BFTII Reference Guide 3.01

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 158

BFT* II Analyzer

Reference Guide 3.01


Not for use in the USA

This document is Siemens Healthcare Diagnostics Reference Guide


version number 3.01 for BFT* II Analyzer. From January 2010, this
document is the only relevant source for assay application information
on the BFT* II Analyzer.
The Reference Guide contains a complete list of all assays evaluated
and released by Siemens. Information in this document supersedes
information in earlier versions of the Reference Guide or Application
Sheets.

Document ID: RG-01-01-3.01


BFT* II Analyzer

Trademarks
Actin, Ci-Trol, Dade, Innovin, ProC and Thromborel are trademarks of Siemens Healthcare Diagnostics.

* BCT, BFA, BFT, InnoCal, Multifibren, Pathromtin and Thromboclotin are trademarks of Siemens
Healthcare Diagnostics.

Sysmex is a registered trademark of SYSMEX CORPORATION.

Note
Siemens Healthcare Diagnostics has validated the provided instructions, reagents, instrument, software and
customizable features for this system to optimize product performance and meet product specifications. User
defined modifications are not supported by Siemens as they may affect performance of the system and test
results. It is the responsibility of the user to validate any modifications made to these instructions, instruments,
reagents or software provided by Siemens.

Siemens Healthcare Diagnostics Products GmbH

Emil-von-Behring-Straße 76
35041 Marburg
Germany

Reference Guide 3.01 - not for use in the USA - 2


BFT* II Analyzer

Revision History

Document Date Changes


Reference Guide 3.01 2010-01 New Application Sheets:
No new Application Sheet.

Deleted Application Sheets:


No Application Sheets deleted.

Revised Application Sheets:


General changes:
1. Disclaimer on product performance revised:
“The parameters defined in this application sheet have been developed by
Siemens Healthcare Diagnostics to provide optimal product performance with the
assay and instrument combination. Any modification to these parameters may
affect performance of this and other assays in use on your system and the
resulting assay values. It is the responsibility of the user to validate any
modifications and their impact on all assay results. Results of this test should
always be interpreted in conjunction with the patient's medical history, clinical
presentation and other findings.
The application sheet lists all combinations of controls and calibrators for use with
the reagent and instrument system; other combinations are not validated or
supported by Siemens.”
2. Disclaimer for on-board stability added:
“The stability data presented here were established under controlled laboratory
conditions. Due to differences in laboratory environmental conditions and reagent
vial fill volumes performance may vary. Please contact your local service
representative to report any issues.”

Changes in detail:
No changes in detail
Reference Guide 3.00 2009-05 New Application Sheets:
No new Application Sheets.

Deleted Application Sheets:


No Application Sheets deleted.

Revised Application Sheets:


General changes:
- New layout for Application Sheets. Information within an Application Sheet may
be moved to other sections in order to be in a more appropriate position with
regards to content.
- Symbols used for reagent kit components instead of translated product names,
according to new labeling model
- Reference to guideline for reference intervals changed to CLSI document
C28-A3
- Parameter-Protocol: Non relevant parameter marked with * and corresponding
note “*This parameter does not have any function in this test.” added under
Remarks. (except: PT INR with Thromborel® S Dade® Innovin® (SW 2.01 and
SW 2.02) and with Dade® Thromboplastin C Plus)

Changes in detail:
PT sec with Thromborel® S and with Dade® Innovin® (SW 2.02)
Reference Interval: New data for reference interval.
® ® ®
PT % with Thromborel S and with Dade Innovin (SW 2.02)
Reference Interval: Data replaced by new data with PT-Multi Calibrator;
Test Protocol: PT-Multi Calibrator above the table deleted.
PT % with Dade® Innovin® (SW 2.01)
Test Protocol: PT-Multi Calibrator above the table deleted.
PT INR with Thromborel® S, Dade® Innovin® (SW 2.01 and 2.02) and with Dade®
Thromboplastin C Plus
Parameter-Protocol: Non relevant parameter marked with ** and corresponding

Reference Guide 3.01 - not for use in the USA - 3


BFT* II Analyzer

note “**This parameter does not have any function in this test.” added under
Remarks.
Fibrinogen with Multifibren* U
Limitations: Note on HES interference revised: “… In a study with a HES solution
(130/0.4) no interference was observed up to 15 g HES per liter plasma.”;
Parameter-Protocol (page 4) - measurement - lot. no.: 5198xx replaced by xxxxxx
Reference Guide 2.8 2008-10 New Application Sheets:
No new Application Sheets.

Deleted Application Sheets:


No Application Sheets deleted.

Revised Application Sheets:


General changes:
Dade Behring Marburg GmbH changed to Siemens Healthcare Diagnostics
Products GmbH.
Dade Behring changed to Siemens Healthcare Diagnostics or Siemens.
Manufacturer and USA Distributor changed to Siemens.
Section Trademarks in Application Sheets deleted.
Trademark ° changed to * .
Test Protocol: “Dade Behring Preset No.” changed to “Preset No.”.

Changes in detail:
PT INR with Dade® Innovin® (SW 2.01 and SW 2.02)
Parameter Protocol – min./max. value: 150.0 % changed to 120.0 %.
Reference Guide 2.7 2008-07 New Application Sheets:
No new Application Sheets.

Deleted Application Sheets:


No Application Sheets deleted.

Revised Application Sheets:


Changes in detail:
Fibrinogen with Multifibren° U
Reference Interval: System specific values for Reference Interval added;
bibliographical reference 1,8-3,5 g/L deleted;
Bibliography: “Thomas, Lothar Clinical Laboratory Diagn. TH-Books
Verlagsgesellschaft mbH“ deleted.
Reference Guide 2.0 to - See respective Reference Guide
2.6
Reference Guide 1.0 - No Reference Guide 1.0 for BFT* II Analyzer

Reference Guide 3.01 - not for use in the USA - 4


BFT* II Analyzer

Table of Contents
Normal printed version number: Current version of the Application Sheet, no changes to the previous
Reference Guide
Bold version number: New Application Sheet version in this Reference Guide

Version Application Sheet Page

PT
07 PT seconds with Thromborel® S ..........................................................................................................7
09 PT % with Thromborel® S.....................................................................................................................10
08 PT INR with Thromborel® S .................................................................................................................14
08 PT INR calibrated with Thromborel® S.................................................................................................17
07 PT seconds (SW 2.01) with Dade® Innovin® ........................................................................................20
06 PT seconds (SW 2.02) with Dade® Innovin® ........................................................................................23
09 PT % (SW 2.01) with Dade® Innovin® ..................................................................................................26
07 PT % (SW 2.02) with Dade® Innovin® ..................................................................................................29
08 PT INR (SW 2.01) with Dade® Innovin® ...............................................................................................32
06 PT INR (SW 2.02) with Dade® Innovin® ...............................................................................................35
07 PT INR calibrated (SW 2.01) with Dade® Innovin®...............................................................................38
06 PT INR calibrated (SW 2.02) with Dade® Innovin®...............................................................................41
08 PT seconds with Dade® Thromboplastin C Plus ..................................................................................44
08 PT INR with Dade® Thromboplastin C Plus .........................................................................................47
08 PT INR calibrated with Dade® Thromboplastin C Plus.........................................................................50

APTT
08 APTT with Dade® Actin® Activated Cephaloplastin Reagent ...............................................................53
08 APTT with Dade® Actin® FS Activated PTT Reagent ..........................................................................56
09 APTT with Dade® Actin® FSL Activated PTT Reagent ........................................................................59
08 APTT with Pathromtin* SL ...................................................................................................................62

Fibrinogen
09 Fibrinogen with Multifibren* U ..............................................................................................................65

Thrombin Time / Batroxobin Time


08 Thrombin Time with Test Thrombin Reagent.......................................................................................69
08 Thrombin Time with Thromboclotin* ....................................................................................................72
08 Batroxobin Time with Batroxobin Reagent...........................................................................................76

Clotting Assays
07 Coagulation Factor II with Thromborel® S............................................................................................79
07 Coagulation Factor II with Dade® Innovin® ...........................................................................................83
07 Coagulation Factor V with Thromborel® S ..........................................................................................87
07 Coagulation Factor V with Dade® Innovin® ..........................................................................................91
07 Coagulation Factor VII with Thromborel® S .........................................................................................95
07 Coagulation Factor VII with Dade® Innovin® ........................................................................................99
07 Coagulation Factor X with Thromborel® S ...........................................................................................103
07 Coagulation Factor X with Dade® Innovin® ..........................................................................................107
07 Coagulation Factor VIII with Dade® Actin® Activated Cephaloplastin Reagent ...................................111
07 Coagulation Factor VIII with Pathromtin* SL .......................................................................................115

Reference Guide 3.01 - not for use in the USA - 5


BFT* II Analyzer

07 Coagulation Factor IX with Dade® Actin® Activated Cephaloplastin Reagent .....................................119


07 Coagulation Factor IX with Pathromtin* SL..........................................................................................123
07 Coagulation Factor XI with Dade® Actin® Activated Cephaloplastin Reagent .....................................127
07 Coagulation Factor XI with Pathromtin* SL .........................................................................................131
07 Coagulation Factor XII with Dade® Actin® Activated Cephaloplastin Reagent ....................................135
07 Coagulation Factor XII with Pathromtin* SL.........................................................................................139
08 Lupus Anticoagulant with LA 1 Screening Reagent / LA 2 Confirmation Reagent..............................143
08 Protein C System with ProC® Global ...................................................................................................147
08 Factor V Leiden with ProC® Global / Factor V Deficient Plasma .........................................................151
08 Protein C with Protein C Reagent ........................................................................................................155

Reference Guide 3.01 - not for use in the USA - 6


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT seconds with
®
Thromborel S
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Thromborel S Reagent Reagent positions 1 -OUHP
4 4/10 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
®
Thromborel S Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 600
Hemoglobin mg/dL 1000
Bilirubin mg/dL 48

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
®
Thromborel S Reagent on BCT* System y = 1.06 x - 0.48 0.973
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 7


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT seconds with
®
Thromborel S

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.3 2.1 2.4
Control Plasma P 2.3 2.5 3.3

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
Comments n Median 2.5th - 97.5th Percentile
— 158 11.6 10.3 - 13.2
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 8


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT seconds with
®
Thromborel S

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit sec cuvette ON
Preset No. 1
1st convers none
Pipetting Scheme falling*
1. Sample 50 µL unit sec.*
decimal place xxx.x*
Incubation time 60 sec
min / max value 5.0 / 150.0% *
Start reagent
time interpolat. linear *
Thromborel® S Reagent 100 µL value interpolat. reciprocal *
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 9


BFT* II Analyzer
Application Sheet for Page 1 of 4
PT % with
®
Thromborel S
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Thromborel S Reagent Reagent positions 1 -OUHP
4 4/10 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
PT-Multi Calibrator OPAT 6 x 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
®
Thromborel S Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 1000
Bilirubin mg/dL 6

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.

Reference Guide 3.01 - not for use in the USA - 10


BFT* II Analyzer
Application Sheet for Page 2 of 4
PT % with
®
Thromborel S

Method Comparison
Test system for the method comparison:
1. Thromborel® S Reagent with Calibration Plasmas
2. Thromborel® S Reagent with PT-Multi Calibrator (comparison of calibrators)
Predicate Device Regression Equation r
®
1. Thromborel S Reagent on BCT* System y = 1.14 x - 2.37 0.986
2. Thromborel® S Reagent with Calibration Plasmas on BFT* // Analyzer y = 1.01 x - 0.59 0.998
r = Correlation Coefficient
Precision
The following studies have been calibrated with Calibration Plasmas.
Within Run Run to Run Total
% % %
Control Plasma N 2.3 4.0 4.5
Control Plasma P 3.1 3.2 4.3

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: % of norm
th th
Comments n Median 2.5 - 97.5 Percentile
— 158 99 78 - 124
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 11


BFT* II Analyzer
Application Sheet for Page 3 of 4
PT % with
®
Thromborel S

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit % cuvette ON
Preset No. 1
1st convers reference curve
Pipetting Scheme falling
1. Sample 50 µL unit %
decimal place xxx.x
Incubation time 60 sec
min / max value 5.0 / 150.0%
Start reagent
time interpolat. linear
Thromborel® S Reagent 100 µL value interpolat. reciprocal
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The standard curve is measured in duplicate.

