MORPHEUS LT/MRI

Anaesthesia Unit
compatible for Magnetic Resonance

code: OM3.A5s/RMN
rev. 4 : dated 01/06/2017

INTENDED USE

The MORPHEUS LT/MRI is an anaesthesia unit that can be used on adult,


children and newborn patients.

The MORPHEUS LT/MRI is an anaesthesia machine compatible for Magnetic


Resonance rooms from 1,5 T (15000 G) and from 3 T (30000 G) and
positionable at a min. distance from the magnet corresponding to a field of 20
mT (200 G).

The MORPHEUS LT/MRI is suitable for administration of Oxygen - Air - Nitrous


Oxide - Halothane - Enflurane - Isoflurane - Sevoflurane mixtures.

GENERAL DESCRIPTION

The MORPHEUS LT/MRI anaesthesia unit is completed with:

 mechanic gas mixing system


 electronic lung ventilator with dual colour display
 valves group: open, semi-closed, closed, heated, with soda lime absorber of
1 Kg. capacity
 SIARETEX rapid connection device, Selectatec compatible for 2 vaporizers
 gas supply group

Morpheus_LT-MRI_te_rev4 1/7
TECHNICAL DATA

Structure Light aluminium alloy and plastic moulds (a magnetic material)

Wheels Pivoting antistatic wheels, diameter 100 mm (2 with brakes)

Cylinder support Not supplied.


Use cylinder aluminium MRI compatible only.

Support for 2 vaporizers On horizontal guide (SIARETEX rapid connection device, Selectatec compatible
for 2 vaporizers)

Auxiliary power supply No. 1 schuko 220 Vac outlet (max. 6 A)


outlets

Work shelf lighting 12Vdc by led

Dimensions 71 x 77 x 138 (L x P x H) cm

Weight 72 kg (without accessories)

Environmental conditions  Temperature from 10 to 40°C


 Relative humidity from 10 to 90% non-condensing

GAS MIXING SYSTEM

It has the function to regulate the capacity and the concentration of gas mixture
(Air, O2, N2O) as well as to deliver it to the anaesthetic gas vaporizer.
It allows to select the mixture to be delivered (Air - O2, or N2O - O2) and the O2
enrichment for delivered mixture in case of emergency.
The anaesthesia module includes a device which guarantees a minimum
concentration of 25% oxygen in all conditions (MIX-LIFE device).

The three pressure gauges on the front panel allow the continuous control of
medical gas feeding pressure coming from the gas pipelines system.

Morpheus_LT-MRI_te_rev4 2/7
Oxygen rotameter Scale: 0.1 - 15 L/min.
Resolution: 0.1 L/min up to 1 L/min and 1 L/min up to 15 L/min
Accuracy:  10% of read value or:  1% of end scale whichever is the worse
case.

Nitrous oxide rotameter Scale : 0.2 - 12 L/min.


Resolution: 0.1 L/min up to 1 L/min and 0.5 L/min up to 12 L/min
Accuracy:  10% of read value or:  1% of end scale whichever is the worse
case.

Air rotameter Scale: 0.1 - 15 L/min.


Resolution: 0.1 L/min up to 1 L/min and 1 L/min up to 15 L/min
Accuracy:  10% of read value or:  1% of end scale whichever is the worse
case.

Low flows oxygen Scale 0.1 - 1 L/min.


rotameter
Resolution: 0.05 L/min
Accuracy:  10% of read value or:  1% of end scale whichever is the worse
case.

Low flow nitrous oxide Scale: 0.1 - 1 L/min.


rotameter
Resolution: 0.05 L/min
Accuracy:  10% of read value or:  1% of end scale whichever is the worse
case.

Medical gas supply OXYGEN


 Pressure included between 280 kPa and 600 kPa (2,8 – 6 bar)
 Max. required flow 90 L/min.

NITROUS OXIDE
 Pressure included between 280 kPa and 600 kPa (2,8 – 6 bar)
 Max. required flow 15 L/min.

MEDICAL COMPRESSED AIR


 Pressure included between 280 kPa and 600 kPa (2,8 – 6 bar)
 Max. required flow 90 L/min.

Gauges No. 3 on front panel (O2 - N2O - AIR ), scale 0 - 6 bar

Alarms Lack or low oxygen pressure with consequent cut-off of nitrous oxide delivery

Morpheus_LT-MRI_te_rev4 3/7
Safety devices AGAINST THE ADMINISTRATION OF HYPOXIC MIXTURES MIX-LIFE: it
always guarantees a minimum concentration of 25 % oxygen on mixtures which
includes nitrous oxide.

IN CASE OF LACK OR LOW OXYGEN PRESSURE CUT-OFF: audible alarm


with immediate cut-off of nitrous oxide delivery.

AGAINST OVERPRESSURE IN FLOWMETER BOX : safety valve calibrated at


0.8 bar for the protection of the glass rotameters.

IN CASE OF LACK OR COMPRESSED AIR LOW PRESSURE : all the devices


(gas feeding) supplied by compressed air are automatically supplied by oxygen.

AGAINST THE SIMULTANEOUS DELIVERY OF AIR AND N2O : selection by


membrane key on the flowmeter front panel.

