Professional Documents
Culture Documents
IH-500_SM_H009231_EN_V2-3
IH-500_SM_H009231_EN_V2-3
IH-500_SM_H009231_EN_V2-3
Service Manual
Identification
Identification
IH-500 Service Manual
Version 2.3 - 12/2020
Document: H009231
n j
o 001500 / 001500RECOND
Manufacturer
a DiaMed GmbH
Pra Rond 23
1785 Cressier FR
Switzerland
Copyright ©
The reproduction, even partially, of this document is forbidden. No part may be copied in any form, and
it may not be used, edited nor transmitted by any electronic means (photocopy, photography, magnetic
supports or other recording processes), without the written authorization of Bio-Rad. All rights and
particularly reproduction, translation, edition, distribution and also industrial property and recording are
reserved.
Version History
2 | IH-500
Table of Contents
Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Chapter 1 Generalities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
1.1 Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
1.2 Warranty Limitation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
1.3 Glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
1.3.1 Persons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
1.3.2 Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
1.4 Typographical Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
1.4.1 Warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
1.4.2 Note. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Chapter 3 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
3.1 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
3.1.1 Dimensions and Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
3.1.2 Electrical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
3.1.3 Environmental Conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
3.1.4 Emissions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
3.2 General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
3.2.1 Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
3.2.2 Rear View. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
3.2.3 Internal View - Working Zone . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
3.2.4 Internal View - Lower Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
IH-500 | 3
3.3 Doors Opening, Closing and Locking (Doors open) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
3.4 Drawers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
3.4.1 Resources Identification (Numbering) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
3.4.2 Right Drawer Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
3.4.3 Left Drawer Configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36
3.5 Transport Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
3.5.1 Transport Arm (Axis Rotation). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
3.5.2 Complete Gripper. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
3.6 Vision Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
3.7 Gel Cards Internal Storage. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
3.8 Gel Cards Piercer / Ionizer Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
3.9 Gel Cards Preparation Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
3.10 Pipetting Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .44
3.10.1 Samples Loading Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
3.10.2 Diluents Area (Pipetting and Piercing) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .46
3.10.3 Gel Cards Pipetting Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .47
3.10.4 Reagents Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .48
3.10.5 Pipettor Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
3.10.5.1 Pipettor Module (Front View). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .49
3.10.5.2 Pipettor Module (Rear View) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
3.10.6 Washing Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
3.11 Liquid Containers Area and Fluidic System Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
3.11.1 Fluidic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
3.11.2 Fluidic System (Diagram) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
3.12 Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .55
3.13 Cooled Centrifuge Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .56
3.14 Solid Wastes Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .58
3.15 Main Frame Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
3.15.1 Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
3.15.2 Main Frame - Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .60
3.16 Samples and Resources Racks and Trays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
3.16.1 Sample Rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
3.16.1.1 Sample Requirements ABO/Rh Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
3.16.1.2 Antibody Screen Testing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61
3.16.1.3 Sample Block Barcode Label Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .62
3.16.1.4 STAT Sample Block Barcode Label Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
3.16.2 Gel Card Tray. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
3.16.3 Diluents Rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .66
3.16.4 Removable Reagents Tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67
3.17 Electronic Boards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
3.17.1 Right View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68
3.17.2 Sample Loading Area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
3.17.3 Pipettor Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69
3.17.4 Centrifuges. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
3.17.5 Vision Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
3.17.6 Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70
3.17.7 Complete Gripper. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
3.17.8 Reagent Module. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71
4 | IH-500
3.18 Electrical System Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .72
3.19 0595077 (M20325) IH-500 Gauge Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
3.20 IH-500 Optional Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .75
Chapter 4 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
4.1 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
4.1.1 General Safety Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
4.1.2 Personnel Qualification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
4.1.3 Documents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
4.1.3.1 Packing List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
4.1.3.2 Documents to fill in . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
4.1.4 Location. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
4.2 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
4.2.1 IH-500 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
4.2.1.1 Installation with Forklift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .83
4.2.1.2 Manual Installation (Lifting Kit) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
4.2.2 Optional Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
4.2.2.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
4.2.2.2 Installation with Forklift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .88
4.3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .89
4.3.1 Leveling of the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .90
4.3.2 Sample Racks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
4.3.3 Sample Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
4.3.4 Removable Reagents Tray . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .92
4.3.5 Diluent Piercing Tips. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
4.3.6 Gel Cards Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .93
4.3.7 Hand-held Barcode Scanner Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94
4.3.8 Keyboard Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .94
4.4 External Liquid Waste Optional Kit Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .95
4.5 Condensate Recovery Kit Installation (Optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
4.5.1 Optional Table with Right Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .97
4.5.2 Optional Table without Right Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .98
4.5.3 Workbench . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99
4.6 Installation with Optional Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
4.6.1 Leveling of the Optional Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .100
4.6.2 Instrument Installation on the Optional Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .101
4.6.3 UPS (Uninterruptible Power Supply) and Battery Pack Installation . . . . . . . . . . . . . . . . . . . . . .102
4.6.4 Optional Table Solid Waste Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .104
4.6.5 Printer Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .108
4.6.6 Workstation Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .109
4.7 Electrical Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .110
4.7.1 Fuses Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .110
4.7.2 Connection to the Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .110
4.8 Blocking Device Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
4.8.1 Pipettor X-Axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
4.8.2 Pipettor Y-Axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
4.8.3 Pipettor Z-Axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .112
4.8.4 Transport Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113
IH-500 | 5
4.9 Centrifuge Trap Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .114
4.10 Maintenance and Pipettor Doors Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .115
4.11 Starting Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .116
4.11.1 Starting up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .116
4.11.2 Software Pack Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117
4.11.3 Language Configuration (OS, Keyboard and Barcode Scanner) . . . . . . . . . . . . . . . . . . . . . . . .117
4.11.4 Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .119
4.11.5 Global Controller Menu Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .120
4.11.6 Hydraulic Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .121
4.11.7 Configuration of the Pediatric and Low Volume Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123
4.11.8 Final Check / QC (Autocheck) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .124
4.11.9 QC Client . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .125
6 | IH-500
5.6 Vision Station Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
5.6.1 Pre-Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
5.6.2 Camera Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .164
5.6.3 Camera Pre-Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .164
5.6.3.1 Camera Azimuth. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .164
5.6.3.2 White Balance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .165
5.6.3.3 Aperture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .166
5.6.3.4 Sharpness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .167
5.6.3.5 Exposure Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .168
5.6.4 Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .169
5.6.5 Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .171
5.7 Transport Arm Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .176
5.7.1 Pre Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .176
5.7.1.1 Gripper - Resources Types. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .177
5.7.1.2 Gripper Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .178
5.7.1.3 Gripper Localization Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .178
5.7.1.4 Gripper Burn-In. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .181
5.7.1.5 Note about Gripper Tolerance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .181
5.7.2 Mapping. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .183
5.7.3 Transfer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .185
5.7.4 Cycles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .188
5.7.4.1 Transfer Cycles - Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .188
5.7.4.2 Transfer Cycles - Reagents and Diluents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .190
5.7.4.3 Cards Piercing Cycles. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .191
5.7.4.4 Diluent Piercing Cycles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .193
5.7.4.5 Cycles Validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .194
5.7.5 Teaching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .195
5.7.5.1 Touch Probe Installation (Probe Mounting) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .196
5.7.5.2 Gripper Localization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .198
5.7.5.3 J1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .203
5.7.5.4 Reading Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .204
5.7.5.5 Front Centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .205
5.7.5.6 Rear Centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .207
5.7.5.7 Front Centrifuge Trap Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .208
5.7.5.8 Rear Centrifuge Trap Door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .209
5.7.5.9 Diluent Piercing Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .210
5.7.5.10 Right Drawer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .210
5.7.5.11 Left Drawer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .213
5.7.5.12 Internal Storage 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .215
5.7.5.13 Piercer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .216
5.7.5.14 Internal Storage 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217
5.7.5.15 Preparation Area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .217
5.7.5.16 Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .218
5.7.5.17 Pipetting Reagent Area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .220
5.7.5.18 Pipetting Cards Block Area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .222
5.7.5.19 Pipetting Diluent Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .223
5.7.5.20 Pipetting Sample Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .224
5.7.5.21 Diluent Swap Rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .224
5.7.5.22 Diluent Intermediate Rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .225
5.7.5.23 Washing Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .226
5.7.5.24 Common Reference Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .227
5.7.5.25 Resources Sensor Localization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .228
5.7.5.26 Gripper Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .230
5.7.5.27 Pipettor Teaching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .231
5.7.5.28 Pipettor Area Update . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .232
5.7.6 Secure Posing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .233
5.7.6.1 Cards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .233
5.7.6.2 Others . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .234
IH-500 | 7
5.8 Centrifuge Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .235
5.8.1 Pre Control – Front Centrifuge and Rear Centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .235
5.8.1.1 Front Centrifuge - Current Position Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .236
5.8.1.2 Rear Centrifuge - Current Position Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .238
5.8.1.3 Trap Door Open/Close Check (Auto-locking) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240
5.8.1.4 Centrifugation Burn-In. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240
5.8.1.5 Centrifugation Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .240
5.8.2 Check Temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .243
5.9 Pipettor Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .244
5.9.1 Pre Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .244
5.9.1.1 Current Position Check. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .245
5.9.1.2 Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .245
5.9.2 Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .247
5.9.2.1 Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .248
5.9.2.2 Card Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .250
5.9.2.3 Diluents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .252
5.9.2.4 Samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .254
5.9.2.5 Washing Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .255
5.9.2.6 Pipettor Position Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .256
5.9.3 Burn In . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .257
5.9.4 Block Adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .258
5.10 Hydraulic Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .262
5.10.1 Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .262
5.10.1.1 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .264
5.10.1.2 Hydraulic Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .266
5.10.2 Fluidic Circuits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .273
5.10.2.1 Intermediate Container Filling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .274
5.10.2.2 Weekly Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .274
5.10.2.3 Needle Change. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .275
5.10.2.4 Needle Washing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .275
5.10.2.5 Fluidic Circuit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .276
5.10.2.6 Decontamination Well . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .279
5.10.2.7 Draining Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .281
5.10.2.8 PDS Activation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .282
5.10.2.9 Fluidic Circuits Validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .282
5.10.3 Level Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .283
5.10.3.1 Level Detection Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .284
5.10.3.2 Level Detection in Reagent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .284
5.10.3.3 Level Detection in Sample Tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .286
5.10.3.4 Level Detection in Washing Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .287
5.10.3.5 Level Detection Validation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .288
5.10.4 Sampling / Dispensing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .289
5.10.4.1 Distribute Dilution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .290
5.10.4.2 Distribute Plasma . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .292
5.10.4.3 Distribute Reagent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .294
5.10.4.4 Distribute Diluent . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .295
5.10.4.5 Sample. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .297
5.10.4.6 Sample CRBC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .300
5.10.4.7 Distribute . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .301
5.10.4.8 Wash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .302
5.10.4.9 Sampling/Dispensing Validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .302
5.11 Incubator Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303
5.11.1 Incubator Temperature Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303
5.11.2 Check Temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .306
8 | IH-500
5.12 Reagent Plate Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .307
5.12.1 Pre Control. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .307
5.12.1.1 Check Shaking Axis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308
5.12.1.2 Current Position . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .308
5.12.1.3 Burn-In . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .309
5.12.2 Cooling Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .310
5.12.3 Check Temperatures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .312
5.13 Final Check / QC Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .313
5.13.1 Final Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .313
5.13.2 QC Cycles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .315
5.14 Tools Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .316
5.14.1 General Tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .316
5.14.1.1 Database Snapshot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .317
5.14.1.2 System Snapshot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .318
5.14.1.3 Parameters - Restore or Update Database . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .318
5.14.1.4 Transport Arm Adjustment Restore. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .320
5.14.1.5 Other Tools. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .321
5.14.1.6 User Management Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .323
5.14.1.7 User Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .323
5.14.2 Check Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .324
5.14.2.1 Instrument Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .324
5.14.2.2 Electronic Boards Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .325
5.14.2.3 Product Line. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .325
5.14.2.4 Software License Keys Activation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .325
5.14.2.5 Euresys License Activation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .326
5.14.2.6 Euresys License Recovery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .330
5.14.2.7 Cooling Reagent Module Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .335
5.14.3 Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .336
5.14.4 Pressure / Level Detection Curves . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .338
5.14.5 Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .340
5.14.6 IP Config . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .341
5.14.6.1 IP Address Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .341
5.14.6.2 How to restore Default IP Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .342
5.14.6.3 Transport Arm IP Address Recovery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .343
5.15 Exit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .344
IH-500 | 9
6.6 Specific Cleaning Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .352
6.6.1 Camera Lens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .352
6.6.2 Backlight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .353
6.6.3 Liquid Container Area and Washing Station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .353
6.6.4 Piercing Tip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .354
6.6.5 Needle Localization Hole . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .354
6.6.6 Centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .355
6.6.7 Gripper. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .356
6.6.8 Computer. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .358
6.6.9 Power Supply and Reagent Cooling Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .359
6.6.10 Transport Arm Controler. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .360
6.6.11 Transport Arm Cover Axis 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .361
6.6.12 Needle Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .361
6.6.13 Decontamination Liquid Container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .362
6.7 Visual Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .363
6.7.1 Gripper Legs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .363
6.7.2 Gel Card and Diluent Piercing Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .363
6.7.3 Centrifuge Rubber Seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .364
6.7.4 TU10, TU11, TU12 & TU13 Liquid Waste Tubing, Quick Connections . . . . . . . . . . . . . . . . . . .364
6.8 Functional Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .365
6.8.1 Gel Card and Diluent Piercing Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .365
6.8.2 Ionizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .365
6.8.3 Reagent Holders Springs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .366
6.8.4 Sample Tubes Holders Springs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .366
6.8.5 Tool Reference Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .367
6.8.6 Left and Right Drawers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .367
6.8.7 Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .368
6.9 Replacement Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .369
6.10 Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .369
6.10.1 Upper Right Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .369
6.10.2 Lower Right Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .370
6.10.3 Upper Left Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .371
6.10.4 Lower Left Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .372
6.10.5 Upper Rear Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .373
6.10.6 Lower Rear Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .374
10 | IH-500
Chapter 7 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 375
7.1 Alarms and Error Management . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .375
7.2 All Modules . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .376
7.3 Transport Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .379
7.4 Centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .384
7.5 Incubator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .387
7.6 Pipetor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .388
7.7 Access to the Pipetting Area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .390
7.8 Ionizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .391
7.9 Consumable Drawer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .391
7.10 Fluidic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .392
7.11 Internal Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .399
7.12 Reagent Cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .400
7.13 Sample Rack . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .401
7.14 Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .402
7.15 Solid Waste . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .405
7.16 Vision Station. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .406
IH-500 | 11
10.7 Electrical Connections (A00227i-EL Schematics) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .426
10.7.1 Automate Schematic (1/8) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .426
10.7.2 I/O Board (2/8) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .427
10.7.3 Pipettor (3/8) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .428
10.7.4 Fluidics (4/8). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .429
10.7.5 Centrifuges (5/8). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .430
10.7.6 Centrifuge Cooling (6/8) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .431
10.7.7 Integration (7/8) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .432
10.7.8 Controller & Power Supply (8/8) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .433
10.8 Electronic Boards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .434
10.8.1 E09982 - Main Power Supply Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .434
10.8.2 0595034 (E09972) - Fluidic Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .436
10.8.3 0595031 (E09978) - I/O Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .438
10.8.4 0595037 (E10904) - Pipetting Applicative Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .440
10.8.5 0590002 (E07722) - Novi-Core Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .441
10.8.6 0595036 (E09974) - PDS Applicative Board. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .442
10.8.7 (E10040) Pressure Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .443
10.8.8 0595029 (E09990) and 0590018 (E08818) Temperature Sensor Board
and 0595179 (E11248) Coated Temperature Sensor Box . . . . . . . . . . . . . . . . . . . . . . . . . . . .443
10.8.9 0595028 (E09980) - Gripper Applicative Board
0595042 (E10026) - Motor Control Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .444
10.8.10 0595035 (E09984) - Sample Rack Detection Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .445
10.8.11 0595030 (E09968) - Centrifuge Applicative Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .446
10.8.12 0595033 (E10928) - Cooling Driver Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .448
10.8.13 0595162 (E11246) - Level Detection Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .450
10.8.14 0595032 (E09986) - Pipetting Connective Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .450
10.9 Power Management (Fuses). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .451
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 453
12 | IH-500
1 Generalities
This chapter contains basic information on the instrument and the document structure.
The User Manager module is required to define the IH-500 users and the related appropriate access
rights. The IH-500 cannot be operated without registered users.
Refer to the appendix B User Manager Module of the IH-500 User Manual for more information.
IH-500 | 13
Generalities
k Any
warranty will, be deemed void if fault is found to have been caused by maltreatment,
misuse, unauthorized maintenance of service or negligence of regular maintenance and service,
accidental damage, incorrect storage or use of the products for operations outside their
specified limitations, outside their specifications, contrary to the instructions given in this manual.
1.3 Glossary
The following terms, among others, are used in this manual.
1.3.1 Persons
Manufacturer
The manufacturer of the IH-500 is:
DiaMed GmbH
Pra Rond 23
1785 Cressier FR
Switzerland
Operator
The operator is the owner of a IH-500 when using it as its owner and when transferring it to a third-
party.
Personnel
The personnel gathers persons who have any kind of activity with the IH-500 and who are qualified in
accordance with the manufacturer’s requirements and who are consequently authorized.
Technical Personnel
The term «technical personnel» designates the duly trained persons, who are permitted to perform
specific tasks on the IH-500.For instance, an electrician is designated as technical personnel for the
activities linked to wiring the IH-500 to the electrical network.
14 | IH-500
Glossary
Personnel Qualifications
The personnel qualifications define the minimum requirements which must be met by the authorized
personnel.
Minor Injury
A Reversible injury that does not require medical treatment.
Moderate Injury
A Reversible injury that does require medical treatment.
Major Injury
The damage that is irreversible, causes handicap or death.
1.3.2 Product
Instrument
This is the IH-500 distributed by the manufacturer.
Sample
Content of any sample tube.
b Only reagents authorized by Bio-Rad may be used in the instrument. Assure that any
operations performed by the user are in compliance with the instruction for use of the
reagents.
Diluent
A Consumable dilution solution to produce a suspension with RBCs from Samples.
Consumable (Resource)
All items placed inside the Instrument by a user except Sample tubes.
Barcode
A Barcode based data tag (an identification tag attached to a Consumable or Sample that can be
automatically read by the Instrument).
IH-500 | 15
Generalities
RBC
Red Blood Cell.
APF
Assay Protocol File.
System
The complete Fully Integrated System hardware (PC integrated), software, Consumables and
operating procedures.
DMS
Data Management Software.
STAT
SAMPLE with high priority to be integrated in the current workflow of SAMPLE processing.
1.4.1 Warning
b Toan designate:
imminent and dangerous situation which, if not avoided, may lead to major injury of the
user; or a potentially dangerous situation which, if not avoided, may lead to:
• moderate injury of user or in the tests being false;
• injury of user or in important delay in delivery of the test results.
1.4.2 Note
k Indicates:
• a preferred procedure or recommended use;
• a general or informative remark.
16 | IH-500
2 Safety and Handling
This chapter sets out safety instructions to ensure safe and trouble-free operation of the IH-500 and its
associated software. It also describes the handling and storage conditions.
2.1 Introduction
2.1.1 Principles
Before carrying out any operation on the IH-500, it is imperative to read this chapter and fully
understand it.
In case of any doubt, consult your Instrument support.
IH-500 | 17
Safety and Handling
b The IH-500 may not be exposed to direct sunlight, heat, dust or excessive humidity (use
The IH-500 must be kept away from potential sources of interference.
The use of the instrument in a dry environment especially if synthetic materials are
present (clothes, synthetic carpets, etc.) may cause electrostatic discharges and lead to
false results.
k Assess
the electromagnetic environment of the site in which the IH-500 is located before
switching it on.
k The
recommended maximum length for the power cable is 2.5 meters.
The use of accessories, transducers and cables other than those specified, with the exception of
transducers and cables sold by Bio-Rad as replacement parts for internal components, may
result in increased emissions or decreased immunity of IH-500.
IH-500 should not be used adjacent to or stacked with other equipment. If adjacent or stacked
use is necessary, IH-500 should be monitored to verify normal operation in the configuration in
which it will be used.
18 | IH-500
Environmental Conditions
IH-500 is intended for use in the electromagnetic environment specified below. The customer or the
user of IH-500 should ensure that it is used in such an environment.
Electromagnetic environment -
Immunity test IEC 60601 test level Compliance level guidance
Electrostatic ±6 kV contact ±6 kV contact Floors should be wood, concrete or
discharge (ESD) ceramic tile. If floors are covered with
IEC 61000-4-2 synthetic material, the relative humidity
±8 kV air ±8 kV air
should be at least 30%.
Electrical fast ±2 kV for power supply lines ±2 kV for power Mains power quality should be that of a
transient burst ±1 kV for lines no input/output supply lines typical commercial or hospital
IEC 61000-4-4 ±1 kV for lines no environment.
input/output
Shock waves ±1 kV line to line ±1 kV line to line Mains power quality should be that of a
IEC 61000-4-5 ±2 kV line to earth ±2 kV line to earth typical commercial or hospital
environment.
Voltage dips and <5% UT <5% UT Mains power quality should be that of a
outages (>95% dip in UT) for 0.5 cycle (>95% dip in UT) for 0.5 cycle typical commercial or hospital
IEC 61000-4-11 environment. If the user of IH-500
40% UT 40% UT
requires continued operation during
(60% dip in UT) for 5 cycles (60% dip in UT) for 5 cycles
power mains interruptions, it is
70% UT 70% UT recommended that IH-500 be powered
(30% dip in UT) for 25 cycles (30% dip in UT) for 25 cycles from an uninterruptible power supply or
<5% UT <5% UT a battery.
(>95% dip in UT) for 5 sec (>95% dip in UT) for 5 sec
Power frequency 3 A/m 3 A/m Power frequency magnetic fields
(50/60 Hz) should be at levels characteristic of a
magnetic field typical location in a typical commercial
IEC 61000-4-8 or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
IH-500 | 19
Safety and Handling
IH-500 is intended for use in the electromagnetic environment specified below. The customer or the
user of IH-500 should ensure that it is used in such an environment.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured
field strength in the location in which the IH-500 is used exceeds the applicable RF compliance level above, the IH-500 should be
observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the IH-500.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
20 | IH-500
Environmental Conditions
The IH-500 is intended for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the IH-500 can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and the IH-500 as recommended below, according to the maximum output power of the
communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can
be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of
the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
b The
IH-500 may only be used indoors and never in the immediate vicinity of patients.
The electrical safety of the IH-500 is guaranteed only if the electrical installation of the
building complies with national regulations applicable to buildings for medical purposes
or laboratories and if this installation is functioning perfectly.
The IH-500 may not be used in premises subject to risks of fire or explosion.
b As
for any other electrical device, the IH-500 may not be located near a water tap.
Do not allow any liquid to penetrate inside the housing of any instrument. In case of major
spillages of liquid inside any instrument, proceed immediately as follows:
A. Switch off the IH-500 at the power switch.
B. Disconnect the plug.
C. Dry the instrument.
D. Clean and decontaminate the instrument.
E. Check all electrical functions.
IH-500 | 21
Safety and Handling
b The IH-500 is not flame resistant. Sparks may cause a fire or an explosion.
FLAME RESISTANCE
b The IH-500 is not protected against water penetrations. Water penetrating the instrument
CONTACT WITH WATER
The sole possession of the service manual does not allow the personnel to perform any
kind of repair on the IH-500.
Take into account all the warnings and follow all instructions provided on the IH-500 and
in the documentation.
The IH-500 may only be connected to the electrical power sources specified.
It is imperative to use the liquids specified in the manual to clean the IH-500. If you intend
using a different liquid, do so only after consulting the manufacturer and receiving his
approval.
The use of materials other than those defined in the Service Manual (unauthorized
hazardous substances such as inflammable, explosive or chemically unstable
substances) is forbidden. Any infringement of this rule will be considered by the
manufacturer as guilty negligence.
Never attempt to use spare parts other than those supplied by the manufacturer.
b Reagents,
Gel cards, samples, liquid and solid wastes should be considered potentially
infectious. Before performing any operation on IH-500 the instrument must be
decontaminated. Local safety regulations and good laboratory practice must be followed.
It is imperative to wear protective gloves when decontaminating the instrument and
working with potentially infectious materials.
For any other materials referred to in this manual (e.g NaOH, Wash Solution A, Microcide
SQ™ or Micro 10®+) corresponding safety data sheets should be consulted before the
first use, adequate protective measures should be applied and any applicable regulations
should be followed.
Always disconnect the power cable before accessing internal parts of the instrument. A
contact with live conductors may cause an electrocution.
22 | IH-500
General Safety Instructions and Signs
Always scan USB devices (USB key, camera...) with adapted anti-virus software prior to
connecting to the instrument.
Make sure that the LIS is equipped with the necessary protection to ensure no virus nor
access is transmitted to the IH-500.
Never modify or suppress any file from any PC. Files are maintained up to date by the
tools provided to the FSE.
Never configure any OS automatic update.
Always perform a backup before any intervention on the instrument.
2.3.2 Signs
The following signs can be found on the IH-500. Associated risks are explained in the table.
IH-500 | 23
Safety and Handling
Do not push Pushing on the side panels deforms the main frame of the instrument.
Keep dry
Keep upright
Hand trucks must not be used on this side to handle the package.
Serial Number
SN
24 | IH-500
3 Description
IH-500 | 25
Description
Commissioning (Crates)
Description Instrument Optional table Battery pack UPS (220 V) UPS (110 V)
Width 1450 mm 1450 mm 570 mm 580 mm 580 mm
(57.09 in) (57.09 in) (22.44 in) (22.83 in) (22.83 in)
Height 1230 mm 388 mm (15.28 in) 225 mm 225 mm 220 mm
(48.43 in) (08.86 in) (8.86 in) (8.66 in)
Depth 1080 mm 1075 mm (42.32 680 mm 825 mm 880 mm
(42.52 in) in) (26.77 in) (32.48 in) (34.65 in)
Weight 326.5 kg 130 kg 36 kg 48 kg 48 kg
(720 Ib) (286 Ib) (79.4 Ib) (105.8 Ib) (105.8 Ib)
Weight (with UPS and - 200 kg - - -
battery pack) (440 Ib)
26 | IH-500
Technical Data
Figure 1.
3.1.4 Emissions
Heat emission 1000 W max.
Noise 65 dB (A) max
Pollution level 2
IH-500 | 27
Description
10
1
8
7
2
6
4
5
3
Figure 2.
28 | IH-500
General Description
12 1
11
3
10
9 4
7
8
5
6
Figure 3.