Remarks
Do not use the “Tables of values” for conversion of the seconds into percent as supplied with the reagent, as the data are only valid for the
old Fibrintimer, not the BFT* II Analyzer.
Do not use the “Quick with factor” method as the factor supplied with the “Tables of values” is only valid for the old Fibrintimer, not the BFT*
II Analyzer.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 12


BFT* II Analyzer
Application Sheet for Page 4 of 4
PT % with
®
Thromborel S

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
PT-Multi Calibrator
Calibrator % sec
Calibrator 1 116 10.6
Calibrator 2 68 14.8
Calibrator 3 47 19.6
Calibrator 4 23 32.4
Calibrator 5 11 61.5

Remarks
®
PT-Multi Calibrator level 6 is not applied to Thromborel S Reagent.

Reference Guide 3.01 - not for use in the USA - 13


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT INR with
®
Thromborel S
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Thromborel S Reagent Reagent positions 1 -OUHP
4 4/10 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
[1]
Standard Human Plasma ORKL 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Optional; only if used to establish the MNPT.
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
®
Thromborel S Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 637
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.

Reference Guide 3.01 - not for use in the USA - 14


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT INR with
®
Thromborel S

Method Comparison
Predicate Device Regression Equation r
®
Thromborel S Reagent on BCT* System y = 1.03 x - 0.01 0.984
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.4 2.3 2.7
Control Plasma P 2.6 2.7 3.6

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
Not applicable.

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 15


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT INR with
®
Thromborel S

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit INR cuvette ON
Preset No. 1
1st convers reference curve
Pipetting Scheme falling
1. Sample 50 µL unit %
decimal place xxx.x
Incubation time 60 sec
min / max value 5.0 / 150.0 %
Start reagent
time interpolat. linear
Thromborel® S Reagent 100 µL value interpolat. reciprocal
Autostart yes
2nd convers INR
100.0% * xx.x s
factor --- **
ISI * x.xx

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
*
System specific MNPT and ISI have to be used.
ISI values for prothrombin time assays must be entered directly as they appear on the current reagent labeling. Any changes of
reagent lot, software upgrades, major servicing, etc., require verification of the ISI value. Failure to enter the correct ISI value will
cause incorrect International Normalization Ratio (INR) results.
In order to report PT % and PT INR “1st conversion” in the BFT*II Analyzer Parameter-Protocol has to be set to “reference curve”. “2nd
conversion” has to be set to “INR”.
To obtain INR without PT %, set “1st conversion” to “none” and “2nd conversion” to “INR”.
The Mean Normal PT (MNPT) can be established with Standard Human Plasma. To use Standard Human Plasma, it has to be measured at
least twice and the obtained clotting time has to be divided by the ratio given for the PT reagent in the “Table of Analytical Values” of the
respective Lot. No. of the plasma.
Do not use the “Tables of values” for conversion of the Prothrombin time into INR as the data are only valid for the Fibrintimer, not the BFT*
II Analyzer.
** This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 16


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT INR calibrated with
®
Thromborel S
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Thromborel S Reagent Reagent positions 1 -OUHP
4 4/10 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
PT-Multi Calibrator OPAT 6 x 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
®
Thromborel S Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 1000
Bilirubin mg/dL 6

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.

Reference Guide 3.01 - not for use in the USA - 17


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT INR calibrated with
®
Thromborel S

Method Comparison
Predicate Device Regression Equation r
®
PT INR calibrated with Thromborel S Reagent on BFT* // Analyzer and PT INR using the y = 1.04 x - 0.15 0.997
ISI value with Thromborel® S Reagent on BFT* // Analyzer.
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 2.1 0.9 2.1
Control Plasma P 4.0 1.4 4.0

Measuring Range
The measuring range is defined by the concentration of the calibrators used.

Reference Interval
Not applicable.

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 18


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT INR calibrated with
®
Thromborel S

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit INR cuvette ON
Preset No. 1
1st convers reference curve
Pipetting Scheme rising
1. Sample 50 µL unit INR
decimal place xx.xx
Incubation time 60 sec
min / max value 0.10 / 10.00 INR
Start reagent
time interpolat. logarithmic
Thromborel® S Reagent 100 µL value interpolat. logarithmic
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The standard curve is measured in duplicate.

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
PT-Multi Calibrator
Calibrator INR sec
Calibrator 1 0.93 10.6
Calibrator 2 1.29 14.8
Calibrator 3 1.71 19.6
Calibrator 4 2.83 32.4
Calibrator 5 5.36 61.5

Remarks
PT-Multi Calibrator level 6 is not applied to Thromborel® S Reagent.
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 19


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT seconds (SW 2.01) with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
® ®
Dade Innovin Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 300
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
® ® ®
Dade Innovin Reagent on Sysmex CA-6000 System y = 1.04 x - 1.79 0.979
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 20


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT seconds (SW 2.01) with
® ®
Dade Innovin

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.6 3.4 3.7
Control Plasma P 2.5 4.0 4.6

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
Comments n Mean Median 5th - 95th Percentile
— 60 9.0 8.9 8.1 - 10.1
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 21


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT seconds (SW 2.01) with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit sec cuvette ON
Preset No. 1
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec. *
decimal place xxx.x *
Incubation time 30 sec
min / max value 5.0 / 150.0% *
Start reagent
time interpolat. linear *
Dade® Innovin® Reagent 100 µL value interpolat. reciprocal *
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 30 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 22


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT seconds (SW 2.02) with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
® ®
Dade Innovin Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 1200
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
®
Dade Innovin Reagent on BCT* System y = 1.06 x - 0.41 0.999
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 23


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT seconds (SW 2.02) with
® ®
Dade Innovin

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.1 2.1 2.4
Control Plasma P 1.7 1.2 2.1

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
Comments n Median 2.5th - 97.5th Percentile
— 156 8.8 7.9 - 10.2
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 24


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT seconds (SW 2.02) with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 9 Method PT-Innovin
Unit sec cuvette ON
Preset No. 9
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec.*
decimal place xxx.x *
Incubation time 30 sec
min / max value 5.0 / 150.0% *
Start reagent
time interpolat. linear *
Dade® Innovin® Reagent 100 µL value interpolat. reciprocal *
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 30 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 3 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 25


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT % (SW 2.01) with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values. 2.01)

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
PT-Multi Calibrator OPAT 6 x 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).
Limitations
Lipemic specimens may affect the results.

On-board Stability
Material Time [h]
® ®
Dade Innovin Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL <200
Hemoglobin mg/dL 600
Bilirubin mg/dL 2.4

Reference Guide 3.01 - not for use in the USA - 26


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT % (SW 2.01) with
® ®
Dade Innovin

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Test system for the method comparison:
1. Dade® Innovin® Reagent with InnoCal* Set
2. Dade® Innovin® Reagent with PT-Multi Calibrator (comparison of calibrators)
Predicate Device Regression Equation r
® ® ®
1. Dade Innovin Reagent on Sysmex CA-6000 System y = 0.92 x + 3.25 0.979
2. Dade® Innovin® Reagent with InnoCal* Set on BFT* // Analyzer y = 0.92 x + 1.03 0.999
r = Correlation Coefficient
Precision
The following studies have been calibrated with InnoCal* Set.
Within Run Run to Run Total
% % %
Control Plasma N 2.7 1.5 3.5
Control Plasma P 3.4 2.6 3.6

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: %
Comments n Mean Median 5th - 95th Percentile
— 60 109.6 110.3 86.6 - 133.6
Data were generated using InnoCal* Set.
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 27


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT % (SW 2.01) with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit % cuvette ON
Preset No. 1
1st convers reference curve
Pipetting Scheme falling
1. Sample 50 µL unit %
decimal place xxx.x
Incubation time 30 sec
min / max value 5.0 / 120.0%
Start reagent
time interpolat. linear
Dade® Innovin® Reagent 100 µL value interpolat. reciprocal
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 30 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The standard curve is measured in duplicate.

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
PT-Multi Calibrator
Calibrator % sec
Calibrator 1 109.0 8.8
Calibrator 2 71.0 11.5
Calibrator 3 47.0 15.1
Calibrator 4 24.0 24.2
Calibrator 5 11 44.1
Calibrator 6 8 59.8

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 28


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT % (SW 2.02) with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
PT-Multi Calibrator OPAT 6 x 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).
Limitations
Lipemic specimens may affect the results.

On-board Stability
Material Time [h]
® ®
Dade Innovin Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL <200
Hemoglobin mg/dL 400
Bilirubin mg/dL 60

Reference Guide 3.01 - not for use in the USA - 29


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT % (SW 2.02) with
® ®
Dade Innovin

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Data were generated using InnoCal* Set.
Predicate Device Regression Equation r
Dade Innovin® Reagent on BCT* System y = 0.90 x + 6.98 0.974
r = Correlation Coefficient
Precision
The following studies have been calibrated with InnoCal* Set.
Within Run Run to Run Total
% % %
Control Plasma N 2.2 4.0 4.6
Control Plasma P 2.0 1.4 2.5

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: % of norm
th th
Comments n Median 2.5 - 97.5 Percentile
— 156 98 75 - 121
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 30


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT % (SW 2.02) with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 9 Method PT-Innovin
Unit % cuvette ON
Preset No. 9
1st convers reference curve
Pipetting Scheme falling
1. Sample 50 µL unit %
decimal place xxx.x
Incubation time 30 sec
min / max value 5.0 / 120.0%
Start reagent
time interpolat. linear
Dade® Innovin® Reagent 100 µL value interpolat. reciprocal
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 30 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 3 s
learn 3 s
lag 1 s

Standard Curve Calibration


The standard curve is measured in duplicate.

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
PT-Multi Calibrator
Calibrator % sec
Calibrator 1 98.0 9.0
Calibrator 2 65.0 11.2
Calibrator 3 42.0 15.1
Calibrator 4 23.0 24.8
Calibrator 5 12.0 42.4
Calibrator 6 9.0 52.8

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 31


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT INR (SW 2.01) with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
[1]
Standard Human Plasma ORKL 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Optional; only if used to establish the MNPT.
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
® ®
Dade Innovin Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 600
Bilirubin mg/dL 6

Reference Guide 3.01 - not for use in the USA - 32


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT INR (SW 2.01) with
® ®
Dade Innovin

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
® ® ®
Dade Innovin Reagent on Sysmex CA-6000 System y = 0.91 x - 0.20 0.951
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.4 3.1 3.4
Control Plasma P 2.3 3.6 4.2

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
Not applicable.

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 33


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT INR (SW 2.01) with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit INR cuvette ON
Preset No. 1
1st convers reference curve
Pipetting Scheme falling
1. Sample 50 µL unit %
decimal place xxx.x
Incubation time 30 sec
min / max value 5.0 / 120.0 %
Start reagent
time interpolat. linear
Dade® Innovin® Reagent 100 µL value interpolat. reciprocal
Autostart yes
2nd convers INR
100.0% * xx.x s
factor --- **
ISI * x.xx

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 30 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
*
System specific MNPT and ISI have to be used.
ISI values for prothrombin time assays must be entered directly as they appear on the current reagent labeling. Any changes of
reagent lot, software upgrades, major servicing, etc., require verification of the ISI value. Failure to enter the correct ISI value will
cause incorrect International Normalization Ratio (INR) results.
In order to report PT % and PT INR “1st conversion” in the BFT*II Analyzer Parameter-Protocol has to be set to “reference curve”. “2nd
conversion” has to be set to “INR”.
To obtain INR without PT %, set “1st conversion” to “none” and “2nd conversion” to “INR”.
The Mean Normal PT (MNPT) can be established with Standard Human Plasma. To use Standard Human Plasma, it has to be measured at
least twice and the obtained clotting time has to be divided by the ratio given for the PT reagent in the “Table of Analytical Values” of the
respective Lot. No. of the plasma.
** This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 34


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT INR (SW 2.02) with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
[1]
Standard Human Plasma ORKL 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Optional; only if used to establish the MNPT.
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
® ®
Dade Innovin Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 337
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Reference Guide 3.01 - not for use in the USA - 35


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT INR (SW 2.02) with
® ®
Dade Innovin

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
®
Dade Innovin Reagent on BCT* System y = 1.06 x - 0.03 0.999
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.1 1.6 1.9
Control Plasma P 1.5 1.1 1.8

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
Not applicable.

