Professional Documents
Culture Documents
Sesi 2 Petra Lewis MERCK
Sesi 2 Petra Lewis MERCK
Sesi 2 Petra Lewis MERCK
impurity profiling
by HPLC
Microbial
Purity Stability
For oral dosage forms: Short-term stability testing
total aerobic microbial count (Formulation Development)
total combined yeast and molds
Stability testing acc. to ICH
For parenteral dosage forms: and forced degradation
sterility experiments
Endotoxins
Microbial
Purity Stability
For oral dosage forms: Short-term stability testing
total aerobic microbial count (Formulation Development)
total combined yeast and molds
Stability testing acc. to ICH
For parenteral dosage forms: and forced degradation
sterility experiments
Endotoxins
MANUFACTURE
• Filtration, Chromatography
• Sterility testing, bioburden,
environmental monitoring
• HPLC analysis
• HPLC Analysis
• Reference materials
• Reference
Materials • Impurity testing
• Impurity
testing
• Cell Analysis
• Cell Analysis • GC analysis
• MultiScreen Filter • Sterility testing, bioburden,
• MultiScreen • MultiScreen Filter
Plates environmental monitoring
Filter Plates Plates
• Life Science • Media Fill Release Testing/
• Life Science • Life Science
Research Tools Dehydrated Culture Media
Research Research Tools • Sterility testing,
Tools • Chemical bioburden, • HPLC analysis
• Chemical synthesis
synthesis environmental
• Chemical • HPLC analysis • Reference materials
Synthesis monitoring
• Bioanalysis
• GMP-grade raw • Impurity testing
materials
• HPLC analysis
• Sterility testing
• Reference
• Life Science • HPLC analysis
materials
Research Tools • Reference materials
• HPLC Analysis
• Impurity testing
• Reference
materials
• Bioanalysis
Following terms are used by various regulatory bodies and ICH to describe the impurities:
1. Intermediate
2. Penultimate intermediate
3. By-products
4. Transformation products
5. Interaction products
6. Related products
7. Degradation products
▪ ICH guidelines “stability testing of new drug substances and products"- Q1A
▪ ICH guidelines “Impurities in New Drug Substances”- Q3A
▪ ICH guidelines “Impurities in New Drug Products”- Q3B
▪ ICH guidelines “Impurities: Guidelines for residual solvents”- Q3C
▪ US-FDA guidelines “NDAs -Impurities in New Drug Substances”
▪ US-FDA guidelines “ANDAs – Impurities in New Drug Substances”
▪ Australian regulatory guideline for prescription medicines, Therapeutic
Governance Authority (TGA), Australia
▪ System repeatability: When a relative standard deviation requirement is specified in an individual monograph, if the requirement is 2.0 or
less the calculation is based on data from five replicate injections of the analyte, if the requirement is more than 2.0% data from six replicate
injections are used.
▪ Precision:
❖ Assay: RSD ≤1% (API) or ≤ 2% (FPP), n ≥ 5, >2% n=6 Compliance with the system suitability
❖ Impurities: in general, RSD ≤ 5% at the limit level, up to 10% or higher at LOQ, n ≥ 6 criteria is required throughout the
chromatographic procedure. No sample
▪ Signal to noise ratio: analysis is acceptable unless the suitability
❖ Detection limit: signal-to-noise ratios are 2:1 or 3:1. of the system has been demonstrated
❖ Quantitation limit: typically acceptable signal-to-noise ratio is 10:1.
▪ Tailing factor/peak asymmetry: Unless otherwise stated, in a test or assay, the symmetry factor (tailing factor) of the peak used for
quantification is 0.8–1.8.