IH-500 | 29
Description
30 | IH-500
General Description
16
15 20
17
18
1
14
19
2
13
3
12 4
5
7
10 8
11 6
Figure 4.
IH-500 | 31
Description
9
6
5 10
7
8
2
3
1
Figure 5.
1 Liquid containers
2 Solid waste
3 Cleaning liquid container (storage area)
4 Front cooled centrifuge (12 gel cards)
5 Rear cooled centrifuge (12 gel cards)
6 Centrifuge cover
7 Centrifuge trapdoor (open)
8 Computer
9 Reagent cooling module
If there is an excessive noise or not enough cold produced, this may be caused by a lack of
cooling liquid in the compressor. Refer to the notification IH-500 NI Filling Cooling v2.1 for more
information.
10 Power supply and transport arm controller
32 | IH-500
Doors Opening, Closing and Locking (Doors open)
5
6
2
3
1
1
4 4
3
Figure 6.
IH-500 | 33
Description
3.4 Drawers
Two drawers on the front of the IH-500 allow the user to load and unload the resources. The drawers
are locked by a magnetic latch and actuated manually. They can not be inverted.
The electromagnetic latch is managed by the I/0 board.
The embedded sensor of the transport arm detects the type, number and position of the resources
loaded in the drawer.
5
6 8
7
4
3
1
Figure 7.
1 Right drawer
2 Left drawer
3 Drawers frame
4 Warning label to inform the user about correct & wrong gel card tray positioning
5 Drawer slide
6 Full solid waste sensor
7 Electromagnetic latch
8 Drawer ejector
34 | IH-500
Drawers
1
13 1 2 3
13 24
14 1 2 3 4
4 5 6
15
16
7 8 9
17 36
25
10 11 12
12
25
13
1 12
14
15
16
17 13 24
36
Figure 8.
IH-500 | 35
Description
1 x 12 2 17
1 x 12 2 11
2 x 12 2 5
2 x 12 0 11
3 x 12 0 5
36 | IH-500
Transport Arm
X Y
Z axis orientation
2
3
1
Figure 9.
1 Transport arm
2 Complete gripper
3 Working zone (outlined red)
IH-500 | 37
Description
J3
J4
J5
J6
J2
J1
Figure 10.
38 | IH-500
Transport Arm
Figure 11.
1 Gripper legs
2 Gripper motor
3 Mapping sensor
4 Gripper applicative board
5 Motor control board
IH-500 | 39
Description
5
4
7
6
3
2
1
Figure 12.
1 Support
2 Camera assembly with lens
3 Internal temperature sensor
4 Protective cover
5 Field of camera vision
6 Samples barcode reader
7 Back-light
40 | IH-500
Gel Cards Internal Storage
3
6 5
2 7
1
Figure 13.
IH-500 | 41
Description
Figure 14.
42 | IH-500
Gel Cards Preparation Area
Figure 15.
IH-500 | 43
Description
Z
X Axis orientation
Y
6
7
2
Figure 16.
44 | IH-500
Pipetting Area
8 9
7 4
6
3
5
2 1
Figure 17.
IH-500 | 45
Description
6
5
4
1
3
Figure 18.
46 | IH-500
Pipetting Area
Figure 19.
IH-500 | 47
Description
5
4 6 6
Figure 20.
1 Gearings box
2 Cooling plate
3 Cooling module input
4 Cooling module output
5 Reagent mixing motor
6 Coated temperature sensor box
7 Reagent holder and reagent holder spring
48 | IH-500
Pipetting Area
5 9
6
7
10
11
2
3 1
Figure 21.
IH-500 | 49
Description
10
11
8
12
7 1
13
6
3
5
Figure 22.
50 | IH-500
Pipetting Area
Figure 23.
IH-500 | 51
Description
1
1
2
5 4
2
3
Figure 24.
52 | IH-500
Liquid Containers Area and Fluidic System Diagram
7
7
7
5 5
6 6
5
5
2
2
1
1
4
4
1
1
1
Figure 25.
IH-500 | 53
Description
Figure 26.
54 | IH-500
Incubator
3.12 Incubator
The incubator module allows to incubate the gel cards at 37°C (98.6°F).
It includes:
• an incubation area with 28 locations for gel cards;
• a cabled heating base;
• two sensors which allow temperature regulation.
Figure 27.
IH-500 | 55
Description
5
4
2
3
Figure 28.
56 | IH-500
Cooled Centrifuge Module
Rotation direction:
• Normal run - positive: counter-clockwise;
• Negative: clockwise.
Figure 29.
IH-500 | 57
Description
8 9
10
10
6 1
2
4 3
Figure 30.
58 | IH-500
Main Frame Assembly
5 4
1
2
6 3
Figure 31.
IH-500 | 59
Description
8
3
7
4 1
6
2
Figure 32.
60 | IH-500
Samples and Resources Racks and Trays
4
2
1 6
1
1
1
Figure 33.
IH-500 | 61
Description
11 22 33 44 55 6 6
77
Figure 34.
62 | IH-500
Samples and Resources Racks and Trays
IH-500 | 63
Description
5 4
2 1
Figure 35. Example using a sample tube holder with flat bottom tubes (T04)
1 P4T04
2 P3T04
3 Tube type orange sticker
STAT (emergency) sample is identified with an S, see section STAT Sample Block Barcode
Label Set on page 65.
4 P2T04
5 P1T04
Never mix barcodes of several sticker sets on the same sample block.
64 | IH-500
Samples and Resources Racks and Trays
Figure 36.
IH-500 | 65
Description
Figure 37.
Figure 38.
4 ID-Diluent rack
5 Additional positioning support
6 Identification sticker
Diluents racks can only be inserted in the left drawer.
k Only
use racks with positioning support (5):
• ID-Diluent 1 (brown plastic);
• ID-Diluent 2 (translucent plastic);
• ID-Titration Rack (translucent plastic).
66 | IH-500
Samples and Resources Racks and Trays
Figure 39.
1 Handle
2 Reagent vial numbering
3 Removable reagents tray
The removable reagents tray can only be inserted in the left drawer.
IH-500 | 67
Description
4
1
Figure 40.
1 Fluidic board
2 White light board (fixed on the top panel)
3 Novi-Core board
4 Pipetting applicative board
5 I/O board
68 | IH-500
Electronic Boards
Figure 41.
4
3
Figure 42.
IH-500 | 69
Description
3.17.4 Centrifuges
1 2
Figure 43.
Figure 44.
3.17.6 Incubator
Figure 45.
70 | IH-500
Electronic Boards
1
2
Figure 46.
3 3
Figure 47.
IH-500 | 71
Description
Figure 48.
72 | IH-500
0595077 (M20325) IH-500 Gauge Kit
6
4 8
5 7
1 2 19 9
11
10
12
18
13
22 20
14
25
21
24 15
17
23 16
IH-500 | 73
Description
The following components are the same as for IH-1000 and can be ordered separately. It is possible to
store them in the kit with the probe 15.
74 | IH-500
IH-500 Optional Table
4
10
2
6
Figure 50.
1 Workstation (left)
2 External solid waste bin
3 Solid waste input
4 Storage area (paper, gel cards)
5 Workstation (right)
6 Printer storage area
7 UPS (Uninterruptible Power Supply)
8 Battery pack
9 Adjustable foot
10 Storage area (sample rack and gel cards)
IH-500 | 75
Description
76 | IH-500
4 Commissioning
This chapter describes the unpacking procedure of the IH-500, its installation, its connections and
installation with optional table. Moreover, this chapter explains all the remaining procedures required to
put the instrument into operation.
4.1 Safety
4.1.1 General Safety Requirements
b The electric installation must comply with the national regulations applicable to buildings
The instrument may only be used indoors and never close to patients.
IH-500 | 77
Commissioning
There is a risk:
• to damage the frame of the instrument;
• to drop the instrument on the floor.
Use the lifting handles or push the optional table to move the instrument.
Figure 51.
4.1.3 Documents
Carefully study all the relevant setup procedures before starting to set up the IH-500.
78 | IH-500
Unpacking
Installation Protocol
To be filled and sent by the technician installing the instrument at the customer’s within 3 weeks.
The official electronic document can be found on the Bio-Rad website:
https://www.bio-radsecretarea.com under Level: I Technical Area - Official forms.
4.1.4 Location
In order to optimize the manipulations and to guarantee an incident-free operation, a location having
the specifications specified in Environmental Conditions on page 27 must be chosen.
4.2 Unpacking
The IH-500 is carefully packed for transport. In order to unpack the instrument correctly, make sure that
the arrows on the crate are oriented upwards.
The elements in the crate are protected but must be handled with care.
k Before
unpacking the IH-500, prepare the lab table or the optional table on which will come the
instrument.
If the option table is available, unpack and install it first.
See chapter Optional Table on page 86.
Store all packagings panels, clips, screws and foams in a storage room for later use (e.g
transport).
b On
external casing, in order to protect the instrument against the shocks and tilt during
the transport, a detector of shock and tilt, and labels of location on packing are added
(spies).
Before anything, check these detectors at reception.
If the equipment shows any sign of damage or any part missing, do not attempt to set it
up, but contact Bio-Rad or your local dealer immediately.
IH-500 | 79
Commissioning
4.2.1 IH-500
A Place the crate on the floor in order to have enough room to open all panels.
C Remove the 8 clips (2) which maintain the upper panel and the side panels.
With a large flat screwdriver.
k Keep these clips in a safe place as they will be used to store the crate after installation.
D Remove the upper panel (1).
F Remove the 6 clips (6) which maintain the front panel marked “FRONT”.
With a large flat screwdriver.
L Remove the lower clips safety screws located on the rear (9) and left (3) panels.
M Remove the 4 clips which maintain the rear panel (9) (marked “REAR”) on the
base.
With a large flat screwdriver
N Remove the 2 clips which maintain the left and the rear panel together.
With a large flat screwdriver.
80 | IH-500
Unpacking
2
2
5 6
6
Figure 53.
IH-500 | 81
Commissioning
3
2
Figure 54.
82 | IH-500
Unpacking
Figure 55.
IH-500 | 83
Commissioning
company following all applicable local regulations and applying all required safety
measures.
k workstation(s).
To lift the instrument on the IH-500 Optional table, ensure to first remove the optional table
B Insert the lifting handle (1) under the instrument in the appropriate locations.
Above the Teflon sliders (2).
b Never use the lifting kit on the instrument without the right and left lower panels
mounted.
D Repeat steps B to C to fix the lifting handle on the left side of the instrument.
G Place the rear central foot and the 2 front feet of the instrument on an appropriate
laboratory table.
84 | IH-500
Unpacking
2 3
3
4x
Figure 56.
IH-500 | 85
Commissioning
4.2.2.1 Unpacking
The optional table with battery and UPS weighs: 170 kg (375 Ib).
A Place the crate on the floor in order to have enough room to remove the elements.
k Do not use cutter to open the box because workstations are just below.
D Remove the 4 lateral foams (4) and the central foam (5).
k The
optional table can be ordered with a second workstation and with UPS and its battery
(inside the table).
F Rise the workstation (11), turn it around and replace it in its location as indicated.
k Ifthetheworkstation
instrument is lifted manually on the option table with the lifting kit, it is mandatory to install
after.
See chapter Installation with Optional Table on page 100.
H Perform the same steps for the second workstation (10, if previously ordered).
k is used instead.
Unpacking figures illustrate packaging with UPS box. If no UPS box was ordered a raising block
J Remove the 8 corner clips (9) and the UPS box (7).
86 | IH-500
Unpacking
4 5
4 6
8
9
10
11
13
12
Figure 57.
IH-500 | 87
Commissioning
B Pull the optional table out of the pallet and move it to the desired location.
Figure 58.
88 | IH-500
Installation
4.3 Installation
If accessories show any sign of damage or if one of them is missing, contact Bio-Rad or your
local dealer.
IH-500 | 89
Commissioning
b push on the top of the instrument and check the rear feet.
Perform this procedure every time the instrument is moved to another location. Do not
k If the procedure is not followed, there is a risk of distortion for the chassis of the instrument.
A Check that the rear central foot (7) is in place.
D With the proper wrenches, adjust alternatively the front feet (2) in order to get a
distance of 65 mm between the instrument bottom and the table or the optional
table.
To level the instrument on the optional table, see section Instrument Installation
on the Optional Table on page 101.
To level the instrument, loosen the nut (5) and adjust the instrument with feet
nuts (6).
Tighten the nut (5) to block the adjustments.
F Place the leveling tool (4) on the Y-axis pipettor rail guide (3).
L When the level is OK, bring the 2 rear feet (1) in contact with the table or the
optional table (with the hand and not with a tool).
The rear central foot must always support the instrument.
90 | IH-500
Installation
7
1
5 8
65 mm
6 9
2
Figure 59.
IH-500 | 91
Commissioning
B Remove the adhesive tape used to maintain the 5 sample racks (3).
B Remove the adhesive tape used to maintain the removable reagents tray (1).
Figure 60.
92 | IH-500
Installation
C Remove the diluent piercing tips from their bubble bag in the solid waste bin.
D Place the 8 gel card blocks (3) in the preparation area for gel cards.
1
2
Figure 61.
IH-500 | 93
Commissioning
D Press strongly the scanner holder (2) for a few seconds on the marked area (1).
F Connect the hand-held barcode scanner (3) to the USB port (4) and place it on its
holder (2).
C Place the Velcro loop strip under the keyboard (5) and the hook strip on the waste
bin area door (7) to fix the keyboard.
3
2
4 6
7
Figure 62.
94 | IH-500
External Liquid Waste Optional Kit Installation
Figure 63.
Figure 64.
Figure 65.
IH-500 | 95
Commissioning
Figure 67.
k During starting up of the instrument, check for leakage around the modified parts.
O Enable the option External liquid waste (GENERAL TOOLS screen, User
options area).
See chapter User Options on page 323.
96 | IH-500
Condensate Recovery Kit Installation (Optional)
3
Figure 70.
IH-500 | 97
Commissioning
B Replace the right handle of the panel by the pierced handle (3).
Included in the kit.
C Extract the pipettor drainage pipe (1) from behind the pipetting applicative board.
D Slide the pipe (1) through the handle of the panel (3).
G Stick the loop part of the Velcro scratch (4) on the container.
Figure 71.
98 | IH-500
Condensate Recovery Kit Installation (Optional)
4.5.3 Workbench
The installation on a workbench is similar as an optional table without right workstation. See chapter
Optional Table without Right Workstation on page 98.
The use of Velcro scratch is optional.
Figure 72.
IH-500 | 99
Commissioning
Excessive heating may cause malfunctions and/or failure. A space of 300 mm (11.8in) on
the left and right sides of the instrument/optional table must be left clear of any obstacle
(50 mm (1.97in) spacers are installed by default on the rear of the instrument).
k the wheels.
Do not act on the feet at a height above 100 mm. The optional table must not be set down on
1
1
Max. 3.94 in
4 (100 mm)
1
Figure 73.
100 | IH-500
Installation with Optional Table
Figure 74.
IH-500 | 101
Commissioning
Figure 75.
102 | IH-500
Installation with Optional Table
b OFF.
Make sure that the power switches of the instrument, the UPS and the battery pack are
Figure 76.
IH-500 | 103
Commissioning
Figure 77.
Figure 78.
Figure 79.
Figure 80.
104 | IH-500
Installation with Optional Table
Figure 81.
Figure 82.
k The extension cables E11032 and E11031 are pre-assembled on the optional table.
M Connect the optional table solid
waste sensor cable E11032 (8)
9
to the extension cable
connector (9).
IH-500 solid waste sensor 8
cable E10194 (7).
7
b
Do not mix up «full solid waste sensor
extension cable E11032» and
«Electromagnetic latch extension cable
E11031».
Figure 83.
IH-500 | 105
Commissioning
Figure 85.
11
Figure 86.
Figure 87.
106 | IH-500
Installation with Optional Table
Figure 88.
V Mount the higher plate and the rear bin plate by proceeding reversely.
Steps I on page 104 to G on page 104.
Figure 89.
AE Select External solid waste (GENERAL TOOLS screen, User options area)
to activate it.
See User Options on page 323 for more information.
IH-500 | 107
Commissioning
AF Check with Input/output menu that the optional table solid waste opening and
closing and waste detection state are functional.
See Solid Waste Open/Close Check on page 157 for more information.
B Connect the printer to the USB connector (2) at the rear of the instrument.
C Plug the power cable supplied with the printed to the mains electrical network.
Figure 91.
108 | IH-500
Installation with Optional Table
3 3 1
Figure 92.
IH-500 | 109
Commissioning
b Disconnect the instrument from the electrical network before changing the fuses.
ELECTROCUTION
The instrument is delivered by default with 2 x T8A-230Vac fuses. Those fuses are
adapted to countries using a “230 VAC” electrical network.
For countries with a “110 VAC” electrical network, it is necessary to replace the mounted
F1 and F2 T8A fuses of the power supply by the 2 x T20A fuses delivered with the
instrument.
B Open the 2 fuse holders F1 (3) and F2 (4) with a flat screwdriver.
On the rear of the instrument.
b The instrument is delivered by default with a power cable for countries using a “230 VAC”
electrical network.
The instrument must be ordered with the adapted power cable of the country.
For countries with a “110 VAC” electrical network, it will be necessary to use a 110 V
adapter or a power cable in accordance with the standards in the country of use. The
minimum capacity for the cable is 20 A at 110 VAC.
k The connector type used for the power cable must be a C19 IEC power socket.
The IEC power cable mounted on the instrument is a C20 connector.
A Check that the power cable is equipped with the ferrite (5) as shown on the
picture.
B Make sure that the main power switch is OFF «O» (2).
C Connect the instrument to the electrical network with the power cable (1).
110 | IH-500
Electrical Connection
5 3
Figure 94.
IH-500 | 111
Commissioning
Z-Axis
Y-Axis
3
2
4
1 X-Axis
Figure 95.
112 | IH-500
Blocking Device Removal
A Loosen the 2 screws (3) on the right side of the blocking device from the main part
(4).
B Loosen the 2 screws (5) of the right side of the blocking device from the frame.
D Loosen the screw (2) on the main part of the blocking device
E Remove the main part (1) of the blocking device from the gripper.
3
3
5
5
Figure 96.
IH-500 | 113
Commissioning
If necessary, adjust the orientation of the traps. The cleat must be fully inserted in the
housing. Keep the flap straight, never at an angle.
Figure 97.
α = 0° α ≠ 0°
Figure 98.
114 | IH-500
Maintenance and Pipettor Doors Adjustment
4
1 5-
8
mm
2 3
3
Figure 99.
IH-500 | 115
Commissioning
C Press the instrument power button (2) until an audible sound is produced.
The Service Software starts automatically. The login screen is displayed.
D Login.
• Username: fsefis.
• Password: fsefis, followed by addition of the day, the month and the year.
Example:
For 3rd December 2020 = 2020+12+3 = 2035.
The correct password is fsefis2035.
E Confirm.
OK
The warning message “Perform decontamination with cleaning liquid
before replacing or adjusting parts inside the instrument.” is
displayed.
b the instrument.
Perform decontamination with cleaning liquid before replacing or adjusting parts inside
Figure 100.
116 | IH-500
Starting Procedure
Figure 101.
Figure 102.
C Confirm.
OK
k Resources
only need to be removed in case of a software update, as this will delete the loaded
resources information.
IH-500 | 117
Commissioning
1
Figure 103.
3
4
Figure 104.
Figure 105.
L Check that the required type of keyboard is configured in the field Installed
services (6).
If not, select on Add, choose the correct keyboard and select OK.
O Shut down and restart the instrument to access the Service software main screen.
118 | IH-500
Starting Procedure
4.11.4 Backup
A Select the TOOLS button from the main screen.
See Database Snapshot on page 317 for more information.
B Select the date of the day. Type the number of previous days to be backed up.
Figure 106.
Figure 107.
IH-500 | 119
Commissioning
Figure 108.
F After global initialization, check that all the modules are functional (green).
Figure 109.
120 | IH-500
Starting Procedure
k In case of spillage or contact, refer to the Wash Solution A Concentrate Instructions for Use.
k Do not activate the pumps and the solenoid valves for too long (as there is no liquid in the
system).
A Inspect the fluidic tubing and the fluidic connections to ensure integrity.
See chapter Fluidic Connection (1/2) on page 424.
C Select UNITS.
Perform a quick check of the main hydraulic I/O to verify if the pumps and
solenoid valves are activated correctly.
See Units on page 262 for more information.
D Check that each container can easily be connected / disconnected from the
instrument.
Figure 110.
IH-500 | 121
Commissioning
Figure 111.
1 2
Figure 112.
K Select on Finish.
Finish
When the flushing process is finished.
L Perform 2 cycles by pressing 2x START PRIME in the Fluidic circuit area (2).
START PRIME
k Priming is only available when the automatic intermediate tank filling is activated.
M When the priming is finished, check:
• the absence of air bubbles in the visible tubing of the pipettor,
• and check that the aspiration of the washing pot has correctly functioned (pot
and well completely empty).
122 | IH-500
Starting Procedure
O Check that the decontamination well is full when the priming process is
completed.
Figure 113.
IH-500 | 123
Commissioning
D Select START.
START See Final Check / QC Menu on page 313 for more information.
Figure 114.
G Check that the procedure is progressing and ends without error messages.
H When the Final Check is completed, verify that all the elements are checked (green
check ), meaning they are functional, and no element is aborted (red cross ).
k Ifpage
one element has a red cross control the element, refer to section Final Check / QC Menu on
313 for more information.
124 | IH-500
Starting Procedure
M SWITCH TO
Select SWITCH TO EXPLOITATION, to finish the starting procedure.
EXPLOITATION
The instrument stops, and after restarting, the instrument automatically starts
in exploitation mode (GUI is launched).
4.11.9 QC Client
A Perform a QC according to local procedures with customer.
IH-500 | 125
Commissioning
126 | IH-500
5 Checks and Adjustments
The present chapter describes procedures to be carried out for particular calibration and controls after
parts replacement or in case of trouble. It contains procedures aimed at checking the functions of the
instrument. They do not require any hardware replacement.
All these checks and adjustments are performed through the dedicated Service software.
b Repair
and service operations must only be performed by a qualified service engineer
appointed by the manufacturer, in accordance with Good Laboratory Practices.
b Before
any technical intervention on the instrument, verify that the instrument is correctly
decontaminated (refer to the decontamination certificate).
If necessary perform a complete decontamination.
See chapter Cleaning and Decontamination Instructions on page 348.
k itself.
The procedures describing hardware replacement are given separately with the replacement part
IH-500 | 127
Checks and Adjustments
b Decontaminate the instrument with cleaning liquid before replacing or adjusting parts.
A Start the IH-500 Software (GUI) from the FIS Launcher.
If not already started.
128 | IH-500
Service Software General Description
Figure 116.
D Validate.
E Select MENU.
k The button STOP ALL will stop all running processes. All running tests will be canceled.
G Select SERVICE SOFTWARE.
The following warning screen is displayed.
Figure 117.
IH-500 | 129
Checks and Adjustments
1 2 3 4 5
7 8 9 10
15
11 12 13 14
17 18 19 20
16
130 | IH-500
Service Software General Description
IH-500 | 131
Checks and Adjustments
Check
Input / Output Input /Output
Burn-in
Check temperature
Block adjustment
Vision station Pre adjustement
Hydraulic Units
Adjustement
Fuidic circuits
Check
Level detection
Transport arm Pre control
Sampling/Dispensing
Mapping
Incubator Adjustement
Transfer
Check temperature
Cycles
Adjustement
Secure posing
Check temperatures
Centrifuge Pre control Front centrifuge
QC Cycles
Check temperatures
IP Config
Open pipetting door
Fluidic curves
Modules
k menu.
The third level is displayed on the main screen header (contextual menu) according to selected
132 | IH-500
Service Software General Description
1 2 3
Figure 120.
5.1.3.2 Buttons
Buttons toggle between available actions.
Figure 121.
4 5 6 5
Figure 122.
4 Active button
5 State information
6 Inactive button (grayed)
Figure 123.
IH-500 | 133
Checks and Adjustments
Figure 124.
To move axis, select corresponding (-Y / +Y / -X / +X) buttons or slide the round cursor.
XY speed (mm/s) displays the maximum speed when using the slide cursor
XY step (mm) displays the movement done at each button selection
To change values, select the current value (a new window appears) and enter the new value.
k Itscreen.
is also possible to use the keyboard arrow keys (left/right/up/down) instead of the touch
5.1.3.5 Validation
Some adjustments need to be validated with the VALIDATE button. If a validation is not OK, use
CANCEL VALIDATION button.
Select the VALIDATE button to change the specified adjustment status of the module.
Modules current states are available in the STATUS screen, from the TOOLS menu. See chapter
Status on page 336.
Figure 125.
k ToInvalidated
ensure proper working of the instrument, all validation buttons need to be OK.
adjustment status of one module can involve several other status changes. Software
manages statuses relationships.
134 | IH-500
Service Software General Description
Figure 127.
5.1.3.7 Levels
Some menus display information on two levels (For example the upper level shows internal storage and
the lower level shows the centrifuges).
The buttons allow to switch from one level to the other.
IH-500 | 135
Checks and Adjustments
1
3
6
5
Figure 129.
1 Current state
2 Close Global controller menu
Press to close the global controller menu.
3 STOP / START button
4 INITIALIZE ALL button
This function initializes all modules.
5 Current speed:
The speed of the transport arm is displayed. Change it with the corresponding + / - buttons.
6 TURN ON/OFF TRANSPORT ARM button
To power off/on the transport arm.
Is ON per default, but it may be necessary to power it OFF.
136 | IH-500
State Module Panel
5
3
1 2 4
Figure 130.
Figure 131.
IH-500 | 137
Checks and Adjustments
Figure 132.
Figure 133.
Figure 134.
E Select Next.
Next
A countdown of 5 seconds starts before the gripper is powered off.
Figure 135.
138 | IH-500
State Module Panel
Figure 136.
Figure 137.
H Select Finish.
Finish
IH-500 | 139
Checks and Adjustments
1
9
8
7
2
6
Figure 138.
140 | IH-500
Firmware Menu
5.4.1.1 Procedure
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
The expected firmware versions are displayed in the Expected versions area (5).
k If a version is not correct, the board is automatically activated for upload (Check box dark red).
k If the Version information displays “???”, no firmware is loaded on the corresponding board.
A CHECK
Display the loaded versions.
VERSIONS
B If a board is not activated for upload after check versions but an update is
required, select the corresponding check box.
Figure 139.
5.4.1.2 Remark
When replacing an electronic board with a board without firmware, an error may occur when trying to
upload the firmware.