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 36


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT INR (SW 2.02) with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 9 Method PT-Innovin
Unit INR cuvette ON
Preset No. 9
1st convers reference curve
Pipetting Scheme falling
1. Sample 50 µL unit %
decimal place xxx.x
Incubation time 30 sec
min / max value 5.0 / 120.0 %
Start reagent
time interpolat. linear
Dade® Innovin® Reagent 100 µL value interpolat. reciprocal
Autostart yes
2nd convers INR
100.0% * xx.x s
factor --- **
ISI * x.xx

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 30 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 3 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
*
System specific MNPT and ISI have to be used.
ISI values for prothrombin time assays must be entered directly as they appear on the current reagent labeling. Any changes of
reagent lot, software upgrades, major servicing, etc., require verification of the ISI value. Failure to enter the correct ISI value will
cause incorrect International Normalization Ratio (INR) results.
In order to report PT % and PT INR “1st conversion” in the BFT*II Analyzer Parameter-Protocol has to be set to “reference curve”. “2nd
conversion” has to be set to “INR”.
To obtain INR without PT %, set “1st conversion” to “none” and “2nd conversion” to “INR”.
The Mean Normal PT (MNPT) can be established with Standard Human Plasma. To use Standard Human Plasma, it has to be measured at
least twice and the obtained clotting time has to be divided by the ratio given for the PT reagent in the “Table of Analytical Values” of the
respective Lot. No. of the plasma.
** This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 37


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT INR calibrated (SW 2.01) with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
PT-Multi Calibrator OPAT 6 x 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
® ®
Dade Innovin Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 453
Hemoglobin mg/dL 600
Bilirubin mg/dL 12

Reference Guide 3.01 - not for use in the USA - 38


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT INR calibrated (SW 2.01) with
® ®
Dade Innovin

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
® ®
PT INR calibrated with Dade Innovin Reagent on BFT* // Analyzer and PT INR using the y = 1.00 x + 0.01 0.999
ISI value with Dade® Innovin® Reagent on BFT* // Analyzer.
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.8 1.3 2.1
Control Plasma P 4.6 1.6 4.6

Measuring Range
The measuring range is defined by the concentration of the calibrators used.

Reference Interval
Not applicable.

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 39


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT INR calibrated (SW 2.01) with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit INR cuvette ON
Preset No. 1
1st convers reference curve
Pipetting Scheme rising
1. Sample 50 µL unit INR
decimal place xx.xx
Incubation time 30 sec
min / max value 0.10 / 10.00 INR
Start reagent
time interpolat. logarithmic
Dade® Innovin® Reagent 100 µL value interpolat. logarithmic
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 30 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The standard curve is measured in duplicate.

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
PT-Multi Calibrator
Calibrator INR sec
Calibrator 1 0.95 8.8
Calibrator 2 1.21 11.5
Calibrator 3 1.56 15.1
Calibrator 4 2.40 24.2
Calibrator 5 4.17 44.1
Calibrator 6 5.52 59.8

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 40


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT INR calibrated (SW 2.02) with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
PT-Multi Calibrator OPAT 6 x 1 mL
BC Vial Kit see Additional Notes OVKE 5 mL see Additional Notes
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Appropriately sized aliquots may be prepared using vials from the BC Vial Kit. The aliquots are stable at 2-8 °C for up to 5 days. Before use
the reagent should be mixed carefully and brought to 37 °C. Each vial can be used once up to 2 hours on the BFT* II Analyzer. Discard the
vial after use.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
® ®
Dade Innovin Reagent 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 637
Hemoglobin mg/dL 600
Bilirubin mg/dL 60

Reference Guide 3.01 - not for use in the USA - 41


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT INR calibrated (SW 2.02) with
® ®
Dade Innovin

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
® ®
PT INR calibrated with Dade Innovin Reagent on BFT* // Analyzer and PT INR using the y = 1.00 x + 0.01 1.000
ISI value with Dade® Innovin® Reagent on BFT* // Analyzer.
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 0.8 1.9 2.1
Control Plasma P 1.6 1.1 1.9

Measuring Range
The measuring range is defined by the concentration of the calibrators used.

Reference Interval
Not applicable.

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 42


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT INR calibrated (SW 2.02) with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 9 Method PT-Innovin
Unit INR cuvette ON
Preset No. 9
1st convers reference curve
Pipetting Scheme rising
1. Sample 50 µL unit INR
decimal place xx.xx
Incubation time 30 sec
min / max value 0.10 / 10.00 INR
Start reagent
time interpolat. logarithmic
Dade® Innovin® Reagent 100 µL value interpolat. logarithmic
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 30 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 3 s
learn 3 s
lag 1 s

Standard Curve Calibration


The standard curve is measured in duplicate.

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
PT-Multi Calibrator
Calibrator INR sec
Calibrator 1 1.01 9.0
Calibrator 2 1.21 11.2
Calibrator 3 1.61 15.1
Calibrator 4 2.56 24.8
Calibrator 5 4.19 42.4
Calibrator 6 5.29 52.8

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 43


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT seconds with
®
Dade Thromboplastin C Plus
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Dade Thromboplastin C Plus Reagent Reagent position 1 B4216 4/10 mL Reagent position 1
Dade® Ci-Trol® Coagulation Control Level 1 B4244-10 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol Coagulation Control Level 2 B4244-20 1 mL
Dade® Ci-Trol® Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Magnetic stir bar — 718.88131 10 —
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Be sure to place a magnetic stir bar in the reagent bottle. Magnetic stir bars can be ordered as spare part (part no. 718.88131) at Siemens
Service.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Before use the reagent should be mixed carefully and brought to 37°C.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Thromboplastin C Plus Reagent 4
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 600
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
® ®
Dade Thromboplastin C Plus Reagent on Sysmex CA-6000 System y = 1.04 x - 2.05 0.956
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 44


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT seconds with
®
Dade Thromboplastin C Plus

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.4 1.4 2.0
Control Plasma P 2.4 0.8 2.4

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
Comments n Mean Median 5th - 95th Percentile
— 60 9.8 9.8 9.2 - 10.6
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 45


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT seconds with
®
Dade Thromboplastin C Plus

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit sec cuvette ON
Preset No. 1
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec. *
decimal place xxx.x *
Incubation time 60 sec
min / max value 5.0 / 150.0% *
Start reagent
time interpolat. linear *
Dade® Thromboplastin C Plus Reagent 100 µL value interpolat. reciprocal *
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 46


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT INR with
®
Dade Thromboplastin C Plus
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Dade Thromboplastin C Plus Reagent Reagent position 1 B4216 4/10 mL Reagent position 1
Dade® Ci-Trol® Coagulation Control Level 1 B4244-10 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol Coagulation Control Level 2 B4244-20 1 mL
Dade® Ci-Trol® Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Magnetic stir bar — 718.88131 10 —
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Be sure to place a magnetic stir bar in the reagent bottle. Magnetic stir bars can be ordered as spare part (part no. 718.88131) at Siemens
Service.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent should be mixed carefully and brought to 37°C.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Thromboplastin C Plus Reagent 4
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 200
Bilirubin mg/dL 6

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
® ®
Dade Thromboplastin C Plus Reagent on Sysmex CA-6000 System y = 0.98 x - 0.00 0.876
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 47


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT INR with
®
Dade Thromboplastin C Plus

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 2.9 2.8 3.9
Control Plasma P 4.8 1.6 4.7

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
Not applicable.

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 48


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT INR with
®
Dade Thromboplastin C Plus

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit INR cuvette ON
Preset No. 1
1st convers none
Pipetting Scheme falling **
1. Sample 50 µL unit % **
decimal place xxx.x **
Incubation time 60 sec
min / max value 5.0 / 150.0 % **
Start reagent
time interpolat. linear **
Dade® Thromboplastin C Plus Reagent 100 µL value interpolat. reciprocal **
Autostart yes
2nd convers INR
100.0% * xx.x s
factor --- **
ISI * x.xx

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
*
System specific MNPT and ISI have to be used.
ISI values for prothrombin time assays must be entered directly as they appear on the current reagent labeling. Any changes of
reagent lot, software upgrades, major servicing, etc., require verification of the ISI value. Failure to enter the correct ISI value will
cause incorrect International Normalization Ratio (INR) results.
“1st conversion” in the BFT* II Analyzer Parameter-Protocol has to be set to “none” and “2nd conversion” to “INR”.
** This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 49


BFT* II Analyzer
Application Sheet for Page 1 of 3
PT INR calibrated with
®
Dade Thromboplastin C Plus
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Dade Thromboplastin C Plus Reagent Reagent position 1 B4216 4/10 mL Reagent position 1
Dade® Ci-Trol® Coagulation Control Level 1 B4244-10 1 mL Beside the BFT* II Analyzer
® ®
Dade Ci-Trol Coagulation Control Level 2 B4244-20 1 mL
Dade® Ci-Trol® Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
PT-Multi Calibrator OPAT 6 x 1 mL
Magnetic stir bar — 718.88131 10 —
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Be sure to place a magnetic stir bar in the reagent bottle. Magnetic stir bars can be ordered as spare part (part no. 718.88131) at Siemens
Service.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent should be mixed carefully and brought to 37°C.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
®
Dade Thromboplastin C Plus Reagent 4
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 400
Bilirubin mg/dL 6

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
®
PT INR using the ISI value with Dade Thromboplastin C Plus Reagent on BFT* // Analyzer y = 1.24 x - 0.43 0.999
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 50


BFT* II Analyzer
Application Sheet for Page 2 of 3
PT INR calibrated with
®
Dade Thromboplastin C Plus

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 10 %.
Within Run Run to Run Total
% % %
Control Plasma N 2.4 1.1 2.5
Control Plasma P 5.5 1.4 5.4

Measuring Range
The measuring range is defined by the concentration of the calibrators used.

Reference Interval
Not applicable.

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 51


BFT* II Analyzer
Application Sheet for Page 3 of 3
PT INR calibrated with
®
Dade Thromboplastin C Plus

Test Protocol Parameter-Protocol


Method store 1 Method PT
Unit INR cuvette ON
Preset No. 1
1st convers reference curve
Pipetting Scheme rising
1. Sample 50 µL unit INR
decimal place xx.xx
Incubation time 60 sec
min / max value 0.10 / 10.00 INR
Start reagent
time interpolat. logarithmic
Dade® Thromboplastin C Plus Reagent 100 µL value interpolat. logarithmic
Autostart yes
2nd convers none
100.0% 11.0 s *
factor --- *
ISI 1.14 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The standard curve is measured in duplicate.

