V6 - XV6 Service Manual - E - v1.04 - 20240109

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 305

V6/XV6

Service Manual
English

Copyright © 2023 by SAMSUNG MEDISON CO., LTD. All rights reserved.


Safety Requirements

■ Classification

– Type of protection against electric shocks: Class I

– Degree of protection against electric shocks (when the patient is in physical contact): Type
BF or type CF mounting

– Degree of protection against the ingress of harmful liquids: General equipment

– Degree of safety of use in the presence of flammable anesthetic agent mixed with air,
oxygen, or nitrous oxide: Not suitable for use near flammable anesthetic agent mixed with
air, oxygen, or nitrous oxide
– Mode of operation: Continuous operation

■ Safety standards the device conforms to


– Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and
Essential Performance [IEC 60601-1:2005/A1:2012]
– Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and
Essential Performance - Collateral Standard: Electromagnetic Compatibility –
Requirements and Tests [IEC 60601-1-2:2007]
– Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and
Essential Performance - Collateral Standard: Usability [IEC 60601-1-6:2010]
– Medical Electrical Equipment - Part 2-37: Particular Requirements for the Basic Safety
and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment
[IEC 60601-2-37:2007]
– Medical Electrical Equipment - Part 1: General Requirements for Safety [IEC 60601-
1:1988, A1:1991, A2:1995]
– Medical Electrical Equipment - Part 1-1: General Requirements for Safety – Collateral
Standards: General Requirements for Medical Electrical Systems [IEC 60601-1-1:2000]
– Medical Electrical Equipment - Part 1-2: General Requirements for Safety – Collateral
Standards: Electromagnetic Compatibility – Requirements and Tests [IEC 60601-1-2:2001,
A1:2004]
– Medical Electrical Equipment - Part 1-4: General Requirements for Safety – Collateral
Standards: Programmable Electrical Medical Systems [IEC 60601-1-4:1996, A1:1999]
– Medical Electrical Equipment - Part 2-37: Particular Requirements for the Basic Safety
and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment
[IEC 60601-2-37:2001, A1:2004, A2:2005]
– Medical Devices – Application of Risk Management to Medical Devices [ISO 14971:2007]
– Medical Electrical Equipment - Part 1: General Requirements for Safety [UL 60601-
1:2003]
– Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and
Essential Performance[CAN/CSA C22.2 No. 60601-1:14]

– Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and
Essential Performance [ANSI /AAMI ES60601-1:2005/(R)2012, AND C1:2009 AND
A2:2010(R)2012]
– Biological Evaluation of Medical Devices – Part 1: evaluation and testing [ISO 10993-1:
2009]
– Standard Means for Reporting the Acoustic Output of Medical Diagnostic Ultrasonic
Equipment [IEC 61157:2007]

 Statements

Nemko-CCL mark with the indicators “C” and “US” means that
the product is certified for both the U.S. and Canadian markets,
to the applicable U.S. and Canadian standards.

This mark certifies that the product conforms to applicable EEC


standards and has been certified by the European certification
agency.

This mark certifies that the product conforms to applicable EEC


standards.

The GMP symbol indicates the Good Manufacturing Practice in


accordance with the Korean quality management regulations.
Precaution for Use
Be sure to read this Service Manual thoroughly to familiarize yourself with the operation of the
product and the relevant safety information before attempting to use the product.

• Keep this service manual near the product and refer to it when using the product.

• Please familiarize yourself with the safety precautions in “Chapter 2. Safety” and “Chapter 10.
Maintenance” in particular.

• This service manual does not include diagnosis results or opinions. Also, check the reference
information for the measured area of the body before using the application’s measurement
results in any diagnosis.

• This product is an ultrasound diagnosis device and cannot be used from the user’s PC. The
manufacturer is not responsible for any problems that may be caused by such attempts.

• This product must only be used by persons who have sufficient knowledge of and/or a
qualification in clinical pathology. Unqualified persons are prohibited from using the product.

• The manufacturer is not responsible for any damage to this product caused by user
carelessness and/or neglect.

• Product orders are based on individually agreed specifications and may not include all
functions specified in this service manual.

• Some functions or options, probes, and the like may not be used in certain countries.

• All reference material on standards, regulations, and related revisions are valid at the time of
the publication of this service manual.

• Screen images in this service manual are examples and may differ from the actual screen or
system.

• The content of this service manual is subject to change without prior notice.

• Products that are not manufactured by Samsung Medison are indicated with the trademarks
of their respective owners.

• The following terms are used to highlight precautions that the user must pay particular
attention to:
Disregarding this instruction may result in death, serious injury, or other dangerous
situations.

DANGER

Follow these instructions to prevent a serious accident or damage to property.

WARNING

Follow these instructions to prevent minor accidents and damage to property.

CAUTION

The accompanying information covers installation, operation, and maintenance


procedures that require careful attention from the user, but have little chance of leading
directly to a dangerous situation.
NOTE
Revision History
The revision history of this service manual is as follows:

VERSION DATE REASON FOR CHANGE

1.04.00 2023.07.28 Initial Release

If You Need Assistance


If you need a service manual or any assistance with the product, please contact the Samsung
Medison customer service department or your local vendor.
Patient Privacy Policy

■ Treatment of Patients’ Personal Information

One of Samsung Medison’s social responsibilities is to recognize the importance of patient


privacy and to handle and protect the personal information provided by our patients in an
appropriate manner.

Patient information stored in repaired and demo products should be handled as follows:

• Patient Information: Refers to information, such as IDs, names, DOBs, and images, which
can be used to identify specific patients.

• Patient Information Management: Personal information provided by patients must be


managed carefully and solely for the purposes of repairs and demonstrations of Samsung
Medison products, and they must not be used for unauthorized purposes, be lost, be revised,
or be shared with a third party.

■ Management of Patient Information on Demo and Trade Products

• Patient information stored on a product during the course of its use must be handled as the
patient desires. (delete or backup)

• When completing a demonstration and/or collecting a trade product, patient information must
be deleted if so requested by the client, to prevent its unintended distribution to outside
parties.

■ Management of Patient Information on Repaired Hospital Products

• Before removing the product from hospital premises for repairs, all patient information should
be backed up to a storage device such as external HDD or memory, and then handed over to
the hospital.

• Before removing the product from hospital premises, each patient should sign the patient
privacy agreement.
Request for Repairs and Patient Privacy Agreement
Date: YYYY/MM/DD
【Customer Information】

Hospital
Handler

【Product Information】

Product name

Software version

S/N

Symptom

□ None □ Main unit □ HDD □ DVD/CD


Stored item
□ USB □ Photo □ Other ( )

□ None
Outsourced product
□ Provide details if applicable ( )

【Administration】

Handling date YY MM DD
Customer service
representative

※ Please sign inside the box below if you agree to the following:

□ Data saved on memory devices of products being repaired (e.g., hard disks) may become
lost during tests and repairs.
□ Data saved on a product being repaired must be backed up by the customer.
□ Samsung will not be held responsible for any loss of data not backed up.
□ Samsung Medison does not use patient information for unauthorized purposes, modify
such information, or provide it to third parties.

【Customer Signature】

I entrust my product to ( ) for repair and agree to the policies above.

Date: YY/MM/DD Name: (Signature)


This document is the technical property of SAMSUNG MEDISON CO., LTD. Using this document without prior permi
ssion or for purposes other than repairs of the products of Samsung Medison without approval may be subject to leg
al actions.
Contents 1

Contents

Chapter 1. Safety and Maintenance


1.1. Safety Guidelines................................................................................................................. 3
1.1.1. Guidelines for Safe Usage .......................................................................................4
1.1.2. Safety Standards ......................................................................................................4
1.1.3. Safety Symbols and Labels ......................................................................................6
1.2. Operational Environment Safety ...................................................................................... 10

1.3. Electrical Safety ................................................................................................................. 11


1.3.1. Prevention of Electric Shocks ................................................................................11
1.3.2. ECG-Related Information .......................................................................................13
1.3.3. ESD (ElectroStatic Discharge) ...............................................................................13
1.3.4. EMI (ElectroMagnetic Interference) .......................................................................14
1.3.5. EMC (ElectroMagnetic Compatibility).....................................................................14
1.3.6. Electrical Safety Check ..........................................................................................23
1.3.7. Accuracy Check .....................................................................................................23
1.4. Mechanical Safety.............................................................................................................. 24
1.4.1. Product Transfer and Brakes .................................................................................24
1.4.2. Safety Precautions When Using the Product .........................................................26
1.5. Biological Safety ................................................................................................................ 31
1.5.1. ALARA Principle .....................................................................................................31
1.6. Maintenance ....................................................................................................................... 44
1.6.1. Cleaning and Frequency ........................................................................................44
1.6.2. Cleaning and Disinfection.......................................................................................45
1.7. Information Maintenance .................................................................................................. 48
1.7.1. User Settings Backup.............................................................................................48
1.7.2. Patient Information Backup ....................................................................................48
1.7.3. Software .................................................................................................................48
1.7.4. Network ..................................................................................................................49
1.7.5. Cybersecurity .........................................................................................................50
1.8. Battery Pack Management ................................................................................................ 51
2 V6/XV6 Service Manual

1.8.1. Replacing the Battery Pack ................................................................................... 51


1.8.2. Recharging the Battery Pack ................................................................................. 51
1.8.3. Disposing of the Battery Pack ................................................................................ 52
1.9. Protecting the Environment ............................................................................................. 53

Chpater 2. Introduction
2.1. Indications for Use .............................................................................................................. 2
2.1.1. Contraindications ..................................................................................................... 2
2.2. Product Specifications ....................................................................................................... 3

2.3. Product Configuration ........................................................................................................ 6


2.3.1. Monitor ..................................................................................................................... 8
2.3.2. Control Panel ......................................................................................................... 10
2.3.3. Console .................................................................................................................. 19
2.3.4. Peripheral Devices ................................................................................................. 21
2.3.5. Probes ................................................................................................................... 25
2.3.6. Accessories ........................................................................................................... 26
2.3.7. Optional Functions ................................................................................................. 28

Chapter 3. Installing Product


3.1. Transporting ........................................................................................................................ 2
3.1.1. Precaution during transport ..................................................................................... 2
3.1.2. Brakes...................................................................................................................... 2
3.1.3. Precaution on Ramp ................................................................................................ 2
3.2. Unpacking the Product ....................................................................................................... 3
3.2.1. Unpacking the product’s box ................................................................................... 3
3.2.2. Accessory ................................................................................................................ 5
3.3. Installation Environment .................................................................................................... 6
3.3.1. Caution .................................................................................................................... 6
3.3.2. Humidity and Temperature ...................................................................................... 7
3.3.3. Standby time before using the product .................................................................... 7
3.4. Installing the Product ......................................................................................................... 8
3.4.1. Installation Safety..................................................................................................... 8
Contents 3

3.4.2. Printer Installation...................................................................................................10


3.4.3. External Keyboard Installation ................................................................................15
3.4.4. Battery Option Installation ......................................................................................16
3.4.5. HV board(CAP board) Installation ..........................................................................17
3.4.6. ECG module Installation ........................................................................................18
3.4.7. Enter Installation Key .............................................................................................19

Chapter 4. Product Inspection


4.1. Inspecting the Functions .................................................................................................... 2
4.1.1. Basic Inspections .....................................................................................................2
4.1.2. Detailed Inspections .................................................................................................3
4.2. Periodic Maintenance Inspection ....................................................................................... 7
4.2.1. Maintenance Cycle by Item ......................................................................................7

Chapter 5. Product Structure


5.1. Overview ............................................................................................................................... 3

5.2. System block diagram and Layout .................................................................................... 4


5.2.1. Overall System block diagram .................................................................................4
5.2.2. System Layout(Body) ...............................................................................................4
5.3. Overall structure description ............................................................................................. 6
5.3.1. Ultrasound System Part ...........................................................................................6
5.3.2. PC Part .....................................................................................................................6
5.3.3. User Interface Part ...................................................................................................7
5.3.4. Power Part................................................................................................................7
5.4. Ultrasound System Part ...................................................................................................... 8
5.4.1. PSA (Probe Select Assembly)..................................................................................8
5.4.2. Beamformer board and CW board .........................................................................10
5.4.3. Back End & Rear board .........................................................................................14
5.5. PC Part ................................................................................................................................ 15
5.5.1. PC Module..............................................................................................................15
5.5.2. Software Digital Scan Converter(DSC) ..................................................................16
5.6. Control Panel ..................................................................................................................... 18
4 V6/XV6 Service Manual

5.6.1. Control Panel ......................................................................................................... 18


5.6.2. Main Monitor .......................................................................................................... 21
5.6.3. Touch Panel........................................................................................................... 22
5.7. Power Part.......................................................................................................................... 23
5.7.1. PSU / Battery / Battery charger .............................................................................. 23
5.7.2. Power Block Diagram ............................................................................................ 24
5.7.3. Power Status LED ................................................................................................. 28
5.8. Interconnect Diagram ....................................................................................................... 30
5.8.1. V6/XV6Interconnect Diagram ................................................................................ 30
5.8.2. Cable Description .................................................................................................. 31

Chapter 6. Service Mode


6.1. Service Mode ....................................................................................................................... 3
6.1.1. How to enter Service Mode ..................................................................................... 3
6.2. Service Information ............................................................................................................. 4

6.3. Configuration ....................................................................................................................... 5


6.3.1. Keyboard/Region ..................................................................................................... 5
6.3.2. TCP/IP ..................................................................................................................... 6
6.3.3. Option ...................................................................................................................... 7
6.3.4. User Account ........................................................................................................... 8
6.3.5. Printer ...................................................................................................................... 9
6.3.6. RMS ....................................................................................................................... 10
6.3.7. RIS Browser ........................................................................................................... 11
6.3.8. AppLocker.............................................................................................................. 12
6.3.9. MobileExport Serve IP ........................................................................................... 13
6.3.10. SonoSync RIO ....................................................................................................... 14
6.3.11. System Status Notifier ........................................................................................... 15
6.3.12. Touch Layout ......................................................................................................... 16
6.3.13. Vet Mode ............................................................................................................... 17
6.4. System................................................................................................................................ 18
6.4.1. Change Information ............................................................................................... 18
6.4.2. Upgrade Software .................................................................................................. 19
6.4.3. Upgrade Hardware ................................................................................................ 20
Contents 5

6.4.4. Window Explorer ....................................................................................................20


6.4.5. Specialist Mode ......................................................................................................21
6.5. Diagnostic .......................................................................................................................... 22
6.5.1. Control Panel Test .................................................................................................22
6.5.2. Keyboard Test ........................................................................................................22
6.5.3. Built in Self Test .....................................................................................................23
6.5.4. Monitor Test ...........................................................................................................24
6.5.5. Binary Verification Tool ..........................................................................................25
6.5.6. Virus Scan ..............................................................................................................26
6.5.7. Probe Element Check ............................................................................................27
6.6. Backup and Restore .......................................................................................................... 28
6.6.1. Backup ...................................................................................................................28
6.6.2. Restore ...................................................................................................................28
6.6.3. Log & Audit Trail Backup ........................................................................................29
6.7. Report ................................................................................................................................. 30
6.7.1. Report.....................................................................................................................30
6.8. Demo Play .......................................................................................................................... 31
6.8.1. Demo Play ..............................................................................................................31
6.8.2. Extended Mode ......................................................................................................32

Chapter 7. Self Diagnostic


7.1. Purpose for usage ................................................................................................................. 2

7.2. Self-Diagnostic Test Item ..................................................................................................... 2


7.2.1 Dummy Probe .............................................................................................................2
7.3. Self-Diagnostic Entry ............................................................................................................ 3
7.3.1 Service mode entry .....................................................................................................3
7.4. Self-Diagnostic Run .............................................................................................................. 4

7.5. Self-Diagnostic Result .......................................................................................................... 6

7.6. Self-Diagnostic Log .............................................................................................................. 7


7.6.1 How to check Self-Diagnostic Log ..............................................................................7
7.7. Self-Diagnostic End .............................................................................................................. 9

7.8. Test Items according to Dummy Probe ............................................................................ 10


6 V6/XV6 Service Manual

7.9. Self-Diagnostic Test Item Description .............................................................................. 11


7.9.1 Power........................................................................................................................ 11
7.9.2 Bus / Analog / CW / Memory / Motor / Fan / PSA .................................................... 12
7.10. Remote Maintenance Sytem ............................................................................................ 13

Chapter 8. Troubleshooting
8.1. Power.................................................................................................................................... 2
8.1.1. Power doesn’t Turn on............................................................................................. 2
8.1.2. Power does not turn off............................................................................................ 2
8.1.3. Power Turns Off by itself ......................................................................................... 2
8.1.3. PSU Stand-Alone Self- diagnostics ......................................................................... 3
8.1.3. BP board Self-diagnostics ....................................................................................... 3
8.2. Monitor ................................................................................................................................. 5
8.2.1. Nothing is displayed on the screen .......................................................................... 5
8.2.2. Screen is discolored ................................................................................................ 5
8.3. Error Message ..................................................................................................................... 6
8.3.1. Error Occurs during Booting .................................................................................... 6
8.4. Image .................................................................................................................................... 7
8.4.1. 2D Mode: There is no Image Echo or Image Format .............................................. 7
8.4.2. Line(noise) Appear in 2D image .............................................................................. 7
8.4.3. M, C, PW, CW Mode failure .................................................................................... 7
8.5. Troubleshooting Tree ......................................................................................................... 8
8.5.1. System does not power on. ..................................................................................... 8
8.5.2. No image on the monitor. ........................................................................................ 9
8.5.3. Noise in the image area. ........................................................................................ 10
8.5.4. Track Ball does not work ....................................................................................... 11
8.5.5. System does not turn off ........................................................................................ 12

Chapter 9. Disassembly
9.1. Caution ................................................................................................................................. 2
9.1.1. Preparation .............................................................................................................. 2
9.2. Disassembling the system ................................................................................................. 3
Contents 7

9.2.1. Front side disassembly ............................................................................................3


9.2.2. Rear side disassembly .............................................................................................4
9.2.3. Keyboard disassembly .............................................................................................6
9.2.4. Control Panel disassembly .......................................................................................7
9.2.5. Touch Panel disassembly ........................................................................................8
9.2.6. Monitor disassembly.................................................................................................9
9.2.7. Monitor ARM disassembly......................................................................................10

Chapter 10. Probe


10.1. Probes ................................................................................................................................... 2
10.1.2. Ultrasound Transmission Gel .................................................................................14
10.1.3. Using Sheaths ........................................................................................................16
10.1.4. Probe Safety Precautions.......................................................................................18
10.1.5. Cleaning, Disinfecting, and Sterilizing Probes ........................................................20
10.1.6. MPTEE Probe (Option) ..........................................................................................35
10.2. Biopsy ................................................................................................................................. 42
10.2.1. Biopsy Kit Components ..........................................................................................43
10.2.2. Using Biopsy Kit .....................................................................................................44
10.2.3. Assembling the Biopsy Kit ......................................................................................47
10.2.4. Cleaning and Disinfecting the Biopsy Kit ................................................................52

Chapter 11. Service parts List


11.1. Overview ............................................................................................................................... 2

11.2. Body Cover Parts ................................................................................................................. 2

11.3. Control Panel Parts ............................................................................................................. 6

11.4. System Cable Parts ........................................................................................................... 20

11.5. System Parts ...................................................................................................................... 23


Chapter 1

Safety and Maintenance


1.1. Safety Guidelines ............................................................... 3
1.1.1. Guidelines for Safe Usage ............................................................... 4
1.1.2. Safety Standards.............................................................................. 4
1.1.3. Safety Symbols and Labels.............................................................. 6
1.2. Operational Environment Safety ..................................... 10
1.3. Electrical Safety ................................................................. 11
1.3.1. Prevention of Electric Shocks ........................................................ 11
1.3.2. ECG-Related Information ............................................................... 13
1.3.3. ESD (ElectroStatic Discharge) ....................................................... 13
1.3.4. EMI (ElectroMagnetic Interference) ............................................... 14
1.3.5. EMC (ElectroMagnetic Compatibility)............................................. 14
1.3.6. Electrical Safety Check .................................................................. 23
1.3.7. Accuracy Check ............................................................................. 23
1.4. Mechanical Safety ............................................................ 24
1.4.1. Product Transfer and Brakes ......................................................... 24
1.4.2. Safety Precautions When Using the Product ................................. 26
1.5. Biological Safety ............................................................... 31
1.5.1. ALARA Principle............................................................................. 31
1.6. Maintenance ...................................................................... 44
1.6.1. Cleaning and Frequency ................................................................ 44
1.6.2. Cleaning and Disinfection .............................................................. 45
1.7. Information Maintenance ................................................. 48
1.7.1. User Settings Backup .................................................................... 48
1.7.2. Patient Information Backup ............................................................ 48
1.7.3. Software ......................................................................................... 48
1 - 2 V6/XV6 Service Manual

1.7.4. Network.......................................................................................... 49
1.7.5. Cybersecurity ................................................................................. 50
1.8. Battery Pack Management ............................................... 51
1.8.1. Replacing the Battery Pack ............................................................ 51
1.8.2. Recharging the Battery Pack ......................................................... 51
1.8.3. Disposing of the Battery Pack ........................................................ 52
1.9. Protecting the Environment ............................................. 53
Chapter 1. Safety and Maintenance 1 - 3

1.1. Safety Guidelines


The following terms are used to emphasize circumstances in which the user must take extra
caution:

Disregarding this instruction may result in death, serious injury, or other


dangerous situations.
DANGER

Follow these warnings to prevent a serious accident or damage to property.

WARNING

Follow these instructions to prevent minor accidents and damage to property.

CAUTION

The accompanying information covers installation, operation, and maintenance


procedures that require careful attention from the user, but have little chance of
leading directly to a dangerous situation.
NOTE
1 - 4 V6/XV6 Service Manual

1.1.1. Guidelines for Safe Usage

■ This product is intended for use by, or by the order of, and under the
supervision of, a licensed physician who is qualified for direct use of medical
devices.
■ This product must not be used for ophthalmological applications, or any other
DANGER use that involves the ultrasound beam passing through the eyeball.

■ The manufacturer is not responsible for any injury or damage that may result
from using the product for purposes not specified or not following the safety
information.
■ Make sure that you use the three-dimensional ultrasound diagnostic imaging
CAUTION system for its intended purposes only, since using it for a long period of time
for purposes other than diagnosing the fetus, such as keepsake images or
videos, may have an adverse effect on the fetus.
■ According to the IEC 60601-2-37 Annex CC, when used for cardiac scanning,
the lung surfaces may be exposed to ultrasonic beam; therefore, recommend
minimizing the MI value.
■ Detergents or disinfectants may cause skin irritation. Avoid leaving residues
on the product.

1.1.2. Safety Standards


■ Classification
– Type of protection against electric shocks: Class I or Internally powered ME
equipment
– Degree of protection against electric shocks (when the patient is in physical contact):
Type BF applied part (Probes), Defibrillation-Proof Type CF applied part (ECG)
– Degree of protection against the ingress of harmful liquids: Ordinary equipment
(Ultrasound System), IPX7 (Probes), IPX8 (Foot Switch)
– Degree of safety of use in the presence of flammable anesthetic agent mixed with air,
oxygen, or nitrous oxide: Not suitable for use near flammable anesthetic agent mixed
with air, oxygen, or nitrous oxide
– Mode of operation: Continuous operation

■ Safety Standards the Device Conforms to


– Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and
Essential Performance [IEC 60601-1]
– Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and
Essential Performance [ANSI/AAMI ES60601-1]
– Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and
Essential Performance [CAN/CSA-22.2 No. 60601-1]
Chapter 1. Safety and Maintenance 1 - 5

– Medical Electrical Equipment – Part 1-2: General Requirements for Basic Safety and
Essential Performance – Collateral Standard: Electromagnetic Compatibility –
Requirements and Tests [IEC 60601-1-2]
– Medical Electrical Equipment – Part 1-6: General Requirements for Basic Safety and
Essential Performance – Collateral Standard: Usability [IEC 60601-1-6]
– Medical Electrical Equipment – Part 2-37: Particular Requirements for the Basic
Safety and Essential Performance of Ultrasonic Medical Diagnostic and Monitoring
Equipment [IEC 60601-2-37]
– Medical Devices – Application of Risk Management to Medical Devices [ISO 14971]
– Biological Safety Evaluation of Medical Devices – Part 1: Evaluation and Testing [ISO
10993-1]

■ Declarations

The Nemko-CCL mark with “C” and “US” identifiers indicates


that the product is compliant with and certified to the applicable
U.S. and Canadian standards.

This mark certifies that the product conforms to applicable EEC


standards and has been certified by the European certification
agency.

This mark certifies that the product conforms to applicable EEC


standards.

The GMP symbol indicates the Good Manufacturing Practice in


accordance with the Korean quality management regulations.
1 - 6 V6/XV6 Service Manual

1.1.3. Safety Symbols and Labels


These are safety and notice symbols for medical electronic equipment. The International
Electrotechnical Commission (IEC) has established the part of the symbols.

Symbols Description

WARNING: Warnings to prevent a serious accident, or damage to


property

CAUTION: Precautions to prevent a minor accident or damage to


property

Refer to the user manual.

Follow the user manual.

CAUTION: Risk of electric shock

Type BF applied part (Classification based on degree of protection


against electric hazard)

Defibrillation-proof type CF applied part (Classification based on degree


of protection against electric hazard)

Power on/off

Power on

Power off

~ Alternating current voltage source


Chapter 1. Safety and Maintenance 1 - 7

Symbols Description

Direct current voltage source

Protective grounding terminal

Equipotential terminal

Data output port

Data input port

Data input/output port

Input port

Output port

Print remote output

Foot switch port

ECG port

USB port

Network port

Microphone port
1 - 8 V6/XV6 Service Manual

Symbols Description

Probe port

IPX 1 Dripping-proof device: Protected against vertically falling water drops

Immersion-proof device: Protected against the effects of non-continuous


IPX 7 immersion in water

Submersion-proof device: Protected against the effects of continuous


IPX 8 immersion in water

CAUTION: Electrostatic sensitive devices (ESD)

Do not sit on the product.

Do not push the product.

DANGER: Do not place your fingers, hands, or any parts of your body in
the space.

Do not place objects heavier than the weight specified next to the
symbol.

Do not place heavy objects on the monitor while the monitor mount is
folded.

Do not push down on the monitor while the monitor mount is folded.
Chapter 1. Safety and Maintenance 1 - 9

Symbols Description

Do not stand on it.

Federal law restricts this device to sale by or on the order of a physician.

The Authorized European Representative’s address is provided.

The manufacturer’s name and address are provided.

This equipment is a medical device.

Consult the latest information.

Indicates the need for the user to consult electronic instructions for use.
(You can access to it by pressing help on the product or download it
from the website.)
www.samsunghealthcare.com

■ Labels
Phrases containing the words ‘WARNING’ and/or ‘CAUTION’ are displayed on the
product’s surface in order to protect it.
1 - 10 V6/XV6 Service Manual

1.2. Operational Environment Safety


The user is responsible for the safety of everyone near the product. Please handle the product
with caution even when it is not in use. The product must be installed by a professional
technician. Please remember the following:

■ Using the system near generators, X-ray machines, or broadcast cables may
result in noise and abnormal visual images. Sharing the power source with
other electrical devices may also cause noise.
■ Do not install or use the product in the Magnetic Resonance(MR)
CAUTION Environment. Doing so may result in improper operation of the product due to
powerful magnetic fields.
■ Optimal conditions for the normal operation of the system are a temperature
of 10–35°C and a humidity of 30–75%.
■ Avoid areas where moisture, direct sunlight, or a heater is near the product;
that are frequently exposed to vibration; that are too dusty or are not properly
ventilated; or where chemicals or gases are present.
■ Avoid using this product close to or on top of another device, as this may
cause malfunctions.

Do not block the vents on the system, peripheral devices, and accessories.

WARNING
Chapter 1. Safety and Maintenance 1 - 11

1.3. Electrical Safety


This product is categorized as a Class I device.

■ As for US requirement, the LEAKAGE CURRENT might be measured from a


center-tapped circuit when the equipment connects in the United States to
240V supply system.
■ To help assure grounding reliability, connect to a “hospital grade” or “hospital
CAUTION only” grounded power outlet.

1.3.1. Prevention of Electric Shocks


In a hospital environment, hazardous current can form due to potential differences between
exposed conductive parts and connected devices. The solution to the problem is consistent
equipotential bonding. Medical equipment is connected with connecting leads made up of
sockets which are angled to the equipotential bonding network in medical rooms.

For more information about an equipotential connection, please refer to


‘Introduction’ in this manual.
NOTE

All connected peripheral devices must comply with the relevant IEC standards (e.g., IEC
60950/EN 60950 for data processing equipment, and IEC 60601-1/EN 60601-1/ANSI AAMI
ES60601-1 for medical devices). Furthermore, all components of the product must comply with
the requirements for medical electrical systems IEC 60601-1/EN 60601-1/ANSI AAMI ES60601-
1. Anybody connecting additional equipment to signal input and output ports of medical
electrical equipment must make sure that the peripherals comply with IEC 60601-1/EN 60601-
1/ANSI AAMI ES60601-1.
1 - 12 V6/XV6 Service Manual

■ Electric shocks may result if this system, including all of its externally
mounted recording and monitoring devices, is not properly grounded.
■ Never open the cover of the product. The interior of this product contains
dangerous high-voltage electricity. Any repairs to the product's internal
WARNING components or replacement of parts must be performed by Samsung
Medison.
■ Always check the product’s housing, cables, cords, and plugs before using
the product. Disconnect the power source and do not use the equipment if
the housing is damaged (for example cracked or chipped), or if the cable is
worn.
■ Always disconnect the system from the wall outlet prior to cleaning the
system.
■ All patient-contact devices, such as probes, must be detached from the
patient prior to using a high-voltage defibrillator.
■ Never use the product in the presence of flammable or anesthetic gas. Doing
so may cause an explosion.
■ Avoid installing the system in such a way that it is difficult for the operator to
disconnect it from the power source.
■ Do not use together with HF surgical equipment. HF surgical equipment may
be damaged, which may result in fire.
■ The product must be connected to a power supply with a protective earth to
prevent electric shock.

■ The system has been designed for 100-240VAC; you should select the input
voltage of any connected printer and VCR. Prior to connecting a peripheral
power cord, verify that the voltage indicated on the power cord matches the
voltage rating of the peripheral device.
CAUTION ■ An isolation transformer protects the system from power surges. The
isolation transformer continues to operate when the system is in standby.
■ Do not immerse the cable in liquids. Cables are not waterproof.
■ Make sure that the inside of the system is not exposed to or flooded with
liquids. In such cases, fire, electric shock, injury, or damage to the product
may occur.
■ The auxiliary socket outlets installed on this system are rated 100-240VAC
with maximum total load of 150VA. Use these outlets only for supplying
power to equipment that is intended to be part of the ultrasound system. Do
not connect additional multiple-socket outlets or extension cords to the
system.
■ Connecting any devices that are not listed in this User Manual to the
auxiliary socket outlet of the system may cause an electrical hazard.
■ Do not touch SIP/SOP and the patient simultaneously. There is a risk of
electric shock from leakage current.
Chapter 1. Safety and Maintenance 1 - 13

1.3.2. ECG-Related Information

■ This device is not intended to provide a primary ECG monitoring function,


and therefore does not have means of indicating an inoperative
electrocardiograph.
■ Do not use ECG electrodes with HF surgical equipment. HF surgical
WARNING equipment may be damaged, which may result in fire.
■ Do not use ECG leads and electrodes in an operating room.

Using ECG electrodes with a cardiac pacemaker or an electrical stimulator at


the same time may cause interference with the ECG signals.
CAUTION

1.3.3. ESD (ElectroStatic Discharge)


An ESD event is generally caused by static electricity that naturally occurs due to friction. ESD
is most prevalent during conditions of low humidity, which can be caused by heating or air
conditioning. The static shock or ESD is a discharge of the electrical energy build-up from a
charged individual to a less or non-charged individual or object. An ESD occurs when an
individual with an electrical energy build-up comes into contact with conductive objects such as
metal doorknobs, file cabinets, computer equipment, and even other individuals.

■ The level of electrical energy discharged from a system user or patient to an


ultrasound system can be significant enough to cause damage to the system
or probes.
■ Always perform the pre-ESD preventive procedures before using connectors
CAUTION marked with the ESD warning label.
– Apply anti-static spray to carpets or linoleum.
– Use anti-static mats.
– Ground the product to the patient table or bed.
■ It is highly recommended that the user be given training on ESD-related
warning symbols and preventive procedures.
1 - 14 V6/XV6 Service Manual

1.3.4. EMI (ElectroMagnetic Interference)


This product has passed the EMI compliance test. However, using the system where an
electromagnetic field is being produced may lower the quality of ultrasound images and/or
damage the product.
If this occurs often, Samsung Medison suggests a review of the environment in which the
system is being used, to identify possible sources of radiated emissions. These emissions could
be from other electrical devices used within the same room or an adjacent room.
Communication devices such as cellular phones and pagers can cause these emissions. The
existence of radios, TVs, or microwave transmission equipment nearby can also cause
interference.

In cases where EMI is causing disturbances, it may be necessary to relocate this


system.
CAUTION

1.3.5. EMC (ElectroMagnetic Compatibility)


EMC testing on this product has been performed according to the international EMC standard
for medical devices, IEC 60601-1-2. This IEC standard was adopted in Europe as the European
norm (EN 60601-1-2).

1.3.5.1. Guidance and Manufacturer’s Declaration – Electromagnetic Emission


This product is intended for use in the electromagnetic environment specified below. The
customer or the user of this product should ensure that it is used in such an environment.

Emission Test Compliance Electromagnetic Environment – Guidance

The Ultrasound System uses RF energy only for its


RF Emission internal functions. Therefore, its RF emissions are very
Group 1
CISPR 11 low and are not likely to cause any interference in nearby
electronic equipment.

RF Emission
Class A
CISPR 11
The Ultrasound System is suitable for use in all
Harmonic Emission establishments other than domestic and those directly
Class A connected to the public low-voltage power supply
IEC 61000-3-2 network that supplies buildings used for domestic
purposes.
Flicker Emission
Complies
IEC 61000-3-3
Chapter 1. Safety and Maintenance 1 - 15

1.3.5.2. Approved Cables, Probes and Peripherals for EMC

■ Cables
Cables connected to this product may affect its emissions; use only the cable types and
lengths listed in the table below.

