Professional Documents
Culture Documents
ACCG XVI eCAT
ACCG XVI eCAT
ACCG XVI eCAT
XVI R5.0
Customer Acceptance Tests
Referenced documents
Elekta does not supply all the documents that we refer to in this document with the equipment.
Elekta reserves the right to make the decision on which of the documents we supply with the
equipment.
Table of Contents
Chapter Title Page
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2 XVI Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
— Blank page —
1 Introduction
Section Title Page
— Blank page —
Note: If Elekta have to do other tests, that are not agreed as part of the Contract of Sale, there are
added costs.
Before you start the tests, make sure that the scale convention of the XVI system is the same as
the host digital accelerator. Refer to XVI Installation Information for more information.
After each test, record the result in this document in the space after the test procedure.
To do the tests in a minimum of time, Elekta recommends that you do them in the sequence of
this document. But, it is possible to do the tests in different sequences.
2 XVI Tests
Section Title Page
— Blank page —
Note: Elekta recommends that you give a minimum of 10 minutes between each scan.
002705_001
1 2 3
180º
4
000851 ©2013 Elekta Limited
Specification ≤1.5%
Note: If you have to comply with the Atomic Energy Regulatory Board (AERB) in India, do not do this
procedure but follow the procedure in the Addendum to the XVI Customer Acceptance Test. This is
applicable in India only.
1 In the G-T direction, align from the gantry end to the fourth alignment marker on the
phantom (see Setting up the Catphan® phantom).
2 Make a new patient ID and name it CAT-Uniformity (see the XVI Instructions for Use to
set up a patient).
3 Make a new treatment name and name it Uniformity.
4 Make a new field name and name it Uniformity.
5 Click Acquire VolumeView™ image.
6 Select the CAT - Image Quality Uniformity preset and acquire an image.
After you acquire the VolumeView™ image, reconstruction starts automatically.
7 Click Accept.
This accepts the reconstruction results.
8 Click View Reconstruction.
9 In the Slice Averaging drop-down list, select 3 slices, (the equivalent to 1.5 mm).
10 In the Transverse image area, select the uniformity module (see Figure 2.3).
Note: This is usually the slice number 32 for the Catphan® CTP 503.
2 cm
NOT TO SCALE 002706 ©2011 Elekta Limited
X Y Z 1 2
Note: Interpolated zooming makes the voxels in the Image probe dialog box appear to be smooth (see
the XVI Instructions for Use manual for more information).
13 To see the Transverse image area, move the Image probe dialog box to the right side of the
Transverse image area.
14 Zoom in until the box size in the Image probe dialog box is 1.00 cm.
a In the Transverse image area, move the mouse pointer approximately 1 cm from the
left edge of the image area.
b Click and hold the right mouse button, and drag down to zoom in until the box size is
1.00 cm.
c If it is necessary, adjust the contrast and brightness.
15 Click the center in the Transverse image area.
16 Record the Mean pixel value that appears in the Image probe dialog box in Table 2.1.
17 In the Transverse image area, click in five different positions of the image (see Figure 2.5).
Record the coordinates in Table 2.1.
5 4.5 cm
Z
2 4
3
001571 ©2011 Elekta Limited
See Formulae for low contrast visibility for more information on the formulae used in this
procedure.
1 To align the treatment table in the G-T direction and set up the Catphan® phantom, do one
of the steps that follow.
a Align the treatment table from the G-end with the first alignment marker on the
Catphan® phantom (see Setting up the Catphan® phantom).
b If you have a license for RATM, move the treatment table in the T-direction. For
Catphan® CTP 503 move the treatment table 11 cm (subtract from the current
position).
2 Make a new patient ID and identify it as CAT - Contrast.
3 Make a new treatment name and identify it as Contrast.
4 Make a new field and identify it as Contrast.
5 Click Acquire VolumeView™ image.
6 Select the CAT - Image Quality Contrast preset and acquire an image.
After the VolumeView™ acquisition is completed, reconstruction starts automatically.