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
PT-Multi Calibrator
Calibrator INR sec
Calibrator 1 0.90 10.1
Calibrator 2 1.16 11.5
Calibrator 3 1.50 13.3
Calibrator 4 2.39 17.2
Calibrator 5 5.12 26.1
Calibrator 6 7.57 32.3

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 52


BFT* II Analyzer
Application Sheet for Page 1 of 3
APTT with
® ®
Dade Actin Activated Cephaloplastin Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Actin Activated Cephaloplastin Reagent B4218-1/-2 2/10 mL Beside the BFT* II Analyzer [1]
Dade® Ci-Trol® 1 291070 1 mL
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Calcium Chloride Solution 24
® ®
Dade Actin Reagent (RT) 24

The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 600
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Reference Guide 3.01 - not for use in the USA - 53


BFT* II Analyzer
Application Sheet for Page 2 of 3
APTT with
® ®
Dade Actin Activated Cephaloplastin Reagent

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 0.73 x + 3.33 0.900
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.0 0.7 1.2
Control Plasma P 1.8 1.1 2.0

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: s
th th
Comments n Mean Median 5 - 95 Percentile
— 110 25.6 25.5 22.2 - 29.8
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 54


BFT* II Analyzer
Application Sheet for Page 3 of 3
APTT with
® ®
Dade Actin Activated Cephaloplastin Reagent

Test Protocol Parameter-Protocol


Method store 2 Method aPTT
Unit sec cuvette ON
Preset No. 2
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec *
® ® decimal place xxx.x *
2. Dade Actin Activated Cephaloplastin Reagent 50 µL
min / max value 0.0 / 0.0 *
Incubation time 180 sec time interpolat. linear *
Start reagent value interpolat. linear *
Calcium Chloride Solution 50 µL
2nd convers none
Autostart yes
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 180 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 55


BFT* II Analyzer
Application Sheet for Page 1 of 3
APTT with
® ®
Dade Actin FS Activated PTT Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Actin FS Activated PTT Reagent B4218-20/-100 2/10 mL Beside the BFT* II Analyzer [1]
Dade® Ci-Trol® 1 291070 1 mL
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Calcium Chloride Solution 24
® ®
Dade Actin FS Reagent (RT) 24

The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 300
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Reference Guide 3.01 - not for use in the USA - 56


BFT* II Analyzer
Application Sheet for Page 2 of 3
APTT with
® ®
Dade Actin FS Activated PTT Reagent

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 1.01 x - 2.18 0.941
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.0 0.7 1.2
Control Plasma P 1.1 1.3 1.7

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
th th
Comments n Mean Median 5 - 95 Percentile
— 110 26.2 26.2 22.4 - 29.3
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 57


BFT* II Analyzer
Application Sheet for Page 3 of 3
APTT with
® ®
Dade Actin FS Activated PTT Reagent

Test Protocol Parameter-Protocol


Method store 2 Method aPTT
Unit sec cuvette ON
Preset No. 2
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec *
® ® decimal place xxx.x *
2. Dade Actin FS Activated PTT Reagent 50 µL
min / max value 0.0 / 0.0 *
Incubation time 180 sec time interpolat. linear *
Start reagent value interpolat. linear *
Calcium Chloride Solution 50 µL
2nd convers none
Autostart yes
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 180 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 58


BFT* II Analyzer
Application Sheet for Page 1 of 3
APTT with
® ®
Dade Actin FSL Activated PTT Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Actin FSL Activated PTT Reagent B4219 2/10 mL Beside the BFT* II Analyzer [1]
Dade® Ci-Trol® 1 291070 1 mL
® ®
Dade Ci-Trol 2 291071 1 mL
Dade® Ci-Trol® 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
Dade® Ci-Trol® Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Calcium Chloride Solution 24
® ®
Dade Actin FSL Reagent (RT) 24
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 600
Hemoglobin mg/dL 200
Bilirubin mg/dL 60

Reference Guide 3.01 - not for use in the USA - 59


BFT* II Analyzer
Application Sheet for Page 2 of 3
APTT with
® ®
Dade Actin FSL Activated PTT Reagent

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 0.84 x + 3.16 0.947
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.6 1.2 1.9
Dade® Ci-Trol® 3 1.4 1.1 1.7

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
th th
Comments n Mean Median 5 - 95 Percentile
— 110 27.8 27.8 24.0 - 32.6
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 60


BFT* II Analyzer
Application Sheet for Page 3 of 3
APTT with
® ®
Dade Actin FSL Activated PTT Reagent

Test Protocol Parameter-Protocol


Method store 2 Method aPTT
Unit sec cuvette ON
Preset No. 2
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec *
® ® decimal place xxx.x *
2. Dade Actin FSL Activated PTT Reagent 50 µL
min / max value 0.0 / 0.0 *
Incubation time 180 sec time interpolat. linear *
Start reagent value interpolat. linear *
Calcium Chloride Solution 50 µL
2nd convers none
Autostart yes
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 180 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 61


BFT* II Analyzer
Application Sheet for Page 1 of 3
APTT with
Pathromtin* SL
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Pathromtin* SL Reagent OQGS 5 mL Beside the BFT* II Analyzer [1]
® ®
Dade Ci-Trol 1 291070 1 mL
® ®
Dade Ci-Trol 2 291071 1 mL
® ®
Dade Ci-Trol 3 291072 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 1 B4244-10 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 2 B4244-20 1 mL
® ®
Dade Ci-Trol Coagulation Control Level 3 B4244-30 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Calcium Chloride Solution 24
Pathromtin* SL Reagent (RT) 24
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 600
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Reference Guide 3.01 - not for use in the USA - 62


BFT* II Analyzer
Application Sheet for Page 2 of 3
APTT with
Pathromtin* SL

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 1.06 x - 1.10 0.969
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 2.3 0.8 2.3
Control Plasma P 3.2 2.3 3.8

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
th th
Comments n Mean Median 5 - 95 Percentile
— 110 32.0 31.7 27.1 - 38.8
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 63


BFT* II Analyzer
Application Sheet for Page 3 of 3
APTT with
Pathromtin* SL

Test Protocol Parameter-Protocol


Method store 2 Method aPTT
Unit sec cuvette ON
Preset No. 2
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec *
decimal place xxx.x *
2. Pathromtin* SL Reagent 50 µL
min / max value 0.0 / 0.0 *
Incubation time 120 sec time interpolat. linear *
Start reagent value interpolat. linear *
Calcium Chloride Solution 50 µL
2nd convers none
Autostart yes
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 64


BFT* II Analyzer
Application Sheet for Page 1 of 4
Fibrinogen with
Multifibren* U
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Multifibren* U Reagent Reagent position 1 OWZG 2/5 mL Reagent position 1
Kaolin Suspension for Fibrintimer OQAB 50 mL Beside the BFT* II Analyzer
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Fibrinogen Calibrator Kit OQVK 6 x 1 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Before use the Kaolin Suspension has to be mixed well to get a homogeneous suspension.
Reconstitute Multifibren* U Reagent with an equal volume of Kaolin Suspension for Fibrintimer.
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).
Limitations
Blood plasma substitutes that contain hydroxyethyl starch (HES) may interfere with the analysis. In a study with a HES solution (130/0.4) no
interference was observed up to 15 g HES per liter plasma.

On-board Stability
Material Time [h]
Multifibren* U Reagent 4
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 600
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Multifibren* U Reagent on BCT* System y = 1.00 x + 0.03 0.968
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 65


BFT* II Analyzer
Application Sheet for Page 2 of 4
Fibrinogen with
Multifibren* U

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
Control Plasma N 1.4 4.0 4.2
Control Plasma P 2.8 3.3 4.2

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: g/L
Comments n Mean Median 5th - 95th Percentile
— 149 2.8 2.7 2.0 - 4.0
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 66


BFT* II Analyzer
Application Sheet for Page 3 of 4
Fibrinogen with
Multifibren* U

Test Protocol Parameter-Protocol


Method store 5 Method Fib. g/L
Unit g/L cuvette ON
Preset No. 5
1st convers reference curve
Pipetting Scheme falling
1. Sample 50 µL unit g/L
decimal place xx.xx
Incubation time 60 sec
min / max value 0.20 / 10.00 g/L
Start reagent
time interpolat. logarithmic
Multifibren* U Reagent 100 µL value interpolat. logarithmic
Autostart yes
2nd convers none
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 40 s
1 500 rpm
2 200 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The standard curve is measured in duplicate.

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Fibrinogen Calibrator Kit
Calibrator g/L sec
Fibrinogen Calibrator 1 0.63 133.0
Fibrinogen Calibrator 2 1.10 66.4
Fibrinogen Calibrator 3 2.50 25.4
Fibrinogen Calibrator 4 3.80 16.9
Fibrinogen Calibrator 5 5.60 11.1
Fibrinogen Calibrator 6 9.30 6.9

Remarks
Do not use calibration values as supplied in the “Table for the evaluation of Fibrinogen determination with Multifibren* U Reagent” for BFT*
II Analyzer, since these values are only for the old Behring Fibrintimer.
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 67


BFT* II Analyzer
Application Sheet for Page 4 of 4
Fibrinogen with
Multifibren* U

Test Protocol for mg/dL:

Test Protocol Parameter-Protocol


Method store 6 Method Fib. mg/dL
Unit mg/dL cuvette ON
Preset No. 6
1st convers reference curve
falling
Pipetting Scheme
unit mg/dL
1. Sample 50 µL
decimal place xxxx.
Incubation time 60 sec
min / max value 20 / 1000 mg/dL
Start reagent time interpolat. logarithmic
Multifibren* U Reagent 100 µL value interpolat. logarithmic
Autostart yes
2nd convers none
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 40 s
1 500 rpm
2 200 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


See previous page.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 68


BFT* II Analyzer
Application Sheet for Page 1 of 3
Thrombin Time with
Test Thrombin Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Test Thrombin Reagent — OWHM Kit —
REAGENT DILUENT — — 50 mL —
TEST THROMBIN REAGENT Reagent positions 1 - 4 — 5 mL Reagent positions 1 - 4
® ®
Dade Ci-Trol 1 291070
Beside the BFT* II Analyzer 1 mL Beside the BFT* II Analyzer
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Test Thrombin Reagent 6
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 600
Hemoglobin mg/dL 1000
Bilirubin mg/dL 36

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Test Thrombin Reagent on BFA* System y = 0.98 x -6.17 0.950
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 10 %.
Within Run Run to Run Total
% % %
Control Plasma N 3.2 2.8 4.1

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Guide 3.01 - not for use in the USA - 69


BFT* II Analyzer
Application Sheet for Page 2 of 3
Thrombin Time with
Test Thrombin Reagent

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
th th
Comments n Mean Median 5 - 95 Percentile
— 76 12.3 11.7 10.4 - 16.1
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 70


BFT* II Analyzer
Application Sheet for Page 3 of 3
Thrombin Time with
Test Thrombin Reagent

Test Protocol Parameter-Protocol


Method store 3 Method Thrombin Time
Unit sec cuvette ON
Preset No. 3
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec *
decimal place xxx.x *
Incubation time 60 sec
min / max value 0.0 / 0.0 *
Start reagent
time interpolat. linear *
Test Thrombin Reagent 100 µL value interpolat. linear *
Autostart yes
2nd convers none
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 71


BFT* II Analyzer
Application Sheet for Page 1 of 4
Thrombin Time with
Thromboclotin*
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Thromboclotin* Reagent Reagent positions 1 -281007
4 10 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070
Beside the BFT* II Analyzer 1 mL Beside the BFT* II Analyzer
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
If expected heparin concentration of the sample is low, the method TT Thromboclotin has to be used.
If expected heparin concentration of the sample is high, the method TT Thromboclotin Hep. high has to be used.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Thromboclotin* Reagent 1
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 300 300*
Hemoglobin mg/dL 600 1000*
Bilirubin mg/dL 60 60*
* Values for TT Thromboclotin / TT Thromboclotin Hep. high

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 10 %.
Within Run Run to Run Total
% % %
For TT Thromboclotin: Control Plasma N 3.8 1.7 3.9
For TT Thromboclotin Hep.high: Control Plasma N 5.9 2.5 6.0

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Guide 3.01 - not for use in the USA - 72


BFT* II Analyzer
Application Sheet for Page 2 of 4
Thrombin Time with
Thromboclotin*