Experimental Conditions SST Stock 0.25 mg/mL of USP Venlafaxine Hydrochloride RS and 0.1 mg/mL of USP
Solution Desvenlafaxine Related Compound A RS prepared as follows. Transfer
Column Purospher® STAR RP-18 endcapped (5µm) 250x4.6 mm suitable amounts of USP Venlafaxine Hydrochloride RS and USP
Detection UV @ 226 nm (analytical flow cell; 10 µL) Desvenlafaxine Related Compound A RS to an appropriate volumetric flask
and add 65% of the total flask volume of Diluent. Dilute with Diluent to
Injection volume 20 µL volume.
Flow rate 1 mL/min
SST Solution 500 μg/mL of USP Desvenlafaxine RS, 2.5 μg/mL of USP Venlafaxine
Column: 35°C Hydrochloride RS, and 1 μg/mL of USP Desvenlafaxine Related Compound A
Temperature RS prepared as follows. Transfer 50 mg of USP Desvenlafaxine RS to a 100-
Autosampler: 20°C
mL volumetric flask. Add 4 mL of citric acid and 20 mL of acetonitrile.
Pressure drop 90-155 bar (1305-2248 psi) Sonicate to dissolve and then add 30 mL of sodium chloride and sonicate
Buffer 5.6 g/L of sodium perchlorate in water for another 5 min. Then add 1 mL of SST stock solution and dilute with
sodium chloride to volume.
Mobile Phase A: Buffer : Acetonitrile (90:10)
B: Acetonitrile Standard 1 μg/mL (1 ppm) of USP Desvenlafaxine RS in sodium chloride solution,
Diluent Mix water and acetonitrile 50:50 (v/v) Solution filter with 0.45-μm pore size. Discard 2 mL and use the filtrate.
Sensitivity 0.25 μg/mL of USP Desvenlafaxine RS from Standard solution in sodium
Gradient see table
Solution chloride passed through a suitable filter of NLT 0.45-μm pore size. Discard
NLT 2 mL and use the filtrate.
Time (min) A(%) B(%)
Test Solution Nominally 500 μg/mL of desvenlafaxine from Tablets prepared as follows.
0 80 20
Transfer a portion of powder from NLT 20 Tablets, equivalent to 125.0 mg
5 80 20 of desvenlafaxine, to a 250-mL volumetric flask. Add 50 mL of acetonitrile
35 60 40 and sonicate for about 5 min. Add 10 mL of citric acid and sonicate for
about 5 min more with occasional shaking. Add 100 mL of sodium chloride
40 20 80
and sonicate for about 20 min with intermittent shaking. Dilute with sodium
42 80 20 chloride to volume. Pass through a filter of 0.45-μm pore size, discard NLT
50 80 20 2 mL, and use the filtrate.
Desvenlafaxine
Retention
Time Tailing
Peak Compound (min) Resolution Factor RRT
1. Specificity: Inject solution and determine the 3. Linearity, LOD & LOQ
retention time of desired analyte in presence of other
Concentration
component like impurities and excipient. Desvenlaflaxine
(μg/mL)
Compound RT (min) 0.25 19085
Desvenlafaxi 0.5 34351
1 8.0
ne
1 72931
Related
2 16.2 5 352585
compound A
12.5 874866
3 Venlafaxine 17.7
25 1748644
40 2838462
2. Standard Repeatability (25 ppm) 50 3553944
STD 5 69126
Mean 69229
Standard Deviation 198.97
RSD (%) 0.3
All acceptance criteria fully met
Speed
Cost
Column: Purospher® STAR RP-18 endcapped (5μm) Hibar® RT 250x4.6 mm Purospher ® STAR RP-18 endcapped (2μm) Hibar ® HR 100x2.1 mm
Injection: 10 μL 2 μL
Cell: 10 μL 2.5 μL (Use 0.1 mm tubing)
Flow Rate: 1.0 mL/min Time saving: 72% 0.3 mL/min
Pressure Drop: 61 to 74 Bar (884 to 1073 psi) 196 to 164 Bar (2827 to 2378 psi)
Solvent saving: 92%
Isocratic Gradient
Column length & particle May be modified, provided that the ratio of the column length (L) to the particle size (dp) remains constant or in the
size (FPP to FPP) range between -25% to +50% of the prescribed L/dp ratio.
