Professional Documents
Culture Documents
tso3-sterizone-vp4-sterilizer-technical-monograph
tso3-sterizone-vp4-sterilizer-technical-monograph
[Texte]
STERIZONE® VP4 Sterilizer - Technical Monograph
The TSO 3 logo and STERIZONE® are registered trademarks of TSO 3 Inc.
The STERIZONE® VP4 Sterilizer is patent pending in the US, corresponding foreign
patent applications pending and/or granted.
2
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table of contents
1.0 INTRODUCTION .............................................................................................. 4
1.1. Single Sterilization Cycle................................................................................. 5
1.2. Components ..................................................................................................... 6
2.0 STERIZONE® VP4 STERILIZER MODE OF OPERATION ............................. 7
3.0 MONITORING OF THE PROCESS .................................................................. 9
3.1. Sterilizer Control ............................................................................................. 9
3.2. Load Control ................................................................................................... 9
3.3. Packaging Control ......................................................................................... 10
4.0 MICROBIAL EFFECTIVENESS ..................................................................... 10
4.1. Demonstration of 10-6 Sterility Assurance Level (SAL) ................................. 10
4.1.1. Selection of the most resistant microorganism ........................................ 11
4.1.2. Load composition ................................................................................... 11
4.1.3. Exposure conditions (Half-cycle definition) ........................................... 12
4.2. AOAC Sporicidal Test ................................................................................... 16
4.3. Simulated-use Tests ....................................................................................... 17
4.3.1. Material study ........................................................................................ 19
4.4. In-use tests (Hospital study) ........................................................................... 21
5.0 MATERIAL COMPATIBILITY ...................................................................... 22
6.0 SAFETY........................................................................................................... 24
6.1. Hydrogen peroxide ........................................................................................ 24
6.2. Ozone ............................................................................................................ 24
6.3. Emission of hydrogen peroxide and ozone at the outlet of the catalytic
converter........................................................................................................ 24
6.4. Determination of hydrogen peroxide and ozone in the breathing zone ............ 26
6.5. Safety of medical instruments (Biocompatibility) .......................................... 27
3
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
1.0 INTRODUCTION
4
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
5
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Validation Load
Load description
load # weight1
Validation load #4 consisted of up to 15 rigid or semi-rigid channeled
instruments in the presence of other packaged medical devices. Three double
channel semi-rigid endoscopes (ureteroscope – 0.7 mm × 500 mm and 1.1
mm × 500 mm) were packaged individually in wrapped trays or containers
4 including appropriate silicone brackets or mats. Additional rigid channeled 19 lb
instruments or stainless steel rigid lumens were added to each package. Two
additional general medical instruments, each packaged in a pouch, were
added.
1.2. Components
The following components are part of the STERIZONE® VP4 Sterilizer system:
• STERIZONE® VP4 Sterilizer
• STERIZONE® Loading Rack
• 125-280 Solution™
• STERIZONE® BI+ Self-contained Biological Indicator
• STERIZONE® CI+ Chemical Indicator
• STERIZONE® VP4 Test Pack
6
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Prior to sterilization, cleaned and dried instruments are packaged using either
aluminum containers compatible with hydrogen peroxide in combination with disposable
polypropylene filters, and either wrapped trays or heat-sealable non-woven, polyethylene
sterilization pouches.
The STERIZONE® VP4 Sterilizer uses dual sterilants, vaporized hydrogen
peroxide (H 2 O 2 ) and ozone (O 3 ), in a multiphase process.
Upon loading the medical devices into the sterilization chamber and closure of the
door, the chamber is subjected to a vacuum of 1 Torr (referred to as pre-conditioning
step). The first cycle phase (Phase 1) is initiated with the Dynamic H 2 O 2 exposure step.
During this step, a 50 weight-percent hydrogen peroxide solution (referred to as 125-280
Solution™) is injected at a fixed injection rate in vapor form into the sterilization
chamber through a continuous micro-pulsed injection until a differential pressure set
point of 19 Torr is reached (i.e., the actual chamber pressure is 20 Torr, less the initial
vacuum of 1 Torr, which is equivalent to a “differential pressure” or “∆P” of 19 Torr).
