AXIOM Artis TC/dTC/TA/dTA (e)
© Siemens AG 2002
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights, in-
cluding rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: AXA4-500.891.02.05.02 Doc. Gen. Date: 04.06
Replaces: AXA4-500.891.02.04.02
0-2 Revision
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho-
rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Page
1 _______General Remarks _______________________________________________ 1 - 1
3 _______Installation Preparation__________________________________________ 3 - 1
Page
Cooling System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Lower Body Radiation Shield (Option) . . . . . . . . . . . . . . . . . . . . . . . 3-2
Upper Body Radiation Shield, Surgery Light, Injector (Option) . . . . . . . . . . . 3-3
Wireless Footswitch (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
On-Floor-mounting of the System . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Installation Directly on Solid Concrete . . . . . . . . . . . . . . . . . . . . . . . 3-5
Installation on Filled Concrete . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
System Ceiling Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
With Movable Upper Body Radiation Shield (Option) . . . . . . . . . . . . . . . 3-7
In-house Electrical Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Remarks regarding the recommendation on Page 3 - 10/11 for the
room electrical installation according to the DIN VDE 0100-710 . . . . . . . . . . 3-9
Recommendation for Application Group 1 . . . . . . . . . . . . . . . . . . . . .3 - 10
Recommendation for Application Group 2 . . . . . . . . . . . . . . . . . . . . .3 - 11
Power Cables by Customer‘s Contractor . . . . . . . . . . . . . . . . . . . . .3 - 12
Siemens Remote Service (SRS) . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 16
Network. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 17
DICOM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 17
Injector Floor Outlet Option. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 18
Page
7 _______Detail Planning_________________________________________________ 7 - 1
Generator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-1
System Control Cabinet #1/2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Image System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
Ceiling Stand: Ceiling Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
Ceiling Stand: Installation of the Longitudinal Rails . . . . . . . . . . . . . . . . . . .7-7
Control Room Distributor (BTEE) . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
Control Console (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-9
Tabletop Opening for Control Modules and ECC . . . . . . . . . . . . . . . . . . . 7 - 10
2nd Data Display (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 11
Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 11
Moveable Upper Body Radiation Shield, Surgery Light (Option) . . . . . . . . . . . 7 - 12
Trolley for Control Consoles (Option) . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 13
Ceiling Outlet for Trolley . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 14
Injector Wall Terminal Box (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 15
Planning Recommendation for Injector Wall Connection . . . . . . . . . . . . . 7 - 16
Injector Rack Mount Unit (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 17
Connecting an Non-Siemens ECG . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 18
System Configuration 1
This Planning Guide (PG) has been prepared for the AXIOM Artis TC/dTC/TA/dTA Sys-
tem. This is comprised of:
Ceiling stand + patient table (4 different versions) + DCS
Remarks 1
• The changes to the previous version of this document are labeled by the following sym-
bol:
• A note must be made on all room plans that are created by the Planning Departments
which makes reference to the General Installation and Delivery Conditions of Siemens
Medical Engineering.
A copy of the General Installation and Delivery Conditions must accompany the plans
when they are turned over to the customer.
• For required spaces, the documents of the application fire safety regulations must be ob-
served.
• The system was developed in accordance with IEC 60 601-1.
• The system does not meet the requirements for AP classification. Use in areas where
there is an explosive atmosphere as defined by IEC 610-1, Main Section 6, is not permit-
ted.
• The AXIOM Artis OR patient table is splash-water protected (IPX4 per IEC 60 601-2-46).
• Unless otherwise specified, all dimensions are given in mm.
Definition of Terms 1
• Orientation points
Points specific to system components to which reference is made when positioning sys-
tem components to each other or in the room.
The isocenter of a radiographic system is always illustrated as the orientation point.
Illustration in the drawings:
• Fixpoints
Clearly marked points on system components, installation ceiling, walls or floor on which
cable outlets are located.
