Professional Documents
Culture Documents
Virtuo Ingles
Virtuo Ingles
BACT/ALERT VIRTUO
Microbial Detection
System User Manual
050574-02 - en - 2022-08 -
EN
IVD
Table of Contents
Safety Information...................................................................................................2-1
System Compliance..................................................................................................... 2-1
Electrical Warnings...................................................................................................... 2-1
Electrical Grounding..........................................................................................2-2
Electrical and Electronic Recycling................................................................... 2-3
USB Devices................................................................................................................2-3
Instrument Labels........................................................................................................ 2-4
Safety Precautions.....................................................................................................2-12
System Configuration.............................................................................................6-1
Accessing the Configuration Screens.......................................................................... 6-2
Configuring Bottle Type Settings..................................................................................6-3
Configuring the Maximum Test Time.................................................................6-4
Sample Volume Measurement..........................................................................6-5
Sample Volume Variation..................................................................................6-5
Setting the Sample Lower Variation..................................................................6-6
Setting the Sample Upper Variation..................................................................6-7
Configuring Bank Settings........................................................................................... 6-7
Naming Instruments.......................................................................................... 6-8
Naming Banks and Labs................................................................................... 6-8
Configuring the Bank Near Full Threshold........................................................ 6-8
Configuring the Incubation Temperature Set Point........................................... 6-8
Configuring the Waste Near Full Threshold...................................................... 6-9
Configuring the Maximum Allowable Time in Retrieval Area............................ 6-9
Configuring Sound Settings............................................................................ 6-10
Configuring Alarm Settings..............................................................................6-11
Configuring General Software Settings...........................................................6-13
Configuring Security Settings..........................................................................6-17
Configuring LIS Settings (standalone configuration only)............................... 6-20
Configuring Bottle Report Settings..................................................................6-22
Managing User Accounts and User Groups.............................................................. 6-24
Accessing the User Management................................................................... 6-24
Password Policy..............................................................................................6-24
Troubleshooting...................................................................................................... 8-1
Resolving Alarms and Alerts........................................................................................8-1
Viewing Alarms and Alerts................................................................................ 8-1
Alarm and Alert Causes and Resolutions......................................................... 8-2
Loading Bottles across a Bank........................................................................8-23
Opening the Instrument for Internal Access....................................................8-24
Closing the Door............................................................................................. 8-29
Unjamming the Robot..................................................................................... 8-32
Audit Trail...................................................................................................................8-35
Accessing the Audit Trail.................................................................................8-35
Audit Trail for Standalone and Site Configuration (MAESTRIA™).................. 8-36
Audit Trail for Site Configuration (MYLA®)...................................................... 8-39
Viewing/Printing Saved Audit on an External PC............................................8-42
Collect Logs............................................................................................................... 8-43
Generating the Log File...................................................................................8-43
Notification Center (Standalone Configuration only)..................................................8-45
Configuring the Notification Center for BACT/ALERT® VIRTUO® Events......8-46
Revision History
Note: The screens and figures shown are intended as illustrations only and must not be interpreted
as actual representations of data, results or equipment.
Screens and equipment are not shown to scale.
IMPORTANT: Please read this document carefully before using the system.
For users in the European Union (Regulation (EU) 2017/746) and in countries with similar
requirements: Should a serious incident occur during the use of this device or as a result of its
use, please report it to the manufacturer and/or their authorized representative as well as to
your national authority.
The user documentation uses several types of statements to alert you to important
information. Important information is labeled in text and identified using symbols.
Statement Types
The statement types are Warning, Caution, Important, and Note. The following examples
define each statement type. The general caution symbol is used in these examples, but other
symbols (see Standard Symbols) may be used instead.
The warning messages in this document mainly refer to:
WARNING
A Warning statement alerts the user to the possibility of injury, death, or
other serious adverse reactions associated with the use or misuse of the
device.
Intellectual Property
Limited Warranty
bioMérieux warrants the performance of the product for its stated intended use provided that
all procedures for usage, storage and handling, shelf life (when applicable), and precautions
are strictly followed as detailed in the instructions for use (IFU).
Except as expressly set forth above, bioMérieux hereby disclaims all warranties, including any
implied warranties of merchantability and fitness for a particular purpose or use, and disclaims
all liability, whether direct, indirect or consequential, for any use of the reagent, software,
instrument and disposables (the "System") other than as set forth in the IFU.
Customer acknowledges and agrees that use of the System for testing of sample types or for
indications other than those described in the IFU is done solely at the Customer’s own risk.
Customer acknowledges and agrees that it is Customer’s sole and exclusive responsibility to
validate the System for any such intended use, and to determine whether the System is
suitable for that intended use. The performance of any validation studies and the subsequent
use of the System based on Customer’s validation studies shall be the Customer’s sole risk
and responsibility.
Product warranty details can be obtained from bioMérieux or your local distributor (contact
information available on www.biomerieux.com).
1 Introduction to the System
BACT/ALERT® VIRTUO® Microbial Detection System and this manual are intended for
laboratory users who are trained in microbiology and good laboratory practices.
WARNING
A potential hazard (false negative test results) may occur if unqualified users
operate the instrument.
Instrument Training
Users should be trained on the system by qualified bioMérieux personnel before
implementing the instrument in their laboratory. This training will include all functions required
to operate the system including accessing the software, daily workflow tasks such as bottle
loading and unloading, and resolving instrument alarms and alerts. Contact your local
bioMérieux representative to schedule training. Retraining may be required following any
instrument update, depending on the specific changes included in the update.
Intended Use
BACT/ALERT® VIRTUO® Microbial Detection System is an automated microbial test system
capable of incubating, agitating, and continuously monitoring for the detection of aerobic,
facultative, and anaerobic microorganism growth from blood and other normally sterile body
fluids and from blood platelet samples.
In blood bank facilities, the BACT/ALERT® VIRTUO® Microbial Detection System will provide
in-process testing for the presence or absence of microorganisms.
CAUTION: bioMérieux shall not be liable as to any defect (including, but not
limited to, incorrect results that may affect patient outcomes) from these
conditions:
• Abuse of the instrument
• Failure to operate and maintain the instrument in accordance with the
User Manual
• Operation of the instrument by a person who has not been trained in its
operation by bioMérieux
• Repair, service, alteration, or modification of the instrument by any person
other than service personnel of bioMérieux
• Modification, change or reuse of the disposables supplied by bioMérieux
for use in the instrument
• Use of disposables (bottles) other than those supplied by bioMérieux
• Use of disposables beyond the expiration date indicated on the label
• Reuse of bottles. Never reuse bottles for multiple patients or samples.
WARNING
The consequences of reusing bottles may be incorrect results
since using bottles for multiple patients or samples is
contradictory to the intended use of the system. See the
Instructions for Use for BACT/ALERT® bottles.
Benefits of Use
General Statements
This section provides important statements that apply to all products. Equipment meets the
requirements and standards stated in the certificate supplied with it.
WARNING
Equipment is intended for professional use only.
Laboratory personnel should be qualified and adhere to the principles of
good laboratory practice.
All the user documents supplied must be read prior to use of the equipment.
Under no circumstance should the user dismantle equipment due to the risk
of touching dangerous parts, including parts that may be infectious or
connected to a source of electric power.
Do not obstruct the equipment and hardware ventilation apertures, and leave
sufficient clearance around the equipment for the circulation of air.
All biological materials should be considered as potentially infectious.
Suitable individual protective equipment is required when handling chemical
or biological substances.
bioMérieux is in no case liable for the harmful consequences of incorrect use
or improper handling of these substances.
WARNING
Electromagnetic Compatibility (EMC):
The EMC class of the equipment is indicated on the certificate supplied with
it.
If equipment is a class A product, it may cause radio interference in a
domestic environment, in which case the user will be required to correct the
interference at his own expense.
Do not use this device near strong sources of electromagnetic radiation (for
example, intentionally unprotected radio-electric sources), which could
interfere with the operation of the equipment.
It is recommended to evaluate the electromagnetic environment before
starting the device.
WARNING
In order to avoid computer viruses or abnormal functioning of your
equipment, never download any software other than those ensuring the
protection of your network and those provided or recommended by
bioMérieux.
It is your responsibility to secure your network and ensure this protection is
appropriate and maintained. It is recommended to use all appropriate means
(including antivirus software, security patches, firewall) to protect your
network from virus intrusion, unauthorized use, alteration, manipulation and
disclosure.
In an effort to reduce the risk of spreading a virus to bioMérieux equipment,
it is recommended that only bioMérieux supplied USB devices are used with
bioMérieux equipment. The use of personal USB devices is not
recommended. To avoid computer viruses and the potential loss of
functionality and/or results, use caution when transferring USB devices
between computers. Do not use USB devices intended for bioMérieux
equipment in other computers that do not have current antivirus software
installed and active.
All computer media (CD, DVD, USB key) supplied with this equipment should
be stored and stocked in a suitable location.
Only modify the software configuration parameters you are authorized to
modify and which are described in the user documentation.
WARNING
Decontamination of equipment at the end of its life cycle:
The following instructions must be followed by all users in countries where
local legislation imposes the treatment and recycling of equipment at the end
of its life cycle.
As a general rule, and as a precautionary measure, any part of the equipment
(including sub-assemblies, components and materials) considered to be
potentially infectious, must be decontaminated, whenever possible, or
removed if decontamination is impossible or presents a risk.
Any part considered to be potentially infectious, which is not
decontaminated, must be removed from the instrument before following the
normal channels for elimination of infectious products, in accordance with
local regulations.
The decontamination instructions in the user documentation correspond to
the parts of the equipment that are potentially infectious according to their
intended use. These operations must be performed before the equipment is
transferred to a third party.
However, bioMérieux cannot exclude that other parts of the equipment have
not been contaminated in other circumstances, in particular as the result of
spillage of infectious substances. In this case, the user is solely responsible
for decontaminating these parts or removing them before they follow the
normal channels for elimination of infectious products.
WARNING
This statement only applies to European countries with regard to the waste
electrical and electronic equipment European directive:
You can play an important role in contributing to reuse, recycling, and other
forms of recovery of waste electrical and electronic equipment. Sorting this
type of waste significantly reduces potential negative effects on the
environment and human health as a result of the presence of hazardous
substances in electrical and electronic equipment.
At the end of the life cycle of this product, do not dispose of the product as
unsorted municipal waste, even if it is decontaminated. It is imperative that
you contact bioMérieux to assure its appropriate disposal.
IMPORTANT: Electrical or other connections should only be made using the accessories supplied
with the equipment.
IMPORTANT: The accuracy of results obtained with this equipment depends on the maintenance
operations described in the user documentation (user maintenance and/or periodic
preventive maintenance performed by bioMérieux).
IMPORTANT: The user should be aware that if the maintenance operations are not performed, are
only partially performed, or are not performed as described in the user
documentation, bioMérieux is in no case liable for any false test results obtained.
IMPORTANT: It is recommended to keep the original packaging materials in case the equipment
needs to be moved. Any damage directly or indirectly resulting from the transport of
the equipment without adequate containers will not be covered by the warranty.
Standard Symbols
The following table defines symbols that may appear in the instructions for use or on the
instrument, package inserts, or packaging. When surrounded by a triangle on a yellow
background, the symbol highlights an immediate warning and is positioned on the instrument
itself.
Irritant Recyclable
Equipotentiality ON (supply)
It is essential that the warnings, cautions and safety requirements contained in this document
are read and understood by the user before operating the system.
Warning symbols have been placed on the system to draw your attention to areas of potential
hazards.
System Compliance
The BACT/ALERT® VIRTUO® System conforms to the relevant European regulations for
electrical safety and electromagnetic compatibility (EMC).
• This IVD instrument complies with the emissions and immunity requirements of IEC
61326.
• This is a Class A product. This instrument was designed and tested against CISPR 11,
class A. In a domestic envrironment this product may cause radio interference, in which
case the user may be required to take adequate measures.
• bioMérieux recommends the evaluation of the electromagnetic environment before
operating the instrument.
Electrical Warnings
BACT/ALERT® VIRTUO® is designed and tested in accordance with these standards and is
supplied in a safe condition. A CB Certification and Construction file is established for the
apparatus.
• EN 61010-2-101:2002, Safety Requirements for Electrical Equipment for Measurement,
Control, and Laboratory Use; Part 2-101: Particular Requirements for In-Vitro Diagnostic
(IVD) Medical Equipment.
• EN 60825-1:2007, Safety of laser products: Part 1: Equipment Classification and
Requirements; and 21 CFR 1040, Performance Standards for Light-Emitting Products.
• EN 55011-2007, Radiated Emissions
• European Parliament Directive 2002/95/EC, 2003, Restriction and Use of Certain
Hazardous Substances in Electrical and Electronic Equipment (RoHS).
WARNING
Adhere to the following warnings to ensure safe operation and to maintain
the apparatus in a safe condition:
• When cycling AC power on BACT/ALERT® VIRTUO® components, wait
several seconds between the Off cycle and the On cycle. Rapid cycling of
the power switch may cause the instrument to malfunction.
• Intentional interruption of the protective conductor inside or outside the
apparatus, or disconnection of the protective ground terminal, is
prohibited.
• Disconnect the apparatus from all voltage sources before it is opened for
any adjustment, replacement, maintenance, or repair.
• Do not perform any adjustments, maintenance, or repairs of the opened
apparatus while under voltage. If this is unavoidable, a skilled person who
is aware of the hazard involved must maintain the instrument.
• If the BACT/ALERT® VIRTUO® instrument has been impaired, disconnect
the power cord to render it inoperative and secure it against any
unintended operation. If moisture is evident, turn off the machine at the
breaker before removing the power cord and fuse.
WARNING
A potential hazard of electrical shock exists during service and maintenance,
electrical checks, or manual load mode.
WARNING
Ensure that product labeling is not corrupted during cleaning and handling
procedures. The consequences of not following the proper cleaning
procedures may be loss of marking and lack of electrical safety/ hazard
awareness.
WARNING
Contact your local bioMérieux representative if any of the following
conditions occur:
• System shows visible damage.
• System fails to perform intended measurements.
• System has been subjected to storage under unfavorable conditions (for
example, above 90% humidity, extreme temperatures, dusty environment,
prolonged storage).
• System has been subjected to severe transport stresses.
Electrical Grounding
An electrical ground is required for this instrument. Before installing the instrument, ensure
that a grounded wall receptacle is available for each instrument, which is factory equipped
with a power supply cord that has the appropriate grounding plug. Plug the instrument into the
UPS and plug the UPS into a mating grounding type wall receptacle in accordance with the
National Electrical Code and applicable local codes and ordinances for this type of
installation.
WARNING
Do not remove the power cord’s ground prong under any circumstances.
1 US Standard
2 220 Volt - European
WARNING
This statement only applies to European countries with regard to the waste
electrical and electronic equipment European directive:
You can play an important role in contributing to reuse, recycling, and other
forms of recovery of waste electrical and electronic equipment. Sorting this
type of waste significantly reduces potential negative effects on the
environment and human health as a result of the presence of hazardous
substances in electrical and electronic equipment.
At the end of the life cycle of this product, do not dispose of the product as
unsorted municipal waste, even if it is decontaminated. It is imperative that
you contact bioMérieux to assure its appropriate disposal.
USB Devices
WARNING
The following potential risks exist if bioMérieux-supplied USB devices are
not used:
• Spreading a virus to bioMérieux equipment
• Loss of function
• Loss of results
Instrument Labels
The BACT/ALERT® VIRTUO® instrument has multiple labels on various parts of the
instrument. Labels may be used to provide information or to make the user aware of potential
hazards. It is important to be familiar with the location and meaning of the labels on the
instrument.
The instrument label is located inside the instrument and is visible upon opening the waste
compartment door. The instrument label contains the part number of the instrument (REF),
the name of the instrument, the serial number (S/N), the instrument voltage, and the
instrument product identity.
Note: The XXXXX in the instrument serial number will be replaced by the assigned serial number.
IMPORTANT: Symbols and information listed on the instrument label are accurate; however, the
placement of the symbols and information on the instrument label are subject to
change.
The instrument contains a Class 1 laser. Laser caution labels appear on the instrument to
inform the user that there is a presence of a Class 1 Laser Product in the instrument. The
expanded Class 1 Laser Product label appears on the rear of the instrument. The simplified
label appears on the door handle.
WARNING
Do not try to access the interior of the laser case for any reason. Failure to
comply may result in exposure to radiation and result in personal injury and
damage to the instrument.
Labels appear on the door handle to inform the user about the presence of a Class 1 laser,
biological risks, and risk of potential pinch points. The user is also cautioned to consult
documentation that accompanies the instrument.
WARNING
All biological samples and quality control (QC) products incubated in this
system, as well as all waste in the waste containers, should be treated as
potentially biohazardous materials. All materials and mechanical
components associated with the waste systems should be handled
according to safe microbiological practices in compliance with the
installation site's biohazard procedures. Use the personal protective
equipment recommended by the facility when handling any of these
components, including gloves, safety glasses, and a lab coat.
The system must be decontaminated before its covers are removed by
service personnel. Ensure that appropriate decontamination is carried out if
hazardous materials are spilled on or into the equipment or surrounding
areas.
WARNING
A potential biohazard exists in these situations:
• A bottle is not correctly identified.
• A bottle leaks from damage or another cause.
• Entrapped biohazard substances are not properly cleaned.
• Proper cleaning guidelines are not correctly followed.
• Surfaces are not cleaned properly.
WARNING
A potential pinch hazard exists for the following scenarios:
• Moving parts during servicing or normal operation
• Moving parts during mechanical check
• Mechanical parts during manual loading of bottles
• Mechanical parts during manual unloading of bottles
Negative bottles unload to the waste container. The waste container can be accessed by
opening the waste container door. A label appears on the waste container to remind the user
that biological risks may still be present in negative bottles.
1 Waste Container
2 Biological Risk
3 Waste Container Door
The instrument contains a Heater Fan that emits hot air. A High Temperature caution label
appears on the interior of the instrument on the Heater Fan Assembly to inform the user that
they should use caution when working around the Heater Fan.
1 High Temperature
The Power Supply Cover Plate covers electrical components. An Electric Shock Warning
label appears on the Power Supply Cover Plate to warn the user that the risk of electrical
shock exists if the Power Supply Cover Plate is not present. The user should not remove the
Power Supply Cover Plate or enter the interior of the instrument when the Power Supply
Cover Plate is not present.
WARNING
The power supply is housed beneath the power supply cover in the power
supply bay. Contact with components in the power supply bay can lead to
exposure to electrical current when power is applied to the instrument. When
performing any procedure where the power supply cover is removed, extra
precautions must be taken to ensure the power supply cover is installed
before returning the instrument to customer use. Failure to comply may
result in serious injury or death to the user.
WARNING
A potential hazard of electrical shock exists during service and maintenance,
electrical checks, or manual load mode.
The door latch cover contains a bottle orientation label to assist the user with the correct
positioning of bottles for the instrument
Place the bottles so that the bottle type information can be read. Do not place the bottles with
their caps down and do not place the bottles on their sides.
The Rack Cover Plate is located on the Rack Mount Assembly. A caution label is located on
the far right side of the rack cover plate to alert the user to consult documentation that
accompanies the instrument.
1 Caution Label
WARNING
Do not use compounds that will harm polycarbonate (for example
CIDEXPLUS® 28 Day solution) to clean the rack cover plate. Failure to comply
results in damage to the rack cover plate.
Safety Precautions
Pay particular attention to the following safety precautions. If these safety precautions are
ignored, injury to the operator or damage to the instrument may occur. Each individual
precaution is important.
WARNING
If the equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
WARNING
All biological samples and quality control (QC) products incubated in this
system, as well as all waste in the waste containers, should be treated as
potentially biohazardous materials. All materials and mechanical
components associated with the waste systems should be handled
according to safe microbiological practices in compliance with the
installation site's biohazard procedures. Use the personal protective
equipment recommended by the facility when handling any of these
components, including gloves, safety glasses, and a lab coat.
The system must be decontaminated before its covers are removed by
service personnel. Ensure that appropriate decontamination is carried out if
hazardous materials are spilled on or into the equipment or surrounding
areas.
WARNING
Treat waste material, including consumable items, and any components
coming into contact with waste material as having the potential hazards of
the samples used.
All service personnel should be familiar with the Material Safety Data Sheet
(MSDS) for all materials used in the procedures relating to this instrument,
and the correct procedures for handling these materials.
WARNING
Even when power is removed from the instrument, the potential exists for
electricity to be generated if components, such as assemblies that are
mounted on belts, are moved too quickly. Components should be moved
slowly to prevent the buildup of electricity. Failure to comply may result in
injury to personnel or damage to the instrument.
WARNING
Do not replace detachable main supply cords by inadequately rated cords.
Only use main supply cords provided by the manufacturer.
Do not replace electric cables. If cables do not have the same technical
specificities, there is a risk of electrical shock.
WARNING
Electronic equipment can be the source of electrical shocks. Installation,
service, and repair should only be performed by authorized and qualified
bioMérieux personnel.