Cable Type Length

VGA Shielded Normal

USB Shielded Normal

LAN(RJ45) Twisted pair Any

S-Video Shielded Normal

Foot Switch Shielded 2.99m

Audio R.L Shielded Normal

Parallel Shielded Normal

HDMI Shielded Normal

ECG 3-lead Shielded Normal

■ Probes
The image probe used with this product may affect its emission. The probe listed in
‘Probes’ when used with this product, have been tested to comply with the group1 Class
A emission as required by International Standard CISPR 11.

■ Peripherals
Peripherals used with this product may affect its emissions.

When connecting other customer-supplied accessories to the system, it is the


user’s responsibility to ensure the electromagnetic compatibility of the system.
CAUTION

The use of cables, probes, and peripherals other than those specified may result
in increased emission or decreased Immunity of the Ultrasound System.
WARNING
1 - 16 V6/XV6 Service Manual

Electromagnetic
Immunity Test IEC 60601 Test Level Compliance Level
Environment – Guidance

Floors should be wood,


concrete, or ceramic tile.
Electrostatic ±8 kV contact ±8 kV contact
Discharge (ESD) If floors are covered with
±2 kV, ±4 kV, ±8 kV, ±15 ±2 kV, ±4 kV, ±8 kV, ±15
synthetic material, a relative
IEC 61000-4-2 kV air kV air
humidity of at least 30% is
recommended.

±2 kV (for power supply ±2 kV (for power supply Mains power quality should
Electrical fast
lines) lines) be that of a typical
transient/burst
±1 kV (for input/output ±1 kV (for input/output commercial or hospital
IEC 61000-4-4 environment.
lines) lines)

±0.5 kV, ±1 kV ±0.5 kV, ±1 kV Mains power quality should


Surge differential mode differential mode be that of a typical
IEC 61000-4-5 ±0.5 kV, ±1 kV, ±2 kV ±0.5 kV, ±1 kV, ±2 kV commercial or hospital
common mode common mode environment.

Mains power quality should


be that of a typical
Voltage dips, 0% Uт (100% dip in Uт) 0% Uт (100% dip in Uт) commercial or hospital
short for 0.5/1 cycles for 0.5/1 cycles environment. If the user of
interruptions, and this product requires
voltage variations 70% Uт (30% dip in Uт) 70% Uт (30% dip in Uт)
continued operation during
on power supply for 25/30 cyclesa for 25/30 cyclesa
power mains interruptions, it
input lines 0% Uт (100% dip in Uт) 0% Uт (100% dip in Uт) is recommended that this
IEC 61000-4-11 for 250/300 cyclesa for 250/300 cyclesa product be powered from an
uninterruptible power supply
or a battery.

Power frequency Power frequency magnetic


magnetic field fields should be at levels
(50/60 Hz) 30 A/m 30 A/m characteristic of a typical
commercial or hospital
IEC 61000-4-8 environment.
NOTE: Uт is the main voltage (AC) prior to the application of the test level.
a
For example, 10/12 means 10 cycles at 50 Hz or 12 cycles at 60 Hz.
Chapter 1. Safety and Maintenance 1 - 17

IEC 60601 Test Compliance


Immunity Test Electromagnetic Environment – Guidance
Level Level

Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications


IEC 61000-4-6 150 kHz – 80 MHz equipment should not be used closer to any part
of the Ultrasound System, including cables, than
Outside ISM the recommended separation distance. This is
bandsc calculated using the equation applicable to the
frequency of the transmitter.

6 Vrms 6 Vrms Recommended Separation Distance


150 kHz – 80 MHz
In ISM bandsc
IEC 60601-1-2: 2007

80 MHz to 800 MHz

800 MHz to 2.5 GHz

IEC 60601-1-2: 2014

80 MHz to 2.7 GHz

Radiated RF 3 V/m 3 V/m Where P is the transmitter’s maximum output


IEC 61000-4-3 80 MHz – 2.7 GHz power rating in watts (W) according to the
transmitter’s manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmittersa, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency rangeb.
Interference may occur in the vicinity of
equipment marked with the following symbol:

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not be applicable in some situations. Electromagnetic propagation is
affected by absorption and reflection of structures, objects, and people.
a Field strengths from fixed transmitters, such as base stations for radio, (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcasts, and TV broadcasts cannot be
predicted with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength, in the location in which
the Ultrasound System is used, exceeds the applicable RF compliance level above, the Ultrasound
System should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Ultrasound System or using a
shielded location with a higher RF shielding effectiveness and filter attenuation.
b When the frequency range exceeds 150 kHz – 80 MHz, the electric field strength should be not higher

than 3 V/m.
c The ISM (Industrial, Scientific and Medical) bands between 150 kHz and 80 MHz are 6.765 MHz to

6.795MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz
1 - 18 V6/XV6 Service Manual

1.3.5.3. Recommended Distance Between Wireless Communication Device and


this Product
This product is intended for use in an electromagnetic environment, in which radiated RF
disturbances are controlled. The customer or the user of this product can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and this product. These distances are
recommended below, according to the maximum output power of the communications
equipment.

Separation Distance in Relation to Frequency of Transmitter [m]


Rated
Maximum IEC 60601-1-2: 2007 IEC 60601-1-2: 2014
Output
Power of 150 kHz – 80 80 MHz – 800 MHz – 2.5 150 kHz – 80 80 MHz – 2.7
Transmitter MHz 800 MHz GHz MHz GHz
[W]

0.01 0.12 0.12 0.23 0.12 0.20

0.1 0.38 0.38 0.73 0.38 0.63

1 1.2 1.2 2.3 1.2 2.0

10 3.8 3.8 7.3 3.8 6.3

100 12 12 23 12 20

For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where p is the maximum output power rating of the transmitter in watts (W), according to
the transmitter’s manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not be applicable in some situations. Electromagnetic propagation is
affected by absorption and reflection of structures, objects, and people.

1.3.5.4. Guidance and Manufacturer’s Declaration – Electromagnetic Immunity


This product is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. Potrable RF communications equipments shoud be used no
closer than 30 cm (12 inches) to any parts of the product. Otherwise, the performance of this
equipment could be degraded.
IEC 60601
Banda Compliance
Immunity Test Servicea Modulationb Test Level
(MHz) Level (V/m)
(V/m)
Pulse modulationb
380 – 390 TETRA 400 27 27
Proximity fields 18 Hz
from RF FMc
wireless GMRS 460
communication 430 – 470 ±5 kHz deviation 28 28
FRS 460
s equipment 1 kHz sine
IEC61000-4-3 LTE Band 13, Pulse modulationb
704 – 787 9 9
17 217 Hz
Chapter 1. Safety and Maintenance 1 - 19

GSM 800/900
TETRA 800
Pulse modulationb
800 – 960 iDEN 820 28 28
18 Hz
CDMA 850
LTE Band 5
GSM 1800
CDMA 1900
GSM 1900 Pulse modulationb
1700 – 1990 28 28
DECT 217 Hz
LTE Band 1, 3,
4, 25 UMTS
Blue Tooth
WLAN
Pulse modulationb
2400 – 2570 802.11 b/g/n 28 28
217 Hz
RFID 2450
LTE Band 7

WLAN 802.11 Pulse modulationb


5100 – 5800 9 9
a/n 217 Hz
NOTE: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is
permitted by IEC 61000-4-3.
a For Some services, only the uplink frequencies are included.
b The carrier shall be modulated using a 50% duty cycle square wave signal.
c As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it

does not represent actual modulation, it would be worst case.

■ Use of this equipment adjacent to or stacked with other equipment should


be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to
verify that they are operating normally.
WARNING
■ Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
■ Portable RF communications equipment (including peripherals such as
antenna cables and external antennas) should be used no closer than 30
cm (12 inches) to any part of the ME EQUIPMENT or ME SYSTEM,
including cables specified by the manufacturer.
■ Failure to use this equipment in the specified type of shielded location
could result in degradation of the performance of this equipment,
interference with other equipment or interference with radio services.
■ This equipment has been tested for radiated RF immunity only at selected
frequencies, and use nearby of emitters at other frequencies could result in
improper operation.
1 - 20 V6/XV6 Service Manual

The EMISSIONS characteristics of this equipment make it suitable for use in


industrial areas and hospitals (CISPR 11 Class A). If it is used in a residential
environment (for which CISPR 11 Class B is normally required) this
equipment might not offer adequate protection to radio-frequency
NOTE communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the equipment.

1.3.5.5. Electromagnetic Environment – Guidance


It is recommended to use ultrasound systems in shielded locations offering RF shielding
effectiveness, with shielded cables. Field strengths outside the shielded location from fixed
RF transmitters, as determined by an electromagnetic site survey, should be less than 3V/m.
It is essential that the actual shielding effectiveness and filter attenuation of the shielded
location be verified to ensure that they meet the minimum specification.

If the system is connected to other customer-supplied equipment, such as a


local area network (LAN), Samsung Medison cannot guarantee that the remote
equipment will work correctly in the presence of electromagnetic emission
CAUTION phenomena.
Chapter 1. Safety and Maintenance 1 - 21

1.3.5.6. Avoiding Electromagnetic Interference


Typical interference on Ultrasound Imaging Systems varies depending on Electromagnetic
phenomena. Please refer to the following table:

Imaging Mode ESD1 RF2 Power Line3

For sector imaging


probes, white radial bands
or flashes in the
centerlines of the image. White dots, dashes,
diagonal lines, or
2D For linear imaging probes, diagonal lines near the
white vertical bands, center of the image.
sometimes more
pronounced on the sides
of the image.

Change of operating White dots, dashes,


mode, system settings, Increase in the image
diagonal lines, or
M or system reset. Brief background noise or white
increase in image
flashes in the displayed M mode lines.
background noise
or recorded image.

Color flashes, radial or


Color flashes, dots,
vertical bands, increase in
Color dashes, or changes in the
background noise, or
color noise level.
changes in color image.

Horizontal lines in the Vertical lines in the


spectral display or tones, spectral display, popping
Doppler
abnormal noise in the type noise in the audio, or
audio, or both. both.

1. ESD caused by discharging of electric charge build-up on insulated surfaces or persons.


2. RF energy from RF transmitting equipment such as portable phones, hand-held radios, wireless
devices, commercial radio and TV, and so on.
3. Conducted interference on powerlines or connected cables caused by other equipment, such as
switching power supplies, electrical controls, and natural phenomena such as lightning.
1 - 22 V6/XV6 Service Manual

A medical device can either generate or receive electromagnetic interference. The EMC
standards describe tests for both emitted and received interference.
Electromagnetic interference generated by Samsung Medison’s ultrasound system does not
exceed the standard level.
An Ultrasound System is designed to receive signals at radio frequency and is therefore
susceptible to interference generated by RF energy sources. Examples of other sources of
interference are medical devices, information technology products, and radio and television
transmission towers. Tracing the source of radiated interference can be a difficult task.
Customers should consider the following in an attempt to locate the source:
1. Is the interference intermittent or constant?
2. Does the interference show up only with one transducer operating at the same frequency
or with several transducers?
3. Do two different transducers operating at the same frequency have the same problem?
4. Is the interference present if the system is moved to a different location in the facility?
The answers to these questions will help to determine if the problem resides with the
ultrasound system or its operating environment. After you have answered all of the questions,
contact your local service representative.
Chapter 1. Safety and Maintenance 1 - 23

1.3.6. Electrical Safety Check


The product’s maintenance status may affect the electrical safety of the product. The product
should be maintained in an electrically safe state to ensure human safety.

Users may request safety inspections as required, but only trained persons are
allowed to perform the safety inspection.
CAUTION

1.3.7. Accuracy Check


The product's maintenance status may affect the measurements obtained using the product.
The product should be maintained in an optimal state to ensure reliable measurements. The
equations and tables related to measurement accuracy are included in the ‘Measurements’.

In case of doubt in the measurements, please contact your local Service


Department for an accuracy test.
CAUTION
1 - 24 V6/XV6 Service Manual

1.4. Mechanical Safety

1.4.1. Product Transfer and Brakes

Special caution should be taken during movement. Careless transportation of


the product may result in product damage or personal injury.
WARNING

■ Before transporting the product, make sure that the caster brakes are
released. Also, make sure that any parts that may move, such as the monitor
arm or keyboard, are secured into position before moving.
■ You can use the caster brakes on the product to control their movement.
CAUTION
Make sure that the brakes are engaged before you use the product.
■ Always use the handles on the console and move the product slowly.
■ Excessive shocks, such as a fall, may seriously damage the product.

■ When you need to move the product up or down an inclined plane, make
sure that the control panel is facing the direction in which you are moving and
that it is moved by at least two people.
■ If the product is temporarily left on a slope, it can still fall over even with the
WARNING caster brakes on. Do not leave the product on a slope.

If the product does not work properly after you move it, please contact the
Customer Service Department.
NOTE

1.4.1.1. Precaution before Moving the Equipment

The keyboard, control panel, touchscreen and monitor arm should be fixed
before moving them.
CAUTION
Chapter 1. Safety and Maintenance 1 - 25

1. Push the keyboard into the control panel for fixation.

2. Press the Lever Ⓐ on the control panel with both hands to fix the control panel at the
lowest height.
3. Push the touchscreen back for fixation.

4. Adjust the monitor location appropriately and then use the lock on the monitor arm Ⓑ
for fixation.
1 - 26 V6/XV6 Service Manual

1.4.2. Safety Precautions When Using the Product

■ Never attempt to modify the product in any way.


■ If using the product after a prolonged period of non-use, always check that it
can be safely operated.
CAUTION ■ Make sure that other objects, such as metal pieces, do not enter the system.
■ Do not block the ventilation slots.
■ Do not pull on the power cord to unplug the product. Doing so might damage
the cord and cause the product to short-circuit, or the cord itself to break.
Unplug the cord by pulling on the plug itself.
■ Excessive bending or twisting of cables on patient-applied parts may cause
failure or intermittent operation of the system.
■ Improper cleaning or sterilization of a patient-applied part may cause
permanent damage.
■ All internal component repairs and part replacements must be done by
qualified Samsung Medison service personnel. Assuming that the product is
used in accordance with the guidelines contained in this manual and
maintained by qualified service personnel, the expected service life of the
product is approximately 7 years.

1.4.2.1. Monitor
When adjusting the height or position of the monitor, be careful of the space in the middle of
the monitor arm. Catching your fingers or other body parts in it may result in injury.

[Figure 1.1 Safety Note for Monitor]


Chapter 1. Safety and Maintenance 1 - 27

[Figure 1.2 Monitor position adjustment]

[Figure 1.3 Monitor height adjustment]

[Figure 1.4 Monitor arm lock]

■ When moving the product, be sure to keep the monitor arm locked.
■ When rotating the monitor, be sure to release the lock first. Rotating
the monitor while it is locked may cause damage to the internal
CAUTION cables.
1 - 28 V6/XV6 Service Manual

1.4.2.2. Control Panel


When adjusting the height or position of the control panel, pay attention to the space
between the lower-hand side of the control panel and the lever. Catching your fingers or
other body parts in it may result in injury.

■ Do not push the control panel with excessive force or lean on it


■ Do not sit on the control panel or exert excessive force on it.
■ Do not lift the product using the handles on the control panel.
CAUTION ■ Press the lever on the control panel all the way to use it.

Lever
[Figure 1.5 Safety Note for Control Panel]
Chapter 1. Safety and Maintenance 1 - 29

1.4.2.3. Touchscreen
Touch screen is a fixed type with no angle adjustment. Do not try to adjust the monitor angle
by confusing it with other models.

 The touch screen is fixed. Do not push back it to adjust the angle.

CAUTION

x
[Figure 1.6 Fixed type Touch Screen with no anble adjustment]
1 - 30 V6/XV6 Service Manual

1.4.2.4. Ventilation Window


When the ventilation window is clogged with dust, it may cause the product to overheat and
generate noise. It may also reduce the reliability and system performance. Ensure enough
room near the ventilation window and clean the dust on the ventilation window frequently
with a soft brush or a vacuum cleaner.

■ Be sure to lock the brakes on the wheels before cleaning the ventilation
window. Otherwise, unexpected movement of the product may result in injury.
■ If moisture is flowed into the ventilation window, it may cause product
CAUTION malfunctions. Make sure to use dry tools for cleaning.

Ventilation Window

[Figure 1.7 Ventilation Window]


Chapter 1. Safety and Maintenance 1 - 31

1.5. Biological Safety


For safety information on the probes and biopsies, please refer to Chapter 9 ‘Probes’ in this
manual.

■ Ultrasound waves may have damaging effects on cells and, therefore, may
be harmful to the patient. If there is no medical benefit, minimize the
exposure time and maintain the ultrasound wave output level at a low setting.
Please refer to the ALARA principle.
WARNING ■ Do not use the product if an error message appears on the screen indicating
that a hazardous condition exists. Note down the message displayed on the
screen, turn off the power, and then contact Samsung Medison’s Customer
Service Department.
■ Do not use a product that exhibits unexpected or abnormal operation.
Discontinuities in the scanning sequence are indicative of a hardware failure
that should be corrected before use.
■ This product limits the temperature that patients are allowed to come into
contact with to 43°C or less. The ultrasound power output (AP&I) is in
compliance with the US FDA’s standards.

1.5.1. ALARA Principle


Guidance for the use of diagnostic ultrasound is defined by the “As Low As Reasonably
Achievable” (ALARA) principle. The decision as to what is reasonable has been left to the
judgment and insight of qualified personnel. No set of rules can be formulated that would be
sufficiently complete to dictate the correct response to every circumstance. By keeping
ultrasound exposure as low as possible, while obtaining diagnostic images, users can minimize
ultrasonic bioeffects.
Since the threshold for diagnostic ultrasound bioeffects is undetermined, it is the sonographer’s
responsibility to control the total energy transmitted into the patient. The sonographer must
reconcile exposure time with diagnostic image quality. To ensure diagnostic image quality and
limit exposure time, the Ultrasound System provides controls that can be manipulated during the
exam to optimize the results.
The ability of the user to abide by the ALARA principle is important. Advances in diagnostic
ultrasound, not only in the technology but also in its applications, have resulted in the need for
increased and improved information to guide the user. The output indices are designed to
provide that important information.
There are a number of variables, which affect the way in which the output display indices can be
used to implement the ALARA principle. These variables include mass, body size, location of
the bone relative to the focal point, attenuation in the body, and ultrasound exposure time.
Exposure time is an especially useful variable, because the user controls it. The ability to limit
the index values over time supports the ALARA principle.

1.5.1.1. Applying ALARA


The system imaging mode used depends upon the information needed. 2D mode and M
mode imaging provide anatomical information, while Doppler, Power, and Color imaging
provide information about blood flow. Scanned modes, like 2D mode, Power, or Color,
disperse or scatter the ultrasonic energy over an area, while an unscanned mode, like M
mode or Doppler, concentrate the ultrasonic energy. Understanding the nature of the imaging
1 - 32 V6/XV6 Service Manual

mode being used allows the sonographer to apply the ALARA principle with informed
judgment. The probe frequency, system set-up values, scanning techniques, and operator
experience aid the sonographer in meeting the ALARA principle. The decision as to the
amount of acoustic output is, in the final analysis, up to the system operator. This decision
must be based on the following factors: type of patient, type of exam, patient history, ease or
difficulty of obtaining diagnostically useful information, and the potential localized heating of
the patient due to probe surface temperatures. Prudent use of the system occurs when
patient exposure is limited to the lowest index reading for the shortest amount of time
necessary to achieve acceptable diagnostic results.
Although a high index reading does not mean that a biological effect is actually occurring, it
should be taken seriously. Every effort should be made to reduce the possible effects of a
high index reading. Limiting exposure time is an effective way to accomplish this goal.
There are several system controls that the operator can use to adjust the image quality and
limit the acoustic intensity. These controls are related to the techniques that an operator
might use to implement ALARA and can be divided into three categories: direct, indirect, and
receiver control.

1.5.1.2. Direct Controls


Application selection and the output intensity control directly affect acoustic intensity. There
are different ranges of allowable intensity or output depending on your selection. Selecting
the correct range of acoustic intensity for the application is one of the first things required
during any exam. For example, peripheral vascular intensity levels are not recommended for
fetal exams. Some systems automatically select the proper range for a particular procedure,
while others require manual selection. Ultimately, the user bears the responsibility for proper
clinical use. Samsung Medison systems provide both automatic and user-definable settings.
Output has direct impact on acoustic intensity. Once the application has been established,
the output control can be used to increase or decrease the output intensity. The output
control allows you to select intensity levels less than the defined maximum. Prudent use
dictates that you select the lowest output intensity consistent with good image quality.

1.5.1.3. Indirect Controls


The indirect controls are those that have an indirect effect on the acoustic intensity. These
controls affect the imaging mode, pulse repetition frequency, focus depth, pulse length, and
probe selection.
The choice of imaging mode determines the nature of the ultrasound beam. 2D mode is a
scanning mode, Doppler is a stationary or unscanned mode. A stationary ultrasound beam
concentrates energy on a single location. A moving or scanned ultrasound beam disperses
the energy over a wide area and the beam is only concentrated on a given area for a fraction
of the time necessary in unscanned mode.
The pulse repetition frequency or rate refers to the number of ultrasound bursts of energy
over a specific period of time. The higher the pulse repetition frequency, the more pulses of
energy in a given period of time. Several controls affect pulse repetition frequency: focal
depth, display depth, sample volume depth, color sensitivity, number of focal zones, sector
width controls, etc.
The focus of the ultrasound beam affects the image resolution. To maintain or increase
resolution at a different focus requires a variation of output over the focal zone. This variation
of output is a function of system optimization. Different exams require different focal depths.
Setting the focus to the proper depth improves the resolution of the structure of interest.
Chapter 1. Safety and Maintenance 1 - 33

Pulse length is the time during which the ultrasonic burst is turned on. The longer the pulse,
the greater the time-average intensity value. The greater the time-average intensity, the
greater the likelihood of temperature increase and cavitations. Pulse length, burst length or
pulse duration is the output pulse duration in Pulsed Doppler. Increasing the Doppler sample
volume, increases the pulse length.
Probe selection affects intensity indirectly. Tissue attenuation changes with frequency. The
higher the probe operating frequency, the greater the attenuation of the ultrasonic energy.
Higher probe operating frequencies require higher output intensity to scan at an increased
depth. To scan deeper at the same output intensity, a lower probe frequency is required.
Using more gain and output beyond a point, without corresponding increases in image
quality, can mean that a lower frequency probe is needed.

■ Receiver Controls
Receiver controls are used by the operator to improve image quality. These controls have
no effect on output. Receiver controls only affect how the ultrasound echo is received.
These controls include gain, TGC, dynamic range, and image processing. The important
thing to remember, relative to output, is that receiver controls should be optimized before
increasing output. For example; before increasing output, optimize gain to improve image
quality.

1.5.1.4. Additional Considerations


Ensure that scanning time is kept to a minimum, and ensure that only medically required
scanning is performed. Never compromise quality by rushing through an exam. A poor exam
will require a follow-up, which ultimately increases the scanning time. Diagnostic ultrasound
is an important tool in medicine, and, like any tool, should be used efficiently and effectively.

1.5.1.5. Output Display Features


The system output display comprises two basic indices: a mechanical index and a thermal
index. The thermal index consists of the following indices: soft tissue (TIs), bone (TIb) and
cranial bone (TIc). One of these three thermal indices will be displayed at all times. The
thermal index to be displayed is determined by the system preset or user choice, depending
upon the application.
The mechanical index is continuously displayed over the range of 0.0 to 1.9, in increments of
0.1. The thermal index consists of the three indices, and only one of these is displayed all the
time. Each diagnostic probe application has a default selection. The TIb or TIs is
continuously displayed over its range of 0.0 to maximum output, based on the probe and
application, in increments of 0.1.
The default setting of the application-specific nature is also an important factor of index
selection. A default setting is a system control state which is preset by the manufacturer or
the operator. The system has default index settings for the probe application. The default
settings are applied automatically by the ultrasound system when the power is turned on,
new patient data is entered into the system database, or a change of application takes place.
The decision as to which of the three thermal indices to display should be based on the
following criteria:
Appropriate index for the application: TIs is used for imaging soft tissue, and TIb for a focus
at or near a bone. Some factors might create artificially high or low thermal index readings
(e.g. presence of fluid or bone, or the flow of blood). A highly attenuating tissue path, for
1 - 34 V6/XV6 Service Manual

example, may cause the potential for local zone heating to be lower than the thermal index
indicates.
The selection of scanned modes or unscanned modes of operation also affect the thermal
index. For scanned modes, heating tends to be near the surface; for unscanned modes, the
potential for heating tends to be deeper in the focal zone.
Always limit ultrasound exposure time. Do not rush the scanning. Ensure that the indices are
kept to a minimum, and that exposure time is limited without compromising diagnostic
sensitivity.

■ Mechanical Index (MI) Display


Mechanical bioeffects are threshold phenomena that occur when a certain level of output
is exceeded. The threshold level varies, however, with the type of tissue. The potential for
mechanical bioeffects varies with peak pressure and ultrasound frequency. The MI
accounts for these two factors. The higher the MI value, the greater the likelihood of
mechanical bioeffects occurring. However, there is no specific MI value that means that a
mechanical bioeffect will actually occur. The MI should be used as a guide for
implementing the ALARA principle.

■ Thermal Index (TI) Display


The TI informs the user of potential temperature increase occurring on the body surface,
within body tissue, or at the point of focus of the ultrasound beam on bone. The TI is an
estimate of the temperature increase in specific body tissues. The actual amount of any
temperature rise is influenced by factors such as tissue type, vascularity, and mode of
operation. The TI should be used as a guide for implementing the ALARA principle.
The bone thermal index (TIb) informs the user about potential heating at or near the focus
after the ultrasound beam has passed through soft tissue or fluid, such as the skeletal
structure of a 2-3 month old fetus. The cranial bone thermal index (TIc) informs the user
about the potential heating of bone at or near the surface, for example, the cranial bone.
The soft tissue thermal index (TIs) informs the user about the potential for heating within
soft homogeneous tissue. TIc is displayed when you select a trans-cranial application.
You can select a TI to display in Setup > Imaging > Preset > Details > Display Thermal
Index.

■ Mechanical and Thermal Indices Display Precision and Accuracy


The Mechanical and Thermal Indices on the system are precise to 0.1 units.
The MI and TI display accuracy estimates for the system are given in the manual's
Acoustic Output Tables. These accuracy estimates are based on the variability ranges of
probes and systems, inherent acoustic output modeling errors, and the measurement
variability, as described below.
The displayed values should be interpreted as relative information to help the system
operator achieve the ALARA principle through prudent use of the system. The values
should not be interpreted as actual physical values of investigated tissue or organs. The
initial data that is used to support the output display is derived from laboratory
measurements based on the AIUM measurement standard. The measurements are then
put into algorithms to calculate the displayed output values.
Many of the assumptions used in the process of measurement and calculation are
conservative in nature. Over-estimation of actual in situ exposure, for the vast majority of
tissue paths, is built into the measurement and calculation process. For example, the
Chapter 1. Safety and Maintenance 1 - 35

acoustic output values measured underwater are derated using a conservative, industry
standard, attenuation coefficient of 0.3dB/cm-MHz.
Conservative values for tissue characteristics were selected for use in the TI models.
Conservative values for tissue or bone absorption rates, blood perfusion rates, blood heat
capacity, and tissue thermal conductivity were selected.
A steady state temperature rise is assumed in the industry standard TI models, and the
assumption is made that the ultrasound probe is held steady in one position long enough
for a steady state to be reached.
A number of factors are considered when estimating the accuracy of display values:
Hardware deviation, algorithm accuracy, and measurement deviation. Deviation among
probes and systems in particular is an important factor. Probe deviation results from
piezoelectric crystal efficiencies, process-related impedance differences, and sensitive
lens focusing parameter variations. Differences in the system pulser voltage control and
efficiencies are also a contributor to variability. There are inherent uncertainties in the
algorithms used for estimating acoustic output values over the range of possible system
operating conditions and pulser voltages. Inaccuracies in laboratory measurements are
related to differences in hydrophone calibration and performance, positioning, alignment
and digitization tolerances, and variability among test operators.
The conservative assumptions of the output estimation algorithms of linear propagation,
at all depths, through a 0.3dB/cm-MHz attenuated medium are not taken into account in
the calculation of the accuracy estimate displayed. Neither linear propagation nor uniform
attenuation at the 0.3dB/cm-MHz rate occurs in underwater measurements, or in most
tissue paths in the body. In the body, different tissues and organs have dissimilar
attenuation characteristics. In water, there is almost no attenuation. In the body, and
particularly in underwater measurements, non-linear propagation and saturation losses
occur as pulser voltages increase.
The display accuracy estimates take into account the variability ranges of probes and
systems, inherent acoustic output modeling errors, and the measurement variability.
Display accuracy estimates are measured according to AIUM measurement standards
but not based on errors caused during the measurement or inherent errors. They are also
independent of the effects of non-linear loss on the measured values.

1.5.1.6. Control Effects – Controls Affecting the Indices


As various system controls are adjusted, the TI and MI values may change. This will be most
apparent as the Power control is adjusted; however, other system controls will also affect the
on-screen output values.

■ Power
Power controls the system acoustic output. Two real-time output values are on the
screen: a TI and a MI. They change as the system responds to Power adjustments.
In combined modes, such as simultaneous Color, 2D mode, and Pulsed Doppler, the total
TI is the sum of TIs of the individual modes. Each mode is a vital contributor to this total;
the displayed MI will be from the mode with the largest peak pressure.
1 - 36 V6/XV6 Service Manual

1.5.1.7. 2D Mode Controls

■ 2D Mode Size
Narrowing the sector angle may increase the frame rate. This will increase the TI. The
pulser voltage may be automatically adjusted down with software controls to keep the TI
below the system maximum. A decrease in pulser voltage will decrease MI.

■ Zoom
Increasing the zoom magnification may increase frame rate. This will increase the TI. The
number of focal zones may also increase automatically to improve the resolution. This
action may change the MI, since the peak intensity can occur at a different depth.

■ Number of Focal Zones


Increasing the number of focal zones may change both the TI and MI by automatically
changing the frame rate or focal depth. Lower frame rates decrease the TI. The MI
displayed will correspond to the focal zone with the largest peak intensity.

■ Focus
In general, higher MI values will occur when the focal depth is near the natural focus of
the probe (transducer).

1.5.1.8. Color and Power Controls

■ Color Sensitivity
Increasing the color sensitivity may increase the TI. More time is required for the
scanning of color images. Color pulses are the dominant pulse type in this mode.

■ Color Sector Width


Narrower color sector width will increase the color frame rate, and so the TI will increase.
The system may automatically decrease the pulser voltage to stay below the system
maximum. A decrease in pulser voltage will decrease the MI. If Pulsed Doppler is also
enabled, then Pulsed Doppler will remain the dominant mode and the TI change will be
small.

■ Color Sector Depth


Deeper color sector depth may automatically decrease color frame rate, or select a new
color focal zone or color pulse length. The TI will change due to the combination of these
effects. Generally, the TI will decrease with increased color sector depth. MI will
correspond to the peak intensity of the dominant pulse type, which is a color pulse.
However, if Pulsed Doppler is also enabled, then Pulsed Doppler will remain the
dominant mode and the TI change will be small.

■ Scale
Using the Scale control to increase the color velocity range may increase the TI. The
system will automatically adjust the pulser voltage to stay below the system maximum. A
decrease in pulser voltage will also decrease MI.
Chapter 1. Safety and Maintenance 1 - 37

■ 2D Mode Size
A narrower 2D mode sector width in Color imaging will increase color frame rate. The TI
will increase. MI will not change. If Pulsed Doppler is also enabled, then Pulsed Doppler
will remain the dominant mode and the TI change will be small.

1.5.1.9. M Mode and Doppler Controls

■ Simultaneous and Update Methods


Use of combination modes affects both the TI and MI through the combination of pulse
types. During Simultaneous mode, the TI is an ancillary element. During auto-update and
duplex, the TI will display the dominant pulse type. The displayed MI will be from the
mode with the largest peak pressure.

■ Sample Volume Depth


When Doppler sample volume depth is increased, the Doppler PRF may automatically
decrease. A decrease in PRF will decrease the TI. The system may also decrease the
pulser voltage to remain below the system maximum. A decrease in pulser voltage will
decrease MI.

1.5.1.10. Other

■ 2D, Color, M Mode, PW and CW


When a new imaging mode is selected, both the TI and the MI will change to their default
settings. Each mode has a corresponding pulse repetition frequency and maximum
intensity point. In combined or simultaneous modes, the TI is the sum of the contribution
from the modes enabled, and the MI is the value for the focal zone of the mode with the
largest derated intensity. If a mode is turned off and then reselected, the system will
return to the previously selected settings.

■ Probe
Each probe model available has unique specifications for the contact area, beam shape,
and center frequency. Defaults are initialized when you select a probe. Samsung
Medison’s manufacturer default settings vary, depending on the probe, application, and
mode selected. Defaults have been chosen below the FDA limits for intended use.

■ Depth
An increase in the 2D mode depth will automatically decrease the 2D mode frame rate.
This would decrease the TI. The system may also automatically choose a deeper 2D
mode focal depth. A change of focal depth may change the MI. The MI displayed is that
of the zone with the largest peak intensity.

■ Application
Acoustic output defaults are set when you select an application. Samsung Medison’s
manufacturer default settings vary, depending on the probe, application, and mode
selected. Defaults have been chosen below the FDA limits for intended use.
1 - 38 V6/XV6 Service Manual

1.5.1.11. Related Guidance Documents


The terms and symbols used in the acoustic output tables are defined in the following
paragraphs.
– Medical Ultrasound Safety (AIUM, 2014).
– AIUM Consensus Report on Potential Bioeffects of Diagnostic Ultrasound: Executive
Summary, J. Ultrasound in Medicine, 2008, Vol. 27, Num. 4.
– WFUMB. Symposium on Safety of Ultrasound in Medicine: Conclusions and
Recommendations on Thermal and Non-thermal Mechanisms for Biological Effects.
Ultrasound in Med. & Biol; 1998, 24: Supplement 1.
– Bioeffects and Safety of Diagnostic Ultrasound (AIUM, 1993)
– Guidelines for the safe use of diagnostic ultrasound equipment. (BMUS, 2009)
– Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound
Systems and Transducers (U.S. FDA - 2008)
– Particular requirements for the basic safety and essential performance of ultrasonic
medical diagnostic and monitoring equipment. (IEC, 2007)
– Acoustic Output Labeling Standard for Diagnostic Ultrasound Equipment (AIUM, 2008)
– Standard Means for the Reporting of the Acoustic Output of Medical Diagnostic Ultrasonic
Equipment. (IEC, 2007)
– Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices On
Diagnostic Ultrasound Equipment (AIUM/NEMA, 2004)
– Ultrasonics - Field characterization - Test methods for the determination of thermal and
mechanical indices related to medical diagnostic ultrasonic fields (IEC, 2005)
– Measurement and Characterization of Medical Ultrasonic Fields up to 40 MHz. (IEC,
2007)
– Ultrasonics-Power Measurements - Radiation Force Balances and Performance
Requirements. (IEC, 2006)
– Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment.
(AIUM/NEMA, 2004).