Column length & particle Other combinations of L and dp can be used, Other combinations of L and dp can be used provided that
size (FPP to SPP or provided that the plate number (N) is within -25% the ratio (tR/Wh)2 is within -25% to +50%, relative to the
monolithic) to +50%, relative to the prescribed column. These prescribed column for all the peaks used to determine the
changes are acceptable, provided that system system suitability parameters. These changes are acceptable
suitability criteria are fulfilled, and selectivity and provided system suitability criteria are fulfilled, and
elution order of the specified impurities to be selectivity and elution order of the specified impurities to be
controlled are demonstrated to be equivalent. controlled are demonstrated to be equivalent.
Inner diameter In absence of a change in particle size and/or length, the internal diameter of the column may be adjusted.
Flow rate When particle size is changed, the flow rate requires Flow rate recalculation using formula.
adjustment, but column performance should not
drop by more than 20%; when column dimension
does not change ±50% change is permitted
pH ± 0.2 units
UV wavelength No adjustment permitted No adjustment permitted
Buffer salts ± 10%
concentration
Mobile phase ± 30% relative or ± 10% absolute whichever is ± 15 % retention window, SST must be met
composition smaller
Column temperature ± 100C ± 50C
Column: Purospher® STAR RP-18 endcapped (5μm) Hibar® RT 250x4.6 mm Purospher ® STAR RP-18 endcapped (2μm) Hibar ® HR 100x2.1 mm
Injection: 10 μL 2 μL
Cell: 10 μL Adjustment of column length 2.5 μL (Use 0.1 mm tubing)
Flow Rate: 1.0 mL/min 0.3 mL/min
Monograph: [L/dp = 50000]
Pressure Drop: 61 to 74 Bar (884 to 1073 psi) 196 to 164 Bar (2827 to 2378 psi)
New: [L/dp = 50000]
Isocratic Gradient
Column length & particle May be modified, provided that the ratio of the column length (L) to the particle size (dp) remains constant or in the
size (FPP to FPP) range between -25% to +50% of the prescribed L/dp ratio.
Column length & particle Other combinations of L and dp can be used, Other combinations of L and dp can be used provided that
size (FPP to SPP or provided that the plate number (N) is within -25% the ratio (tR/Wh)2 is within -25% to +50%, relative to the
monolithic) to +50%, relative to the prescribed column. These prescribed column for all the peaks used to determine the
changes are acceptable, provided that system system suitability parameters. These changes are acceptable
suitability criteria are fulfilled, and selectivity and provided system suitability criteria are fulfilled, and
elution order of the specified impurities to be selectivity and elution order of the specified impurities to be
controlled are demonstrated to be equivalent. controlled are demonstrated to be equivalent.
Inner diameter In absence of a change in particle size and/or length, the internal diameter of the column may be adjusted.
Flow rate When particle size is changed, the flow rate requires Flow rate recalculation using formula.