WARNING
All power switches should be off when connecting or disconnecting cables
to power outlets to reduce the risk of electrical shock.
WARNING
bioMérieux recommends connecting this instrument to a main power outlet
that is protected with a ground fault circuit interrupter to reduce the risk of
electrical shock.
WARNING
As with any mechanical system, certain precautions must be taken when
operating the instrument. The instrument has a protective cover intended to
prevent the operator from coming into contact with any moving parts and
aerosols. When servicing the instrument, take special care, as there are
moving parts that can cause injury.
WARNING
Cleaning and disinfecting solutions have corrosive properties. Always wear
protective (chemical resistant) gloves and safety glasses when handling
cleaning and disinfecting solutions.
WARNING
Hot surfaces can cause injury.
CAUTION: The computer and its operating system have been carefully
configured for optimal system performance. Altering the configuration may
severely hamper the usability of the instrument.
Note: Before performing electrical safety or other compliance testing on the instrument, contact
bioMérieux or your local distributor.
WARNING
The user must only perform the maintenance operations described in this
document and rigorously follow each of the steps.
The use of tools not specified by bioMérieux is forbidden.
Powder-free gloves, a lab coat and protective glasses or goggles must be
worn when performing maintenance operations.
Always use personal protective equipment, including gloves, a lab coat and
safety glasses or goggles when handling reagents.
WARNING
Use caution when operating the instrument in a high EFT (Electrical Fast
Transient) environment. Alarms or alerts may occur.
Instrument Configuration
The BACT/ALERT® VIRTUO® instrument is modular and scalable, and can be upgraded to
accommodate increased capacity requirements. Each unit contains 16 racks with 27 cells
each for bottle monitoring. Four cells are dedicated for calibration standards. An instrument
has a total capacity of 432 cells.
System Configuration
A BACT/ALERT® VIRTUO® instrument bank may consist of a single A-unit or an A-unit and up
to three B-units attached with a common loading area and Large Display screen.
BACT/ALERT® VIRTUO®bank configurations include:
• Standalone Configuration
• Multi-Instance Site Configuration
• Primary/Secondary Site Configuration
WARNING
Failure to follow the instructions to configure instruments may lead to
incorrect results.
Standalone Configuration
Standalone configuration allows an instrument or bank to stand alone or be connected to a
user network (LIS, VILINK® remote support, and network printers).
Note: Users with Multi-Instance site configuration may need to refer to the MYLA® V4 User Manual
(or later) or the MAESTRIA™ V5 User Manual (or later).
Limitations of Use
Multi-Instance site configuration has the following limitations of use:
• Up to four BACT/ALERT® VIRTUO® A units can be connected in Multi-Instance site
configuration. Each A unit can be connected to three B units.
• Bottles cannot be reloaded across different BACT/ALERT® VIRTUO® banks in Multi-
Instance site configuration.
• Multi-Instance site configuration is not intended for use by blood banks.
• Multi-Instance site configuration is only for clinical users.
Bottle Management
WARNING
Moving bottles between banks may cause false negative results.
Bottles must not be moved between banks in Multi-Instance site configuration. When a bottle
is reloaded into a different bank, the bottle record does not transfer to that bank and analysis
is interrupted and incomplete. No warning appears on the Bottles in Progress screen when a
bottle is moved between banks.
A bottle must be reloaded into the same bank from which it was unloaded. To identify the
appropriate bank, go to the bioMérieux middleware dashboard, search by disposable ID, and
view the Patient Visit or Bottle Record screen.
Note: Users with Primary/Secondary site configuration may need to refer to the MYLA® V4 User
Manual (or later) or the MAESTRIA™ V5 User Manual (or later).
Limitations of Use
Primary/Secondary site configuration has the following limitations of use:
• Up to sixteen BACT/ALERT® VIRTUO® instruments (A units and B units) can be
connected in Primary/Secondary site configuration. Each A unit can be connected to three
B units.
Bottle Management
Bottles can be moved between banks in Primary/Secondary site configuration. When a bottle
is reloaded into a different bank, the bottle record transfers to that bank and analysis
continues. For example, when a bottle is unloaded from Bank 1 and reloaded into Bank 2, the
bottle record transfers to Bank 2. You can no longer view the bottle in Bank 1 and must view it
in Bank 2 instead. The bottle information on the bioMérieux middleware dashboard only
includes the bottle's current location.
You can access the audit trail in the bioMérieux middleware for the complete history of the
bottle record.
Communication Loss
In a Primary/Secondary site configuration, secondary A-units communicate with the primary
A-unit to receive information. When communication is lost, secondary A-units display the alert
in the figure below. This alert indicates that information displayed on the secondary A-units is
either unavailable or outdated.
You can identify whether an instrument bank is primary or secondary on the System
Information screen of the bank's A unit.
On the A unit's Home screen, select Application and Instrument Information.
The System Information screen appears. On the System Information screen, A(P) or A
may appear in the Type column for the A unit.
• A(P) indicates that the A unit is in a primary bank.
• A indicates that the A unit is in a secondary bank.
Data Management
bioMérieux data management is a separate software system that manages the data
generated by BACT/ALERT® VIRTUO®. Both predefined and user-customized queries and
reports are available. This data management software provides a high level of flexibility to
store data for a definable length of time, edit, query, sort, and produce reports.
Reagents
To get ordering information, contact bioMérieux or your local distributor (contact information
available on www.biomerieux.com).
Note: Refer to the BACT/ALERT® bottle Instructions for Use before using the bottles.
• Clinical Bottles
◦ BACT/ALERT® SA
◦ BACT/ALERT® SN
◦ BACT/ALERT® FA PLUS
◦ BACT/ALERT® FN PLUS
◦ BACT/ALERT® PF PLUS
• Industry Blood Bank Bottles
◦ BACT/ALERT® BPA
◦ BACT/ALERT® BPN
◦ BACT/ALERT® BPA EU
◦ BACT/ALERT® BPN EU
◦ BACT/ALERT® BPA US
◦ BACT/ALERT® BPN US
Note: The clinical bottles listed above support fill volume except BACT/ALERT® PF PLUS.
BACT/ALERT® PF PLUS does not support fill volume because it is a low specimen volume
clinical bottle. See the Instructions for Use.
WARNING
A potential hazard (false negative test results) exists if non-bioMérieux
and/or unsupported and/or substandard bottles are used with the instrument.
Use only bottles manufactured by bioMérieux.
WARNING
A biohazard risk may occur if glass culture bottles are loaded in the
instrument. Never load glass bottles in the instrument.
WARNING
Use only plastic bottles with the instrument. A potential hazard (false
negative test results) exists if plastic bottles are not used with the
instrument.
Additional Supplies
BACT/ALERT® VIRTUO® has an incubation temperature port at the top of the instrument. To
perform independent temperature monitoring, use a probe up to 8 mm wide and install in the
probe holder located at the top of the instrument. The probe reading should match the
temperature on the instrument screen within +/- 2 °C.
The BACT/ALERT® VIRTUO® also supports temperature mapping by an independent
mapping tool. Contact your local bioMérieux representative for assistance.
WARNING
Do not step on an unstable device or platform to place an independent
temperature probe in its correct external location to perform an independent
instrument calibration check as the consequence may be a physical injury or
crush hazard.
WARNING
The consequences of the failure to use an independent temperature
monitoring probe that meets specifications may be false negative results.
WARNING
A bioMérieux representative must perform independent temperature
checking. The consequences of changing the offset may be false negative
results.
WARNING
The consequences of using (and leaving) an independent temperature
mapping probe inside the instrument may be delayed results.
Theory of Operation
Principle of Detection
According to the Principle of Detection, if microorganisms are present in the test sample,
carbon dioxide (CO2) is produced as the microorganisms metabolize the substrates in the
culture medium. When growth of the microorganisms produces CO2, the color of the sensor in
the bottom of each culture bottle changes from dark to light.
A light-emitting diode (LED) projects light onto the sensor. A photodetector measures the light
reflected. As more CO2 is generated, more light is reflected. This information is compared to
the initial sensor reading. A high rate of CO2 production and/or a sustained production of CO2
indicate the sample is positive.
If the CO2 level does not change significantly after a specified number of days at optimal
conditions, the sample is negative.
WARNING
A potential hazard of false negative test results exists if:
• The system is used for release testing based on preliminary or user-
modified results.
• The system is used off label.
• The operator fails to follow intended use guidelines.
Each instrument contains 16 racks with 27 bottle cells each, for a total of 432 cells.
WARNING
A potential of user injury exists if accessories are used that have sharp
edges or are otherwise not UL compliant.
1 Warning Light Tower (Optional) Indicates a visual status for the instrument from any angle in
the lab.
4 Barcode Scanner Scans data from a barcode into the software, and displays
information for a bottle.
5 Loading Area Automatically loads from one to 40 bottles into the instrument
at a time.
6 Waste Container Holds all negative bottles that are unloaded. Periodically empty
the waste container based on instructions on the Large
Display.
7 RJ45 Connectors (2) Connects the PC and the MCB with an external router or
switch.
11 Bottle Indexer and Transfer Arm with Transfers the bottles to the inside of the instrument where the
Cover bottle label is scanned and the bottle is positioned for loading
or, if configured, transfers to the next connected instrument.
12 Bottle Return Chutes (2) and Bottle Unloads positive bottles to the Bottle Retrieval Area, and also
Retrieval Area unloads selected bottles at user request.
14 Audible Alarm Speaker Flags instrument failures, operator errors, and positive bottle
results.
15 System Status Indicator (SSI) Indicates a visual status for the instrument. Illuminates in
yellow, green, or red depending on the level of alarm or alert
(for example, yellow or red can indicate a positive bottle).
2 Rack Detection Assembly (RDA) Holds four racks and controls agitation (rocking).
4 Wheels (4) and Stabilizing Feet (2) Wheels move the instrument. Stabilizing feet adjust to prevent
movement.
6 Rack Contains 27 cells. The racks are designated with a letter from A
to P.
7 Cell Each cell is designated with a number from 1 to 27. Each cell
holds and monitors one culture bottle.
1 Indexer
2 Transfer Arm
3 Gripper Assembly
Note: The Safe to Move sensor turns green when it is in the parked position.
B Unit
Each A unit can be connected to up to 3 B units to extend the capacity of a bank. B units have
the same specifications, only without a PC and large display screen.
2 Waste Container Holds all negative bottles that are unloaded. Periodically empty
the waste container based on instructions on the Large
Display.
3 RJ45 Connectors (2) Connects the PC and the MCB with an external router or
switch.
5 Bottle Indexer and Transfer Arm with Transfers the bottles to the inside of the instrument where the
Cover bottle label is scanned and the bottle is positioned for loading
or, if configured, transfers to the next connected instrument.
6 Bottle Return Chutes (2) and Bottle Unloads positive bottles to the Bottle Retrieval Area, and also
Retrieval Area unloads selected bottles at user request.
8 Audible Alarm Speaker Flags instrument failures, operator errors, and positive bottle
results.
9 System Status Indicator (SSI) Indicates a visual status for the instrument. Illuminates in
yellow, green, or red depending on the level of alarm or alert
(for example, yellow or red can indicate a positive bottle).
Note: If an A unit indexer is unable to load bottles, the cover can be removed from the B unit for
bottle loading.
1 A Unit
2 B Units
WARNING
A potential hazard (false negative test results) exists if the instrument is
operated in an environment that is out of specification, or if proper electrical
power is not maintained.
WARNING
A potential for property damage exists if the instrument is not stored in the
recommended environmental conditions.
Physical Characteristics
Electrical Power 100/120 VAC, 12A (50/60 Hz)
Services Requirements
220/240 VAC, 6A (50/60 Hz)
Power Consumed in A Unit: @120 VAC (1440 Watts typical); B Unit: @120 VAC
Watts (1320Watts typical)
A Unit: @220 Volts (1320 Watts typical); B Unit: @220 Volts
(1200 Watts typical)
Heat Dissipated A Unit: 1287 BTU/h at instrument set point of 37 °C and ambient
temperature of 20 °C; B Unit: 1129 BTU/h at instrument set
point of 37 °C and ambient temperature of 20 °C
Sound Emission 60 dBA (normal operation)
67 dBA (loading bottles)
Dimensions (Instrument Height: 196 cm (77.2 in.)*
Door Closed)
Width: 73 cm (28.7 in.)
Depth: 91 cm (35.8 in.)
Dimensions (Instrument Height: 196 cm (77.2 in.)*
Door Open)
Width: 94 cm (37 in.)
Depth: 165 cm (65 in.)
Mass (Empty) A Unit: 292 kg (643 lb); B Unit: 284 kg (625 lb)
Mass (Loaded with A Unit:321 kg (707 lb); B Unit: 313 kg (689 lb)
Bottles)
Clearance • Instrument can be placed with its back against the wall.
• Minimum unobstructed clearance of 25 cm (10 in.) must be
provided on the left side of the left most instrument.
• Minimum of 30 cm (12 in.) clearance above the instrument
Environmental Characteristics
Installation Location Indoor usage only. Do not install under or near a HVAC register
or in direct sunlight or bright overhead lights. Install on a level
surface.
Ambient Room 15 °C to 28 °C (59 °F to 82 °F), ambient temperature must be at
Temperature least 9 °C below the incubator set point as measured 10-15 cm
(4-6 in) above the instrument door.
Shipping/Storage −17 °C to 57 °C (0 °F to 135 °F)
Temperature
Operating Humidity 20% to 80% relative humidity, non-condensing
Shipping/Storage 20% to 85% relative humidity, non-condensing
Humidity
Maximum Operating and ≤ 2000 meters (6562 feet)
Storage Altitude
Pollution Degree Degree 2 in accordance with IEC 664
Overvoltage Category II per IEC 664
Laser Class 1 laser product. Complies with 21 CFR 1040.10 and
1040.11 except for deviations pursuant to Laser Notice No. 50,
dated June 24, 2007.
Note: Some text in the software's user interface is not translated and appears in English.
Large Display
The table names and describes the screens that are available on the Large Display.
Note: The BACT/ALERT® VIRTUO® PC performs a reboot the first day of each month at 1:00 AM.
During the reboot, the Large Display is unavailable. The reboot does not interrupt bottle
processing.
Note: If you experience slow PC performance, reboot the PC. See Rebooting the Internal PC
(Large Display). If the slowness persists after reboot, contact your local bioMérieux
representative for assistance.
Small Display
The Small Display is located above the bottle return chutes. Use this display to access
specific instrument status information and maintenance functions.
The Small Display screen presents system information and icons as listed below.
Open Door Request Sends a request for the robot to finish its activity so
that the door can be safely opened.
Door OK to Open Displays when robot has finished activity and door is
OK to open.
Icons
Icons are graphic symbols with text to convey information and concepts.
Buttons
Buttons are rectangular shapes. Select a button to make a choice or activate a function.
The icon on the button face indicates the function of a button.
A gray button is inactive, or disabled, and its associated function is not available.
1 Enabled button
2 Disabled button
IMPORTANT: The external barcode scanner used to scan data from a barcode into the software and
display information for a bottle does not beep or provide any sound notification that
the barcode label has been scanned. Monitor the display to ensure that the barcode
has been scanned and is not scanned more than once.
To enter text with the on-screen keyboard, select the field. A cursor appears in the field, and
the on-screen keyboard appears at the bottom of the screen.
Edit Mode
To edit information on a screen, select Edit.
Group MTT Edit Edit the max test time for multiple bottles at one
time.
CAUTION: Once the instrument is moved to its final location, the wheels
should always be turned parallel to the front of the instrument to avoid a
tipover/crush hazard.
At a minimum, place the instrument in a contained environment with controlled access. The
locations should have surfaces that can be easily decontaminated using an appropriate topical
disinfectant. Do not place the instrument in an open corridor or hallway that is accessible to
the general public.
WARNING
A potential tipover/crush hazard exists in the following situations:
• During transit (if the instrument is not secured properly)
• During installation (if the instrument is not secured properly)
• If the instrument is placed on an inadequately designed support surface
(for example, not level)
• From the robotic arm (when the cabinet door is opened)
• During (general) service and maintenance
WARNING
A potential pinch hazard exists for the following scenarios:
• Moving parts during servicing or normal operation
• Moving parts during mechanical check
• Mechanical parts during manual loading of bottles
• Mechanical parts during manual unloading of bottles
Note: Only a user with bioMérieux administrator access for the Internal PC can change the date
and time.
Windows Environment
WARNING
If you install your own anti-virus software, follow your own laboratory
information systems policy. The consequences of not following the correct
procedure may be delayed results.
WARNING
Changes to the Windows® environment must be validated before proceeding
and may include consequences of delayed results.
1 Close button
Introduction
The basic functions are performed during your daily workflow by using the Large Display.
These functions include:
• Monitoring the system (viewing alarms and alerts, bottle capacity, temperature)
• Managing bottle information
• Configuring the system
• Loading and unloading bottles
• Resolving anonymous bottles
• Searching for specific bottle information
• Viewing, editing, and printing bottle information
Home Screen
The Home screen displays the system status for all instruments connected within the bank,
including the number of cells available, number of positive and negative bottles, and active
alarms or alerts. The Home screen also displays the buttons that navigate to other function
screens, and a button to silence alarms and alerts for a defined period of time. The Home
screen is divided into five zones, and can be configured.
1 Zone 1
2 Zone 2
3 Zone 3
4 Zone 4
5 Zone 5
Zone 1
Zone 1 is at the top of the screen and includes the bioMérieux corporate logo and navigation
buttons.
Zone 2
Zone 2 is located just under Zone 1 on the Home screen. Zone 2 displays all active alarms
and alerts that have not been resolved. Zone 2 also indicates when bottles are loading and
unloading.
Alarms and alerts are divided into five categories. A specific icon represents each category of
alarm or alert.
To view an alarm or an alert, select the icon.
Note: The number of occurrences of each type of alarm and alert appears below the icon.
IMPORTANT: Alarms and alerts remain on the screen until they are resolved.
In addition, severity symbols may appear with the alarm or alert icon.
The Snooze button quiets an alarm or alert to allow extra time for resolution. The Snooze
button is in Zone 2 and appears on every screen.
Note: The Snooze button cannot be used to quiet an alarm or alert while the instrument is
experiencing a communication loss.
The length of time the snooze feature is enabled is configurable. See Configuring Snooze
Time Period.
Zone 3
Zone 3 is in the center of the Home screen. Zone 3 displays the number of cells available and
the total number of positive and negative bottles currently loaded.
WARNING
Remove positive bottles in a
timely manner. The
consequences of delays may
be delayed results.
IMPORTANT: The number of negative and positive bottles appears on every screen.
Zone 4
Zone 4 is located at the bottom of the Home screen. Zone 4 displays the user name, the bank
name, the lab name, and the software release version.
Zone 5
Use the navigation buttons on the left side of the Home screen to access additional function
screens. See Large Display.
System Log In
A user account is required to log in to the system. Establish user accounts through the
bioMérieux User Management System. Select User Management in Zone 1 to access the
User Management Application. See System Configuration for more information.
WARNING
A potential hazard (incorrect result) exists if passwords are not strong
enough to prevent an unauthorized user from gaining access to the system.
See Password Policy.
Inactivity Timeout
If no operator activity has been recorded within a user-configured period of time, the
instrument automatically logs you off. To perform functions that require a user login, enter a
user name and password again.
Settings on the General Software Configuration screen determine the number of barcode
labels on the bottle and may include an alternate bottle ID, an accession barcode, or a patient
barcode, for a maximum of two user applied barcodes.
The BACT/ALERT® VIRTUO® supports the following user applied 1D barcode symbologies:
• Code 128
• Code 39
• Codabar
• Interleaved 2of5 with check-digit
• UPC
• Code 93
The following specifications are required for user applied 1D barcodes listed above:
• Grade B or better
• Minimum Element Size: 6.7 mil (0.17 mm)
• Minimum Barcode Height: 0.25 in (6.35 mm)
• Minimum Digits (Interleaved 2of5 ONLY): 6
Note: Since the PDF417 barcode is rectangular, the size dimensions refer to the smaller of the
two barcode dimensions.
WARNING
The system’s internal barcode scanner requires a check-digit for Interleaved
2of5 (I2of5) by default. bioMérieux has identified a greater risk in using
barcodes without a form of checksum, and strongly encourages using
barcodes with checksums before enabling I2of5 without a check-digit. Inform
the bioMérieux Field Service Engineer upon installation to enable I2of5
without check-digit ONLY if required. Allowing this symbology without
check-digit without need puts other barcodes at risk of incorrect
interpretation and may result in a delay of patient results.
Properly place the barcode label on the blood culture bottles so the instrument can scan the
label accurately.
WARNING
If the proper barcode format is not used, the consequences may be false
negative results.
WARNING
The potential for incorrect results exists if a barcode is not properly applied
to a bottle.
WARNING
If barcode information is not applied to the correct bottle, or if proper
barcode format is not used, the consequences may be false negative results.
If labels are not positioned correctly on the blood culture bottles, the barcode scanner may not
be able to read the barcode. Accession barcodes not scanned trigger the Missing Accession
ID alert. Manually enter the missing accession ID after the bottle is loaded.
