET in Inverter module B has broken down.
Action:
Replace the inverter.
Error Code: ER013
Description: Beam_kV Fault
Appears when: During exposure
Information/Symptom: Excessive kV output is detected.
7K3254 – 2023-06-19 71
Service Manual
Possible cause:
kV generation circuit defect.
kV feedback circuit defect.
Action:
1. Replace MCB
2. Replace the tank.
Error Code: ER014
Description: Beam_IR Fault
Appears when: During exposure
Information/Symptom: Excessive resonant current is detected.
Possible cause:
Tube is not rotating
Tube arcing
Tank arcing
Generator overload
High voltage cable arcing
Action:
1. If error happened during installation, verify the tube rotor cable connection and make sure tube
anode is rotating at proper speed.
2. Confirm the tube "seasoning" procedure is complete.
3. Otherwise, the tube needs "seasoning," then reset the error.
4. Otherwise, check the arcing trace at the high voltage plug.
5. If yes, clean the plug, coat the silicon grease and connect the plug reliably.
6. Otherwise, replace the tube or tank or the high voltage cable if it's suspected.
72 2023-06-19 – 7K3254
Action:
1. Check the waveform of the TP406 and TP407 when the exposure is at 150 kV/20 mA/32 ms.
Oscilloscope setting: CH1 is TP406, CH2 is TP407,voltage scale to 2 V/div, time base 10 ms/
div
The normal waveform is shown as below.
7K3254 – 2023-06-19 73
Service Manual
Action:
1. Avoid long time delay between pressing Prep hand switch and Exposure hand switch.
2. If error still happens, replace the hand switch if hand switch is damaged or replace the
communication cable if communication cable (see below table) is damaged.
Error Code: ER019
Description: Hand switch is not released in 30 seconds after exposure
Appears when: After exposure
Information/Symptom: Hand switch is not released in 30 seconds after exposure.
Possible cause:
Incorrect operation
Hand switch failure
Communication cable failure
Action:
1. Release the hand switch when exposure is finished.
2. If error still happens, replace the hand switch if hand switch is damaged or replace the
communication cable if communication cable is damaged.
74 2023-06-19 – 7K3254
Action:
1. Check the waveform of TP408 and TP409 on MCB, the signal's frequency varies with the exposure
parameters. Oscilloscope setting: voltage scale to 5 V/div, time base 5 ms/div.
7K3254 – 2023-06-19 75
Service Manual
Replace tank.
Replace tube
Escalate to technical support
Error Code: ER022
Description: mA too low
Appears when: During exposure
Information/Symptom: The mA feedback signal is 20 % smaller than setting.
Possible cause:
mA short jump plate on E1 and E2 of tank lid missing
HT cables connected in wrong polarity
Calibration data not correct
Tube aging
FCB failure
Tank failure
Action:
Check mA short jump plate on tank lid
Verify HT cables are connected in right polarity
Measure mA feedback, compare with mA setting.
Re-calibrate the tube.
Replace FCB.
Replace tank.
Escalate to technical support
76 2023-06-19 – 7K3254
3. Check TP8 and TP10 on AEC board. Oscilloscope setting: TP10's voltage scale 5 V/div, TP8's
voltage scale 2 V/div, time base depends on settings.
7K3254 – 2023-06-19 77
Service Manual
2. Check cable connection from tube rotor to J4 on RIB, check the pin1 and pin3 of J4 on RIB after
powering off the generator.
3. If open, wait for tube to cool down or replace the TUBE.
4. If it's short circuit, replace the RIB.
78 2023-06-19 – 7K3254
Possible cause:
Door open
RIB failure
Action:
1. Close the door.
2. If error still happens, check the cable connected on the J3 on RIB.
3. If the cable connection is OK, check the pin1 and pin3 of J4 on RIB after powering off the generator
and closing the door.
4. If open, replace the door switch.
5. Otherwise, replace the RIB.
3. If the error still happens, measure the resistance between the pin7 and pin8 of J1 on RIB after
powering off the generator.