1.5.1.12. Acoustic Output and Measurement


Since the first usage of diagnostic ultrasound, the possible human biological effects
(bioeffects) of ultrasound exposure have been studied by various scientific and medical
institutions. In October 1987, the American Institute of Ultrasound in Medicine (AIUM) ratified
a report prepared by its Bioeffects Committee (Bioeffect Considerations for the Safety of
Diagnostic Ultrasound, J Ultrasound Med., Sept. 1988: Vol.7, No.9 Supplement), sometimes
referred to as the Stowe Report, which reviewed available data on possible effects of
ultrasound exposure. Another report, “Bioeffects and Safety of Diagnostic Ultrasound”, dated
January 28, 1993, provides more up to date information. In addition, periodically updated
reports on biological effects, results, and guidelines on safe usage have been published by
groups such as the WFUMB (World Federation of Ultrasound in Medicine and Biology),
AIUM, and BMUS.
The Acoustic output for this system has been measured and calculated in accordance with
the Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices On
Chapter 1. Safety and Maintenance 1 - 39

Diagnostic Ultrasound Equipment (AIUM/NEMA, 2004) and Acoustic Output Measurement


Standard for Diagnostic Ultrasound Equipment (AIUM/NEMA, 2004).

1.5.1.13. In Situ, Derated, and Water Value Intensities


All intensity parameters are measured in water. Since water does not absorb acoustic energy,
these water measurements represent the largest possible value. Biological tissue absorbs
acoustic energy. The true value of the intensity at any point depends on the amount and type
of tissue, and the frequency of the ultrasound that passes through the tissue. The intensity
value in the tissue, In Situ, has been estimated using the following formula:
 ( 0.23 alf )
In Situ = Water [ e ]
Where: In Situ = In Situ Intensity Value
Water = Water Value Intensity
e = 2.7183
a = Attenuation Factor
Tissue a (dB/cm-MHz)
Brain 0.53
Heart 0.66
Kidney 0.79
Liver 0.43
Muscle 0.55
l = Skin line to measurement depth (cm)
f = Center frequency of the transducer/system/mode combination (MHz)
Since the ultrasonic path during an examination is likely to pass through varying lengths and
types of tissue, it is difficult to estimate the true In Situ intensity. An attenuation factor of 0.3
is used for general reporting purposes; therefore, the In Situ value is commonly reported
using this formula:
 ( 0.069 lf )
In Situ (derated) = Water [ e ]
Since this value is not the true in situ intensity, the term “derated” is used.
The maximum derated and maximum water values do not always occur under the same
operating conditions. Therefore, the reported maximum water and derated values may not
be related to the In Situ (derated) formula. For example, a multi-zone array transducer has
the greatest water value intensities at its deepest zone. The same transducer may have its
largest derated intensity at one of its shallowest focal zones.

1.5.1.14. Terms and Symbols Related to Acoustic Output and Measurement


The terms and symbols used in the acoustic output tables are defined in the following
paragraphs.

fawf ACOUSTIC WORKING FREQUENCY is the frequency of an acoustic signal


based on the observation of the output of a hydrophone placed in an
acoustic field at the position corresponding to the spatial-peak temporal-
1 - 40 V6/XV6 Service Manual

peak acoustic pressure. (unit: MHz)

Ipa, at zpii,α The ATTENUATED PULSE-AVERAGE INTENSITY at the depth of


maximum ATTENUATED SPATIAL-PEAK TEMPORAL-AVERAGE
INTENSITY. (unit: Watts per square centimeter)

Ita (z) The TEMPORAL-AVERAGE INTENSITY at axial distance z. (unit: Milliwatts


per square centimeter)

Ita,  (z) The ATTENUATED TEMPORAL-AVERAGE INTENSITY at axial distance z.


(unit: Milliwatts per square centimeter)

MI MECHANICAL INDEX is a displayed parameter representing potential


cavitation bioeffects. (unit: N/A)

npps The number of acoustic pulses travelling along a particular ULTRASONIC


SCAN LINE.

P OUTPUT POWER is time-average ultrasonic power emitted by an ultrasonic


transducer into an approximately free field under specified conditions in a
specified medium, preferably water. (unit: mW)

P1x1 Maximum value of the time average acoustic output power emitted from any
one-centimetre square region of the active area of the transducer, the one-
centimetre square region having 1 cm dimensions in the x- and y-directions.

Pr at zpii The PEAK-RAREFACTIONAL ACOUSTIC PRESSURE at the depth of


maximum SPATIAL-PEAK TEMPORAL-AVERAGE INTENSITY. (unit:
Megapascals)

Pr,α at zMI The ATTENUATED PEAK-RAREFACTIONAL ACOUSTIC PRESSURE


associated with the transmit pattern giving rise to the reported value of MI.
(unit: Megapascals)

pii Time integral of the instantaneous intensity at a particular point in an


acoustic field integated over the acoustic pulse waveform. (unit: Watts per
square centimeter)

pii α (Z) Value of the pulse-intensity integral after attenuation, at a specified distance
from the external transducer aperture. (unit: Watts per square centimeter)

prr PULSE REPETITION RATE is reciprocal of time interval between


equivalent points on successive pulses or tone-bursts. (unit: Hz)

srr SCAN REPETITION RATE is reciprocal of time interval between identical


points on two successive frames, sectors or scans, applying to automatic
scanning systems with a periodic scan sequence only. (unit: Pulses per
second or Hz)
Chapter 1. Safety and Maintenance 1 - 41

TIB BONE THERMAL INDEX is a themal index for applications such as foetal
(second and third trimester), in which the ultrasound beam passes through
soft tissue and a focal region is in the immediate vicinity of bone. (unit: N/A)

TIC CRANIAL BONE THERMAL INDEX is a thermal index for applications, in


which the ultrasound beam passes through bone near the beam entrance
into the body, such as paediatric and adult cranial or neonatal cephalic
applications. (unit: N/A)

TIS SOFT TISSUE THERMAL INDEX is a thermal index related to soft tissues.
(unit: N/A)

zb,ns DEPTH FOR BONE THERMAL INDEX for non-scanning modes, the
distance along the beam axis from the external transducer aperture to the
plane where the product of attenuated output power and attenuated spatial-
peak temporal-average intensity is a maximum over the distance range
equal to, or greater than, the break-point depth, zbp. (unit: cm)

zMI Depth on the BEAM-AXIS from the EXTERNAL TRANSDUCER


APERTURE to the plane of maximum ATTENUATED PULSE INTENSITY
INTEGRAL (pii,α). (unit: cm)

zs,ns DEPTH FOR SOFT-TISSUE THERMAL INDEX for non-scanning modes,


the distance along the beam axis from the external transducer aperture to
the plane at which the lower value of the attenuated output power and the
product of the attenuated spatial-peak temporal-average intensity and 1 cm 2
is maximized over the distance range equal to, or greater than, the break-
point depth, zbp. (unit: cm)

zpii,α Position of maximum ATTENUATED SPATIAL-PEAK TEMPORAL-


AVERAGE INTENSITY for NONSCANNING MODE components,
determined beyond the BREAK-POINT DEPTH, zbp, on the BEAM-AXIS.
(unit: cm)

1.5.1.15. Acoustic Measurement Precision and Uncertainty


The Acoustic Measurement Precision and Acoustic Measurement Uncertainty are described
below.

Quantity Precision Total Uncertainty

pii  (attenuated pulse intensity integral) 3.2% +21% to -24%

P (acoustic power) 6.2% ±19%

Pr,α (attenuated rarefaction pressure) 5.4% ±15%

fawf (acoustic working frequency) < 1% ±4.5%

■ Systematic Uncertainties
1 - 42 V6/XV6 Service Manual

For the pulse intensity integral, attenuated rarefaction pressure, center frequency, and
pulse duration, the analysis includes considerations of the effects on accuracy of:
Hydrophone calibration drift or errors.
Hydrophone/Amp frequency response.
Spatial averaging.
Alignment errors.
Voltage measurement accuracy, including.
– Oscilloscope vertical accuracy.
– Oscilloscope offset accuracy.
– Oscilloscope clock accuracy.
– Oscilloscope Digitization rates.
– Noise.
The acoustic power is measured using a radiation force for systematic uncertainties
through the use of calibrated NIST acoustic power sources. We also refer to a September
1993 analysis conducted by the working group of the IEC technical committee 87 and
prepared by K. Beissner, as a first supplement to IEC publication 1161. The document
includes analysis and discussion of the sources of error/measurement effects due to:
– Balance system calibration.
– Absorbing (or reflecting) target suspension mechanisms.
– Linearity of the balance system.
– Extrapolation to the moment of switching the ultrasonic transducer (compensation for
ringing and thermal drift).
– Target imperfections.
– Absorbing (reflecting) target geometry and finite target size.
– Target misalignment.
– Ultrasonic transducer misalignment.
– Water temperature.
– Ultrasonic attenuation and acoustic streaming.
– Coupling or shielding foil properties.
– Plane-wave assumption.
– Environmental influences.
– Excitation voltage measurement.
– Ultrasonic transducer temperature.
– Effects due to nonlinear propagation and saturation loss.
Chapter 1. Safety and Maintenance 1 - 43

1.5.1.16. Training
The users of this ultrasound system must familiarize themselves with the ultrasound system
to optimize the performance of the device and to detect possible malfunctions. It is
recommended that all users receive proper training before using the device. You can receive
training on the use of the product from the Samsung Medison service department, or any of
the customer support centers worldwide.
1 - 44 V6/XV6 Service Manual

1.6. Maintenance
1.6.1. Cleaning and Frequency
Cycle Component Cleaning Method

Clean, disinfect, and sterilize according to


“ Cleaning disinfecting, and sterilizing probes” in
After patient • Probe lens and Case
Chapter9.
exmaination • Biopsy Bracket
- Biopsy Accessories must be cleaned and
disinfected or discarded for each patient.

Clean according to Chapter 1, “Safety and


• Control Panel
Maintenance”.
• Touch Panel
Daily - Cross-contamination may occur, so please check
• Probe Holder
daily.
• Front and rear Handle
(Touch screen : Only model with it)

• Console (Body) Clean according to Chapter 1, “Safety and


• Storage Accessories Maintenance”.
• Trackball - Trackball : rotate the Rim counterclock-wise and
Weekly • Air Filter separate it, and then wipe it with a soft cloth.
• Foot Switch
• B/W Printer - Air filter : Dust off with a soft brush or a vaccum
• Monitor젤 cleaner.

• Cable and Connector


• Probe Switch Box Clean according to Chapter 1, “Safety and
• Power cable Maintenance” .
• Power cable of Clean, disinfect, and sterilize according to
Monthly
peripheral
“ Cleaning disinfecting, and sterilizing probes” in
• Control panel keyboard
• Caster Chapter9.
• External monitor

Since a pH neutral soap solution does not contain strong chemical ingredients, it
will not irritate the skin. The solution must not contain any fragrances, oil, or
alcohol.
NOTE

■ Make sure that the touchscreen is not contaminated by an electrically


conductive substance.
■ Contamination of the LCD screen by ultrasound gel or other substances may
degrade the sensitivity of the touchscreen and cause malfunctions. This
CAUTION phenomenon is caused by the contact between the capacitive touchscreen
and the conductive substance. If you find foreign substances on the screen,
clean it with a dry cloth, and then restart the system.
Chapter 1. Safety and Maintenance 1 - 45

1.6.2. Cleaning and Disinfection


The ultrasound system, probes, and other components that are used with the product may
come in direct contact with the user and patient. Hence, to prevent infections, clean and
disinfect the product in accordance with the procedures specified in this user manual or the
manufacturer's user manual that comes with each product.

■ Always wear a protective face mask and gloves when cleaning and
disinfecting the product.
■ Turn off the system and disconnect the power cord from the wall outlet before
cleaning and disinfecting. Otherwise, there is a risk of electric shock or fire.
WARNING
■ Do not clean the electrical contacts or connectors.
■ Clean and disinfect the product every day or after each exam.

To learn how to clean and disinfect probes, please refer to Chapter 9 ‘Probes’ in
this manual.
NOTE

1.6.2.1. Information on Detergents and Disinfectants


The device care method determines the appropriate disinfectant for your device. An
appropriate detergent or disinfectant should be used for all products. For the latest details
about compatible detergents and disinfectants, please see the user guide and the Samsung
Medison website.
– User Guide: This is provided as a booklet upon purchase of the product.
– Website: http://www.samsunghealthcare.com

■ Using an inappropriate disinfection procedure may damage the system.


Make sure that you check the disinfectant’s expiry date.
■ Mix the disinfectant solution to the strength specified on the instruction of the
disinfectant manufacturer.
CAUTION ■ When using a disinfectant, wipe it carefully with a soft cloth.
■ If the disinfectant solution remains on the product, it may cause the product
to malfunction. After using the disinfectant on the product, please be sure to
wipe away any remaining disinfectant with a soft cloth before using the
product.
1 - 46 V6/XV6 Service Manual

1.6.2.2. Cleaning
The device must be cleaned according to this instruction. Cleaning is an important procedure
that must be carried out before disinfection.
1. Turn off the system and disconnect the power cord from the wall outlet.
2. Wipe off any gel or biological material with a soft, lint-free cloth moistened with detergent
solution (made using manufacturer’s instructions) or tap water, taking care to remove all
visible contamination.
Dispose of the used cloths appropriately according to the local medical waste regulations.
3. For a list of compatible cleaning or disinfectant agents, see the user guide or
‘Disinfectants for System Surfaces, Disinfectants for Monitor Surface’ on the Samsung
Medison website (http://www.samsunghealthcare.com).
Follow the detergent manufacturer’s instructions for preparation, temperature, solution
strength, and duration of contact. If a premixed solution is used, be sure to observe the
solution expiration date.
4. Wipe off the device’s surface according to the disinfectant manufacturer’s instructions for
temperature, wipe durations, and duration of disinfectant contact. For difficult parts to
wipe such as crevices and edges of the device, use a cotton swab wrapped with a cloth
soaked with the detergent. Make sure that liquids or other objects do not enter the inside
of the product.
5. Visually inspect for any remaining soil and if necessary repeat the steps, starting at step 4.
6. According to the detergent manufacturer’s instruction, air dry the device or remove water
from all surface of the device using a soft and clean disposable, lint-free cloth.
7. Examine the device for damage such as cracks, splitting, sharp edges, or projections. If
such damage is found, discontinue use of the device and contact your Samsung Medison
representative.

■ When you use a detergent, follow the recommendations of the detergent


manufacturer.
■ Do not spray detergent directly onto the product’s exterior. Doing so may
discolor or crack the surface.
CAUTION ■ Repeated use of disinfectants may cause discoloration.
■ When using a spray cleaner, do not spray it directly onto the product
surface to prevent liquid from entering the product.
■ Do not use chemicals such as wax, benzene, alcohol, paint thinner,
insecticide, air freshener, or lubricant.
■ Do not use glass cleaners or products containing bleach.
■ Make sure that liquids or other objects do not enter the inside of the
product.
■ Do not use any liquid containing glycol components such as Propylene
glycol, Butylene glycol, or Polyethylene glycol, including cosmetics, on the
external surface of the product. Doing so may discolor or crack the
external surface of the product.
Chapter 1. Safety and Maintenance 1 - 47

1.6.2.3. Disinfection
In general, an ultrasound system has an irregular surface, and disinfection of the entire
surface of the product is not recommended. Therefore, users should follow the instructions of
reprocessing method suggested.

 When you use a disinfectant, follow the recommendations of the disinfectant


manufacturer.
 Use only the disinfectants recommended in the product user manual to disinfect
the system’s surfaces.
CAUTION

1. Clean the device according to the procedures in ‘Cleaning’ of this chapter.


2. After cleaning, choose the disinfectant that is compatible with your device. For a list of
compatible disinfectant agents, see the user guide or ‘Disinfectants for System Surfaces,
Disinfectants for Monitor Surface’ on the Samsung Medison website
(http://www.samsunghealthcare.com).
Follow disinfectant manufacturer’s instructions for preparation, temperature, solution
strength, and duration of contact. If a premixed solution is used, be sure to observe the
solution expiration date.
3. Wipe off the device’s surface according to the disinfectant manufacturer’s instructions for
temperature, wipe durations, and duration of disinfectant contact. For difficult parts to
wipe such as crevices and edges of the device, use a sterile cotton swab wrapped with a
sterile cloth soaked with the disinfectant. Make sure that liquids or other objects do not
enter the inside of the product.
4. Wipe off the device surface using a sterile cloth soaked with pure water according to the
disinfectant manufacturer’s rinsing instruction.
5. According to the disinfectant manufacturer’s instruction, air dry the device or remove
water from all surface of the device using a sterile, lint-free cloth.
6. Examine the device for damage such as cracks, splitting, sharp edges, or projections. If
such damage is found, discontinue use of the device and contact your Samsung Medison
representative.
1 - 48 V6/XV6 Service Manual

1.7. Information Maintenance

You may lose user settings or patient information files because of physical
shocks to the product or internal errors. Therefore, you should back up this
information on a regular basis.
CAUTION

1.7.1. User Settings Backup


Always keep a backup copy of all information related to the user settings in case of data loss.
Users should not attempt to back up the product’s User Settings themselves. A Samsung
Medison Customer Service representative should perform all backups upon customer request.

1.7.2. Patient Information Backup


The SonoView program can be used to back up patients’ basic information and scanned images.
They can be saved by users, but only to external media. For more information, please refer to
‘Starting Modes of Operation’. If the system needs to be reinstalled because of an issue with the
product, then a Samsung Medison Customer Service representative should restore the basic
information and scanned images of patients saved on the system.

1.7.3. Software
The software may be changed to improve the product’s performance. Users should not attempt
to modify the software by themselves; they must consult staff at Customer Service Department
and ask them to guide them through the modification process.

Minor software updates may be carried out without prior notice from the
manufacturer.
CAUTION

If an error occurs in the operating system (Windows), or you need to upgrade it, please follow
the operating system manufacturer’s instructions.

This product uses the Windows firewall to prevent any hackers or malicious
software from accessing the system through the internet or network.
NOTE
Chapter 1. Safety and Maintenance 1 - 49

1.7.4. Network

1.7.4.1. Wireless LAN (WLAN) Specifications


IEEE 802.11ac, IEEE 802.11a, IEEE 802.11n, IEEE 802.11g, IEEE
Wireless Standards
802.11b
Wireless Security 64/128 bit WEP, WPA-PSK/WPA2-PSK, WPA2-Enterprise
Frequency 5GHz, 2.4GHz
Certification KC, CE, FCC, RoHS
Modulation Technology DBPSK, DQPSK, CCK, 16-QAM, 64-QAM, 256-QAM

1.7.4.2. Protocols for Interhost Communication


– DNS – Only used as client.
– DHCP – Only used as client.

1.7.4.3. Required Network Characteristics


53 DNS Client
68 DHCP Client
Open port for DICOM (configurable on the system) Image file transfer (Minimum 100
104
Mbit/sec, recommended 1Gbit/sec)
135 System Log (Only used as log source)

1.7.4.4. Potential Hazardous Situations resulting from network security


vulnerabilities

■ The following situations may occur in connection of ultrasound system to a less


secure network.
– Delayed or impaired access to scan image
– Delayed or impaired access to patient data or exam information
– Temporary damage or permanent loss of all the data stored

■ Problems may occur in changes to network environment such as those listed


below, which require consistent analysis and risk management (identification,
analysis, evaluation and control).
– Changes in network configuration (add or disconnect items)
– Update or upgrade of items connected to network
1 - 50 V6/XV6 Service Manual

1.7.5. Cybersecurity
In recent years, cybersecurity threats have intensified.
Threats manifest in a variety of forms, including system breakdowns caused by virus infections
or leaking of patient data stored on the system, and unfortunately, no security measure can
prevent such threats 100%. Therefore, please comply with the following to prepare for cyber
threats.
■ Do not connect the product to an unsecured public wired or wireless network, and under no
circumstances do you connect to the Internet via this product.
■ To start functions that require network connection (Export, Remote Control etc.), use firewall
to restrict access to the network connected with product.
■ Place this product in an area where it cannot be operated or tampered with physically by
non-designated persons, and shut off power when it is not in use.
■ Only external devices (e.g., USB drive) that have been scanned for viruses and malwares
should be connected to the product.
■ If you suspect that the product is infected, stop using it immediately, shut off the power, and
make sure that you contact the Security Manager (Technical Support).
■ If a new security patch has been released, adjust the schedule with the Service Team to
apply the patch promptly.
■ Samsung Medison provides remediation support for known cybersecurity vulnerabilities of its
products until the end of their lifecycle.
Chapter 1. Safety and Maintenance 1 - 51

1.8. Battery Pack Management

The battery pack is a consumable, and will lose performance over time. If the battery life
becomes less than half of what it was when first purchased, it is time for a replacement.

To purchase a battery pack, please contact Samsung Medison Service


Department.
NOTE

1.8.1. Replacing the Battery Pack


To replace a battery pack, please contact Samsung Medison Service Department.

Disconnect the power cord if you are not planning to use the product. Leaving
the product unused and not plugged in to a power outlet for an extended period
of time may deplete the battery completely, making it impossible to recharge the
batteries. In addition, allowing the battery pack to become completely depleted
WARNING may cause communication problems for the product.

1.8.2. Recharging the Battery Pack


Connecting the AC adapter automatically begins charging the battery pack. The battery pack will
be charged faster if system is powered off or in power saving mode.

■ If the low battery message appears while you are using the product,
immediately save the diagnosis information and connect the AC adapter.
■ Before connecting the AC adapter, make sure it’s the right way up. Forcing
the adapter into the product in the wrong way can damage the product.
WARNING
■ Do not recharge the battery pack using a method other than that described in
this manual. Doing so may lead to a fire or an explosion.

Below is an operating environment temperature for Battery Pack. If the battery temperature
goes beyond a certain degree, stop recharging for safety.
– Operating environment temperature: 10 - 35C

The battery pack can overheat if the operating environment temperature is too
high, or can take much longer than normal to recharge if the temperature is too
low.
CAUTION
1 - 52 V6/XV6 Service Manual

If using the battery pack as the power source, check the battery icon shown on
the screen to find out how much battery charge is left. For more information on
battery icons, refer to the ‘Monitor > Screen Layout’ section of ‘Introduction’.
NOTE

1.8.3. Disposing of the Battery Pack


A service representative of Samsung Medison or an authorized dealer must replace and
dispose of the battery.

Do not dispose of the battery pack carelessly. Do not incinerate the battery, as
this may cause an explosion or a fire.
WARNING
Chapter 1. Safety and Maintenance 1 - 53

1.9. Protecting the Environment

■ For disposal of the system or accessories that have come to the end of their
service lives, contact the vendor or follow appropriate disposal procedures.
■ Waste must be disposed of in compliance with the local regulations.
CAUTION ■ The lithium ion battery used in the product must be replaced by a service
representative or an authorized dealer.

1.9.1.1. EU WEEE Directive

For Correct disposal of this product.


Please visit: http://www.samsung.com/weee.pdf

1.9.1.2. EU REACH Regulation


For information on Samsung’s environmental commitments and product specific regulatory
obligations e.g. REACH visit:
www.samsung.com/uk/aboutsamsung/sustainability/environment/our-commitment/data/

1.9.1.3. EU Battery Directive

Correct disposal of batteries in this product


(Applicable in countries with separate battery return systems.)
The marking on the battery, manual or packaging indicates that the battery in this
product should not be disposed of with other household waste. Where marked,
the chemical symbols Hg, Cd or Pb indicate that the battery contains mercury,
cadmium or lead above the reference levels in EC Directive 2006/66.
The battery incorporated in this product is not user replaceable. For information
on its replacement, please contact your service provider. Do not attempt to
remove the battery or dispose it in a fire. Do not disassemble, crush, or puncture
the battery. If you intend to discard the product, the waste collection site will take
the appropriate measures for the recycling and treatment of the product,
including the battery.
Chapter 2

Introduction
2.1. Indications for Use ............................................................. 2
2.1.1. Contraindications ............................................................................. 2
2.2. Product Specifications ....................................................... 3
2.3. Product Configuration ........................................................ 6
2.3.1. Monitor ............................................................................................. 8
2.3.2. Control Panel ................................................................................. 10
2.3.3. Console .......................................................................................... 19
2.3.4. Peripheral Devices ......................................................................... 22
2.3.5. Probes ........................................................................................... 26
2.3.6. Accessories.................................................................................... 27
2.3.7. Optional Functions ......................................................................... 28
2 - 2 V6/XV6 Service Manual

2.1. Indications for Use


The ultrasound diagnostic system and probes are designed to obtain ultrasound images and
analyze body fluids.
The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative,
Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal,
Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric,
Thoracic, Trans-esophageal (Cardiac) and Peripheral vessel.
It is intended for use by, or by the order of, and under the supervision of, an appropriately
trained healthcare professional who is qualified for direct use of medical devices. It can be used
in hospitals, private practices, clinics and similar care environment for clinical diagnosis of
patients.

For more information on applications and presets, please refer to User


manual’s Chapter4, Probes.
NOTE

2.1.1. Contraindications
This product must not be used for ophthalmological applications, or any other use that involves
the ultrasound beam passing through the eye.

■ Federal law restricts this device to sale by or on the order of a physician.


■ The method of application or use of the device is described in Chapter5
‘Starting Modes of Operation’ and ‘Modes of Operation’.
CAUTION
Chapter 2. Introduction 2 - 3

2.2. Product Specifications

Height: 1,370-1,698mm (with monitor)


Width: 559mm
Physical
Depth: 964mm
Dimensions
Weight: 80kg (with monitor)
Weight: Approx.105kg (with Safe Working Load)

2D Mode
M Mode
Color Doppler Mode
Pulsed Wave (PW) Spectral Doppler Mode
Continuous Wave (CW) Spectral Doppler Mode
Tissue Doppler Imaging (TDI) Mode
Imaging Modes Tissue Doppler Wave (TDW) Mode
Power Doppler (PD) Mode
ElastoScan™ Mode
3D/4D Mode
Dual Mode
Quad Mode
Combined Mode

Gray Scale 256 (8 bits)

Transmit focusing, maximum of eight points (four points simultaneously


Focusing selectable)
Digital dynamic receive focusing (continuous)

Linear Array
LA2-14A, LA3-22AI

Curved Array
CA1-7S, CA1-7SD, CA3-10A, CA4-10M

Endocavity
EA2-11AR, EA2-11ARD, EA2-11AV, EA2-11AVD, miniER7

Probes 3D
(Type BF/IPX7) CV1-8A, CV1-8AD, EV2-10A

Phased Array
PA1-5A, PA3-8

MPTEE
MMPT3-7
Pencil
CW6.0, DP2B
2 - 4 V6/XV6 Service Manual

3 Active Probe Ports


Probe
4 Active Probe Ports (Option)
Connections
CW Probe Port (Option)

Main Monitor
Number of Pixel: 1920 x 1080
21.5 inch LCD Monitor
(LED Backlight unit, hereafter referred to as “LCD monitor”)
23.8 inch LCD Monitor (Option)
Monitor
(LED Backlight unit, hereafter referred to as “LCD monitor”)

Touchscreen Monitor
Number of Pixel: 1920 x 1080
14 inch LCD Monitor
(LED Backlight unit, hereafter referred to as “LCD monitor”)

ECG USB Type (Defibrillation-proof type CF)

Audio Jack Stereo Output (Option)


Rear Panel S-VHS Output (Option)
Input/Output LAN
Connections USB Port (USB3.0 compatible)
HDMI Output

Maximum 23,910 frames for Cine memory


Image Storage Maximum 13,146 Lines for Loop memory
Image filing system

Obstetrics, Gynecology, Urology, Abdomen, Cardiac, Vascular, Small Parts, MSK,


Application
Pediatric, TCD, Thoracic, Intraoperative

77V - 289V, 800VA, 47/60Hz


Electrical
19V, MAX 6A
Parameters
Battery: 14.4V, 6900mAh

Obstetrics, Fetal Heart, Gynecology, Cardiac, Abdomen, Pediatric Hips, Urology,


Measurement Small Parts, MSK, Vascular, TCD
Packages
* Refer to the ‘Chapter 8. Measurements’ for additional information.

TGC/LGC control
Mode-independent gain control
Acoustic power control (adjustable)
Signal Processing
Dynamic aperture
(Pre-processing)
Dynamic apodization
Dynamic range control (adjustable)
Image view area control
Chapter 2. Introduction 2 - 5

M Mode sweep speed control


Dynamic LPF Control

Digital TGC Control


Slider TGC Control
Mode-Independent Gain Control
Black Hole/Noise Spike Filtering
Signal Processing 2D Edge/Blurring Filtering
(Post-processing) M/D Mode Sweep Speed Control
Zoom
Image Orientation (left/right and up/down, rotation)
Dynamic Range Control
Reject Level Control

Trackball operation of multiple cursors


– 2D Mode: Linear measurements and area measurements using elliptical
Measurement approximation or trace
– M Mode: Continuous readout of distance, time, and slope rate
– Doppler Mode: Velocity and trace

Digital B/W Video Printer


Digital Color Video Printer
USB Printer
Auxiliary Foot switch (IPX8)
USB Storage Device
ECG (USB Type)
Monitor

User Interface English, German, French, Spanish, Italian, Chinese, Portuguese, Russian

Operating: 700 – 1060hPa


Pressure Limits
Storage: 700 – 1060hPa

Operating: 30 – 75%
Humidity Limits
Storage & Shipping: 20 – 90%

Temperature Operating: 10 – 35°C


Limits Storage & Shipping: -25 – 60°C

Altitude ≤3000m
2 - 6 V6/XV6 Service Manual

2.3. Product Configuration


This product consists of monitor, control panel, console, peripheral devices and probes.

① Monitor
② Monitor arm
③ Touch Panel
④ Speaker
⑤ Control Panel & Handle
⑥ USB port
⑦ Probe holder
⑧ Lift & Lift Lever
⑨ ECG port
⑩ Outlet for internal peripheral
devices
⑪ Probe port
⑫ CW Probe Port
⑬ Wheel

[Figure 2.1 Front of the Product]


Chapter 2. Introduction 2 - 7

① Storage compartments
② Ventilation
③ Cable holder
④ Rear panel
⑤ ID label
⑥ Power connection

[Figure 2.2 Back of the Product]


2 - 8 V6/XV6 Service Manual

2.3.1. Monitor

2.3.1.1. Screen Layout

[Figure 2.3 Monitor Display]

① Title Area
Displays patient information, hospital name, application, frame rate, depth, probe
information, acoustic output information, and the current date and time.
② Menu Area
Displays Preset Change. You can quickly change the preset of a probe. The EzExam+™
menu will also appear if being used. The Left menu will also appear if being used.

For more information on EzExam+™, please refer to User manual’s Chapter 3 ,


Utility > Setup > System > EzExam+™.
NOTE

③ Image Area
Displays ultrasound images. TGC, Image information, annotation, and measurement
information are also displayed.
④ Thumbnail Area
Images saved by pressing the preset Store button are shown in the thumbnails. Place a
pointer on a Thumbnail image to enlarge it in the preview image.
⑤ User Information and Status Information Area
Chapter 2. Introduction 2 - 9

Information that is useful to the user, such as current system status, image information,
selectable items, user defined key settings, etc., is displayed.

For more information about user keys, please refer to User manual’s Chapter
3, Utility > Setup > Customize.
NOTE

TIP BatteryAssist™
BatteryAssist™ provides battery power to the system, enabling users to perform
scans when AC power is temporarily unavailable. It also allows to transport the
ultrasound system to another location and start to scan right away.
Please check the battery capacity before use. A ‘Battery’ icon will be displayed on a
product that has a battery in it. The user can check the battery level as shown in the
table below:
Icon Description Tool Tip Message

Charged (at least 81% charged)

Charging (61–80%)

Battery/‘Valid Capacity %’
Charging (41–60%)

Charging (21–40%)

Remaining battery level (11–20%)

Remaining battery level (10% or below)

⑥ Contextual Button Area


Settings for Context Keys, including the positions of Set and Context, are displayed.
Depending on the ultrasound inspection items, the functions assigned to the buttons
around the trackball can be utilized to reduce the hassle of menu selection.

For more information, please refer to User manual’s Chapter 3,Utility > Setup >
Customize.
NOTE
2 - 10 V6/XV6 Service Manual

2.3.2. Control Panel

[Figure 2.4 Control Panel]

2.3.2.1. Adjusting the Control Panel


Use the two levers on the handle to move it vertically. Release the levers to fix it in the
current position. You can move it horizontally without pressing any levers.

■ Do not apply excessive force to the control panel.


■ Use the handles on the console when moving the product.
CAUTION

Functions of some buttons on the Control Panel can be changed in Setup >
Customize. Functions configured will appear in User Information and Status
Information area and Contextual Button area on the monitor screen.
NOTE
Chapter 2. Introduction 2 - 11

2.3.2.2. Functions of the Control Panel

1 Button Turns the system on/off.

On/Off

Adjusts the angle of the sample volume in


2 Dial-button Spectral Doppler mode. It is also used to
adjust the probe angle for a BodyMarker.

Allows the user to enter a BodyMarker over


an image.
3 Dial-button
You can adjust the probe angle by turning the
dial-button.

It uses the functions available in a current


4 Dial-button scan mode such as EzExam+™, Left Menu,
and Preset Change.

In Scan Mode, an arrow-shaped pointer


5 Button
appears on the screen

Pointer

Exits the function currently being used and


6 Button
returns to the previous screen.

Exit
Enter or exit M Mode.
Rotate this dial-button to adjust the M Gain
7 Button value. In 3D View and S-Fusion™, turning
this dial-button will rotate the image around
the x-axis.

Enter or exit Power Doppler Mode.


Rotate this dial-button to adjust the PD Gain
8 Dial-button value or the Gain value in MV-Flow™. In 3D
View and S-Fusion™, turning this dial-button
will rotate the image around the y-axis.
2 - 12 V6/XV6 Service Manual

Enter or exit PW Spectral Doppler Mode.


