Download as pdf or txt
Download as pdf or txt
You are on page 1of 21

INTERNATIONAL JOU RNAL OF LABO RATO RY HEMATO LOGY

Spurious counts and spurious results on haematology


analysers: a review. Part II: white blood cells, red blood
cells, haemoglobin, red cell indices and reticulocytes
M. ZANDECKI, F. GENEVIEVE, J. GERARD, A. GODON

Haematology Laboratory, SUMMARY


University Hospital of Angers,
Angers, France Haematology analysers provide quick and accurate results in most
situations. However, spurious results, related either to platelets (part I
Correspondence:
M. Zandecki, Haematology Labor-
of this report) or to other parameters from the cell blood count (CBC)
atory, University Hospital of Angers, may be observed in several instances. Spuriously low white blood cell
4, rue Larrey, 49000 Angers, (WBC) counts may be observed because of agglutination in the
France. Tel.: +33 241 35 53 53; Fax: presence of ethylenediamine tetra-acetic acid (EDTA). Cryoglobulins,
+33 2 41 35 55 99; E-mail:
mazandecki@chu-angers.fr lipids, insufficiently lysed red blood cells (RBC), erythroblasts and
platelet aggregates are common situations increasing WBC counts. In
doi:10.1111/j.1365-2257.2006.00871.x most of these instances flagging and/or an abnormal WBC differential
scattergram will alert the operator. Several situations lead to abnormal
Received 30 January 2006;
accepted for publication 30 July haemoglobin measurement or to abnormal RBC count, including
2006 lipids, agglutinins, cryoglobulins and elevated WBC counts. Mean (red)
cell volume (MCV) may be also subject to spurious determination,
Keywords because of agglutinins, excess of glucose or salts and technological
Haematology analysers, automated considerations. In turn, abnormality related to one measured param-
count, cell blood count, spurious
count, white blood cells, haemo- eter will lead to abnormal calculated RBC indices: mean cell haemo-
globin, red blood cells, mean cell globin content (MCHC) is certainly the most important RBC indices to
volume consider, as it is as important as flags generated by the haematology
analysers (HA) in alerting the user to a spurious result. In many
circumstances, several of the measured parameters from CBC may be
altered, and the discovery of a spurious change on one parameter
frequently means that the validity of other parameters should be
considered. Sensitive flags now allow the identification of several
spurious counts, but only the most sophisticated HA have optimal
flagging and more simple HA, especially those without a WBC
differential scattergram, do not possess the same sensitivity for
detecting anomalous results. Reticulocytes are integrated now into
the CBC in many HA, and several situations may lead to abnormal
counts.

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41 21
22 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

scattergrams will be most prone to overlook several


INTRODUCTION
factitious counts.
Many haematology analysers (HA) that enumerate
WBC generate a WBC differential scattergram, which
Spuriously low WBC counts
is the basis for the automated WBC differential. Such
WBC scattergrams are also pivotal for the generation
Polymorphonuclear neutrophil aggregates
of flags or alarms, detecting that the WBC count (and
the WBC differential) is possibly erroneous and giving Polymorphonuclear neutrophil (PMN) aggregates may
some information about the possible origin of the be observed on blood samples drawn into EDTA. The
anomaly. As mentioned in part I of this report, the incidence is low but certainly underestimated, corre-
WBC differential scattergram is also of crucial import- sponding to 2/65 000 full blood counts in USA
ance to identify anomalies of platelet (PLT) counting, (Epstein & Kruskall, 1988), 1/9500 in Italy (Bizzaro,
PLT clumps being the most obvious example. 1993), 1/7500 in France (Lesesve et al., 2000). Both
Red blood cell (RBC) counts, haemoglobin (Hb), male and female show the anomaly. Although no
and mean (Red) cell volume (MCV) are other pathology or no specific disease is associated with
parameters measured by HA, which are also subject clustering of PMN, an acute or chronic inflammatory
to the spurious values in several situations. Red cell context (Luke, Koepke & Siegel, 1971; Savage, 1989;
(Wintrobe) indices other than MCV and packed cell Kahlil, 1991; Robbins, Conly & Oettinger, 1991;
volume (haematocrit; Hct) are usually calculated Imbing et al., 1996; Lesesve et al., 2000), liver diseases
from these measured parameters, and in most cir- (Epstein & Kruskall, 1988; Savage, 1989; Kobayashi
cumstances erroneous directly measured RBC param- et al., 1991; Vinatier et al., 1994; Imbing et al., 1996),
eters will subsequently cause in turn erroneous or circumstances associated with the generation of
calculated RBC indices. These indices can be used as cold agglutinins (Guibaud, Plumet-Leger & Frobert,
a flag for abnormal counts. Reticulocytes may be 1983; Epstein & Kruskall, 1988; Robbins, Conly &
considered now as a part of the CBC, and many HA Oettinger, 1991; Deol, Hernandez & Pierre, 1995;
have integrated reticulocyte counts under a fully Imbing et al., 1996; Lesesve et al., 2000) have been
automatized method. reported in many instances. The phenomenon may or
may not be transitory (Antonsen & Beyer, 1989; Bizz-
aro, 1993; Schinella, Kojikara & Curci, 1995).The
WHITE BLOOD CELLS
decrease may be either moderate or clinically signifi-
For blood cell analysis, each large sized particle cant, leading at times to suspect agranulocytosis and
(greater than the size of a PLT) that is not destroyed to generate unneeded investigations (bone marrow
by haemolytic agents will be identified as a WBC on aspiration or biopsy) or therapy (antibiotics; Epstein &
most HA. After enumeration, and according to the Kruskall, 1988; Vinatier et al., 1994; Schinella, Kojikara
type, impedance with low- and high-frequency elec- & Curci, 1995). There is no relationship between the
tromagnetic or direct current, laser light scattering (at phenomenon described here, which is a pure in vitro
one or at various angles), or peroxidase staining anomaly related to sampling in EDTA anticoagulant,
intensity are used, either individually or together, to and the in vivo anomaly occurring in diseases such as
generate a five-, six-, or even seven- part differential adult respiratory distress syndrome or leukostasis, in
(for review, see Bain & Bates, 2001). It is not in the which PMN tend to aggregate as a consequence of
scope of this report to study how the various WBC membrane interactions with complement (Jacob et al.,
are classified but it must be kept in mind that scatter- 1980).
grams generated by the HA to display the WBC differ- The mechanism leading to agglutination is not
ential must be fully understood by operators (Bain & fully elucidated (Carr et al., 1996). It is always an in
Bates, 2001). In many instances, WBC scattergrams vitro phenomenon, mainly EDTA dependent (Epstein
allow the detection of abnormalities related to spur- & Kruskall, 1988), although agglutination after the
ious counts and/or help to explain them. As a rule, use of sodium citrate or heparin as anticoagulants
instruments that do not generate WBC differential has also been reported in some instances (Rohr &
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 23

Rivers, 1990; Robbins, Conly & Oettinger, 1991; Vin-


atier et al., 1994; Deol, Hernandez & Pierre, 1995;
Carr et al., 1996). Using transfer experiments, either
plasma from the affected patients in contact with
WBC from normal patients or normal plasma in con-
tact with PMN from the affected patients, the anom-
aly has proved to be related to a plasma component,
and incubation of plasma with an anti-IgM antibody
or with dithiothreitol abolished or quantitatively
decreased PMN aggregation (Bizzaro, 1993). Using
flow cytometry, IgM was found on the PMN surface
in a patient (Carr et al., 1996). Some authors have
reported that the size of the aggregates was larger at
low temperature and that they disappeared at 37 C Figure 1. Aggregate of polymorphonuclear neutroph-
(Bizzaro, 1993; Carr et al., 1996). Several other ils observed in an infected patient.
authors reported that WBC agglutination in EDTA-
anticoagulated blood was not corrected by warming
the EDTA blood sample, at least in some cases, min- NY, USA; Abbott, Abbott Park, IL, USA). White blood
imizing the implication of a cold agglutinin in the cell aggregates are destroyed after drastic lysis of WBC
genesis of the clusters (Robbins, Conly & Oettinger, membranes (Galifi et al., 1993). Blood films may show
1991; Vinatier et al., 1994; Deol, Hernandez & Pierre, small (up to five cells), moderate (up to 50 cells), or
1996). A high level of integrin expression (CD11b– large (>100 cells) clusters of PMN (Figure 1). Careful
CD18) on PMN membrane was proposed to be rela- morphological examination of these aggregates shows
ted to the generation of PMN clumps (Galifi et al., that a few lymphocytes or monocytes may be at times
1993). entrapped within the aggregates (Hillyer, Knopf &
White blood cell differential scattergrams (using Berkman, 1990). Immature granulocytes (myelocytes,
either size and complexity or size and peroxidase con- metamyelocytes) and band cells are not infrequently
tent) may demonstrate anomalies, but it must be kept reported as a part of the clustered cells, and it was
in mind that the largest aggregates are overlooked, proposed that aggregates might develop around
and only the smallest will trigger flagging. On imped- myelocytes, whereas PMN alone failed to cluster
ance-type HA, the anomaly is suspected when events together (Deol, Hernandez & Pierre, 1995). Aggregates
are present above the area of PMN (top of the WBC of PMN are devoid of PLT, in contrast to PLT–PMN
graph), or when it is difficult to separate WBC classes aggregates (discussed in part I). Eventually, as aggre-
(Galifi et al., 1993), and the most frequent flags gener- gation of PMN in the presence of EDTA leads to a
ated using such HA are ‘immature granulocytes’ or reduction in their number, HA may generate flags
‘band cells’ (Lippi et al., 1994). On Bayer-Technicon corresponding to spuriously abnormal WBC differen-
HA, the smallest clusters may appear as a band of dots tials, e.g. spurious agranulocytosis or spurious lympho-
in the upper right of the WBC differential scattergram: cytosis (Epstein & Kruskall, 1988; Vinatier et al., 1994;
as these aggregates contain many PMN they are Schinella, Kojikara & Curci, 1995).
reported as peroxidase-rich particles, and an alarm Home-made anticoagulants were proposed to over-
corresponding to high peroxidase content is gener- come the agglutination (Schinella, Kojikara & Curci,
ated. On some HA, analysis of WBC nuclei is per- 1995). As mentioned above, warming the sample at
formed on a dedicated channel after the use of a 37 C may reduce both size and number of clusters in
drastic solution ensuring lysis of WBC membranes: some instances, but full disappearance is far from
number of free nuclei is used as a control for WBC being a consistent finding, and that method cannot be
count. Discrepancy between WBC count obtained proposed to overcome the anomaly. Finger prick and
from that channel and that from the fluorescence or immediate dilution of the blood sample prevents the
peroxidase channel is observed (Bayer, Tarrytown, agglutination.
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
24 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

containing excess of K3-EDTA (but not K2-EDTA)