Method Comparison
Predicate Device Regression Equation r
®
For TT Thromboclotin with Thromboclotin* Reagent on Sysmex CA-6000 System y = 1.42 x - 8.91 0.849
For TT Thromboclotin Hep.high with Thromboclotin* Reagent on Sysmex® CA-6000 y = 0.90 x - 3.13 0.941
System
r = Correlation Coefficient
Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
th th
Comments n Mean Median 5 - 95 Percentile
for TT Thromboclotin 76 12.7 13.0 10.3 - 15.0
For TT Thromboclotin Hep.high 76 11.4 11.0 9.3 - 14.2
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 73


BFT* II Analyzer
Application Sheet for Page 3 of 4
Thrombin Time with
Thromboclotin*

TT Thromboclotin:

Test Protocol Parameter-Protocol


Method store 3 Method Thrombin Time
Unit sec cuvette ON
Preset No. 3
1st convers none
falling *
Pipetting Scheme
unit sec *
1. Sample 100 µL
decimal place xxx.x *
Incubation time 60 sec
min / max value 0.0 / 0.0 *
Start reagent time interpolat. linear *
Thromboclotin* Reagent 100 µL value interpolat. linear *
Autostart yes
2nd convers none
100.0% 0 s*
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 74


BFT* II Analyzer
Application Sheet for Page 4 of 4
Thrombin Time with
Thromboclotin*

TT Thromboclotin Hep. high:

Test Protocol Parameter-Protocol


Method store 3 Method Thrombin Time
Unit sec cuvette ON
Preset No. 3
1st convers none
falling *
Pipetting Scheme
unit sec *
1. Sample 80 µL
decimal place xxx.x *
Incubation time 60 sec
min / max value 0.0 / 0.0 *
Start reagent time interpolat. linear *
Thromboclotin* Reagent 140 µL value interpolat. linear *
Autostart yes
2nd convers none
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 140 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 75


BFT* II Analyzer
Application Sheet for Page 1 of 3
Batroxobin Time with
Batroxobin Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Batroxobin Reagent Reagent positions 1 -OUOV
4 5 mL Reagent positions 1 - 4
Dade® Ci-Trol® 1 291070
Beside the BFT* II Analyzer 1 mL Beside the BFT* II Analyzer
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Batroxobin Reagent 8
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 1800
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Batroxobin Reagent on BCT* System y = 0.98 x - 2.26 0.961
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 10 %.
Within Run Run to Run Total
% % %
Dade® Ci-Trol® 1 2.4 2.1 3.1

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Guide 3.01 - not for use in the USA - 76


BFT* II Analyzer
Application Sheet for Page 2 of 3
Batroxobin Time with
Batroxobin Reagent

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
th th
Comments n Mean Median 5 - 95 Percentile
— 69 12.9 12.8 11.9 - 14.1
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 77


BFT* II Analyzer
Application Sheet for Page 3 of 3
Batroxobin Time with
Batroxobin Reagent

Test Protocol Parameter-Protocol


Method store 4 Method Batroxobin Time
Unit sec cuvette ON
Preset No. 4
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec *
decimal place xxx.x *
Incubation time 60 sec
min / max value 0.0 / 0.0 *
Start reagent
time interpolat. linear *
Batroxobin Reagent 100 µL value interpolat. linear *
Autostart yes
2nd convers none
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 78


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor II with
®
Thromborel S
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Thromborel S Reagent Reagent positions 1 -OUHP
4 4/10 mL Reagent positions 1 - 4
Coagulation Factor II Deficient Plasma OSGR 1 mL Beside the BFT* II Analyzer [1]
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Thromborel® S Reagent 8
FII Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 8
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Thromborel® S Reagent on BCT* System y = 1.19 x - 2.39 0.977
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 79


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor II with
®
Thromborel S

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 6.6 6.7 9.1
Control Plasma P 4.2 3.9 5.6

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 80


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor II with
®
Thromborel S

Test Protocol Parameter-Protocol


Method store 7 Method Extr. Factor
Unit % cuvette ON
Preset No. 7
1st convers reference curve
Predilution 1:20 falling
Sample 10 µL unit %
Imidazole Buffer Solution 190 µL decimal place xxx.x
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor II Deficient Plasma 50 µL 2nd convers none
Incubation time 60 sec 100.0% 0.0 s *
Start reagent factor --- *
Thromborel® S Reagent 100 µL ISI 0.00 *
Autostart yes measurement lot. no.: xxxxxx
Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:16 10 --- 150 ---
1.0 1:20 10 --- 190 ---
0.8 1:25 20 --- 480 ---
0.6 1:33.3 --- 150 50 ---
0.4 1:50 --- 100 100 ---
0.15 1:133.3 --- 30 130 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 81


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor II with
®
Thromborel S

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
145.5 18.8
97.0 21.9
72.8 25.2
48.5 30.0
32.0 36.1
24.3 42.2
12.1 59.7

Reference Guide 3.01 - not for use in the USA - 82


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor II with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Coagulation Factor II Deficient Plasma OSGR 1 mL Beside the BFT* II Analyzer [1]
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Innovin® Reagent 3
FII Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 3
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Thromborel® S Reagent on BCT* System y = 1.06 x + 0.11 0.985
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 83


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor II with
® ®
Dade Innovin

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 3.8 2.4 4.3
Control Plasma P 3.6 2.5 4.3

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 84


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor II with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 7 Method Extr. Factor
Unit % cuvette ON
Preset No. 7
1st convers reference curve
Predilution 1:20 falling
Sample 10 µL unit %
Imidazole Buffer Solution 190 µL decimal place xxx.x
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor II Deficient Plasma 50 µL 2nd convers none
Incubation time 60 sec 100.0% 0.0 s *
Start reagent factor --- *
Dade® Innovin® Reagent 100 µL ISI 0.00 *
Autostart yes measurement lot. no.: xxxxxx
Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:16 10 --- 150 ---
1.0 1:20 10 --- 190 ---
0.8 1:25 20 --- 480 ---
0.6 1:33.3 --- 150 50 ---
0.4 1:50 --- 100 100 ---
0.15 1:133.3 --- 30 130 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 85


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor II with
® ®
Dade Innovin

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
145.5 17.3
97.0 21.3
72.8 24.8
48.5 30.1
32.0 37.3
24.3 45.7
12.1 74.7

Reference Guide 3.01 - not for use in the USA - 86


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor V with
®
Thromborel S
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Thromborel S Reagent Reagent positions 1 -OUHP
4 4/10 mL Reagent positions 1 - 4
Coagulation Factor V Deficient Plasma ORSM 1 mL Beside the BFT* II Analyzer [1]
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Thromborel® S Reagent 3
FV Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 3
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 1000
Bilirubin mg/dL 24

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Thromborel® S Reagent on BCT* System y = 0.87 x + 1.10 0.944
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 87


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor V with
®
Thromborel S

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 7.9 3.4 8.1
Control Plasma P 8.3 5.1 9.3

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 88


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor V with
®
Thromborel S

Test Protocol Parameter-Protocol


Method store 7 Method Extr. Factor
Unit % cuvette ON
Preset No. 7
1st convers reference curve
Predilution 1:20 falling
Sample 10 µL unit %
Imidazole Buffer Solution 190 µL decimal place xxx.x
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor V Deficient Plasma 50 µL 2nd convers none
Incubation time 60 sec 100.0% 0.0 s *
Start reagent factor --- *
Thromborel® S Reagent 100 µL ISI 0.00 *
Autostart yes measurement lot. no.: xxxxxx
Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:16 10 --- 150 ---
1.0 1:20 10 --- 190 ---
0.8 1:25 20 --- 480 ---
0.6 1:33.3 --- 150 50 ---
0.4 1:50 --- 100 100 ---
0.15 1:133.3 --- 30 130 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 89


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor V with
®
Thromborel S

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
138.0 21.2
92.0 23.6
69.0 25.6
46.0 28.8
30.4 32.5
23.0 36.1
11.5 43.7

Reference Guide 3.01 - not for use in the USA - 90


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor V with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Coagulation Factor V Deficient Plasma ORSM 1 mL Beside the BFT* II Analyzer [1]
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Innovin® Reagent 6
FV Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 6
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 1000
Bilirubin mg/dL 6

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Thromborel® S Reagent on BCT* System y = 0.89 x + 1.03 0.920
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 91


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor V with
® ®
Dade Innovin

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 6.9 6.6 9.2
Control Plasma P 7.3 9.9 12.1

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 92


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor V with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 7 Method Extr. Factor
Unit % cuvette ON
Preset No. 7
1st convers reference curve
Predilution 1:20 falling
Sample 10 µL unit %
Imidazole Buffer Solution 190 µL decimal place xxx.x
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor V Deficient Plasma 50 µL 2nd convers none
Incubation time 60 sec 100.0% 0.0 s *
Start reagent factor --- *
Dade® Innovin® Reagent 100 µL ISI 0.00 *
Autostart yes measurement lot. no.: xxxxxx
Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:16 10 --- 150 ---
1.0 1:20 10 --- 190 ---
0.8 1:25 20 --- 480 ---
0.6 1:33.3 --- 150 50 ---
0.4 1:50 --- 100 100 ---
0.15 1:133.3 --- 30 130 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 93


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor V with
® ®
Dade Innovin

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
138.0 15.9
92.0 17.5
69.0 19.0
46.0 21.2
30.4 23.9
23.0 25.7
11.5 31.4

Reference Guide 3.01 - not for use in the USA - 94


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor VII with
®
Thromborel S
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Thromborel S Reagent Reagent positions 1 -OUHP
4 4/10 mL Reagent positions 1 - 4
Coagulation Factor VII Deficient Plasma OTXV 1 mL Beside the BFT* II Analyzer [1]
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Thromborel® S Reagent 3
FVII Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 3
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 3038
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Thromborel® S Reagent on BCT* System y = 1.11 x + 1.07 0.984
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 95


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor VII with
®
Thromborel S

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 6.5 4.4 7.5
Control Plasma P 5.8 6.6 8.6

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 96


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor VII with
®
Thromborel S

Test Protocol Parameter-Protocol


Method store 7 Method Extr. Factor
Unit % cuvette ON
Preset No. 7
1st convers reference curve
Predilution 1:20 falling
Sample 10 µL unit %
Imidazole Buffer Solution 190 µL decimal place xxx.x
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor VII Deficient Plasma 50 µL 2nd convers none
Incubation time 60 sec 100.0% 0.0 s *
Start reagent factor --- *
Thromborel® S Reagent 100 µL ISI 0.00 *
Autostart yes measurement lot. no.: xxxxxx
Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:16 10 --- 150 ---
1.0 1:20 10 --- 190 ---
0.8 1:25 20 --- 480 ---
0.6 1:33.3 --- 150 50 ---
0.4 1:50 --- 100 100 ---
0.15 1:133.3 --- 30 130 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 97


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor VII with
®
Thromborel S

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
139.5 23.2
93.0 26.1
69.8 28.5
46.5 32.3
30.7 36.6
23.3 40.3
11.6 50.2

Reference Guide 3.01 - not for use in the USA - 98


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor VII with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Coagulation Factor VII Deficient Plasma OTXV 1 mL Beside the BFT* II Analyzer [1]
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Innovin® Reagent 3
FVII Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 3
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 2400
Hemoglobin mg/dL 1000
Bilirubin mg/dL 48

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Thromborel® S Reagent on BCT* System y = 0.98 x + 1.60 0.986
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 99


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor VII with
® ®
Dade Innovin

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 6.1 5.8 8.1
Control Plasma P 5.9 3.3 6.5

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 100


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor VII with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 7 Method Extr. Factor
Unit % cuvette ON
Preset No. 7
1st convers reference curve
Predilution 1:20 falling
Sample 10 µL unit %
Imidazole Buffer Solution 190 µL decimal place xxx.x
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor VII Deficient Plasma 50 µL 2nd convers none
Incubation time 60 sec 100.0% 0.0 s *
Start reagent factor --- *
Dade® Innovin® Reagent 100 µL ISI 0.00 *
Autostart yes measurement lot. no.: xxxxxx
Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:16 10 --- 150 ---
1.0 1:20 10 --- 190 ---
0.8 1:25 20 --- 480 ---
0.6 1:33.3 --- 150 50 ---
0.4 1:50 --- 100 100 ---
0.15 1:133.3 --- 30 130 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 101


