1(7)

Outotec PSI® 500i Datasheet 10000007663e Revision A

Outotec PSI® 500i Particle size analyzer

®
1. PSI 500i analyzer

PSI 500 particle size analyzer is an on-line size


measurement system for mineral slurries. It is used in
controlling wet mineral processes, primarily grinding,
classification re-grinding and thickening. PSI 500i
automatically takes samples from one to three process
streams and measures their particle size distribution in
the range of 1-500 µm. The analyzer displays the
analysis results on touch screen display and reports
them to the plant’s automation system for real-time
process control.
The system operates on a continuous basis 24 hours a
day, with minimal manual maintenance needs. The
daily maintenance routines include checking the
sampling system and cleaning possible trash buildup in
the secondary sampler. Otherwise the maintenance
needs are covered by preventive maintenance calls by
Outotec Customer Service.
The particle size analysis is based on laser diffraction
using a patented multiple scatter calculation method to
enable the measurement with dense slurries. Due to the
fundamental nature of the analysis principle, no
calibration against a reference analysis is required for
volumetric size distribution measurement A linear
calibration model can be used to convert the results to
approximate other size measurement methods (sieve,
sedigraph etc.) based on regression modeling of
calibration results.
There are three options for the optical focal length in
the system, to optimize measurement for the particle
size range of interest. There is a wide variety of
particle size distribution characteristics that can be
selected as outputs to the plant automation system
through the interactive analyzer configuration displays.
The primary sampling from the process stream is
normally engineered in the analyzer delivery project
using the range of primary samplers, to ensure
satisfactory representativeness and trouble-free
transport of the sample to the analyzer. Figure 1. PSI 500i particle size analyzer and
primary sampler
2(7)

Outotec PSI® 500i Datasheet 10000007663e Revision A

2. Analyzer system structure


The PSI 500i particle size analyzer is shown in Figure In addition to PSI 500i itself, the complete analyzer
1. This unit includes secondary and calibration system requires also primary sampling and a dilution
sampling, dilution unit, pneumatics, I/O, local control water tank. These items are also supplied as a part of
panel and optical sensor head. PSI 500i delivery by Outotec. Figure 2 shows a
simplified flow sheet of the complete system.

Figure 2. PSI 500i analyzer system flow sheet


2.1. Primary sampling of the system. Samples will be typically returned back
PSI 500i can handle at most three sample lines and to process by gravity.
supports automatic flushing for the following Outotec 2.2. Dilution water tank
sampler types: LSA, NLA, PSA and CPS.
The dilution water tank is needed to ensure that the
Representative sampling with straight sample transfer
diluted sample entering the measurement head is free
piping are essential for dependable particle size
of small air bubbles that would appear as particles in
measurement. PSI 500i can be located close to the
the analysis. Suitable de-aeration is achieved by 1.2 m3
sampling points so that the sample transport distance is
water volume free surface tank. The water tank can be
short and continuous gravity flow can be used.
located within 20 m distance from the analyzer. A
Recommended primary sample flow is 50-100 l/min.
centrifugal pump lifts the dilution water pressure up to
Provisions for automatic flushing and cleaning of the
4 bar.
sampler and the sample line guarantee high availability
3(7)