Aggregation of WBC other than polymorphs
anticoagulant, resulting from insufficient blood drawn
in the presence of EDTA
after vein puncture (Goossens, Van Duppen & Verwil-
Clusters of normal (nonmalignant) lymphocytes were ghen, 1991).
reported in a patient with urinary tract infection, in a
patient with a B-cell lymphoma without bloodstream
Spuriously high WBC counts
involvement, in a patient with escarres, and in
another one with chronic myelomonocytic leukaemia
PLT aggregates and large platelets
(Deol, Hernandez & Pierre, 1995; Lesesve et al., 2001).
Lymphocyte aggregation was also reported to occur in Pseudoleukocytosis may be secondary to PLT clumps
chronic lymphocytic leukaemia patients, spontane- large enough to mimic WBC size (Solanki & Blackburn,
ously when lymphocyte counts are extremely high 1983; Cornbleet, 1983; Savage, 1984; Payne & Pierre,
(>400 · 109/l; O’Flaherty, Kreutzer & Ward, 1978), or 1984; Lombarts & de Kieviet, 1988; Schrezenmeier
not (Bizzaro & Piazza, 1991). Aggregates of three to et al., 1995). All modern HA that analyse WBC sub-
50 lymphoma cells were observed in two cases of populations detect this anomaly: PLT clumps are
Splenic Lymphoma with Villous Lymphocytes (SLVL; localized as a rocket shaped area of dots at the lower
Juneja et al., 1992; Imbing et al., 1995) and in one left hand corner of the WBC differential scattergram,
case to non-Hodgkin lymphoma mimicking SLVL and a flag corresponding to inability to discriminate
(Shelton & Frank, 2000). In these situations, the lar- among the WBC categories, namely lymphocytes,
gest aggregates are overlooked by the HA and WBC is generated (see Figure 2, part I). Several authors
count is spuriously but variably low. The smallest have reported that HA without WBC differential
aggregates may disturb WBC differential and may or scattergrams are unable to detect such anomaly
may not generate a flag, depending on the type of HA (Payne & Pierre, 1984; Lombarts & de Kieviet, 1988).
used (Shelton & Frank, 2000). In all cases reported so Some very large PLT may be encountered in myelo-
far EDTA was implicated, although small clusters of proliferative and in myelodysplastic disorders, whose
cells were also observed on heparinized samples size and volume may reach those of WBC and which
drawn as controls (Deol, Hernandez & Pierre, 1996) may be enumerated as WBC: flags are usually gener-
and sodium citrate did not appreciably change the ated, that mention the presence of pathological parti-
clustering tendency of the lymphocytes in one case cles that cannot be classified as WBC (usually ‘giant
(Shelton & Frank, 2000). Fingerpricking and immedi- PLT’, or ‘PLT aggregates’, depending on the HA).
ate dilution of blood seems the best way to avoid
aggregates (Lesesve et al., 2001). Heating (37 C) was
Nucleated red blood cells (NRBC)
reported to reduce clumping either partly (Shelton &
Frank, 2000) or had no significant effect on the size They may be found in the blood stream in physiologi-
of clumps (Lesesve et al., 2001). As the number of cal (newborns) and in pathological circumstances, and
cases reported is low, only hypotheses on the mech- at times they may be much more numerous than
anism(s) leading to lymphocyte agglutination have WBC. Using HA, these NRBC are in contact with lysis
been proposed, implicating various molecules such as agents that destroy their membrane, leaving nuclei
adrenalin (epinephrin), arachidonic acid, or leukotrien free, the latter being responsible for anomalies. Free
B4 (Villa et al., 1984; Shelton & Frank, 2000). In one NRBC nuclei are usually particles <40–50 fl, a size
instance aggregates involving all WBC classes was that is superimposable to that of PLT clumps or large
reported, in a patient with a long-standing history of PLT. Unsurprisingly, NRBC and PLT clumps disturb
alcohol abuse and alcoholic cirrhosis (Savage, 1989). WBC counts identically and both artifacts generate
equally the corresponding flags (Figure 2). According
to the HA, NRBC are either included within the WBC
Nature and amount of anticoagulant
count (instruments that do not generate any WBC dif-
A decrease in WBC count not related to agglutination ferential scattergram, but also some that do), or not
was reported in situations corresponding to samples enumerated as WBC, or counted separately. On
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 25

Figure 2. Nucleated red blood cells (NRBC ¼ 85 · 109/l) in a neonate (haemolytic anaemia): the WBC scattergram
(right) shows a large number of dots, corresponding to NRBC nuclei, as compared to a normal WBC scattergram
(left; Bayer Advia 120).

several recent and sophisticated HA, NRBC are specif- scattergrams may fail to detect the anomaly, and in
ically identified and may be enumerated using fluor- patients undergoing chemotherapy it may be of cru-
escence technology (Kim et al., 1998; Wang et al., cial importance, as it leads to normal or even elevated
2003). WBC counts in otherwise leukopenic patients: com-
parison with previous results (delta check) and exami-
nation of blood smear are necessary in such instances.
Red blood cells resistant to lysis
Some HA generate a message « RBC resistant to
Red blood cell may not be destroyed by lysis reagents lysis » and give access to an extended lysis mode
in various circumstances, either physiological (neo- which lengthens contact of blood cells with the lysis
nates) or pathological (abnormal haemoglobins, liver reagent (Abbott CellDyn 4000). On HA using a dedi-
disease, uraemia, chemotherapy): they are considered cated channel to enumerate WBC nuclei, discrepancy
as WBC and falsely increase the WBC count (Luke, between full WBC count and WBC nuclei count is
Koepke & Siegel, 1971; Cornbleet, 1983; Griswold & obvious: the channel analysing WBC nuclei uses lysis
Champagne, 1985; Vinatier et al., 1994; Mellors & agents strong enough to destroy all membranes (WBC
McArdle, 1995; Elghetany & Hudnall, 1996). White and RBC) and gives the accurate WBC count.
blood cell differential scattergrams demonstrate dots of
small size located near or within the lymphocyte win-
Cryoglobulins
dow, and inability to perform a WBC differential leads
to the generation of a flag. The picture is more or less The presence of cryoglobulin was first reported as
superimposable to that observed with the presence of causing erroneous WBC counts (Emori, Bluestone &
NRBC. This phenomenon seems to be a consistent Goldsberg, 1973; Taft et al., 1973; Haeney, 1976; Gul-
finding in haemoglobin C – containing RBC (either liani, Hyun & Gabaldon, 1977), but it also leads in
CC, Cbthal, or CS) but not in b thalassemic patients many instances to spuriously elevated PLT counts,
(either trait or major) or in haemoglobin S disease and at times to altered RBC counts or Hb measure-
(Booth & Mead, 1983). HA without WBC differential ments (cf., see Fohlen-Walter et al., 2002). Abnormal-
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
26 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

Figure 3. Several morphological aspects of cryoglobulins may be observed on peripheral blood samples, and two
among them are shown here: small lucent precipitates scattered around RBC (left), or invisible precipitates that
change morphology of RBC into a ‘moth-eaten’ aspect (right).

ities that occur on HA are neither consistent nor rela- an alarm is usually generated (see Figure 8, part I). If
ted to the amount of cryoglobulin or to its nature cryoglobulin particles cluster together or if their size is
(each type may be involved in spurious counts). How- larger, reaching that of leukocytes, WBC count is spu-
ever, anomalies generated must be considered because riously increased: a cloud of small dots may be
they have some peculiar characteristics, and their observed on the WBC differential scattergram, near
recognition may be the first clue leading to the diag- the lymphocytes or above the threshold for PLT, or at
nosis of cryoglobulinaemia (Zandecki et al., 1989; Foh- the top of the PMN cloud, depending on the HA used
len-Walter et al., 2002). Cryoprecipitates may be (Fohlen-Walter et al., 2002). When samples are
observed in some instances on stained blood smears warmed to 37 C for at least 30 min and reanalysed
(Figure 3), and in nearly all cases after examination of promptly anomalies disappear, but it is not a fully
fresh blood sample using a phase contrast microscope consistent finding and not infrequently a new sample
(more obvious if performed at low temperature). Var- must be taken and kept at 37 C until analysis to
ious morphological aspects were reported, including obtain accurate CBC counts (Zandecki et al., 1989;
dense amorphous clusters or flake-like particles, nee- Infanti et al., 1998; Figure 4). Hb determination and
dle-shaped crystals, and pinkish globules (cf., see Zan- RBC counts may at times be altered in the presence of
decki et al., 1989). As cryoglobulins are cryoglobulins (see later). In some situations cryoglob-
immunoglobulins that precipitate at temperature ulins may form a gel, leading to inaccurate aspiration,
lower than 37 C, abnormal counts are observed and an alarm may be generated on some HA (insuffi-
mainly on HA that both use reagents and perform cient sampling, or related).
analysis at room temperature. However, HA that use
prewarmed reagents, even with low pH reagents that
Cryofibrinogen or related
normally avoid precipitation of cryoglobulins, are not
free of anomalies and spurious counts are also Particles related to cryofibrinogen or to fibrin have
observed (Fohlen-Walter et al., 2002). According to been reported to elevate WBC counts and at times
their size, cryoprecipitates may only affect PLT counts, PLT counts and, according to the reports, PLT and
leading to spuriously elevated ones, up to eight times WBC counts may be found up to twice and up to 16
(Fohlen-Walter et al., 2002): histograms of PLT vol- times the accurate value, respectively (Gulliani, Hyun
umes show large number of small-sized particles, and & Gabaldon, 1977; Griswold & Champagne, 1985;
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 27

Corberand et al., 1991). Careful examination of the


corresponding blood smears shows thin fibrils, related
to fibrin strands after electron microscopic study
(Figure 5). These fibrils might correspond either to
cryofibrinogen, although biochemical research may be
negative in some instances (Griswold & Champagne,
1985), or to a mixture of cryoglobulin and cryofibri-
nogen (Gulliani, Hyun & Gabaldon, 1977). However,
analysis of new EDTA blood samples from the patients
frequently fail to reproduce the phenomenon,
hypothesizing that such particles might be the result
of fibrin polymerization after a difficult venepuncture,
initiating coagulation in vitro before contact of the
blood with EDTA (Corberand et al., 1991). After
warming the samples at 37 C abnormal counts are
either less pronounced or fully disappear.

