Professional Documents
Culture Documents
Quality System Procedure - 32 humidity
Quality System Procedure - 32 humidity
Quality System Procedure - 32 humidity
References : NABL – 141 , IEC 60068-3-5 , The soc. Of Environmental Engineers Publish
paper. DKD-R-5-7 .Euramet –cg-20 vol 1
Equipment used for calibration :
Set up preparation:
1. The UUC under test should be cold & empty but with usual accessories
2. Selection of no. of sensor if the chamber size is < 2000 Liter use nine senor .
Above 2000 liter position one sensor @ 1 meter. OR If chamber size is < 3 ft 3 use 5
no. sensor.
3. The flexible Sensor positioned at the corners of working space of UUC
4. One sensor for determine reference temperature in the geometric center
5. The working space of UUC need to be well specified , usually the space at least 5cm from
any wall
6. Mounting of the survey sensors can be facilitated by using rigid wire frame to hold the
sensors close to their sites.
7. Wires can shift with heating if they are not secured.
8. One more sensor to monitor the ambient temperature.
9. Hot junctions shall then be cleaned & thinly.
10. The UUC door joint & all other possible sources of leakage shall be effectively shield
11. The leads may be brought out a split rubber stopper inserted
12. Refer fig. No.1
Environmental Conditions:
Parameter Requirement
Temperature 25 0 C +/- 4 0 C
RH 50 ± 20%
Earth to Neutral <1V
voltage
Document No:- NI Procedure for calibration of Humidity Chamber and uncertainty of Sect. No:- Thermal
/LAB/CPM measurement.
Issue No:- Issue Date:- 01- Revision No :- Revision Date:- Prepared By & Approved By:-
04 02-2022 V.B. Hingmire ( TD )
1
Note :
1. On site calibration temperature may be allowed up to 35 0 C. If temperature correction
required this contribution shall be taken in to uncertainty budget.
2. If above environmental condition not observed then stop the calibration
3. Environmental condition should be record in the calibration data sheet
Method of calibration :
1. Refer operation manual o both UUC and Master where UUC = unit under calibration.
2. Before starting survey , ensure that the Humidity chamber controller is working properly
3. Check supply cord & other cable ensure this cables should not cut any where.
4. The controller should be set to Humidity and not subsequently adjusted to obtain the right
Humidity.
5. Tuning of the controller may be needed.
6. Ensure that both controller & data logger functioning correctly are they calibrated.
7. Measurements shall not commence until at least one hour after the chamber has been
switched on
8. Select the temperature point as per customer’s requirement & if customer has not
suggested calibration point, then calibrate the instrument at approximately 15%, 50% and
90% of F.S. of UUC
Document No:- NI Procedure for calibration of Humidity Chamber and uncertainty of Sect. No:- Thermal
/LAB/CPM measurement.
Issue No:- Issue Date:- 01- Revision No :- Revision Date:- Prepared By & Approved By:-
04 02-2022 V.B. Hingmire ( TD )
2
7 8
1 1
6 9
1
5
3
2
4
1
Fig. 1 Set up diagram of equipment to assessment the Humidity uniformity – The numbered
circles indicate to the sensors location –within the chamber
Humidity Gradient :
Expressed as the difference between highest & lowest average Humidity value of the sensors
on different location within the working space after achieving stability
Error calculation:
Uniformity :
Maximum deviation from average between the center & other points
Uncertainty of measurement
Document No:- NI Procedure for calibration of Humidity Chamber and uncertainty of Sect. No:- Thermal
/LAB/CPM measurement.
Issue No:- Issue Date:- 01- Revision No :- Revision Date:- Prepared By & Approved By:-
04 02-2022 V.B. Hingmire ( TD )
3
Type A contribution:
Repeatability:
Type B Contribution:
2. Uncertainty due to temp. effect arrives from temp. coefficient of temp. measuring
Instrument electronics & the temp. variation it experiences during use from manufacture
specification. It is multiplied by 4.5 to convert it to humidity (U 2)= Temp. coefficient / √3
(Considering rectangular distribution, degrees of freedom = ∞)
3. Contribution due to Drift to the standard equipment can be estimated by considering the
current & pervious calibration certificate. (U 3) = Drift of std. instrument / √3
(Considering rectangular distribution, degrees of freedom = ∞)
5. The Humidity gradient can be derived from the measurement at each measurement the
average & std. deviation (SD) of nine sensor has been calculated the std deviation is a
measure of Humidity gradient .The safest assumption is to take largest value of std
deviation. The input budget is a std. Deviation from repeated measurement & so can be
assume to be normal & the divisor is 1. (U 5) = largest value of std deviation / 1
(Considering normal distribution, degrees of freedom = ∞)
6. Fluctuations are also derived from the data for each sensor the average & std. deviation
over the period of the test has been calculated. The std. deviation is a measure of the
fluctuation. The safest assumption is to take the largest value of std. deviation again it is
std. deviation so it can be assume to be normal & divisor is 1 (U 6) = largest value of std
deviation / 1 (Considering normal distribution, degrees of freedom = ∞)
Document No:- NI Procedure for calibration of Humidity Chamber and uncertainty of Sect. No:- Thermal
/LAB/CPM measurement.
Issue No:- Issue Date:- 01- Revision No :- Revision Date:- Prepared By & Approved By:-
04 02-2022 V.B. Hingmire ( TD )
4
7. Over all mean of all measurement is statically based & allowance must be made for
possibility that it may not present the true average the value input in to the uncertainty
budget is the over all std. deviation divided by the square root of the no. of measurements
This component referred to as std. deviation of the mean may appear negligible but it
induced for completeness. If only a few measurements were make it could become
significant . (U7) = over all std. deviation / √ no. of measurements. (Considering normal
distribution, degrees of freedom = ∞)
Records:
1. The calibration report should be issued. . A copy of the same shall be retained in the file with
subject “Record of the calibration certificates issued”
2. The original observations should be retained in the file with subject “Observation sheets”
3. All calibration request formats duly filled by customers and Cal lab should be retained in the
file with subject “Calibration request forms”
Document No:- NI Procedure for calibration of Humidity Chamber and uncertainty of Sect. No:- Thermal
/LAB/CPM measurement.
Issue No:- Issue Date:- 01- Revision No :- Revision Date:- Prepared By & Approved By:-
04 02-2022 V.B. Hingmire ( TD )
5
Prepared By Approved By
Technical Officer Head of WC
NISHITRONICS INSTRUMENTATION PUNE
CALIBRATION PROCEDURE MANUAL Procedure No. NI/CP/08
Issue: 03 Date: 03.04.2012 REVISION: 00 Date: 30.04.2012 Page: 1 of 4
Prepared By Approved By
Technical Manager Technical / Quality Manager