Courier® 6 HX Sampling Datasheet 10000004520e Revision A

Courier® 6 HX On-stream XRF Analyzer System

1. Sampling for Courier® 6 HX extraction and reduction, or zinc calcine leach,


solution purification and electrowinning.
The Courier 6 HX system analyzes samples of This datasheet introduces the standard sampling
process solutions from hydrometallurgical devices available, and some examples of typical
processes, using the field-tried high-performance sampling connections.
X-ray fluorescence technology to make elemental
analyses, supplemented by OTI titrators to 1.1. Sampling from the process
analyze ionic contents, such as acid. Hydrometallurgical Samplers HSA are available
which protect the other sampling equipment and
To bring to the analyzer solutions which may be the analyzer from process pressure.
hot, corrosive, liable to crystallize, and to remove
solids form the solution before analysis, a suite of Sample Transport
sampling devices have been developed. These
can be combined to bring samples to the analyzer Samples can be transported to the analyzer in a
reliably, and to assure the continued operation of number of ways. Courier 6 HX sampling design
the analyzer equipment, and the safety of plant minimizes the amount of sample discharged at the
personnel. analyzer location.

Figure 1 shows a number of typical sampling Samples from locations near to the analyzer can
setups which can be applied in processes such as be taken through a Controlled Direct Sample
copper heap-leach, solvent extraction and Connection CDS configured to operate safely at
electrowinning, copper electrorefining and impurity the actual process pressure.
removal, nickel/cobalt acid leach, solvent
PROCESS PROCESS
GRAVITY FLOW
IN PIPE

WATER
WATER
RETURN
RETURN

P
WATER
TANK OR CCS TANK OR
LAUNDER CLOSED CIRCUIT SET LAUNDER
WITH PUMPED CIRCULATION
(FAST LOOP) CDS CCS RETURN
TO ANALYZER CONNECTION DIRECT
GRAVITY FLOW CLOSED CIRCUIT SET
SET FROM SAMPLE WITH PUMPED CIRCULATION
IN PIPE
POINT TO ANALYZER (FAST LOOP)
TO SFS NEAR ANALYZER

WATER
AIR PUMPED CIRCULATION TO
SAMPLE FILTER SET
SDS NEAR ANALYZER
SAMPLE DISPATCH STATION
RETURN PNEUMATIC BATCH
TRANSPORT TO ANALYZER P
SFS
SAMPLE FILTER SET

SFS
SAMPLE FILTER SET

FREE FLOW TO
SAMPLE FILTER SET
AND PNEUMATIC
PUMPED WATER BATCH TRANSPORT PUMPED
FLOW TO ANALYZER FLOW
SAMPLE INLET

WATER
AIR RETURN
SFS

RETURN
FLS SDS
FAST LOOP SET SAMPLE DISPATCH STATION
FREE-FLOWING PNEUMATIC BATCH
OTI 16 OTI 16
CIRCULATION TRANSPORT TO ANALYZER
TITRATORS
(FAST LOOP)
TO ANALYZER WATER HSA
HYDROMETAL-
RETURN
LURGICAL
ELECTRICITY
SAMPLER

PROCESS PRIMARY SECONDARY SAMPLING / HXM 4540 Courier 6 HX PCS 4000 PRIMARY PROCESS
SAMPLING ANALYZER SHELTER SOLUTION ANALYZER PROBE SAMPLING
MULTIPLEXER PROBE CONTROL SET