Lipids
As previously discussed (part 1), lipids may generate
droplets large enough to disturb PLT counts, but also
WBC counts, together with or independently of PLT
Figure 4. Low grade B-cell lymphoma and cryoglobu- count. Abnormal WBC differential scattergrams are
lin type I (Ig M Kappa). Analysis at room tempera- generated, displaying various changes, at times more
ture (20 C; left) shows that WBC count was not or less superimposable to those outing from the
performed and that PLT count was inaccurate, and presence of cryoglobulins (Figure 6). Comments on
several flags were generated. After warming the lipids disturbing Hb determination will be discussed
same sample at 37 C and prompt analysis all abnor-
later.
malities disappear (37 C; right; STKS II Coulter).

Figure 5. In some instances (see text), fibrin strands may be observed on peripheral blood smears (MGG staining;
left); according to their size they may be enumerated as PLT or/and as WBC. Electron microscopic analysis (right;
a) shows such strands intermixed with PLT and WBC (arrows), and inset (right; b) shows peculiar periodic ultra-
structure corresponding to fibrin strands.

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
28 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

Figure 6. Lipids in a patient with liver disease. When compared with a normal scattergram (left), lipids generated
here a strand of dots elongating PLT region (arrow; Sysmex XE2100).

Microorganisms Overfilling blood collection vacuum tubes


Bacteria or fungi may be observed on peripheral blood As previously mentioned together with spurious PLT
smears and, as previously discussed (part I), may lead counts, overfilling the collection tube leads to inade-
to spuriously elevated PLT counts. They were also quate sample mixing and setting of cellular contents:
reported as disturbing WBC differentials if they all the parameters are altered (Pewarchuk, Vander-
clumped together, and in vitro studies adding large Boom & Blajchman, 1992).
number of such microorganisms were shown to
increase WBC count (Marshall et al., 1990).
Concluding remarks on spurious WBC counts
Haemozoin contained in PMN and monocytes in
malarial cases may cause depolarization of the laser As for spurious PLT counts, it is not always possible to
light and subsequently misclassification of WBC types find or to identify the phenomenon that leads to spur-
on some HA (Sysmex, Kobe, Japan), although full ious WBC counts, even if something is visible on per-
WBC count does not clearly change (Huh et al., 2005) ipheral blood smears (Savage, 1989). The WBC
scattergram from HA is useful and must be interpreted
carefully; examination of stained films must also be
Adipose tissue
considered.
Total WBC count was reported as artefactually eleva-
ted, caused by contamination with subcutaneous adi-
HAEMOGLOBIN, RBC COUNT AND RBC
pose tissue of the blood sample obtained by traumatic
PARAMETERS
femoral vein puncture: peculiar scatter plots also dem-
onstrated an abnormal WBC differential (Whiteway & HA measure several RBC parameters, including RBC
Bain, 1999). count and MCV both determined on the same chan-

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 29

nel(s), and Hb concentration which is measured on Jimenez, 1996). After the study of various types of
another channel(s). Haematocrit (Hct), mean cellular hyperlipoproteinemias, it was observed that erroneous
Hb (MCH), and mean corpuscular Hb concentration Hb and MCHC >36 g/dl were observed in patients
(MCHC) are usually calculated from the measured with at least 20 g/l of triglycerids, corresponding to
parameters. Circumstances leading to erroneous meas- type I and part of type V hyperlipoproteinemias (rich
ured RBC parameters will subsequently generate in chylomicrons) but not to type IV hyperlipoprote-
abnormal calculated RBC indices. So, an anomaly of inemias (rich in very low density lipoproteins; Gagne
at least one of the RBC indices, beside its pathophysi- et al., 1977). Samples taken after a meal may at times
ological significance, must also be considered as a flag demonstrate superimposable spurious Hb measure-
for the detection of an error of at least one among the ment. Even the most recent HA are sensitive to
red cell parameters measured (RBC, MCV, Hb). As hyperlipemia, although to a variable extent, as for
well as the classical Wintrobe RBC indices, other example Abbott Cell Dyn 4000 that is defined as giv-
parameters derived from the technology have been ing true Hb values for up to 13 and 9 g/l of triglycerid
introduced, which differ according to the HA, but and cholesterol levels, respectively (M.C. Chrieten,
their use may also be helpful for the detection of personal communication). Although excess of lipids
abnormal counts or measurements. One example is usually spuriously increases Hb, a spurious fall of Hb
the assessment of Hb concentration of individual RBC, was also reported once (Savage, 1989). Similar to the
also named cellular haemoglobin concentration mean presence of cold agglutinins, anomaly related to lipids
(CHCM), which is subsequently compared with the is suspected when MCHC is >36 g/dl, or when WBC
calculated MCHC, generating a flag for discordant data scattergrams demonstrate high number of particles of
(Abbott, Bayer). low to moderate size (Figure 6). For laser-beam HA
that measure Hb level within each RBC and generate
the so-called « measured MCHC » or CHCM, a differ-
Haemoglobin
ence between the CHCM and the calculated MCHC is
Hb concentration is measured using a colorimetric observed. Various methods have been proposed to
(spectrophotometric) method, either by a modification obviate the abnormality on the relevant sample,
of the manual cyanmethaemoglobin method or by including isovolumetrical replacement of hyperlipemic
addition of various reagents, such as sodium lauryl plasma with isoosmotic diluent, or ether extraction of
sulphate or imidazole. According to the manufactur- lipids. As previously mentioned, such methods may in
ers, various nonionic detergents are included to insure turn lead to erroneous PLT and WBC counts.
rapid RBC lysis and to reduce turbidity because of cell
membranes and plasma lipids.
High WBC counts
White blood cell may induce excessive turbidity and
Lipids and hyperchylomicronemia
disturb Hb measurement if their number is sufficiently
Lipemia may cause erroneous PLT (see part I) and high. There is no clear-cut threshold for WBC counts
WBC counts (see above), but may also induce inter- associated with spuriously elevated Hb, but one must
ferences by turbidity (Creer & Ladenson, 1983; Sand- be careful with all samples with WBC count over 50
berg, Sonstabo & Christensen, 1989). An abnormally or 100 · 109/l (Cornbleet, 1983; Sandberg, Sonstabo
high MCHC (>36 g/dl) corresponding to an errone- & Christensen, 1989), although some HA define
ously high Hb has been reported for patients with threshold at 250 · 109/l or even seem to be fully
severe constitutional or acquired hypertryglyceridemia insensitive to WBC count because of entire WBC lysis
(Gagne et al., 1977; Mayan et al., 1996), and for performed before Hb measurement (Sysmex). As for
patients receiving intravenous administration of fat lipid disturbance, CHCM determined on some HA
emulsions (Nosanchuck, Roark & Wanser, 1974; may help to find the true Hb value (McVeigh, Faim &
Shah, Patel & Rao, 1975; Nicholls, 1975, 1977; Creer van der Weyden, 1989). However, RBC count may be
& Ladenson, 1983; Artiss & Zak, 1987; Sandberg, also disturbed by high WBC counts (see later), and
Sonstabo & Christensen, 1989; Cantero, Conejo & one must pay attention to the limits in recalculating
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
30 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

RBC indices, and a packed cell volume (centrifuged 2002). However, spurious Hb measurement is far from
Hct) should be considered. being a consistent finding in the presence of cryoglob-
ulins (Fohlen-Walter et al., 2002).

Immunoglobulins
Haemolysis
They have been reported to interfere with a number
of clinical laboratory tests (cf., see Roberts, Fontenot Hb free within plasma is measured together with that
& Lehman, 2000). False elevation of Hb measurement from the RBC, but its amount ranges from 10 to
by automated methods was observed for several 40 mg/l in normal conditions and does not affect total
patients with Waldenström’s macroglobulinemia and Hb measurement. However, in situations related to
monoclonal IgM, or with multiple myeloma and major intravascular haemolysis, including chemicals,
monoclonal IgA or IgG (Wallis & Ford, 1987; McMul- mechanical haemolysis associated with heart valves,
lin, Wilkin & Elder, 1995; Roberts et al., 2000). This and haemolytic anaemias associated with blood trans-
anomaly is related to high levels of Ig that interact fusion, free plasma Hb may be elevated enough to
with reagents of the lysis solution. For IgM, this phe- affect total Hb measurement. MCHC may be >36 g/dl.
nomenon has been related to the amount of mono- Centrifuged Hct shows a pink or red plasma tinge,
meric component within the circulating paraprotein namely if free plasma Hb is >200 mg/l, and is, in some
(Goodrick et al., 1993). HA employing Hb conversion instances, the only reliable RBC parameter. Some
to cyanmethaemoglobin seem to be more affected laser-beam HA directly determine Hb within each
than others, because of addition of surfactants for RBC, the Cellular Haemoglobin Concentration Mean
cyanide-free methods. Red blood cell count and MCV (CHCM), which allows the accurate Hb value to be
are unaffected in this situation but, as Hb is overesti- calculated. A short time lapse between venepuncture
mated, MCHC usually exceeds 36 g/dl. In order to and analysis is of crucial importance because haemo-
obtain more accurate results in this instance, it was lysis may continue in vitro, leading to a spurious
proposed to determine ‘plasma Hb’ after centrifugation decrease of RBC count and total Hb with a spurious
of the sample (which gives turbidity because of the increase of free plasma Hb.
paraprotein) and to withdraw it from Hb of the whole
blood (Roberts et al., 2000). For Sysmex instruments,
Chemical structure of haemoglobin and bilirubin
it seems that if the sample is half diluted by the oper-
ator before analysis on the HA, the phenomenon does In physiological situations, Hb is more or less coupled
not occur (Roberts et al., 2000). For laser-beam HA to oxygen or to carbon-di-oxide and, according to
that determine the measured MCHC (CHCM) a clear- which molecule is coupled to Hb, the peak of optimal
cut difference with the calculated MCHC generates an light absorbance differs slightly. The addition of
alarm, and accurate Hb from the sample may be several reagents leave Hb free from coupled molecules
obtained from measured MCHC. and changes it into one stable molecule (cyanmethae-
moglobin is an example), the latter demonstrating a
narrow peak of light absorbance, which allows accu-
Cryoglobulins
rate determination of Hb concentration. However,
Spurious Hb values have been reported in some high amounts of carbon monoxide coupled to Hb
instances, and several mechanisms are proposed to may not be fully transformed, and in such situations
explain these abnormal findings. In some cases spuri- a spuriously high Hb concentration is reported
ously high Hb values were related to a mechanism (Cornbleet, 1983; Vinatier & Flandrin, 1993). In con-
similar to that described above for immunoglobulins trast, sulfhaemoglobin in high amount has been
(Taft et al., 1973; Cornbleet, 1983) or to the distur- reported as lowering Hb measurement (Cornbleet,
bance of light transmittance, whereas in other cases a 1983). Bilirubin: although one must pay attention to
slight decrease of both Hb measurement and RBC very high amounts of bilirubin within the plasma,
count was related to a flow anomaly (Taft et al., 1973; most HA do not presently demonstrate any interfer-
Bremmelgaard & Nygard, 1991; Fohlen-Walter et al., ence with bilirubin, at least for concentrations up to
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 31

250 mg/l. Above theses values, however, attention is MCV, and subsequently lead to abnormal calculated
needed. MCHC (discussed later).