Rotate this dial-button to adjust the PW Gain
9 Dial-button value. In 3D View and S-Fusion™, turning
this dial-button will rotate the image around
the z-axis.

Enter or exit Color Doppler Mode.


10 Dial-button Rotate this dial-button to adjust the C Gain
value. In 3D View, turning this dial-button will
move the reference slice horizontally.

Enter 2D Mode.
11 Dial-button Rotate this dial-button to adjust the Gain
value.

12 Button Enter or exit CW Spectral Doppler Mode.

13 Button Enter or exit 3D Mode.

14 Button Enter or exit 4D Mode.

15 Button Turn the QuickScan™ function on.

16 Switch Adjusts the scanning depth of the image.

Moves the focus to the target area for


17 Switch
observation.
Chapter 2. Introduction 2 - 13

Makes the Zoom Box appear. In order to exit


18 Dial-button
Zoom Mode, press Exit.

Only the image is displayed on the screen.


19 Button
(Single Mode)

Compares four independent images. (Quad


20 Button
Mode)

Compares two independent images. (Dual


21 Button Mode) The active image mode is shown on
the left.

Compares two independent images. (Dual


22 Button Mode) The activated image mode is shown
on the right.

Stands for Peripheral Key; the user can


23 Button
assign the desired functions to each button.

Stands for User Key; the user can assign the


24 Button
desired functions to each button.

Deletes text, Arrows, BodyMarkers,


25 Button measurement results, etc., displayed on an
image.
Clear

26 Button Allows the user to place text over an image.


Text

Changes to a different function that can be


27 Button
supported by the current trackball setting.
Change

28 Button Starts measurements by application.


Calculator
2 - 14 V6/XV6 Service Manual

Starts to measure distance, circumference,


29 Button
area, and volume.
Caliper
The user can set up and use Context/Set or
Set/Context.
– Set: Selects an item or value using the
trackball. Alternatively, this can be used to
change the function of the trackball.
30 Button
– Context (Contextual Button): Depending
on the ultrasound inspection items, the
① functions assigned to the buttons around
the trackball can be utilized to reduce the
hassle of menu selection.
Contextual Button: Depending on the
ultrasound inspection items, the functions
31 Button assigned to the buttons around the trackball
can be utilized to reduce the hassle of menu
② selection.

Pauses an image being scanned or


32 Button
reactivates a paused image.
Freeze

Moves the cursor on the screen. Also, scrolls


33 Trackball
through Cine images.

Trackball
Chapter 2. Introduction 2 - 15

■ Keyboard (Option)
A keyboard, an optional item that you can connect via a USB port, is available. You can
use the keyboard to enter text, etc.

[Figure 2.5 Keyboard]


2 - 16 V6/XV6 Service Manual

2.3.2.3. Touchscreen
Items that are available for use in each Operation mode will be enabled on the touchscreen.
You can change the image format or optimize an image to facilitate your diagnosis.
– In a combined mode that uses more than one Operation mode, tap the relevant tab on
the touchscreen to specify settings for each mode.
– You can change the tabs on the touchscreen by dragging them as if you are turning
pages. Press the touch button or use the dial-button on the control panel to select a value.

[Figure 2.6 Touchscreen Display]

① These buttons are always displayed on the touchscreen.

Displays the Patient Information screen, which is used for selecting a patient
Patient
ID from the list or entering new patient information.

Displays the Probe Selection screen to select or change the probe and
Probe
application.

Displays the Report screen that shows the measurement results of the current
Report
application and other information.

SonoView Launches SonoView, an image filing program.

Finishes the exam of the patient currently being diagnosed and resets the
End Exam
related data.

Utility Configure the system settings.


Chapter 2. Introduction 2 - 17

② TouchEdit: A customizable touchscreen interface that allows the user to move


frequently used functions to the first page, keeping the focus on the patient instead of
the system.

③ Displays the mode that is currently used.

④ Displays the buttons of functions relevant to the operation mode that is currently in use.

⑤ This is the TGC (Time Gain Compensation) area. It displays sliders where you can
control Gain.

An excessively large difference between the Gain value settings of adjacent


TGC sliders may cause stripes to appear in an image.
CAUTION

– Penetration of ultrasound becomes weaker as the depth of the site to study


increases. Use of TGC can compensate for this phenomenon.
– TGC can adjust the Gain based on the depth. The depth increases gradually from
the top to the bottom sliders and moving the slider to the right increases the Gain,
which makes the image brighter.
User TGC Settings: User can set and save the TGC form.

[Figure 2.7 TGC – Touchscreen]

1. Use the TGC sliders to adjust the Gain, and press Save as Preset.
2. With the User TGC button, you can save the pattern you want.
3. Select the TGC Preset button to complete the User TGC settings.
4. Press the first button of TGC Preset for center alignment.

⑥ The Preset saved in the Probe field is displayed as a button.


2 - 18 V6/XV6 Service Manual

QuickPreset: With one touch, the user can select the most common transducer and
preset combinations. QuickPreset increases efficiency to make a full day of scanning
simple and easy.

For more information, please refer to User manual’s Chapter 5, “Starting


Modes of Operation”.
NOTE

⑦ Soft Menu Area: The soft menu items that are available in the current input mode are
shown. The menu in use is shown in the border color. Press or rotate the dial-buttons
right below each menu.

TIP When there are Two Soft Menus


When there are two menus available - upper and lower, both menus can be
adjusted with the corresponding dial-button. Or tap the button for the menu you
want to use on the touchscreen and then use the dial-button.
Chapter 2. Introduction 2 - 19

2.3.3. Console
The product consists mainly of ultrasound imaging components on the inside, and various
connectors, probe holders, a storage compartment, handles, and wheels on the outside.

2.3.3.1. Rear Panel


A monitor and other peripheral devices, are connected via the rear panel at the back of the
system.

[Figure 2.8 Rear Panel Part]

① Network port: Connect to a network. You can transfer patient information to another
server via DICOM network.
② HDMI port (Output): Outputs digital signals to the monitor. (Input: Not supported.)
③ USB port: Used to connect to peripheral USB devices.
④ S-VHS Port (Output): Provides a S-VHS connection for a VCR. (Option)
⑤ Audio port (Output): Used for audio signal output. (Input: Not supported.) (Option)
2 - 20 V6/XV6 Service Manual

2.3.3.2. Power Connection Part


The power connection part is located at the right-hand side of the rear panel.

[Figure 2.9 Power Connection Part]

① Power Switch: Turns on the power of the product, and cuts off power.
② Power Inlet: Accepts the power cord, which connects to an external power supply.
③ Equipotential Terminal: This should be connected to the equipotential connection part in
the exam room.
Chapter 2. Introduction 2 - 21

2.3.3.3. Probe Holder


Holders, which can hold and store probes, are attached to the left and right-hand sides of the
control panel. These can be detached for cleaning.

TIP Endocavity Probe Holder


Install this to fix the vaginal probe onto the holder. It is provided as an option.

[Figure 2.10 Probe Holder Part]


2 - 22 V6/XV6 Service Manual

2.3.4. Peripheral Devices


Peripheral devices can be connected to their corresponding ports on the left/right or rear sides
of the console, as needed.

■ Do not install peripheral devices that are not listed in this user manual in the
patient environment. If you install an unlisted device in the patient
environment, it may cause an electrical hazard.
■ Do not connect additional external peripheral devices to the auxiliary socket
CAUTION outlet. Doing so may decrease safety level.

[Figure 2.11 Patient Environment]

Refer to the user manual of the peripheral device for its operating information.
NOTE

The following products are recommended:

When using a peripheral device via a USB port, always turn the power off
before connecting/disconnecting the device. Connection/disconnection of USB
devices while the power is on may lead to malfunction of the system and the
USB devices.
CAUTION
Chapter 2. Introduction 2 - 23

■ Digital Thermal Printer


– Black and White: Sony UP-D897, UP-X898MD, UP-D898MD
– Color: Sony UP-D25MD

Sony UP-D897, and Sony UP-D897MD are not available in the US and Canada.

NOTE

■ USB Printer
– Samsung CLP-620NDK, ML-2950

■ Be sure to check the English OS version first to install a compatible printer


and driver. Contact Samsung Medison customer support division for inquiries
about printer driver installation.
■ When connecting the printer, ensure that the printer is configured under
CAUTION Microsoft Windows or system setup and has been chosen as the default
printer.
■ Please check the port that the printer uses before connecting. Digital Thermal
Printers should be connected to the printer port while the USB printer should
be connected to the USB port.

■ Foot Switch
– 3 Pedals HID Type

Set the functions in User manual Chapter 3, Utility > Setup > Customize >
General > Foot Switch.
NOTE

Connecting or removing the Foot Switch while the product is turned on may
cause the system to malfunction.
CAUTION
2 - 24 V6/XV6 Service Manual

■ Wi-Fi Adaptor
– Wi-Fi adaptor (USB Type): Archer T4UH (TP-LINK AC1750)

For more information about how to connect to wireless networks, please refer to
User manual Chpater3 ‘Utility > Setup > Connectivity > Network> Wireless’.
NOTE

The device’s transmitter/receiver protocol specifications may vary depending on the Wi-Fi
Adaptor in use.
Model Archer T4UH (TP-LINK AC1750)
Protocol IEEE 802.11ac
IEEE 802.11a
IEEE 802.11n
IEEE 802.11g
IEEE 802.11b
Signal Rate 5 GHz
11ac: Up to 867 Mbps(dynamic)
11n: Up to 300 Mbps(dynamic)
11a: Up to 54 Mbps(dynamic)

2.4 GHz
11n: Up to 400 Mbps(dynamic)
11g: Up to 54 Mbps(dynamic)
11b: Up to 11 Mbps(dynamic)
Frequency 2.4 GHz, 5 GHz
Transmit Power <20 dBm (EIRP)
Wireless Security Support 64/128 bit WEP, WPA-PSK/WPA2-PSK, 802.1x
Chapter 2. Introduction 2 - 25

The transmitter/receiver specifications, such as frequency or bandwidth, may vary


depending on the Wi-Fi Adaptor in use and the protocol of the connected device.

IEEE 802.11 Network PHY Standards

Frequency Bandwidth
Protocol Stream Data Rate (Mbit/s) Modulation
(GHz) (MHz)

802.11a 5 20 6, 9, 12, 18, 24, 36, 48, 54 OFDM

802.11b 2.4 22 1, 2, 5.5, 11 DSSS

802.11g 2.4 20 6, 9, 12, 18, 24, 36, 48, 54 OFDM

20 Up to 288.8
802.11n (Wi-Fi 4) 2.4/5
40 Up to 600

20 Up to 346.8
MIMO-OFDM
40 Up to 800
802.11ac (Wi-Fi 5) 5
80 Up to 1733.2

160 Up to 3466.8

■ Misc.
– USB Memory Media

■ The system cannot recognize USB memory. Remove the USB memory from
the console and equip again with an appropriate device.
■ Regarding file formats that are not ordinarily saved: Please check first to see
if it is possible to save the file format on a desktop PC before trying to save
NOTE the file on USB memory.

■ Do not use USB memory media which contain anti-virus programs or are
defective. Otherwise, the product may fail to work properly.
■ Please use a USB drive compatible with Windows 10. Otherwise, some
features may not work correctly.
CAUTION
2 - 26 V6/XV6 Service Manual

2.3.5. Probes
Probes are devices that generate ultrasound waves and process reflected wave data for the
purpose of image formation.

For information on probes, refer to User Manual Chapter 4, Probes.

NOTE

2.3.5.1. Connecting Probes


Be sure to connect or disconnect probes when the power is off to ensure the safety of the
system and the probes.
1. Connect probes to the probe ports on the front panel of the system. The CW probe
should be connected to CW-only port.

2. Turn the probe’s lockdown switch clockwise to lock it in place.

■ When connecting a probe, make sure that the connector is not exposed to
any liquid or foreign substance. Otherwise, fire, electric shock, injury, or
damage to the product may occur.
■ The probe may experience connection problems caused by foreign objects. If
CAUTION such a problem occurs, remove any foreign objects and try again.
Chapter 2. Introduction 2 - 27

2.3.6. Accessories
An accessory box containing the items below is supplied with the product.

[Figure 2.12 Accessories]

Supplied accessories vary by region.

NOTE
2 - 28 V6/XV6 Service Manual

2.3.7. Optional Functions


For more information about optional features, please refer to the relevant sections in User
Manual.
2D Follicle™ HDVI™
5D CNS+™ HQ-Vision™
5D Follicle™ IOTA-ADNEX
5D Heart Color™ LaborAssist™
5D LB™ LumiFlow™
5D Limb Vol.™ Mobile Export
5D NT™ MV-Flow™
ADVR™ NeedleMate+™
ArterialAnalysis™ NerveTrack™
AutoIMT+ Panoramic+
AutoEF QUS™ (TAI™, TSI™)
BiometryAssist™ RealisticVue™
Cardiac Measurement S-Detect™ for Breast
CEUS+ S-Detect™ for Thyroid
CrystalVue™ S-Shearwave Imaging™
CrystalVue Flow™ Smart 4D
CW Function SonoSync™
DICOM Strain+
E-Strain™ StressEcho
ElastoScan+™ System Activation
EzExam+™ XI STIC
EzHRI™ Expanded Storage (1TB)
HeartAssist™

Optional features vary by country and version. Please check the product to
confirm which features are available.
NOTE
Chapter 3
3. 제품 설치

제3장. 제품 설치

Installing Product
3.1. Transporting.............................................................................................. 2
3.1.1. Precaution during transport ................................................................................... 2
3.1.2. Brakes.................................................................................................................... 2
3.1.3. Precaution on Ramp .............................................................................................. 2

3.2. Unpacking the Product ............................................................................ 3


3.2.1. Unpacking the product’s box ................................................................................. 3
3.2.2. Accessory .............................................................................................................. 5

3.3. Installation Environment .......................................................................... 6


3.3.1. Caution .................................................................................................................. 6
3.3.2. Humidity and Temperature .................................................................................... 7
3.3.3. Standby time before using the product .................................................................. 7

3.4. Installing the Product ............................................................................... 8


3.4.1. Installation Safety .................................................................................................. 8
3.4.2. Printer Installation ................................................................................................ 10
3.4.3. External Keyboard Installation ............................................................................. 15
3.4.4. Battery Option Installation ................................................................................... 16
3.4.5. HV board(CAP board) Installation ....................................................................... 17
3.4.6. ECG module Installation ...................................................................................... 18
3.4.7. Enter Installation Key........................................................................................... 19
Chapter 3. Installing Product 3 – 2

3.1. Transporting
This product is finely-tuned medical electronic equipment; careful attention is required when
transporting it.

Use extra caution when you move the product. Careless transportation or
transfer of the product may result in product damage or personal injury.
WARNING

3.1.1. Precaution during transport


This product and its packaging are designed to protect the product from physical impacts.
However, drops and external impacts may cause serious damage to the product.
• Before transporting the product, make sure that the caster brakes are released.
Also, make sure that you secure any parts that may move, such as the monitor arm or
keyboard, into position before you move them.
• Always use the handles at the back of the console and move the product slowly.

3.1.2. Brakes
You can use the caster brakes on the product to control their movement. Make sure that the
brakes are engaged before you use the product.

3.1.3. Precaution on Ramp


Always make sure that the control panel is facing the direction of movement. If you leave the
product on an incline, the product may fall over even if the brakes are engaged. Do not rest the
product on a ramp.

The product weighs more than 80 kg. Be extra careful when transporting it.
Careless transportation of the product may result in product damage or personal
injury.
WARNING

Be especially aware of the castors when moving the system.


Samsung Medison recommends that you practice transporting the product up or
down ramps in advance of any significant relocation.
WARNING
Chapter 3 Installing Product 3-3

3.2. Unpacking the Product

3.2.1. Unpacking the product’s box


1. Remove the Packing wire and the master box guide.
2. Remove the master box.
3. Remove the two cross members pad.
4. Remove the two CP side paper pad.
5. Remove the two CP rear paper pad.
6. Remove the monitor paper pad.
7. Remove the monitor front paper pad.
8. Remove the CP front paper pad.
9. Remove the two 2D probe boxes, two 3D probe boxes, accessory box and gel-warmer box.
10. Remove the accessory out paper pad.
11. Remove the plate-slope.
12. Remove the slope paper pad.
13. Remove the Bottom front paper pad.
14. Remove the Bottom rear paper pad.
15. Remove the protective vinyl.
16. Unload the system from the pallet.
3–4 V6/XV6 Service Manual

[Figure 3.1 System Unpacking]


Chapter 3 Installing Product 3-5

3.2.2. Accessory
An accessory box containing the items is supplied with the product.
If it does not match your order, please contact the Samsung Medison customer service
department or your local vendor.

[Figure 3. 2 Accessory]

Accessory configuration may vary by region.

NOTE
3–6 V6/XV6 Service Manual

3.3. Installation Environment

3.3.1. Caution
When installing the product, please pay attention to the following: For more information on use
and setup, please refer to the accompanying manual for this product.

Placing the system near generators, X-ray machines or broadcast cables may
result in abnormal images on the screen due to electrical noise.
Sharing the power source with other electrical devices may also produce the
CAUTION similar problem.

• Optimum conditions for system operation are 10-35 temperatures and 30-75% humidity.
• Avoid excessive humidity.
• Avoid direct sunlight.
• Avoid excessive fluctuations in temperature.
• Avoid installing the product near a heating system.
• Avoid places that are dusty or poorly ventilated.
• Avoid locations that are subject to vibration.
• Avoid locations where chemical substances or harmful gases are present.
Chapter 3 Installing Product 3-7

3.3.2. Humidity and Temperature


Table 3-1. Temperature and Humidity Limits shows information on temperature and humidity
ranges suitable for transporting, storing and operating the product.

Category Temperature [°C] Humidity [%]


Transporting -25 ~ 60 20 ~ 90
Storage -25 ~ 60 20 ~ 90
Operating 10 ~ 35 30 ~ 75

[Table 3.1 Temperature and Humidity Limits]

3.3.3. Standby time before using the product


When using the product after storage or transport at a temperature outside of `10~35C, use it
after allowing the standby time to reach the operating environment temperature guaranteed by
our company. A standby time of 2 hours per 5C is required before turning on the power.

If the product has been transported or stored outside the operating environment
temperature (10~35C), do not turn on the product immediately. A sudden
change in temperature may cause product malfunction.
CAUTION Refer to the table below and allow standby time before using the product.

Temperature 10
[°C]
-40 -35 -30 -25 -20 -15 -10 -5 0 5 40 45 50 55 60
~35
Standby
20 18 16 14 12 10 8 6 4 2 0 2 4 6 8 10
time[h]

[Table 3.2 Standby Time to reach the operating temperature]


3–8 V6/XV6 Service Manual

3.4. Installing the Product

3.4.1. Installation Safety


The power connection part is located at the right-hand side of the rear panel.


[Figure 3.3 Power Connection Part]

① Power switch: Turns on the power of the product, and cuts off power.

② Power inlet: Accepts the power cord, which connects to an external power supply.

③ Equipotential Terminal: This should be connected to the equipotential connection part in


the exam room
Chapter 3 Installing Product 3-9

3.4.1.1. Power Cord Connection


Prior to connecting a power cord, verify that the voltage indicated on the power cord matches
the voltage rating of the installation point.

The product may be shipped with the power cable already connected to the
console.
NOTE

■ If the product needs to be transported or stored for an extended period, the


temperature and humidity of the environment must be checked.
■ A sudden change in temperature may cause condensation and lead to
CAUTION product failure.

3.4.1.2. Connecting the Equipotential Terminal


In a hospital environment, dangerous electric currents may occur as a result of the potential
difference between a contactable conductive part and connected equipment in treatment rooms.
The solution to the problem is consistent equipotential bonding. The equipotential terminal of an
item of medical equipment must be connected to the equipotential bonding network in its
treatment room.

3.4.1.3. Probe Connection


Be sure to turn off the power before connecting or disconnecting a probe to ensure the safe use
of the system and the probes.
1. Connect the probes to the probe connectors on the front panel of the system. Up to four
probes may be connected.
2. To install it, turn the probe connector’s knob clockwise.
3 – 10 V6/XV6 Service Manual

3.4.2. Printer Installation

3.4.2.1. Thermal Printer Cable Installation


1. ① Unscrew 8 screws and separate the Rear Case.
2. ② Unscrew 2 screws and separate the Cable Bracket.
3. ③④ Connect BW Printer or Color Printer Signal Cable to the product. You can insert the
cables either of the two ports.
4. ⑤ Insert BW Printer or Color Printer Power Cable to the product.
5. ⑥ Install the Cable Bracket and fix it with 2 screws.
6. ⑦ Pull out the signal cables and power cables to the front side via the hole of the product.

[Figure 3.4 Thermal Printer Cable Installation]


Chapter 3 Installing Product 3 - 11

3.4.2.2. Thermal Printer Total Tray Installation


1. ① Unscrew 4 screws and separate the Printer Total Tray module.
2. ⑥ Remove the Rubber Caps and 4 screws from the product.
3. ⑦ Install the Printer Total Tray and fix it with screws.

[Figure 3.5 Thermal Printer Total Tray Installation]


3 – 12 V6/XV6 Service Manual

3.4.2.3. BW Thermal Printer Installation


1. ①② Turn over the BW Printer. Put the BW printer tray on it and fix it with 4 screws.
2. ③④⑤ Install the BW Printer on the total tray and fix it with 4 screws.
3. ⑥ Lastly, connect the power cable and the signal cable to the printer.

[Figure 3.6 BW Thermal Printer Installation]


Chapter 3 Installing Product 3 - 13

3.4.2.4. Color Thermal Printer Installation


1. ①② Turn over the color printer. Put the printer tray on it and fix it with 4 screws.
2. ③④⑤ Install the color printer on the Total Tray and fix it with 3 screws.
3. ⑥ Connect the color printer and power cable to the product.

[Figure 3.7 Color Thermal Printer Installation]


3 – 14 V6/XV6 Service Manual

3.4.2.5. LaserJet Color Printer Installation


1. Connect the USB cable between USB port on Real panel and Printer USB port.
2. Connect Power Cord to wall power.

[Figure 3.8 LaserJet Color Printer Installation]


Chapter 3 Installing Product 3 - 15

3.4.3. External Keyboard Installation


1. ③ Unfold the rail of the external keyboard.
2. ④⑤ Hang the rail frames on both side of the external keyboard to the hooks under the
control panel and pull them forward.
3. ⑥ Fix the keyboard with 4 screws.
4. ⑦ Connect the USB cable on the Key board to USB Port.

[Figure 3.9 External Keyboard Installation]


3 – 16 V6/XV6 Service Manual

3.4.4. Battery Option Installation


1. ① Unscrew 8 screws and separate the Rear Cover.
2. ③④ Insert 3 Battery Packs into the product and fix it with 8 screws.
3. ⑥ Close the Rear cover and fix it with 8 screws.

[Figure 3.10 Battery Option Installation]


Chapter 3 Installing Product 3 - 17

3.4.5. HV board(CAP board) Installation


1. ① Unscrews 12 screws and separates PSU Top case.
2. ③ Unscrews 8 screws and separates Side Cover.
3. ④ Unscrews 8 screws and separates Non- SWE HV board.
4. ⑥ Installs SWE HV board and fix it with 9 screws. When inserting the board, be careful to
insert the connector correctly.
5. Last of all, close the cover and fix it with 12screws.

[Figure 3.11 HV board (CAP board) installation ]


3 – 18 V6/XV6 Service Manual

3.4.6. ECG module Installation


1. ① Prepare the components of ECG option.
2. ②③ Separate Foothold and Front Cover.
3. ④ Insert the ECG cable into the ECG Module.
4. ⑤ Tie the cable with a cable tie so that the cable does not disconnect from the ECG module.
5. ⑥⑦ Last of all, fix the module with a screw and reassemble the cases.

[Figure 3.12 ECG module Installation ]


Chapter 3 Installing Product 3 - 19

3.4.7. Enter Installation Key


After install the system, register system information on the GSIS.

1. Once the system reboots after installation, the System Installation screen will be displayed.
2. Enter the Installation Key.
You can get the Installation Key from GSIS (Global Service Information System) website
after register the hospital information and entering the system information.
GSIS URL: https://gsis.samsunghealthcare.com

[Figure 3.13 GSIS’s Installation Key creation]

3. Clink the [Authentication] button.


Chapter 4
4. Product Inspection

Chapter 4. Product Inspection

Product Inspection
4.1. Inspecting the Functions ......................................................................... 2
4.1.1. Basic Inspections ...................................................................................................... 2
4.1.2. Detailed Inspections.................................................................................................. 3

4.2. Periodic Maintenance Inspections .......................................................... 7


4.2.1. Maintenance Cycle by Item ...................................................................................... 7
4-2 V6/XV6 Service Manual

4.1. Inspecting the Functions


After installation, inspect the device and use the major functions to verify that the product works
properly.

4.1.1. Basic Inspections

4.1.1.1. Monitor
Check the screen for color, focus, dots, afterimage, stain, and blurriness.
Check the screen while applying impact on the monitor and check the signal while shaking the
cable to the left and right.

■ Control Panel and LED Status

Press any keys on the control panel and see if text appears or breaks.
Verify that the Keyboard LEDs turn on.

■ BodyMark Key

Check if a BodyMarker image appears correctly and whether selection works properly.

■ Indicator Key

Check whether the trackball works properly by rolling it up, down, left, and right.

■ Clear Key

Check whether all texts and measurements are properly cleared.

■ Check Zoom Functioning

Check the zoom function for proper action and make sure that there are no errors.

■ Check SonoView

Save Images and Cine Images in each mode.


Check for errors in saving images.
Check whether Backup & Restore function properly.

■ Measure

Check the operation of Distance, Caliper, Calc, etc.


Chapter 4. Product Inspection 4-3

■ Patient

Enter information in Patient and check whether the same Patient information is displayed in
Report, SonoView, etc.

■ End Exam

Make an arbitrary measurement in New Patient and see whether the measurement is erased
when you press End Exam.

■ Probe Key

Perform a Probe Change and check whether it functions correctly.

4.1.2. Detailed Inspections

■ 2D Mode

1. Perform a Knife Test to check whether part of the image is not displayed.
2. Perform a Phantom to check for errors in image.
3. Check the Freeze Cine actions (broken image, Auto Run, Auto Run Speed, Trackball
Cine).
4. Check for changes in image brightness when you adjust Gain.
5. Check whether TGC Gain operates correctly by adjusting it and check for changes in
image brightness in accordance with depth.
6. Perform Left/Right Flip, Up/Down Direction, and Rotation to check whether the image
changes its orientation correctly.
7. Check whether the Select Image menus (EE, DR, View Area, Tissue, Frame Rate) work
properly.
8. Check for errors in frequency (Phantom, Res, Pen, Gen).
9. Check for changes in image in accordance with the changes in depth.
10. Check for changes in image by depth when you change the focus.
11. Check whether or not Image Compensation Mode works.
4-4 V6/XV6 Service Manual

■ Dual Mode

1. Perform a Phantom to check for errors in image.


2. Perform Left/Right Flip, Up/Down Direction, and Rotation to check whether the image
changes its orientation correctly.
3. Check the proper operation of Select Image menus (EE, DR, View Area, Tissue, Frame
Rate, Power).
4. Check for errors in frequency (Phantom, Res, Pen, Gen).
5. Check for changes in image in accordance with the changes in depth.
6. Check for changes in image by depth when you change the focus.
7. Check for errors in operations related to left and right image Cine (number of pages, Cine
progression order, broken image, Auto Run, Auto Run Speed, Track Ball Cine).

■ M Mode

1. Perform a Phantom to check for errors in image.


2. Check whether the M-Line’s information is displayed in the Image area.
3. Check for changes in image brightness when you adjust Gain.
4. Perform Left/Right Flip, Up/Down Direction, and Rotation to check whether the image
changes its orientation correctly.
5. Check whether the Select Image menus (EE, DR, View Area, Tissue, Frame Rate, Same
Angle, Power) operate correctly.
6. Check for changes in image in accordance with the changes in depth.
7. Check for changes in image by depth when you change the focus.
8. Check for changes in speed when you change Speed gradually and for errors in the
information.
9. Check whether the image becomes inverted when Negative is performed.
10. Check for errors in Top Down Format and Side by Side Format images when Loop
Format is selected.
11. Check for changes in Format B-Mode and M Line area sizes.
12. Check the Freeze Cine actions (broken image, Auto Run, Auto Run Speed, Trackball
Cine).
Chapter 4. Product Inspection 4-5

■ Color Doppler/Power Doppler Mode

1. Perform a Phantom to check for errors in image.


2. Check the operations related to the Select Image menus (Balance, Sensitivity, Color
Mode, Display, CFR).
3. Check for changes in image in accordance with the changes in depth.
4. Check the Freeze Cine actions (broken image, Auto Run, Auto Run Speed, Trackball
Cine).
5. Check for changes in image brightness when you adjust Color Gain.
6. Check for noises and breaks in image (B or C Mode Noise) when you move the ROI Box.
7. Check for noises and breaks in image (B or C Mode Noise) when you resize the ROI Box.
8. Adjust the Scale up and down to check whether or not the frequency changes and if the
speed range of blood flow is adjusted. (Check by actually performing a scan.)
9. Operate Filter to check whether small signals are removed by stage.
10. Check whether or not Color Bar is inverted when you operate the Invert key.
11. Move the Baseline up and down to check whether the speed range of blood flow moves
to the “+” or “-” position.

■ PW Spectral Doppler Mode

1. Perform a Phantom to check for errors in image.


2. Check whether Doppler’s PRF value changes when you set Simultaneous to on/off.
3. Check for errors in the Doppler spectrum.
4. Change the Scale to check the change in velocity range.
5. Move the Baseline up and down to check whether the spectrum range moves to the “+” or
“-” position.
6. Check whether changing Filter removes low signals from the Spectrum.
7. Check whether the Doppler wave form is inverted when you operate Invert.
8. Perform Angle action.
9. Change the position and size of SV, and then check for errors.
10. Check whether Spectrum’s image changes when you change Spectrum’s Type.
11. Check for errors in Sound Volume.
12. Perform Auto Calc and check whether the line is displayed without breaks; check whether
the calculated results are carried out automatically.
13. Check for errors in Top Down Format and Side by Side Format images when Loop
Format is selected.
14. Check the Cine/Loop (broken image, Auto run, Auto run speed, Trackball Cine)
operations.
4-6 V6/XV6 Service Manual

■ 3D/4D Mode

1. Check whether loading is performed correctly when you proceed with Free Hand 3D
SCAN and when you skip to Freeze; check for broken images and noise while proceeding.
2. Check whether loading is performed correctly when you proceed with Static 3D Scan;
check for broken images and noise while proceeding. Check the probe for noise and
whether the probe’s motor works normally.
3. Check whether loading is performed correctly when you proceed with Live 3D Scan;
check for broken images and noise while proceeding. Check the probe for noise and
whether the probe’s motor works normally.
4. Check for errors in ROI 3D, ABC 3D, and Full images.
5. Check whether the 3D image changes to the selected angle.
6. Check whether the 3D image’s contrast changes to the selected value.
7. Check for errors in the image when you change the size of the image.
8. Check the Display Format Image (ACB, Volume CT Image)
9. Step Angle, Rotation Angle, Rot. Axis, and then proceed with Cine; check whether Cine
Loading works in accordance with the Setting items and check for breaks and errors in
the image.
Chapter 4. Product Inspection 4-7

4.2. Periodic Maintenance Inspections


This chapter describes Periodic maintenance inspections.
These procedures are needed to maintenace the ultrasound system’s performance in the best
condition. After reading this chapter, please be aware of the procedure before performing the
inspections.

■ It is the customer’s responsibility to perform maintenance activities on a


regular basis to ensure that the product maintains its best performance.

■ The Periodic Maintenance Inspections are not mandatory. Inspection items

NOTE and cycle may be added depending on the MSA program.

For details, please contact Samsung Medison’s professional service engineers.

4.2.1. Maintenance Cycle by Item

Item Service activity Cycle

Check whether the Power outlet has a ground.


Power Noise & Ground Check if there is are any devices that generate Monthly
eletromagnetic wave around.

AC POWER Check and record the current voltage of the power outlet. Monthly

FAN Check Bottom Fan, Rear Fan and CPU cooling Fan Monthly

Booting Status Check the system’s Booting time. Monthly

Monitor/Touch panel Check Brightness, Contrast, Stain and Dot of the display Weekly

TGC Volume Check 8 TGC volume’s contamination/defect Weekly

Track Ball Check the movement of Track Ball up,down,left and right Daily

Caster(Wheel) Check caster Rolling and Locking status Monthly

Probe appearance Check for lens damage, Cap cracks and S/R separation Daily

Probe element Check for element status through Knife Test Monthly

Probe performance Check the image qualtity through the Phantom Annually

System Function Check 2D/PW/C/M/CW/3D/4D/Measure function Annually

Peipheral Device Check Printer/External monitor function Annually


4-8 V6/XV6 Service Manual
Chapter 5
5. 제품 구조

제5장. 제품 구조

Product Structure
5.1. Overview ................................................................................................... 3
5.2. System block diagram and Layout ......................................................... 4
5.2.1. Overall System block diagram ............................................................................... 4
5.2.2. System Layout(Body) ............................................................................................ 4

5.3. Overall structure description .................................................................. 6


5.3.1. Ultrasound System Part......................................................................................... 6
5.3.2. PC Part .................................................................................................................. 6
5.3.3. User Interface Part ................................................................................................ 7
5.3.4. Power Part ............................................................................................................. 7

5.4. Ultrasound System Part ........................................................................... 8


5.4.1. PSA (Probe Select Assembly) ............................................................................... 8
5.4.2. Main board and CW board .................................................................................. 10
5.4.3. Rear I/O board ..................................................................................................... 14

5.5. PC Part .................................................................................................... 15


5.5.1. PC Module ........................................................................................................... 15
5.5.2. Software Digital Scan Converter(DSC) ............................................................... 16

5.6. Control Panel .......................................................................................... 18


5.6.1. Control Panel ....................................................................................................... 18
5.6.2. Main Monitor ........................................................................................................ 21
5.6.3. Touch Panel ......................................................................................................... 22

5.7. Power Part ............................................................................................... 23


5.7.1. PSU / Battery / Battery charger ........................................................................... 23
5.7.2. Power Block Diagram .......................................................................................... 24
5-2 V6/XV6 Service Manual

5.7.3. Power Status LED ............................................................................................... 28

5.8. Interconnect Diagram............................................................................. 30


5.8.1. V6/XV6Interconnect Diagram ............................................................................. 30
5.8.2. Cable Description ................................................................................................ 31
Chapter 5. Product Structure 5-3

5.1. Overview
Chapter 5 describes the overall system structure of V6/XV6and the function of detailed
components.
5-4 V6/XV6 Service Manual

5.2. System block diagram and Layout

5.2.1. Overall System block diagram

[Figure 5.1 V6/XV6 System block diagram]

5.2.2. System Layout(Body)

[Figure 5.2 V6/XV6 System Layout (Body)]


Chapter 5. Product Structure 5-5

[Figure 5.3 V6/XV6System Layout(CP)]


5-6 V6/XV6 Service Manual

5.3. Overall structure description


This product is largely composed of 4 blocks such as Ultrasound System part, PC part, User
Interface part and Power part. The functional description of each part is as follows.

