Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 7

Code: EPQ0

Operation procedure for metal detector Section: 4.3


On the packaging line. Revision: 00

Date: 141/08/20
STARTUP AND OPERATION VERIFICATION FOR
THE METAL DETECTOR Page: 1 of 7

Keeping radios, cell phones, keys and any metallic object more than one meter away
from the detector can interfere with detection.

Under no circumstances should you put your hand or any object other than the
product inside the metal detector and the rejection piston.

1. RESOURCES
1.1 Tools – Materials – Equipment
Verification witnesses:
- Blue (Stainless Steel 3.0mm).
- Yellow (Non-ferrous material 2.0mm).
- Black (1.5mm Ferrous Material).
1.2 Reagents
Does not apply.
1.3 Corrective actions
Corrective actions are executed by the Maintenance department.
2. INSTRUCTIONS

2.1 Turn on the equipment and confirm that the conveyor belt starts rolling.

2.2 Carry out a general check of the equipment, engine, pneumatic system, rejection cylinder, to
verify that everything is in perfect condition.

2.3 Verify that the conveyor belt is clean.


Code: EPQ0
Operation procedure for metal detector Section: 4.3
On the packaging line. Revision: 00

Date: 141/08/20
STARTUP AND OPERATION VERIFICATION FOR
THE METAL DETECTOR Page: 2 of 7

2.4 Control Panel.


Screen, three lights, indicators (red, green and yellow), green Enter button, red Exit button and four blue
triangular navigation buttons.

Figure No. 1. Control Panel.

• Indicator lights: Provide a quick overview of how the detector works.


Red: When it flashes it indicates an excess of product effect; If it remains
illuminated, it indicates a fault.
Green: Indicates that there is a product in the detection head.
Yellow: Indicates that a contaminant has been detected in the product.

• Navigation Buttons: These allow you to navigate through the detector's menus and
submenus and are used to increase or decrease numbers, as well as to select characters
and settings in various menus and input screens.

• Enter Button: Press the enter button to select or start one of the detector functions.
Code: EPQ0
Operation procedure for metal detector Section: 4.3
On the packaging line. Revision: 00

Date: 141/08/20
STARTUP AND OPERATION VERIFICATION FOR
THE METAL DETECTOR Page: 3 of 7

• Exit Button: Press the exit button to stop one of the detector functions or to return to
the previous menu screen. Press it several times to return to the main menu.

2.5 Unlocking the panel: To perform this action, you must initially enter the key-shaped icon
highlighted in Figure No. 2. It should be noted that to locate said icon, you must use the
navigation buttons and then press the enter button to enable the selected option.

Figure No. 2. Control panel located on the password icon.

2.6 Enter the access menu: After entering the access menu, you must select the icon highlighted
in Figure No. 3 and then press the enter button.

Figure No. 3. Control panel located in the access option.


Code: EPQ0
Operation procedure for metal detector Section: 4.3
On the packaging line. Revision: 00

Date: 141/08/20
STARTUP AND OPERATION VERIFICATION FOR
THE METAL DETECTOR Page: 4 of 7

2.7 Enter access password: In this menu you must write the access password “SUPE”,
it is important to note that to locate the letters you must use the navigation buttons
and press the enter button to select each letter. Finally, to return, locate the door-
shaped icon, highlighted in Figure No. 4, also by using the navigation buttons and
then press the enter button.

Figure No. 4. Control panel located in the access password option.

2.8 Product selection menu: To perform this action, you must locate the option
highlighted in Figure No. 5. Once there, to change the product, you must press the
navigation buttons to the left or right.

Figure No. 5. Control panel located in the product selection.


Operation procedure for Code: EPQ0
Section: 4.3
metal detector
Revision: 00
On the packaging line.
STARTING AND VERIFICATION Date: 141/08/20
OPERATION FOR THE DETECTOR Page: 5 of 7
METALS
2.9 Enter the exit menu: To enter the exit menu, select the key-shaped icon located on the main panel
explained above (see Figure
No. 2), after entering said menu, you must select the highlighted icon on the
Figure N° 6 and press the enter key to lock the panel again.

Figure No. 6. Control panel located in the output menu.

2.10 Visually verify the correct selection of the type of cheese on the detector screen and then begin to
introduce the product into the device.

2.11 Place the Yellow control (Non-ferrous Material) on the cheese and hold it to the product with a
rubber band (see Figure No. 7).
Code: EPQ0 O
p Section: 4.3 e
mtemu Revision: 00
r Device input 884 a
massa
ti address Date: 141/08/20

KNOLL Witness
Page: 6 of 7 o

Figure No. 7 . Correct position of the control in the


cheese.
n procedure for
metal detector
On the packaging line.
STARTING AND VERIFICATION
OPERATION FOR THE DETECTOR
METALS

2.12 Pass the cheese through the metal detector as indicated in figure No. 7. Verify that the equipment is
capable of detecting the gauge or indicator and also ensure that the piston rejection function is
executed correctly.

2.13 Repeat steps 2.12 and 2.13 using the Red and Blue indicator lights.

2.14 If when passing the tokens with the cheeses through the metal detector, they are rejected, the start-up
and verification process ends. Therefore, the normal packaging operation can continue.

2.15 If the detector does not reject the product with any of the three indicators even after having carried
out the calibration procedure, therefore carry out the complete calibration procedure.

2.16 Calibration of X and R noise thresholds. This calibration establishes a monitoring baseline for the
noise detector and must be performed with the belt running and no product present in the detector
head. (see Figure No. 8).

• Check that the main menu is displayed.


• Highlight the Frequency and Gain menu.
• Press the enter button.
• Scroll to the “Noise Threshold Calibration” menu. (Note: The background of the function
calibration should be highlighted in black, as shown in the figure).

Figure No. 8. Control panel located at the Calibration detection level.


Operation procedure for Code: EPQ0
4.3
metal detector Section:

On the packaging line. Revision: 00

STARTING AND VERIFICATION Date: 141/08/20


OPERATION FOR THE DETECTOR
Page: 7 of 7
METALS

• Press the enter button and a monitor screen will appear. The calibration is complete when the
previous screen reappears showing the calibrated values for the X and R thresholds. Typical
values for these thresholds are as follows. X threshold < 300 and Rthereshold < 100 .

2.17 Full Product Calibration: Full product calibration allows the detector to learn to identify
uncontaminated product and establishes the following basic parameters for the product. Adjustment
of phase angle and detection level. (See Figure No. 9).

• Make sure the product name is highlighted in the main menu.


• Press the enter button and the “Complete Product Calibration” function will be highlighted
automatically.

Figure No. 9. Control panel located in the Complete Product Calibration.

• Make sure the conveyor belt is running and passing uncontaminated product through the detector
head.
• Press the Enter button and a monitor screen will appear. During the calibration process, a series of
secondary frequency/gain screens may appear that pause the calibration process while the detector
searches for the optimal frequency/gain setting for the product. When this occurs, press the Enter
button to restart the calibration process. The main product screen appears when the calibration is
complete.

You might also like