Figure 1.Typical Primary Sampling for Courier 6 HX Analyzer System


2(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

A process pressure or gravity driven free-flowing operation at up to 1000 kPa (10 bar), 150 psi
Controlled Fast Loops Set FLS brings a process pressure.
continuous or semi-continuous flow of sample
2.2. HSA 2570 Hydrometallurgical Sampler
close to the analyzer, from which it takes a small
sample. HSA 2570 poppet type valve for sampling from
process pipe, diameter 63 mm and greater.
Controlled Circulation Pump Set CCS circulates • spring loaded, normally closed valve. The
the sample from the process sample point close effect of the process pressure is also to close
by the analyzer and back to the process. An SFS the valve, thus ensuring safe operation.
Filter Set may be integrated in the loop. • operated by a pneumatic actuator, which is
connected through a remote VBS solenoid
The discharge from a low pressure sample point box to the Courier 6 HX.
(a gravity pipeline or low pressure pump outlet) or • all AISI 316 construction, with TEFLON (R)
from an SFS filter set can be transported by a actuator rod seals.
pneumatic Sample Dispatch Station SDS. This • two sample ports: One connects the sampler
transfers a small quantity of clear solution, and to the Courier 6 HX sampling system, the
minimizes the amount of process solution that other is available for flushing the sample line
leaves the immediate process area. through the valve body.
Sample preparation The HSA 2570 is inserted through a flange into
The Sample Filtering Set SFS which is dedicated the pipeline
to one sample produces a small quantity of filtered
sample, returning the bulk of the sample intake to
the process.
Sample multiplexing
All samples are taken into the Courier 6 HX
analyzer through a Hydrometallurgical
Multiplexer HXM, located in the analyzer room or
shelter. This is described more closely in the
Courier 6 HX Analyzer System Datasheet.

2. Standard Sampling Devices


2.1. HSA Hydrometallurgical Samplers
HSA samplers are intended for sampling from
pump pressure lines where the pressure is higher
than the permissible inlet pressure of the Courier
6 HX HXM Hydrometallurgical Sample
multiplexer. These are typically pipes transporting
solution a long distance, or pipes feeding directly
into autoclaves or press filters. The HSA sampler
provides a reliable closure against process Figure 2. HSA 2570 Hydrometallurgical Sampler
pressure, and bypass connection for washing the
sample line and valve body even when the Dimensions
washing water pressure is lower than process • HSA 2570 (short), HSA 2571 (long):
pressure, or washing water cannot for some other − Flange DIN 25766 PN 10 DN 50, 165
reason be admitted to the process flow. (diam.)
The HSA valve closes the sampling connection in − 288 mm (h from flange) + 220 mm (max.
such a way that process pressure holds the valve depth from flange into pipe)
closed. The valves are fitted with spring-loaded − 11.4” (h from flange) + 8.7” (max. depth
single-acting actuators, so that their operation is from flange into pipe)
fail-safe. All wetted parts are acid-resistant steel, − weight 5.5 kg, 12.2 lbs
PVDF, or equivalent. HSA valves are designed for
3(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

Pressures The HSA 2580 is installed between flanges into


− Process pressure max.: 1000 kPa (10 the pipeline
bar), 150 psi Dimensions
− Control pressure min.: 450 kPa (4.5 bar),
• HSA 2580:
73 psi
− Flange DIN 25766 PN 10 DN 25, 115
Connections (diam.)
• Sample outlet of valve: 2 x R 1/2” DIN pipe − 115 mm (w) x 200 mm (l) x 242 mm (h)
thread, female − 4.5” (w) x 7.8” (l) x 7.8” (h)
• Control inlet: 6/4 mm connector for PVDF or − weight 5.5 kg, 12.2 lbs approx.
PA tube. Connect both inputs for added Pressures
protection against airborne contaminants
• Sample pipe connection: Pipe size − Process pressure max.: 1000 kPa (10
bar), 150 psi
− OD: 16 mm, 3/8 or 0.315 inch
− Control pressure min.: 450 kPa (4.5 bar),
− ID: 13 mm, 5/16 or 0.394 inch
73 psi
• Water supply
− Potable, sand-filtered water, pressure Connections
300 - 600 kPa, 45 - 90 psig • Sample outlet: 2 x R 1/2” DIN pipe thread,
Materials female
• Control inlet: 6/4 mm connector for PVDF or
• Wetted parts: acid resistant steel AISI 316 or
PA tube. Connect both inputs for added
equivalent
protection against airborne contaminants
• Actuator: plastic, max. temp 80° C
• Sample pipe connection: Pipe size
2.3. HSA 2580 Hydrometallurgical Sampler − OD: 16 mm, 3/8 or 0.315 inch
HSA 2580 plunger type valve for sampling from − ID: 13 mm, 5/16 or 0.394 inch
process pipe, diameter DN 25 mm. (min. 20 mm, • Water supply
max. 32 mm) − Potable, sand-filtered water, pressure
• spring loaded, normally closed valve. The 300 - 600 kPa, 45 - 90 psig
effect of the process pressure is also to close
Materials
the valve, thus ensuring safe operation.
• operated by a pneumatic actuator, which is • Wetted parts: acid resistant steel AISI 316 or
connected through a remote VBS solenoid equivalent, PVDF seals
box to the Courier 6 HX. • Actuator: plastic, max. temp 80° C
• two sample ports: One connects the sampler
to the Courier 6 HX sampling system, the 3. CDS Controlled Direct Sample
other is available for flushing the sample line Connection
through the valve body.
The CDS Controlled Direct Sample Connection is
used to deliver small quantities of clear sample
from a sample point to the HXM multiplexer for the
COURIER 6 HX system.