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor VII with
® ®
Dade Innovin

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
139.5 19.5
93.0 22.3
69.8 25.1
46.5 29.9
30.7 34.9
23.3 40.2
11.6 54.2

Reference Guide 3.01 - not for use in the USA - 102


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor X with
®
Thromborel S
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
Thromborel S Reagent Reagent positions 1 -OUHP
4 4/10 mL Reagent positions 1 - 4
Coagulation Factor X Deficient Plasma OTXY 1 mL Beside the BFT* II Analyzer [1]
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Thromborel® S Reagent 6
FX Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 6
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 1000
Bilirubin mg/dL 48

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Thromborel® S Reagent on BCT* System y = 0.95 x - 0.72 0.986
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 103


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor X with
®
Thromborel S

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 4.6 8.3 9.4
Control Plasma P 3.7 7.9 8.7

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 104


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor X with
®
Thromborel S

Test Protocol Parameter-Protocol


Method store 7 Method Extr. Factor
Unit % cuvette ON
Preset No. 7
1st convers reference curve
Predilution 1:20 falling
Sample 10 µL unit %
Imidazole Buffer Solution 190 µL decimal place xxx.x
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor X Deficient Plasma 50 µL 2nd convers none
Incubation time 60 sec 100.0% 0.0 s *
Start reagent factor --- *
Thromborel® S Reagent 100 µL ISI 0.00 *
Autostart yes measurement lot. no.: xxxxxx
Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:16 10 --- 150 ---
1.0 1:20 10 --- 190 ---
0.8 1:25 20 --- 480 ---
0.6 1:33.3 --- 150 50 ---
0.4 1:50 --- 100 100 ---
0.15 1:133.3 --- 30 130 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 105


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor X with
®
Thromborel S

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
148.5 22.9
99.0 26.8
74.3 30.3
49.5 35.7
32.7 41.7
24.8 48.6
12.4 68.5

Reference Guide 3.01 - not for use in the USA - 106


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor X with
® ®
Dade Innovin
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Innovin Reagent Reagent positions 1 -B4212
4 4/10/20 mL Reagent positions 1 - 4
Coagulation Factor X Deficient Plasma OTXY 1 mL Beside the BFT* II Analyzer [1]
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Innovin® Reagent 3
FX Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 3
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Thromborel® S Reagent on BCT* System y = 0.93 x - 3.60 0.982
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 107


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor X with
® ®
Dade Innovin

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 5.8 9.1 10.6
Control Plasma P 3.2 5.3 6.1

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 108


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor X with
® ®
Dade Innovin

Test Protocol Parameter-Protocol


Method store 7 Method Extr. Factor
Unit % cuvette ON
Preset No. 7
1st convers reference curve
Predilution 1:20 falling
Sample 10 µL unit %
Imidazole Buffer Solution 190 µL decimal place xxx.x
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor X Deficient Plasma 50 µL 2nd convers none
Incubation time 60 sec 100.0% 0.0 s *
Start reagent factor --- *
Dade® Innovin® Reagent 100 µL ISI 0.00 *
Autostart yes measurement lot. no.: xxxxxx
Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:16 10 --- 150 ---
1.0 1:20 10 --- 190 ---
0.8 1:25 20 --- 480 ---
0.6 1:33.3 --- 150 50 ---
0.4 1:50 --- 100 100 ---
0.15 1:133.3 --- 30 130 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 109


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor X with
® ®
Dade Innovin

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
148.5 18.6
99.0 22.1
74.3 24.6
49.5 28.5
32.7 32.9
24.8 38.6
12.4 53.1

Reference Guide 3.01 - not for use in the USA - 110


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor VIII with
® ®
Dade Actin Activated Cephaloplastin Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Actin Activated Cephaloplastin Reagent B4218-1/-2 2/10 mL Beside the BFT* II Analyzer [1]
Coagulation Factor VIII Deficient Plasma OTXW 1 mL
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
FVIII Deficient Plasma (RT) 8
® ®
Dade Actin Reagent (RT) 8
Imidazole Buffer Solution (RT) 8
Calcium Chloride Solution 8
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 1200
Hemoglobin mg/dL 400
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 0.97 x 5.43 0.946
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 111


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor VIII with
® ®
Dade Actin Activated Cephaloplastin Reagent

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 7.2 3.1 7.4
Control Plasma P 7.3 3.8 7.8

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 150 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 112


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor VIII with
® ®
Dade Actin Activated Cephaloplastin Reagent

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 8 Method Intr. Factor
Unit % cuvette ON
Preset No. 8 1st convers reference curve
Predilution 1:5 falling
unit %
Sample 40 µL
decimal place xxx.x
Imidazole Buffer Solution 160 µL
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor VIII Deficient Plasma 50 µL 2nd convers none
3. Dade® Actin® Activated Cephaloplastin Reagent 50 µL 100.0% 0.0 s *
Incubation time 120 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 1 s
learn 3 s
lag 15 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:4 30 --- 90 ---
1.0 1:5 30 --- 120 ---
0.75 1:6.7 30 --- 170 ---
0.5 1:10 20 --- 180 ---
0.25 1:20 10 --- 190 ---
0.05 1:100 --- 30 120 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 113


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor VIII with
® ®
Dade Actin Activated Cephaloplastin Reagent

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
136.5 42.4
91.0 48.4
68.3 52.4
45.5 59.8
30.0 66.4
22.8 69.2
11.4 78.2

Reference Guide 3.01 - not for use in the USA - 114


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor VIII with
Pathromtin* SL
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Pathromtin* SL Reagent OQGS 5 mL Beside the BFT* II Analyzer [1]
Coagulation Factor VIII Deficient Plasma OTXW 1 mL
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Pathromtin* SL Reagent (RT) 8
FVIII Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 8
Calcium Chloride Solution 8
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 1.10 x - 1.61 0.956
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 115


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor VIII with
Pathromtin* SL

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 9.9 4.7 10.4
Control Plasma P 6.3 2.5 6.4

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 150 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 116


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor VIII with
Pathromtin* SL

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 8 Method Intr. Factor
Unit % cuvette ON
Preset No. 8 1st convers reference curve
Predilution 1:5 falling
unit %
Sample 40 µL
decimal place xxx.x
Imidazole Buffer Solution 160 µL
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor VIII Deficient Plasma 50 µL 2nd convers none
3. Pathromtin* SL Reagent 50 µL 100.0% 0.0 s *
Incubation time 120 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 1 s
learn 3 s
lag 15 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:4 30 --- 90 ---
1.0 1:5 30 --- 120 ---
0.75 1:6.7 30 --- 170 ---
0.5 1:10 20 --- 180 ---
0.25 1:20 10 --- 190 ---
0.05 1:100 --- 30 120 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 117


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor VIII with
Pathromtin* SL

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
136.5 41.4
91.0 45.4
68.3 47.7
45.5 52.3
30.0 57.6
22.8 60.7
11.4 69.6

Reference Guide 3.01 - not for use in the USA - 118


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor IX with
® ®
Dade Actin Activated Cephaloplastin Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Actin Activated Cephaloplastin Reagent B4218-1/-2 2/10 mL Beside the BFT* II Analyzer [1]
Coagulation Factor IX Deficient Plasma OTXX 1 mL
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Actin® Reagent (RT) 8
FIX Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 8
Calcium Chloride Solution 8
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 400
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 0.99 x + 2.59 0.964
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 119


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor IX with
® ®
Dade Actin Activated Cephaloplastin Reagent

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 6.8 3.6 7.3
Control Plasma P 6.3 8.5 10.4

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 120


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor IX with
® ®
Dade Actin Activated Cephaloplastin Reagent

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 8 Method Intr. Factor
Unit % cuvette ON
Preset No. 8 1st convers reference curve
Predilution 1:5 falling
unit %
Sample 40 µL
decimal place xxx.x
Imidazole Buffer Solution 160 µL
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor IX Deficient Plasma 50 µL 2nd convers none
3. Dade® Actin® Activated Cephaloplastin Reagent 50 µL 100.0% 0.0 s *
Incubation time 120 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 1 s
learn 3 s
lag 15 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:4 30 --- 90 ---
1.0 1:5 30 --- 120 ---
0.75 1:6.7 30 --- 170 ---
0.5 1:10 20 --- 180 ---
0.25 1:20 10 --- 190 ---
0.10 1:50 --- 30 120 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 121


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor IX with
® ®
Dade Actin Activated Cephaloplastin Reagent

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
150.0 38.5
100.0 43.2
75.0 46.8
50.0 51.7
33.0 57.7
25.0 61.5
12.5 70.3

Reference Guide 3.01 - not for use in the USA - 122


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor IX with
Pathromtin* SL
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Pathromtin* SL Reagent OQGS 5 mL Beside the BFT* II Analyzer [1]
Coagulation Factor IX Deficient Plasma OTXX 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Standard Human Plasma ORKL 1 mL
Imidazole Buffer Solution OQAA 15 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Pathromtin* SL Reagent (RT) 6
FIX Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 6
Calcium Chloride Solution 6
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 1.00 x + 0.26 0.981
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 123


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor IX with
Pathromtin* SL

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 7.1 2.2 7.1
Control Plasma P 6.4 3.1 6.8

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 124


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor IX with
Pathromtin* SL

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 8 Method Intr. Factor
Unit % cuvette ON
Preset No. 8 1st convers reference curve
Predilution 1:5 falling
unit %
Sample 40 µL
decimal place xxx.x
Imidazole Buffer Solution 160 µL
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor IX Deficient Plasma 50 µL 2nd convers none
3. Pathromtin* SL Reagent 50 µL 100.0% 0.0 s *
Incubation time 120 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 1 s
learn 3 s
lag 15 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:4 30 --- 90 ---
1.0 1:5 30 --- 120 ---
0.75 1:6.7 30 --- 170 ---
0.5 1:10 20 --- 180 ---
0.25 1:20 10 --- 190 ---
0.10 1:50 --- 30 120 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 125


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor IX with
Pathromtin* SL

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
150.0 38.1
100.0 42.0
75.0 45.1
50.0 50.1
33.0 55.0
25.0 59.9
12.5 69.7

Reference Guide 3.01 - not for use in the USA - 126


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor XI with
® ®
Dade Actin Activated Cephaloplastin Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Actin Activated Cephaloplastin Reagent B4218-1/-2 2/10 mL Beside the BFT* II Analyzer [1]
Coagulation Factor XI Deficient Plasma OSDF 1 mL
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Actin® Reagent (RT) 8
FXI Deficient Plasma (RT) 6
Imidazole Buffer Solution (RT) 8
Calcium Chloride Solution 8
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 284
Hemoglobin mg/dL 600
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 0.91 x + 2.98 0.972
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 127


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor XI with
® ®
Dade Actin Activated Cephaloplastin Reagent

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 6.2 5.1 7.7
Control Plasma P 5.9 7.5 9.4

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 128


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor XI with
® ®
Dade Actin Activated Cephaloplastin Reagent

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 8 Method Intr. Factor
Unit % cuvette ON
Preset No. 8 1st convers reference curve
Predilution 1:5 falling
unit %
Sample 40 µL
decimal place xxx.x
Imidazole Buffer Solution 160 µL
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor XI Deficient Plasma 50 µL 2nd convers none
3. Dade® Actin® Activated Cephaloplastin Reagent 50 µL 100.0% 0.0 s *
Incubation time 120 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 1 s
learn 3 s
lag 15 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:4 30 --- 90 ---
1.0 1:5 30 --- 120 ---
0.75 1:6.7 30 --- 170 ---
0.5 1:10 20 --- 180 ---
0.25 1:20 10 --- 190 ---
0.10 1:50 --- 30 120 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 129