This error is due to an issue with the “Automation Controller” that may not be stopped correctly if a
board has no firmware.
IH-500 | 141
Checks and Adjustments
5.4.1.3 Workaround
To perform the firmware upload to an empty electronic board, it is required to verify that certain
processes are stopped.
Open the task manager and kill the following processes if they are running:
Figure 140.
142 | IH-500
Firmware Menu
4
12 1 2 3 5
11
10 6
9 7
Figure 141.
1 Application area
2 Name of the firmware installed
3 Installed version area
4 Contextual menu to display this screen
5 Expected version area
6 Current progress bar
7 Overall progress bar
8 Information filtered by: area
9 Upload firmware command
10 Upload gripper firmware command
11 Upload application button
12 Version check command
k APN application concerns gripper software, see chapter Gripper Firmware Update on page 145.
IH-500 | 143
Checks and Adjustments
5.4.2.1 Procedure
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A CHECK
Display the current firmware versions.
VERSIONS
k Ifcorresponding
the Installed version information displays “???”, no firmware is loaded on the
board.
k The APN application corresponds to the gripper software. See chapter Gripper Firmware
Update on page 145.
C If one of the installed version differs from the expected version, select:
Current progress bar (6) shows the upload status of the current task and
Overall progress bar (7) shows the overall status.
Steps are logged in the Information filtered by: area (8).
Figure 142.
F Shut down and start again the instrument to complete the upgrade.
The Installed version and Expected version areas should display the
same values.
144 | IH-500
Firmware Menu
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
k Use the IH-500 Gripper board prog cable (E10349) to upgrade the firmware.
A Disconnect the gripper cable (1) from the transport arm.
Figure 143.
B Connect the IH-500 Gripper board prog cable (E10349) (2) to the transport arm
and to the gripper spiral cable (3).
Figure 144.
Figure 145.
IH-500 | 145
Checks and Adjustments
E Remove 2 x screws (4) and the protective plate to access the computer.
4 4
Figure 146.
F Connect the serial port of the IH-500 Gripper board prog cable (E10349) to an
available port on the computer.
COM2 (5) is recommended.
Figure 147.
6
7
Figure 148.
H From the Communication settings tab, select the serial Port (6) (computer
port connected to the IH-500 gripper board prog cable E10349) and 38400 in
Baud Rate (7).
146 | IH-500
Firmware Menu
I From the Application tab, use the Browse button (8) and select APN version
available in C:\ProgramData\IH-500\Firmwares\APN01.xx.yy.zz.sw.
9 10
Figure 149.
L Disconnect the IH-500 gripper board prog cable (E10349) and plug the gripper
cable back on the robot.
k Disconnecting cable makes board reset which is necessary to start last downloaded version.
M Put back the protective plate.
IH-500 | 147
Checks and Adjustments
1
2 4
Figure 150.
148 | IH-500
Input / Output Menu
1 2
3 4 5 6
15 7
14 12 8
13 11 9
10
Figure 151.
IH-500 | 149
Checks and Adjustments
5.5.1.1 Ionizer
Figure 152.
b ELECTROCUTION
High voltage > 2500 V.
Figure 153.
B Confirm.
YES
A hazard symbol appears in the Ionizer area to remind about the high
voltage.
Figure 154.
150 | IH-500
Input / Output Menu
5.5.1.2 Fan
Figure 155.
B Check the internal light switches on over the pipetting area door.
IH-500 | 151
Checks and Adjustments
Figure 156.
Figure 157.
B Check that the external light switches on with the selected colour.
5.5.1.5 Temperatures
A Check the temperature values displayed for each module in the Temperatures
area.
Refer to the table below.
152 | IH-500
Input / Output Menu
Figure 158.
B When the LED is ON, check that three check boxes in the Resource sensor
area turn active together.
Figure 159.
D Check the Led switches OFF and that the check boxes in the Resource sensor
area are all inactive.
Figure 160.
IH-500 | 153
Checks and Adjustments
Figure 161.
154 | IH-500
Input / Output Menu
Figure 162.
IH-500 | 155
Checks and Adjustments
5.5.1.10 Drawers
RIGHT DRAWER
Figure 163.
5.5.1.11 Doors
Open
Closed
Figure 164.
156 | IH-500
Input / Output Menu
Absent
Present
Figure 165.
b adjustment.
Place carefully the IH-500 solid waste detection tool on its position to avoid incorrect
k table.
Refer to the appropriate procedure to check the full solid waste sensor when using the optional
IH-500 | 157
Checks and Adjustments
D Place the IH-500 solid waste detection tool (M20881) on the drawer back face
upper edge.
Figure 166.
G Check that the detection is active when the sighting point (1) is on the tool central
part.
Is full
The orange led (2) is ON and steady and the Waste state status is full.
I Check that the detection is inactive when the sighting point (3) is on the tool right
part.
Is empty
The orange led (4) is OFF and the Waste state status is empty.
158 | IH-500
Input / Output Menu
In the Solid waste area, the Waste state detected by the sensor is displayed.
b Place carefully the IH-500 solid waste detection tool on its position to avoid incorrect
adjustment.
D Place the IH-500 solid waste detection tool (M20881) on the drawer back face
upper edge.
Figure 169.
F Open the pipetting door and the right drawer to reach the full solid waste sensor.
Figure 170.
IH-500 | 159
Checks and Adjustments
2
1
Figure 171.
Figure 172.
k If no action is made on sensor within 5 minutes, the sensor programming mode turns OFF.
K Insert a metallic tool (e.g. screwdriver) at the rear of sensor.
Figure 173.
1Hz
Figure 174.
160 | IH-500
Input / Output Menu
1Hz
Figure 175.
N Within a delay of 4 seconds, briefly insert the metallic tool to change the operating
mode.
Successful calibration is confirmed if rear led lights up blue.
ᅛ
Figure 176.
8Hz
ᅟ
Figure 177.
IH-500 | 161
Checks and Adjustments
1 3 4
8 7 6 5
Figure 178.
162 | IH-500
Vision Station Menu
5.6.1 Pre-Adjustment
4
1 2
Figure 179.
k Protection cover needs to be removed to adjust the azimuth, aperture and sharpness.
IH-500 | 163
Checks and Adjustments
k Ifcompletely
the graduations are not visible, rotate the lens to see the scale. Never remove the lens
though.
b Place
carefully the camera azimuth adjustment tool on its position to avoid incorrect
adjustment.
F Check that the black shape is between the red template (4).
If it is not the case, follow the next steps to adjust the camera.
2 3
Figure 180.
164 | IH-500
Vision Station Menu
J Check that the red line adjustment is still OK (4). Tighten all screws.
L VALIDATE
VALIDATE the operation.
Azimuth
A green check mark displays by the validate button.
5
6
Figure 181.
IH-500 | 165
Checks and Adjustments
Figure 182.
5.6.3.3 Aperture
Figure 183.
166 | IH-500
Vision Station Menu
5.6.3.4 Sharpness
b Place carefully the sharpness setting tool on its position to avoid incorrect adjustment.
b Place assembly in front of the backlight.
A Open:
• the pipettor door;
• the maintenance doors.
C Insert a gel card (1) in the IH-500 Sharpness setting tool (2).
Tool reference M18979.
E Adjust the focus ring (3) to reach the highest sharpness value.
k Highest value is a peak. By turning one direction it is reached but will then decrease.
F VALIDATE
VALIDATE the operation.
VisionStationSha...
A green check mark displays by the validate button.
G Remove the gel card and the IH-500 Sharpness setting tool (M18979).
1
2
Figure 184.
IH-500 | 167
Checks and Adjustments
Figure 185.
D VALIDATE
Validate the pre-adjustment of the vision station (camera parameters).
ExposureTime
The camera parameters area displays: Calibrated.
168 | IH-500
Vision Station Menu
5.6.4 Adjustment
This menu allows the computer to calculate the mm/pixel ratio, and to detect height and diameter of
samples.
3 2
5
Figure 186.
Procedure
b Place carefully the sample camera calibration tool on its position to avoid incorrect
adjustment.
Figure 187.
IH-500 | 169
Checks and Adjustments
Figure 188.
Figure 189.
170 | IH-500
Vision Station Menu
5.6.5 Checks
This menu verifies the reading and interpretation of the different resources used on the instrument.
2
10
9
8
1
7
6 3
Figure 190.
k The
check can be performed with various configuration of resources: gel card and/or reagent
and/or diluent and/or sample tube.
IH-500 | 171
Checks and Adjustments
Procedure
11
12 12
11
Figure 191.
Figure 192.
F Load a gel cards basket partially or completely full in position 1 of the right drawer.
k Samples block and samples tubes are equipped with barcode label identification.
172 | IH-500
Vision Station Menu
Figure 193.
Figure 194.
M Check the quality of the picture and available information displayed in Results
area for each card.
Figure 195.
IH-500 | 173
Checks and Adjustments
N Check the quality of the picture and information displayed in Results area for
each sample.
Figure 196.
O Check the quality of the picture and information displayed in Results area for both
diluents.
Figure 197.
P Check:
• the quality of the picture and information displayed in Results area for the
three reagents;
• that the cap detection and volume are correct.
Figure 198.
174 | IH-500
Vision Station Menu
R VALIDATE
Select VALIDATE (4) to memorize status.
VisionStationChe...
Remarks
According to the selected option, the following features are verified by the instrument during the vision
check:
• Identification: the instrument checks that the ressources barcodes are correctly identified and the
gel control is correct for gel cards;
• Result analysis: the instrument checks and interprets the result of a reading.
• Brightness analysis: the instrument verifies the quality of the image, checks for the brighness of
each well and calculates the variance of brightness between the wells.
IH-500 | 175
Checks and Adjustments
4
3
5
2
11 10 8 7
9 6
Figure 199.
176 | IH-500
Transport Arm Menu
11 Command to move the transport arm to the resource sensor adjustment position button
The function moves the transport arm to the resource sensor adjustment position.
D E G
Figure 200.
IH-500 | 177
Checks and Adjustments
This step is to be performed with no tool nor resource loaded in the instrument.
b Place carefully the localization tool on its position to avoid incorrect adjustment.
b The localization tool is delicate, a crash or a fall can destroy it.
A Start the localization.
CHECK LOCALIZATION
A request window appears.
Figure 201.
Figure 202.
178 | IH-500
Transport Arm Menu
Figure 203.
Figure 204.
k The
localization tool needs a process in at least 3 steps to ensure that the tool is OK. Each step
consists in operating a 180° rotation.
Figure 205.
IH-500 | 179
Checks and Adjustments
Figure 206.
Figure 207.
Figure 208.
180 | IH-500
Transport Arm Menu
Figure 209.
IH-500 | 181
Checks and Adjustments
Figure 210.
Note that the field procedure which consists to add a layer of tape on the gel card used to adjust the
gripper has a direct impact on the pickup of the gel card.
The tape increases the gripping distance reference used to check the correct pick-up of gel card.
182 | IH-500
Transport Arm Menu
5.7.2 Mapping
This menu checks the resources sensor detection in all available positions.
Figure 211.
Procedure
k It is possible to check the desired modules selection boxes for mapping only these modules.
B Start the process.
START MAPPING
IH-500 | 183
Checks and Adjustments
Figure 212.
Otherwise the resource sensor localization shall be done again. See chapter
Resources Sensor Localization on page 228.
k case, the instrument checks with the gripper the presence of the vial.
On the reagent plate, the mapping sensor may detect a vial whereas no vial is present. In this
Figure 213.
H VALIDATE
If the check is successful, select VALIDATE to validate the mapping check status.
TransportArmMa...
Else, resource sensor localization shall be done again. See chapter Resources
Sensor Localization on page 228..
184 | IH-500
Transport Arm Menu
5.7.3 Transfer
This menu allows to remove resources from one position to the next. Only logical moves (which are also
performed by the instrument in the user software) can be performed.
Figure 214.
Procedure
IH-500 | 185
Checks and Adjustments
Figure 215.
Figure 216.
Figure 217.
Figure 218.
186 | IH-500
Transport Arm Menu
k
If a move is impossible an Error on moving resource window appears.
Figure 219.
For example, moving a gel card directly in pipettor area provokes an error (the right way is to
move cards in cards block and then cards block in pipettor area).
Figure 220.
IH-500 | 187
Checks and Adjustments
5.7.4 Cycles
This menu allows performing movement cycles for the different resources. In addition piercing cycles
for cards and diluents can be performed.
6
5 2
3
4
Figure 221.
k reagent or diluent.
Card offset, reagent offset or diluent offset refer to position number of corresponding card,
188 | IH-500
Transport Arm Menu
Parameter Description
Number of cards Indicates the number of gel cards to be used
Card offset Positions on internal storage and preparation area to ignore during the cycle test.
For example, if the parameter is 5, the positions 1 to 5 are not checked by during the cycle.
Number of cycles Number of cycles to perform
7 8
8 8
8 8
Figure 223.
IH-500 | 189
Checks and Adjustments
C Enter the corresponding values in Number of cards: and Card offset: fields.
Figure 224.
The parameter Number of cards indicates the number of gel cards used for
the cycling test.
The setting Card offset indicates that the positions on the internal storage to
ignore during the cycling test.
Figure 225.
Figure 226.
k Itused.
is possible to use only reagents OR diluents by entering «0» in number of resource filed not
190 | IH-500
Transport Arm Menu
B SELECT ORIGIN
Select SELECT ORIGIN CARD PLACES (2) to select the available resources.
CARD PLACES
The following screen displays the cards location.
Figure 227.
IH-500 | 191
Checks and Adjustments
E MODIFY INTERMEDIATE
Select MODIFY INTERMEDIATE POSITION.
POSITION
Figure 228.
J Use the gel card and diluent piercing tool (M16140) to check all pierced wells.
The tool should go through the wells only by its own weight and without
pressure.
Figure 229.
192 | IH-500
Transport Arm Menu
1
2
Figure 230.
B Move the diluent completely to the right to have the same position as the
instrument puts the diluent.
Figure 231.
k Use
a Diluent 1 for the diluent tip which is responsible for diluent 1.
Use a Diluent 2 for the diluent tip which is responsible for diluent 2.
J At the end of the process, open the door and remove the diluent racks.
IH-500 | 193
Checks and Adjustments
K Use the Gel card and diluent piercing tool (M16140) to check all pierced wells.
Figure 232.
k The tool should go through the wells only by its own weight and without pressure.
5.7.4.5 Cycles Validation
A VALIDATE
Validate the cycles check status.
Transport Arm C...
194 | IH-500
Transport Arm Menu
5.7.5 Teaching
This menu allows palpating the different modules of the instrument after a replacement or for
readjustment.
1
5
2 3 4
7
6
10 9
Figure 233.
k Inperform
case of a module change or a module part change, refer to the replacement procedure to
teaching.
IH-500 | 195
Checks and Adjustments
b•
Before using the touch probe, check the following points:
the body (M16637) of the touch probe is correctly tightened on the touch probe
support (M18681);
Use the probe body tool (M17799) from the gauge kit to fix the touch probe body.
• the head (M17540) is correctly magnetized on the body;
• the tip (M20667) is correctly tightened on the head.
Use the Probe tip tool (M17798) from the gauge kit to fix the touch probe tip.
k Several teachings without use of the probe can be performed at the same time.
3
1
1
Figure 234.
Figure 235.
196 | IH-500
Transport Arm Menu
Figure 236.
E Manually and carefully move the touch probe tip (4) several times.
Do at least 3 times.
Figure 237.
G Validate.
OK
The following message appears.
Figure 238.
I Validate.
OK
IH-500 | 197
Checks and Adjustments
b•
Before using the touch probe, check the following points:
the body (M16637) of the touch probe is correctly tightened on the fixed probe support
(M18275);
Use the Probe body tool (M17799) from the gauge kit to fix the touch probe body (M16637).
• the head (M17540) is correctly magnetized on the body;
• the tip (M20667) is correctly tightened on the head.
Use the Probe tip tool (M17798) from the gauge kit to fix the touch probe tip (M20667).
1 4
1
Figure 239.
b adjustment.
Place carefully the localization tool (M18741) on its position to avoid incorrect
b The IH-500 localization tool is delicate, a crash or a fall can destroy it.
b Verify the correct assembly of the touch probe.
A Start the teaching of the gripper position.
LOCALIZE GRIPPER
A confirmation request is displayed.
198 | IH-500
Transport Arm Menu
B Continue.
OK
Pipetting door and right drawer open.
A Transport arm request message is displayed.
Figure 240.
Figure 241.
F Manually and carefully move the touch probe tip (3) several times.
Do at least 3 times.
Figure 242.
IH-500 | 199
Checks and Adjustments
H Validate.
OK
A Transport arm request message is displayed.
Figure 243.
I Insert and center the IH-500 localization tool (M18741) in the gripper legs (4).
Figure 244.
Figure 245.
Figure 246.
200 | IH-500
Transport Arm Menu
Figure 247.
Figure 248.
U Perform another 180° rotation of the tool and repeat steps above.
A Transport arm request message is displayed.
V Remove the tool.
IH-500 | 201
Checks and Adjustments
Figure 249.
Figure 250.
202 | IH-500
Transport Arm Menu
5.7.5.3 J1 Calibration
The first operation performed by the manufacturer is the J1 robot calibration. The transport arm robot
(reference face (1)) has to identify its position relatively to the casing (first touched area (2)).
Figure 251.
1 J1 Robot axis
2 Diluent location used as reference axis
b Ifthetheinstrument.
J1-axis is modified, it is necessary to make a complete teaching and adjustment of
This operation is reserved to manufacturer. Perform this operation only
after confirmation from instrument support.
Figure 252.
IH-500 | 203
Checks and Adjustments
E Confirm.
OK
A transport arm request message is displayed.
The pipetting door opens automatically.
204 | IH-500
Transport Arm Menu
Figure 253.
k Before the teaching, remove dust from the interior of the centrifuges and on the cards holders.
A Install the touch probe.
See chapter Touch Probe Installation (Probe Mounting) on page 196.
E Validate to continue.
OK
A check is required to ensure the localization is possible.
F Check that:
• the touch probe is above the hole (1);
• the card holder N°1 is directed towards the front of the instrument.
IH-500 | 205
Checks and Adjustments
I Open the doors and put the internal storage back in place.
206 | IH-500
Transport Arm Menu
Figure 254.
D Open the door. Turn the internal storage up. Uncover the rear centrifuge. Remove
the reading station cover. Close the doors.
E Validate to continue.
OK
A check is required to ensure the localization is possible.
F Check that:
• the touch probe is above the hole (1);
• the card holder N°1 is directed towards the front of the instrument.
IH-500 | 207
Checks and Adjustments
I Open the doors. Put the rear centrifuge, the internal storage and the reading
station covers back in place.
208 | IH-500
Transport Arm Menu
IH-500 | 209
Checks and Adjustments
210 | IH-500
Transport Arm Menu
1
2
Figure 255.
1
2
Figure 256.
IH-500 | 211
Checks and Adjustments
1
2
Figure 257.
212 | IH-500
Transport Arm Menu
b Place carefully the gripper calibration tool on its position to avoid incorrect adjustment.
k Remove dust from the left drawer.
A Install the touch probe.
See chapter Touch Probe Installation (Probe Mounting) on page 196.
1
2
Figure 258.
IH-500 | 213
Checks and Adjustments
1
2
Figure 259.
1
2
Figure 260.
214 | IH-500
Transport Arm Menu
IH-500 | 215
Checks and Adjustments
5.7.5.13 Piercer
B Select Piercer.
In Teaching zones - Transport arm area.
216 | IH-500
Transport Arm Menu
IH-500 | 217
Checks and Adjustments
k palpation).
Two palpations are performed so it can take a long time (gripper turns 180° for the second
5.7.5.16 Incubator
B Select Incubator.
In Teaching zones - Transport arm area.
218 | IH-500
Transport Arm Menu
IH-500 | 219
Checks and Adjustments
1 2
2nd row
b Place
carefully the IH-500 reagent area teaching tool on its position to avoid incorrect
adjustment.
F Move the reagent module insulating body (2) to the rear to release it from the
retaining clips.
220 | IH-500
Transport Arm Menu
3
2
Figure 262.
b Do not damage the motor connector when removing the reagent module insulating body.
H Place the five IH-500 reagent area teaching tools M19892 in the middle of the
second row.
K Move the five IH-500 reagent area teaching tools M19892 to the middle of the
third row.
N Move the five IH-500 reagent area teaching tools M19892 to the middle of the
fourth row.
Q Move the five IH-500 reagent area teaching tools M19892 to the middle of the fifth
row.
IH-500 | 221
Checks and Adjustments
b Do not damage the motor connector when installing the reagent module insulating body.
U Remove the probe.
REMOVE PROBE
The transport arm moves to its touch probe removal position.
The instrument doors open. A message is displayed.
b Dispensing in the gel cards is critical. The adjustment must be very accurate.
A Install the touch probe.
See chapter Touch Probe Installation (Probe Mounting) on page 196.
222 | IH-500
Transport Arm Menu
J Check the Pipettor position in the card block for all position. Adjust if needed.
See chapter Check on page 247.
IH-500 | 223
Checks and Adjustments
224 | IH-500
Transport Arm Menu
IH-500 | 225
Checks and Adjustments
226 | IH-500
Transport Arm Menu
Figure 263.
b Place carefully the pipettor localization tool on its position to avoid incorrect adjustment.
A Install the touch probe.
See chapter Touch Probe Installation (Probe Mounting) on page 196.
E Place the pipettor localization tool M20093 (1) on the diluent pipetting area.
Use 4 screws CHC M3x8.
b This step is mandatory and must be performed directly after this procedure.
K Perform all the pipettor points update.
See chapter Pipettor Area Update on page 232.
IH-500 | 227
Checks and Adjustments
D Insert the gripper calibration and mapping sensor location tool (M20873) in the
third position.
The arrow must point to the back.
1
2
Figure 264.
G Validate.
OK
A Transport arm request message is displayed.
The pipetting door opens automatically.
I Adjust the resources sensor screw (3) until the LED flashes (unstable detection).
Figure 265.
228 | IH-500
Transport Arm Menu
K Validate.
OK
Figure 266.
O Validate.
OK
After successful teaching, the red cross mark is replaced by a green check
mark for the Resources sensor localization status.
k case, the instrument checks with the gripper the presence of the vial. 2 wrong detections of
On the reagent plate, the mapping sensor may detect a vial whereas no vial is present. In this
IH-500 | 229
Checks and Adjustments
Figure 267.
k New manufactured gel cards are identifiable with letter A or B marked between wells 3 and 4.
E Insert the gripper calibration and mapping sensor location tool (M20873) in the
third position.
The arrow must point to the back.
1
2
Figure 268.
I Validate.
OK
A Transport arm request message is displayed.
The left drawer opens automatically.
230 | IH-500
Transport Arm Menu
L Validate.
OK
After successful teaching, the red cross mark is replaced by a green check
mark for the Gripper calibration status.
Figure 269.
k one.
If only one gel cards tool is available, perform the procedure by selecting gel cards block one by
D Place 4 gel cards tool M20579 (1) on the gel cards pipetting area.
A Transport arm request message is displayed.
E Follow the instructions on screen.
F Validate.
OK
After successful teaching, the red cross mark is replaced by a green check
mark for all update statuses.
IH-500 | 231
Checks and Adjustments
k selected area. This update is necessary after a new transport arm teaching of an area shared
The pipettor area points update operates a new calculation of the points coordinates of the
A Select the desired points area to update in the Teaching zones - Pipettor area
Washing station points for example.
Figure 270.
Figure 271.
232 | IH-500
Transport Arm Menu
3 5
Figure 272.
b The secure posing is done at factory. It is not required on the field unless clearly
expressed during the replacement of a part.
Refer to the Replacement Procedure document (RP).
IH-500 | 233
Checks and Adjustments
Figure 273.
5.7.6.2 Others
Figure 274.
234 | IH-500
Centrifuge Menu
k Rear Centrifuge.
The description applies for the screens Pre Control - Front Centrifuge and Pre Control -
1 2
6 7
Figure 275.
IH-500 | 235
Checks and Adjustments
b must be checked.
After an intervention on centrifuges, the correct lock of centrifuges covers and trapdoors
Figure 276.
C Lift up the cards internal storage (1) to access the front centrifuge.
k The
Origin sensor is Active only when the angle current position is near 0°. It indicates the
real initialization position of the motor encoder.
E Check that the gel card holder N°1 (2) is oriented to the front of the instrument.
G Manually rotate the centrifuge head to orientate the holder N°1 in front of the
instrument (2).
J VALIDATE
Validate.
FrontCentrifugeB...
236 | IH-500
Centrifuge Menu
Figure 277.
IH-500 | 237
Checks and Adjustments
b must be checked.
After an intervention on centrifuges, the correct lock of centrifuges covers and trapdoors
Figure 278.
k The Origin sensor is Active only when the angle current position is near 0°. It indicates the
real initialization position of the motor encoder.
H Check that the gel card holder N°1 (2) is oriented to the front of the instrument.
J VALIDATE
Validate.
RearCentrifugeB...
238 | IH-500
Centrifuge Menu
Figure 279.
IH-500 | 239
Checks and Adjustments
b must be checked.
After an intervention on centrifuges, the correct lock of centrifuges covers and trapdoors
D Check that the number of cycles performed is equal to the one defined in Cycle
number.
Displayed by the field Cycle passed:.
E VALIDATE
Validate.
FrontCentrifugeB...
Figure 280.
240 | IH-500
Centrifuge Menu
B Close the rear centrifuge cover or/and the lower the cards internal storage.
Figure 281.
Figure 282.
IH-500 | 241
Checks and Adjustments
Figure 283.
90°
Figure 284.
242 | IH-500
Centrifuge Menu
8 7 6 5 4
Figure 285.
IH-500 | 243
Checks and Adjustments
k Make sure the pipetting area access door is closed. Otherwise the menu is not available.
5.9.1 Pre Control
3
4
Figure 286.
244 | IH-500
Pipettor Menu
Figure 287.
5.9.1.2 Procedure
A Initialize each axis to make sure that the initialization position is available.
INIT X
Use the function keys to initialize each axis in the XY check and Z check
areas.
INIT Y
INIT Z
k The Origin sensor indicates the real initialization position of the motor encoder.
C Expand the State module panel. Initialize the pipettor.
INITIALIZE
Wait until the pipettor status is Functional.
k The hole and the needle must be clean to perform this operation.
D Check the needle localization.
CHECK CONTACT
DETECTION
k During the check, the pipettor localization is repeated three times at least.
IH-500 | 245
Checks and Adjustments
F Align the needle tip with the center of the hole, without touching the bottom.
