Download as pdf or txt
Download as pdf or txt
You are on page 1of 694

QDR

Reference Manual

Caution: Federal (U.S.A.) law restricts this device to sale by or on the


order of a physician (or properly licensed practitioner).

Document No. MAN-00732 Revision 002


October 2007
The information contained in this Manual is confidential and proprietary to Hologic, Inc.
This information is provided only to authorized representatives of Hologic’s customers
solely for the purpose of facilitating the use of Hologic’s products. No information
contained herein may be disclosed to any unauthorized person for any purpose
whatsoever without the prior written consent of Hologic, Inc.
The information in this document is subject to change without notice.
©
Copyright 2000-2007 by Hologic, Inc. All rights reserved.
Discovery™, and Explorer™ are trademarks of Hologic, Inc.
QDR® and the Hologic logo are registered trademarks of Hologic, Inc.
HSA® is a registered trademark of The Johns Hopkins University Applied Physics
Laboratory.
Protected by at least some of the following U.S. Patents and foreign counterparts:
4,811,373; Re34,511; 5,040,199; 5,138,553; 5,165,410; 5,432,834; 5,483,960;
5,572,998; 5,657,369; 5,687,211; 5,715,820; 5,717,735; 5,748,705; 5,762,608;
5,771,272; 5,778,045; 5,835,555; 5,838,562; 5,838,765; 5,850,836; 5,891,033;
6,002,959; 6,009,147; 6,230,036; 6,233,473; 6,385,283.
Printed in the U.S.A.
Windows® and Windows® XP are either registered trademarks or trademarks of
Microsoft Corporation in the United States and/or other countries.
All trademarks, registered trademarks, and product names used within this document are
the property of their respective owners.

Hologic, Inc.
35 Crosby Drive
Bedford, MA 01730
USA
Tel: (781) 999-7300
Fax: (781) 280-0669
Service: (800) 321-HOLX (321-4659)
User Support: (800) 321-HOLX (321-4659

Hologic Europe
Hologic N.V.
Leuvensesteenweg 250A
1800 Vilvoorde
Belgium
Tel: 32.2.711.4680
Fax: 32.2.725.2087
Table of Contents
Chapter 1 -
Introducing the QDR Series System ...................................................................................... 1-1
Introducing the QDR Series Models .................................................................................... 1-2
Introducing the System Hardware Components .................................................................. 1-3
Operator Console ........................................................................................................... 1-4
Computer Hardware and Peripherals ....................................................................... 1-4
4500 Power Module ....................................................................................................... 1-4
Delphi AC Power Isolation Module .............................................................................. 1-4
Examination Table ......................................................................................................... 1-4
Table Pad ................................................................................................................. 1-4
C-Arm ............................................................................................................................ 1-5
X-Ray Source and Detectors .................................................................................... 1-5
Laser ......................................................................................................................... 1-5
Control Panel for QDR Systems .................................................................................... 1-6
The Discovery and Explorer Control Panels ........................................................... 1-6
Discovery/Explorer Control Panel Switches and Indicators ............................... 1-6
The QDR 4500 and Delphi Control Panels ............................................................ 1-7
The QDR 4500 Control Panels ........................................................................... 1-7
The Delphi Control Panels .................................................................................. 1-7
QDR 4500/Delphi Control Panel Switches and Indicators ................................. 1-8
Introducing the APEX Software .......................................................................................... 1-9
Main Window Overview ............................................................................................... 1-9
Main Window Features .................................................................................................. 1-9
Introducing the Quality Control Spine Phantom ............................................................... 1-10
Introducing the Positioning Aids ....................................................................................... 1-11
Knee Positioner ............................................................................................................ 1-11
Foot Restraint (Hip Positioner) .................................................................................... 1-11
Head Positioner for A and SL Models ......................................................................... 1-12
Introducing the Product Support Materials ........................................................................ 1-12

Chapter 2 -
Understanding QDR Series Technology ................................................................................ 2-1
Introducing Key Elements ................................................................................................... 2-2
Presenting DXA Technology ......................................................................................... 2-2
Accuracy and Precision ................................................................................................. 2-3
Principles of Operation .................................................................................................. 2-3
Introducing the Scan Sites ................................................................................................... 2-4
Lumbar Spine ................................................................................................................. 2-4
Proximal Femur ............................................................................................................. 2-5
Forearm .......................................................................................................................... 2-6
QDR Reference Manual

Whole Body ................................................................................................................... 2-7


Understanding the Scan Results .......................................................................................... 2-8
Z-Score ........................................................................................................................... 2-8
T-Score ........................................................................................................................... 2-8
Reference Database ........................................................................................................ 2-8

Chapter 3 -
APEX Software ........................................................................................................................ 3-1
Understanding the Mouse .................................................................................................... 3-2
Holding the Mouse ......................................................................................................... 3-2
Mouse Tasks .................................................................................................................. 3-2
Using the Mouse with APEX Software ............................................................................... 3-3
Pointing with the Mouse ................................................................................................ 3-3
Clicking the Mouse ........................................................................................................ 3-3
Double-Clicking the Mouse ........................................................................................... 3-3
Dragging and Dropping ................................................................................................. 3-3
Understanding the User Interface ........................................................................................ 3-4
User Interface Windows ................................................................................................ 3-4
Main Window .......................................................................................................... 3-4
Tab Windows ........................................................................................................... 3-5
Dialog Windows ...................................................................................................... 3-6
Understanding Main Window Components ........................................................................ 3-7
Main Window Buttons ................................................................................................... 3-7
Using Menus ............................................................................................................ 3-8
Choosing Menu Items .............................................................................................. 3-8
Using Sub-Menus .................................................................................................... 3-9
Closing Menus ......................................................................................................... 3-9
Understanding Other Window Components .................................................................. 3-9
Entering Text ........................................................................................................... 3-9
Tabs ........................................................................................................................ 3-10
Drop-down Lists .................................................................................................... 3-10
Radio Buttons ........................................................................................................ 3-10
Command Buttons ................................................................................................. 3-10
Check Boxes .......................................................................................................... 3-11
Scroll Bars .............................................................................................................. 3-11
Sorting List Items ................................................................................................... 3-11
Identifying the Sort Order ...................................................................................... 3-11
Changing the Sort Order ........................................................................................ 3-12
Selecting Items From a List ......................................................................................... 3-12
Selecting One Item ................................................................................................. 3-12
Selecting Multiple Items ........................................................................................ 3-12
Selecting a Range of Items .................................................................................... 3-12
Introducing the System Software Reminders .................................................................... 3-13

iv Table of Contents
QDR Reference Manual

Flashing Buttons .......................................................................................................... 3-13


Daily QC ................................................................................................................ 3-13
System Backup ...................................................................................................... 3-13
Introducing the System Software Help .............................................................................. 3-14
Tool Tips ...................................................................................................................... 3-14
Online Context Sensitive Help .................................................................................... 3-14
Ticker Tape .................................................................................................................. 3-15

Chapter 4 -
Performing Daily System Operations .................................................................................... 4-1
Discovery Systems ............................................................................................................... 4-2
Discovery Circuit Breaker and Indicator ....................................................................... 4-2
Discovery-C, -Ci, -W and -Wi Control Panel ................................................................ 4-2
Discovery-A and -SL Control Panel .............................................................................. 4-4
Turning Discovery Power On and Off ........................................................................... 4-5
Beginning of the Workday ....................................................................................... 4-5
End of the Workday ................................................................................................. 4-6
Discovery Emergency Procedures ................................................................................. 4-7
Operator Action During a Power Failure ................................................................. 4-7
Shutting Down During a Power Failure ............................................................. 4-7
After Power is Restored ...................................................................................... 4-7
If Equipment Fails While in Operation .................................................................... 4-7
If AC Power has been Turned Off ........................................................................... 4-8
Explorer Systems ................................................................................................................. 4-9
Explorer Circuit Breaker and Indicator ......................................................................... 4-9
Explorer Control Panel .................................................................................................. 4-9
Turning Explorer Power On and Off ........................................................................... 4-10
Beginning of the Workday ..................................................................................... 4-10
End of the Workday ............................................................................................... 4-11
Explorer Emergency Procedures ................................................................................. 4-11
Operator Action During a Power Failure ............................................................... 4-12
Shutting Down During a Power Failure ........................................................... 4-12
After Power is Restored .................................................................................... 4-12
If Equipment Fails While in Operation .................................................................. 4-12
If AC Power has been Turned Off ......................................................................... 4-13
QDR 4500 Systems ............................................................................................................ 4-14
QDR 4500 Power Module ........................................................................................... 4-14
QDR 4500 Power Module Side Panel ................................................................... 4-14
QDR 4500 Power Module Rear Panel ................................................................... 4-15
QDR 4500 Control Panel ............................................................................................. 4-15
QDR 4500C and QDR 4500W Control Panel ....................................................... 4-16
QDR 4500A and QDR 4500SL Control Panel ...................................................... 4-17
Turning QDR 4500 Power On and Off ........................................................................ 4-18
Beginning of the Workday ..................................................................................... 4-18

Table of Contents v
QDR Reference Manual

End of the Workday ............................................................................................... 4-19


QDR 4500 Emergency Procedures .............................................................................. 4-20
Operator Action During a Power Failure ............................................................... 4-20
Shutting down during a power failure .............................................................. 4-20
After power is restored ..................................................................................... 4-21
If QDR 4500 Equipment Fails While in Operation ............................................... 4-21
If QDR 4500 AC Power has been Turned Off ....................................................... 4-21
Delphi Systems .................................................................................................................. 4-23
Delphi AC Power Isolation Module ............................................................................ 4-23
Delphi Control Panels .................................................................................................. 4-23
Delphi-C and -W Control Panel ........................................................................ 4-23
Delphi-A and -SL Control Panel ........................................................................... 4-25
Turning Delphi Power On and Off .............................................................................. 4-26
Beginning of the Workday ..................................................................................... 4-27
End of the Workday ............................................................................................... 4-27
Delphi Emergency Procedures ..................................................................................... 4-28
Operator Action During a Power Failure ............................................................... 4-28
Shutting Down During a Power Failure ........................................................... 4-29
After Power is Restored .................................................................................... 4-29
If Delphi Equipment Fails While in Operation ...................................................... 4-29
If Delphi AC Power has been Turned Off ............................................................. 4-30

Chapter 5 -
Quality Control ........................................................................................................................ 5-1
The Daily QC Procedure ...................................................................................................... 5-2
Start Daily QC ............................................................................................................... 5-2
Position the Spine Phantom ........................................................................................... 5-3
Automatic System Test .................................................................................................. 5-4
Auto QC Passed ............................................................................................................. 5-5
Review Analysis ...................................................................................................... 5-6
QC Plot .................................................................................................................... 5-7
About the QC Plot ............................................................................................... 5-7
Auto QC Failed .............................................................................................................. 5-9
Details ...................................................................................................................... 5-9
Review Analysis ...................................................................................................... 5-9
Plot ......................................................................................................................... 5-10
OK .......................................................................................................................... 5-10
About the QC Spine Phantom ............................................................................................ 5-10
QC Data Management ....................................................................................................... 5-11
Access Data Management ............................................................................................ 5-11
Set QC Plot Parameters ................................................................................................ 5-11
Select scans to be included in the QC Plot .................................................................. 5-13
Exclude Data Points from the QC Plot Screen ...................................................... 5-15
Create a New Phantom Record .................................................................................... 5-15

vi Table of Contents
QDR Reference Manual

Set up the QC Plot ....................................................................................................... 5-17


Plot Regression .................................................................................................................. 5-20
Archiving QC Phantom Scans ........................................................................................... 5-20

Chapter 6 -
Managing Patient Records ...................................................................................................... 6-1
Introducing Patient Records ................................................................................................. 6-2
Accessing the Records ................................................................................................... 6-2
Working with Patient Records ............................................................................................. 6-4
Overview of Patient Record ........................................................................................... 6-4
Biography Tab ......................................................................................................... 6-4
Insurance Tab ........................................................................................................... 6-6
................................................................................................................................. 6-6
Retrieving a Patient Record ........................................................................................... 6-7
Searching the Patient Drawer .................................................................................. 6-7
Creating/Editing a Patient Record ................................................................................. 6-9
New Patient ............................................................................................................ 6-10
Updating a Patient Record ..................................................................................... 6-12
Patient Biography ............................................................................................. 6-13
Patient Questionnaire ........................................................................................ 6-14
Introducing Scan Records .................................................................................................. 6-16
Scan Drawer ................................................................................................................. 6-16
Working with Scan Records .............................................................................................. 6-17
Scan Details ................................................................................................................. 6-17
Details Tab ............................................................................................................. 6-18
Identification Tab ................................................................................................... 6-19
Deleting Scan Files ...................................................................................................... 6-21
Archived Scans Tab ............................................................................................... 6-22
UnArchived Scans Tab .......................................................................................... 6-22
Retrieving the Scan Records ........................................................................................ 6-22
Performing an Analysis ......................................................................................... 6-23
Adding Scan Information from Other Manufacturer’s Systems ........................................ 6-24

Chapter 7 -
The Patient Examination ......................................................................................................... 7-1
The Patient Examination ...................................................................................................... 7-2
Preparing the Patient ...................................................................................................... 7-2
Patient Interview ...................................................................................................... 7-2
Patient Questions ................................................................................................ 7-2
Patient Preparation ................................................................................................... 7-3
Creating/Retrieving a Patient Biography ....................................................................... 7-4
Creating a Patient Biography ................................................................................... 7-4

Table of Contents vii


QDR Reference Manual

Retrieving a Patient Biography ................................................................................ 7-4


Using the Select A Patient Tab ........................................................................... 7-5
Confirming the Patient Information ......................................................................... 7-6
Filling out the Patient Questionnaire ....................................................................... 7-7
Using the Worklist Tab ....................................................................................... 7-8
Using Modality Worklist with a Study .................................................................. 7-13
Modality Worklist Operations ............................................................................... 7-16
Adding a Patient ................................................................................................ 7-17
Adding a Study ................................................................................................. 7-18
Matching an Existing Patient ............................................................................ 7-18
Selecting the Scan Type and Mode .............................................................................. 7-21
Positioning the Patient and C-arm ............................................................................... 7-21
Placing the Patient on the Table ............................................................................ 7-21
Positioning for the Specific Examination .............................................................. 7-22
Performing the Examination ........................................................................................ 7-22
Exiting the Examination .............................................................................................. 7-23
Adding Scan Comments ........................................................................................ 7-24
Performing the Analysis .............................................................................................. 7-24
Contrast and Brightness ......................................................................................... 7-25
Exiting the Analysis ..................................................................................................... 7-27
Using a Study ............................................................................................................... 7-28
Generating and Printing Reports .................................................................................. 7-32
Configuring the Report .......................................................................................... 7-32
Editing Physician’s Comments .............................................................................. 7-33

Chapter 8 -
Performing and Analyzing the AP Lumbar Spine Scan ...................................................... 8-1
The AP Lumbar Spine Examination .................................................................................... 8-2
Preparing the Patient ...................................................................................................... 8-2
Creating/Retrieving a Patient Biography ....................................................................... 8-2
Selecting the Scan Type and Mode ................................................................................ 8-2
Choosing the Scan Mode ......................................................................................... 8-3
Confirming Scan Parameters ................................................................................... 8-3
Positioning the Patient and C-arm ........................................................................... 8-4
Placing the Patient on the Table .............................................................................. 8-5
Positioning the Patient ............................................................................................. 8-5
Positioning the C-arm .............................................................................................. 8-6
Performing the Examination .......................................................................................... 8-7
Starting the AP Lumbar Spine Scan ........................................................................ 8-8
Repositioning the Scan ............................................................................................ 8-9
Exiting the Examination .............................................................................................. 8-11
Performing the Analysis .............................................................................................. 8-11
Choosing the Method of Analysis ......................................................................... 8-11
Methods of Analysis ......................................................................................... 8-12
Performing the Analysis ........................................................................................ 8-13

viii Table of Contents


QDR Reference Manual

One-Time™ Auto Analysis ................................................................................... 8-14


Manual Analysis .................................................................................................... 8-15
Performing the Manual Analysis ...................................................................... 8-16
Adjusting the Image Display ............................................................................ 8-16
Defining the Region of Interest (ROI) .............................................................. 8-16
The Global ROI Toolbox .................................................................................. 8-18
Viewing/Editing the Bone Map ........................................................................ 8-19
Marking Intervertebral Spaces .......................................................................... 8-21
Labeling The Vertebral Bodies ......................................................................... 8-23
Including or Excluding Vertebrae from Analysis ............................................. 8-23
Viewing Results ................................................................................................ 8-25
Exiting the Analysis ..................................................................................................... 8-27
Generating and Printing Reports .................................................................................. 8-27
Alternate Scan Modes .................................................................................................. 8-27
Evaluating the Image ................................................................................................... 8-28
Scoliotic Spine Scans ......................................................................................................... 8-30
Adjusting the ROI Box ................................................................................................ 8-30
Marking Intervertebral Spaces of Scoliotic Spines ...................................................... 8-33
Pivoting an Intervertebral Line .............................................................................. 8-33

Chapter 9 -
Performing and Analyzing the Hip Scan ............................................................................... 9-1
The Hip Examination ........................................................................................................... 9-2
Preparing the Patient ...................................................................................................... 9-2
Creating/Retrieving a Patient Biography ....................................................................... 9-2
Selecting the Scan Type and Mode ................................................................................ 9-3
Choosing the Scan Mode ......................................................................................... 9-3
Left or Right Hip ................................................................................................. 9-3
Dual-Hip ............................................................................................................. 9-4
Confirming Scan Parameters ................................................................................... 9-5
Positioning the Patient and C-arm ................................................................................. 9-6
Placing the Patient on the Table .............................................................................. 9-7
Positioning the Patient ............................................................................................. 9-7
Positioning the C-arm .............................................................................................. 9-8
Positioning for the Left Hip or Right Hip Scan .................................................. 9-8
Positioning for the Dual-Hip Scan ...................................................................... 9-8
Performing the Examination .......................................................................................... 9-8
Starting the Hip Scan ............................................................................................... 9-9
Repositioning the Scan .......................................................................................... 9-10
Additional Steps for the Dual-Hip Scan ................................................................ 9-12
Exiting the Examination .............................................................................................. 9-13
Performing the Analysis .............................................................................................. 9-14
Choosing the Method of Analysis ......................................................................... 9-14
Methods of Analysis ......................................................................................... 9-15
Performing the Analysis ........................................................................................ 9-16

Table of Contents ix
QDR Reference Manual

One-Time™ Auto Analysis ................................................................................... 9-16


Manual Analysis .................................................................................................... 9-17
Performing the Manual Analysis ...................................................................... 9-17
Adjusting the Image Display ............................................................................ 9-18
Defining the Region of Interest (ROI) .............................................................. 9-18
The Global ROI Toolbox .................................................................................. 9-20
Viewing/Editing the Bone Map ........................................................................ 9-21
Positioning the Neck Box ................................................................................. 9-24
Adjusting the Neck Box .................................................................................... 9-26
Viewing Results ................................................................................................ 9-27
Exiting the Analysis ..................................................................................................... 9-29
For All Scans Except Dual-Hip ............................................................................. 9-29
For Dual-Hip Scans ............................................................................................... 9-29
Generating and Printing Reports .................................................................................. 9-29
Alternate Scan Modes .................................................................................................. 9-29
Evaluating the Image ................................................................................................... 9-30
Modifying Other Regions of the Hip ................................................................................. 9-33
About the Midline ........................................................................................................ 9-33
About Ward’s Triangle ................................................................................................ 9-35
About the Trochanter Region ...................................................................................... 9-37
Hip Structure Analysis® .................................................................................................... 9-38
HSA Toolbox ............................................................................................................... 9-39
Manually Adjusting the HSA Region Boxes .................................................... 9-39
Tips for Adjusting the HSA Region Boxes ...................................................... 9-40
Viewing Results ................................................................................................ 9-41

Chapter 10 -
Performing and Analyzing the Forearm Scan .................................................................... 10-1
The Forearm Examination ................................................................................................. 10-2
Prepare the Patient ....................................................................................................... 10-2
Creating/Retrieving a Patient Biography ..................................................................... 10-2
Measuring the Patient's Forearm .................................................................................. 10-2
Selecting the Scan Type and Mode .............................................................................. 10-4
Choosing the Scan Mode .................................................................................. 10-4
Confirming Scan Parameters ................................................................................. 10-4
Positioning the Patient and C-arm ............................................................................... 10-6
Placing the Forearm in the Scan Field ................................................................... 10-6
Positioning the C-arm ............................................................................................ 10-7
Performing the Examination ........................................................................................ 10-7
Starting the Forearm Scan ...................................................................................... 10-8
Repositioning the Scan .......................................................................................... 10-9
Exiting the Examination ............................................................................................ 10-11
Performing the Analysis ............................................................................................ 10-12
Choosing the Method of Analysis ....................................................................... 10-12

x Table of Contents
QDR Reference Manual

Methods of Analysis ....................................................................................... 10-13


Performing the Forearm Analysis ........................................................................ 10-14
Entering the Forearm Length .......................................................................... 10-15
Adjusting the Image Display .......................................................................... 10-15
Defining the Region of Interest (ROI) ............................................................ 10-16
The Global ROI Toolbox ................................................................................ 10-17
About the 1/3 Distal Region ........................................................................... 10-21
Viewing/Editing the Bone Map ...................................................................... 10-22
Adjusting the MID/UD Regions ..................................................................... 10-25
The MID/UD Toolbox .................................................................................... 10-26
Viewing Results .............................................................................................. 10-28
Exiting the Analysis ................................................................................................... 10-32
Generating and Printing Reports ................................................................................ 10-32
Evaluating the Image ................................................................................................. 10-32

Chapter 11 -
Performing and Analyzing the Whole Body Scan .............................................................. 11-1
The Whole Body Examination .......................................................................................... 11-2
Auto Whole Body ........................................................................................................ 11-2
Preparing the Patient .................................................................................................... 11-2
Creating/Retrieving a Patient Biography ..................................................................... 11-3
Selecting the Scan Type and Mode .............................................................................. 11-3
Choosing the Scan Mode ....................................................................................... 11-3
Confirming Scan Parameters ................................................................................. 11-4
Positioning the Patient and C-Arm .............................................................................. 11-4
Placing the Patient on the Table ............................................................................ 11-5
Positioning the Patient ........................................................................................... 11-5
Performing the Examination ........................................................................................ 11-6
Starting the Whole Body Scan ............................................................................... 11-6
Exiting the Examination .............................................................................................. 11-7
Performing the Analysis .............................................................................................. 11-7
Choosing the Method of Analysis ......................................................................... 11-8
Choosing the Compare to Previous Analysis ........................................................ 11-9
Performing the Whole Body Analysis ................................................................... 11-9
Adjusting the Image Display .......................................................................... 11-10
Defining the Regions ...................................................................................... 11-10
Regions Toolbox ............................................................................................. 11-11
Viewing Results ................................................................................................... 11-15
Exiting the Analysis ................................................................................................... 11-17
Generating and Printing Reports ................................................................................ 11-18
Evaluating the Image ................................................................................................. 11-18

Chapter 12 -
Performing and Analyzing the AP/Lateral Spine Scan ...................................................... 12-1
The AP/Lateral Spine Examination ................................................................................... 12-2

Table of Contents xi
QDR Reference Manual

Preparing the Patient .................................................................................................... 12-2


Creating/Retrieving a Patient Biography ..................................................................... 12-2
Placing the Patient on the Table .................................................................................. 12-2
Selecting the Scan Type and Mode .............................................................................. 12-3
Choosing the Scan Mode .................................................................................. 12-3
Positioning the Patient ................................................................................................. 12-4
Positioning the C-arm .................................................................................................. 12-5
Confirming Scan Parameters ................................................................................. 12-6
Performing the AP Scan .............................................................................................. 12-7
Starting the AP Scan .............................................................................................. 12-8
Repositioning the Scan .......................................................................................... 12-9
Analyzing the AP Scan ................................................................................................ 12-9
Performing the Analysis ........................................................................................ 12-9
Choosing the Method of Analysis ......................................................................... 12-9
Methods of Analysis ....................................................................................... 12-10
Performing the AP Analysis ...................................................................................... 12-11
One-Time™ Auto Analysis ................................................................................. 12-11
Manual Analysis .................................................................................................. 12-12
Performing the Manual Analysis .................................................................... 12-13
Adjusting the Image Display .......................................................................... 12-13
Defining the Region of Interest ...................................................................... 12-14
Viewing/Editing the Bone Map ...................................................................... 12-16
Marking Intervertebral Spaces ........................................................................ 12-18
Labeling The Vertebral Bodies ....................................................................... 12-20
Including or Excluding Vertebrae from Analysis ........................................... 12-20
Viewing Results .............................................................................................. 12-21
Verifying the Start Position ............................................................................ 12-22
Performing the Lateral Scan ...................................................................................... 12-25
Rotating the C-Arm for the Lateral Scan ............................................................. 12-25
Table Safety Feature ....................................................................................... 12-26
Confirming Scan Parameters ............................................................................... 12-26
Starting the Lateral Scan ...................................................................................... 12-26
Removing the Patient From the Table ................................................................. 12-28
Exiting the Examination ............................................................................................ 12-28
Analyzing the Lateral Scan ........................................................................................ 12-28
Choosing the Method of Analysis ....................................................................... 12-29
Performing the Lateral Analysis .......................................................................... 12-31
Adjusting the Image Display .......................................................................... 12-32
Defining the Region of Interest (ROI) ............................................................ 12-32
The Global ROI Toolbox ................................................................................ 12-33
Reanalysis Warning ........................................................................................ 12-33
Adjusting the Vertebral Boundaries ............................................................... 12-34
Adjusting Vertebral Bodies ............................................................................ 12-35
Confirming Lumbar Vertebrae Labels ............................................................ 12-36
Viewing/Editing the Bone Map ...................................................................... 12-37
Adjusting the Mid Regions ............................................................................. 12-38

xii Table of Contents


QDR Reference Manual

Viewing Results .............................................................................................. 12-39


Exiting the Analysis ................................................................................................... 12-42
Generating and Printing Reports ................................................................................ 12-42
Alternate Scan Modes ................................................................................................ 12-42

Chapter 13 -
Instant Vertebral Assessment and Quantitative Morphometry ........................................ 13-1
Introduction ........................................................................................................................ 13-2
Specification ................................................................................................................ 13-2
Patient Position and Scanning ...................................................................................... 13-2
Patient Positioning on A or SL System ................................................................. 13-2
Patient Positioning on C, Ci, W or Wi System ...................................................... 13-3
The IVA Examinations ...................................................................................................... 13-4
IVA Imaging Exam ...................................................................................................... 13-4
IVA with BMD Exam .................................................................................................. 13-4
IVA Imaging on an A or SL System .................................................................................. 13-4
Preparing the Patient .................................................................................................... 13-5
Creating/Retrieving a Patient Biography ..................................................................... 13-5
Starting the Examination ............................................................................................. 13-5
Selecting the Scan Type and Mode .............................................................................. 13-6
The AP IVA Scan ........................................................................................................ 13-9
Confirming Scan Parameters for AP IVA Scan ................................................... 13-10
Positioning the Patient for the AP IVA Scan ....................................................... 13-10
Performing the AP IVA Scan .............................................................................. 13-11
Repositioning the Scan ................................................................................... 13-12
Using the IVA Image Viewer for AP IVA Scan ................................................. 13-14
Exiting the IVA Image Viewer ............................................................................ 13-14
If Only AP IVA Scan Selected ............................................................................ 13-14
If Both AP and Lateral IVA Scans Selected ................................................... 13-14
The Lateral IVA Scan ................................................................................................ 13-15
Confirming Scan Parameters for Lateral IVA Scan ............................................ 13-15
Performing the Lateral IVA Scan ........................................................................ 13-16
Repositioning the Scan ................................................................................... 13-17
Using the IVA Image Viewer for Lateral IVA Scan ........................................... 13-19
Returning the C-arm to the AP Position .............................................................. 13-20
IVA Imaging on a C, Ci, W or Wi System ...................................................................... 13-21
Preparing the Patient .................................................................................................. 13-21
Creating/Retrieving a Patient Biography ................................................................... 13-21
Starting the Examination ........................................................................................... 13-21
Selecting the Scan Type and Mode ............................................................................ 13-22
The AP IVA Scan ...................................................................................................... 13-24
Confirming Scan Parameters for AP IVA Scan ................................................... 13-25
Positioning the Patient for the AP IVA Scan .................................................. 13-25
Performing the AP IVA Scan .............................................................................. 13-26

Table of Contents xiii


QDR Reference Manual

Repositioning the Scan ................................................................................... 13-27


Using the IVA Image Viewer for AP IVA Scan .................................................. 13-29
Exiting the IVA Image Viewer ............................................................................ 13-29
The Lateral IVA Scan ................................................................................................ 13-30
Confirming Scan Parameters for Lateral IVA Scan ............................................ 13-30
Positioning the Patient for the Lateral IVA Scan ................................................ 13-31
Placing the Patient on the Table .......................................................................... 13-31
Positioning the Patient .................................................................................... 13-31
Performing the Lateral IVA Scan ........................................................................ 13-33
Repositioning the Scan ................................................................................... 13-34
........................................................................................................................ 13-34
Using the IVA Image Viewer for Lateral IVA Scan ........................................... 13-36
IVA with BMD ................................................................................................................ 13-38
Preparing the Patient .................................................................................................. 13-38
Creating/Retrieving a Patient Biography ................................................................... 13-38
Starting the Examination ........................................................................................... 13-39
Selecting the Scan Type and Mode ............................................................................ 13-40
The AP IVA Scan ...................................................................................................... 13-43
Confirming Scan Parameters for AP IVA Scan ................................................... 13-44
Performing the AP IVA Scan .............................................................................. 13-44
Repositioning .................................................................................................. 13-46
Using the IVA Image Viewer for AP IVA Scan .................................................. 13-46
The AP BMD Scan .................................................................................................... 13-46
Confirming Scan Parameters for AP BMD Scan ................................................. 13-47
Performing the AP BMD Scan ............................................................................ 13-47
Analyzing the AP BMD Scan .............................................................................. 13-47
Exiting the AP BMD Analysis ............................................................................. 13-47
The Lateral BMD Scan .............................................................................................. 13-48
Confirming Scan Parameters for Lateral BMD Scan .......................................... 13-48
Performing the Lateral BMD Scan ...................................................................... 13-49
Exiting the Lateral BMD Scan ............................................................................. 13-50
The Lateral IVA Scan ................................................................................................ 13-51
Confirming Scan Parameters for Lateral IVA Scan ............................................ 13-51
Performing the Lateral IVA Scan ........................................................................ 13-52
Using the IVA Image Viewer for the Lateral IVA Scan ..................................... 13-53
Returning the C-arm to the AP Position .............................................................. 13-54
IVA Image Viewer ........................................................................................................... 13-55
The IVA Image Viewer ............................................................................................. 13-55
The Image ............................................................................................................ 13-55
The Patient Data ................................................................................................... 13-56
The Control Bar ................................................................................................... 13-56
Mouse Pointer Controls ....................................................................................... 13-58
DE Scan ..................................................................................................................... 13-59
Using Markers .................................................................................................................. 13-61

xiv Table of Contents


QDR Reference Manual

The Object of QM Markers ....................................................................................... 13-61


Using QM Markers .................................................................................................... 13-61
Flipping the Image ............................................................................................... 13-62
Adding QM Markers ............................................................................................ 13-63
Selecting Markers ................................................................................................ 13-64
Moving Markers .................................................................................................. 13-64
Editing Markers ................................................................................................... 13-64
Labeling Markers Sets ......................................................................................... 13-65
Deleting Marker Sets ........................................................................................... 13-66
QM Result Block ................................................................................................. 13-66
QM Results Block for Lateral IVA Scans ...................................................... 13-66
QM Results Block for AP IVA Scans ............................................................ 13-68
Generating and Printing Images and Reports .................................................................. 13-69
Printing IVA Images from the Reports Button .......................................................... 13-69
Sample Printouts ........................................................................................................ 13-72
Sample Image ...................................................................................................... 13-72
Sample Report ...................................................................................................... 13-73
Interpreting the IVA Image .............................................................................................. 13-74

Chapter 14 -
Compare and Follow-up ........................................................................................................ 14-1
Introducing Compare ......................................................................................................... 14-2
Steps ............................................................................................................................. 14-2
Restoring the Patient’s Baseline Scan ............................................................................... 14-3
Restoring a Baseline Scan ............................................................................................ 14-3
Evaluating the Patient’s Baseline Scan .............................................................................. 14-4
Evaluating Baseline Scan Patient Positioning ............................................................. 14-4
Reviewing Operator or Physician Comments ........................................................ 14-4
Baseline Positioning Acceptable ........................................................................... 14-5
Baseline Positioning Unacceptable ........................................................................ 14-5
Scanning Once to Duplicate Incorrect Baseline Positioning ............................ 14-5
Scanning Twice ................................................................................................. 14-5
Scanning Once with Correct Positioning .......................................................... 14-5
Evaluating Baseline Scan Analysis .............................................................................. 14-6
Performing the Follow-up Scan ......................................................................................... 14-7
Updating the Patient’s Biography Record ................................................................... 14-7
Performing the Follow-Up Scan .................................................................................. 14-7
Analyzing the Follow-up Scan .......................................................................................... 14-8
Generating a Rate of Change Report ............................................................................... 14-12
Rate of Change Report for Unlike Scan Modes ........................................................ 14-13
Rate of Change Report for Spine Scans with Different Regions ............................... 14-14

Table of Contents xv
QDR Reference Manual

Chapter 15 -
Reports .................................................................................................................................... 15-1
Introducing Reports ........................................................................................................... 15-2
Report Types ................................................................................................................ 15-2
Filing Report .......................................................................................................... 15-2
Interpreting Report ................................................................................................. 15-2
Referring Report .................................................................................................... 15-2
Rate of Change Report ........................................................................................... 15-3
Report Information Blocks .......................................................................................... 15-3
Header .................................................................................................................... 15-4
Patient Information ................................................................................................ 15-4
Scan Information .................................................................................................... 15-4
Image ..................................................................................................................... 15-5
DXA Results .......................................................................................................... 15-5
Reference Graph ............................................................................................... 15-5
Rate of Change Graph ....................................................................................... 15-5
Physician’s Comment ............................................................................................ 15-5
Printing Reports ................................................................................................................. 15-6
Printing a Report After Analysis ................................................................................. 15-6
Printing a Report Using the Report Button .................................................................. 15-6
Printing the Patient Questionnaire ......................................................................... 15-8
Editing the Physician Comment ............................................................................ 15-9
Print Dialog Box ........................................................................................................ 15-10
Previewing Reports .......................................................................................................... 15-11
Report Examples .............................................................................................................. 15-13
Filing Report .............................................................................................................. 15-13
Interpreting Report ..................................................................................................... 15-14
Referring Report ........................................................................................................ 15-15
Rate of Change Report ............................................................................................... 15-15
Patient Questionnaire ................................................................................................. 15-17
DICOM Reports ............................................................................................................... 15-18
Choosing the Report Type ......................................................................................... 15-19
Viewing a Scan’s Details ........................................................................................... 15-22
Entering the Accession Number and User Definable Entries .................................... 15-24
Previewing a DICOM Report .................................................................................... 15-25
Printing a DICOM Report .......................................................................................... 15-26
Saving a DICOM Report ........................................................................................... 15-27
Sending a DICOM Report ......................................................................................... 15-29
Sorting the Scan List .................................................................................................. 15-30
Selecting/Deselecting All Scans ................................................................................ 15-31
Viewing the Queue .................................................................................................... 15-32
Viewing a History of Sent Reports ............................................................................ 15-32
Refreshing DICOM Reports in the Queue ................................................................. 15-33

xvi Table of Contents


QDR Reference Manual

Deleting a DICOM Report from the Queue ............................................................... 15-34


