Data Sheet PSI 500i
Data Sheet PSI 500i
®
1. PSI 500i analyzer
Figure 3. Analyzer status display on local control Figure 5. The operation of the analyzer is controlled
panel by interactive configuration displays.
Figure 7. Measuring head models are optimized for specific size ranges
5(7)
3. Specifications 4. Options
3.1. Sample streams 4.1. Additional Sample Inlet
The basic version of PSI 500i can be connected to one PSI 500i can be upgraded with one or two additional
primary sample line. With additional sample inlet sample inlet options to enable sampling from
options PSI 500i is capable of taking up to three maximum of three process lines.
primary sample streams. 4.2. Demagnetizer and ultrasound dispersion option
Primary sample flow rate: 50…100 l/min For samples with significant magnetite content,
typical, depends on sample solids content and demagnetization is required together with ultrasound
size distribution dispersion to de-agglomerate the particles.
Maximum solids content: up to 70% w/w, Fine particles with an agglomeration tendency can be
sample has to flow freely dispersed with the ultrasound dispersion option.
Primary sample line: Inlet OD 60 mm / G2”
Primary sample reject fitting: OD 76 mm / 4.3. Analyzer Remote Station
G2½” The analyzer remote station is a Windows® PC for:
Secondary sample reject fitting: OD 25 mm
defining the run time parameters remotely
3.2. Sample handling managing analysis data
Sample dilution to 0.5% (vol.) typical, dilution is collecting diagnostic data from the analyzer
controlled automatically for suitable laser beam displaying the analyzer status and analysis results
transmission through the diluted sample remote diagnostics
Composite sampling of undiluted sample with Normally the analyzer remote station is set up in the
timed intervals control room.
Cleaning of primary sampler with automatic 4.4. Acid Window Wash Option
sample/flush valves by timed intervals or manual
For contaminating samples, PSI 500i can be equipped
command
with a washing system to assure that optical sensor
Adjustable secondary and composite sampler
head windows and dilution mechanism remain clean.
cutter openings
Automatic secondary sampler and manually 4.5. Salt Water Accessories
controlled composite sampler cutter cleaning by If salt water or other corrosive medium is used for
integrated water sprays dilution and washing, PSI 500i can be supplied with
Demagnetizer and dispersion options acid proof steel valves
3.3. Analytical 5. Installation
Size range: 1- 500 micrometers, size range can
be optimized according to application. 5.1. Max dimensions
Measurement data output: 40 fractions (%vol. PSI 500 analyzer unit:
±x µm or mesh), particle size percentiles (µm at - 2071 (H) x 1200 (W) x 930 mm (D)
x %), Blaine number (m2/kg), specific surface - 82” (H) x 47” (W) x 37” (D)
area (m2/cm3) Dilution water tank:
Precision: 1-2% (rel.) for Dv(50) particle size - 1304 (H) x 1022 (W) x 1206 mm(D),- 51 “ (H) x
based on a glass bead sample. 40” (W) x 47” (D)
Accuracy: when compared to other particle Water pump with cover
sizing methods than laser diffraction, differences - 520 (H) x 820 (W) x 320 (D) mm, 21 “ (H) x 32”
are noticed due to different influence of particle (W) x 13” (D)
shape, particle specific gravity etc. to the Pump control box 430 (H) x 300 (W) x 240 (D)
analysis. The system allows correlation of the mm ,17 “ (H) x 12” (W) x 9,5” (D)
measured particle size data to other
5.2. Weight
measurements by independent calibration
samples; however, this correlation is application PSI 500i analyzer: 250 kg, 550 lbs
dependent Dilution water tank (empty): 165 kg, 365 lbs
Measurement interval: typically 3 min / sample Dilution water tank (full): 1365 kg, 3010 lbs
line Water pump and control box: 35 kg, 80 lbs
6(7)
Figure 8 PSI 500i analyzer dimensions in mm. NOTE Analyzer with 300 mm optics requires 1200 mm width.
See installation manual for full details and instructions.
7(7)
5.10. Standards
Plant equipment conforms with European Union PSI® Particle Size Analyzer is a registered trademark
machine directive, low voltage directive and of Outotec Oyj.
EMC directive for heavy industrial environment
© 2011 Outotec Oyj. All rights reserved. Information
EMC: heavy industrial environment, EN 50082-2
is subject to change without notice.
& EN 50081-2
Environmental tolerance: IP56/NEMA 4X
Laser source meets Class II specifications