adjustment, but column performance should not
drop by more than 20%; when column dimension
does not change ±50% change is permitted
pH ± 0.2 units
UV wavelength No adjustment permitted No adjustment permitted
Buffer salts ± 10%
concentration
Mobile phase ± 30% relative or ± 10% absolute whichever is ± 15 % retention window, SST must be met
composition smaller
Column temperature ± 100C ± 50C
• A change in column dimensions, and thus in column volume, impacts the gradient volume, which controls selectivity
• Gradients are adjusted to the column volume by changing the gradient volume in proportion to the column volume. This applies to every
gradient segment volume
• Since the gradient volume is the gradient time (tG) multiplied by the flow rate (F), the gradient time for each gradient segment must be
adjusted to maintain a constant ratio of the gradient volume to the column volume (expressed as L × dc2)
Column: Purospher® STAR RP-18 endcapped (5μm) Hibar® RT 250x4.6 mm Purospher ® STAR RP-18 endcapped (2μm) Hibar ® HR 100x2.1 mm
Injection: 10 μL 2 μL
Cell: 10 μL 2.5 μL (Use 0.1 mm tubing) Injection Volume
Flow Rate: 1.0 mL/min 0.3 mL/min
Pressure Drop: 61 to 74 Bar (884 to 1073 psi) 196 to 164 Bar (2827 to 2378 psi) [0.83 µL]
2.1² x 5
x 1 = 0.52 mL/min
4.6² x 2
▪ Full loadability
Fused-Core®
Superficially porous silica particles (SPP)
▪ Ascentis Express®
▪ BIOshell®
▪ Analytical scale Micro /UHPLC / HPLC
▪ Highest efficiency (Resolution)
Macropores
Monolithic silica 2µm
▪ Chromolith®
▪ Chromolith® WP
1.5µm
▪ Scalable from Micro-LC to Semi-Preparative
1.15µm
▪ Outstanding matrix tolerance / lifetime 120Å
Mesopores
300Å
Isocratic Gradient
Column length & particle May be modified, provided that the ratio of the column length (L) to the particle size (dp) remains constant or in the
size (FPP to FPP) range between -25% to +50% of the prescribed L/dp ratio.
Column length & particle Other combinations of L and dp can be used, Other combinations of L and dp can be used provided that
size (FPP to SPP or provided that the plate number (N) is within -25% the ratio (tR/Wh)2 is within -25% to +50%, relative to the
monolithic) to +50%, relative to the prescribed column. These prescribed column for all the peaks used to determine the
changes are acceptable, provided that system system suitability parameters. These changes are acceptable
suitability criteria are fulfilled, and selectivity and provided system suitability criteria are fulfilled, and
elution order of the specified impurities to be selectivity and elution order of the specified impurities to be
controlled are demonstrated to be equivalent. controlled are demonstrated to be equivalent.
Inner diameter In absence of a change in particle size and/or length, the internal diameter of the column may be adjusted.
Flow rate When particle size is changed, the flow rate requires Flow rate recalculation using formula.
adjustment, but column performance should not
drop by more than 20%; when column dimension
does not change ±50% change is permitted
pH ± 0.2 units
UV wavelength No adjustment permitted No adjustment permitted
Buffer salts ± 10%
concentration
Mobile phase ± 30% relative or ± 10% absolute whichever is ± 15 % retention window, SST must be met
composition smaller
Column temperature ± 100C ± 50C
3000
Particle size
distribution 2500
Fused-Core particle size distribution
Average = 2.77 µm;
2000 standard deviation = 6% of mean
500
Amount
Particle Diameter, µm
-500
1 2 3 4 5 6 7
Eddy
diffusion
Poorly-ordered HPLC packed bed
Diffusion
path/Mass
Transfer
Common Fully Porous C18, 3 µm
72.5 % ACN 2
N = 140,600 N/m, N = 21,090 N/col.
Pressure = approx. 4,000 psi
1 4
3
1. p-Hydroxy ethylbenzene
2. Napthalene
3. p-Xylene
4. Biphenyl 0 2 4
2
Ascentis Express C18, 2.7 µm
65 % ACN
N = 237,700 N/m, N = 35,655 N/col.
Pressure = approx. 4,000 psi 1
4
0 2 4
Column: 150 x 4.6 mm; Mobile phase: ACN/water; Flow rate: 1.5 mL/min; Injection: 2.0 mL; Detection: 220 nm
0 0 0
Phosphate buffer Dissolve 0.9 g of dipotassium hydrogen phosphate and 2.7 g of potassium dihydrogen
phosphate in
900 mL of water and mix well. Adjust to pH 6.0 with phosphoric acid, dilute to 1000 mL
with water and filter.
Test solution (a) Dissolve 50.0 mg of the substance to be examines in the solvent mixture and dilute to
100 mL with the solvent mixture.
Test solution (b) Dilute 5.0 mL of the test solution (a) to 100 mL with the solvent mixture.