7 Click Accept to accept the reconstruction results.
8 Click the View Reconstruction.
9 In the Slice Averaging list, select 3 slices.
10 In the Transverse image area, go to the contrast resolution module.
Note: This is usually slice number 32 for the Catphan® CTP 503.
11 Right-click in the Transverse image area, and select Pixel value Loc.
12 Use the PAGE UP and PAGE DOWN keys until the alignment markers are aligned vertically
and horizontally.
For Catphan® CTP 503 (see Figure 2.7):
– The Teflon rod (8) is in the top right position.
– The keyway (11) is at the bottom center position.
5 2 6
8
9
4 7
10
9 9
3 11 2 1
002735_001
This is an example of the top element of the equation, with the values from step 21 :
Note: Record the values in this test in full, this includes the numbers with a decimal point. For example
000.00.
The test measurements show that the XVI system can sense the difference between water and
fat. To do the tests, Elekta use a Catphan® phantom with the materials that follow:
• Polystyrene, to simulate water
• LDPE, to simulate fat.
An object is defined as visible when the average CT numbers of the object and the background
are different by more than the local average Standard Deviation (SD). The SD is a measurement
of the noise.
1 2
Pixel count
If the result of the above equation is greater than one, the object is visible.
Where:
• CT polystyrene is the mean CT number measured locally for polystyrene.
• CT LDPE is the mean CT number, measured locally for LDPE.
The formula to calculate the contrast in CT numbers is:
When the background is water, the formula to calculate the percentage contrast is:
The formula to calculate the percentage low contrast visibility of an object is:
Include the CT number for polystyrene and LDPE in this formula to get the value for low contrast
visibility:
If the result of the above equation is greater than one, the object is visible.
1 To align the treatment table in the G-T direction and set up the Catphan® phantom, do one
of the steps that follow.
a Align the treatment table from the G-end with the applicable alignment marker on the
Catphan® phantom. For the Catphan® CTP 503, align the treatment table with the
second alignment marker.
b If you have a license for RATM, move the treatment table 3 cm in the G-direction from
the alignment marker position (add to current position) on the Catphan® CTP503.
2 Make a new patient ID and name it CAT - Spatial.
3 Make a new treatment name and name it Spatial.
4 Make a new field name and name it Spatial.
5 Click Acquire VolumeView™ image.
6 Select the CAT - Image Quality Spatial preset and acquire an image.
After the VolumeView™ acquisition is completed, reconstruction starts automatically.
7 Click Accept.
8 Click the View Reconstruction button.
9 In the Slice Averaging list, select 3 slices.
10 In the Transverse image area, go to the spatial resolution module.
Note: This is usually slice number 32 for the Catphan® CTP 503.
11 Use the PAGE UP key, and count the slice number where the spatial resolution module does
not appear. Record the value in Table 2.5.
12 Use the PAGE DOWN key and count the slice number where the spatial resolution module
does not appear. Record the value in Table 2.5.
13 Select the slice in the middle of the recorded range of slices.
14 Zoom in or zoom out of the image until the spatial resolution module fills the image area.
15 Hold down the W key and move the pointer on line pair 10 until you get the best brightness
and contrast. For Catphan® CTP 503 (see Figure 2.11).
2
1
002709_001
16 Record the highest number line pair that you can see in Table 2.5.
17 Select Dismiss
This closes the VolumeView Reconstruction window.
Results
Slice number step 11 30
Slice number step 12 42
Spatial resolution (lp/cm) 12
1 Make sure that the S20 collimator and the F0 filter cassettes are in the kV source arm.
2 To align the treatment table in the G-T direction and set up the Catphan® phantom, do one
of the steps that follow.
a Align the treatment table from the G-end to the third alignment marker on the
Catphan® phantom. This alignment marker is applicable to a Catphan® CTP 503.
b If you have a license for RATM, move the treatment table 4 cm from the third
alignment marker position in the G-direction (add to current position).
3 Make a new patient ID and name it CAT - Geometric.
4 Make a new treatment name and name it Geometric.
5 Make a new field name and name it Geometric.
6 Click Acquire VolumeView™ image.
7 Select the CAT - Image Quality Geometric preset and acquire an image.
8 Click Accept.
9 Click the View Reconstruction button.
10 In the Slice Averaging list, select 3 slices.
11 In the Transverse image area, go to the contrast resolution module. This is usually slice
number 120.