It is applied in the following circumstances:


• the sample will flow by gravity or by process
pressure to the analyzer location, and
• the pressure is within the range that can be
fed to the HXM multiplexer, and
• the distance from the sample point to the
analyzer is short and flushing the pipe from
the analyzer location to the sample point is
feasible.

Figure 3. HSA 2580 Hydrometallurgical Sampler


4(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

The CDS Controlled Direct Sample Connection is • the sample will flow by gravity or by process
a compact unit which can be installed in an pressure to the analyzer location, and
industrial environment close to the sample point. • the distance from the sample point to the
analyzer is very long and flushing the pipe up
It includes - to the analyzer location would use a large
• necessary valves for volume of process solution or take an
− sample intake and supply of washing unacceptably long time.
water and transport of the sample along a
pipe to the analyzer location The FLS Controlled Fast Loop Set is a compact
• a pressure sensor and manual pressure unit which can be installed in an industrial
regulators to ensure a suitable environment close to the sample point.

The CDS Controlled Direct Sample Connection is It includes -


controlled by the COURIER 6 HX analyzer and • necessary valves for
standard routines for taking and transporting − sample intake and supply of washing
samples, and for washing the flow system after water and transport of the sample along a
sample transport, can be chosen in the pipe to the analyzer location
configuration of the analyzer. • a pressure sensor and manual pressure
3.1. CDS 4580 Controlled Direct Connection regulators to ensure a suitable
Set
The FLS Controlled Fast Loop Set is controlled by
Pipe Connections the COURIER 6 HX analyzer and standard
• Minimum pressure : 75 kPa 11 psi routines for taking and transporting samples, and
• Maximum pressure : 250 kPa 38 psi for washing the flow system after sample
• Maximum temperature: 97 degrees C transport, can be chosen in the configuration of
• Solids content: no solids the analyzer.
• Pipe size 4.1. FLS 4581 Controlled Fast Loop Set
− OD: 8 mm, (5/16 or 0.394 inch)
Pipe Connections
− ID: 6 mm, (15/64 or 0.236 inch)
Sample Circulation
• Air supply
− Instrument air (dry, oil-free compressed • Minimum pressure : 100 kPa 15 psi
air) pressure 600 - 1000 kPa (90 - 150 • Maximum pressure: 1000 kPa 150 psi
psig) if required • Maximum temperature: 97 degrees C
• Water supply • Flow rate: 6 l/min, 1.5 US gpm
− Potable, sand-filtered water, pressure • Solids content: no solids
300 - 600 kPa, 45 - 90 psig • Pipe size
• Pressure lines
Materials − OD: 8 mm, (5/16 or 0.394 inch)
Wetted parts
− ID: 6 mm, (15/64 or 0.236 inch)
• Module blocks, valve bellows and pistons, • Gravity lines
piping, fittings: PVDF − Nominal diameter
• Valve seals, O-rings: Viton − 13 mm, 1/2”
Structural parts
− 16 mm, 5/8”
• Frame: Acid resistant steel AISI 316 − 25 mm, 1”
− 32 mm, 1 1/4”
4. FLS Controlled Fast Loop Set Outlet
The FLS Controlled Fast Loop Set is used to • Maximum back pressure: 200 kPa 30 psi
deliver small quantities of clear sample from a • Pipe size
sample point to a filtering device close to the − OD: 10 mm, 3/8 or 0.315 inch
analyzer or to the HXM multiplexer for the − ID: 8 mm, 5/16 or 0.394 inch
COURIER 6 HX system. • Air supply