4. If it's open circuit, RIB is good, fix the issue of Bucky Ready signal source.
5. Otherwise, replace the RIB.
7K3254 – 2023-06-19 79
Service Manual
Action:
1. Verify the cable connection to Room Interface Board – J2.
2. Verify the Input signal setting at generator Service Tool > Station Setting > Input.
3. If the error still happens, measure the resistance between the pin7 and pin8 of J2 on RIB after
power off the generator.
4. If it's open circuit, RIB is good, fix the issue of Bucky Ready signal source.
5. Otherwise, replace the RIB.
80 2023-06-19 – 7K3254
Communication cable connection definition Signal
MCB_J10-1 Control box-GEN-6 PCU_TXD
MCB_J10-2 Control box-GEN-1 PCU_RXD
MCB_J10-3 Control box-GEN-2 COM
MCB_J10-4 Control box-GEN-7 PREP
MCB_J10-5 Control box-GEN-8 EXP
MCB_J10-6 Control box-GEN-3 PWR_ON
MCB_J10-7 Control box-GEN-4 PWR_OFF
MCB_J10-8 Control box-GEN-5 24 V
7K3254 – 2023-06-19 81
Service Manual
7K3254 – 2023-06-19 83
Service Manual
84 2023-06-19 – 7K3254
Information/Symptom: No tube file
Possible cause:
Tube file is not downloaded
Action:
Download the tube file.
7K3254 – 2023-06-19 85
Service Manual
Caution
High voltage on DC bus.
1. Check DIP switch S101-6 S101-7 if the setting meet actual power supply.
2. Check the V (dc) on rectifier bridge DC side.
3. If measured DC bus voltage is out of range, check the input power supply.
4. Check DC-bus voltage at Service Tool->Tool->View DC bus voltage.
7K3254 – 2023-06-19 87
Service Manual
5. If voltage in view DC bus voltage window is same with measured value, replace the MCB.
6. Otherwise, replace the RCB.
88 2023-06-19 – 7K3254
Action:
1. Verify tube setting and shift capacitor is correct.
2. Measure the resistance between tube stator connectors.
3. If measured value is not match with tube specification.
4. Otherwise, replace RCB.
7K3254 – 2023-06-19 89
Service Manual
7K3254 – 2023-06-19 91
Service Manual
92 2023-06-19 – 7K3254
Error Code: ER115
Description: Dual energy no calibration curve
Appears when: At DE exposure mode
Information/Symptom: Dual energy is not calibrated
Possible cause:
Dual energy is not calibrated
Action:
Calibrate dual energy exposure.
7K3254 – 2023-06-19 93
Service Manual
94 2023-06-19 – 7K3254
Error Code: ER206
Description: Generator mAs Limit
Error type: Warning
Appears when: Parameter setting
Information/Symptom: Requesting mAs is not allowed as the generator mAs limit has been reached
Possible cause:
Incorrect setting
Action:
Check the mAs.
7K3254 – 2023-06-19 95
Service Manual
Possible cause:
Incorrect setting
Action:
Reset the error.
96 2023-06-19 – 7K3254
Error Code: Console Error
Description: Fail to get the Generator tube statistics, reason: False
Error type: Warning
Appears when: At any moment
Information/Symptom: N/A
Possible cause:
Communication error
Action:
Reset the error.
LED Indicators
The following table describes the normal status and definition of the status and diagnostic LEDs on the
printed circuit boards in the generator. The definition of LED is described when LED is ON, unless
specific explanation is present.
7K3254 – 2023-06-19 97
Service Manual
DS101 Green Indicates DSP is functioning when flashing DSP in normal mode: Flashing in
1Hz
DSP in bootloader mode:
Flashing 2 times, delay for 1
second and repeat
DS102 Green Indicates data is transmitted when flashing Flashing when transmitting data
DS103 Green Indicates data is received when flashing Flashing when receiving data
DS201 Green Indicates the presence of large focus On when large focus selected
DS202 Green Indicates the presence of small focus On when small focus selected
DS203 Green Indicates the presence of high speed for anode On when rotor HS selected
rotor
DS204 Green Indicates the presence of low speed for anode On when rotor is selected
rotor
DS301 Red Indicates the filament current too low error* OFF
98 2023-06-19 – 7K3254
DS602 Yellow Indicates the anode rotor is running OFF
*When generator is in error phase, filament voltage will be inhibited to protect the filament. So it's
normal that DS301 is ON when generator is in error phase.