Illustration in the drawings: circle with letter/number combination.
Example: PU1
The cable lengths establish the maximum fixpoint distances and thus the maximum dis-
tances between the individual system components.
• Room height
The room height is the distance measured from the top surface of the finished floor to the
bottom surface of the ceiling structural elements (Unistrut rails) (bottom surface of drop
ceiling).
• PG: DCS
AXA4-700.891.01...
• PG: Klüver Cooling Unit
RA71-020.891.01...
• Measuring station: see the component PG
• PG: SIEMENS Remote Services (SRS)
TD00-000.891.01...
• PG: DICOM
TD00-000.891.02...
• PG: AXIOM Artis Emergency Power Supply
AXA4-770.891.01...
• PG: Ultrasound Integration
AXA4-810.891.01...
• PG: Multi Modality Viewing
AXA4-100.891.14...
• PG: Voice Control
AXA4-100.891.15...
Room Plan 2
Recommendation 2
Control
console
Equipment
room
Patient table
Room height:
2700 - 2740 mm
(see Page 2 - 7)
• Plan the control room at the foot-end of the patient table (optimum visual contact to the
examination room).
• Plan the radiation protection window in terms of size and position so that the examina-
tion room can be clearly viewed from the control room.
• According to DIN 6868-57 (international standard in preparation), lighting in rooms in
which diagnosis is made on image playback/display devices (monitors) must meet the
following requirements:
- Adjustable, no-glare screens, reproducible adjustment of luminous intensity (e.g. dim-
mer with a scale).
- No glare or reflection from windows, lights and light boxes in the vicinity in which the
monitors are normally used.
• Plan the entrance area as illustrated.
• Plan the equipment room to the left of the examination room.
• If at all possible, do not plan the generator and cooling unit within the examination room
(because of generation of noise and heat).
• OR patient table (option)
When the OR table is used for surgical applications, an on-site UPS should be planned.
For further information, see Page 2 - 8.
Optimum Room 2
Full rotation of the C-Arm system (+/- 135°) possible in the head-end park position.
*2
*4 *2
Park position
*3
*1
*1
*5
Park position
1:50
Park position of the C-Arm systems at the head end possible only laterally (+90° or -90°).
Side change of the C-Arm system possible after moving the ceiling stand out of the park-
ing position in the foot-end direction.
*2
*5 *2
*3 *2
Park position
*1
*1
*4
Park position
1:50
*2
*1
1:50
*1 Room height measured from the highest point of the finished floor (including floor
covering) to the lowest point of the ceiling substructure Page 2 - 7)
*2 Required minimum height in the drop ceiling for corrugated hose holder and lay-
ing cables; for details regarding the ceiling outlet, see Page 7 - 6
*2
*1
1:50
*1 Room height measured from the highest point of the finished floor (including floor
covering) to the lowest point of the ceiling substructure Page 2 - 7)
*2 Required minimum height in the drop ceiling for corrugated hose holder and lay-
ing cables; for details regarding the ceiling outlet, see Page 7 - 6
Planning Information 2
When planning external components in the examination room, pay attention to the
movement ranges of the system components (risk of collision!).
Ceiling stand
movement range *1
Ceiling stand
movement range
1:50
*1 With the minimum room dimensions (see Page 2 - 3), a side change of the
C-Arm system in the park position is not possible. To change the side, the ceiling
stand must be moved out of the park position.
Ceiling substructure
2700 - 2740mm *1
(e.g. Unistrut)
Room height
Finished floor
(incl. floor covering)
*2
1:20
*1 Room height measured from the highest point in the cross hatched area on the
finished floor (incl. floor covering) to the lowest point of the ceiling substructure.
Here, the area of the ceiling substructure in which the longitudinal rails for the
ceiling stand are installed is the determining factor.
Recommended room height: 2710 mm
*2 The cross-hatched area is determined by the movement range of the ceiling
stand C-Arm.