2 Bottle ID barcode
4 2D (Aztec) code
5 Fill-to mark
WARNING
To ensure that bottle and accession barcodes are properly scanned by the
BACT/ALERT® VIRTUO® and that other important bottle information is not
obscured:
• Do not place barcode labels horizontally or at an angle. Instead, place
barcode labels vertically on the bottle.
• Do not cover the entire bottle ID barcode.
• Do not cover the bottle lot number and expiration date.
• Do not heavily overlabel the bottle.
• Do not cover the 2D barcode.
• Do not cover the sample fill indicator.
• Do not cover the open area between the edges of the label.
• Do not cover the bottle sensor on the bottom of the bottle.
• Do not write on the bottom of the bottle.
• Ensure that the exterior of the bottles are clean and do not contain any
sticky substances or residues.
• Do not damage or obscure bottle labels.
WARNING
Failure to maintain cleanliness of bottles, or to avoid sticky substances or
residue on external surface of bottles, or to avoid exposed adhesive from
backing of applied labels, may lead to no results.
WARNING
If the bottle sensor at the bottom of the bottle is obscured during a manual or
auto load, the consequences may be false negative results.
WARNING
Damaged or obscured bottle labels may cause delayed results.
WARNING
Heavily overlabeled bottles may delay results. Heavy labeling impedes
loading the bottle into the cell and unloading the bottle from the cell. A rack
assembly retains bottles in a cell when the bottle diameter is between 34.34
and 35.56 mm.
WARNING
If large numbers of bottles are loaded into BACT/ALERT® VIRTUO®, there
may be a very large heat transfer within the racks. This may trigger the
detection algorithm to erroneously flag positive. To avoid triggering these
conditions, follow these recommendations:
1. Avoid loading over 70 bottles within one hour.
2. Limit the number of bottles loaded in a single batch to no more than 40
bottles.
3. Allow the batch to completely load before adding more bottles to the
conveyor.
Batches of 20 bottles is optimal. If the bank contains multiple instruments,
the probability of false positives due to heat transfer is reduced.
This warning applies to bottles loaded on a single A unit. If a bank contains 1
or more B units, the risk of false positives due to bulk loading is reduced
since bottles are loaded across all units.
The instrument automatically loads up to 40 bottles placed in the loading area at one time. Do
not overload the conveyor.
CAUTION: When the bottle loading area is filled completely , the indexer
cannot reverse bottles (in special circumstances such as instrument full, for
example) to properly re-initialize. An indexer jam alarm results.
WARNING
A biohazard risk may occur if glass culture bottles are loaded in the
instrument. Never load glass bottles in the instrument.
WARNING
Always wear protective gloves when loading bottles into the instrument to
prevent transmitting infectious agents from a bottle leak.
WARNING
Bottles must be at room temperature before use. An erroneous test result
(such as false positive or false negative) may occur if bottles are not at room
temperature when loaded.
CAUTION: Only bottles provided by bioMérieux are authorized for use with
BACT/ALERT® VIRTUO®. bioMérieux shall not be liable as to any defect
(including, but not limited to, incorrect results that may affect patient
outcomes) arising from the use of unauthorized bottles. Users who use
bottles not authorized by bioMérieux do so at their own risk.
WARNING
Do not remove the indexer cover from any instrument when bottles are
loading or transferring between instruments. Removing an indexer cover
may cause instrument jams.
WARNING
A potential hazard (false negative result) exists if the bottle sensor at
the bottom of the bottle is obscured during loading.
WARNING
A potential hazard (false negative result) exists if there is a delay in
loading a bottle into the instrument and growth has already occurred.
WARNING
The potential consequence of a delayed result exists if a bottle is
loaded upside down on the conveyor.
3. As the bottle moves through the indexer, the imager scans the bottle label and
decodes bottle information from the 1D and 2D barcodes. The instrument also reads
up to two user applied barcodes .
Note: If necessary, you can improve a poor quality bottle image by unloading and reloading
the bottle. The bottle image is updated upon reload.
4. The indexer will move the bottle to the load position or transfer the bottle to the next
instrument for loading.
5. The robot arm (gripper) retrieves the bottle from the bottle indexer and loads the bottle
into an available cell.
The available capacity of the bank is updated on Available Cells.
6. Before leaving the instrument, verify that the instrument is operating properly and that
the bottles remain upright and are being properly loaded into the bottle indexer.
WARNING
If correct loading is not verified, bottles may not be loaded into the
instrument, resulting in delayed patient results.
If the instrument identifies the bottle as a new bottle, a bottle record is created. View
the scanned bottle image and bottle information in the bottle record.
IMPORTANT: If a bottle cannot be scanned and identified, the bottle is loaded in the
instrument as an anonymous bottle and an alert appears. A bottle record is
created.
WARNING
Subculture any bottle that is reloaded anonymously. The
consequences of failing to subculture may be incorrect results.
WARNING
Never reload bottles anonymously. The consequences of not following
these instructions may be false negative results.
IMPORTANT: If an alarm or alert appears, select the applicable alarm or alert icon in Zone 2
and follow the instructions on the screen to resolve the issue. If the problem
persists, contact your local bioMérieux Representative.
The instrument automatically unloads all identified positive and negative bottles from any
instrument in a bank when the instrument is configured to automatically unload bottles after
testing and a final test result is available.
Anonymous bottles must be identified before they can be automatically unloaded.
WARNING
Process unloaded bottles in a timely manner. The consequences of delays
may be no results.
WARNING
Remove positive bottles in a timely manner. The consequences of delays
may be delayed results.
WARNING
Be sure to account for any bottles that are removed from the system. The
consequences of not accounting for bottles removed from the system may
be missed or delayed results.
Fully remove a bottle from the return chute. If a bottle is only partially removed, the bottle may
appear to the instrument to be completely removed. The result may be that multiple bottles
unload into that chute. If three bottles stack up, and the bottom bottle is not touching the
sensor, the return chute jam alarm sounds.
CAUTION: Never force a bottle back into an empty return chute after it has
been previously or partially unloaded. The consequences may be delayed
results.
WARNING
Always open the cabinet and check for dropped bottles inside the instrument
when a bottle does not reach its intended destination. The consequences of
a missed bottle may be delayed or no results.
Automatic unloading stops and an alarm appears in the following situations. Resolve the
alarm to continue automatic unloading.
• Waste container is full.
• Waste container is missing.
• Bottle retrieval area is full.
• Robot jam or other instrument malfunction occurs.
Note: Up to three bottles can unload to each chute in the retrieval area at one time. No additional
positive bottles can unload from the cells until the other bottles are removed from the
retrieval area.
4. Select OK.
WARNING
Canceling a bottle under test may lead to no result.
The test is canceled, and the bottle unloads to the bottle retrieval area.
Note: Bottles that have been canceled using Delete cannot be re-loaded on to the
instrument.
Note: If a reloaded positive bottle signals positive a second time, that bottle cannot be
reloaded.
WARNING
A potential hazard (false negative test result) exists if a bottle is
reloaded into a different bank at the site. Bottles can only be moved
between banks in Primary/Secondary site configuration.
WARNING
A potential hazard (incorrect test result) exists if the bottle is not
reloaded in a timely manner.
When the instrument is configured to unload bottles by user request, the instrument unloads
positive and negative bottles in two ways:
• Unload all positive or negative bottles that have a final test result.
• Unload specific bottles through a user-selected search.
WARNING
A potential hazard (false negative test result) exists if a bottle is removed
from the instrument before testing is complete.
WARNING
Failure to subculture any negative to date bottle that is out of the instrument
longer than the missed readings period may lead to a hazard of incorrect
results.
WARNING
If a negative bottle completes the protocol, the bottle is not unloaded,
and analysis continues, the bottle result could change from negative
to positive. If a report is printed before unloading negative bottles, the
report may contain incorrect results.
CAUTION: For optimal results, bottles should remain testing within the same
instrument or bank where they were originally loaded until the test is
complete. Bottles with a negative-to-date result should not be unloaded or
relocated to a different instrument. This action interrupts bottle readings and
may cause a delay in positive determination or an inaccurate result. Only if
necessary, follow the instructions below to unload or relocate a bottle with a
negative-to-date result.
Note: Unload is also available from the Bottle Status Information screen and the Edit Bottle Record
screen.
Note: If you select Cancel Unload as a bottle is being removed from its cell, the instrument
unloads the bottle, and then cancels the remaining bottles to be unloaded.
Note: Bottle data may not associate to the accession ID or to the patient if the data in the LIS
message is invalid (such as an invalid birth date or maximum testing time). Correct the data
in the LIS and resend the message.
time period configured in the Data Retention/Accession Reuse setting. For more
information about this setting, see Configuring Data Retention and Accession ID Reuse.
1 Search button
2 Search criteria
Searching by Patient ID
Note: Patient data can be deleted from the system. All bottles associated to the patient must be
canceled first, see Canceling a Bottle Test, then contact your local bioMérieux representative
to delete the patient.
If more than one patient ID is found, the search results are displayed in a table.
2. From the search results, select your desired patient ID.
Note: Select Cancel to close the Add Search Criteria screen and return to the Search
screen.
3. Select Search.
4. Select the patient name from the list.
5. Select OK.
The Search screen shows the patient name.
6. Select Search.
See Accessing the Bottle Status Information Screen.
Searching by an Accession ID
From the Search screen:
1. Select Accession ID.
2. Enter the accession ID.
3. Click the Search for Bottle Information button.
Searching by a Bottle ID
From the Search screen:
1. Select Bottle ID.
2. Enter the bottle ID.
3. Select Search for Bottle Information.
Note: To search for all bottles that have a bottle ID with the same first two characters (for
example, SA), enter the first two characters using the on-screen keyboard. In this
example, all bottles with a bottle ID that starts with SA would be listed.
Figure 51: Add Search Criteria - Criteria Values for Bottle Type
2 OK button
3 Cancel button
3. Select OK.
The selected bottle types appear in the Bottle Type field.
4. Select Search for Bottle Information.
WARNING
The potential for delayed result exists if missing accession IDs
(orphan) or anonymous bottles are not properly handled and
addressed.
Note: Depending on your configuration, some of the search criteria may not be available.
Note: The text field is not case sensitive. For example, searching by blood or Blood will
return the same results.
2. Optional: Select the Negation checkbox if searching for a Sample Type other than
what is entered in the Sample Type field.
Deselect the Negation checkbox if searching for the Sample Type that is entered in
the Sample Type field.
Search Examples
Scenario 1: Search for bottles that have a Sample Type that is blood.
Enter blood in the Sample Type field. Click Search for Bottle Information.
Results are displayed that have a Sample Type of blood.
Scenario 2: Search for bottles that have a Sample Type that is not blood.
Enter blood in the Sample Type field. Select the Negation Checkbox to
search for bottles that have a Sample Type that is not blood. Click Search for
Bottle Information. Results are displayed that do not have a Sample Type of
blood.
1 Instrument name
Option Description
Underfilled The sample volume falls below the configured minimum.
Unknown The bottle was loaded before the instrument was updated to
software release 3.1.
Unable to Read The sample volume could not be read when the bottle was loaded
because there is no fill-to line on the bottle, the fill-to line is
obstructed, or the bottle is anonymous.
Pending The sample volume could not be read when the bottle was loaded -
for example, because there was foam in the bottle and the bottle
was scheduled for rescan within 2 hours of loading.
Not Supported The bottle type does not support a fill-to line.
Disabled Sample volume measurement is disabled.
Label Issue The sample volume could not be read when the bottle was loaded
because a label is obstructing the area around the fill-to line. For
correct label placement, see Applying a Barcode Label to a Bottle
Note: To narrow this search to one instrument in multi-instrument configurations, select the
instrument of interest in addition to rack and/or cell.
Note: Use the Page Scroll buttons at the bottom of the screen to access additional search results.
Note: If more than one bottle was selected, use the Page Scroll buttons at the bottom of the
screen to access the bottle status information for the additional bottles.
The table describes the buttons and symbols on the Bottle Status Information screen
and their functions.
3. (Site configuration only) If necessary, click the Upload button on the Bottle Details
screen to view patient demographic information.
Note: In site configuration, there may be a delay in the transfer of patient demographic
information from the bioMérieux middleware to BACT/ALERT® VIRTUO® while bottles
are processing. When the bottle results are final, patient demographic information
downloads to BACT/ALERT® VIRTUO®.
Note: Any edits made to the bottle record will override any information supplied by the LIS.
WARNING
Assigning bottle record information by using multiple methods may lead to
incorrect results.
WARNING
Always save the desired changes when editing a bottle record. Failure to
save the changes may cause incorrect results.
WARNING
If a physical keyboard is installed, be sure it matches the language of the
system installed. Failure to install a keyboard in the configured language
may cause delayed results.
Note: The Bottle ID, Bottle Type, Lot Number, and Expiration Date cannot be edited if the
information was read from a scanned bioMérieux bottle code when loaded.
WARNING
If you must use alternate bottle IDs during Edit, the IDs must be unique for
the duration of your data retention period. Duplicating alternate bottle IDs
within this time period can result in incorrect results.
Note: The ability to edit patient information is only allowed if the instrument is configured as
Standalone without an LIS connection.
WARNING
The consequences of inadvertently clearing an accession ID may be
delayed results.
2. If the accession ID is not associated with patient information, enter a patient ID and
click Look Up to display the associated patient name.
Figure 64: Edit Bottle Record - Accession ID Not Associated with Patient
3. If the patient ID is not associated with a patient name, enter the patient's last name
and first name if desired.
4. Click Save.
Note: The scanner does not provide an audible indication when a barcode has been
scanned.
WARNING
Failure to properly follow instructions for reducing maximum test time
may cause false negative results.
WARNING
Failure to properly follow instructions for correctly increasing
maximum test time may cause delayed or negative results.
Note: See the appropriate bottle Instructions for Use for specific maximum test time
recommendations.
WARNING
Failure to properly follow instructions for changing bottle test results
may lead to incorrect results.
WARNING
A potential hazard (incorrect test result) exists if the wrong test status
is assigned.
WARNING
bioMérieux shall not be liable as to any defect (including, but not limited to,
incorrect results that may affect patient outcomes or cause a loss of product)
arising from the use of expired bottles, or the communication of the results
obtained using expired bottles. Users who use expired bottles do so at their
own risk.
Note: Only a user with LabSuper access is authorized to process expired bottles. See Managing
User Accounts and User Groups.
When the bottle expired alarms sounds, the following message appears:
Select one.
• Select Continue Test to release results on all expired listed in the Expired Bottle
alarm. The bottle results indicate that the bottles are expired.
• Select Cancel Test to cancel processing and unload all bottles that are listed in the
alarm as expired.
Note: The alarm message remains on the screen and the Large Display cannot be used
until the alarm is addressed. All currently loaded bottles are not affected and continue
to process, but other functions are not accessible.
Once the alarm is addressed, the message disappears and the Large Display is
accessible.
• Select Cancel to cancel the reload. The bottle status will remain positive and the
bottle will be unloaded.
Note: A user name and password are required to resolve the reload positive message.
If a bottle is loaded into the instrument without being identified (because of a hardware
malfunction or damage to the bottle label), the bottle is called an anonymous bottle because it
is not associated with a bottle ID or bottle type. Identify an anonymous bottle to ensure that
the bottle is correctly associated with the patient.
WARNING
Address anonymous positive bottles immediately to ensure correct results
are reported. Failure to address anonymous positive bottles immediately
may lead to delayed results.
WARNING
Be sure to identify anonymous bottles before unloading from the instrument.
Note: Another option is to use the Search function to locate anonymous bottles. See Using
Additional Search Criteria.
2. Enter your user name and password, if not already logged in.
3. Select OK.
Anonymous bottles appear one record at a time for editing. Navigate to the next
record, if necessary.
WARNING
Failure to unload anonymous bottles one at a time may lead to
incorrect results.
WARNING
Associating a bottle with the incorrect accession ID may lead to
incorrect results.
WARNING
A potential hazard (false negative test result) exists if the algorithm
does not have sufficient information to calculate the correct bottle test
result.
Note: If an anonymous bottle is reloaded (for example, an unconfirmed positive), the bottle
remains anonymous. When identified, the two data records are not combined. Do not
unload the bottle at this time. The bottle continues to process.
Note: The bottle result may change after saving the bottle type.
Note: Another option is to use the Search function to locate anonymous bottles. See Using
Additional Search Criteria.
2. Enter your user name and password, if not already logged in.
3. Select OK.
Anonymous bottles appear one record at a time for editing. Navigate to the next
record, if necessary.
WARNING
Failure to unload anonymous bottles one at a time may lead to
incorrect results.
6. Select Save.
7. Enter a bottle type, if it is visible on the image.
8. Select Save.
9. Select Unload to unload and relabel the bottle. To continue unloading and relabeling
this bottle, see Unloading, Relabeling, and Reloading a Previously Identified
Anonymous Bottle.
Note: If the bottle type is not visible on the image, enter the bottle type once the bottle is
unloaded and the bottle type is determined.
Note: The bottle result may change after saving the bottle type.
WARNING
Incorrectly associating a bottle record may lead to incorrect results.
When an anonymous bottle has been identified as described above, see Unloading,
Relabeling, and Reloading a Previously Identified Anonymous Bottle to understand the
sequence of events to unload this bottle.
It is not necessary to unload and relabel a bottle if the bottle ID and bottle type could be
determined from the image and were entered into the bottle record. If a bottle has no image or
has an obstructed barcode, remove the obstruction or relabel the bottle.
1. Select Unload.
2. Select OK.
The bottle returns to the bottle retrieval area.
Note: The bottle may also automatically return to the bottle retrieval area if it is determined
to be positive.
4. Ensure that the bottle ID barcode is present and correct any obstructions, if possible.
5. If the barcode obstruction cannot be corrected, affix a scannable barcode label to the
bottle.
6. If the bottle ID in the bottle record does not match the bottle ID in the affixed barcode,
select Edit.
WARNING
The bottle ID on the label must match the bottle ID entered in the
record.
WARNING
Failure to affix the corresponding barcode label or to remove any
obstruction of the barcode label before reload will cause the reloaded
bottle to be anonymous.
WARNING
Bottle IDs must be unique, and must be unique for the duration of the
data retention period on their BACT/ALERT® VIRTUO® system. Do not
reuse a bottle ID within the data retention period as duplicate bottle
IDs can lead to false negative results.
7. Enter the bottle ID from the affixed barcode in the bottle ID field.
8. Select Save.
9. Reload the bottle by placing it on the conveyor.
10. Select Dismiss to close the message and resolve the alert.
11. Wait for the Bottle(s) Successfully Reload confirmation message.
12. Select Dismiss to close the message and resolve the alert.
An alternate bottle ID is a barcode for bottle ID replacement and is usually created by a LIS.
WARNING
If you use alternate bottle IDs, the IDs must be unique for the duration of
your data retention period. Duplicating alternate bottle IDs within this time
period can result in delayed or inaccurate results.
Note: A bottle label image is available for four weeks after the associated bottle has been unloaded
from the instrument.
Note: A bottle readings graph is available for four weeks after the associated bottle is unloaded
from the instrument.
1 Positive curve
2 Print button
3 Zoom in button
4 Close button
5 Time (In days)
6 Bottle reflectance readings
2. Select Zoom In to display an expanded view of the readings.
Note: Zooming into a graph provides an expanded view of bottle readings in a given area of
the graph. When zooming in, always refer back to the full graph.
Zooming in automatically adjusts the X and Y axis to fit the available data.
1 Positive curve
2 Print button
3 Zoom out button
4 Close button
5 Time (In days)
6 Bottle reflectance readings
3. Optional: Select Print.
Bottle Reports
The following bottle reports are available.
• Positive Bottles
• Negative Bottles
• Negative to Date Bottles
• Orphan Bottles
• Anonymous Bottles
• Loaded Bottles
IMPORTANT: The following procedure for generating bottle reports is for instruments in standalone
and site configuration (MAESTRIA™) only.
Note: Standalone Bottle Reports can be configured in the System Configuration Report tab.
1 Calendar button
2 Refresh button
3 Export button
4 Print button
5 Close button
IMPORTANT: The following procedure for generating bottle reports is for instruments in site
configuration (MYLA®) only.
Note: There may be a slight delay as the instrument communicates with bioMérieux middleware
when accessing the reports for the first time.
Note: The Critical Information icon may disappear from the Bottle Reports screen when a bottle is
unconfirmed positive. You can locate this information on both the Bottle Details screen and in
the PDF report.
1 Calendar buttons
2 Sorting arrows
3 Critical Information icon
4 Search button
5 Refresh button
6 Print button
7 Expand and collapse button
1 Trash Button
The Bottles in Progress Report lists bottles that are currently loaded with a Negative to
Date status.
The Trash Button is available on some reports. To remove a line from the report, select
the line, then select Trash.
Instrument Reports
Instrument reports contain incubator temperature and calibration information based on a date
range.
Three instrument reports are available.
• The Hourly Temperature Detail Report provides information about hourly temperature
readings in the instrument including trends in the temperature readings.