Outotec PSI® 500i Datasheet 10000007663e Revision A

2.3. Secondary sampling 2.6. Optical Sensor Head


The secondary sampler cuts representative 0.01-0.03 l The optical sensor head is equipped with a solid-state
slices out of the primary sample flow, by moving the diode laser, a transmission flow cell with sapphire
sample line across a stationary secondary sample windows to pass the laser beam through the sample and
cutter. This cutter has a programmable wash spray to a ring detector to measure the scattered light
prevent clogging and cross contamination of samples. distribution from the other side of the flow cell.
The cutter opening and cutting speed can be adjusted to The flow cell provides a smooth duct channel for the
set the sample volume to a suitable range according to sample flow. The flow is turbulent to provide good
the primary sample flow rate and sample properties. presentation of large particles. The diameter of the
The sample is fed to the dilution unit through optional measurement area is 10 mm at the center of the flow.
demagnetizer and dispersion unit. The flow cell outlet tube has a pinch shutoff valve to
leave the cell full of water when the analyzer is not in
The number of secondary sample cuts is automatically
operation.
controlled using the optical transmission signal in the
sensor head, to ensure optimal measurement 2.7. Probe Control Set
conditions. If the transmission is too low the number of The analyzer control cabinet has an integrated
cuts is decreased to allow more water to be mixed to industrial PC hosting a controller application for the
the flow, and vice versa. sampling/dilution, a local user interface for displaying
Another cutter in the secondary sampler cuts manually the equipment status and analysis results on the field as
requested calibration samples or programmable well as for setup and maintenance operations.
composite samples at the original (process) solids The PC hosts software to control and configure the
content for laboratory analysis. The calibration sample optical sensor head. The control application
cutter has a manually controlled wash spray to prevent synchronizes the measurement with the sampling
clogging. control sequence and provides user interfaces for
2.4. Sample demagnetizer and dispersion options measurement and diagnostic data transfer to the plant
automation system.
If samples contain more than 2 – 4% ferromagnetic
minerals, e.g. magnetite, the particles tend to The measurement sequence starts with a sample cutting
agglomerate and are measured as larger particles. The phase where the secondary samples are introduced into
sample demagnetizer option applies an electronically the diluter, and ends with a draining phase where all
controlled oscillating magnetic field to the samples, particles from the secondary sample are passed to the
and the demagnetized particles are dispersed with optical sensor head. The particle size signal is an
ultrasound dispersion. average over the measurement cycle, weighted by the
apparent %solids of the diluted sample. Before the next
The ultrasound option can also be used alone to measurement, the diluter is completely flushed, and a
disperse slurries with fine agglomerating particles to background check for the optical measurement head is
retrieve a particle size distribution resembling the done at set intervals.
original distribution.
The control application provides a status/data display
2.5. Diluter Unit and records the data and events into a history database
The patented diluter unit adds water to the slurry for viewing and diagnosing.
sample sufficiently so that the slurry entering the
A simplified remote HMI view with a web browser or
measurement flow cell passes enough light. The top
full control of the analyzer with a remote desktop
portion of the dilution unit has relatively still flow
software are available. Both choices utilize an Ethernet
conditions in order to let entrained air to rise to the
connection. The remote desktop connection can be
surface. A froth killing water spray is introduced from
used for local operator monitoring, parameter setup and
the top to break large bubbles.
service diagnostics. Outotec service contract remote
The water is mixed to the slurry from two feed points. diagnostics/services are available through optional
The spray water addition is controlled using the level secured internet or dial up connection.
signal. A screen stops trash and oversized particles
before they block the flow cell.
The total sample volume in the dilution unit is ca. 9
liters. The water flow at the outlet is 10 l/min.
4(7)

Outotec PSI® 500i Datasheet 10000007663e Revision A

Figure 3. Analyzer status display on local control Figure 5. The operation of the analyzer is controlled
panel by interactive configuration displays.

Figure 4. Particle size distribution display on local


control panel Figure 6. It is possible to measure manually
introduced external sample.

Measuring head Focal length D50 range D2 – D98 range

OLE 5112 100 mm 2.5 -125 µm 0.5 – 200 µm

OLE 5212 200 mm 5.0 – 250 µm 1.0 – 400 µm

OLE 5312 300 mm 7 – 350 µm 1,5 – 600 µm

Figure 7. Measuring head models are optimized for specific size ranges
5(7)

Outotec PSI® 500i Datasheet 10000007663e Revision A

3. Specifications 4. Options
3.1. Sample streams 4.1. Additional Sample Inlet
The basic version of PSI 500i can be connected to one PSI 500i can be upgraded with one or two additional
primary sample line. With additional sample inlet sample inlet options to enable sampling from
options PSI 500i is capable of taking up to three maximum of three process lines.
primary sample streams. 4.2. Demagnetizer and ultrasound dispersion option
Primary sample flow rate: 50…100 l/min For samples with significant magnetite content,
typical, depends on sample solids content and demagnetization is required together with ultrasound
size distribution dispersion to de-agglomerate the particles.
Maximum solids content: up to 70% w/w, Fine particles with an agglomeration tendency can be
sample has to flow freely dispersed with the ultrasound dispersion option.
Primary sample line: Inlet OD 60 mm / G2”
Primary sample reject fitting: OD 76 mm / 4.3. Analyzer Remote Station
G2½” The analyzer remote station is a Windows® PC for:
Secondary sample reject fitting: OD 25 mm
defining the run time parameters remotely
3.2. Sample handling managing analysis data
Sample dilution to 0.5% (vol.) typical, dilution is collecting diagnostic data from the analyzer
controlled automatically for suitable laser beam displaying the analyzer status and analysis results
transmission through the diluted sample remote diagnostics
Composite sampling of undiluted sample with Normally the analyzer remote station is set up in the
timed intervals control room.
Cleaning of primary sampler with automatic 4.4. Acid Window Wash Option
sample/flush valves by timed intervals or manual
For contaminating samples, PSI 500i can be equipped
command
with a washing system to assure that optical sensor
Adjustable secondary and composite sampler
head windows and dilution mechanism remain clean.
cutter openings
Automatic secondary sampler and manually 4.5. Salt Water Accessories
controlled composite sampler cutter cleaning by If salt water or other corrosive medium is used for
integrated water sprays dilution and washing, PSI 500i can be supplied with
Demagnetizer and dispersion options acid proof steel valves
3.3. Analytical 5. Installation
Size range: 1- 500 micrometers, size range can
be optimized according to application. 5.1. Max dimensions
Measurement data output: 40 fractions (%vol. PSI 500 analyzer unit:
±x µm or mesh), particle size percentiles (µm at - 2071 (H) x 1200 (W) x 930 mm (D)
x %), Blaine number (m2/kg), specific surface - 82” (H) x 47” (W) x 37” (D)
area (m2/cm3) Dilution water tank:
Precision: 1-2% (rel.) for Dv(50) particle size - 1304 (H) x 1022 (W) x 1206 mm(D),- 51 “ (H) x
based on a glass bead sample. 40” (W) x 47” (D)
Accuracy: when compared to other particle Water pump with cover
sizing methods than laser diffraction, differences - 520 (H) x 820 (W) x 320 (D) mm, 21 “ (H) x 32”
are noticed due to different influence of particle (W) x 13” (D)
shape, particle specific gravity etc. to the Pump control box 430 (H) x 300 (W) x 240 (D)
analysis. The system allows correlation of the mm ,17 “ (H) x 12” (W) x 9,5” (D)
measured particle size data to other
5.2. Weight
measurements by independent calibration
samples; however, this correlation is application PSI 500i analyzer: 250 kg, 550 lbs
dependent Dilution water tank (empty): 165 kg, 365 lbs
Measurement interval: typically 3 min / sample Dilution water tank (full): 1365 kg, 3010 lbs
line Water pump and control box: 35 kg, 80 lbs
6(7)