5.3.1. Ultrasound System Part


Front End recognizes the inserted probe and creates the High Voltage based on the system
information and application, and then generates 128 channel strong ultrasound through TX
focusing and radiates it into the human body.
The Echo signal returned from the human body is amplified by LNA (Low Noise Amp) and
compensated to the same size by TGC(Time Gain Compensator) and converted to digital signal
by ADC(Analog to Digital Converter).
Lastly, the 128 channel echo signal is added as one powerful signal in the RX focusing stage
and transmits to BE part.
BE part creates images for each mode such as 2D mode, Doppler spectrum and Color Doppler
with inputted echo signal and transmits it to PC.
Ultrasound System Part consists of the following components.
• PSA(Probe Select Assembly) board
• Main(Beamformer + Back End) board
• BP (Back Plane) board
• Rear IO board (Optional)
• CW board (Optional)

5.3.2. PC Part
Digital Scan Converter program of the PC integrates the three mode imaged such as B mode, C
mode and D mode into one image with diagnostic meaning and then corrects the image.
The Pre/Post processing part of Digital Scan Converter program processes images according to
various menus and add various digital information to the diagnosis image.
Finally, VGA part converts the digital image data to a video signal for peripheral devices.
PC Part consists of the following components.
• PC Module: Advantech, Com Express PC Module (SOM-5899)
• CPU: Intel i5-8400H
• RAM: 16GB (DDR4 8GB*2)
• Storage: M.2 NVMe SSD 512GB
• OS: Windows 10 64bit
• VGA Card : nVidia Geforce GTX1650
Chapter 5. Product Structure 5-7

5.3.3. User Interface Part


User I/F part consists of a LCD monitor and a Touch panel that users can check the ultrasound
image and operate the system.
The control panel is composed of Encoder Knob, Switch button, Touch key, External keyboard
and Track ball that perform user commands.
These control signals are transmitted to PC through MCU/HUB.
User Interface Part consists of the following components.
• Monitor: Kortek 21.5 inch LCD Monitor
• Touch panel: Kortex 14 inch Touch Screen
• Control Panel
• Trackball
• Key Matrix Board
• Alphanumeric K/B(Optional)

5.3.4. Power Part


The power part rectifies 77~286V AC voltage and converts it to DC 17V first.
Then, HV, LV, ATX Power Blocks convert this voltage into various DC voltages and supply them
to each board.

Power Part consists of the following components.


• Power
• Battery(Optional)
5-8 V6/XV6 Service Manual

5.4. Ultrasound System Part

5.4.1. PSA (Probe Select Assembly)


PSA board links the system and the probe and support 4 2D/3D probe ports and 1 CW probe
port. It reads Probe ID and transmit to PC and checks probe insertion status if it is inserted or
not. Also It controls the probe port selection.

[Figure 5.4 PSA Board Image]

[Figure 5.5 PSA Board Block Diagram]


Chapter 5. Product Structure 5-9

5.4.1.1. Specification
• 3/4 Probe Port Support (Optional)
• 1CW Pencil Probe Port Support(Option)
• Probe Insert Check, ID read, Port selection

5.4.1.2. Description

■ Probe Selection

It selects one of four probe ports and the probe port is selected through the user control
transmitted from BE board.

■ CW Pencil Probe Connection

The Static CW Probe port uses a LEMO Connector.


The connector is located on a separate connector board and is connected with a cable.
5 - 10 V6/XV6 Service Manual

5.4.2. Main board and CW board


Beamformer block is located between PSA and Back-End. It transmits the focused ultrasound
wave on a specific location of the human body and It amplifies the signal received from the
human body and converts it into a digital signal.
CW board removes the vessel wall’s Doppler noise in the CW data from Beamformer block and
separated the pure Red Blood Cell’s Doppler signal. And then It transfers the data to
Beamformer block after converting into digital data.
Lastly, the beamformed signal is transmitted to the Back End board to create 2D/C/D image.

[Figure 5.6 Main board Top side and CW board picture]

[Figure 5.7 Main board Bottom side picture]


Chapter 5. Product Structure 5 - 11

[Figure 5.8 Main board Layout]

[Figure 5.9 Main board block diagram]


5 - 12 V6/XV6 Service Manual

5.4.2.1. Specification

■ Main board – BF part

• TX Pulser + TR Switch 8ea


• AFE (LNA+TGC+ADC) 8ea
• BF FPGA 1ea
– RX Dynamic Aperture
– RX Apodization
– Synthetic Aperture
– Trapezoidal imaging
– Multi-line receiving
– TX Focal point
• CWD (Daughter board type, Optional)

■ Main board – BE part

• Focused RF data Interface


• BW Data Processing
• Color Doppler Data Processing
• IQ + Digital Signal Processing
• Digital Signal Processing
• Direct Memory Access
• PCI Express Interface
• RTC
• 3D Probe motor driver
• Diagnostic Signal Processing

■ Main board – IO part

• USB3.0 CP 3ea
• USB2.0 ECG 1ea
• USB2.0 DriveBay 1ea
• USB2.0 BW/Color Printer 2ea
• DP 1ea
• HDMI 2ea
• LAN 1
Chapter 5. Product Structure 5 - 13

5.4.2.2. Description
• Pulser
It transmits the High Voltage Pulse to the Probe after amplifying 128 channel Firing pulse
clock up to ±100V.
• TR Switch
It removes HV inputted from PSA in order not to pass to AFE.
• AFE
After amplifying the input echo signal to a sufficient size, it compensates to a certain size.
Then, it converts it into a digital signal for digital beamforming.
• BF FPGA
It consists of 128 TX channels and creates Firing pulse clock. It performs the digital
beamforming for 128 RX channel.
Lastly, it sums 128 RX channel data and transmits it to BE part.
• CWD
It removes only the clutter noise among CW I,Q signal inputted from LNA of the BF part.
• BE Part
MID processor creates BW Image, Spectral Doppler Image, and Color Doppler Image.
PC Host Interface transfers image data to PC through PCIe GEN3 x4 high speed serial bus.
• Motor controller
It consists of Step motor control part and Motor Drive part to drive the 3D probe.
When driving 3D Probe, the motor drive part drives the probe motor with SIN, COS and the return
path of this signal.
• BE - Internal I/O Part
It supports 1 USB 3.0 port and 1 DP port for CP & ECG Interface, 2 USB2.0 ports for printer.
• BE - User I/O Part
It supports 2 USB3.0 ports, 2 HDMI ports and 1 LAN port for the user.
It can be converted to an analog video signal using HDMI to DVI adapter.
5 - 14 V6/XV6 Service Manual

5.4.3. Rear I/O board


The rear I/O board supports S-VHS and Audio port. It is applied to some countries that require
analog signals as a default option, and other countries are provided as an option.

[Figure 5.10 Rear I/O board Image (Option)]

[Figure 5.11 Rear I/O board block diagram]

5.4.3.1. Specification

■ Rear I/O board (Optional)

• S-VHS 1ea
• Audio

■ Rear I/O board is optional.

■ Only certain countries that require analog signals are applied by default.
NOTE
.
Chapter 5. Product Structure 5 - 15

5.5. PC Part

5.5.1. PC Module
It supports High speed serial bus to receive BE image data and peripheral port for User
interface. It makes various application images with the inputted image data from BE board and
add all kind of digital information on the image and converts into the digital video signal for
monitor.

5.5.1.1. Specification
• PC Module: Advantech, Com Express Module (SOM-5899)
• CPU: Intel(8th core) i5-8400H
• RAM: 16GB DDR4(8GB*2)
• Storage: M.2 NVMe SSD 512GB
• OS: Windows 10 64bit
• VGA: nVidia Geforce GTX1650
• Monitor: 21.5 inch LG display(1920X1080)

[Figure 5.12 PC Module / VGA Card / M.2 NVMe SSD ]


5 - 16 V6/XV6 Service Manual

5.5.2. Software Digital Scan Converter(DSC)


DSC integrates image data for each mode into one image, processes various images according
to the application, and outputs them to VGA card.
The functions made by each part are as follows.

[Figure 5.13 DSC Block Diagram]

5.5.2.1. Specification
• Digital Scan Conversion
• Cine
• Ezexam+
• Zoom
• Edge Enhancement
• Frame Average
• MultiVision(SCI)
• ClearVision(SDMR)
• S-QuickScan
• Elastoscan
• Auto IMT
• Post Measurement
Chapter 5. Product Structure 5 - 17

5.5.2.2. Description
DSC consists of 4 software blocks.
Acquisition block acquire image data from BE FPGA.
Pre/Post filter makes application images such as Quick Scan, Multi vision, Frame Average,
CINE and etc.
Scan Conversion block integrates three mode images such as BW Image, Spectral Doppler
Image, Color Doppler Image as one image with diagnosis meaning .
5 - 18 V6/XV6 Service Manual

5.6. Control Panel

5.6.1. Control Panel


Control panel interfaces between user and the system. MCU receives input data from various
input devices such as 15 Encoders, 21 key switches, Track ball, and Alphanumeric keyboard,
and then transmits it to PC. It also controls on/off of 102 LEDs.

[Figure 5.14 Control Panel’s composition ]


Chapter 5. Product Structure 5 - 19

[Figure 5.15 Control Panel Block Diagram]


5 - 20 V6/XV6 Service Manual

5.6.1.1. Specification

■ CP- board

• Encoder 15EA
• Tack- Switch 21EA
• LED 102EA
• Trackball

■ CP- Docking board

• USB2.0 HUB
• PIC18F96J94(8bit Microcontroller)
• MSP(Multi-Stream Transport) HUB
• Encoder Interface
• Tack SW Interface
• Trackball Interface
• Keyboard Interface
• LED backlight control

■ CP-Switch board

• Tack SW of button around Track Ball


• LED Output of button around Track Ball

■ CP-USB board

• User USB 2 Port

5.6.1.2. Description
MCU (PIC18F96J94) scans the key values of Track Ball, Encoder, and Key SW while
interfacing between the control panel and the PC and transmits it to the PC. It controls
On/Off of Button Backlight LED.
MST (Multi-Stream Transport) HUB receives the Display Port signal from the PC and divides
the output to the connected Main monitor and Touch monitor.
Chapter 5. Product Structure 5 - 21

5.6.2. Main Monitor


23.8 Inch KORTEX Full HD LCD Panel is applied.

[Figure 5.16 Main Monitor]

5.6.2.1. Specification

Item SPECIFICATION UNIT

Display area 21.5 Inch Full HD panel (KORTEX) inch

Number of Pixels 1920 x 1080 pixels

Pixel Pitch 247.95(per one triad) x 247.95 um

Electrical Interface Dual channel LVDS (8 bit)

Display colors 16.7Million colors (6bit + A-FRC)

White Luminance (Center) 250 (Typ) Cd/m2


5 - 22 V6/XV6 Service Manual

5.6.3. Touch Panel


14 Inch KORTEX LCD Panel is applied.

[Figure 5.17 Touch Panel]

5.6.3.1. Specification

Item SPECIFICATION UNIT

Display area 14 Inch Full HD KORTEX Panel inch

Number of Pixels 1920 x 1080 pixels

Pixel Pitch 0.1611(H) x 0.1611(V) mm

Electrical Interface eDP1.3

Display colors 16.7M(6 bit + FRC)

Viewing angle −85∼+85°(H), −85∼+85°(V)


Chapter 5. Product Structure 5 - 23

5.7. Power Part

5.7.1. PSU / Battery / Battery charger

[Figure 5.18 PSU Image]

[Figure 5.19 Battery Image]

[Figure 5.20 Battery charger Image]


5 - 24 V6/XV6 Service Manual

5.7.2. Power Block Diagram

[Figure 5.21 V6/XV6 Overall Power Block Diagram]


Chapter 5. Product Structure 5 - 25

[Figure 5.22 V6/XV6PSU HV part Layout]

[Figure 5.23 V6/XV6 PSU LV+ATX part Layout]


5 - 26 V6/XV6 Service Manual

5.7.2.1. Specification and Description

■ Input

V6/XV6Power is designed for use from AC 100V to 240V Input voltages.

Input Voltage Input Frequency Input Power

100 ~ 240V 50/60Hz 800VA

[Table 5.1 V6/XV6 Input part specification]

■ Output

PSU(AC-DC power Module) makes a variety of DC voltages for the entire system.
The output voltages generated by PSU is as follows.

Part Output Output Current

Live SB_5V(5.1) 0.2A

12VD2(12.2) 10A

ATX 12VD(12.2) 9A

24VM(24) 1.5A

3.8VA(3.9) 12.5A

LV 5.5VA(5.4) 6.5A

N5.5VA(-5.7) 0.5A

CW(18V) 0.7A

80V(80V) 0.2A

N80V(-80V) 0.2A

HV 70V(70V) 0.2A

N70V(-70V) 0.2A

100V(100V) 0.05A

N100V(-100V) 0.05A

[Table 5.2 V6/XV6 Output voltage specification ]


Chapter 5. Product Structure 5 - 27

■ Battery

V6/XV6has two 6900 mAh Lithium-ion batteries and you can run for 40 minutes under the
worst condition like 3D live scan.
• 3D Live scan time : Min 60 minutes
• Sleep mode sustaining time with battery : 70 hour
• Fast Boot time : Max 25 second

Out Voltage Capacity Power

DC 14.4V 6900mAh 99W

[Table 5.3 V6/XV6 Battery 1ea specification]


5 - 28 V6/XV6 Service Manual

5.7.3. Power Status LED

V6/XV6PSU is equipped with 9 Power Satus LEDs that show the overall voltage status of the
system. The service engineer can diagnose whether there is a problem in the system power by
the LED status.

5.7.3.1. Power Status LED Composition and Location

The Power Status LED consists of 9 and displays the status of the 8 voltages supplied from the
PSU to the system. It is located next to the input switch on the back of the system, and you can
conveniently check the LED status from the outside without disassembling the device.

[ Figure 5.24 V6/XV6 PSU’s Power Status LED ]


Chapter 5. Product Structure 5 - 29

5.7.3.2. Check Point by Power Status LED status

LED Status Check


LED
On(Green) OFF Blink Point

Live Input AC/STB Power OK Input AC/STB Power fail 1. PSU

· 1 blink : UVP/OCP
· 2 blink : OVP 1. PSU
Status PSU output OK PSU output fail
· 3 blink : OTP 2. Each board
· 4 blink : Fan Error

· Blink every1.5sec: Charging


BATT #1 Battery OK Battery fail or not installed 1. Battery
· Blink every0.2sec: Fail

· Blink every1.5sec: Charging


BATT #2 Battery OK Batteruy fail or not installed 1. Battery
· Blink every0.2sec: Fail

· Blink every1.5sec: Charging


BATT #3 Battery OK Batteruy fail or not installed 1. Battery
· Blink every0.2sec: Fail

· On when using Mux Probe If the LED is off, · 1 blink : UVP/OCP 1. PSU
Mux
· FHV OK for PSA/Probe FHV(±100V) fail · 2 blink : OVP 2. Probe

· On during Bmode If the LED is off, · 1 blink : UVP/OCP 1. PSU


B mode
· HV1 OK for B mode HV1(±80V) fail · 2 blink : OVP 2. BF board

· On during C/D mode If the LED is off,, · 1 blink : UVP/OCP 1. PSU


C/D mode
· HV2 OK for C/D mode HV2(±70V) fail · 2 blink : OVP 2. BF board

1. PSU
· On during CW mode If the LED is off, · 1 blink : UVP/OCP
CW mode 2. BF board
· CW power OK CW Power fail · 2 blink : OVP
3. CW board

[Table 5.4 Check Point by the Power Status LED ]


5 - 30 V6/XV6 Service Manual

5.8. Interconnect Diagram

5.8.1. V6/XV6Interconnect Diagram

[Figure 5.25 Interconnect Diagram]


Chapter 5. Product Structure 5 - 31

5.8.2. Cable Description


No Parts Code Cable From To Remark Q'ty

1 MI39-02581A USB3.0_LEFT BE board CP Docking board New 1

2 MI39-02522A DP BE board CP Docking board New 1

3 MI39-02526A GEL_PWR_R CP Docking board Gel Warmer New 1

4 MI39-02525A PRINTER_USB BE board BW/Color printer New 2

5 MI39-02523A CP PWR BE board CP Docking board New 1

6 MI39-02025A ECG_USB_CBL BE board ECG Common 1

7 MI39-02229A COLOR PRINT_PWR POWER(SMPS) BW/Color printer Common 2

8 MI39-02290A TRACKBALL_DATA CP board Track Ball Common 1

9 MI41-01741A TRACKBALL_FPCB CP board Common 1

10 MI39-02306A GEL_PWR CP Docking board Gel Warmer Common 1

11 MI39-02298A ALPHAKEY_DATA_CBL CP Docking board ALPHAKEY Board Common 1

12 MI39-01378A AC CORD (POWER) POWER(SMPS) Wall Power Common 1

13 MI39-02459A BOTTOM_FAN EXT POWER(SMPS) BOTTOM FAN Common 1

14 MI39-02475A SPEAKER_EXT CP Docking board Speaker Common 2

15 MI39-02480A DRIVE USB(OPTION) BE board Drive bay Option 1

16 MI39-02518A DRIVE POWER(OPTION) BE board Drive bay Option 1

17 MI39-02462A MID_FAN_ASSY1 POWER(SMPS) MID FAN Common 1

18 MI39-02461A MID_FAN_ASSY2 POWER(SMPS) MID FAN Common 1

19 MI39-02460A BOTTOM_FAN BOTTOM FAN EXT BOTTOM FAN Common 1

20 MI39-02483A TOUCH_USB CP Docking board Touch Monitor Common 1

21 MI39-02482A TOUCH_POWER CP Docking board Touch Monitor Common 1

22 MI39-02529A TOUCH HDMI CP Docking board Touch Monitor New 1

23 MI39-02524A MONITOR PWR ARM Main Monitor CP Docking board New 1

24 MI39-02521A MAIN ARM HDMI Main Monitor CP Docking board New 1

[Table 5.5 Signal cable description]


Chapter 6
6. Service Mode

제6장. Service Mode

Service Mode
6.1. Service Mode ............................................................................................ 3
6.1.1. How to enter Service Mode ................................................................................... 3

6.2. Service Information .................................................................................. 4

6.3. Configuration ............................................................................................ 5


6.3.1. Keyboard/Region ................................................................................................... 5
6.3.2. TCP/IP ................................................................................................................... 6
6.3.3. Option .................................................................................................................... 7
6.3.4. User Account ......................................................................................................... 8
6.3.5. Printer .................................................................................................................... 9
6.3.6. RMS ..................................................................................................................... 10
6.3.7. RIS Browser ........................................................................................................ 11
6.3.8. AppLocker ........................................................................................................... 12
6.3.9. MobileExport Serve IP ......................................................................................... 13
6.3.10. SonoSync RIO ..................................................................................................... 14
6.3.11. System Status Notifier ......................................................................................... 15
6.3.12. Touch Layout ....................................................................................................... 16
6.3.13. Vet Mode ............................................................................................................. 17

6.4. System ..................................................................................................... 18


6.4.1. Change Information ............................................................................................. 18
6.4.2. Upgrade Software ................................................................................................ 19
6.4.3. Upgrade Hardware .............................................................................................. 20
6.4.4. Window Explorer ................................................................................................. 20
6.4.5. Specialist Mode ................................................................................................... 21
6–2 V6/XV6 Service Manual

6.5. Diagnostic ............................................................................................... 22


6.5.1. Control Panel Test ............................................................................................... 22
6.5.2. Keyboard Test ..................................................................................................... 22
6.5.3. Built in Self Test ................................................................................................... 23
6.5.4. Monitor Test ......................................................................................................... 24
6.5.5. Binary Verification Tool ........................................................................................ 25
6.5.6. Virus Scan ........................................................................................................... 26
6.5.7. Probe Element Check ......................................................................................... 27

6.6. Backup and Restore ............................................................................... 28


6.6.1. Backup ................................................................................................................ 28
6.6.2. Restore ................................................................................................................ 28
6.6.3. Log & Audit Trail Backup ..................................................................................... 29

6.7. Report ...................................................................................................... 30


6.7.1. Report ................................................................................................................. 30

6.8. Demo Play ............................................................................................... 31


6.8.1. Demo Play........................................................................................................... 31
6.8.2. Extended Mode ................................................................................................... 32
Chapter 6. Service Mode 6–3

6.1. Service Mode


This chapter describes about V6’s Service Mode.

6.1.1. How to enter Service Mode


① Press [Setup] and Select [Service] in the menu on the left and click [service] button.
② Set User type to [Service] on the Service Login box.
③ Run GSIS App on your smart phone and Select [Receptionist management]
④ Click [OTP] in Option/OTP
⑤ Choose s reason for entering the Service Mode.
⑥ Click QR code icon in QR/Bar code search.
⑦ When your smartphone camera is turned on, take the QR code of the system.
⑧ Then you will see 6 digit OTP password.
⑨ On the system screen, insert your GSIS ID and 6-digit OTP code and enter Service mode.

[Figure 6.1 Service Mode entering procedure]


6–4 V6/XV6 Service Manual

6.2. Service Information


Service Information shows H/W & S/W version information of the system and version
information of each board.

[Figure 6.2 Service Information Screen ]

• Workstation: shows H/W&SW version information of the system.


• License Information: shows the license information used in the SW.
• Atom Information: shows Atom information.
Chapter 6. Service Mode 6-5

6.3. Configuration

6.3.1. Keyboard/Region

[Figure 6.3 Keyboard/Region screen]

• Keyboard Type: International, French, German...

When the keyboard type is changed, a restart of the system is required.

CAUTION

• Region: Regional setting(WW, US, CN, RU)


• Apply : the set region is applied.
6–6 V6/XV6 Service Manual

6.3.2. TCP/IP
• Commands: PC’s IP config information is displayed on the right screen.
• LAN: Set wired network connection properties.
• WLAN: Set wireless network connection properties.
• Support: New device addition, driver installation and device manager execution.

[Figure 6.4 TCP/IP screen]


Chapter 6. Service Mode 6-7

6.3.3. Option
This is the menu to manage the application that the License is registered.
• Import Options: Install Option using the USB that option information is saved.
• Install: Install the selected Option.
• Uninstall: locks the option that it is already installed.

[Figure 6.5 Option screen]


6–8 V6/XV6 Service Manual

6.3.4. User Account


This is a function to limit system login authority and menus that can be used.
• UserAccount Reset: reset User Account information. (Hospital administrator)
• Administrator password Reset: reset Administrator password (HQ)
• User Account Manager : A window to set User Account information is created.

[Figure 6.6 User Account screen]


Chapter 6. Service Mode 6-9

6.3.5. Printer
• Printer: set default printer
• Properties: set the printing method for the selected printer.
• Add Printer: Add new printer driver except default printers.
Reboot is required to install the new printer driver.

[Figure 6.7 Printer screen]


6 – 10 V6/XV6 Service Manual

6.3.6. RMS
Remote Maintenance System : Remotely diagnose system defects and set up the system.
RMS ServiceApplication Run: run RMS Application by pressing [Run] button..

[Figure 6.8 RMS screen]


Chapter 6. Service Mode 6 - 11

6.3.7. RIS Browser


RIS Browser: A browser that allows you to view patient information remotely accessing the
medical information system (worklist).
RIS Browser On/Off: Adds or removes RIS Browser button in the patient information window.

[Figure 6.9 RIS browser screen]


6 – 12 V6/XV6 Service Manual

6.3.8. AppLocker
AppLocker : This is a function that restricts files not to execute to prevent arbitrary software
installation.
• Default setting is set to Off (Tender specification).

[Figure 6.10 AppLocker]


Chapter 6. Service Mode 6 - 13

6.3.9. MobileExport Serve IP


It is a function to set the LAN cart when installing Hello Mom.
• MobileExport Server IP Setting

[Figure 6.11 MobileExport Serve IP screen]


6 – 14 V6/XV6 Service Manual

6.3.10. SonoSync RIO


It is a function that can support image setting remotely.
• SonoSync Remote Image Optimization
• Remotely receive the exam file and push the file

[Figure 6.12 SonoSync RIO screen]


Chapter 6. Service Mode 6 - 15

6.3.11. System Status Notifier


It is a function that transmits information with a specific format(TCP/IP) when changing
Preset/App or when an error occurs.
• It is used to monitor multiple systems in hospitals.(Tender specification)

[Figure 6.13 System Status Notifier screen]


6 – 16 V6/XV6 Service Manual

6.3.12. Touch Layout


It is a function that can map Knob button according to application.

• When changing the default layout of the touch button to GI or OB/GYN Layout, delete the
user setting and change the default touch layout to the set layout.

[Figure 6.14 Touch Layout screen]


Chapter 6. Service Mode 6 - 17

6.3.13. Vet Mode

it switches from Human Mode to Vet mode or from Vet mode to Human mode.
• Press the Vet mode Button and enter the Enter Mode Key.

[Figure 6.15 Vet Mode]


6 – 18 V6/XV6 Service Manual

6.4. System

6.4.1. Change Information


• System S/N: System Serial Number change

Service Core’s authority is required to change the system S/N..

NOTE

• Model Name: Model Name change

[Figure 6.16 Change Information]


Chapter 6. Service Mode 6 - 19

6.4.2. Upgrade Software


Upgrade the ultrasound UAI SW.

[Figure 6.17 Upgrade Software screen]

1. Insert USB memory containing UAI file.


2. Press [Start] button and run software upgrade.
3. When UAI Manager window pops up, click Install button.

[Figure 6.18 UAI Manager window]

4. SW installation takes about 15 minutes.

When rebooting the system after completing the UAI installation,


USB memory must be removed.
CAUTION
6 – 20 V6/XV6 Service Manual

6.4.3. Upgrade Hardware


Upgrade system Hardware FPGA.

[Figure 6.19 Upgrade Hardware screen]

6.4.4. Window Explorer


Run Window Explorer.

[Figure 6.20 Window Explorer screen]


Chapter 6. Service Mode 6 - 21

6.4.5. Specialist Mode


Specialist Mode

[Figure 6.21 Window Explorer screen]


6 – 22 V6/XV6 Service Manual

6.5. Diagnostic

6.5.1. Control Panel Test


Test all Buttons of Control Panel.

[Figure 6.22 Control Panel Test screen]

6.5.2. Keyboard Test


Test all buttons of Keyboard.

[Figure 6.23 Keyboard Test screen]


Chapter 6. Service Mode 6 - 23

6.5.3. Built in Self Test


It is a function that can diagnose the system hardware automatically with a software.
For more detailed information, please refer to “ Chapter7, Self Diagnostic “ in Service manual.

[Figure 6.24 BIST screen]


6 – 24 V6/XV6 Service Manual

6.5.4. Monitor Test


Test the monitor color and brightness.

[Figure 6.25 Monitor screen]


Chapter 6. Service Mode 6 - 25

6.5.5. Binary Verification Tool


In order to check whether the software is malfunctioning or not, the file integrity of the SSD is
checked. It checks the Windows OS file in the C drive and the Ultrasound application file in the
D drive. Data files are not checked.
If the diagnosis result shows that the software is damaged, the location where the Fail occurred
and the final result are displayed as Fail on the screen.

[Figure 6.26 Binary Verification Tool Menu screen]

[Figure 6.27 Test result screen]


6 – 26 V6/XV6 Service Manual

6.5.6. Virus Scan


Run Window defender.

[Figure 6.28 Virus Scan screen]


Chapter 6. Service Mode 6 - 27

6.5.7. Probe Element Check


• Test: Inspect the element of the currently set probe.
• Log Export: Save the log to USB.

Available only when Region is set to US.

CAUTION

• Alert Popup: When [On] is checked, an error message pops up when an error occurs as a
result of the probe element diagnosis.

[Figure 6.29 Probe Element Check Screen


6 – 28 V6/XV6 Service Manual

6.6. Backup and Restore

6.6.1. Backup
Back up the selected USER file. External storage such as USB is required.

[Figure 6.30 Backup screen]

6.6.2. Restore
Restore the selected USER data file. You should have previously backed up data.

[Figure 6.31 Restore screen]


Chapter 6. Service Mode 6 - 29

6.6.3. Log & Audit Trail Backup


Copies all log and audit trail files saved in the system to the storage device. Before you start,
you need to have an external storage device such as USB..

[Figure 6.32 Log Backup screen]


6 – 30 V6/XV6 Service Manual

6.7. Report

6.7.1. Report
It shows the temperature, BIST, connectivity, and error log information stored in the system.

[Figure 6.33 Report screen]


Chapter 6. Service Mode 6 - 31

6.8. Demo Play

6.8.1. Demo Play


• Play: Play the already saved image.
• Import: Load images saved in external storage devices such as USB.
• Export: Saves the image to the path specified in an external storage device such as USB.

[Figure 6.34 Demo Play screen]


6 – 32 V6/XV6 Service Manual

6.8.2. Extended Mode


Function to activate Preset for bidding
• activate Preset for bidding
• Some of the setting values of the existing preset are changed for bidding.
• Changed to Off when rebooting.

[Figure 6.35 Extended mode screen]


Chapter 7

Self Diagnostic
7.1. PURPOSE FOR USAGE ......................................................... 2

7.2. SELF-DIAGNOSTIC TEST ITEM ............................................ 2


7.2.1 Dummy Probe ................................................................................... 2

7.3. SELF-DIAGNOSTIC ENTRY ................................................... 3


7.3.1 Service mode entry ........................................................................... 3

7.4. SELF-DIAGNOSTIC RUN ....................................................... 4

7.5. SELF-DIAGNOSTIC RESULT ................................................. 6

7.6. SELF-DIAGNOSTIC LOG ....................................................... 7


7.6.1 How to check Self-Diagnostic Log .................................................... 7

7.7. SELF-DIAGNOSTIC END........................................................ 9

7.8. TEST ITEMS ACCORDING TO DUMMY PROBE ................. 10

7.9. SELF-DIAGNOSTIC TEST ITEM DESCRIPTION ................ 11


7.9.1 Power .............................................................................................. 11
7.9.2 Bus / Analog / CW / Memory / Motor / Fan / PSA .......................... 12

7.10 REMOTE MAINTENANCE SYSTEM.................................... 13


7 - 2 V6/XV6 Service Manual

7.1. Purpose for usage


The V6/XV6's Self Diagnostic function automatically diagnoses abnormalities for each board in
the system and displays the results. Based on the results, the service engineer can accurately
diagnose the condition of the system.

■ Test results may differ depending on the timing and number of


Self-Diagnostics. Even if the result is displayed as Check, it does
not necessarily mean that the system is defective.
■ The service engineer uses the results of Self-Diagnostic only as a
CAUTION reference for service, and the final decision must be made by the
engineer.

7.2. Self-Diagnostic Test Item


Self-Diagnostic consists of 13 test items to check the operation status of the system's power,
bus, and board.

7.2.1 Dummy Probe

Dummy Probe is a Jig (Tester) used for PSA Board inspection. In case of diagnosing system
without dummy probe, only other items except PSA board can be diagnosed.
For details, refer to page 11 “ Test Items According to Dummy Probe”.

Dummy Probe has not been provided since 2020. For related
matters, please contact the Area Service Manager of the head
office.
NOTE
Chapter 7. Self Diagnostic 7-3

7.3. Self-Diagnostic Entry


Self- Self-Diagnostic can be run in Service Mode. Since OTP Password is required to enter the
Service Mode, please contact the Service Engineer who has the password.

7.3.1 Service mode entry

① Enter Service Mode

② Setup  Service  Diagnostic  Built In Self Test


7 - 4 V6/XV6 Service Manual

7.4. Self-Diagnostic Run


Install the dummy probe after removing all probes from the system.
If there is no dummy probe, run the Self-Diagnostic without the dummy probe.

∙ No Dummy Probe : When testing without the dummy probe, continue to click
[Enter] key when a dummy probe insertion message pops up.

∙ Dummy Probe 1EA : When the dummy probe message pops up, insert the dummy probe
into the probe port.

① In Built In Self Test menu, press [Start] Buton.

② When the Built-In Self Test box pops up, select the test item from the Menu and press
[RUN] button.

■ You cannot test PSA board without the Dummy probe.


■ Do not select the PSA test for 3 Port and PSA test for 4 Port in
the menu.
CAUTION
Chapter 7. Self Diagnostic 7-5

[PASS]

[FAIL]

When the Self-Diagnostic is running, the test result for each test item is displayed on the
screen. The test result is displayed in real time, and the test result is summarized and
displayed when the final test is completed .
7 - 6 V6/XV6 Service Manual

7.5. Self-Diagnostic Result

■ The result of Self-Diagnostic may be different depending on the


timing and frequency of the testing.
■ Even if the diagnosis result is displayed as “Check”, it does not
necessarily mean a defect in the system. After confirming the
CAUTION detailed measurement value, check whether it is defective after
checking the degree out of the reference range.

Self-Diagnostic’s diagnosis result means as follows.


∙ Pass : The board or part corresponding to the test item operates normally.
∙ Check : The board or part is out of the reference range.
The diagnosis result is saved in the system path as below.

→ Location : E: / Log / BIST

Enter the Storage path by pressing [Explorer] Button.


Chapter 7. Self Diagnostic 7-7

7.6. Self-Diagnostic Log


7.6.1 How to check Self-Diagnostic Log

① Restart the system after completing Built In Self Test.


② Run Window Explorer in Service mode
③ In the Window Explorer screen, go to [E:/Log/BIST] and check detailed results.