Reference solution (a) Dissolve 50.0 mg of Lopinavir CRS in the solvent mixture and dilute to 100 mL with the
solvent mixture.
Dilute 5 mL of this solution to 100 mL with the solvent mixture.
Injection: 12 µL
Temperature: 50 °C
https://www.sigmaaldrich.com/US/en/technical-documents/protocol/analytical-chemistry/small-molecule-hplc/lopinavir-european-
pharmacopeia-hplc-assay-method
33 Time,
KGaA,min
0 ©2023 HPLC Impurity profiling - Petra Lewits, Merck Darmstadt
2
Maximum Resolution and Speed
USP Monograph for Itraconazole
• Itraconazole is an antifungal medication
FPP C18, 3 µm, 100 mm x 4.6 mm I.D. • Suitability Requirements
10 µL, 1.5 mL/min, 30 C • Tailing factor: NMT 2.0
Gradient: 20−50% B in 12 min • RSD: NMT 2.0%
Back pressure: 242 bar
SPP
10 µm
Monolith
Particles
Pressure (bar)
Monolith
Matrix tolerance
5 µm 5µm
3 µm
2.7µm
2µm
<2 µm
100 000 N/m 150 000 N/m 200 000 N/m 300 000 N/m
Separation efficiency
36 ©2023 HPLC Impurity profiling - Petra Lewits, Merck KGaA, Darmstadt
Column material selection Loadability
Matrix tolerance vs. Efficiency +++
FPP
SPP +
10 µm
Monolith ++
Particles
Pressure (bar)
Monolith
Matrix tolerance
5 µm 5µm
3 µm
2.7µm
2µm
<2 µm
100 000 N/m 150 000 N/m 200 000 N/m 300 000 N/m
Separation efficiency
37 ©2023 HPLC Impurity profiling - Petra Lewits, Merck KGaA, Darmstadt
Chromatographic Performance
Comparison Monolithic vs. FPP vs. SPP
Silica monolith Fully porous silica particle (FPP) Superficially porous silica particle (SPP)
Chromolith® HighResolution RP-18e C18, 3 µm C18, 2.7 µm
Pressure = 36 bar Pressure = 146 bar Pressure = 170 bar
Chromatographic Conditions
Column: Silica monolith: Chromolith® HighResolution RP-18e 100-4.6 mm
FPP: PurospherTM STAR RP18e 3 µm, 100-4.6mm
SPP: Ascentis® Express C18, 2.7µm 100-4.6 mm
Mobile Phase: A: 100% acetonitrile
B: 20 mM Phosphate buffer pH 4.5
Gradient: Time/min %A %B
0 20 80
12.0 80 20 Sample: 1. Ascorbic acid
Flow rate: 1.0 mL/min 2. 4-Hydroxybenzoic acid
Detection: UV 230 nm 3. Benzoic acid
Detection cell: Standard 11 µL 4. Sorbic acid
Temperature: 22 °C 5. Methyl 4-hydroxy benzoic acid
Injection volume: 2.0 µL 6. Methyl 4-hydroxy benzoic acid
7. Methyl 4-hydroxy benzoic acid
Mesopores:
Mesopores form a fine porous structure
with a large uniform surface area on which
adsorption takes place, thus enabling high-
performance chromatographic separation.
[130 Å; 150 Å; 300 Å]
Macropores:
The large Macropores allow rapid flow of the
mobile phase at very low back pressure with
maximum robustness and selectivity.
[1.15 µm; 1.5 µm, 2 µm]
It is stated that it is not allowed to change chromatographic *”The chapter does not say it is not
support – does that mean that it is not allowed to change allowed. Only indicates that the
particulate material to monolithic even if both materials
are made of pure silica?
support should be similar to the one
Is it possible to change from type A silica to Type B silica prescribed. All adjustments require
material? verification before implementation.”