Potentiometers
Fuses
7K3254 – 2023-06-19 99
Service Manual
Test Point
[]
What-to-do Table
MCB
Prerequisites
Backup the generator settings (if available).
To remove
1. Disconnect and lock out the main power.
Important
If emergency off connector P54 is short connected, don't forget to move short connector to new MCB.
Without P54 pin 1 and pin 3 short connected, generator can't be powered up.
To install
Important
Before installing the new MCB, you must set the dip switch and jumpers to match the old board.
1. On the new MCB, check the correct settings for dip switch S101, S102 and jumper JW101.
- S101dip switch setting:
FCB
Prerequisites
None.
To remove
1. Disconnect and lock out the main power.
To install
• Reverse the steps in the removal process to install the new FCB.
• Power on the system and make an exposure to verify generator function.
Prerequisite:
None.
To install
• Reverse the steps in the removal process to install new RCB and safety cover.
• Power on the system and make an exposure to verify generator function.
Prerequisite:
None.
Inverter-Assembly
Prerequisites
None.
To remove
1. Disconnect and lock out the main power.
To install
1. Reverse the steps in the removal process to install new AEC board.
2. Verify that the jumper setting is the same with old AEC board.
3. Power on the system and check the AEC functions.
Prerequisite:
None.
To install
1. Reverse the steps in the removal process to install new tank.
2. Power on the generator.
3. Perform kV adjustment, low mA adjustment and high mA adjustment. See my original file
4. Calibrate the tube.
Prerequisite:
None.
Other Parts
Adjustment
Adjustment Specification
Purpose: To provide accurate output of kV and mA
Specification: Three pots which are adjusted in the procedure are respectively used to control the
output accuracy of kV, H_mA and L_mA.
Tools required to carry out this adjustment procedure.
1. UNFORS dose meter
2. mAs meter
3. Potentiometer adjustment tool
3. Open Tools->calibration mA, and set filament current to 3.3. Press Enter after keying in the
value.
4. Make an exposure, and read the kV value from the dose meter. If it is in the range of 99-101 kV,
there is no adjustment needed.
5. If the kV value is out of the range, adjust potentiometer R444 (See Figure: Potentiometer
Layout).Turn the potentiometer clockwise if value >101 kV, or turn it clockwise if value< 99 kV.
6. Repeat steps 4 and 5, until KV value from the dose meter is in the range of 99-101 KV.
Note
The whole range of potentiometer is five turns..
Low mA Adjustment
1. Power off the generator.
2. Remove the link between E1 and E2 connectors on the top of Tank. Serial connect two maximum
meter probes to the plug connectors of E1,E2 to measure mA.
3. Power on the generator. Open the service tool, and set the exposure technique to mA/ms mode,
select small focus, the exposure parameters at 100 kV,100ms. Any mA value is acceptable, it will
not impact on adjustment result.
4. Open Tools->calibration mA->set filament current to 3.45, and press Enter after key in the
value.
5. Make an exposure. Read and record the mAs value from mAs meter. mA value can be calculated
from divide mAs value by 0.1, If 15< calculated mA< 25, Record this value for later use,go to step
8, otherwise go to step 6.
Note
It’s recommend to adjust the filament current by 0.01 or 0.02. The maximum step to adjust filament
current is 0.1. DO NOT try to make a large step.
7. Repeat steps 5 and 6 until calculated mA is within 15-25. Record this value for later use.
8. Read and record the mA feedback from calibration mA form window. Compare the mA value with
the calculated mA value if difference <=1 mA, no adjustment is needed.
9. If out of the range, adjust potentiometer R497 (See Figure: Potentiometer Layout) and make
exposures. If feedback mA < calculated mA, turn R497 counter-clockwise, otherwise turn it
clockwise.
Note
The whole range of potentiometer is five turns.
10. Repeat steps 8 and 9 until feedback mA is in the range.
High mA adjustment
Note
It’s recommend to adjust the filament current by 0.01 or 0.02. If low mA adjustment is just finished
before this adjustment, go to step 4.