Use the cursors of XY check and Z check areas to move the needle tip.
k Also possible to use the keyboard arrow keys (left/right/up/down) instead of the touch screen.
G SAVE Save the new position.
LOCALIZATION
POSITION
X
Y
Figure 288.
246 | IH-500
Pipettor Menu
5.9.2 Check
This menu allows to check the position of the needle in every possible position. Single points can also
be adjusted here.
10 2
7 6 5 4
9 8
Figure 289.
b Perform a first approach with APPROACH XY button, then the Z adjustment manually
There is a risk of crash in Z axis if the tool position or adjustments are not correct.
with cursor.
IH-500 | 247
Checks and Adjustments
5.9.2.1 Reagents
b Place carefully the reagent tool on its position to avoid incorrect adjustment.
k Always install the reagent tool in the longitudinal direction of the reagent module.
A Select Macro selection - Reagents (10 - Figure 289).
B Insert the IH-500 reagent tool (M20880) in position 28 (11) on the reagent module.
k Use the Z-axis adjustment function keys to make easier the position check.
k The adjustment is available only for the selected position.
F Select Macro selection - Idle back (10 - Figure 289) to move back the
pipettor.
To get an access to the IH-500 reagent tool.
G Repeat the procedure for the 4 corner positions (12) on the reagent module.
248 | IH-500
Pipettor Menu
11
12
Figure 290.
IH-500 | 249
Checks and Adjustments
b Dispensing in the gel cards is critical. Adjustments and checks must be very accurate.
b Place carefully the gel card tool on its position to avoid incorrect adjustment.
A Select Macro selection - Card blocks (10 - Figure 289).
B Install the IH-500 Gel card tool (M20579) in the card blocks area (1) of the pipettor.
C Select the well 1 of card 1 (6) corresponding to the position of the IH-500 gel card
tool.
If the option GO ON CLICK is selected (ON), the pipettor automatically moves
to the selected position.
E Place the IH-500 Z needle control tool (M20882) (2) between the top of IH-500 Gel
card tool (M20579) and the needle.
k Use the Z-axis adjustment function keys to make easier the position check.
k The adjustment is available only for the selected position.
I Select Macro selection - Idle back (10 - Figure 289) to move back the
pipettor.
J Repeat the procedure for the 96 positions of the gel card block.
Adjust only the wells 1 and 6 for each line (card).
The wells 2, 3, 4 and 5 are automatically calculated.
250 | IH-500
Pipettor Menu
2 1
Figure 291.
IH-500 | 251
Checks and Adjustments
5.9.2.3 Diluents
b Place carefully the diluent rack tool on its position to avoid incorrect adjustment.
A Select Macro selection - Diluents (10 - Figure 289).
B Place the diluent rack tool (M20158) (1) in position 1 of the pipettor area.
k rack.
Move the tool completely to the right to insert it the same way the instrument inserts a diluent
D Select well 1 of the rack 1 (4) corresponding to the position of the tool.
If the option GO ON CLICK is selected (ON), the pipettor automatically moves
to the selected position.
k Use the Z-axis adjustment function keys to make easier the position check.
k The adjustment is available only for the selected position.
G Select Macro selection - Idle back to move back the pipettor.
H Repeat the procedure for the extreme positions of each rack (5).
252 | IH-500
Pipettor Menu
4 5
Figure 292.
IH-500 | 253
Checks and Adjustments
5.9.2.4 Samples
b Place carefully the sample tube on its position to avoid incorrect adjustment.
A Select Macro selection - Samples (10 - Figure 289).
B Place the tube Ø12x75 (M12450) (1) in position 1 on sample tube rack 1.
k Use the Z-axis adjustment function keys to make easier the position check.
k The adjustment is available only for the selected position.
F Select Macro selection - Idle back to move back the pipettor.
Figure 293.
254 | IH-500
Pipettor Menu
F Repeat the procedure for the decontamination (2) and overflow (3) wells.
Top and bottom.
2 4
Figure 294.
IH-500 | 255
Checks and Adjustments
Figure 295.
k Also possible to use the keyboard arrow keys (left/right/up/down) instead of the touch screen.
Modify the XY speed and Z speed values to move the pipettor faster.
Fields located in XY adjustment and Z adjustment areas.
k Visible when selecting the direction button (+XY, -XY, +Z, -Z).
C Move the needle until the tip is centered and flushed with the target.
~ 0 mm
Figure 296.
E Confirm.
YES
256 | IH-500
Pipettor Menu
G Confirm.
YES
5.9.3 Burn In
Figure 297.
D Check that the CYCLE PASSED number is equal to the predefined CYCLE
NUMBER.
E VALIDATE
Validate.
PipettorBurninVa...
IH-500 | 257
Checks and Adjustments
k If the user wants to sample into other tubes, the FSE must set new parameters for the tubes.
k block of this type of tubes is introduced in the instrument, the tubes are not well identified, a
If the configuration of the pediatric or low volume tube is not performed, and if one sample tube
0x0118902 error will be displayed, and all the tubes will be considered as absent.
2 3
Figure 298.
258 | IH-500
Pipettor Menu
Sampling Parameters
B Sampling offset Serum (mm) lowest Serum sampling position compared to the bottom of the tube
C Sampling offset CRBC (mm) RBC sampling position compared to the bottom of tube
D Tube’s depth (mm) internal height of the tube, between the bottom (value declared above)
and the top of the tube
E Tube’s internal diameter (mm) the diameter of the tube measured on the top of the tube
E
D
B
A
B
E D
C
Figure 299.
IH-500 | 259
Checks and Adjustments
1
2
Figure 300.
E OPEN PIPETTING
Open the pipetting door to reach the needle.
DOOR
260 | IH-500
Pipettor Menu
F Move down the needle until it touches the bottom of the tube.
Figure 301.
I Enter the internal height of the pediatric tube in the field Tube’s depth.
In the example:
• for pediatric tube = 42 mm;
• for low volume tube = 74 mm.
Otherwise an error message is displayed.
J Enter the new value in the field Sampling offset Serum (mm).
In the example:
• for pediatric tube = 10 mm;
• for low volume tube = 4 mm.
L Enter the internal diameter in the field Tube’s internal diameter (mm).
In the example:
• for pediatric tube = 8 mm;
• for low volume tube = 11 mm.
N Confirm.
YES
IH-500 | 261
Checks and Adjustments
5.10.1 Units
This screen allows to activate hydraulic components individually.
1
6
5
3
4
Figure 302.
262 | IH-500
Hydraulic Menu
Select the valves to activate/deactivate. The status is shown by the continuous line.
EV04 EV04
Figure 303.
Select the pumps to start/stop them. Next available action changes accordingly.
PO05 PO05
Figure 304.
When the hydraulic sensors state changes, the corresponding colour and state information change
accordingly.
CP36 CP36
Connect Disconn
Figure 305.
IH-500 | 263
Checks and Adjustments
5.10.1.1 Sensors
Connection Sensors
A Press down the detection handle of the system liquid container 1 corresponding to
sensor CP36.
C Repeat the procedure for the sensors CP37, CP30 and CP31.
Level sensors
A Approach the hand to the level sensor CP33 (waste container 2).
C Repeat procedure with CP32, CP41, CP42, CP40 and CP70 (overflow sensor).
CP31
CP37
CP30
CP36
CP33
CP32
CP41
CP42
CP40
CP70
Figure 306.
264 | IH-500
Hydraulic Menu
A Insert the waste container 1 (2) in position 1 to check the sensor CP34.
C Repeat the procedure for the waste container position 2 and the sensor CP35.
D Insert a cleaning liquid container (black cap) (1) in the first system liquid container
position to check the sensor CP38.
F Repeat the procedure for the system liquid container in position 2 and the sensor
CP39.
CP35
CP34
CP39
CP38
Figure 307.
IH-500 | 265
Checks and Adjustments
EV05
EV06
EV03
PO04
EV04 EV07
PO05
FT02
FT01
FT05
PO01
PO03
Figure 308.
b Each component can be driven individually, BUT when using liquid make sure dispensing
is sucked in waste circuit.
k All components can be activated when dry (purged circuit), except PDS pump PO02.
k Activation can be validated by state change noise or electrical motor start.
k Sound for solenoid valves can be very low.
266 | IH-500
Hydraulic Menu
A Insert at least:
• a full system liquid container;
• an empty liquid waste container.
Figure 309.
Figure 310.
IH-500 | 267
Checks and Adjustments
Decontamination well
Overflow well
Collect well
Overflow
well
EV05
Collect well
EV06
Decontamination
well
PO05
EV03
CP30 CP31
Not full Not full
CP32 CP33
Not full Not full
CP34 CP35
Present Present
Figure 311.
A Check that the EV03 drives the flow into the present container.
Otherwise select the EV03 icon to change direction.
268 | IH-500
Hydraulic Menu
CP21
Filled
Intermediate
container
PO02
EV01
EV02
Needle
PO04
Figure 312.
B Select PIPETTOR TO COLLECT WELL (3) to move the pipettor to the washing
station.
C Check that the EV03 drives the flow into the present container.
Otherwise select the EV03 icon to change direction.
E Listen to the suction noise to make sure the waste circuit is OK.
IH-500 | 269
Checks and Adjustments
CP21
Filled
Intermediate
container
PO02
EV01
EV02
Needle
PO04
Figure 313.
B Select PIPETTOR TO COLLECT WELL (3) button to move the pipettor to the
washing station.
C Check that the EV03 drives flow into the present container.
Otherwise select EV03 icon to change direction.
270 | IH-500
Hydraulic Menu
Overflow
well
Collect well
Decontamination
well
Decontamination well
Overflow well
Collect well
PO03 EV05
EV06
PO05
EV03
CP30 CP31
Not full Not full
CP32 CP33
Not full Not full
Decontamination
Waste container 1 Waste container 1
solution
Figure 314.
C Listen to the suction noise to make sure that the waste circuit is OK.
IH-500 | 271
Checks and Adjustments
Decontamination well
Overflow well
Collect well
PO03 EV05
EV06
PO05
EV03
CP30 CP31
Not full Not full
CP32 CP33
Not full Not full
Decontamination
Waste container 1 Waste container 1
solution
Figure 315.
272 | IH-500
Hydraulic Menu
2 3 4
10 5
9 6
8 7
Figure 316.
IH-500 | 273
Checks and Adjustments
Figure 317.
Figure 318.
274 | IH-500
Hydraulic Menu
B Confirm.
Next
A warning about the mandatory use of gloves is displayed.
C Confirm.
Next
k Make sure that the localization hole and the needle are clean.
E Wait during the finalization.
The automate operates a localization of the needle.
Figure 319.
IH-500 | 275
Checks and Adjustments
N VALIDATE
Validate.
Fluidic Circuits
Priming
The priming is used to fill the whole hydraulic circuit between the intermediate tank and the needle.
All bubbles are eliminated.
CP21
Filled
Intermediate
container
PO02
EV01
EV02
Needle
PO04
Figure 320.
k The START PRIME button is active only if the Intermediate container filling is started.
A Start the priming.
START In the Fluidic circuit area.
A successful priming enables the START PRIME button.
276 | IH-500
Hydraulic Menu
Purge
The purge allows to empty the hydraulic circuit between the intermediate tank and waste container.
CP21
Filled
Intermediate
container
PO02
EV01
EV02
Needle
PO04
Decontamination well
Overflow well
Collect well
EV05
EV06
PO05
EV03
CP30 CP31
Not full Not full
CP32 CP33
Not full Not full
CP34 CP35
Present Present
Figure 321.
IH-500 | 277
Checks and Adjustments
Flushing
The flushing allows to evacuate particles in the hydraulic circuit.
CP21
Filled
Intermediate
container
PO02
EV01
EV02
Needle
PO04
Figure 322.
k The flushing is necessary when the circuit is filled for the first time with liquid and after
intervention on:
• the intermediate tank;
• the tubing between the intermediate tank and the complete sampling module;
• the washing pump;
• the complete sampling module;
• the faradized vein.
Figure 323.
278 | IH-500
Hydraulic Menu
Figure 324.
Priming
The decontamination priming brings the NaOH to the top of decontamination well.
All bubbles are eliminated.
Decontamination well
Overflow well
Collect well
PO03
EV05
Decontamination
solution
CP40
Filled
Figure 325.
IH-500 | 279
Checks and Adjustments
Purge
The purge allows to empty the hydraulic circuit.
CP21
Filled
Intermediate
container
PO02
EV01
EV02
Needle
PO04
Decontamination well
Overflow well
Collect well
EV05
EV06
PO05
EV03
CP30 CP31
Not full Not full
CP32 CP33
Not full Not full
CP34 CP35
Present Present
Figure 326.
280 | IH-500
Hydraulic Menu
PO02
EV01
EV02
Needle
PO04
Decontamination well
Overflow well
Collect well
EV05
EV06
PO05
EV03
CP30 CP31
Not full Not full
CP32 CP33
Not full Not full
CP34 CP35
Present Present
Figure 327.
IH-500 | 281
Checks and Adjustments
A VALIDATE
Validate the fluidic circuits checks.
FluidicCircuitsVal...
282 | IH-500
Hydraulic Menu
b detection checks.
It is required to perform a priming of hydraulic circuits before performing the level
10 3
9 2
8
7 4
6
Figure 328.
IH-500 | 283
Checks and Adjustments
3 ml
Figure 329.
Figure 330.
284 | IH-500
Hydraulic Menu
F Check that the ten level detections are performed without error, under the status of
detection cycle area.
Figure 331.
Figure 332.
The maximum divergence between Max value and Min value must be less than
0.3 mm.
IH-500 | 285
Checks and Adjustments
1 ml
Figure 333.
Figure 334.
286 | IH-500
Hydraulic Menu
G Check that the ten level detections are performed without error, under the status of
detection cycle area.
Figure 335.
Figure 336.
The maximum divergence between Max value and Min value must be less than
0.3 mm.
Figure 337.
IH-500 | 287
Checks and Adjustments
D Check that the ten level detections are performed without error, under the status of
detection cycle area.
Figure 338.
Figure 339.
The maximum divergence between Max value and Min value must be less than
0.3 mm.
A VALIDATE
Validate the level detection checks.
LevelDetectionVal...
288 | IH-500
Hydraulic Menu
2
11 3
10 7 6 5 4
8
Figure 340.
IH-500 | 289
Checks and Adjustments
Figure 341.
Figure 342.
290 | IH-500
Hydraulic Menu
Figure 343.
Figure 344.
K Check that the needle is centered on targets (sample tube, diluent well, gel card
well).
During the process
Sample liquid
Air gap
Figure 345.
IH-500 | 291
Checks and Adjustments
Figure 346.
Figure 347.
292 | IH-500
Hydraulic Menu
Figure 348.
H Select START.
START
I Check that the needle is centered on targets (sample tube, diluent well, gel card
well).
During the process
Sample liquid
Air gap
Figure 349.
IH-500 | 293
Checks and Adjustments
Figure 350.
Figure 351.
Figure 352.
294 | IH-500
Hydraulic Menu
H Select START.
START
I Check that the needle is centred on targets (sample tube, diluent well, gel card
well).
During the process
Sample liquid
Air gap
Figure 353.
Figure 354.
IH-500 | 295
Checks and Adjustments
Figure 355.
Figure 356.
H Select START.
START
I Check that the needle is centred on targets (sample tube, diluent well, gel card
well).
During the process.
Sample liquid
Air gap
Figure 357.
296 | IH-500
Hydraulic Menu
5.10.4.5 Sample
Figure 358.
C Select Reagents.
In the Macro selection area.
Figure 359.
F Select START.
START
IH-500 | 297
Checks and Adjustments
Figure 360.
H Select Diluents.
In the Macro selection area.
Figure 361.
K Select START.
START
L Check that the needle is centered on the diluent well and no error occurs.
During the process.
Figure 362.
M Select Samples.
In the Macro selection area.
298 | IH-500
Hydraulic Menu
Figure 363.
P Select START.
START
Q Check that the needle is centered on the sample tube and no error occurs.
During the process.
Figure 364.
IH-500 | 299
Checks and Adjustments
Figure 365.
Figure 366.
E Select START.
START
F Check that needle dips in the tube and stops near the bottom.
Figure 367.
300 | IH-500
Hydraulic Menu
5.10.4.7 Distribute
Figure 368.
Figure 369.
D Select START.
START
F Check that the needle is centered on gel card wells and no error occurs.
During the process.
IH-500 | 301
Checks and Adjustments
5.10.4.8 Wash
CP21
Filled
Intermediate
container
PO02
EV01
EV02
Needle
PO04
Figure 370.
C Select START.
START
D Check that the needle is centered on the washing station and no error occurs.
During the process.
A VALIDATE
Validate the sampling/dispensing checks.
SamplingandDisp...
302 | IH-500
Incubator Menu
2
3
6 5
Figure 371.
IH-500 | 303
Checks and Adjustments
B Wait until the temperatures at front and rear sensors are stable.
G Place the thermometer sensor under the instrument, near the external sensor (3).
I Place a thermometer sensor in the instrument, near the internal sensor (4).
Next to the visual station.
M Confirm.
YES
The compensation is automatically calculated.
The value is displayed in the Offset field.
N VALIDATE
Validate the adjustment.
IncubatorAdjust...
304 | IH-500
Incubator Menu
Figure 372.
IH-500 | 305
Checks and Adjustments
7 8
6
5 4 3 2
Figure 373.
k The front and rear temperature have to be the target temperature (38.5°C ± 1°C).
k The maximal difference between front and rear temperatures should be 2°C during regulation.
A VALIDATE
Validate the temperature check.
IncubatorCheckV...
306 | IH-500
Reagent Plate Menu
Figure 374.
IH-500 | 307
Checks and Adjustments
Figure 375.
Figure 376.
k•
A typical cycle is alternately:
4x (10s in clockwise, 1s short downtime, 10 seconds in counter-clockwise);
• 1x 420 seconds long downtime.
G VALIDATE
Validate the shaking pre-control.
Shaking Precontr...
308 | IH-500
Reagent Plate Menu
5.12.1.3 Burn-In
D Check that the Cycle number value is equal to the required Cycle number and
that there are no errors.
F VALIDATE
Validate the reagent plate burn-in.
Reagent Area Bur...
IH-500 | 309
Checks and Adjustments
7 2
5 4
3
Figure 377.
Procedure
B Wait until the Regulated and Control temperatures (4) are stable.
k The
top sensor is used for the regulation.
Expected temperature: 13±1 °C.
310 | IH-500
Reagent Plate Menu
H Confirm.
YES
The compensation is automatically calculated.
The value is displayed on the Offset field.
I VALIDATE
Validate the adjustment.
Cooling Adjustm...
Figure 378.
IH-500 | 311
Checks and Adjustments
5 4 3 2
Figure 379.
k The Regulated and Control cooling temperatures have to be target temperature (13 ± 1 °C).
A VALIDATE
Validate the temperature check.
Cooling Check Va...
312 | IH-500
Final Check / QC Menu
1
2
4 5
Figure 380.
b Ifbethea crash
diluent locking device is in closed position when starting a Final check, there will
while loading the diluent rack and a breakage of the gripper legs.
IH-500 | 313
Checks and Adjustments
Figure 381.
Figure 382.
314 | IH-500
Final Check / QC Menu
5.13.2 QC Cycles
k
This section is used by the manufacturing service for QC qualification before instrument
release. It is not recommended to use this function.
IH-500 | 315
Checks and Adjustments
2 9
4
8
5
7
6
Figure 383.
316 | IH-500
Tools Menu
A Set the date from which the data should be collected in the Selected target
date.
Figure 384.
B Enter the number of previous days to collect data related to the selected target
date in the field Number of (previous) days.
Maximum 60 days.
IH-500 | 317
Checks and Adjustments
k
The snapshot must be created on an external device.
The name of the created snapshot follows this format:
[sn]_[yyyy-mm-dd]_[hh]h[mm]_IH500_Snapshot.
G Confirm.
OK
A description line describes the backup process.
k This is only possible after a snapshot has been created on an external drive.
5.14.1.2 System Snapshot
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Open the Acronis software.
ACRONIS
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
k For details refer to the installation procedure IH-500 IP-Backup & Restore.
This section is used by the manufacturing service. It is not recommended to use this function.
Figure 385.
318 | IH-500
Tools Menu
B Confirm.
YES
Figure 386.
Figure 387.
Figure 388.
IH-500 | 319
Checks and Adjustments
Figure 389.
Figure 390.
C Confirm.
YES
The transport arm adjustment and parameters are restored.
The instrument prompts to restart the instrument.
Figure 391.
320 | IH-500
Tools Menu
Figure 392.
k Removing resources is only necessary if a software update is performed. The loaded resources
information are deleted.
C Confirm.
OK
D Log in.
Username: Biorad
Password: bior@d500
The Windows classic desktop is displayed. The parameters like regional
settings, keyboard language or barcode scanner can be configured.
See the chapter Language Configuration (OS, Keyboard and Barcode
Scanner) on page 117 to modify parameters or see the appropriate Released
Note for an update.
IH-500 | 321
Checks and Adjustments
Reset Statistics
Cleanup System
The function allows:
• to create an archive on an external drive;
• to clean the database;
• to remove all APF’s and the user data;
• to remove all archives;
• to remove all files on the desktop;
• to remove all the event windows.
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Clean the operating software system.
CLEANUP SYSTEM
Refer to the installation procedure of IH-500_CleanupDatabaseDescription
to realize a cleanup system and a system re-installation.
The function creates an archive on an external drive and cleans the database.
b Previous
analyses, logs and APFs, Windows trash and recent files are erased.
The Windows desktop is also cleaned except for the short-cuts to FIS Launcher and to
User Management.
A PREPARE FOR
Select to:
SHIPMENT
• clean and purge the hydraulic circuit;
• prepare the instrument for shipment.
322 | IH-500
Tools Menu
A Modify the Server name field to allow user management from a remote server.
Figure 393.
k Remove server is not functional yet. Only use FIS as server name.
B Update the configuration.
SAVE
B Select the option External solid waste for the use of the optional table solid
waste bin.
Refer to Optional Table Solid Waste Installation on page 104 for more details.
C Select the option External liquid waste for the use of an external waste drain.
Refer to External Liquid Waste Optional Kit Installation on page 95 for more
details.
IH-500 | 323
Checks and Adjustments
2 4
3
5
7 6
Figure 394.
b The
instrument serial number change involves that the status turns FAULT.
The instrument has to be completely adjusted.
k This section is used by the manufacturing service. It is not recommended to use this function.
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Enter the new instrument serial number (2).
If necessary.
324 | IH-500
Tools Menu
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Select a new product line.
In the drop-down menu.
B CHANGE PRODUCT
Confirm the selection.
LINE
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Enter the license keys on the fields (6).
The software license number for Windows and Acronis are visible on the
sticker of the computer.
The license number for EURESYS (2D barcode) have to be added manually.
Refer to document IH-500 IP-V2.2 Installation Update V1.4, chapter 3, step 4
the point A.
There must be at least one digit entered in all licenses fields.
IH-500 | 325
Checks and Adjustments
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Make sure that the instrument is connected to Internet.
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Make sure that the instrument is connected to Internet.
326 | IH-500
Tools Menu
Figure 395.
Figure 396.
IH-500 | 327
Checks and Adjustments
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Start the Euresys license manager.
On the PC with the internet connection.
Figure 397.
328 | IH-500
Tools Menu
Figure 398.
Figure 399.
IH-500 | 329
Checks and Adjustments
O Plug the USB memory key on the PC with the internet connection.
W Close the license manager application. Unplug the USB memory key.
Online Recovery
Restoring a license may be necessary in the event of any major changes to the computer as:
• a computer replacement;
• a hard disk replacement;
• an operation system re-installation (e.g. after Acronis).
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Make sure that the instrument is connected to Internet.
330 | IH-500
Tools Menu
Figure 400.
F Select:
• No, if a Acronis backup image was performed (hardware not changed);
• Yes, if the computer (or the hard drive) was replaced.
Figure 401.
IH-500 | 331
Checks and Adjustments
Figure 402.
332 | IH-500
Tools Menu
Figure 403.
b Create
a folder on the USB memory key to avoid some antivirus warning (An “autorun.ini”
file is created during this action).
Figure 404.
K Select:
• No, if a Acronis backup image was performed (hardware not changed);
• Yes, if the computer (or the hard drive) was replaced.
IH-500 | 333
Checks and Adjustments
Figure 405.
Figure 406.
Q Plug the USB memory key on the PC with the internet connection.
334 | IH-500
Tools Menu
Y Close the license manager application. Unplug the USB memory key.
Figure 407.
Figure 408.
B Validate.
IH-500 | 335
Checks and Adjustments
5.14.3 Status
This screen shows the detailed state of each module.
Figure 409.
Figure 410.
k Inarm,the6example
above, there are 47 of 50 validate adjustments and 3 invalidate on the transport
of 7 validate adjustments and 1 invalidate on the vision station, as well 0 of 1 adjustment
invalidate on the auto-check.
336 | IH-500
Tools Menu
C Confirm.
YES
The green check mark is replaced by a red cross mark for the specified
statuses adjustments, and the number of validated status is refreshed.
Figure 411.
k Itarrow.
is possible to see the list of statuses that are getting invalidated by selecting the Details
k Ifinvalidated.
an adjustment is invalidated on any module, the status of QCFinalCheckValidated is
IH-500 | 337
Checks and Adjustments
4 5
Figure 412.
Figure 413.
338 | IH-500
Tools Menu
IH-500 | 339
Checks and Adjustments
5.14.5 Modules
This screen allows to activate or to deactivate modules on the instrument.
Figure 414.
340 | IH-500
Tools Menu
5.14.6 IP Config
This screen allows:
• to check the communication between the computer, the DMS and the robot;
• to set up the IP address for the computer and the transport arm.
2 3
Figure 415.
k User in Admin mode required. Refer to Software Update / Admin Mode on page 321.
A Enter the new IP address in the field.
b An incorrect IP address input for the transport arm may cause an permanent loss of
communication.
IH-500 | 341
Checks and Adjustments
Figure 416.
C Confirm.
OK
The IP address is updated.
The instrument prompts to restart the instrument.
Figure 417.
A RESTORE DEFAULT
Select to restore the factory configuration.
CONFIG
A warning message appears.
Figure 418.
B Confirm.
OK
The IP address is set to factory configuration.
The instrument prompts to restart the instrument.
Figure 419.
342 | IH-500
Tools Menu
k controller display.
Requires the Emergency IP address of the transport arm which is accessible on the robot
C Unplug the network cable the transport arm controller port J205.
H When done, plug back the ethernet cable from J204 to J205.