12 Move the pointer to align with the top edge of the upper air insert.
13 Hold down the ALT key and the mouse button and drag the pointer to the top edge of the
lower air insert.
14 Release the ALT key and the mouse button, and measure the distance between the two air
inserts (see Figure 2.12).
117 mm
Air
117 mm
LPDE Delrin
Air
016118_001
Figure 2.12 Measure the distance between the two inserts - Catphan® CTP 503
15 Multiply the value on the screen by 10 to get the result in millimeters. Record the result in
Table 2.6.
1 Load the reconstructed images from the test of the 3D transverse vertical scale.
2 In the Transverse image area, go to the contrast resolution module. This is usually slice
number 120.
3 Move the pointer to align with the left edge of the Delrin insert.
4 Hold down the ALT key and the mouse button and drag the pointer to the left edge of the
LDPE insert.
5 Release the ALT key and the mouse button, and measure the distance between the two
inserts.
6 Multiply the value on the screen by 10 to give the result in millimeters. Record the result in
Table 2.7.
1 Load the reconstructed images from the test of the 3D transverse vertical scale.
2 Open the Sagittal image area (see Figure 2.13).
Note: It is only possible to see the alignment marker lines in the isocenter slice.
110
110 mm
Figure 2.14 Catphan® CTP 503 sagittal scale measurement (Distance of 110 mm)
4 If it is necessary, adjust the settings for contrast and brightness.
5 Multiply the value on the screen by 10 to give the result in millimeters. Record the value in
Table 2.8.
Accepted by: Flamarion Barbosa Goulart Signature: Barbosa Goulart Date: 2020.09.22 15:16:34
-03'00'
CAUTION 2.2
Do not rotate the gantry before you make sure that the path of the beam limiting device, kV
source arm, kV and MV detector arms is clear. If the path is not clear, a collision of the parts
and the treatment table can occur. If you ignore this caution, you can cause damage to the
equipment.
1 Do the setup for the tests.
a Put the ball bearing in a position that is an offset by 15 mm of the isocenter in all three
axis.
Note: The Table Move Assistant dialog box is set to one decimal point.
l On the FKP, press and hold the ENABLE and TABLE ASU keys on the FKP to move the
treatment table to the new position.
Note: A tolerance of >35 mm on the Table Move Assistant automatically disables the TABLE ASU and
ENABLE keys on the FKP. It is then only possible to move the treatment table from the controls in
the treatment room (see the XVI Instruction for Use manual).
XVI convert to
Manual XVI calculated shifts
Axis Direction corrections after Direction ≤ spec
offset (Test1)
move (Test 2)
cm mm cm mm
Lateral (X) 15 mm 1 0.48 A 0 0.1 A Y
Longitudinal (Y) 15 mm 1 0.47 G 0 0 - Y
Height (Z) 15 mm 1 0.48 U 0 0.7 D Y
This test uses the results from the Table Move Assistant accuracy tests.
1 Make sure that the accuracy tests for the Table Move Assistant are completed.
2 Do the test for the 3D Registration accuracy.
1 Copy the Test 2 results from the VolumeView™ treatment table accuracy tests to the
applicable results table in the Results of registration accuracy measurement section.
Note: For IEC 61217, +Longitudinal = G-direction, +Lateral = B-direction, +Height = Up. For IEC
60601-2-1, +Longitudinal = G-direction, +Lateral = B-direction, +Height = Down.
2 Make sure that you record the direction of error, for example, 0.23 mm in the A-direction.
Note: A license for IMRT is a prerequisite for this measurement. If you do not have an IMRT license,
acquire the images manually (see the XVI Installation Information manual).
17 Click OK.
The acquired IMRT images are exported to the specified folder.
18 On XVI, select the patient Bill Ballbearing, treatment 1, and field kV images.
19 Select Service > kV Service Functions > Flexmap Functions > Full Flexmap Calibration.
20 Do the on screen wizard instructions until the Isocenter Location window opens.
21 Click Repeat Scan.
22 Click Next.
The Movement Instruction wizard opens.