1. Remove the link between E1 and E2 connections on the top of Tank. Connect the Mas meter to the
plug connections of E2 to measure Ma.
2. Power on the generator. Open service tool, and set the exposure technique to mA/ms mode, select
small focus, the exposure parameters at 100 KV,100ms. Any mA value is acceptable, it will not
impact on adjustment result.
3. Open Tools -> calibration mA -> set filament current to 4.33, and press Enter after key in the
value.
4. Make an exposure. Read and record the mAs value from mAs meter. mA value can be calculated
from divide mAs value by 0.1. If 195 < calculated mA < 205, Record this value for later use, go to
steps 8, otherwise go to steps 6.
5. Adjust the filament current value in the calibration mA window.
Note
The maximum step to adjust filament current is 0.1. DO NOT try to make a large step.
6. Repeat steps 5 and 6 until calculated mA is within 195-205. Record this value for later use.
7. Read and record the mA feedback from calibration mA form window. Compare the mA value with
the calculated mA value if difference <= 5 mA, no adjustment is needed.
8. If out of the range, adjust potentiometer R488 and make exposures. If feedback mA < calculated
mA, turn R488 counter-clockwise, otherwise turn it clockwise.
Note
The whole range of potentiometer is five turns.
9. Repeat steps 8 and 9 until feedback mA is in the range.
10. Remove mAs meter and reconnect the link between connectors E1 and E2 on the tank lid.
11. Make an exposure to verify that generator is operational.
5. Check the Port Name. If the Port Name is incorrect, type in the correct Port Name.
6. Click the UpLoad button, generator will restart automatically. If upgrade failed, click the UpLoad
button again.
7. When firmware upgrade is successfully finished, click Close button. Generator will restart
automatically.
3. Select File > Backup Setting and click Backup All to choose a destination for generator backup
file.
Note
The generator data will normally be remain, except instructed, don’t need to restore the data back after
firmware upgrade. The backup file is only used in case the generator data abnormal or crash happened.
4. Select Utility > Upgrade Firmware to open upgrade firmware window.
Note
During firmware upgrade process an Error: ‘communication is disconnected’ will be shown on top of
Service Tool and Service Tool GUI will grey out. This is normal. No action is required.
7. The message ‘Firmware upgraded successfully’ indicates upgrade process complete. Close the
Upgrade Firmware window.
8. When firmware upgrade is successfully finished, click Close button. The generator will restart
automatically.
This section provides a recommended schedule for periodic maintenance of the X-ray generators. The
initial installation date and location, and all service performed on the generator, should be recorded in
the table below. The record should be as thorough as possible, detailing the scope and type of work
that was performed (all service and a record of all replacement parts that were installed). Additionally,
the person performing the work should date and sign the record. This information will be invaluable in
the future for traceability and to ensure continued compatibility of the generator.
If a major component (such as the HT oil tank or a main control board) is replaced, recalibration will be
needed. A separate procedure will be included with the spare in those cases, detailing the required
calibration procedure. The acceptance test procedure should then be performed prior to placing the
generator back into service.
Caution
Maintenance is to be performed only by competent, trained personnel who are familiar with the potential
hazards associated with this equipment.
Note
Maintenance schedule frequency may be dictated by certain regulatory requirements of the country or
state in which the installation is located. Always check the local codes and regulations when setting the
maintenance schedule.
Equipment Maintenance
Caution
Do not attempt mechanical or electrical repair of the System. Contact your authorized Service Provider
if any unit does not perform to your satisfaction.
The System must be maintained in good operating order at all times to provide safe conditions for
operating personnel and patients. This section provides a recommended schedule for periodic
maintenance of the generator.
Semi-annually and Whenever a Related Certifiable X-ray Component is Replaced:
1. Check that all HV connections are tight.
2. Clean the main cabinet as needed.
3. Perform the X-ray tube auto calibration at the initial 6 months, and then every 12 months thereafter.
4. Perform any additional tests required by laws governing this installation.
Annually:
1. Pull out HT cable connector from tank and tube side, check if any sign of arcing exist. Clean all HV
connections and silicon washers using alcohol.
2. Verify the KV, mA accuracy of the generator according to Section 6 Adjustment. Adjust the
potentiometers if needed.