It is applied in the following circumstances -


5(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

− Instrument air (dry, oil-free compressed standard routines for taking and transporting
air) pressure 600 - 1000 kPa (90 - 150 samples, and for washing the flow system after
psig) if required sample transport, can be chosen in the
• Water supply configuration of the analyzer.
− Potable, sand-filtered water, pressure 5.1. CCS 4590 for clear samples only
300 - 600 kPa, 45 - 90 psig Pump type
Materials • Diaphragm pump, electrically driven
Wetted parts • Capacity, max. 5.8 l/min, 1.67 US gpm,
• Module blocks, valve bellows and pistons, manually variable
piping, fittings: PVDF Pipe Connections - Sample Circulation
• Valve seals, O-rings: Viton • Maximum inlet pressure : 500 kPa 75
Structural parts psi
• Frame: Acid resistant steel AISI 316 • Maximum back pressure: 1000 kPa 150
psi
5. CCS Controlled Circulation • Maximum temperature: 97 degrees C
• Pipe size
Pump Set − OD: 8 mm, (5/16 or 0.394 inch)
The CCS Controlled Circulation Pump Set is used − ID: 6 mm, (15/64 or 0.236 inch)
to deliver small quantities of clear sample from a
sample point to a filtering device close to the 5.2. CCS 4591 for samples with solids, to feed
analyzer or to the HXM multiplexer for the SFS Sample Filter Set
Pump type
COURIER 6 HX system.
• Peristaltic (hose) pump, electrically driven
It is applied in the following circumstances - • Capacity 10 l/min, 2.88 US gpm, fixed
• the sampling point is below the level of the capacity
analyzer location, or the sample will not flow • Solids content: maximum 100 g/l (10%)
by gravity or by process pressure to the • Hose size 32 mm 1 1/4”
analyzer location, or Pipe Connections - Sample Circulation
• the distance from the sample point to the • Maximum inlet pressure : 500 kPa 75
analyzer is very long and flushing the pipe up psi
to the analyzer location would use a large • Maximum back pressure: 1000 kPa 150
volume of process solution or take an psi
unacceptably long time, or • Maximum temperature: 97 degrees C
• the sample contains solids and must be • Intake - Hose or AISI 316 pipe, nominal
filtered before analysis diameter
− 16 mm, 5/8”, or
The CCS Controlled Circulation Pump Set is a − 19 mm, 3/4”, or
compact unit which can be installed in an − 25 mm, 1”
industrial environment close to the sample point. • Outlet - Hose, Nominal diameter
− 16 mm, 5/8”, or
It includes -
− 19 mm, 3/4”
• a small peristaltic pump or diaphragm pump,
designed maintain a suitable circulation of 5.3. CCS 4592 Heavy Duty Set for samples with
sample solution solids, to feed SFS Sample Filter Set
• necessary valves for Pump type
− sample intake and supply of washing • Peristaltic (hose) pump, electrically driven
water and transport of the sample along a • Capacity 20 l/min, 5.76 US gpm, fixed
pipe to the analyzer location capacity
• a pressure sensor and manual pressure • Solids content: maximum 250 g/l (25%)
regulators to ensure a suitable • Hose size 40 mm 1 9/16”
Pipe Connections - Sample Circulation
The CCS Controlled Circulation Pump Set is • Maximum inlet pressure : 500 kPa 75
controlled by the COURIER 6 HX analyzer and psi
6(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

• Maximum back pressure: 1000 kPa 150 to the analyzer location would use a large
psi volume of process solution or take an
• Maximum temperature: 97 degrees C unacceptably long time.
• Intake - AISI 316 pipe nominal diameter 50
mm, 2” The Sample Dispatch Station SDS is a compact
• Outlet - Hose, Nominal diameter unit which can be installed in an industrial
− 25 mm, 1”, or environment close to the sample point.
− 32 mm, 1 1/4”
It includes -
Common Data • a collection vessel, designed to be easy to
Outlet to HXM wash clean after sample dispatch
• Maximum back pressure: 1000 kPa 150 • necessary valves for
psi − sample intake form the process or filtering
• Pipe size device
− OD: 8 mm, (5/16 or 0.394 inch) − sample reject to drain
− ID: 6 mm, (15/64 or 0.236 inch) − supply of washing water and or air or inert
gas for propelling the sample
• Air supply − transport of the sample along a pipe to
− Instrument air (dry, oil-free compressed the analyzer location
air), pressure 600 - 1000 kPa (90 - 150 • a sensor to detect the presence of solution
psig) if required • the necessary pressure regulators and filters
• Water supply for air and water supply
− Potable, sand-filtered water, pressure 300 • a sturdy frame and drip shield
- 600 kPa, 45 - 90 psig
Materials
Wetted parts
• Pumping head: AISI 316, PVDF, PP, PVC,
depending on sample
• Module blocks, valve bellows and pistons,
piping, fittings: PVDF
• Hoses of peristaltic pumps, NBR, for high
temperatures EPDM, other hoses available
on case by case basis
• Valve seals, O-rings: EPDM, VITON
Structural parts
• Frame: Acid resistant steel AISI 316

