Professional Documents
Culture Documents
1230066Q 3D SCANNER Hardware Reference Guide s
1230066Q 3D SCANNER Hardware Reference Guide s
Page ii
Preface
Intended Use
The 3D SCANNER system, Model 1230, is intended for radiotherapy dosimetry measurements
and export of those measurements for commissioning a treatment planning system (TPS)
computer. It is also intended for periodic beam quality assurance (QA) tests as defined by the
medical physicist responsible for the QA program.
CAUTION: Federal laws restrict this software to sale by or on the order of a qualified
radiation therapy professional, such as a medical dosimetrist, medical physicist, or
radiation oncologist, for use in radiation oncology QA.
Application
The 3D SCANNER, Model 1230, is a three dimensional dosimetry scanning system with
applications in:
Radiotherapy LINAC beam and Cobalt 60 Acceptance Data collection for the TPS
Commissioning of the beam models.
Monthly and annual QA protocol testing of radiotherapy delivery systems.
Delivery system bunker testing at the manufacturer’s facility.
Sun Nuclear meets the requirements of FDA 21 CFR 820 and is certified to relevant country
regulations, including ISO 13485:2016. Sun Nuclear places the CE mark upon Class I medical
devices under its sole discretion. Sun Nuclear utilizes the involvement of a Notified Body when
placing the CE mark on medical devices requiring Notified Body involvement.
Follow these health and safety instructions when using the 3D SCANNER system:
WARNING: Using the wrong power supply or power cord with the 3D SCANNER
system can damage the equipment and pose a potential safety hazard. Only use the
provided power supply (APX Model SM130P924SRR), PDI 3.0 power supply (GlobTek
Model GTM21089-1818-T3), and power cords.
WARNING: Use of accessories, transducers and cables other than those specified or
provided by Sun Nuclear for use with the 3D SCANNER system could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
Two person lift required for the 3D SCANNER Water Tank, Automatic Leveling
Platform, and 3D MiniLift. Use proper lifting and handling techniques to prevent injury
and equipment damage. The weight of each component is provided in Specifications
on page 65.
Use the 3D MiniLift push handle to move the 3D SCANNER system. Do not push or
lean on the Automatic Leveling Platform of the 3D SCANNER Water Tank.
• Ensure that high voltage is disabled on the FIELD and REF ports before connecting or
disconnecting a radiation detector.
• Use the power supply extension cable to keep the 3D SCANNER power supply at a
sufficient distance from the facility’s water supply to minimize the risk of water ingress
in the power supply and exposure to high voltage.
• Never pull on a cable to disconnect it. Always grasp the connector to protect insulation.
• When the 3D SCANNER system is positioned under the beam or in storage, engage the
3D MiniLift or 3D Lift caster locks to prevent unintended movement.
• Adjust the 3D MiniLift or 3D Lift caster locks by foot, never by hand.
• Do not route cables under or through the treatment room door, as this could cause a trip
hazard. Route the cables through a conduit from the control room to the treatment room.
• Do not transport the 3D SCANNER system with water in the Water Tank.
• Before activating motion on the 3D SCANNER, ensure that nearby equipment will not be
impacted and that personnel are at a safe distance from the 3D SCANNER.
• Do not remove the electronics enclosure housing or a motor housing unless specifically
instructed to do so by Sun Nuclear personnel. Use caution when re-installing the cover
or housing to prevent pinching of fingers/hands.
• Before transporting or storing the 3D SCANNER system, fully retract the lift table.
• Do not fill the water tank beyond the maximum fill line.
• Do not spill liquids on any 3D SCANNER parts that are outside of the water tank, such as
the control unit, the power supply, the PDI 3.0, and cables.
• The 3D SCANNER is not intended to be used in an oxygen rich environment.
For instructions to report health or safety related concerns, see Reporting Health or Safety Related
Issues or Concerns on page 85.
Refer to the following sources for additional information about the 3D SCANNER System.
• SNC Dosimetry Software Reference Guide
• SNC Dosimetry Software User Guide
• 3D Reservoir User Guide
• 3D MiniLift User Guide
• Technical Bulletin 02-13, Setting Up 3D TPR
• Technical Bulletin 06-13, Setting Up WaterProof PROFILER
• Technical Bulletin 09-13, Updating 3D SCANNER System Firmware
• Technical Bulletin 10-13, Updating 1D SCANNER System Firmware
• SNC Dosimetry Help Index
All of the above documents can be accessed in SNC Dosimetry software. Click the Help button
in the upper right corner of the main window, then select System Guides.
For software, firmware, and documentation updates, see the Sun Nuclear Support website
(sunnuclear.com/support). See Support Website on page 49 for access instructions.
The following parts are included with the 3D SCANNER system (Figure 1):
12 3 1
11
10
8
9
2
5
7
6
13
WARNING: Using the wrong power supply or power cord with 3D SCANNER can
damage the equipment and pose a potential safety hazard. Only use the provided
power supply (APX Model SM130P924SRR), PDI 3.0 power supply (GlobTek Model
GTM21089-1818-T3), and power cords.
Note: Vertical detector holders (7 mm P/N 1230465Z and 10 mm 1230445Z) are not
included in the detector holder kit. However, they can be ordered separately. See
Detector Holders on page 4.
Note: Setup caps are not included in the detector holder kit. Contact Sun Nuclear
Sales to order setup caps specific to the detectors that you intend to use.
1 2 3 4
Cutout
for H
Offset or
6 Inverted
Offset
Arm
OH
Vertical
Cutouts
for
Vertical
Detector
8 Cutouts
for Setup
Holders Caps
1 2 3 4 5
7
8
1 2 3 4 5
6 7 8
1.2.6 3D MiniLift
The 3D MiniLift supports the Automatic Leveling Platform and 3D SCANNER Water Tank. It allows
the user to position the 3D SCANNER system relative to the delivery system and set the Water
Tank to the correct vertical position. The 3D MiniLift vertical position is controlled using the
pendant (see Pendant on page 57.)
The 3D MiniLift legs can be adjusted to three positions.
• Measurement — The 3D MiniLift straddles the Floor Ring on the treatment room floor
and provides maximum stability for the 3D SCANNER system.
• Transport — For moving the 3D SCANNER system within the facility.
• Storage — Creates a small footprint for long-term storage, loading into a vehicle, or
placing in optional shipping case (see Shipping Cases on page 5).
For more information about the 3D MiniLift and detailed instructions for use, see the 3D MiniLift
User Guide, PN 1230037.
Figure 7. 3D Reservoir
1.2.8 3D TPR
The 3D TPR sensor is used in conjunction with the 3D Reservoir and 3D SCANNER system to
collect TPR measurements for the Delivery System under test. For 3D TPR hardware setup
instructions, see Technical Bulletin 02-13, Setting Up 3D TPR, P/N 1230054.
WaterProof PROFILER hardware setup instructions are provided in Technical Bulletin 06-13,
Setting Up WaterProof PROFILER, P/N 1203009. WaterProof PROFILER software setup and
measurement instructions are provided in the SNC Dosimetry Software Reference Guide, P/N
1230067.
Before using the WaterProof PROFILER with the 3D SCANNER system, SNC Dosimetry software
must be upgraded to version 3.3 or higher and the following firmware must be upgraded: 3D
SCANNER Controller, 3D Electrometer, and 3D SCANNER Accessory Controller (WaterProof
PROFILER). For firmware upgrade instructions, see Technical Bulletin 09-13, Updating 3D
SCANNER System Firmware, P/N 1203010.
For physical dimensions and characteristics of the 3D SCANNER hardware components, see
Hardware Reference on page 51 and Specifications on page 65.
Sun Nuclear personnel will perform the initial install of the 3D SCANNER at your facility and
provide training.
1 With the 3D MiniLift legs in Measurement position for stability, lock the casters then attach
the Automatic Leveling Platform to the 3D MiniLift. See Attach Automatic Leveling Platform
to 3D MiniLift on page 9.
2 Attach the 3D SCANNER to the Automatic Leveling Platform. See Attach Water Tank to
Automatic Leveling Platform on page 11.
3 Align the 3D SCANNER to the Delivery System lasers. See Position 3D SCANNER to Delivery
System on page 12.
4 Connect 3D SCANNER cables. See Connect Cables in Control Room on page 13 and Connect
Cables in Treatment Room on page 14.
5 Fill the Water Tank to the maximum fill line and adjust SSD. See Fill Water Tank on page 14.
6 Install a reference detector, if required. See Install Reference Detector (Optional) on page 17.
7 Install the field detector holder and field detector. See Detector Setup on page 20, Install Field
Detector on page 18, and Detector Setup on page 20.
8 Set the maximum height. See Set the Vertical Limit (Maximum Height) on page 22.
For 3D TPR hardware setup instructions, see Technical Bulletin 02-13, Setting Up 3D TPR, P/N
1230054.
For WaterProof PROFILER hardware setup instructions, see Technical Bulletin 06-13, Setting Up
WaterProof PROFILER, P/N 1203009.
WARNING: Two person lift required for Automatic Leveling Platform. Use proper
lifting and handling techniques to prevent injury or equipment damage. For weight
specifications, see Specifications on page 65.
CAUTION: To avoid pinching fingers when changing the 3D MiniLift leg positions,
keep hands clear of the leg joints and frame.
Keyed White
corner supports
Power
cord
Adapter
plate
Power
connector
Gantry
label
4 Gently place the Automatic Leveling Platform on the Adapter Plate. The Adapter Plate is keyed
for proper placement. Ensure the motor control cables and power cord are not pinched.
5 From the underside of the adapter plate, install four M10 x 1.5, 25 mm bolts to secure the
Automatic Leveling Platform (Figure 11). Turn bolts to finger tight and then an additional 1/8th
turn using 8 mm hex wrench.
