Professional Documents
Culture Documents
6.3-01.04-32_checklist_24_month_hs100_200(2)
6.3-01.04-32_checklist_24_month_hs100_200(2)
Required Material
Hardware:
1 x 168902 "HumaStar 100/200 Service Manual"
1 x 16890/250-24 "Maintenance kit 24 month for HS100 and 200"
4 x 17400/56 “HC-Acute (50ml)”
1 x 1000 ml Vessel for preparation of the decontamination solution
The length of the inner tube between the pinch valve and the silicon-to-TEFLON adapter must be
approx. 10 mm.
Replace the tubing "7" labelled as B4 between the wash well of the sampling arm, bottom output
connector and the input connector (inner side) "14" of the peristaltic pump 7. This tubing includes a
check valve.
Replace the tubing “6” labelled as A3 from the wash station manifold to the inlet of pump 6, fitting
11.
Replace the Y-fitting silicon tubes (4) labelled as B1 which are connected to the fittings 12 and
13 (inner side of the pump panel) to the fitting 7 of the pumps panel manifold. Please check the
Service Schematic in the Service Manual.
Note: The tubing labelled as B2 is not required.
Step 11: Waste tubing check valve (15) and air filters (16)
Place/replace the check valve (15) at waste tubing, with the black O-ring side in direction to the
instrument connector (see the picture below).
Replace the air filters (16) on top of the blue (system solution) and green (wash solution)
containers (see the picture below).
Step 13: Sampling needle (18), needle holder (19) and O-ring (20)
Note: It is necessary to check all the adjustments for the mechanical calibration. A detailed
instruction can be found in the Service Manual.
Replace the needle (18), needle holder (19) and black O-ring (20) (see SM "12.5 Needle
replacement").
Note: In case of any needed support please do screenshots of the result pages and sent them to
Human.
A) F1 Hardware Test
Check the vertical collision sensor by moving the needle manually up and down and watching the
second indicator (marked red) for changing between 0 and 1. OS-F1-F5
Level Sw.:10110001
Check liquid level sensing by touching/untouching the needle manually close to the tip and watch
the first indicator (marked red) for changing between 0 and 1. OS-F1-F5
Level Sw.:10110001
YES
IF the difference between Asymmetrical backlash
COUNT1 and COUNT2 is > 2
NO
YES
IF GAP < 4 No actions to be taken
NO
NO YES
IF GAB = 4 re-schedule test in 6
months
NO
B) F2 MECHANICAL CALIBRATION
Check all the mechanical calibrations.
C) F3 MECHANICAL CHECK
Verify the belts and the mechanics therefore run. OS-F3-F4
D) F5 WASHINGS MENU
Note: perform fill tubes up to 3 times to restore the normal condition. (OS-F5-F5-2)
Pumps test. OS-F5-F5-1
Adjust pumps. OS-F5-F5-5
Additional notes
Instrument data
Following mandatory data can be found in the INFO window of the HI program.
After the maintenance has been performed, please submit a scanned copy of this checklist
page duly signed and stamped to extend or confirm the warranty of your device. Conditions
apply.