󠇍 Folder name
• The folder name is created in the form of YYYYMMDD_HHMMSS .
• Every time the test is run, one folder is automatically created, and Log files are saved in
the folder..
7 - 8 V6/XV6 Service Manual

󠇍 File name:
• [HWINFO].ini : the system’s H/W & S/W information.
• 00_STDOUT.txt : Test result - details
• 01_RESULT.txt : Test result - Result
• 02_SUMMARY.txt : Test result - Summary
• 03_ANALYSIS.txt : Test result – Analysis content for each test
Chapter 7. Self Diagnostic 7-9

7.7. Self-Diagnostic End

■ If the system does not shut down normally after completing the
test, the system may not boot normally.
■ Be sure to shut it down according to normal shutdown
CAUTION procudure.

When Built-In Self Test is complete, click [Close] button to end the test.
7 - 10 V6/XV6 Service Manual

7.8. Test Items according to Dummy Probe


○ : Items that can be tested. X : Items that cannot be tested.

No One
No Test Item Sub Test Item
Dummy Dummy

1 Power Voltage ○ ○

2 Power SMPS ○ ○

3 Power CAP SW ○ ○

4 Battery Battery Communication ○ ○

5 Battery Battery Operation ○ ○

6 Bus FPGA Configuration ○ ○

7 Bus Board Temperature ○ ○

8 Bus I2C ○ ○

9 Bus System bus ○ ○

10 Bus GTH Bus ○ ○

11 Analog ADC ○ ○

12 Analog RX ○ ○

13 Analog TX ○ ○

14 Analog ATGC ○ ○

15 CW CW ○ ○

16 Memory DDR3 Memory ○ ○

17 PSA PSA Port for 4 port X ○

18 PSA PSA Port for 3 port X ○

19 Motor Motor Driver ○ ○

20 FAN System FAN ○ ○


Chapter 7. Self Diagnostic 7 - 11

7.9. Self-Diagnostic Test Item Description


7.9.1 Power

Parameter Check
Sub Item Description Parameter
Description Point
FPGA0_VCCINT 0.95V Main
FPGA1_VCCINT 0.95V Main
FPGA2_VCCINT 0.95V Main
FPGA1_AUX_1.8V 1.8V Main
FPGA2_AUX_1.8V 1.8V Main
FPGA0_AUX_1.8V 1.8V Main
FPGA_IO_1.8V 1.8V Main
FPGA_IO_3.3V 3.3V Main
VCCINT_0.95V 0.95V Main
MGTAVCC_1.0V 1.0V Main
Measure Main
Voltage BE_DCDC_1.2V 1.2V Main
powers.
BE_DCDC_1.8V 1.8V Main
BE_DCDC_2.5V 2.5V Main
BE_DCDC_3.3V 3.3V Main
MAIN_12V2 12V Main
FIX_HV_100V 100V PSU
FIX_HV_N100V -100V PSU
BHV_80V 80V PSU
BHV_N80V -80V PSU
CHV_70V 70V PSU
CHV_N70V -70V PSU
HV_80V 80V PSU
HV_N80V -80V PSU
HV_70V 70V PSU
HV_N70V -70V PSU
12VD2 12V PSU
5VD 5V PSU
Check PSU output 24VM 24V PSU
SMPS
voltage SB_5V 5V PSU
1.6VA 1.6V PSU
2.3VA 2.3V PSU
3.8VA 3.8V PSU
5.5VA 5.5V PSU
N5.5VA -5.5V PSU
12VA 12V PSU
Battery

Battery Communication RSOC, Temperature, Voltage, Current check Battery


Battery Operation
7 - 12 V6/XV6 Service Manual

7.9.2 Bus / Analog / CW / Memory / Motor / Fan / PSA

Check
Test Item Sub Item Description Parameter
Point

BF/BE FPGA Configuration check


FPGA - Main
BF/BE FPGA Version check

Temperature Main /PSA temperature check - Main, PSA


Bus
I2C PSA/BF/BE EEPROM I2C Bus check - Main, PSA

System Bus BF/BE FPGA System Bus check - Main, PC

GTH Bus BF/ BE FPGA GTH Bus Sync check - Main

ADC Test AFE IC output data check - Main

RX Test 192 channel’s signal check Rx signal 5MHz Main

Analog HV1 Tx signal


TX Test 192 channel’s Pulser check Main
HV2 Tx signal

ATGC Test 192 channel’s ATGC check 192CH ATGC Main

CW CW Test AFE IC’s CW mode & ADC check I, Q data CW

Memory Memory Write & Read the pattern in DDR Main

MOT_SIN_DRV_V

MOT_SIN_RTN_V
Motor Motor Test Motor drive DAC/AMP check Main
MOT_COS_DRV_V

MOT_COS_RTN_V

FAN FAN Test System’s 4 Fan check Fan current FAN

PSA PSA Test PSA’s signal line check - PSA


Chapter 7. Self Diagnostic 7 - 13

7.10 Remote maintenance System


V6 is equipped with SAMSUNG Remot Maintenance System(RMS). This function allows
SAMSUNG’s service engineers to diagnose the entire system remotely. If the system is
connected to the Internet, you can do various service work such as Board defect diagnosis ,
Function check, and Log file collection through the RMS.
However, to use the RMS function, Customer’s agreement and Access password are required.
For detailed information, please contact your Samsung’s service engineer.

■ RMS function is activated after a customer sighs the agreement in


the system.
■ Remote access is only available when the system is connected to
the Internet and the service platform is properly configured.
NOTE
■ Only Samsung’s professional engineers can access this RMS.

[Custome Agreement Signing Screen for RMS Function Use]


7 - 14 V6/XV6 Service Manual

[ Remote Access(Teamviewer) Execution Screen]


Chapter 8
7. 고장 진단

제7장. 고장 진단

Troubleshooting
8.1. Power......................................................................................................... 2
8.1.1. Power doesn’t Turn on........................................................................................... 2
8.1.2. Power does not turn off ......................................................................................... 2
8.1.3. Power Turns Off by itself ....................................................................................... 2
8.1.4. PSU Stand-Alone Self-diagnostics ........................................................................ 3
8.1.5. BP board Self-diagnostics ..................................................................................... 4

8.2. Monitor ...................................................................................................... 5


8.2.1. Nothing is displayed on the screen ....................................................................... 5
8.2.2. Screen is discolored .............................................................................................. 5

8.3. Error Message........................................................................................... 6


8.3.1. Error Occurs during Booting .................................................................................. 6

8.4. Image ......................................................................................................... 7


8.4.1. 2D Mode: There is no Image Echo or Image Format ............................................ 7
8.4.2. Line(noise) Appear in 2D image ............................................................................ 7
8.4.3. M, C, PW, CW Mode failure .................................................................................. 7

8.5. Troubleshooting Tree ............................................................................... 8


8.5.1. System does not power on. ................................................................................... 8
8.5.2. No image on the monitor. ...................................................................................... 9
8.5.3. Noise in the image area....................................................................................... 10
8.5.4. Track Ball does not work ..................................................................................... 11
8.5.5. System does not turn off...................................................................................... 12
8-2 V6/XV6 Service Manual

8.1. Power

8.1.1. Power doesn’t Turn on


The power cord is not connected, or the output voltage and power connection of the PSU
(Power Supply Unit) are expected to be defective.

1. Make sure the power cord is properly plugged and the power switch is on.
2. Plug another device into the power outlet to see if it works.
– If the another appliance works, the power connection part has failed.
– If the another appliance does not work, the power outlet has failed
3. Turn on the power switch and check if the System Power Button is turned on.
– If the Power Button LED is on, it is more likely due to something other than the
connection.
– If the Power Button LED does not turn on, a power failure is expected.
4. Check PC and PSU.

8.1.2. Power does not turn off


The software is running or a software error may have occurred. PC and Main board failure is
expected.
1. Press and hold the Power button for at least 3 seconds and the power will be turned off
automatically. However, if the operation of the printer software does not stop or an OS error
occurs, the power does not turn off.
2. If the power doesn’t turn off after trying the method described in 1, it is possible that the PC e
and/or Main board failure is expected.

8.1.3. Power Turns Off by itself


It is likely that the Power Cord, PC, Main, Cable & Connector Pin (shorted), and/or PSU may be
defective.
1. Check the power code is properly plugged in and the power switch is on.
2. Connect a different appliance to the power outlet and check the power outlet whether it
works or not.
– If the different appliance works, the power connection part has failed.
– If the different appliance doesn’t work, the power outlet has failed.
3. Check the LED state of PSU.
4. If the symptom still appears even with methods 1 and 2, Main board, CP board and PSU are
expected to be defective.
Chapter 8. Troubleshooting 8-3

8.1.4. PSU Stand-Alone Self-diagnostics


V6 can check whether PSU is OK or not in a single unit state.

1. Turn off the circuit breaker of the PSU after turning off the system.
2. Pull out PSU after unplugging the power cord from the power outlet.
3. Select Dip switch No. 2 of the PSU to “ON”.
4. Turn on the circuit breaker of the PSU single unit after plugging the power code into the
power outlet.
5. If only one LED blinks continuously or no LED lights up, the PSU is fail.
※If the four LEDs below continue to light up sequentially, It’s a normal.
6. After checking the PSU, select Dip switch No.2 to “OFF” again.
※ The system will be automatically turned off unless you return the Dip Switch to it’s
original position.
8-4 V6/XV6 Service Manual

8.1.5. BP board Self-diagnostics


You can check if BP board or Main board is bad by the state of the 3 LEDs on the back side of
the system.
- Bothe the BP board and the Main board are normal, the LEDs doesn’t turn on.
- If all 3 LEDs are turned on, BP board is defective, but only 2 LEDs are turned on, Main
board is defective.
Chapter 8. Troubleshooting 8-5

8.2. Monitor

8.2.1. Nothing is displayed on the screen


It is expected that the HDMI cable or Monitor or PC is defective.
1. Check the state of the system with output of the printer.
2. Check the monitor connection state.
3. If the symptom still appears even with methods 1 and 2, the monitor and Main board are
expected to be defective. Check the PC and Main board.

8.2.2. Screen is discolored


It is expected that the Monitor or Monitor cable or VGA card is defective.
1. Check the monitor cable state.
2. If the symptom still appears even with method 1, check the Monitor and VGA.
8-6 V6/XV6 Service Manual

8.3. Error Message

8.3.1. Error Occurs during Booting


A temporary software error or product defect is expected.
1. Turn power off, and then turn it back on in 1-2 minutes.
3. If method 1 does not solve the problem, check when the error message occurred.
If the error occurs while WINDOWS is initiating, it is likely that the OS and/or PC have
(has) failed. If the error occurs after the logo is displayed, it is likely that the System
Software or Ultrasound System part has failed.
Chapter 8. Troubleshooting 8-7

8.4. Image

8.4.1. 2D Mode: There is no Image Echo or Image Format


Contact between the probe and the system may be poor, or PSU and Main board failure are
expected.
1. Check whether the contact between the probe and the system is poor or not.
2. Check the vibrating sound of the probe. If you can’t hear the sound, it is likely that the PSU
has failed.
3. If the problem is not solved even by the method 1 and 2 above, it is likely that the Main board
has failed.

8.4.2. Line(noise) Appear in 2D image


Power noise or PSA/Main board noise are expected.
1. Check whether the power outlet is shared with other products or not.
2. Noise may occur when used with a product that uses an electric motor
3. Check if the symptom also occurs in another room with a power outlet.
4. If noise occurs, it is a symptom of the power noise.
5. If the problem is not solved even by the method 1 and 2 above, it is likely that the Main board
has failed.

8.4.3. M, C, PW, CW Mode failure


Main board failure is expected.
8-8 V6/XV6 Service Manual

8.5. Troubleshooting Tree

8.5.1. System does not power on.

Power cord plug in NO Plug in the power


correctly? cord to the outlet.

YES

Circuit breakers NO
Circuit breaker ON
is on ?

YES
YES NO
Status LED is on ? Replace PSU

YES

NO Power cable YES


Power button LED Replace PC
is on ? connection is OK ?

NO

YES Reconnect Power cable

NO NO
CP LED turned orange? CP LED blinking? Check PC or SSD

YES YES

System boot normally Replace PSU


Chapter 8. Troubleshooting 8-9

8.5.2. No image on the monitor.

No Monitor/Touch image

NO NO
Monitor power/HDMI signal
LED Backlight is on ? Replace the cable or PSU
is OK ?

YES YES

Replace Monitor Replace Monitor


8 - 10 V6/XV6 Service Manual

8.5.3. Noise in the image area.

Noise on the image

YES Use a different power


Are there devices that line or remove noise
generate noise around? equipment

NO

Check BIST report


Move the probe to next port

NO

YES
Test the analog part
Is there still noise? Is BIST result pass ?
with BIST

NO YES

Check PSA board Check the probe failure


Chapter 8. Troubleshooting 8 - 11

8.5.4. Track Ball does not work

Trackball not working /


Cursor jumps

Cleaning the inside of the


trackball

NO
Does it work normally? Replace the trackball

YES

NO

Does it work normally? Replace the trackball cable


Normal

YES

NO
Normal Does it work normally?

YES

Normal

Check CP Module
8 - 12 V6/XV6 Service Manual

8.5.5. System does not turn off

Press the power button

NO Suspected software conflicts. reinstall SW.


Window closed? If the operating speed is slow, replace the PC and SSD.

YES

System shut down in NO System stop during shut NO


Blue screen error happen?
progress? down?

YES YES YES

System turn off Reinstall system SW - Check PC module


- Check SSD,VGA,RAM
Chapter 9
8. 분해 및 조립

제8장. 분해 및 조립

Disassembly
9.1. Caution ...................................................................................................... 2
9.1.1. Preparation ............................................................................................................ 2

9.2. Disassembling the system ...................................................................... 3


9.2.1. Front side disassembly .......................................................................................... 3
9.2.2. Rear side disassembly .......................................................................................... 4
9.2.3. Keyboard disassembly .......................................................................................... 6
9.2.4. Control Panel disassembly .................................................................................... 7
9.2.5. Touch Panel disassembly ...................................................................................... 8
9.2.6. Monitor disassembly .............................................................................................. 9
9.2.7. Monitor ARM disassembly ................................................................................... 10
9-2 V6/XV6 Service Manual

9.1. Caution
This chapter describes the procedure for disassembling V6 system. Refer to this chapter when
upgrading or servicing the system’s hardware.

■ Samsung Medison's customer service department or authorized engineers


are allowed to repair or replace the parts of the products.
WARNING ■ The interior of this product contains dangerous high-voltage electricity. Do not
disassemble the product. Otherwise, short-circuit may occur.

■ Do not wear an antistatic wrist strap while working with the product in
powered on state. Otherwise, you may sustain electrical injury.

9.1.1. Preparation
You will need the following items: Phillips (+) screwdriver, antistatic gloves, and antistatic wrist
strap.
Turn off the power of the product and disconnect the battery.

Wear antistatic gloves and wrist strap when disassembling or reassembling the product.
They help prevent accidents for the engineer and product failure caused by static
electricity
NOTE

[Figure 9.1 Antistatic Gloves and Wrist Strap]


Chapter 9. Disassembly 9-3

9.2. Disassembling the system

9.2.1. Front side disassembly


1. ①② First, lift up while pulling forward two hooks of the Foot Cover using the index finger on
both sides. Since there is no screws, it can be easily separated.
③ Hold the front cover with both hands as shown in the picture below and pull it forward.
Since there are no screws, it can be easily removed.
④ Unscrew 10 screws, then hold the handle of PSA board and pull it forward to separate it.
9-4 V6/XV6 Service Manual

9.2.2. Rear side disassembly


1. ① Unscrew 8 screws and separate Rear Cover. ② Disconnect 2 connectors first.
③ Unscrew 4 screws and separate 2 brackets and 2 cables.

⑤ Among 3 USB ports, Only very left port is USB3.0 and the other two are
USB2.0 Port. When assembling, USB3.0 LEFT cable must be inserted at the
very left port. If you insert it to other port, Power button may be out of order.
NOTE

⑥ Remove 4 cables. If there is no printer option, the cable are not provided.
Chapter 9. Disassembly 9-5

2. ⑦ Unscrew 11 screws and separate a plate.


⑨ Separate Main Board.
⑩ Unscrew 6 screws and separate Battery + Charger module.
⑪ Remove 4 screws and 1 cable. And then separate the Power module.
9-6 V6/XV6 Service Manual

9.2.3. Keyboard disassembly


①② First, open the Keyboard. ③ remove 1 USB cable and 4 screws at the bottom of the
control panel. ④ hold the keyboard rail with both hands and push it back to separate the
keyboard from a hanger and take it out to the front.
Chapter 9. Disassembly 9-7

9.2.4. Control Panel disassembly


1. Unscrew 10 screws at the bottom of the Control Panel and lift the control panel up.

2. ①② unscrew 8 screws at the bottom side to separate the Control Panel Top Cover.
③ separate Monitor Arm cover.
④ separate Top cover.

3. Remove 4 screws and 2 connectors. Then separate the Speaker module.


9-8 V6/XV6 Service Manual

9.2.5. Touch Panel disassembly


①② first, unscrew 8 screws on the bottom side as shown in the picture below.
④⑤ Hold the Monitor Arm Holder and separate it as shown in the picture.
⑥ Lift the Top cover up and separate it.
⑦ Unscrew 2 screws.
⑧ Hold on the Touch panel with both hands and pull it forward to separate it.
⑨ Hold the Touch panel with your left hand and disconnect the three connectors as shown
in the picture.
Chapter 9. Disassembly 9-9

9.2.6. Monitor disassembly


①② First, unscrew 2 screws. Tilt the monitor as shown below and remove a cover.
③ Remove 2 screws and 1 connector. Disconnect the monitor power cable.
④⑤⑥ Unscrew 4 screws and separate the monitor using your both hands.
9 - 10 V6/XV6 Service Manual

9.2.7. Monitor ARM disassembly


① Disconnect 3 connectors as shown in the picture below.
② Unscrew 2 screws of the cable holders and separate it.
③ Unscrew 2 screws in the monitor ARM.
④⑤Hold the monitor arm with your left hand and unscrew 2screws on the opposite side.

When working without holding the monitor arm, the arm may fall and hurt
the worker.
For safety, be sure to remove the screw while holding the monitor arm
CAUTION
with your hand.
Chapter 10

Probes
10.1. Probes ................................................................................. 2
10.1.2. Ultrasound Transmission Gel ......................................................... 14
10.1.3. Using Sheaths................................................................................ 16
10.1.4. Probe Safety Precautions .............................................................. 18
10.1.5. Cleaning, Disinfecting, and Sterilizing Probes ................................ 20
10.1.6. MPTEE Probe (Option) .................................................................. 35
10.2. Biopsy ............................................................................... 42
10.2.1. Biopsy Kit Components .................................................................. 43
10.2.2. Using Biopsy Kit ............................................................................. 44
10.2.3. Assembling the Biopsy Kit.............................................................. 47
10.2.4. Cleaning and Disinfecting the Biopsy Kit........................................ 52
10 - 2 V6/XV6 Service Manual

10.1. Probes
The probe is a device that sends and receives ultrasound for acquiring image data.
The system limits patient contact temperature to 43 degrees Celsius, and acoustic output values
to their respective U.S. FDA limits. A power protection fuse circuit protects against over-current
conditions. If the power monitor protection circuit senses an over-current condition, then the
drive current to the probe is shut off immediately, preventing overheating of the probe surfaces
and limiting acoustic output.

10.1.1.1. Probe List


The ultrasound image scanner uses probes to obtain graphic data of the human body and
then displays it on the screen. Always use application-specific probes in order to obtain the
best quality images. It is also important to use a preset that is best suited to the particular
organ being scanned.

■ Thermal Index (TI) Table


The Thermal Index (TI) is displayed in the Title area of the screen, and represents the
probability of temperature rise in a particular area of the body. The thermal index consists
of the following indices: Soft tissue (TIs), bone (TIb), and cranial bone (TIc). This product
automatically displays an appropriate thermal index for the current probe and application.
Probe types, applications, presets and thermal index available for this product are as
follows:
Thermal
Probes Application Preset
Index

Abdomen Abdomen, Renal, Bowel, Aorta TIs, TIb


MSK General, Superficial, Deep, Spine TIs, TIb
LA2-9S Small Parts Breast, Thyroid, Testicle TIs, TIb
Vascular Carotid, Arterial, Venous TIs, TIb
Pediatric Ped Abd, Ped Hip TIs, TIb
Abdomen Abdomen, Renal, Bowel, Aorta TIs, TIb
MSK General, Superficial, Deep, Spine TIs, TIb
LA3-14AD Small Parts Breast, Thyroid, Testicle TIs, TIb
Vascular Carotid, Arterial, Venous TIs, TIb
Pediatric Ped Abd, Ped Hip TIs, TIb
MSK General, Superficial, Superficial1, Deep TIs, TIb
LA3-22AI
Intraoperative General TIs, TIb
MSK General, Superficial, Superficial1, Deep TIs, TIb
Small Parts Breast, Thyroid, Testicle TIs, TIb
L3-22
Vascular Carotid, Arterial, Venous TIs, TIb
Pediatric Neo Head, Ped Abd TIs, TIb
CA1-7AD Abdomen Abdomen, Renal, Bowel, Aorta, Biopsy TIs, TIb
Chapter 10 Probes 10 - 3

Thermal
Probes Application Preset
Index

MSK General, Spine TIs, TIb


Vascular Arterial, Venous TIs, TIb
1st Trimester, 2nd Trimester, 3rd Trimester, Fetal Heart,
OB TIs, TIb
NT
GYN Uterus, Adnexa TIs, TIb
Pediatric Ped Abd, Ped Hip TIs, TIb
Urology Prostate, Bladder TIs, TIb
Abdomen Abdomen, Renal, Bowel, Aorta, Biopsy, Penetration TIs, TIb
MSK General, Spine TIs, TIb
Vascular Arterial, Venous TIs, TIb
1st Fetal Heart, 1st Trimester, 2nd Trimester, 3rd
OB TIs, TIb
CA1-7S Trimester, Fetal Heart, NT
GYN Uterus, Adnexa TIs, TIb
Thoracic Lung TIs, TIb
Pediatric Ped Abd, Ped Hip TIs, TIb
Urology Prostate, Bladder TIs, TIb
Abdomen Abdomen, Renal, Bowel, Aorta, Biopsy, Penetration TIs, TIb
MSK General, Spine TIs, TIb
Vascular Arterial, Venous TIs, TIb
1st Fetal Heart, 1st Trimester, 2nd Trimester, 3rd
OB TIs, TIb
CA1-7SD Trimester, Fetal Heart, NT
GYN Uterus, Adnexa TIs, TIb
Thoracic Lung TIs, TIb
Pediatric Ped Abd, Ped Hip TIs, TIb
Urology Prostate, Bladder TIs, TIb
Abdomen Abdomen, Renal, Bowel, Aorta, Biopsy TIs, TIb
MSK General, Spine TIs, TIb
Vascular Arterial, Venous TIs, TIb
1st Trimester, 2nd Trimester, 3rd Trimester, Fetal Heart,
OB TIs, TIb
CA3-10A NT
GYN Uterus, Adnexa TIs, TIb
Thoracic Lung TIs, TIb
Pediatric Ped Abd, Ped Hip TIs, TIb
Urology Prostate, Bladder TIs, TIb
Abdomen Abdomen, Renal, Bowel, Aorta TIs, TIb
CA4-10M
Vascular Carotid, Arterial, Venous TIs, TIb
10 - 4 V6/XV6 Service Manual

Thermal
Probes Application Preset
Index

Neo Head TIs, TIc


Pediatric
Ped Abd, Ped Hip TIs, TIb
OB 1st Trimester, 2nd Trimester, 3rd Trimester TIs, TIb
EA2-11AR GYN Uterus, Uterus2, Adnexa, Cervix, Penetration TIs, TIb
Urology Prostate, Bladder TIs, TIb
OB 1st Trimester, 2nd Trimester, 3rd Trimester TIs, TIb
EA2-11ARE GYN Uterus, Uterus2, Adnexa, Cervix, Penetration TIs, TIb
Urology Prostate, Bladder TIs, TIb
OB 1st Trimester, 2nd Trimester, 3rd Trimester TIs, TIb
EA2-11AV GYN Uterus, Uterus2, Adnexa, Cervix, Penetration TIs, TIb
Urology Prostate, Bladder TIs, TIb
OB 1st Trimester, 2nd Trimester, 3rd Trimester TIs, TIb
EA2-11AVE GYN Uterus, Uterus2, Adnexa, Cervix, Penetration TIs, TIb
Urology Prostate, Bladder TIs, TIb
OB 1st Trimester, 2nd Trimester, 3rd Trimester TIs, TIb
miniER7 GYN Uterus, Adnexa, Cervix, Penetration TIs, TIb
Urology Prostate, Bladder TIs, TIb
Abdomen Abdomen, Renal, Bowel, Aorta, Penetration TIs, TIb
1st Fetal Heart, 1st Trimester, 2nd Trimester, 3rd
OB TIs, TIb
CV1-8A Trimester, Fetal Heart, NT
GYN Uterus, Adnexa, Penetration TIs, TIb
Urology Prostate, Bladder TIs, TIb
Abdomen Abdomen, Renal, Bowel, Aorta, Penetration TIs, TIb
1st Trimester, 2nd Trimester, 3rd Trimester, Fetal Heart,
OB TIs, TIb
CV1-8AE NT
GYN Uterus, Adnexa, Penetration TIs, TIb
Urology Prostate, Bladder TIs, TIb
OB 1st Trimester, 2nd Trimester, 3rd Trimester TIs, TIb
EV2-10A GYN Uterus, Uterus1, Uterus2, Adnexa, Cervix, Penetration TIs, TIb
Urology Prostate, Bladder TIs, TIb
Abdomen Abdomen, Renal, Bowel, Aorta TIs, TIb
Cardiac Adult Echo, Ped Echo, Aortic Arch TIs, TIb
PA1-5A Neo Head TIs, TIc
Pediatric
Ped Abd, Ped Hip TIs, TIb
Thoracic Lung TIs, TIb
Chapter 10 Probes 10 - 5

Thermal
Probes Application Preset
Index

Vascular Carotid, Arterial, Venous TIs, TIb


TCD TCD TIs, TIc
Abdomen Abdomen, Renal, Bowel, Aorta TIs, TIb
Cardiac Adult Echo, Ped Echo, Aortic Arch TIs, TIb
Neo Head TIs, TIc
PA3-8B Pediatric
Ped Abd, Ped Hip TIs, TIb
Vascular Carotid, Arterial, Venous TIs, TIb
TCD TCD TIs, TIc
Abdomen Abdomen, Renal, Bowel, Aorta TIs, TIb
Cardiac Adult Echo, Ped Echo, Aortic Arch TIs, TIb
Neo Head TIs, TIc
PA4-12B Pediatric
Ped Abd, Ped Hip TIs, TIb
Vascular Carotid, Arterial, Venous TIs, TIb
TCD TCD TIs, TIc
MMPT3-7 Cardiac Adult Echo, Ped Echo, Aortic Arch TIs, TIb
Cardiac Adult Echo, Ped Echo TIs, TIb
CW6.0 Vascular Carotid, Arterial, Venous TIs, TIb
TCD TCD TIs, TIc
Cardiac Adult Echo, Ped Echo TIs, TIb
DP2B Vascular Carotid, Arterial, Venous TIs, TIb
TCD TCD TIs, TIc

■ In addition to the system-optimized settings, the user can configure the


desired settings in UserPreset.
■ For more information on probe selection and settings, please refer to User
Manual Chapter 5 ‘Starting Modes of Operation’.
NOTE
■ The default thermal index may vary, corresponding to the presets.
■ You can change the thermal index in Setup > Imaging > Preset.
10 - 6 V6/XV6 Service Manual

■ Function List
The functions available in this product for various probes and applications are as follows:
Multi Panoramic Freehand
Probes Application Har S-Har. CEUS+ ElastoScan+™
Vision + 3D

Abdomen X O O O X X O

MSK X O O O X X O

LA2-9S Small Parts X O O O X O O

Vascular X O O O X X O

Pediatric X O O O X X X

Abdomen X O O O X X O

X O O X O O
MSK O
(Only General)
LA3-14AD Small Parts X O O O X O O

Vascular X O O O X X O

Pediatric X O O O X X X

MSK X O O O X X O
LA3-22AI
Intraoperative X O O O X X O

MSK X O O O X X O

Small Parts X O O O X X O
L3-22
Vascular X O O O X X O

Pediatric X O O O X X X

Abdomen O O O O X X O

MSK O O O O X X X

Vascular O O O O X X X

CA1-7AD OB O O O O X X O

GYN O O O O X X X

Pediatric O O O O X X X

Urology O O O O X X X

O
(Only
Abdomen O O O O
Abdomen,
X O
Penetration)

MSK O O O O X X X

CA1-7S Vascular O O O O X X X

OB O O O O X X O

GYN O O O O X X X

Thoracic O O O O X X X

Pediatric O O O O X X X
Chapter 10 Probes 10 - 7

Multi Panoramic Freehand


Probes Application Har S-Har. CEUS+ ElastoScan+™
Vision + 3D

Urology O O O O X X X

O
(Only
Abdomen O O O O
Abdomen,
X O
Penetration)

MSK O O O O X X X

Vascular O O O O X X X
CA1-7SD
OB O O O O X X O

GYN O O O O X X X

Thoracic O O O O X X X

Pediatric O O O O X X X

Urology O O O O X X X

Abdomen O O O O X X O

MSK O O O O X X X

Vascular O O O O X X X

OB O O O O X X O
CA3-10A
GYN O O O O X X X

Thoracic O O O O X X X

Pediatric O O O O X X X

Urology O O O O X X X

Abdomen X O O O X X X

CA4-10M Vascular X O O O X X X

Pediatric X O O O X X X

OB X O O X X X X

EA2-11AR GYN X O O X X O O

O
Urology X O O X X
(Only Prostate)
O

OB X O O X X X X

EA2-11ARE GYN X O O X X O O

O
Urology X O O X X
(Only Prostate)
O

OB X O O X X X X

EA2-11AV GYN X O O X X O O

O
Urology X O O X X
(Only Prostate)
O

OB X O O X X X X
EA2-11AVE
GYN X O O X X O O
10 - 8 V6/XV6 Service Manual

Multi Panoramic Freehand


Probes Application Har S-Har. CEUS+ ElastoScan+™
Vision + 3D
O
Urology X O O X X
(Only Prostate)
O

OB X O O X X X X

miniER7 GYN X O O X X O O

O
Urology X O O X X
(Only Prostate)
O

Abdomen O O O O X X X

OB O O O O X X X
CV1-8A
GYN O O O O X X X

Urology O O O O X X X

Abdomen O O O O X X X

OB O O O O X X X
CV1-8AE
GYN O O O O X X X

Urology O O O O X X X

OB X O O X X X X

O
O
(Only Uterus,
EV2-10A GYN X O O X (Only
Adnexa, Cervix,
X
Uterus1)
Penetration)
O
Urology X O O X X
(Only Prostate)
X

Abdomen O O X X X X X

Cardiac O O X X X X X

Pediatric O O X X X X X
PA1-5A
Thoracic O O X X X X X

Vascular O O X X X X X

TCD O O X X X X X

Abdomen O O X X X X X

Cardiac O O X X X X X

PA3-8B Pediatric O O X X X X X

Vascular O O X X X X X

TCD O O X X X X X

Abdomen O O X X X X X

Cardiac O O X X X X X

PA4-12B Pediatric O O X X X X X

Vascular O O X X X X X

TCD O O X X X X X
Chapter 10 Probes 10 - 9

Multi Panoramic Freehand


Probes Application Har S-Har. CEUS+ ElastoScan+™
Vision + 3D

MMPT3-7 Cardiac O O X X X X X

Cardiac X X X X X X X

CW6.0 Vascular X X X X X X X

TCD X X X X X X X

Cardiac X X X X X X X

DP2B Vascular X X X X X X X

TCD X X X X X X X

■ The meaning of each symbol in the table is described below:


– Har: Harmonic Imaging
– S-Har™: S-Harmonic™ Imaging
NOTE – CM: Color M
■ MultiVision is only available in 2D, C, M, PD, and S-Flow™ Modes.