IH-500 | 343
Checks and Adjustments
5.15 Exit
This menu allows to:
• exit to Software menu;
• switch to exploitation mode (IH-500 user software);
• shut down the instrument.
Exit Options
2
3
4
Figure 420.
344 | IH-500
6 Maintenance
This chapter describes the preventive maintenance tasks and the replacement of defective parts.
The maintenance table contains all the maintenance procedures performed as well as the inspections
and replacements. A copy of the original report sheet can be found in the Appendix.
See chapter Maintenance Table on page 420.
k Every
time a maintenance operation is performed on the IH-500, both for a preventive
maintenance task and for the defective part replacement, the maintenance table must be filled
in.
6.1 Warnings
b Leaving the IH-500 powered on with its cover removed is dangerous and should be
ELECTROCUTION
avoided During such circumstances, the instrument should not be left unattended.
b to switch off the instrument (except when connecting the transport arm touch probe).
Before any operation requiring an electrical connection or disconnection, it is necessary
Always disconnect the power cable before accessing internal parts of the instrument. A
contact with live conductors may cause an electrocution.
Always perform a backup before any intervention on the instrument. See chapter Backup
on page 119.
See:
• Cleaning and Decontamination Instructions on page 348;
• Decontamination Certificate on page 418.
IH-500 | 345
Maintenance
Replacement
Cleaning
346 | IH-500
Periodical Maintenance
Functional control
Decontamination
Visual control
Replacement
Cleaning
Preventive maintenance Chapter or Procedure number
Gel card piercer / Ionizer
Gel card piercing tip X X 6.5.1 on page 348
Gel card piercing tip X 6.7.2 on page 363
Gel card piercing tip X 6.8.1 on page 365
Ionizer X 6.8.2 on page 365
Diluents piercing
Diluent piercing tip X X 6.5.1 on page 348
Diluent piercing tip X 6.7.2 on page 363
Diluent piercing tip X 6.8.1 on page 365
Pipetting area
Reagent holder spring X 6.8.3 on page 366
Sample tube holder spring V2 X 6.8.4 on page 366
Washing station X 6.6.3 on page 353
Needle localization hole X X 6.6.5 on page 354
Sampling needle X 6.6.12 on page 361
Sampling needle X IH-500 RP-0595004
Diluent locking device O-ring X IH-500 RP-12006920
Level detection cable X IH-500 RP-0595048
Faradized vein X IH-500 RP-0595129
Liquid container area and fluidic system
Intermediate tank area X X 6.3 on page 348
Liquid container area X X 6.6.3 on page 353
Decontamination liquid container X 6.6.13 on page 362
TU10, TU11, TU12 & TU13 liquid waste tubing X 6.7.4 on page 364
Quick connections X 6.7.4 on page 364
Intermediate tank X IH-500 RP-0595116
Fluidic filter 25 μm X IH-500 RP-0595058
Tube connection liquid waste X IH-500 RP-0595125
Tube connection system liquid X IH-500 RP-0595126
Centrifuges module with cooling
Rubber seal X 6.7.3 on page 364
Centrifuge X 6.6.6 on page 355
Centrifuge X 6.7.3 on page 364
Consumable drawers
Left drawer X 6.8.6 on page 367
Right drawer X 6.8.6 on page 367
Fan
Computer X 6.6.8 on page 358
Power supply X 6.6.9 on page 359
Reagent cooling X 6.6.9 on page 359
Transport arm controller X 6.6.10 on page 360
IH-500 | 347
Maintenance
Do not use other cleaning agents or products before consulting Bio-Rad and receiving
their agreement.
k Clean the housing and the doors with a lint-free cloth soaked in 70% ethanol.
6.4 Cleaning
6.4.1 Touch Screen, Doors and Cladding
b FIRE
AND EXPLOSION
Ethanol is flammable and can lead to explosions if used improperly.
Proper laboratory safety precautions must be respected.
Never use acetone, as it will damage the instrument.
k Clean the housing, the doors and the touch screen with a lint-free cloth soaked in 70% ethanol.
6.5 Decontamination
6.5.1 Suitable Products
b FIRE
AND EXPLOSION
Ethanol is flammable and can lead to explosions if used improperly. Proper laboratory
safety precautions must be respected.
b with electronics.
The disinfectant can influence the performance of the instrument if it comes into contact
348 | IH-500
Decontamination
A Apply the suitable disinfectant either directly on the part or the area, or directly on
a lint-free cloth.
B Spray and clean the area to be treated with the lint-free cloth.
Figure 421.
C Do not rinse.
k Alltreated
parts of the instrument that come into contact with potentially infectious material must be
as potentially infectious.
• Doors, windows,
• Touch screen,
• Hand-held barcode scanner.
• Left drawer,
• Right drawer,
• Removable reagents tray,
• Full solid waste sensor.
IH-500 | 349
Maintenance
• Quick-connections,
• Washing station with protection cover,
• Overflow area.
b Verify
that at the end of the cleaning the
internal storage is completely screwed.
• Complete gripper.
350 | IH-500
Decontamination
• Protective cover,
• Samples barcode reader
• Incubator.
IH-500 | 351
Maintenance
D Use a dry air spray to remove dusts and particles susceptible to damage the lens.
Figure 422.
352 | IH-500
Specific Cleaning Instructions
6.6.2 Backlight
b backlight.
Do not use a lint-free cloth soaked in 70% ethanol or disinfectant, it could damage the
A Use a computer suitable cleaning product with a lint-free cloth to clean the
backlight.
Figure 423.
Figure 424.
IH-500 | 353
Maintenance
Figure 425.
B Use demineralized water with a lint-free cloth to perform a cleaning of the needle
localization hole.
For decontamination, refer to section Decontamination on page 348.
Figure 426.
354 | IH-500
Specific Cleaning Instructions
6.6.6 Centrifuge
b must be checked.
After intervention in centrifuges, the correct locking of centrifuges covers and trapdoors
A Use a dry air spray to remove dusts and particles susceptible to create an
abnormal working of the centrifuges.
B Use demineralized water with a lint-free cloth to perform a cleaning of the vat.
Figure 427.
IH-500 | 355
Maintenance
6.6.7 Gripper
A Check the position of the gripper before the gripper legs removal:
• right leg (2) flush to the gripper;
• left leg (1) with a tooth to the edge of the gripper.
b Do
not move the pinions when cleaning internally the gripper.
Lock the pinion with one gripper leg:
• use the left leg (5) to block the pinion and clean the rear groove;
• use the right leg (6) to block the pinion and clean the front groove.
F Check that the position of the legs is identical as the one checked on step A.
356 | IH-500
Specific Cleaning Instructions
1 flush
1 1 tooth
4
6
Figure 428.
IH-500 | 357
Maintenance
6.6.8 Computer
A Shutdown and switch off the instrument.
Figure 429.
358 | IH-500
Specific Cleaning Instructions
C Clean the fan grid and the entire reagent cooling module.
Figure 430.
IH-500 | 359
Maintenance
Figure 431.
360 | IH-500
Specific Cleaning Instructions
Figure 432.
Figure 433.
IH-500 | 361
Maintenance
Figure 434.
362 | IH-500
Visual Checks
Figure 435.
C Replace if necessary.
Figure 436.
IH-500 | 363
Maintenance
b must be checked.
After intervention in centrifuges, the correct locking of centrifuges covers and trapdoors
A Check that the rubber seals of the rear and front centrifuges are not damaged.
B Replace if necessary.
C Check that there is no space between the vat and the cover.
Figure 437.
6.7.4 TU10, TU11, TU12 & TU13 Liquid Waste Tubing, Quick Connections
A Check that the Tygon tubing and the quick connections are not damaged (cracks),
and that there is no white tracks or residue.
B Check that tubings are still transparent and have not become discolored or
opaque.
C Replace if necessary.
Figure 438.
364 | IH-500
Functional Checks
B Replace if necessary.
Figure 439.
6.8.2 Ionizer
A Check the ionizer,
See chapter Ionizer on page 150.
B Replace if necessary.
Figure 440.
IH-500 | 365
Maintenance
E Replace if necessary.
Figure 441.
D Replace if necessary.
Figure 442.
366 | IH-500
Functional Checks
B Check if there is no backlash in the sliding system. When the drawers are open,
check for lateral backlash.
Figure 443.
IH-500 | 367
Maintenance
6.8.7 Camera
Perform a complete camera pre-adjustment.
See chapter Camera Pre-Adjustment on page 164.
368 | IH-500
Replacement Procedures
6.10 Disassembly
Only the technical personnel is authorized to disassemble an IH-500.
k after obtaining the manufacturer’s authorization. The objectives of the future application must be
The parts and assemblies of the IH-500, disassembled by qualified personnel, may be reused
described in details.
E Remove the 2 screws and washers (2) outside the instrument on the left of the
panel.
F Remove the screw, the washer (3) and the spacer (4) outside the instrument on the
right of the panel.
2 3
Figure 444.
IH-500 | 369
Maintenance
2
1
Figure 445.
370 | IH-500
Disassembly
F Remove the screw, the washer (2) and the spacer (3) outside the instrument.
On the left of the panel.
Figure 446.
IH-500 | 371
Maintenance
An important pressure may slightly move the centrifuges module. This may cause errors
during transfers.
B Remove the screw and washer (2) located on the right of the panel.
Figure 447.
372 | IH-500
Disassembly
Figure 448.
IH-500 | 373
Maintenance
2
Figure 449.
374 | IH-500
7 Troubleshooting
b Make
sure to have read and understood this documentation before carrying out any
troubleshooting on the IH-500. In case of unclear information, please contact the
distributor.
Then select OK/acknowledge and execute the required procedure to solve the problem.
For more detailed description about error messages, refer to the document DC-CP-039A01-07.
IH-500 | 375
7.2 All Modules
376 | IH-500
Troubleshooting
ALL MODULES
Error code Error description Possible cause Corrective actions
0x018B2F During initialization, access door is (1) Front door opened (1) Check that the front door is correctly closed
opened (2) Sensor failure (2) In case of sensor failure, see error code 0x018B43.
0x018B38 During initialization, weekly (1) Weekly maintenance interrupted (1) After default correction, perform again weekly maintenance with user or service
maintenance is on undefined status after power switching software
or weekly maintenance canister in (2) Weekly maintenance canister (2a) Replace the canister with the system liquid container.
detected instead of the system detected (2b) Check the detection sensor for the canister.
liquid container
0x018B3C During initialization, needle change Needle change procedure interrupted After default correction, perform again needle change with user or service software
is in progress after power switching
0x018B3F During initialization, presence (1) Sensor failure (1) Check sensor (LED on the sensor OK) encoder
detection auto-test failure (2) Sensor not adjusted (2) If OK: Perform again detection sensor adjustment (see Resources Sensor
Localization on page 228)
(3) If not OK: Check wiring (gripper side) and change the sensor if necessary
0x018E2A During initialization, an embarked Shut down during clot procedure (1) In service mode, check there is no clot on the needle.
clot default is present (2) Clear default and perform fluidic priming.
(3) If the needle is clogged, change it
0x019003 The internal ambient temperature (1) High internal temperature (1) Internal temperature > 30°C
reach the early warning limit (30°C) (2) Temperature sensor failure (2a) Check temperature sensor calibration
(2b) Check temperature sensor wiring
(2c) Change the temperature sensor
0x019004 The internal ambient temperature (1) Too high internal temperature (1) Internal temperature > 31°C
reach the warning limit (31°C) (2) Temperature sensor failure (2a) Check temperature sensor calibration
(2b) Check temperature sensor wiring
(2c) Change the temperature sensor
0x019005 The external ambient temperature (1) High external temperature (1) External temperature > 28°C or check if there is no heat source just near the
reach the warning limit (28°C) (2) Temperature sensor failure sensor
(2a) Check temperature sensor calibration
(2b) Check temperature sensor wiring
(2c) Change the temperature sensor
0x019006 Difference between internal (1) Temperature sensor failure (1) Check external temperature: environmental conditions (opened windows,
temperature sensor and external (2) One of the sensor failed heating, air flow)
temperature sensor (> 10°C) (2a) Check temperature sensors calibration
(2b) Check temperature sensors wiring
(2c) Change the temperature sensor
0x019007 Internal ambient temperature too (1) External temperature too low (1) Check external temperature and environmental conditions of the device (open
low (18°C). (2) Malfunction of a temperature window, heating, air current, etc...)
sensor (2a) Check temperature sensor calibration
(2b) Check temperature sensor wiring
(2c) Change the temperature sensor
ALL MODULES
Error code Error description Possible cause Corrective actions
0x103001 APF checker error. The system APF Checker error. Process not Contact the Bio-Rad customer support. Send the snapshot for investigation.
detect an error between the APF compliant with the APF folder
and the process perfomed on the
instrument,
0x104B03 Card block detected as missing (1) Card block missing (1) Check that all card blocks are correctly installed.
during the instrument initialization. (2) Mapping sensor failure (2) Check the mapping sensor adjustment.
0x104C01 APF Checker failure. Error during APF Checker error. Process not Contact the Bio-Rad customer support. Send the snapshot for investigation.
the check between the APF and compliant with the APF folder
the dispense set used.
0x104D06 Expiry date of the reagent reached. Reagent vial expired. (1) Check the expiry date of the product.
(2) Check the date of the computer.
0x104D07 Expiry date of the diluent reached. Diluent rack expired. (1) Check the expiry date of the product.
(2) Check the date of the computer.
0x104D08 Expiry date of the gel card reached. Gel card expired. (1) Check the expiry date of the product.
(2) Check the date of the computer.
0x104D09 On board time of the reagent OBT of the reagent vial expired. Remove the product.
reached.
0x104D10 On board time of the gel card OBT of the gel card expired. Remove the product.
reached.
0x104D11 On board time of the diluent rack OBT of the diluent rack expired. Remove the product.
reached.
0x104D22 Test performed when the Test performed in service mode. Switch to exploitation mode.
instrument is not in exploitation
mode.
0x104F02 The maximum duration (10min) The card did not go to the incubator Check for failure on transport arm, pipettor and incubator.
between the pipetting and the (following an other error). For more details, contact the Bio-Rad customer support. Send the snapshot for
incubation is reached. investigation.
0x104F05 The maximum duration between 2 Pipetting issue. Check for failure on pipettor.
dispenses is reached. For more details, contact the Bio-Rad customer support. Send the snapshot for
investigation.
0x104F06 The maximum duration (60min) Instrument issue. Check for failure on the system.
between the end of the For more details, contact the Bio-Rad customer support. Send the snapshot for
centrifugation and the reading is investigation.
reached.
0x108B16 Diluent piercing pin found missing (1) Piercing pin removed by the user. (1) Check the presence of the piercing pin.
during the instrument initialization. (2) Gripper failure. (2) Check the gripping of the piercing pin, If necessary, redo the gripper calibration.
0x108B41 Software version are not correct Server, application or files not compliant Check the version on the service software.
after a check during the instrument with the current software version. Check the firmware and all applications.
IH-500 | 377
initialization.
All Modules
If necessary, install the software.
378 | IH-500
Troubleshooting
ALL MODULES
Error code Error description Possible cause Corrective actions
0x108C07 Adjustment status found missing Adjustment status not present. According the status, perform the correct adjustment and validate the status. Switch
after a check during the instrument to exploitation and try to initialize the system.
initialization.
0xmm0101 Error on electronic boards. Electronic boards failure. “mm” linked to the module:
0xmm0102 • 02: X pipettor axis;
0xmm0103 • 03: Y pipettor axis;
0xmm0104 • 04: Z pipettor axis;
0xmm0105 • 05: Front centrifuge;
0xmm0106 • 06: Rear centrifuge;
0xmm0107 • 07: I/O fluidic board;
0xmm0108
0xmm0109 Check the wiring, and the communication between the PC and the power supply.
0xmm010A Check all the fuse on the power supply.
0xmm010B If necessary, change the electronic board
0xmm010C
0xmm010D
0xmm010E
0xmm010F
0xmm0110
0xmm0111
0xmm0112
0xmm0113
0xmm0114
0xmm0115
0xmm0309
0xmm030D
0xmm030F
0x0B8904
7.3 Transport Arm
For more detailed description about error messages related to the transport arm, refer to the document IH-500 NI Transport Arm Troubleshooting.
TRANSPORT ARM
Error code Error description Possible cause Corrective actions
0x018010 Communication failure with the Connection failure (TCP/IP port). Start again the instrument and check communication between PC and transport arm.
0x018011 transport arm controller. Check connection between PC and transport arm (RJ45 cable).
0x018013 Transport arm controller failure. Contact Instrument support.
0x018014
0x018003
0x018103
0x01811C
0x018106 During empty movement, the (1) Contact with resource or object. (1) Check there is no contact with an object which could block moves.
0x018110 transport arm does not position (2) Drag overshoot of transport arm. (2) In case of frequent default:
properly. • Change the main power supply.
0x018B1A
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks).
• Analyze error.log file to find on which axis occurs the default. If necessary,
change power module linked to this (only in case of frequent default).
0x018106 During initialization, the transport (1) Contact with resource or object. (1) Check there is no contact with an object which could block moves.
0x018110 arm does not position properly. (2) Drag overshoot of transport arm. (2) In case of frequent default:
0x018B1B • Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks).
• Analyze error.log file to find on which axis occurs the default. If necessary,
change power module linked to this (only in case of frequent default).
0x018106 During resources detection, the (1) Contact with resource or object. (1) Check there is no contact with an object which could block moves.
0x018110 transport arm does not position (2) Drag overshoot of transport arm. (2) In case of frequent default:
properly. • Change the main power supply.
0x018B1F
• Check connexions on the DSI box (especially on the motor breaks)
• Analyse error.log file to find on which axis occurs the default. If necessary,
change power module linked to this (only in case of frequent default)
0x018116 Power failure on the transport arm. (1) Emergency stop. (1a) Check that instrument access are correctly closed (electromagnetic latch and
(1) Security chain: Impossibility to security sensor of the left door and LED board connector).
power transport arm. (1b) Check J100 connector on transport arm controller.
(1c) Check power function (24V_ST_A & 24V_ST_B). Change the main power if
necessary.
(1d) Contact Instrument support.
Transport Arm
IH-500 | 379
380 | IH-500
Troubleshooting
TRANSPORT ARM
Error code Error description Possible cause Corrective actions
0x018119 During movement with (1) Contact with resource or object: (1) Check there is no contact with an object which could block moves. If there is no
0x01811A consumable, sample transfer or secure posing outbreak. object and in case of frequent default, check gripper localization and perform
0x018B19 diluent reconstitution, the transport (2) Drag overshoot of transport arm. secure posing.
arm does not position properly. (2) In case of frequent default:
• Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks)
• Analyse error.log file to find on which axis occurs the default. If necessary,
change power module linked to this (only in case of frequent default)
0x01811B Error during the positioning of the (1) Contact with an object (1) Check that there is no contact with an object which could block moves. If there is
diluent lock (2) Positioning fault of the diluent rack no object and in case of frequent default, check gripper localization and perform
secure posing.
(3) Drag overshoot of transport arm
(2) Check the position of the diluent rack or if the diluent lock is not too hard
(3) In case of frequent default:
• Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks).
• Analyze error.log file to find on which axis occurs the default.If necessary,
change power module linked to this (only in case of frequent default).
0x01811D Position error of the transport arm (1) Contact with an object (1) Check that there is no contact with an object which could block moves. If there is
while moving a resource. (2) Drag overshoot of transport arm no object and in case of frequent default, check gripper localization and perform
secure posing.
(2) Check the position of the diluent rack or if the diluent lock is not too hard
(3) In case of frequent default:
• Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks).
• Analyze error.log file to find on which axis occurs the default.If necessary,
change power module linked to this (only in case of frequent default).
0x01811E Position error of the transport arm (1) Contact with an object (1) Check that there is no contact with an object which could block moves. If there is
while piercing a diluent rack. (2) Drag overshoot of transport arm no object and in case of frequent default, check gripper localization and perform
secure posing.
(2) Check the position of the diluent rack or if the diluent lock is not too hard
(3) In case of frequent default:
• Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks).
• Analyze error.log file to find on which axis occurs the default.If necessary,
change power module linked to this (only in case of frequent default).
TRANSPORT ARM
Error code Error description Possible cause Corrective actions
0x01811F Position error of the transport arm (1) Contact with an object (1) Check that there is no contact with an object which could block moves. If there is
when unloading the sample rack. (2) Drag overshoot of transport arm no object and in case of frequent default, check gripper localization and perform
secure posing.
(2) Check the position of the diluent rack or if the diluent lock is not too hard
(3) In case of frequent default:
• Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks).
• Analyze error.log file to find on which axis occurs the default.If necessary,
change power module linked to this (only in case of frequent default).
0x018120 Position error of the transfer robot (1) Contact with an object (1) Check that there is no contact with an object which could block moves. If there is
when performing a centrifuge to (2) Position error of the centrifuge trap no object and in case of frequent default, check gripper localization and perform
trap operation secure posing.
(3) Drag overshoot of transport arm
(2) Check the position of trap door and check if the movement of the trap is not too
hard.
(3) In case of frequent default:
• Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks).
• Analyze error.log file to find on which axis occurs the default. If necessary,
change power module linked to this (only in case of frequent default).
0x018121 Unable to open the gripper after a Hardware failure Check if the movement of the gripper is smooth or not.
0x018417 gripper control position/torque Clean the gripper. See chapter Gripper on page 356.
check failure
0x018B34
0x018130 Position error of the transport arm (1) Contact with an object (1) Check that there is no contact with an object which could block moves. If there is
while piercing a gel card. (2) Position error of the piercing pin no object and in case of frequent default, check gripper localization and perform
secure posing.
(3) Drag overshoot of transport arm
(2) Check the position of piercing pin
(3) In case of frequent default:
• Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially
on the motor breaks).
• Analyze error.log file to find on which axis occurs the default.If necessary,
change power module linked to this (only in case of frequent default).
0x018210 Timeout error during the access to Module failure in the shared area Error due to an issue on the pipetting area.
the shared area (with no resources Analyze previous default.
on the gripper)
0x018212 Unexpected object detected in the The system previously saved the state Blocking fault during the initialization.
gripper during the transport arm of the gripper. On the service software, acknowledge the error and switch on the exploitation mode.
initialization
Transport Arm
IH-500 | 381
0x018213 Timeout error during the access to Module failure in the shared area Error due to an issue on the pipetting area.
the shared area (with resources on Analyze previous default.
the gripper)
382 | IH-500
Troubleshooting
TRANSPORT ARM
Error code Error description Possible cause Corrective actions
0x018220 An unexpected diluent rack is Transport arm failure. On the service software, remove the diluent rack on the intermediate diluent position.
detected in the intermediate diluent Acknowledge the error.
position during the transport arm Initialize the transport arm.
initialization.
0x018224 The gripper does not close (1) Hardware failure (1) Check if the gripper is smooth. If not, clean the gripper according the procedure
correctly when picking a (2) Wrong picking position of the of the service manual.
consumable. consumable (2) Check the positioning of the consumable.
(3) Bad calibration (3) Redo the gripper calibration.
0x018226 Error sent by the transport arm Communication failure Check the communication.
0x018227 controller during the power For more details, contact the Bio-Rad customer support. Send the snapshot for
calibration of the transport arm investigation.
motor.
0x018410 Communication error between the Communication failure (1) Check the connection of the gripper spiral cable (if necessary, change the spiral
0x018411 transport arm controller and the cable).
0x01841B gripper. (2) Check the voltage on the gripper board (if LED is ON or not). If not, change the
gripper board and/or the gripper spiral cable.
(3) Check the connection of the gripper cable on the TA controller (connector J203)
0x018412 The gripper does not open Hardware failure Check if the movement of the gripper is smooth or not.
correctly when placing a Clean the gripper. See chapter Gripper on page 356.
consumable.
0x018412 The gripper does not open Hardware failure Check if the movement of the gripper is smooth or not.
correctly when placing a Clean the gripper. See chapter Gripper on page 356.
consumable in the solid waste.
0x018412 Position error of the gripper when Hardware failure Check if the movement of the gripper is smooth or not.
opening/closing a centrifuge Clean the gripper. See chapter Gripper on page 356.
trapdoor or diluent lock.
0x018413 Inserted diluent rack not detected (1) Diluent rack removed by the user (1) Check the mapping sensor adjustment, and ask if the user remove the diluent
during a pickup for identification. (2) Gripper failure rack.
(2) Check the gripping of the diluent rack. If necessary, redo the gripper calibration.
0x018413 Piercing tip not detected when (1) Piercing tip removed by the user (1) Check if the user removed the piercing pin.
0x018B16 picked by the gripper (clamping (2) Gripper failure (2) Check the gripping of the piercing pin. If necessary, redo the gripper calibration.
force)
0x018413 Resource (sample rack excluded) (1) Resource removed by the user (1) Check if the user removed the resources.
0x018B18 not detected when picked by the (2) Gripper failure (2) Check the gripping of the resources. If necessary, redo the gripper calibration.
transport arm.
0x018413 Sample rack not detected when (1) Resource removed by the user (1) Check the presence of the resource.
0x018B1D picked by the transport arm. (2) Gripper failure (2) Check the gripping of the resources. If necessary, redo the gripper calibration.
0x018413 Resource (sample rack excluded) (1) Resource not detected (1) Check the adjustment of the mapping sensor.
0x018B30 not detected when picked by the (2) Gripper failure (2) Check the gripping of the resources. If necessary, redo the gripper calibration.
transport arm, but detected by the
mapping sensor.
TRANSPORT ARM
Error code Error description Possible cause Corrective actions
0x018413 Card block detected by the (1) Resource not detected (1) Check the adjustment of the mapping sensor.
0x018B37 mapping sensor but not detected (2) Gripper failure (2) Check the gripping of the resources. If necessary, redo the gripper calibration.
by the gripper when picked by the
gripper.
0x018414 Excessive torque for the gripper to (1) Gripper failure (1) Check if the gripper is smooth. If not, clean the gripper. See chapter Gripper on
pick up a resource (value out of (2) Resource not correctly installed in page 356.
range). the instrument (2a) The consumable is not compliant about the specification.
Not applicable for the drawers. (2b) The resource is not correctly installed on the instrument.
(2c) Redo the calibration of the gripper
0x018414 Excessive torque for the gripper to (1) Gripper failure (1) Check if the gripper is smooth. If not, clean the gripper according the procedure
0x01841C pick up a resource from the (2) Resource not correctly installed in of the service manual
drawers (value out of range). the instrument (2a) The consumable is not compliant about the specification.
(2b) The resource is not correctly installed on the instrument.