Max. adm. floor unevenness in the shaded area: 8 mm
Patient Table 2
Versions 2
There are 4 different versions of the AXIOM Artis patient table that can be shipped:
1. Basic table
2. Peri table (same as the basic table, but also with motorized longitudinal move-
ment of the tabletop)
3. Tilting table (same as the Peri table, but also with a tilting function in the longitudi-
nal direction)
4. OR table (same as the tilting table, but also with the tilting function in the trans-
verse direction)
The AXIOM Artis patient tables all have the same external dimensions.
• An additional control cabinet (SC2) is shipped along with the OR table. For the most part,
the cabinet contains the motor controller. Unlike with the other AXIOM Artis patient ta-
bles, the motor controller cannot be installed in the table base because the tilt and cant
functions would have limited movement.
• The OR table is splash-water protected (IPX4 per IEC 60 601-2-46).
• Use in the OR environment
If the table is used for surgical applications, it must still be possible to use the table even
if there is a power failure, as defined by IEC 60 601-2-46. In this case, the SC1 system
control cabinet must be connected to an on-site, uninterruptible power supply (UPS).
Since all system components, with the exception of the generator, are supplied with
power via the system control cabinet, the units, the image system, etc. must still be oper-
ational if there is a power failure.
For detailed information about the UPS, see AXIOM Artis Emergency Power Supply PG,
Print No. AXA4-770.891.01...
*6
View X *4
*5
*5 X
*1 *2
*7
*3
*8
*10
*9
*9
*1 0
1:50
T1 Patient table, floor fixpoint/orientation /rotation point: for cable routing, see Page 3 - 5/6
*1 Socket (Option) *6 Table tilt, transverse (option)
(see Page 3 - 15)
*2 Panel with injector connector *7 Table lift
*3 Panel with nipple for compressed air connec- *8 Table longitudinal movement
tion
*4 Accessory rail *9 Table transverse movement
*5 Table tilt, longitudinal (option) *10 Table rotation range
Recommendation 1 2 Recommendation 2 2
Cooling unit
cabinet
• Recommendation 1:
Generator (PU1), system control cabinets (SC1/SC2) and cooling unit (CU1) located in equip-
ment room (due to generation of noise and heat).
• Recommendation 2:
Generators (PU1) and system control cabinets (SC1/SC2) located in the examination room,
cooling unit (CU1) located in the equipment room.
Caution: Take note of noise and heat generated by the generators (PU1)!
• Image system (IS):
is regularly positioned within the control room (as standard). Its optional positioning in the equip-
ment room is possible only if this was explicitly indicated to Logistics already in the time of order.
• Observe the heat dissipation data and temperature ranges of the cabinets and components (see
Chapter 5). If necessary, air condition the equipment room.
• Systems with flat panel detector: the max. admissible height difference between the detector
and the SC1 control cabinet is 10 m.
• The cabinets can be set up as illustrated. Take note of the following:
- The cabinets can be set up separated from each other or corner-to-corner only at the places
marked with arrows. When this is done, take note of the cable list on Page 4 - 2.
- If set up is corner-to-corner, maintain at least 40 mm space to the next cabinet (needed to
open the cabinet doors).
- PU1 and cable cabinet can also be positioned to the right of the SC1/SC2.
- For additional information, see Page 7 - 1 to Page 7 - 3.
• Cable cabinet (option)
- The cable cabinet is required if the excess cable lengths can not be stored in some other man-
ner (e.g. under computer flooring). It has the same dimensions as the generator cabinet. Ca-
ble lead-in is possible from the top left/right (see Page 4 - 9), into the cabinet base from below,
and from all sides (except the front).
- If a cable cabinet is required, please mark this on the checklist that is sent from Logistics. The
cable cabinet is provided at no charge.
Control Room 2
Hardcopy
Image system
• Plan the control room in a way that it is at the foot-end to the examination room.
• The control console and second display (not shown; for details see
Page 7 - 9) can be ordered as options.