6 Connect the Automatic Leveling Platform power cord to the Lift Connection port on the 3D
MiniLift. The connector is keyed. Ensure the connector is completely seated then turn the
collar clockwise until finger tight.
Lift
Connection
WARNING: Two person lift required for 3D SCANNER Water Tank. Use proper lifting
and handling techniques to prevent injury or equipment damage. See Specifications on
page 65 for weight.
1 Locate the keyed corner of the Automatic Leveling Platform. The keyed corner has two black
supports, and the other three corners have one white support each
2 Locate the extended Water Tank handle closest to the electronics enclosure. The extended
handle mates to the keyed corner of the Automatic Leveling Platform.
3 With two people holding the handles at the base of the Water Tank, orient the Water Tank
with the extended handle aligned over the keyed corner of the Automatic Leveling Platform.
4 Gently place the Water Tank on the Automatic Leveling Platform (Figure 13).
Extended
handle
5 Ensure the Water Tank is seated flush and the three white supports on the Automatic Leveling
Platform fit into the reliefs under the Water Tank handles. The white supports on the
Automatic Leveling Platform fit closely with the reliefs in the Water Tank handles, as shown
in Figure 14.
White
support
White
support
Figure 14. White Supports Fit Closely to Reliefs in Water Tank Handles
6 Connect the pendant cable to the PENDANT port on the electronics enclosure, as shown in
Figure 55 on page 53.
7 Connect the ALP1 and ALP 2 cables to the ALP1 and ALP2 ports on the electronics enclosure,
as shown in Figure 55 on page 53.
CAUTION: Keep hands away from the Automatic Leveling Platform when lowering the
3D MiniLift. A crush hazard exists between the Automatic Leveling Platform and
nearby objects, such as the treatment couch.
CAUTION: To avoid pinching fingers when changing the 3D MiniLift leg positions,
keep hands clear of the leg joints and frame.
Use the 3D MiniLift push handle to move the 3D SCANNER system. Do not push or
lean on the Automatic Leveling Platform of the 3D SCANNER Water Tank.
5 Move the 3D SCANNER system to its new location, and then lock the 3D MiniLift or 3D Lift
casters to prevent unintended movement.
CAUTION: To avoid pinching fingers when changing the 3D MiniLift leg positions,
keep hands clear of the leg joints and frame.
2 Place 3D MiniLift legs in Measurement position, lock leg position using provided knob
screws, then lock casters.
1 Connect the PDI power cord (ii) to mains power (i) and to the PDI power supply (iii).
2 Connect the PDI power supply (iii) to the PDI 3.01 (iv).
3 Connect the USB cable (v) to the PDI 3.0 (iv) and to the computer.
Note: The computer that is connected to the 3D SCANNER must comply with the IEC
60950-1, Safety of Information Technology Equipment safety standard
4 Connect the PDI power and data cable (vi) to the PDI 3.0.
5 Verify the LED beside the PDI 3.0 DC port is illuminated.
i v
iv
ii x
iii
vi
ix
vii
viii
1. The previous version of the PDI, labeled PDI 2.0, is compatible with 3D SCANNER. However, the first
generation PDI, labeled P/DI, is not compatible and cannot be used with the 3D SCANNER system. For more
information about PDI device types, see Technical Bulletin 2-09, Power Data Interface (PDI) 3.0, P/N 0224011.
WARNING: Keep the power supply at a distance from the water source to minimize
the risk of water ingress in the power supply and exposure to high voltage.
3 Connect the power supply power cord (ix) to the 3D SCANNER power supply (viii) and to
mains power (x).
4 Turn the 3D SCANNER power supply (viii) switch to the ON position.
5 Verify LEDs A and B on the electronics enclosure are blinking.
Perform these steps from the Pendant. Reference the pendant button controls described in
Pendant on page 57.
1 Press the pendant Lift Table/Automatic Leveling Platform (B) button twice to activate
Advanced Functions Mode 2.
2 Press and release the Level 2 + button to move the Water Tank motors to the home position.
LED B blinks rapidly while homing.
3 When Water Tank homing is complete, press and release Lift + button to move the
Automatic Leveling Platform motors to the home position. LED B blinks rapidly while homing.
4 When Automatic Leveling Platform homing is complete, press the Lift Table/Automatic
Leveling Platform (B) button twice to deactivate Advanced Functions Mode 2.
5 Align the crosshairs on the Water Tank base to the linac crosshairs. It is more important to be
parallel to the marks on the Water Tank base than to be centered. AutoSetup aligns the ring
center to the delivery system.
6 Gently lock 3D MiniLift casters and verify that crosshair alignment is unchanged.
CAUTION: After adding water to the water tank, do not move the 3D
SCANNER system.
Note: The Water Transfer Tube must be removed before performing AutoSetup.
6 Connect the hose to the 3D Reservoir. The 3D Reservoir and Water Transfer tube have
identical hose couplers.
a. On the 3D Reservoir, press both house coupler latch release buttons and pull the latches
out to open. Press the two latch release buttons on the coupler and pull the latches out.
b. Insert the hose fitting into the 3D Reservoir hose coupler and push in until fully seated.
c. Keeping the pull rings clear of the latches, push both latches toward the 3D Reservoir
until they lock.
7 Connect the hose to the Water Transfer Tube.
a. On the Water Transfer Tube, press both house coupler latch release buttons and pull the
latches out to open. Press the two latch release buttons on the coupler and pull the
latches out.
b. Insert the hose fitting into the Water Transfer Tube hose coupler and push in until fully
seated.
c. Keeping the pull rings clear of the latches, push both latches toward the Water Transfer
Tube until they lock.
8 Slide the vent valve Closed by rotating it clockwise and then pull the foot valve UP.
9 On the 3D Reservoir control panel, connect the 6-pin Mini-DIN cable to the 3D SCANNER
Control port.
10 On the 3D SCANNER Water Tank electronics enclosure, connect the 6-pin Mini-DIN cable to
the Reservoir port.
11 Connect power to the 3D Reservoir:
a. Connect one end of the power supply extension cable to the power supply cable.
b. Connect the other end of the power supply extension cable to the 3D Reservoir control
panel Power Input.
c. Connect power supply line power cable to power supply and then to mains power.
d. Move power supply switch to On, and on 3D Reservoir control panel, confirm power indi-
cator A is illuminated.
12 On the 3D Reservoir control panel, set the Fill/Drain switch to Fill.
Note: Do not store the 3D Reservoir power supply inside the 3D Reservoir end
pocket.
13 Move the 3D Reservoir out of the rotation path of the linac, or to a storage location.
WARNING: Set bias voltage to 0 Volts in SNC Dosimetry software before connecting
or disconnecting a radiation detector to or from the FIELD or REF inputs.
If using EDGE Detector as a reference detector, turn it upside-down to increase the signal-to-noise
ratio.
1 Insert the sensor end in the reference detector holder. Secure it using the thumbscrew.
2 Secure the cable to the reference detector arm using the hook and loop fasteners.
Reference Detector
Holder
Fasteners
3 Route the cable through the clip on the back of the motor housing (Figure 17).
CAUTION: Do not connect a coaxial plug to the triaxial electrometer input. This will
damage the electrometer. Use an appropriate adapter. See Detector Adapters on
page 3.
+X
Redete
fe ct
re or
nc
e
Figure 18. Positioning the Reference Detector
Note: Do not shadow the planned field detector measurement path with the reference
detector.
Detector Holder
Base
A B
For instructions to install other detector holders, see the following subsections:
• Installing Offset Detector Holder on page 71.
• Installing Inverted Offset Detector Holder on page 72.
• Installing Farmer Type Detector Holder and Farmer Type Ionization Chamber on page 74.
• Installing Parallel Plate Detector Holder and Parallel Plate Ionization Chamber on page 78.
• Installing the Universal Vertical Detector Holder on page 82.
Note: For a list of supported detectors for each detector holder, see Supported
Detectors on page 69.
Use a scanning detector for AutoSetup, such as SNC125c or EDGE Detector. Do not
use Farmer type, Parallel Plate type, or large volume detectors during AutoSetup.
1 Insert the detector in the detector holder, pointed toward the center of the water tank.
2 If you have a Setup Cap for the detector, see Setup Cap on page 21.
WARNING: Set bias voltage to 0 Volts in SNC Dosimetry software before connecting
or disconnecting a radiation detector to or from the FIELD or REF inputs.
Note: Verify a sufficient length of cable is available for 40 cm depth at the Water Tank
edge.
1 Use the pendant or software to rotate the ring drive until the dual motor housing is 180° from
the electronics enclosure, as shown in the figure below.
Electronics
enclosure
Cable strap
Field detector
extension cable
Clip
Clips
Strap
CAUTION: Do not connect a coaxial plug to the triaxial electrometer input. This will
damage the electrometer. Use an appropriate adapter. See Detector Adapters on
page 3.
4 Verify the opposite end of the field detector extension cable is connected to the FIELD input
on the electrometer (Figure 22).
Field
Ref
5 Rotate the ring drive and verify the field and reference detector cables do not tangle or snag.
16
mm
Ring Center
Figure 23. Cylindrical Detector Setup
Longitudinal
markings show
chamber’s
centerline
Figure 26. Using Setup Cap to Position Detector in Detector Holder, SNC 125c Shown
If desired, test the vertical range of movement to ensure the detector will not collide with any
objects. Use the pendant or SNC Dosimetry software to set the Maximum Height.
• To set the vertical limit using the pendant, move to the desired Z position, press A twice
to select Advanced Functions, and then press Pump - to set the new vertical limit.
