Morpheus - LT MRI Data Sheet
Morpheus - LT MRI Data Sheet
Anaesthesia Unit
compatible for Magnetic Resonance
code: OM3.A5s/RMN
rev. 4 : dated 01/06/2017
INTENDED USE
GENERAL DESCRIPTION
Morpheus_LT-MRI_te_rev4 1/7
TECHNICAL DATA
Support for 2 vaporizers On horizontal guide (SIARETEX rapid connection device, Selectatec compatible
for 2 vaporizers)
Dimensions 71 x 77 x 138 (L x P x H) cm
It has the function to regulate the capacity and the concentration of gas mixture
(Air, O2, N2O) as well as to deliver it to the anaesthetic gas vaporizer.
It allows to select the mixture to be delivered (Air - O2, or N2O - O2) and the O2
enrichment for delivered mixture in case of emergency.
The anaesthesia module includes a device which guarantees a minimum
concentration of 25% oxygen in all conditions (MIX-LIFE device).
The three pressure gauges on the front panel allow the continuous control of
medical gas feeding pressure coming from the gas pipelines system.
Morpheus_LT-MRI_te_rev4 2/7
Oxygen rotameter Scale: 0.1 - 15 L/min.
Resolution: 0.1 L/min up to 1 L/min and 1 L/min up to 15 L/min
Accuracy: 10% of read value or: 1% of end scale whichever is the worse
case.
NITROUS OXIDE
Pressure included between 280 kPa and 600 kPa (2,8 – 6 bar)
Max. required flow 15 L/min.
Alarms Lack or low oxygen pressure with consequent cut-off of nitrous oxide delivery
Morpheus_LT-MRI_te_rev4 3/7
Safety devices AGAINST THE ADMINISTRATION OF HYPOXIC MIXTURES MIX-LIFE: it
always guarantees a minimum concentration of 25 % oxygen on mixtures which
includes nitrous oxide.
Control for activation of Setting of MANUAL modality on ventilator (MAN) with automatic deviation of
exit of fresh gas for fresh gas to the manual system of anaesthesia unit valves group, or to a TO-
manual ventilations. AND-FRO circuit with visual indicator.
O2 emergency by-pass By apposite membrane key on the front shelf, max flow 35 L/min.
IN gas sockets on gas No. 3 sockets for distribution system (O2 - N2O - AIR)
supply group
No. 2 sockets for cylinder (O2 - N2O)
No. 1 fresh gas connector for external use for ex. TO AND FRO (selectable
by apposite membrane keyboard on the front shelf - AUX).
Morpheus_LT-MRI_te_rev4 4/7
BREATHING SYSTEM
It allows the ventilation in modality: real open circuit, semi-closed circuit, closed
circuit at low flows.
The system also allows the spontaneous and manual ventilation in case of
anaesthesia unit breakdown or machine off.
Top special CO2 absorber canister of 1 Kg with rapid connection: this allows
canister replacement also during interventions (the canister is autoclavable and
reusable).
The recycling system is a selective type, hence the soda lime and fresh gas
consumption are reduced to the minimum.
The heated valves group reduces the condensation and heats the fresh gas.
LUNG VENTILATOR
User’s interface Dual colour display 165 x 145 mm with membrane keyboard and encoder
Gas feeding Medical compressed air or Oxygen with pressure included between 280 kPa and
600 kPa (2,8 – 6 bar)
Morpheus_LT-MRI_te_rev4 5/7
Autotest Primary test: at anaesthesia unit’s start-up, a control test of medical gas supply,
sensors operation, back-up battery, oxygen cell, integrity of the alarm audible
indicator, led on CPU board. This test takes around 15 seconds.
Subtest: this subtest permits to verify the dead space and losses or to perform
the oxygen cell calibration.
Alarms Fan Failure, Air Failure, High / Low Airways Pressure, Low O2 concentration, O2
cell into operation or not present, Apnoea, Expired Tidal Volume, Power Failure,
Low Battery.
Morpheus_LT-MRI_te_rev4 6/7
ELECTRIC POWER SUPPLY
CONFORMITY TO DIRECTIVES
ACCESSORIES
SIARE applies the UNI EN ISO 13485:2004 Quality System and 93/42 EEC Dir.ve.