• Control modules (option), ECC (option)
- The control modules for ceiling stand, collimator and patient table are not suitable for
free setup. The modules must be installed in the console. The opening must be made
on-site by a carpenter using a router. For details see Page 7 - 10.
- An emergency stop switch is shipped along with the control modules (not with the
ECC). This must be positioned within easy reach for the operator.
• Position the components in the vicinity of the control room distributor (see Page 4 - 5).
• The location of the image system ought to be planned in the control room (as standard).
Its optional positioning in the equipment room is possible only if this was explicitly indi-
cated to Logistics already at the time of order.
• For the Multi Modality Viewing Container (Option), see PG AXA4-100.891.14...
UPS *1 *3
*1 Option
*2 Image system in the control room = standard, optional in the equipment room; must be specified
at the time the order is placed; for details, see Page 7 - 5
*3 UPS (option) for the image system has to be located in close vicinity to the Image system. To
ensure proper air circulation a minimum free space of 25 mm all around is needed.
*4 An RPAU is necessary, if the image system is located outside of the control room.
*5 Emergency stop switch
For details regarding the control room distributor, see Page 7 - 8.
Static Requirements 2
Ceiling Substructure 2
Patient Table 2
The max. pull forces are exerted with the table rotated (+/-45o) and with the tabletop
moved fully out towards the head end. The max. pull forces were determined with a
patient (250 kg) and table accessories (40 kg).
The installation plate is anchored with 4 Hilti heavy-duty expansion bolts in proven con-
crete.
Minimum concrete quality: C20/25.
Load at the foot end per mounting point: max. pull forces: 4.5 kN
Note:
With the "standard tabletops", the max. admissible patient weight is 200 kg. The tabletop
with a max. admissible patient weight of 250 kg is shipped exclusively for the patient table
The shifting center of gravity with rotation of the table was determined under the follow-
ing conditions:
- Tabletop fully moved out to the head end.
- Tabletop moved in the transverse direction to the right up to stop.
- Tabletop not tilted.
- Tabletop loaded with a patient weight of 250 kg.
- Table accessories of 40 kg hooked onto the right accessory rail.
Orientation point/
rotation point
General Remarks 3
Required Aids 3
Ceiling Stand 3
Patient Table 3
DCS 3
• Place the control room distributor in the control room either under the control console
(option) or mount it on the wall. For details, see Page 7 - 8.
• If configured, plan the optional second footswitch and/or control modules (floor stand,
patient table) in the control room so that when they are operated, there is visual and
acoustical contact to operating personnel, to the patient and to the injector in the exami-
nation room.
• When setting up the DVD recorder, make sure there is adequate heat venting. The clear
space to the walls, to built-in fixtures or to other components must be at least 3 cm.
• Positioning is optional in the DCS (standard), and also in the control room as an option
(see Page 2 - 11).
• Positioned in the control room or in the examination room on the wall or on a console.
• For details, see Page 7 - 11.
Control Cabinets 3
• For floor and wall mounting, as well as for cable lead-in, see the chapter "Detail Plan-
ning", Page 7 - 1 to Page 7 - 3.
• Store excess cable lengths in the cable cabinet (option), under computer flooring or in
cable conduits, etc. (see Page 4 - 9).
Cooling System 3
The lower body radiation shield is mounted on the left or right on the table.
1. Overview
Approved Attachments
Ceiling column mounted - Upper body radiation shield
on the carriage and/or surgery light
(for Dimensional Drawing, see Page 7 - 12;
for positioning, see
Page 3 - 7)
4. Injector (Option)
• For details Injector Rack Mount Unit see Page 7 - 17.
Patient Table
The mounting plate must be installed on a solid base that has sufficient load capacity. Cut
away the floor covering. If necessary, any material in the location of the mounting plate
that does not have the required load capacity must be replaced with filled concrete.