Spurious RBC counts and red cell parameters Giant platelets


According to the technology used, spurious results High number of giant PLT may lead to spuriously low
may or may not be observed using some HA, and PLT count (see part 1) but, as they are enumerated as
some knowledge about the methods analysing blood RBC, they may also affect RBC count in a way similar
parameters is necessary. On impedance – type HA an to that for WBC; however, RBC count is usually only
aliquot of the blood sample is diluted isoosmotically slightly affected in this instance (Cornbleet, 1983;
and the number and height of electric pulses gener- Bain & Bates, 2001).
ated by the electrical resistance of RBC that pass
through a small orifice allow the determination of
Spuriously decreased RBC counts
both the RBC count and the MCV. Using laser-beam
methods, scattering at least at two angles allows the
Cold agglutinins
determination of RBC count and MCV. In order to
improve accuracy, pretreatment of RBC with a specific Cold agglutinins aggregate RBC when the temperature
reagent that changes them isovolumetrically from dis- is lower than 37 C. Unsurprisingly, peculiar anomal-
coid to a sphere is performed on some HA. Setting dis- ies of RBC parameters in the presence of cold aggluti-
crimination thresholds is an important consideration: nins were reported first on counters acting at room
discriminating the smallest RBC from the largest PLT temperature (Hattersley et al., 1971; Petrucci, Duanne
is at times a challenge, already discussed (part I). On & Chapman, 1971; Bessman & Banks, 1980). Accord-
the contrary, even in extreme pathological situations, ing to the HA, the upper threshold that may consider
MCV does not exceed 150–160 fl and, as there is no particles as RBC into the RBC channel(s) is located
particle above that size in the blood stream in health between 200 and 300 fl. So, only particles correspond-
or in disease, HA do not analyse any particle above ing either to isolated RBC or to small RBC clumps
200–300 fl in volume. (two or three RBC), are analysed, whereas large RBC
clumps are fully neglected by the HA. This leads to
spuriously low RBC counts and to abnormally high
Spuriously elevated RBC counts
MCV (each small RBC clump is considered as one sin-
gle particle). Haematocrit (RBC · MCV) is erroneous
High WBC counts
and spuriously low, contrasting with Hb that is meas-
Most HA enumerate RBC and WBC together within ured after RBC lysis and is unaffected by agglutinins.
the same channel(s), and the RBC count reported is As a rule, the MCHC is spurious, usually >36 g/dl. The
the sum of both RBC and WBC counts. In physiologi- peculiar association of low RBC count, high MCV and
cal conditions, it is not of any importance, as it corres- MCHC >36 g/dl is almost pathognomonic of the cold
ponds to overestimate RBC count by 0.1% (if we induced artifact on instruments that work at laborat-
consider a WBC count of 5 · 109/l and a RBC count ory temperature, and not infrequently helps to diag-
of 5 · 1012/l). However, high WBC counts nose the cold agglutinin in the patient analysed
(>100 · 109/l) may lead to a significant change in the (Petrucci, Duanne & Chapman, 1971). HA working
RBC count, particularly if the patient is also anaemic with reagents at temperatures near 37 C are not fully
(Bain & Bates, 2001). Moreover, in the latter insensitive to cold agglutinins, however, but changes
instances the MCV reported corresponds to the mean are less obvious and may remain undiscovered in some
volume of RBC and of WBC from the sample and instances (Solanki & Blackburn, 1985). HA that mea-
may be spurious, variably according to the nature and sure directly the amount of Hb within each RBC (mea-
the number of WBC from the relevant sample. So, sured MCHC, or CHCM) usually show discrepancy
high WBC counts may induce several abnormal find- between the measured MCHC and the calculated one.
ings, including Hb (discussed above), RBC count, Whatever the HA used, the common finding is that Hb
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
32 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

value is unaffected and that anomalies disappear when


In vitro haemolysis
the sample is warmed at 37 C and analysed promptly
afterwards. Amplification of the anomalies after the As mentioned above, in some situations, namely hae-
sample has been cooled at 4 C for 1 or 2 h reinforces molysis related to chemicals or blood transfusion, RBC
the diagnosis. As viscosity of the sample may be high, may continue to lyse within the sample, leading to
leading to inaccurate aspiration, an alarm may be spuriously low RBC count, together with abnormal
generated on some HA (insufficient sampling, or Hb measurement and abnormal red cell indices (see
related). Co-existence of RBC agglutination with Hb measurement).
EDTA-dependant thrombocytopenia was reported but
antibodies directed against RBC and those directed
Mean cell volume and haematocrit
against PLT differed (Bizzaro & Fiorin, 1999).
Some situations leading to spurious MCV (see Table 1)
are already discussed above [high WBC counts, cold
Warm autoimmune haemolytic anaemia
(warm) agglutinins]. Spurious Hct is a frequent find-
In some instances warm autoantibodies were also ing, related to abnormal MCV or/and to abnormal
reported as inducing RBC agglutination, leading to RBC count as it is a parameter that is usually calcula-
spurious MCV and RBC count, in a situation superim- ted by the HA: situations leading to abnormal Hct,
posable to that observed for cold agglutinins, but not either increased or decreased, will not be discussed in
reversible by warming (Weiss & Bessman, 1984). details as they may be deducted from MCV changes
(Table 1).

Very small RBC and discrimination with PLT


MCV and the technology used for measurement
As discussed in part 1, PLT counts may be disturbed
by microcytic cells, namely if RBC volume is <36 fl: Using impedance-type HA, disc-shaped RBC become
some HA consider particles over this threshold as elongated into a cigar shape as they pass through the
RBC, and in severe microcytic anaemias PLT count aperture. Elongation is more pronounced for particles
may be altered. Underestimation of RBC counts may flowing at the periphery of the orifice than at the cen-
be observed in turn, because of the elimination of the tral part, and also when Hb content is low within the
smallest erythrocytes from the RBC count. In such RBC. Mean (red) cell volume is spuriously lowered
instances up to 6% lowered RBC counts were repor- and MCH is spuriously elevated, mainly in hypochro-
ted (Savage et al., 1985). mic disorders, and for MCV of 55 fl a spurious
decrease of up to 7 fl may be observed (Mohandas
et al., 1980; Paterakis et al., 1994; Bain & Bates, 2001).
Cryoglobulins
Hydrodynamic focusing, a technical change that
As previously mentioned, a slight decrease of both Hb directs the flow of particles at the centre of the aper-
measurement and RBC count was reported in a few ture, improves MCV measurement. For laser-beam
instances, related to a flow anomaly (Taft et al., 1973; HA, the discoid shape of RBC was a major drawback
Cornbleet, 1983; Bremmelgaard & Nygard, 1991; Foh- for peer analysis of RBC until the isovolumetric spher-
len-Walter et al., 2002). isation was applied, allowing current instruments to
generate accurate MCV (Mohandas et al., 1986). Some
impedance-type HA use both technical improvements
Considerations related to the quality of the sample.
to analyse RBC (Abbott), whereas others measure the
Clotting, or abnormal mixing secondary to overfilling Hct and calculate MCV using the HCT and RBC (Sys-
of the sample (Pewarchuk et al., 1992), may lead to mex; ABX, Kyoto, Japan; some Beckman instruments,
an aspirated aliquot unrepresentative of the blood Miami, FL, USA). In that latter instance however,
sample. Cryoglobulins and cold agglutinins may lead the presence of cold agglutinins, which disturbs RBC
to high viscosity of the sample, leading to inadequate count (see above) leads to erroneous MCV (Sysmex
aspiration. XE-2100 operator’s manual – revised April 2004).
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 33

Table 1. Situations leading to erroneous results on haematology analysers

Other parameters altered

WBC: spurious decrease


Agglutination of PMN (EDTA - related)
Agglutination of WBC other than PMN
(lymphocytes, lymphoma cells, leukaemic blasts)
Excess amount of K3-EDTA anticoagulant Shrinkage of RBC: MCV and Hctfl
Coagulation within the sample All parameters
WBC: spurious increase
PLT aggregates PLTfl
(Very) large PLT PLTfl
Nucleated red blood cells
RBC resistant to lysis (newborns, abnormal Hb,
chemotherapy, uraemia, liver disease…)
Cryoglobulin, cryofibrinogen, immunoglobulins PLT›
Lipids PLT›, Hb›, MCH›
Microorganisms (bacterial aggregates) PLT›
Others (adipose tissue, overfilling vacuum tubes) All parameters
RBC: spurious decrease
Cold agglutinins, warm agglutinins MCV›, MCH›
Very small RBC PLT›
Cryoglobulins (fl flow, inadequate aspiration) Hbfl, WBCfl, PLTfl
In vitro haemolysis Hb›, MCH›
Coagulation All parameters
RBC: spurious increase
High WBC counts MCV›, Hct›, MCH›
Giant PLT PLTfl
Haemoglobin: spurious increase
Lipids WBC›, Hb›, MCH›
High WBC counts MCH›, RBC›
Immunoglobulins (and cryglobulins) WBC›, PLT›, MCH›,
In vitro haemolysis MCH›
Carboxyhaemoglobin (high amount)
Bilirubin (>250–300 mg/l) MCH›
Haemoglobin: spurious decrease
Coagulation within the sample All parameters
Overfilling vaccum tube All parameters
Veinipuncture near a drip MCV› (glucose drip)
Sulfhaemoglobin
MCV
Cold agglutinins, warm agglutinins: › MCH›, RBCfl, PLTfl
High WBC counts: › RBC›
Hyperglycemia: › MCH›
K2 EDTA in excess: › MCHfl
Hyper- or hyponatremia: › or fl MCHfl or ›
Technology: impedance without hydrodynamic focusing MCH›
(MCVfl in hypochromic anaemias)
MCH > 36 g/dl (not related to spurious counts in some disorders: spherocytosis, xerocytosis, abnormal Hb: see text)
Cold agglutinins, warm agglutinins RBC, MCV›
Lipids WBC›, Hb›
Immunoglobulins Hb›
In vivo and in vitro haemolysis Hb altered, Hct
Carboxyhaemoglobin (>10–20%)
Bilirubin (>300 mg/l, at times less)
Immunosuppressive drugs (see also spuriously high Hb, spuriously low MCV)
MCH <32 g/dl
Hyperglycemia (see also spuriously high MCV, spuriously low Hb) MCV›