• The Daily Average Temperature Report provides information about daily temperature
readings in the instrument.
• The Instrument Calibration Report provides information about calibration events over a
period of time.
3 Selected day
4 Print
5 Export
6 Clear
7 Close
Field Description
Date Date the report was selected.
Time of Day Time of day by hours.
Average °C Average temperature during the previous
hour.
Low / High °C Low and high temperatures during the
previous hour.
In Range "Yes" if the temperature is in range
"No" if temperature is out of range
1 Download button
Field Description
Date Dates in the date range.
Average °C Average temperature for that date.
Low/High °C Low and high temperatures for that date.
Out of Range/Day Number of hours on that date when the
temperature is out of range.
Note: Temperature is measured 30 times
per hour. Out of Range/Day is the
number of hours when at least one of the
30 readings is out-of-range. An out-of-
range reading may occur if the instrument
door was opened during that hour.
7. Select Export.
The report is exported to a CSV.
8. Use the PDF functions to print the report or save the report to another location.
9. Select Download to save the report to local media or a remote location.
Event Description
QC check Determines whether a cell requires full
calibration. Occurs every time a bottle is
unloaded from the cell.
Full calibration Calibrates a cell using the calibration
standards. Occurs every 6 months as routine
maintenance or if the cell fails the QC check.
The instrument cell calibration report displays QC check and full calibration results for each
cell, including the date and time when the event occurred. Event results display on the report
as follows:
• If a cell fails the QC check, Status displays as "Out of Service," Full Calibration
Performed displays as "No," and then full calibration occurs. If a cell passes the QC
check, it is not included on the report.
• If a cell fails full calibration, Status displays as "Out of Service," and Full Calibration
Performed displays as "Yes." If a cell passes full calibration, Status displays as
"Available."
Note: "No information found" displays on-screen when no reportable calibration events occurred for
the cell(s) during the selected time period.
In the following example, the instrument cell calibration report indicates that:
• Cell M1 passed routine full calibration and is available.
• Cell M2 failed routine full calibration and has been taken out of service.
• Cell M3 failed the QC check and was taken out of service pending full calibration. The cell
then passed full calibration and became available again for loading bottles.
• Cell M4 failed both the QC check and full calibration and has been taken out of service.
Note: It is unusual for cells to fail the QC check. Usually, routine full calibration keeps cells well
calibrated.
Negative bottles unload to the waste container. Alerts display when the capacity of the waste
container is full or almost full. Bottles stop unloading if the waste container is full or missing.
The waste container capacity level appears on the Small Display.
The waste bin is considered full if the weight reaches 7 kg or the bottle count reaches 50 or if
the internal waste chute is blocked open.
Alerts display when the capacity of the waste container is full or almost full.
1. Pull the waste container door handle.
The waste container door is located below the bottle loading area.
2. Remove the yellow waste container from the compartment.
3. Remove the liner bag from the waste container.
4. Dispose of the bag in an approved biohazard container.
5. Inspect the waste container and compartment for debris and sample spills.
6. If required, clean with an approved cleaning agent.
See Approved Disinfection Agents for approved cleaning agents.
7. Put a new liner bag in the waste container.
CAUTION: Use only 23- to 25-inch flat bottom biohazard liner bags in
the waste container.
8. Inspect the waste area for any bottles that may have fallen out of the waste container.
9. Return the waste container to the compartment and close the door.
10. If any alerts remain unresolved, see Resolving Alarms and Alerts
1. Select Configuration.
2. Select the appropriate screen tab to access the associated settings.
WARNING
Configuration settings must be carefully implemented in an authorized
manner. Failure to properly address authorization of access to system
configuration settings may lead to incorrect results. Contact
bioMérieux if necessary.
The tabs for the configuration screens are across the top of the screen. A highlighted
tab indicates an active screen.
2 Bank configuration
3 Sound configuration
4 Alarm configuration
6 Security configuration
2 Edit button
3 Save button
4 Cancel button
IMPORTANT: If the drop-down menu does not display the desired bottle type, use the scroll
bar to scroll through the entire list.
Note: The default maximum test time for BACT/ALERT® SA and BACT/ALERT® SN bottles
is 7 days.
The default maximum test time for BACT/ALERT® FA PLUS, BACT/ALERT® FN
PLUS, and BACT/ALERT® PF PLUS bottles is 5 days.
Note: Change the maximum test time for a specific bottle by editing the associated record.
Note: Changes to the maximum test time per bottle type do not impact bottles already under
test.
6. Select Save.
WARNING
Failure to follow appropriate bottle instructions for use for reducing
maximum test time may cause false negative results.
Note: Failure to save the desired changes to the bottle configuration settings may lead to
false results.
Sample volume measurement is supported for the listed clinical bottle types and whole blood
samples only:
• BACT/ALERT® SA
• BACT/ALERT® SN
• BACT/ALERT® FA PLUS
• BACT/ALERT® FN PLUS
Note: Sample volume measurement is for whole blood samples only. Sample volume appears for
other sample types, but will not be supported.
When fill-to volume is enabled, any bottles with a fill-to line will include fill volumes with bottle
information. However, fill volume will not be reported if the internal imager detects foam when
the bottle is loaded. In this event, the instrument will continue to load the bottle and perform a
rescan in approximately 2 hours to measure fill volume.
WARNING
Failure to assure the correct bottle fill volume may lead to incorrect results.
WARNING
If the amount of sample added to a bottle varies too much from the sample
volume target, either higher or lower, incorrect test results may occur.
IMPORTANT: The sample volume target (mL) is set at the factory and cannot be changed.
Use the Bank Configuration screen to configure settings for banks containing one or more
instruments.
The following settings are configured from the Bank Configuration screen:
• Instrument, Bank, and Lab names
• Bank Near Full Threshold (%)
• Incubation Temperature Set Point (°C)
• Waste Near Full Threshold (%)
• Maximum Allowable Time in Retrieval Area (min)
Naming Instruments
From the Bank Configuration screen:
1. Select Instrument Name.
2 Instrument's name
IMPORTANT: Use only letters and/or numbers when naming banks. Do not use special
characters.
3. Select Save.
3. Select Save.
WARNING
Do not adjust incubation temperature if bottles are under test.
Note: The incubation temperature set point range is 30.0 to 40.0 in increments of 0.5. The
usual test range encountered in clinical laboratories is 35°C to 37°C.
3. Select Save.
Note: The default setting is 75%. Available options are 50, 75, or 90% .
1. In the Maximum Allowable Time in Retrieval Area (min) field, select Display Menu.
Note: The maximum allowable time in the retrieval area range is 15 to 60 in increments of
15.
3. Select Save.
Note: The Positive Bottle, Alarm, and Alert volumes can be turned off.
WARNING
Use caution when disabling positive bottle alarms or instrument
alarms. Disabling the positive bottle alarm can lead to delayed results
if the positive bottle is not addressed. If these alarms are disabled and
a positive bottle or instrument malfunction occurs, then there will only
be a visual alarm. There will not be an audible alarm to indicate the
alarm condition.
Note: The snooze range is 1 to 5 minutes in 1 minute increments and then 5 minute
increments up to 30 minutes.
3. Select Save.
2 Save button
3 Checkbox enabled
4 Checkbox disabled
WARNING
A potential hazard (false negative test result) exists if you disable an audible
alarm.
2. Select Save.
• Yellow
2. Select Save.
Note: When this setting is enabled, patient data displays on the Bottle Status Information screen.
Additionally, in order to search by Patient ID in the Search screen, Use Patient Data must
be enabled.
WARNING
Failure to properly configure an accession ID can lead to delayed
results, which may be critical in some applications such as blood
bank.
• Alternate bottle ID barcode on bottle (Alternate bottle ID barcode uses the alternate
ID as the bottle ID in the system and for messages sent to the LIS.)
• Patient ID barcode on bottle
If more than one barcode on bottle setting is enabled (accession ID, alternate bottle
ID, or patient ID), define a validation pattern for either the accession ID or the patient
ID barcode.
2. Enter the validation pattern using regular expressions.
Expression Definition
\d Matches any digit
. Matches any character
* 0 or more of previous expression
+ 1 or more of previous expression
^ Start of string
$ End of string
Expression Definition
[] Explicit set of characters to match
[A-Z] any upper case letter
[a-z] any lower case letter
Description Patterns
Barcode starts with 3 upper case letters, ^[A-Z][A-Z][A-Z]\d\d.*
followed by 2 numbers, followed by any
characters
Barcode starts with any characters but *\d\d$
ends with two digits
Barcode always starts with the upper ACC\d+
case letters ACC followed by 1 or more
digits
Note: Multiple validation patterns are allowed. Use the pipe symbol (“|”) to separate multiple
patterns.
3. Select Save.
IMPORTANT: The following procedure for configuring data retention and accession ID reuse is for
instruments in standalone configuration only.
Note: This setting determines when accession IDs and associated bottle data are removed from
the system, thus allowing for reuse of the accession IDs.
Note: Remote Notification must be enabled in order to access the Notification Center
Dashboard.
2. Select Save.
Note: Refer to the Notification Center Software User Manual for more information.
Note: The archive occurs daily once the Archive Bottles checkbox is selected.
Note: By default, the system saves the archive file to the F drive, but it can be configured to
save in another location.
3. If you wish to archive bottle graphs and images, select Archive Graphs and Images.
4. If you wish to encrypt the archive file, complete the following steps.
a) Select Encrypt Archive.
b) Enter a password.
Note: This password is used to view the archive file on another system. If you lose or
forget this password, you cannot view the archive file.
c) Re-enter the password to confirm.
d) Select Save.
2 21 CFR 11 field
2. Select OK.
2. Select OK.
3. Review settings and select the checkbox to enable or disable settings.
Note: The range is 5 to 30 minutes in 10 minute increments. The default inactivity duration
is 10 minutes.
3. Select Save.
Using a LIS
From the LIS Configuration screen:
1. Select the Use LIS checkbox.
2. Click Save.
Note: When enabled, consolidated messages sent to the LIS include accession IDs and all
associated bottles.
From the LIS Configuration screen:
The following Bottle Report Configuration settings are available for each bottle report:
• Setting the date range
• Adding and removing columns
• Changing a column's position
• Sorting a column in ascending or descending order
3. Click Edit.
4. Enter your user name and password.
5. Click OK.
Sorting a Column
From the Bottle Report Configuration screen:
1. From the Sort Column drop-down menu, select the column you want to sort on the
report.
2. Select the Ascending or Descending checkbox as your desired sort order.
The User Management Users icon shows user identification data (first name, last name,
username), the groups to which the user belongs, and the status of the account (enabled,
locked, password expired).
The User Management Group icon shows the group name, application name, group
description, and group members.
The Lab Admin role sets up users initially and then can grant the permission to the Lab Super
role.
Password Policy
bioMérieux User Management applies a password policy which determines the rules for use
of passwords with the bioMérieux software. The policy includes three security levels:
• High security, in compliance with 21 CFR11, defined by the following rules:
◦ A password is valid for 3 months. The user receives a warning 2 weeks before his
password expires.
◦ The password must be at least 8 characters long and include at least one uppercase
letter and a digit (example: PasswØrd).
◦ The user’s last three passwords are saved for 1 year and cannot be reused during this
period of time.
◦ If the user enters an incorrect password 3 times, the session will not open.
◦ The user has to change his password at his first connection and after his account has
been unlocked.
• Medium security, in compliance with 21 CFR11, defined by the following rules:
◦ A password is valid for 6 months. The user receives a warning 2 weeks before his
password expires.
◦ The password must be at least 8 characters long (example: password).
◦ The user’s last three passwords are saved for 1 year and cannot be reused during this
period of time.
◦ If the user enters an incorrect password 10 times, the session will not open.
◦ The user has to change his password at his first connection and after his account has
been unlocked.
• Low security, no rules imposed.
This password policy cannot be modified from the three security levels described above.
The high level security policy is configured by default.
Note: No security rules are imposed for password creation as the password is temporary
and must be modified at first connection to BACT/ALERT® VIRTUO®.
3. Select Save.
4. Select Yes to save all changes.
A confirmation message displays. The new account appears in the User list.
4. Select Save.
5. Select Yes to save all changes.
Note: After editing a user, log out to allow changes to become effective.
4. Select Save.
5. Select Yes to save all changes.
The user account is removed from the group. A confirmation message displays.
Note: After editing a user, log out to allow changes to become effective.
Note: After editing a user, log out to allow changes to become effective.
Note: After editing a user, log out to allow changes to become effective.
Note: The user must change this new password at the next login.
Note: After editing a user, log out to allow changes to become effective.
Note: After editing a user, log out to allow changes to become effective.
IMPORTANT: There is no connection between User Management and an external Active Directory.
External users that are added to local groups may not exist in the external Active
Directory.
When external users are added to a local group, the user is not created in the local LDAP, but
is added as a member of the group.
Note: The external user name is the same user name that is used to login.
2. Add groups to the external user. See Modifying User Identification Data.
4. Select Save.
Note: Refer to the bioMérieux User Management User Manual for more information.
Note: A blank cell in the table below indicates the user group is not supported for that
configuration.
VIRTUO LabSuper
MYLA Administrator
MYLA Manager
MYLA Tech
MAESTRIA Administrator
MAESTRIA Configuration
MAESTRIA Review
MAESTRIA Routine
◦ Support
◦ Routine
The following groups may be assigned to the LabTech role:
• VIRTUO LabTech
• Audit Trail User (ATuser)
• BCI Connect User
The following groups may be assigned to the LabSuper role:
• VIRTUO LabSuper
• Audit Trail Admin (ATadmin)
• BCI Connect Admin
• User Management Admin (UMadmin)
To ensure that all manually loaded bottles are recognized by the instrument, keep a
handwritten log of all bottles loaded manually. Document the bottle IDs and cell locations.
Minimally, note the number of bottles loaded during each loading sequence. The Emergency
Manual Load Worksheet is an optional template to record bottles loaded and unloaded.
WARNING
In the Emergency Manual Bottle Loading procedure, if you reload a bottle that
was already in the system without scanning its barcode, the bottle is treated
as a new bottle. Data previously collected for that bottle is disregarded.
WARNING
A potential pinch hazard exists for mechanical parts during manual loading
and unloading of bottles.
Manually load and unload bottles if the following hardware failures occur:
• The conveyor belt is jammed or not functioning properly.
• The bottle indexer is jammed or not functioning properly.
• The robot is jammed or not functioning properly.
• The automatic scanner is not functioning properly and is causing anonymous bottles.
CAUTION: Do not manually load and unload bottles at the same time.
Manually load bottles and manually unload bottles as separate actions.
WARNING
The potential exists for bottle damage and leaking. Damaged and leaking
bottles may cause delayed or false positive test results.
Note: The best way to circumvent false positives after the door has been opened is to keep the
door closed for 3.5 hours before reopening. This ensures a stable temperature environment
after the door has been opened.
Barcode Scanner Not Indicates the handheld barcode scanner has not
Detected been detected.
WARNING
Follow these steps to manually unload positive bottles. Failure to follow
these steps may cause clinical false negative results.
Note: Do not manually load bottles while in the Unload Positive mode.
WARNING
Step on the power supply cover plate when manually loading or unloading
bottles, if necessary to achieve more height. Do not step on a separate
device, as the consequence may be a crush hazard.
Note: The number of positive bottles to be unloaded appears prominently on the Home screen.
Note: If automated unloading is interrupted, do not unload final negative bottles. Keep these bottles
in the instrument until the automated unloading is repaired. If the automated unloading repair
is delayed and unloading negative bottles is necessary, contact bioMérieux support for
guidance on how to perform this task.
Note: The bottles on the screen are color coded. Positive bottles are yellow or red
(depending on user configuration), negative to date bottles are blue, and negative
bottles are gray.
CAUTION: Do not step on the bottom door hinge arm when the
instrument door is open as this may damage the instrument.
Note: Ensure no gap exists between the cabinet and the door.
Note: The bottles on the screen are color coded. Positive bottles are yellow or red
(depending on user configuration), negative to date bottles are blue, and negative
bottles are gray.
6. Verify the list on the screen with the actual bottles unloaded. Use the Emergency
Manual Load Worksheet as an optional template to manually record bottle information.
The number of positive bottles on the Home screen should be 0.
WARNING
To minimize temperature loss, limit the amount of time the instrument
door is open to once every two hours. Load bottles in batches so that
the door is not open more than 10 minutes. Shut the door when the
Door Open Too Long alarm sounds. The consequences of not
following these instructions may be false negative test results.
WARNING
Leaving the door open past the Door Open Too Long alarm may cause
missed readings and incorrect or no results and the need to
subculture bottles.
If a bottle still under test (negative to date status) is unloaded, a Negative to Date
Bottle Removed in Manual Loading alarm sounds. See Emergency Manual Loading
Bottles to reload this bottle.
7. Verify the actual bottles unloaded against the bottles listed on the Positive Bottles
Report. See Generating Bottle Reports in Site Configuration (MYLA).
8. Select Exit.
If the system is still alarming for positives because a positive bottle was not removed, then
repeat the process starting from step 1.
WARNING
Follow these steps to enter and scan barcodes correctly. Failure to follow
these steps may cause clinical false negative results.
WARNING
Failure to account for loaded or unloaded bottles during manual loading may
lead to a consequence of no results or false negative results.
Note: To minimize temperature loss, limit the amount of time the instrument door is open. Load
bottles in batches so that the door is not open more than 10 minutes. Shut the door when the
Door Open Too Long alarm sounds.
WARNING
Leaving a bottle out of the system too long may lead to false negative
results.
WARNING
Step on the power supply cover plate when manually loading or unloading
bottles, if necessary to achieve more height. Do not step on a separate
device, as the consequence may be a crush hazard.
CAUTION: Do not step on the bottom door hinge arm when the instrument
door is open as this may damage the instrument.
Note: Any bottle that is manually loaded because of disabled automatic loading will be
automatically rescanned after instrument automation is repaired.
CAUTION: Do not manually load and unload bottles at the same time.
Manually load bottles and manually unload bottles as separate actions.
WARNING
Subculture any bottle before reloading it manually. The consequences of not
subculturing a bottle before reloading it manually may be false negative
results.
WARNING
Subculture any bottle without a final result if the bottle will not be reloaded.
The consequences of not following this instruction may be no results.
WARNING
Do not manually load a bottle into a cell from which you just manually
unloaded a bottle without scanning the bottle in accordance with Emergency
Manual Loading Bottles. This could result in erroneously creating a new
bottle record if the same bottle was unloaded and immediately reloaded; and
incorrect results may be a consequence.
WARNING
Allow the instrument to image manually loaded anonymous bottles rather
than resolving the anonymous bottles (unless the bottle goes positive before
automation is restored). The consequences of not following these
instructions may be false results.
Note: Select Manual Load and scan the barcode before loading, or the loaded bottle will be
anonymous. See Emergency Manual Unloading Anonymous Bottles to resolve any
manually loaded anonymous bottles.
Note: When manually loading bottles, avoid placing bottles into cells that are unavailable.
4. Retrieve the handheld barcode scanner from its intended storage location to the right
of the waste container inside the instrument.
5. Open the instrument door.
6. Plug the handheld barcode scanner into the USB port on the back of the opened
instrument door.
The Barcode Scanner Detected icon appears on the Manual Load screen. If the
scanner is not detected, the Barcode Scanner Not Detected icon appears on the
Manual Load screen. Ensure the barcode scanner is plugged in properly.
Note: If the handheld barcode scanner is unresponsive or not detected then unplug and
reinsert the scanner into the barcode scanner port.
WARNING
Incorrectly entering and scanning barcodes to load bottles manually
may cause no results or false negative results.
7. Scan the bottle ID and other barcodes on the bottle in any sequence.
• The scanner beeps once for each barcode scanned and the green light on the
scanner blinks.
• The recognized barcode appears on the Manual Load Barcode screen if the
barcode is valid and is not a duplicate barcode.
• When the configured number of barcodes have been recognized, the scanner
beeps two more times, vibrates, and the green light on the scanner blinks.
• The Manual Load Bottle icon appears on the Small Display.
Note: If the barcodes were not successfully scanned or if more than the acceptable number
of barcodes were scanned, the scanner beeps seven times. Select Clear Barcodes
and rescan the barcodes to correct any scanning mistakes or to begin scanning a
new bottle.
Normally, blue and red LEDs illuminate throughout all empty cells in a continuous
motion. In the Manual Load mode, after the acceptable number of barcodes is
confirmed by the barcode scanner, the blue and red LEDs alternate between
illuminating all empty available cells in sequence or one suggested cell per rack.
8. Insert the bottle completely into the suggested cell, paying attention to the following:
• Place bottles into the cell with the bottle sensor inside the cell and the bottle top
outside the cell.
WARNING
Failure to properly orient the bottle when loading may lead to
incorrect results.
WARNING
Do not touch or move other bottles previously loaded that do not
need to be moved; the consequences of not following these
instructions may be false positive results.
WARNING
Incorrect results may be a consequence of incorrectly loading bottles,
or of not fully seating bottles when manually loading into cells.