Viewing a History of Sent Storage Commitment Reports ......................................... 15-36
Printing the HxReport ................................................................................................ 15-37
Closing DICOM ............................................................................................................... 15-38
Alternate Navigation ........................................................................................................ 15-39
Drag and Drop Records ............................................................................................. 15-39
Drag and Drop Patient Records ........................................................................... 15-39
Drag and Drop Scan Records ............................................................................... 15-40

Chapter 16 -
Archiving, Locating and Restoring Scans ........................................................................... 16-1
Overview ............................................................................................................................ 16-2
Handling Read/Write CDs ..................................................................................... 16-2
Formatting a Read/Write CD ................................................................................. 16-2
Removing a Read/Write CD .................................................................................. 16-3
Archiving Scans ........................................................................................................... 16-3
Locations ................................................................................................................ 16-4
Archive Labels ....................................................................................................... 16-5
Locating Scans ............................................................................................................. 16-5
Restoring Scans ............................................................................................................ 16-6
Copying Scans ............................................................................................................. 16-6
About Archiving and Restoring Scans ......................................................................... 16-6
How Scans are Transferred .................................................................................... 16-6
Transfer Progress Message .................................................................................... 16-7
Canceling a Transfer .............................................................................................. 16-7
Errors During Transfer ........................................................................................... 16-7
Notification of Successful Transfers ...................................................................... 16-7
How the System Handles Data Inconsistencies ..................................................... 16-7
How the System Handles Archive Data ................................................................ 16-7
Archiving Scans ................................................................................................................. 16-8
Archive Location ......................................................................................................... 16-8
Archive Process ........................................................................................................... 16-8
Locating Scans ................................................................................................................. 16-12
Restoring Scans ................................................................................................................ 16-16
Copying Scans ................................................................................................................. 16-19
Query/Retrieve Scans ...................................................................................................... 16-22
Query Parameters ....................................................................................................... 16-23
Study Level Filters ..................................................................................................... 16-24
Retrieve Queue .......................................................................................................... 16-25
Retrieve Queue Log ............................................................................................. 16-26
Managing Archive Locations ........................................................................................... 16-27
Troubleshooting Transfer Errors ..................................................................................... 16-28

Table of Contents xvii


QDR Reference Manual

Problems Accessing an Archive Location ................................................................. 16-28


Transferring Scans ..................................................................................................... 16-28
Archive Locations ...................................................................................................... 16-29
Advanced User Information ....................................................................................... 16-30

Chapter 17 -
System Backup and Recovery ............................................................................................... 17-1
Overview ............................................................................................................................ 17-2
Files Saved ................................................................................................................... 17-2
Media Formatting .................................................................................................. 17-3
Management of Media ................................................................................................. 17-3
Floppy Diskette ...................................................................................................... 17-3
SuperDisk™ ........................................................................................................... 17-3
Read/write CD ....................................................................................................... 17-3
Network Drive ....................................................................................................... 17-3
When to Run System Backup ...................................................................................... 17-3
Naming the Backup File .............................................................................................. 17-4
Running the System Backup .............................................................................................. 17-5
Errors and Exceptions ........................................................................................................ 17-6
Backup file with the same name .................................................................................. 17-6
Drive is not accessible ................................................................................................. 17-7
Drive does not have enough space ............................................................................... 17-7
Drive is not accessible ................................................................................................. 17-8
System Recovery ............................................................................................................... 17-8

Chapter 18 -
Configuring the System ......................................................................................................... 18-1
Introduction to System Configuration ............................................................................... 18-2
System Tab ........................................................................................................................ 18-3
Date Format ................................................................................................................. 18-3
Patient Biography ........................................................................................................ 18-3
Autopositioning for Hip Scans .................................................................................... 18-3
Report Dose Area Product ........................................................................................... 18-3
Language ...................................................................................................................... 18-4
Customer ...................................................................................................................... 18-4
Default Ethnicity .......................................................................................................... 18-4
Model, S/N, System ID ................................................................................................ 18-4
Data Directory .............................................................................................................. 18-4
Software ....................................................................................................................... 18-4
Last System Backup Date ............................................................................................ 18-4
Last Service Date ......................................................................................................... 18-4
Analyze Tab ....................................................................................................................... 18-5
Default ......................................................................................................................... 18-5

xviii Table of Contents


QDR Reference Manual

General Options ........................................................................................................... 18-6


Display Region of Interest Coordinates ................................................................. 18-6
Enable Body Composition Analysis ...................................................................... 18-6
Image Display Options ................................................................................................ 18-6
Initially display unanalyzed Lateral Image as Single Energy ................................ 18-6
Initially display unanalyzed Hip image as Single Energy ..................................... 18-6
Display pseudo color image ................................................................................... 18-6
Image Pro ............................................................................................................... 18-6
Display Hip Axis Length ....................................................................................... 18-6
Lateral Options ............................................................................................................ 18-7
Display MID regions ............................................................................................. 18-7
Prevent vertical motion of ROI .............................................................................. 18-7
Analysis Option ........................................................................................................... 18-7
Disable Auto Low Density .................................................................................... 18-7
Disable Auto Analysis ........................................................................................... 18-7
Disable Auto Whole Body ..................................................................................... 18-7
Enable NHANES BCA .......................................................................................... 18-7
Image Style .................................................................................................................. 18-7
CAD BMD Tab ................................................................................................................. 18-8
Report Tab ....................................................................................................................... 18-10
Report - General Tab ................................................................................................. 18-10
Report - WHO Criteria Tab ....................................................................................... 18-12
Report - Least Significant Change Tab ...................................................................... 18-12
Configuring Reports .................................................................................................. 18-14
Configuring the Reports ....................................................................................... 18-14
Configure the Patient, Scan and Results Blocks ............................................. 18-14
Configure the Optional Reports ...................................................................... 18-16
Configure the Common Block ........................................................................ 18-19
Details of the Patient Block ................................................................................. 18-23
Details of the Scan Block ..................................................................................... 18-24
Details of the Results Block ................................................................................. 18-25
Details of the Reference Block for Configure Common Block Option ............... 18-26
Reference Block Hip Configuration ............................................................... 18-27
Reference Block Forearm Configuration ........................................................ 18-28
Reference Block Lateral Configuration .......................................................... 18-29
Configuring the Ancillary Report ........................................................................ 18-30
Archive Tab ..................................................................................................................... 18-31
Adding a New Location ............................................................................................. 18-31
Changing the Default Location .................................................................................. 18-32
Deleting a Location from the Archive List ................................................................ 18-32
Including Raw Data Files .......................................................................................... 18-33
Automatically Deleting Scans After Archiving ......................................................... 18-33
HL7 Tab ........................................................................................................................... 18-34
Entering HL7 Fields ................................................................................................... 18-34

Table of Contents xix


QDR Reference Manual

Changing HL7 Fields ................................................................................................. 18-36


Removing HL7 Fields ................................................................................................ 18-37
Scan Conversion .............................................................................................................. 18-39
History ............................................................................................................................. 18-41
Default Questionnaire Template ................................................................................ 18-41
Location of Completed Questionnaire ....................................................................... 18-42
Port ............................................................................................................................. 18-42
Print Patient Name and ID ......................................................................................... 18-43
DICOM Tab ..................................................................................................................... 18-44
Modality Worklist ...................................................................................................... 18-44
Configuring Modality Worklist ........................................................................... 18-45
Configure Worklist Provider .......................................................................... 18-49
Map Worklist Fields ....................................................................................... 18-51
DICOM Send Destinations ....................................................................................... 18-53
Add / Edit DICOM Send Destination Dialog ...................................................... 18-55
Configure DICOM Send ..................................................................................... 18-57
DICOM Storage Commitment Destinations ............................................................. 18-59
Add / Edit Storage Commitment Destination Dialog .......................................... 18-61
Configure Storage Commitment ......................................................................... 18-62
DICOM Query/Retrieve Destinations ....................................................................... 18-63
Add / Edit DICOM Query/Retrieve Destination Dialog ..................................... 18-65
Host ............................................................................................................................ 18-66
Privacy Tools ................................................................................................................... 18-67
Study Tab ......................................................................................................................... 18-68
Adding a Study .......................................................................................................... 18-68
Editing an Existing Study .......................................................................................... 18-72
Adding Items to an Existing Study ...................................................................... 18-72
Deleting Items from an Existing Study ................................................................ 18-74
Deleting a Study ......................................................................................................... 18-77
Renaming a Study ...................................................................................................... 18-78
Locating a Study from a Long List ............................................................................ 18-81

Chapter 19 -
Utilities .................................................................................................................................... 19-1
Starting a Utility ................................................................................................................ 19-2
System Configuration ........................................................................................................ 19-2
Usage ................................................................................................................................. 19-3
Usage Program Menu .................................................................................................. 19-4
Usage Summary Report ............................................................................................... 19-4
Technically Unacceptable Studies ......................................................................... 19-6
Usage Detail Report ............................................................................................... 19-6
Claims for Technically Unacceptable Studies ....................................................... 19-8

xx Table of Contents
QDR Reference Manual

Database Tools ................................................................................................................... 19-9


Patient Management .................................................................................................... 19-9
Starting Patient Management ................................................................................. 19-9
Managing Patients ................................................................................................ 19-10
Managing Patient Information ............................................................................. 19-12
Biography ........................................................................................................ 19-12
Insurance ......................................................................................................... 19-14
QC Phantom Details ............................................................................................ 19-15
Managing Patient Baseline Scans ........................................................................ 19-17
Removing a Patient From the System .................................................................. 19-18
Managing Scans ................................................................................................... 19-18
Moving Scans ................................................................................................. 19-21
Erasing Scans ................................................................................................. 19-23
Export/Import ............................................................................................................. 19-25
Types of Databases .............................................................................................. 19-26
Types of Data ....................................................................................................... 19-26
Types of Transfers ............................................................................................... 19-26
Export ......................................................................................................................... 19-27
Import ......................................................................................................................... 19-29
Reconcile ................................................................................................................... 19-31
Patient Callback ......................................................................................................... 19-33
Auto Baseline Utility ....................................................................................................... 19-36
Scan File Look ................................................................................................................. 19-36
Scan File Plot ................................................................................................................... 19-39
Emergency Motion .......................................................................................................... 19-40
AP Reposition .................................................................................................................. 19-40
Reference Curve .............................................................................................................. 19-42
Starting Reference Curve Editor ................................................................................ 19-43
Viewing Reference Curve Data ................................................................................. 19-44
Creating New Reference Curve Records ................................................................... 19-44
Copying a Reference Curve Record .......................................................................... 19-49
Editing Reference Curve Records .............................................................................. 19-51
Deleting Reference Curve Records ............................................................................ 19-52
Adding New Ethnic Groups ....................................................................................... 19-53
Restoring Reference Curve Database ........................................................................ 19-54
Rebuild Archive Index ..................................................................................................... 19-55
Install Options .................................................................................................................. 19-57

Chapter 20 -
Digital Imaging and Communications in Medicine (DICOM) .......................................... 20-1
DICOM Selections ............................................................................................................. 20-2
Configuring DICOM .......................................................................................................... 20-4

Table of Contents xxi


QDR Reference Manual

Using DICOM .................................................................................................................... 20-4


Troubleshooting ................................................................................................................. 20-5

Appendix A -
Options ..................................................................................................................................... A-1
Options List ......................................................................................................................... A-2
Forearm ......................................................................................................................... A-2
IVA ............................................................................................................................... A-3
IVA HD ......................................................................................................................... A-3
DICOM Store ................................................................................................................ A-3
DICOM Modality Worklist .......................................................................................... A-3
General Region of Interest ............................................................................................ A-3
Body Composition Analysis ......................................................................................... A-3
Decubitus Lateral .......................................................................................................... A-3
Small Animal ................................................................................................................ A-3
Prosthetic Hip ............................................................................................................... A-4
Enterprise Data Management ........................................................................................ A-4
Hip Axis Length ............................................................................................................ A-4
Infant Whole Body ........................................................................................................ A-4
Low Density Spine ........................................................................................................ A-4
International .................................................................................................................. A-4
DAP Meter .................................................................................................................... A-4
Hip Structure Analysis® (HSA) ................................................................................... A-4
Mobility ........................................................................................................................ A-4

Appendix B -
Operator Maintenance ............................................................................................................ B-1
Periodic Cleaning ................................................................................................................. B-2
Cleaning the QDR and Computer Components ............................................................. B-2
Cleaning the Table Pad .................................................................................................. B-2
Disinfecting the Table Pad ............................................................................................. B-2
Periodic Visual Inspection ................................................................................................... B-3
Checking for Adequate Clearance ................................................................................. B-3
Checking Cables and Outlets ......................................................................................... B-3
Inspecting the Table Pad ................................................................................................ B-3
Cleaning Accidental Spills ................................................................................................... B-4

Glossary

Index

xxii Table of Contents


Chapter 1
Introducing the QDR Series
System
This chapter introduces the operator to the QDR Series bone
densitometers. This series includes fifteen models of multiple detector,
fan-beam, Dual-energy X-ray Absorptiometry (DXA) bone
densitometers. Each model offers some unique scanning abilities that
meet the diagnostic needs of a particular clinical environment.

Content
Subject Page
Introducing the QDR Series Models 1-2
Introducing the System Hardware Components 1-3
Introducing the APEX Software 1-9
Introducing the Quality Control Spine Phantom 1-10
Introducing the Positioning Aids 1-11
Introducing the Product Support Materials 1-12

1-1
QDR Reference Manual

Introducing the QDR Series Models


Discovery-A
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, supine lateral spine, hip, dual hip, forearm,
whole body, AP and lateral imaging (single and dual energy) and
visualization of abdominal aortic calcification.
Discovery-SL
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, supine lateral spine, hip, dual hip, forearm,
AP and lateral imaging (single and dual energy) and visualization
of abdominal aortic calcification.
Discovery-W
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip, dual hip, forearm, whole body, AP and
lateral imaging (single and dual energy) and visualization of
abdominal aortic calcification.
Discovery-Wi
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip, dual hip, forearm and whole body.
Discovery-C
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip, dual hip, forearm, AP and lateral
imaging (single and dual energy) and visualization of abdominal
aortic calcification.
Discovery-Ci
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip, dual hip and forearm.
Explorer
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip and dual hip.
4500A
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, supine lateral spine, hip, dual hip, forearm
and whole body.
4500SL
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, supine lateral spine, hip, dual hip, and
forearm.

1-2 Introducing the QDR Series System


QDR Reference Manual

4500W
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip, dual hip, forearm, and whole body.
4500C
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip, dual hip, and forearm.
Delphi-A
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, supine lateral spine, hip, dual hip, forearm,
whole body, AP and lateral imaging (single and dual energy) and
visualization of abdominal aortic calcification.
Delphi-SL
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, supine lateral spine, hip, dual hip, forearm,
AP and lateral imaging (single and dual energy) and visualization
of abdominal aortic calcification.
Delphi-W
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip, dual hip, forearm, whole body, AP and
lateral imaging (single and dual energy) and visualization of
abdominal aortic calcification.
Delphi-C
This model provides for the acquisition and analysis of scans of
the AP lumbar spine, hip, dual hip, forearm, AP and lateral
imaging (single and dual energy) and visualization of abdominal
aortic calcification.

Introducing the System Hardware Components


The QDR Series provides DXA technology in a modular design
that includes the Operator Console and Examination Table.
Figure 1-1 Printer
QDR System C-Arm
Components Monitor
Rear
Table Pad Safety
Note: Figure depicts Discovery-A. Stop
Keyboard

Computer

Front
Safety
Stop
Control Panel KP1068A-0101

Operator’s Console Examination Table

Introducing the QDR Series System 1-3


QDR Reference Manual

Operator Console
The Operator Console acts as the system controller and contains
the Computer Hardware and Peripherals.

Computer Hardware and Peripherals


The computer system consists of an IBM compatible PC, monitor,
keyboard and printer.

4500 Power Module


The Power Module is used on QDR 4500 systems only and is
located in the lower back section of the Operator Console. The
Power Module contains switches on the rear and side panels for
turning the computer system and instrument power on and off. It
also includes a key for enabling generation of X-rays.

Delphi AC Power Isolation Module


The AC Power Isolation Module is used on the Delphi System
only. It is a stand-alone unit that receives AC power from the
facility and isolates it using a transformer. This protects the
system from unwanted noise and spikes.
The only control on the AC Power Isolation Module is the circuit
breaker located on the front. It provides overload protection for
the system. The only indicator is a green lamp that shows when
the circuit breaker is on and power is applied to the module.

Examination Table
The Examination Table is the part of the system where the scans
are performed and the data is sent to the computer. The major
components are described below.

Table Pad
The Table pad provides a cushion for the patient during the exam
with indicators to help align the patient during the exam.

1-4 Introducing the QDR Series System


QDR Reference Manual

Figure 1-2
Table Pad (typical)

KP1068A-0102

Table 1-1. Item Description Function


Table Pad Guides
1 Scan Limit Indicates the maximum scan range of
border the unit. Patient positioning must be
Note: The Table Pad markings within this border.
vary with the different models.
2 Center Lines Used to center the patient on the table.

C-Arm
Note: Explorer C-arm also contains The C-arm and table drive motors provide the table and arm
its Control Panel. motion required to scan the X-ray beam along the patient’s body.
The C-arm contains the X-ray source, X-ray detectors and laser.

X-Ray Source and Detectors


The X-rays are generated underneath the table and projected
upward. At the top of the C-arm are detectors that convert the X-
rays into electronic data. This data is then sent to the computer
where it is converted to scan images.

Laser
The laser projects a red visual cross-hair indicator onto the patient
during the patient positioning procedure. The cross-hair indicator
shows the starting position of the X-ray beam.

Introducing the QDR Series System 1-5


QDR Reference Manual

Control Panel for QDR Systems


Note: Refer to Chapter 4 for a The control panel for the QDR Systems contain a set of switches
detailed explanation of the and indicators that control the movement of the table and C-Arm.
functions of the control panel.
This allows the operator to help the patient on and off the table,
and permits accurate positioning of the laser.
Also, an Emergency Stop button allows the operator to halt the
scan at any point that the operator feels that the patient’s safety is
in jeopardy.

The Discovery and Explorer Control Panels


The Discovery control panel is located on the right front of the
examination table. The C, Ci, W and Wi control panel is shown in
Figure 1-3 and the A and SL control panel is shown in Figure 1-4.
The Explorer control panel is located on the C-Arm and is shown
in Figure 1-5.
Figure 1-3
Discovery C/Ci/W/Wi
Control Panel

KP1068A-0402

Figure 1-4
Discovery A/SL
Control Panel

KP1068A-0403

Figure 1-5
Explorer
Control Panel

00158-001

Discovery/Explorer Control Panel Switches and Indicators


The Control Panel contains the following switches and indicators:
• C-Arm Controls: Arm (left) and Arm (right) switches
• Table Controls: Table (in) and Table (out) switches
• Laser switch - light blinks when laser is on

1-6 Introducing the QDR Series System


QDR Reference Manual

• Center Table switch (also centers the C-arm)


• Patient On/Off switch
• X-ray On indicator
• Emergency Stop button
• Power switch (all models except Explorer)
• Enable Lateral switch (Discovery-A and -SL models only)
• Table Lift switch (Discovery-A and -SL models only)

The QDR 4500 and Delphi Control Panels


Note: Refer to Chapter 4 for a The control panel allows the operator to control the movement of
detailed explanation of the the C-arm and table. This helps the patient to easily get on or off
functions of the control panel.
the table, and permits the operator to accurately position the
laser.
Also, an Emergency Stop button allows the operator to halt the
scan at any point that the operator feels that the patient’s safety is
in jeopardy.

The QDR 4500 Control Panels


The QDR 4500 Control Panels are located on the right front of the
examination table unit.
Figure 1-6
Table
QDR 4500 C and W Arm Laser
Center
Table
Patient
On/Off Stop
Control Panel

X-Ray

KP0886-0105

Figure 1-7 Table


Emergency
QDR 4500 A and SL Arm Laser
Enable
Lateral
Center
Table
Patient
On/Off Table Stop
Control Panel

X-Ray

KP0886-0106

The Delphi Control Panels


The Location of the control panel for the Delphi-C and -W is on
the front of the C-arm, the location of the control panel for the
Delphi-A and -SL is on the right front of the examination table.
The Delphi Control Panels contains a series of switches, buttons,
and indicator lights. The switches control the movements and the

Introducing the QDR Series System 1-7


QDR Reference Manual

indicator lights inform the operator these activities are in


progress.
Figure 1-8 Emergency
Delphi-C and -W Stop Button
Control Panel

KP0886-0107

Figure 1-9
Delphi-A and -SL Patient
On/Off
X-Ray EMERGENCY
Power
Control Panel Table Table Lift Stop

A A
R R Center
M M Table

Table

Enable R

Laser
Lateral

KP0886-0108

QDR 4500/Delphi Control Panel Switches and Indicators


The Control Panels contain the following switches and indicators:
This control panel includes:
• C-arm Left and Right switches
• Table In and Out switches
• Laser switch On/Off (light blinks when laser is on)
• Center Table switch (also centers the C-arm)
• Patient On/Off switch
• X-ray On indicator
• Emergency Stop button
• Laser On/Motion Enabled indicator (light blinks when
laser is on, Delphi-C and -W systems only)
• Instrument On/Off (Power) switch (Delphi systems only)
• Enable Lateral switch (4500A and SL, and Delphi-A and -
SL only)
• Emergency Table (Table Lift) switch (4500A and SL, and
Delphi-A and -SL only)

1-8 Introducing the QDR Series System


QDR Reference Manual

Introducing the APEX Software


Note: Refer to Chapter 3 for The QDR Series uses the Windows® XP operating system. The
detailed information about using
APEX software.
computer screens use a graphical user interface (GUI) that are
called windows. The primary interface consists of a main window
with buttons or menus that allow the operator to choose specific
tasks or actions.

Main Window Overview


Figure 1-10 1 2 3
The QDR Series
Main Window
Note: Figure depicts Discovery
Main Window, other models have
slightly different appearance but
the same functions.

KP1068A-0103

6 5 4 7

Main Window Features


The important features of the main window include:
1 - Menu Bar
The Menu Bar can be used with the mouse or keyboard to
perform or analyze a scan, create or edit patient records and
perform system maintenance functions.
2 - Main Work Area
The Main Work area acts as the holder for all the other windows
or dialog boxes the operator uses when acquiring or analyzing a
scan.

Introducing the QDR Series System 1-9


QDR Reference Manual

3 - Filing Cabinet buttons


The Filing Cabinet buttons open the file cabinets that store patient
and scan records.
4 - System Function buttons
The System Function buttons start the major system maintenance
activities that the operator perform such as archiving (long-term
storage).
5 - System Message ticker tape
The System Message ticker tape provides a moving message
about activities that the operator may need to perform such as
reminding the operator to perform daily QC.
6 - Daily Function buttons
The Daily Function buttons begin five daily activities that the
operator may use to acquire and analyze a scan.
7 - Help
The APEX Online Help system provides step-by-step directions
for using the APEX software. The operator can access online help
by clicking on the Help button (either at the top or bottom of the
main window or on most dialog boxes), or by pressing the F1 key.

Introducing the Quality Control Spine Phantom


Note: Refer to Chapter 5 for Daily quality control procedures ensure that the QDR Series
more information about using system functions properly. When the operator performs daily
the QC Spine Phantom.
QC, the spine phantom is scanned. The system adds the results of
the scan to a database, and plots this on a graph. This records a
daily comparison to 10 separate measurements taken at the time
of the system’s installation, and provides the basis for the system
calibration.

Continuous, self-calibration is a feature of the QDR series. The


scanner has a calibration drum permanently installed inside the
unit that acts as an internal reference to automatically compensate
for variations in the emission from the X-ray source.1 A
calibration factor (CF) is listed on an individual scan report and is
used to identify if a scan was performed before or after a
calibration change or instrument failure.

1. pg. 63, Fogelman, I, Wahner, The Evaluation of Osteoporosis: Dual


Energy X-ray Absorptiometry in Clinical Practice, Martin Duntz, UK,
1994

1-10 Introducing the QDR Series System


QDR Reference Manual

Introducing the Positioning Aids


The QDR Series includes several aids to help the operator
position patients for specific exams. Positioning aids maintain the
correct patient position during the acquisition of a scan. They
include:
• Knee Positioner
• Foot Restraint (hip positioner)
• Head Positioner

Knee Positioner
Figure 1-11 Knee
The Knee Positioner
Positioner

KP1068A-0105

The large knee positioner is placed under the patient’s lower legs.
This allows positioning of the femurs so that they are as close to
90o to the spine as possible in order to flatten the back. The
operator can rotate the pillow to one of three sides to adjust for
the height of the patient and length of their legs.

Foot Restraint (Hip Positioner)


Figure 1-12
The Foot Restraint

R
Velcro
Strap
KP1068A-0106

The Foot Restraint (Hip positioner) maintains the correct position


of the femur and minimizes movement during the acquisition of a
hip scan. It is placed between the feet. The foot and leg of the side
being examined are rotated inwards, with the foot against the

Introducing the QDR Series System 1-11


QDR Reference Manual

fixture. A strap is placed around the foot, secured in the correct


position.

Head Positioner for A and SL Models


This Head Positioner is supplied on A and SL models and is
placed under the patient’s head for correct positioning during an
AP/Lateral Spine scan. It provides patient comfort to the elevated
arms. The arms must be moved away from the spine for both the
AP and lateral scans.
Figure 1-13
The Head Positioner
for A and SL Models
Head
Positioner

KP1068A-0108

Introducing the Product Support Materials


The QDR Series systems come with an array of support resources
such as manuals, online help and internet support. These include:
QDR Reference Manual
The QDR Reference Manual (this manual) provides system
technology.
APEX Online Help
The online help system provides step-by-step directions for using
the APEX software. Access online help by clicking on the Help
button (either at the top or bottom of the main window or on most
dialog boxes), or by pressing the F1 key.
QDR Series Technical Specifications Manual
The Technical Specifications Manual provides specifications and
regulatory information for QDR Systems.

1-12 Introducing the QDR Series System


QDR Reference Manual

Operator’s Manual
The Discovery and Explorer Operator’s Manuals provide user
instructions and are available in English, French, Italian, German,
Spanish, Brazilian Portuguese, and Polish.
Hologic Internet Site
The Hologic web site (http://www.hologic.com) provides a wide
variety of customer support information including service and
training offerings, FAQ, and contact information.
Information about cyber-security for QDR systems is available on
the Hologic web site at the following URL:
http://www.hologic.com/cc/netwrksec.htm

Introducing the QDR Series System 1-13


QDR Reference Manual

1-14 Introducing the QDR Series System


Chapter 2
Understanding QDR Series
Technology
This chapter provides explanations of the Quantitative Digital
Radiography application of DXA technology and the scan results
produced by the QDR Series systems.

Content
Subject Page
Introducing Key Elements 2-2
Introducing the Scan Sites 2-4
Lumbar Spine 2-4
Proximal Femur 2-5
Forearm 2-6
Whole Body 2-7
Understanding the Scan Results 2-8

2-1
QDR Reference Manual

Introducing Key Elements


To help provide the user with a fundamental understanding of the
operation of the QDR Series system, the key elements are
explained in the following sections:
• Presenting DXA technology (brief description)
• Introducing the scan sites (anatomical sites of the skeleton
areas that are scanned)
• Understanding the scan results (results produced by a scan
that can be interpreted by a physician)

Presenting DXA Technology


Dual Energy Absorptiometry (DXA or DEXA) is a method
utilized to measure bone density. Developed by Hologic in 1987,
these X-ray instruments emit alternating high, 140kVp, and low,
100kVp, X-rays. The first generation Hologic QDR instruments
used a single, or pencil X-ray beam, and a single detector. The
scanner worked by moving the detector and X-ray source in a
rectilinear pattern to acquire data of the lumbar spine and hip.
Continued development of the pencil beam single detector
configuration enabled a second generation of QDR instruments to
measure tissue composition, fat and lean, as well as bone. The
third generation QDR densitometers employ multiple detectors
and a dual energy X-ray fan-beam. The arm moves in a single
direction decreasing scan times from minutes to seconds with
improved image quality and equivalent precision. The beam
sweeps across a region of interest on the scan area in a fan-shaped
pattern, and is detected by a high-resolution multi-detector array
to form a high quality image.

B
C

KP0886-0201

2-2 Understanding QDR Series Technology


QDR Reference Manual

Accuracy and Precision


Accuracy is the ability to measure the bone mineral density
(BMD) of an object the same as by using an alternate method
such as ashing. Precision is the ability to reproduce the
measurement repeatedly. Independent tests scanning the same
patients on both pencil and fan beam instruments have
demonstrated no significant differences in accuracy. Studies have
reported a precision of 1% or better for spine, hip, and whole
body measurements when scanned on Hologic instruments.

Principles of Operation
The basic principle of DXA data acquisition is based on the
different bone and soft tissue attenuation characteristics at the two
pulsed X-ray levels. A calibration drum, comprised of known
amounts of bone and soft tissue equivalent materials is placed in
the beam. As the beam passes through the patient, more lower
energy X-rays than higher energy X-rays are absorbed by the
anatomical structures in the patient. The beam is then registered
by the detectors in the C-arm. The raw scan data, containing the
attenuation values of tissue, bone, and the calibration drum are
captured and transferred to a computer. An algorithm interprets
each pixel, creates an image and quantitative measurement of the
bone and body tissues.

Understanding QDR Series Technology 2-3


QDR Reference Manual

Introducing the Scan Sites


The scan areas that provide the most information about patterns
of bone loss include:
• Lumbar Spine
• Proximal Femur
• Forearm
• Whole Body

Lumbar Spine
The spine consists of a number of vertebrae categorized as
cervical (upper column), thoracic (mid column), lumbar (mid to
lower), and sacrum and coccyx (at the end). Significant bone loss
due to aging or osteoporosis is often observed in the vertebrae,
and particularly in the lumbar area.

KP0886-0202

The AP Lumbar Spine is the most widely used anatomical site for
the evaluation of osteoporosis. The vertebral bodies, L1 through
L4, contain approximately 40% cortical and 60% trabecular bone.
The high amount of trabecular bone and the relative ease of

2-4 Understanding QDR Series Technology


QDR Reference Manual

reproducibility of positioning of the spine provide an important


source for baseline bone density data.

Proximal Femur
The femur is the skeletal site where the most serious
consequences of osteoporosis (fractures) often occur.

Femoral Neck

Trochanter
Ward’s
Triangle

Inter-trochanter
Total Hip

KP0886-0203

BMD results for proximal femur (Hip) scans, are reported for five
different anatomical areas: Total Hip, Femoral Neck, Trochanter,
Inter-Trochanter, and Ward’s Triangle.
• The Total Hip is the most reproducible measurement of
the hip and has become the preferred clinical
measurement of the hip bone mineral density.
• The Femoral Neck region contains a large proportion of
trabecular bone and provides an easily reproducible area
in the hip for DXA scanning. The analysis is completed
by placing a specific region at the anatomical landmark.
That landmark, called the “greater trochanter notch”
permits reproducibility for longitudinal scanning.
• The Trochanter is a triangular region whose boundaries
are the lateral edge of the femur and the inferior edge of
the neck box and the solid line where the edge of the
femur changes curvature below the trochanter.

Understanding QDR Series Technology 2-5


QDR Reference Manual

• The Inter-Trochanteric region is below the Femoral


Neck to the bottom edge of the ROI.
• The Ward’s Triangle is located automatically by the
system and places a small box approximately 1cm x 1cm
at the area of minimum density in the femoral neck
region. For some patients, an area of minimum density
cannot be found. When this occurs, the Ward’s Triangle is
automatically centered at the intersection of the femoral
midline and the initial position of the bottom edge of the
femoral neck box.

Forearm
Bone loss may be notable in the forearm and for this reason it can
be used as a scan site. The forearm scan includes the region of the
radius and ulna and the wrist bones (carpals).

Carpal Bones

Ulna

Radius

KP0886-0204

The Ulna is the smaller of the two bones and is located laterally,
or on the “small finger side” of the forearm.
The Radius is larger than the Ulna and allows the rotation of the
forearm. This bone includes an important anatomical structure
described as the “Distal 1/3.” This is an active metabolic bone
site.
The Carpal bones are two rows of small bones also known as the
“Wrist” bones.

2-6 Understanding QDR Series Technology


QDR Reference Manual

Whole Body
Note: Whole Body is available as Discovery-A, -W and Wi instruments allow regional and global
an option on Explorer. measurement of the Whole Body. An advanced application of
QDR technology can directly measure fat and skeletal mass and is
useful in research and clinical studies.

The anatomical
areas analyzed are:
• the head
• the left arm
• the right arm
• the left rib
• the right rib
• the T-spine
• the L-spine
• the pelvis
• the left leg
• the right leg
The subregion
analyzed includes
all of the above
except the head.
KP0886-0205

Understanding QDR Series Technology 2-7


QDR Reference Manual

Understanding the Scan Results


The QDR DXA technology provides accurate and precise
measurement of small changes in bone mineral density (BMD)
measured in grams per centimeter squared (g/cm2). Presently
BMD measurements offer the physician the most reliable means
of recording the rate of bone loss or gain and estimating a
patient’s risk of fracture.

Z-Score
The Z-score is a measure of the difference between the patient’s
BMD and that of healthy people of the same sex, age and
ethnicity.

T-Score
The T-score is a measure of the difference between the patient’s
BMD and that of a young adult population of the same sex and
ethnicity.

Reference Database
The reference database represents the average results as a
function of age, sex, and ethnicity for a matched population.
Reference curves specify average BMD, and standard deviation
as a function of age. Each curve applies to a specific scan type,
analysis type, bone region, patient sex, and ethnic group.
Reference database reports use these reference curves for graphic
display, and for the calculation of Z and T scores. Each of these
curves compares a patient scan, or a series of patient’s scans, with
the reference database.

2-8 Understanding QDR Series Technology


Chapter 3
APEX Software
The APEX software displays a user interface on the computer screen
that contains both words and pictures. This chapter provides
information on the use of this user interface and its features.

