NATIONAL METROLOGY INSTITUTE OF ETHIOPIA

(NMIE)

CERTIFICATE
This calibration certificate documents the
traceability to national standards, which realize
the units of measurement according to the

OF CALIBRATION international system of units





Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail: info@nmie.net
website: http://www.nmie.net

Date of Issue: 2024-06-26 Certificate number: PPS-0594 Page 1 of 3




Object Blood Pressure Monitor

Manufacturer Omron Health Care Co, Ltd

Type/Model HEM-7121

Serial Number DBP-01

Customer DOT DENTAL CLINIC


Registration No.: 16-0766

Number of pages of the certificate 3

Date of Calibration: 2024-06-25

This calibration certificate may not be reproduced other than in full except with the permission of
the issuing authority. Calibration certificates without signature and seal are not valid.
Seal Date Approved By Calibrated By

2024-06-26 Dawit Yilma Ephrem W/gebreal


_____________________________________________________________________________________________
NATIONAL METROLOGY INSTITUTE OF ETHIOPIA
(NMIE)

CERTIFICATE
This calibration certificate documents the
traceability to national standards, which realize
the units of measurement according to the

OF CALIBRATION international system of units





Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail: info@nmie.net
website: http://www.nmie.net

Date of Issue: 2024-06-26 Certificate number: PPS-0594 Page 2 of 3




Calibrated Object
Manufacturer Omron Health Care Co, Ltd
ID. Number DBP-01
Measuring range: 0 to 300 mmHg
Calibration range: 0 to 260 mmHg
Division: 0.1 mmHg

Standards Type: Master test gauge Serial Number: PS-S-005


Resolution 1 mmHg

Calibration procedure The instrument is calibrated using calibrated Master test gauge.
For selected measurement points, three tests are carried out &
reported on the certificate.

Place of Calibration In NMIE Pressure Lab., Addis Ababa.

Environmental Condition ---

Type of Calibration In house Calibration.

Validity of calibration The values in this certificate are correct at the time of calibration.
Subsequently the accuracy will depend on such factors as the care
exercised in handling and use of the instrument and the frequency of
use. Recalibration should be performed after a period, which has been
chosen to ensure that the instrument’s accuracy remains within the
desired limits.

Measuring results
NATIONAL METROLOGY INSTITUTE OF ETHIOPIA
(NMIE)

CERTIFICATE
This calibration certificate documents the
traceability to national standards, which realize
the units of measurement according to the

OF CALIBRATION international system of units





Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail: info@nmie.net
website: http://www.nmie.net

Date of Issue: 2024-06-26 Certificate number: PPS-0594 Page 3 of 3





Instrument Standard
Reading Reading Correction
[mmHg] [mmHg] [mmHg]
0 0 0
40 40 0
60 60 0
80 80 0
100 100 0
120 119 -1
140 139 -1
160 159 -1
180 179 -1
200 198 -2
220 218 -2
240 239 -1
260 259 -1
Note: Standard (Actual) value = Instrument reading + Correction
Estimated uncertainty of measurement: ± 2 mmHg
The uncertainties are based on root sum square of the contributions with a confidence interval
Not less than 95% and coverage factor k=2.

____________________________End of Certificate_________________________________