Hilti heavy-duty expansion bolts are included in the shipment for installing the mounting
plate:
• Longer heavy-duty expansion bolts are available from the Hilti Co.
• If necessary, the mounting plate can also be installed using M12 threaded studs, min.
hardness rated 8.8 per the ISO norm, which are inserted through the ceiling of the room
(threaded studs, etc. must be obtained locally).
Patient table
installation plate
Ceiling stand
longitudinal
ceiling rails
Orientation point
patient table *1
1:50
*1 *3 *1
*2
Head end
*5
*1 *1
*1 Patient Table
installation plate
Solid concrete
*4
1:10
*4 Cable led in through an opening in the floor from the room below
*1 *1
Cable duct,
e.g. 200 x 70
*3
Head end
*2
*4
*1 *1
Size of area for
filled concrete
*1
Patient table installation plate
Self-leveling grout
1:10
*2 Area for cable lead-in to the table base; lay cables in the cable channel or through an opening
in the floor from the ceiling in the room below.
If laying cables in the cable channel: to keep the cables from being too sharply bent in the area
where they are led into the table base, plan only cable channels with a min. height of 70 mm.
*3 Opening in the mounting plate
Grid spacing
*2
*1
*3 *2
1:50
Do not plan any lighting fixtures, ventila-
tion inlets, ventilation outlets, sprinkler
elements, etc. between the rails.
Ceiling stand
Radiation shield
Recommendation for
room distributor
Legend:
1 Potential equalization leads to external, conductible parts
2 Power line to permanent or portable operating devices
3 Power line to permanent or portable operating devices, e.g. developer machine
4 GFI (ground fault interrupter)
4a GFI for system power distributor, see note on Page 3 - 9
4b GFI for generator, see note on Page 3 - 9
4c Back-up fuse for generator, see Page 3 - 9
5 System breaker
6 Isolation transformer with overload monitoring per DIN VDE 0100-710, Section
710.512.1.6.2 (output power 3.15 - 8 kVA, isolation test voltage 4kV)
7 For connection of the MMV, see PG: AXA4-100.891.14... (Option)
EAT On-Off switch with control lamp for the X-ray system
AT1 Emergency OFF switch in examination room
AT2 Emergency OFF switch in control room
AT3 Emergency OFF switch in equipment room
[] Figures are free cable lengths in m
Note: The ratings for the short-circuit fuses apply only for SIEMENS GFIs
5 x per
calculation
NOTE - The system control cabinet #1 and generator require only Phase L1 - L3
and Gnd.
- In some countries, N is required to conform to country-specific regulations.
- If a fault current switch is available or is required, N must also be connected.
Angio Generator M
P
UN
100 kW 80 kW
400 V* ≤ 0.10 Ohm ≤ 0.14 Ohm
415 V** ≤ 0.14 Ohm ≤ 0.17 Ohm
440 V** ≤ 0.16 Ohm ≤ 0.21 Ohm
480 V** ≤ 0.20 Ohm ≤ 0.27 Ohm
POLYDOROS A100
P
UN
100 kW 80 kW
380 V ≤ 0.08 Ohm ≤ 0.10 Ohm
400 V ≤ 0.09 Ohm ≤ 0.11 Ohm
420 V ≤ 0.09 Ohm ≤ 0.12 Ohm
440 V ≤ 0.10 Ohm ≤ 0.14 Ohm
480 V ≤ 0.12 Ohm ≤ 0.16 Ohm
Resistance values in Ohms with UN+/- 10%
Room distributor *1
*2a *2b
max. 250 V AC
max. 2A
max. 250 V AC
max. 2A
*4 *4
- If the socket has been ordered, this is already installed in the table in the factory.
- All patient table are already delivered ex works with drawn-in power cable
2 x 1.5 mm 2.