The total amount of hydrogen peroxide introduced into the sterilization chamber and thus
the duration of the injection varies depending on load composition (e.g. surface area),
weight and temperature.
The second step of the cycle phase is the H 2 O 2 Reduction Step. During this step, a
set concentration of ozone is injected into the chamber, which reacts with residual
hydrogen peroxide to form hydroxyl radicals, further enhancing lethality.
During the second cycle phase (Phase 2), the same sequence is repeated, including
the Dynamic H 2 O 2 exposure and H 2 O 2 reduction steps. The full Cycle (Figure 1) is then
completed with an evacuation and ventilation, through a catalytic converter, to the
atmosphere, at which point the chamber door can be safely opened.
The cycle process parameters are summarized in Table 2.
Table 2. STERIZONE® VP4 Sterilizer – Cycle process parameters
Hydrogen peroxide exposure Ozone exposure
Chamber Nb of
Vaporizer /
Hydrogen differential Sterilant O3 O3
Time Chamber phases
peroxide solution pressure injected injection dwell
temperature
set point
125-280
210-600 120°C /
Solution™ 19 Torr 8.4-24 g* 2 mg/L 5 min 2
sec* 41 ± 3°C
(H2 O 2 50 wt%)
* Vaporized hydrogen peroxide injection/exposure time (Dynamic H2 O 2 exposure step) varies with load
composition and conditions. The quantity of vaporized hydrogen peroxide injected is directly related to
the time required to reach a pressure differential of 19 Torr in the chamber, for load temperature ranging
from 20°C to 26°C. If the H2 O 2 injection time is less than 210 seconds, or greater than 600 seconds, the
cycle will abort.
7
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
8
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
The STERIZONE® VP4 Test Pack has equivalent to or greater resistance than the
worst case devices and loads in any load configuration, and is designed to be more
resistant than the full half-cycle, including exposure to hydrogen peroxide and ozone.
9
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
10
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
11
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
The type and quantity of instruments included in the validation loads highlights the
utility of the STERIZONE® VP4 Sterilizer. It is the only sterilization process using
hydrogen peroxide that can sterilize three flexible endoscopes in a single load,
particularly with additional medical devices (Figure 4), or sterilize semi-rigid endoscopes
and flexible endoscopes in the same load. The STERIZONE® VP4 Sterilizer is also the
only low temperature sterilizer using hydrogen peroxide capable of processing load
weights of up to 75 lb.
Figure 4. Validation load #3: three single channel flexible endoscopes with additional
general medical devices
12
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 3. Half-cycle validation results for general instruments (loads #1 and #2)
Number sterile/
Instrument name Inoculation site Number tested
Handle 3/3
Camera (head)
ON/OFF button 3/3
Battery pack One side 3/3
Kelly Hemostats A (Open position) Clamp (both sides) 3/3
Kelly Hemostats B (Open position) Serrated surface (both sides) 3/3
Mayo scissors (Open position) Pivot hinge 3/3
Crile hemostats (Open position) Box-lock hinge 3/3