• To set the vertical limit using SNC Dosimetry software, move to the desired Z position,
right-click Set the Vertical Limit AutoSetup step, select Set, then click Yes to confirm.
Use the pendant to adjust the tank position until the SSD, as
indicated by the ODI or the front pointer, is at the desired distance,
typically 100 cm. If you are planning to perform TPR measurements,
leave enough space to install the Water Transfer Tube.
See the SNC Dosimetry Software Reference Guide, P/N 1230067, or the SNC Dosimetry Help Index
for instructions to perform the following steps.
1 Open SNC Dosimetry software and verify the 3D SCANNER is recognized, as indicated by the
presence of the AutoSetup steps and the Hardware tab.
2 Define the environment, such as the institution, delivery system, and detectors, before
acquiring a measurement. See the options in the Environment tab.
3 Before initiating AutoSetup or acquiring measurements, ensure that the pendant is powered
off by verifying that LEDs A and B are not illuminated. To turn off power to the pendant, click
the Pendant button next to the accessories panel.
4 Perform AutoSetup.
5 Define scans, collect and analyze measurements, create reports.
When a 3D SCANNER Water Tank is used with a Halcyon delivery system, a Couch Positioning
Plate is placed directly on the treatment couch and then the 3D SCANNER Water Tank is placed
on the Couch Positioning Plate. The Automatic Leveling Platform and 3D MiniLift are not used in
this configuration.
The following are required to use a 3D SCANNER with a Halcyon delivery system.
• SNC Dosimetry software version 3.6 or higher
• A modified 3D SCANNER Lift Ring, P/N 1231406Z
• Halcyon Kit P/N 1231450Z, which includes a Couch Plate (P/N 1231410Z), Halcyon
Bracket (P/N 1231405Z), and a Halcyon SSD label (P/N 1231407Z).
If you received a Halcyon kit that includes 3D SCANNER replacement parts, Kit P/N
1231550Z or 1231650Z, a Sun Nuclear associate will install the replacement parts on
your 3D SCANNER. Do not attempt to perform the instructions in this section until a Sun
Nuclear associate has installed the replacement parts.
If the Halcyon SSD Label has not been installed on the 3D SCANNER Water Tank, install the
Halcyon SSD label that is provided with the Halcyon Kit. Remove the label backing, mate the top
edge of the label to the Maximum Fill line on the 3D SCANNER, and then carefully apply the label.
Halcyon
90 cm SSD
Halcyon
100 cm SSD
The Couch Positioning Plate (CPP) provides a protective surface between the couch and the 3D
SCANNER. The cutouts on the top surface of the CPP mate to the 3D SCANNER Lift Ring, and the
features on the bottom surface of the CPP align with couch positions F1, H1, and H3. The wedged
design compensates for couch sag associated with the weight of the 3D SCANNER and water and
reduces the magnitude of additional corrections controlled by software during measurements.
P/N 1230066, Rev Q Page 23 Using 3D SCANNER with a Halcyon Delivery System
Align the CPP to couch positions F1, H1, and H3 as shown, place the CPP on the couch, and then
verify that the features on the bottom surface of the CPP are fully seated.
F1 H1 H3
3D SCANNER
orientation
Cutout on
top surface
of CPP
Cutout on
top surface
of CPP
F1 H1 H3
Figure 28. Couch Positioning Plate for use with Halcyon Delivery System
1 Disconnect the following cables from the 3D SCANNER electronics enclosure, PDI, Power,
Pendant. If you have an Automatic Leveling Platform, disconnect the ALP1 and ALP2 cables
also.
2 With two people holding two Lift Ring handles each, lift the 3D SCANNER, orient it as shown
on the CPP and in the figure below, and then lower the Lift Ring into the CPP.
When the Lift Ring is fully seated in the CPP cutouts, the 3D SCANNER’s positive x axis aligns
to the Halcyon's negative y axis.
Place holder.
Couch
Position
Label F1 H3
Figure 29. Place 3D SCANNER on Couch Positioning Plate
CAUTION: IMPORTANT: Verify that the power cable is disconnected from the
electronics enclosure before disconnecting the sensor and motor cables.
1 Disconnect the sensor and motor cables from the dual motor housing.
Using 3D SCANNER with a Halcyon Delivery System Page 24 P/N 1230066, Rev Q
’
Figure 30. Disconnect Sensor and Motor Cables from Dual Motor Housing
2 Place the Halcyon Bracket over the front panel of the dual motor housing.
3 Secure the Halcyon Bracket to the dual motor housing with the provided locking nut. Using
fingers only, no tools, turn the nut clockwise until tightened.
Locking
Nut
Halcyon Bracket
Figure 31. Place Halcyon Bracket on Dual Motor Housing and Secure with Locking Nut
4 Install the connectors labeled A and B on the dual motor housing. Angle the connectors as
shown.
5 Install the sensor and motor cables that you removed in Step 1 on the right side of the
Halcyon Bracket, as shown in the figure below.
1 Install a Field Detector holder as described in Install Field Detector Holder on page 18.
2 Install a Field Detector as described in Install Field Detector on page 18.
3 Install a Reference Detector as described in Install Reference Detector (Optional) on page 17.
Follow the 3D SCANNER cabling instructions in Connect Cables in Control Room on page 13 and
Connect Cables in Treatment Room on page 14.
P/N 1230066, Rev Q Page 25 Using 3D SCANNER with a Halcyon Delivery System
3.8 Move the 3D SCANNER into the Bore
A label on the Couch Positioning Plate provides the coordinates that are recommended to prevent
collisions when moving the couch and the 3D SCANNER into the Halcyon bore. See the label
location in Figure 29 on page 24. Move the couch to the position indicated on the label.
CAUTION: If you are using a 3D Reservoir, insert the hose directly into the 3D
SCANNER Water Tank and support the hose by hand while filling and draining.
Do not use the Water Transfer Tube with a Halcyon delivery system!
Only use potable (drinkable, clean, filtered) water in the 3D SCANNER. Fill the 3D SCANNER to the
H90 or H100 mark on the SSD label.
If you installed a reference detector, position the detector so it will be in the beam but not in the
penumbra. AutoSetup for Halcyon uses a 10 cm x 14 cm (x,y) field.
Note: Do not shadow the planned field detector measurement path with the reference detector.
+Y
+X
or e
ct c
te ren
deefe
R
If you have not upgraded to SNC Dosimetry software version 3.6 or higher, follow the instructions
in the SNC Dosimetry Software Reference Guide or the SNC Dosimetry Help index to upgrade the
software.
1 Once the software is installed, double-click on the SNC Dosimetry desktop icon to open the
application.
2 Log in to the application, verify that 3D SCANNER is connected, as shown in the status bar,
and verify that the AutoSetup steps are displayed.
Perform the following steps in SNC Dosimetry software to define the Halcyon delivery system. For
detailed instructions to define a delivery system in SNC Dosimetry software, see the SNC
Dosimetry Software Reference Guide or the SNC Dosimetry Help Index.
1 In SNC Dosimetry software, select Environment > Delivery System > New For >
<Institution> and then click OK.
2 Select Manufacturer/Model = Varian Halcyon, assign a delivery system name, and then
enter the serial number.
3 The suggested energy is 6 MV FFF. Click Add Selected to move this energy to Assigned.
Using 3D SCANNER with a Halcyon Delivery System Page 26 P/N 1230066, Rev Q
4 On the Field Size Selection tab, configure a field size and then click Add. Repeat until all
relevant field sizes are added.
5 Click OK to save the delivery system.
Perform the following steps in SNC Dosimetry software to initiate AutoSetup with a 3D SCANNER
and Halcyon delivery system. For detailed AutoSetup instructions, see the SNC Dosimetry
Software Reference Guide or the SNC Dosimetry Help Index.
1 In the Hardware Configuration panel in SNC Dosimetry software, select the Halcyon delivery
system, select the AutoSetup SSD relevant to your measurements (90 cm or 100 cm), select
field and reference detectors, and then set a voltage appropriate for the selected detectors.
If using high sensitivity diode detectors, select the Use High Current Mode check box. Leave
the other two check boxes unchecked for AutoSetup.
Note: When a Halcyon delivery system is selected, and the 3D SCANNER is
disconnected from the Automatic Leveling Platform, the AutoSetup steps are updated
to support operation without a lift table. Also, the ring drive is automatically offset to
90° from the standard configuration.
2 In the AutoSetup panel, click Start to begin AutoSetup. Follow the on-screen prompts until
all of the AutoSetup steps are complete and a green check mark appears to the left of each
step, as shown in the figure below.
For the steps where a beam is required, set up a 10 cm x 14 cm (x,y) field.
Find the Water Surface AutoSetup Step: A message prompts you to set the SSD using the
imaging system. For instructions to set the SSD, see Varian document Setting up a Water Tank
for SSD Verification Rev 1.0.
Adjust the Ring Center Position to the Central Axis AutoSetup Step: One crossline scan
and one inline scan are acquired and then a message prompts you to adjust the couch LAT
and LNG based on the measured beam centers. The crossline and inline scans are repeated
until the residual centering is within tolerance. When complete, the AutoSetup panel displays
the total X and Y axis offset that were required to properly center the tank. The adjustment
magnitudes are based on the beam center found during the crossline and inline scans.
3D SCANNER Level AutoSetup Step: PDD measurements are acquired at three horizontal
positions within the tank and the results of these measurements are used to characterize the
alignment of the system with respect to the water surface. When the step completes,
evaluate the x and y axis tilts displayed in the AutoSetup panel. If the alignment is acceptable,
measurements may be collected using continuous scan mode. Otherwise, collect
measurements using step-by-step scan mode, which corrects the detector motions during
measurements based on the leveling results. For instructions to set up and collect scans in
continuous or step-by-step mode, see the SNC Dosimetry Software Reference Guide or the
SNC Dosimetry Help Index.