Control for activation of Setting of MANUAL modality on ventilator (MAN) with automatic deviation of
exit of fresh gas for fresh gas to the manual system of anaesthesia unit valves group, or to a TO-
manual ventilations. AND-FRO circuit with visual indicator.

O2 emergency by-pass By apposite membrane key on the front shelf, max flow 35 L/min.

IN gas sockets on gas  No. 3 sockets for distribution system (O2 - N2O - AIR)
supply group
 No. 2 sockets for cylinder (O2 - N2O)

OUT gas sockets on gas  No. 1 sockets for O2


supply group
 No. 1 sockets O2 - AIR for active scavenger feeding

 No. 1 fresh gas connector for external use for ex. TO AND FRO (selectable
by apposite membrane keyboard on the front shelf - AUX).

Other  Socket for recycle of exhaust monitor gas


 Connection for anaesthetic gas scavenging (optional device: active type, or
passive type)

Morpheus_LT-MRI_te_rev4 4/7
BREATHING SYSTEM

Compact system with automatic connections, easy dismountable and


autoclavable

It allows the ventilation in modality: real open circuit, semi-closed circuit, closed
circuit at low flows.

The system also allows the spontaneous and manual ventilation in case of
anaesthesia unit breakdown or machine off.

Top special CO2 absorber canister of 1 Kg with rapid connection: this allows
canister replacement also during interventions (the canister is autoclavable and
reusable).

The recycling system is a selective type, hence the soda lime and fresh gas
consumption are reduced to the minimum.

The heated valves group reduces the condensation and heats the fresh gas.

The transition from one ventilation modality to another is completely controlled by


the ventilator without any user’s action on valves group.

LUNG VENTILATOR

User’s interface Dual colour display 165 x 145 mm with membrane keyboard and encoder

Control modality Electronic by microprocessor

Dead space Automatic


compensation system

Automatic compensation Present (max. 5000 mt)


of atmospheric pressure
on measured pressure

Respiratory parameters Present (newborn, children, adult)


default setting

Flow generation Electronic system

Gas feeding Medical compressed air or Oxygen with pressure included between 280 kPa and
600 kPa (2,8 – 6 bar)

Morpheus_LT-MRI_te_rev4 5/7
Autotest Primary test: at anaesthesia unit’s start-up, a control test of medical gas supply,
sensors operation, back-up battery, oxygen cell, integrity of the alarm audible
indicator, led on CPU board. This test takes around 15 seconds.

Subtest: this subtest permits to verify the dead space and losses or to perform
the oxygen cell calibration.

Ventilation modalities MANUAL, VC-VAC, PCV-APCV, P-SIMV, V-SIMV

Breathing rate From 5 to 90 bpm (step 1 bpm)

Breathing rate (SIMV) From OFF, 1 ÷ 89 bpm (step 1bpm)

I:E Ratio 1:4 ÷ 4:1

Inspiratory time From 0.2 to 5 sec.

Tidal Volume From 50 to 1500 ml

Minute Volume From 1 ÷ 30 liters

PEEP OFF, 3 ÷ 30 cmH2O

Inspiratory Flow From 1 ÷ 80 L/min.

Oximeter Minimum resolution 1% / Automatic calibration procedure

High pressure limit From 10 to 80 cmH2O

Bronchomanometer Electronic: from -10 ÷ 80 cmH2O

Flow trigger From OFF, 1 to 15 L/min (step 1 L/min)

Pressure trigger From OFF, -1 to -9 cmH2O under the PEEP level

Safety Electronic and mechanical limit of airways pressure / Self-diagnosis system

Flow sensor Pressure relief single patient sensor

Alarms Fan Failure, Air Failure, High / Low Airways Pressure, Low O2 concentration, O2
cell into operation or not present, Apnoea, Expired Tidal Volume, Power Failure,
Low Battery.

Measured parameters PAW, FiO2 , VM, Tinsp, RATE, Vte

Morpheus_LT-MRI_te_rev4 6/7
ELECTRIC POWER SUPPLY

Electric power supply 100 - 240Vac / 50 - 60Hz

Maximum power 100 Watt

Back-up battery 12Vdc - 3 Ah Pb battery which guarantees an autonomy of around 120


minutes

Charging time Around 10 hours

CONFORMITY TO DIRECTIVES

Class and type according with IEC 601-1 Class I Type B

Class according with 93/42/ EEC Dir.ve. Class IIb

EN 60601-1:2006 , EN 60601-1-2:2007, IEC 601-1-6:2010, IEC 601-1-8:2007, EN 60601-2-13:2011, EN


62304:2006, EN ISO 5356-1:2015, EN ISO 4135:2001, DIR. 93/42/EEC, DIR. 2011/65/CE, D.Lgs 49/2014

ACCESSORIES

Standard accessories  User’s Manual


 O2 supply hose
 N2O supply hose
 Air supply hose
 Top Special CO2 absorber canister of 1 kg (no. 1 / code 30900)
 O2 cell
 Flow sensor (no. 2)
 Adult PVC patient circuit (240cm) (No. 2)
 Adult Mapleson C adult patient circuit
 Manual ventilation KIT
 Electric power supply cable

Other optional accessories See current export price list

SIARE applies the UNI EN ISO 13485:2004 Quality System and 93/42 EEC Dir.ve.