MERCK
How above-described situation changes *”As soon as we have data showing column equivalency, we
if Ph method mentioning that column can add to the monograph that monolith columns can be used.
is packed with 5µm particles (and We need to have data demonstrating column equivalency
monoliths has no particles)? for the particular method.”
If Ph method is mentioning 150 cm *”The user needs to verify in a case-by-case approach if two
column, for whatever reason is it columns in series is equivalent to the column mentioned in
allowed to use 10 and 5 cm the monograph. If data is shared with USP we can add this
columns coupled together instead? information in the particular monograph”
What kind of test users must perform in order to implement *”The elution order must be the same
such change? and system suitability requirements
must be met.”
*Horacio N. Pappa, CQE, Ph. D., Senior Director – General Chapters, Global Science and Standards Division, Email: hp@usp.org
& Margareth R. C. Marques, M.Sc., Ph.D. Sr. Principal Scientist
Detection:
Cell:
UV, 230 nm
11 μL
“when column dimension does not change
Flow Rate: 1.0 mL/min [A] and 1.5 mL/min [B] ±50% change is permitted”
Mobile Phase: Water and methanol 40:60 (v/v)
Temperature: Ambient
Diluent: Acetonitrile/Water 1:1 (v/v)
Sample Solution: 0.5 mg/mL (500 ppm) of Spironolactone RS in diluent
Pressure Drop: 67 bar (972 psi) [A] and 110 bar (1595 psi) [B]
A B
250
150
Impurity A
200
120
Intensity (mV)
150
Intensity (mV)
90
Impurity A
100
60
50
30
0 0
0 5 10 15 20 0 2 4 6 8 10
Retention time (min) Retention time (min)
* no specific particle size is mentioned in the current monograph - thus any type of column backbone can be used without additional verification!
Chromatographic Conditions
Monolithic HighResolution RP-18e, FPP C18, 1.7 µm,
Column: Chromolith® HighResolution RP-18e 50-2 mm 50-2.1 mm
50-2 mm Pressure = 74 bar Pressure = 476 bar
FPP C18, 1.7 µm, 50-2.1 mm
Mobile phase: A: 100% acetonitrile
B: 20 mM Phosphate buffer, pH 4.5
Gradient: Time/min %A %B
0 40 60
0.1 40 60
2.0 60 40
Flow rate: 600 µL/min
Pressure: 74 / 476 bar
Detection: DAD 20Hz UV, 220 nm
Detector cell: LightPipe 10 mm
Temperature: 23 °C
Injection volume: 0.2 µL
Sample: Indoprofen 5.0 mg
Ketoprofen 5.9 mg
Fenoprofen 5.2 mg
Fluorbiprofen 3.4mg
Ibuprofen 6.1mg
dissolved in 100 mL A/B 50/50
The Stability test was performed to demonstrate the resistance to column clogging rep. matrix-tolerance and stability of a
monolithic silica column in comparison to a 1.7 µm fully porous particulate (FPP) column which is typically used in UHPLC
applications. Both columns have been treated the same way with frequent overloading to stress both columns to a maximum.
Both columns were tested with an RP-Test after stressing the columns 10 times. No pre-columns were used for this test.
Experimental Conditions for Stability test
®
Column_1: Chromolith HighResolution RP-18 endcapped 100 x 2 mm
Column_2: FPP Competitor column (hybrid silica C18 1.7 µm, 100 x 2.1 mm)
Flow Rate: 200 µL/min for Chromolith; 221 µL/min for FPP 1.7 µm column (due to laminar flow conditions (2.0 mm I.D. for monolithic / 2.1 mm I.D. for FPP)
Temperature: Ambient
Injection volume: 3 times 1 µL; 10 times 10 µL; 3 times 1 µL; 10 times 10 µL etc.
Detection: UV @254 nm
Sample preparation: Shake a quantity of the gel containing 50 mg of diclofenac sodium with 50 mL acetone for 10 minutes, filter and evaporate the filtrate to dryness under a gentle
nitrogen flow. Dissolve the residue in 100 mL of a mixture of 40 volumes of water and 60 volumes of methanol, dilute 1 volume of this solution to 10 volumes with
the mobile phase and filter through a glass fiber filter.