- In systems without MMV, the power cable must be connected to an on-site isolating
transformer (with overload monitoring) according to DIN VDE 0100-710 Section
710.512.1.6.2 (output power 3.15 - 8 kVA, insulation test voltage 4 kVA). Secondary-
side fuse 115 V: 5 A; at 230 V: 2.5 A
Public tele-
phone system
ISDN router
New modality
Switched or cus-
tomer’s own network
Any existing modality
Remote diagnostics and "Pro-active Event Management" require remote LAN access to
our systems. A router is required for this which we provide to our customers at no charge
during the warranty period and beyond that period if a service contract is concluded.
The router remains the property of Siemens AG. The required telephone connection (pref-
erably ISDN) and the power supply must be provided on site. Even if several Siemens sys-
tems are connected to the same data net, only a single central router is required.
Of course, we observe the confidentiality of patient data in the framework of remote diag-
nostics and have provided security against access to it. Further information on the topic of
data security can be found in the system-specific publications.
Network 3
The image system is equipped with a network card for Fast Ethernet (100 BaseT).
Network connection with System Switch (in the scope of supply)
-Install the System Switch preferably in the control room distributor or on the wall (refer to
manual supplied)
1X 2X 3X 4X 5X 6X 7X 8MDI-X 8MDI
Sw itch
Sw itch
Sy stem-Switch
(Ethernet / DICOM) Artis Image System
1X HOST PC.LAN_A
L1
N
PE 2X Leonardo (3D)
3X Sensis ACQ
Sensis Postprocessing
4X WS
/Report WS
7X ACOM PC
R-DIAGRouter
DICOM 3
Head end
Medrad:
Gary Bastin (Cardiovascular Product Manager)
Medrad Inc.
One Medrad Drive
Indianola, PA 15051-0780
gbastin@medrad.com
+1 (412) 767 8012
Typco:
Pat Dyer
Site Planning Dept.
2111 E. Galbraith Rd.
Cincinnati, Ohio 45237
pat.dyer@tycohealthcare.com
+1 (513) 761-2700 ext.5721
You can receive detailed information for Planning/Project Planning from the listed
addresses.
E.g. the floor outlet from Medrad:
General Remarks 4
• All the system-specific cables (including ground wires) are included in the regular scope
of delivery, with the shipment.
• The system cabling comprises the standard cables and standard wiring harnesses with
fixed lengths. These lengths may not be shortened.
• Store excess cable lengths in the cable cabinet (option), in cable conduits, under com-
puter flooring, etc.
• For remarks regarding the room electrical installation, see Page 3 - 9.
RD
RD
RD RD
Ethernet
Remarks:
• RD = room distributor
• Room distributor and power cables are not included in the shipment.
• The illustration includes the cable cabinet (option).
The wiring harnesses can be laid directly to the fixpoints, i.e. without first going
through the cable cabinet. Excess cable lengths must then be stored in some other
manner (in cable conduits, under computer flooring, etc.).
• For the definition of the fixpoints, see Page 1 - 3.
Keyboard 3.5 m
Mouse 3.5 m
Options:
2m Reference display** 3.5 m*
2m Hand switch
2m Control modules
2m ECC
4m Emergency stop
1.7 m Foot switch
2.5 m DVD recorder 3.5 m*
Image system,
located in the
control room
Control room
distributor
IS
CR1
* video cables
* Video cables
** Monochrome and color displays can be shipped:
Max. distance to
Type of display image system in m
17" color display 7.5
18" color display 3.5
18" monochrome display 3.5
If the image system is in the equipment room, the video cables are laid through the control
room distributor to the display's, etc.
Standard:
3m Remote Power and Archive Unit
2m Flat screen monitor (Display) *
2m Keyboard
2m Mouse
Options:
2m Reference display *
2m Hand switch
2m Control modules
2m ECC
4m Emergency stop
1.7 m Foot switch
2.5 m DVD recorder
CR1
*4
internal 2.5 m 2
SC1 Room light
max. 2.5 mm² flexible
K4
max. 250 V AC
max. 2 A *2
*4
internal 2.5 m 2 external radiation
SC1
max. 2.5 mm² flexible display
K5
max. 250 V AC
max. 2 A
24 V
24 V
Potential-free contact
internal 2.5 m 2 (24 VDC / 20mA) of UPS *3;
SC1 emergency power
max. 1.5 mm² flexible
"on"= contact closed
*1 If needed, a switch for radiation blocking can be installed in series with the door
contact:
- The switch (24 VDC / 20 mA) must be provided by customer’s contractor.