Content
Subject Page
Understanding the Mouse 3-2
Using the Mouse with APEX Software 3-3
Understanding the User Interface 3-4
Understanding Main Window Components 3-7
Introducing the System Software Reminders 3-13
Introducing the System Software Help 3-14

3-1
QDR Reference Manual

Understanding the Mouse


A graphical user interface (GUI) is designed so that you can move
easily from step to step with the help of a mouse. The mouse acts
like a keyboard.
The mouse allows you to move a bit faster around a computer
screen than using just the keyboard. You can use a mouse to move
the cursor (arrow) around on the computer screen, and “point” to
different objects.
The “mouse” is shaped so that it is easy for your hand to hold. As
a result, it looks like a mouse with a tail. The tail is the wire that
attaches the device to the computer. Although the look of this
device has changed for some computers over the last several
years, they are still generally referred to as a mouse.

Holding the Mouse


You hold a mouse by covering it with your hand. The mouse-tail
(wire) is positioned up at your fingers with your thumb resting on
the side. Your first finger rests lightly on the left mouse button
and your second finger rests lightly on the right mouse button.
The heel of your hand rests lightly on the back end of the mouse.
This helps you move the mouse more easily.

Mouse Tasks
You can use the mouse:
To... Move the mouse And...
and point...
Enter text Inside a field Press and release (click) the
in a field left mouse button once to
insert the cursor into a field
so that you can type text.
Perform an Anywhere on a Press and release the left
action button in a mouse button (click).
window
Move an To the object Press and hold the left mouse
object on button down.
the desktop Move the mouse, dragging
the icon to the spot that you
want it to remain.
Release the mouse button.
This is referred to as
dragging and dropping.

3-2 APEX Software


QDR Reference Manual

Using the Mouse with APEX Software


The mouse can have two or three buttons. You only need to use
the left mouse button with the APEX software.
Left Mouse
Button KP0886-0301

Pointing with the Mouse


As you move the mouse, an arrow moves on the screen. You must
point to an object before you can click on it.

Clicking the Mouse


Clicking the mouse means quickly pressing and releasing the left
mouse button. For example, when the instructions indicate to
click the OK button, it means positioning the arrow on the button,
and quickly pressing and releasing the left mouse button once.

Double-Clicking the Mouse


Double-clicking means clicking the left mouse button twice
quickly. Use the double-click function to open another window.
For example, double-clicking on a patient’s name displays the
Patient Record window instead of having to click on the name
once, and then click the OK button.

Dragging and Dropping


Dragging and dropping means pressing and holding the left
mouse button on an object or item, dragging that item to another
area of the screen, and releasing the mouse button.

APEX Software 3-3


QDR Reference Manual

Understanding the User Interface


The user interface refers to the screen displayed on your computer
monitor which is called a “window.” A window contains various
information which can be “opened and closed,” multiple
windows can be opened at the same time.
The User Interface windows contain words and picture such as
buttons and icons. These pictures represent sets of instructions to
the computer that are activated with the mouse.

User Interface Windows


The User Interface windows provide access to different scan
functions using buttons, menus, and a task bar. There are three
basic types of windows:
• Main window
• Tab windows
• Dialog windows

Main Window
Figure 3-1 Menu bar Main work area
The Main Window
Note: Figure depicts Discovery
Main Window, Explorer has
slightly different appearance but
the same functions.

KP1068A-0302
Function Buttons Ticker Tape

The main window includes function buttons, menu bar and a


ticker tape. It also has a main work area as shown in Figure 3-1.

3-4 APEX Software


QDR Reference Manual

Tab Windows
These windows are generally a series of windows with tabs that
are grouped together because they contain related information.
For example, the “Select a Scan” window is a series of three tabs
that allow you to select unanalyzed scans, analyzed scans, or all
scans.
Figure 3-2 Window Tabs Search Text Field
Tab Window

KP0886-0304
Sort List Function Button Scroll Bar

These windows can include:


• Window tabs
• Search Text fields
• Sort lists
• Function buttons
• Scroll Bar

APEX Software 3-5


QDR Reference Manual

Dialog Windows
Figure 3-3 Radio Button Drop Down Menu
Dialog Windows

Check Box

Text Field

Dialog windows allow you to perform certain actions such as


entering or editing information. For example, the Confirm
window allows you to enter patient’s height, weight, and the
operator’s initials.
These windows include a number of features such as:
• Radio buttons - allow you to choose one, and only one,
item from two or more items
• Drop down lists - allow you to select from a list of items

3-6 APEX Software


QDR Reference Manual

• Check boxes - allow you to choose the item (box checked)


or not (box unchecked)
• Text fields - allow you to enter text (from keyboard)

Understanding Main Window Components


Each window has unique features that allow you to acquire and
analyze a scan, and store the records for future use.

Main Window Buttons


The main window (see Figure 3-1) provides a series of buttons
that help you perform the primary tasks of the system.
Function Buttons

Button Description
This button initiates the daily QC procedure.
When flashing, indicates that daily QC needs to
be performed before you can begin scanning
KP1035A-0307 patients.

This button initiates the patient exam by


displaying a list of existing patients. You can also
add a new patient record, or edit an existing
KP1035A-0309 patient record.

This button allows you to locate archived scans


and restore them onto the computer. You can also
select a patient to get their biography and
KP1035A-0311 insurance details.

This button initiates Analyze of a specific scan to


generate BMD results.

KP1035A-0313

This button allows you to create, configure, or


print a report for a scan that has already been
analyzed.
KP1035A-0315

APEX Software 3-7


QDR Reference Manual

Filing Cabinet Buttons

Button Description
This button opens the Patient drawer and displays
a list of all patients on the system. You can also
add a new patient record, edit an existing patient
KP1035A-0317
record or input data from any DXA system for
patient follow up.
This button opens the Scan drawer and displays a
list of scans currently on the system. You can also
view scan details of a selected scan.
KP1035A-0319

System Management Buttons

Button Description
This button initiates saving scan data to selected
media so that it may be restored for later use.

KP1035A-0321

This button initiates back ups of the database in


case of computer failure.

KP1035A-0323

This button initiates exiting the software.

KP1035A-0325

Using Menus
The menus available in the software are listed in the menu bar at
the top of the window. In most cases, you can use the buttons to
perform the same tasks as some of the menu options.
Choosing Menu Items
You can choose a menu selection by:
1. Pointing to it with the mouse and clicking on it; or,
2. Pressing and holding the Alt key while typing the underlined
letter in the menu title. For example, from the main window
press Alt+F to open the Filing Cabinet menu, then press
Alt+P to open the Patient drawer.

3-8 APEX Software


QDR Reference Manual

Using Sub-Menus
Some menu selections display a sub-menu that appears alongside
the parent menu. Menu items that open a sub-menu display with a
right arrow (→) or left arrow (←) symbol. Select an item from a
sub-menu the same way that you select an item from a regular
menu. For example, press Alt + U to open the Utilities menu.
Scroll down to Database Tools, then scroll across to select an item
from the sub-menu.
Closing Menus
To close a menu, click anywhere outside of the menu.

Understanding Other Window Components


There are a number of other window components located on the
Tab and Dialog windows that allow you to perform actions that
include:
• Enter text
• Finding records stored in folders with tabs.
• Selecting an item from a list that drops down from a field.
• Selecting functions by turning on and off radio buttons.
• Performing tasks using command buttons.
• Moving up and down through a window using a scroll bar.
• Sorting a list.
• Selecting an item in a list.

Entering Text

KP0886-0327

Use the Tab key or your mouse to position the cursor into the text
field. Click once inside the text field to begin entering text.

APEX Software 3-9


QDR Reference Manual

Tabs

KP0886-0328

Some dialog boxes in the software have a series of tabs across the
top that let you view specific information. The tabs are similar to
folder tabs in a filing cabinet. Click once on the tab to open that
“folder.”

Drop-down Lists

The dialog box shows a drop-down list for the Spine Analysis.
Clicking on the down arrows displays the list of options from
which to choose. Highlight the desired field and click once.

Radio Buttons

KP0886-0330

Radio buttons allow you to pick one choice from a number of


options. Click in the small circle next to the option to select it. In
this example, a “filled” circle indicates that Choose Analysis
Method is the selected choice.

Command Buttons

KP0886-0331

Buttons activate an operation. In a dialog box, for example,


clicking on the Cancel button cancels the operation and closes the
dialog box. Some examples of buttons used in the software
include OK, SelectAll, DeselectAll, Cancel, Close, Next, Back.

3-10 APEX Software


QDR Reference Manual

Check Boxes

KP1035A-0331a

Check Boxes appear in various menus and are used to select one
or more items in a list.

Scroll Bars
Scroll bars appear on the right and bottom edges of some screens.
Use the scroll bars to view portions of the window that are not
entirely visible.
View hidden portions of the screen by clicking on the arrows at
either end of the scroll bar, or by dragging the slider box between
the arrows.
The horizontal scroll bar moves the contents within the window
left and right. The vertical scroll bar moves the contents of the
window up and down.

Sorting List Items


The software displays screens with listings such as patient data,
scan data, and so on. This section provides helpful hints on
changing the sort order of lists, as well as selecting items from a
list.

Identifying the Sort Order

KP0886-0332

APEX Software 3-11


QDR Reference Manual

An asterisk (*) in the heading identifies the sort order. For


example, an asterisk in the Patient heading indicates the sort order
is alphabetical by the patient’s last name.

Changing the Sort Order


You can change the sort order of listed items by clicking on the
heading above the item. For example, to sort the list by Patient
ID, click on Patient ID.

Selecting Items From a List


Selecting One Item
To select one item from a list, click on the item and the item
becomes highlighted. Then click OK to select that item.

Selecting Multiple Items


To select multiple items from a list, click on the first item, then
press and hold down the Ctrl key while clicking on subsequent
items. Selected items are highlighted. Release the Ctrl key and
click OK.

Selecting a Range of Items


To select a range of items, click on the first item in the range, then
press and hold the Shift key and click on the last item. Selected
items are highlighted. Release the Shift key and click OK.

3-12 APEX Software


QDR Reference Manual

Introducing the System Software Reminders


The software interface includes visual software aids and functions
to:
• Remind you to perform a certain task
• Show you the function of a button before you click on it
• Let you get help on a certain function or task
• View important system messages

Flashing Buttons
Flashing buttons indicate that you need to perform a task before
you can proceed with daily operations or that you are in danger of
losing unsaved information.

Daily QC
When the Daily QC button is flashing, it indicates
that you must perform QC before performing an
examination. Refer to Chapter 5, Quality Control,
KP1035A-0307 for detailed instructions for performing daily QC.

System Backup
When the System Backup button is flashing, it
indicates that you need to perform a system
backup. Backing up the system is a weekly
KP1035A-0323
procedure for ensuring that your data is protected
in the event of a system failure. You can still
perform your daily operations, but you should
perform the system backup (by clicking the
System Backup button), as soon as possible.

APEX Software 3-13


QDR Reference Manual

Introducing the System Software Help


APEX has many tools to help you run the software. Some of them
are:
• Tool Tips
• Online context sensitive help
• Ticker Tape messages

Tool Tips
Tool tips appear when you position the mouse pointer over a
button. A small rectangle appears with a description of the button.
You do not need to click the button to view the tool tip.

Online Context Sensitive Help


This feature gives you help on the active window, that is the
window or dialog box you are currently using.
To use Context Sensitive Help:
Step Action
1 If the blue Help button (see left) is visible, click it to get
KP0886-0333 help on the window that you are working on.

2 If the blue Help button is not visible, look for a Help button
KP0886-0334 on the currently open window.
3 Click Help if the button is there.
In either case, a screen similar to the one below appears.
The online help is standard Windows help and operates the
same.

KP0886-0335

You can learn more about Windows help by clicking Start


KP0886-0336
on the Windows Task bar and selecting Help.

3-14 APEX Software


QDR Reference Manual

Ticker Tape

KP0886-0337

The ticker tape appears at the bottom of the main window. It


displays helpful information, or instructions to remind you to
perform a specific task. For example, the ticker tape message
“Click on the System Backup button to start the...” displays,
along with a flashing System Backup button to remind you to
perform a system backup.

APEX Software 3-15


QDR Reference Manual

3-16 APEX Software


Chapter 4
Performing Daily System
Operations
This chapter provides information about the basic operating
procedures for the Discovery, Explorer, QDR 4500 and Delphi
systems including:
• Turning the System On and Off.
• Switches and indicators for daily system operation.
• Emergency procedures.
Content
Subject Page
Discovery Systems 4-2
Discovery Circuit Breaker and Indicator 4-2
Discovery-C, -Ci, -W and -Wi Control Panel 4-2
Discovery-A and -SL Control Panel 4-4
Turning Discovery Power On and Off 4-5
Discovery Emergency Procedures 4-7
Explorer Systems 4-9
Explorer Circuit Breaker and Indicator 4-9
Explorer Control Panel 4-9
Turning Explorer Power On and Off 4-10
Explorer Emergency Procedures 4-11
QDR 4500 Systems 4-14
QDR 4500 Power Module 4-14
QDR 4500 Control Panel 4-15
Turning QDR 4500 Power On and Off 4-18
QDR 4500 Emergency Procedures 4-20
Delphi Systems 4-23
Delphi AC Power Isolation Module 4-23
Delphi Control Panels 4-23
Turning Delphi Power On and Off 4-26
Delphi Emergency Procedures 4-28

4-1
QDR Reference Manual

Discovery Systems
Discovery Circuit Breaker and Indicator
The Discovery circuit breaker and indicator are located on the
back of the left pedestal (see Figure 4-1). Table 4-1 gives the
description and function of the circuit breaker and indicator.
Figure 4-1
Location of Discovery Circuit
Breaker and Indicator

KP1068_002-0401

Table 4-1 Description Function


Discovery Circuit Breaker
and Indicator Circuit Breaker Turns the power to the Discovery
system off (0) and on (1).
Indicator Glows green when the circuit breaker is
in the on (1) position.

Discovery-C, -Ci, -W and -Wi Control Panel


The Discovery-C, -Ci, -W and -Wi Control Panel (see figure
below) is located on the right front of the examination table. Table
4-2 gives the description and function of the switches and lamps.
Figure 4-2
Discover-C, -Ci, -W and -Wi
Control Panel

MAN-00175_002-0401

4-2 Performing Daily System Operations


QDR Reference Manual

Table 4-2 Switch/Lamp Function Lamp Indicators


Discovery-C, -Ci, -W and -Wi
Control Panel Switches and Patient On/Off The table moves to the
Lamps front-most position and
the C-arm moves to the
left-most position
Center Moves the table and C-
arm to the center
position
Arm (right) Moves the C-arm to the
right (towards the head
of the table)
Table (in) Moves the table toward
the back
Laser Turns the laser cross- When the laser is
hair indicator ON or ON the Laser On/
OFF Motion Enabled
green indicator
blinks
Arm (left) Moves the C-arm to the
left (towards the foot of
the table)
Table (out) Moves the table toward
the front
EMERGENCY Pressing this button
Stop Button stops scanning
immediately. Pressing
the button down and
turning it to the right
resumes normal
operation.
X-Ray When X-rays are
(amber indica- being produced this
tor) indicator is ON
Power Turns the power to the When the indicator
instrument on or off is green, power is
ON in the
instrument

Performing Daily System Operations 4-3


QDR Reference Manual

Discovery-A and -SL Control Panel


The Discovery-A and -SL Control Panel (see Figure 4-3) is
located on the right front of the examination table. Table 4-3 gives
the description and function of the switches and lamps.
Figure 4-3
Discovery-A and -SL
Control Panel

MAN-00175_002-0403

Table 4-3 Switch/Lamp Function Lamp Indicators


Discovery-A and -SL Control
Panel Switches and Lamps Patient On/Off The table moves to the
front-most position and
the C-arm moves to the
left-most position
Enable Lateral Enables the C-arm and Light Blinking ON
table to move for the = You need to
acquisition of the AP/ perform a task as
Lateral Spine scan. The instructed.
system instructs you
how and when to use
this switch.
Center Moves the table and C-
arm to the center
position
Arm (right) Moves the C-arm to the
right (towards the head
of the table)
Table (in) Moves the table toward
the back
Laser Turns the laser cross- When the laser is
hair indicator ON or ON the Laser On/
OFF Motion Enabled
green indicator
blinks
X-Ray (amber When X-rays are
indicator) being produced this
indicator is ON

4-4 Performing Daily System Operations


QDR Reference Manual

Arm (left) Moves the C-arm to the


left (towards the foot of
the table)
Table (out) Moves the table toward
the front
EMERGENCY Moves the table up Light ON = Enabled
Table Lift when pressed. when you press the
Emergency Stop
switch.
EMERGENCY Pressing this button
Stop Button stops scanning
immediately. Pressing
the button down and
turning it to the right
resumes normal
operation.
Power Turns the power to the When the indicator
instrument on or off is green, power is
ON in the
instrument

Turning Discovery Power On and Off


Do not use the circuit breaker on the back of the left pedestal (see
Figure 4-1) to turn the power on or off. If the green indicator light
above the circuit breaker goes out while the system is running or
if the indicator is not on at the beginning of the day, refer to the
“Discovery Emergency Procedures” on page 4-7.

Beginning of the Workday


Perform the following steps at the start of each workday to turn
the Discovery System on.

Step Action
1 Verify that the green indicator on the back of the left
pedestal is on (Figure 4-1).
This light indicates that the system is receiving AC power.
2 On the Control Panel (see “Discovery-C, -Ci, -W and -Wi
Control Panel” on page 4-2 or “Discovery-A and -SL
Control Panel” on page 4-4). press the Power button.
The Power green indicator illuminates.

Performing Daily System Operations 4-5


QDR Reference Manual

Step Action
Note: Because the computer and 3 On the Operator Console, turn the computer on.
peripherals may vary, refer to The monitor (in power save mode) and printer(s) should
documentation that came with the
already be on during a normal day’s power-up.
Discovery System for controls
and indicators on the computer, The computer boots itself and starts Windows XP. It then
monitor and printers. starts the APEX software automatically. If there are any
problems, warning messages will tell you what to do.

End of the Workday


Perform the following steps at the end of each workday to turn the
Discovery System off.

Step Action
1 Make sure only the Main Window appears on the monitor.
This is to ensure that you are not in the middle of any
function and all activity has stopped.
2 Click Exit.
The Exit button is in the lower right-hand corner of the
screen.
KP1035A-0325
3 In the Exit QDR System dialog box that appears, select
“Exit QDR with shutdown?” and click OK.

KP1068A-0404

This shuts down the APEX software, and the Windows XP


operating system
4 On the computer, press the power on/off switch to power
off the computer.
5 Turn off the Discovery System by pressing the Power
switch on the control panel (see “Discovery-C, -Ci, -W
and -Wi Control Panel” on page 4-2 or “Discovery-A and
-SL Control Panel” on page 4-4).
The green indicator in the Power button goes out.
Important: Do not turn off the Do not turn off the printer(s).
circuit breaker on the back of the
left pedestal (see Figure 4-1). The Do not turn off the monitor. It goes into power save mode
green indicator should remain on. when the computer shuts down.

4-6 Performing Daily System Operations


QDR Reference Manual

Discovery Emergency Procedures


While operating the Discovery there are three emergency
situations that may require your action:
• A facility power failure occurs
• Equipment fails while in operation
• System is turned off

Operator Action During a Power Failure


If the facility power fails, it is a good idea to turn off all the
equipment. When the power returns, frequently it is not stable and
can spike up and down for a few moments. You should wait a few
minutes after power returns to turn on the equipment.

Shutting Down During a Power Failure


Step Action
1 If the system was in operation when the power failure
occurred, assist the patient from the table
The Patient On/Off switch will be inoperative at this time.
2 Turn off the computer.
Note: Refer to documentation 3 Turn off the circuit breaker (see Figure 4-1).
that came with the system for Since this is a power failure condition, the green Indicator
controls and indicators on the
should already be off.
computer, monitor and printer.

After Power is Restored


Step Action
1 After waiting a few minutes for the facility power to
stabilize,
Turn on the circuit breaker.
The green indicator light above the circuit breaker goes
on.
2 Turn the computer on.
3 Perform the procedure in “Beginning of the Workday” on
page 4-5).

If Equipment Fails While in Operation


If the operator determines that the equipment has malfunctioned
(e.g., X-Ray is on and C-arm is not moving) perform the
following procedure.

Performing Daily System Operations 4-7


QDR Reference Manual

Emergency Stop Procedure:


Step Action
1 On the Control Panel, press the red Emergency Stop
Button.
The table and the C-arm immediately stop moving, and the
X-rays and laser turn off.
2 Assist the patient off the table.
Note: The power should not be 3 Turn off the circuit breaker (see Figure 4-1).
restored without consulting
Hologic. 4 Disconnect the power cord from the AC outlet (if
possible).
5 Call Hologic customer service.

If AC Power has been Turned Off


If for any reason the circuit breaker has been turned off (not
because of an equipment failure), or the plug has been pulled
from the wall, use the following procedure to apply power to the
system:
Applying power to the system:
Step Action
1 If necessary, place the power cord into the AC outlet.
Note: If the system does not turn 2 Place the circuit breaker in the 1 (on) position (see
on, make sure that the power cord Figure 4-1).
is plugged into the wall. If it is,
The green Indicator above the circuit breaker illuminates.
call Hologic customer service.
3 Perform the procedure in “Beginning of the Workday” on
page 4-5

4-8 Performing Daily System Operations


QDR Reference Manual

Explorer Systems
Explorer Circuit Breaker and Indicator
The circuit breaker and indicator, on Explorer, are located on the
Power Module (see Figure 4-4). Table 4-4 gives the description
and function of the circuit breaker and indicator.
Figure 4-4
Location of Explorer Circuit
Breaker and Indicator

00132-001

Table 4-4 Description Function


Explorer Circuit Breaker
and Indicator Circuit Breaker Turns the power to the Explorer system
off (0) and on (1).
Indicator Glows green when the circuit breaker is
in the on (1) position.

Explorer Control Panel


The Explorer Control Panel (see Figure 4-5) is located on the C-
Arm. Table 4-5 gives the description and function of the switches
and lamps.
Figure 4-5
Explorer Control Panel

MAN-00175_002-0401

Table 4-5 Switch/Lamp Function Lamp Indicators


Explorer Control Panel
Switches and Lamps Patient On/Off The table moves to the
front-most position
Center Moves the table to the
center position

Performing Daily System Operations 4-9


QDR Reference Manual

Arm (right) Moves the C-arm to the


right (towards the head
of the table)
Table (in) Moves the table toward
the back
Laser Turns the laser cross- When the laser is
hair indicator ON or ON the Laser On/
OFF Motion Enabled
green indicator
blinks
Arm (left) Moves the C-arm to the
left (towards the foot of
the table)
Table (out) Moves the table toward
the front
EMERGENCY Press this button to
Stop Button immediately stop
scanning. Press the
button down and turn it
to the right to resume
normal operation.
X-Ray When X-rays are
(amber indica- being produced this
tor) indicator is ON

Turning Explorer Power On and Off


Do not use the circuit breaker on the Power Module (see
Figure 4-4) to turn the power on or off. If the green indicator light
above the circuit breaker goes out while the system is running or
if the indicator is not on at the beginning of the day, refer to the
“Explorer Emergency Procedures” on page 4-11.

Beginning of the Workday


Perform the following steps at the start of each workday to turn
the Explorer on.

Step Action
1 Verify that the green indicator on the Power Module is on
(Figure 4-4).
This light indicates that the system is receiving AC power.

4-10 Performing Daily System Operations


QDR Reference Manual

Step Action
Note: Because the computer and 2 On the Operator Console, turn the computer on.
peripherals may vary, refer to The monitor (in power save mode) and printer(s) should
documentation that came with the
already be on during a normal day’s power-up.
Explorer System for controls and
indicators on the computer, The computer boots itself and starts Windows XP. It then
monitor and printers. starts the APEX software automatically. If there are any
problems, warning messages will tell you what to do.

End of the Workday


Perform the following steps at the end of each workday to turn the
Explorer off.

Step Action
1 Make sure only the Main Window appears on the monitor.
This is to ensure that you are not in the middle of any
function and all activity has stopped.
2 Click Exit.
The Exit button is in the lower right-hand corner of the
screen.
KP1035A-0325

3 In the Exit QDR System dialog box that appears, select


“Exit QDR with shutdown?” and click OK.

KP1068A-0404

This shuts down the APEX software, and the Windows XP


operating system
4 On the computer, press the power on/off switch to power
off the computer.
Important: Do not turn off the Do not turn off the printer(s).
circuit breaker on the Power
Module. The green indicator Do not turn off the monitor. It goes into power save mode
should remain on. when the computer shuts down.

Explorer Emergency Procedures


While operating the system there are three emergency situations
that may require your action:

Performing Daily System Operations 4-11


QDR Reference Manual

• A facility power failure occurs


• Equipment fails while in operation
• System is turned off

Operator Action During a Power Failure


If the facility power fails, it is a good idea to turn off all the
equipment. When the power returns, frequently it is not stable and
can spike up and down for a few moments. You should wait a few
minutes after power returns to turn on the equipment.

Shutting Down During a Power Failure


Step Action
1 If the system was in operation when the power failure
occurred, assist the patient from the table
The Patient On/Off switch will be inoperative at this time.
2 Turn off the computer.
Note: Refer to documentation 3 Turn off the circuit breaker (see Figure 4-4).
that came with the system for Since this is a power failure condition, the green Indicator
controls and indicators on the
should already be off.
computer, monitor and printer.

After Power is Restored


Step Action
1 After waiting a few minutes for the facility power to
stabilize,
Turn on the circuit breaker (see Figure 4-4).
The green indicator light above the circuit breaker goes
on.
2 Turn the computer on.
3 Perform the procedure “Beginning of the Workday” on
page 4-10.

If Equipment Fails While in Operation


If the operator determines that the equipment has malfunctioned
(e.g., X-Ray is on and C-arm is not moving) perform the
following procedure.

4-12 Performing Daily System Operations


QDR Reference Manual

Emergency Stop Procedure:


Step Action
1 On the Control Panel, press the red Emergency Stop
Button (see Figure 4-5).
The table and the C-arm immediately stop moving, and the
X-rays and laser turn off.
2 Assist the patient off the table.
Note: The power should not be 3 Turn off the circuit breaker.
restored without consulting
Hologic. 4 Disconnect the power cord from the AC outlet (if
possible).
5 Call Hologic customer service.

If AC Power has been Turned Off


If for any reason the circuit breaker has been turned off (not
because of an equipment failure), or the plug has been pulled
from the wall, use the following procedure to apply power to the
system:
Applying power to the system:
Step Action
1 If necessary, place the power cord into the AC outlet.
Note: If the system does not turn 2 Place the circuit breaker in the 1 (on) position (see
on, make sure that it has power Figure 4-4).
(that is, the cord is plugged into
The green Indicator above the circuit breaker illuminates.
the wall). If it is, call Hologic
customer service. 3 Perform the procedure “Beginning of the Workday” on
page 4-10.

Performing Daily System Operations 4-13


QDR Reference Manual

QDR 4500 Systems


QDR 4500 Power Module
The Power Module supplies electrical power to the QDR 4500
system and is located below the workstation table. The Power
Module right side and back panel include on/off switches, circuit
breakers, X-ray key and lamp indicators.

QDR 4500 Power Module Side Panel


The Power Module side panel contains the computer on/off
switch, the instrument power switch, the X-ray enable key, and
most of the circuit breakers (See Figure 4-6 and Table 4-6). These
circuit breakers protect different parts of the system. Under
normal operating conditions, these circuit breakers should remain
on. Under normal circumstances you will not have to take any
action with the circuit breakers.
Figure 4-6
Key Switch
QDR 4500 Power Module
Side Panel Indicator

Instrument X-Ray Enable ON


Computer Power
Power
OFF
CB1
10A
Computer

CB2
4A
24/28VDC Supply

CB3
1A
n15VDC Supply

3A CB4
24VDC

CB5
10A
28VDC

X-Ray Supply Pedestals X-Ray Emission Ind.


CB6 CB7 CB8 CB9 15VAC max 3A

6A 3A 3A 3A

KP0886-0401

4-14 Performing Daily System Operations


QDR Reference Manual

The functions of the switches and indicators are:


Table 4-6 Description Function
QDR 4500 Power Module
Components Computer Power Switch Turns the computer, monitor, and
printer ON and OFF.
Instrument Power Turns the QDR 4500 instrument ON
Switch and OFF.
X-ray Enable Key Enables or disables X-ray generation.
X-ray Enable Lamp When lit, indicates that X-rays can be
generated.
Circuit breakers Only used by the operator if there are
problems with the power supply to
different areas of the system.
X-ray Emissions Ind. You can use this outlet to enable an
indicator light (an X-ray warning light)
such as a red light bulb.

QDR 4500 Power Module Rear Panel


Note: Refer to page 4-20 for more The main power switch (CB1) is located on the rear panel of the
information about operator power module. It is generally used only after a power failure, to
actions during a power failure.
turn the system OFF and then ON.
Figure 4-7
Power Module Rear
Panel

CB1
Main
Power

Main
Power Power 115VAC max 10A
Indicator Cord To Power Strip
KP0886-0402

QDR 4500 Control Panel


Note: The QDR 4500C and The Control Panel, located on the right of the examination table
QDR 4500W Control Panels do unit, allows you to direct the movement of the C-arm and table, and
not include an Enable Lateral
turn on the laser. This means you can move the C-arm to correctly
or Emergency Table switch.
position the laser. You can also position the C-arm and table so that
the patients can easily get on or off the table.

Performing Daily System Operations 4-15


QDR Reference Manual

QDR 4500C and QDR 4500W Control Panel


Figure 4-8
Table
QDR 4500C and W Center Patient
Arm Laser Table On/Off Stop
Control Panel

X-Ray

KP0886-0105

This Control Panel has a series of buttons to move the table, C-


arm, and turn the laser cross-hair indicator on or off. Indicator
lights verify that components are on or enabled (see Figure 4-8
and Table 4-7). An Emergency Stop button allows the operator to
halt the scan at any point that the operator feels that the patient’s
safety is in jeopardy.
Table 4-7 Switch/Lamp Function Lamp Indicators
QDR 4500C and W Control
Panel Switches and Lamps Arm ← → Moves the C-arm left or Light ON = C-arm is
right. moving either left or
right.
Table ↑↓ Moves the table in or Light ON = Table is
out moving in or out.
Laser Turns the laser cross- Light Blinking ON
hair indicator on or off = The laser is on. It
will automatically
turn off after several
seconds.
Center Table Moves the table and C- Light Blinking ON
arm to the center =Table and C-arm
position moving to the center
position.
Patient On/On Moves the table and C- Light Blinking ON
arm to the patient on/off = Table and C-arm
position moving to the
patient on/off
position.
X-ray on Lamp Light ON = X-rays
being produced.

4-16 Performing Daily System Operations


QDR Reference Manual

Emergency Stop Pressing button stops


button scanning immediately.
Pressing the button
down and turning it to
the right resumes
normal operation.

QDR 4500A and QDR 4500SL Control Panel


This Control Panel has a series of buttons to move the table, C-
arm, and turn the laser cross-hair indicator on or off (see Figure 4-
9 and Table 4-8). This control panel also includes an Enable
Lateral button to move the C-arm and table into position for the
acquisition of the AP/Lateral Spine scan. Indicator lights verify
that components are on or enabled. There are two emergency stop
buttons. One button shuts down the system whenever the operator
feels the patient’s safety is in jeopardy. The other stops table
movement.
Figure 4-9
Table
QDR 4500A and SL Enable Center Patient Emergency
Arm Laser Lateral Table On/Off Table Stop
Control Panel

X-Ray

KP0886-0404

Table 4-8 Switch/Lamp Function Lamp Indicators


QDR 4500A and SL Control
Panel Switches and Lamps Arm ← → Moves the C-arm left or Light ON = C-arm is
right. moving either left or
right.
Table ↑↓ Moves the table in or Light ON = Table is
out. moving in or out.
Laser Turns the laser cross- Light Blinking ON
hair indicator on or off. = The laser is on. It
will automatically
turn off after several
seconds.

Performing Daily System Operations 4-17


QDR Reference Manual

Enable Lateral Enables the C-arm and Light Blinking ON


table to move for the = You need to
acquisition of the AP/ perform a task as
Lateral Spine scan. The instructed.
system instructs you
how and when to use
this switch.
Center Table Moves the table and C- Light Blinking ON
arm to the center =Table and C-arm
position. moving to the center
position.
Patient On/Off Moves the table and C- Light Blinking ON
arm to the patient on/off = Table and C-arm
position. moving to the
patient on/off
position.
Emergency Moves the table up Light ON = Enabled
Table when pressed. when you press the
Emergency Stop
switch.
X-ray on Lamp Light ON = X-rays
being produced.
Emergency Stop Pressing button stops
button scanning immediately.
Pressing the button
down and turning it to
the right resumes
normal operation.

Turning QDR 4500 Power On and Off


Note: The Instrument (Figure 4- These procedures are performed at the beginning and end of each
6) and Main Power Switch work day.
(Figure 4-7) should always
remain on.

Beginning of the Workday


Each of the steps that you complete at the beginning of the
workday, or whenever you need to restart the system, are critical
to the correct functioning of the system. For example, turning the
computer system on without enabling X-ray generation causes the
system to power up as a workstation, which makes it incapable of
scanning.

4-18 Performing Daily System Operations


QDR Reference Manual

Turning the QDR 4500 System On:


Perform the following steps at the start of each workday to turn
the system on and enable X-ray generation:
Step Action
1 On the rear panel, verify that the Power indicator is lit (see
Figure 4-7).
This light indicates that the system is in standby mode.
2 On the right side panel, verify that the Instrument Power
switch is in the ON position (see Figure 4-6).
3 On the right side panel, turn the Computer Power Switch
to the ON position.
This switch turns on the computer, monitor, table, and
printer.
4 On the right side panel, turn the X-ray Enable key
clockwise to enable X-ray production (see Figure 4-6).
The X-ray Enable indicator light, when lit, indicates the
system can produce X-rays.

End of the Workday


At the end of workday, verifying that the instrument and main
power switches remain in the ON position is important. When
you turn off the computer and disable the X-ray key, the
instrument power goes into a standby mode. Standby mode means
that power is maintained to the X-ray detector. This eliminates
warming up the detector when the system is turned on which can
take a considerable amount of time.
Turning the QDR 4500 System Off:
Perform the following steps, in the order shown, at the end of
each workday to turn the system off:

Step Action
1 On the main window, click the Exit button.
The system displays the following message:

KP0886-0405

4-19
QDR Reference Manual

Step Action
2 Select “Exit QDR with shutdown” and click OK.
The system displays a message indicating that it is OK to
shut the computer down.
3 On the right side panel of the Power Module, turn the X-
ray Enable key counterclockwise (vertical) to disable
production of X-rays (see Figure 4-6).
The X-ray enable indicator light goes off when you turn
the key in the disabled (upright) position.
Note: Do not turn the Instrument 4 On the right side panel, turn the Computer Power switch
Power Switch off. to the OFF position (see Figure 4-6).
This shuts off the computer, table, monitor, and printer.
Note: Leaving the Instrument 5 On the rear panel, verify that the Power indicator light is
Power Switch on eliminates lit (see Figure 4-7).
warming up the detector when the
This light indicates that the system is in standby mode and
system is turned back on.
power is maintained to the X-ray detector.