1400
1200
column backpressure/bar
1000
800
200
0
0 1000 2000 3000 4000 5000 6000 7000 8000
Injection volume/µL
Stability test data on Chromolith HighResolution RP-18 endcapped 100 - 2 mm I.D. column and FPP 1.7 µm Hybrid silica column from competitor 100 –
2.1 mm I.D. Data points with 1 µL injection were used for measurement.
The 1.7 µm FPP column clogged after 520 injection volumes.
The monolithic Chromolith® HighResolution column started to clog after 7000 injection volumes.
Chromatographic Data [Olive oil spiked with 1.25 µg/mL BPA] Experimental Conditions
®
Column Chromolith HighResolution RP-18e 100x2 mm
Olive oil spiked Detection Dionex Ultimate 3000 FLD; @ = Ex 275 nm Em 305 nm
2.E-01
Isocratic A/B 51/49 (v/v)
2
1.E-01 Flow Rate: 190 µL/min
1 Temperature: 30 °C
0.E+00
Injection 1 µL
volume:
-5.E-02
0 1 2 3 4 5 Dilue Ethanol
nt
Retention Time (minutes)
sample solution Weigh 2.0 g of olive oil into a 15 mL screw-cap test tube, add 1
spiked mL of standard solution, seal with screw cap and shake for 30
No. Compound Retention Time (min) S/N Area (Counts) Tailing Factor
(1.25 µg/mL sec. Add 5 mL of diluent seal with screw cap and shake for 1 min.
1 t0 void volume 0.8 BPA) Sonicate the solution for 15 min. at room temperature, centrifuge
at 5000 rev/min for 25 min. Take an aliquot of the supernatant
2 Bisphenol A 2.5 17.4 120337 1.3 and pass it through a 0.45 µm PTFE filter into a HPLC vial.
Too much
Poor Peak Shape Aromatic
Isomers retention Retention too short or inadequate Separation on RP-18
(Basic Compounds) compounds
on C18
Ascentis® Express
Chromolith®
C18e C8e
Diol Cyano Amino Si
/ (HR) / (HR)
Purospher®STAR
Discovery®
RP- F5 Diol Amino Si ZIC
C18 C8 Phenyl Cyano
Ascentis® Amide (PFP)
SeQuant®
Gradient: 5% A to 95% A
180
in15 min.
160
Inject. Vol.: 20 µL,
Reversed Phase C 18 tR = t0
cps
140
Flow: 0.6 mL/min
120
100 HILIC tR = 8.2 min
80
60
40
20
0
2 4 6 8 10 12 14 16 18 20 22 24
Time, min
Reversed phase
HILIC
51 ©2023 HPLC Impurity profiling - Petra Lewits, Merck KGaA, Darmstadt
Separation of polar compounds
Comparison of HILIC columns
k U – absolute retention
α (CH2) – degree of
hydrophobicity
α (OH) – degree of
hydrophilicity
α (2dG/3dG) – separation
factor for positional isomers
α (Tb/Tp) – acidic-basic
nature of the stationary
phase
Chromatographic Conditions
Column: SeQuant® ZIC®-cHILIC (3μm, 100Å) 150x4.6 mm
Injection: 5 μL
Detection: UV 218 nm
Cell: 8 μl
Flow Rate: 1.5 mL/min
Mobile Phase: Buffer: Dissolve 4.62 g of ammonium acetate in 1000 ml water (60 mM).
Adjust buffer to pH 5 using glacial acetic acid. Mix Acetonitrile and Buffer 90:10 (v/v)
Temperature: 30 °C
Diluent Mobile phase
Sample: 5000 ppm metformin & 1 ppm of each impurity: A, C and melamine &
5 ppm of impurity B in mobile phase
Pressure Drop: 105 bar (1522 psi)