3. Open the generator main cabinet and examine the unit for any visible damage: missing or loose
ground connections, oil leaks, damaged cables, etc.
Cleaning
Caution
Before using any disinfectant, please obtain SDS (Safety Data Sheet) or AIS(Article information sheet)
and familiarize with it.
• Never use anything other than mild soap and water to clean plastic surfaces. Other cleaners may
damage the plastic.
• Never use any corrosive, solvent or abrasive detergents or polishes.
• Ensure that no water or other liquid can enter any equipment. This precaution prevents short
circuits and corrosion forming on components.
• Methods of disinfection used must conform to legal regulations and guidelines regarding
disinfection and explosion protection.
• If disinfectants are used which form explosive mixtures of gases, these gases must have dissipated
before switching on the equipment again.
• Disinfection by spraying is not recommended because the disinfectant may enter the X-ray
equipment.
• If room disinfection is done with an atomizer, it is recommended that the equipment be switched
OFF, allowed to cool down, and covered with a plastic sheet. When the disinfectant mist has
subsided, the plastic sheet may be removed and the equipment be disinfected by wiping.
②
①
③
⑤ ⑥
⑧ ⑦
①
②
⑥
ŀ
①
②
③ ⑤
⑧ ⑨
Figure 4 Cover
Numerical Index
Index
The following functional schematics and PCB key point index are included in this manual.
1 2 3 4 5 6 7 8
Tube
E E
AC MAIN 1L1 S1 2T1 3 LF1 3 1L1 K1 2T1 C D1 B E2 E3 Cathode Anode
INPUT
IN1
380/400 3L2 4T2 2 2 3L2 4T2 D Inverter A
/480 VAC EMC Main 3 Phase
3 PHASE 5L3 6T3 1 Filter 1 5L3 Contactor 6T3 E Bridge A E1 E4
P1 P2
IN2
HV
TANK Ion
D T1-*V Chamber D
T1
Auxilary
Transformer
T1-0V
P1 P2
E2 E3
P1
Inverter B
E1 E4
C C
Emergency
P54 Stop
Inverter P9
Temp P1 P71 Room
P61 P3 Interface
P8
Board
P73 J4
A Generator AEC A
P1 Board J1/J2
1 2 3 4 5 6 7 8
E1+
S1 LF1 K1 D2 INVERTER A
E2-
1L1 2T1 2T1
AC MAIN 3 3
1L1 C
INPUT 3L2 4T2 EMC 3L2 4T2 A
2 2 D E1+
380/400 5L3 Filter 6T3
6T3 1 1 5L3 INVERTER B
/480 VAC E E2-
E 3 PHASE
B E
83 84
P6-4 P10-3 P10-1 TP225
PE P6-3 DC BUS + DC BUS +
A1 A2 MAIN FL1 ˄To 9H4407,5D) P1-1
P6-2CONTACTOR F1
CONTROL(From P2-2 DC bus detection
F1 F2
P6-1 9H4401,1B) +12VAA
-12VAA P2-1
P2-3 DS1
K4 P1-2 } (To 9H4401,2B)
+24V P2-4 F2
P7-1 F3 K3 K4 DC BUS -
DC bus +12VAA
discharge (To 9H4407,6B) TP226
K4 ROTOR CONTROL BOARD DC BUS -
P7-2 F4
+12VAA