(2c) Redo the calibration of the gripper
0x018415 Resource lost during the transfer. (1) Resource not correctly picked by (1) Check the gripper and the resources.
0x018B20 the gripper (2) Check if the gripper is smooth. If not, clean the gripper according the procedure
(2) Gripper failure of the service manual
0x018416 Gripper initialization failure (home (1) Effort not detected (1) Check if the gripper is smooth. If not, clean the gripper according the procedure
position not detected). (1) Motor control board failure of the service manual.
(2) Check the gripper board, and the connection between the motor and the board.
0x018418 Unable to open the gripper after a Gripper failure Check if the gripper is smooth. If not, clean the gripper according the procedure of
gripper failure to pick resources. the service manual.
0x018B32 Transport arm failure after placing a Hardware failure Check that there is no contact with an object which could block moves. If there is no
reagent vial. object and in case of frequent default, check the gripper localization and perform a
secure posing.
In case of frequent default:
• Change the main power supply.
• Check connexions on the Transport Arm Controller Interface box (especially on the
motor breaks).
• Analyze error.log file to find on which axis occurs the default.
If necessary, change power module linked to this (only in case of frequent default)
0x018B32 During initialization, the transport is The transport arm is not correctly With the service software:
in the shared area. initialized after a needle change or a • acknowledge the error;
clot detection.
• initialize the transport arm and then the pipetor.
0x104B04 During initialization, a gel card is After abnormal stop: (1) User error (check gel card presence in pipetting area)
detected in the pipetting or (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
preparation area. (2) Mapping default
Transport Arm
0x104B04 During pipetting area mapping after After abnormal stop: (1) User error (check gel card presence in pipetting area)
IH-500 | 383
access, a gel card is detected (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
present
(2) Mapping default
384 | IH-500
Troubleshooting
TRANSPORT ARM
Error code Error description Possible cause Corrective actions
0x104B04 During initialization, a reagent vial is After abnormal stop: (1) User error (check reagent vial presence in pipetting area)
detected in the pipetting area. (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
(2) Mapping default
0x104B04 During initialization, a diluent rack is After abnormal stop: (1) User error (check diluent rack presence in pipetting area)
detected in the pipetting area. (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
(2) Mapping default
0x104B04 During pipetting area mapping after After abnormal stop: (1) User error (check diluent rack presence in pipetting area)
access, a diluent rack is detected (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
present
(2) Mapping default
0x104B05 During pipetting area mapping after After abnormal stop: (1) User error (check gel card presence in pipetting area)
access, a gel card is detected no (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
more present
(2) Mapping default
0x104B05 During pipetting area mapping after After abnormal stop: (1) User error (check gel card presence in pipetting area)
access, a diluent rack is detected (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
no more present (2) Mapping default
0x104B05 During initialization, a diluent rack is After abnormal stop: (1) User error (check diluent rack presence in pipetting area)
detected no more present (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
(2) Mapping default
0x104B05 During initialization, a gel card is After abnormal stop: (1) User error (check gel card presence in pipetting area)
detected no more present in (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
pipetting or preparation area
(2) Mapping default
0x104B05 During initialization, a reagent is After abnormal stop: (1) User error (check reagent presence in reagent area)
detected no more present in (1) Manipulation during stop (2) In case of frequent default, perform mapping sensor adjustment
pipetting area (2) Mapping default
7.4 Centrifuge
CENTRIFUGE
Error code Error description Possible cause Corrective actions
0x014B04 During initialization of the (1) Resource left in instrument. (1) Check the absence of gel card in the centrifuges.
0x018B30 instrument, a gel card was (2) Presence detection sensor failure. (2) Check the adjustment of the presence detection sensor.
detected present in one centrifuge.
0x014B05 During initialization of the (1) Manual handling of the resources (1) Check the absence of gel card in the centrifuges.
0x018B30 instrument, a gel card was while instrument was off. (2) Check the adjustment of the presence detection sensor.
detected absent in one centrifuge. (2) Presence detection sensor failure.
CENTRIFUGE
Error code Error description Possible cause Corrective actions
0x018217 During position detection of the (1) Trapdoor absent - detection sticker (1) Check the integrity of the centrifuge trapdoor.
0x058217 trapdoor, the trapdoor is detected on the trapdoor absent. (2) Check the integrity of the gripper.
0x068217 NOT CLOSED. (2) Gripper is damaged. (3) Check the adjustment of the resource detection sensor.
(3) Failure of the resource detection
sensor.
0x01821E During position detection of the (1) Trapdoor absent - detection sticker (1) Check the integrity of the centrifuge trapdoor.
0x05821E trapdoor, the trapdoor is detected on the trapdoor absent. (2) Check the integrity of the gripper.
0x06821E NOT OPENED. (2) Gripper is damaged. (3) Check the adjustment of the resource detection sensor.
(3) Failure of the resource detection
sensor.
0x019008 The centrifuge temperature Centrifuge temperature regulation (1) Check the centrifuge temperature regulation.
reached warning value (out of failure. (2) Check the electrical connection on the centrifuge board and on the cooling
range 60.8°F (16°C) / 77°F (25°C)). centrifuge board.
(3) Change the centrifuge module if cooling is out of order.
0x019009 The centrifuge temperature is out of Centrifuge temperature regulation (1) Check the centrifuge temperature regulation.
range. failure. (2) Check the electrical connection on the centrifuge board and on the cooling
centrifuge board.
(3) Change the centrifuge module if cooling is out of order.
0x050404 Error during origin research on (1) Axis fault (hard spot). (1) Check the absence of hard spot on the centrifuge axis.
0x050411 centrifuge axis. (2) Interference between the (2) Check the absence of foreign body in the centrifuge, and check the absence of
0x050413 centrifuge head and an element contact between the centrifuge head and the centrifuge cover.
0x05FD07 (foreign body). (3) Check that the centrifuge head is correctly inserted.
0x05FD12 (3) Centrifuge head missing or not (4) Check the centrifuge board, the motor, the encoder, the electrical connection,
correctly inserted. and if necessary, replace the centrifuge module.
0x050404
(4) Motor, encoder, or electronic board (5) Check the presence and the integrity of the joint on the centrifuge axis.
0x060411
failure.
0x060413
(5) Joint on the centrifuge axis
0x06FD07 defective or absent.
0x06FD12
0x050411 Positioning error of the centrifuge (1) Axis fault (hard spot). (1) Check the absence of hard spot on the centrifuge axis.
0x050413 (static or dynamic) during motion (2) Interference between the (2) Check the absence of foreign body in the centrifuge, and check the absence of
before a gel card is put down. centrifuge head and an element contact between the centrifuge head and the centrifuge cover.
0x05FD07
(foreign body). (3) Check the centrifuge board, the motor, the encoder, the electrical connection,
0x05FD12
(3) Motor, encoder, or electronic board and if necessary, replace the centrifuge module.
0x060411
failure.
0x060413
0x06FD07
0x06FD12
IH-500 | 385
Centrifuge
386 | IH-500
Troubleshooting
CENTRIFUGE
Error code Error description Possible cause Corrective actions
0x050411 Positioning error of the centrifuge (1) Axis fault (hard spot). (1) Check the absence of hard spot on the centrifuge axis.
0x050413 (static or dynamic) during motion (2) Interference between the (2) Check the absence of foreign body in the centrifuge, and check the absence of
0x05FD07 before a gel card taking. centrifuge head and an element contact between the centrifuge head and the centrifuge cover.
0x05FD12 (foreign body). (3) Check the centrifuge board, the motor, the encoder, the electrical connection,
0x060411 (3) Motor, encoder, or electronic board and if necessary, replace the centrifuge module.
failure.
0x060413
0x06FD07
0x06FD12
0x050411 Positioning error of the centrifuge (1) Axis fault (hard spot) (1) Check the absence of hard spot on the centrifuge axis
0x050413 (static or dynamic) during (2) Interference between the (2) Check the absence of foreign body in the centrifuge, and check the absence of
0x05FD07 centrifugation ending. centrifuge head and an element contact between the centrifuge head and the centrifuge cover
0x05FD12 (foreign body) (3) Check the centrifuge board, the motor, the encoder, the electrical connection,
0x060411 (3) Motor, encoder, or electronic board and if necessary, replace the centrifuge module
failure (4) Check the presence and the integrity of the joint on the centrifuge axis
0x060413
(4) Joint on the centrifuge axis
0x06FD07
defective or absent
0x06FD12
0x050411 During initialization / resource (1) Axis fault (hard spot) (1) Check the absence of hard spot on the centrifuge axis
0x050413 detection, a positioning error (static (2) Interference between the (2) Check the absence of foreign body in the centrifuge, and check the absence of
or dynamic) occurred centrifuge head and an element contact between the centrifuge head and the centrifuge cover
0x05FD11
(foreign body) (3) Check the centrifuge board, the motor, the encoder, the electrical connection,
0x05FD12
(3) Motor, encoder, or electronic board and if necessary, replace the centrifuge module
0x060411
failure
0x060413
0x06FD11
0x06FD12
0x05FD04 Error on the control of the (1) Axis fault (hard spot) (1) Check the absence of hard spot on the centrifuge axis.
0x05FD09 centrifugation speed (potentially (2) Interference between the (2) Check the absence of foreign body in the centrifuge, and check the absence of
due to overload error on axis or due centrifuge head and an element contact between the centrifuge head and the centrifuge cover
0x06FD04
to a speed control error, or due to a (foreign body) (3) Check the centrifuge board, the motor, the encoder, the electrical connection,
0x06FD09 current limitation error)
(3) Motor, encoder, or electronic board and if necessary, replace the centrifuge module.
failure (4) Check the presence and the integrity of the joint on the centrifuge axis
(4) Joint on the centrifuge axis (5) The instrument should not be shocked, especially during centrifugation
defective or absent
(5) Shock on the instrument during
the centrifugation
7.5 Incubator
INCUBATOR
Error code Error description Possible cause Corrective actions
0x014301 The temperature of the incubator is The temperature of the incubator is out (1) Check the incubator temperature regulation
out of range (36.5 ± 2.7 °C) in the of range (2) Change the incubator module or the analytic board
well
0x014B04 During initialization of the (1) Manual handling of the resources (1) Check the absence of ID-card in the incubator
instrument, a gel card was while instrument was off. (2) Check the adjustment of the presence detection sensor
detected present in the incubator (2) Failure of the presence detection
sensor
0x014B05 During initialization of the (1) Manual handling of the resources (1) User error
instrument, a gel card was while instrument was off (2a) Check the adjustment of the presence detection sensor
detected absent in the incubator (2) Failure of the presence detection (2b) If fault often occurs, check and adjust the presence detection sensor, and check
sensor the integrity of the background under the incubator (background must be
uniform)
0x019001 Discordance between the values of Failure of the temperature Check the temperature sensors adjustment.
the temperature sensors of the measurement function (sensor, board, If temperatures are incoherent, replace the temperature sensor(s).
incubator (dispersion > 5,4°C etc.)
If the incubator does not warm up, replace the incubator or the electronic board
between the 2 values)
0x104F03 The maximum duration (10min) of Centrifuge not available (following Check the absence of failure on the centrifuge(s)
incubation at 37°C of the well of an another failure)
ID-card was reached (overshoot)
0x104F03 The maximum duration (5min) of (1) Centrifuge not available (following Check the absence of failure on the centrifuge(s)
incubation at ambient temperature another failure)
of the well of an ID-card was (2) Incubation control failure
reached (overshoot)
0x104F03 The maximum duration (10min) of (1) Centrifuge not available (following Check the absence of failure on the centrifuge(s)
incubation of the well of a gel card another failure)
without incubation was reached (2) Incubation control failure
(overshoot)
IH-500 | 387
Incubator
7.6 Pipetor
388 | IH-500
Troubleshooting
PIPETOR
Error code Error description Possible cause Corrective actions
0x018B21 Transport arm in default status still Transport arm failure in common area Analyze previous default
present on common area or
element lost in this area
0x018B32 Pipettor X, Y or Z position error (1) Mechanical failure (1) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
while the transport arm is in the (2) Driving failure mechanism
reagents protective position in the (2) Check motor driving and wiring
(3) Position control chain failure
shared area (3) Check encoder information returned with service software. Check encoder wiring
and cleaning of the optical scales
0x018E01 After default, pipetor Z axis doesn't (1) Mechanical failure (1) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
reach its clearance position static (2) Driving failure mechanism
error or drag overshoot of Z axis) (2) Check motor driving and wiring
(3) Position control chain failure
(3) Check encoder information returned with service software. Check encoder wiring
and cleaning of the optical scales
0x018E14 Needle localization or target Contact failure between needle and Clean correctly with demineralized water needle extremity and target area (and also
teaching invalid (pipetor out of target gauges used for teaching). Then perform again localization.
range)
0x018E25 During needle localization, teaching Contact default between needle and Clean correctly with demineralized water needle extremity and target area (and also
quantity for one point reaches the target used gauges for teaching). Then perform again localization
limit
0x018E26 Capacitive detection signal failure (1) Level detection failure (1a) Check connection between faradized vein and level detection wire (change them
during needle localization if necessary): move manually pipetor Z axis, the signal fluctuation must be low
and when touching the needle, the signal must be zero
(1b) Check the level detection board (change it if necessary)
(1c) If level detection signal is OK, change PDS applicative board and check wiring
between PDS applicative board and pipetor applicative board
0x020404 During X, Y or Z initialization, the (1) Mechanical failure (1) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
0x020411 transport arm doesn't reach its (2) Driving failure mechanism
position (static error or drag (2) Check motor driving and wiring
0x020413 (3) Position control chain failure
overshoot or origin sensor failure (3) Check encoder information returned with service software. Check encoder wiring
0x028E28 for X, Y or Z axis)
and cleaning of the optical scales
0x030404
0x030411
0x030413
0x038E28
0x040404
0x040411
0x040413
0X048E28
PIPETOR
Error code Error description Possible cause Corrective actions
0x020411 During X or Y movement, transport (1) Mechanical failure (1) Check there is no hard spot on pipetor axis
0x020413 arm doesn't reach its position (2) Driving failure (2) Check motor driving and wiring
0x028E28 (static error or drag overshoot) (3) Position control chain failure (3) Check encoder information returned with service software. Check encoder wiring
0x030411 and cleaning of the optical scales
0x030413
0x038E28
0x020411 Pipetor static error or drag (1) Mechanical failure (1) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
0x020413 overshoot during needle (2) Driving failure mechanism
localization (2) Check motor driving and wiring
0x030411 (3) Position control chain failure
0x030413 (4) Level detection failure (3) Check encoder information returned with service software. Check encoder wiring
0x040411 and cleaning of the optical scales
(4a) Check connection between faradized vein and level detection wire (change them
0x040413
if necessary): move manually pipetor Z axis, the signal fluctuation must be low
0x018E27 and when touching the needle, the signal must be zero
(4b) Check the level detection board (change it if necessary)
(4c) If level detection signal is OK, change PDS applicative board and check wiring
between PDS applicative board and pipetor applicative board
0x040411 Static error or drag overshoot of Z (1) Mechanical failure (1a) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
0x040413 axis during liquid detection or (2) Driving failure mechanism
during movement to sampling (1b) Check there is no caps on the tube
0x018E1E (3) Position control chain failure
position (1c) Check that the tube is conform to rack type
(2) Check motor driving and wiring
(3) Check encoder information returned with service software. Check encoder wiring
and cleaning of the optical scales
0x040411 During vial bottom detection, (1) Mechanical failure (1) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
0x040413 pipetor Z axis doesn't reach its (2) Driving failure mechanism
position (static error or drag (2) Check motor driving and wiring
0x048E28 (3) Position control chain failure
overshoot of Z axis out of bottom (3) Check encoder information returned with service software. Check encoder wiring
detection area)
and cleaning of the optical scales
0x040411 During movement, pipetor Z axis (1) Mechanical failure (1) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
0x040413 doesn't reach its position (static (2) Driving failure mechanism
error or drag overshoot of Z axis) (2) Check motor driving and wiring
0x048E28 (3) Position control chain failure
(3) Check encoder information returned with service software. Check encoder wiring
and cleaning of the optical scales
0x018E08 Bottom detection error Reagent vial bottom not detected by Perform the teaching of the reagent plate (the Z positions are not correctly set).
the pipetor
IH-500 | 389
Pipetor
7.7 Access to the Pipetting Area
390 | IH-500
Troubleshooting
ACCESS TO THE DRAWER / PIPETTING AREA
Error code Error description Possible cause Corrective actions
0x018B31 Reagent suspension stop time Agitation system failure Analyze previous defaults
exceeds the limit See defaults on reagent area:
• 0x080411 on page 390
• 0x080413 on page 390
• 0x088B22 on page 390
0x018B43 During opening driving, the (1) Blocked door (1) Check that the door is not blocked or closed to fast (wait 1s after opening)
pipetting right door is not detected (2) Driving failure (2) Check if the electromagnetic latch is driven correctly
as opened
(3) Sensor failure (3) Check sensor status with service software
(4) Check status change on the applicative board (driving and sensor status)
(5) Check electromagnetic latch wiring
(6) Change electromagnetic latch or applicative board
0x080404 Reagent plate initialization failure Agitation system failure: (1) Check there is no hard spot on the reagent motor axis
0x080411 (1) Mechanical default (2) Check motor, encoder or applicative board wiring. If necessary, change the
0x080413 (2) Electrical default complete module or the applicative board
0x088B22
0x080411 During reagent suspension, reagent Agitation system failure (1) Check there is no hard spot on the reagent area
0x080413 area position or speed is not (2) Check motor, encoder or applicative board wiring. If necessary, change the
conform complete module or the applicative board
0x088B2A
0x088B51
0x104B04 During initialization, a sample rack (1) Rack insertion. (1) User error (rack insertion whereas software didn't start)
is detected present (2) Rack insertion outside of the (2) Rack detection board failure. See error code 0x018B35 on page 401
process
0x108B2E During operation, the pipetor right Forced opening of the door (1) Door closed by user
door is detected opened (2) Front door electromagnetic latch failure
If OK, see error code 0x018B43 on page 390
0x108B42 During the clot resolution, the User validates door closure whereas it Close the front door
pipetting door is detected opened is detected opened If OK, see error code 0x018B43 on page 390
after pressing the button Next.
7.8 Ionizer
IONIZER
Error code Error description Possible cause Corrective actions
0x018B28 Ionizer failed before gel card Ionizer failure (1) Check tension on the extremity of the ionizer with the high voltage tester pen and
piercing check with service software the ionizer status
(2) Check ionizer wiring (between ionizer power and fluidic applicative board then
between ionizer power and the pin)
(3) Change the complete ionizer or the fluidic board
Ionizer
7.10 Fluidic
392 | IH-500
Troubleshooting
FLUIDIC
Error code Error description Possible cause Corrective actions
0x018E06 Level detection signal control failure Level detection failure (1) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2) Check level detection board (change it if necessary
(3) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E07 Level detection error (level lost not (1) Foam (1) Check there is no foam or bubble in the reagent
detected during level detection) (2) Level detection failure (2a) Check connection between faradized vein and level detection cable. Change
even after many tries them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2b) Check level detection board (change it if necessary)
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E0A During level detection before a (1) Empty sample tube (1a) Check there is liquid in the sample tube
sample tube sampling, the bottom (2) Level detection failure (1b) Check the sample bottom level for this rack and tube type. If necessary, adjust it.
level is reached (No level detection (2a) Check connection between faradized vein and level detection cable. Change
in the expected range)
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2b) Check level detection board (change it if necessary)
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E0C Filling of the decontamination well (1) Decontamination circuit empty (1a) Check function and wiring of decontamination well filling pump
failure (level detection height under (2) Level detection failure (1b) Check function and wiring of washing station solenoid valve
the limit)
(1c) Check the decontamination circuit (filter and tubings)
(1d) Level sensor for NaOH bottle failure
(2a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2b) Check level detection board (change it if necessary)
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E0E Sampling set default: (1) APF definition error APF definition default. Reload them.
supplementary volume not null (2) Software bug If it is not OK, send LOG files and pictures to instrument support to perform complete
expertise
0x018E0F Sampling set default: drying (1) APF definition error APF definition default. Reload them.
volume not null (2) Software bug If it is not OK, send LOG files and pictures to instrument support to perform complete
expertise
FLUIDIC
Error code Error description Possible cause Corrective actions
0x018E10 Unpriming control at the end of the (1) Bubble in reagent during sampling (1) Check blood quality (less than 5 days)
sampling failed (for reagent) (2) Level detection failure (2a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2b) Check level detection board (change it if necessary)
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E12 Not enough reagent during (1) Not enough reagent (1a) Check if there is liquid in the reagent
0x104B06 sampling, no sampling can be (2) Detection of the bottom before (1b) Check the reagent bottom level and adjust it if necessary
done (volume calculated during liquid detection (2) Check there is no hard spot on pipetor axis and no backlash in rack-and-pinion
level detection)
(3) Level detection failure mechanism
(3a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(3b) Check level detection board (change it if necessary)
(3c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E13 Not enough reagent during (1) Not enough reagent (1a) Check if there is liquid in the reagent
sampling, at least on sampling can (2) Detection of the bottom before (1b) Check the reagent bottom level and adjust it if necessary
be done (volume calculated during liquid detection (2) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
level detection) (3) Level detection failure mechanism
(3a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(3b) Check level detection board (change it if necessary)
(3c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E16 Unpriming control at the end of the (1) Embarked filament (1) Check blood quality (less than 5 days)
sampling failed (embarked filament) (2) Level detection failure (2a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2b) Check level detection board (change it if necessary)
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E17 Unpriming control failed at the end (1) Embarked filament then broken (1) Check blood quality (less than 5 days)
of the sampling (level lost in the during sampling (2a) Check connection between faradized vein and level detection cable. Change
height range - filament broken) (2) Level detection failure them if necessary: move manually pipetor Z axis, signal variation must be low
(3) Needle position default in the and if you touch the needle, signal must be equal to zero
sample tube (2b) Check level detection board (change it if necessary)
IH-500 | 393
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
Fluidic
394 | IH-500
Troubleshooting
FLUIDIC
Error code Error description Possible cause Corrective actions
0x018E18 Not enough liquid in the sample (1) Not enough reagent (1a) Check there is liquid in the sample tube
tube during sampling (volume (2) Level detection failure (1b) Check the sample tube bottom level and adjust it if necessary
calculated during level detection) (1c) In case of «low volume» and «pediatric» tubes, check bottom level adjustment
(see Pediatric / Low Volume Tube Adjustment on page 260).
(2a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2b) Check level detection board (change it if necessary)
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E19 Purge default of the overflow well (1) Purge pump and solenoid valve (1) Check function and wiring of washing station solenoid valve and purge pump
or the collect well (Liquid present in failure (2) Check that there is no pinched or clogged tubings or sediment int the tubing
the collect well during periodical (2) Clogged purge circuit after the washing station and the waste external tubing
control with level detection)
(3) If this error occurs again, check the position of the needle on the washing station.
If necessary, perform a Change Pipette Needle to adjust the position.
0x018E1B Out of range or missing level (1) Purge pump failure. (1) Check function and wiring of washing station pump
detection in the overflow well at the (2) Solenoid valve blocked in purge (2) Check function and wiring of washing station solenoid valve
end of rinsing or pre-washing position
0x018E1D Level lost detection error after (1) The needle is not centered in the (1) Check needle position in the decontamination well
many tries in the decontamination decontamination well (2a) Check connection between faradized vein and level detection cable. Change
well (2) Level detection failure them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2b) Check level detection board (change it if necessary)
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x018E23 Collect well filling error (level (1) Washing circuit flow too low (1) Check function and wiring of washing pump (check there is no foreign object in
detection height in the collect well (2) Washing station leak the circuit and change the washing pump if necessary)
under the limit, during periodical (2) Check function and wiring of washing station solenoid valve
control)
0x018E24 Purge of the decontamination well (1) Pump or solenoid valve default. (1a) Check function and wiring of washing station purge pump
error (with level detection) Purge circuit clogged (1b) Check function and wiring of washing station solenoid valve
(2) Level detection failure (1c) Check that there is no pinched or clogged tubings or sediment int the tubing
after the washing station and the waste external tubing
(1d) Check the purge circuit (filter and tubings)
(2a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(2b) Check level detection board (change it if necessary)
(2c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
FLUIDIC
Error code Error description Possible cause Corrective actions
0x018E29 Volume control failure during (1) Dispense error: In case of more than 3% of failed dispense:
dispense • drop under foil (1a) Check needle position in gel card (drop under foil detected or contact between
• foam dispense needle and foil)
• non dispense
(1b) Check there is no foam in the sample tubes (detection of foam dispense)
(1c) Deviation of the PDS calibration: change the PDS module (low additional volume)
0x018E2B During weekly maintenance, before (1) Liquid detection in the overflow vat (1a) Fluidic leak / bottle leak / washing station overflow / intermediate tank overflow
dispense in washing station, liquid (2) Leak detection sensor failure (1b) Washing station overflow due to waste circuit clogged
is detected in overflow vat
(1c) Bottle overflow (system or NaOH solution) during filling
(2) Check function and wiring of leak detection sensor
0x018E2C Not enough liquid in the reagent (1) Reagent vial empty (1a) Check that there is liquid in the reagent vial.
vial (2) Level detection failure (1b) Check the reagent bottom level and adjust it if necessary.
(1c) Check the connection between the faradized vein and the level detection cable:
• Replace them if necessary.
• Move manually the pipetor Z axis, the signal variation must be low.
• If you touch the needle, the signal must be equal to zero.
(2d) Check the level detection board (change it if necessary).
(2e) If the level detection signal is correct, change the PDS applicative board and
check the wiring between the PDS applicative board and the I/O board.
0x018E2D During the washing, the state of the Washing pump failure Check the washing pump, and if necessary change it.
intermediate tank sensor did not
change
0x018E2E Time between the error occurrence Response of the user too long. Clean or replace the needle.
and the acknowledgment of this
error is too long. The needle is no
longer usable (blood dried inside
the needle)
0x018E33 Not enough volume on the (1) Not enough volume on the sample (1) Check the volume on the tube with the user.
pediatric or low volume tube tube (2a) Check the level detection on the pediatric or low volume tube (in some case, the
(during sampling) (2) The volume on the sample is not level detection must be deactivated on these tubes).
detected (2b) Check the level detection signal:
• Check connection between the faradized vein and the level detection cable.
Change them if necessary: move manually the pipetor Z axis, the signal
variation must be low and if you touch the needle, the signal must be equal
to zero.
• Check level detection board (change it if necessary).