P/N 1230066, Rev Q Page 27 Using 3D SCANNER with a Halcyon Delivery System
Figure 34. Example of Completed AutoSetup for 3D SCANNER with Halcyon Delivery System
If you plan to acquire small field PDDs, it is recommended to acquire Ray Trace scans to ensure
that the detector follows the beam center.
Ray Trace setup must be completed before executing Ray Trace measurements. Ray Trace setup
determines the positions of beam center as a function of depth and the parameters that are used
to determine the Begin and End positions for a Ray Trace measurement.
Perform the following steps in SNC Dosimetry software to set up for Ray Trace measurements.
For detailed Ray Trace setup instructions, see the SNC Dosimetry Software Reference Guide or the
SNC Dosimetry Help Index.
a. In SNC Dosimetry software, right-click the Find the Ray Trace Angle AutoSetup step then
select Run AutoSetup Step.
b. In the dialog box, select the desired field size, shallow depth, and deepest achievable depth,
and then click OK.
c. A message asks you to set the selected field size and turn on the beam. Follow the instructions
and then click OK.
Crossline and Inline scans are executed at the selected shallow and deep depths with 3.00 cm
additional scan length for the selected field size. When complete, the angles and beam
centers are calculated and a message displays the ray trace parameters.
Using 3D SCANNER with a Halcyon Delivery System Page 28 P/N 1230066, Rev Q
3.15.2.1 Define Scans Using New > Scan Option
1 In the Queue tab Scan Plans panel, add a scan group. Right-click Scan Plans then select New
> Scan Group, or New > Quick Scan Group.
2 Right-click a scan group and then select New > Scan. Or, right-click a scan group and then
select New Quick Scan and scan type. Fill in the relevant fields and then click OK.
3 In the Scan Definition Properties dialog box, select the scan type, fill in relevant fields in each
tab, and then click OK.
Perform the following steps in SNC Dosimetry software to execute scans. For detailed
instructions, see the SNC Dosimetry Software Reference Guide or the SNC Dosimetry Help Index.
1 If desired, change scan mode by right-clicking on a scan in the queue and selecting QuickSet.
2 Select a scan group or a subset of scans in a group and then click Add Selected Items to
Queue.
3 Choose a project for the completed measurements or select New Project.
4 Click Start.
5 Perform analysis on completed measurements as described in the SNC Dosimetry Software
Reference Guide or the SNC Dosimetry Help Index.
3.17 Disassembly
1 Move the couch and 3D SCANNER out of the Halcyon bore and then drain water from the 3D
SCANNER.
2 Disconnect the PDI, Power, and Pendant cables from the electronics enclosure.
3 Disconnect the field detector cable from the field detector extension cable and then remove
the field detector from the detector holder.
4 Disconnect the reference detector from the REF port on the electronics enclosure and then
remove the reference detector from the detector holder.
5 Disconnect cables A and B from the left side of the Halcyon Bracket and then disconnect
cables A and B from the right side of the bracket.
6 Hold the Halcyon Bracket while turning the locking nut counter-clockwise. When the locking
nut is removed, remove the Halcyon Bracket from the dual motor housing.
7 With two people holding two Lift Ring handles each, lift the 3D SCANNER/Lift Ring assembly
from the Couch Positioning Plate.
8 Place the 3D SCANNER/Lift Ring assembly on a level surface, or, if you have a 3D MiniLift and
Automatic Leveling Platform (ALP), place the 3D SCANNER/Lift Ring assembly on the ALP, as
follows:
a. Orient the extended handle of the Lift Ring, the one closest to the electronics enclosure,
over the keyed corner of the ALP, and then place the 3D SCANNER/Lift Ring assembly
on the ALP.
b. Ensure that the Lift Ring is seated flush to the ALP and that the white supports on the
ALP are mated to the reliefs below each lift handle.
c. Connect the ALP1 and ALP2 cables to the ALP1 and ALP2 ports on the electronics
enclosure.
9 Connect the Pendant cable to the Pendant port on the electronics enclosure.
10 Connect the motor and sensor cables to the dual motor housing.
P/N 1230066, Rev Q Page 29 Using 3D SCANNER with a Halcyon Delivery System
11 Remove the Couch Positioning Plate from the couch.
12 Ensure that all 3D SCANNER parts are clean and dry before storing. For cleaning and storage
instructions, see Post-Measurement on page 39.
Using 3D SCANNER with a Halcyon Delivery System Page 30 P/N 1230066, Rev Q
4 Using 3D SCANNER
with CyberKnife
4.1 Overview
This procedure provides instructions for aligning the CyberKnife® so the beam is perpendicular to
the ring drive of the 3D SCANNER system. When this procedure is complete, CyberKnife beam
measurements can be collected using the 3D SCANNER system. This procedure includes the
following steps:
1 Prepare CyberKnife.
2 Prepare 3D SCANNER.
3 Prepare SNC Dosimetry Software for Scanning Stereotactic Fields.
4 Perform 3D SCANNER AutoSetup.
5 Set CyberKnife SSD.
6 Set CyberKnife Delivery Parallel to Gravity.
7 Verify 3D SCANNER / CyberKnife Setup.
8 Exporting to TPS.
Note: Video instructions for this procedure are available on the 3D SCANNER Support
website. Select the video titled 3D SCANNER Setup with CyberKnife.
Note: For additional information about CyberKnife, refer to your CyberKnife user
documentation or contact Accuray customer support.
This procedure prepares the CyberKnife accelerator to receive the 3D SCANNER by setting up the
reference detector port access and recording motor positioning alignment.
1 Enable access to the reference port so a reference detector can be connected later in this
procedure. The reference port may be located under the external housing (Figure 35).
External
housing
5 Adjust A5 and A6 so the beam is parallel to gravity (Figure 37). This can be done by placing a
cup of water on the floor underneath the laser, and looking back up at the port from which
the laser emits. Adjust A5 and A6 until the reflection on the laser’s exit port surface is
centered.
6 Record all motor positions from the CyberKnife Pendant, A1, A2, A3, A4, A5, and A6. Record
motor positions in the World Reference frame as well as in the reference frame of each
individual motor.
CAUTION: Be careful when moving the 3D MiniLift into the treatment room. Keep the
casters clear of imaging panels in the floor.
Robot
base
3 Connect 3D SCANNER cables. See Connect Cables in Control Room on page 13.
4 Using the pendant, home the 3D SCANNER Water Tank and Automatic Leveling Platform
positions:
a. Press Lift Table / Leveling Platform twice to select Advanced Functions mode.
LED “B” blinks slowly.
b. Press and hold Level 2 + for one second and then release to home the 3D
SCANNER Water Tank positions.
c. When Water Tank homing is complete, press and hold Lift + for one second and
then release to home the Automatic Leveling Platform positions.
d. When Leveling Platform homing is complete, press Lift Table / Leveling Platform
to exit Advanced Functions mode.
5 Using the pendant, fully retract the 3D MiniLift or 3D Lift.
a. Press and release Lift Table / Leveling Platform to select Lift Table / Leveling
Platform mode. LED “B” turns on.
b. Press and hold Lift - until the Lift Table is fully retracted.
6 Verify there is sufficient room beneath the CyberKnife gantry head to position the 3D
SCANNER.
• If there is not enough room under the gantry, make appropriate adjustments using the
CyberKnife’s Earth Coordinate system.
• Leave approximately 30 cm between the top of the 3D SCANNER and the bottom of the
accelerator head. This prevents collision while setting up the 3D SCANNER, and allows
you to set the SSD at precisely 80 cm after the Water Tank has been filled. See Set
CyberKnife SSD on page 36.
7 Check 3D SCANNER positioning. When positioned correctly, the CyberKnife laser should be
centered in the crosshairs in the bottom of the 3D SCANNER Water Tank (Figure 39).
Note: The 3D SCANNER is positioned correctly when the laser moves within and
parallel to the markings on the bottom of the 3D SCANNER Water Tank.
8 To correlate CyberKnife motor movement with the 3D SCANNER inline and crossline
directions, move the A5 motor and verify the laser stays parallel to the crosshairs on the
bottom of the tank.
If the laser does not stay parallel, physically rotate the 3D SCANNER by slightly adjusting the
angle of the 3D MiniLift or 3D Lift with respect to the A5 motor axis, and then check the laser
again. Deviations from parallelism will make altering the CyberKnife motors to deliver a beam
parallel with gravity more difficult.
Precise X / Y centering will be performed in Verify 3D SCANNER / CyberKnife Setup on
page 37.
9 When the 3D SCANNER is correctly positioned, lock the casters on the 3D MiniLift or 3D Lift.
10 Connect the Field and Reference detectors. Be sure to fully insert the Reference detector
cable in the reference port on the CyberKnife and secure it to the side of the CyberKnife with
appropriate tape (Figure 40).
11 If using the Sun Nuclear EDGE Detector as a Reference detector, ensure the back of the
detector is facing up before you insert it into the reference detector port. This places the
diode closer to dMax to obtain a better signal-to-noise ratio.
12 If using the Edge Detector as a Field detector, use the Edge Detector Setup Cap to ensure
proper positioning (Figure 41).
If you do not have a Setup Cap for your Field detector, set the geometric center of the active
volume 16 mm from the edge of the detector holder base. See Detector Setup on page 20.
Use a ruler or caliper to ensure proper alignment.
13 Fill the 3D SCANNER tank with water to the recommended fill line. See Fill Water Tank on
page 14.
1 In the control room, launch SNC Dosimetry software and ensure that a connection to the 3D
SCANNER is established.