S-
HQ- Shadow Nerve
Probes Application CM MV-Flow™ Shearwave
Vision™ HDR™ Track™
Imaging™

Abdomen X X X X X X

O
MSK X X (Only X X O
General)

LA2-9S Small Parts X X X X X X

O
Vascular X (Only Carotid, X X X X
Arterial)

Pediatric X X X X X X

Abdomen X X X X X X

O O
MSK X (Only (Only X O O
General) General)

LA3-14AD O
O
Small Parts X (Only Breast, X X
(Only Breast)
X
Thyroid)

Vascular X X X X X X

Pediatric X X X X X X

O O
(Only (Only
LA3-22AI MSK X General, General, X X O
Superficial, Superficial,
Superficial1) Superficial1)
10 - 10 V6/XV6 Service Manual

S-
HQ- Shadow Nerve
Probes Application CM MV-Flow™ Shearwave
Vision™ HDR™ Track™
Imaging™

Intraoperative X X X X X X

O O
(Only (Only
MSK X General, General, X X O
Superficial, Superficial,
Superficial1) Superficial1)
L3-22
Small Parts X X X X X X

Vascular X X X X X X

Pediatric X X X X X X

O
O O
(Only
Abdomen X
Abdomen,
(Only (Only X X
Abdomen) Abdomen)
Renal)

MSK X X X X X X

Vascular X X X X X X

O
CA1-7AD (Only 1st
O O
Trimester,
OB (Only Fetal
2nd
X (Only 2nd X X
Heart) Trimester)
Trimester, 3rd
Trimester)

GYN X X X X X X

Pediatric X X X X X X

Urology X X X X X X

O
O
(Only O O
(Only
Abdomen X Abdomen, (Only (Only
Abdomen,
X
Penetration Abdomen) Abdomen)
Penetration)
Renal)

MSK X X X X X X

Vascular X X X X X X

O
O (Only 1st O
CA1-7S (Only Fetal Trimester, (Only 2nd
OB Heart, 1st 2nd
X
Trimester, 3rd
X X
Fetal Heart) Trimester, 3rd Trimester)
Trimester)

GYN X X X X X X

Thoracic X X X X X X

Pediatric X X X X X X

Urology X X X X X X

O O
O O
(Only (Only
CA1-7SD Abdomen X (Only (Only
Abdomen,
X
Abdomen, Abdomen) Abdomen)
Penetration, Penetration)
Chapter 10 Probes 10 - 11

S-
HQ- Shadow Nerve
Probes Application CM MV-Flow™ Shearwave
Vision™ HDR™ Track™
Imaging™
Renal)

MSK X X X X X X

Vascular X X X X X X

O
O (Only 1st O
(Only Fetal Trimester, (Only 2nd
OB Heart, 1st 2nd
X
Trimester, 3rd
X X
Fetal Heart) Trimester, 3rd Trimester)
Trimester)

GYN X X X X X X

Thoracic X X X X X X

Pediatric X X X X X X

Urology X X X X X X

Abdomen X X X X X X

MSK X X X X X X

Vascular X X X X X X

O
(Only 1st
O O
Trimester,
OB (Only Fetal
2nd
X (Only 2nd X X
CA3-10A Heart) Trimester)
Trimester, 3rd
Trimester)

GYN X X X X X X

Thoracic X X X X X X

Pediatric X X X X X X

Urology X X X X X X

Abdomen X X X X X X

Vascular X X X X X X
CA4-10M O
(Only Neo
Pediatric X
Head, Ped
X X X X
Abd)

OB X O X X X X

O O
EA2-11AR GYN X (Only Uterus, X (Only Uterus, X X
Uterus2) Uterus2)

Urology X X X X X X

OB X X X X X X

O
EA2-11ARE GYN X X X (Only Uterus, X X
Uterus2)

Urology X X X X X X
10 - 12 V6/XV6 Service Manual

S-
HQ- Shadow Nerve
Probes Application CM MV-Flow™ Shearwave
Vision™ HDR™ Track™
Imaging™

OB X O X X X X

O O
EA2-11AV GYN X (Only Uterus, X (Only Uterus, X X
Uterus2) Uterus2)

Urology X X X X X X

OB X X X X X X

O
EA2-11AVE GYN X X X (Only Uterus, X X
Uterus2)

Urology X X X X X X

OB X X X X X X

O O
GYN X
(Only Uterus)
X
(Only Uterus)
X X
miniER7
O
Urology X O X (Only X X
Prostate)

Abdomen X X X X X X

O
O (Only 1st O
O
(Only Fetal Trimester, (Only 2nd
OB Heart, 1st 2nd
(Only 1st
Trimester, 3rd
X X
CV1-8A Trimester)
Fetal Heart) Trimester, 3rd Trimester)
Trimester)

GYN X X X X X X

Urology X X X X X X

Abdomen X X X X X X

O O
OB X X (Only 1st (Only 2nd X X
CV1-8AE Trimester) Trimester)

GYN X X X X X X

Urology X X X X X X

OB X O X X X X

O O
EV2-10A GYN X
(Only Uterus)
X
(Only Uterus)
X X

Urology X X X X X X

Abdomen X X X X X X

Cardiac O X X X X X

Pediatric X X X X X X
PA1-5A
Thoracic X X X X X X

Vascular X X X X X X

TCD X X X X X X
Chapter 10 Probes 10 - 13

S-
HQ- Shadow Nerve
Probes Application CM MV-Flow™ Shearwave
Vision™ HDR™ Track™
Imaging™

Abdomen X X X X X X

Cardiac O X X X X X

PA3-8B Pediatric X X X X X X

Vascular X X X X X X

TCD X X X X X X

Abdomen X X X X X X

Cardiac O X X X X X

PA4-12B Pediatric X X X X X X

Vascular X X X X X X

TCD X X X X X X

MMPT3-7 Cardiac O X X X X X

Cardiac X X X X X X

CW6.0 Vascular X X X X X X

TCD X X X X X X

Cardiac X X X X X X

DP2B Vascular X X X X X X

TCD X X X X X X
10 - 14 V6/XV6 Service Manual

10.1.2. Ultrasound Transmission Gel


For successful acoustic signal transmission, ultrasound gels or couplant must be applied.

■ The use of inappropriate ultrasound gels could result in damages in the


probe. Using damaged probe could cause injuries such as electric shock in
users or patients.
■ Only use the ultrasound gels approved by Samsung Medison. Using
WARNING unapproved gels may result in damage to the probe and void the warranty.
■ Completely remove the remaining ultrasound gel on the probe surface after
using the probe. Residues from the ultrasound gel may result in
contamination and discoloration.
■ Do not use ultrasound gels or couplant that contain the following ingredients:
– Oil based lotion type couplant
– Mineral oil
– Olive oil
– lubricant oil
– Paraffin based grease
– Dimethylsilicone
– Alcohol such as methanol, ethanol, isopropanol
– Acetone
– Dioctylphtalate
– Aromatic substances
– Iodine compounds
– Aloe Vera.

■ The tip of injection nozzle may cause damage to the probe lens.
■ When applying the ultrasound gel to the probe, make sure that the tip of the
ultrasound gel injection nozzle does not touch the surface of the probe lens.
CAUTION
Chapter 10 Probes 10 - 15

10.1.2.1. Gel Warmer (Option)


The Gel Warmer keeps the ultrasound gel warm. Warming up ultrasound gel takes
approximately 5 minutes.

■ Always turn the Gel Warmer off when it is not in use.


■ Do not put your hands inside the Gel Warmer, as it may burn your skin.
WARNING

■ Do not use the Gel Warmer for purposes other than to control the
temperature of ultrasound gel.
■ Do not place the probe or any other equipment inside the Gel Warmer.
CAUTION ■ Do not touch the power terminal of the Gel Warmer while you are examining
a patient. There is a risk of electric shock from leakage current.
■ Do not apply excessive force to the Gel Warmer, or you may damage its
support parts.
10 - 16 V6/XV6 Service Manual

10.1.3. Using Sheaths


Sheaths are recommended for clinical applications of an invasive nature, including
intraoperative, transrectal, transvaginal, and biopsy procedures. Using a sheath also prevents
contamination from blood or other bodily fluids during operations or biopsy.
Samsung Medison does not supply sheaths, so appropriate sheaths should be purchased
independently.

■ For more information about sheaths, please contact the manufacturer.


■ Always use sheaths certified and approved by relevant national/local
authorities.
■ Use a sterile sheath for intraoperative or biopsy procedures.
WARNING
■ Use a sterile, single-use, latex sheath for transesophageal procedures.
■ Use a sterile sheath (preferred) or single-use sheath for transrectal or
transvaginal procedures.
■ Sheaths are disposable. Do not reuse them.
■ Make sure that you clean and disinfect the probes after every use.
■ Check the exterior of the Sheath before use. If a sheath is split or
contaminated, do not use it.
■ In neurosurgical applications, a disinfected probe must be used with sterile
gel and a sterile pyrogen-free sheath.
■ If the sterile sheath becomes compromised during neurosurgical applications
involving a patient with Creutzfeldt-Jakob disease, the probe cannot be
successfully sterilized by any disinfection method.
■ Some sheaths contain natural rubber latex and talc, which can cause allergic
reactions in some individuals. Please confirm whether the specific patient is
allergic to those materials before using such sheaths. Please refer to the FDA
Medical Alert released on March 29, 1991.
Chapter 10 Probes 10 - 17

10.1.3.1. Installing the Sheath

1. Remove the sheath from the packaging, and fill it with ultrasound gel. Be sure to put on
sterilized surgical gloves.
2. Insert the probe into the sheath and pull the latex tip to cover the probe completely. If
possible, cover the probe cable as well.
3. Ensure that there are no air bubbles trapped within the ultrasound gel. If necessary,
secure the sheath to the probe and the probe cable.
4. Dispose of the sheath after use.
10 - 18 V6/XV6 Service Manual

10.1.4. Probe Safety Precautions


The probe can be damaged by improper use or by coming into contact with certain chemical
substances. Always follow the instructions in the user manual to inspect the probe cable, case
and lens before and after each use.
Check the probe for cracks, broken parts, leaks and sharp edges. If there is any damage, stop
using the probe immediately and contact the Samsung Medison Customer Support Department.
Using damaged probes may result in electric shocks and other hazards to the patients and/or
users.

■ The user is responsible for cleaning, disinfecting, and sterilizing probes.


■ Probes are not sterilized at the time of shipment.
■ Do not apply mechanical shock to the probe.
CAUTION ■ If a probe hits the ground or a hard surface, disconnect the probe
immediately from the console of the ultrasound system. Damaged electrical
insulation may cause an electrical shock.
■ Do not place the probe cable on the floor where the cable can be run over by
equipment wheels, etc. Do not apply excessive force to bend or pull the
cable.
■ Do not immerse the probe into any inappropriate chemical substances that
are not compatible.
■ Do not expose it to temperatures of 50°C or higher.
■ 3D and 4D probes may make a slight noise if they are used for a long time or
if a certain mode is in use.
■ Probes should be cleaned and disinfected prior to replacement or disposal.
■ Always inspect the lenses, cables, and connectors after using probes. If
damaged, stop using them immediately.
Chapter 10 Probes 10 - 19

10.1.4.1. Using and Infection Control of the Probe


The ultrasonographic image scanner uses ultrasound, and makes direct contact with the
patient when in use. Depending on the types of examinations, such contact can be made to
a wide variety of locations, including the ordinary skin or the location of blood transfusion
during a surgical procedure.
The most effective method of preventing infection among patients is to use each probe only
once. However, probes may need to be reused, as they are complex in design and
expensive. Consequently, the safety instructions must be followed carefully in order to
minimize the risk of infection among patients.

No neurosurgical treatments or examinations should be carried out on a patient


with Creutzfeldt-Jakob disease. If the probe has been used on such a patient, it
cannot be sterilized by any method whatsoever.
If the probe or product is contaminated, dispose of it according to the local
WARNING biohazardous waste disposal procedures.

Sufficient washing and disinfecting must be carried out to prevent infection. This
is the responsibility of the user who manages and maintains the disinfection
procedures for the equipment. Always use legally approved cleansing solutions
and sheaths.
CAUTION

10.1.4.2. Electric Shocks


The probe uses electrical energy. If it touches conductive materials, there are risks of electric
shocks to the patient or the user.

■ If damage to the product (lens, cable) that can cause electric shock is found,
discontinue use of the device immediately and contact your Samsung
Medison representative.
■ Do not immerse the probe into liquid.
WARNING
■ Do not touch the probe connectors or the connection pins on the product.
■ Do not drop the probe or apply mechanical shocks.
■ Inspect the housing, strain relief, lens, and seal for damage, and check for
any functional problem before and after each use.
■ Do not apply excessive force to twist, pull or bend the probe cable.
■ The power protection fuse protects the probe and the product from excess
current. If the power monitoring protection circuit detects excess current, it
immediately shuts off the current to the probe in order to prevent the probe
surface from overheating and to restrict the ultrasound power output.
■ The temperature of the product for making contact with patients is limited
under 43°C. The ultrasound power output (AP&I) is in compliance with US
FDA standards.
10 - 20 V6/XV6 Service Manual

10.1.5. Cleaning, Disinfecting, and Sterilizing Probes


All probes must be cleaned and disinfected after each use. Cleaning is an important procedure
that must be carried out before disinfecting the probe. For information on cleaning and
disinfecting the probe, please refer to the appropriate chapter. Using an inappropriate
disinfectant may damage the probe.

■ Always use protective equipment such as face mask, eyewear, and gloves
when cleaning, disinfecting, and sterilizing probes.
■ Inspect the housing, strain relief, lens and seal for damage, and check for
WARNING any functional degradation before and after cleaning and disinfecting the
probe.
■ Using an inappropriate cleaning or disinfecting agent may damage the
probe.

10.1.5.1. Information on Cleaner, Disinfectant, and Ultrasound Gel

■ Reprocessing Method by Probe Type


To maintain the performance of ultrasound probes, proper maintenance is required.
As ultrasound probes are classified into critical, semi-critical or non-critical devices based
on the standards of FDA guidance* and the Hygiene Requirements for the Reprocessing
of Medical Devices from Germany guideline of Robert Koch Institute (RKI), proper
cleaning and disinfection or sterilization methods for that classification should be used.
Table for Choosing the Correct Probe-Care Method

Classification
Contact Area Application Probe Level Selection
Criteria

Non-critical Curved, Linear, and


Intact skin Low level disinfection
device Phased array probes

Semi-critical Mucous membrane and High level disinfection or


Endocavity, MPTEE
device damaged skin sterilization

Critical device Blood, sterile tissue, etc. Intraoperative Sterilization

* Guidance for Industry and FDA Staff – Marketing Clearance of Diagnostic Ultrasound
Systems and Transducers - Appendix E
* The FDA reprocessing guidance “Reprocessing Medical Devices in Health Care
Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug
Administration Staff” March 17, 2015, (https://www.fda.gov/media/80265/download)
The care method for your probes determines the appropriate disinfectant for your probe.
An appropriate cleaner, disinfectant or ultrasound gel should be used for all probes. For
the latest details about compatible cleaner, disinfectants, and ultrasound gel, please see
Samsung Medison website and the User Guide.
– User Guides: This is provided as a booklet upon purchase of the product.
– Website: http://www.samsunghealthcare.com
Chapter 10 Probes 10 - 21

10.1.5.2. Warnings and Cautions During Cleaning, Disinfecting, Sterilizing Probes


These general cleaning instructions must be followed for all probes.
All probes must be cleaned after each use. Cleaning the probe is an essential step before
effective disinfection or sterilization. Before cleaning a probe, read “Warnings and Cautions”
below.
After cleaning, you must disinfect or sterilize probes by following the appropriate procedures.

■ Always use protective equipment such as face mask, eyewear, and gloves
when cleaning, disinfecting, and sterilizing probes.
■ If a premixed solution is used, be sure to observe the solution expiration
date.
WARNING
■ Probes must be cleaned after each use. Cleaning the probe is an essential
step before effective disinfection or sterilization. Be sure to follow the
manufacturer’s instructions when using cleaner, disinfectant, disinfection
system and sterilization system.

■ Attempting to disinfect a cable or connector by using a method other than the


one included here can damage the device and voids the warranty.
■ Do not allow sharp objects, such as scalpels or cauterizing knives, to touch
probes or cables.
CAUTION
■ When handling a probe, do not dump the probe on hard surfaces.
■ Do not use a brush on the connector label.
■ During cleaning, disinfection, and sterilization, orient the parts of the probe
that must remain dry higher than the wet parts until all parts are dry. This will
help keep liquid from entering unsealed areas of the probe.
■ Do not allow any type of fluid to enter the connector. Ensure that fluid does
not enter through the strain relief, connector, electrical contacts, or areas
surrounding the locking-lever shaft and the strain relief. Fluid in the connector
may void the device warranty.
■ Ensure the solution contains only 70% or less alcohol when using an alcohol
for disinfection. Solutions of more than 70% alcohol can cause product
damage.
■ Do not use any alcohol or alcohol-based products on the cable or strain relief.
■ Avoid disinfectant contact with the connector label.
■ Be sure to use the proper concentration of enzymatic cleaner and rinse
thoroughly.
■ Do not use paper products or products that are abrasive when cleaning the
probe. They damage the lens of the probe.
■ Aggressive wiping or scrubbing can damage the probe. Use a gentle wiping
motion for the probe lens, the strain relief, the areas surrounding the strain
relief, and the cables.
■ When wiping the probe, hold the housing. Do not suspend the probe by the
cable. This can damage probes.
■ Do not apply excessive force to twist, pull or bend the probe cable. This can
damage probes.
10 - 22 V6/XV6 Service Manual

■ Before storing probes, ensure that they are thoroughly dry. If it is necessary to
dry the probe lens after cleaning, use a soft cloth and a blotting motion,
instead of a wiping motion.
Chapter 10 Probes 10 - 23

10.1.5.3. Precautions for Cleaning and Disinfecting MPTEE Probes


Proper cleaning and disinfection prevent infectious diseases. The user must clean and
disinfect the probe by using an effective method, such as the one described in “Cleaning,
Disinfecting, and Sterilizing Probes” in this user manual.
Before cleaning a MPTEE probe, read “Warnings and Cautions During Cleaning, Disinfecting,
Sterilizing Probes” and “Precautions for Cleaning and Disinfecting MPTEE Probes” below.
After cleaning, the user must disinfect probes by following the appropriate procedures.

■ Do not allow the probe connector to come in contact with cleaner or


disinfectant.
■ Do not allow the control handle to be immersed into cleaner or disinfectant.
Clean and disinfect the control handle and the cable with a moist cloth
CAUTION dampened with compatible cleaner or disinfectant; clean and disinfect only
the area between, and including, the scan head at the end of the probe and
the area marked as 100 cm.

■ Cleaning

Never clean the probe with methanol or ethanol. Doing so may cause serious
problems to the product.
CAUTION

■ Disinfection

Do not use iodine, steam, heat, or ethylene oxide to disinfect the product.

WARNING

■ Do not leave the probe in disinfectant solution for more than one hour.
■ The probe must be rinsed immediately after disinfection.
■ For MPTEE Probes, only use the Immersion Method for disinfection.
CAUTION
10 - 24 V6/XV6 Service Manual

10.1.5.4. Structure and Name of Probes


 Probe (Non-MPTEE)

1 Lens

2 Housing

3 Transducer

4 Strain Relief

5 Cable

6 Connector

[Figure 10.1 Structure and Name of a Probe]


Chapter 10 Probes 10 - 25

■ MPTEE Probe

1 Lens

2 Tip

3 Transducer

4 Tube

5 Shaft

6 Control handle

7 Dial

8 Strain Relief

9 Cable

10 Tube

[Figure 10.2 Structure and Name of a MPTEE Probe]


10 - 26 V6/XV6 Service Manual

10.1.5.5. Cleaning Probes


All probes must be cleaned according to this instruction after each use. Cleaning is an
important procedure that must be carried out before disinfecting and sterilizing the probe.
The cleaning method is applicable for all probes provided by Samsung Medison.

■ Only use a surgeon’s brushes on the housing. Do not use brushes on the
other parts, including the probe lens, the strain relief, the areas surrounding
the strain relief, and the cables, than the housing when cleaning probes.
Using any types of brushes can damage the probe lens, the strain relief, the
CAUTION areas surrounding the strain relief, and the cables. For MPTEE probes, use
surgeon’s brushes only on the area between the scan head at the end of the
probe and the area marked as 100 cm but excluding the lens.
■ During cleaning and disinfection, keep the parts of the probe that must
remain dry higher than the other parts during wetting until all parts are dry.
■ Do not bend or crimp the gastroscope or cable.
■ Do not use isopropyl alcohol on the strain relief/housing joint, the strain relief,
or the cable.

■ Wiping Method
1. Disconnect the probe from the system.
2. Remove the biopsy bracket or needle guide from the probe. Collect the reusable parts,
reusable biopsy bracket, or needle guide that need to be disinfected or sterilized, and
discard the disposable parts in the clinical waste bin (For more information on cleaning
and disinfection/sterilization of the reusable parts, please refer to the ‘Cleaning and
Disinfecting Biopsy Kits’ in this chapter.).
3. For a list of compatible cleaners, see the User Guide or Samsung Medison website:
http://www.samsunghealthcare.com.
Follow the cleaner manufacturer's instructions for preparation, temperature, solution
strength, and duration of contact. Ensure that the solution strength and duration of
contact are appropriate for the intended clinical use of the device. If a premixed solution is
used, be sure to observe the solution expiration date.
4. Remove any visible gel or biological material on the surface with wipes or soft cloth
soaked with the cleaner (made using manufacturer's instructions). For difficult to reach
areas such as seams, cracks and crevices of the device, use a surgeon’s brush wrapped
with a wipe or cloth soaked with the cleaner.
Use a gentle wiping motion for the probe lens, the strain relief, the areas surrounding the
strain relief, and the cables. When wiping the probe, hold the housing. Do not suspend
the probe by the cable. Do not apply excessive force to twist, pull or bend the probe cable.
These can damage probes.
Dispose of the used wipes or cloth appropriately according to the local medical waste
regulations.
5. Wipe off the probe surface with fresh wipes or cloth soaked with the cleaner, following the
cleaner manufacturer’s instruction on the temperature, wiping time, and cleaner contact
time.
6. Visually inspect for any remaining soil and if necessary repeat the steps, starting at step 4.
Chapter 10 Probes 10 - 27

7. Make sure to follow the rinse (and neutralization) instructions from the cleaner
manufacturer for rinsing the probe. If insufficient instructions are supplied, then, as a
minimum, thoroughly rinse the probe with a clean lint-free cloth dampened with pure
water until all signs of residual debris and cleaning solution are removed. (at minimum of
two minutes) For difficult to reach areas such as seams, cracks and crevices of the
device, use a surgeon’s brush wrapped with a clean lint-free cloth dampened with pure
water.
8. Repeat the step 7 (rinsing) at least three times with a fresh clean lint-free cloth dampened
with pure water each time.
9. Following the dry instructions from the cleaner manufacturer, air-dry the probe or remove
water from all surfaces of the probe with a clean, dry, disposable soft, lint-free cloth or
wipe. If it is necessary to dry the probe lens after cleaning, use a soft cloth and a blotting
motion, instead of a wiping motion.
10. Examine the probe and cable for damage such as cracks, splitting, sharp edges, or
projections. If such damage is found, discontinue use of the device and contact your
Samsung Medison representative.
10 - 28 V6/XV6 Service Manual

■ Immersion Method
1. Disconnect the probe from the system.
2. Remove the biopsy bracket or needle guide from the probe. Collect the reusable parts,
reusable biopsy bracket, or needle guide that need to be disinfected or sterilized, and
discard the disposable parts in the clinical waste bin (For more information on cleaning
and disinfection/sterilization of the reusable parts, please refer to the ‘Cleaning and
Disinfecting Biopsy Kits’ in this chapter.).
Rinse off any visible gel or biological material with running tap water. If it is hard to
remove materials with running tap water, wipe off any gel or biological material with a
moist cloth or sponge moistened with cleaner (made using manufacturer's instructions) or
tap water, taking care to remove all visible contamination.
Dispose of the used cloth or sponge appropriately according to the local medical waste
regulations.
3. For a list of compatible cleaner s, see the User Guide or Samsung Medison website:
http://www.samsunghealthcare.com.
Follow the cleaner manufacturer’s instructions for preparation, temperature, solution
strength, and duration of contact. Ensure that the solution strength and duration of
contact are appropriate for the intended clinical use of the device. If a premixed solution is
used, be sure to observe the solution expiration date.

4. Fill a sink or bowl with freshly-made cleaner solution.


5. Immerse the probe up to the immersion point shown in the figure below.
– Some probes can be cleaned up to the cable. Do not clean the cable that is within 50
cm of the end of the cable from the connector. For a list of probes that can be cleaned
up to the cable, see the User Guide or Samsung Medison website:
http://www.samsunghealthcare.com.

[Figure 10.3 Immersion Point]


Chapter 10 Probes 10 - 29

– For MPTEE probes, immerse only the area between the scan head at the end of
the probe and the area marked as 100 cm.
The area between the Scan Head and the area marked 100 cm

[Figure 10.4 Immersion Point for MPTEE Probe]

6. Continue soaking the probe until the total cleaner contact time specified by the cleaner
manufacturer is reached.
7. After the cleaner contact time, brush the probe housing with a surgeon’s brush at
minimum of one minute to remove any visible soil. Pay close attention to seams, cracks
and crevices during brushing time. Do not use brushes on the other parts than the
housing; Using any types of brushes can damage the other parts, including the probe
lens, the strain relief, the areas surrounding the strain relief, and the cables, than the
housing.
– For MPTEE probes, use surgeon’s brushes only on the area between the scan head
at the end of the probe and the area marked as 100 cm but excluding the lens.
8. Visually inspect for any remaining soil and if necessary repeat the steps, starting at step 4.
9. Make sure to follow the rinse (and neutralization) instructions from the cleaner
manufacturer for rinsing the probe. If insufficient instructions are supplied, then, as a
minimum, thoroughly rinse the probe with tap water at 22ºC to 43ºC (72ºF to 110ºF) to
prevent the coagulation of blood and proteins until all signs of residual debris and
cleaning solution are removed (at minimum of one minute)
10. Repeat the step 9 (rinsing) at least three times.
11. Following the dry instructions from the cleaner manufacturer, air-dry the probe or remove
water from all surfaces of the probe with a clean, dry, disposable soft, lint-free cloth. If it
is necessary to dry the probe lens after cleaning, use a soft cloth and a blotting motion,
instead of a wiping motion.
12. Examine the probe and cable for damage such as cracks, splitting, sharp edges, or
projections. If such damage is found, discontinue use of the device and contact your
Samsung Medison representative.
10 - 30 V6/XV6 Service Manual

10.1.5.6. Disinfecting Probes


The probe must be cleaned and disinfected after each use. Disinfection of the probe is an
important step that must be carried out. User must follow the reprocessing method
appropriate for probe type as described in “Reprocessing Method by Probe Type” in this
chapter.

■ Always use protective equipment such as face mask, eyewear, and gloves
when cleaning, disinfecting, and sterilizing probes.
■ Inspect the housing, strain relief, lens and seal for damage, and check for
any functional problem before and after disinfecting the probe.
WARNING
■ Never use disinfectants after the expiration date.
■ The area the probe will contact during use dictates the level of disinfection
required for a probe. Ensure that the solution strength and duration of contact
are appropriate for disinfection.

■ Low-Level Disinfection
Low-level disinfection of probes uses the wipe method with a low-level disinfectant.
Before performing this procedure, read the warnings and cautions in the “Warnings and
Cautions During Cleaning, Disinfecting, Sterilizing Probes” section and “Table for
Choosing the Correct Probe-Care Method” in this chapter.
1. Clean the probe according to the procedures in “Cleaning Probes” in this chapter.
2. After cleaning, choose low-level disinfectants that are compatible with your probe. For
a list of compatible disinfectants, see the User Guide or Samsung Medison website:
http://www.samsunghealthcare.com.
Follow disinfectant manufacturer’s instructions for preparation, temperature, solution
strength, and duration of contact. Ensure that the solution strength and duration of
contact are appropriate for the intended clinical use of the device. If a premixed
solution is used, be sure to observe the solution expiration date.
3. Wipe the probe surfaces, following disinfectant manufacturer’s instructions for
temperature, wipe durations, and duration of disinfectant contact. Pay close attention
to seams, cracks and crevices during wiping time. Ensure that the disinfectant solution
does not enter the device or the connector.
– Some probes can be disinfected up to the cable. Do not disinfect the cable that is
within 50 cm of the end of the cable from the connector. For a list of probes that
can be disinfected up to the cable, see the User Guide or Samsung Medison
website: http://www.samsunghealthcare.com.
– When disinfecting the connector, wipe only the outer surfaces of the connector; do
not allow any type of fluid to enter through other parts than the connector surfaces,
including the strain relief, electrical contacts, or areas surrounding the locking-lever
shaft.
4. Make sure to follow the rinse (and neutralization) instructions from the disinfectant
manufacturer for rinsing the probe. If insufficient instructions are supplied, then, fill a
sink or a bowl with pure water and rinse the probe. Immerse the probe up to the
immersion point shown in the figure below. (at minimum of one minute)
Chapter 10 Probes 10 - 31

– Some probes can be immersed up to the cable. Do not immerse the cable that is
within 50 cm of the end of the cable from the connector. For a list of probes that
can be immersed up to the cable, see the User Guide or Samsung Medison
website: http://www.samsunghealthcare.com.

[Figure 10.5 Immersion Point]

5. Repeat the step 4 (rinsing) at least three times with a fresh batch of pure water each
time.
6. Following the dry instructions from the disinfectant manufacturer, air-dry the probe or
remove water from all surfaces of the probe with a sterile, lint-free cloth and store it
properly (e.g. in a sterile bag) to prevent any further contamination. If it is necessary to
dry the probe lens after cleaning, use a soft cloth and a blotting motion, instead of a
wiping motion.
7. Examine the probe and cable for damage such as cracks, splitting, sharp edges, or
projections. If such damage is found, discontinue use of the device and contact your
Samsung Medison representative.

■ High-Level Disinfection
High-level disinfection of probes typically uses the immersion method with a high level
disinfectant or the disinfection system method. Before performing this procedure, read the
warnings and cautions in the “Warnings and Cautions During Cleaning, Disinfecting,
Sterilizing Probes” section and “Table for Choosing the Correct Probe-Care Method” in
this chapter.

■ Inspect the housing, strain relief, lens and seal for damage, and check for
any functional problem before and after disinfecting the probe.
■ Ensure that the cycle is appropriate for the desired disinfection level.
WARNING

■ Using a probe that is not compatible with a disinfection system or not


following the instructions of the disinfection system manufacturer can
damage the probe. This could also void the probe warranty.
CAUTION ■ Disinfection wipes are not FDA-cleared for use on semi-critical devices,
10 - 32 V6/XV6 Service Manual

which require high-level disinfection.


■ For MPTEE Probes, only the Immersion Method is recommended for
disinfection.

<Immersion Method>
1. Clean the probe according to the procedures in “Cleaning Probes” in this chapter.
2. After cleaning, choose high-level disinfectants that are compatible with your probe. For
a list of compatible disinfectants, see the User Guide or Samsung Medison website:
http://www.samsunghealthcare.com.
Follow disinfectant manufacturer’s instructions for preparation, temperature, solution
strength, and duration of contact. Ensure that the solution strength and duration of
contact are appropriate for the intended clinical use of the device. If a premixed
solution is used, be sure to observe the solution expiration date.
3. Immerse the probe into the appropriate disinfectant solution for your probe as shown
in the figure below. Make sure that there are no air bubbles on the probe surfaces.
Follow the instructions on the disinfectant label for the duration of probe immersion.
Do not immerse probes longer than the minimum time needed for your level of
disinfection.
– Some probes can be disinfected up to the cable. Do not disinfect the cable that is
within 50 cm of the end of the cable from the connector. For a list of probes that
can be cleaned up to the cable, see the User Guide or Samsung Medison website:
http://www.samsunghealthcare.com.

[Figure 10.6 Immersion Point]

– For MPTEE probes, immerse only the area between the scan head at the end of
the probe and the area marked as 100 cm.
The area between the Scan Head and the area marked 100 cm

[Figure 10.7 Immersion Point for MPTEE Probe]


Chapter 10 Probes 10 - 33

4. Make sure to follow the rinse (and neutralization) instructions from the disinfectant
manufacturer for rinsing the probe. If insufficient instructions are supplied, then, fill a
disinfected sink or a disinfected bowl with pure water and rinse the probe at minimum
of one minute.
– For MPTEE probes, use sterile pure water.
5. Repeat the step 4 (rinsing) at least three times with a fresh batch of pure water each
time.
– For MPTEE probes, use sterile pure water.
6. Following the dry instructions from the disinfectant manufacturer, air-dry the probe or
remove water from all surfaces of the probe with a sterile, lint-free cloth and store it
properly (e.g. in a sterile bag) to prevent any further contamination. If it is necessary to
dry the probe lens after cleaning, use a soft cloth and a blotting motion, instead of a
wiping motion.
7. Register the disinfection data for traceability according to local regulations.
8. Examine the probe and cable for damage such as cracks, splitting, sharp edges, or
projections. If such damage is found, discontinue use of the device and contact your
Samsung Medison representative.
<Disinfection System Method>
1. Clean the probe and cable according to the procedures in “Cleaning Probes” in this
chapter. Observe all warnings and cautions.
2. Refer to the user instructions of the disinfection system for details of proper use,
storage and disposal of waste.
3. Place the transducer of the probe in the disinfection system, following the user
instructions of the disinfection system.
4. Start the desired disinfection process and wait until the end of the cycle.
5. Remove the probe at the end of the cycle from the disinfection system and store it
properly (e.g. in a sterile bag) to prevent any further contamination.
6. Register the disinfection data for traceability according to local regulations.
7. Examine the probe and cable for damage such as cracks, splitting, sharp edges, or
projections. If such damage is found, discontinue use of the device and contact your
Samsung Medison representative.

10.1.5.7. Sterilizing Probes


Sterilization is required for probes when contact bloodstream, sterile tissue or body-space
regardless the use of a sterile cover.

■ Always use protective equipment such as face mask, eyewear, and gloves
when cleaning, disinfecting, and sterilizing probes.
■ In intraoperative applications, sterilized probes should be used with sterile gel
and a sterile probe cover certified and approved by relevant national/local
WARNING authorities.
■ Sterile probe covers are disposable and must not be reused.
10 - 34 V6/XV6 Service Manual

■ Probes must be cleaned after each use. Cleaning the probe is an essential
step before effective disinfection or sterilization. Be sure to follow the
manufacturer’s instructions when using disinfectant, disinfection system, and
sterilization system.
CAUTION ■ Use low temperature plasma sterilization system to sterilize probes. Using
sterilization solution, autoclave, gas (EtO), or other methods not approved by
Samsung Medison will damage your probe and void your warranty.
■ Do not allow sharp objects, such as scalpels and cauterizing knives, to touch
probes or cables.
■ For MPTEE probes, do not use any method of sterilization, including
methods using materials such as iodine, steam, heat, or ethylene oxide.

■ Sterilization System Method


Using a probe that is not compatible with a sterilization system may damage the probe.

■ Inspect the housing, strain relief, lens and seal for damage, and check for
any functional problem before and after sterilizing the probe.
■ Ensure that the cycle is appropriate for the desired disinfection or sterilization
WARNING level.

Using a probe that is not compatible with a sterilization system or not following
the instructions of the sterilization system manufacturer can damage the probe.
This could also void the probe warranty.
CAUTION

1. Clean the probe and cable according to the procedures in “Cleaning Probes” in this
chapter. Observe all warnings and cautions.
2. Refer to the user instructions of the sterilization system for details of proper use,
storage and disposal of waste.
3. Place the probe in the sterilization system, following the user instructions of the
sterilization system.
4. Start the desired sterilization process and wait until the end of the cycle.
5. Remove the probe at the end of the cycle from the sterilization system and store it
properly (e.g. in a sterile bag) to prevent any further contamination.
6. Register the sterilization data for traceability according to local regulations.
7. Examine the probe and cable for damage such as cracks, splitting, sharp edges, or
projections. If such damage is found, discontinue use of the device and contact your
Samsung Medison representative.
Chapter 10 Probes 10 - 35

10.1.6. MPTEE Probe (Option)


The Multi-Planar Transesophageal Echocardiography (MPTEE) probe is used by inserting it
inside the body.

