Please refer to the official BTL CardioPoint Trolley User manual, chapters 4.2 – 5.11.
Safety precautions!
Never open the gas cylinder with its valve without a pressure regulator connected to
it.
Close the gas cylinder with its valve after every calibration.
4.3 Devices
Ergometer or Treadmill
ECG
Gas analyzer
Ambient sensor
Blood pressure monitor
Note: Connect the devices to the USB SS ports (blue ones) which can transfer data ten
times faster than the normal ones. These are found on the back side of the computer (HP
Prodesk).
The CPET set uses two computer softwares – BTL CardioPoint and PistonXP. The
installation of the BTL CardioPoint software is described in its official manual. Installation of
the PistonXP is described below. It doesn’t matter in which order the softwares are installed.
11
11
1
1
11
22
11
11
11
Before starting to do the tests, certain settings need to be applied to the PistonXP software.
Follow the instructions below to set your software properly and prepare it for the tests.
1) Make sure that the PRE-201 analyzer is connected directly to the computer by means of
USB cable (not via USB HUB).
13) The software will now ask you to calibrate the gas analyzer. You may proceed
according to the CPET quick guide, chapter 1 COMPLETE CALIBRATION.
14) Please note that it is necessary to run the O2 sensor characteristics measurement
each time you change the O2 sensor (approx. every 1 – 2 years).