2 In the Home ribbon, select System Preferences, then click Make Changes and ensure that
the selected device is 3D SCANNER.
3 Select the Processing tab then uncheck Perform Beam Center Averaging to provide more
accurate analysis parameter calculations for small field sizes.
4 Also in the Processing tab, select Analysis Protocol > CyberKnife.
5 Click Save to apply the changes.
Note: In older versions of the Incise, this field size may not be available. If so, select
the closest available field size.
5 Click Start, follow the on-screen instructions, and complete all of the AutoSetup steps.
• After the Find the Water Surface step, you will be prompted to set SSD. For instructions,
see Set CyberKnife SSD below.
• If you selected a Circular AutoSetup type, the Ring Angle Offset is automatically set to
~4.1 cm. Otherwise, the Ring Angle Offset AutoSetup step is performed to determine
the appropriate angle.
6 When AutoSetup is complete, define all scans required for the TPS and set up all scans with
scan speed of ≤1mm/second or Step by Step mode. For scan definition instructions, see the
SNC Dosimetry Help Index or the SNC Dosimetry Software Reference Guide, P/N 1230067.
2 Verify CyberKnife is parallel to the floor by checking the laser reflection. The laser should
reflect directly into the center of the laser port.
3 Verify that the flat plate by the A6 designator remains
parallel to the surface of the floor.
1 Right-click the Find the Ray Trace Angle step and select Run AutoSetup™ Step.
a. Set the depth and field size for Profiles to be collected during the Ray Trace procedure.
Perform both of the following procedures to verify the Water Tank is correctly aligned to the
delivery system.
If MLCs not attached: Define the field size using the primary collimator.
2 Define two scans, one at 95 degrees and the other at 130 degrees, and choose a slow scan
speed such as ≤1mm/second, or Step by Step mode.
3 Verify the status bar displays CyberKnife as the selected analysis protocol. If not, click
Current Protocol in the status bar, select CyberKnife, and then click OK.
4 Execute the scans then record the field size parameter displayed in the analysis panel for
each.
5 Enter the field size parameters in the following equation and calculate the result.
FieldSize 95° 180
CalcOffset = atan 1.428 – ---------------------------------- 1.743 ---------
FieldSize 130°
6 Enter CalcOffset in the following equation and calculate the result. The AngleOffset is found
in the results of the Determine the Angle Offset Relative to the Delivery System AutoSetup
step
AngleAlignmentError = 95 – CalcOffset – AngleOffset
7 The AngleAlignmentError indicates the degree of uncertainty. If adjustment is needed, do one
of the following:
• Repeat the Determine the ring angle offset relative to the delivery system AutoSetup
step. Right-click the step and select Run AutoSetup Step.
• Manually adjust the ring angle offset. In the Hardware Status panel Position tab, select
Polar, enter the angle that you just calculated, AngleAlignmentError, then click Move.
Then, right-click the Determine the ring angle offset relative to the delivery system
AutoSetup step and select Set.
For detailed MultiPlan export instructions, see the SNC Dosimetry Help Index or the SNC
Dosimetry Software Reference Guide, P/N 1230067.
5.1.1 Retract 3D MiniLift or 3D Lift and Move Water Tank to Drain Position
Keep hands away from the Automatic Leveling Platform when lowering the 3D
SCANNER Water Tank. A crush hazard exists between the Automatic Leveling Platform
and nearby objects, such as the treatment couch.
1 Using the pendant, fully retract the 3D MiniLift or 3D Lift. See Pendant on page 57.
a. From the pendant, press and release the Lift Table / Leveling Platform B button once to
select Lift Table/Leveling Platform mode. LED B turns ON.
b. Press and hold the Lift - button until the 3D MiniLift or 3D Lift is fully lowered.
2 Using the pendant, move the Water Tank to the drain position. See Pendant on page 57.
a. Press the Lift Table / Leveling Platform B button twice to select Advanced Functions
Mode 2. LED B blinks slowly.
b. Press and release the Level 1 + button.
CAUTION: Do not move the 3D SCANNER system until all water has been
removed.
Note: Only use microfiber and lint free cloth to dry the 3D SCANNER. Ensure threads
and debris do not fall into the 3D SCANNER drive components.
5.1.3.2 Procedure
1 Dry all surfaces of the Water Tank using a microfiber towel for the first pass and a microfiber
mitt for the second pass (Figure 43).
Delrin Bearing
Assembly
7 Dry the Pinion Gear on the Ring Drive then apply a small amount of lubricant to the Pinion
Gear (Figure 48).
Keep hands away from the motor housings when rotating the 3D SCANNER Ring
Drive. Hands/fingers can be pinched between the motor housings.
a. Using the pendant, rotate the Ring Drive completely in both directions.
• If LED “A” on the pendant is not illuminated, press the Water Tank A button to select
Water Tank mode. LED A turns ON.
• Press and hold the Ring + or Ring - button to rotate the Ring Drive.
b. Disconnect power from the 3D SCANNER and then manually rotate the Ring Drive in both
directions.
Ring Drive
Gears
10 Store the 3D SCANNER system in a location that meets the storage conditions described in
Storage or Transport on page 46.
Two person lift required for the 3D SCANNER Water Tank. Use proper lifting and han-
dling techniques to prevent injury or equipment damage. Lift the Water Tank using the
four lift handles. For weight specifications, see Specifications on page 65.
6 With two people grasping the handles at the base of the Water Tank, lift it away from the
Leveling Platform. Place the Water Tank on a level surface.
7 Using the provided 8 mm hex wrench, loosen the four bolts that secure the Automatic
Leveling Platform to the 3D MiniLift adapter plate, as shown in Figure 11 on page 10, and then
remove the bolts. Bolts are accessed from underside of the adapter plate.
8 Place the hex wrench and four bolts in a plastic bag and store them with other 3D SCANNER
ancillary parts.
Two person lift required for the Automatic Leveling Platform. Use proper lifting and
handling techniques to prevent injury or equipment damage. For weight specifica-
tions, see Specifications on page 65.
9 With two people supporting the Automatic Leveling Platform from the sides, lift it away from
the 3D MiniLift.
10 For 3D MiniLift disassembly and storage instructions, see the 3D MiniLift User Guide, P/N
1230037.
Do not lift an assembled 3D MiniLift. Remove the Lift Subassembly from the Frame,
then lift each piece separately.
To ensure ongoing support after the 3D SCANNER warranty expires, Sun Nuclear offers a Standard
Maintenance Agreement and Preventive Maintenance Inspection Agreement. For optimal
performance, it is recommended to schedule the Preventive Maintenance Inspection prior to
using the 3D SCANNER system for annual or periodic QA. For more information about 3D
SCANNER maintenance agreements, contact Sun Nuclear Sales at +1 321-259-6862, or fax +1
321-757-0066.
6.2.1 Repair
Field repairs may only be performed by Sun Nuclear personnel or by persons authorized by Sun
Nuclear to perform repairs on the 3D SCANNER.
To ensure operator safety and interference-free operation, field repairs must use parts provided
by or approved by Sun Nuclear. Violation will result in loss of warranty.
If you suspect product damage that could prevent the 3D SCANNER system from meeting its
intended use, contact Sun Nuclear. See Contacting Sun Nuclear Support on page 49.
6.2.2 Inspection
Periodically inspect the 3D SCANNER, cables, and ancillary parts for damage. If damage is visible,
discontinue use and contact Sun Nuclear Support. See Contacting Sun Nuclear Support on
page 49.
6.2.3 Cleaning
If additional cleaning is needed after performing the steps in Support and Maintenance on
page 45, observe the following guidelines:
• Use compressed air to dust gears and motorized parts.
• Use a soft sponge or a damp cloth to clean surfaces and cables.
• For spot cleaning, use a mild soap solution. Do not use abrasives or solvents, such as
alcohol, to clean the equipment.
• Disconnect cables from devices before cleaning.
6.3 Shipping
The environmental conditions for shipping or transport, temperature, relative humidity, and
atmospheric pressure, are listed in 3D SCANNER Specifications on page 65.
When shipping or transporting the 3D SCANNER system between facilities, it is recommended to
use the accessory shipping cases for the 3D SCANNER Water Tank, Automatic Leveling Platform,
and 3D MiniLift. See Shipping Cases on page 5.
For shipping case packing or unpacking instructions, see the following Technical Bulletins.
• Technical Bulletin 15-12, 3D SCANNER Shipping Cases Packing/Unpacking Instructions,
P/N 1230043.
• Technical Bulletin 16-12, 3D MiniLift Shipping Case Packing/Unpacking Instructions, P/N
1230044.
Two person lift required. Use proper lifting and handling techniques to prevent injury
or equipment damage. Lift the Water Tank using the four lift handles. For the weight of
each component, see Specifications on page 65.
6.5 Troubleshooting
* Other device data includes calibration file, ADCL calibration file, K L Values, Bucket DAC Settings, and Pulse
Height Variable.
The Sun Nuclear Product Warranty can be found on the Sun Nuclear Support website,
www.sunnuclear.com/support.
WaterProof
PROFILER
connector and
"+X" Label Water Sensor
connector
3D TPR bracket*
+Y Gantry
Label
Laser Alignment
Field Detector
Holder Base
* The Water Transfer Tube bracket is used with the 3D Reservoir Water Transfer Tube. The 3D TPR bracket is used
with the 3D TPR Sensor. See Options and Accessories on page 3.
+ -
Vertical
Drive
Diameter
Drive
Scanning Belt
Electrometer
Status LED
Reference
Detector Status Pendant
LED
Reference
Detector
Thermistor Input
(Reserved)
3D SCANNER Automatic
Status LED Leveling
Platform 1
Sensor for TPR
Accessory
External
Electrometer
Reserved
3D Reservoir
Automatic
Leveling
PDI Platform 2
Power
Connector Function
FIELD Triaxial BNC or TNC port for the field detector.