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor XI with
® ®
Dade Actin Activated Cephaloplastin Reagent

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
139.5 41.7
93.0 47.2
69.8 51.2
46.5 57.3
30.7 64.1
23.3 69.4
11.6 79.4

Reference Guide 3.01 - not for use in the USA - 130


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor XI with
Pathromtin* SL
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Pathromtin* SL Reagent OQGS 5 mL Beside the BFT* II Analyzer [1]
Coagulation Factor XI Deficient Plasma OSDF 1 mL
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Pathromtin* SL Reagent (RT) 8
FXI Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 8
Calcium Chloride Solution 8
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 1.04 x + 0.89 0.982
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 131


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor XI with
Pathromtin* SL

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 4.3 2.6 4.8
Control Plasma P 5.9 4.2 7.0

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 132


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor XI with
Pathromtin* SL

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 8 Method Intr. Factor
Unit % cuvette ON
Preset No. 8 1st convers reference curve
Predilution 1:5 falling
unit %
Sample 40 µL
decimal place xxx.x
Imidazole Buffer Solution 160 µL
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor XI Deficient Plasma 50 µL 2nd convers none
3. Pathromtin* SL Reagent 50 µL 100.0% 0.0 s *
Incubation time 120 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 1 s
learn 3 s
lag 15 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:4 30 --- 90 ---
1.0 1:5 30 --- 120 ---
0.75 1:6.7 30 --- 170 ---
0.5 1:10 20 --- 180 ---
0.25 1:20 10 --- 190 ---
0.10 1:50 --- 30 120 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 133


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor XI with
Pathromtin* SL

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
139.5 40.7
93.0 45.0
69.8 48.9
46.5 54.7
30.7 60.0
23.3 64.1
11.6 74.6

Reference Guide 3.01 - not for use in the USA - 134


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor XII with
® ®
Dade Actin Activated Cephaloplastin Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
® ®
Dade Actin Activated Cephaloplastin Reagent B4218-1/-2 2/10 mL Beside the BFT* II Analyzer [1]
Coagulation Factor XII Deficient Plasma OSDG 1 mL
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Dade® Actin® Reagent (RT) 2
FXII Deficient Plasma (RT) 4
Imidazole Buffer Solution (RT) 2
Calcium Chloride Solution 2
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 400
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 0.91 x - 3.16 0.951
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 135


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor XII with
® ®
Dade Actin Activated Cephaloplastin Reagent

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 6.1 3.4 6.7
Control Plasma P 7.4 3.7 7.9

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 120 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 136


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor XII with
® ®
Dade Actin Activated Cephaloplastin Reagent

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 8 Method Intr. Factor
Unit % cuvette ON
Preset No. 8 1st convers reference curve
Predilution 1:5 falling
unit %
Sample 40 µL
decimal place xxx.x
Imidazole Buffer Solution 160 µL
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor XII Deficient Plasma 50 µL 2nd convers none
3. Dade® Actin® Activated Cephaloplastin Reagent 50 µL 100.0% 0.0 s *
Incubation time 120 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 1 s
learn 3 s
lag 15 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:4 30 --- 90 ---
1.0 1:5 30 --- 120 ---
0.75 1:6.7 30 --- 170 ---
0.5 1:10 20 --- 180 ---
0.25 1:20 10 --- 190 ---
0.10 1:50 --- 30 120 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 137


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor XII with
® ®
Dade Actin Activated Cephaloplastin Reagent

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
142.5 37.7
95.0 41.5
71.3 45.3
47.5 50.5
31.4 57.4
23.8 62.9
11.9 77.6

Reference Guide 3.01 - not for use in the USA - 138


BFT* II Analyzer
Application Sheet for Page 1 of 4
Coagulation Factor XII with
Pathromtin* SL
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Pathromtin* SL Reagent OQGS 5 mL Beside the BFT* II Analyzer [1]
Coagulation Factor XII Deficient Plasma OSDG 1 mL
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
Pathromtin* SL Reagent (RT) 8
FXII Deficient Plasma (RT) 8
Imidazole Buffer Solution (RT) 8
Calcium Chloride Solution 8
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 576
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Pathromtin* SL Reagent on BCT* System y = 1.05 x - 5.38 0.982
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 139


BFT* II Analyzer
Application Sheet for Page 2 of 4
Coagulation Factor XII with
Pathromtin* SL

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 15 %.
Within Run Run to Run Total
% % %
Control Plasma N 4.7 3.2 5.4
Control Plasma P 7.8 5.2 9.0

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
70 - 150 % [2]
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
[2]
Thomas, Lothar, “Clinical Laboratory Diagnostics”, TH-Books Emil-von-Behring-Str. 76
Verlagsgesellschaft mbH, Frankfurt/Main, Germany, 1998: 607. 35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 140


BFT* II Analyzer
Application Sheet for Page 3 of 4
Coagulation Factor XII with
Pathromtin* SL

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 8 Method Intr. Factor
Unit % cuvette ON
Preset No. 8 1st convers reference curve
Predilution 1:5 falling
unit %
Sample 40 µL
decimal place xxx.x
Imidazole Buffer Solution 160 µL
min / max value 0.5 / 200.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. Coagulation Factor XII Deficient Plasma 50 µL 2nd convers none
3. Pathromtin* SL Reagent 50 µL 100.0% 0.0 s *
Incubation time 120 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 120 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 1 s
learn 3 s
lag 15 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.

Standard Human Prediluted Standard


Factor [1] Dilution Plasma Human Plasma 1:25 Imidazole Buffer
µL µL µL —
1.25 1:4 30 --- 90 ---
1.0 1:5 30 --- 120 ---
0.75 1:6.7 30 --- 170 ---
0.5 1:10 20 --- 180 ---
0.25 1:20 10 --- 190 ---
0.10 1:50 --- 30 120 ---
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 141


BFT* II Analyzer
Application Sheet for Page 4 of 4
Coagulation Factor XII with
Pathromtin* SL

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
142.5 37.3
95.0 41.7
71.3 44.4
47.5 49.4
31.4 54.8
23.8 57.4
11.9 69.8

Reference Guide 3.01 - not for use in the USA - 142


BFT* II Analyzer
Application Sheet for Page 1 of 4
Lupus Anticoagulant with
LA 1 Screening Reagent / LA 2 Confirmation Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
LA 1 Screening Reagent OQGP 2 mL Reagent positions 1 - 4
LA 2 Confirmation Reagent OQGR 1 mL
Control Plasma N ORKE 1 mL Beside the BFT* II Analyzer
LA Control High OQWD 1 mL
LA Control Low OQWE 1 mL
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
Be sure to swirl the reagent bottle to assure that condensation on the bottle walls is re-mixed into the solution.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
LA 1 Screening Reagent 4
LA 2 Confirmation Reagent 4
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 1200
Hemoglobin mg/dL 1000
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
LA 1 Screening Reagent on BCT* System y = 1.03 x - 1.38 0.990
LA 1 Screening Reagent / LA 2 Confirmation Reagent on BCT* System y = 0.99 x - 0.01 0.984
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 143


BFT* II Analyzer
Application Sheet for Page 2 of 4
Lupus Anticoagulant with
LA 1 Screening Reagent / LA 2 Confirmation Reagent

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 5 %.
Within Run Run to Run Total
% % %
LA 1: Control Plasma N 2.2 1.0 2.3
LA 1: LA Control Low 1.3 2.1 2.4
LA 1: LA Control High 1.6 1.3 2.0
LA 2: Control Plasma N 1.8 0.4 1.8
LA 2: LA Control Low 1.2 0.8 1.4
LA 2: LA Control High 1.3 1.2 1.7
LA 1 / LA 2 Ratio: Control Plasma N 2.7 0.8 2.6
LA 1 / LA 2 Ratio: LA Control Low 1.1 2.0 2.3
LA 1 / LA 2 Ratio: LA Control High 1.7 0.5 1.7

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: sec
th th
Comments n Mean Median 5 - 95 Percentile
LA 1 Screening Reagent 75 31.1 30.5 27.8 - 35.1
LA 2 Confirmation Reagent 75 31.2 31.2 29.0 - 34.3
LA 1 / LA 2 Ratio (no unit) 75 0.99 0.99 0.91 - 1.11
Reference intervals vary from laboratory to laboratory depending on the population served and the technique and reagent lot used.
Therefore, each laboratory must establish its own reference intervals or verify them whenever one or more of the aforementioned variables
are changed.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 144


BFT* II Analyzer
Application Sheet for Page 3 of 4
Lupus Anticoagulant with
LA 1 Screening Reagent / LA 2 Confirmation Reagent

Test Protocol Parameter-Protocol


Method store 14 Method LA 1
Unit sec cuvette ON
Preset No. 14
1st convers none
Pipetting Scheme falling *
1. Sample 100 µL unit sec *
decimal place xxx.x *
Incubation time 60 sec
min / max value 0.0 / 0.0 *
Start reagent
time interpolat. linear *
LA 1 Screening Reagent 100 µL value interpolat. linear *
Autostart yes
2nd convers none
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
Refer to the Instructions for Use for information regarding result evaluation.
All parameters have to be entered manually.
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 145


BFT* II Analyzer
Application Sheet for Page 4 of 4
Lupus Anticoagulant with
LA 1 Screening Reagent / LA 2 Confirmation Reagent

Test Protocol Parameter-Protocol


Method store 15 Method LA 2
Unit sec cuvette ON
Preset No. 15
1st convers none
Pipetting Scheme falling *
1. Sample 100 µL unit sec *
decimal place xxx.x *
Incubation time 60 sec
min / max value 0.0 / 0.0 *
Start reagent
time interpolat. linear *
LA 2 Confirmation Reagent 100 µL value interpolat. linear *
Autostart yes
2nd convers none
100.0% 0.0 s *
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 100 µl
incubat (0 = off) 1st 0 s
2nd 60 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


The assay is not calibrated.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 146


BFT* II Analyzer
Application Sheet for Page 1 of 4
Protein C System with
®
ProC Global
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
ProC Global Assay — OQLS Kit —
REAGENT APTT — 5 mL Beside the BFT* II Analyzer [1]
ACTIVATOR — 2 mL
BUFFER — 2 mL
Standard Human Plasma ORKL 1 mL
®
ProC Control Plasma OQKE 1 mL
Control Plasma N ORKE 1 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
REAGENT APTT (RT) 24
ACTIVATOR (RT) 24
BUFFER (RT) 24
Calcium Chloride Solution 24
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 453
Hemoglobin mg/dL 200
Bilirubin mg/dL 60

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
ProC® Global Assay on BFA* System y = 0.91 x + 0.01 0.945
r = Correlation Coefficient

Reference Guide 3.01 - not for use in the USA - 147


BFT* II Analyzer
Application Sheet for Page 2 of 4
Protein C System with
®
ProC Global

Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 10 %.
Within Run Run to Run Total
% % %
Control Plasma N 3.0 1.7 3.3
ProC® Control Plasma 4.3 0.9 4.1

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: NR
Comments n Mean Median 5th - 95th Percentile
— 65 1.05 1.01 0.78 - 1.40
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 148


BFT* II Analyzer
Application Sheet for Page 3 of 4
Protein C System with
®
ProC Global

Test Protocol Parameter-Protocol


Method store 10 Method PCAT
Unit sec cuvette ON
Preset No. 10
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec *
decimal place xxx.x *
2. ACTIVATOR 50 µL
min / max value 0.0 / 0.0 *
3. REAGENT APTT 50 µL time interpolat. linear *
Incubation time 180 sec value interpolat. linear *
Start reagent
2nd convers none
Calcium Chloride Solution 50 µL
100.0% 0.0 s *
Autostart no
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 180 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


Refer to the Instructions for Use regarding calibration. The calibration factor must be determined every month for each analyzer and lot of
reagent.