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
34 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

RBC RBC

50 100 200 300 fL 50 100 200 300 fL

12 12
RBC = 1. 99 10 /l Hb = 6.4 g/dl RBC = 3.78 10 /l Hb = 12.5 g/dl
MCV = 141.5 fl MCH = 22.9 g/dl MCV = 97.1 fl MCH = 34.1 g/dl

Figure 7. Venepuncture performed near a glucose infusion. Blood sample from that patient was diluted (Hb low)
and excess of glucose led to a swelling of RBC: MCV was spuriously high and in turn MCHC spuriously low. Sam-
ple drawn correctly by the next morning showed normal values (no transfusion had been performed). RBC histo-
gram on diluted sample (left) and on new sample (right; Beckman Coulter STKS II).

present (blood drawn near a glucose drip). Studies


MCV and high WBC counts
performed on RBC at various concentrations of glu-
As mentioned above, WBC are counted together with cose have shown that within a few seconds after dilu-
RBC, and MCV reported is the mean of both RBC and tion into the HA (diluents from the manufacturers)
WBC volumes. In situations related to high WBC water penetrates within RBC rich in glucose, inducing
counts and low RBC counts (e.g. acute leukaemias), swelling: return to a normal size is progressively
MCV is altered. Up to a 15–20 fl increase may be observed and is complete after up to 5 min. According
observed in some instances. HA which calculate MCV to the time spent between dilution and MCV meas-
using the HCT and RBC (Sysmex, ABX, some Beck- urement, the anomaly will be more or less conspicu-
man instruments) are also sensitive to leukocytosis ous (Holt, DeWandler & Arvan, 1982). The anomaly
over 100 · 109/l (Sysmex XE-2100 operator’s manual is more obvious if RBC are isovolumetrically sphered,
– revised april 2004). because stabilization of RBC volume occurs within a
few seconds after the dilution of the sample (Van
Duijnhoven & Treskes, 1996). For the above-men-
Severe hyperglycemia
tioned reasons HA are more or less sensitive to this
Mean (red) cell volume is reported as spuriously high phenomenon, but glucose concentrations over
in blood samples containing high levels of glucose, 20 mmol/l may begin to increase MCV, and blood
either related to severe hyperglycemia in the relevant glucose >35 mmol/l may overestimate MCV up to
patient or to a sample, which has been drawn near an 50 fl (Holt, DeWandler & Arvan, 1982; Planas, Van
intravenous glucose infusion (Morse et al., 1981; Voolen & Kelly, 1985; Van Duijnhoven & Treskes,
Strauchen et al., 1981; Holt, DeWandler & Arvan, 1996; Figure 7). Increased MCV leads to an increase
1982; Evan-Wong & Davidson, 1983; Savage & Hoff- in the calculated Hct and to a spuriously low MCHC.
man, 1983; Planas, Van Voolen & Kelly, 1985; Van CHCM measured on some HA is also altered. CBC
Duijnhoven & Treskes, 1996). In the latter situation, demonstrating macrocytic hypochromia may be con-
the sample is diluted leading at times to a severe spur- sidered first as related to hyperglycemic sample.
ious macrocytic and hypochromic anaemia (Figure 7).
The calculated Hct is spuriously high and the MCH is
Considerations related to the anticoagulant
spuriously low: RBC count and Hb are unaffected in
diabetic patients and variably decreased if samples are The use of either K2- or K3-EDTA salt does not
diluted because of glucose infusion (Figure 7). This induce any difference in the CBC, in optimal condi-
phenomenon is related to plasma glucose concentra- tions of sampling. However, when concentration of
tion, which balances that of intraerythrocytic, either anticoagulant is increased, because of insufficient
in the body or in the sample if an excess of glucose is volume of blood drawn after venepuncture (or in
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 35

neonates), some changes may be observed. Decrease MCHC >36 g/dl is infrequent on most impedance-
of centrifuged Hct resulting from shrinking of RBC type HA, whereas it may be occasionally observed on
was reported in conditions related mainly to excess of laser-beam HA, in several constitutive disorders in
K3 salt (Goossens, Van Duppen & Verwilghen, 1991; which RBC are ‘dehydrated’, including hereditary
Hinchliffe, Bellamy & Lilleyman, 1992), in contrast spherocytosis, various haemoglobin disorders (CC, SC,
with data obtained on HA: no influence of K3-EDTA Cb thal), and some rare RBC disorders (xerocytosis).
concentration was observed on MCV, while K2-EDTA Some acquired conditions mimic constitutive ones,
at high concentrations resulted in a slight increase namely acquired immune haemolytic anaemias
in MCV, and such phenomenon could be observed caused by warm agglutinins, in which RBC coated
using several different HA (Goossens, Van Duppen & with warm antibodies may transform gradually into
Verwilghen, 1991). spheres in vivo or in vitro after venepuncture, leading
to more or less dehydrated and spherised cells. Many
situations corresponding to abnormally high Hb val-
Hyper- and hypo-natremia
ues, and/or abnormally low RBC, and/or spuriously
Macrocytic and hypochromic RBC changes were low MCV, also lead to increase MCHC (see the corres-
observed on blood samples in conditions related to ponding paragraphs and Table 1). Although the mech-
hypernatremia, whereas hyponatremia generated a anism is unknown, some immunosuppressive drugs
tendency towards microcytic and hyperchromic RBC may slightly increase MCHC, usually not above
(Cornbleet, 1983). Such situations were also reported 37.5 g/dl (Cornbleet 1983).
in animals (Boisvert, Tvedten & Scott, 1999). Hyper- MCHC <32 g/dl. As mentioned above, hyperglyce-
and hypo-natremia are both situations also reported mia leads to artefactual hypochromic macrocytosis.
as leading to spurious centrifuged Hct (Cornbleet, Storage of blood samples has been reported to cause a
1983). facticious increase in percentage of hypochromic cells,
and become conspicuous after 24 h of storage
(Murphy, Spaven & Casey 2002). Although changes
Storage of the sample and MCV
are not major, the number of hypochromic cells may
EDTA used as anticoagulant allows the accurate deter- be an important element to consider, as for example
mination of the CBC up to 24 h after the sample has in iron deficiency anaemias or during erythropoietin
been drawn. However, after that time, MCV may therapy (Richardson, Bartlett & Will, 2001).
increase, namely if the sample is stored at room tem-
perature. This coupled with low haemoglobin could
RETICULOCYTES
cause the operator to suspect that a slightly microcytic
anaemia is normocytic and that a normocytic could be Enumeration of peripheral blood reticulocytes is
mistaken for a macrocytic anaemia (Cohle, Saleem & essential in the diagnosis and management of anaemic
Makkaoui, 1981). patients and may be considered now as a part of the
CBC. If manual counting by light microscopy remains
the standard of reticulocyte enumeration, automated
Mean cell haemoglobin content (MCHC)
methods developed during the past two decades are
Measured parameters allow the calculation of mean now more accurate, precise, and cost-effective than
cell haemoglobin of individual RBC [MCH as manual counting and, in addition, provide a variety of
expressed in pg ¼ Hb (g/l)/RBC (1012/l)] and MCHC reticulocyte-related parameters, such as volume, hae-
[MCHC in g/dl ¼ Hb (g/l) · 100/MCV (fl) · RBC moglobin concentration, and maturity, which are
(1012/l)]. Some HA measure Hb concentration directly unavailable with light microscopy (Cavill, 1993; Van
within RBC, named cellular Hb concentration mean Petegem et al., 1993; Davis et al., 1994; Koepke, 1999;
(CHCM): discordant values (usually difference over Brugnara, 2000; Siekmeier, Bierlich & Jaross, 2000;
1.5 g/dl) between MCHC and CHCM allow in many Riley et al., 2001; Bain & Bates, 2001; Pierre, 2002;
instances the detection of anomalies related to one of Riley et al., 2002). Automated reticulocyte counts can
the measured RBC parameters. be performed using general purpose flow cytometers,
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
36 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

