CALIBRATION PROCEDURE
for the
Kodak DirectView CR 800/900/825/850/950/500/975
SYSTEMS
and the
CLASSIC/ELITE CR SYSTEMS
Service Codes: 3519, 3520, 5634, 4825, 4826, 4366,
1883, 5258, 5259
Table of Contents
Description Page
2 03NOV08 – XP3519-1
Calibration for the CR SYSTEMS
PHOSPHOR CASSETTE
SCREEN Important
You must start with the largest size CASSETTE.
[1] Move the LATCH to the right to release the
PHOSPHOR SCREEN.
T
ub
e
S
id
e
[2] Remove the PHOSPHOR SCREEN from the
CASSETTE.
[3] Check that the PHOSPHOR SCREEN:
• is correctly fastened to the CASSETTE.
• edges are not lifted.
• does not have changes in color.
[4] Clean the PHOSPHOR SCREENS.
hole
LATCH [5] Install the PHOSPHOR SCREEN into the
CASSETTE.
[6] Insert a tool into the hole to snap and close the
CASSETTE.
[7] Check that the LATCH on the CASSETTE is
Snap
Close operating correctly.
[8] From the Main Menu, touch [Erase Cassette].
To
Open [9] Insert the CASSETTE into the CR SYSTEM.
[10] Do Step 1 - Step 9 again for all size CASSETTES.
H177_2443GCA
H177_2443GC
XP3519-1 – 03NOV08 3
MISCELLANEOUS PUBLICATION
X-RAY Caution
TUBE • Dangerous Voltage
• Do not use a BUCKY or a GRID during this
COLLIMATOR procedure.
LIGHT
• The exposure to air kerma conversion factor is 8.76
microGy per mR.
• 100 mR corresponds to 100 x 8.76 = 876 microGy
or 0.876 mGy.
[1] Move the X-RAY TUBE to 180 cm (72.0 in.) above
the floor.
180 cm
(72.0 in.)
Important
CASSETTE A LEAD SHEET and the help of a licensed Radiological
Technologist (RT) is necessary for this procedure.
tape
[2] Ask the RT to energize the X-ray equipment and
the COLLIMATOR LIGHT on the X-RAY TUBE.
[3] Place the CASSETTE on the floor under the beam
of light.
[4] Align the beam of light with a 5.0 cm (2.0 in.) area
of light on the floor around the CASSETTE.
H177_0004GCC 5.0 cm
H177_0004GC (2.0 in.)
Important
You must place the tape correctly.
[5] Place the tape on the floor around the edges of
the beam of light.
[6] Use tape to fasten the COPPER and ALUMINUM
FILTERS TL-5259 to each other.
X-RAY TUBE
Important
The COPPER FILTER must be directly next to the X-
RAY TUBE.
[7] Insert both of the FILTERS into the bottom of the
X-RAY TUBE. If necessary, use tape to adhere the
COPPER FILTER FILTERS to the bottom of the X-RAY TUBE.
tape
H149_0013ACC
ALUMINUM FILTER H149_0013AC
4 03NOV08 – XP3519-1
Calibration for the CR SYSTEMS
Important
A LEAD SHEET and the help of a RT is necessary for
DOSIMETER
this procedure.
[8] Use the following values for the first set up of the
X-ray equipment:
• 80 kVp - ± 1
• 80 mAs
[9] Remove the CASSETTE from the floor.
Important
You must use a LEAD SHEET or a LEAD APRON under the DOSIMETER. If a LEAD SHEET or a LEAD APRON is
not used, the values on the DOSIMETER will be false by a maximum of 20%.
[10] Place a LEAD SHEET or a LEAD APRON and the DOSIMETER TL-5930 in the center of the area.
LCD DISPLAY
READ/RESET
SWITCH
F
OF
ON
POWER
SWITCH
H187_2596BCA
H187_2596BC
XP3519-1 – 03NOV08 5
MISCELLANEOUS PUBLICATION
Note
You will use the recorded number when you do the calibration procedure.
[20] Ask the RT to expose the DOSIMETER 3 times to check that the values remain within
± 0.2 mR of the value that you recorded in Step 19.
Note
If the values are not within the specification, use another X-ray source.