REF Triaxial BNC or TNC port for the reference detector.
ºC LEMO thermistor port for external temperature measurement. For future use.
TPR TPR Sensor.
Auxiliary External electrometer. For future use.
USB For future use.
Reservoir 3D Reservoir accessory.
PDI Power Data Interface Cable (from PDI 3.0).
Power Power cable from power supply.
ALP1 Connects to an Automatic Leveling Platform cable.
ALP2 Connects to an Automatic Leveling Platform cable.
Pendant Connects to the Pendant.
Connects to an external computer to upgrade firmware. The computer that is
Ethernet connected to the 3D SCANNER must comply with the IEC 60950-1, Safety of
Information Technology Equipment safety standard.
* Other device data includes calibration file, ADCL calibration file, K L Values, Bucket DAC Settings, and Pulse
Height Variable.
7.1.3 3D Electrometer
The 3D Electrometer is housed in the electronics enclosure. The electrometer has two triaxial BNC
or triaxial TNC1 inputs for connection to the field and reference detectors. If your detector has a
coaxial connector, install a coaxial to triaxial adapter..
Coaxial Triaxial
Figure 56. Coaxial Center Pin larger than Triaxial Center Pin
1. The electrometer input type, BNC or TNC, is selected when ordering the 3D SCANNER from Sun Nuclear. The
part number for a 3D SCANNER with TNC inputs is 1230000-0TZ.
WARNING: Set bias voltage to 0 Volts in SNC Dosimetry software before connecting
or disconnecting a radiation detector to or from the FIELD or REF inputs.
Note: Leave the protective caps on the electrometer FIELD and REF inputs when not
in use. The interior of the connectors has high impedance, meaning if contamination
enters it will cause leakage current to increase and possibly make the electrometer
unusable. Do not spray electronic contact cleaner into the connectors, as this could
increase leakage.
Ionization Chamber
When an ion chamber is connected to the electrometer, the center conductor carries the bias
voltage and the outer braid is grounded.
To
measurement
circuit
To
measurement
circuit
Leveling Platform X
motor
Ring drive rotation
Leveling Platform Y
Vertical (Z) drive up or motor
down
Leveling Platform Z1
Diameter drive motor
Leveling Platform Z2
Pump drain/fill motor
Pendant LEDs
Control State LED B, Lift Table or
LED A, Water Tank Automatic Leveling
Platform
Normal Mode, Pendant Idle OFF OFF
Lift Table or Automatic Leveling Platform Control OFF ON
Mode
Water Tank Control Mode ON OFF
Advanced Functions Mode 1 Slow Blink, 2 Hz OFF
Advanced Functions Mode 2 OFF Slow Blink, 2 Hz
Home 3D SCANNER positions, home Automatic
Leveling Platform positions, or move to Drain Fast Blink, 4 Hz Fast Blink, 4 Hz
Position
Press and hold to rotate the Ring Drive Press and hold to rotate the Ring Drive
clockwise (-Release to stop. a counterclockwise (-. Release to stop. a
Press and hold to move the Vertical Drive Press and hold to move the Vertical Drive
down (+SSD). Release to stop. a up (-SSD). Release to stop.a
Press and hold to move the Diameter Drive Press and hold to move the Diameter Drive
(-Dia). Release to stop. a (+Dia). Release to stop. a
3D Reservoir Pump drains Water Tank. b 3D Reservoir Pump fills Water Tank b
Press and hold an Increment button while Press and hold an Increment button while
pressing and releasing a Pump button to pressing and releasing a Pump button to
activate and deactivate pump. activate and deactivate pump.
Press & hold Increment + and press and release Pump + = 300 sec, (5 min.) FILL
Press & hold Increment – and press and release Pump + = 30 sec, FILL
Press & hold Increment + and press and release Pump – = 300 sec, (5 min.) DRAIN
Press & hold Increment – and press and release Pump – = 30 sec, DRAIN
Press and hold to select fine increment for Press and hold to select coarse increment
Ring, Vertical, Diameter Drive motion, and for Ring, Vertical, Diameter Drive motion,
for 3D Reservoir pump fill or drain. and for 3D Reservoir pump fill or drain.
Default Fine Increment c Default Coarse Increment c
Diameter: Dia = 1 mm Diameter: Dia = 10 mm
Vertical: Z/SSD = 1 mm Vertical: Z/SSD = 10 mm
Ring: = 15° Ring: = 90° (Cardinal Points)
Pump: 30 seconds Pump: 5 minutes (300 seconds)
Figure 61. Pendant - Water Tank Mode (Continued)
a. If the 3D SCANNER positions have not been homed, the Ring, Diameter, and Vertical drive controls are
disabled.
b. If the Valve Control Switch on the 3D Reservoir is not already set to the correct position, FILL or DRAIN, it may
take up to 20 seconds before the pump is activated. It is not necessary to touch the valve control switch on
the 3D Reservoir because the pendant will change the valve position automatically.
c. The default fine and coarse increments for Diameter, Vertical, and Ring drive motion can be changed in SNC
Dosimetry software system preferences. For details, see the SNC Dosimetry Software Reference Guide, P/N
1230067, or the SNC Dosimetry Help Index.
Press and hold to tilt left side of Press and hold to tilt left side of
Leveling Platform down. Leveling Platform up. Release
Release to stop. to stop.
Press and hold to tilt right side Press and hold to tilt right side
of Leveling Platform down. of Leveling Platform up.
Release to stop. Release to stop.
Press and hold to lower the Lift Press and hold to raise the Lift
Table. Release to stop. Table. Release to stop.
Figure 62. Pendant - Lift Table or Leveling Platform Mode (Continued)
Note: The above functions are only available if the Leveling Platform positions have
NOT been homed. For instruction to bypass the Leveling Platform Home Positions, see
Automatic Leveling Platform Home Positions Bypass Function on page 49.
Advanced Functions Mode 1, LED A Blinks Slowly Advanced Functions Mode 2, LED B Blinks Slowly
a. Ring Center (RC) is defined by software or the Pendant (default (RC):255 mm. The valid range is 240 mm RC
265 mm. The Ring Center is not necessarily equivalent to the crosshair position. To set the Ring Center manually,
see Manual Setup Step 3 - Set the Ring Center in the SNC Dosimetry Software Reference Guide or the SNC
Dosimetry Help Index.
b. This function is only available after homing the 3D SCANNER positions.
c. When complete, set the ring center.
d. Angle Offset valid range: 0° 10°.
e. When complete, set the water level.
f. This function is only available after homing the Automatic Leveling Platform positions.
Water Sensor
Field Detector
Holder Base
Figure 64. Example of Detector Holder Mounted on Field Detector Holder Base
For a description of the PDI 3.0 connectors and LEDs as well as PDI 3.0 troubleshooting
information, see Technical Bulletin 2-09, Power Data Interface (PDI) 3.0, P/N 0224011, which is
provided with the PDI 3.0.
’
Note: The previous version of the PDI, labeled PDI 2.0, is compatible with 3D
SCANNER. However, the first generation PDI, labeled P/DI, is not compatible and
cannot be used with the 3D SCANNER system. For more information about PDI device
types, contact Sun Nuclear Support.
The Automatic Leveling Platform attaches to the top of the Lift Table or 3D MiniLift and provides
a stable platform for the 3D SCANNER Water Tank. The Automatic Leveling Platform is required
to use the 3D SCANNER AutoSetup feature.
White
support
White
White support
support
Keyed corner
supports
7.3 3D MiniLift™
The 3D MiniLift positions the 3D SCANNER system in the beam under test and sets the vertical
position of the 3D SCANNER system.
The 3D MiniLift legs can be adjusted to three preset positions, as shown in the figure below. With
the legs in the Storage position, the 3D MiniLift can be easily transported in a vehicle. When the
3D MiniLift legs are in the Measurement position, the 3D MiniLift straddles the Floor Ring on the
treatment room floor and provides maximum stability for the 3D SCANNER system.
CAUTION: To avoid pinching fingers when changing the 3D MiniLift leg positions,
keep hands clear of the leg joints and frame.
Characteristic Specification
• Temperature: 10° to 40° C (50° to 104° F)
Storage • Relative Humidity: 10 to 90%, non-condensing
• Pressure: 57.2 to 106 kPa
• Ambient Temperature:18° to 30° C (64° to 86° F)
Operating • Relative Humidity: 20 to 80%
• Pressure: 57.2 to 106 kPa
• Temperature: –30 to 50 °C (–22 to 122 °F)
Shipping and Transport • Relative Humidity: 10 to 90%, non-condensing
• Pressure: 18.7 to 106 kPa
10 years, assuming the equipment is operated as described in Intended Use
on page iii, the equipment is operated and stored in an environment that
Service Life meets the temperature and humidity conditions listed below, and the equip-
ment maintenance agreement is renewed annually.
676
mm
673
mm
875
mm
Characteristic Specification
Channels 2
Charge 10 pC to no upper limit
Current 12 pA to 7.2 nA
Collection interval 25 ms to 100 ms
LINAC pulse count Included. With threshold detection
Bias voltage 0 V to ±400 V *
Tested to assure compliance with the relevant performance
Medical Electrical Equipment characteristics as documented in the IEC 60731 standard. The 3D
Classification Electrometer was tested as part of a scanning class dosimetry system.
* Although a bias voltage from -400 V to +400 V can be selected, the precision of the voltage is ±5%. For
example, if -350 V is selected, the electrometer sets the value from -367 V to -333 V.
* Height is with the Leveling Platform in nominal position with the leveling motors centered.