9. Wait for the scanner to beep twice and vibrate to signify that the instrument recognizes
the bottle loaded. An attempt to scan another bottle before the previously scanned
bottle is recognized results in 7 beeps, the red light on the scanner flashing, and the
scanner vibrating.
The barcode scanner does not scan the next bottle until this bottle is recognized as
being loaded.
Note: The instrument does not detect a bottle that is loaded in an unavailable cell. If this
occurs, See Discrepancy in the Manually Loaded Bottle Count to resolve.
Note: In order to avoid potential false positives, do not load bottles into the same rack or
into cells adjacent to previously loaded bottles during the same load event.
11. Remove the handheld barcode scanner from inside the instrument door when
scanning is complete.
An alarm sounds if the door is closed with the scanner still attached.
12. Replace the handheld barcode scanner in its intended storage location to the right of
the waste container inside the instrument.
WARNING
Each bank has only one handheld barcode scanner. The
consequences of not replacing the handheld barcode scanner in its
intended location when scanning is complete may be no results.
CAUTION: Do not step on the bottom door hinge arm when the
instrument door is open as this may damage the instrument.
WARNING
The consequences of not correctly following these steps after the
anonymous bottle alert appears may be no results.
If the anonymous bottle alert appears on the Large Display after manual loading,
complete one of the following actions:
• If the bottle has been incubating for less than four hours and the bottle ID and cell
ID are recorded, unload the bottle. See Emergency Manual Unloading Anonymous
Bottles.
• If the bottle has been incubating for more than four hours and instrument repair is
expected soon, let the bottle continue to test as anonymous.
• If the bottle has been incubating for more than four hours and instrument repair is
not expected for some time, unload and then reload the bottle. See Emergency
Manual Unloading Anonymous Bottles and Emergency Manual Loading Bottles.
If anonymous bottle records were created during the manual load procedure:
Note: Do not use this procedure to resolve any anonymous bottles that loaded automatically.
Note: The following procedure requires a handheld barcode scanner. If you reload an anonymous
bottle without scanning its barcode, it remains anonymous.
1. Select Information.
2. Select Manual Unload Anonymous.
A list of anonymous bottles appears showing the cell locations.
Note: These bottles are all anonymous; however, if they were just loaded they are all likely
negative to date (blue).
CAUTION: Do not step on the bottom door hinge arm when the
instrument door is open as this may damage the instrument.
4. Remove the bottles from the cells listed, one bottle at a time. As you remove each
bottle, record the bottle ID, accession ID, and the cell location.
5. Close the instrument door.
A list of the recently loaded and unloaded bottles appears on the Small Display.
6. Verify the list on the Small Display with the actual bottles unloaded and then select
Exit.
The Bottle Label Is Required message appears on the Large Display for the recently
unloaded anonymous bottles.
IMPORTANT: Do not dismiss the message before identifying the anonymous bottles.
Dismissing the message will cause the bottle records to be lost, and the bottles
will need to be subcultured as a result.
7. On the Large Display, perform the following steps to identify the anonymous bottles.
a) Select the Edit Bottles Requiring a Label button.
b) Enter your user name and password if not already logged in.
c) On the Edit Bottle Record screen, enter the correct bottle ID based on the cell
location recorded in Step 4, save the change, and then enter the accession ID.
Note: A bottle image is not available to obtain information from because the botttle was
loaded manually.
Use this worksheet as a template to manually record bottle information and confirm that all
manually loaded and unloaded bottles have been accounted for when you perform the
Emergency Manual Load procedure. Compare this list to the Manually Loaded and Unloaded
Bottles list on the Small Display.
WARNING
The consequences may be no results if all the bottles have not been
accounted for.
Any active alarms and alerts that are unresolved display in Zone 2 on the Home screen.
In addition, symbols may appear in the Severity column to indicate the severity of the alarm or
alert.
Column Description
Severity An orange triangle indicates an alert. This
does not prevent the system from
operating properly. A red circle indicates
an alarm. This problem must be resolved
immediately for the system to operate
properly.
Category Pictorial image indicates the category of
the alarm or alert: bottle, instrument,
communication, or PC
Type Text describes the specific alarm or alert.
Number The number of occurrences of the
specific alarm or alert.
Actions The means to resolve an alarm or alert.
Three buttons are available depending on
the alarm or alert.
Details Provides details about the alarm or alert
and the affected bottles. Select Bottle
Status Information to go directly to the
Bottle Status Information screen.
Column Description
Dismiss Clears the alarm or alert from the screen.
Resolve Additional instructions about how to
resolve the alarm or alert appear. See
Alarm and Alert Causes and Resolutions.
WARNING
If you must open the door to resolve an alarm or alert, do not lean on the
conveyor belt. Leaning on the conveyor belt when the door is open (or any
other time as when the door is closed) may damage the door or the conveyor
belt and may result in a dropped bottle and no results.
CAUTION: Do not keep the door open for more than 10 minutes as that may
result in temperature loss and affect bottle test results.
Note: When the instrument door is open, follow the proper procedure. See Opening the Instrument
for Internal Access.
WARNING
Resolve all alarms immediately. The consequences of delays may be delayed
results.
WARNING
A potential hazard (delayed or false negative results) exists in the following
situations:
• System alarms are ignored or dismissed without corrective action.
• Corrective action and troubleshooting procedures are not correctly
followed.
Failure to act appropriately may result in a subculture recommendation.
Follow the instructions for the specific alarm in the Instrument Alarms and Alerts table to help
resolve the issue.
Anonymous The bottle type and/or See Identifying an Anonymous Bottle without a Visible bioMérieux Label and Identifying
bottle found. bottle ID are missing from an Anonymous Bottle with a Visible bioMérieux Label to resolve the alert.
[3] the bottle record.
In Alternate Bottle Mode If the bioMérieux bottle ID is visible, unload the bottle and apply an alternate bottle ID
— Alternate Bottle ID label that matches the Alternate Bottle ID pattern configuration. See Applying a
label is missing from the Barcode Label to a Bottle.
bottle.
In Alternate Bottle Mode If the alternate bottle ID label is visible on the bottle image, unload the bottle and
— Alternate Bottle ID correctly apply the alternate bottle ID label to the bottle. See Applying a Barcode Label
label is incorrectly applied to a Bottle.
to the bottle.
In Alternate Bottle Mode Update the alternate bottle ID barcode configuration pattern to match the alternate
— Alternate Bottle ID bottle IDs being used. See Enabling Barcodes on Bottles and Defining Validation
label does not match the Patterns.
configured Alternate
Bottle ID pattern.
In Alternate Bottle Mode 1. On the PC, search for the bottle and view the image. A list of scanned barcodes
— Multiple bottle labels appears next to the image.
match the configured 2. View the scanned barcode list and determine if more than one barcode
Alternate Bottle ID matches the Alternate ID barcode pattern.
pattern.
3. If there is more than one barcode in the scanned barcode list that matches the
Alternate ID barcode pattern, update the alternate ID barcode configuration
pattern to be more specific to match the Alternate ID barcode label on the
bottle. See Enabling Barcodes on Bottles and Defining Validation Patterns.
In Alternate Bottle Mode Edit the bottle and assign a bottle type.
— No Bottle Type
assigned.
Bottle data One or more barcodes 1. View the Bottle Status Information screen for the changes.
changed on were changed on a bottle 2. If necessary, correct the bottle data, and possibly remove the bottle.
the listed before reload.
3. Select Dismiss.
bottle(s). [6]
® ®
A currently A bottle has been loaded A bottle is under test and the results are visible in BACT/ALERT VIRTUO .
expired bottle that has already reached
Note: Only a user with LabSuper access is authorized to process expired bottles. See
has been its expiration date.
Managing User Accounts and User Groups.
loaded. [8] Results may be
unreliable. Select one of the following options:
WARNING
bioMérieux shall not be liable as to any defect (including, but not
limited to, incorrect results that may affect patient outcomes)
arising from the use of expired bottles, or the communication of
the results obtained using expired bottles. Users who use expired
bottles do so at their own risk.
Bottle(s) left A bottle has been left in Remove the bottle from the bottle retrieval area so that additional bottles can be
in the bottle the bottle retrieval area unloaded from the incubator.
retrieval area beyond the configured
beyond the time period.
configured
time period.
[9]
The The barcodes scanned Unload the bottle and verify that the bottle ID does not exist for another bottle in the
barcode(s) on on the bottle matched database. If the bottle ID already exists in the database, it will be necessary to relabel
the listed one or more other bottle the bottle.
bottle(s) IDs.
match the
Bottle ID of
other bottle(s)
in the system.
[12]
Canceled A bottle test was 1. Reestablish the communication between the MCB and embedded PC.
bottle(s) could canceled during a loss of 2. Select Details to display the Alarm Details.
not be communication between
3. Select Search to display the Bottle Details.
unloaded the MCB and the
4. Select Unload to unload the bottle.
automatically. embedded PC.
5. If the problem persists, contact your local bioMérieux representative.
[14]
The Bottle ID The bottle ID for a 1. Access the Bottle Status Information screen for the bottle just loaded.
for a recently recently loaded bottle 2. Unload the bottle.
loaded bottle matches the bottle ID of a
3. Enter a new bottle ID.
matches the bottle previously known.
4. Apply a new label.
Bottle ID of a
5. Reload the bottle.
bottle
previously Note: If both bottles are available, use them to verify the correct bottle ID.
known to the
instrument. WARNING
[19]
Incorrect results may occur if a duplicate bottle alert is not
resolved.
WARNING
If it is determined that two bottles actually have the same bottle
ID, then contact your local bioMérieux representative to report the
issue.
A bottle was reloaded Unload the bottle. The alarm will resolve once the bottle is unloaded.
more than 10 days after it
was last unloaded. WARNING
If it is determined that two bottles actually have the same bottle
ID, then contact your local bioMérieux representative to report the
issue.
Negative A negative bottle was 1. Extend the maximum test time. See Changing the Maximum Test Time.
bottle(s) was reloaded.
reloaded. [32] Note: Do not wait more than 15 minutes to extend the maximum test
time. If automatic unloading is enabled, the bottle will unload to
the waste container after 15 minutes.
Note: If needed, access the Bottle Status Information screen to view the bottle details.
No accession No accession ID 1. Add the missing information on the Bottle Edit screen.
ID assigned assigned for the bottle. 2. Unload the bottle if it is necessary to apply an accession barcode.
for the bottle.
3. Apply a label to the bottle. See Applying a Barcode Label to a Bottle.
[33]
4. Reload the bottle.
Multiple bottle barcode 1. On the PC, search for the bottle and view the image. A list of scanned barcodes
labels match the appears next to the image.
configured accession ID 2. View the scanned barcode list and determine if more than one barcode
pattern. matches the Accession ID barcode pattern.
3. If there is more than one barcode in the scanned barcode list that matches the
Accession ID barcode pattern, update the accession ID barcode configuration
pattern to be more specific to match the accession barcode label on the bottle.
See Enabling Barcodes on Bottles and Defining Validation Patterns.
Accession ID barcode Update the accession ID barcode configuration pattern to match the accessions being
label does not match the used. See Enabling Barcodes on Bottles and Defining Validation Patterns.
configured accession ID
pattern.
Accession ID barcode is Accession ID barcodes must be placed in a vertical position in order to be decoded by
not placed in the correct the imager.
orientation.
• Select Cancel. The bottle status remains positive. The bottle unloads.
Unmatched A bottle was manually 1. Record the cell ID of the bottle and unload the bottle.
reloaded reloaded without being 2. Perform a subculture, as the readings for the bottle are split across two different
bottle has scanned and therefore bottle records.
resulted in a anonymous. Upon
3. (Site configuration only) To remove the bottle from the Anonymous Bottle report:
duplicate rescan, the data is sent to
a. Locate the bottle on the report using the cell ID recorded in Step 1.
bottle record. the software and
[36] determines that there are b. Select the bottle and then click the Waste Can icon to delete the record.
Sample Sample volume is out of Dismiss the alert. See Sample Volume Variation
volume is out configured range for the
of the listed bottle.
configured
range for the
listed
bottle(s). [43]
bioMérieux The subculture The Subculture Recommended Alarm is generated for bottles that have a gap of
recommends recommended alarm is readings of at least four hours. The algorithm may also generate the alarm after a gap
subculture of generated for bottles that of two hours, depending on the magnitude in the change of readings. bioMérieux
the bottle(s). have a gap of readings of recommends that the bottles are unloaded and subcultured.
[47] at least four hours. The
1. Unload the bottle.
algorithm may also
2. Subculture the bottle.
generate the alarm after a
3. Reload the bottle.
gap of two hours,
depending on the
magnitude in the change WARNING
of readings. bioMérieux
Reloaded bottles must be subcultured. The consequences of
recommends that the
failing to subculture may be incorrect results.
bottles are unloaded and
subcultured.
The number The number of non- A non-bioMérieux bottle has been loaded. bioMérieux cannot guarantee results
of non- bioMérieux bottles loaded obtained with non-bioMérieux bottles. Contact your local bioMérieux representative for
bioMérieux is approaching the assistance.
bottles loaded maximum allowed.
is
approaching
the maximum
allowed. [58]
A label needs A bottle ID barcode label 1. Relabel the bottle to match the entered bottle ID before reloading this bottle.
to be applied is required. 2. Label the bottle with a generic barcode.
to this bottle.
3. Contact your local bioMérieux representative for assistance.
[72]
Canceled Canceled bottle(s) was Remove the canceled bottle from the bottle retrieval area.
bottle(s) was reloaded. Canceled
reloaded. [74] bottle(s) cannot be
reloaded for testing.
Bottle(s) of an A bioMérieux bottle was 1. Unload the bottle from the bottle retrieval area.
unsupported loaded that is not 2. Dismiss the alert.
type will not supported by
® ®
3. Process the bottle manually.
be loaded. BACT/ALERT VIRTUO .
[75] The bottle will unload to
the bottle retrieval area.
There was a There was a conflict 1. Review the bottle record and the image with the listed scanned barcodes to
conflict between data from a determine which data values are correct.
between data scanned barcode and 2. Edit if necessary.
from a data in the bottle record.
scanned
barcode and
data in the
bottle record.
[76]
A bottle with a A bottle with a duplicate Resolve the bottle ID duplication before reloading the bottle. Failure to resolve the
duplicate bottle ID of another duplication will result in merged records for these two bottles and potentially incorrect
bottle ID of loaded bottle was results.
another unloaded to the retrieval
loaded bottle area.
was unloaded
to the retrieval
area. [78]
• Examine the bottle for any unnecessary labeling that may have caused the
bottle to get stuck in the cell and remove if found.
• Close the instrument door.
• Place the bottle on the conveyor.
5. If the cell contained a calibration standard:
A bottle was A bottle was found in the 1. Open the instrument door.
found in the robot gripper when the 2. Remove the bottle from the robot gripper.
gripper when instrument was started.
3. Place the gripper mechanism in a safe to move position, ensuring the gripper
the instrument
jaws are horizontal and do not extend beyond the end of the bracket.
was started.
4. Close the instrument door.
[84]
5. Place the bottle on the conveyor to reload.
A bottle was A bottle was manually Unload the bottle. See Emergency Manual Unloading Positive Bottles or Emergency
manually loaded and determined to Manual Unloading Anonymous Bottles.
loaded and have an unsupported
determined to bottle type or to be a
have an duplicate.
unsupported
bottle type or
to be a
duplicate. [88]
A bottle was A bottle was manually 1. Verify that this bottle should be unloaded.
manually unloaded during the 2. Dismiss the alarm on the screen.
unloaded manual loading process.
3. If the bottle should not be unloaded, reload the bottle. See Emergency Manual
during the
Loading Bottles.
manual
loading
process. [89]
A bottle found A bottle found in an 1. Perform a subculture of the bottle if it has been more than four hours since the
in a non- unavailable cell was bottle has been manually loaded, or the load time is unknown.
operational unloaded to the bottle 2. Place the bottle in the loading area.
cell was retrieval area.
unloaded to
After closing the door 1. If it is a bottle, reload it.
the bottle
following a Bottle 2. If it is a calibration standard:
retrieval area.
Dropped Alarm [7], the
[90] • Open the door.
robot discovered an
• Place it back in its reserved location. Locate the other calibration standards
unseated bottle or in the instrument (the default location is at the bottom right in rack P), and
calibration standard in a place it back in empty reserved cell. See the Figure 102 for placement.
The The bioMérieux bottle ID 1. Compare the bottle ID on the image to the bottle ID on the record.
bioMérieux conflicts with the bottle ID 2. If the bottle IDs are different, unload the bottle and subculture. The bottle(s)
bottle ID manually scanned. This cannot be safely reloaded.
conflicts with indicates that bottle(s)
3. Contact your local bioMérieux representative for assistance.
the bottle ID were not manually loaded
manually correctly and should be
scanned. [92] subcultured.
Duplicate The system has detected 1. Unload the duplicate bottle from the other instrument. The bottle cannot be
bottle was that bottles with duplicate loaded into the system until it has a unique bottle ID.
detected in bottle IDs have been 2. If unable to resolve, contact your local bioMérieux representative for assistance.
another loaded across different
instrument. instruments. A
[99] corresponding alarm "An
unexpected bottle has
been discovered and
unloaded by the
instrument.[97]" should
already exist on the
instrument that has the
first of the duplicate
bottles loaded.
A non- A non-bioMérieux bottle Contact your local bioMérieux representative for assistance.
bioMérieux has been loaded.
bottle has bioMérieux can not
been loaded. guarantee results
[102] obtained with non-
bioMérieux bottles.
An agitation An agitation malfunction 1. Open the instrument door. See Opening the Instrument for Internal Access.
malfunction has occurred in the listed 2. Inspect the racks for problems that might inhibit movement. IF a problem is
has occurred instruments. Test results found, follow the robot jam recovery procedure. See Unjamming the Robot.
in the listed for bottles in the RDA
3. If no problem is found, close the door and wait about 15 minutes for the
instrument. [2] may be impacted.
instrument to initialize.
4. If a problem persists, contact your local bioMérieux representative.
Bottle(s) on This takes into account 1. Ensure conveyor cover is on all B units and attempt to reload at the A unit.
the conveyor not only if the bank is full 2. Unload bottles with a final negative or positive status.
cannot be but also if all
3. Attempt to load bottles on each individual B unit. See Loading Bottles across a
loaded due to downstream instruments
Bank.
lack of bank are full. The upstream
4. Consider unloading the bottles nearing completion. See Unloading Specific
capacity or a may not be full.
Bottles by User Request.
malfunction/
jam. [5] Insufficient capacity to Bottles must be unloaded from the listed instrument or from any instrument to the right
load or transfer bottle in of the listed instrument. Consider unloading bottles nearing completion to add capacity.
listed instrument See Unloading Specific Bottles by User Request.
The instrument cannot 1. Open and close the blue flap to initialize the loader.
load bottles, even 2. Place the bottle(s) on the conveyor to load.
though capacity is
available and no other
alarm or alert displays.
Bottle(s) The bottle was not 1. At the Large Display, select Details from the alarm message to view the cell
dropped inside detected at its intended location. The cell location listed indicates the last known location of the bottle. If
the instrument. location and occrred the cell location is listed as P24, then the bottle dropped before it was loaded.
[7] during one of the 2. Open the instrument door.
following actions:
3. Inspect the following areas that may contain a bottle or calibration standard:
• Loading a bottle
• bottom of instrument
• Unloading a bottle
• robot gripper (bottle may be across from the gripper)
to the waste
container • retrieval chute
• Unloading a bottle • cell containing an unseated bottle or calibration standard
to the retrieval chute • waste chute
• Rescanning a bottle • rescan station
4. Inspect the listed cell location to determine if the bottle has been loaded.
Confirm the bottle is fully seated in the cell. If it is a bottle, then remove it from
the cell. If it is a calibration standard in cell P24 through P27, then push the
bottle further into the cell. If a calibration standard is found to be dropped inside
of the instrument, then place it back into the appropriate cell position: CAL 1 =
P24, CAL 2 = P25, CAL 3 = P26, CAL 4 = P27.
5. Close the door.
6. Reload the bottle on the conveyor.
WARNING
The consequences of not following instructions to resolve a bottle
dropped alarm may be no test results.
Note: If no bottle or calibration standard was found, then it is possible that it was placed
in a cell, but was not seated well enough to be detected by the instrument. After closing
the door, the robot will attempt to unload from the intended cell. If a bottle or calibration
standard is found by the robot, the Bottle Found In Unavailable Cell Alarm occurs and it
is returned to the retrieval area. If the dropped bottle is a calibration standard, see
Figure 102 for placement.
A bottle was A bottle has been placed 1. Access the Bottle Status Information screen to view the bottle details and
placed in a cell in a cell, but not fully determine the cell location.
that has seated. 2. Open the instrument door.
become non-
3. Remove the bottle from the cell.
operational.
4. Close the instrument door.
[10]
5. Place the bottle on the conveyor to reload.
A cell has become non- 1. Access the Bottle Status Information screen to view the bottle details and
operational while a bottle determine the cell location.
is under test. 2. If a bottle unloaded to the return chute when the alarm occurred, reload the
bottle on the conveyor.