QDR 4500 Emergency Procedures


During operation of the QDR System there are three emergency
situations that may require operator action:
• A facility power failure occurs
• Equipment fails while in operation
• System is turned off

Operator Action During a Power Failure


In the event that your facility experiences a power failure, it is
important to unplug the QDR system from the wall outlet even
when you have used the switch to turn it OFF. This protects it
from damage due to power surges when power is restored.

Shutting down during a power failure


Step Action
1 On the side panel, turn the X-ray Enable key
counterclockwise to the OFF position (see Figure 4-6).
2 On the side panel, turn the Computer Power switch to the
OFF position.
3 Unplug the QDR system from the wall outlet.

4-20
QDR Reference Manual

After power is restored


Step Action
1 When power is restored, plug the QDR system into the
wall outlet.
2 On the side panel, turn the Computer Power switch to the
ON position.
3 On the side panel, turn the X-ray Enable key clockwise to
enable X-ray production (see Figure 4-6).
The X-ray Enable indicator, when lit, indicates the key is
in the proper position to generate X-rays.
4 Allow the system to warm up for the same amount of time
that the power was off or for one hour, whichever is
shorter.
Note: Refer to Chapter 5 for 5 Perform the daily QC procedure to verify that the system
detailed instructions for is functioning properly.
performing daily QC.

If QDR 4500 Equipment Fails While in Operation


If the operator determines that the equipment has malfunctioned
(e.g., X-Ray is on and C-arm is not moving, equipment sparks or
smokes, etc.) perform the following procedure.
Emergency Stop Procedure:
Step Action
1 On the Control Panel, press the red Emergency Stop
Button.
The table and the C-arm immediately stops moving, and
the X-rays and laser turn off.
2 Assist the patient off the table.
Note: The power should not be 3 Go to the back of the Power Module (Figure 4-7) and turn
restored without consulting off the Main Circuit Breaker.
Hologic.
4 Remove power to the Power Module (if possible).
5 Call Hologic customer service.

If QDR 4500 AC Power has been Turned Off


If for any reason the circuit breaker on the Power Module
(Figure 4-7) has been turned off (not because of an equipment
failure), or the plug has been pulled from the wall, use the follow-
ing procedure to apply power to the system:

4-21
QDR Reference Manual

Applying power to the system:


Step Action
Note: If the Power Module Power 1 On the Power Module, place the Main Circuit Breaker
Indicator does not turn on, make circuit breaker in the 1 (on) position.
sure that it has power (that is, the
The Power Module Power Indicator illuminates.
cord is plugged into the wall). If it
is, call Hologic customer service. 2 Perform the procedure titled “Turning QDR 4500 Power
On and Off” on page 4-18.

4-22
QDR Reference Manual

Delphi Systems
Delphi AC Power Isolation Module
The AC Power Isolation Module contains a power transformer,
circuit breaker and power indicator (see Figure 4-10). Table 4-9
gives the description and function of the switches and indicators.
Figure 4-10
AC Power Isolation Module

KP0886-0406
Circuit Breaker Indicator
.
Table 4-9 Description Function
AC Power Isolation Module
Circuit Breaker Turns the power to the Delphi System
off (0) and on (1).

Indicator Glows green when the circuit Breaker


is in the on (1) position.

Delphi Control Panels


Delphi-C and -W Control Panel
The Delphi-C and -W Control Panel (see Figure 4-11) is located
on the front of the C-arm. It contains the controls and indicators
that allow the operator to move the table and the C-arm to
properly position the patient for a scan. Table 4-10 describes the
switches, indicator lights and Emergency Stop Button on the
control panel.

4-23
QDR Reference Manual

Figure 4-11 Emergency


Delphi-C and -W Stop Button
Control Panel

KP0886-0107

Table 4-10 Switch/Lamp Function Lamp Indicators


Control Panel Switches and
Indicators Table In Moves the C-arm left or
right.

Table Out Moves the table toward


the front.

Arm Right Moves the C-arm to the


right (towards the head
of the table).

Arm left Moves the C-arm to the


left (towards the foot of
the table).

Laser Turns the laser cross- When the laser is


hair indicator ON or ON the Laser On/
OFF. Motion Enabled
green indicator
blinks.

Patient On/Off The table moves to the


front-most position and
the C-arm moves to the
left-most position.

Center Table Moves the table and C-


arm to the center
position.

X-Ray On When X-rays are


(amber indica- being produced this
tor) indicator is ON.

4-24
QDR Reference Manual

Laser On/ When this indicator


Motion Enabled is blinking, the laser
(green indicator) is on;
when this indicator
is not blinking, the
table and C-arm can
be moved.

Instrument On/ Turns the power to the When the indicator


Off instrument on or off. is green, power is
ON in the
instrument.

Emergency Stop Pressing this button


Button stops scanning
immediately. Pressing
the button down and
turning it to the right
resumes normal
operation.

Delphi-A and -SL Control Panel


The Delphi-A and -SL Control Panel (see Figure 4-12) is located
on the right front of the examination table. It contains the controls
and indicators that allow the operator to move the table and the C-
arm to properly position the patient for a scan. Table 4-11
describes the switches, indicator lights and Emergency Stop
Button on the control panel.
Figure 4-12
Delphi-A and -SL Patient X-Ray EMERGENCY
On/Off
Control Panel Table Table Lift Stop Power

A A
R R Center
M M Table

Table

Enable R

Laser
Lateral

KP0886-0108

Table 4-11 Switch/Lamp Function Lamp Indicators


Delphi-A and -SL Control
Panel Switches and Lamps Arm Moves the C-arm left or Light ON = C-arm is
right. moving either left or
right.

4-25
QDR Reference Manual

Table Moves the table in or Light ON = Table is


out. moving in or out.
Laser Turns the laser cross- Light Blinking ON
hair indicator on or off. = The laser is on. It
will automatically
turn off after several
seconds.
Enable Lateral Enables the C-arm and Light Blinking ON
table to move for the = You need to
acquisition of the AP/ perform a task as
Lateral Spine scan. The instructed.
system instructs you
how and when to use
this switch.
Center Table Moves the table and C- Light Blinking ON
arm to the center =Table and C-arm
position. moving to the center
position.
Patient On/Off Moves the table and C- Light Blinking ON
arm to the patient on/off = Table and C-arm
position. moving to the
patient on/off
position.
Emergency Moves the table up Light ON = Enabled
Table Lift when pressed. when you press the
Emergency Stop
switch.
X-ray on Lamp Light ON = X-rays
are being produced.
Emergency Stop Pressing button stops
button scanning immediately.
Pressing the button
down and turning it to
the right resumes
normal operation.
Power Switch Turns equipment power
on and off.

Turning Delphi Power On and Off


The procedure to turn power on and off the Delphi System is
described below.

4-26
QDR Reference Manual

Do not use the circuit breaker on the AC Power Isolation Module


(Figure 4-10) to turn power on or off. If the indicator light on the
AC Power Isolation Module goes out while the Delphi System is
running or if the light is not on at the beginning of the day, refer to
the Emergency Power Procedure below.

Beginning of the Workday


Perform the following steps at the start of each workday to turn
the Delphi System on.

Step Action
Note: If the AC Power Isolation 1 Verify that the indicator on the AC Power Isolation
Module light is not on, refer to Module is on (Figure 4-10).
Emergency Power Procedures
This light indicates that the system is receiving AC power.
below.
2 On the Control Panel, press the Instrument On/Off
(Power) button.
The Instrument On/Off (Power) green indicator
illuminates.
Note: Because the computer and 3 On the Operator Console, turn the computer on.
peripherals may vary, refer to The monitor (in power save mode) and printer(s) should
documentation that came with the
already be on during a normal day’s power-up.
Delphi System for controls and
indicators on the computer, The computer boots itself and starts Windows. It then
monitor and printers. starts the APEX software automatically. If there are any
problems, warning messages will tell you what to do.

End of the Workday


Perform the following steps at the end of each workday to turn the
Delphi System off.

Step Action
1 Make sure only the Delphi main window appears on the
monitor.
This is to ensure that you are not in the middle of any
function and all activity has stopped.
2 Click Exit.
The Exit button is in the lower right-hand corner of the
screen.
KP1035A-0325

4-27
QDR Reference Manual

Step Action
3 In the Exit QDR System dialog box that appears, select
“Exit QDR with shutdown?” and click OK. (See below.)

KP0886-0405
This shuts down the APEX software, and the Windows
operating system
4 If the computer did not automatically shut itself down,
press the power on/off switch on the computer.
5 On the control panel press the button to turn power off:
• Instrument On/Off button on the -C and -W
control panel (see Figure 4-11)
• Power button on the -A and SL control panel (see
Figure 4-12)
Important: Do not turn off the The green indicator in the Instrument On/Off (Power)
circuit breaker on the AC Power switch goes out when power to the instrument is off.
Isolation Module. The green
indicator should remain on. Do not turn off the printer(s).
Do not turn off the monitor. It goes into power save mode
when the computer shuts down.

Delphi Emergency Procedures


During operation of the Delphi System there are three emergency
situations that may require operator action:
• A facility power failure occurs
• Equipment fails while in operation
• System is turned off

Operator Action During a Power Failure


If the facility power fails, it is a good idea to turn off all the
equipment. When the power returns, frequently it is not stable and
can spike up and down for a few moments. You should wait a few
minutes after power returns to turn on the equipment.

4-28
QDR Reference Manual

Shutting Down During a Power Failure


Step Action
1 If the Delphi was in operation when the power failure
occurred, assist the patient from the table
Remember that the Patient On/Off switch is inoperative at
this time.
Note: Refer to documentation 2 Turn the circuit breakers on the AC Power Isolation
that came with the Delphi System Module (see Figure 4-10) off.
for controls and indicators on the
The green indicator to the right of it goes off.
computer, monitor and printer.

After Power is Restored


Step Action
1 After waiting a few minutes for the facility power to
stabilize,
Turn the circuit breaker on the AC Power Isolation
Module (Figure 4-10) on.
The green indicator light to the right of it goes on.
2 Turn the computer on.
3 Perform the procedure “Beginning of the Workday” on
page 4-27.

If Delphi Equipment Fails While in Operation


If the operator determines that the equipment has malfunctioned
(e.g., X-Ray is on and C-arm is not moving, equipment sparks or
smokes, etc.) perform the following procedure.
Emergency Stop Procedure:
Step Action
1 On the Operator Control Panel, press the red Emergency
Stop Button.
The table and the C-arm immediately stop moving, and the
X-rays and laser turn off.
2 Assist the patient off the table.
Note: The power should not be 3 Go to the AC Power Isolation Module (Figure 4-10) and
restored without consulting turn off the circuit breaker.
Hologic.
4 Remove the power the AC Power Isolation Module (if
possible).
5 Call Hologic customer service.

4-29
QDR Reference Manual

If Delphi AC Power has been Turned Off


If for any reason the circuit breaker on the AC Power Isolation
Module (Figure 4-10) has been turned off (not because of an
equipment failure), or the plug has been pulled from the wall, use
the following procedure to apply power to the Delphi System:
Applying power to the system:
Step Action
Note: If the AC Power Isolation 1 On the AC Power Isolation Module, place the circuit
Module Indicator does not turn breaker in the 1 (on) position (see Figure 4-10).
on, make sure that it has power
The AC Power Isolation Module Indicator illuminates.
(that is, the cord is plugged into
the wall). If it is, call Hologic 2 Perform the procedure “Beginning of the Workday” on
customer service. page 4-27.

4-30
Chapter 5
Quality Control
The QDR system uses the daily Quality Control (QC) procedure to
ensure that the system and software are performing properly prior to
scanning patients. Since Bone Mineral Density (BMD) changes slowly
in humans, it is essential that measured BMD changes are attributable
to biological change rather than densitometer drift.
The QC procedure must be performed once per day prior to using the
system to scan patients. This chapter provides information and
instructions for performing the QC procedure.
This chapter does not include the QC procedures for Step Phantom
Calibration for Body Composition. Refer to Appendix A for
information on this Option.

Content
Subject Page
The Daily QC Procedure 5-2
Start Daily QC 5-2
Position the Spine Phantom 5-3
Automatic System Test 5-4
Auto QC Passed 5-5
Auto QC Failed 5-9
About the QC Spine Phantom 5-10
QC Data Management 5-11
Plot Regression 5-20
Archiving QC Phantom Scans 5-20

5-1
QDR Reference Manual

The Daily QC Procedure


Important: This procedure must Daily QC is used to verify that the QDR system meets system
be performed each day prior to specifications.
scanning patients.
The Daily QC procedure includes all activities that begin when
you start QC, and ends when the QC plot has been evaluated and
determined to be acceptable. The Daily QC procedure takes
approximately 2 minutes and consists of the following steps:

• Start QC
• Position the Spine Phantom
• Automatic System Test
• Automatic Scan and Analysis of the Spine Phantom
• Add the QC Scan to the Plot
• Evaluate the QC Plot
• Exit QC

Start Daily QC
To start the daily QC procedure click on the Daily QC button in
the main window.
The system displays the “Daily QC Setup” window, prompting
you to place the specified phantom on the table.

KP1035A-0307

Note: Alternatively, you can select


Daily QC from the QC pull-down
menu in the main window to start
the daily QC procedure.

KP0886-0502

5-2 Quality Control


QDR Reference Manual

Position the Spine Phantom


To position the spine phantom on the table refer to Figure 5-1 and
Figure 5-2 and follow the procedure below.

Figure 5-1
Spine Phantom Registration
Mark A

Figure 5-2
Spine Phantom and Laser
Cross Hair Position
B

Quality Control 5-3


QDR Reference Manual

To position the spine phantom on the table:


Step Action
1 Place the spine phantom on the table with the registration
mark (see "A" in Figure 5-1) to the left, foot end.
Note: Use the horizontal line of 2 Position the phantom parallel to the back of the table.
the laser to confirm that the
phantom is straight on the table
3 Align the laser cross hair (see "B" in Figure 5-2) with the
registration mark.
4 Once the spine phantom is properly positioned, click the
Continue button.
The systems starts performing an automatic system test.

Automatic System Test


The QDR system automatically performs a system test to verify
proper operation of its X-ray subsystem prior to scanning the
spine phantom. The following screen is displayed during system
test:

KP0886-0504

5-4 Quality Control


QDR Reference Manual

If the automatic system test succeeds, a message appears on the


screen indicating that the system test passed.

KP0886-0505

If the automatic system test fails, the system displays a message


stating that the test failed, and information on how to correct the
error.

KP0886-0506

To correct a failed system test:


Step Action
1 Read the message in the Error window.
Note: The error message varies 2 Perform the steps as instructed in the Error window to
depending on why the system test correct the error and repeat the QC procedure.
failed.

Auto QC Passed
Upon successful completion of the system test, the system runs
Auto QC. When Auto QC has been completed, a window appears
giving either a passed or failed message.
When Auto QC passes, the message is “Daily QC has passed”
(see Figure 5-3).

Quality Control 5-5


QDR Reference Manual

Figure 5-3
QC Results Window for Auto QC
Passed

KP0886-0507

The QC Results window for a passed Auto QC contains 3


buttons: Review Analysis, Plot, and OK.

Note: If QC passes, performing If Auto QC passed, click on OK to start the daily operation of the
Review Analysis and/or Plot is System.
not required.
The Review Analysis button and the Plot button are described
below (for informational purposes).

Review Analysis
To review the analysis of the Auto QC scan, click on the Review
Analysis button from the QC Results window for a passed Auto
QC. The Analysis window appears (see Figure 5-4).
Figure 5-4
Auto QC Analysis Window

KP1035-0507a

Verify that the phantom has been properly positioned and that the
scan was properly analyzed. If these items appear to be
satisfactory, click Close to continue.
If not, click Close and repeat the procedure, making sure to
properly position the phantom.
From the QC Results window (see Figure 5-3) click OK to start
the daily operation of the System.

5-6 Quality Control


QDR Reference Manual

QC Plot
To review the QC plot, click on the Plot button from the QC
Results window for a passed Auto QC. The QC Plot window
appears (see Figure 5-5).
Figure 5-5
QC Plot Window

KP1035-0507b

When the QC Plot window is displayed, the BMD values are


shown along the y axis (left side) of the QC plot. The example
above shows the mean value at calibration time (green horizontal
center line) as 0.987.
The BMD tab is displayed by default. To show BMC or Area
values, select the appropriate tab by clicking on it.
The QC Plot Window contains four buttons

Button Function
<< Back Returns to the Auto QC passed window
(see Figure 5-3).
Plot Regression Refer to Plot Regression on page 5-20
Print Prints the QC Plot
Close Returns to the system main window

About the QC Plot


A QC plot is a graph of the results of a series of scans of the same
QC Spine Phantom over time. The graph shows a number of
statistical indexes designed to help you to determine if there have
been significant changes in the measurements generated from the
scans.
Scan Dates
Scan dates are displayed along the horizontal axis. The example
above shows the scan dates from August, 2000 to October, 2002.

Quality Control 5-7


QDR Reference Manual

Data Points
The small green circles are called data points. One data point
displays for each analyzed QC scan that is included in the QC
plot. The most recent data point is shown as a filled in green
circle. Any data point that falls off the graph (above or below) is
displayed as a red cross at the top (for data point above the scale)
or bottom (for data point below the scale) of the graph.
Mean Value at Calibration
The green horizontal line indicates the mean value at calibration.
This value is set by Hologic and is located at the mean value of
the measured parameter (in this case BMD) when your system
was calibrated. It is used as a guideline to compare the results of
your Daily QC scans to your QDR system when it was last
calibrated.
QC Limit Values
Note: An obvious drift, or points The red dashed lines represent the upper and lower QC limits for
outside the limit lines, should be your system. The limits are set by Hologic and are located at ±
reported to Hologic.
(plus or minus) 1.5% of the mean value at calibration

Setup
The setup area shows the type of QC scan performed, phantom
type (Spine, Block or Hip) and ID, and system serial number.
Reference Values
The Reference Values area shows the values that were
determined during the most recent database setup. It includes the
limits, the mean BMD, and the standard deviation (SD) for the
mean.
Plot Statistic
The Plot Statistics area shows the following:
• Number of Points
Number of scans contained in the QC plot.
• Mean
The mean is the average of all QC scans plotted.
• S.D.
The S.D. is the Standard Deviation from the mean for the
QC scans in the plot. The standard deviation indicates the
spread of plotted values from the mean.
• CV
The CV is the coefficient of variation for the values in the

5-8 Quality Control


QDR Reference Manual

graph. The smaller the CV percentage, the better the


precision of the results.

Auto QC Failed
If Auto QC fails, the message is “Daily QC failed” and a series of
steps is provided as shown in the example in Figure 5-6.
Figure 5-6
QC Results Window for Auto QC
Failure

KP1035-0508

If Auto QC fails, the QC Results Window contains four buttons:


Details, Review Analysis, Plot and OK.

Details
Clicking Details provides additional information about the QC
failure as shown in the example in Figure 5-7.

Figure 5-7
QC Results Details Window

Note: If the phantom is


positioned correctly and QC
fails a second time, contact
Hologic Customer Service.

KP1035-0508a

Review Analysis
Clicking the Review Analysis button displays an analysis
window similar to the one shown in the example in Figure 5-4.
From this window you will be able to verify that the phantom has
been properly positioned and that the scan was properly analyzed.
Click Close to return to the QC Results Window.

Quality Control 5-9


QDR Reference Manual

Plot
Clicking Plot from the QC Results window for a failed Auto QC
displays the Analysis window of the QC Plot (similar to the one
in Figure 5-5).

OK
Clicking OK, from the QC Results window for a failed Auto QC,
returns to the system main window so Daily QC can be re-run.

About the QC Spine Phantom


Your QDR system includes a DXA spine phantom that is used
during the Daily QC procedure. It is used to test the system’s
calibration and precision. The phantom is designed to produce
scan data that resembles patient data. Other phantom types are
available from Hologic.
Each QC Spine Phantom is assigned a name in the patient name
field. The information about the QC Spine Phantom is stored in
the patient database. If a patient name for the QC Spine Phantom
does not exist, the daily QC procedure will create a record.
The format of the patient name for the QC Spine Phantom is as
follows:

Phantom Number

<Type> Phantom # < Phantom ID>

Type: Spine, Block, or Hip KP0886-0512

5-10 Quality Control


QDR Reference Manual

QC Data Management
Occasionally there may be a need to manage the QC data
contained in your QDR system’s database. This section provides
information and instructions for managing the quality control
data. It includes the following:
• Access Data Management
• Set QC plot parameters
• Select scans to be included in the QC plot
• Create a new phantom record
• Set up the QC plot

Access Data Management


Follow the procedure below to gain access to the QC data
management functions.
To access QC data management functions:
Step Action
1 In the main window, click on QC in the menu bar.
The system displays the QC sub-menu.
2 Click on the QC Data Management sub-menu item to
display the QC Data Management sub-menu.
The system displays the QC Data Management functions
installed in your system, see below.

KP0886-0513

Set QC Plot Parameters


The “Quality Control Plot Parameters” window, shown below,
allows you set the parameters needed for the QC Plot graph.

Quality Control 5-11


QDR Reference Manual

KP0886-0514

The following table describes the fields in the Quality Control


Plot Parameters window.

Field Description
QC Plot Setup Indicates which QC Spine Phantom to
include in the QC plot. Choose from a list
of valid QC Spine Phantoms used with your
system.
Note: The North American format Start Date The date of the first scan you want to
is mm/dd/yyyy; the International include in the QC plot.
format is dd.mm.yyyy (for
example: 07/04/2002 is July 4,
2002; 04.07.2002 is 4 July 2002).
Leaving the date fields blank
results in all QC scans being
included in the QC plot.
End Date The date of the last, or most recent scan you
want to include in the QC plot.
Plot button Generates a QC plot based on the
parameters you specified.
Cancel Button Cancels the operation and returns you to the
main window.
To set plot parameters:
Step Action
1 In the main window, click on QC in the menu bar.
The system displays the QC sub-menu.

5-12 Quality Control


QDR Reference Manual

Step Action
2 Click on the QC Data Management sub-menu item to
display the QC Data Management sub-menu. Click on
Plot.
The system displays the “Quality Control Plot Parameters”
window.
3 To change which QC Spine Phantom is used to plot the QC
graph, click on the down arrow in the QC Plot Setup field
and select the QC Spine Phantom.
Note: If you include a large span 4 To change the Start and/or End dates, enter the starting and
between dates, or leave the date ending dates using the following format:
fields blank, the graph may be
difficult to read, as it will include a mm/dd/yyyy (North America) or dd.mm.yyyy
large number of data points. (International).
If you leave both dates blank, all QC scans stored on your
system will display on the plot. If you leave the End date
blank, all scans from the start date on will display.
5 Click the Plot button to display the QC plot with the new
parameters. Click the Cancel button to cancel the operation
and not save any changes to the Plot Parameters.

Select scans to be included in the QC Plot


The “Select Quality Control Scans” window, shown below,
allows you select the scans to be available for plotting in the QC
Plot graph.

KP0886-0515

Quality Control 5-13


QDR Reference Manual

The following table describes the fields in the “Select Quality


Control Scans” window.

Field Description
Included Scans Lists scans that will be used to calculate the
baseline data for the QC plot.
Excluded Scans Lists scans that will not be included in the
calculations to determine the baseline.
Include Scans Use this button to move selected scans from
button the Excluded list area to the Included list
area.
Exclude Scans Use this button to move selected scans from
button the Included list area to the Excluded list
area.
All QC scans performed on your QDR system will be
automatically listed under scans to be included for plotting of the
QC Plot graph. You should only exclude scans that you know to
be invalid from this list.
To include or exclude scans:
Step Action
1 In the main window, click on QC in the menu bar.
The system displays the QC sub-menu.
2 Click on the QC Data Management sub-menu item to
display the QC Data Management sub-menu. Click on
Select Scans.
The system displays the “Select Quality Control Scans”
window.
Note: To select multiple scans, 3 To include scans, select each scan from the “Excluded”
hold the Ctrl key while clicking on section of the window and click the Include Scans button.
each scan. To select a range of
scans, click on the first scan in the
range, and hold the Shift key while
clicking on the last scan in the
range.

4 To exclude scans, select each scan you want to exclude


from the “Included” section of the window and click the
Exclude Scans button.
5 Make sure there are at least twenty scans listed in the
Included area.

5-14 Quality Control


QDR Reference Manual

Step Action
6 Click the OK button when done to generate the baseline
data for QC plot. Click the Cancel button to cancel the
operation.

Exclude Data Points from the QC Plot Screen


In addition to the method described above, QC scans can also be
excluded directly from the Quality Control Plot screen. Only
scans that you know to be invalid should be excluded.
To exclude scans from the QC Plot screen:
Step Action
1 From the QC Plot screen, click on the data point to be
excluded,
or,
click and drag a box around the data points to be excluded.
The system displays the Scan Information screen.

KP0886-0516

Note: Only scans that you know to 2 Highlight the scan, or scans, to be excluded by clicking on
be invalid should be excluded. them.
Note: Excluded scans are not
deleted from the database.
3 Click OK to exclude the scans.

Create a New Phantom Record


This section provides information and instructions for creating a
new phantom record in the patient database. Every phantom used

Quality Control 5-15


QDR Reference Manual

in the QC procedures should have only one record describing it in


the patient database.
The “Add Quality Control Phantom” window, shown below,
allows you add new phantoms to the patient database.
The following table lists and describes the fields in the “Add
Quality Control Phantom” window.

KP0886-0517

Field Description
Phantom Type Indicates the type of phantom. The drop
down list includes standard types: Spine
Phantom, Hip Phantom, and Block
Phantom as well as non-standard phantom
types: Other1 Phantom, Other2 Phantom,
and Other3 Phantom. The phantom type
you select appears in the Phantom Name
area at the top of the window.
Phantom Number Indicates the number or name of the
phantom. A valid string can include
numbers and letters. The number (name)
you enter appears in the Phantom Name
area at the top of the window.
Comment Stores any comments you enter about the
phantom in the patient database.
To add a new QC Spine Phantom:
Step Action
1 In the main window, click on QC in the menu bar.
The system displays the QC sub-menu.

5-16 Quality Control


QDR Reference Manual

Step Action
2 Click on the QC Data Management sub-menu item to
display the QC Data Management sub-menu. Choose New
Phantom.
The system displays the “Add Quality Control Phantom”
window.
3 Click the down arrow in the Phantom Type field and select
the type of phantom you want to add.
4 Enter the name of the phantom in the Phantom Number
field.
If the name you enter already exists in the patient database,
the system displays an error message. Respond to the error
message and enter a unique name in the Phantom Number
field.
5 Enter any relevant comments in the Comment field.
6 Click the OK button to add the new phantom to the patient
database, or click the Cancel button to return to the main
window without adding a new phantom.

Set up the QC Plot


This section provides information and instructions for specifying
the phantom, scan type, and serial number that will be used in
plotting. This information generates the baseline data used in the
QC plot.
Setting up a QC plot involves two steps:
1. Specifying data in the “Quality Control Plot Setup” window
2. Including or excluding scans in the “Select Quality Control
Scans” window
Use the “Quality Control Plot Setup” window, shown below, to
begin setting up the QC plot.

KP0886-0518

Quality Control 5-17


QDR Reference Manual

The following table lists and describes the fields in “Quality


Control Plot Setup” window.
Field Description
QDR Serial Number Indicates the system’s serial number.
Phantom Name Lists phantoms used in analyzed QC
scans performed on the system
specified by the serial number.
Changing the system’s serial number
changes the list of phantoms.
Scan Type Lists scan types that have been used in
analyzed QC scans of the phantom
specified in the Phantom Name field.
Changing the phantom name changes
the scan types for the plot setup.
When you finish selecting information in the “Quality Control
Plot Setup” window, click the Next button to display the “Select
Quality Control Scans” window, shown below.

KP0886-0519

This window allows you select QC scans to be included in the QC


Plot. The scans listed are based upon the criteria you selected in
the previous window. Only scans matching the criteria you
entered displays. For example, the window above only shows
scans matching the following criteria:
QDR Serial Number = 3009
Phantom Name = SPINE PHANTOM #6124
Scan Type = a Lumbar Spine

5-18 Quality Control


QDR Reference Manual

The following table lists and describes the buttons in “Select


Quality Control Scans” window.

Field Description
Select All button Selects all scans contained in the list of
QC scans for the system and phantom
you selected.
Deselect All button Deselects all selected QC scans.
Finish button Instructs the system to calculate QC
data associated with the selected scans
for display in QC Plot.
Cancel button Cancels the operation and returns you
to the main window.
To setup a new QC plot:
Step Action
1 In the main window, click on QC in the menu bar.
The system displays the QC sub-menu.
2 Click on the QC Data Management sub-menu item to
display the QC Data Management sub-menu. Click on
Setup Plot.
The system displays the “Quality Control Plot Setup”
window.
3 Verify the system’s serial number in the QDR Serial
Number text entry area.
4 Click the down arrow in the Phantom Name field, and
select the phantom you want to use to generate baseline
data.
5 Click the down arrow in the Scan Type field, and select the
scan type you want to use to generate baseline data.
6 Click the Next button to continue.
The “Select Quality Control Scans” window is displayed.
Note: A minimum of twenty scans 7 Select the scans to be used for the QC plot.
are recommended for sufficient • Use the Select All button to select all QC scans.
statistical information to set up a
new plot. • Use the Deselect All button to deselect all selected
Note: To select multiple scans, scans
hold the Ctrl key while clicking on 8 Click the Finish button when done so that the system can
each scan. To select a range of calculate data for display in QC Plot. Click the Cancel
scans, click on the first scan in the
range, and hold the Shift key while
button to cancel the operation.
clicking on the last scan in the
range.

Quality Control 5-19


QDR Reference Manual

Plot Regression
The QDR system provides a way to perform linear regression and
display it on the QC plot.
To display linear regression on the Quality Control Plot, click the
Plot Regression button. Slope is shown in the lower portion of
the graph.
The example below shows a BMD linear regression plot.

Figure 5-8
Regression Plot Example

KP0886-0521

The linear regression option calculates a least squares fit that


represents the change in the plotted QC parameter with time. The
slope of the line is calculated and displayed.
Regression can also be performed on BMC and Area plots by
selecting that tab and clicking the Plot Regression button.

Archiving QC Phantom Scans


The QC Phantom Scan provides information on how the QDR
system is functioning. Having access to these scans allows
Hologic to verify the operation of your system. All QC Phantom
scans should be archived when you archive the patient scans (see
“Archive Process” on page 16-8 of Chapter 16).

5-20 Quality Control


Chapter 6
Managing Patient Records
This chapter provides information about accessing the patient and scan
records in the QDR system.

Content
Subject Page
Introducing Patient Records 6-2
Working with Patient Records 6-4
Introducing Scan Records 6-16
Working with Scan Records 6-17
Adding Scan Information from Other 6-24
Manufacturer’s Systems

6-1
QDR Reference Manual

Introducing Patient Records


The QDR system stores patient and scan information in a database
that is also referred to as the filing cabinet. The three types of
records stored include:
• Patient Records
Each patient has a record of biographical and insurance
information. Recall this record from the Patient Drawer
each time a patient returns for a scan.
• Scan Records
A scan is data gathered from a patient, or phantom, during
each exam. The scan data, along with additional
information about the scan characteristics and the patient,
is stored in the Scan Drawer of the filing cabinet. These
can include both analyzed scans and unanalyzed scans.
Note: Refer to Chapter 14 for • Analysis Reports
more information about Reports.
When a scan is analyzed, a report can be generated, saved
(hardcopy), and retrieved later.

Accessing the Records


The main window provides access to records in the filing cabinet
several different ways. The method selected depends upon where
you are during an exam, or what types of tasks are being
performed. For example, when starting to acquire a scan, select
the Perform Exam button.
Or, if the scan has been performed but the analysis is not
complete, select the scan from the Scan Drawer by clicking on the
Scans button.
You can find and retrieve an archived scan (not stored on the
computer) by clicking the Locate Scans button. You can select a
patient record by clicking the Patients button.
The following table lists the methods of finding records in the file
cabinet.

6-2 Managing Patient Records


QDR Reference Manual

Note: Refer to Chapter 3 for The following table lists the methods of finding records in the file
additional information about using cabinet.
the buttons or menu on the main
window. When... And needed Then retrieve from...
is...
Performing a A patient Patient Drawer by:
new scan record • Clicking the Patients
button in the main window,
selecting the desired
patient, and dragging them
to the Perform Exam
button; or,
• Clicking the Perform
Exam button in the main
window; or,
• Choosing Exam, then
Perform Exam from the
menu bar at the top of the
main window.
Analyzing a The scan Scan Drawer by:
previously results • Clicking the Scans button
performed in the main window and
scan dragging the desired scan to
the Analyze Scan button at
the bottom of the main
window; or,
• Clicking the Analyze Scan
button in the main window;
or,
• Choosing Exam, then
Analyze Scan from the
menu bar at the top of the
main window.
Looking for a An archived Main window by:
scan that is scan • Clicking the Locate Scans
not stored on button
the computer

Managing Patient Records 6-3


QDR Reference Manual

Working with Patient Records


One important aspect of acquiring and analyzing scans includes
ensuring that you have a current patient record. This record
remains in the Patient Drawer so that it readily accessible each
time a scan is performed or analyzed, or if it needs to be updated.

Overview of Patient Record


The patient record is divided into two sections, the biography tab,
and the insurance tab.

Biography Tab

Field Description
Last Name Patient’s last name and suffix, if applicable (e.g.
III)
First Name Patient’s first name
MI Patient’s middle initial (no punctuation is
necessary)

6-4 Managing Patient Records


QDR Reference Manual

Field Description
Sex Click the drop-down list to choose F (Female) or
M (Male)
Ethnicity Click the drop-down list to choose Asian,
Hispanic, White or Black.
DOB Date of Birth. Click the arrows to select the
month, day and year.
Patient ID Medical record number/X-ray #/SS #
Identifier 2 Department number
Referring Click the drop-down list to choose name of the
Physician physician that referred the patient or type a new
name. Entering a new name stores that name on
the system.
Click Delete followed by YES to remove the
name from the patient’s record and from list of
physician names stored on the system.
Menopause Patient’s age at the onset of menopause
Age
Height/Weight Today’s height and weight. Body Mass Index is
automatically calculated as height and/or weight
is entered or changed.
Body Mass Automatically calculated from Height/Weight
Index entries.
Patient Information that is specific to the patient such as
Comment “Premenopausal study”

Managing Patient Records 6-5


QDR Reference Manual

Insurance Tab

KP0886-0602

Field Description
Name Name of the patient (filled in by the system)
Patient ID Patient ID (filled in by the system)
Plan Patient’s insurance plan
Group Patient’s insurance group
Insurance Patient’s insurance company
Address 1 Patient’s street address
Address 2 Additional address information, such as
apartment number or PO Box
City Patient’s city
State Patient’s two letter state code
Postal Code Patient’s zip code

6-6 Managing Patient Records


QDR Reference Manual

Retrieving a Patient Record


The QDR system allows the operator to retrieve a record several
different ways from the patient drawer. The method often depends
on the task being performed. The primary method involves using
the Patients button located on the main window.