619 mm (24.37
inches)
619 mm
(24.37
inches)
For 3D MiniLift specifications, see the 3D MiniLift User Guide, P/N 1230037.
For 3D Reservoir specifications, see the 3D Reservoir User Guide, P/N 1230014.
Supported Detectors
This section lists the ionization chambers and diode detectors that are supported for use with 3D
SCANNER detector holders.
Table A-1. Detectors Supported by the EDGE Detector Holder Kit, P/N 1230370Z
Table A-2. Detectors Supported by the 6.35 mm Horizontal Detector Holder, P/N 1230339Z and the
6.35 mm Vertical Detector Holder P/N 1230344Z
Table A-3. Detectors Supported by the 7 mm Detector Holder, P/N 1230375Z and the 7 mm
Vertical Detector Holder P/N 1230465Z
* These radiation detectors cannot be used in the 7 mm Horizontal or Offset Detector Holders.
Table A-4. Detectors Supported by the 10 mm Detector Holder, P/N 1230340Z and the 10 mm
Vertical Detector Holder, P/N 1230445Z
Table A-6. Detectors Supported by the Universal Detector Holder, P/N 1230247Z
Table A-7. Detectors Supported by the Universal Vertical Detector Holder, P/N 1230274Z*
Table A-8. Detectors Supported by the 30 mm Parallel Plate Detector Holder, P/N 1230272Z
Table A-9. Detectors Supported by the 44 mm Parallel Plate Detector Holder, P/N 1230269Z
The Offset Detector Holder extends the 3D SCANNER diameter drive scanning range from 500
mm to 650 mm.
• Measurements using the Offset Detector Holder travel in the negative axis of the
diameter drive. See Scanning Mechanism on page 51.
• The Offset Detector Holder is not for use with vertically held detectors or Farmer type
chambers.
• The Offset Detector Holder does not support Exradin detectors or the EDGE Detector.
For instructions to set up SNC Dosimetry software to acquire measurements using the Offset
Detector Holder, see the SNC Dosimetry Software Reference Guide or the SNC Dosimetry Help
Index.
1 Attach the Offset Detector Holder to the 3D SCANNER detector base as shown. Tighten the
thumbscrews clockwise, finger tight.
Detector
Holder Base 7 mm or 10 mm
Horizontal
Detector Holder
2 Connect the detector to the electrometer as described in Install Field Detector Holder on
page 18.
3 Insert the detector in the 7mm or 10 mm horizontal detector holder, as shown in Figure A-2.
• Insert the detector in the detector holder from the center of the Water Tank such that the
active area is pointed toward the horizontal rail.
• If necessary, loosen the clamp thumbscrew (rotate counterclockwise), insert the detector
in the detector holder clamp, and push the detector through the detector holder until the
stem is visible.
• If the detector has a setup marking, orient it toward the radiation source.
• If you have a setup cap for the detector, use it to position the detector in the detector
holder as described in Detector Setup on page 20.
Optional: For better cable management, install a detector holder (7 mm or 10 mm) on the detector
holder base and route the detector cable through the strap and the detector holder.
Figure A-2. Installing Detector in the Offset Detector Holder (SNC125c Shown)
The Inverted Offset Detector Holder allows you to obtain large field scans (>50 cm) in the 145-
179 degree region as needed for certain treatment planning systems, such as Monaco CMS,
which requires a >50 cm scan at each 10 degree interval. For an example use case, see Monaco:
Large Scans (>50 cm Length) at 150-180 (330-360) Degrees in the SNC Dosimetry Software
Reference Guide or the SNC Dosimetry Software Help Index.
• The Inverted Offset Detector Holder is only for scans at 145 - 175 degrees.
• The Inverted Offset Detector Holder is not for use with vertically held detectors or Farmer
type chambers.
• The Inverted Offset Detector Holder does not support Exradin detectors.
For instructions to set up SNC Dosimetry software to acquire measurements using the Inverted
Offset Detector Holder, see the SNC Dosimetry Software Reference Guide or the SNC Dosimetry
Help Index.
1 Attach the Inverted Offset Detector Holder to the 3D SCANNER detector base as shown.
Tighten the thumbscrews clockwise, finger tight.
7 mm or 10 mm
Horizontal
Detector Holder
Inverted Offset Detector
Holder
2 Connect the detector to the electrometer as described in Install Field Detector Holder on
page 18.
3 Insert the detector in the 7 mm or 10 mm detector holder starting from the center of the
Water Tank such that the active area is pointed toward the horizontal rail.
• If necessary, loosen the clamp thumbscrew (rotate counterclockwise), insert the detector
in the detector holder clamp, and push the detector through the detector holder until the
stem is visible.
• If the detector has a setup marking, orient it toward the radiation source.
• If you have a setup cap for the detector, use it to position the detector in the detector
holder as described in Detector Setup on page 20.
Optional: For better cable management, install a detector holder (7 mm or 10 mm) on the detector
holder base and route the detector cable through the strap and the detector holder.
Figure A-4. Installing Detector in Inverted Offset Detector Holder (SNC125c Shown)
Note: SNC Dosimetry AutoSetup cannot be completed with a Farmer type ionization
chamber. AutoSetup should be completed with a small volume thimble chamber, Sun
Nuclear EDGE™ detector, or other diode detector prior to acquiring measurements
with a Farmer type ionization chamber.
When the SNC600c Farmer type ionization chamber is properly installed in the Farmer type
detector holder, the intersection of the geometric center of the active volume and the central axis
is located at the Water Tank’s ring center, 16 mm radially from the front face of the detector holder
base, and 30 mm vertically above the tip of the water sensor. The vertical offset from the water
surface must be set manually using the 3D SCANNER pendant or SNC Dosimetry™ software.
See Supported Detectors on page 69 for a list of Farmer type ionization chambers that can be used
with the Farmer type detector holder.
The Farmer type detector holder is required to support the Sun Nuclear 30 mm (P/N 1230272Z)
and 44 mm (P/N 1230269Z) parallel plate detector holders used with parallel plate ionization
chambers. See Installing Parallel Plate Detector Holder and Parallel Plate Ionization Chamber on
page 78.
1 For ease in setup, use the pendant or SNC Dosimetry software to raise the detector holder
base above water level. Remove any currently installed detector holders.
2 Using two captive thumbscrews, install the Farmer type detector holder in the 3D SCANNER
detector holder base. Note the orientation symbol on the detector holder to ensure the
detector is installed in the correct orientation to the Water Tank center. Rotate thumbscrews
clockwise until finger tight.
Align thumbscrews
over holes in Tighten
detector block thumbscrews
finger tight
Orientation
symbol
3 Set the 3D Electrometer’s voltage to 0 volts and verify one end of the field detector extension
cable is connected to the 3D Electrometer’s Field input.
4 Remove the connector’s protective cap and connect the Farmer type detector to the other
end of the 3D SCANNER field detector extension cable.
5 Install the Farmer type detector in the detector holder as shown in Figure A-6.
a. Loosen the thumbscrew that secures the detector in the detector holder.
b. Referencing the detector holder orientation symbol, insert the detector into the detector
holder so the thimble is pointed toward the center of the Water Tank (away from the hor-
izontal rail), then push the detector through the detector holder until the stem is visible.
c. Orient the rotational alignment marking toward the delivery system.
Loosen
thumbscrew
Figure A-6. Installing Farmer Type Ion Chamber in Farmer Type Detector Holder (SNC600c shown)
Note: Ensure a sufficient length of cable is available for 40 cm depth at the Water Tank
edge.
10 Rotate the ring drive and verify the cables do not tangle or snag.
11 Verify the gantry is in the fully upright position.
16
mm
Ring center
Figure A-8. Farmer Type Ionization Chamber Alignment in 3D SCANNER (SNC600c shown)
12 Referencing the three figures below, visually verify that the ionization chamber is aligned to
the delivery system, that the delivery system crosshairs align with the buildup cap markings,
and that the ion chamber is parallel with the water surface.
• The lateral effective point of measurement should be located 16 mm from the front face
of the detector holder base; alignment can be observed with congruence of crosshairs
and setup markings.
• The vertical effective point of measurement (central axis) is nominally located 30 mm
above the tip of water sensor when the detector is parallel with the detector holder base
(if the detector pitch is adjusted, reset the water level).
Check alignment of
delivery system crosshairs
to buildup cap markings
c. When the water surface bisects the dot marking or central axis, use SNC Dosimetry or
pendant to manually set the water level.
14 Confirm the setup by observing congruence of delivery system crosshairs with buildup cap
markings at 90°, 180°, and 270° rotation at the water surface, and at the intended
measurement depth, then return to 0° position. If congruence is not observed, check the
setup, and also investigate possible issues with gantry position, lasers, or crosshairs.
15 Remove the buildup cap.
16 Place the ODI Check on the water surface, then set SSD.
17 Use SNC Dosimetry software to select the Farmer type detector as the field detector, and
then apply voltage, collect background, and perform normalization.
Note: In the 3D SCANNER system, parallel plate ionization chambers are only used
for PDD scans or point measurements. SNC Dosimetry AutoSetup cannot be
completed with a parallel plate ionization chamber. Perform AutoSetup with a small
volume thimble chamber, Sun Nuclear EDGE™ detector, or other diode detector prior
to acquiring PDD scans or point measurements with a parallel plate ionization
chamber.
The 30 mm and 44 mm Parallel Plate detector holders (P/N 1230272Z and 1230269Z) are designed
to be installed in the Sun Nuclear Farmer type detector holder (P/N 1230267Z). For Farmer type
detector holder installation instructions, see Installing Farmer Type Detector Holder and Farmer
Type Ionization Chamber on page 74.