6. SDS Sample Dispatch Station


The Sample Dispatch Station SDS is used to
deliver small quantities of clear sample from a
sample point or a filtering device close to the
process to the HXM multiplexer for the COURIER Figure 4. Sample Dispatch Station SDS 4500
6 HX system.
The SDS is controlled by the COURIER 6 HX
It is applied in the following circumstances - analyzer and standard routines for taking and
• the quantity of sample available is small, or transporting samples, and for washing the flow
the sample must be filtered to remove solids system after sample transport, can be chosen in
the configuration of the analyzer.
• the sampling point is below the level of the
analyzer location, or the sample will not flow 6.1. Sample Dispatch Station SDS 4500
by gravity or by process pressure to the
analyzer location Dimensions
• the distance from the sample point to the • Overall
analyzer is very long and flushing the pipe up − 503 mm (w) x 467 mm (d) x 500 mm (h)
7(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

− 19.8” (w) x 18.4” (d) x 19.7” (h) Washing the flow system is performed
− Weight: 24.9 kg, 54.8 lbs automatically at suitable intervals.
Intake It includes -
• Minimum pressure: 25 kPa 4 psi • a bypass filter with an exchangeable filter
• Maximum pressure : 500 kPa 75 psi cartridge using commercially available filter
• Maximum temperature: 97 degrees C cloth
• Solids content: solids free • manual valves for
• Pipe size − shut off of the sample inlet from the
− OD: 8 mm, (5/16 or 0.394 inch) process or circulating flow
− ID: 6 mm, (15/64 or 0.236 inch) − shut of the sample exit to the process or
circulating flow
Outlet
− regulation of inlet pressure
• Maximum back pressure: 200 kPa 30 − adjustment of short circuit to reduce flow
psi through the filter, if the available flow
• Pipe size exceeds the filter capacity
− OD: 8 mm, (5/16 or 0.394 inch) • a pressure sensor
− ID: 6 mm, (15/64 or 0.236 inch) • a sturdy frame and drip shield
Air supply 7.1. Sample Filter Set SFS 4520
• Dry, oil-free instrument air • Overall:
(Inert gas e.g. nitrogen may also be used if − 459 mm (w) x 200 mm (d) x 1151 mm (h)
required) − 18.1” (w) x 7.9” (d) x 45.3” (h)
• Pressure − weight 18.5 kg 40.8 lbs
− max. 600 kPa, 90 psi
− min. 300 kPa, 45 psi
Water supply
• Potable, sand-filtered water,
− max. pressure 600 kPa, 90 psi
− min. pressure 300 kPa, 45 psi
Materials
Wetted parts
• Module blocks, valve bellows and pistons,
piping, fittings: PVDF
• Valve seals, O-rings: Viton
Structural parts
• Frame and drip shield: Acid resistant steel
AISI 316
• Support for vessel: HDPE Figure 5. SFS 4520 Sample Filter Set Dimensions

7. SFS Sample Filter Set Sample Circulation Connections


The Sample Filter Set SFS is used to produce • Maximum inlet pressure: 600 kPa 90 psi
small quantities of clear sample by filtration from a • Maximum exit pressure: 500 kPa 75 psi
circulation of solution containing solid particles at Maximum temperature: 97 degrees C
a point close to the process for delivery to the • Flow rate: 6 - 10 l/min, 1.5 - 2.5 US gpm
HXM multiplexer for the COURIER 6 HX system. • Solids content: max. 200 g/l 20%
• Pipe size
The Sample Filter Set SFS 4520 is a compact unit − Nominal 1/2” (12 mm) R 1/2” male thread
designed for wall mounting that can be installed in
an industrial environment close to the sample
point.
8(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