Electrophysiology cable Patient lead connector 3/3
Cystoscope stopcock Pivot mechanism 3/3
Defibrillator handles Handle 3/3
One black side of the paddle
Internal defibrillator paddles (2) 3/3
spoon
Open screw 3/3
Reuter tip deflecting handle
Closed screw 3/3
Stryker System 6 (drill) Cannula (5.0 mm x 12 cm) 3/3
Stryker System 6 (drill attachment) Drill bit mechanism (jaws) 3/3
Stryker System 6 (wire collet) Cannula (0.7/1.8 mm × 9.0 cm) 3/3
Bulldog clamp Spring (basis of the spring) 3/3
Forcep micro cup end Distal end only (cup) 3/3
Rongeur punch A (Kerrison) Hinge with screw 3/3
Rongeur punch B (Kerrison) Gliding mechanism 3/3
Resectoscope sheath Locking bridge mechanism 3/3
Haemorrhoidal ligator Distal end (swivel parts) 3/3
Fiberoptic lightcable 10 cm2 of cord 3/3
Rigid telescope (no lumen) Distal end (lens + 2 cm2 distal 3/3
tip)
Proximal lens (lens only) 3/3
Eyepiece (surface) 3/3
Syringe (unlock) Luer-lock 3/3
13
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 4. Half-cycle validation results for short single channel flexible endoscopes
(load #3)
Number sterile/
Instrument name Inoculation site Number tested
Table 5. Half-cycle validation results for semi-rigid or rigid endoscope (load #4)
Number sterile/
Instrument name Inoculation site Number tested
Channel: 0.7 mm × 500 mm 9/9
Semi-rigid ureteroscope
Channel: 1.0 mm × 500 mm 9/9
Stainless steel tubing Channel: 2.0 mm × 575 mm 6/6
Resectoscope Channel: 2.2 mm × 173 mm 3/3
(working element) Channel: 4.7 mm × 270 mm 3/3
Channel: 7.0 mm × 227 mm 3/3
Rotary resectoscope with irrigation
Channel: 7.8 mm × 198 mm 3/3
Thoracoscope Channel: 4.0 mm × 370 mm 3/3
Channel: 5.5 mm × 166 mm 3/3
Trocar
Channel: 7.0 mm × 105 mm 3/3
Cystourethroscope short bridge
Pivot mechanism 3/3
double-horn
Table 6. Half-cycle validation results for the mixed load (load #5)
Number sterile/
Instrument name Inoculation site Number tested
Channel: 1.0 mm × 850 mm
Flexible ureteroscope 3/3
Channel: 0.7 mm × 500 mm 3/3
Semi-rigid ureteroscope
Channel: 1.0 mm × 500 mm 3/3
Channel:1.8 mm × 830 mm
Flexible Bronchoscope 3/3
Table 7. Half-cycle validation results for the light weight load evaluation (Load #6)
Number sterile/
Instrument name Inoculation site Number tested
Haemorrhoidal ligator Distal end (swivel parts) 3/3
Rongeur punch A (Kerrison) Hinge with screw 3/3
Stryker System 6 (wire collet) Cannula (0.7/1.8 mm × 9.0 cm) 3/3
14
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 8. Half-cycle validation results for the heavy weight load evaluation (Load #7)
Number sterile/
Instrument name Inoculation site Number tested
Mayo scissors Pivot hinge 9/9
Crile hemostats Box-lock hinge 9/9
Syringe (unlock) Luer-lock 9/9
15
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 9. AOAC Sporicidal Screening Test Results with the STERIZONE® VP4
Sterilizer
Number of positive/Number tested
Carriers B. subtilis C. sporogenes
Validation test
Penicylinders 0/180 0/180
Dacron suture loops 0/180 0/180
Confirmatory test
Penicylinders 0/30 0/30
Dacron suture loops 0/30 0/30
Based on the results obtained during the validation and testing of the AOAC
sporicidal activity, the conditions found in the STERIZONE® VP4 Sterilizer meet the
requirements of the Official Methods of Analysis of AOAC International to qualify as a
sterilant.