- Control is +24 V from the system control cabinet.
- Plan in the switch so that unintentional operation is not possible.
- The switch must be provided with an appropriate symbol (or label) so that its
function is easily recognizable.
*2 If the current load of 2 A is insufficient, an external power relay must be planned,
over which the room light can be switched.
*3 The signal "emergency power ON" disables/interlocks the radiography mode;
thereafter only the emergency fluoroscopy mode remains enabled.
*4 With systems shipped prior to 6/2005, the K4/K5 relay must be provided on site.
Control of the relay is made with 24 V DC (max. 0.2 A). A free-wheeling diode
must be circuited over the relay coil.
SC1 System controller fixpoint
Laying Cables 4
Remarks:
• Lay cables / wiring harnesses separately in pipes or in enclosed cable conduits.
• If cables are laid in open cable conduits, separate cables using metal spacer plates or a
similar method.
• The depth of the cable conduit should be at least 70 mm; if there are intersections, plan a
greater depth.
For all cable conduit / pipe sections, calculate the sum of the cross sections of all cables/
wiring harnesses. The individual cross sections of the cables/wiring harnesses can be
found in the cable list. The required fill factors are already included in this.
If excess cable lengths are to be stored inside cable conduits, this fact must be taken into
account when calculating the size.
Electrical Data 5
Examination Room 5
each display
D1 DCS see PG DCS see PG DCS
< 0.1
Equipment Room 5
Dimensions Heat
Weight
Fixpoint Subsystem (W x D x H) dissipation
[kg]
[mm] [kW]
PU1 Angio generator (Power Unit 1) 800 x 434 x 2202 350 1.0
PU1 POLYDOROS A100 800 x 434 x 2202 300 1,0
Control Room 5
Dimensions Heat
Weight
Fixpoint Subsystem (W x D x H) dissipation
[kg]
[mm] [kW]
n.a. UPS for image system (option) 170 x 439 x 216 19 0.03
Remote Power Archive Unit (RPAU);
n.a. 216 x 368 x 222 9 < 0.1
only if Image system in equipment room
n.a. Control module patient table (option) 91 x 161 x 65/85 <1 n.a.
Environmental Conditions 5
Examination and control Temperature range: 10° C ... 35° C *1 (recomm. temperature 22° C)
room 15° C ... 30° C *2
Relative humidity: 20 - 75% non-condensing
Image System (IS) Temperature range: 10° C ... 35° C
Relative humidity: 20- 75% non-condensing
max. temperature gradient: 10 C/h
Air flow volume: 657 m3/h
max. noise generation: 50 dB(A)
Angio generator M/POLY- Temperature range: 10° C ... 35° C
DOROS A100 (PU1) Relative humidity: 20 - 75% non-condensing
max. temperature gradient: 5° C/h
Air flow volume: 800 m3/h
max. noise generation: 55 dB(A)
System Control Cabinet #1 Temperature range: 10° C ... 35° C *1
(SC1) 15° C ... 30° C *2
Relative humidity: 20 - 75% non-condensing
max. temperature gradient: 5° C/h
Air flow volume: 500 m3/h
max. noise generation: 48 dB(A)
System Control Cabinet #2 Temperature range: 10° C ... 35° C
(SC2) Relative humidity: 20 - 75% non-condensing
max. temperature gradient: 5° C/h
Air flow volume: n.a.
max. noise generation: n.a.