23 Record the values in the Movement Instruction wizard in the MV registration accuracy
table. Make sure that you record the direction of error. For example, 0.5 mm to A.
24 Click Cancel to close the Full FlexMap Calibration wizard.
25 Make sure that the MV registration accuracy tests are ≤1.0 mm.
26 Calculate the difference between the kV and MV registration accuracy results. Record the
result in the Difference between kV & MV column). Make sure that you record the direction
of the difference correctly.
Note: The VolumeView™ values are in centimeters and the MV values in millimeters.
27 Make sure that the difference between the kV and MV registration accuracy values are
≤ 0.5 mm.
Note: VolumeView™ show the isocenter values in centimeters and the MV system values in millimeters.
Make sure that you multiply the VolumeView™ values by 10 to get the results in millimeters.
Table 2.12 Results and difference between kV and MV registration accuracy measurements
Note: For IEC 61217, +Longitudinal = G-direction, +Lateral = B-direction, +Height = Up. For IEC
60601-2-1, +Longitudinal = G-direction, +Lateral = B-direction, +Height = Down.
Digitally signed by Flamarion
Flamarion
Flamarion Barbosa Goulart Barbosa Goulart
Accepted by: Signature: Barbosa Goulart Date: 2020.09.22 15:17:38
-03'00'
Difference Image
Image Gantry
Ball bearing center Image center center and ball
number angle (°)
bearing center
X Y X Y X Y
1 180º 518 510 518 509 0 1
2 90º 519 514 518 513 1 1
3 0º 518 511 517 510 1 1
4 -90º 515 509 515 507 0 2
This test is to show the 2D low contrast visibility of the full X-ray imaging chain.
1 Warm up the X-ray tube and make sure that it is at the correct temperature.
2 Put the TOR 18FG Leeds phantom (2) on the carbon fiber table top at the isocenter at 45° to
the room lasers. The top of the phantom is at isocenter (see Figure 2.16).
45º
1
Note: Make sure that the phantom is approximately at 45° to the room lasers (see step 2 ).
5 A
B
3
NOT TO SCALE 002714 ©2011 Elekta Limited
Note: The more disks you can see, the better the low contrast visibility is.
1 Load the image from the test of the 2D Low contrast visibility.
2 Set the brightness and contrast so that you see the two disks clearly in the squares (1) and
(2) (see Doing a test of the 2D Low contrast visibility).
3 Adjust the quality of the image until you can see the brightness and contrast boxes.
4 Monitor the resolution grid and count the number of spatial frequency groups that you can
see clearly (see Figure 2.18).
Note: A frequency group is a group of five bars. Count the frequency groups from the left column, and in
each column from the top down.
Count
Digitally signed by
Flamarion
Accepted by: Flamarion Barbosa Goulart Signature: Barbosa Goulart Date:
Flamarion Barbosa Goulart
2020.09.22 15:19:10
-03'00'
The installation and commissioning of the Elekta products were completed. The applicable tests in the customer
acceptance test plan were completed in compliance with the specification. A user manual (Instructions for Use)
was supplied.
Hospital
ACCG - Hospital Araujo Jorge
Address
Rua 239, n. 206 Setor Universitario
Country
Brazil
If the Customer Acceptance Tests are completed using an electronic Customer Acceptance Tests (CAT) document,
upload the approved electronic document to the Elekta Customer Lifecycle Management (CLM) system for
equipment registration.
For Customer Acceptance Tests using hardcopies, the BU or Distributor keeps two copies, and the customer keeps
one copy. A copy of the approved Certificate of Acceptance is uploaded to the CLM for equipment registration.
1 This is not the serial No. of the control cabinet. The serial No. for the system can be downloaded from CLM.
— Blank page —
Asia Pacific
T +852 2891 2208
F +852 2575 7133
Japan
T +81 3 6722 3800
F +81 3 6436 4231
China
T +86 10 5669 2800
F +86 10 5669 2900
elekta.com
/elekta
@elekta
/company/elekta