D K3 P52-2 +12VAA +12VAA P73-13 D
PRE-CHARGE P52-1 -12VAA
DRIVER(From 9H4401,2C) P52-3 P1-13
+24V -12VAA
P52-4 -12VAA P73-11
P1-11
+12V
P73-12,14,17 P1,14,17
+24V +24V
P2-1 P1-5 +24V
TP516 +12V
ROOM P2-3 P1-7 +12VAA +12VAA
INTERFACE P1-2 P53-4 AEC BOARD
BOARD +12V SWITCHING
+24V +24V ON/OFF REGULATOR CIRCUIT
P5-1 P1-1 P1-5,6 P1-4 P53-2
COMMAND `
(U511,U510 AND
P5-3 P1-3
(From 9H4401,7E) CR521 ASSOCIATED
C P1-1 P53-3 COMPONENTS) C
FILAMENT CONTROL BOARD
P1-3 P53-1
TP13
+35VDC +12VAA TP504 +24V
P5-1 FH3 FH5 +5V TP522 +12VAA
25V CR6 VR501
C8 DS3
TP12
220V -12V SWITCHING
+24V
REGULATOR CIRCUITķ Note:
110V (VR504 AND 1. (6A) -12V switching regulator
TP514
+1.8V ASSOCIATED
+5V VR503 COMPONENTS)
circuit is switched on by +12VAA
0V
+1.8V
REGULATOR
T1
A A
MAIN CONTROL BOARD
System ON (9H4401)
9B,62 9
9
21 8
( - 4 (
4
212))
2)) - &200$1'
7R+&
4
- 4
,62/$7,21
*1'B,62
*1'B,62 4
6
21
' '
6
2))
9
(0(5*(1&<6723
1RWH
3
① & .LVRSHQZKHQ9
3 . SUHVHQWV\VWHPFDQQRWSRZHURQ
ZKHQ.FORVH
8
35(&+$5*( ,2
& '5,9(5 3 3 '5,9(5 &
7R+'
3 3
8'
9$$ 9
% 3 %
9 9$
0$,1 5 5
$ 8
&217$&725 3
,2
'5,9(5
7R+( .
3
3 3
8
3 3 ,2
'5,9(5
52725&21752/%2$5'
0$,1&21752/%2$5' &38
$ $
5220,17(5)$&(%2$5' )
- 3
9$&287 `
IRUFROOLPDWRU - 3
⑧ (
9$& 3
)URP$X[LOLDU\
FRQWURO
( 9$& 3 WUDQVIRUPHU
( - ( (
%XFN\UHWXUQ 3
%XFN\RXW -
%8&.<3:5 ) ⑨ ( 9$&
(
- . 9
%XFN\6WDUW 9 ) ),/$0(17&21752/
3 3
- ( %2$5'
%XFN\5'< 9
9 3 3
- (
%XFN\*1'
8
① 5
73
3 3 ,2
8
,2
5 3 3
9 4
' 9 9 '
%XFN\UHWXUQ - 8
%XFN\RXW - ⑤
. 73 5
- .
%XFN\6WDUW
8
- 3 3 ,2
%XFN\5'<
- 5
8
② '5,9(5
%XFN\*1' 73
3 3 ,2
9
8
. 9 9
-
&
5220/,*+7
`
- 8
⑥ &
73 5
.
- 5
'225
,17(5/2&.
`
- 9
③ 3 3
8
'5,9(5
,2
73
8 ,2
3 3
9
9 9
78%( - 8
7+(50$/ `
- 5
⑦ 5
6:,7&+ 73 5
.
9 ④ 8
7202 - 73 3 3 ,2
% 8 '5,9(5 %
(;32685(
- 3 3 ,2
,1387
3 3 ,2
9
3 3
- 8
9'& 3 3
` 5
2XW - 0$,1&21752/%2$5'
&38
( (
% %
$ $
1 2 3 4 5 6 7 8
+12VAA
R561 ķ
E E
P42-9 +3.3V TP303
P1-G
DRIVER ENABLE
P1-H P42-5 (To 9H4404,7B)
R562 R563
U318
I/O X-Ray Enable
(From 9h4403,3C) Q317
H.T. TANK
D D
FILFAULT R352
Q312
+3.3V +3.3V (From 9H4405,2B) +5V
+5V R357
DS105 DS104
+5V DS302
R356
R133 R131
U112
R354 U314B
KV CHARGE
P11-8 11 6 I/O (To 9H4403,6B
C 9H4404,3B) C
P11-7 12 5 X-Ray Enable
R355
(To 9H4403,6E) Q314
P11-5 I/O R353
Q313
ISOLATION
KV CHARGE
GND_ISO GND_ISO CLR
(From 9H4403,6C)
Latch
I/O PT STOP
IN
EXP (From 9H4408,1B)
X-Ray
J10-5
Prepare
PREP U316B
B J10-4 B
R364
D /Q
PR Q315
HANDSWITCH
GND_ISO
I/O
Note:
ķ:”HIGH” (APPROXIMATE +5DC) = GATE DRIVE PULSES ENABLED.