dedicated flow cytometers, or integrated with the observed in severe haemolytic anaemias, thalassaemia
other parameters from the CBC under a fully automa- major, congenital Heinz-body anaemia, or post-sple-
tized method. RNA from reticulocytes is stained using nectomy disorders, are also situations reported as dis-
the classical new methylene blue (part of Abbott and turbing automated reticulocyte counts (Hinchliffe,
Beckman Coulter instruments), oxazine (Bayer), or 1993; Espanol, Pedro & Remacha, 1999; listed in Riley
one among various fluorescent dyes, including thiaz- et al., 2001), as well as sickle cells (Pappas, Owens &
ole orange (Horiba ABX), auramine O (Sysmex), Flick, 1992; Ghevaert et al., 1997), spherocytes
CD4K530 (Abbott), or coriphosphine O (Beckman (Ghevaert et al., 1997), and haemoglobin H inclusion
Coulter). According to the HA and to the reagent bodies (Riley et al., 2001).
used, flow cytometric analysis is performed, using Although they are only faintly stained using fluor-
either the measurement of volume, conductivity, and escent dyes, intraerythrocytic parasites (malaria, babe-
light scatter (VCS) or the fluorescence detection (for sia) may interfere with automated reticulocyte count,
review see Riley et al., 2002). and value six times higher than the manual count
Careful gating of reticulocytes is crucial in deter- was reported in a patient demonstrating 70% of RBC
mining accurate counts, because dyes stain also RNA infected with plasmodium falciparum (Laurencet,
from PLT and WBC, and may also combine with the Martinez & Beris, 1997).
DNA of nucleated cells. Red blood cells and reticulo- Intensity of intracellular staining or fluorescence of
cytes being larger than PLTs and smaller than WBC, reticulocytes is assessed in most HA, the brightest reti-
both the former are discriminated from both the latter culocytes being the youngest, leading to peculiar
according to the size. Unsurprisingly, gating may be indices such as the immature reticulocyte fraction
difficult in the presence of blood components of (IRF; for review see Brugnara, 2000). White blood
abnormal size and containing stained RNA/DNA. cells displaying intense staining were reported as lead-
Giant PLT, PLT clumps, abnormal WBC, abnormal ing to an erroneous estimation of reticulocyte matur-
number of WBC, and WBC fragments, are situations ation index, the error being directly correlated with
reported as a potential source of interference with the WBC count (Villamor et al., 1996). Very mature
automated methods of reticulocyte analysis reticulocytes contain only a few coloured dots and
(Oyamatsu et al., 1989; Pappas, Owens & Flick, 1992; may be insufficiently detected in neonates using flow
Riley et al., 2001). NRBC have been reported as dis- cytometry, possibly due to the low concentration of
turbing reticulocyte counts (Pappas, Owens & Flick, the colouring matter used in some HA (Bayer; Wieg-
1992; Riley et al., 2001), but, as NRBC themselves are and et al., 2004). Although reticulocyte counts are s
not counted as reticulocytes in the HA, it has been after storing blood samples for 72 h at 4 C and 24–
hypothesized that abnormal reticulocyte count 48 h at room temperature (Brugnara et al., 1994;
observed in that instance could be due to increased Rudensky, 1997; Lacombe et al., 1999; Bain & Bates,
number of very young reticulocytes which frequently 2001), immature reticulocyte fraction was observed as
appeared with NRBC (Oyamatsu et al., 1989). stable for only 8 h at 4 C and 6 h at room tempera-
After gating of RBCs, reticulocytes are identified ture (Lacombe et al., 1999).
within the RBC population according to the coloured Various other situations are mentioned as gener-
particles or fluorescent material they contain. Cyto- ating occasionally spurious reticulocyte counts, inclu-
plasmic particles other than RNA that can be stained ding agglutinated cells in the presence of cold
by supravital dyes, including Howell–Jolly bodies, agglutinins (Oyamatsu et al., 1989; Riley et al., 2001),
Pappenheimer bodies, or basophilic stippling, may be autofluorescence of RBC (porphyria, drugs) and diag-
confused with reticulum granules using automated nostic intravenous fluorescent dyes (Riley et al.,
techniques in HA, as well as they interfere in manual 2001), high amount of paraproteins, and haemolysis
techniques (Pappas, Owens & Flick, 1992; Ghevaert (Riley et al., 2001).
et al., 1997; Riley et al., 2001), although such interfer- Although depending on the software gate correc-
ences have been found to be more or less conspicuous tions, in most of the above described situations the
according to the machines used (Oyamatsu et al., HA show abnormal flags, prompting the technologist
1989; Lofsness, Kohnke & Geier, 1994). Heinz bodies to perform a manual count. However, exceptions to
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 37

performed after RBC lysis, leaving intraglobular com-


Table 2. Situations leading to possible interference
with automated methods for reticulocyte analysis ponents free and enumerated together with PLT.
Improvements in analysis of PLT and RBC parameters
Inaccurate gating of RBCs: giant PLTs, PLT clumps, has led to an increase in the accuracy of results and
abnormal WBCs, abnormal number of WBCs, WBC has also led to the generation of several flags in
fragments, nucleated red blood cells
abnormal situations. The enumeration of PLT in
Intraerythrocytic particles: Howell-Jolly bodies,
Pappenheimer bodies, basophilic stippling, Heinz bodies, thrombocytopenias, the identification of NRBC leading
sickle cells, spherocytes, Haemoglobin H inclusions, to spurious PLT and WBC counts, and other situa-
plasmodium, babesia tions, have led manufacturers to optimise computer-
Others: cold agglutinin disease, autofluorescence of RBCs ized analysis for the detection and analysis of every
(drugs, porphyria), paraproteins, haemolysis, diagnostic blood cell, and moreover to develop further methods
intravenous fluorescent dyes
for their specific identification and their specific enu-
meration on the same sample and on the same HA.
The improvement of WBC analysis and the careful
study of WBC differential scattergrams have led to
flagging have been reported, at least in cold agglutinin another major improvement in blood cell analysis,
disease (Oyamatsu et al., 1989) and in beta-thalas- allowing WBC differential to be performed automatic-
saemia major (Ghevaert et al., 1997; see Table 2). ally but also to exhibit and to give explanation for
several anomalous CBC results. Several peculiar WBC
scattergrams generated were progressively identified,
CONCLUDING REMARKS
and related to spurious counts. As discussed all along
The widespread use of HA has led to major improve- both reports, inaccurate identification, analysis, or
ment of cellular haematology, because of quick and enumeration of one or several components from the
accurate results found in most instances, and now CBC leads in many instances to an abnormal WBC
preanalytical and analytical variables should be con- differential scattergram. By the rule the WBC differen-
sidered first within the laboratory when spurious tial is flagged or invalidated, so a blood smear and an
results from the HA are found (see also concluding optical count are often needed. Automated abnormal
remarks from part I). In most situations, HA generate WBC differentials related to spurious counts must be
flags or peculiar scattergrams in response to clinical or included together with the various but insufficiently
artifactual parameters, independent of the technology reported situations related to the inability of the HA
used. Inadequate blood sample may be responsible for to perform WBC differential or to identify specifically
various anomalies, including initiation of coagulation one or several cells from the WBC differential. All
within the tube because of difficult venepuncture or these situations should certainly need a specific
to low blood flow, excess of EDTA salt result of insuf- report.
ficient filling of the tube, which generate PLT and/or So, if some spurious counts were found to be
WBC aggregates, fibrin precipitates, or others. Over- numerous enough to generate technical improve-
filling of blood sample and time elapsed from samp- ments to identify them clearly and create other meth-
ling to analysis must also be considered at times. In ods for measurement, several other spurious counts
several instances the disease itself generates changes either do not generate specific flags to identify the
that disturb measurements performed by HA (RBC anomaly or even do not generate any flag at all.
agglutination, cryoprecipitates). Some anomalies, Moreover, if the most recent and powerful HA are
however, are intrinsic to the technology used for ana- able to generate several flags related to at least a part
lysis, and have led manufacturers to develop technical of spurious counts, it is stressed that simpler HA,
changes in analysing blood cells, one example corres- namely those without any WBC differential scatter-
ponding to the use of hydrodynamic focusing, which gram, will not. So, if acquiring a new HA is a personal
dramatically improved measurement of MCV in choice, each haematologist must know how his
impedance-type HA. Some protocols for measurement machine will react when (at least) the situations
of blood cells have been discarded, such as PLT counts reported here occur.
 2007 The Authors
Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
38 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

ACKNOWLEDGEMENTS CONFLICT OF INTEREST


We are indebted to Mathilde LINARD for kindly No conflict of interest.
reviewing English manuscript.

REFERENCES of Clinical and Laboratory Investigation interlaboratory correlation study. Amer-


51, 489–492. ican Journal of Clinical Pathology 102,
Antonsen S. & Beyer J.M. (1989) In vitro Brugnara C. (2000) Reticulocyte cellular 468–477.
granulocyte aggregation. Scandinavian indices: a new approach in the diagno- Deol I., Hernandez A. & Pierre R.V.
Journal of Clinical and Laboratory sis of anemias and monitoring of ery- (1995) Ethylenediamine tetraacetic
Investigation 49, 685–688. thropoietic function. Critical Reviews in acid-associated leukoagglutination.
Artiss J.D. & Zak B. (1987) Problems with Clinical Laboratory Sciences 37, 93–130. American Journal of Clinical Pathology
measurements caused by high con- Brugnara C., Hipp M.J., Irving P.J., 103, 338–340.
centrations of serum lipids. Critical Lathrop H., Lee P.A., Minchello E.M. & Deol I., Hernandez A. & Pierre R.V.
Reviews in Clinical Laboratory Sciences Winkelman J. (1994) Automated reticu- (1996) Ethylenediamine tetraacetic
25, 19–41. locyte counting and measurement of acid-associated leukoagglutination
Bain B.J. & Bates I. (2001) Basic haema- reticulocyte cellular indices. Evaluation (comments). American Journal of Clin-
tological techniques. In: Dacie and of the Miles H*3 blood analyzer. Amer- ical Pathology 105, 133–134.
Lewis Practical Haematology (eds S.M. ican Journal of Clinical Pathology 102, Elghetany M.T. & Hudnall S.D. (1996)
Lewis, B.J. Bain & I. Bates), pp. 19–46. 623–632. Spurious automated white cell count
Churchill, Livingston. Cantero M., Conejo J.R. & Jimenez A. with Coulter STKS in the myelodysplas-
Bessman J.D. & Banks D. (1980) Spurious (1996) Interference from lipemia in cell tic syndromes suggests the presence of a
macrocytosis, a common clue to ery- count by hematology analyzers. Clinical red cell membrane defect. American
throcyte cold agglutinins. American Chemistry 42, 987–988. Journal of Hematology 52, 69.
Journal of Clinical Pathology 74, 797– Carr M.E., Whitehead J., Carlson P., Todd Emori H.W., Bluestone R. & Goldsberg
800. W., Orcutt J. & Wallace H. (1996) Case L.S. (1973) Pseudo-leukocytosis asso-
Bizzaro N. (1993) Granulocyte aggrega- report: immunoglobulin M-mediated, ciated with cryoglobulinemia. American
tion is edetic acid and temperature temperature-dependent neutrophil Journal of Clinical Pathology 59, 202–
dependent. Archives of Pathology and agglutination as a cause of pseudoneu- 204.
Laboratory Medicine 117, 528–530. tropenia. American Journal of the Med- Epstein H.D. & Kruskall M.S. (1988)
Bizzaro N. & Fiorin F. (1999) Coexistence ical Sciences 311, 92–95. Spurious leukopenia due to in vitro
of erythrocyte agglutination and EDTA- Cavill I. (1993) The rejected reticulocyte. granulocyte aggregation. American
dependent platelet clumping in a British Journal of Haematology 84, Journal of Clinical Pathology 89, 652–
patient with thymoma and plasmocy- 563–565. 655.
toma. Archives of Pathology and Cohle S.D., Saleem A. & Makkaoui D.G. Espanol I., Pedro C. & Remacha A.F.
Laboratory Medicine 123, 159–162. (1981) Effect of storage of blood cells (1999) Heinz bodies interfere with
Bizzaro N. & Piazza I. (1991) Lymphocytic on stability of hematologic parameters. automated reticulocyte counts. Haema-
clusters in peripheral blood: an atypical American Journal of Clinical Pathology tologica 84, 373–374.
morphologic pattern of chronic lympho- 76, 67–69. Evan-Wong L.A. & Davidson R.J. (1983)
cytic leukemia. Acta Haematologica 86, Corberand J.X., Laharrague P.F., Fillola Raised Coulter mean corpuscular
209–211. G., Charlet J.P. & Cambus J.P. (1991) volume in diabetic ketoacetosis, and its
Boisvert A.M., Tvedten H.W. & Scott Discovery of unsuspected pathological underlying association with marked
M.A. (1999) Artifactual effects of hyper- states using a new haematology analy- plasma hyperosmolarity. Journal of
natremia and hyponatremia on red cell ser. Medical Laboratory Sciences 48, Clinical Pathology 36, 334–336.
analytes measured by the Bayer H*1 80–83. Fohlen-Walter A., Jacob C., Lecompte T.
analyser. Veterinary Clinical Pathology Cornbleet J. (1983) Spurious results from & Lesesve J.F. (2002) Laboratory identi-
28, 91–96. automated hematology cell counters. fication of cryoglobulinemia from auto-
Booth F. & Mead S.V. (1983) Resistance Laboratory Medicine 14, 501–514. mated blood cell counts, fresh blood
to lysis of erythrocytes containing hae- Creer M.H. & Ladenson J. (1983) Analy- samples, and blood films. American
moglobin C detected in a differential tical error due to lipemia. Laboratory Journal of Clinical Pathology 117, 606–
white cell counting system. Journal of Medicine 14, 351–355. 614.
Clinical Pathology 36, 816–818. Davis B.H., Bigelow N.C., Koepke J.A., Gagne C., Auger P.L., Moorjani S., Brun
Bremmelgaard A. & Nygard J. (1991) Borowitz M.J., Houwen B., Jacobberger D. & Lupien P.J. (1977) Effect of
Interference by cryoglobulins with J.W., Pierre R.V., Corash L., Ault K.A. hyperchylomicronemia on the measure-
white blood cell measurements on & Batjer J.D. (1994) Flow cytometric ment of hemoglobin. American Journal
Coulter Counter. Scandinavian Journal reticulocyte analysis. Multiinstitutional of Clinical Pathology 68, 584–586.

