1230262G-1 3D SCANNER Quick Start Guide p
1230262G-1 3D SCANNER Quick Start Guide p
Note: For detailed setup instructions and safety information, see the 3D SCANNER Hardware Reference Guide, SNC
Dosimetry Software Reference Guide, and 3D SCANNER instructional videos, all of which are available in SNC
Dosimetry software by selecting Help > System Guides. The Help button, , is located in the upper right corner.
WARNING: HEAVY OBJECTS. Two person lift required for the 3D MiniLift Lift Subassembly, Automatic Leveling
Platform, and Water Tank.
© 2015-2022, Sun Nuclear Corp.; 1230262, Rev G-1, 13 September 2022; www.sunnuclear.com
4. Position to Delivery System
a. Position 3D SCANNER with ‘Gantry’ labels on Water Tank and 3D MiniLift toward gantry, as shown.
b. Secure 3D MiniLift legs in Measurement position.
a. Press Automatic Leveling Platform/Lift Table (B) button twice to activate Advanced
Functions Mode 2.
b. Press and release Level 2 + button to move Water Tank motors to home position. LED B blinks
rapidly while homing.
c. When Water Tank homing is complete, press and release Lift + button to move Automatic
Leveling Platform motors to home position. LED B blinks rapidly while homing.
d. When Automatic Leveling Platform homing is complete, press Automatic Leveling Platform/
Lift Table (B) button twice to deactivate Advanced Functions Mode 2.
e. Align marks on Water Tank base to linac crosshairs. It is more important to be parallel to marks Step c. Step b.
on Water Tank base than centered. AutoSetup aligns the ring center to the delivery system. Press once Press once
f. Gently lock 3D MiniLift casters and verify that crosshair alignment is unchanged.
page 2
9. Install Reference Detector (optional) 10. Install Field Detector
If using EDGE Detector as a reference detector, turn it Use a scanning detector for AutoSetup, such as SNC125c
upside-down to increase signal-to-noise. or EDGE Detector. Do not use Farmer type, Parallel Plate
a. Insert sensor end in reference detector holder. Secure type, or large volume detectors during AutoSetup.
with thumbscrew.
a. Install detector
b. Secure cable Detector Setup cap
Reference Detector Holder holder on detector Holder
using hook holder base, then
and loop insert detector
(Velcro-type) through detector
straps.
holder pointed
c. Route cable toward center of Detector
through clip on Holder
water tank. Base
back of motor Straps
housing. b. If you have a setup
cap for the detector, gently push setup cap over
d. Connect cable to Ref input on 3D electrometer. detector to positive stop. Then, push setup cap and
CAUTION: Do not connect a coaxial plug to the detector toward detector holder until setup cap is flush
triaxial electrometer input. This will damage to detector holder. Tighten thumbscrew (rotate
the electrometer. Use an appropriate adapter. clockwise) to finger tight.
e. Position reference detector so it will +Y If you do not have a
be in the flat region of the beam, not setup cap for the
in the penumbra. AutoSetup typically
detector, set the
uses a 10 x 15 cm field size.
geometric center of
Note: Do not shadow the +X the active volume 16
mm
planned field detector exactly 16 mm from
measurement path with the the edge of the
Redete
fe ct
base.
e
page 3
14. Select Hardware Configuration
a. Set delivery system, AutoSetup SSD, field detector, and reference
detector (if applicable).
Use Effective Point of Measurement: shifts field detector position prior
to measurement. Shift is based on detector properties (diameter or
offset). See Environment > Detectors. Leave unchecked for AutoSetup.
Use Offset Detector Holder: for profile measurements >50 cm in length.
Leave unchecked for AutoSetup.
Use High Current Mode: for high dose rate delivery with high sensitivity diode detector(s). Select if applicable; otherwise
leave unchecked.
b. Bias voltage applies to both electrometer inputs. Set the lower of the bias voltages recommended by the manufacturer
when using two different ionization chambers or when using an ionization chamber and a diode detector such as the Sun
Nuclear EDGE detector, which is not damaged by bias voltage.
15. AutoSetup *
AutoSetup instructions and behaviors are specific to selected delivery system
and AutoSetup type (circular vs. rectangular collimation, TPR vs. standard).
Select AutoSetup type, click Start, and then follow on-screen instructions.
Green check marks indicate completed steps. ‘Details’ column shows results.
Note: If ‘Require Collimator Rotation’ system preference enabled, AutoSetup
requests 180° collimator rotation. For Varian delivery systems, rotate collimator
between 90° - 270° to take advantage of 180° rotation. For small field setups or
delivery systems where the collimator cannot be rotated, set Rotate
Collimator = No in AutoSetup panel.
*For instructions to perform manual setup, TPR AutoSetup, or post setup verification, see
the SNC Dosimetry Software Reference Guide or SNC Dosimetry Help Index.