3. If the bottle details show the bottle as loaded, press the unload button. Reload
the bottle on the conveyor.
4. Subculture the bottle if a subculture alarm occurs after the reload.
5. If the alarm occurs again for the same cell, contact your local bioMérieux
representative.
Cell calibration Cell calibration failed Contact your local bioMérieux representative if multiple cells are out of service.
failed and the and the cell was
cell was removed from service.
removed from
service. [15]
Conveyor is The conveyor is not 1. Remove the indexer cover and retrieve bottles from the conveyor belt and
not functioning functioning properly. indexer.
properly. [17] 2. Inspect the conveyor belt for foreign objects or damage.
Instrument Instrument door or door Close the door and allow the instrument temperature to equilibrate.
door open too latch cover has been
long. [18] open more than 2
WARNING
minutes.
Leaving the door open past the Door Open Too Long alarm may
cause missed readings and incorrect or no results and the need to
subculture bottles.
The Bottle The bottle imager is not Contact your local bioMérieux representative for assistance.
Imager is not functioning. Bottles
functioning. loaded automatically
[21] while the bottle imager is
not functioning are
loaded as anonymous
bottles. Do not reload
bottles while this alarm
displays.
The Bottle The bottle indexer is 1. Verify that all bottles are upright on the conveyor.
Indexer is jammed. 2. Remove the indexer cover.
jammed on the
3. Verify that no bottles are wedged between the conveyor and the indexer
listed
assembly.
instrument(s).
4. Replace the indexer cover.
[22]
5. Verify that the message clears.
6. If the problem persists, check the Robot Activity on the Small Display.
7. If Robot Activity is not on the screen, open and close the blue flap.
8. If the problem still persists, contact your local bioMérieux representative for
assistance.
WARNING
Using a damaged indexer cover may lead to delayed results or
physical injury.
The Bottle The bottle indexer is not 1. Verify that all bottles are upright on the conveyor.
Indexer is not functioning. 2. Remove the indexer cover.
functioning on
3. Verify that no bottles are wedged between the conveyor and the indexer
the listed
assembly.
instrument(s).
4. Replace the indexer cover.
[23]
5. Verify that the message clears.
6. If the problem persists, check the Robot Activity on the Small Display.
7. If Robot Activity is not on the screen, open and close the blue flap.
8. If the problem persists, contact your local bioMérieux representative.
Power was Power was lost and the 1. Check the instrument power connections.
lost in the system was operating on 2. If the problem persists, contact your local bioMérieux representative for
listed UPS power. assistance.
instrument(s).
[37]
WARNING
A potential hazard (false negative results) exists if power is off for
longer than two hours.
Bottle retrieval The bottle retrieval area Remove bottles from the bottle retrieval area.
area is full for is full of bottles. Bottles
listed cannot be unloaded from
instrument(s). the instrument until
[39] bottles are removed from
the retrieval area.
A robot Robot calibration has 1. Open the instrument door. See Opening the Instrument for Internal Access.
calibration failed. 2. Follow the robot jam recovery procedure. See Unjamming the Robot.
failure has
occurred in the
WARNING
listed
instrument(s). The consequences of failing to assure proper rack re-
[40] installation may be delayed results.
3. Close the instrument door. See Opening the Instrument for Internal Access.
4. Wait about 15 minutes for the robot to recalibrate and the robot activity indicator
disappears from the small touchscreen.
5. If the problem persists, contact your local bioMérieux representative for
assistance.
A robot jam The robot is jammed. 1. Open the instrument door. See Opening the Instrument for Internal Access.
has occurred 2. Follow the robot jam recovery procedure, including a visual inspection of the
in the listed bottom of the instrument for fallen bottles. See Unjamming the Robot.
instrument(s).
3. Close the instrument door. See Opening the Instrument for Internal Access.
[41]
4. If unable to resolve, contact your local bioMérieux representative.
A robot A robot malfunction has 1. Open the instrument door. See Opening the Instrument for Internal Access.
malfunction occurred. 2. Follow the robot jam recovery procedure. See Unjamming the Robot.
has occurred
3. Close the instrument door. See Opening the Instrument for Internal Access.
in the listed
4. If unable to resolve, contact your local bioMérieux representative.
instrument(s).
[42]
Temperature is Instrument temperature 1. If the temperature is too low, ensure the door properly closes.
out of range is one degree outside of 2. Monitor the instrument temperature, allow to stabilize.
for the listed the instrument set point
3. If the temperature does not stabilize, contact bioMérieux Customer Service.
instrument(s). for more than 20
[48] minutes.
CAUTION: Loss of incubator temperature may adversely
affect bottle testing. Address immediately.
4. If the temperature is too high, verify the ambient temperature is within normal
operating range. See Environmental Characteristics.
The UPS The UPS battery is Contact your local bioMérieux representative for assistance.
battery is bad defective.
in the listed
instrument(s).
[50]
The UPS The UPS battery is low. Check that the UPS is properly plugged in to AC power. If the connections are correct,
battery is low and there is no power outage, contact your local bioMérieux representative for
in the listed assistance.
instrument(s).
[51]
WARNING
A potential hazard (false negative test result) exists if power is off
for longer than two hours.
Waste Waste container door is Replace the waste container or close the waste door.
container door open or the waste
If the problem persists,contact your local bioMérieux representative for assistance
is open or the container is missing.
waste Negative bottles cannot
container is unload to the waste
missing in the container.
listed
instrument(s).
[52]
Waste Waste container is full. Empty the waste container. See Emptying the Waste Container.
container is Negative bottles cannot
If the problem persists,contact your local bioMérieux representative for assistance.
full in unload to the waste
instrument(s). container until it is
[53] emptied.
Waste The waste container is Empty the waste container as soon as possible or negative bottles will not unload to the
container is nearly full. waste container. See Emptying the Waste Container and Configuring the Waste Near
near full in the Full Threshold.
listed
If the problem persists,contact your local bioMérieux representative for assistance.
instrument(s).
[54]
Problem with a A bottle has fallen on the 1. Set the bottle upright or remove the bottles from the loading area.
bottle(s) on conveyor. 2. Open and close the door flap if the alarm does not resolve.
the conveyor
3. Contact your local bioMérieux representative for assistance.
of the listed
instrument(s).
[59]
A general A general hardware fault Contact your local bioMérieux representative for assistance.
hardware fault has occurred.
has occurred
WARNING
on the listed
instrument(s). A potential hazard (delayed or false negative results) exists in the
[60] following situations:
The bottle The bottle retrieval area 1. Remove bottles from the bottle retrieval area.
retrieval area is jammed. 2. Open the instrument door and manually remove the bottles from the bottle
is jammed on return chute.
the listed
3. Close the instrument door.
instrument(s).
4. If unable to resolve, contact your local bioMérieux representative.
[62]
An instrument An instrument rack fault 1. Manually unload bottles from the affected rack.
rack fault has has occurred on the 2. Place bottles on conveyor to reload.
occurred on listed instrument rack.
3. Contact your local bioMérieux representative for assistance.
the listed
instrument
rack(s). [63]
WARNING
A potential hazard (delayed or false negative results) exists in the
following situations:
A spinner A spinner malfunction 1. Verify that all bottles are upright on the conveyor.
malfunction has occurred. 2. Remove the indexer cover.
has occurred
3. Verify that no bottles are wedged between the conveyor and the indexer
on the listed
assembly.
instrument(s).
4. Replace the indexer cover.
[67]
5. Verify that the message clears.
6. If the problem persists, check the Robot Activity on the Small Display.
7. If Robot Activity is not on the screen, open and close the blue flap.
8. If the problem persists, contact your local bioMérieux representative for
assistance.
A spinner jam A spinner jam has 1. Verify that all bottles are upright on the conveyor.
has occurred occurred. 2. Remove the indexer cover.
on the listed
3. Verify that no bottles are wedged between the conveyor and the indexer
instrument(s).
assembly.
[68]
4. Replace the indexer cover.
5. Verify that the message clears.
6. If the problem persists, check the Robot Activity on the Small Display.
7. If Robot Activity is not on the screen, open and close the blue flap.
8. If the problem still persists, contact your local bioMérieux representative for
assistance.
The conveyor Conveyor belt jammed 1. Remove the indexer cover and retrieve bottles from the conveyor belt and
is jammed on when attempting to load indexer.
the listed bottles. 2. Inspect the conveyor belt for foreign objects or damage.
instrument(s).
a. If no problem is evident, replace the indexer cover and place bottles on
[69]
conveyor to load.
b. If the problem persists, contact your local bioMérieux representative for
assistance.
The The instrument transfer 1. Remove the indexer cover of the affected unit, checking the transfer arm.
instrument mechanism inside the 2. Remove any obstruction and ensure that the bottle is upright, and replace the
transfer indexer cover has indexer cover.
mechanism jammed.
3. If the bottle cannot be located, remove the conveyor cover of the unit to the
inside the
right, remove any obstruction, and replace the conveyor cover.
indexer cover
4. Reload the bottle on the conveyor.
has jammed.
[71]
The ambient The ambient room 1. Ensure the ambient room temperature is within specifications (see
temperature is temperature is above the Environmental Characteristics).
above the recommended
recommended specifications.
WARNING
specifications.
[79] Bottle readings are not taken if the internal temperature is
outside specifications. The instrument stops taking
readings when the internal temperature is more than 1 C
from the setpoint for 20 minutes. The instrument begins
taking readings when the internal temperature is less that
0.5 C from setpoint for at least 30 minutes.
The light The light source for the Contact your local bioMérieux representative.
source for the imager is not working.
imager is not
working. [85]
The bottle The indexer cover is Replace or reseat the indexer cover.
indexer cover missing or not seated
is missing or properly.
WARNING
not seated
properly. [91] Failure to address an indexer not seated alarm may lead to
incorrect results.
The The instrument could not 1. Remove the conveyor cover of the unit that is affected.
instrument confirm that the bottle 2. Remove any obstruction and ensure the bottle is upright.
could not was successfully
3. If the bottle cannot be located, remove the indexer cover on the transferring
confirm that transferred to the next
instrument, remove any obstruction, and replace the cover.
the bottle was instrument.
4. If the issue is not resolved, contact your local bioMérieux representative for
successfully
assistance.
transferred to
the next
instrument.
[93]
One of the One of the instruments Contact your local bioMérieux representative for assistance.
instruments in in the bank has been
the Bank has configured incorrectly.
been
configured
incorrectly.
[95]
The transfer The transfer gate is open 1. Remove the conveyor cover for the alarming instrument.
gate is open and must be closed to 2. Remove any obstruction and ensure the gate is closed.
and must be continue. See the
closed to Transfer Gate figure in
continue. [96] BACT/ALERT VIRTUO
Hardware.
An An unexpected bottle 1. Reload any bottles that were unloaded due to alarm 97 as soon as possible.
unexpected has been discovered 2. If the system has been restored with a System Image, contact your local
bottle has and unloaded by the bioMérieux representative for assistance
been instrument.
.
discovered
The instrument has
and unloaded
unloaded a bottle that it
by the
has determined that the
instrument.
PC software does not
[97]
have knowledge of. This
can happen as a result
of the PC being restored
with a System Image but
no data backup or a
duplicate bottle loaded
across instruments. The
SW thinks the duplicate
is a reload of the original
bottle.
Note: If the instrument is configured for user request unload mode and there is a
communication loss for more than 15 minutes, the instrument will automatically
unload any final positive bottles to the chutes and any negative bottles to the
waste container. When communication is reestablished, the instrument will revert
back to user request unload mode.
Note: If the instrument is configured for user request unload mode and there is a
communication loss for more than 15 minutes, the instrument will automatically
unload any final positive bottles to the chutes and any negative bottles to the
waste container. When communication is reestablished, the instrument will revert
back to user request unload mode.
® ®
MYLA A MYLA general software error Contact your local bioMérieux representative for assistance.
general occurred (site configuration
software only).
error. [30]
® ®
Communicati The MYLA service was 1. Wait 15 minutes for MYLA to fully initialize.
on has been recently restarted (site 2.
®
The alarm will automatically resolve itself once MYLA is fully started.
lost with configuration only).
®
3. If the alarm does not resolve, reboot the internal PC. See Rebooting the
MYLA . [31]
Internal PC (Large Display).
4. If the problem persists, contact your local bioMérieux representative.
Purge old Purge old data has not recently 1. Perform a manual backup to trigger another purge. See Performing a
data has not completed successfully. Manual Data Backup.
recently 2. If the problem persists, contact your local bioMérieux representative for
completed assistance.
successfully.
[38] Note: The impact of not addressing the alarm can result in slow
PC performance, files growing too large, and backup
failures.
Scheduled Scheduled system backup 1. Select My Computer from the Start menu.
backup failed. failed. 2. Verify a USB device is installed on the PC and is mapped as drive F.
[44]
Note: If the system is configured for a network backup, verify
the network backup location is available.
Messages A message from the instrument Contact your local bioMérieux representative for assistance.
failed failed validation and was
validation and rejected by the PC.
were
rejected. [45]
An An unexpected software error Contact your local bioMérieux representative for assistance.
unexpected has occurred.
software error
has occurred.
[46]
Time is out of The time on the PC was Contact your local bioMérieux representative for assistance.
synchronizati changed. This does not affect
on with the the test time of bottles under
PC in the test. It could affect the test time
listed of bottles reloaded after
instrument(s). changing the PC time.
[49]
An An unexpected software error 1. Contact your local bioMérieux representative for assistance.
unexpected occurred where the software 2. Dismiss the alarm.
database found a discrepancy in the
integrity error record.
has occurred.
An unexpected database Contact your local bioMérieux representative for assistance.
[55]
integrity error has occurred.
Firmware Firmware version is • If updating the system, finish updating all connected firmware and the alarm
should automatically resolve.
version is incompatible with software
• If the alarm is not resolved or no upgrade has taken place, collect all
incompatible version. The PC software has software and firmware logs and contact your local bioMérieux representative
with software detected that a connected for assistance.
version. [98] instrument is using an
incompatible version of
firmware. This can happen
during an upgrade, the software
is upgraded first and will throw
this alarm until all connected
firmware is upgraded.
Unexpected An unexpected error occurred in 1. Open BCI CONNECT and review the error(s).
error the BCI CONNECT software. 2. Contact your local bioMérieux representative for assistance.
occurred in
BCI
CONNECT.
[103]
Archive Archive failed. Verify that the archive path is valid and writable.
failed. [104]
Communicati An unexpected error occurred in 1. Open the BCI CONNECT software and review the error(s).
on has been the BCI CONNECT software. 2. Contact your local bioMérieux representative for assistance.
lost with BCI
CONNECT.
[105]
A bottle has A bottle has been matched to a 1. Verify that the bottle has been matched to the correct test order using the
been potentially incorrect test order. accession number.
matched to a The test order may be outdated 2. If the test order is incorrect, unload the bottle and apply the appropriate
potentially or the bottle may have been accession barcode.
incorrect test loaded more than 30 days ago.
order. [106]
Bottle(s) were Bottle data could not be sent to 1. Navigate to the bottle record(s) and send the bottle data again.
unable to be the BCI CONNECT software. 2. If the problem persists, contact your local bioMérieux representative for
sent to BCI assistance.
CONNECT.
[107]
Notification A Notification Center software 1. Contact your local bioMérieux representative for assistnace.
Center error occurred (standalone 2. Dismiss the alarm.
Software configuration only).
Error [111]
Condition Resolution
Unloading indicator does not Follow the instructions in Performing System Maintenance for shutting down the software.
resolve after bottles have
been unloaded.
Bottles do not load or unload 1. Remove the indexer cover, and inspect the bottle indexer and conveyor for obstructions.
as expected. 2. Install the indexer cover.
3. If step 1 does not work, follow the instructions for Unjamming the Robot.
Robot Activity stays on the 1. Remove the indexer cover, and inspect the bottle indexer and conveyor for obstructions.
Small Display more than three 2. Install the indexer cover.
minutes.
3. If step 1 does not work, follow the instructions for Unjamming the Robot.
The conveyor belt does not 1. Remove the indexer cover, and inspect the bottle indexer and conveyor for obstructions.
start or stop. 2. Install the indexer cover.
3. If step 1 does not work, follow the instructions for Unjamming the Robot.
The indexer wheel does not 1. Remove the indexer cover, and inspect the bottle indexer and conveyor for obstructions.
move. 2. Install the indexer cover.
3. If step 1 does not work, follow the instructions for Unjamming the Robot.
The conveyor cover is missing [94] alert will occur when the cover is removed. Dismiss
this from the large display of the A unit.
2. Load bottles directly onto the conveyor as you would for an A unit and ensure bottles
are moving to the Indexer.
3. Replace the conveyor cover when bottles have been successfully loaded.
WARNING
Not replacing the conveyor cover after loading bottles may cause
delayed results.
Do not open the door, including the blue door handle cover above the conveyor, if the Robot
Activity indicator appears on the Small Display screen or when the Loading or Unloading
indicators appear on the Home screen. If the Robot Activity indicator remains on the Small
Display screen for 3 to 5 minutes, reboot the instrument. See Rebooting the Instrument
(Small Display).
WARNING
Opening the door when the Robot Activity icon appears on the Small Display
screen creates the potential of a crush hazard.
Note: The robotics and the agitation mechanism are disabled while the door is open.
WARNING
Step on the power supply cover plate when accessing the inside of the
instrument if necessary to achieve more height. Do not step on a separate
device, as the consequence may be a crush hazard.
WARNING
Flashing lights are used inside the instrument, which may not be suitable for
photosensitive epilepsy.
Note: Instrument damage may result and patient test results may be lost or negatively affected if
the instrument door is not opened and closed properly.
Note: The best way to circumvent false positives after the door has been opened is to keep the
door closed for 3.5 hours before reopening. This ensures a stable temperature environment
after the door has been opened. A conservative bottle loading workflow after a door open
event would be to limit loading to less than 15 bottles every 10 minutes for the 3.5 hours after
the door is closed. After 3.5 hours, the user can return to their normal workflow. The open
door and batch loading guidelines above are intended to help reduce false positives after
instrument interventions. Due to the number of variables involved, each circumstance will be
different therefore false positive results and subculture recommendations should be
thoroughly evaluated.
To minimize the potential of a false positive or a subculture recommendation, do not open the
door, including the door latch above the conveyor, if the Robot Activity icon appears on the
Small Display screen or if the Loading or Unloading indicators appear on the Home screen.
Note: The robotics and the agitation mechanism are disabled while the door is open.
The Do Not Enter icon will appear on the main status screen while the robot returns to
its home position.
Note: If the Entry Allowed icon does not appear after 5 minutes, call bioMérieux customer
service.
3. If the instrument is a B unit, grasp the conveyor cover and gently push toward the
instrument, and then push upward to release the conveyor cover from the standoffs.
1 Tabs
2 Conveyor Cover
3 Standoffs
4. Pull backward gently to release the conveyor cover from the tabs securing it to the
instrument.
5. On the grooved area on the door latch cover, press and release to open the door latch
cover.
1 Grooved Area
6. Fold the door latch cover down to access the door latch.
7. Before opening a B unit, ensure that the transfer arm of the instrument to left is not
extended into the B unit.
1 Correct position
2 Incorrect position
8. Grasp the door latch and swing it outward and to the right to the unlocked position.
IMPORTANT: The door latch is in the locked position when it is flush with the instrument. The
door latch is in the unlocked position when it is perpendicular to the
instrument. Do not pull on the instrument door when the door latch is in the
locked position.
CAUTION: The door latch should be used to pull the door outward. Do
not use the sides of the instrument to pull the door of the instrument
forward. Failure to comply may result in damage to the exterior surface
of the instrument.
WARNING
Instrument damage may result and patient test results may be lost or
negatively affected if the instrument door is not opened and closed properly.
Note: The robotics and the agitation mechanism are disabled while the door is open.
WARNING
Failure to place the robot gripper in the safe to move position prevents robot
automation, which could delay results. For information on the safe to move
sensor, see BACT/ALERT VIRTUO Hardware.
1. Place the gripper mechanism in the safe to move position, ensuring the gripper jaws
are horizontal and do not extend beyond the end of the bracket.
2. Using the door latch, swing the door inward to position the door directly in front of the
instrument.
CAUTION: The door latch should be used to swing the door inward. Do
not use the sides of the instrument to push the door of the instrument
forward. Failure to comply may result in damage to the exterior surface
of the instrument.
WARNING
If the door is not fully secured, the consequences may be delayed
results.
Note: Once the door is squared with the chassis, it is acceptable to push on the top sides of
the instrument to close the door. Do not push on the large display screen or small
display screen when moving the door to the closed position. Failure to comply may
result in damage to the exterior surface of the instrument.
The kickstand will raise and stow inside the door when the door is pushed forward to
the fully closed position.
4. Ensure the door is fully aligned against the instrument.
5. While pushing firmly on the top of the instrument above the Bank Display, grasp the
door latch and push it flush against the instrument to the locked position.