X-RAY
TUBE Important
• The ARROW on the CASSETTE must be parallel to
the X-RAY TUBE.
• Do not use a LEAD SHEET during this procedure.
6 03NOV08 – XP3519-1
Calibration for the CR SYSTEMS
[2] Enter the number from the LCD DISPLAY from Step 19 of Preparing the Equipment for GEN RAD CASSETTES
in the “MR VALUE” field.
[3] Insert the first CASSETTE into the system.
[4] Touch [Start Calibration Scan].
[5] Insert the exposed CASSETTES every 2 minutes.
Important
If the calibration is not successful, an error code occurs.
[6] Was the calibration successful?
Yes No
The procedure is complete. Continue with Step 7.
[7] At the “Last Calibrated Date” screen, locate the existing date and time for the corresponding CASSETTE.
[8] Locate and correct the corresponding error code in the error log.
Note
For error codes 16019 and 16020, you must see the adjustment procedure for the GALVO ROTATION for your CR
SYSTEM.
XP3519-1 – 03NOV08 7
MISCELLANEOUS PUBLICATION
“View” and “Plot COLLECTOR PROFILES” for the GEN RAD CASSETTES
[1] From the Main Menu, touch:
• [Service Functions]
• [Diagnostics]
• [View Collector Profiles]
[2] Select the 35 x 43 cm CASSETTE.
Important
“SPIKES” might be the result of dust in the COLLECTOR or the illumination of the laser beam on the COLLECTOR.
Pixel Value
No Spikes
Note
Correct values displays:
• no large differences in the pixel values - 10 or more or “Spikes”
• “Offset” is between 500 and 2000
• “Amplitude” is less than 4000
[4] Is the COLLECTOR PROFILE from Step 3 correct?
Yes No
Advance to Step 7. Continue with Step 5.
Caution
Use caution when you clean the COLLECTOR.
[6] If large differences appear in the pixel values “Spikes”, use a LINT-FREE CLOTH to clean dust that might be
blocking the beam of light to the COLLECTOR.
[7] Do:
• Exposing the GEN RAD CASSETTE
• Calibration for the Scan using GEN RAD CASSETTES
• Kodak DirectView TOTAL QUALITY TOOL on all size CASSETTE
• view “Logs” and check that the “Linearity” is less than 0.5 on the 18 x 24 cm CASSETTE
XP3519-1 – 03NOV08 9
MISCELLANEOUS PUBLICATION
Caution
• Dangerous Voltage
• The help of a licensed Radiological Technologist
(RT) is necessary for this procedure.
COLLIMATOR
LIGHT
H187_2599GCA
H187_2599GC
X-RAY Important
TUBE
Do not place the tape in the image area.
[3] Use TAPE to fasten the 2 ALUMINUM FILTERS
6C3015 onto the bottom of the X-RAY TUBE.
2 ALUMINUM
FILTERS
TAPE
H187_2600ACA
H187_2594AC
10 03NOV08 – XP3519-1
Calibration for the CR SYSTEMS
Important
The help of a RT is necessary for this procedure.
[4] Use the following values for the set up of the X-ray
equipment:
• 28 kVp
• 160 mAs
Note
The mAs value above is a recommended starting point
for doing calibration.
PATIENT
SUPPORT [5] Set the “Target/Filter” to [Mo/Mo].
[6] Place the DOSIMETER:
• on the PATIENT SUPPORT
• 6.0cm (2.4 in.) from the edge
6.0 cm
(2.4 in.)
edge
H187_2601GCA
H187_2601GC DOSIMETER
LABEL
20.0m = 3300 CV GenRaD Calibration
10.0m = 3000 CV GenRaD Flat/Phantom
LCD DISPLAY
F m = 3000 CV Mammo Calibration
OF m = 2700 CV Mammo Flat/Phantom
ON
POWER
SWITCH
READ/RESET H187_2598BCA
SWITCH H187_2598BC
[7] Move the POWER SWITCH on the DOSIMETER to the “ON” position.
[8] Press the “READ/RESET” SWITCH.
[9] Check that “0.00” displays on the LCD DISPLAY.
[10] Ask the RT to expose the DOSIMETER.