Outotec PSI® 500i Datasheet 10000007663e Revision A

5.3. Typical shipping dimensions 5.6. Remote connection options


PSI 500i analyzer unit: 350 kg, 3.0 m3 Remote desktop diagnostics/service connection
(2170x1140x1220 mm) via ethernet (LAN/Internet) or dial up modem
Auxiliares box: 150 kg, 0.85 m3 line
(1200x800x860 mm) Remote display: Web server
Dilution water tank (empty): 200 kg, 1.6 m3
(1200x1020x1300 mm) 5.7. Water
Particle-free (sand filtered, potable) water
5.4. Power supply Water consumption: 10 l/min (2.7 GPM)
PSI 500i analyzer unit: average, max 30 l/min (8 GPM)
- Single phase AC, 230V±10% at 2A Water temperature: max. 40 °C
50/60Hz or 115V ± 10 % at 4A 50/60Hz (wash spray water can be raised up to 80 °C with
Dilution water pump: AISI 316 diluter unit options)
- Single phase AC, 230V±10% at 4A Inlet in dilution water tank: 1” hose coupling
50/60Hz or 115V ± 10 % at 8A 50/60Hz for water hose of 25 mm inner diameter
A double conversion (AC/DC/AC) UPS 1 kW / 10 min
is highly recommended to secure trouble-free and 5.8. Instrument air
stable operation. Oil-free dry instrument air 5…10 bar (75…145
psi), no condensation
5.5. DCS connection options
Air consumption: 17 Nl/min (0.6 cuft/min)
Modbus RTU (PSI 500i is master or slave ) average
Modbus TCP/IP (PSI 500i is slave) Inlet: G½”
OPC (PSI 500i is server)
Profibus DP (PSI 500i is slave)

Figure 8 PSI 500i analyzer dimensions in mm. NOTE Analyzer with 300 mm optics requires 1200 mm width.
See installation manual for full details and instructions.
7(7)

Outotec PSI® 500i Datasheet 10000007663e Revision A

5.9. Environment Quality assurance


Ambient temperature: 5-45 °C (no direct
PSI® 500i Particle Size Analyzer is CE marked
sunlight), air conditioned cabin recommended
according to the European Union Directives.
for temperatures above +35 deg C and heated
cabin below +5 deg C. The quality system of Outotec (Finland) Oy is ISO
Installation altitude: 0 to 5000 m (16000 ft) 9001 certified. The company has a quality management
above sea level policy to ensure that the expectations of customers are
Sample temperature: 5-50 °C met.
(with AISI 316 funnel option 80 °C)
Outotec is a leading supplier of minerals processing
Corrosion: AISI 316L, HDPE, POM, sapphire, technologies including grinding, flotation, thickening,
rubber, neoprene and silicone when in contact filtering, physical separation, and automation
with sample, frame phosphated AISI 316L, equipment. The company manufactures a wide range of
enclosures powder epoxy coated (60-80 µm) samplers for on-line analysis, composite, and
AISI 304 metallurgical sampling of mineral slurries, solutions
Moisture: non condensing and dry powders.
Shock: 15 G equivalent / 11 ms
Vibrations: 1 G / 5-150 Hz You can find more information at www.outotec.com

5.10. Standards
Plant equipment conforms with European Union PSI® Particle Size Analyzer is a registered trademark
machine directive, low voltage directive and of Outotec Oyj.
EMC directive for heavy industrial environment
© 2011 Outotec Oyj. All rights reserved. Information
EMC: heavy industrial environment, EN 50082-2
is subject to change without notice.
& EN 50081-2
Environmental tolerance: IP56/NEMA 4X
Laser source meets Class II specifications