KP1035A-0317

To go to the Patient Drawer:


Step Action
Note: Refer to the table on page 1 Click the Patients button in the main window.
6-3 in this chapter for suggestions The system displays the Patient Drawer window with a list
about alternative methods for
of existing patient records.
accessing the Patient Drawer.

Searching the Patient Drawer


Note: It is important to avoid Once you have accessed the patient drawer, you can search the
creating duplicate patient records. Patient Drawer for a record by one of the following methods:
Therefore, you should search for
the correct patient record before • Using the scroll bar
creating a new record.
• Using a search criteria such as patient’s last name, ID, or
date of birth.

Managing Patient Records 6-7


QDR Reference Manual

Search Field Scroll Bar

KP0886-0604
Patient Data

To search the Patient Drawer using the scroll bar:


Step Action
Note: An asterisk (*) in the 1 To move down through the list of patients, click on the
heading identifies the sort order. lower portion of the vertical scroll bar or on the down
arrow on the right of the patient window.
List of patients moves up.
2 To move up through the list of patients, click on the upper
portion of the vertical scroll bar or on the up arrow on the
right of the patient window.
List of patients moves down.
3 Select the patient’s name by clicking on it.

6-8 Managing Patient Records


QDR Reference Manual

Patient Name

KP0886-0605
Scroll Bar
To search the Patient Drawer using the search text field:
Step Action
Note: An asterisk (*) in the 1 To search by last name, click the correct heading;
heading identifies the sort order. Asterisk displays near heading title (i.e., Patient Name,
Patient ID, Birth Date, or whatever search criteria you
elect to use) and search criteria displays next to text field.
2 Point to the text field with the mouse, and click with the
left mouse button.
3 Type the search criteria.
Patient list moves, displaying those names that match the
search criteria.
4 Select the patient’s record by clicking on it.
5 To search by another criteria, began at step #1, clicking on
the heading (for example, Birth Date).
Asterisk displays near heading title.
6 Complete steps 2 through 4 for each of the search criteria,
as needed.

Creating/Editing a Patient Record


From the Patient Drawer you can create a patient record for a new
patient or edit the patient record for an existing patient.

Managing Patient Records 6-9


QDR Reference Manual

New Patient
Before entering a new patient record, search the database for an
existing record to avoid duplicate entries. If a record does not
exist, create a new record, adding biographical and insurance
information.

Click on New Patient to create new


biography and insurance data.

KP0886-0606

To create a new patient record:


Step Action
Note: Refer to page 6-3 in this 1 Click the Patients button in the main window.
chapter for more information The system displays the Patient Drawer window with a list
about retrieving an existing
of existing patient records.
record.
2 Click the New Patient button to add a new patient record.
The system displays the Patient window with Biography
and Insurance tabs at the top.
Note: Refer to pages 6-4 and 6-6 3 Click the Biography tab, enter the patient’s biography
for field descriptions of the information then click OK.
Biography and Insurance tabs.
The Patient Drawer window appears.
4 Click the Insurance tab, enter the patient’s insurance
information, and click OK.

6-10 Managing Patient Records


QDR Reference Manual

Figure 6-1
Patient Screen, Biography Biography Tab
Insurance Tab

To enter patient biographical information (see Figure 6-1):


Step Action
1 Place the cursor in a field by pointing with the mouse to
the text box.
2 Click in the text box.
Blinking cursor displays in the field.
3 Type the appropriate information.
Use the keyboard Tab key or the mouse to move to the
next field.
4 Click on the Insurance tab.
Insurance dialog window displays.

Managing Patient Records 6-11


QDR Reference Manual

Figure 6-2
Patient Screen, Insurance

KP0886-0608

To enter Insurance information (see Figure 6-2):


Step Action
1 Place the cursor in a field by pointing with the mouse to
the text box.
2 Click in the text box.
Blinking cursor displays in the field.
3 Type the appropriate information.
4 Move to the next field.
5 Click OK when finished.
Patient’s name displays in Select a Patient window.

Updating a Patient Record


You can change the information on a patient’s record any time,
simply by retrieving the record, and entering the new information.
You can also fill out the patient Questionnaire.

6-12 Managing Patient Records


QDR Reference Manual

Figure 6-3 Edit Patient button


Patient Drawer, Edit Patient

KP0886-0609

Patient Biography
To edit a patient record:
Step Action
1 Click the Patients button in the main window.
The system displays the window with a list of patients
sorted alphabetically by last name.
Note: Refer to page 6-7 for 2 Search for and select the patient’s name.
instructions on searching the
Patient Drawer.

Managing Patient Records 6-13


QDR Reference Manual

Step Action
3 With the patient’s name highlighted, click the Edit
Patient button (see Figure 6-3).
The system displays the Patient window with Biography
and Insurance tabs.

Note: Refer to pages 6-4 and 6-6 4 To edit biography information, click the Biography tab
for field descriptions of the and change information, as needed.
Biography and Insurance tabs.
5 To edit insurance information, click the Insurance tab and
change information, as needed.
6 Click OK when done to return to the Patient Drawer
window.

Patient Questionnaire
The Questionnaire button appears if the Mobility option is
installed.
The Patient Questionnaire (see Figure 6-4) is a method of
electronically entering and saving bone density related patient
information.

6-14 Managing Patient Records


QDR Reference Manual

Figure 6-4 Header


Patient Questionnaire Window Today’s Date

Patient
information

Questionnaire

Update Header button

The Header comes from the Header Block placed on Reports (see
“To configure the Header Block” on page 18-21).
Patient Information comes from the Patient Biography screen (see
“New Patient” on page 6-10). If the Patient Biography has been
changed (see “Updating a Patient Record” on page 6-12) then the
patient information (as well as the Header) can be updated by
clicking on the Update Header button.
To fill out a New or edit an Existing Patient Questionnaire
Step Action
1 On the Patient screen, Biography tab, click the
Questionnaire button.
The Questionnaire window appears (see Figure 6-4).

2 The following operations can be performed from the


Questionnaire:
• Clear button. Updates the Header, Patient
information, and changes the date to today while
clearing answers to all questions.
• Update Header. Updates the Header, Patient
information and changes the date to today.
• Print button. Prints the current questionnaire.
• Cancel button. Exits the Questionnaire without
saving data and returns to the previous window.
3 Click OK to save the Questionnaire and return the to
previous screen.

Managing Patient Records 6-15


QDR Reference Manual

Introducing Scan Records


Scan data is gathered from a patient, or phantom, during each
exam. The QDR system stores the scan data, along with
additional information about the scan characteristics and the
patient. These can include both analyzed scans and unanalyzed
scans.

Scan Drawer
The Scan Drawer window allows you to display scans currently
stored in the QDR database. This includes analyzed, unanalyzed,
or a list of both (all scans).
The Scan Drawer window also allows you to delete scans from
the database and retrieve scan details about the selected scan and
the associated patient.

KP1035A-0319

From the Scan Drawer, you can also:


• initiate an exam
• analyze the scan
• initiate a report

6-16 Managing Patient Records


QDR Reference Manual

Working with Scan Records


Click Scans in the main window to display the Scan Drawer
window.
The system displays the following information in the headings of
the Scan Drawer window. An asterisk (*) next to the heading
specifies the sort order of the list. For example, an asterisk in the
Scan Date heading displays the list of scans in order by date.
Field Description
Patient Name The patient’s name, or phantom, for which the scan
was performed and analyzed.
Patient ID The patient’s ID number.
Scan Date The date on which the scan was performed.
Scan Type The type of scan performed.
Scan ID The scan identification number.
Analysis Date The date the scan was analyzed.

Scan Details
When a scan is highlighted, you can click the Scan Details button
to view details and identification information about the scan.

Scan Details button


(only active when a
scan is highlighted)

KP0886-0611

When Scan Details is selected, the Scan Property window


displays with two tabs, Details and Identification.

Managing Patient Records 6-17


QDR Reference Manual

Details Tab
The Details tab allows you to edit the Accession Number, Height,
Weight, Operator, Scan Comment, and three additional user
defined fields (shown as HL7 Field 1, HL7 Field 2, and HL7
Field 3) in the dialog box below.
The user defined field labels shown are examples of user
definable label names. The labels and entry fields may not appear
on the dialog box. See “HL7 Tab” on page 18-34 for details on
enabling/disabling the displaying of these fields and defining
their label names.
Field Description
Patient Name The patient’s name, or phantom, for which
the scan was performed and analyzed.
Accession Number Number of your choice.
Study Instance UID A DICOM attribute which uniquely
identifies a study. Worklist option must be
installed on your system and you must be
logged in as Administrator for this label and
field to appear.
HL7 Field 1 Sixteen characters of your choice.
HL7 Field 2 Sixteen characters of your choice.
HL7 Field 3 Sixteen characters of your choice.
Height Height at the time of the scan.
Weight Weight at the time of the scan.
Operator Initials of person who performed the scan.
Scan Comment Information added to scan by operator.

6-18 Managing Patient Records


QDR Reference Manual

Identification Tab
The Identification tab lists information specific to the scan. You
cannot edit any of the fields in the Identification tab.
Field Description
Patient Name Name of patient as entered on the patient
record.
Patient ID The patient’s ID number.
Scan ID ID number assigned by the system.
Scan Date Day, Month, Year, and time the scan was
performed.
Scan Mode Identifies speed at which scan was performed.
Scan Type Scan Type
Serial Number Number assigned by the software to this scan
after it is completed
Machine Type QDR model number the scan was performed
on

6-19
QDR Reference Manual

Field Description
Analysis Date Day, Month, Year and time the scan was
analyzed
Archive 1 Date The date the scan was initially archived. If the
scan has not been archived, this field displays
as blank
Archive 2 Date The date the scan was next archived. If the
scan has not been archived to a secondary
location, this field displays as blank.

KP0886-0613

6-20
QDR Reference Manual

Deleting Scan Files


Scan files can be deleted from the Scan Drawer by clicking the
Delete Files button.

KP0886-0614
Delete Files button

Clicking the Delete Files button closes the Scan Drawer and
opens the Delete Scans window.

KP0886-0615

The Delete Scans window displays with two tabs, Archived Scans
and UnArchived Scans. Click on the Archived Scans tab to see all
the archived scans in the database.

6-21
QDR Reference Manual

Archived Scans Tab


To delete an archived scan, highlight the desired scan and click on
the Delete Scan Files button. The scan will be deleted. Multiple
scans may be selected by using the Shift or Ctrl keys.
To select all of the scans in the database click the Select All
button. To deselect all highlighted scans click the Deselect All
button.

UnArchived Scans Tab


Caution: Unarchived scans To delete unarchived scan(s), click on the UnArchived Scans tab
cannot be recovered after they are to see all the unarchived scans in the database. Highlight the
deleted.
desired scan and click on the Delete Scan Files button.
Multiple scans may be selected by selecting them while pressing
the Shift or Ctrl keys.

Retrieving the Scan Records


Storing scan information makes it possible to recall previous
scans when comparing studies. It also allows the operator to
analyze the day’s scans in one sitting, rather than after each
individual exam. Generally, the following tasks are performed
with a scan record:
• Retrieve a scan to view the analysis results
• Retrieve a scan to perform an analysis
• Retrieve a scan to compare scan results during analysis
The scan must be selected and dragged onto one of the following
buttons:
• Perform Exam
• Analyze Scan
• Report
To go to the Scan Drawer:
Step Action
Note: Refer to the table on page 6- 1 Click the Scans button in the main window.
3 of this chapter for suggestions The system displays the Scan Drawer window with a series
about alternative methods.
of tabs listing analyzed, unanalyzed, or all scans stored on
the system.

6-22
QDR Reference Manual

Unanalyzed Scans tab


Analyzed Scans tab
All Scans tab

KP0886-0616

To view a scan:
To View... Click...
analyzed scans the Analyzed Scans tab to view a list of all
analyzed scans currently stored in the
database.
unanalyzed scans the Unanalyzed Scans tab to view a list of
all unanalyzed scans currently stored in the
database.
all scans the All Scans tab to view a list of both
analyzed and unanalyzed scans currently
stored in the database.

Performing an Analysis
If the operator is unable to analyze a scan immediately after it has
been performed, it is stored on the computer as an unanalyzed
scan. The analysis can be performed later.
To complete an analysis:
Step Action
Note: Refer to page 6-22 for a 1 Click the Unanalyzed tab.
review of retrieving scans.
2 Select the scan from the list.
3 Drag the scan to the Analyze Scan button on the main
window.
The Analyze Setup window displays so that you can select
an analysis method.

6-23
QDR Reference Manual

Adding Scan Information from Other


Manufacturer’s Systems
The QDR system has the capability of adding data from other
manufacturer’s systems into its database for use on various
reports (such as Rate of Change). To do this, the data from the
other manufacturer’s scans must first be converted (see “Scan
Conversion” on page 18-39). Information for other
manufacturer’s scans can be added to the database for an AP
Lumbar Spine or Left, or Right, Hip scan.
In order to enter this information two things are required:
1. A hardcopy of a patient scan report of the AP Lumbar Spine
or Left, or Right, Hip from the other manufacturer, and
2. The patient must have a Patient Record (see “Creating/Edit-
ing a Patient Record” on page 6-9).
Entering data from other manufacturer’s scans:
Step Action
1 Click Patients to bring up the Patient Drawer window.
2 Highlight the patient’s name and click the Patient Scan tab
to bring up a list of scans for that patient.
KP0886-0617_1

Note: If the patient’s name does


not appear on the Patient window
a Patient Record must be created
(see “Creating/Editing a Patient
Record” on page 6-9).

KP0886-0617_2

6-24
QDR Reference Manual

Step Action
3 Click Scan Entry to bring up the Manual Scan Entry
window.
KP0886-0617_3

KP0886-0617_4

Note: If the other manufacturer is 4 Click the drop down menu next to BMD Results from and
not on the drop down list contact select the other manufacturer.
your MIS or RIS department to
have them create this information
in the QDR database.

KP0886-0617_5

Note: Only information from an 5 Click the drop down menu next to Scan Site and select the
AP Lumbar Spine or Left, or Right, type of scan from the other manufacturer’s hardcopy scan
Hip can be entered into the
report.
database.

KP0886-0617_6

6-25
QDR Reference Manual

Step Action
6 In the Scan Date field, enter the Month, Day and Year of
the other manufacturer’s scan report.

KP0886-0617_7

7 If entering data for:


• spine scan see “Entering data from other
manufacturer’s scans for spine” below.
• hip scan see “Entering data from other
manufacturer’s scans for hip:” on page 6-27.
Entering data from other manufacturer’s scans for spine:
Step Action
1 In the BMD field enter the total BMD on the scan report
into the Total field.

KP0886-0617_8

2 Ensure that the Regions of the scan report reflect the


regions in Regions included in Total.
If one or more of the regions are missing on the scan report,
click the region (L1, L2, L3, and/or L4) to remove the
check mark () from the box, excluding that region.
3 Click Region Details to bring up the Region Details Entry
window.
KP0886-0617_9

KP0886-0617_10

6-26
QDR Reference Manual

Step Action
4 Enter the data from the scan report for the BMC and Area
for the appropriate regions (L1, L2, L3 and L4).
5 Click OK to close the Region Details Entry window.
6 Click OK to close the Manual Scan Entry window
Entering data from other manufacturer’s scans for hip:
Step Action
1 In the BMD field enter the Total, Neck and Trochanter
data from the scan report.

KP0886-0617_11

2 Click OK to close the Manual Scan Entry window


With the information from the other manufacturer’s scan report
added to the QDR database, this scan will be reported under the
patient’s name on the Patient Drawer and the scan can be selected
for various reports (such as Rate of Change).

6-27
QDR Reference Manual

6-28
Chapter 7
The Patient Examination
The QDR system is used to perform an X-ray scan of the AP Lumbar
Spine, Hip, Forearm, AP/Lateral Spine, or Whole Body (depending
upon which model you have). Optional examinations are explained in
Appendix A.
To complete a patient examination the operator must perform a series
of different procedures. Some of these procedures are unique to the
anatomy being examined and are reported in subsequent chapters;
some of these procedures are the same for all types of examinations and
are described in this chapter.

Content
Subject Page
The Patient Examination 7-2
Preparing the Patient 7-2
Creating/Retrieving a Patient Biography 7-4
Selecting the Scan Type and Mode 7-21
Positioning the Patient and C-arm 7-21
Performing the Examination 7-22
Exiting the Examination 7-23
Performing the Analysis 7-24
Exiting the Analysis 7-27
Using a Study 7-28
Generating and Printing Reports 7-32

7-1
QDR Reference Manual

The Patient Examination


The goal of performing an examination is to acquire a quality
scan that produces accurate results when analyzed. The reports
generated from the analysis of the scan are used by a physician as
part of an ongoing treatment plan for osteoporosis and other
related conditions.
The following is a list of procedures used to perform a patient
examination:
1. Preparing the patient
2. Creating/Retrieving a patient biography
3. Selecting the scan type and mode
4. Positioning the patient and C-arm
5. Performing the Examination
6. Exiting the Examination
7. Performing the Analysis
8. Exiting the Analysis
9. Generating and printing a report

Preparing the Patient


To prepare the patient the operator must
• interview the patient
• prepare the patient for the examination

Patient Interview
The answers to the following questions may help to setup the
scanning procedure, could impact the interpretation of the test
results, or even postpone the exam.

Patient Questions
The following is a list of suggested questions to ask the patient.
Some questions may not apply to all examinations, but questions
such as possible pregnancy should be asked before any
examination.
• Is there any chance of pregnancy?
If the patient is female, and within childbearing age, you
must ask if she is pregnant. If the patient is pregnant, or
may be pregnant, postpone the scan until pregnancy is
ruled out.

7-2 The Patient Examination


QDR Reference Manual

• Has the patient had any procedure using the following


substances within the last 7 days:
– Iodine
– Barium
– Nuclear medicine isotope study
The isotope activity may affect the results of the bone
densitometer exam. The nuclear medicine department
should be contacted to determine if the bone densitometry
exam is feasible.
• Is the patient wearing any objects in the scan area such
as an ostomy device, metal buttons or snaps, or
jewelry?
This may affect the position of the patient as well as
interfere in the scanning of the patient.
• Has the patient had any surgery in the area being
scanned?
If surgery was performed in the area of examination a
decision whether to perform the examination must be
made. If the patient had surgery on a hip or forearm, then
the uninjured hip or forearm should be scanned.
• Does the patient have any internal objects that could
interfere with the scan such as:
– Pacemaker leads
– Radioactive seeds
– Metal implants
– Surgical staples
– Foreign bodies, e.g., shrapnel
– Radio-opaque catheters or tubes

Patient Preparation
The next step is to prepare the patient for the examination:
• Instruct the patient to remove or adjust their clothing in
the area to be scanned. Be sure that any metal, e.g., zipper,
snap, belt, etc., is out of the scan field. If necessary, ask
the patient to change into a gown for the examination.
• For AP lumbar spine, hip, or whole body examinations,
instruct the patient to remove their shoes as they may
interfere with some scans. This also helps reduce wear on
the table mat.

The Patient Examination 7-3


QDR Reference Manual

• Measure and record the patient’s height and weight. The


weight limit of the table is 450 pounds for Discovery, 350
pounds for Explorer and 300 pounds for Delphi and QDR
4500. If a patient exceeds the weight limit you may scan
the forearm with the patient seated in a chair.

Creating/Retrieving a Patient Biography


Note: Remember that daily QC Information concerning the patient is stored in a database on the
must be performed before computer and is called the Patient Biography. This information is
acquiring the first scan of the day.
entered into the computer on the patient’s first examination and
For more information about QC
refer to Chapter 5. can be retrieved on subsequent visits.
A patient record can be created or retrieved after selecting the
Perform Exam button.

Creating a Patient Biography


A new Patient Biography must be created for a new patient on
their first examination. Refer to the Creating a Patient Record in
Chapter 6 to create the patient biography and optional insurance
information.

Retrieving a Patient Biography


Prior to retrieving the patient’s biography, weigh and check the
height of the patient. This information, as well as the operator’s
identification, will be entered during this procedure.
Each patient must have a record of biographical information that
is stored in the computer database. Every time the patient returns
for an examination, this information can be recalled.
To retrieve a patient’s file:

Click the Perform Exam button.


The Select a Patient for this Exam window appears with the Select
A Patient tab displayed.

KP1035A-0309

7-4 The Patient Examination


QDR Reference Manual

Note: For information on the


Worklist tab refer to“Using the
Worklist Tab” on page 7-8.

KP1035A-0702

Using the Select A Patient Tab

1 Click on the patient’s name.


Patient’s name is highlighted.
2 Click OK.
The Patient Confirmation window displays.

Searching for a Patient’s Name


If the patient’s name is not in the Select a Patient window, type in
the patient’s name using the computer keyboard. If the name is in
the computer database it will appear highlighted at the top of the
Select a Patient window.
If the patient’s name is not found in the computer database then
this is a new patient and a patient biography must be created, refer
to Creating/Editing a Patient Record in Chapter 6.

The Patient Examination 7-5


QDR Reference Manual

Confirming the Patient Information


Once a patient’s name is selected the Patient Confirmation
window will appear.

Note: Accession number label and


data entry field appears on the
dialog box only when the DICOM
option is installed on your QDR
system.
Note: HL7 Field 1, HL7 Field 2,
and HL7 Field 3 are user
definable label names. These
labels and their data entry fields
appear on the dialog box when
enabled in System Configuration.
See “HL7 Tab” on page 18-34 of
Chapter 18 for details.

The Patient Confirmation window is used to confirm the patient’s


name, ID and date of birth. This window is also used to update the
patient’s biography for height and weight, to identify the operator
performing the examination and the referring physician, and to
enter the DICOM accession number and up to three user
definable data entries when those fields are present on the dialog
box.
Body Mass Index is automatically calculated as height and/or
weight is entered or changed.
This window also allows the user to enter information into the
patient Questionnaire and to edit the patient’s biography and
insurance information.
To update the patient’s height, weight and the operator ID via
the Patient Confirmation window:
Step Action
1 Verify the patient name, ID and date of birth.
2 Enter the height and weight of the patient, operator’s
identification, and enter or select referring physician’s
identification.
3 Click the OK button.
The Select Scan Type window displays.

7-6 The Patient Examination


QDR Reference Manual

Step Action
4 Continue with “Selecting the Scan Type and Mode” on
page 7-21

To edit the patient’s biography and insurance information via


the Patient Confirmation window:
Step Action
1 Click Edit Patient.
The Patient window displays.
Note: Refer to pages 6-4 and 6-6 2 Click the Biography tab, enter the patient’s biography
for field descriptions of the information.
Biography and Insurance tabs.
3 Click the Insurance tab, enter the patient’s insurance
information, and click OK to return to the Patient
confirmation window.
4 Click the OK button.
The Select Scan Type window displays.
5 Continue with “Selecting the Scan Type and Mode” on
page 7-21

Filling out the Patient Questionnaire


The Questionnaire button appears if the Mobility option is
installed.
The Patient Questionnaire (see Figure 7-1) is a method of
electronically entering and saving bone density related patient
information.
Figure 7-1 Header
The Questionnaire Widow Today’s Date

Patient
information

Questionnaire

Update Header button

The Header comes from the Header Block placed on Reports (see
“To configure the Header Block” on page 18-21).

The Patient Examination 7-7


QDR Reference Manual

Patient Information comes from the Patient Biography screen (see


“Confirming the Patient Information” on page 7-6). If the Patient
Biography has been changed then the patient information (as well
as the Header) can be updated by clicking on the Update Header
button.
To fill out a New or Existing Patient Questionnaire
Step Action
1 On the Patient screen Biography tab click on the
Questionnaire button.
The Questionnaire window appears (see Figure 7-1).

2 The following operations can be performed from the


Questionnaire:
• Clear button. Updates the Header, Patient
information, and changes the date to today while
clearing answers to all questions.
• Update Header. Updates the Header, Patient
information and changes the date to today.
• Print button. Prints the current questionnaire.
• Cancel button. Exits the Questionnaire without
saving data and returns to the previous window.
3 Click OK to save the Questionnaire and return the to
previous screen.
Using the Worklist Tab
After the Modality Worklist has been configured by the HIS/RIS
administrator, the operator can use the Modality Worklist through
this interface.
Use the following procedure to display the Modality Worklist tab.
Step Action
1 Click Perform Exam.
The Select a Patient for this Exam window appears with
the Select A Patient tab displayed.

KP1035A-0309

7-8 The Patient Examination


QDR Reference Manual

Step Action
Note: For information on the 2 Click on the Worklist tab.
Select A Patient tab refer Worklist tab
to“Using the Select A Patient
Tab” on page 7-5 in this user’s
guide.

KP1043A-02

The Worklist tab appears (Figure 7-2) with studies for


different patients to be performed on the QDR System.
Figure 7-2
Modality Worklist Patient studies

The labels across the top of the task list include:


Note: Clicking on any of these Patient Name
labels will sort the task list Patient ID
according to that column (an
Start Date
asterisk indicates the current
sort column). Clicking the Study Name
column label will toggle the sort Procedure Code
between ascending and Accession Number
descending order.

The Patient Examination 7-9


QDR Reference Manual

The modality worklist for the operator contains the following


areas:
Patient Name

KP1043A-17

This is a text box search engine that allows the operator to find
any patient’s name, within the list of tasks, by typing it into the
text box.
Query

KP1043A-18

Clicking on this button will update the worklist database on the


QDR System. The Query may take a while (during this time the
database is updated). The following screen appears.

KP1043A-22

The Abort button stops the query.

7-10 The Patient Examination


QDR Reference Manual

Display Range

KP1043A-19

The list of tasks shown on the worklist tab are contained in a


database on the QDR System. This database is controlled by how
worklist is configured.
The Display Range controls how much of the database is
displayed in a range of days. The operator can set this range of
days by using the Days Back and Days Forward spin counters.
• Days Back spin counters - this control sets the limit of
days in the past to display, 0 stands for today’s date. The
range for the spin counter is 100 days.
• Days Forward spin counters - this control sets the limit of
days in the future to display, 0 stands for today’s date. The
range for the spin counter is 100 days.
An example: If the Days Back spin counter is set to 7 and the
Days Forward spin counter is set to 0 (zero) then the list will
display only those entries that fall within this date range.
Clicking Apply will activate the settings in Days Back and Days
Forward.
Manual Study

KP1043A-20

This check box determines if the QDR system will perform


manual examinations, or is programmed to perform examinations
determined by studies, on patients in the task list. If there is a
check mark in the box then manual examinations will be
performed (see “Confirming the Patient Information” on page 7-
6); if the check box contains no check mark, then the QDR
System will perform examinations determined by studies in the
task list (see “Study Tab” on page 18-68).
Use Default Mode

KP1043A-21

This check box determines if the QDR system uses the default
scan mode for the selections in the study. If there is no check

The Patient Examination 7-11


QDR Reference Manual

mark in the box, the operator must enter the scan mode for the
scan (see Selecting the Scan Type and Mode in the appropriate
chapter of the User’s Guide for the scan being performed); if the
check box contains a check mark, the QDR System will
automatically select the default scan mode.
For a manual examination (see Manual Study on page 7-11) the
“default scan mode” is the default mode for that scan type. For a
study the “default scan mode” is the scan mode from the scan
definition.
Extended Details

Clicking on this button displays extended details pertaining to the


exam selected in the patient studies list.

Click Additional Data to display additional data pertaining to the


exam selected in the patient studies list.

7-12 The Patient Examination


QDR Reference Manual

Click Close to return to the Select a Patient for this Exam -


Worklist tab.

Using Modality Worklist with a Study


Use the following procedure to perform a task from the worklist.
Step Action
Note: If the worklist indicates 1 Highlight the patient’s name on the worklist.
that the patient is new, go to
Adding a Patient on page 7-17.
Note: If the worklist tries to
match another patient because
information is different, go to
Matching an Existing Patient
on page 7-18.

KP1043A-23

Note: If the patient’s name is not 2 Click OK.


in the QDR database, see The Patient Confirmation window appears. (Refer to
“Adding a Patient” on page 7-
“Confirming the Patient Information” on page 7-6 for
17.
details.)
3 Enter the patient’s weight and height and your operator’s
number (if necessary) then click OK.
Note: If an error message 4 The Study will begin by displaying the Scan Parameter
appears stating that the study is window of the first scan type in the study. (The following
not defined, go to Adding a
is an example.)
Study on page 7-18.

KP1043A-36

The Patient Examination 7-13


QDR Reference Manual

Step Action
5 The scan type examination will start. The examination is
the same as described in the User’s Guide for the scan
type being performed. When the examination is complete
the Exit Study window appears.

KP1043A-37

The operator has four choices from this window.


• Analyze Scan - This is described in “Performing
the Analysis” on page 7-24.
• New Scan - This is described in “Selecting the
Scan Type and Mode” on page 7-21.
• Exit Study - This button has two functions:
If this is the last examination of the study. This is
described in “Exiting the Examination” on
page 7-23.
If this is not the last examination of the study, the
follow confirmation message appears.

KP1043A-38

• Add Scan Comments - This is described in


“Adding Scan Comments” on page 7-24.

7-14 The Patient Examination


QDR Reference Manual

Step Action
6 From the Exit Study window select Analyze Scan to
continue the study. The Analysis of the examination will
begin. The analysis is the same as described in the User’s
Guide for the scan type being performed. When the
analysis is complete the Exit Study Analysis window
appears.

KP1043A-39

The operator has four choices from this window.


• Report - If this is not the last examination in the
study, initiating a report will force exiting from
the study and the following confirmation window
will appear:

KP1043A-38

• Analyze Another Scan- This selection is grayed


out.
• Next Scan - This starts the next examination in
the study.
• Exit Study- If this is not the last examination of
the study, the follow confirmation message
appears.

KP1043A-38

The Patient Examination 7-15


QDR Reference Manual

Step Action
7 The study will continue, repeat steps 4 through 6 for each
examination in the study.
8 When the analysis of the final examination of the study is
complete, the Exit Analysis window appears.

KP1043A-40

The operator has four choices from this window.


• Report - A report for each examination of the
study will be developed.This is described in
“Generating and Printing Reports” on page 7-32.
• Analyze Another Scan- This selection is grayed
out.
• Next Scan - This is described in “Selecting the
Scan Type and Mode” on page 7-21.
• Exit - This will exit the study. This is described in
“Exiting the Examination” on page 7-23.

Modality Worklist Operations


Using modality worklist, the operator can perform the following
operations:
• Adding new patients to the QDR database
• Adding new studies to the Study Tab under System
Configuration
• Matching an Existing Patient

7-16 The Patient Examination


QDR Reference Manual

Adding a Patient
If the operator selects a patient that is not in the QDR System
database, the following screen appears:

Clicking New Patient adds the patient to the QDR System


database using the HIS/RIS information for that patient. The
KP1043A-26
Patient window opens so the operator can verify information or
add missing data.

Click OK.
The Patient Confirmation window appears.

Enter the patient’s weight and height and your operator number
(if necessary) and click OK.

The Patient Examination 7-17


QDR Reference Manual

The study will continue.

Adding a Study
If the study in the task list is not defined in the QDR database the
following message appears:

KP1043A-28

The parameters of the study must be placed into the database on


the QDR System. This information is provided by the physician
or HIS/RIS. Once the information has been acquired, the operator
can manually enter it into the database (see “Study Tab” on
page 18-68).

Matching an Existing Patient


If the patient clicks on a patient whose name is not in the database
of the Patient Biography, a screen similar to the following will
appear.

Information concerning new patient

The operator should proceed with caution before performing the


following procedure because if information from the HIS/RIS
database is accepted, it will overwrite information in the Patient
Biography database. Even if the HIS/RIS information is incorrect
(such as a misspelled name), if all other information matches, the
operator must accept the HIS/RIS information in order to run the

7-18 The Patient Examination


QDR Reference Manual

study. The operator can contact the HIS/RIS administrator later to


have the information changed.
Use the following procedure to match an existing patient:
Matching an Existing Patient
Step Action
1 When the Select/Create Patient for Exam window
appears, it will highlight one patient in the Patient
Biography that it believes is the closest match to the
patient name from the HIS/RIS study.

Matching names

2 The operator now has a choice:


• If this is a new patient the operator should click
on the New Patient button and go to Adding a
KP1043A-26
Patient on page 7-17.
• If the operator believes that there is a match
between the two patients, click on the OK button.
KP1043A-32
A window appears that provides QDR Patient
Info and the Worklist Patient Info information.
QDR Patient Info

Worklist Patient Info

The Patient Examination 7-19


QDR Reference Manual

Step Action
3 The operator should check each field in the QDR Patient
Info with each entry in the Worklist Patient Info,
• If the information the QDR Patient Info does not
match the Worklist Patient Info then these are
KP1043A-34
not the same patient, click on the Reject button to
return to the Select/Create Patient for Exam
window to select another patient to match or
create a new patient.
• If most of the fields match and it is determined
that the patient in the Worklist Patient Info is the
KP1043A-32
patient in the QDR Patient Info, click OK. A
confirmation window appears.

KP1043A-35

4 This confirmation window is that last warning before


information in the Patient Biography database is changed.
If the information in the QDR Patient Info matches the
information in the Worklist Patient Info, click OK.
The information in the Worklist Patient Info will
overwrite the information in the Patient Biography and
the study will continue.

7-20 The Patient Examination


QDR Reference Manual

Selecting the Scan Type and Mode


Note: If using a Study for the The scan type and mode are specific to the examination being
examination and the analysis refer performed. Refer to the Selecting the Scan Type and Mode
to Using a Study on page 7-28.
heading in the chapter indicated:
AP Lumbar Spine—Chapter 8
Hip—Chapter 9
Forearm—Chapter 10
Whole Body—Chapter 11
AP Lateral—Chapter 12
IVA and QM—Chapter 13

Positioning the Patient and C-arm


Except for the Forearm examination, the first step in positioning
the patient is to place the patient on the table. Once the patient is
on the table, positioning for the specific examination can be
performed.

Placing the Patient on the Table


Follow the procedure below to maximize patient comfort and
safety while assisting the patient onto the table.

Figure 7-3
Patient On/Off Position

KP1068A-0704

The Patient Examination 7-21


QDR Reference Manual

To place the patient on the table:


Step Action
1 Press the Patient On/Off switch on the Control Panel.
This moves the C-arm to the far left position, and extends
the table out from the base, making it easier for the
patient to get on the table.
2 After the C-arm and table stop moving, seat the patient on
the table pad so that their right shoulder is adjacent to the
C-arm (see Figure 7-3).
3 Lay the patient back with their head at the right end of the
table.
4 Continue with positioning the patient for the specific
examination.