Clear Sample Outflow and Back-wash


Connection
• Maximum back pressure: 200 kPa, 30 psi
lower than exit pressure of sample circulation
• Pipe size
− OD: 8 mm, (5/16 or 0.394 inch)
− ID: 6 mm, (15/64 or 0.236 inch)
Air supply
• Instrument air (dry, oil-free compressed air)
pressure 600 - 1000 kPa (90 - 150 psig) if
required
Water supply
• Potable, sand-filtered water,
− max. pressure 600 kPa, 90 psi
− min. pressure 300 kPa, 45 psi
Materials
Wetted parts
• Module blocks, valve bellows and pistons,
piping, fittings: PVDF
• Valve seals, O-rings: Viton
• Filter capsule and support: HDPE, PP,
depending on application
• Filter cloth: depending on application
Structural parts
• Back plate: acid-resistant steel AISI 304
9(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

8. Typical Examples of Sample is 14 m, 46”. These parameters are suitable for a


CDS connection.
Connections If the pressure at the process is higher than 200
8.1. Connection Using a CDS Controlled Direct kPa , 30 psi, using an HSA Sampler to isolate the
Sample Connection HXM from possible pressure peaks in the process
is advisable. Compare Figure 7.
GRAVITY FLOW
IN PIPE

PRESSURE
FLOW IN PIPE
HSA
SAMPLER
WATER

MAX. 15 m
CDS WATER
DIRECT
CONNECTION
TO ANALYZER
MAX. 15 m
CDS
DIRECT
CONNECTION
TO ANALYZER

SAMPLE INLET

SAMPLE INLET

ANALYZER HXM 4540 Courier 6 HX


SHELTER SOLUTION ANALYZER
MULTIPLEXER PROBE

Figure 6. A Controlled Direct Sample Connection ANALYZER HXM 4540 Courier 6 HX


SHELTER SOLUTION ANALYZER
MULTIPLEXER PROBE
This connection (see Figure 6) is suitable where:
• the process pressure is within the limits for Figure 7. A Controlled Direct Sample Connection
the HXM inlet Using a HSA Sampler
• the distance from the sample points permits
washing the sample line either from the
analyzer to the sample point, or from the
sample point to the analyzer
For example the process pressure is 120 kPa (1.2
bar), 18 psi, and the sample point is 5 m, 16.5”,
above the analyzer. The specific gravity of the
solution is 1.25. The total head at the analyzer
location is 18.25 m, equivalent to 182 kPa, 27 psi.
The distance along the pipe from the sample point
10(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

8.2. Free-flowing Controlled Fast Loop to the If the pump pressure is sufficient to maintain a
Analyzer flow of about 6 l/min, 1.5 US gpm a 13 mm, 1/2’
This connection shown in Figure 8 is suitable if pipe the pressure at the analyzer location will be
the pressure or static head at the sample points is usually be within the limits for the HXM inlet.
sufficient to transport the sample to the analyzer,
so that the pressure at the analyzer location is It may be convenient to return the solution to a
within the limits for the HXM inlet. The bulk of the common point close to the analyzer, in which case
fast loop flow is returned to the process, e.g. to the connection is as shown in Figure 9. below
the pump feed tank or the another suitable return
point. The same connection can be used with an HSA
Sampler if the sample point is close to the
analyzer but the pressure is too high for the HXM
inlet, see Figure 10.
15 - 200 m

PUMPED WATER
FLOW
SAMPLE INLET

FLS
FAST LOOP SET
FREE-FLOWING
CIRCULATION
(FAST LOOP)
TO ANALYZER
RETURN

PROCESS PRIMARY ANALYZER HXM 4540 Courier 6 HX


SAMPLING SHELTER SOLUTION ANALYZER
MULTIPLEXER PROBE

Figure 8. Controlled Fast Loop Set sampling form a remote pump pressure line

15 - 200 m

PUMPED WATER
FLOW
SAMPLE INLET

FLS
FAST LOOP SET
FREE-FLOWING
CIRCULATION
(FAST LOOP)
TO ANALYZER
RETURN

PROCESS PRIMARY ANALYZER HXM 4540 Courier 6 HX


SAMPLING SHELTER SOLUTION ANALYZER
MULTIPLEXER PROBE

Figure 9. Fast loop from remote pump pressure line terminating at analyzer location
11(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