Note: The door latch is in the locked position when it is flush with the instrument. The door
latch is in the unlocked position when it is perpendicular to the instrument. Do not pull
on the instrument door when the door latch is in the locked position.
1 Grooved Area
7. Address any alarms or messages related to the retrieved bottles.
Note: If the alarm persists after two attempts, contact your local bioMérieux representative
for assistance.
8. Attempt this recovery a second time if the alarm remains after completing all of the
steps.
9. Replace the conveyor cover.
10. Ensure the red dot from the conveyor cover sensor is visible on the right tab.
11. If the red dot from the conveyor cover sensor is not visible, reposition the conveyor
cover until the red dot is visible.
Note: Instrument damage may result and patient test results may be lost or negatively affected if
the instrument door is not opened and closed properly.
The robotics and the agitation mechanism are disabled while the door is open.
WARNING
When the instrument door is open, take care of the critical laser component
required for robotic alignment. Do not grasp, pull on, or impact the laser. Do
not touch the laser optics window.
3. Pull the gripper and the bottle away from the cell.
4. Gently move the bottle from side to side to ease it out of the gripper.
Note: A bottle should not be in the gripper when the door is closed.
7. If any bottles or calibration standards are on the floor of the instrument, remove them.
8. Move the robot to the left and then to the right to look for bottles on both sides.
9. Place bottles on the loading station conveyor belt to be processed by the system.
10. Verify that four calibration standards are present in their reserved cells. CAL 1 must be
in cell P24, CAL 2 must be in cell P25, CAL 3 must be in cell P26, and CAL 4 must be
in cell P27.
Note: Calibration standards must always be located in cells P24, P25, P26, and P27.
1 CAL1
2 CAL2
3 CAL3
4 CAL4
11. Find missing calibration standards located elsewhere in the incubator and move them
to correct cells.
12. Place the gripper mechanism in its safe to move position, ensuring the gripper jaws
are horizontal and do not extend beyond the end of the bracket.
13. Check that the green LED turns on when in the proper position.
3 Gripper Assembly
Audit Trail
The Audit Trail contains auditable actions that affect bottle test results such as user
modifications to records and system configuration.
Access system status information while a bottle is under test.
Figure 156: Audit Trail Screen for Standalone and Site Configuration (MAESTRIA™)
Menu Access the audit trail about screen and sign out.
Filter Description
Date Range (Start Date - End Date) Select a date range for which to display Audit
Trail records.
Record State Select a record state for which to display
Audit Trail records.
Data Content Type Select a data content type for which to
display Audit Trail records.
User Select a user for which to display Audit Trail
records.
System Select a system name for which to display
Audit Trail records.
Action Select an action for which to display Audit
Trail records.
The available events can include:
• Auto logoff
• Create
• Delete
• Login
• Logout
• Update
Record Details
Highlight a record on the Audit Trail screen. The record appears in the Record Details screen.
Note: Select the All radio button to select all of the displayed Audit Trail records for export.
1 Filter area
Filter Description
Date range (from - to) Select a date range for which to display Audit Trail records.
Event type Select an event for which to display Audit Trail records.
The available events may include auto logoff, create, delete,
login, logout, or update.
Audit type Select an audit type for which to display Audit Trail records.
The available audit types depend on the connected systems
and instruments and may include alarm or bottle.
System name Select a system name type for which to display Audit Trail
records.
Identifier Enter text to search for available identifiers.
Expand to display or collapse to hide further search criteria
for the display of Audit Trail records.
Advanced Criteria
Record type Select a record type for which to display Audit Trail records.
Available record types are:
• Auditchange - electronic data change that must be
audited for 21CFR11 compliance.
• Auditinformation - Additional information of an audit entry
for support and investigation purposes.
• Auditlogin - Login audit entry.
Audit hierarchy Select an audit hierarchy type for which to display Audit Trail
records.
System ID Select a system ID for which to display Audit Trail records.
Filter Description
Audit hierarchy ID Enter text to search for available audit hierarchy IDs.
User ID Enter text to search available user IDs.
Audit details Enter text to search available audit details.
Search Select to display Audit Trail records according to the filter
settings.
Clear Filter Select to reset all filter settings.
Column header Select a column header to sort the records according to the
column value (by date, alphabet, or number).
Select to sort the records in descending order (by date,
alphabet, or number).
Select to sort the records in ascending order (by date,
alphabet, or number).
Date Creation date and time of the Audit Trail record.
Record type Record type of the Audit Trail record.
Event type type Event type for the Audit Trail record.
System name System name for the Audit Trail record.
Audit type Audit type of the Audit Trail record.
Audit ID Audit ID of the Audit Trail record.
Audit hierarchy Audit hierarchy of the Audit Trail record.
Audit hierarchy ID Audit hierarchy ID of the Audit Trail record.
User ID User ID related with the Audit Trail record.
Select to open the Hierarchical View menu for the record.
Record Details
Highlight a record on the Audit Trail screen. The record appears in the Record Details section.
1. Access the external device (USB device, network drive) and open the exported file in a
program such as Excel (.csv) or Adobe Acrobat (.pdf).
2. To print, use the print function provided by the appropriate program
Note: Refer to the bioMérieux User Management User Manual applicable to the system
configuration.
Collect Logs
The BACT/ALERT® VIRTUO® logs can be collected and provided to bioMérieux support for
issue investigation.
Note: The collected logs may contain patient data, so a password may be entered in order
to encrypt the log files. If an encryption password is entered, then the password must
be provided to bioMérieux support so that the log files can be accessed.
4. If the log files needs to be encrypted, enter a password and select OK.
If the log files do not need to be encrypted, do not enter a password and select
Cancel.
The command will continue to run for several minutes. Do not close the command
screen. When the log collection is complete, a dialog box displays indicating the file is
on the Desktop.
5. Select OK.
The command screen closes. The VIRTUO_logs_YYYYMMDDHHMMSS.zip folder is
on the Desktop and can be copied to a USB to be used by bioMérieux support. If an
encryption password was entered in the previous step, provide the password to
bioMérieux support.
The system can notify the user through email when positive bottles or alarms occur on the
system.
Remote Notification must be enabled in General Software Configuration in order to access
the Notification Center.
The following events are available for the BACT/ALERT® VIRTUO® system:
1 Conjunction
2 Criterion name
3 Operator
4 Value
5 Delete
Event Criteria
Positive Bottle The criteria corresponds to the
information available for the positive
bottle.
• System: Corresponds to the system
currently incubating the bottle
• Bank Name
• Bottle ID
• Accession ID: Corresponds to the Lab
ID
• Instrument Name
• Instrument Serial Number
• Sample Type
• Bottle Result Date
• Generated Date and Time
• Bottle Type
• Time to Detection: Corresponds to the
duraction between load date and time
and positive date and time
• Load Date and Time
• Contains
• Does not contain
10. Enter the Value associated to the criteria.
11. Select APPLY CRITERIA to save the criteria.
The applied criteria displays in the Summary pane.
The figure below is an example of a Positive Bottle event, where an email is sent each
time a bottle becomes positive that is incubating in a System named VIRTUO or if the
bottle is incubating in a bank with a Bank Name that contains the word Bank.
The figure below is an example of an Alarm Event, where an email is sent each time a
new alarm is generated on an instrument that is named INSTRUMENT1 or
INSTRUMENT2.
12. Optional: Add additional criteria that are separated by the conjuctions OR or AND.
15. Select the Send notification to Mailing List toggle to activate the email notification.
16. Define the following for the email notification:
• Mailing List
• Subject of the email
• Body of the email
17. Select Finish to finalize the rule creation.
The rule displays on the RULES home screen.
Refer to the Notification Center Software User Manual for more information.
A false positive is a bottle that flags positive on the instrument, but the Gram stain and
subculture are negative. All bottles that are flagged positive by the instrument must have a
Gram stain and subculture performed.
Investigation into the cause of false positive rates for BACT/ALERT® VIRTUO® systems have
resulted in a list of best practices that, if performed, will reduce the occurrence of false
positives. This chapter is divided into two main sections: preventing false positives by the user
and causes of false positives that are due to the instrument. Detailed descriptions of
conditions that may cause false positives are provided, along with best practices and
recommendations to prevent these conditions.
Preventing False Positives - User
This section describes conditions that may cause false positives that can be prevented by the
user.
Figure 176: Cell Graph Example - High Amount of White Blood Cells
Best Practice #1 When no organisms are seen on the Gram stain, it is appropriate to reload
the bottle on the instrument. If the system determines that the bottle does not support a
positive result after the positive bottle is reloaded, then the status will revert to Negative-to-
Date and continue incubating for the duraction of the Maximum Test Time. See Handling
Unconfirmed Positive Bottles and Reloading Positive Bottles.
Best Practice #2 Do not fill the bottles with more than the volume recommended in the
Instructions for Use. See the Instructions for Use for each bottle type to determine the
recommended volume.
CAUTION: Do not fill the bottles with more than the recommended volume of
sample.
2. Failure to close the door properly. See Opening the Instrument for Internal Access and
Closing the Door.
3. Positioning the instrument under an air conditioning or heating vent.
4. Heating or cooling differences in the lab. For example, the lab technicians on a different
shift turn up or down the heat or air conditioning, there are open windows in the
laboratory, or there are drafts that are directed at the instrument. The ambient room
temperature must be at least 9° cooler than the instrument temperature set-point
(usually 35-37°C).
5. By-passing the A unit conveyor when loading bottles and placing them directly on to a B
unit. Bottle loading should originate from the A unit and move to downstream B units to
evenly disperse the bottle across all units. Loading bottles in one area can cause a large
heat transfer to occur and cause false positives.
6. Not following the correct steps or door open recommendations when performing
Emergency Manual Bottle Loading and Unloading. The Emergency Manual Bottle
Loading and Unloading procedure should only be performed in the event that one or
more of the identified hardware failures listed in the procedure has occurred on the
instrument and there is an urgent need.
Best Practice #4 Ensure that temperature changes in the laboratory are minimized. Under
normal operation, the instrument door is not opened. To minimize the potential of false
positive determinitations or subculture recommendations, follow these temperature guidelines
when opening the instrument door:
• Do not keep the instrument door open more than 10 minutes.
• If the door has been open for less than 10 minutes, ensure the door is closed for a
minimum of 10 minutes before reopening the door.
• If the door has been open for more than 10 minutes, then the door should be closed as
soon as possible and remain closed for a minimum of 40 minutes before reopening the
door.
• The best way to avoid false positives after the door has been opened is to keep the door
closed for 3.5 hours before reopening. This ensures a stable temperature environment
after the door has been opened.
IMPORTANT: Bottle reflectance and output of electronics varies with temperature. Limiting
temperature fluctuations reduces the false positive rate.
This section describes causes of false positives that are due to the instrument.
Cell Noise
Cells may exhibit electronic "noise" that is visible in the graph. There are three main types of
instrument noise:
1. The first type of noise is when cell graphs show irregular fluctuating signals and sharp
peaks and valleys, or large spikes in the graph. Contact you bioMérieux representative
to help determine if there are electronic issues that may be isolated and corrected.
Electro-mechanical noise above a certain threshold should be detected by the
instrument. A Bottle Placement Failure alarm, code [10], will occur if this type of noise is
detected. Every 6 hours, bottle readings are evaluated for electrical noise in the data.
This window will start over if the instrument is rebooted, if there was a gap in the data, or
if the bottle was reloaded. If electrical noise is sensed by the instrument while a bottle is
loaded into the cell, an error code [10] will be flagged for the affected cell, and the cell
will be made unavailable. If Bottle Placement Failure Alarms are triggered, verify that the
cells where the alarms are occurring as they may display a pattern. Some examples may
include: if the electro-mechanical noise issues are specific to a particular rack or if
electro-mechanical noise is infrequent.
2. The second type of noise reverts to a normal graph when the bottle is reloaded into
another location. Contact your bioMérieux representative if this occurs.
3. A third type of cell noise involves a characteristic signal pattern in the instrument that
repeats (sometimes simultaneously in multiple cells) and can cycle on a daily basis at
particulat times. It can correlate to particular events such as temperature, lighting, or
facility power events. These cyclic blips trigger a false positive event. Verify that there is
a dedicated circuit for the instrument. For example, verify that there is not a centrifuge or
a refrigerator on the same circuit as the instrument. Ensure that there are no voltage
drops at the power outlet. Also follow Best Practice #4 to minimize temperature
fluctuations.
WARNING
When performing maintenance on a BACT/ALERT® VIRTUO® system that is
processing bottles in the racks, the door can be opened once for up to one
hour while performing maintenance and/or to decontaminate the instrument if
needed. The door cannot be opened again for a minimum of two hours to
allow the system to get back to the proper temperature. Multiple openings of
the door for short periods of time can cause issues with the temperature
controls inside the incubator. Not following these instructions can lead to
delayed results.
Only your local bioMérieux representative can shut down the instrument. Users can reboot
the firmware using the Small Display.
CAUTION: Rebooting the instrument in any manner other than the order
listed in this procedure could result in a malfunction of the instrument.
Note: Before rebooting the instrument, ensure the bottles have been unloaded from the bottle
retrieval area.
1. Ensure the power cord is securely connected to the power cord connector. The power
cord and power switch are located at the bottom front of the instrument behind the
blue kick plate.
Note: The software, firmware, and Internal PC operating system will initialize.
IMPORTANT: A screen may appear with a message that states that a valid instrument bank
must be selected. For stand-alone configurations, it is only required to press
the refresh button.
Restoring Data
CAUTION: The following data will be lost when you attempt to restore data.
• All edits (such as the following) that have been done since the backup was
made will be lost:
◦ Anonymous bottles that were identified
◦ Bottles where the accession ID was changed
◦ Positive results that were user-changed
• The instrument temperature data since the date of the backup being
restored will be lost.
• In standalone mode, the following will be lost:
◦ All audits that were made since the backup
◦ Any users created in the bioMérieux User Management System since
the backup
Note: Audits that were made since the backup and related to instrument data will be rebuilt. The
PC and firmware can restore some of the data and send to the PC (for example, bottle loads
and unloads). The firmware only stores up to 7 days of this data, however.
Automated Backups
Cobian Backup is the software used to backup the BACT/ALERT® VIRTUO® data to a
dedicated hard disk, an external USB drive, or a network directory.
Every day, the Cobian Backup software is configured to systematically back up all the
BACT/ALERT® VIRTUO® data:
• It produces a backup file every Monday, Wednesday and Friday.
• It produces another backup file every Tuesday, Thursday and Saturday.
• It produces a final backup file every Sunday.
This backup method ensures all the data for a week are backed up.
Note: Ensure that an external USB device is inserted in the PC for scheduled automated
backups.
1. Ensure that an external USB drive is inserted, assigned to the F drive, and available
for the manual backup.
2. Select Expand in the system notification tray to show the hidden icons.
3. Right-click Cobian.
4. Select Open.
5. Select Backup Manual.
8. After the Cobian manual backup is completed, Backup done appears on the Backup-
Manual Log screen.
Disinfection Procedures
WARNING
A potential biohazard exists in these situations:
• A bottle is not correctly identified.
• A bottle leaks from damage or another cause.
• Entrapped biohazard substances are not properly cleaned.
• Proper cleaning guidelines are not correctly followed.
• Surfaces are not cleaned properly.
Bottle Leaks
WARNING
A consequence of bottle leaks may be delayed or false positive results.
Note: Refer to the product labeling for any cleaning or disinfection agents that may be used to
ensure proper handling of any potentially caustic substances.
• 10% sodium hypochlorite (bleach) of 700 PPM free chlorine solution (13 mL of 5.25%
bleach to 1 liter of tap water)
• 3–25% hydrogen peroxide (wipe down)
• hydrogen peroxide (vapor phase)
• EPA registered tuberculocidal disinfectants (safe for use with polycarbonate)
WARNING
Use only polycarbonate-safe agents to avoid the risk of leaking bottles or
incorrect results.
WARNING
Do not use compounds that will harm polycarbonate (for example
CIDEXPLUS® 28 Day solution) to clean the rack cover plate. Failure to comply
results in damage to the rack cover plate.
Note: When performing tuberculocidal decontamination, some cleaning agents may cause
discoloration on instrument surfaces over time.
Cleaning a Spill
Where appropriate, follow your laboratory's spill cleanup procedures.
1. Gently cover the spill area with a paper towel. Apply a 10% bleach solution or other
EPA registered tuberculocidal disinfectant.
2. Using the bleach solution, wet down all surfaces where the spill may have come in
contact.
3. Allow all surfaces adequate contact time with the bleach solution (15–30 minutes)
before cleanup.
4. Treat all materials used in the cleanup as biohazardous waste.
WARNING
Always wear protective gloves when handling blood, blood components, or
bloodcontaminated material.
WARNING
Handle samples and inoculated culture bottles as though capable of
transmitting infectious agents. Autoclave all inoculated bottles, sample
collection needles, and blood drawing devices before discarding.
WARNING
The potential exists for bottle damage and leaking. Damaged and leaking
bottles may cause delayed or false positive test results.
2. After decontamination, wipe with damp (water only) towel and thoroughly dry.
CAUTION: Do not keep the door open for more than 10 minutes as that
may result in temperature loss and affect bottle test results.
2. Visually inspect the extent of the leakage or spill. Determine if one or more racks is
contaminated.
3. Remove the leaking bottle if possible.
Note: If a bottle becomes lodged in the cell, call bioMérieux. Do not try to dislodge a bottle
by pulling on the rack. You may need to subculture the bottle.
Note: Do not reload bottles affected by the spill until they have been decontaminated.
Follow your institution's recommended procedure for decontamination or the
procedure described in the latest revision of Clinical and Laboratory Standards
Institute (CLSI® guideline "Protection of Laboratory Workers from Instrument
Biohazards and Infectious Disease Transmitted by Blood, Body fluids, and Tissue”
CLSI® Document M29-A.
7. If an affected cell contains a large amount of liquid, carefully aspirate it using a pipette
or similar device, and dispose of it in an appropriate biohazardous waste container.
8. If the spill is confined to one or a few cells of one rack, clean and disinfect the affected
cells with an approved disinfection agent, following the steps in Cleaning the Bottle
Racks.
WARNING
Do not use compounds that will harm polycarbonate (for example
CIDEXPLUS® 28 Day solution) to clean the rack cover plate. Failure to
comply results in damage to the rack cover plate.
Note: For more extensive spills, decontaminate bottles, cells, and racks. Contact your local
bioMérieux representative for instructions to relocate or disable affected cells or
racks.
9. Although a plate covers the cables, verify that the leak did not contaminate the cables
at the bottom of the instrument. If the spill leaked onto the cables, contact your local
bioMérieux representative.
10. Close the instrument door.
Clean the robot gripper pads when they are visibly dirty or when instructed by a bioMérieux
representative. Wipe the gripper pads using gauze and the approved cleaning agent to
remove dirt, debris, adhesive buildup, or contaminants.
Dampen the robot gripper pads when cleaning. Do not soak the pads.
Clean the loader conveyor belt surface when it is visibly dirty or when bottles start tipping
over. Use silicone wipes (418007) to wipe the conveyor surface.
1. Wipe the conveyor surface that is visible with silicone wipes (418007) to lubricate.
2. Place a hand above the conveyor to activate the light curtain sensors.
The conveyor advances forward.
3. Wipe this section of the conveyor surface that is visible.
4. Continue in this manner until the entire conveyor is lubricated.
WARNING
Operating the instrument with any of the outer instrument panels removed
may lead to incorrect results.
Clean the outer instrument panels when they are visibly dirty or when directed by a
bioMérieux respresentative.
Wipe the outer instrument panels with gauze and an approved cleaning agent, except
hydrogen peroxide (vapor phase). See Approved Disinfection Agents.
Before removing a bottle rack for cleaning, unload all bottles from the bottle rack.
1. Locate the bottle rack fasteners (screws) below bottle cells 2, 7, and 12.
1 Screws
WARNING
Carefully put the bottle rack assembly back in the instrument so as not
to damage the bottle rack. The consequences of damaging or
distorting the bottle rack may be delayed results.
Clean the rack cover plate if it is visibly dirty or when instructed by a bioMérieux
representative.
1. Remove the bottle rack following the steps in Removing a Bottle Rack Cover Plate.
2. Wipe the rack cover plate with gauze and a cleaning agent.
See Approved Disinfection Agents.
WARNING
Do not use compounds that will harm polycarbonate (for example
CIDEXPLUS® 28 Day solution) to clean the rack cover plate. Failure to
comply results in damage to the rack cover plate.
Note: Saturate the gauze but not so much that liquid drips.
3. Wipe the rack cover plate with gauze and distilled water.
Note: Saturate the gauze but not so much that liquid drips.
4. Use a dry gauze to lightly remove any residual water spots on the rack cover plate.
5. Reinstall the bottle rack assembly over the rack cover plate.
6. Tighten the three bottle rack assembly fasteners, ensuring proper orientation.
WARNING
The consequences of not tightening the screws may be delayed
results.
WARNING
Carefully put the bottle rack assembly back in the instrument so as not
to damage the bottle rack. The consequences of damaging or
distorting the bottle rack may be delayed results.