16
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
17
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 11. Simulated-use validation results for short single channel flexible
endoscopes (load #3)
Number sterile/
Instrument name Inoculation site Number tested
Table 12. Simulated-use validation results for semi-rigid or rigid endoscope (load #4)
Number sterile/
Instrument name Inoculation site Number tested
Channel: 0.7 mm × 500 mm 9/9
Semi-rigid ureteroscope
Channel: 1.0 mm × 500 mm 9/9
Stainless steel tubing Channel: 2.0 mm × 575 mm 6/6
Resectoscope Channel: 2.2 mm × 173 mm 3/3
(working element) Channel: 4.7 mm × 270 mm 3/3
Channel: 7.0 mm × 227 mm 3/3
Rotary resectoscope with irrigation
Channel: 7.8 mm × 198 mm 3/3
Thoracoscope Channel: 4.0 mm × 370 mm 3/3
Channel: 5.5 mm × 166 mm 3/3
Trocar
Channel: 7.0 mm × 105 mm 3/3
Cystourethroscope short bridge
Pivot mechanism 3/3
double-horn
Table 13. Simulated-use validation results for the mixed load (load #5)
Number sterile/
Instrument name Inoculation site Number tested
Channel: 1.0 mm × 850 mm
Flexible ureteroscope 3/3
Channel: 0.7 mm × 500 mm 3/3
Semi-rigid ureteroscope
Channel: 1.0 mm × 500 mm 3/3
Channel:1.8 mm × 830 mm
Flexible Bronchoscope 3/3
Table 14. Simulated-use validation results for the light weight load evaluation
(Load #6)
Number sterile/
Instrument name Inoculation site Number tested
Haemorrhoidal ligator Distal end (swivel parts) 3/3
Rongeur punch A (Kerrison) Hinge with screw 3/3
Stryker System 6 (wire collet) Cannula (0.7/1.8 mm × 9.0 cm) 3/3
18
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 15. Simulated-use validation results for the heavy weight load evaluation
(Load #7)
Number sterile/
Instrument name Inoculation site Number tested
Mayo scissors Pivot hinge 9/9
Crile hemostats Box-lock hinge 9/9
Syringe (unlock) Luer-lock 9/9
Data confirm that the STERIZONE® VP4 Sterilizer reproducibly and repeatedly
sterilizes challenging medical instruments under worst-case simulated-use conditions.
19
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
20
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
In-use studies confirm that devices used in actual surgical procedures are sterilized to the
same degree as inoculated devices in a controlled environment.
21
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Material and device compatibility testing is an integral and fundamental part of the
verification and validation of the STERIZONE® VP4 Sterilizer. Exhaustive testing
ensures that the material and devices are compatible with the sterilization process, and
that the reusable medical devices remain safe, sterile, and work as designed.
Compatibility testing ensures that the devices can be safely and repeatedly
processed, and continue to meet performance specifications as determined by the
manufacturer. TSO 3 Inc. performs testing, as part of its Compatibility Testing Services
(CTS), in an ongoing and rigorous program, with the cooperation of the device
manufacturers. This testing is based on the intended use and manufacturer’s instructions
for reprocessing.
Functionality is determined by the manufacturer and is based on the number of
cycles selected for testing of a device. It is dependent on many factors, determined by the
manufacturer, such as:
• Industry standards
• Historical data
• Intended use
• Device criticality
• Warranty
• Expected life of device
Device compatibility is assessed after the predetermined number of sterilization
cycles to ensure that the device is not adversely affected by exposure to repeated
STERIZONE® VP4 Sterilizer cycles. The compatibility is assessed based on visual and
mechanical criteria for a specific device, and is based on functional and cosmetic
characteristics such as:
• Overall appearance and integrity
• Optics
• Mechanical properties and proper function of moving parts
• Visual observations
Table 18 shows the categories of medical devices whose compatibility has been
tested. All devices were found to be compatible with Cycle 1 of the STERIZONE® VP4
Sterilizer.
22
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 18. Compatible device categories for the STERIZONE® VP4 Sterilizer Cycle 1
Medical Instruments Category
Battery
Bipolar cord
Bipolar forceps
Bridge
Camera
Defibrillator paddles and handles
Doppler probe
Electrocautery instrument
Electrode
Electrophysiology cable
Electrosurgical cable
Fiberoptic light cable
ENT flexible endoscope
Flexible bronchoscope
Flexible cystoscope
Flexible hysteroscope
Flexible nephroscope
Flexible ureteroscope
Forceps
Laryngoscope blade
High frequency cable
Lacrimal probe
Laser fiber
Lighted retractor
Ophthalmic surgical instrument (without channel)
Resectoscope and resectoscope sheath
Rigid endoscope (without channel)
Rigid hysteroscope
Scalpel handle
Semi-rigid and rigid endoscope
Speculum
Stopcock
Temperature probe
Thoracoscope
Trocar and trocar sheath
23
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
6.0 SAFETY
The toxicity of hydrogen peroxide (H 2 O 2 ) and ozone are well documented in the
scientific literature, with relevant standards having been established by the United States
Occupational Safety and Health Administration (OSHA).