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 39

Galifi M., Schinella M., Nicoli M. & Lippi macrocytosis induced by excessive K2- Journal of Clinical Hematology 32, 205–
G. (1993) Instrumental reports and EDTA concentration. Clinical and 211.
effect of anticoagulants in a case of Laboratory Haematology 14, 268–269. Koepke J.A. (1999) Update on reticulo-
neutrophil agglutination in vitro. Hae- Holt J.T., DeWandler M.J. & Arvan D.A. cyte counting. Laboratory Medicine 30,
matologica 78, 364–370. (1982) Spurious elevation of the elec- 339–343.
Ghevaert C., Fournier M., Reade V., tronically determined mean corpuscular Lacombe F., Lacoste L., Vial J.P., Briais A.,
Mehiaoui L. & Goudemand J. (1997) volume and hematocrit caused by Reiffers J., Boisseau M.R. & Bernard P.
Laboratory evaluation of the Coulter hyperglycemia. American Journal of (1999) Automated reticulocyte counting
STKS reticulocyte method in a chil- Clinical Pathology 77, 561–567. and immature reticulocyte fraction
dren’s hospital. Laboratory Hematology Huh J., Jung J., Yoon H. & Chung W. measurement. Comparison of ABX
3, 92–97. (2005) Pseudoeosinophilia associated PENTRA 120 retic, Sysmex R-2000,
Goodrick M.J., Boon R.J., Bishop R.J.D., with malaria infection determined in flow cytometry, and manual counts.
Copplestone J.A. & Prentice A.G. (1993) the Sysmex XE-2100 hematology ana- American Journal of Clinical Pathology
Inaccurate haemoglobin estimation in lyser. Annals of Hematology 84, 400– 112, 677–686.
Waldenström’s macroglobulinaemia: 402. Laurencet F.M., Martinez T. & Beris P.
unusual reaction with monomeric IgM Imbing F. Jr, Kumar D., Kumar S., Yuoh (1997) Spurious extreme reticulocytosis
paraprotein. Journal of Clinical Pathol- G. & Gardner F. (1995) Splenic lym- with an antomated reticulocyte analy-
ogy 46, 1138–1139. phoma with circulating villous lympho- zer. The New England Journal of Medi-
Goossens W., Van Duppen V. & Verwil- cytes. Journal of Clinical Pathology 48, cine 337, 1922–1923.
ghen R.L. (1991) K2- or K3- EDTA: the 584–587. Lesesve J.F., Haristoy Y., Thouvenin M.,
anticoagulant of choice in routine hae- Imbing F.D. Jr, Adegboyega P.A., McLu- Latger-Cannard V., Buisine J. &
matology? Clinical and Laboratory Hae- cas E. & Elghetany T. (1996) EDTA- Lecompte T. (2000) Pseudoleucopénie
matology 13, 291–295. associated leukoagglutination. American par leuco-agglutination in vitro des
Griswold D.J. & Champagne V.D. (1985) Journal of Clinical Pathology 105, 133– polynucléaires neutrophiles: expérience
Evaluation of the Coulter S plus IV 134. d’un laboratoire, revue de la littérature
three part differential in an acute care Infanti L., Spengeler V., Hoffmann T., et conduite proposée. Annales de Biolo-
hospital. American Journal of Clinical Favre G., Passweg J., Trendelenburg N. gie Clinique 58, 417–424.
Pathology 84, 49–57. & Tichelli A. (1998) Correction of spur- Lesesve J.F., Haristoy X., Fisher B., Goupil
Guibaud S., Plumet-Leger A. & Frobert Y. ious blood cell counts in a patient with J.J. & Lecompte T. (2001) Agglutination
(1983) Transient neutrophil aggregation cryoglobulinemia reversed by the use of in vitro EDTA-dépendante des lympho-
in a patient with infectious mono- citrate-anticoagulated blood. Schweizer- cytes. Annales de Biologie Clinique 59,
nucleosis. American Journal of Clinical ische Medizinische Wochenschrift. Jour- 497–501.
Pathology 80, 883–884. nal Suisse de Medecine 128, Lippi U., Bellavite P., Schinella M., Nicoli
Gulliani G.L., Hyun B.H. & Gabaldon H. 1608–1610. M. & Lippi G. (1994) Assessment of
(1977) Falsely elevated automated leu- Jacob H.S., Craddock P.R., Hammersch- neutrophil aggregation by Coulter STKR
kocyte counts on cryoglobulinemic and/ midt D.E. & Moldow C.F. (1980) Com- and STKS haematological analysers.
or cryofibrinogenemic blood samples. plement-induced granulocyte Clinical and Laboratory Haematology
Laboratory Medicine 8, 14–16. aggregation. An unsuspected mechan- 16, 43–55.
Haeney M.R. (1976) Erroneous values for ism of disease. New England Journal of Lofsness K.G., Kohnke M.L. & Geier N.A.
the total white cell count and ESR in Medicine 302, 789–794. (1994) Evaluation of automated reticu-
patients with cryoglobulinaemia. Jour- Juneja S., Wolf M. & McLennan R. locyte counts and their reliability in the
nal of Clinical Pathology 29, 894–897. (1992) Clumping of lymphoma cells in presence of Howell–Jolly bodies. Ameri-
Hattersley P.G., Gerard P.W., Caggiano V. peripheral blood induced by EDTA. can Journal of Clinical Pathology 101,
& Nash D.R. (1971) Erroneous values Journal of Clinical Pathology 45, 538– 85–90.
on the model S Coulter counter due to 540. Lombarts A.J.P.F. & de Kieviet W. (1988)
high titer cold agglutinins. American Kahlil S.H. (1991) EDTA-dependent leu- Recognition and prevention of pseudo-
Journal of Clinical Pathology 55, 442– koagglutination. American Journal of thrombocytopenia and concomitant
446. Clinical Pathology 96, 147–148. pseudoleukocytosis. American Journal
Hillyer C.D., Knopf A.N. & Berkman E.F. Kim Y.R., Yee M., Metha S., Chupp V., of Clinical Pathology 89, 634–639.
(1990) EDTA-dependent leukoaggluti- Kendall R. & Scott CS. (1998) Simulta- Luke R.G., Koepke J.A. & Siegel R.R.
nation. American Journal of Clinical neous differentiation and quantification (1971) The effects of immunosuppres-
Pathology 94, 458–461. of erythroblasts and white blood cells sive drugs and uremia on automated
Hinchliffe R.F. (1993) Errors in automated on a high throughput clinical haematol- leukocyte count. American Journal of
reticulocyte counts due to Heinz bodies. ogy analyser. Clinical and Laboratory Clinical Pathology 56, 503–507.
Journal of Clinical Pathology 46, 878– Haematology 20, 21–29. Marshall B.A., Theil K.S. & Brandt J.T.
879. Kobayashi S., Seki K., Yamaguchi M., (1990) Abnormalities of leukocyte his-
Hinchliffe R.F., Bellamy G.J. & Lilleyman Maruta A. & Kodama F. (1991) Studies tograms resulting from microorganisms.
J.S. (1992) Use of the technicon H1 on EDTA-dependent pseudoneu- American Journal of Clinical Pathology
hypochromia flag in detecting spurious tropenia. Rinsho Ketsueki. Japanese 93, 526–532.