WARNING
Carefully place the bottle rack assembly back in the instrument so as
not to damage the laser alignment features. The consequences of
damaging or distorting the laser alignment may be delayed results.
Clean the bottle racks when they are visibly dirty or when instructed by a bioMérieux
representative.
Note: If the rack cleaning takes longer than two hours, some of the bottles will be marked for
subculture when they are reloaded. If the rack cleaning takes longer than four hours, all the
reloaded bottles will be marked for subculture.
Note: Follow your institution's recommended procedure for decontamination or the procedure
described in the latest revision of Clinical and Laboratory Standards Institute (CLSI®)
guideline "Protection of Laboratory Workers from Instrument Biohazards and Infectious
Disease Transmitted by Blood, Body fluids, and Tissue” CLSI® Document M29-A.
WARNING
Do not use compounds that will harm polycarbonate (for example
CIDEXPLUS® 28 Day solution) to clean the rack cover plate. Failure to comply
results in damage to the rack cover plate.
CAUTION: Do not expose the cell or rack to the approved disinfection agent
for an extended period. Do not use an unapproved disinfectant or a bleach
solution stronger than 10% or damage to cell components may occur.
WARNING
Failure to properly adhere to the bottle rack cover plate cleaning instructions
may lead to delayed results.
1. Remove the bottle rack following the steps in Removing a Bottle Rack Cover Plate.
2. Wipe the exterior and interior cells regions with gauze and recommended cleaning
agents. See Approved Disinfection Agents
Note: Saturate the gauze but not so much that liquid drips.
5. Reinstall the bottle rack assembly over the rack cover plate.
6. Tighten the three bottle rack assembly fasteners, ensuring proper orientation.
WARNING
The consequences of not tightening the screws may be delayed
results.
WARNING
Carefully put the bottle rack assembly back in the instrument so as not
to damage the bottle rack. The consequences of damaging or
distorting the bottle rack may be delayed results.
CAUTION: The bottle rack assembly may fall or prevent proper robot
alignment if not reinstalled correctly.
WARNING
Carefully place the bottle rack assembly back in the instrument so as
not to damage the laser alignment features. The consequences of
damaging or distorting the laser alignment may be delayed results.
To decommission an instrument:
1. Follow your institution's recommended procedure for decontamination or the
procedure described in the latest revision of Clinical and Laboratory Standards
Institute (CLSI®) guideline "Protection of Laboratory Workers from Instrument
Biohazards and Infectious Disease Transmitted by Blood, Body fluids, and Tissue ”
CLSI® Document M29-A.
WARNING
A potential biohazard exists if proper cleaning guidelines are not
correctly followed.
Generic Bottle ID
A bottle ID that does not indicate one of the predefined bottle types.
Handheld Barcode Scanner
Scanner that is plugged into the instrument and scans data from a barcode into the software
and displays information for a bottle.
Internal Barcode Scanner
Scanner that is plugged into the instrument and scans data from a barcode into the software
and displays information for a bottle.
Laboratory Information System (LIS)/Blood Establishment Computer System (BECS)
A customer maintained computer (usually a mainframe) or computerized system that collects
information from laboratory instruments.
Large Display
The large touchscreen that provides bottle and system information for the associated banks.
Loading Area
Automatically loads from one to 40 bottles into the instrument at a time.
Negative
Describes a culture bottle that has no evidence of microbial activity.
Negative to Date
Describes a culture bottle that is under test and a final result has not yet been determined.
Positive
Describes a culture bottle that has evidence of microbial activity.
Rack
Contains 27 cells, each of which can hold and monitor an individual culture bottle. Racks
gently rock to agitate blood culture bottles. The racks are designated with a letter from A to P.
Rack Detection Assembly (RDA)
Holds four racks and controls agitation (rocking).
Robot Arm
A gripper that retrieves bottles from the bottle indexer and automatically loads bottles to an
available cell and unloads bottles to either the waste container or the bottle return chutes
(depending on bottle status and how the instrument is configured).
Small Display
Small touchscreen that displays instrument temperature and provides a means of manually
loading and unloading bottles. This display is primarily for individual instrument status and
maintenance.
Stabilizing Feet
Feet that adjust to prevent movement.
Unconfirmed Positive
Describes a bottle the system determined to be positive but no confirmation is available by
gram stain or subculture.
Waste Container
The receptacle that holds all negative bottles that are automatically unloaded.
Release Part
Change Type Change Summary
Date Number
Safety Information:
Technical
2022-09 050574-02 • Instrument Labels: Replaced A and B unit instrument label figures.
change
• Safety Precautions: Changed powder-less to powder free gloves.
System Overview:
• Introduction to BACT/ALERT® VIRTUO®: Reword for clarity. Clarify
that the fill-to line is not etched on the bottle. Removed statement that
sample unloads to the waste container.
• Multi-Instance Site Configuration: Replaced figure. Removed system
version limitation.
• Primary Secondary Site Configuration: Replaced figure. Added and
removed limitations.
• Communication Loss: Replaced figure.
• Identifying Primary and Secondary Banks: Replaced figure and
Technical added callouts. Removed Secondary A Unit System Information
change figure.
• Bottles Supported for Use: Added BPA US and BPN US.
• Physical Characteristics: Added sound emission for loading bottles.
Clarified above sea level.
• Large Display: Replace table icons.
• Small Display: Reword for clarity. Removed wording below table.
• Buttons: Replaced figure.
• Common System Buttons: Replaced icons. Added calendar icon.
• Common Edit and Configure Screen Elements: Replaced figure.
• Search Criteria Screen Icons: Replaced icons. Added Group MTT
Edit icon.
Release Part
Change Type Change Summary
Date Number
Basic Functions:
• Introduction: Corrected records and data to information.
• Zone 1: Corrected User Security to User Management.
• Zone 3: Replaced icons.
• Display Bank Screen: Added Number of Available Cells.
• System Log In: Corrected User Security to User Management.
• Applying a Barcode Label to a Bottle: Added 2D barcodes. Removed
old bottle label. Replaced figure. Corrected fill-to line to fill mark.
Added Information for Use reference number. Added do not write on
bottom of bottle to warning.
• Loading Bottles Automatically: Modified warning to include
application. Replaced figures.
• Canceling a Bottle Test: Moved step 4. Replaced figures.
• Handling Unconfirmed Positive Bottles: Added guidance for reloading
positive bottles.
• Unloading Specific Bottles by User Request: Added caution.
• Searching for Specific Bottle Information: Replaced figures.
• Searching by Patient ID: Added note for deleting patient data.
• Searching by Bottle Type: Updated to reflect current interface.
Replaced figure.
• Searching by Sample Type: New section.
• Using Additional Search Criteria: Updated to reflect current
interface.Replaced figure.
• Searching by Instrument Name: Replaced figure.
Technical
• Search Results Screen: Replaced figure.
change
• Accessing the Bottle Status Information Screen: Replaced figure.
Added and replaced icons.
• Editing Bottle Record: Replaced figure.
• Editing Patient Information: Clarified note. Replaced figures.
• Processing Expired Bottles: Replaced figure.
• Identifying an Anonymouse Bottle with a Visible bioMérieux Label:
Replaced figure. Added note.
• Identifying an Anonymouse Bottle without a Visible bioMérieux Label:
Replaced figures. Added note.
• Unloading, Relabeling, and Reloading a Previouslt Identified
Anonymouse Botte: Replaced figure.
• Zooming and Rotating Bottle Label Image: New section.
• Generating Bottle Reports in Standalone and Site Configuration
(MAESTRIA™): Updated title to specify MAESTRIA™ site
configuration. Added Note.
• Generating Bottle Reports in Site Configuration (MYLA®): Updated
title to specify MYLA® site configuration. Updated to reflect current
interface.
• Generating an Hourly Temperature Detailed Report: Replaced
figures.Updated to reflect current interface.
• Generating a Daily Average Temperature Report: Replaced
figures.Updated to reflect current interface.
• Generating an Instrument Cell Calibration Report: Replaced
figure.Updated to reflect current interface.
Release Part
Change Type Change Summary
Date Number
System Configuration
• Updated to reflect current interface.
• Configuring Bottle Type Settings: Replaced figure. Updated to reflect
current interface.
• Sample Volume Measurement: Updated note for default
configuration.
• Setting the Sample Upper Variation: Updated to reflect current
interface.
• Configuring Bank Settings: Replaced figure.
• Naming Instruments: Replaced figure.
• Configuring Sound Settings: Added note. Replaced figure.
• Configuring Sound and Volume: Removed note. Added warning for
disabling alarms.
• Configuring Alarm Settings: Replaced figure. Updated to reflect
current interface.
• Configuring General Software Settings: Added Note. Replaced
figure.Updated to reflect current interface.
Technical • Using Patient Data: Updated note.
change
• Measuring Sample Volume: New section.
• Configuring Remote Notification: New section.
• Accessing the Notification Center Dashboard: New section.
• Archiving Bottle Data: Removed Important statement. Added note.
Updated to reflect current interface.
• Accessing the BCI Connect Dashboard: Updated to reflect current
interface.
• Enabling 21 CFR 11 Settings: Replaced figure.
• Configuring LIS Settings: Added standalone configuration only.
Replaced figure.Updated to reflect current interface.
• Uploading Bottle Results: Updated to reflect current interface.
• Setting Date Range: Added date range options.
• Adding Columns to a Report: Added minimum and maximum
requirements.
• Managing User Account and User Groups: Updated to reflect current
interface.
• Configuring External Users in User Management: New section.
Troubleshooting
• Updated the following bottle alarms and alerts: 3, 8, 14, 19, and 47.
• Updated the following instrument alarms and alerts: 7, 10, 20, 41, 59,
and 97.
• Updated the following software and communication alarms and alerts:
Technical 31 and 38. Added the following alarms and alerts: 108, 109, 110, and
change 111.
• Opening the Instrument for Internal Access: Added warning.
• Audit Trail: Created separate sections for Standalone and Site
Configuration (MAESTRIA™) and for Site Configuration (MYLA®).
• Collect Logs: New section.
• Notification Center: New section.
Technical
Best Practices for Preventing False Positives: New section.
change
Release Part
Change Type Change Summary
Date Number
Technical
Glossary: Added bioMérieux middleware.
change
Throughout:
• Added MAESTRIA™.
Administrative • Replaced MYLA® with bioMérieux middleware when referring to
MYLA® and MAESTRIA®.
• Changed customer to user.
General Information:
• Added Serious Incident statement.
Administrative • Warning and Safety Messages: Moved from Introduction to the
System to General Information.
• Intellectual Property: Corrected Virtual Patent Marking web address.
Standard Symbols:
Administrative
• Added Importer symbol and definition.
System Configuration:
Administrative • Sending Negative to Date Messages to the LIS upon Loading Bottles:
Added Negative to Date to the title.
Troubleshooting:
• Opening the Instrument Door: Reworded for clarity.
Administrative
• Closing the Door: Removed duplicate note.
• Unjamming the Robot: Reword for clarity.
System Overview:
• System Configuration: Updated definition of instrument bank and
available bank configurations.
• Multi-Instance Site Configuration: Added section and sub-sections.
• Primary/Secondary Site Configuration: Added section and sub-
sections.
• Bottles Supported for Use: Added BACT/ALERT® BPA and BPN
bottles for industry blood bank use. Updated warning about use of
non-bioMérieux and/or unsupported bottles to indicate that such
bottles are no longer rejected in R3.0.
• Additional Supplies: Removed MYLA® remote notification tower.
Technical • Principle of Detection: Edited information about CO2 production for
change accuracy. Updated warning to remove industry blood bank use as an
off label use of the system.
• BACT/ALERT® VIRTUO® Hardware: Updated call-outs for A-Unit and
B-Unit "Front View" figures.
• BACT/ALERT® VIRTUO® Screens: Added note about untranslated
text in user interface.
• Large Display: Added note about slow PC performance.
• Small Display: Added note about system power loss or reboot.
Updated "Manual Load Bottle" description. Added "Ready to Load"
icon and description. Updated "Bottle Return Chutes" figure.
• Search Criteria Screen Icons: Updated "Upload" icon description to
include LIS.
Basic Functions:
• Zone 2: Added note about the Snooze button.
• Zone 5: Updated to reflect user interface in R3.0.
• Remote Notification: Removed section because it pertains to MYLA®
• Applying a Barcode Label to a Bottle: Added "Correct Label
Placement" figures for old and new bottle labels.
• Loading Bottles Automatically: Updated warning about use of
unauthorized bottles to indicate that test result for such bottles are no
longer suppressed in R3.0. Added warning about removing index
cover. Added note about poor quality bottle images. Updated
"Example of Scanned Bottle Image" figure to show new bottle label.
• Unloading Positive Bottles Automatically: Moved warning about
subculturing negative to date bottles to "Unloading Bottles by User
Request" section.
Technical • Handling Unconfirmed Positive Bottles: Updated warning about
change reloading bottles across banks with information about Primary/
Secondary.
• Data Retention Period for Bottle Information: Updated to reflect
functionality in R3.0.
• Searching for Specific Bottle Information: Updated sub-sections,
including figures, to reflect search functionality in R3.0.
• Searching by Patient ID: Added sub-section to "Searching for Specific
Bottle Information" section.
• Search Results Screen: Added explanation of bottle location. Added
notes about search criteria and Page Scroll buttons.
• Accessing the Bottle Status Information Screen: Updated "Upload"
icon description to include the LIS. Added "Revert" icon and
description. Added step to view patient demographic information in
site configuration.
• Editing a Bottle Record: Removed note about Revert icon.
Release Part
Change Type Change Summary
Date Number
System Configuration:
• Chapter introduction: Added new system configuration settings.
Moved warning about configuring accession IDs to "Enabling
Barcodes on Bottles and Defining Validation Patterns" sub-section.
• Sample Volume Measurement: Added note about enabling sample
volume measurement.
• Naming Instruments: Added "Location" column call-out to "Naming
Instruments" figure.
• Naming Banks and Labs: Updated important note about naming
banks.
• Configuring General Software Settings: Added new general software
Technical
settings.
change
• Configuring Data Retention and Accession ID Reuse: Added sub-
section to Configuring General Software Settings.
• Archiving Bottle Data: Added sub-section to Configuring General
Software Settings.
• Configuring LIS Settings: Added section and sub-sections.
• Configuring Bottle Report Settings: Added section and sub-sections.
• Configuring LIS/Export: Removed section because it pertains to
MYLA®.
• Password Policy: Added important note regarding system
configuration. Updated maximum password age for high security
level.
Release Part
Change Type Change Summary
Date Number
Troubleshooting
• Updated the following alarm or alert codes:
◦ Negative bottle(s) was reloaded. [32]
◦ Unmatched reloaded bottle has resulted in a duplicate bottle
record. [36]
◦ Bottle unable to be removed from cell. [83]
◦ Duplicate bottle was detected in another instrument. [99]
◦ Bank capacity has reached the configured threshold. [4]
◦ Bottle(s) on the conveyor cannot be loaded due to lack of bank
capacity or a malfunction/jam. [5]
◦ A bottle was placed in a cell that has become non-operational.
[10]
◦ The embedded software has rebooted on the listed instrument(s).
[65]
◦ Bottle readings are not being taken because the instrument is out
of range. [80]
◦ Communication has been lost with the listed instrument(s). [28]
◦ Communication has been lost with the listed instrument(s). [29]
Technical
• Added the following alarm or alert codes:
change
◦ Problem with a bottle(s) on the conveyor of the listed
instrument(s). [59]
◦ BCI CONNECT general error. [103]
◦ Archive failure. [104]
◦ Loss of communication with BCI CONNECT. [105]
◦ Bottle matched an old accession number. [106]
◦ Bottle send to BCI CONNECT failed. [107]
• Removed the following alarm or alert codes:
◦ An unknown LIS code was received. [24]
◦ Communication is unavailable between the instrument and the
LIS. [25]
◦ Unexpected LIS error. [26]
• Opening the Instrument for Internal Access: Added further
explanation about Robot Activity indicator.
• Closing the Door: Updated warning about placing robot gripper in
safe to move position.
• Unjamming the Robot: Updated note about reserved cells for
calibration standards.
Technical
Updated warranty and liability disclaimer.
change
Technical Introduction: Added chapter per template and changed all references to
change package insert to IFU.
Technical
Safety Information: Added chapter and updated label
change
System Overview:
• Physical Characteristics: Added B Unit weight, heat transfer, and
power consumed specifications.
• System Configuration: Removed network and site configuration.
• General Instrument Configuration: Added B Unit specifications.
• Bottles Supported for Use: Changed description of PF bottle.
Technical • Bottles Not Supported for Use: Removed MB bottle.
change • BACT/ALERT® VIRTUO® Hardware: Added B unit images and
specifications and images for transfer arm, transfer gate, and safe to
move sensor.
• Modified temperature probe warning.
• Small Display: Updated images and icons. Removed paragraph
under table.
• Search Criteria Screen Icons: Added new icons.
• Additional Supplies: Added list of additional components.
Release Part
Change Type Change Summary
Date Number
Basic Functions
• Home Screen: Updated images and icons.
• System Login: Updated Figure 21.
• Password Policy: Removed section.
• Applying a Barcode Label to a Bottle: Added color specifications for
bottle labels. Added fill-line to Figure 32.
• Handling Unconfirmed Positive Bottles: Added step to login. Moved
warning about transfer of bottle. Added note to step 3.
• Loading Bottles Automatically: Modified warning on loading bottles;
added instructions for loading batches of bottles; modified steps in
procedure.
• Unloading Positive Bottles Automatically: Changed wording around
"removed" and "unloaded" and modified note.
Technical
• Canceling a Bottle Test: Modified step 3.
change
• Updated Figure: 26, 29, 33, 34, 35, 36, 43, 45, 47, 48, 54, 61, 63, 65,
67
• Using Additional Search Criteria: Added criteria and note.
• Searching by Contaminated: Added instructions on search criteria.
• Reloading Positive Bottles: Modified procedure.
• Editing a Bottle Record: removed note about Labsuper role and
modified precedence rules note. Added waning about physical
keyboard.
• Altering Bottle Reports: Added section.
• Viewing and Printing Bottle Reports: Added definition for anonymous
bottles; added steps to view reports.
• Emptying the Waste Container: Added note.
• Added section Loading Bottles Across a Bank
System Configuration
• Removed general software configuration setting.
• Identifying an Anonymous Bottle Without a Visible bioMérieux Label:
Replaced image.
• Added sections on Exporting Archive Records and Viewing Archive
Records
• Naming Banks and Labs: Added note on special characters.
Technical • Replaced Figure 81 and 83.
change • System Configuration: Remove sections on MYLA® import and
export.
• Disabling 21 CFR 11 Settings: Added instructions of how to configure
settings for individuals.
• Sample Volume Measurement: Added paragraphs on how sample
volume measurement works.
• Viewing Archive Records: Added step to procedure.
• User Management Groups: Added resolving LIS codes to MYLA®
LabTech role.
Troubleshooting:
• Updated Figure 95; added entry to Table 18.
• Added caution and warnings about time and temperature loss.
• Removed the following error codes:
◦ Instrument Automatically Rebooted Alarm [65]
◦ LIS Patient Conflict Alarm [27]
◦ MYLA® Rejected Message [101]
• Added the following codes:
◦ Bottle Transfer Error [93]
◦ Conveyor Cover Off [94]
◦ Transfer Gate Open [96]
◦ Bank Capacity Exceeded changed to No Space Available [5]
◦ Bottle Not Know by PC [97]
◦ Duplicate Bottle in another instrument [99]
Technical ◦ Incompatible Firmware version [98]
change ◦ Knowledge Base update [100]
• Added definition for jam and malfunction.
• Modified the following alarms: 2, 7, 13, 18, 33, 12, 20, 25, 28, 29, 35,
40, 44, 48, 53, 55, 59, 63, 67, 68, 69, 70, 71, 72, 74, 80, 81, 82, 83,
86, 89, 90, 93, 94, 96, 97, 98, 99.
• Unexpected Conditions: Modified all steps in resolutions.
• Added procedure for loading bottles across a bank.
• Opening the Instrument for Internal Access and Opening the Door:
Modified procedure.
• Closing the Door: Modified procedure and added warning for safe to
park sensor.
• Shutting Down the Instrument: Changed to rebooting the instrument
and updated procedure.
• Starting Up the Instrument: Removed Important note. Added note
about location of power cord and switch.
• Unjamming the Robot: Added notes.
• Added Exporting Audit Trail Reports and Saving to an External Drive
Technical
Glossary: Added term bank
change
Introduction
System Overview
Basic Functions:
• Zone 1: Removed navigation from table title.
Administrative • Searching by the Load Status: Added clarification to load and unload
status.
• Processing Expired Bottles: Removed second duplicate note.
System Configuration:
• Configuring General Software Settings: Made minor changes to
Administrative defining pattern bullet and Enabling Barcodes on Bottles and Defining
Validation Patterns to add Patient ID.
• Moved introductory text on configuration after user management
section.
Troubleshooting:
Administrative • Changed troubleshooting table into 4 separate tables. Added error
code numbers.