The by-products from reaction of hydrogen peroxide with ozone are oxygen and
hydroxyl radicals. Hydroxyl radicals, in turn, can react with ozone or hydrogen peroxide
to produce either oxygen or water (non-toxic by-products).
Both hydrogen peroxide and ozone decompose with the formation of water (H 2 O)
and oxygen gas (O 2 ).
2 H2 O2 → 2 H2 O + O2
2 O3 → 3 O2
6.2. Ozone
In accordance with OSHA (Occupational Safety and Health Administration)
standards (29 CFR part 1910.1000, Table Z-1), an individual must not be exposed to a
concentration of ozone higher than:
• An average of 0.1 ppm (0.2 mg ozone/m3 air) over a time-weighted-average (TWA)
of eight hours.
• A maximum of 0.3 ppm (0.6 mg ozone/m3 air) as ceiling limit (CL) for a short-term
exposure limit (STEL) of 15 minutes.
24
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 19. H 2 O 2 and O 3 concentration at the outlet of the catalytic converter after
every 100 cycles*
Cycle number H 2 O 2 Concentration O 3 Concentration
(ppm) (ppm)
0 0.00 0.003
100 0.00 0.010
200 0.00 0.000
300 0.01 0.013
400 0.02 0.015
500 0.00 0.001
600 0.00 0.013
700 0.00 0.011
800 0.00 0.010
900 0.00 0.014
1000 0.00 0.020
1100 0.06 0.017
1200 0.02 0.008
1300 0.02 0.004
1400 0.00 0.012
1500 0.07 0.012
1600 0.08 0.011
1700 0.05 0.015
1800 0.04 0.017
1900 0.06 0.027
2000 0.10 0.012
2100 0.00 0.014
2900 0.00 0.024
3000 0.03 0.022
3100 0.00 0.019
3200 0.00 0.021
3300 0.02 0.009
* The data of seven points are missing, the monitors were in calibration
25
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
Table 21. Ozone peak reached from the opening of the door to the end of the load
manipulation (minimum 15 minutes) in mg/m3
26
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
27
MA-200-045 r0
STERIZONE® VP4 Sterilizer - Technical Monograph
BIBLIOGRAPHY
Chaunet, M., S. Dufresne and S. Robitaille. 2007. The Sterilization Technology for the
21st century. white paper. TSO 3 .
European Commission: Health & Consumer Protection Directorate-General. 2005.
Opinion on Hydrogen Peroxide in Tooth Whitening Products. Adopted by the
Scientific Committee on Consumer Products the 15th of March 2005. 50 pages
European Chemical Bureau. 2003. European Union Risk Assessment Report -
hydrogen peroxide. Vol. 38.Helsinki, Finland, Office for Official Publications of
the European Communities. 246 pages
Ikarashi, Y., T. Tsuchiya and A. Nakamura. 1995. Cytotoxicity of medical materials
sterilized with vapor-phase hydrogen peroxide. Biomaterials 16(3): 177-183.
ISO 10993-1: 2009 Biological evaluation of medical devices — Part 1: Evaluation and
testing within a risk management process. ISO Geneva, Switzerland, 21 pages
Smith, D.F. 2008. STERRAD® 100 NX™ Sterilization System Technical Information.
White paper AD-54083-001 Rev B. Advanced Sterilization Products. Irvine,
CA. 16 pages
STERIS. 2008. Technical Data Monograph: Amsco® V-PRO™ 1 Low Temperature
Sterilization System. STERIS Corporation.
TSO 3 Inc.
2505, avenue Dalton
Québec (Québec)
Canada G1P 3S5
Telephone: 418.651.0003
Fax: 418.653.5726
Technical Services: 1.888 .651.8763
www.tso3.com
28
MA-200-045 r0