8.3. Pumped Circulation in a Fast Loop


The controlled pumped circulation shown in
Figure 11 can be used for remote samples from
PUMPED
FLOW
WATER points where the pump or static pressure is not
SAMPLE INLET
sufficient to transport the sample. Using this
connection only the small quantity of sample
actually used in the analyzer system is discharged
FLS
FAST LOOP SET
in the analyzer location.
FREE-FLOWING
CIRCULATION
(FAST LOOP)
TO ANALYZER
RETURN The controlled pumped circulation provides the
PROCESS PRIMARY ANALYZER HXM 4540 Courier 6 HX
possibility to flush the sample circulation either at
SAMPLING SHELTER SOLUTION
MULTIPLEXER
ANALYZER
PROBE regular intervals or at events like starting or
shutting down sampling from that point.
Figure 10. HSA Sampler and Fast Loop used to
protect analyzer and secondary sampling form
process pressure

PROCESS

WATER

RETURN
P

TANK OR
CCS
LAUNDER
PUMPED CIRCULATION
(FAST LOOP)
TO ANALYZER

SAMPLE INLET

ANALYZER HXM 4540 Courier 6 HX


SHELTER SOLUTION ANALYZER
MULTIPLEXER PROBE

Figure 11. Controlled Pumped Circulation feeding the Analyzer


12(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

This set-up is shown in Figure 12.


8.4. Removal of Solids
If the sample is available at sufficient pressure,
The SFS Sample Filtration Set can be used on its but remote from the analyzer location, the
own if the pressure drop available from the analyzer can be fed directly into an SFS Sample
sample point to the return point is sufficient to Filtration Set, from which the clear filtrate is fed
drive the filtrate through the filter, and the sample into an SDS Sample Dispatch Station, which can
point is so close to the analyzer that the filtered transport a small amount of clear sample to the
sample will flow into the HXM multiplexer. This is analyzer. This set-up is shown in Figure 13.
not always the case.
These two set-ups avoid any unnecessary
If the pressure at the sample point is not sufficient duplication of sample pressurization.
to filter the sample the sample can be transported
to the location of the analyzer in a CCS controlled
pumped circulation fast loop, which feeds the SFS
sample filtration set. This is so close to the
analyzer shelter that the filtered sample will flow
into the HXM without further pressurization.

WATER

RETURN

TANK OR
LAUNDER
SAMPLE INLET
CCS
CLOSED CIRCUIT SET
WITH PUMPED CIRCULATION
(FAST LOOP)
TO SFS NEAR ANALYZER

PUMPED CIRCULATION TO
SAMPLE FILTER SET
NEAR ANALYZER

SFS RETURN
SAMPLE FILTER SET

ANALYZER HXM 4540 Courier 6 HX


SHELTER SOLUTION ANALYZER
MULTIPLEXER PROBE

Figure 12. Filtering a sample from a point where pressure is low, using a Controlled
Pumped Circulation Set and Sample Filter Set
SFS
SAMPLE FILTER SET

FREE FLOW TO
SAMPLE FILTER SET
AND PNEUMATIC
BATCH TRANSPORT PUMPED
TO ANALYZER FLOW
SAMPLE INLET

WATER
AIR RETURN
SFS

RETURN
SDS
SAMPLE DISPATCH STATION
PNEUMATIC BATCH
TRANSPORT TO ANALYZER

HSA
HYDROMETAL-
LURGICAL
SAMPLER

ANALYZER HXM 4540 Courier 6 HX


SHELTER SOLUTION ANALYZER
MULTIPLEXER PROBE

Figure 13. Filtering a sample from a point under pressure in an SFS Sample Filter Set, and
transporting the clear filtrate using an SDS Sample Dispatch Station
13(13)

Courier® 6 HX Sampling Datasheet 10000004520e Revision A

Outotec quality assurance


Courier® 6 HX analyzer is CE marked according to the
European Union Directives.

The quality system of Outotec Oyj is ISO 9001


certified. The company has a quality management
policy to ensure that the expectations of
customers are met.

Outotec Minerals is a leading supplier of minerals


processing technologies including grinding,
flotation, thickening, physical separation, and
automation equipment. The company
manufactures a wide range of samplers for on-line
analysis, composite, and metallurgical sampling of
mineral slurries, solutions and dry powders.
You can find more information at
www.outotec.com