Positioning for the Specific Examination


Positioning the patient and the C-arm are specific to the
examination being performed. Refer to the Positioning the
Patient and C-Arm heading in the chapter indicated:
AP Lumbar Spine—Chapter 8
Hip—Chapter 9
Forearm—Chapter 10
Whole Body—Chapter 11
AP Lateral—Chapter 12
IVA and QM—Chapter 13

Performing the Examination


Performing the examination is specific to the examination being
performed. Refer to the Performing the Examination heading in
the chapter indicated:
AP Lumbar Spine—Chapter 8
Hip—Chapter 9
Forearm—Chapter 10
Whole Body—Chapter 11
AP Lateral—Chapter 12
IVA and QM—Chapter 13

7-22 The Patient Examination


QDR Reference Manual

Note: When performing the examination, the Scan Parameter


window and the Scan and Reposition screens display baseline
scan. If the patient has a baseline scan of the same type as the
currently selected scan, that baseline scan will be displayed. If
the patient does not have a baseline scan of the selected scan
type a default scan of the scan type is displayed. For
information on selecting baseline scans refer to “Managing
Patient Baseline Scans” on page 19-17.

Exiting the Examination


After acquiring an acceptable scan, the Exit Exam window
displays.

KP0886-0705

This window provides four options for the operator.


Option Description
Analyze Scan This button continues with the exam by
performing the analysis (see
Performing the Analysis in this
chapter).
Note: If the analysis of the scan New Scan This button starts a new examination
is postponed to a later time, the on the same patient by displaying the
scan can be searched for in the
Select Scan Type window. The
Scan Drawer. Refer to Chapter 6
for information about the scan previous scan is stored on the computer
drawer. and can be analyzed at a later time.

Add Scan Comments This button allows the operator to


insert comments into the computer
database which will remain with the
scan and be printed on the report. A
Comment window will display (see the
instructions that follow).

The Patient Examination 7-23


QDR Reference Manual

Adding Scan Comments


Comments can be added to a scan and stored with the scan and
patient’s record.

KP0886-0706

To add scan comments:


Step Action
Note: To locate comments on a 1 Click the Add Scan Comments button.
scan report, refer to Chapter 15. A comments window displays.
2 Place the cursor in the comments field, and type the
comment, up to 50 characters.
3 Clicking the Add patient comment to scan comments
button will cause any comments previously placed in the
Patient Comments of the Patient Biography to appear.
4 If more than one scan exists for this patient, clicking the
<< Previous Scan or Next Scan >> buttons will show
comments on the other scan(s).
5 Click OK to save the comment.
Exit Exam window displays.

Performing the Analysis


Performing the analysis is specific to the examination
being performed. Refer to the Performing the Analysis
heading in the chapter indicated:
AP Lumbar Spine—Chapter 8

7-24 The Patient Examination


QDR Reference Manual

Hip—Chapter 9
Forearm—Chapter 10
Whole Body—Chapter 11
AP Lateral—Chapter 12
IVA and QM—Chapter 13

Contrast and Brightness


The contrast and brightness of the image can be adjusted to obtain
the best definition of important anatomical features. Adjusting the
contrast and brightness has no effect on calculations; it is used to
KP0886-0817 optimize the display and for visual adjustments only.

When the Sun/Moon button is clicked, the Image Toolbox


(Figure 7-4) displays. From this toolbox the operator can:
• Change the contrast and brightness of the image.
• Invert the image.
• Change the image from dual energy to single energy, or
from single energy to dual energy.
• Change the zoom of the image.

Figure 7-4 Black dot

Image Toolbox

Length

Default Width

KP0886-0707

The Patient Examination 7-25


QDR Reference Manual

To adjust the contrast and brightness:


Note: Contrast and brightness are The contrast and brightness of the image is represented by the
at the preference of the operator, location of the black dot in the Image Toolbox triangle (see
there are no right or wrong
Figure 7-4). The contrast level is represented by the number in the
settings.
L box (L stands for Length), and the brightness level is
represented by the number in the W box (W stands for Width).

The operator can change the contrast and brightness in the


following ways:
• Click and drag the black dot in the triangle. Dragging the
dot to the right or left in the triangle adjusts brightness,
and dragging it up or down in the triangle adjusts the
contrast. The numbers in the W and L boxes change as the
black dot is moved.
• Click in either the W or L boxes and highlight the number.
Change the value to a number between 0 and 100. Repeat
this action in the other box. The black dot moves inside
the triangle to the point represented by the number(s)
entered.
The D button is the default button. Clicking the D button places
the black dot in the center of the triangle and changes both the L
KP0886-0708
and W box numbers to 50.

To Invert the Image:


The Invert button inverts the colors of the scan image, white
becoming black, and black becoming white. To invert the image,
KP0886-0709
click the Invert button.
To change the image to/from Single or Dual Energy:
The SE/DE button is used to switch between single and dual
energy views of the scan image. To toggle the image between
KP0886-0710
single and dual energy, click the SE/DE button
The SE/DE button is not available for Forearm or Whole Body
scans, and this button does not appear in the Image Toolbox for
those scans.
To Zoom in and out on an Image:
The Zoom tool is used to either shrink or enlarge the scan image.
Use the button to enlarge the image. Use the button to
KP0886-0711 shrink the image.

7-26 The Patient Examination


QDR Reference Manual

Exiting the Analysis


After performing the analysis, the Exit Analysis window displays.

KP0886-0712

To continue:
To ... Click on the ...
Select a report type and Report button
print.
Select a different scan to Analyze Another Scan button
analyze.
Perform another scan on New Scan button
the same patient
Exit and return to the main Exit button
window.

When Exit is select, if the Dose Area Product (DAP) Meter


option is installed on the system, the following window displays
the radiation dose applied to the patient during the exam:

The Patient Examination 7-27


QDR Reference Manual

Using a Study
Note: Studies must first be When a study (or studies) have been added to the APEX software,
configured before they can be the operator can select a study on the Select Scan Type window
used, see “Study Tab” on page 18-
instead of a specific scan type. The QDR system proceeds directly
68 of Chapter 18.
to the scan types listed in the study.

Using a Study
Step Action
1 When you are finished entering patient data or
confirming patient data, the Select Scan Type window
appears.

KP1035A-1753

The Studies that have been entered for this QDR system
appear at the top of the list of scan types.
Highlight the desired study and click the Next>> button.
2 To start the first scan type of the study, the system
prompts you first to place the patient on the table, then to
take the scan.

7-28 The Patient Examination


QDR Reference Manual

Step Action
3 When the scan is completed, the Exit Study window
appears.

KP1035A-1754

The Exit Study window contains the following buttons:


The Analyze Scan button continues with the
examination by analyzing the scan just taken.
The Next Scan button continues with the next
scan in the study. If this is the last scan of the
study then this button becomes New Scan (see
New Scan button in “Exiting the Examination”
on page 7-23).
The Exit Study button has two functions:
1. If the study has not completed all scans,
clicking the Exit Study button brings up
the following window asking you to con-
firm exiting the study.

KP1035A-1759

2. Click Yes to exit the study and return to


the main window, or No to return to the
Exit Study window.
The Add Scan Comments button allows com-
ments to be added to the scan (see “Adding Scan
Comments” on page 7-24).

The Patient Examination 7-29


QDR Reference Manual

Step Action
4 Click the Analyze Scan button to continue the
KP1035A-1755 examination. When you click the Close button from the
analysis, the Exit Study Analysis window appears.

KP1035A-1760

The Exit Study Analysis window contains 4 but-


tons:
The Report button generates a report (see “Gen-
erating and Printing Reports” on page 7-32).
1. If this is not the last scan in the study, gen-
erating a report exits the study. The fol-
lowing window appears asking
confirmation of generating a report.

KP1035A-1762

2. Click the Yes button to generate a report


and exit the study, or the No button to
return to the Exit Study Analysis window.
The Analyze Another Scan button is grayed out
and inactive.

7-30 The Patient Examination


QDR Reference Manual

Step Action
The Next Scan button starts the next scan in the
study.
The Exit Study button has two functions:
1. If the study has not completed all scans,
clicking the Exit Study button brings up
the following window asking you to con-
firm exiting the study.

KP1035A-1766

2. If this is the last scan in the study, clicking


the Exit Study button returns to the main
screen.
5 Click the Next Scan button to continue with the next scan
KP1035A-1764
in the study.
6 When the analysis of the last scan in the study is
complete, the Exit Analysis window appears.
6

KP1035A-1766

The Exit Analysis window contains 4 buttons:


The Report button generates a report of each scan
in the study (see “Generating and Printing
Reports” on page 7-32).
The Analyze Another Scan button is grayed out
and inactive.
The New Scan button returns to the Select Scan
Type window to start a new scan on the same
patient.
The Exit button exits the study and returns to the
main window.

The Patient Examination 7-31


QDR Reference Manual

Generating and Printing Reports


After completing analysis, the Report button is available on the
Exit Analysis window. Click Report to display the Print window.

KP0886-0713

Note: Refer to “Report Tab” on A variety of report types may be printed singly or as multiple
page 18-10 for more information copies. Reports can be uniquely configured, or chosen from a list
about configuring reports.
of default reports already set up on the system.
To generate a report:
Step Action
1 Click the Report button.
The Print window displays.
Note: Refer to “DICOM 2 Select the type of report by clicking its associated check
Reports” on page 15-18 for box. Multiple report types can be selected.
information on generating and
A check mark is displayed in the box next to each selected
sending DICOM reports.
report type.
3 Select the number of copies by clicking om the up or
down arrows. The increases the number of copies, the
decreases the number of copies.
The number of copies to be printed displays in the text
area.
4 Click the Print button.
The report prints.
5 Click the Close button.
The Print window closes, returning to the main window.

Configuring the Report


Refer to “Report Tab” on page 18-10 in Chapter 18, for
procedures on how to configure the report according to the type
of examination.

7-32 The Patient Examination


QDR Reference Manual

Editing Physician’s Comments


Refer to “Physician’s Comment” on page 15-5 for procedures to
edit any physician’s comments to be included on the report.

The Patient Examination 7-33


QDR Reference Manual

7-34 The Patient Examination


Chapter 8
Performing and Analyzing the
AP Lumbar Spine Scan
This chapter provides procedures to perform an AP Lumbar Spine
scan and analysis on the QDR Series systems. It also provides
information about scanning patients that exhibit lateral curvature,
or scoliosis of the spine, and low density spine analysis.

Content
Subject Page
The AP Lumbar Spine Examination 8-2
Preparing the Patient 8-2
Creating/Retrieving a Patient Biography 8-2
Selecting the Scan Type and Mode 8-2
Performing the Examination 8-7
Exiting the Examination 8-11
Performing the Analysis 8-11
Exiting the Analysis 8-27
Generating and Printing Reports 8-27
Alternate Scan Modes 8-27
Evaluating the Image 8-28
Scoliotic Spine Scans 8-30

8-1
QDR Reference Manual

The AP Lumbar Spine Examination


The AP Lumbar Spine examination includes all activities that
begin when the patient enters the exam room, and ends when an
acceptable image is obtained. The activities involved in
performing the exam include:
• Preparing the Patient
• Creating/Retrieving a Patient Biography
• Selecting the Scan Type and Mode
• Positioning the Patient and the C-arm
• Performing the Examination
• Exiting the Examination
• Performing the Analysis
• Exiting the Analysis
• Generating and Printing Reports

Preparing the Patient


Note: Remember to ask the Prepare the patient by asking the pre-examinations questions and
patient if they are, or might be, ensuring that they are not wearing anything that will interfere
pregnant. Postpone the exam
with the scan. Refer to Preparing the Patient on page 7-2 of
until pregnancy is ruled out.
Chapter 7 for detailed instructions.

Creating/Retrieving a Patient Biography


If this is a new patient, create a new patient biography (refer to
Creating/Editing a Patient Record on page 6-9 of Chapter 6 for
detailed instructions). If this patient already has a biography,
retrieve it now (refer to Retrieving a Patient Record on page 6-7
of Chapter 6).

Selecting the Scan Type and Mode


After a patient biography has been created, or retrieved, the Select
Scan Type window displays.

8-2 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

KP0886-0801

The scan type for the AP Lumbar Spine scan is AP Lumbar


Spine, selected in the Select Scan Type window. A check mark
() in the Use Default Scan Mode box designates the Express
mode for Discovery-A, -SL, -W, -C models, Fast Array mode for
Discovery-Wi, Ci models, or Explorer mode for Explorer.

Choosing the Scan Mode


To choose an alternate scan mode:
To perform the scan in a different mode, remove the check mark
and a window will display that allows selection of an alternate
mode.
To choose the default scan mode (check mark in box):
Step Action
1 Click on the AP Lumbar Spine in the scan type list.
The scan type is highlighted.
2 Click the Next >> button.
The Scan Parameters window displays.

Confirming Scan Parameters


The only scan parameter that can be modified is the scan length
(see Figure 8-1). The scan length for an AP Lumbar Spine exam
is approximately 8 inches (20 cm). If the patient is taller than six
feet, it may be necessary to increase the scan length to 9.4” (23.9
cm). The system may change the number entered to match the
pre-programmed step size of the scanning mechanism.

Performing and Analyzing the AP Lumbar Spine Scan 8-3


QDR Reference Manual

Figure 8-1
AP Lumbar Spine
Scan Parameters

KP1035-0802

Verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan Length
field and type the new length.
3 Stop here and position the patient and the C-arm.

KP0886-stop

Positioning the Patient and C-arm


The goal for positioning the patient on the table is to ensure that
the spine is as straight as possible for the scan (refer to Figure 8-
2). Notice that the figure shows the patient straight on the table
with their lower legs on the positioner (helping to keep the spine
flat). Also note that the area to be scanned starts at about one inch
below the iliac crest.

8-4 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Figure 8-2
Patient Positioning Goal

C C

Knee positioner Iliac Crest

L5
L4
L3
L2
L1
1 inch Area to be scanned
KP0886-0803

The remainder of this section provides detailed information for


positioning the patient, the table, and the C-arm for an AP
Lumbar Spine exam. It is important to follow each set of
directions carefully since they are designed to help acquire an
ideal scan by describing:
• How to maximize patient comfort and safety.
• How to assist the patient onto the table for proper alignment.
• Suggestions for correct positioning techniques.
• Proper use of positioning aids.

Placing the Patient on the Table


Note: Refer to Chapter 7, page Press the Patient On/Off switch on the Control Panel and seat
7-21, for a detailed description the patient on the table between the scan field indicators on the
of placing the patient on the
table pad. Their right shoulder should be adjacent to the C-arm.
table.
Continue with positioning the patient below.

Positioning the Patient


Positioning the patient for an AP Lumbar Spine scan involves
using the positioner (see Figure 8-1), supplied by Hologic. This
positioner helps to reduce the lordotic curve of the patient’s spine
on the table pad by keeping the femurs vertically aligned. The
positioner accommodates three different patient heights (short,
medium, or tall). For example, if the patient is short, place the
positioner to the shorter side to adjust to the size of the patient’s
legs.

Performing and Analyzing the AP Lumbar Spine Scan 8-5


QDR Reference Manual

A pillow can be used under the patient’s head to make them more
comfortable.
To position the patient:
Step Action
1 Lie the patient on their back with their head at the right end
of the table (see Figure 8-1).
2 Position the patient’s body so that the spine is straight on
the table pad.
The patient’s shoulders should be at the upper scan limit
line.
3 Verify that the patient’s pelvis and shoulders are aligned
straight on the table pad and centered to the marks on the
table pad.
Note: On Explorer the Center 4 Press the Center Table button on the Control Panel.
Table button must be held and only The table and C-arm move to the center.
the table moves.
5 Place the positioner under the patient’s lower legs.
6 Adjust the positioner by rotating it until the femurs are as
vertical as possible. This will help reduce the lordotic curve
of the lumbar spine.

o
90

KP0886-0805

7 Place patient’s arms in the most comfortable position,


either by their sides or up over their head.
8 Instruct the patient to remain still and to breathe normally.

Positioning the C-arm


The C-arm and table drive motors provide the table and C-arm
motion required to scan the X-ray beam along the patient’s spine.

8-6 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

To position the C-arm using the Laser:


Step Action
1 Determine the position of the patient’s iliac crest.
2 Press the Laser button on the Control Panel.
3 Use the C-Arm Controls on the Control Panel to place the
cross hair of the laser 1” to 2” below the iliac crest and
centered in the mid-line of the patient.

Performing the Examination


An acceptable AP Lumbar Spine scan (see Figure 8-3) includes
the following:
• The scan starts in the middle of L5.
• The iliac crest is evenly displayed in both lower corners of
the image area.
• The AP Lumbar Spine is centered in the middle of the
scan window.
• There are even amounts of soft tissue on each side of the
spine.
• The scan stops where ribs are attached to T12 (usually the
middle of T12).
Figure 8-3
AP Lumbar Spine Scan

KP0886-0806

Performing and Analyzing the AP Lumbar Spine Scan 8-7


QDR Reference Manual

Starting the AP Lumbar Spine Scan


Once the patient and C-arm are positioned correctly, return to the
computer to start the AP Lumbar Spine scan. At this time, the
Scan Parameters window should be displayed.
To start the AP Lumbar Spine scan:

Step Action
Note: If the Control Panel X-ray 1 Click the Start Scan button
lamp fails to extinguish within 10 The Scan window displays with the image appearing on the
seconds after the end of the scan,
left side. Flashing X-rays On indicator at the top of the
press the red Emergency Stop
button on the Control Panel window continues until the scan stops.
immediately. Then call Hologic
Service before resuming operation.

2 Inspect the image.

3 When you see where ribs are attached to T12 on the image,
click the Stop Scan button
Exit Exam window displays.

While the scan is being acquired the operator has four options:

Option Description

(let the scan complete) The scan completes the length


designated on the Scan Parameters
window and then the Exit Exam
window displays.

8-8 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Option Description

Stop Scan button Terminates the scan at the point at


which the button is pressed and the
KP0886-0808 Exit Exam window displays.

Abort Scan button Exits the scan immediately. An exit


window displays then proceeds to
KP0886-0809 the main window. To start another
scan the C-arm must be
repositioned.

Reposition Scan button Stops the scan in progress and


enables the operator to reposition
KP0886-0810 either the window image or the
patient. Refer to the following
instructions.

Repositioning the Scan


If the scan is positioned correctly (see Figure 8-3) it will not be
necessary to reposition. If it needs to be repositioned, follow the
procedure below.
Two basic activities of repositioning include:
• Repositioning the image
If the image starts too far above or below L5 or is not
centered, move the image up, down, left, or right using the
scroll bars or cursor hand to reposition the scan for the
correct starting point. Refer to “Evaluating the Image” on
page 8-28 of Chapter 8 for image positioning examples.
• Repositioning the patient
If the spine is not straight, move the patient’s upper torso
either left or right to straighten the spine.
To reposition the image:
Step Action
Note: In order to reposition the 1 Click the Reposition Scan button
scan, the Reposition Scan button The image acquired so far displays with scroll bars on the
must be clicked before the scan is
right and bottom.
completed.
Note: You can also use the scroll 2 Position the cursor over the spine image.
bars to reposition the image. The arrow cursor changes to a hand.

Performing and Analyzing the AP Lumbar Spine Scan 8-9


QDR Reference Manual

Step Action

Cursor
Hand

Vertical
Positioning
Lines

Horizontal
Positioning
Scroll
Lines
Bars

KP0886-0811

Note: On Explorer, the scan can be 3 To reposition the lumbar spine:


repositioned a maximum of 1 inch
(2.54 cm.) towards the foot end of • Click and drag the image so that the iliac crest is at
the table and any distance towards or below the blue horizontal positioning line and
the head end. within the lower portion of the scan field.
• The center of the lumbar spine should be aligned
with the blue vertical positioning line.
4 When the spine is positioned correctly, click the Restart
Scan button.
The Scan Parameters window displays.
Note: If the Control Panel X-ray 5 Click the Start Scan button to start a new scan at the new
lamp fails to extinguish within 10 position.
seconds after the end of the scan,
The Scan window displays with a flashing X-rays On
press the red Emergency Stop
button on the Control Panel message. The image displays.
immediately. Then call Hologic 6 When you see where the ribs are attached to T12 on the
Service before resuming operation.
image, click the Stop Scan button.
When the scan completes, the Exit Exam window displays.

8-10 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

While repositioning the scan the following buttons are displayed


in the window.
Button Description
Restart Scan button Proceeds to the Scan Parameters
window to commence X-ray scan
KP0886-0909
from the new position.
Accept Scan button Saves the scan in progress without
KP0886-0910
repositioning the scanner.
Cancel Scan button Cancels the exam and returns to the
KP0886-0911
main window.

Exiting the Examination


After acquiring an acceptable image, the Exit Exam window
displays. To perform the analysis of the scan just completed, click
the Analyze Scan button in the Exit Exam window. Refer to
Exiting the Examination on page 7-23 in Chapter 7 for detailed
instructions.

Performing the Analysis


Note: If the analysis of the scan The analysis of the AP Lumbar Spine scan includes the
is to be postponed to a later following:
time, the scan can be retrieved
• Choosing the analysis method
from the Scan Drawer. Refer to
Chapter 6 for information about Note: If the patient has an AP Lumbar Spine baseline
the scan drawer. scan, the software proceeds directly to the manual
compare analysis function instead of choosing the
analysis method. Compare analysis is discussed in detail
in Chapter 14. Manual AP Lumbar Spine scan analysis is
discussed beginning on page 8-16.
• Performing the analysis

Choosing the Method of Analysis


The first analysis activity involves selecting the analyze method.
The analyze method is the method that the system uses to perform
the analysis. It can perform the analysis either as a single new
scan, or as a comparison to one that has been previously
performed.
If this scan is the patient’s first visit, select the default method
(Spine) and the system presents a single scanned image for
analysis. If this scan is a follow-up visit, select the Compare to
Previously Analyzed Scan radio button. The system then
presents two images. The left image is the unanalyzed scan, and

Performing and Analyzing the AP Lumbar Spine Scan 8-11


QDR Reference Manual

the right image is a previous scan with the region of interest


(ROI) defined.
Compare allows the system to perform the new analysis on the
same ROI as the previous one and maximizes accurate reporting
of the rate of change. Compare and follow-up visits are discussed
in detail in Chapter 14.
To select analyze setup for a first visit:
Step Action
1 From the Exit Exam window, click the Analyze Scan
button.
The Analyze Setup window displays.

Note: The Analyze Setup window 2 Click the Choose Analysis Method radio button, if
displays with the Choose Analysis necessary.
Method radio button selected as
the default. Comparison is covered
in detail in Chapter 14.
3 Verify the analysis method (see Methods of Analysis
below).
4 Click the Next >> button.
The Analysis window displays.

Methods of Analysis
The Spine analysis is the standard method for analyzing AP
Lumbar Spine scans. It includes an automatic low density
analysis (unless disabled).
To select a method of analysis (other than default):
Step Action
1 Click the drop down list arrow on the Analyze Setup
window.
List of methods displays.

8-12 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Step Action
Note: Subregion Array Spine
analysis is available as an option.

2 Click on the desired method.


List disappears, and selected method displays in field.
3 Click the Next >> button to continue.
The Analysis window displays.

Performing the Analysis


Unless disabled, once the method of analysis is chosen the
software will automatically perform the analysis (excluding
Subregion Array Spine analysis). This is called One-Time™ Auto
Analysis and is a feature of all models in the QDR Series.
After the One-Time™ Auto Analysis is completed the operator
can choose to perform a manual analysis of the acquired scan.
Both analysis methods are described below.

Performing and Analyzing the AP Lumbar Spine Scan 8-13


QDR Reference Manual

One-Time™ Auto Analysis


Note: Hologic’s auto-analysis The AP Lumbar Spine scan auto analysis includes all activities
algorithms have been optimized that begin with a satisfactory image and ends with the calculation
for use on patients. You may be
and summary of the results of the scan. After completing a scan,
required to manually analyze
scans of phantoms that were the system automatically saves the scan, in the Scan Drawer, for
scanned as if they were patients. analysis.
Once the method of analysis is selected, the system will
automatically perform an analysis of the scan (excluding
Subregion Array Spine analysis). While this operation is being
performed, no operator interaction is allowed. If system software
determines it is needed, an Automatic Low Density Spine
analysis will be used (unless disabled). Upon completion of Auto
Analysis, analysis results are displayed. If the Automatic Low
Density Spine analysis was used, it will be noted in the analysis
results label.
Note: Legacy spine scans that
used the older version of Low If the system was upgraded from an older version of operating
Density Spine software may software in which Low Density Spine was listed in the analysis
continue to use this software. If menu, that option can be turned on by changing the Analysis
the compare mode is used on
follow-up scans the Low Density
Configuration in the Utilities Tool Bar. This is only
Spine software will be applied to recommended for patient spine scans that used this option
the newer scan. previously. All newly acquired spine scans should be analyzed
with One-Time Auto Analysis.

Figure 8-4
AP Lumbar Spine
One-Time Auto Analysis
Results

Note: Evaluate the analysis for


correct placement of regions and
bone outline. It may be
necessary to adjust the image
display.

KP1068_002-0814
Analysis results

8-14 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Manual Analysis
The operator may perform a manual analysis of the lumbar spine
scan after the auto analysis is completed. In most cases it will not
be necessary to perform manual analysis. However, upon
examination of the auto analysis results, if one of the following
conditions is seen, manual analysis may be required:
• regions are positioned incorrectly
• poor bone mapping
To perform manual analysis after auto analysis:
Step Action
1 Click on the Global ROI button (Figure 8-4).
2 Perform the Manual Analysis starting with Performing the
Manual Analysis (page 8-16).

A specific set of tools, the Analysis Step Buttons, located on the


left side of the window (see Figure 8-5), allow the operator to
proceed step-by-step through each task. These tools can
maximize image quality and accuracy, preventing the need for re-
scanning.
Figure 8-5 Step buttons
AP Lumbar Spine Analysis
Window

Toolbox KP1035A-0816_1

Brightness/Contrast button

Performing and Analyzing the AP Lumbar Spine Scan 8-15


QDR Reference Manual

Performing the Manual Analysis


To perform manual analysis from the Analyze Window, click on
the Global ROI button. The steps to manually analyze an AP
Lumbar Spine include:
1. Adjusting the image display brightness and contrast (Bright-
ness/Contrast button).
2. Defining the Global Region of Interest (Global ROI button).
3. Viewing/editing the bone map (Bone Map button).
4. Marking the Intervertebral spaces (Vertebral Line button).
5. Labeling the Vertebral bodies and view analysis results
(Results button).
6. Exiting the Analysis window (Close button).

Adjusting the Image Display


The contrast and brightness of the image can be adjusted to
obtain the best definition of important anatomical features.
Adjusting the contrast and brightness has no effect on
KP0886-0817 calculations; it is used to optimize the display and for visual
adjustments only. Refer to Contrast and Brightness on page 7-
25, in Chapter 7, for detailed instructions.

Defining the Region of Interest (ROI)


The global ROI refers to the defined boundaries of the image that
is being analyzed. The ROI appears on the image as a box: the
dashed line is active and can be adjusted; the solid lines are
KP0886-0818
inactive and cannot be adjusted.

8-16 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Figure 8-6 Dashed yellow Line Area to be analyzed


ROI Image in the
Analysis Window

KP0886-0819

Solid red lines

A properly positioned Global ROI includes:


• The spine centered within the ROI.
• A preset width of 116 lines (automatically preset).
• The top line of the ROI positioned between T12 and L1.
The bottom line of the ROI positioned between L4 and L5

Performing and Analyzing the AP Lumbar Spine Scan 8-17


QDR Reference Manual

The Global ROI Toolbox


The Global ROI Toolbox is used to define the region of interest:
• Line Mode
When selected, one line on the ROI box initially displays
as a dashed yellow line. Any of the ROI box lines can be
selected and then moved. Top or bottom (horizontal) lines
can be moved up or down. Side (vertical) lines can be
moved left or right.
• Whole Mode
When selected, the entire ROI box initially displays as a
dashed yellow line. This allows the entire box to be
moved to center the spine while maintaining the 116 line
width.
• Point Mode
When selected, a plus sign displays in each corner of the
ROI box, as well as at the center of the top and bottom
lines. A corner or center point can be selected to angle the
top and bottoms lines to match lateral curvature or
KP0886-0820 scoliosis (explained in Scoliotic Spine Scans on page 8-
30 of this chapter).

To define the ROI in Line Mode:


Step Action
1 Click the Global ROI button in the top left corner of the
window, if not already selected.
The ROI box displays in the image area.
2 The Line Mode button is the default and should be active
(click if not already selected).
The top line of the ROI box displays as an active line
(dashed yellow).
Note: Dashed yellow lines are 3 Click and drag the top line to its position between T12 and
active and can be edited. Solid red L1 intervertebral space. (Evenly spaced, if possible).
lines are inactive.
Note: The top line remains dashed 4 Click and drag the bottom line to its position between L4
until the bottom line is clicked. and L5 Intervertebral space. (Evenly spaced, if possible.)

To define the ROI in Whole Mode:


Step Action
1 Click the Global ROI button in the top left corner of the
window, if not already selected.
The ROI box displays in the image area.

8-18 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Step Action
2 Click the Whole Mode button in the Global ROI Toolbox.
The ROI box displays as active lines (dashed yellow).
3 Position the cursor arrow anywhere within the ROI box.
The crossed arrows cursor displays.
4 Center the ROI box over the spine by clicking and
dragging it to the correct position.
Point Mode
The point mode is used to adjust the ROI to match the lateral
curvature or scoliosis of the spine. This is explained in the
Scoliotic Spine Scans on page 8-30 of this chapter.

Viewing/Editing the Bone Map


The system creates a yellow bone map of the spine over the ROI
defined in the previous steps. It automatically fills holes in the
yellow bone map and removes bone artifacts.
KP0886-0821

To view the bone map:


Step Action
1 Click the Bone Map button.
The bone map displays in yellow.

KP0886-0822

2 Verify whether the bone map is acceptable or needs to be


edited.

Performing and Analyzing the AP Lumbar Spine Scan 8-19


QDR Reference Manual

Evaluate the bone map:


It is strongly recommended that the operator not alter the bone
edges of the spine that are displayed in yellow. This degrades the
reproducibility of the spine image. If large areas of the bone map
are automatically excluded, the operator must edit the bone map.
To edit the bone map:
To add or delete bone from the image, use the Bone Map Toolbox
portion of the Analysis window. The operator can also adjust the
cursor size, fill holes, undo changes, or reset the image to the
original bone map if unintentionally changed.

KP0886-0823

To add bone to the bone map:


Step Action
Important: Do not alter the bone 1 Click the Add Bone radio button.
edges of the spine, as it degrades The cursor displays as a small green rectangular paint
the reproducibility of the scan. If
brush when placed inside the global ROI on the image.
necessary, use the optional Low
Density method to improve bone 2 Select the cursor size using the Cursor Size tool.
mapping. See the Scanning Special Hologic recommends the default paint brush (size 2) be
Conditions section of this chapter.
used.
3 Move the cursor to the starting point.

8-20 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Step Action
Note: Be careful to follow the bone 4 Click and drag the cursor to connect outside edges to
edge to bridge all gaps. If there are complete the bone map.
any gaps, the Fill Holes function
will not work. 5 To automatically fill the remaining holes, click the Fill
Holes button.
To delete bone from the bone map:
Step Action
Note: For spurs or prominent 1 Click the Delete Bone radio button.
Transfers Processes, retain the The cursor displays as a small green rectangular paint
image as painted by the software
brush when placed inside the global ROI on the image.
to maintain the reproducibility of
the scan. 2 Select the cursor size using the Cursor Size tool.
Hologic recommends the default paint brush (size 2) be
used.
3 Move the cursor to the starting point.
Note: Be careful to delete a 4 Click and drag the cursor to erase a line through the yellow
complete line through the bone bone map.
map until the image separates.
This creates a small island—an isolated section of bone.
Note: If you leave any connection, 5 To automatically remove the small island, click on the Sink
only the line you deleted is Islands button.
removed.
Small island is removed.

Marking Intervertebral Spaces


Marking each intervertebral space with a line allows each
individual vertebra to be analyzed separately. The Vertebral
Lines button is used to mark the intervertebral spaces within the
KP0886-0824
region of interest. Up to three lines appear on the image that can
be used to mark the spaces. You may insert up to three lines or
delete lines as necessary.

To mark intervertebral lines using Line Mode:


Step Action
1 Click the Vertebral Lines button, if not already selected.
Three horizontal lines display in the image area.
Note: Dashed yellow lines are 2 The Line Mode button is the default and should be active
active and can be edited. Solid red (click if not already selected).
lines are inactive.
The top line displays as an active line (dashed yellow).
Note: The yellow dashed line is 3 Click and drag the top dashed yellow line and place it in
active until the next line is the L1/L2 space, evenly spaced between vertebrae.
selected.
The top line displays as active until you select the next line.

Performing and Analyzing the AP Lumbar Spine Scan 8-21


QDR Reference Manual

Step Action
Note: Active yellow dashed lines 4 Then click and drag the middle line and place it in the L2/
can not cross a solid line except in L3 space, evenly spaced between vertebrae.
point mode.
The middle line displays as active until you select a
different line.
5 Click and drag the lowest intervertebral line and place it in
the L3/L4 space, evenly spaced between vertebrae.
The lowest line displays as active until you select a
different line.

KP0886-0825

To remove a line:
Step Action
Note: Use if analysis is three 1 Click on the line.
vertebrae or less. Line displays as active (dashed yellow).
2 Click the Delete Line button.
Line disappears.

To insert a line:
Step Action
Note: A maximum of three 1 Click the Insert Line button.
vertebral lines can appear on the A new line displays as active (dashed yellow), just above
image.
the previous active line, if any.

8-22 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Point Mode
Point mode is used to mark intervertebral spaces of the Scoliotic
space and is explained in Scanning Special Conditions in this
chapter.

Labeling The Vertebral Bodies


The Results button is used to automatically label the marked
vertebral bodies. Vertebra labels are automatically placed in
areas outside the bone region when possible.
KP0886-0825

Labeling Guidelines
Use the following guidelines to verify labels:
• Labels are automatically assigned numbers starting at the
top with L1.
• The top-most vertebra within the ROI is always labeled
L1.

Including or Excluding Vertebrae from Analysis


Once the operator verifies that the labels correspond to the correct
vertebral bodies, specific vertebrae can be included or excluded
from the analysis. Some reasons for excluding vertebral bodies
include: foreign objects (artifacts) within the scan area, the scan
area is a surgical location, or a physician requested the exclusion
(physician’s discretion).
To automatically label vertebrae
Step Action
1 Click the Results button, if not already selected.
The vertebral bodies are automatically labeled starting
with L1.
2 Verify that the labels correspond correctly to the anatomy.
Usually no changes are necessary.

