NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)

INDUSTRIAL METROLOGY DIRECTORATE

CERTIFICATE
This calibration certificate documents the traceability
to national standards, which realize the units of
measurement according to the international system of

OF CALIBRATION units
PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail:info@nmie.net
website: http://www.nmie.net

Date of Issue: 2024-06-26 Certificate number: OTH-0755 Page 1 of 3

Object Dental Steam Sterilizer Autoclave

Manufacturer GST Corporation Ltd.

Type/Model NUT LOCKING

Serial Number 6814

Customer Dot Dental Clinic

Registration No.: 16-0766

Number of pages of the certificate 3

Date of Calibration: 2024-06-25

This calibration certificate may not be reproduced other than in full except with the permission of
the issuing authority. Calibration certificates without signature and seal are not valid.

Seal Date Approved By Calibrated by

2024-06-26 Dawit Yilma Ephrem W/gebreal


NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)
INDUSTRIAL METROLOGY DIRECTORATE

CERTIFICATE
This calibration certificate documents the traceability
to national standards, which realize the units of
measurement according to the international system of

OF CALIBRATION units
PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail:info@nmie.net
website: http://www.nmie.net

Date of Issue: 2024-06-26 Certificate number: OTH-0755 Page 2 of 3

Calibrated Type: NUT LOCKING


Object Identification No.: 6814

Measuring range: 50 to 200 oC


Calibration range: See Next Page
Scale interval: 5 oC

Standards Type - Standard Type K thermocouple Serial No;PTH-2646,2021-07


Calibration procedure  The instrument is calibrated with Standard thermocouple. For selected
measurement points the standard and instrument readings are reordered
& reported on the certificate.
Place of Calibration A.A on the clients Laboratory
Validity of calibration The values in this certificate are correct at the time of calibration.
Subsequently the accuracy will depend on such factors as the care
exercised in handling and use of the instrument and the frequency of use.
Recalibration should be performed after a period, which has been chosen
to ensure that the instrument’s accuracy remains within the desired limits.

Measuring result

Instrument Standard
Correction
Reading Reading
[oC]
[oC] [oC]
121 122.371 1.4
135 136.264 1.3
NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)
INDUSTRIAL METROLOGY DIRECTORATE

CERTIFICATE
This calibration certificate documents the traceability
to national standards, which realize the units of
measurement according to the international system of

OF CALIBRATION units

PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail:info@nmie.net
website: http://www.nmie.net

Date of Issue: 2024-06-26 Certificate number: OTH-0755 Page 3 of 3

Note: Standard (Actual) value = Instrument reading + Correction

Estimated uncertainty of measurement: ± 0.81 oC


The uncertainties are based on root sum square of the contributions with a confidence interval
Not less than 95% and coverage factor k=2.

_____________________________End of Certificate______________________________