• If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x019101 Intermediate tank filling time-out (1) Filling flow for intermediate tank (1a) Check flow, if they are low, change the intermediate tank
(for complete filling or during too low (1b) Check filling pump flow, change it if necessary
priming or weekly maintenance,)
IH-500 | 395
(2) Level sensor failure (2) Check the level sensor in the intermediate tank
Fluidic
396 | IH-500
Troubleshooting
FLUIDIC
Error code Error description Possible cause Corrective actions
0x019102 Intermediate tank filling time-out (1) Filling flow for intermediate tank (1a) Check flow, if they are low, change the intermediate tank
(for continuous filling) too low (1b) Check filling pump flow, change it if necessary
(2) Level sensor failure (2) Check the level sensor in the intermediate tank
0x019103 One system solution bottle is out of Status OK if the bottle is not Check the system solution bottle (it has no magnet on the rear)
order (not connected or empty or connected, empty or not the correct Check function and wiring of presence level and connexion sensors. If necessary,
presence of a maintenance bottle) one change them and/or the fluidic board.
0x019104 Two system solution bottles are out Status OK if the bottle is not Check the system solution bottle (it has no magnet on the rear)
of order (not connected or empty connected, empty or not the correct Check function and wiring of presence level and connexion sensors. If necessary,
or presence of a maintenance one change them and/or the fluidic board.
bottle)
0x019105 On run, during priming, the low Status OK is the volume is less than 0.5 Check the filling of the decontamination bottle
level of the decontamination bottle liter If it's OK, check function and wiring of level sensor. If necessary, change it and/or the
is reached fluidic board.
0x019107 During weekly maintenance, no Status OK if the bottle is not Check the maintenance bottle (it has a magnet on the rear)
maintenance solution bottle is connected, empty or not the correct Check function and wiring of presence, level and connexion sensors. If necessary,
operational one change them and/or the fluidic board.
0x019110 Before pipetting, liquid is detected (1) Liquid detection in the overflow vat (1a) Fluidic leak / bottle leak / washing station overflow / intermediate tank overflow
in the overflow vat (2) Leak detection sensor failure (1b) Washing station overflow due to waste circuit clogged
(1c) Bottle overflow (system or NaOH solution) during filling
(2) Check function and wiring of leak detection sensor
0x019201 One waste bottle is not operational Status OK if the bottle is not Check the waste bottle (it has a magnet on the rear)
(not connected or not present or connected, empty or not the correct Check function and wiring of presence, level and connexion sensors. If necessary,
full) one change them and/or the fluidic board.
0x019202 Two waste bottles is not Status OK if the bottle is not Check the waste bottle (it has a magnet on the rear)
operational (not connected or not connected, empty or not the correct Check function and wiring of presence, level and connexion sensors. If necessary,
present or full) one change them and/or the fluidic board.
0x019203 During purge, one waste bottle is Status OK if the bottle is not Check the waste bottle (it has a magnet on the rear)
not operational (not connected or connected, empty or not the correct Check function and wiring of presence, level and connexion sensors. If necessary,
not present or full) one change them and/or the fluidic board.
0x0C0411 Liquid not detected during reagent (1) Vial is empty (1a) Check there is liquid in the reagent
0x018E0A evaluation (drag overshot during (2) Detection of the bottom before (1b) Check the reagent bottom level and adjust it if necessary
liquid detection or bottom level liquid detection (2) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
reached) (3) Level detection failure mechanism
(3a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(3b) Check level detection board (change it if necessary)
(3c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
FLUIDIC
Error code Error description Possible cause Corrective actions
0x0C0411 Liquid not detected during first (1) Vial is empty (1a) Check there is liquid in the reagent
0x018E0A reagent sampling (drag overshot (2) Detection of the bottom before (1b) Check the reagent bottom level and adjust it if necessary
during liquid detection or bottom liquid detection (2) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
level reached)
(3) Level detection failure mechanism
(3a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(3b) Check level detection board (change it if necessary)
(3c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x0C0411 During sampling except washing, (1) PDS pump or solenoid valve wiring (1) Check PDS pump or solenoid valve wiring.
0x0C0413 the PDS pump doesn't reach its default (2) Change PDS module.
position (static or dynamic error) (2) Particles in the pump
0x0C0411 During sampling except washing (1) PDS pump or solenoid valve wiring (1) Check PDS pump or solenoid valve wiring.
0x0C0413 and priming, the PDS pump default (2) Change PDS module.
doesn't reach its position (static or (2) Particles in the pump
dynamic error)
0x0C0411 During sampling or dispense with (1) PDS pump or solenoid valve wiring (1) Check PDS pump or solenoid valve wiring.
0x0C0413 washing, the PDS pump doesn't default (2) Change PDS module.
reach its position (static or dynamic (2) Particles in the pump
error)
0x0C0411 During dispense with priming or (1) PDS pump or solenoid valve wiring (1) Check PDS pump or solenoid valve wiring.
0x0C0413 draining or flushing, the PDS pump default (2) Change PDS module.
doesn't reach its position (static or (2) Particles in the pump
dynamic error)
0x0C0411 During weekly maintenance, on (1) PDS pump or solenoid valve wiring (1) Check wiring of PDS module and solenoid valve.
0x0C0413 sampling to fill washing station, the default (2) Change the PDS module.
PDS module pump doesn't reach (2) Particles in the pump
its position (static or dynamic error)
0x0C0411 Liquid non detected before reagent (1) Empty reagent (1a) Check there is liquid in the reagent
0x0C0413 sampling (drag overshot during (2) Reagent bottom level detected (1b) Check the reagent bottom level. If necessary, adjust it.
level detection or bottom level before liquid detection
0x018E0A (2) Check there is no hard spot on pipetor axis and no play in rack-and-pinion
reached)
(3) Level detection failure mechanism
(3a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(3b) Check level detection board (change it if necessary
(3c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x0C0411 During unblocking sequence, the (1) PDS pump or solenoid valve wiring (1) Check PDS pump or solenoid valve wiring
IH-500 | 397
0x0C0413 PDS pump doesn't reach its default (2) Change PDS module
position (static or dynamic error)
Fluidic
0x018E22 (2) Particles in the pump
398 | IH-500
Troubleshooting
FLUIDIC
Error code Error description Possible cause Corrective actions
0x0C0416 Overpressure of the PDS pump (1) Sampled liquid too viscous (1) Check blood quality (less than 5 days).
during gel card or diluent rack (2) Needle clogged (2) Check that needle or faradized vein are not clogged. Change them if necessary..
dispense (3) Pressure sensor failure (3) Control and/or change PDS module
0x0C0416 During weekly maintenance, PDS (1) Sampled liquid too viscous (1) Check liquid: no foreign object
0x018E1F module pump overpressure default (2) Needle clogged (2) Check that needle or faradized vein are not clogged. Change them if necessary
for dispense to fill washing station
(3) Pressure sensor failure (3) Control and/or change PDS module
or priming or purge
0x0C0416 Overpressure of the PDS module (1) Sampled liquid too viscous (1) Check liquid: no foreign object.
0x018E20 pump during dispense with priming (2) Needle clogged (2) Check that needle or faradized vein are not clogged. Change them if necessary.
or flushing or purge
(3) Pressure sensor failure (3) Control and/or change PDS module
0x0C0416 Overpressure of the PDS module (1) Sampled liquid too viscous (1) Check liquid: no foreign object.
0x018E21 pump during purge with service (2) Needle clogged (2) Check that needle or faradized vein are not clogged. Change them if necessary.
software.
(3) Pressure sensor failure (3) Control and/or change PDS module.
0x0C0416 Pressure sensor failure, pressure (1) Liquid to slimy (1) Check the liquid (not too old).
0x018E31 out of range during the sampling (2) Needle clogged (2) Check the needle.
(3) Issue with the pressure sensor (3) Change the PDS module.
0x0C0416 Overpressure of the PDS pump (1) Sampled liquid too viscous (1) Check there is no sediment in the washing fluidic circuit (change tubings and
0x0C8E1A during washing (2) Needle clogged intermediate tank if necessary).
(3) Pressure sensor failure (2) Check that needle or faradized vein are not clogged. Change them if necessary.
(3) Control and/or change PDS module.
0x0C4011 Positioning error of the PDS pump (1) Wiring issue of the PDS module (1) Check the wiring.
0x0C4013 (during titration) (2) Dust on the pump (2) Change the PDS Module.
0x018E2F
0x018E30
0x0CED07 Excessive negative pressure when (1) Viscosity of liquid out of range (1) Check blood quality (less than 5 days).
withdrawing liquid, except for (blood over 5 days) (2) Check the needle position on tube bottom (especially for «low volume»).
washing (air or liquid) (2) Needle too close to the bottom of (3a) Check that needle or faradized vein are not clogged. Change them if necessary
the sample tubes
(3b) Control and/or change PDS module
(3) Pressure sensor failure
0x0CED07 Excessive negative pressure when (1) Viscosity of liquid out of range (1) Check liquid: no foreign object.
0x0CED08 withdrawing decontamination (2) The needle is too close to bottom (2) Check needle position in the tube.
solution during washing (air or of the decontamination well
0x018E1C (3a) Check that needle or faradized vein are not clogged. Change them if necessary
liquid)
(3) Pressure sensor failure (3b) Control and/or change PDS module
0x0CED08 Hydraulic circuit clogged (not Clogged circuit (1) Check that the needle or the faradized vein are not clogged. Change them if
solved) without samples necessary.
(2) Check and/or the change PDS module.
0x0CED08 Hydraulic circuit clogged (not (1) Clot detection or out of range (1) Check the blood quality (less than 5 days).
0x018E15 solved) for samples liquid viscosity (2a) Check that needle or faradized vein are not clogged. Change them if necessary.
(2) Clogged circuit (2b) Control and/or change PDS module
FLUIDIC
Error code Error description Possible cause Corrective actions
0x0CED0B Pressure sensor default (out of (1) Unplugged sensor (1) Check wiring of PDS module, pressure sensor and PDS applicative board
range static pressure at the (2) Sensor damaged (2) If necessary, change the PDS applicative board and/or the PDS module
beginning of the sampling or no
signal variation at the beginning of
the sampling)
0x104D12 After reagent vial pipetting, there is Not enough volume on the reagent vial Insert a new reagent vial.
no enough volume to process the
next batch
0x108E2C Liquid volume estimated less than (1) Not enough liquid (1a) Check the presence of liquid in the bottle
1 dose (2) Detection of the bottom before (1b) Check the low rating bottle (if necessary, reset)
liquid detection (2) Verify the absence of hard spot on the Z axis and / or backlash (if necessary,
(3) Level detection failure repeat the adjustment procedure of the toothed rack)
(3a) Check connection between faradized vein and level detection cable. Change
them if necessary: move manually pipetor Z axis, signal variation must be low
and if you touch the needle, signal must be equal to zero
(3b) Check level detection board (change it if necessary)
(3c) If the level detection signal is correct, change the PDS applicative board and
check wiring between PDS applicative board and I/O board
0x109109 During instrument initialization, (1) Liquid detection in the overflow vat (1a) Fluidic leak / bottle leak / washing station overflow / intermediate tank overflow
liquid is detected in overflow vat (2) Leak detection sensor failure (1b) Washing station overflow due to waste circuit clogged
(1c) Bottle overflow (system or NaOH solution) during filling
(2) Check function and wiring of leak detection sensor
Internal Storage
IH-500 | 399
400 | IH-500
Troubleshooting
INTERNAL STORAGE
Error code Error description Possible cause Corrective actions
0x104B04 Expected gel card detected during (1) Mapping sensor failure. (1a) Check the adjustment of the mapping sensor.
the instrument initialization. (2) Gel card removed by the user. (1b) Check if the mapping works correctly on the internal storage, and check the
backplane (no dust and check the black sticker under the internal storage).
(2) Check the presence of the gel card.
0x104B05 Expected gel card not detected (1) Mapping sensor failure. (1a) Check the adjustment of the mapping sensor.
during the instrument initialization. (2) Gel card removed by the user. (1b) Check if the mapping works correctly on the internal storage, and check the
backplane (no dust and check the black sticker under the internal storage).
(2) Check the presence of the gel card.
Sample Rack
IH-500 | 401
0x108B3E A sample rack is missing Sample rack removed by user (1) Sample rack removed voluntary
(2) If the rack is present, rack detection board failure
See error code 0x018B35 on page 401
7.14 Software
402 | IH-500
Troubleshooting
SOFTWARE
Error code Error description Possible cause Corrective actions
0x018B01 Software exception. Unmanaged (1) Unmanaged error (1) Unmanaged default: The instrument stops after an error (check previous error)
default. (2) Bug software (2) Otherwise, contact instrument support to ask analyses
0x018B02 Sequence forbidden states: {0} The status of the instrument can not Check the status of the instrument.
allow the action. Acknowledge all errors.
0x018B03 Missing required states: {0} The status of the instrument can not Check the status of the instrument.
allow the action. Acknowledge all errors.
0x018B04 Not authorized states: {0} The status of the instrument can not Check the status of the instrument.
allow the action. Acknowledge all errors.
0x018B06 Position not adjusted Position not adjusted (status not Check and validate the adjustment.
validated)
0x018B07 Position not found Position not adjusted (status not Check and validate the adjustment.
validated)
0x018B09 Destination not allowed. Transfer by the transport arm not Use an other position to do the transfer.
possible (the trajectories do not exist).
0x018B0A Drawer open, impossible to start Drawer open during a transport arm Close and do not open the drawer during the transport arm transfer.
the sequence. transfer.
0x103002 APFChecker: The maximum Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
duration between the dispense and investigation.
the incubation is reached
(controlled at this end of the
process)
0x103003 APFChecker: The minimal duration Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
of the incubation is not reached investigation.
(controlled at this end of the
process)
0x103004 APFChecker: The maximum Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
duration between the incubation investigation.
and the centrifugation is reached
(controled at this end of the
process)
0x103005 APFChecker: The maximum Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
duration between the dispense and investigation.
the centrifugation is reached
(controled at this end of the
process)
0x103006 APFChecker: The maximum Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
waiting time on the centrifuge is investigation.
reached (controlled at this end of
the process)
SOFTWARE
Error code Error description Possible cause Corrective actions
0x103007 APFChecker: The minimal duration Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
of the centrifugation is not reached investigation.
(controlled at this end of the
process)
0x103008 The maximum time for an open well Process issue For more details, contact the Bio-Rad customer support. Send the snapshot for
is reached. investigation.
0x103009 APFChecker: An expired reagent Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
vial was used during the test investigation.
0x103010 APFChecker: An expired gel card Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
was used during the test investigation.
0x103011 APFChecker: An expired diluent Non-compliant APF For more details, contact the Bio-Rad customer support. Send the snapshot for
rack was used during the test investigation.
0x104921 Length of the sample bar code is Too long bar code Used conform bar code (less than 30 digits)
upper to parameters (30 digits)
0x104923 Image not found during the Image corrupted Check the integrity of the archive folder.
0x104924 transmission of the picture well
0x104D01 Test deleted (1) Test deleted with Stop all Test deleted done by user
0x104D02 (2) Test to repeat launch (first test
0x104D03 deleted with 4D02 and new test
created)
0x104D04
(3) Test deleted by DMS
(4) Test with missing resources
deleted by sample rack removing
0x104D10 On-board time for the gel card On-Board Time management Remove the gel card.
expired
0x104D11 On-board time for the diluent rack On-Board Time management Remove the diluent.
expired
0x104D13 QC flag disabled. QC flag disabled by the user. Enable the QC flag and start a new test.
0x104D14 The user sent the interpretation No issue. -
with the function “Send result to
DMS”.
0x104D15 The user disables the QC flag for No issue. -
QC.
0x104D16 On board time option disabled for No issue. -
gel cards
0x104D17 On board time option disabled for No issue. -
diluent racks
0x104D18 On board time option disabled for No issue. -
IH-500 | 403
reagents
Software
404 | IH-500
Troubleshooting
SOFTWARE
Error code Error description Possible cause Corrective actions
0x104D19 On board time status for the No issue. -
reagent not sent by the DMS, but
validated by the user.
0x104D20 On board time allowed for the No issue. -
reagent is reached, but validated by
the user.
0x104D21 On board time status for the No issue. -
reagent not sent by the DMS, but
validated by the user.
0x105001 Incorrect file name Non-compliant APF Non-compliant file. Start again with a new file
0x105002 Error during APF file import. File Non-compliant APF Non-compliant file. Start again with a new file
doesn't concern IH-500
0x105003 Error during APF file import. Non-compliant APF Non-compliant file. Start again with a new file
Incorrect file version
0x105004 Error during APF file import. File Non-compliant APF Non-compliant file. Start again with a new file
doesn't concern consumables
used on the instrument
0x105005 Error during APF file import. Same Non-compliant APF Non-compliant file. Start again with a new file
or lower version of the APF file still
imported
0x105006 Error during APF file import. Invalid Non-compliant APF Non-compliant file. Start again with a new file
XML file (corrupted file)
0x105007 Error during APF file import. Invalid Non-compliant APF Non-compliant file. Start again with a new file
data (corrupted file)
0x105009 Error during APF file import. Invalid Non-compliant APF Non-compliant file. Start again with a new file
checksum invalid (corrupted file)
0x105010 Error during APF file import. Data Non-compliant APF Non-compliant file. Start again with a new file
insert failed
0x105010 Error during APF file import. Non-compliant APF Non-compliant file. Start again with a new file
0x105011 Pipetting weighting are not
compatible with the data from the
instrument.
0x105012 Error during APF file import. APF incompatible Non-compliant file. Start again with a new file
Incompatible NPAC with generic
gel card
0x105014 Error during vision files import. Corrupted file Non-compliant file. Start again with a new file.
0x105015 Error during flag files import. Corrupted file Non-compliant file. Start again with a new file.
0x105016 Error during APF files import. Vision Non-compliant APF Non-compliant file. Start again with a new file.
parameters not found.
0x105017 Error during APF files import. Non-compliant APF Non-compliant file. Start again with a new file.
0x105018 External media used for a snapshot Media full. Check the external device
full.
SOFTWARE
Error code Error description Possible cause Corrective actions
0x105019 External media used for a snapshot Name of the external media not found Check the connection of the external device
not found.
0x105200 Error during translation files import. Corrupted file Non-compliant file. Start again with a new file.
0x105201 Error during translation files import. The file does not contain translation Non-compliant file. Start again with a new file.
data.
0x105202 Error during translation files import. Incorrect file version. Non-compliant file. Start again with a new file.
0x108B3D Weekly maintenance expired Weekly maintenance expired Perform the weekly maintenance.
Solid Waste
7.16 Vision Station
406 | IH-500
Troubleshooting
VISION STATION
Error code Error description Possible cause Corrective actions
0x018504 During first identification the gel Not conform inserted gel card Check that the gel card corresponds to an APF managed by the instrument
0x104902 card bar code is in incorrect format
/ type
0x01880B During image acquisition, the Camera image acquisition default (1) Check camera integrity and USB connexion between camera and PC
camera can not make it (black (2) If not OK, change the camera, the USB cable then the computer
picture or camera default) even
after many try and camera reset
0x018901 During sample bloc identification (1) Wrong bar code orientation (1) Check sample tube orientation in the bloc
many different codes are read on (2) Bad sticker (2) Check integrity and quality of the bar code used (see specifications in user
the same sample tube place manual)
(3) Unknown bar code type used
(4) Bar code reader failure (3) Check that bar code is supported by the reader (see specification in the user
manual)
0x018902 During sample bloc identification, (1) Unknown bar code type used (1) Check presence, order and integrity of the separation stickers on the sample
sample tube type of the sample (2) Missing or damaged separation bloc
bloc is unreadable or unknown stickers (2) Check that used bar code is conform to user manual specification
(3) Bar code reader failure (3) Check bar code reader (wiring, clean)
0x018903 During sample bloc identification, (1) Missing or damaged separation (1) Check presence, order and integrity of the separation stickers on the sample
presence or order of the separators sticker bloc
is not conform (2) Bar code reader failure (2) Check bar code reader (wiring, clean,...)
0x018B52 Can not perform the sample Settings for low volume or pediatric Adjust the position for the low volume tubes and for the pediatric tubes.
identification on rack type 2 or 6. tubes not defined.
0x018F04 During air gap control, gel card bar (1) Gel card moved manually during (1) User error
code is different to the one process (2) Send LOG files and pictures to instrument support to perform complete
expected (2) Software bug expertise
0x018F04 During well reading, gel card bar (1) Gel card moved manually during (1) User error
code is different to the one process (2) Send LOG files and pictures to instrument support to perform complete
expected (2) Software bug expertise
0x018F0C During recalibration, a problem is (1) An object is present (1) Check there is no dust or spot on the backlight or the lens of the camera
0x018F0E detected on the lighting (back (2) Lighting failure (2) Check backlight homogeneity (LED out of order)
ground brightness average or CV (3) Camera or lens failure (3) Check Camera clearness and aperture adjustment
out of range or reference
brightness out of range)
0x018F0D During first identification the gel (1) Labeling faulty (1) Check gel card sticker
card bar code is unreadable (2) Camera clearness adjustment (2) Check Camera clearness adjustment
default (3) If the gel card vibrates, it could be a bad gel card taking into the store (the card
(3) Significant vibration during slides into the legs of the gripper). Check store adjustments and clean the
identification gripper legs
VISION STATION
Error code Error description Possible cause Corrective actions
0x018F0D During the air gap control, the (1) Labeling faulty (1) Check gel card sticker
barcode of the gel card is illegible (2) Camera clearness adjustment (2) Check Camera clearness adjustment
(even after the second attempt) default (3) If the gel card vibrates, it could be a bad gel card taking into the store (the card
(3) Significant vibration during slides into the legs of the gripper). Check store adjustments and clean the
identification gripper legs
0x018F0D During the interpretation of the (1) Labeling faulty (1) Check gel card sticker
wells, the barcode of the gel card is (2) Camera clearness adjustment (2) Check Camera clearness adjustment
illegible (even after the second default (3) If the gel card vibrates, it could be a bad gel card taking into the store (the card
attempt)
(3) Significant vibration during slides into the legs of the gripper). Check store adjustments and clean the
identification gripper legs
0x018F10 During recalibration, lighting is (1) Lens or backlight dusty (1) Check there is no dust or spot on the backlight or the lens of the camera
0x018F14 found failed (Exposure out of range (2) Backlight failure (2) Check backlight homogeneity (LED out of order)
or brightness gap of one of the
0x018803 (3) Camera failure (3) Check Camera clearness and aperture adjustment
color channels or impossible
0x018804 background management)
0x018F11 During vision station adjustment, (1) Camera not enough adjusted (1a) Check there is no dust or spot on the backlight or the lens of the camera
0x018805 white balance adjustment is (2) Lighting failure (1b) Check there is no external lighting source which could disturb the vision station
impossible (expected exposure out
0x018806 (3) Camera or lens failure (1c) Check that adjustments are still correct (thread sealing present)
of range or impossible white
(2) Check backlight homogeneity (LED out of order)
balance management)
(3) Check Camera clearness and aperture adjustment
0x018F12 During vision station adjustment, (1) Camera not enough adjusted (1a) Check there is no dust or spot on the backlight or the lens of the camera
0x018808 exposure time adjustment is (2) Lighting failure (1b) Check if no external lighting source could disturb the vision station
impossible (expected exposure out
0x018809 (3) Camera or lens failure (1c) Check that adjustments are still correct (thread sealing present)
of range or impossible white
balance management) (2) Check backlight homogeneity (LED out of order)
(3) Check Camera clearness and aperture adjustment
0x018F13 During vision station adjustment, (1) Camera not enough adjusted (1a) Check there is no dust or spot on the backlight or the lens of the camera
0x018F15 lighting control failed (back ground (2) Lighting failure (1b) Check if no external lighting source could disturb the vision station
brightness or expected exposure (3) Camera or lens failure (1c) Check that adjustments are still correct (thread sealing present)
out of range)
(2) Check backlight homogeneity (LED out of order)
(3) Check Camera clearness and aperture adjustment
0x0B8807 Camera communication failure Camera communication failure (1) Check camera integrity and USB connexion between camera and PC
during initialization or run (2) Check that camera is found by windows
If not OK, change the camera, the USB cable then the computer
0x104901 During first identification the gel Out of date gel card Check that the expiry date is coherent, If yes, check date configuration of the
card is out of date computer
0x104903 A cap is detected on a sample tube (1) Cap present on a sample tube (1) Check there is no cap on the sample tube
(2) Cap detection function failure (2) Check top of sample tube (no default, dried blood or sticker on the top)
Vision Station
See user manual
IH-500 | 407
408 | IH-500
Troubleshooting
VISION STATION
Error code Error description Possible cause Corrective actions
0x104904 During sample bloc identification, a (1) Wrong bar code orientation (1) Check sample tube orientation in the bloc
sample tube bar code is (2) Bad sticker (2) Check integrity and quality of the bar code used (see specifications in user
unreadable (3) Unknown bar code type used manual)
(4) Bar code reader failure (3) Check that bar code is supported by the reader (see specification in the user
manual)
(4) Check wiring between bar code reader and applicative board
(5) Change the bar code reader
0x104907 A cap is detected on a reagent vial (1) Cap present on a reagent vial (1) Check there is no cap on the reagent vial
(2) Cap detection function failure (2) Check the reagent vial neck (no default)
See user manual
0x104908 During reagent vial identification, its (1) Wrong orientation of the reagent (1) Check vial position in the drawer
bar code is unreadable (even after vial (2) Check vial sticker integrity
2nd try) (2) Bad sticker (3) Check Camera clearness adjustment
(3) Camera clearness adjustment (4) External vibration
default
(4) Significant vibration during
identification
0x104909 During diluent rack identification, its (1) Bad sticker (1) Check vial sticker integrity
bar code is unreadable (even after (2) Camera clearness adjustment (2) Check Camera clearness adjustment
2nd try) default (3) External vibration
(3) Significant vibration during
identification
0x104910 During first identification the diluent Out of date diluent rack (1) Check that the expiry date is coherent,
rack is out of date (2) Check date configuration of the computer
0x104910 During instrument initialization, the Expiry date reached (1) Check that the expiry date is coherent,
diluent rack is detected out of date (2) If yes, check date configuration of the computer
0x104911 During first identification the Out of date reagent (1) Check that the expiry date is coherent,
reagent is out of date (2) Check date configuration of the computer
0x104912 During gel control, a well is (1) Still used well (1) Check that the well is empty
detected not empty (2) Foreign object in the gel or on the (2) Check there is no foreign object (dust in or out of the well)
card
0x104913 During first identification the Not conform inserted reagent Check that the reagent corresponds to an APF managed by the instrument
reagent is in incorrect format / type
0x104914 During first identification the diluent Not conform inserted diluent rack Check that the diluent rack corresponds to an APF managed by the instrument
rack is in incorrect format / type
0x104915 During start, vision station is not Calibration not performed With service software, check the camera calibration (see Vision Station Menu on
calibrated page 163 in the service manual)
0x104917 During identification after an Sample bloc inversion by user Forbidden action during access to the pipetting area
access, a sample bloc is detected
different from its definition before
access
VISION STATION
Error code Error description Possible cause Corrective actions
0x104918 When identifying a gel card, this Gel card already used. (1) Check the gel card barcode.
one is detected as already used in (2) Discard the gel card if it was used,
the instrument (This case includes
balancing cards. The software
must return the gel card if this one
has already been used as a
balancing card in the instrument.).