Performing and Analyzing the AP Lumbar Spine Scan 8-23


QDR Reference Manual

KP1035A-0826

To change vertebrae labels:


Step Action
1 To change the label numbering, click the or arrows in
the First Label area of the Label Toolbox.
The label numbering on the image increments or
decrements accordingly. The first (top-most) label displays
in the First Label box.
2 To include or exclude a vertebra, click anywhere within the
boundary of the labeled vertebra on the image.
The ROI boundary for the selected vertebra displays as
active (dashed yellow).
3 To exclude, click the Exclude button.
The label on the image turns red indicating that the
vertebra has been removed.
4 To include previously excluded vertebra, click the Include
button.
The label on the image turns yellow indicating that the
vertebra has been included.

8-24 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Viewing Results
Results of the analysis display in the lower right portion of the
Analysis window (see Figure 8-7).
KP0886-0827

The spine results area of the Analysis window shows:


• Pertinent Statistical Information
The headings of the results section indicate the statistical
figures that are used in maintaining accuracy and
diagnostic precision for this machine, and are set at the
factory.
• Regions
Indicates individual vertebra included in the analysis and
points to the results.
• Area
Area expressed in cm² (centimeters squared).
• BMC
Bone mineral content expressed in grams.
• BMD
Bone mineral density expressed in g/cm² (grams per
centimeter squared).
• Auto Low Density Spine Analysis
The scan is labeled Auto Low Density Spine Analysis if
the auto low density analysis software was used.

Performing and Analyzing the AP Lumbar Spine Scan 8-25


QDR Reference Manual

Figure 8-7 Results button


AP Lumbar Spine
Analysis Window

Auto Analyze button KP1068_002-0828

Results area
To view results:
Step Action
1 The Results button should have already been selected
when labeling vertebrae. (Click if not already selected).
The results display in lower right corner of the Analysis
window.
To reanalyze the current scan with different settings:
Step Action
Note: Once an Analysis Step 1 Click the appropriate Analysis Step Button (see Figure 8-
Button is modified, each 5) and modify.
subsequent button (analysis step)
must be performed or results will 2 Repeat for each subsequent Analysis Step Button in the
not be accurate. Global ROI Toolbox.
To reanalyze a scan using Auto Analysis:
Step Action
1 Click the Auto Analyze button.
Auto Analyze overrides any manual changes made and
uses the original scan data.
2 Evaluate the image.
3 Click the Close button.

8-26 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

To reanalyze the current scan from the main window:


Step Action
1 Click the Analyze Scan button.
2 Click on the Analyzed Scans tab.
3 Select the desired scan.
4 Click the Next >> button.
5 Continue with analysis as described earlier in this
chapter.

Exiting the Analysis


To exit the analysis:
Step Action
Note: This applies to both auto 1 Click the Close button in the lower left corner of the
and manual analysis of the scan. Analysis window.
The Exit Analysis window displays. Refer to Exiting the
Analysis in Chapter 7 for detailed instructions.

Generating and Printing Reports


A variety of reports can be generated from the results of the
analysis, customized to fit the needs of the examination. Generate
and print the desired report. Refer to Generating and Printing
Reports on page 7-32 of Chapter 7 for more information.

Alternate Scan Modes


You may need to use an alternate scan mode if:
• The patient was previously scanned using a scan mode
other than the default.
• The patient is very large (a longer scan time may improve
the image quality).
To choose a specific scan mode from the Select Scan Type
window:
Step Action
1 Click on the AP Lumbar Spine exam type in the list.
The selected scan type is highlighted.
2 Remove the default scan mode check mark by clicking in
the check box.
Check mark removed.

Performing and Analyzing the AP Lumbar Spine Scan 8-27


QDR Reference Manual

Step Action
3 Click the Next >> button to continue.
The Select AP Lumbar Spine Scan Mode window displays.

KP0886-0829

4 Click on the desired scan mode in the list.


The selected mode is highlighted.
5 Click the Next >> button to continue.
The Scan Parameters window displays. Continue with
Confirming Scan Parameters on page 8-3.

Evaluating the Image


This section provides examples that illustrate unacceptable scans
of the lumbar spine.
Figure 8-8
Unacceptable Lumbar Spine
Scan: Off Center

KP0886-0830

8-28 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Figure 8-9
Unacceptable
Lumbar Spine Scan:
Starting Point Too Low
and Scan Stopped Too Soon

KP0886-0831

Figure 8-10
Unacceptable Lumbar Spine
Scan: Starting Point Too High

KP0886-0832

Performing and Analyzing the AP Lumbar Spine Scan 8-29


QDR Reference Manual

Scoliotic Spine Scans


Some patients may exhibit lateral curvature or scoliosis of the
spine on the scan image. Analysis of the AP Lumbar Spine with
scoliosis may require two modifications to the standard AP
analysis.

Note: Be sure to maintain the 1. The yellow ROI box may need to be angled along the top or
automatically preset distance of bottom of the box, possibly both, depending on the curva-
“116” between the left and right
ture.
lines.
2. The intervertebral line may need to be “tilted” to match the
angle of the intervertebral spaces.

Adjusting the ROI Box


Auto Analyze does not adjust for spine curvature, so manual
analysis must be performed.
The point mode button allows the operator to adjust the ROI to
match the curve of the patient’s spine. Adjust the ROI box so that
the top-most and bottom-most angled lines are parallel to the
desired inter-space, making sure that both vertical lines and
portions of the top and bottom lines remain straight.

8-30 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Figure 8-11
Scoliotic Spine: ROI

KP0886-0833

To define the ROI for a scoliotic spine in Line Mode (top line):
Step Action
1 Click Line Mode in the Global ROI Toolbox.
2 Click the top line of the ROI box, and slide it to the
highest point of L1, or the highest vertebra included in
the analysis.
3 Click the bottom line of the ROI box and slide it to the
lowest point of L4, or the lowest vertebra included in the
analysis.
KP0886-0834

Performing and Analyzing the AP Lumbar Spine Scan 8-31


QDR Reference Manual

Figure 8-12
Scoliotic Spine: Point Mode

KP0886-0835

To define the ROI for a Scoliotic Spine in the Point Mode:


Step Action
Note: Do not bend sides of ROI 1 Click Point Mode in the Global ROI Toolbox.
box.
Note: Make sure there is at least 2 Click and drag the middle point on the top line left or
one 90o corner on top and right to the higher side of L1. Do not drag the point up or
bottom of the ROI. down.
3 Click and drag the corner point of the lower side of L1
down so that the line is placed in the intervertebral space,
parallel to the vertebral bodies. Do not drag this point left
or right.
4 Click and drag the bottom middle point, left or right, to
the lower side of L4. Do not drag this point up or down.
5 Click and drag the corner point of the higher side of L4 so
that the line is placed in the intervertebral space, parallel
to the vertebral bodies. Do not drag this point left or right.

8-32 Performing and Analyzing the AP Lumbar Spine Scan


QDR Reference Manual

Marking Intervertebral Spaces of Scoliotic Spines


Marking intervertebral spaces of scoliotic spines is similar to
marking spaces in normal spines except you must tilt the line to
keep the intervertebral lines parallel to the spaces.

Pivoting an Intervertebral Line


To pivot an Intervertebral Line use the Point Mode button to
match the intervertebral lines to the curve of the spine. A line can
be pivoted at either end, but cannot be bent in the middle.
Make sure the lines are kept in their original order and they do not
pass over one another.
Figure 8-13
Scoliotic Spine:
Marking Intervertebral Spaces

KP0886-0836

To mark the intervertebral lines of a scoliotic spine:


Step Action
1 Click the Vertebral Lines button to mark the
intervertebral spaces within the ROI.
The system automatically inserts three lines.
KP0886-0824

Performing and Analyzing the AP Lumbar Spine Scan 8-33


QDR Reference Manual

Step Action
2 Click and drag to select and move the line up or down
close to the intervertebral space. Start with the bottom
line and move up.
3 Click the Point Mode button to change to point mode.
Two points appear, one at each end of the 3 lines.
Note: An intervertebral line 4 Drag the point up or down to match the angle of the
cannot completely cross over intervertebral space. You can pivot the line from either
another line or it will mislabel at
end, but you cannot bend the line in the middle.
the next step.
5 Click the Delete Line button to delete a line.
6 Click the Insert Line button to insert a new active line
above a previous line (max 3 lines).
7 Repeat steps 3 through 7 to mark each space.
Figure 8-14
Scoliosis Spine: Marking
Intervertebral Spaces in Point
Mode

KP0886-0837

8-34 Performing and Analyzing the AP Lumbar Spine Scan


Chapter 9
Performing and Analyzing the
Hip Scan
This chapter provides procedures to perform a Total Hip DXA
scan and analysis for the QDR systems. The anatomical sites that
are measured for the Total Hip scan are: the femoral neck, the
greater trochanter, the intertrochanter and the Ward’s triangle.

Content
Subject Page
The Hip Examination 9-2
Preparing the Patient 9-2
Creating/Retrieving a Patient Biography 9-2
Selecting the Scan Type and Mode 9-3
Positioning the Patient and C-arm 9-6
Performing the Examination 9-8
Exiting the Examination 9-13
Performing the Analysis 9-14
Exiting the Analysis 9-29
Generating and Printing Reports 9-29
Alternate Scan Modes 9-29
Evaluating the Image 9-30
Modifying Other Regions of the Hip 9-33
About the Midline 9-33
About Ward’s Triangle 9-35
About the Trochanter Region 9-35
Hip Structure Analysis® 9-38

9-1
QDR Reference Manual

The Hip Examination


The Hip examination includes all activities that begin when the
patient enters the exam room, and end when an acceptable image
is obtained. The activities involved in performing the exam
include:
• Preparing the Patient
• Creating/Retrieving a Patient Biography
• Selecting the Scan Type and Mode
• Positioning the Patient and C-arm
• Performing the Examination
• Exiting the Examination
• Performing the Analysis
• Exiting the Analysis
• Generating and Printing Reports

Preparing the Patient


Note: Remember to ask the Prepare the patient by asking the pre-examination questions and
patient if they are, or might be, ensuring that they are not wearing anything that will interfere
pregnant. Postpone the exam
with the scan. Refer to Preparing the Patient on page 7-2 of
until pregnancy is ruled out.
Chapter 7 for detailed instructions.

Creating/Retrieving a Patient Biography


If this is a new patient, create a new patient biography (refer to
Creating/Editing a Patient Record on page 6-9 of Chapter 6 for
detailed instructions). If this patient already has a biography,
retrieve it now (refer to Retrieving a Patient Record on page 6-7
of Chapter 6).

9-2 Performing and Analyzing the Hip Scan


QDR Reference Manual

Selecting the Scan Type and Mode


After a patient biography has been created, or retrieved, the Select
Scan Type window displays.

Note: Hologic suggests scanning


the patient’s non-dominant hip
(does not apply to Dual-Hip
scans).

KP0886-0901

The scan type for the Hip scan is either Left Hip or Right Hip
and is selected in the Select Scan Type window. A check mark
() in the Use Default Scan Mode box designates the Express
mode for Discovery-A, -SL, -W, -C models, Fast Array mode for
Discovery-Wi, Ci models, or Explorer mode for Explorer.

Choosing the Scan Mode


To choose an alternate scan mode:
If you need to perform the scan in a different mode, remove the
check mark and a window displays allowing you to select an
alternate scan mode.

Left or Right Hip


To choose the default scan mode for Left or Right Hip (check
mark in box):
Step Action
1 Click Left Hip or Right Hip in the scan type list.
The scan type is highlighted.
2 Click the Next >> button to continue.
The Scan Parameters window appears.

Performing and Analyzing the Hip Scan 9-3


QDR Reference Manual

Dual-Hip
In Dual-Hip mode, you must position the patient before you
select the scan mode (see Positioning the Patient and C-arm on
page 9-6).
To choose the default scan mode for Dual-Hip (check mark in
box):
Step Action
1 Click Dual-Hip in the scan type list.
The scan type is highlighted.

KP0886-0902

2 Click the Next >> button to continue.


A massage appears telling you to center the laser on the
patient.
3 Center the laser on the patient by moving the table in or
out.
You can position the laser anywhere on the midline of the
patient.
4 Click Continue.
The Scan Parameters window appears.

9-4 Performing and Analyzing the Hip Scan


QDR Reference Manual

Confirming Scan Parameters


The only scan parameter that can be modified is the scan length
(see Figure 9-1 for left hip or Figure 9-2 for right hip). The scan
length for a hip exam is approximately 6 in. (15cm). The system
may change the number entered to match the pre-programmed
step size of the scanning mechanism.
Figure 9-1
Left Hip Scan Parameters

KP1035A-0903

Figure 9-2
Right Hip Scan Parameters

KP1035A-0903a

To verify the scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.

Performing and Analyzing the Hip Scan 9-5


QDR Reference Manual

Step Action
2 To change scan length, place the cursor in the Scan Length
field and type the new length.
3 Stop here and position the patient and the C-arm.

KP0886-stop

Positioning the Patient and C-arm


The goal for positioning the patient on the table is to ensure that
the hip is as straight as possible for the scan (refer to Figure 9-3).
Figure 9-3 Left Hip Patient Positioning Goal
Hip Patient Positioning Goal

Foot Positioner

Area to be scanned
KP0886-0904

The remainder of this section provides detailed information on


positioning the patient, foot positioner, table, and C-arm for a hip
exam. It is important to follow each set of directions carefully
since they are designed to help acquire an ideal scan by
describing:
• How to maximize patient comfort and safety.
• How to assist the patient onto the table for proper
alignment.
• Suggestions for correct positioning techniques.
• Proper use of positioning aids.

9-6 Performing and Analyzing the Hip Scan


QDR Reference Manual

Placing the Patient on the Table


Note: Refer to Chapter 7, page Press the Patient On/Off switch on the Control Panel and seat
7-21, for a detailed description the patient on the table between the scan field indicators on the
of placing the patient on the
table pad. Their right shoulder should be adjacent to the C-arm.
table.
Continue with positioning the patient below.

Positioning the Patient


Note: The Dual-Hip scan begins Positioning the patient for a hip scan involves using the Foot
with a scan of the left hip. Positioner (see Figure 9-1 for left hip or Figure 9-2 for right hip),
supplied by Hologic. This positioner helps to align the patient’s
hip and holds the foot firmly in place.

A pillow can be used under the patient’s head to make them more
comfortable.
To position the patient:

Step Action
Note: Patient must remove their 1 Lie the patient on their back on the table with their head at
shoes. the right end of the table (see Figure 9-1).
Note: On Explorer the Center 2 Press the Center Table button on the Control Panel.
Table button must be held and only The table and C-arm move to the center.
the table moves.

3 Place the Foot Positioner under the patient’s legs, and


align the center of the positioner with the patient’s midline.
Note: The foot should not be 4 Rotate the patient’s entire leg to be scanned (from hip
elevated above the positioner nor socket to foot) 25° inward and place the medial edge of the
hanging off the edge.
foot against the triangle. The foot should be flexed towards
the ceiling (see Figure 9-1).

5 Adjust the Velcro® strap to hold the foot in the correct


position.
Velcro
Strap 6 Align the femur to be parallel with the table edge to
KP0886-0906 provide adequate space for the neck box. Abduct the leg
from the midline of the body in order to straighten the
femur.
To ensure the femur is parallel with the long axis of the
table, you can offset the foot positioner while keeping it
parallel with the front edge of the table pad, if necessary.

Performing and Analyzing the Hip Scan 9-7


QDR Reference Manual

Positioning the C-arm


Note: There are different The C-arm and table drive motors provide the table and C-arm
procedures for the Left Hip and motion required to scan the X-ray beam along the patient’s hip
Right Hip scans versus the Dual-
area.
Hip scan (see below).

Positioning for the Left Hip or Right Hip Scan


To position the C-arm and align the Laser:
Step Action
Tips: To help identify the greater 1 Press the Laser button on the Control Panel.
trochanter: The laser cross-hair indicators display on the table pad.
• Place your thumb on the
iliac crest, spread your
fingers. Direct your little 2 Use the Arm Controls on the Control Panel to place the
finger toward the knee. cross hair of the laser 3 inches below the greater trochanter
The greater trochanter is and 1 inch medial to the shaft of the femur.
located under your little
finger.
3 Instruct the patient to remain still and to breathe normally.
• Palpate for the symphysis
pubis. The greater
trochanter is at the same
level.

Positioning for the Dual-Hip Scan


To position the C-arm and align the Laser:
Step Action
1 Turn on the laser and position the C-arm over the center of
the patient.
2 Click Continue.
Note: In Dual-Hip mode, the left 3 Palpate the patient’s left leg to identify the greater
hip is always scanned first. trochanter (see Tips above).
4 Use the Arm Controls to place the cross hairs of the laser
3 inches below the greater trochanter and 1 inch medial to
the shaft of the femur.
5 Click Continue.
6 Instruct the patient to remain still and to breathe normally.

Performing the Examination


An acceptable hip scan (see Figure 9-4) includes the following:

9-8 Performing and Analyzing the Hip Scan


QDR Reference Manual

• The femoral shaft is straight and parallel to the edge of the


scan image.
• The greater trochanter is centered vertically in the
window.
• The entire femoral head is visible.
• A 25° internal rotation of the hip showing minimal or no
lesser trochanter on the scan image.
• For Hip Axis Length (HAL) or Hip Structure Analysis
(HSA) options, ensure that a small amount of pelvic inlet
is shown on the scan image.
Figure 9-4
Right Hip Scan

KP0886-0907

Starting the Hip Scan


Once the patient and C-arm are positioned correctly, return to the
computer to start the Hip scan. At this time, the Scan Parameters
window should be displayed.

Performing and Analyzing the Hip Scan 9-9


QDR Reference Manual

To start the Hip scan:


Step Action
Note: If the Control Panel X-ray 1 Click the Start Scan button.
lamp fails to extinguish within 10 The Scan window displays with the image appearing on the
seconds after the end of the scan,
left side. Flashing X-rays On indicator at the top of the
press the red Emergency Stop
button on the Control Panel window continues until the scan stops.
immediately. Then call Hologic 2 Inspect the image as it is generated.
Service before resuming operation.
• If the hip is positioned correctly, allow the scan to
complete.
The Exit Exam window displays.
• If the hip is not positioned correctly, click the
Reposition Scan button.
The scan stops to allow repositioning (see
Repositioning the Scan below).
While the scan is being acquired the operator has four options:
Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the Exit Exam
window displays.
Stop Scan button Terminates the scan at the point at
which the button is pressed and the
KP0886-0808 Exit Exam window displays.
Abort Scan button Exits the scan immediately. An exit
window displays then proceeds to
KP0886-0809 the main window. To start another
scan the C-arm must be
repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP0886-0810 either the window image or the
patient. Refer to the following
instructions.

Repositioning the Scan


If your scan is positioned correctly (see Figure 9-4) it is not
necessary to reposition. If you need to reposition, follow the
procedure below.
Two basic activities of repositioning include:
• Repositioning the image
Reposition the image up, down, left or right using the

9-10 Performing and Analyzing the Hip Scan


QDR Reference Manual

scroll bars or cursor hand to reposition the scan to include


the entire femoral head, neck, and approximately 3 inches
of the shaft.
• Repositioning the patient
If the femoral shaft is not straight and parallel with the
edge of the scan image, move the patient’s leg away from,
or towards, the midline (as appropriate).
Re-rotate the leg if it is over or under rotated depending on the
lesser trochanter visibility.
To reposition the image:
Step Action
Note: In order to reposition the 1 Click the Reposition Scan button
scan, the Reposition Scan button The image acquired so far displays with scroll bars on the
must be clicked before the scan is
right and bottom.
completed.
Note: You can also use the scroll 2 Position the cursor over the hip image.
bars to reposition the image. The arrow cursor changes to a hand.

Vertical
Positioning
Line
Greater
Trochanter
Horizontal
Positioning
Line
Scroll
Cursor Bars
Hand
KP0886-0908

Note: On Explorer, the scan can be 3 To position the hip, click and drag the image to the
repositioned a maximum of 1 inch intersection of the two blue positioning lines so that:
(2.54 cm.) towards the foot end of
• The vertical positioning line touches the outer edge
the table and any distance towards
the head end. of the Greater Trochanter
• The horizontal position
4 When the hip is positioned correctly, click the Restart
Scan button.
The Scan Parameters window displays.
5 Click the Start Scan button to start a new scan at the new
position.
The Scan window displays with a flashing X-rays On
message. The image displays.
While repositioning the scan the following buttons are displayed
in the window:

Performing and Analyzing the Hip Scan 9-11


QDR Reference Manual

Button Description
Restart Scan button Proceeds to the Scan Parameters
window to commence X-ray scan
KP0886-0909
from the new position.
Accept Scan button Saves the scan in progress without
KP0886-0910
repositioning the scanner.

Cancel Scan button Cancels the exam and returns to


KP0886-0911
the main window.

To reposition the patient:


Step Action
Note: In order to reposition the 1 Click the Reposition Scan button.
scan, the Reposition Scan button The image acquired so far displays.
must be clicked before the scan
is completed. 2 If the femoral shaft is not straight and parallel with the
edge of the scan image, move the patient’s leg away
from, or towards, the midline (as appropriate).
3 When the hip is positioned correctly, click the Restart
Scan button.
The Scan Parameters window displays.
Note: If the Control Panel X-ray 4 Click the Start Scan button to start a new scan at the new
lamp fails to extinguish within position.
10 seconds after the end of the
The Scan window displays with a flashing X-rays On
scan, press the red Emergency
Stop button on the Control Panel message. The image displays.
immediately. Then call Hologic 5 Evaluate the image as it displays. If the image is
Service before resuming
satisfactory, allow the scan to complete.
operation.
When the scan completes, the Exit Exam window
displays.

Additional Steps for the Dual-Hip Scan


If you are performing a Dual-Hip scan, the scanner automatically
moves from the left hip to the right hip after finishing the first
scan.
As the scanner moves to the right hip, the following message box
appears on the computer screen.

9-12 Performing and Analyzing the Hip Scan


QDR Reference Manual

KP0886-0912

When this happens, do the following:

Step Action
1 Move the foot positioner from the left foot to the right
foot.
2 Align the patients hip, leg and foot as you did for the left
hip scan.
3 Fasten the patients foot in the foot positioner.
4 Turn on the laser and make sure that the scanner is
properly positioned over the patient’s right hip.
5 Click Continue.
6 Verify an acceptable right hip scan has been acquired. If
not, reposition the image or the patient as necessary and
rescan.

When the scan has finished, exit the exam.

Exiting the Examination


After acquiring an acceptable image, the Exit Exam window
displays. To perform the analysis of the scan just completed, click
the Analyze Scan button in the Exit Exam window. Refer to
Exiting the Examination on page 7-23 in Chapter 7 for detailed
instructions.

Performing and Analyzing the Hip Scan 9-13


QDR Reference Manual

Performing the Analysis


Note: If the analysis of the scan The analysis of the Hip scan includes the following:
is to be postponed to a later • Choosing the analysis method
time, the scan can be retrieved
Note: If the patient has a respective (left or right) Hip
from the Scan Drawer. Refer to
“Scan Drawer” on page 6-16, baseline scan, the software proceeds directly to the
for information about the scan manual compare analysis function instead of choosing
drawer. the analysis method. Compare analysis is discussed in
detail in Chapter 14. Manual Hip scan analysis is
discussed beginning on page 9-17.
• Performing the analysis

Choosing the Method of Analysis


The first analysis activity involves selecting the analysis method.
The analyze method is the method that the system uses to perform
the analysis. It can perform the analysis either as a single new
scan, or as a comparison to one that has been previously
performed.
If this scan is the patient’s first visit, select the default method
(Hip) and the system presents a single scanned image for
analysis. If this scan is a follow-up visit, select the Compare to
Previously Analyzed Scan radio button. The system then
presents two images. The left image is the unanalyzed scan, and
the right image is a previous scan with the region of interest
(ROI) defined.
Compare allows the system to perform the new analysis on the
same ROI as the previous scan and maximizes accurate reporting
of changes. Follow-up visits are discussed in detail in
“Performing the Follow-up Scan” on page 14-7 of Chapter 14.

9-14 Performing and Analyzing the Hip Scan


QDR Reference Manual

To select analyze setup for a first visit:


Step Action
1 From the Exit Exam window, click the Analyze Scan
button.
The Analyze Setup window displays.

Note: The Analyze Setup window 2 Click the Choose Analysis Method radio button, if
displays with the Choose necessary.
Analysis Method radio button
selected as the default. 3 Verify the analysis method (see Methods of Analysis
The Comparison method is below).
covered in detail on page 4 Click the Next >> button.
“Performing the Follow-up
Scan” on page 14-7 of Chapter
The Analysis window displays.
14.

Methods of Analysis
The Hip analysis is the standard method for analyzing Hip scans.
To select a method of analysis (other than default):
Step Action
Note: Subregion Array Hip is 1 Click the drop down list arrow on the Analyze Setup
available as an option. window.
List of methods displays.

2 Click on the desired method.


List disappears, and selected method displays in field.
3 Click the Next >> button to continue.
The Analysis window displays.

Performing and Analyzing the Hip Scan 9-15


QDR Reference Manual

Performing the Analysis


Note: Refer to “Analyze Tab” on Unless disabled, once the method of analysis is chosen the
page 18-5 of Chapter 18 on how software will automatically perform the analysis (excluding
to disable One-Time™ Auto
Subregion Array Spine analysis). This is called One-Time™
Analysis.
Auto Analysis and is a feature of all models in the QDR Series.
After the One-Time™ Auto Analysis is completed the operator
can choose to perform a manual analysis of the acquired scan.
Both analysis methods are described below.

One-Time™ Auto Analysis


Note: Hologic’s auto-analysis The Hip scan auto analysis includes all activities that begin with
algorithms have been optimized a satisfactory image and ends with the calculation and summary
for use on patients. You may be
of the results of the scan. After completing a scan, the system
required to manually analyze
scans of phantoms that were automatically saves the scan, in the Scan Drawer, for analysis.
scanned as if they were patients. Once the method of analysis is selected, the system will
automatically perform an analysis of the scan (excluding
Subregion Array Hip analysis). While this operation is being
performed, no operator interaction is allowed. While the analysis
Note: Refer to “Analyze Tab” on
page 18-5 of Chapter 18 on how
is being performed, system software will use an automatic low
to disable Auto Low Density Hip density hip analysis (unless disabled) if it is needed. Upon
analysis. completion of Auto Analysis, analysis results are displayed. If
the software determined automatic low density hip analysis was
required, it will be noted in the analysis results label.
Figure 9-5
Hip Auto Analysis Results

9-16 Performing and Analyzing the Hip Scan


QDR Reference Manual

Manual Analysis
The operator may perform a manual analysis of the Hip scan after
the auto analysis is completed. In most cases it will not be
necessary to perform manual analysis. However, upon
examination of the auto analysis results, if one of the following
conditions is seen, manual analysis may be required:
• regions are positioned incorrectly
• poor bone mapping
To perform manual analysis after auto analysis:
Step Action
1 Click on the Global ROI button (Figure 9-6).
2 Perform the Manual Analysis (see “Performing the
Manual Analysis” on page 9-17 of this chapter).

Figure 9-6
Step buttons
Hip Analysis Window
Note: Refer to “Hip Structure
Analysis®” on page 9-38 for
details pertaining to the optional
HSA step button.

Toolbox
Brightness/Contrast button

Performing the Manual Analysis


Once the setup method has been selected, continue to the
Analysis window to adjust the quality of the image and properly
define the area of analysis called the region of interest (ROI).

Performing and Analyzing the Hip Scan 9-17


QDR Reference Manual

A specific set of tools called the Analysis Step Buttons, located


on the left side of the window (see ), allow the operator to
proceed step-by-step through each task. These tools can
maximize image quality and accuracy, preventing the need for re-
scanning.
The steps to analyze the hip scan include:
1. Adjusting the image display brightness and contrast (Bright-
ness/Contrast button).
2. Defining the Global Region of Interest (Global ROI button).
3. Viewing/editing the bone map (Bone Map button).
4. Adjusting the Neck box (Neck button).
5. Viewing analysis results (Results button).
6. Exiting the Analysis window (Close button).

Adjusting the Image Display


The contrast and brightness of the image can be adjusted to
obtain the best definition of important anatomical features.
Adjusting the contrast and brightness has no effect on
KP0886-0817 calculations; it is used to optimize the display and for visual
adjustments only. Refer to Contrast and Brightness on page 7-
25, in Chapter 7, for detailed instructions.

Defining the Region of Interest (ROI)


The global ROI refers to the defined boundaries of the image that
is being analyzed. The ROI appears on the image as a box: the
dashed line is active and can be adjusted; the solid lines are
KP0886-0818
inactive and cannot be adjusted.

9-18 Performing and Analyzing the Hip Scan


QDR Reference Manual

Figure 9-7
ROI Image in the Analysis
Window

KP0886-0818

A properly positioned Global ROI includes:


• Proximal femur in its entirety
• Lesser trochanter
• Top of the femoral head
• Lateral side of the greater trochanter

Performing and Analyzing the Hip Scan 9-19


QDR Reference Manual

The Global ROI Toolbox


The Global ROI Toolbox includes tools to help define the region
of interest:

• Line Mode
When selected, one line on the ROI box initially displays
as a dashed yellow line. Any of the ROI box lines can be
selected and then moved. Top or bottom (horizontal) lines
can be moved up or down. Side (vertical) lines can be
moved left or right.
• Whole Mode
When selected, the entire ROI box initially displays as a
dashed yellow line. This allows the entire box to be
moved. This tool is usually used during a compare scan.

KP0886-0819

To define the ROI in Line Mode:


Step Action
1 Click the Global ROI button in the top left corner of the
window, if not already selected.
The ROI box displays in the image area.
Note: Dashed yellow lines are 2 The Line Mode button is the default and should be active
active and can be edited. Solid (click if not already selected).
red lines are inactive. Dashed
The top line of the ROI box displays as an active line
blue lines are analysis aids.
(dashed yellow) and blue dashed lines as analysis aid
lines.
3 Using the mouse pointer, drag the top yellow line so that
the blue analysis aid line is just above the femoral head.
This indicates the landmark that the system uses to place
the actual line a set distance away from the landmark.
4 Select the right vertical line (for a left hip), then click and
drag it so that the blue analysis aid line is just touching
the lateral side of the greater trochanter.
The line selected in step 3 appears as a solid line a set
distance away from your original line.

9-20 Performing and Analyzing the Hip Scan


QDR Reference Manual

Step Action
Note: If the lesser trochanter can 5 Select the bottom line, then click and drag it so that the
not be identified, place the blue analysis aid line is at the base of the lesser
bottom line two times the
trochanter.
distance of the greater
trochanter.
6 Select the left line (for a left hip), then click and drag it so
that the blue analysis aid line is just lateral to the femoral
head.
To define the ROI in Whole Mode:
Step Action
1 Click the Global ROI button in the top left corner of the
window, if not already selected.
The ROI box displays in the image area.
2 Click the Whole Mode button in the Global ROI
Toolbox.
The ROI box displays as active lines (dashed yellow).
3 Position the cursor arrow anywhere within the ROI box.
The crossed arrows cursor displays.
4 Click and drag the ROI to the correct position.

Viewing/Editing the Bone Map


The system creates a yellow bone map of the hip over the ROI
defined in the previous steps. It automatically fills holes in the
yellow bone map.
KP0886-0821

To view the bone map:


Step Action
1 Click the Bone Map button.
The Bone map displays in yellow.
2 Verify whether the bone map is acceptable or needs to be
edited.

Performing and Analyzing the Hip Scan 9-21


QDR Reference Manual

Step Action

KP0886-0820

Evaluate the bone map:


Evaluate the image to determine if any erroneous bone mapping
should be excluded or any excluded bone should be added.
• Unacceptable Bone Map
Look for any bone that was erroneously excluded from the
bone map, such as holes or edges that the bone map
process did not properly fill. If this is the case, proceed to
To Edit the Bone Map below.
• Acceptable Bone Map
The bone map should include the entire hip region, as
specified during the process of defining the ROI. There
are no holes, or divots, or sections that should be filled in.
If the bone map is acceptable, proceed to Positioning the Neck
Box.
To edit the bone map:
The “Edit Bone Map Feature” is intended to correct small, local
aberrations in the bone map. Some examples include small holes
in the bone that did not fill in properly or irregular regions of soft
tissue that were misidentified as bone. It should not be used to fill
in significant portions of the bone map. As a general rule no more
than about 5% of the bone in a given region should be filled in
manually.

9-22 Performing and Analyzing the Hip Scan


QDR Reference Manual

If significant portions of the bone in the femur were not


identified, increase the global ROI height by clicking on the
“Global ROI” button and moving the upper global ROI up 10-20
lines toward the top of the image. Then click the “Bone Map”
button and verify that the resulting bone map has improved.
Usually, increasing the ROI height by 10-20 lines will generate an
acceptable bone map but sometimes further adjustments are
necessary.
To add or delete bone from the image, use the Bone Map Toolbox
portion of the Analysis window. The operator can also adjust the
cursor size, fill holes, undo changes, or reset the image to the
original bone map if unintentionally changed.
In some cases, it may be necessary to delete bone from the bone
map. It may be necessary to delete the pelvis (ischium) if it is too
close to the femoral shaft. Otherwise, there may be difficulty
placing the Neck box in the next step of the analysis.

KP0886-0821

To Add bone to the Bone Map:


Step Action
1 Click the Add Bone radio button.
The cursor displays as a small green rectangular paint
brush when placed inside the global ROI on the image.
2 Select the cursor size using the Cursor Size tool.
Hologic recommends the default paint brush (size 2) be
used.
3 Move the cursor to the starting point.

Performing and Analyzing the Hip Scan 9-23


QDR Reference Manual

Step Action
Note: Be careful to follow the 4 Click and drag the cursor to connect outside edges to
bone edge to bridge all gaps. If complete the bone map.
there are any gaps, the Fill
Holes function will not work.
5 To automatically fill the remaining holes, click the Fill
Holes button.
How to determine a starting point when deleting bone
If the pelvis (ischium) is too close to the femoral shaft, it
interferes with the next step, which is placing the Neck box. To
delete the pelvis, you need to draw a line starting from the inside
edge of the femoral shaft where it joins with the femoral neck and
trace a line up to the level of the base of the femoral head.
Continue the line horizontally (laterally) to the edge of the ROI.
To delete bone from the bone map:
Step Action
1 Click the Delete Bone radio button.
The cursor displays as a small green rectangular paint
brush when placed inside the global ROI on the image.
2 Select the cursor size using the Cursor Size tool.
Hologic recommends the default paint brush (size 2) be
used.
Note: Refer to How to determine 3 Move the cursor to the starting point.
a starting point when deleting
bone for information on the
starting point.
Note: Be careful to delete a 4 Click and drag the cursor to erase a line through the
complete line through the bone yellow bone map.
map until the image separates.
This creates a small island—an isolated section of bone.
Note: If you leave any 5 To automatically remove the small island, click on the
connection, only the line you Sink Islands button.
deleted is removed.
Small island is removed.

Positioning the Neck Box


This step of the analysis allows the operator to position the
femoral Neck Box, which is usually the only adjustment for a hip
scan, other than placing the ROI.
KP0886-0822

In this step, verify the following:


• The Neck Box covers only the femoral neck.