A MAIN CONTROL BOARD CPU “LOW”(APPROXIMATELY 0V DC) = GATE DRIVE PULSE NOT A
ENABLED DUE TO A FAULT OR NO VALID EXPOSURE REQUEST
P/N Name Version Page
9H4403 X-Ray Exposure - radiographic A 1 of 1
1 2 3 4 5 6 7 8
,QYHUW³6+227 ,QYHUW³6+227
7+58´6(16( 7+58´6(16(
9 9$$ )URP+3% )URP+3(
5 &5
& 8& 8$ &
5 8$ 8' (5525$03/,),(56 5
&5 5 8$ 9&2
8' &5 5
,1&/8'(6
8$
5 8888 8%
8& 84
5
.9&+$5*(
6RIW6WDUW '5,9(5(1$%/(
)URP+& )URP+(
'
9$$
5 8
5
5 7 7
5 5
8
% 8 %
,2 %($0B,5
'DWDODWFK
8 5
8
&38
5 5 0$,1&21752/%2$5'
$ $
1 2 3 4 5 6 7 8
+12VAA +12VAA U2
P71-3 P1-3 P42-8
P1-2 P71-2 Temperature KV FEEDBACK P42-7
ADC Sensor SIGNAL
P1-1 P71-1 (To 9H4404 P1,8E) P42-5 J1
P42-6 ANODE
INVERTER TEMPERATURE BOARD
E E
C S
T2 P2-1
MOSFET INVERTER P51-1 L
(INCLUDES Q1-Q16) E1:DC BUS+ Invert “SHOOT
THRU” SENSE
P2-3 (To 9H4404 P1,4E) *
P51-3
HV ANODE
P41-1 P1-1 BOARD ANODE
P41-2 SAMPLE TUBE
P1-2 BOARD
CPU
E3 P1-L
D D
P1-U
P41-3 P1-3
P1-H
P41-4 P1-4
P1-T
E4 INV1
INV2
E2: DC BUS -
INVERTER BOARD B
TANK LID
BOARD
C C
T2 P2-1 P50-1
MOSFET INVERTER
E1:DC BUS +
(INCLUDES Q1-Q16) Invert “SHOOT
THRU” SENSE
P2-3 (To 9H4404 P1,4E)
P50-3
P40-1 P1-1
B P40-2 P1-2 * B
E3 CATHODE
SAMPLE
BOARD
HV CATHODE
P40-3 P1-3 BOARD
P40-4 P1-4
J2
E4
HV TANK CATHODE
P32-1,2
L
C
P32-3,4
E2:DC BUS - S
Resonant current sensor
A (To 9H4404 P1,6A) A
MAIN CONTROL INVERTER BOARD A
BOARD
0$;,080),/$0(17&855(17
/,0,7&,5&8,7
0$;$
( 73 (
5 5 9$$
9 $ 5
),/B,B5()
8' 3
3
8
5 5 &5
'$& 3 3
'$& 8& 5
8$ 3 3
5 8
9'& 9'&
5 5
73
' 9 $
3 3 '
),/B,B)(('%$&. 8& 3 3' /
3:0 3 3 78%(
5(*8/$725 &$7+2'(
5
3 3(
/
5
&
. 3&
3 3 3
8& 4 4 6
$'&
5
3 35
5
6
9'& 9'& +97$1.
8
&
&
8%
506 3 3 7 &
&219(57(5
9 8$ 5
5
3 3
'6
9
5 5()
,2
8'
4
9
),/)$8/7
% %
7R+'
.
8
,2 ),/B6(/(&7 3 3
'5,9(5
),/$0(17&21752/%2$5'
$ &38 $
0$,1&21752/%2$5'
73
+B0$B)%
9 0$
( (
5
5
+B0$B)% $'&
$12'(08/7,%2$5' 5 5
- 8'
8$
$12'( 5 5
6 5 5
& 73
/ /B0$B)%
9 0$
5 5
30 3 5 5 5
5 5 /B0$B)% $'&
8'
5
' 8% '
5 8&
31 3 5 5 5
5 5
5
78%( 9$$
P$7(67 5 5
-$&.