0x104919 During diluent rack first Still used diluent rack User error. Still used diluent rack insertion forbidden
identification, the rack is detected
still used in the instrument
0x104920 During identification after an Rack inversion by user Forbidden action during access to the pipetting area
access, a diluent rack is detected
different from its definition before
access
0x104920 During instrument initialization, the Manipulation during stop Forbidden action when the instrument is stopped
diluent rack identification is
detected different from its definition
before stop
0x104920 During instrument initialization, the (1) Damaged sticker (1) Check integrity of the bar code used
bar code diluent rack is unreadable (2) Reading default on the vision (2) Check clearness adjustment and check there is no dust or spot on the backlight
(even after the 2nd try) station or the lens of the camera
0x104925 Gel control not activated. Gel control disabled by the user. -
0x104926 Structure of the reagent vial Incorrect or unknown barcode (1) Check the vial barcode.
barcode incorrect. (2) Check if the corresponding APF is correctly loaded.
0x104B05 During identification after an Rack inversion, adding, or removing by Forbidden action during access to the pipetting area
0x104B04 access, a sample tube is detected user
0x104917 different from its definition before
access or no more present or new
present
0xmm030F During sample bloc identification, Bar code reader failure (1) Check wiring between bar code reader and applicative board
0x018904 communication with bar code (2) Change the bar code reader
reader failed
Vision Station
IH-500 | 409
Troubleshooting
410 | IH-500
8 Spare Parts List and
Accessories
The content of this chapter was extracted to create a specific and separate manual.
The spare parts manual describes in detail all available spare parts and their location in the instrument.
See Spare Parts Book H009345.
IH-500 | 411
Spare Parts List and Accessories
412 | IH-500
9 Decommissioning, Storage and
Transport
This chapter describes the measures to take in order to decommission a IH-500, store it and dispose
of it.
For decommissioning, short term or long term storage, clean and purge the hydraulic circuit.
See chapter Prepare for Shipment on page 322.
9.1 Decommissioning
In a general manner, the IH-500 is taken out of operation before being moved, repaired or stored for a
further use.
b Indecontaminate
order to prevent any health hazard during the procedure, it is required to
the IH-500.
C Proceed in the reverse order of the chapter Installation on page 89 to store all the
accessories.
See Equipment and Accessories on page 417 to find what is equipment and
what is optional accessories.
See Installation on page 89 to follow the procedure in reverse order.
D Proceed in the reverse order of the section Blocking Device Removal on page 112
to block the 3 axes.
IH-500 | 413
Decommissioning, Storage and Transport
9.2 Storage
When it is planned not to use the instrument for a prolonged period, specific operations must be
considered.
Conditions:
• drawers and doors closed.
• temperature: 18°C (64.4°F) to 28°C (82.4°F);
• relative humidity: 5 to 80% (without condensation);
• protected against direct sunlight and dust.
b not used!
Weekly maintenance operations must be performed every week even if the instrument is
9.3 Packing
Packing is essential before shipping or when the IH-500 must be stored outside laboratory conditions
or for a long term storage.
b frame and the 2 crossbars, and not with the instrument’s 4 feet.
Make sure that the contact between the IH-500 and the pallet is done with the instrument
414 | IH-500
Disposal
b instrument.
Make sure that the 2 packing flange are in a correct position and maintains correctly the
B Prepare and place the wrapping elements on each side of the IH-500.
C Proceed in the reverse order of the section Installation on page 89 to place the
IH-500 on the pallet and to assemble the different panels.
D Place the IH-500 in the crate box. Over it, place the intermediate cardboard
element, the cables and all the technical documentation.
9.4 Disposal
This chapter describes the appropriate disposal of the IH-500.
Prerequisites
b The
IH-500 must be decontaminated before disposal.
See Cleaning on page 348 and Decontamination on page 348.
The IH-500 must be taken out of service (see Decommissioning on page 413).
9.4.1 Disassembly
Only the technical personnel is authorized to disassemble the instrument.
k The
pieces and assemblies of the IH-500, disassembled by qualified personnel, may be reused
after obtaining the manufacturer’s authorization. The objectives of the future application must be
described in details.
IH-500 | 415
Decommissioning, Storage and Transport
9.4.4 Dismantling
The IH-500 must be returned to an authorized point of destruction.
k IH-500 materials.
Refer to the IH-500 dismantling manual for further instructions. This dismantling manual lists all
9.5 Transport
9.5.1 Recommended Means of Transportation
Domestic
The appropriate means for domestic transportation are by truck, by postal parcel and by train.
b During transportation, ensure that the packaging instructions are respectfully followed.
See chapter Packing on page 414.
Figure 450.
416 | IH-500
10 Appendix
This chapter contains the additional and helpful documents to be used during maintenance on IH-500
as schematic diagrams, maintenance record, decontamination certificate, and more.
IH-500 | 417
Appendix
10.2 CE Compliance
This instrument is and in-vitro diagnostic medical device and complies with all applicable European
Community Directives and associated harmonized standards, including but not limited to the standards
pertaining to the electrical safety as well as the emission and immunity requirements as specified in IEC
61326-2-6.
k For the following documents, excepting the maintenance table, the official electronic document
can be found on the Bio-Rad website: https://www.bio-radsecretarea.com under Official
forms.
b conspicuous place.
Fill in a copy of the decontamination certificate and place it on the IH-500, in a
The absence of decontamination certificate may extend the time needed for revision and maintenance.
Bio-Rad shall bill any cleaning and decontamination operation.
418 | IH-500
'&+%"&+"'&()')%
/
',&+)/
%"$
!'&
&*+),%&+
+)"$'
&*+),%&+'
$&"& '&+%"&+"'&()')%/
)-"*#"$()',+*,((')+)**")"')'%'). 1
*.00123405
!"#$% 105
$&'()"*)''("*+ (,( 6''$"(%"&7"(88.4
Appendix
Type Page
Serial number
420 | IH-500
Documents / Forms
IH-500 | 421
Appendix
IH-500
k just near the cleaning liquid container and USB
A second ID-plate is placed into the instrument, 1
RO00007 C
o
(21)0501330
Manufacturer’s reference
p Serial number
This class (B) digital apparatus complies with Canadian ICES-003
x
j European Conformity marking
Intended use
u
WEEE directive symbol
422 | IH-500
List of Documents
IH-500 | 423
Appendix
424 | IH-500
10.6
10.6.1
CP10 – E11246
Level detection board
M18990 CP52
PO02
Liquid waste Intermediate tank Pressure sensor
Pump Pa
EV01 EV02
System liquid IN solenoid valve OUT solenoid valve
M20231
Decontamination liquid Intermediate tank
NC NO NO NC Faradized vein
Overflow Pbar
Peristaltic pump tubing FT04
1 2 3 3 2 1
10 μm filter
2 x 16006249
CP21 – E10591
Intermediate tank Float Manifold
16005915 M18341
Level sensor
M20126 Security ring Needle
TU07 – M19788 – 835 mm
12010702
FT03 Complete sampling module M19700
10 μm filter Collect well Washing pot
M18436
Clip
well
TU08 – M06119 – 650 mm
M18436
Overflow well
Decontamination
TU06 – M06119 – 188 mm Clip
20 mm
PO05
M08816 NC NO NO NC
PO03 Waste pump EV06
M19672 EV05
TU05 – M06119 – 162 mm M18436 M16311
Decontamination pump M16311
Clip Common/Overflow
Collect / decontamination
solenoid valve
solenoid valve
TU04 – M06119 – 25 mm
20 mm
TU03 – M06119 – 25 mm TU23 – M06119 TU11 – M06119 – 282 mm
Length to be defined
during installation
3 x 16006249
M19708
FT02 M18436 3 x 16006249
TU15 – M19997 – 20 mm
Switching solenoid
valve NO NC
FT05
M19263
25 μm filter
TU18 – M04693 – 100 mm
CR-07-
f 2017-124
Fittings M13579 replaced by 16006249 and TU03 & TU04 changed from 20 to 25mm FSE 2020-08-26 MLE 2020-08-26
CR-07-
e 2018-005
PDS module M19999 (0595085) replaced by 12010702 (12010701) FSE 2020-01-15 MLE 2020-01-15 Title:
d
CR-07-
2017-122
Security ring M06039 replaced by 16005915 FSE 2019-05-28 MLE 2019-06-11
IH-500 fluidic connection
CR-07-
d 2018-071
TU08 length : 590mm changed to 650mm GLE 2019-05-10 MLE 2019-06-11
In
M20418 With insulating foam
M20418 Out
TU27 – M19405 - 57 mm
Reagent Cooling (2/2)
M20419
M20285
M20380
Reagent cooling tubing kit
M20419
In Reagent plate
TU24 – M20046 - 365 mm With insulating foam
M20285
With insulating foam
M20268
Reagent module
TU25 – M20046 - 420 mm
With insulating foam
M20144 M20145
Cooling pump
IH-500 | 425
Appendix
426 | IH-500
10.7
5 4 3 2 1
10.7.1
USER ACCESS
Touchscreen IHM
+ loud speaker
E11272
D ON/OFF D
Vision station
Ethernet Computer USB
USB Camera
Ethernet E10030 RS232
barcode Barcode reader
12V
ON/OFF Backlight
PSON
RS232
E10092
(Transport Arm)
Power management board
Security switch
M20720 E09982 Sheet 7
C C
Sheet 8
BSB
Detection
sensor
Teach
RS232
Mapping
sensor
CA05 CA02 CA04 CA01
CA06
Incubator Pipetting applicative X,Y,Z axies Centrifuge applicative
board Fluidic board Solid waste board
stepper motor
(with 3 Novicore) management
Gripper applicative I/O Board
board Reagent
(agitation + cooling) Teach Liquid
E10904 waste E09968
E09980 E09972
Security
(doors, ..) Fluidics
B CA03 Centrifuge 1 B
DC motor
Consummable PDS applicative board
loading (x2) Ionisor
Robot arm Centrifuge 2
DC motor
Int/ext temp. CA13
management E09974 CA09
Cooling driver board
Level
ON/OFF detection Fan E10928
Backlight External status PDS module board
Barcode Centrifuge
CA07 Sample rack Pipetting robot
cooling
detection board
A E09984 A
Sheet 3 Sheet 4 Sheet 5
/&50 !3'6PRGXOH &0$3%2 TITLE :
N&5$50&RQWUROHU0!0 &0$3%5 Automate schematic
N&5&RUHFWLRQ(UUHXUSODQ( !( 3%23%5
j 2017-011 TouchScreen E10546 replaced by E11272 CMA 07/07/2017 PBO 07/07/2017
E09978 i 2016-081 Label USB between computer and power supply replace by 12V CMA 12/04/2017 PBO 12/04/2017 SHEET : 1/8
Sample status h 2015-297 E09998 replacedby E11246 PBR 17/05/2016 PBO 17/05/2016
Add cable E11252 REF. :
GNO 03/01/2012 JNE 11/07/2014
E10232 c 2014-051 E09970 board replaced by E10904
E10063 board replaced by E10928 A00227 l
Sheet 2 Ind AM DESCRIPTION AUTHOR DATE REVIEWER DATE
Document history No copy, nor communication allowed
5 4 3 2 1
10.7.2
5 4 3 2 1
CA05
Power supply unit I/O board Reagent agitation
Power supply (24V, +VMOT) (JP91) E10704
JP39 E10113
JP76 (MO11) Motor (stepper)
Cooling power supply (JP92) E10705 JP41 (CC11) Encoder sensor E10610
D D
Step by step μcontroller E10552
JP75
E10706 JP40
Incubator power supply (JP93) Motor Shaking
I/O Board (2/8)
Solid Waste Full sensor (CP68) E10194 E10556 E10125 Option (CP60) Internal temperature sensor for
JP48 JP63 E08818
E10660 general machine
A A
i 2016-081 Reagent plate: 2xE08818 => E11248 PBO 07/01/2016 PBR 08/01/2017
TITLE :
I/O board
g 2016-052 Cordon Switch E10625=>E11255; Suppression Switch porte E008761 (CP76) PBO 15/03/2016 PBR 17/03/2016
2015-210 E10886 => E10886 or E11191
c 2014-051 Temp board E08818 replaced by E09990 for external and internal areas
SHEET : 2/8
E10929 cable deleted
PBO 27/05/2012 JNE 11/07/2014 REF. :
E10115 cable modification
E10996 cable added
Ind AM DESCRIPTION AUTHOR DATE DATE
A00227 l
REVIEWER
Document history No copy, nor communication allowed
5 4 3 2 1
Electrical Connections (A00227i-EL Schematics)
IH-500 | 427
Appendix
428 | IH-500
10.7.3
5 4 3 2 1
CA03
Needle Safety CA09
PDS applicative
D
JP20 D
board Level detection board
CA02 E10294
JP19 (CP47)
Pipettor (3/8)
E11252 E11246
Pipetting applicative board Needle
JP12
Logic
JP03
E09982 E10099
JP14
JP25
CA12 16005360
CA04 E07722 E10564 JC02
Presure and temperature board
JP09
JP28
Logic extension
Fluidic board NOVI CORE AXE Z NC Z Pressure/vacuum measurement
E10200 E10624 JC01 functions + signal set up SPI
Pump control JP35 JP07 Z axis step by step motor E10565 JP24 communication with PDS board
Power
JP13
Z axis optical encoder
JP08
Z axis origin JP16
Level detection (CP29) I2C temperature sensor
E09972 Touch probe E09986 JP11 JP15 E09974 JP17 JP26 JP18
PDS synchronisation Board power supply +/- 5V
16004804
B PDS Module B
E07722 12010702
+5V power
+7V power
Main frame CA15
X axis step by step motor (MO09) Z axis step by step motor (MO08) OPTION
Y axis step by step motor (MO10)
E10172 E10979 E10608
A A
5 4 3 2 1
IH-500 | 429
Appendix
430 | IH-500
10.7.5
5 4 3 2 1
D D
CA08 CA01
Power management board Front centrifuge
CP027 PT100 sensor N°2 (Sheet 6) (CP27)
Centrifuges (5/8)
E10908 (MO06)
JP02 J3
J4 E09748
CP01
J5
CP02 Main frame
C C
AC01
Rear centrifuge
ON/OFF Cooling: see Centrifuge Cooling (Sheet 6) AC02
CP028 PT100 sensor N°2 (Sheet 6) (CP28)
Motor + encoder
ET01
E10192
CC01 (CP46) J2
ET02
(CP45) J1
Not mounted
JP01 E10908 (MO07) J3
J4 E09748
J5
Main frame
B B
A A
TITLE :
Centrifuges
SHEET : 5/8
REF. :
c 2014-051 First edition GNO 03/01/2012 JNE 11/07/2014
Ind AM DESCRIPTION AUTHOR DATE REVIEWER DATE
A00227 l
Document history No copy, nor communication allowed
5 4 3 2 1
10.7.6
5 4 3 2 1
CA13
JC01
Temperature monitoring
CP02 sensor area 1 (Reagent
option)
DIP SWITCH: SW1
1 = OFF Front centrifuge
2 = ON Front cooling Peltier element
3 = ON E10992
4 = OFF
Centrifuge Cooling (6/8)
CC01
Fan AC05
CP08 E10824
Rear centrifuge
Cooling area 2
(Front centrifuge) AC03 Front cooling Peltier element
E10992
Cooling area 3
(Rear centrifuge) Rear cooling Peltier element
AC04 E10992
CP11
Temperature monitoring E10822
(Rear centrifuge) PT100 sensor N°1 (CP80)
E10825
CP04
On/OFF Cooling: see Centrifuges (Sheet 2) E10827
CP05 see Centrifuges (See Sheet 5) PT100 sensor N°2 (CP28)
B
AC01 B
CP07
Power supply
E10826 unit
JP125
A A
TITLE :
Centrifuge cooling
SHEET : 6/8
REF. :
c 2014-051 E10823 Peltier element replaced by E10992 PBO 02/07/2014 JNE 11/07/2014
IH-500 | 431
Appendix
432 | IH-500
10.7.7
5 4 3 2 1
E10251
&RPSXWHU E10030
COM2 Camera TouchScreen
12V
USB E10669
D D
DVI
E10265 USB
COM1
Sound
E11272
Integration (7/8)
E10252
E11041
Rear panel
RJ45
J205
E10858
USB E11033 USB USB option
E10348
J101 E10885
RJ45 E07498 RJ45 Ethernet option
C C
6 Axis Robot Controler
(Transport Arm) J103 Teach pendant connection
E10792
E10716
E10715
OPTION
Internal panel
JP105 JP100 JP95 E10253
USB E11033 USB Keyboard
RS232 PS-ON 12V
E10249
J1240 J203 J211 JP94 12V
Left panel
E10253 Bar code reader
USB E11033 USB
+24V E11096
J2 Laser sensor
JP96
Touch probe 3RZHUVXSSO\XQLW
E10983
J400
E10102
6 Axis harness JP79 (CC12) Encoder 12010406
E10795
JP83 JP82
JP81 (CP65) Mapping sensor
IPOS3602
E10658
6 axis arm
E10026 E10104 16008287
JP80
5 4 3 2 1
E10707
E10830 E10982 Door JP102 JP58 E09978
E10864 Cover
JP121 DC_OK
LINE
C C
Controller & Power Supply (8/8)
E10856
Filter Power supply
LOAD
750W/24V JP96 See Sheet 7
150W/12V
E10708
+24V J406 24V3
SLOT V+ E11002
+24V DSI
VAC_IN +24V JP108 24V2
+24V Fuses
V- E11005
0V
See SM
for more E10709 J030 200VDC_IN
SLOT V+ E11003
+12V details +24V JP106 J1164 24V3
+12V E10710
V- 0V +24V JP99 J213 24V1
E11005
E10896 E10894
ON/OFF switch G AC Fail Main PWR OK
ON/OFF switch connector J120
JP109 6 Axis Robot Controler
E07560 SLOT V+ E11004
+VMot 200VDC Fail (Fieldbus) Vbus OK
+24V ou E10712 (Transport Arm)
V- E11005
0V Laser Laser
+36V JP107 J212
E10893 Touch probe Touch probe
E10713 Door
B Door JP111 J100 B
M20720
+12V E10249
JP94 DC power Screen
Global CTRL E10877 Global enable
JP101 & AC fail
Vmot Enable E10715
E10878 +12V JP95 12V Computer
E10792
E10865 RS232_2 JP105 RS232
E10829 E10716
JP103 Enable VMOT PS_ON PS_ON
JP100
PC Enable PC_ON/OFF E10030
Junction
box E10987
Capacitiv load board JP122
E09988 JP124 E10852 E10853
BP ON JP98 Start button E09417
Option
JP97 E10248
Fan Touch probe JP104 Fixed touch probe
A E09982 A
M16108 Option
TITLE :
e 2014-097 E10867+E10868=>E11001; E10871+E10872=>E11002; PBO 16/09/2014 JNE 02/12/2014
Controler & power supply
Outlet E10873+E10874=>E11003; E10875+E10876=>E11004; Ajout 4xE11005;
E10866=>E11007; Modif câblage des terres. E10838=>E10983 SHEET : 8/8
E10547 d 2014-051 E10985 replace E10857 ICB-PLE 10/09/2014 FSE 10/09/2014
REF. :
c 2014-051 E10987 cable added PBO 27/05/2012 JNE 11/07/2014
Ind AM DESCRIPTION AUTHOR DATE REVIEWER DATE A00227 l
Document history No copy, nor communication allowed
5 4 3 2 1
Electrical Connections (A00227i-EL Schematics)
IH-500 | 433
Appendix
434 | IH-500
Figure 452.
JP88 JP96 JP107 JP104
JP95 D26
JP94
D12
D1 JP89 JP100 JP109 JP111
JP93 D4
D3
D9 D10
D2 JP91 JP108 JP106 JP98
JP92 D11 JP105
D7 JP123
D6 D25
JP125 D5 D8 JP90 JP102 JP99
D43
D42
Electronic Boards
IH-500 | 435
Appendix
Firmware Configuration
436 | IH-500
Electronic Boards
Figure 453.
IH-500 | 437
Appendix
Firmware Configuration
438 | IH-500
Electronic Boards
JP42
JP70
JP71
JP75
JP72
JP62 JP64 JP65 JP66 JP67 JP68
Figure 454.
IH-500 | 439
Appendix
D10
D4
D2
D5
D1
Figure 455.
440 | IH-500
Electronic Boards
D1
Figure 456.
IH-500 | 441
Appendix
Firmware Configuration
D15 D7
D16
D2
D11
D18
D10
D17
D3
Figure 457.
442 | IH-500
Electronic Boards
D2
Figure 458.
Figure 459.
IH-500 | 443
Appendix
Firmware Configuration
D1
Figure 460.
444 | IH-500
Electronic Boards
Figure 461.
IH-500 | 445
Appendix
Firmware Configuration
446 | IH-500
Electronic Boards
LD9
LD8
LD6 LD7
LD5
LD3
LD1
LD4
LD2
Figure 462.
IH-500 | 447
Appendix
Switch configuration
ON
1 2 3 4
OFF
Figure 463.
448 | IH-500
Electronic Boards
LD9
LD8
LD6
LD7
LD4
LD5
LD1
LD3
LD2
Figure 464.
IH-500 | 449
Appendix
Figure 465.
Figure 466.
450 | IH-500
10.9
Door
On power
Figure 467.
+12V_IN +Vmot_IN +24V_IN management
board
E10619 E10617 E10617 E10618 E10619 E11125 E10619 E10618 E10618 E10618
Polyswitch
4 1A
4
+24V_INT
+12V_PC +12V_SCREEN +Vmot +24V_BSB +24V_A +24V_COOL_R +24V_COOL_C +24V_INCUB +24V_ST_A +24V_ST_B
5V autoprotected
2.5A
Power Management (Fuses)
Board
internal
power
JP95 JP94 JP88 JP90 JP89 JP91 JP96 JP92 JP125 JP93 JP99 JP106 JP108 JP98 JP97 JP101
PC Screen Pipetting Fluidic Centrifuge I/O board Gripper I/O board Cooling I/O board 6 axis 6 axis 6 axis Start button Fan Global control
applicative board Applicative For reagent Applicative For reagent Driver For Incubator controler controler controler LED
board board agitation board cooling board Stäubli_A Stäubli_B Stäubli_C
IH-500 | 451
Appendix
452 | IH-500
Index
Index
A I
Alarm Identification plate 422
Behavior 375 IH-500 gauge kit 73
Antibody Screen Testing 61 Installation 89
Instrument 15
B Intended Use 13
Back (instrument) Internal view
General description (rear view) 29
General description 30
Backup 119
L
C Lifting kit 84
Centrifuge
Loading area
12 Gel Cards (general description) 57 General description
General description (rear view) 29
Diluent rack 66
Maintenance
Sample rack 61
Decontaminate 355, 363, 364, 365, 367 Maintenance
Preventive maintenance table 347
Preventive maintenance table (racks) 347
Commissioning Troubleshooting 376
Backup 119
Localization module
General safety requirements 77 Maintenance (table) 347
Global checks 120
Hydraulic checks 121
Removing blocking devices 112
M
Starting procedure 116 Maintenance 345
Complete gripper Major Injury 15
General description 39 Materials 416
Minor Injury 15
D Moderate Injury 15
Decommissioning 413
Decontamination instructions 348 N
Definition Note 16
Warning 16
Disposal 415 P
Doors Packing 414
General description 33 Piercer module
Drawers General description 42
General description 34 Pipettors
Hydraulic checks
E Commissioning 121
Electrical connections (A00227d schematics) 426
Electronic boards Q
General description 68 QC client 125
F R
Forms and lists Rear centrifuge
Accessories list 418 General description 32
Decontamination certificate 420 Removable reagents tray 67
Equipment list 417
List of documents 423 S
Maintenance table 420
Safety
Front
Environmental conditions 18
General description (front view) 28
General instructions 22
Front centrifuge Importance of the instructions 17
General description 32 Non-compliance with rules 17
Sample block barcode label set 62
G Sample Requirements ABO/Rh Testing 61
Global checks (commissioning) 120
Schematics and connection lists 426
Spare parts list and accessories 411
H
Starting procedure 116
Hydraulic connections (A00227-FL schematics) 424
Start-up 116, 117
IH-500 | 453
Index
Storage
Long term 414
Storage drawer
General description 28
T
Transport 416
Transport arm
General description 37
Troubleshooting 375
U
Unpacking 79
Remove from the crate 80
Fork truck only 80
V
Vision station
General description 40
W
Warning
Definition 16
Warranty
Limitations 14
454 | IH-500
Notes
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
.......................................................................................................................................................................................
IH-500 | 455
Bio-Rad
Laboratories
Website www.bio-rad.com/diagnostics Australia 61-2-9914-2800 Austria 43-1-877-8901 Belgium +32 (3)710-53-00 Brazil +55 (31)3689-6600 Canada 1-514-334-4372
Clinical China 86-21-61698500 Czech Republic 420-241-430-532 Denmark +45-4452-1000 Finland 358-9-804-22-00 France 33-1-47-95-60-00 Germany +49 (0)89-318-840
Greece 30-210-7774396 Hong Kong 852-2789-3300 Hungary +36-1-459-6100 India 1800-180-1224 Israel 972-3-9636050 Italy +39-02-216091 Japan 81-3-6361-7070
Diagnostics Group Korea 82-2-3473-4460 Mexico +52 (55)5488-7670 The Netherlands +31-318-540666 New Zealand 64-9-415-2280 Norway +47-23-38-41-30 Poland 48-22-3319999
Portugal 351-21-472-7700 Russia +7-495-721-1404 Singapore 65-6415-3170 South Africa 27-11-442-85-08 Spain 34-91-590-5200 Sweden +46-8-555-127-00
Switzerland +41 (0)26-674-55-05/06 Taiwan 886-2-2578-7189 Thailand 662-651-8311 United Kingdom +44 (0)20-8328-2000