Remarks
To assure inter-laboratory comparability of the results, it is recommended that the Normalized Ratio (NR) of each sample should be
calculated. Refer to Instruction for Use for further information regarding result evaluation.
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 149


BFT* II Analyzer
Application Sheet for Page 4 of 4
Protein C System with
®
ProC Global

Test Protocol Parameter-Protocol


Method store 11 Method PCAT/ 0
Unit sec cuvette ON
Preset No. 11
1st convers none
Pipetting Scheme falling *
1. Sample 50 µL unit sec*
decimal place xxx.x*
2. BUFFER 50 µL
min / max value 0.0 / 0.0 *
3. REAGENT APTT 50 µL time interpolat. linear *
Incubation time 180 sec value interpolat. linear *
Start reagent
2nd convers none
Calcium Chloride Solution 50 µL
100.0% 0.0 s *
Autostart no
factor --- *
ISI 0.00 *

measurement lot. no.: xxxxxx


Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 180 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


Refer to the Instructions for Use regarding calibration. The calibration factor must be determined every month for each analyzer and lot of
reagent.

Remarks
To assure inter-laboratory comparability of the results, it is recommended that the Normalized Ratio (NR) of each sample should be
calculated. Refer to Instruction for Use for further information regarding result evaluation.
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 150


BFT* II Analyzer
Application Sheet for Page 1 of 4
Factor V Leiden with
®
ProC Global / Factor V Deficient Plasma
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
®
ProC Global Assay — OQLS Kit —
REAGENT APTT — 5 mL Beside the BFT* II Analyzer [1]
ACTIVATOR — 2 mL
BUFFER — 2 mL
Coagulation Factor V Deficient Plasma ORSM 1 mL
Standard Human Plasma ORKL 1 mL
®
ProC Control Plasma OQKE 1 mL
Control Plasma N ORKE 1 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).

On-board Stability
Material Time [h]
REAGENT APTT (RT) 24
ACTIVATOR (RT) 24
BUFFER (RT) 24
FV Deficient Plasma (RT) 8
Calcium Chloride Solution 24
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 600
Hemoglobin mg/dL 1000
Bilirubin mg/dL 48

Reference Guide 3.01 - not for use in the USA - 151


BFT* II Analyzer
Application Sheet for Page 2 of 4
Factor V Leiden with
®
ProC Global / Factor V Deficient Plasma

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
®
ProC Global / Factor V Assay on BCT* System y = 1.00 x - 0.02 0.978
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 10 %.
Within Run Run to Run Total
% % %
Control Plasma N 4.0 1.2 3.9
ProC® Control Plasma 3.1 2.4 3.8

Measuring Range
Refer to the Instructions for Use of the analyzer for additional information.

Reference Interval
In a study with ostensibly healthy subjects the following data were obtained:
Unit: NR
th th
Comments n Mean Median 5 - 95 Percentile
— 66 1.02 1.01 0.91 - 1.14
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 152


BFT* II Analyzer
Application Sheet for Page 3 of 4
Factor V Leiden with
®
ProC Global / Factor V Deficient Plasma

Test Protocol Parameter-Protocol


Method store 12 Method FV PCAT
Unit sec cuvette ON
Preset No. 12
1st convers none
Predilution 1:5 falling *
Sample 20 µL unit sec *
Coagulation Factor V Deficient 80 µL decimal place xxx.x *
min / max value 0.0 / 0.0 *
Pipetting Scheme time interpolat. linear *
1. Prediluted sample 50 µL value interpolat. linear *
2. ACTIVATOR 50 µL 2nd convers none
3. REAGENT APTT 50 µL 100.0% 0.0 s *
Incubation time 180 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 180 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


Refer to the Instructions for Use regarding calibration. The calibration factor must be determined every month for each analyzer and lot of
reagent.

Remarks
To assure inter-laboratory comparability of the results, it is recommended that the Normalized Ratio (NR) of each sample should be
calculated. Refer to Instruction for Use for further information regarding result evaluation.
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 153


BFT* II Analyzer
Application Sheet for Page 4 of 4
Factor V Leiden with
®
ProC Global / Factor V Deficient Plasma

Test Protocol Parameter-Protocol


Method store 13 Method FV PCAT/ 0
Unit sec cuvette ON
Preset No. 13
1st convers none
Predilution 1:5 falling *
Sample 20 µL unit sec *
Coagulation Factor V Deficient 80 µL decimal place xxx.x *
min / max value 0.0 / 0.0 *
Pipetting Scheme time interpolat. linear *
1. Prediluted sample 50 µL value interpolat. linear *
2. BUFFER 50 µL 2nd convers none
3. REAGENT APTT 50 µL 100.0% 0.0 s *
Incubation time 180 sec factor --- *
Start reagent ISI 0.00 *
Calcium Chloride Solution 50 µL measurement lot. no.: xxxxxx
Autostart no Start Reag. 50 µl
incubat (0 = off) 1st 0 s
2nd 180 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 120 digits
adjust 5 s
learn 3 s
lag 15 s

Standard Curve Calibration


Refer to the Instructions for Use regarding calibration. The calibration factor must be determined every month for each analyzer and lot of
reagent.

Remarks
To assure inter-laboratory comparability of the results, it is recommended that the Normalized Ratio (NR) of each sample should be
calculated. Refer to Instruction for Use for further information regarding result evaluation.
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 154


BFT* II Analyzer
Application Sheet for Page 1 of 4
Protein C with
Protein C Reagent
For additional information, limitations and interferences please refer to
the Instructions for Use of the analyzer and check current Instructions
for Use for reagents, controls and calibrators and tables of assigned/
analytical values.

The parameters defined in this application sheet have been developed by Siemens Healthcare Diagnostics to provide optimal product
performance with the assay and instrument combination. Any modification to these parameters may affect performance of this and other
assays in use on your system and the resulting assay values. It is the responsibility of the user to validate any modifications and their impact
on all assay results. Results of this test should always be interpreted in conjunction with the patient's medical history, clinical presentation
and other findings.
The application sheet lists all combinations of controls and calibrators for use with the reagent and instrument system; other combinations
are not validated or supported by Siemens.

Materials Required
Material Catalog # Size On-board position
Protein C Reagent — OQYG Kit —
ACTIVATOR — 3 mL Beside the BFT* II Analyzer [1]
DEFICIENT — 1 mL
REAGENT APTT — 10 mL
Standard Human Plasma ORKL 1 mL
Control Plasma N ORKE 1 mL
Control Plasma P OUPZ 1 mL
Imidazole Buffer Solution OQAA 15 mL
Calcium Chloride Solution Reagent positions 1 -ORHO
4 15 mL Reagent positions 1 - 4
Eppendorf ep T.I.P.S. — 003.0000.870 — —

Additional Notes
[1]
Do not place on a reagent position of the BFT* II Analyzer, since all reagent positions of the BFT* II Analyzer are at 37 °C. These reagents
have to be placed beside the BFT*II Analyzer for use.
Keep cuvettes in the incubation block during the incubation. After incubation is complete place the cuvettes into the measuring channel.
Before use the reagent has to be mixed carefully.
Materials positioned beside the BFT* II Analyzer have to be kept at room temperature (+15 °C to +25 °C, RT).
Limitations
Hemolyzed samples are not suitable for protein C determination.

On-board Stability
Material Time [h]
ACTIVATOR (RT) 8
DEFICIENT (RT) 4
REAGENT APTT (RT) 8
Calcium Chloride Solution 4
The stability data presented here were established under controlled laboratory conditions. Due to differences in laboratory environmental
conditions and reagent vial fill volumes performance may vary. Please contact your local service representative to report any issues.

Interference Studies
No interferences up to ...
Triglycerides mg/dL 582
Hemoglobin mg/dL 1000
Bilirubin mg/dL 24

Reference Guide 3.01 - not for use in the USA - 155


BFT* II Analyzer
Application Sheet for Page 2 of 4
Protein C with
Protein C Reagent

Performance Characteristics
The following studies were conducted using specimens collected in 3.2 % sodium citrate solution.
Method Comparison
Predicate Device Regression Equation r
Protein C Reagent, coagulometric on BCT* System y = 1.03 x + 1.99 0.986
r = Correlation Coefficient
Precision
Acceptable variability (imprecision) should be such that the total coefficient of variation (CV) of the analytical system on the same lot of
control plasma is less than 20 %.
Within Run Run to Run Total
% % %
Control Plasma N 3.4 2.4 4.0
Control Plasma P 3.6 1.3 3.6

Measuring Range
The measuring range is defined by the concentration of the calibrators used. Refer to the Instructions for Use of the analyzer for additional
information.

Reference Interval
Refer to the Instructions for Use of the reagent.
Reference intervals vary from laboratory to laboratory depending on e.g. the technique used; therefore, each laboratory should establish its
own reference interval.
For more information on establishing reference intervals see CLSI document C28-A3, “How to Define and Determine Reference Intervals in
the Clinical Laboratory; Approved Guideline.”

Bibliography Manufacturer
Refer to the Instructions for Use of the reagent. Siemens Healthcare Diagnostics Products GmbH
Emil-von-Behring-Str. 76
35041 Marburg / Germany

USA Distributor
Siemens Healthcare Diagnostics Inc.
Newark, DE 19714 USA

Reference Guide 3.01 - not for use in the USA - 156


BFT* II Analyzer
Application Sheet for Page 3 of 4
Protein C with
Protein C Reagent

The measurement has to be started manually by pressing the start button with the pipetting of Calcium Chloride since the autostart does not
work with this assay.

Test Protocol Parameter-Protocol


Method store 14 Method Protein C Clot
Unit % cuvette ON
Preset No. 14 1st convers reference curve
Predilution 1:10 rising
unit %
Sample 20 µL
decimal place xxx.x
Imidazole Buffer Solution 180 µL
min / max value 10.0 / 150.0 %
Pipetting Scheme time interpolat. logarithmic
1. Prediluted sample 50 µL value interpolat. logarithmic
2. DEFICIENT 50 µL 2nd convers none
3. ACTIVATOR 50 µL 100.0% 0.0 s *
4. REAGENT APTT 50 µL factor --- *
Incubation time 240 sec ISI 0.00 *
Start reagent measurement lot. no.: xxxxxx
Calcium Chloride Solution 50 µL Start Reag. 50 µl
Autostart no incubat (0 = off) 1st 0 s
2nd 240 s
mixer 1 -> 2 10 s
1 500 rpm
2 500 rpm
min value 150 digits
adjust 1 s
learn 3 s
lag 1 s

Standard Curve Calibration


The results are evaluated using a reference curve established with Standard Human Plasma. The standard curve is measured in duplicate.
1. dilution step 2. dilution step
Standard Human
Factor [1] Dilution Plasma Deficient Plasma SHP dil. with Imidazole Buffer
µL µL deficient plasma µL µL
1.50 1:6.7 60 --- --- 340
1.00 1:10 50 --- --- 450
0.75 1:13.3 150 50 50 450
0.50 1:20 50 50 50 450
0.33 1:30 50 100 50 450
0.10 1:100 20 180 50 450
[1]
Factor: Multiply the analytical value of the calibrator by the specified factors.
The analytical value is lot specific, refer to the table of analytical values.

Remarks
* This parameter does not have any function in this assay.

Reference Guide 3.01 - not for use in the USA - 157


BFT* II Analyzer
Application Sheet for Page 4 of 4
Protein C with
Protein C Reagent

Standard Curve (example only)


A new reference curve should be established with each change of reagent lot or with any deviation from controls or proficiency testing limits.
Do not use the calibration curve shown here or in the software, as it is an example only.
Standard Human Plasma
% sec
148.5 102.7
99.0 85.8
74.3 72.3
49.5 58.2
32.7 50.3
10.0 38.5

Remarks
Calibrators are measured in duplicate using 50 µl of the respective final calibrator dilution.

Reference Guide 3.01 - not for use in the USA - 158

You might also like