When a parallel plate detector holder is installed in 3D SCANNER the geometric center of the
active volume is located 16 mm from the 3D SCANNER detector holder base. The vertical EPOM
and the water surface must be set manually using the pendant or SNC Dosimetry software.
See Supported Detectors on page 69 for a list of parallel plate ionization chambers that can be
used with the 30 mm and 44 mm Parallel Plate detector holders. The 30 mm parallel plate detector
holder supports the Markus® and Advanced Markus® detectors’ protection cap used for in-water
measurements.
1 For ease in setup, use the pendant or SNC Dosimetry software to raise the detector holder
base above water level before beginning installation.
2 Install the Farmer type detector holder in the 3D SCANNER detector holder base (see
Installing Farmer Type Detector Holder and Farmer Type Ionization Chamber on page 74).
3 Insert the parallel plate detector holder in the Farmer type detector holder, then rotate the
thumbscrew clockwise until finger tight. If necessary, loosen the Farmer type detector holder
thumbscrew (rotate counterclockwise), before installing the parallel plate detector holder
When installed, the parallel plate detector holder must be flush with the Farmer type detector
holder to ensure correct alignment in relation to the 3D SCANNER detector holder base.
Insert
flush to
holder
Tighten
thumbscrew
Figure A-12. Installing Parallel Plate Detector Holder in Farmer Type Detector Holder (44 mm
Shown)
Figure A-13. Rotate Parallel Plate Ionization Chamber Underwater to Dissipate Air Bubbles Before
Installing (SNC350p shown)
Tighten thumbscrew
finger tight
Use
thumbscrew
to adjust roll
Turn knob to
adjust pitch
Figure A-14. Installing Parallel Plate Ionization Chamber in Detector Holder in 3D SCANNER Water
Tank
Note: Ensure a sufficient length of cable is available for 40 cm depth at the Water Tank
edge.
11 Rotate the ring drive and verify the cables do not tangle or snag.
12 Use SNC Dosimetry software or the pendant to raise the horizontal drive so the ionization
chamber’s top surface is level with the water surface.
Figure A-15. SNC Dosimetry Detector Details, Enter Manual Offset (SNC350p shown)
Note: If the detector properties offset is set to -0.11, as shown in the figure above, the
software assumes the user will manually align the top surface of the parallel plate
chamber to the water surface.
16 mm
• The geometric center of the active volume is located 16 mm from the 3D SCANNER
detector holder base when the ionization chamber and holder are correctly installed.
Alignment can be observed with congruence of crosshairs and setup markings.
1.07 mm
16
mm
Figure A-17. SNC350; Vertical and Radial EPOM Setup in 3D SCANNER (side view)
Note: If the ionization chamber is raised above the water surface while acquiring
measurements, the air must be dissipated from the ion chamber’s body cavity before
acquiring additional in-water measurements.
18 Use SNC Dosimetry software to select the parallel plate ionization chamber as the field
detector, and then apply voltage, collect background, and perform normalization.
Uninstalling
1 In SNC Dosimetry software, set the 3D electrometer to 0 volts.
2 Disconnect the ionization chamber from the 3D SCANNER’s field detector extension cable.
3 Use the pendant or SNC Dosimetry software to raise the diameter drive above water level.
4 Loosen the parallel plate detector holder thumbscrew (rotate counterclockwise) and remove
the ionization chamber from the detector holder following these precautions:
• Never push on the bottom center of a parallel plate ionization chamber to remove it from
the detector holder, as the slightest compression on the enclosure could affect
calibration.
• Do not pull on the cable to remove the ionization chamber from the detector holder.
Remove by grasping
ion chamber body or
stem only
DO NOT pull on
cable to remove
from holder
Figure A-18. Removing Detector from Parallel Plate Detector Holder
5 Remove the parallel plate detector holder from the Farmer type detector holder,.
6 Remove the detector cable from the cable clips and then remove the Farmer type detector
holder from the detector holder base.
7 Dry all parts thoroughly before storing.
For additional instructions, see the SNC Dosimetry Software Reference Guide or the SNC
Dosimetry Help Index, and the manufacturer’s instructions.
The universal vertical detector holder, PN 1230276Z, supports a range of detectors from 4 to 12.7
mm diameter. See Supported Detectors on page 69. Typical use is with a diode detector (e.g., IBA
diodes), or a detector that does not fit in one of the other vertical detector holders (6.35, 7 or 10
mm.)
The following symbols are used in this guide or in Sun Nuclear Corporation’s product labels.
WARNING: Indicates a hazard that could result in major injury or equipment damage.
(EN ISO 7010, W001)
CAUTION: Indicates a potential hazard that could result in minor injury or equipment
damage. (EN ISO 15223-1, 5.4.4)
CAUTION: Indicates possible injury to hands due to the closing motion of mechanical
parts or equipment. (EN ISO 7010, W024)
Indicates the user must not push on the equipment. (EN ISO 7010, P017)
Indicates the user must not step on the surface of the equipment. (EN ISO 7010, P019)
Indicates a floor level obstacle or trip hazard. (EN ISO 7010, W007)
By prescription only. Federal law restricts this product to sale by or on the order of a
qualified radiation therapy professional, such as a medical dosimetrist, medical
physicist, or radiation oncologist for use in radiation oncology QA.
(FDA 21CFR801.109)
Consult instructions for use. This equipment must be used in accordance with the
instructions in this manual. Read all instructions and safety labels before use. (EN ISO
15223-1, 5.4.3)
Operator Responsibility
This guide is intended for an operator who has a working knowledge in radiation physics and
dosimetry and experience working with treatment machines and accessories.
The operator of the 3D SCANNER bears the full responsibility for validating measurement results.
The operator is recommended to periodically (annually) compare measurement results against an
independent measurement device to ensure consistency of output. If the 3D SCANNER’s
measurement results are inconsistent or if you suspect damage to the equipment, contact Sun
Nuclear immediately (see Contacting Sun Nuclear Support on page 49.)
The instrument must not be used for any other purpose than described in this guide (see Intended
Use on page iii.) Violation could result in loss of warranty.
Should the need arise to report any safety or health related issues or concerns regarding the use
of Sun Nuclear products, contact Sun Nuclear directly.
EMERGO EUROPE
Prinsessegracht 20
2514 AP The Hague
The Netherlands
Modifications to Equipment
Any changes or modifications to the 3D SCANNER that are not expressly approved by Sun Nuclear
Corporation could void your warranty.
The instrument can only be connected to external computer equipment that is compliant with
IEC 60950-1, Safety of Information Technology Equipment.
EMC Precautions
The instrument complies with IEC 60601-1-2, Electromagnetic Performance standards, and the
operator should ensure that it is used in such an environment.
IMMUNITY Test IEC 60601 Test Compliance Levels Electronic Environment — Guidance
Level
Electrostatic ±8 kV Contact Complies Floors should be wood, concrete or
Discharge (ESD) ceramic tile. If floors are covered with
±15 kV Air
Immunity synthetic material, the relative humidity
should be at least 30%.
IEC 61000-4-2
Electrical Fast ±2kV on AC Mains Complies Mains power quality should be that of a
Transient / Burst typical commercial or hospital
±1 kV on I/O Ports
Immunity environment.
IEC 61000-4-4
Surge Immunity ±0.5 kV, ±1 kV, Complies Mains power quality should be that of a
±2kV CM Line- typical commercial or hospital
IEC 61000-4-5 Gnd environment.
±0.5 kV, ±1 kV,
DM Line-Line
NA on I/O Ports
Voltage Dips / C @ >100% 250 Complies Mains power quality should be that of a
Voltage periods (5 sec) typical commercial or hospital
Interruptions environment. If the user of the instrument
100% dip, 0.5 requires continued operation during power
IEC 61000-4-11 period mains interruptions, it is recommended
100% dip, 1 period that the instrument be powered from an
uninterruptible power supply or a battery.
30% dip, 25/30
periods
Power Frequency 30 A/m, 50 or 60 Complies Power frequency magnetic fields should be
Magnetic Field Hz at levels characteristic of a typical
Immunity Test commercial or hospital environment.
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Table B-3. Manufacturer’s Declaration—Electromagnetic Immunity for Systems that Are Not Life
Supporting
IMMUNITY Test IEC 60601 Test Compliance Levels Electronic Environment — Guidance
Level
Portable and mobile RF communications
equipment should be used no closer to any
part of the instrument including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Conducted 3Vrms, 0.15 - Complies Recommended separation distance
Immunity 80 MHz
IEC 61000-4-6 6Vrms on ISM d = 1.2 P
bands, 0.15 -
80 MHz
AC Mains and I/O
Ports
Radiated RF 3V/m, Complies Recommended separation distance
IEC 61000-4-3 80 MHz to 2.7 GHz
d = 1.2 P 80 MHz to 800 MHz
IMMUNITY Test IEC 60601 Test Compliance Levels Electronic Environment — Guidance
Level
NOTE 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency applies.
NOTE 2. These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a Field strengths from fixed transmitters such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
instrument is used exceeds the applicable RF compliance level above, the instrument should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such
as re-orienting or relocating the instrument.
b Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Essential Performance
Essential Performance of the 3D SCANNER system is defined as the collection of relative charge
measurements accurate to within 0.5% of the maximum charge per update. A failure to complete
a data collection scan will not violate Essential Performance if the data that is collected and used
by the user software meets the 0.5% requirement.
If Essential Performance is lost or degraded due to EM disturbances, the operator can expect to
see distortions in measured dose profiles in the form of reduced signal to noise or elevated or
depressed detector response in various regions of a scanned measurement. These distortions will
be apparent to the operator either through their masking of key measurement features, or by data
comparison to operator or manufacturer established baselines and standards for the applicable
radiotherapy delivery systems.
Environmental Requirements