■ The MPTEE probe should be used only by a qualified physician who has
received appropriate training in proper operation of the probe and in
endoscopic techniques as dictated by current relevant medical practices.
■ Familiarize yourself with the use and safety instructions for the MPTEE probe
WARNING completely before using it.
■ Inspect the MPTEE probe for any sharp areas or rough surfaces both with
the tilting section straight and tilted. If damage is found, discontinue use of
the product and contact Samsung Medison’s Customer Service Department.
Using damaged probes may result in electric shocks and other hazards to the
patients and/or users.
■ The use of a bite-guard is mandatory. Failure to use the bite-guard may result
in damage to the probe, which could result in electric shocks and other
hazards to the patients and/or users. This could also void the probe warranty.
■ In order to minimize the risk of "U-turning" of the tilting section in the
esophagus, inspect if the tilting of the tip is properly manipulated before use.
(The up/down and/or left/right tilting may after prolonged use develop an
unwanted amount of free play.) If any abnormal operation is found including
an unwanted amount of free play of the tilting section or excess of the
maximal tilting angles, discontinue use of the product and contact Samsung
Medison’s Customer Service Department.
■ Only use the probe as an internal insertion probe.
■ Do not apply excessive force when inserting or operating the
transesophageal probe.
■ Do not transport the product while the MPTEE probe is placed in the probe
holder. The probe is not secured, and it may fall.
■ The image obtained with the MPTEE probe may differ from the image
displayed on screen. It is necessary to perform an inspection in advance to
reduce the risk of difference in images.
10 - 36 V6/XV6 Service Manual

■ The user is responsible for the correct configuration and use. Please make
sure to read the User Manual thoroughly.
■ The ability of a patient to swallow or accommodate the probe should be
considered. Any history of gastro-esophageal diseases must be determined
CAUTION and considered as well as the possible effects of other therapies the patient
is undergoing. All gastro-esophageal abnormalities must be considered as
well.
■ Only use water-soluble coupling gel for the MPTEE probe.
■ Do not use ultrasound gels that contain any of the following agents for the
MPTEE probe:
– Oils such as mineral oil and lanolin
– Alcohols such as ethanol and denatured alcohol
– Iodine
– All lotions or gels that contain aromatic substances
– Gels that contain aloe vera, methyl or ethyl para benzoic acid
■ Never rub or spray the MPTEE probe with an anesthetic agent.
■ The MPTEE probe must be handled with care. It may be damaged when
dropped or abused. In particular keep sharp objects away from the lens. Do
not touch the lens unnecessarily. Never exert force onto the lens.
■ Avoid forceful manipulations and excessive force in using the probe, which
could result in patient injury.
■ To protect the patient and the probe, keep the probe in brake inactive state
on both knobs when introducing or withdrawing the probe and make sure the
tilting section is straight.
■ Never manually tilt the distal tip of the probe directly; always use the dials for
controlling tilting of the tip.
■ To avoid the risk of electric shock, the MPTEE probe must only be connected
to a supply with protective earth.
■ The connector is not watertight, and should always be kept dry.
■ The control handle should not be immersed into liquid.
■ All internal component repairs and part replacements must be performed by
qualified Samsung Medison’s Customer Service personnel. To prevent
electric shock, never open the cover or panels of the product.
■ Do not twist, tightly coil, or apply excessive force on the probe cable or shaft.
Insulation failure may result.
Chapter 10 Probes 10 - 37

10.1.6.1. How to Use MMPT3-7 Probe

1 Dial rotation (clockwise)

2 Small dial (left and right)

3 Large dial (up and down)

4 Brake button (left and right)

5 Brake button (up and down)

6 Scan Plane rotation button of lens (counter clockwise)

7 Scan Plane rotation button of lens (clockwise)

[Figure 10.8 MMPT3-7 Probe Control Handle]


10 - 38 V6/XV6 Service Manual

■ Tip Tilting
Use dials to tilt the tip. Rotate the small dial (①) clockwise to tilt the tip to the right. Rotate
the small dial (①) counterclockwise to tilt the tip to the left. Rotate the large dial (②)
clockwise to tilt the tip upward. Rotate the large dial (②) counterclockwise to tilt the tip
downward. When rotating the dials a click is felt at the central position, indicating the tip is
in a straight position (no tilting).

1 Left 3 Up

2 Right 4 Down

[Figure 10.9 Tip Tilting]

■ Dial Brake Mode


In Brake Mode, the movement of the tip is restrained by generating friction to the dials.
The Brake Button ③ restrains the left/right movement of the tip, while the Brake Button ④
restrains the up/down movement of the tip. If the Brake button is protruding and shows
the blue mark as depicted in the left picture below, the probe is in the Brake Active state.

A Brake Inactive

B Brake Active

[Figure 10.10 Brake Mode]


Chapter 10 Probes 10 - 39

■ Lens Scan Plane Rotation


The scan plane of the lens can be rotated from 0° to 180°. Button ⑤ rotates the lens
angle counterclockwise. Button ⑥ rotates the lens angle clockwise.

A Counterclockwise

B Clockwise

[Figure 10.11 Lens Scan Plane Rotation]


10 - 40 V6/XV6 Service Manual

10.1.6.2. Precautions for Using the Probe

Do not use the product with a probe temperature outside the range of 25°C to
42°C. It may cause damage to the human body.
DANGER

■ Prolonged pressure on the esophagus by the tip of the probe may lead to a
Pressure Necrosis phenomenon. Thus, the tip should be removed from the
esophagus wall when you are not scanning by releasing the tip in the straight
position (no tilting). If continuous monitoring is required, the probe tip should
CATUION often be re-positioned.
■ Whenever active scanning is not desired, ‘freeze’ the image and release the
tip in the straight position (no tilting).
■ Position the lens such, that good acoustical contact is established. Before
rotating the scan plane, actuate the Brake Mode on the up/down movement.
If this is not done, rotating the scan plane can mean loss of acoustical
contact.

■ Temperature Change Warning Message


When the temperature of the MPTEE probe exceeds 41°C, the following warnings are
displayed:

Temperature Range Warning Message

41.0°C ≤ MPTEE
Estimated surface temperature is xx.x°C.
Temperature < 42.0°C

42.0°C ≤ MPTEE Estimated surface temperature is xx.x°C. Please reduce the power
Temperature < 42.5°C value.

42.5°C ≤ MPTEE Estimated surface temperature is xx.x°C, near the thermal limit of
Temperature < 43.0°C 43.0°C. Please reduce the power value.

As the surface temperature reached 43.0°C, the power value was


automatically reduced.
MPTEE Temperature = (After the fourth failed attempt) Critical surface temperature. The
43.0°C System has stopped transmitting and returned to the Probe
Selection Dialog. When you reselect the probe, please reduce the
power value if necessary.

Critical surface temperature. The System has stopped transmitting


43.0°C < MPTEE
and returned to the Probe Selection Dialog. When you reselect the
Temperature ≤ 50.0°C
probe, please reduce the power value if necessary.

50.0°C < MPTEE Error condition detected. The System has stopped transmitting and
Temperature or MPTEE returned to the Probe Selection Dialog. Please contact a Service
Temperature < 5.0°C Representative.
Chapter 10 Probes 10 - 41

For the MPTEE probe, the following warnings are displayed depending on the status of
the circuit:
– Short circuit or open circuit: Error condition detected. The System has stopped
transmit and returned to the Probe Selection Dialog. Please contact a Service
Representative.

The probe is intended to be inserted into the human body; always keep it clean.
Never place the probe on the floor.
CAUTION
10 - 42 V6/XV6 Service Manual

10.2. Biopsy
A biopsy is an examination method that surgically extracts tissue from the patient for
examination. The probe and the biopsy kit are used together when conducting a biopsy with the
ultrasonographic image scanner.
The ultrasound imaging system shows the needle, which penetrates through the skin surface
and veins, along with the examined body part, minimizing the potential risks to the patient.

■ Ultrasonographic studies using a biopsy kit must be conducted by a physician


or qualified, experienced healthcare professional. Make sure that you comply
with the preventive safety measures and sterilization procedure in all
environments.
CAUTION ■ Do not use the biopsy kit for purposes other than ultrasonographic studies
such as human anatomical dissection.
■ If the biopsy kit is not used according to the user manual or is used
inappropriately, the user has a responsibility.
■ Please make sure to read the User Manual thoroughly.
Chapter 10 Probes 10 - 43

10.2.1. Biopsy Kit Components


The biopsy kit consists of the bracket, needle guide, and needle. The components vary
depending on the probe type.

[Figure 10.12 Biopsy Kit Components]

① Bracket: Prevents the needle guide from shaking by securing it to the probe tightly.

② Needle Guide: Guides the angle (direction) of the needle so that it may reach the biopsy
target accurately. It also secures the needle so that the needle does not waver.

③ Needle: This is the needle that is inserted into the patient’s body. The biopsy kit supplied
by Samsung Medison does not contain a needle.
Sheath: Prevents the probe and bracket from being contaminated by any unwanted
substances (blood or other bodily fluids) discharged during examinations.
Ultrasound gel: Ensures the best quality images by filling the air space between the probe
and the sheath with the ultrasound gel.
10 - 44 V6/XV6 Service Manual

10.2.2. Using Biopsy Kit

■ Only use needles approved for use in your country.


■ Verify the condition of the biopsy needle before use. Do not use a bent biopsy
needle.
WARNING ■ The needle and guide must be sterilized before use.
■ The needle may bend while penetrating a tissue. Therefore, the precise
location of the needle must be checked by monitoring the echo generated
from the needle.

For more information about conducting ultrasonographic studies using the


biopsy kit, please contact the biopsy kit manufacturer.
CAUTION

10.2.2.1. Before Using Biopsy Kit


Check all the components of the biopsy kit. Ensure that the biopsy kit you are using is
compatible with the probe, system, and system software.

■ Biopsy kits not supplied by Samsung Medison may not be compatible with
the probes. Always use biopsy kits supplied by Samsung Medison, since
improper setup may adversely affect the patient.
WARNING ■ Be sure to read the instructions for installing a sheath and aligning the needle
guide thoroughly before performing a biopsy.
■ Always ensure that the probe and the needle guide are secured on both the
left and the right sides.
Chapter 10 Probes 10 - 45

10.2.2.2. Biopsy Procedure


The system generates a needle guideline through the displayed real-time ultrasound images
to indicate the anticipated path of the needle. Use this guideline to ensure that the needle or
other instrument follows the correct path.
1. Ready the patient according to the procedure appropriate for the examination objectives.
2. Install the sheath and the biopsy kit.
3. Set the system controls for the biopsy procedure. If necessary, apply acoustic gel to the
patient.
4. Scan the patient. Adjust the patient so that the location for examination fits into the
needle guideline on the screen.
5. Insert the needle into the needle guide. Perform the puncture by sliding the needle
through the groove in the guide until the needle intercepts the target.
To prevent the needle from shaking, press down on the top of the needle guide with your
index finger during examinations.

6. When the examination site is reached, take the needle out of the needle guide.
7. Detach the needle guide, bracket, and sheath from the probe.
8. Dispose of the components that are not designed for reuse.

■ Dispose of non-reusable components according to infectious waste


regulations.
■ Sterilize reusable components before disposing of them.
CAUTION
10 - 46 V6/XV6 Service Manual

10.2.2.3. Needle Guide Alignment


Alignment of the needle guide displayed on the system is for the purpose of verifying
whether the needle and the needle guide are properly installed. This must be done prior to
biopsy. If the needle fails to follow an accurate path while verifying the alignment of needle
guide, stop using the product and contact Samsung Medison Customer Support Department.
Reverberation or other tissue artifacts may produce false needle images on the screen.
Ensure the needle path is along the guideline and that you are not using a false needle
image to locate the needle.

■ The needle used for alignment verification must not be used for the actual
procedure. Always use a sterile needle for each biopsy procedure.
■ To assist in accurate projection of the needle, use a straight, new needle for
WARNING each alignment procedure.
■ The needle guideline only displays the projected path of the needle. Since
the default line may be different from the actual line, the actual location of the
needle must be checked by monitoring the echo from the needle.

1. Attach the biopsy kit.


2. Set the system depth for the procedure to be performed and select the Biopsy menu.
3. Immerse the probe into the water bath, and insert the needle into the needle guide.
4. Confirm that the needle image is on the needle guidelines. If it is, then the needle guide is
properly aligned.
5. If the needle image displayed on the screen is out of the needle guideline, check the
needle guide or the probe bracket, etc.
Chapter 10 Probes 10 - 47

10.2.3. Assembling the Biopsy Kit

10.2.3.1. Stainless Steel Biopsy Kit


1. Place a sheath up to the top of the probe’s handle.

2. Mount the bracket onto the probe. If the surface of probe is fluted, mount the bracket in
accordance with it.

3. Insert the needle into the needle guide and start the exam.
10 - 48 V6/XV6 Service Manual

10.2.3.2. Plastic Biopsy Kit


1. Mount the bracket on the probe.

2. Cover the bracket and probe completely with the sheath.

3. Install the needle guide onto the bracket.


Chapter 10 Probes 10 - 49

4. Install the Needle Guide Clip if it is included in the components.

5. Insert the needle into the needle guide and start the exam.

TIP Using Multi Angle


Use the angle adjuster.
10 - 50 V6/XV6 Service Manual

10.2.3.3. Biopsy Specifications

Biopsy
Probes
Bracket Reusable/ Needle Multi Angle
Model Name Components
Material Disposable Gauge Depth

Bracket Reusable
BP-KIT-053 0.591,
Acetal 16, 18, 20,
LA3-14AD 0.984,
[BP-KIT-053-NG] Needle Guide & Copolymer Disposable
22
1.575 (in)
Probe Cover Kit

Bracket Reusable 1.969,


Acetal 14, 16, 18, 3.937,
CA1-7AD BP-KIT-058
Needle Guide & Copolymer 20, 22, 25 5.906,
Disposable
Probe Cover Kit 7.874 (in)

Bracket Reusable 14, 15, 16,


17, 18, 20, 2, 3.79,
CA1-7S Acetal
BP-KIT-094 21, 22, 25G 6.15, 10, 15
CA1-7SD Needle Guide & Copolymer
Disposable 7, 8, 9, 10, (cm)
Probe Cover Kit
12FR

Bracket Reusable 14, 15, 16,


Acetal 17, 18, 20, 3.5, 6.5,
CA3-10A BP-KIT-071
Needle Guide & Copolymer 21, 22, 23, 8.5, 12 (cm)
Disposable
Probe Cover Kit 8.5FR

BP-KIT-079 Needle Guide ABS Disposable 16-18 2°

Stainless
BP-KIT-080 Needle Guide Reusable 16-18 2°
steel

BP-KIT-081 Needle Guide ABS Disposable 16-18 0°


EA2-11AR
EA2-11ARE Stainless
BP-KIT-082 Needle Guide Reusable 16-18 0°
EA2-11AV steel
EA2-11AVE
BP-KIT-088 Needle Guide PC Disposable 16-18 2°

BP-KIT-089 Needle Guide PC Disposable 16-18 0°

Stainless
BP-KIT-090 Needle Guide Reusable 16-18 0°
steel

Bracket Reusable
CV1-8A BP-KIT-059 Acetal 14, 16, 18, 1.969,
CV1-8AE [BP-KIT-059-NG] Needle guide & Copolymer 20, 22, 25 3.937 (in)
Disposable
Probe Cover Kit

BP-KIT-085 Needle Guide PC Disposable 16-18 1.8°


EV2-10A
Stainless
BP-KIT-086 Needle Guide Reusable 16 0°
steel

Bracket Reusable 14, 15, 16, 2, 3.22,


Acetal
PA1-5A BP-KIT-102 17, 18, 20, 4.98, 8.28,
Copolymer
Needle guide & Disposable 21, 22, 25G 13.65 (cm)
Chapter 10 Probes 10 - 51

Biopsy
Probes
Bracket Reusable/ Needle Multi Angle
Model Name Components
Material Disposable Gauge Depth
Probe Cover Kit 7, 8, 9, 10,
12FR

Bracket Reusable 14, 15, 16,


Acetal 17, 18, 20, 1.5, 2, 3, 4
PA4-12B BP-KIT-067 21, 22, 23G
Copolymer (cm)
Needle Guide Disposable
8.5FR

Bracket Reusable 14, 15, 16,


17, 18, 20, 2, 3.22,
Acetal 21, 22, 25G
LA2-9S BP-KIT-103 4.76, 7.09,
Needle guide & Copolymer
Disposable 7, 8, 9, 10, 9.82 (cm)
Probe Cover Kit
12FR

miniER7 BP-KIT-105 Needle Guide ABS Disposable 16–18G 0°

Bracket Reusable 14, 15, 16,


17, 18, 20, 1.00, 1.73,
Acetal 21, 22, 25G
L3-22 BP-KIT-107 2.56, 3.65,
Needle guide & Copolymer
Disposable 7, 8, 9, 10, 4.76 (cm)
Probe Cover Kit
12FR

When installing a biopsy kit for use in the S-Fusion™ Prostate preset, do not use
the two biopsy kits (BP-KIT-079, BP-KIT-080) of EA2-11AR, EA2-11ARD, EA2-
11AV and EA2-11AVD probes as it may cause damage to the probe lenses or
cases.
CAUTION

None of the model numbers of biopsy kit is available in Canada.

NOTE
10 - 52 V6/XV6 Service Manual

10.2.4. Cleaning and Disinfecting the Biopsy Kit


Reusable Biopsy brackets must be cleaned and disinfected after each use. Read the biopsy kit
user manual carefully before cleaning and disinfecting.

■ Always use protective equipment such as face mask, eyewear and gloves
when cleaning and disinfecting biopsy kit.
■ Disposable biopsy brackets and disposable biopsy kit parts must be disposed
WARNING of as medical waste after each use.

10.2.4.1. Cleaning Biopsy Kit


Cleaning is an important procedure that must be carried out before disinfecting or sterilizing
reusable biopsy brackets.
1. After use, remove the biopsy kit from the probe.
2. Disassemble the biopsy kit into its component parts and discard the disposable parts in
the clinical waste bin.
3. Rinse with water to remove excess dust from the device.
4. Prepare an enzymatic solution (e.g. Enzol), following the manufacturer’s instructions for
proper dilution and immersion duration. Immerse the parts in the detergent and allow
soaking for the prescribed time.
5. Clean the parts thoroughly using a soft brush. Clean the lumen using an appropriately-
sized brush. If any visible parts remain, continue brushing until clean.
6. Rinse the device thoroughly with water and dry the components carefully. Dispose of the
used drying cloths appropriately according to the local medical waste regulations.

10.2.4.2. Disinfecting Biopsy Kit


It must be disinfected after each use. Refer to the user instructions of the disinfectant for
details of proper storage, use, and disposal of the disinfectant.

■ Always refer to the biopsy kit user manual for the latest information on the
supported methods and disinfectants.
■ A stainless steel biopsy bracket is reusable; a plastic biopsy bracket may be
reusable. A plastic biopsy needle guide is in general disposable.
CAUTION
■ Plastic biopsy kits can only be disinfected by using a chemically compatible
cold-disinfectant, either manually or with a medical washer-disinfector. Do not
use an autoclave, gas or radiation.

1. Clean the reusable parts of the biopsy bracket, according to the “Cleaning Biopsy Kits” in
this chapter.
2. After cleaning, choose high-level disinfectants that are compatible with your reusable
biopsy bracket. For the compatible disinfectants, see biopsy kit manufacturer’s
instructions.
Follow disinfectant manufacturer’s instructions for preparation, temperature, solution
strength, and duration of contact. Ensure that the solution strength and duration of
contact are appropriate for the intended clinical use of the device.
Chapter 10 Probes 10 - 53

3. Immerse the parts in high-level disinfectants and allow soaking for the prescribed time
following the manufacturer’s instructions.
4. Rinse the device thoroughly with sterile water and dry the components. Dispose of the
used drying cloths appropriately according to the local medical waste regulations.
5. Inspect the components for damage such as cracks, rust, or breakage; and if damage is
found, stop using the device and contact the Samsung Medison Customer Support
Department.
6. Properly pack the disinfected bracket and label it according to the local regulations.

10.2.4.3. Sterilizing Stainless Steel Biopsy Kit


The main difference between sterilization and high-level disinfection is the length of time that
the device is immersed and the type of disinfectant or sterilant that is used.
1. Clean the reusable parts of the biopsy bracket, according to the “Cleaning Biopsy Kits” in
this chapter.
2. After cleaning, choose high-level disinfectants/sterilants that are compatible with your
reusable biopsy bracket. For the compatible disinfectants/sterilants, see biopsy kit
manufacturer’s instructions.
Follow sterilant/HLD disinfectant manufacturer’s instructions for preparation, temperature,
solution strength, and duration of contact. Ensure that the solution strength and duration
of contact are appropriate for the intended clinical use of the device.
3. Immerse the parts in high-level disinfectants/sterilants and allow soaking for the
prescribed time, or steam (autoclave) following the manufacturer’s instructions.
Follow the instructions on the sterilization label for the duration of immersion or autoclave
for sterilization. Do not immerse or steam the biopsy bracket longer than the minimum
time needed for your level of sterilization.
4. Remove the biopsy bracket from the sterilization solution or steam (autoclave) after the
recommended sterilization time has elapsed.
5. Only for immersion method, rinse the device thoroughly with sterile water and dry the
components. Dispose of the used drying cloths appropriately according to the local
medical waste regulations.
6. Inspect the components for damage such as cracks, rust, or breakage; and if damage is
found, stop using the device and contact the Samsung Medison Customer Support
Department.
7. Properly pack the sterilized bracket and label it according to the local regulations.
Chapter 11
11. 서비스 부품

제11장. 서비스 부품 리스트

Service Parts List


11.1. Overview .................................................................................................. 2

11.2. Body Cover Parts .................................................................................... 2

11.3. Control Panel Parts ................................................................................. 6

11.4. System Cable Parts .............................................................................. 20


11.5. System Parts ......................................................................................... 23
11 – 2 V6/XV6 Service Manual

11.1. Overview
These parts codes may not be the latest Version. If you want the latest Version, refer to CM.
The latest version of the CM(Compatibility Matrix table) can be downloaded from GSIS.

11.2. Body Cover Parts


Categ Part
Description Part Image
ory Code

Body
MI63-
Cover COVER-SIDE-R
02464A
Parts

Body
MI63-
Cover COVER-SIDE-L
02465A
Parts

Body
MI63-
Cover COVER-MONITOR CABLE
03039A
Parts
Chapter 11. Service Parts List 11 – 3

Categ Part
Description Part Image
ory Code

Body
MI63-
Cover COVER-BODY REAR UPPER
03102B
Parts

Body
MI63-
Cover COVER-FOOT
03135A
Parts

Body
MI63-
Cover COVER-LIFT TOP
03136A
Parts

Body
MI64-
Cover HANDLE-REAR
02266A
Parts

Body
MI66-
Cover WHEEL Lock(Front, Back Left)
01438A
Parts
11 – 4 V6/XV6 Service Manual

Categ Part
Description Part Image
ory Code

Body
MI66-
Cover WHEEL Free(Back Right)
01439A
Parts

Body
MI63-
Cover COVER-CP PROBE HOLDER L
03105B
Parts

Body
MI63-
Cover COVER-CP PROBE HOLDER R
03106B
Parts

Body
MI97-
Cover ASSY COVER-PSA
04710A
Parts

Body
MI97-
Cover ASSY COVER-TOP BODY
04712A
Parts
Chapter 11. Service Parts List 11 – 5

Categ Part
Description Part Image
ory Code

Boddy
MI97-
Cover ASSY COVER-REAR BODY
05690A
Parts
11 – 6 V6/XV6 Service Manual

11.3. Control Panel Parts


Part
Category Description Part Image
Code

Control
MI61-
Panel HOLDER-PROBE S
04057B
Parts

Control
MI61-
Panel HOLDER-PROBE M
04058B
Parts

Control
MI61-
Panel HOLDER-PROBE L
04059B
Parts

Control
MI61-
Panel HANGER-CP BOTTOM
04748A
Parts
Chapter 11. Service Parts List 11 – 7

Part
Category Description Part Image
Code

Control
MI63-
Panel COVER-CP SPEAKER
03357A
Parts

Control
MI63-
Panel COVER-CP REAR
02798B
Parts

Control
MI63-
Panel COVER-CP REAR PARTIAL
02799B
Parts

Control
MI59-
Panel TRACK BALL
01351A
Parts

Control
MI61-
Panel HOLDER-CP KNOB
03227A
Parts

Control
MI63-
Panel COVER-USB RUBBER
02459A
Parts
11 – 8 V6/XV6 Service Manual

Part
Category Description Part Image
Code

Control
MI63-
Panel COVER-CP TOP MAIN
03107B
Parts

Control
MI64-
Panel KNOB-ENCODER SINGLE
02144A
Parts

Control
MI64-
Panel BUTTON-TOGGLE SWITCH
02331A
Parts

Control
MI64-
Panel KNOB-ENCORDER
02163A
Parts

Control
MI64-
Panel BUTTON-MAIN RUBBER
02334A
Parts
Chapter 11. Service Parts List 11 – 9

Part
Category Description Part Image
Code

Control
MI67-
Panel RUBBER-CAP KEY
01109A
Parts

Control
MI92-
Panel ASSY BOARD-CP_SWITCH
03176A
Parts

Control
MI92-
Panel ASSY BOARD-CP_USB
03178A
Parts

Control
MI92-
Panel ASSY BOARD-CP
03294A
Parts

Control
MI95-
Panel ASSY CAP KEY-POWER
01629A
Parts

Control
MI64-
Panel BUTTON-ENCORDER M
02014A
Parts
11 – 10 V6/XV6 Service Manual

Part
Category Description Part Image
Code

Control
MI64-
Panel BUTTON-ENCORDER 2D
02017A
Parts

Control
MI64-
Panel BUTTON-ENCODER PD(Y
02235A
Parts

Control
MI64-
Panel BUTTON-ENCODER PW(Z)
02237A
Parts

Control
MI64- BUTTON-ENCODER
Panel
02239A C(REF.SLICE)
Parts

Control
MI64-
Panel BUTTON-ENCODER MENU
02241A
Parts
Chapter 11. Service Parts List 11 – 11

Part
Category Description Part Image
Code

Control
MI63-
Panel COVER-LIGHT DEPTH
02814A
Parts

Control
MI63-
Panel COVER-LIGHT FOCUS
02912A
Parts

Control
MI63-
Panel COVER-LIGHT ZOOM
02913A
Parts

Control
MI63-
Panel COVER-LIGHT BODY MARKER
03143A
Parts
11 – 12 V6/XV6 Service Manual

Part
Category Description Part Image
Code

Control
MI63-
Panel COVER-LIGHT ANGLE
03211A
Parts

Control
MI64-
Panel BUTTON-TRACKBALL ERASER
02148A
Parts

Control
MI64- BUTTON-TRACKBALL
Panel
02149A MEASURE
Parts

Control
MI64-
Panel BUTTON-FREEZE
02150A
Parts
Chapter 11. Service Parts List 11 – 13

Part
Category Description Part Image
Code

Control
MI64-
Panel BUTTON-TRACKBALL CALIPER
02151A
Parts

Control
MI64-
Panel BUTTON-TRACKBALL CHANGE
02155A
Parts

Control
MI64- BUTTON-TRACKBALL ARC
Panel
02157A FLAT
Parts

Control
MI64-
Panel BUTTON-TRACKBALL SET
02158A
Parts
11 – 14 V6/XV6 Service Manual

Part
Category Description Part Image
Code

Control
MI64-
Panel BUTTON-SINGLE
02187A
Parts

Control
MI64-
Panel BUTTON-QUAD
02189A
Parts

Control
MI64-
Panel BUTTON-U1
02192A
Parts

Control
MI64-
Panel BUTTON-U2
02193A
Parts
Chapter 11. Service Parts List 11 – 15

Part
Category Description Part Image
Code

Control
MI64-
Panel BUTTON-P1
02195A
Parts

Control
MI64-
Panel BUTTON-P2
02196A
Parts

Control
MI64-
Panel BUTTON-P3
02197A
Parts

Control
MI64-
Panel BUTTON-3D
02204A
Parts
11 – 16 V6/XV6 Service Manual

Part
Category Description Part Image
Code

Control
MI64-
Panel BUTTON-4D
02205A
Parts

Control
MI64-
Panel BUTTON-QSCAN
02207A
Parts

Control
MI64-
Panel BUTTON-TRACKBALL TEXT
02251A
Parts

Control
MI64-
Panel BUTTON-POINTER
02252A
Parts
Chapter 11. Service Parts List 11 – 17

Part
Category Description Part Image
Code

Control
MI64-
Panel BUTTON-EXIT SQUARE
02253A
Parts

Control
MI64-
Panel BUTTON-LEFT SQUARE
02254A
Parts

Control
MI64-
Panel BUTTON-RIGHT SQUARE
02255A
Parts

Control
MI64-
Panel BUTTON-CW CIRCLE
02256A
Parts

Control
MI61-
Panel HANGER-REAR CABLE
01658C
Parts
11 – 18 V6/XV6 Service Manual

Part
Category Description Part Image
Code

Control
MI63-
Panel ABSORBER-SPEAKER
02458A
Parts

Control
MI96-
Panel ASSY MONITOR UNIT 23.8”
01527D
Parts

Control
MI96-
Panel ASSY MONITOR UNIT 21.5”
01558B
Parts

Control
MI96-
Panel ASSY TOUCH PANEL 14”
01598A
Parts

Control
6502-
Panel CABLE CLAMP
001244
Parts

Control
MI61-
Panel MAGNET
01705A
Parts
Chapter 11. Service Parts List 11 – 19

Part
Category Description Part Image
Code

Control
MI97-
Panel ASSY ARM-MECHANISM
05248B
Parts

Control
MI67-
Panel RUBBER-DAMPER LCD
01060A
Parts

Control
MI95- ASSY ALPHA KEYBOARD-WR
Panel
01850A (Option, WR)
Parts

Control
MI95- ASSY ALPHA KEYBOARD-WR
Panel
01851A (Option, DE)
Parts

Control
MI95- ASSY ALPHA KEYBOARD-WR
Panel
01852A (Option, RU)
Parts
11 – 20 V6/XV6 Service Manual

11.4. System Cable Parts


Categor Part
Description Part Image
y Code

System
MI39-
Cable CBF HARNESS-ECG_USB
02025A
Parts

System
MI39-
Cable CBF HARNESS - PRINTER_USB
02525A
Parts

System
MI39-
Cable CBF HARNESS-PRINTER_POWER
02229A
Parts

System
MI39-
Cable CBF HARNESS - SPEAKER_EXT
02475A
Parts
Chapter 11. Service Parts List 11 – 21

Categor Part
Description Part Image
y Code

System
MI39-
Cable CBF-TRACKBALL_DATA
02290A
Parts

System
MI39-
Cable CBF CABLE-SPEAKER
01718A
Parts

System
MI39-
Cable CBF HARNESS-CP PWR
02523A
Parts

System
MI39-
Cable CBF HARNESS- USB 3.0 MID
02527A
Parts

System
MI39-
Cable CBF HARNESS - DP
02522A
Parts
11 – 22 V6/XV6 Service Manual

Categor Part
Description Part Image
y Code

System
MI39- CBF HARNESS-GEL PWR_R
Cable
02526A 140mm
Parts

System
MI41-
Cable TRACKBALL_FPCB
01741A
Parts

System
MI39-
Cable WIRE HARNESS-AC CORD(FILTER)
01378A
Parts

System
MI39-
Cable CBF HARNESS-GEL PWR
02306A
Parts

System
MI39- CBF HARNESS ALPHAKEY_DATA_CB
Cable
02298A (Option)
Parts

System
MI39-
Cable CBF HARNESS - BOTTOM_FAN EXT
02459A
Parts

System

Cable MI39- CBF HARNESS - DRIVE USB

Parts 02480A (OPTION)

System
MI39-
Cable CBF HARNESS - DRIVE POWER
02518A
Parts (OPTION)
Chapter 11. Service Parts List 11 – 23

11.5. System Parts


Part
Category Description Part Image
Code

System MI92-
ASSY BOARD-BP
Parts 03304A

System MI92-
ASSY BOARD-CP DOCKING
Parts 03168C

System 4301-
BATTERY-LI(2ND)
Parts 001104

System MI96-
ASSY STORAGE-NVME,512
Parts 01596A

System MI96- ASSY STORAGE-NVME,1T

Parts 01553A (Option)

System MI59-
CARD-VGA;GTX 1650
Parts 01270A

System MI96-
ASSY PC-PC MODULE
Parts 01562A
11 – 24 V6/XV6 Service Manual

Part
Category Description Part Image
Code

System MI92-
ASSY BOARD-PSA 3P
Parts 03207B

System MI92-
ASSY BOARD-PSA 3P-PEN
Parts 03202B

System MI92- ASSY BOARD-PSA 4P

Parts 03201B (Option)

System MI92- ASSY BOARD-PSA 4P-PEN

Parts 03184B (Option)

System MI82-
ASSY BOARD-MAIN
Parts 02624A

System MI92- ASSY BOARD-CW

Parts 02884A (Option)

System MI43-
BATTERY
Parts 01010A

System MI96-
ASSY Battery Charger
Parts 01514A
Chapter 11. Service Parts List 11 – 25

Part
Category Description Part Image
Code

System MI96- ASSY POWER-DDM

Parts 01600A (No CAP BD)

ASSY POWER-DDM
System MI96-
include CAP BD
Parts 01599A
(S-Shearwave)

System MI92- ASSY BOARD-

parts 03188A POWER_BASE;V6,0A,POWER_BASE

System MI92- ASSY BOARD-

parts 03194A POWER_HV;V6,0A,Power_HV

System MI92- ASSY BOARD-POWER_HV_BASIC

parts 03388A (NO SWE);V6,A,P

System MI96-
ASSY FAN-BOTTOM FAN
Parts 01511A

System MI96-
ASSY FAN-TOP;HS80,MID FAN
Parts 01512A

System MI97-
ASSY LIFT
Parts 05653A
11 – 26 V6/XV6 Service Manual

Part
Category Description Part Image
Code

System MI92- ASSY BOARD-REAR

Parts 02882A (Option)


Chapter 11. Service Parts List 11 – 27
11 – 28 V6/XV6 Service Manual

V6 Service Manual

Manufacturer : SAMSUNG MEDISON / SERVICE SOLUTION TEAM


Address : East Central Tower, 1077, Cheonho-dae-ro, Gangdong-gu, Seoul, Korea
Tel : 02-2194-1000
Fax : 02-2194-1028
Website: www.samsungmedison.com

All information in this manual is owned by Samsung Medison Co., Ltd. and is protected by related laws.
This manual is a textbook designed for the training of the service engineers at Samsung Medison Training
Center.
Therefore, unauthorized use of this manual is prohibited, and legal sanctions are imposed if the contents
are reproduced, or distributed.

You might also like