( 8
8
B 5 %($0B$12'(
'DWDODWFK
,2
(
& &
9$$
5 5
5
3. 3 5 5
- 8
&$7+2'( 8' 8 %($0B&$7+2'(
5 ,2
5 'DWDODWFK
/ & 5 5
3- 3
6 5
% %
&38
&$7+2'(08/7,
%2$5'
7$1./,'%2$5'
0$,1&21752/%2$5'
+97$1.
$ $
( (
6KLIW&XUUHQW
3 3 6HQVRU
3 3
0DLQ&XUUHQW
6HQVRU
,19(57(5
&855(17
8
'(7(&725
9
5
' '&%86 /6&$3 '
3 /&
)URP+( . /&
'6
3 +& +&
5RWRURYHUFXUUHQW 3 3
8 3 +6&$3 +&
,*%76:,7&+ ,*%76:,7&+
4 4
,2 5272567$7(
-
,2 35(3 - . 3 3 6+,)7 3
& /RJLF&LUFXLWDQG
,VRODWLRQ &
,*%7 7UDQVIRUPHU
)DXOW&LUFXLW 3 0$,1 3
'5,9(5 'ULYHU&LUFXLW 3
,2 3:0 88
77 3 3
3 &20
,2 3:0
,*%76:,7&+ ,*%76:,7&+
4 4
5220
,17(5)$&( ;UD\7XEH5RWRU
%2$5'
9
% '&%86 %
)URP+'
9$$
.
8
,2 3 3
'5,9(5
.
8
,2 3 3
'5,9(5
52725&21752/%2$5'
&38 0DLQ&RQWURO%RDUG
$ $
( (
$ $
8
,2 3
( &+$0%(56(/(&7 (
,2 3
&+$0%(56(/(&7
,2 3
&+$0%(56(/(&7
,2 3
&+$0%(56(/(&7
'5,9(5
,2 3
67$57
' '
3
,2
5,*+7),(/'6(/(&7
3
,2
0,''/(),(/'6(/(&7
,2 3
/()7),(/'6(/(&7
8
& &
,2
3
376723
3
$(&B)%B&/5B'59
,2
4
&38
$ 0$,1&21752/%2$5' $
( (
9 9
5 5
'6 '6
8
,62/$7,21
3
9
5
& &
'6
+($57B%($7
-
7['
&38
-
5['
% %
3
$ $
( (
7XEH
' '
,QWHUIDFH%R[ 7$1.
& &
3URWHFWLYHHDUWK
FHQWHU
6LJQDOJURXQGLQJRQO\
+9*HQHUDWRU
% %
,RQ&KDPEHU $(&%RDUG
$ $
Input Output
Bucky 1 ready Bucky 1 start
Bucky 2 ready Bucky 2 start
Extern sync
Input Output
Enable signal Start signal
Trigger signal
Enable signal: When enable signal is send from system to generator, generator is notified to detect
trigger signal. The trigger signal is send to generator by system, when exposure is allowed. Enable
signal is a low active signal.
Trigger signal: The system send this signal to generator to allow exposure when enable signal is
present. The trigger signal can be configured to raise edge or fall edge. If it is configured as raise edge,
the generator will start to exposure when the trigger signal change from low level to high level. If it is
configured as fall edge, the generator will start to exposure when the trigger signal change from high
level to low level.
Configuration
Select the enable signal and start signal in station settings, set the time out limit refer to the timing
figure.
7K3254 152
Service Manual
Note
Don`t select enable signal and trigger signal in same input source.
Note
Make sure tube is calibrated before performing dual energy calibration.
1. Open Service Tool, Tool-> Dual energy calibration, calibration window will pop up.
2. Select first exposure curve: 60 kV 560 mA 630 ms, don't revise the current value, click save
calibration.
3. Don't close the calibration window, make an exposure.
4. Select second exposure curve: 60 kV 20 mA 630 ms, click save calibration.
5. Make an exposure, wait for 5 seconds, close calibration window if not error pop up. Calibration
finished.
6. Calibration finished.
7K3254 154
Service Manual
Manual History