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
40 M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS

Mayan H., Gurevitz O., Mouallem M. & foreign surface adhesiveness on human Riley R.S., Ben-Ezra J.M., Goel R. & Tid-
Farfel Z. (1996) Multiple spurious leukocytes. American Journal of Pathol- well A. (2001) Reticulocytes and reticu-
laboratory results in a patient with ogy 90, 537–550. locyte enumeration. Journal of Clinical
hyperlipemic pancreatitis treated by Oyamatsu T., Shimizu N., Takeuchi K., Laboratory Analysis 15, 267–294.
plasmapheresis. Israel Journal of Medi- Yamamoto M., Kawai Y., Watanabe K. Riley R.S., Ben-Ezra J.M., Tidwell A. &
cal Sciences 32, 762–766. & Iri H. (1989) Automated measure- Romagnoli G. (2002) Reticuocyte analy-
McMullin M.F., Wilkin H.J. & Elder E. ment of reticulocyte count by flow sis by flow cytometry and other techni-
(1995) Inaccurate haemoglobin estima- cytometry. II: analysis of the blood con- ques. Hematology/Oncology Clinics of
tion in Waldenstrom’s macroglobuline- taining abnormal erythrocytes or giant North America 16, 373–420.
mia. Journal of Clinical Pathology 48, platelets. Rinsho Byori. Japanese Jour- Robbins S.H., Conly M.A. & Oettinger J.
787. nal of Clinical Pathology 37, 807–812. (1991) Cold-induced granulocyte agglu-
McVeigh D.J., Faim L.S. & van der Wey- Pappas A.A., Owens R.B. & Flick J.T. tination. A cause of pseudoleukopenia.
den M.B. (1989) Correction of spurious (1992) Reticulocyte counting by flow Archives of Pathology and Laboratory
haematological results using Technicon cytometry. A comparison with manual Medicine 115, 155–157.
H 1-derived data. Clinical and Labora- methods. Annals of Clinical and Labora- Roberts W.L., Fontenot J.D. & Lehman
tory Haematology 11, 369–373. tory Science 22, 125–132. C.M. (2000) Overestimation of hemo-
Mellors I. & McArdle B. (1995) Improved Paterakis G.S., Laoutaris N.P., Alexia S.V., globin in a patient with an IgA-kappa
cell counting in osmotically resistant Siourounis P.V., Stamoulakatou A.K., monoclonal gammopathy. Archives of
erythrocytes. Clinical and Laboratory Terzoglou G.N., Papassotiriou I.G. & Pathology and Laboratory Medicine
Haematology 17, 23–30. Loukopoulos D. (1994) The effect of red 124, 616–618.
Mohandas N., Clark M.R., Kissinger S., cell shape on the measurement of red Rohr L.R. & Rivers F.M. (1990) Spurious
Bayer C. & Shohet S.B. (1980) Inac- cell volume. A proposed method for the automated leukopenia due to in vitro
curacies associated with the automated comparative assessment of this effect granulocyte aggregation. American
measurement of mean cell hemoglobin among various haematology analysers. Journal of Clinical Pathology 93, 572–
concentration in dehydrated cells. Blood Clinical and Laboratory Haematology 574.
56, 125–128. 16, 235–245. Rudensky B. (1997) Comparison of a
Mohandas N., Kim Y.R., Tycko D.H., Orlik Payne B.A. & Pierre R.V. (1984) Pseudo- semi-automated new Coulter methylene
J., Wyatt J. & Groner W. (1986) Accu- thrombocytopenia: a laboratory artifact blue method with fluorescence flow
rate and independent measurement of with potentially serious consequences. cytometry in reticulocyte counting.
volume and hemoglobin concentration Mayo Clinic Proceedings 59, 123–125. Scandinavian Journal of Clinical and
in individual red cells by laser light Petrucci J.V., Duanne P.A. & Chapman Laboratory Investigation 57, 291–296.
scattering. Blood 68, 506–513. C.C. (1971) Spurious erythrocyte indi- Sandberg S., Sonstabo K. & Christensen
Morse E.E., Kalache G., Germino W. & ces as measured by the model S Coulter N.G. (1989) Influence of lipid and leu-
Stockwell R. (1981) Increased electronic counter due to cold agglutinins. Amer- kocytes on the haemoglobin determina-
mean corpuscular volume induced by ican Journal of Clinical Pathology 56, tion by Coulter Counter S Plus III,
marked hyperglycemia. Annals of Clin- 500–502. Technicon H6000, Technicon H1,
ical and Laboratory Science 11, 184– Pewarchuk W., VanderBoom J. & Blajch- LK540, Reflotron, and Hemocap. Scan-
187. man M.A. (1992) Pseudopolycythemia, dinavian Journal of Clinical and
Murphy P.T., Spaven M. & Casey M. pseudothrombo-cytopenia, and pseudo- Laboratory Investigation 49, 145–148.
(2002) Storage of full blood count sam- leukopenia due to overfilling of blood Savage R.A. (1984) Pseudoleukocytosis
ples causes a factitious increase in per- collection vacuum tubes. Archives of due to EDTA-induced platelet clumping.
centage of hypochromic cells. American Pathology and Laboratory Medicine American Journal of Clinical Pathology
Journal of Kidney Diseases 39, 1124. 116, 90–92. 81, 317–322.
Nicholls P.D. (1975) Erroneous model S Pierre R.V. (2002) Reticulocytes. Their Savage R.A. (1989) Analytic inaccuracy
Counter values on patients undergoing usefulness and measurement in periph- resulting from hematology specimen
parenteral nutrition with intravenous eral blood. Clinics in Laboratory Medi- characteristics. Three cases of clinically
lipid emulsions. Medical Laboratory cine 22, 63–79. misleading artifacts affecting white
Technology 30, 293–295. Planas A.T., Van Voolen G.A. & Kelly L.A. blood cell and platelet counts. American
Nicholls P.D. (1977) The erroneous hae- (1985) Hyperglycemic macrocytosis in Journal of Clinical Pathology 92, 295–
moglobin–hyperlipidaemia relationship. electronically determined mean corpus- 299.
Journal of Clinical Pathology 30, 638– cular volume. Use of three different Savage R.A. & Hoffman G.C. (1983) Clin-
640. automatic cell counters. Annals of Clin- ical significance of osmotic matrix errors
Nosanchuck J.S., Roark M.F. & Wanser C. ical and Laboratory Science 15, 286– in automated hematology. The fre-
(1974) Anemia masked by triglycer- 291. quency of hyperglycemic matrix errors
idemia. American Journal of Clinical Richardson D., Bartlett C. & Will E.J. producing spurious macrocytosis. Amer-
Pathology 62, 838–839. (2001) Optimizing erythropoietin ther- ican Journal of Clinical Pathology 80,
O’Flaherty J.T., Kreutzer D.L. & Ward apy in hemodialysis patients. American 861–865.
P.A. (1978) Chemotactic factor influ- Journal of Kidney Diseases 38, 109– Schinella M., Kojikara P. & Curci V.
ences on the aggregation, swelling and 117. (1995) Prevention of polymorpho-

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41
M. ZANDECKI ET AL. SPURIOUS COUNTS ON HAEMATOLOGY ANALYSERS 41

nuclear leukocyte agglutination in vitro. Pseudoleukocytosis due to cryoprotein Wallis J.P. & Ford J.M. (1987) Incorrect
Haematologica 80, 196–197. crystals. American Journal of Clinical haemoglobin estimation on the Coulter
Schrezenmeier H., Muller H., Gunsilius Pathology 60, 669–671. S + IV in some patients with IgM para-
E., Heimpel H. & Seifried E. (1995) Van Duijnhoven H.L.P. & Treskes M. proteins. Clinical and Laboratory Hae-
Anticoagulant-induced pseudothrombo- (1996) marked interference of hypergly- matology 9, 95–96.
cytopenia and pseudoleucocytosis. cemia in measurements of mean (red) Wang F.S., Itose Y., Tsuji T., Hamagushi
Thrombosis Haemostasis 73, 506–513. cell volume by Technicon H analyzers. Y., Hirai K. & Sakata T. (2003) Develop-
Shah P.C., Patel A.R. & Rao K.R. (1975) Clinical Chemistry 42, 76–80. ment and clinical application of nucle-
Hyperlipemia and spuriously elevated Van Petegem M., Cartuyvels R., de Scho- ated red blood cell counting and staging
hemoglobin values. Annals of Internal uwer P., van Duppen V., Goossens W. on the automated haematology analyser
Medicine 82, 383. & van Hove L. (1993) Comparative eva- XE-2100. Clinical and Laboratory Hae-
Shelton J.B. Jr & Frank I.N. (2000) Sple- luation of three flow cytometers for matology 25, 17–23.
nic B cell lymphoma with lymphocyte reticulocyte enumeration. Clinical and Weiss G.B. & Bessman J.D. (1984) Spur-
clusters in peripheral blood smears. Laboratory Haematology 15, 103–111. ious automated red cell values in warm
Journal of Clinical Pathology 53, 228– Villa S., Colotta F., de Gaetano G. & autoimmune hemolytic anemia. Amer-
230. Semerano N. (1984) Arachidonic acid ican Journal of Hematology 17, 433–
Siekmeier R., Bierlich A. & Jaross W. and leukotriene B4 induce aggregation 435.
(2000) Determination of reticulocytes: of human peripheral blood mono- Whiteway A.J. & Bain B.J. (1999) Arte-
three methods compared. Clinical nuclear leukocytes in vitro. British Jour- factual elevation of an automated white
Chemistry and Laboratory Medicine 38, nal of Haematology 58, 137–146. cell count following femoral vein punc-
245–249. Villamor N., Kirsch A., Huhn D., Vives- ture. Clinical and Laboratory Haematol-
Solanki D.L. & Blackburn B.C. (1983) Corrons J.L. & Serke S. (1996) Inter- ogy 21, 65–68.
Spurious leukocytosis and thrombocyto- ference of blood leukocytes in the mea- Wiegand G., Effenberger-Klein A., Weber
penia. A dual phenomenon caused by surements of immature red cells R., Kosanke W., Niethammer D. & Bru-
clumping of platelets in vitro. Journal of (reticulocytes) by two different (semi-) chelt G. (2004) Potential pitfalls of com-
the American Medical Association 250, automated flow-cytometry technologies. parative measurements of reticulocytes
2514–2515. Clinical and Laboratory Haematology using flow cytometry and microscopy in
Solanki D.L. & Blackburn B.C. (1985) 18, 89–94. premature and infants. Clinical Chem-
Spurious red cell parameters due to Vinatier I. & Flandrin G. (1993) Avan- istry and Laboratory Medicine 42,
serum cold agglutinins: observations on tages et limites de l’hémogramme au- 1150–1154.
Ortho ELT-8 cell counter. American tomatisé. La Revue du Praticien 7, 69– Zandecki M., Dupriez B., Fenaux P., Jude
Journal of Clinical Pathology 83, 218– 73. B., Watel A., Dracon M., Demory J.L. &
222. Vinatier I., Capiod J.C., Sassier P., Roussel Cosson A. (1989) Cytological and ultra-
Strauchen J.A., Alston W., Anderson J., B. & Delobel J. (1994) Four cases of structural assessment of free crystals or
Gustafson Z. & Fajardo L.F. (1981) Inac- spuriously low WBC count due to in precipitates associated with pseudoleu-
curacy in automated measurement of vitro leukocyte agglutination: contribu- kocytosis and pseudothrombocytosis in
hematocrit and corpuscular indices in tion of the hematology analyzer Coulter cryoglobulinemia. Nouvelle Revue
the presence of severe hyperglycemia. STKS in detecting this clinically mis- Francaise D Hematologie 31, 397–402.
Blood 57, 1065–1067. leading artefact. Pathologie–Biologie 42,
Taft E.G., Grossman J., Abraham G.N., 775–780.
Leddy J.P. & Lichtman M.A. (1973)

 2007 The Authors


Journal compilation  2007 Blackwell Publishing Ltd, Int. Jnl. Lab. Hem. 2007, 29, 21–41

You might also like