Cooling unit (CU1) Cooled air: 5° C - 30° C (frost-free room)
(for Megalix tube units) Air volume: 1100 m3/h
max. noise developed: 55 dB(A) at 50 Hz; 59 dB(A) at 60 Hz
Stand with flat panel detector max. temperature gradient: 5° C/h
Atmospheric pressure: 700hPa ... 1040hPa
Shocks: max. 10 g / 16 ms
Vibrations: max. 0.1 g / 10 - 200 Hz
Transport/Storage 5
NOTE In systems with flat panel detectors, the detector is removed from
the stand for transport to the customer. The limited transport and
storage conditions apply for the detector.
Paint Surfaces 5
Largest crate with packaging, weight W 2630 x D 1180 x H 2070 mm, approx. 1175 kg
Largest single part with transport W 2470 x D 1000 x H 1900 mm, approx. 910 kg
carriage (min. door opening), weight
Generator 7
*4 Exhaust *4
air
*2
*3 *3
Air intake
*1 *1
1:20
Exhaust air *2
*1
*2 *2
1:20
Exhaust air
*4
*4
*3 *2
Air intake
1:20
*1 *1 *1
Cabinet base
Top-down view
Exhaust air
*2
*1
*3 *3
1:20
Image System 7
Exhaust air
*1
Wall
*2 *2
M10 threaded
*2
studs *3
*1
Round
cover *4
panel
Unistrut
ceiling Opening in Cover housing
ceiling
Corrugated hose holder
Bracket can be
installed in the
25 mm radius range
*1 145 mm min. height between lower surface of profile and ceiling (for corrugated hose holder and cable lead-in)
*2 Ceiling mount, tension per screw, less than 1500 N; screws + screw anchors from customer's contractor
*4 Cable run on ceiling. Cable inlet to be cut into the cover housing. Cable duct inner dimensions, min. 60 x 160 mm
The longitudinal rails already familiar from the DCS are used for the ceiling stand. For
detailed information about mounting of the longitudinal rails, see the DCS PG.
Console plate
*1
*1
CR1
CR1 Lead-in of cables coming from the
system is possible from both, left Cable inlet
or right sides (see arrows)
Cable feeding (from above) for monitors, ECC, Mouse and key-
board, for optional control modules and for Remote Power Archive
Unit (only if Image system in equipment room)
1:10
*1 Distances by fixation.
Height adjustable
1:20
Table opening
Work counter
Make the opening in the work counter to fit the configured control modules, and if
applicable, the ECC:
- Table control module (TCM)
- Ceiling stand control module (SCM)
- Collimator control module (CCM)
- ECC
The ECC can also be installed in a different location, independent of the control
modules. Unlike the control modules, the ECC is also suitable for free setup, i.e.,
it does not have to be recessed into the tabletop.
The data display unit can be positioned either on a console provided on-site, or it can be
attached to the wall.
Wall mounting 7
Fixation points *1
Carriage
Longitudinal rail
Surgery light
750 1- oder 3-strahlig
0
91
750
Ceiling column
RH - 700
Large pane
Cable holder
Corrugated hose
ø 30 (6 m up to the
ceiling) outlet)
1 : 20
e.g. Unistrut
profile
Mounting material
included in shipment
Cable outlet
Angled bracket
(shipped)
Mounting of angled
brackets to profiles
(on-site)
*1
*1
Cable duct
*1
1:50
• Versions:
- only table stand:
For space reasons, the control cable between the table stand-injector controller can
not be led through the table.
• Permissible injectors:
Injector Remarks
ANGIOMAT ILLUMENA - Table stand: With tilting or OR table, injector with air de-
tector required
MEDRAD MARK V PROVIS - with ODU connector
l1
NOTE The current data for allowed injectors are described in the
SIR PM AX 39....
- HIGH pulse width (min. ... max.) of the trigger signal: 1 msec ... 50 msec
• Connection of the ECG signal cable is made in the system control cabinet, SC1/3, on the
RTC.