The OFFSET for a calibrated microbolometer is typically
�2�C (3.6�F) across its full temperature detection range, and usu- ally �1�C (1.8�F) at physiologic temperatures. If the imager�s OFFSET is greater than the specified maxi- mum error, off to professional recalibration it goes! Be sure to tell the technicians to calibrate it carefully at the 18�40�C (~65�105�F) physiologic range and be less concerned about other temperatures. Image flatness at room temperature Although checking image FLATNESS is becoming less important with the newer low-power imagers that develop less internal heat, it should be checked when the unit is first put into service and at least annually. This test can be performed using the same image taken for the OFFSET measurement (Figure 4.5b). Equipment Needed: � Same as that required for the OFFSET measurement described above. � The NIST-traceable bulb or digital thermometer is not required here. � A stable room-temperature blackbody device may be used (optional). Procedure: Using the image already taken for the OFFSET measurement, make four additional area measurements, one in each corner of the image. The absolute FLATNESS for the image is the lowest of the five measurements subtracted from the highest measurement. Some image interpretation software applications can automatically find and measure the highest and lowest temperatures detected across an image, making this measurement simple and more universal. Tflatness Tmax Tmin Troom= - @ where � Tflatness is the maximum temperature difference over the image. � Tmax is the maximum temperature reading among the measurements. � Tmin is the minimum temperature reading among the measurements. � @Troom means the image is taken at a nominal room temperature. In this example (Figure 4.6a), Tflatness = 72.2�71.9 = 0.3�F (0.17�C) @Troom. A FLATNESS value of 0.17�C (0.3�F) with this simple test is acceptable. There will often be a slight temperature gradient across the foam board due to air convection. This same imager gave a FLATNESS value of 0.0�C (0.0�F) using a Blackbody Calibrator as shown in the five-image sequence (Figure 4.6b). Bucket calibration Another calibration verification method suitable for a human thermography clinic is referred to as �Bucket Calibration.�6 A variation of Bucket Calibration is explained here. This method is inexpensive, semi-automated, and requires little equipment. Bucket Calibration tests both the imager GAIN and OFFSET, but not image FLATNESS. A common thin-walled metal bucket with the outside painted flat black is filled with hot tap water at about 43�C (~110�F). The water temperature is monitored by a NIST-calibrated remote- reading, long-stem thermometer. The infrared imager under test is aimed to measure an area temperature on the black portion of the bucket, while the visual camera in the unit records the remote readout of the immersed thermometer (Figure 4.7). Imager Emissivity should be set to 0.95 (approximately equal to the blackened bucket surface, see reference 9) and the Background Temperature set to the room temperature. The envi- ronment should be at nominal room temperature with little air movement and no radiant infrared sources or sunlight.