[11] Record:
• value displayed in the LCD DISPLAY of the DOSIMETER
• value recorded next to “Mammo Calibration” on the LABEL
[12] Is the value displayed +/- 10% of the “Mammo Calibration Value” recorded on the LABEL?
Yes No
Advance to Step 17. Do Step 13 - Step 16.
XP3519-1 – 03NOV08 11
MISCELLANEOUS PUBLICATION
Note
You will use the recorded number when you do the calibration procedure.
[18] Ask the RT to expose the DOSIMETER 3 times to check that the values remain within
± 0.5 of the value that you recorded in Step 17.
12 03NOV08 – XP3519-1
Calibration for the CR SYSTEMS
Important
• The CHEST WALL edge of the CASSETTE must
be 5.1 cm (2.0 in.) off the edge of the PATIENT
SUPPORT.
TUBE • Some systems have a RAISED EDGE at the TUBE
SIDE side.
• If your system has a RAISED EDGE, the
2 L-BRACKETS CASSETTE and the 2 L-BRACKETS cannot be
placed more than 5.1 cm (2.0 in.) off the edge
PATIENT
SUPPORT • Do not place the CASSETTE or the
2 L-BRACKETS on the raised edge.
Important
• After exposing the CASSETTE, you must wait 5
minutes before starting this procedure.
• Every minute that you wait beyond 5 causes the
exposure index value to decrease by 5.
[5] Wait 5 minutes from the time of exposure of the
first CASSETTE.
XP3519-1 – 03NOV08 13
MISCELLANEOUS PUBLICATION
[3] Enter the “Scan Calibration Value” from Step 1 in the “MR VALUE” field.
[4] Insert the first CASSETTE into the system.
[5] Touch [Start Calibration Scan].
[6] Insert the next exposed CASSETTE.
Important
If the calibration is not successful, an error code occurs.
[7] Was the calibration successful?
Yes No
The procedure is complete. Continue with Step 8.
Important
Under “Last Calibrated Date”, the date and time displays only if the test passes.
[8] At the “Last Calibrated Date” screen, locate the existing date and time for the corresponding CASSETTE.
[9] locate and correct the corresponding error code in the error log.
Note
For error codes 16019 and 16020, you must see the adjustment procedure for the GALVO ROTATION for your CR
SYSTEM.
14 03NOV08 – XP3519-1
Calibration for the CR SYSTEMS
Important
“Spikes” might be the result of dust in the COLLECTOR or the illumination of the laser beam on the COLLECTOR.
[3] Observe the “Collector Profile” on the graphic. Correct values:
• have a pixel value of less than 150
• will not have large differences in the pixel values - 10 or more or “spikes”
[4] Is the “Collector Profile” correct for the newly calibrated CASSETTES?
Yes No
The calibration procedure is complete. a. Check:
• Flat Field Test
• Phantom Test
b. Continue with Step 5.
[5] Check the error log for the error code failure.
[6] locate and correct the corresponding error code in the error log.
XP3519-1 – 03NOV08 15
Calibration for the CR SYSTEMS
Important
For error codes 16019 and 16020, see the adjustment procedure for the GALVO ROTATION for your CR SYSTEM.
[7] Is the pixel value above 150?
Yes No
• Check: Continue with Step 8.
– Servo Gain is set correctly on the GALVO
– COLLECTOR spacing to the plate is correct
– beam alignment through the COLLECTOR
• Do:
– Adjustment procedure for the FOLD MIRROR for your CR SYSTEM.
– “Exposing the EHR CASSETTE” on Page 13
– “Calibration for the Scan using EHR CASSETTES” on Page 14
– ““View” and “Plot the COLLECTOR PROFILES” of each EHR
CASSETTE” on Page 15
• Kodak DirectView TOTAL QUALITY TOOL on all EHR CASSETTE sizes
Caution
Use caution when you clean the COLLECTOR.
[8] Do the following if large differences appear in the pixel values of “spikes”.
(a) Clean the COLLECTOR using a LINT-FREE CLOTH.
(b) Do again:
• “Exposing the EHR CASSETTE” on Page 13
• “Calibration for the Scan using EHR CASSETTES” on Page 14
• “View” and “Plot the COLLECTOR PROFILES” of each EHR CASSETTE on Page 15
Publication History