9-24 Performing and Analyzing the Hip Scan


QDR Reference Manual

• The box should not include any area of the ischium


femoral head, or greater trochanter.
When positioning the Neck Box, move the box as a whole along
the midline until the upper outer corner is positioned at the notch
of the greater trochanter. The remaining three corners of the Neck
Box should remain in soft tissue.

KP1035A-0823

To recalculate the locations of all hip regions after manually


editing the global ROI or bone map:
Step Action
1 Click the Neck button, if not already selected.
The femoral Neck Box and related lines display in the
image area.
2 Click the Auto Position button.
The system recalculate all hip regions based on the
current bone map.

Performing and Analyzing the Hip Scan 9-25


QDR Reference Manual

To position the Neck Box in Whole Mode:


Step Action
1 Click the Neck button, if not already selected.
The femoral Neck Box and related lines display in the
image area.
Note: Dashed yellow lines are 2 The Whole Mode button is the default and should be
active and can be edited. Red active (click if not already selected).
lines are inactive.
The Neck Box displays as active lines (dashed yellow).
3 Click and drag the Neck Box along the femoral midline
(dashed red line) until the upper outer corner of the box is
positioned at the edge of the inside curve of the greater
trochanter. The remaining three corners should remain in
soft tissue. The two lower corners should not overlap the
ischium.

Adjusting the Neck Box


There are two reasons why it would be necessary to adjust the
Neck Box:
• The Neck Box overlaps the ischium.
• The Neck Box overlaps the femoral head.
To move the Neck Box away from the ischium in Line Mode:
Step Action
1 Click the Neck button, if not already selected.
The femoral Neck Box and related lines display in the
image area.
Note: Dashed yellow lines are 2 Click the Line Mode button, if not already selected.
active and can be edited. Red One of the Neck Box lines displays as active (dashed
lines are inactive.
yellow).
3 Select the line on the short side of the Neck Box that is
between the femoral shaft and the ischium.
The selected line displays as active (dashed yellow).
4 Click and drag the selected line towards the neck (the
midline) until the two lower corners of the Neck box are
in tissue (the dark area of the image between the ischium
and femoral shaft). This shortens the length of the box.

9-26 Performing and Analyzing the Hip Scan


QDR Reference Manual

Step Action
5 If the two lower corners clear the ischium, it is not
necessary to perform steps 6-8 below. If the two lower
corners cannot clear the ischium, click the Undo button
to undo the last move. It is not necessary to edit the bone
map and delete the ischium from the bone map image.
Note: Refer to To Edit the Bone 6 Click the Bone Map button and edit the bone map.
Map for instructions.
7 Perform steps 1-3 of the To Position the Neck Box in
Whole Mode procedure on page 9-26.
8 Repeat steps 2-4 of this procedure.
To move the Neck Box away from the femoral head in Line
Mode:
Step Action
1 Click the Neck button, if not already selected.
The femoral Neck Box and related lines display in the
image area.
Note: Dashed yellow lines are 2 Click the Line Mode button, if not already selected.
active and can be edited. Red lines One of the Neck Box lines displays as active (dashed
are inactive.
yellow).
3 Select the top line (long line) on the Neck Box.
The selected line displays as active (dashed yellow).
Note: This is done as a last resort. 4 Click and drag the selected line down the midline until the
By making the box narrower, the Neck Box clears the femoral head.
bone sample is smaller reducing
The Neck Box becomes narrower as you drag the line.
the accuracy of the results.

Viewing Results
Results of the analysis display in the lower right portion of the
Analysis window (see Figure 9-8).
KP0886-0825

The hip results area of the Analysis window shows:


• Pertinent Statistical Information
The headings of the results section indicate the statistical
figures that are used in maintaining accuracy and
diagnostic precision for this machine, and are set at the
factory.

Performing and Analyzing the Hip Scan 9-27


QDR Reference Manual

• Regions
Indicates individual regions of the hip included in the
analysis and points to the results.
• Area
Area expressed in cm² (centimeters squared).
• BMC
Bone mineral content expressed in grams.
• BMD
Bone mineral density expressed in g/cm² (grams per
centimeter squared).
• Auto Low Density Hip Analysis
The scan is labeled Auto Low Density Left, or Right, Hip
Analysis if auto low density analysis software was used.
Figure 9-8 Results button
Analysis Window

KP1068_002-0824
Analysis results

To view results:
Step Action
1 Click the Results button.
The results display in lower right corner of the Analysis
window.

9-28 Performing and Analyzing the Hip Scan


QDR Reference Manual

To reanalyze the current scan with different settings:


Step Action
Note: Once an Analysis Step 1 Click the appropriate Analysis Step Button (see on page
Button is modified, each 9-17) and modify.
subsequent button (analysis
step) must be performed or 2 Repeat for each subsequent Analysis Step Button in the
results will not be accurate. Global ROI Toolbox.

To reanalyze the current scan from the main window:


Step Action
1 Click the Analyze Scan button.
2 Click on the Analyzed Scans tab.
3 Select the desired scan.
4 Click the Next >> button.
5 Continue with analysis as described earlier in this
chapter.

Exiting the Analysis


For All Scans Except Dual-Hip
After performing the analysis, click Close and the Exit Analysis
window displays. To select a report type to print, click on the
Report button in the Exit Analysis window. Refer to Exiting the
Analysis on page 7-27 of Chapter 7 for detailed instructions.

For Dual-Hip Scans


After performing the analysis on the right hip, click Close and the
Exit Analysis window displays. To analyze the left hip just
scanned (as part of the Dual-Hip exam), click Analyze Another
Scan. The software chooses the left hip from the Dual-Hip exam.

Generating and Printing Reports


A variety of reports can be generated from the results of the
analysis, customized to fit the needs of the examination. Generate
and print the desired report. Refer to Generating and Printing
Reports on page 7-32 of Chapter 7 for more information.

Alternate Scan Modes


You may need to use an alternate scan mode if:

Performing and Analyzing the Hip Scan 9-29


QDR Reference Manual

• The patient was previously scanned using a scan mode


other than the default.
• The patient is very large (a longer scan time may improve
the image quality).
To choose a specific scan mode from the Select Scan Type
window:
Step Action
1 Click on the Left Hip or Right Hip exam type in the list.
The selected scan type is highlighted.
2 Remove the default scan mode check mark by clicking in
the checkbox.
Check mark removed.
3 Click the Next >> button to continue.
The Select Left or Right Hip Scan Mode window displays.

KP1068A-0834

4 Click on the desired scan mode in the list.


The selected mode is highlighted.
5 Click the Next >> button to continue.
The Scan Parameters window displays. Continue with
Confirming Scan Parameters on page 9-5.

Evaluating the Image


This section provides examples that illustrate unacceptable scan
images.

9-30 Performing and Analyzing the Hip Scan


QDR Reference Manual

Figure 9-9
Unacceptable Hip Scan: Clipped
the Head

KP0886-0826

Figure 9-10
Unacceptable Hip Scan: Started
Too Low

KP0886-0827

Performing and Analyzing the Hip Scan 9-31


QDR Reference Manual

Figure 9-11
Unacceptable Hip Scan: Poor
Interior Rotation

KP0886-0828

Figure 9-12
Unacceptable Hip Scan: Shaft
not Parallel to Edge of Scan
Window

KP0886-0829

9-32 Performing and Analyzing the Hip Scan


QDR Reference Manual

Modifying Other Regions of the Hip


This section provides instructions for modifying other regions of
the hip. When the Neck button is first pressed, there are two
buttons in the Neck Toolbox: the Neck button and the Other
Regions button. By clicking on the Other Regions button, the
Neck Toolbox expands with additional buttons that allow the
following regions to be modified:

Note: In most cases, it is not • Midline (Midline button)


necessary to adjust these
regions. • Ward’s Triangle (Ward’s Triangle button)
• Trochanter (Trochanter button)

Clicking the Reset button causes the Neck Toolbox to display


only the Neck and Other Regions buttons.

About the Midline


The femoral midline is computed from the outline of the femoral
neck. The system locates the narrowest portion of the neck and
draws the midline midway between the sides of the neck at that
point. The femoral midline is aligned so it is parallel to the long
axis of the femoral neck.

KP0886-0830

KP0886-0831

Performing and Analyzing the Hip Scan 9-33


QDR Reference Manual

To adjust the midline


Step Action
Note: Complete the Global ROI 1 Click the Other Regions button in the Neck Toolbox, if
and Bone Map procedures prior necessary.
to performing this procedure.
The Neck Toolbox displays with additional buttons for
adjusting the midline, Ward’s triangle, and the trochanter.
2 Click the Midline button.
The Neck Toolbox displays additional buttons (Whole
Mode and Line Mode) for adjusting the midline.
Note: Dashed yellow lines are 3 The Whole Mode button is the default and should be
active and can be edited. Solid active.
red lines are inactive.
The midline displays as an active line (dashed yellow).
4 Use the following buttons to select a method for
modifying the midline:
• Whole Mode to move the line left or right of the
midline.
• Point Mode to pivot the midline by selecting
either end (the upper or lower point).
5 Adjust the midline according to the method selected.

9-34 Performing and Analyzing the Hip Scan


QDR Reference Manual

About Ward’s Triangle


Ward’s triangle is the area of initial bone loss in the femoral
neck. It can be identified on standard radiographs once bone loss
has become significant.
The system automatically locates Ward’s triangle, and places a
small box at the area of minimum density in the femoral neck.
The system can search for this region (approximately 2.7 by 3.5
cm) by centering on the lower edge of the femoral neck box to
locate the area of minimum density.
For some patients (especially young patients), it is difficult to
locate an area of minimum density. When this occurs, the Ward’s
triangle is automatically centered at the intersection of the
femoral midline and the initial position of the bottom edge of the
femoral neck box.

KP0886-0832

KP0886-0833

To adjust Ward’s triangle:


Step Action
Note: Complete the Global ROI 1 Click the Other Regions button in the Neck Toolbox, if
and Bone Map procedures prior necessary.
to performing this procedure.
The Neck Toolbox displays with additional buttons for
adjusting the midline, Ward’s triangle, and the trochanter.

Performing and Analyzing the Hip Scan 9-35


QDR Reference Manual

Step Action
2 Click the Ward’s Triangle button.
The Neck Toolbox displays additional buttons specific to
the Ward’s Triangle tool.
Note: Dashed yellow lines are 3 The Whole Mode button is the default and should be
active and can be edited. Solid active.
red lines are inactive.
The Ward’s Triangle displays as active (dashed yellow
lines).
4 Use the following buttons to select a method for
modifying the midline:
• Click the Search radio button and then the Whole
Mode button to have the system search for the
lowest amounts of bone density and automatically
position Ward’s triangle.
• Click the Search radio button and then the Line
Mode button to manually adjust Ward’s Triangle.
• Click the Auto Position button to have the system
automatically locate the position of Ward’s
triangle.
5 If manually adjusting Ward’s Triangle, select one of the
sides of the Ward’s Triangle Box.
The line displays as active (dashed yellow line).
6 Click and drag the line to the desired position. Repeat as
necessary for the other sides of the box.

9-36 Performing and Analyzing the Hip Scan


QDR Reference Manual

About the Trochanter Region


The system automatically identifies the greater trochanter region.
This is a triangular region whose boundaries are defined as the
bottom edge of the Neck Box and a solid line connecting the
intersection of the bottom edge of the Neck Box and femoral
midline to the point where the edge of the femur changes
curvature below the trochanter. You can only adjust the lower
end of this line.

KP0886-0834

KP0886-0835

To adjust the trochanter region:


Step Action
Note: Complete the Global ROI 1 Click the Trochanter button.
and Bone Map procedures prior A movable yellow plus sign appears at the lower end. A
to performing this procedure.
fixed red plus sign appears at the origin.
2 Click the Trochanter button.
A movable yellow plus sign appears at the lower end. A
fixed red plus sign appears at the origin.
3 Click on the yellow plus sign and drag to the correct
position.

Performing and Analyzing the Hip Scan 9-37


QDR Reference Manual

Hip Structure Analysis®


The Hip Structure Analysis (HSA) option measures the
distribution of bone mineral mass at specific cross sections of the
hip and allows the physician to estimate structural properties of
the hip, such as CSA, CSMI, Z and Buckling Ratio.
HSA locates the following three analysis regions in their correct
positions on image:
1. Narrow Neck: located across the femoral neck at its narrow-
est point.
2. Intertrochanter: located across the bend of the femur along
the bisector of the neck and shaft axes.
3. Femur Shaft: across the femur shaft at a distance of 1.5 times
the narrowest diameter of the neck, distal to the intersection
of the neck and shaft axes.
The regions are automatically located by placing boxes around
them but each can be adjusted if necessary.

Figure 9-13
Hip Structure Analysis button
Hip Structure Analysis
Results

Note: Before performing Hip


Structure Analysis read “Tips for
Adjusting the HSA Region
Boxes” on page 9-40.

Hip Structure Analysis results

9-38 Performing and Analyzing the Hip Scan


QDR Reference Manual

HSA Toolbox
The HSA Toolbox includes tools to manually adjust analysis
regions.

• Narrow Neck, Intertrochanter, Femur Shaft


Selects the respective region box.
• Auto Position
Returns position and size of all region boxes to original
setting.
• Whole Mode
When selected, the entire selected region box initially
displays as a dashed yellow line. This allows the entire
box to be moved as a whole unit.
• Line Mode
When selected, one line on the selected region initially
displays as a dashed yellow line. Any of the region lines
can be selected and then moved. Top or bottom
(horizontal) lines can be moved up or down. Side
(vertical) lines can be moved left or right.
• Select a line
Appears when Line Mode is selected. Click the
respective button to select the next clockwise or counter-
clockwise line of the selected region box.
• Recalculate
Recalculates results of NN, IT, FS (only active for Auto
Position).

Manually Adjusting the HSA Region Boxes


The automatically located Narrow Neck, Intertrochanter, or
Femur Shaft region boxes can be moved, resized or reshaped if
adjustment if necessary.
To undo all manual changes returning the position and size of all
region boxes to original setting click Auto Position.
To recalculate the structural properties of the hip after manually
adjusting any region click Recalculate.
To position a region box in Whole Mode:
Step Action
1 Click the respective region (Narrow Neck,
Intertrochanter, or Femur Shaft) button.

Performing and Analyzing the Hip Scan 9-39


QDR Reference Manual

Step Action
2 The Whole Mode button is the default and should be
active (click if not already selected).
The respective region box displays as active lines (dashed
yellow).
3 Click and drag the selected region box to the desired
position.
To resize or reshape a region box in Line Mode:
Step Action
1 Click the respective region (Narrow Neck,
Intertrochanter, or Femur Shaft) button.
Note: Dashed yellow lines are 2 Click the Line Mode button.
active and can be edited. Red One of the selected region box lines displays as active
lines are inactive.
(dashed yellow).
4 Click and drag on the region box line to move.

Tips for Adjusting the HSA Region Boxes


Narrow Neck (NN) - Place the ROI in the middle of the femoral
neck with approximately 2mm of tissue on the short sides of the
NN ROI. The NN ROI usually is placed higher up the midline
then the neck box BMD measurement. In general, the NN should
not be widened.
Intertrochanter (IT) - Verify that the center of the IT ROI box
intersects the HAL (dotted yellow midline) and the solid red
vertical shaft line. If the medical (inside corner) of the IT ROI
box overlaps the lesser trochanter, press the Line Mode button, to
activate the short inside edge of the IT ROI, and move it up until
the ROI is above the less trochanter. This ROI moves as a
parallelogram (opposite sides stay parallel to each other). Verify
that there is approximately 2 mm of tissue on the short ends of the
IT ROI. If necessary, elongate to include the correct amount of
tissue. The IT ROI cannot extend outside the Global ROI and, in
general, should not be widened.
Femoral Shaft (FS) - Verify that the FS ROI is centered (left to
right) to the red vertical shaft line with approximately 2 mm of
tissue on both sides of the short sides of the ROI. This box can be
slightly skewed so that it stays perpendicular to the red vertical
shaft line. The red femoral shaft vertical line is automatically
placed and cannot be adjusted.

9-40 Performing and Analyzing the Hip Scan


QDR Reference Manual

Viewing Results
Results of the analysis display in the lower right portion of the
Analysis window (see Figure 9-8). Click Recalculate to
recalculate the structural properties of the hip after manually
adjusting any region.

The Hip Structure Analysis results area of the Analysis window


shows:
• Regions
Indicates individual regions of the hip included in the
analysis and points to the results.
• CSA
Area of bone material in a given cross section of the bone
expressed in cm².
• CSMI
Cross Sectional Moment of Inertia Bone mineral content
expressed in cm4.
• Z (Section Modulus)
Indicator of bending strength for maximum bending stress
in the image 0plane expressed in cm3.
• CORT
Estimate of inside diameter of cortex expressed in cm.
• BR (Buckling Ratio)
Relative thickness of cortex as an estimate of cortical
stability in buckling.

Performing and Analyzing the Hip Scan 9-41


QDR Reference Manual

9-42 Performing and Analyzing the Hip Scan


Chapter 10
Performing and Analyzing the
Forearm Scan
Note: Forearm is an option on This chapter provides procedures to perform a Forearm scan and
Explorer. analysis. The anatomical sites that are measured for the forearm
scan are: the distal portion of the radius and ulna, the wrist bones,
and the first row of carpal bones.

Content
Subject Page
The Forearm Examination 10-2
Prepare the Patient 10-2
Creating/Retrieving a Patient Biography 10-2
Measuring the Patient's Forearm 10-2
Selecting the Scan Type and Mode 10-4
Positioning the Patient and C-arm 10-6
Performing the Examination 10-7
Exiting the Examination 10-11
Performing the Analysis 10-12
Exiting the Analysis 10-32
Generating and Printing Reports 10-32
Evaluating the Image 10-32

10-1
QDR Reference Manual

The Forearm Examination


The Forearm examination includes all activities that begin when
the patient enters the exam room, and ends when an acceptable
image is obtained. The activities involved in performing the exam
include:
• Preparing the Patient
• Creating/Retrieving a Patient Biography
• Measuring the Patient’s Forearm
• Selecting the Scan Type and Mode
• Positioning the Patient and the C-arm
• Performing the Examination
• Exiting the Examination
• Performing the Analysis
• Exiting the Analysis
• Generating and Printing Reports

Prepare the Patient


Note: Remember to ask the Prepare the patient by asking the pre-examinations questions and
patient if they are, or might be, ensuring that they are not wearing anything that will interfere
pregnant. Postpone the exam
with the scan. Refer to Preparing the Patient on page 7-2 of
until pregnancy is ruled out.
Chapter 7 for detailed instructions.

Creating/Retrieving a Patient Biography


If this is a new patient, create a new patient biography (refer to
Creating/Editing a Patient Record on page 6-9 of Chapter 6 for
detailed instructions). If this patient already has a biography,
retrieve it now (refer to Retrieving a Patient Record on page 6-7
of Chapter 6).

Measuring the Patient's Forearm


Measure and record the length of the patient's forearm in
centimeters. During analysis, the system uses this measurement to
determine the length of the global ROI.

10-2 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Figure 10-1
Measuring the Left Forearm

Step Action
1 Place a chair that has no wheels, arms or cushions, in
front of the table.
For Left Forearm Scan patient faces the table's right
(head end)
For Right Forearm Scan patient faces the table's left
(foot end)
2 Place elbow on the table surface, holding the forearm
vertical (see Figure 10-1).
3 Place a ruler on the table surface and measure up to the
ulna styloid [A] in centimeters. Record the measurement.

Performing and Analyzing the Forearm Scan 10-3


QDR Reference Manual

Selecting the Scan Type and Mode


After a patient biography has been created, or retrieved, the Select
Scan Type window displays.

KP0886-1001

The scan type for the Forearm scan is either Left Forearm or
Right Forearm and is selected in the Select Scan Type window.

Choosing the Scan Mode


Alternate scan mode:
Note: The Use Default Scan There are no alternate scan modes for the Forearm examination.
Mode check box cannot be
unchecked.

To choose the only scan mode:


Step Action
1 Click on Left Forearm or Right Forearm in the scan
type list.
The scan type is highlighted.
2 Click the Next >> button.
The Scan Parameters window displays.

Confirming Scan Parameters


The only scan parameter that can be modified is scan length. The
scan length for a forearm is approximately 6 inches (15 cm). The
system may change the number entered to match the pre-
programmed step size of the scanning mechanism.

10-4 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Figure 10-2
Left Forearm Scan Parameters

Figure 10-3
Right Forearm Scan Parameters

To verify the scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, double click in the Scan Length
field and type the new length.
3 Stop here and position the patient and the C-arm.

KP0886-stop

Performing and Analyzing the Forearm Scan 10-5


QDR Reference Manual

Positioning the Patient and C-arm


The goal for positioning the patient is to ensure that the forearm
bones are as straight as possible for the scan.
For a Left Forearm scan, the scan begins from the mid-forearm to
the first row of carpal bones. For a Right Forearm scan, the scan
starts at the first row of carpal bones and scans up the forearm.

Figure 10-4
Patient Positioned for
Scanning the Forearm

Note: For a left forearm scan,


fingers point towards the right
side (head end) of the table. For
a right forearm scan, the arm Laser Position - Left Forearm
fingers point towards the left
side (foot end) of the table.

Laser Position - Right Forearm

Placing the Forearm in the Scan Field


Positioning the forearm is easiest when the table is centered.
To position the patient’s forearm:
Step Action
1 Press the Center Table button to center the table.
2 Press the Laser button on the Control Panel to activate
the laser cross-hair indicator.
3 Instruct the patient to lean into the table so that their rib
cage is resting against the table edge.
4 Align the long bones of the forearm with the long axis of
the laser.
5 Have the patient make a loose fist to flatten the wrist.
6 Bend the patient’s elbow to approximately 105°.

10-6 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Positioning the C-arm


To position the C-arm and fine tune the starting point:
Step Action
Helpful Hint for aligning the 1 Press the Laser button on the Control Panel a second
laser with the left forearm: time to activate the laser cross-hair indicator.
• Position the laser in the
center of the patient’s
2 Use the Arm Controls on the Control Panel to fine tune
forearm. the starting point on the forearm.
• Press the C-arm 3 Verify the starting points:
buttons on the Control • Left Forearm Scan: Starts at the mid-forearm.
Panel and move the Verify that the first row of carpal bones are within
beam down to the wrist
6 inches (15 cm) of the starting point.
to verify the centering
position. If the beam is • Right Forearm Scan: Starts at the first row of
centered on the wrist, it carpal bones.
indicates the laser is
properly aligned.
• If the cross hairs are 4 Instruct the patient to remain still and breathe normally.
not centered on the
wrist, adjust the laser
using the Table buttons
on the Control Panel
and repeat the
alignment verification.
• Press the C-arm
buttons to move the
laser back to the
starting point.

Performing the Examination


An acceptable forearm scan (see Figure 10-5) includes the
following:
• The forearm appears straight and centered in the scan
image window.
• The image includes at least the first row of carpal bones.
• The ulna side contains at least enough air to equal the
shaft width of the ulna.

Performing and Analyzing the Forearm Scan 10-7


QDR Reference Manual

Figure 10-5
Forearm Scan

KP0886-1006

Starting the Forearm Scan


Once the patient and C-arm are positioned correctly, return to the
computer to start the scan. At this time, the Scan Parameters
window should be displayed.
To start the forearm scan:
Step Action
Note: If the Control Panel X-ray 1 Click the Start Scan button
lamp fails to extinguish within The Scan window displays with the image appearing on
10 seconds after the end of the
the left side. Flashing X-rays On indicator at the top of
scan, press the red Emergency
Stop button on the Control Panel the window continues until the scan stops.
immediately. Then call Hologic
Service before resuming
operation.

KP0886-1007

10-8 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Step Action
2 Inspect the image as follows:
• The radius and ulna appear straight and centered.
• The image shows at least the first row of carpal
bones near the bottom for a right forearm, or near
the top for a left forearm.
• The ulna side of the scan contains at least enough
air to equal the shaft width of the ulna.
3 If the image is not acceptable, click the Reposition Scan
button.
4 If the image is acceptable, allow the scanning to
complete.
When the scan completes, the Exit Exam window
displays.
While the scan is being acquired the operator has four options:

Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the Exit Exam
window displays.
Stop Scan button Terminates the scan at the point at
which the button is pressed and the
KP0886-0808 Exit Exam window displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP0886-0809 proceeds to the main window. To
start another scan the C-arm must
be repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP0886-0810 either the window image or the
patient. Refer to the following
instructions.

Repositioning the Scan


Two basic activities of repositioning include:
• Repositioning the image
Use the mouse and scroll bars to center the image in the
window.

Performing and Analyzing the Forearm Scan 10-9


QDR Reference Manual

• Repositioning the patient


Adjust the patient’s forearm so that it presses against the
back of the Positioner.
To reposition the image:
Step Action
Note: In order to reposition the 1 Click the Reposition Scan button
scan, the Reposition Scan button The image acquired so far displays with scroll bars on the
must be clicked before the scan
right and bottom.
is completed.
Note: You can also use the scroll 2 Position the cursor over the forearm image.
bars to reposition the image. The arrow cursor changes to a hand.

Carpal
Bones

Horizontal
Positioning
Line
Vertical
Positioning
Lines
Scroll
Cursor Bars
Hand
KP0886-1008

3 To position the forearm:


• Move the first row of carpal bones within the
horizontal blue positioning line and the outer limit
of the scan field.
• The radius and the ulna should be parallel
between the two blue vertical positioning lines.
4 When the forearm is positioned correctly, click the
Restart Scan button.
The Scan Parameters window displays.
5 Click the Start Scan button to start a new scan at the new
position.
The Scan window displays with a flashing X-rays On
message. The image displays.

10-10 Performing and Analyzing the Forearm Scan


QDR Reference Manual

While repositioning the scan the following buttons are displayed


in the window:
Button Description
Restart Scan button Proceeds to the Scan Parameters
window to commence X-ray scan
KP0886-0909
from the new position.
Accept Scan button Saves the scan in progress without
KP0886-0910
repositioning the scanner.

Cancel Scan button Cancels the exam and returns to


KP0886-0911
the main window.

To reposition the patient:


Step Action
Note: The patient’s forearm can 1 To reposition the patient if forearm is not straight on the
be moved up or down to include table, move the patient’s forearm so that it presses against
or exclude more of the carpal
the back of the Positioner.
bones.
2 When the forearm is positioned correctly, click the
Restart Scan button.
The Scan Parameters window displays.
Note: If the Control Panel X-ray 3 Click the Start Scan button to start a new scan at the new
lamp fails to extinguish within position.
10 seconds after the end of the
The Scan window displays with a flashing X-rays On
scan, press the red Emergency
Stop button on the Control Panel message. The image displays.
immediately. Then call Hologic 4 Evaluate the image as it displays. If the image is
Service before resuming
satisfactory, allow the scan to complete.
operation.
When the scan completes, the Exit Exam window
displays.

Exiting the Examination


After acquiring an acceptable image, the Exit Exam window
displays. To perform the analysis of the scan just completed, click
the Analyze Scan button in the Exit Exam window. Refer to
Exiting the Examination on page 7-23 in Chapter 7 for detailed
instructions.

Performing and Analyzing the Forearm Scan 10-11


QDR Reference Manual

Performing the Analysis


Note: If the analysis of the scan The Forearm analysis includes all activities that begin with a
is to be postponed to a later satisfactory image and ends with the calculation and summary of
time, the scan can be searched
the results of the scan. After completing a scan, the system
for in the Scan. Refer to Chapter
6 for information about the Scan automatically saves the scan, in the Scan Drawer, for analysis.
Drawer. The three parts of analysis include:
• Choosing the analysis method.
Note: If the patient has a respective (left or right)
Forearm baseline scan, the software proceeds directly to
the compare analysis function instead of choosing the
analysis method. Compare analysis is discussed in detail
in Chapter 14. Forearm scan analysis is discussed
beginning on page 10-14.
• Performing the analysis.
• Generating a report.

Choosing the Method of Analysis


The first analysis activity involves selecting the analyze setup.
The analyze setup determines the method that the system uses to
perform the analysis. It can perform the analysis either as a single
new scan, or as a comparison to one that has been previously
performed.
If this scan is the patient’s first visit, select the default method
(Left or Right Forearm analysis) and the system presents a single
scanned image for analysis. If this scan is a follow-up visit, select
the Compare to Previously Analyzed Scan radio button. The
system then presents two images. The left image is the
unanalyzed scan, and the right image is a previous scan with the
region of interest (ROI) defined.
This allows the system to perform the new analysis on the same
ROI as the previous scan and maximize accurate reporting of
changes. Follow-up visits are discussed in detail in Chapter 13.

10-12 Performing and Analyzing the Forearm Scan


QDR Reference Manual

To select analyze setup for a first visit:


Step Action
1 From the Exit Exam window, click the Analyze Scan
button.
The Analyze Setup window displays.

KP0886-1009

Note: The Analyze Setup window 2 Click the Choose Analysis Method radio button.
displays with the Choose
Analysis Method radio button 3 Verify the analysis method.
selected as the default.
4 Click Next >> button.
Comparison is covered in detail
in Chapter 13. The Analysis window displays.

Methods of Analysis
To select a method of analysis (other than default):
Step Action
1 Click the drop down list arrow on the Analyze Setup
window.
List of methods displays.
From the Exit Exam window, click the Analyze Scan
button.
The Analyze Setup window displays.
Note: Subregion Forearm
analysis is available as an
option.

KP0886-1010

2 Click on the desired method.


List disappears, and selected method displays in field.
3 Click the Next >> button to continue.
The Analysis window displays.

Performing and Analyzing the Forearm Scan 10-13


QDR Reference Manual

Performing the Forearm Analysis


Once the setup method has been selected, continue to the analysis
window to adjust the quality of the image and properly define the
area of analysis called the region of interest (ROI).
A specific set of tools called the Analysis Step Buttons, located
on the left side of the window (see Figure 10-6), allow the
operator to proceed step-by-step through each task. These tools
can maximize image quality and accuracy, preventing the need
for re-scanning.
Figure 10-6 Step buttons
Forearm
Analysis Window

KP0886-1011
Toolbox
Brightness/Contrast button

The steps to analyze the forearm scan include:


1. Entering the patient’s forearm length, measured earlier, in
centimeters (Length Toolbox).
2. Adjusting the image display brightness and contrast (Bright-
ness/Contrast button).
3. Defining the Global Region of Interest (Global ROI button).
4. Viewing/editing the bone map (Bone Map button).
5. Adjusting the ultra distal region and divider (MID/UD but-
ton).
6. Viewing analysis results (Results button).
Exiting the analysis window (Close button).

10-14 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Entering the Forearm Length


Enter the forearm length, measured during the patient
positioning, to proceed with the analysis.
KP0886-1012

KP0886-1013

To enter the forearm length:


Step Action
Note: The forearm length must 1 Click the Length button.
be between 12.0 and 42.0
centimeters. If a value is entered 2 Enter the forearm length in centimeters into Forearm
outside of this range, the system length (cm).
displays a message prompting a
correct value be entered. The
entered forearm length value is
displayed throughout the
analysis.

Adjusting the Image Display


The contrast and brightness of the image can be adjusted to
obtain the best definition of important anatomical features.
Adjusting the contrast and brightness has no effect on
KP0886-0817 calculations; it is used to optimize the display and for visual
adjustments only. Refer to Contrast and Brightness on page 7-
25, in Chapter 7, for detailed instructions.

Performing and Analyzing the Forearm Scan 10-15


QDR Reference Manual

Defining the Region of Interest (ROI)


The global ROI refers to the defined boundaries of the image that
is being analyzed. The ROI appears on the image as a box: the
dashed line is active and can be adjusted; the solid lines are
KP0886-0818
inactive and cannot be adjusted.
Figure 10-7
Left Forearm Scan Right Forearm Scan
ROI Image in the Analysis
Window

Note: The images show the Left


and Right Forearm scans with a
yellow dashed box around the
Global ROI. Note that the Right
Forearm scan is the reverse of
the Left Forearm scan because
of the positioning of the forearm
during the examination.

Hints for Identifying Soft


Tissue and Air: Soft tissue
displays as the lightly shaded
area surrounding the radius and
ulna. Air is the darker area
outside the forearm soft tissue. KP1068_002-1014

The system automatically calculates the length of the ROI box


based on the forearm length previously entered. It calculates the
box length by dividing the forearm length by 3 and adding 10mm
to allow for the 1/3 Distal Region.
The Forearm ROI should include:
• a small amount of air on the ulna side
• radius and ulna bones

10-16 Performing and Analyzing the Forearm Scan


QDR Reference Manual

The Global ROI Toolbox


The Global ROI Toolbox includes tools to help define the region
of interest.
• Whole Mode
When selected, the entire ROI box initially displays as a
dashed yellow outline with three horizontal solid red
lines. Use this mode to position the top or bottom lines of
the ROI on the Styloid Process. The ROI moves as a
whole unit.
• Line Mode
When selected, the left border line on the ROI box
initially displays as a dashed yellow line. Use this mode
to precisely position the left and right sides of the ROI
box on landmarks, as indicated in the Step/Action table.
• 1/3 Distal
When selected, the top or bottom border line on the ROI
box initially displays as a dashed yellow line. Use this to
adjust the size of the 1/3 Distal region if necessary (it is
KP0886-1015 rarely necessary to adjust this region).

To define the ROI using Whole and Line Mode:


Step Action
Note: This procedure is for a 1 After the length of the forearm is entered the program
Left Forearm scan, the will automatically select the Global ROI button and the
procedure for the Right Forearm
Whole Mode button (as defaults).
scan is the same except the
image of the scan is the reverse The system displays a yellow dashed box around the
(see Figure 10-7 on page 10-16), forearm image (see Figure 10-7 on page 10-16).

Performing and Analyzing the Forearm Scan 10-17


QDR Reference Manual

Step Action

KP1068_002-1033

Note: For a Right Forearm scan 2 Click in the yellow dashed box and hold the button to
the yellow dashed line would move the box. Position the top line on the ulnar styloid
position the bottom line on the
process.
ulnar styloid process.

3 Click the Line Mode button.


The ROI box outline displays with two dashed lines on
KP1068_002-1035
the left side, a line with large yellow dashes and a line
with small white dashes.

10-18 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Step Action
Yellow Large Dashed Line
Blue Small Dashed Line

KP1068_002-1038

The distance between the two dashed lines has been


calculated for the proper analysis and cannot be changed.
The Yellow dashed line is active, it can be moved using
the mouse.
4 Click on the yellow dashed line and move it and the blue
dashed line until the blue dashed line is just touching
bone.
5 Click either button under Select a line.
The yellow and blue dashed lines now appear on the right
side of the scan.

KP1068_002-1039

Performing and Analyzing the Forearm Scan 10-19


QDR Reference Manual

Step Action

Yellow Dashed Line


Blue Dashed Line

6 Click on the yellow dashed line and move it and the blue
dashed line until the blue dashed line just touches bone.

KP1068_002-1041

10-20 Performing and Analyzing the Forearm Scan


QDR Reference Manual

About the 1/3 Distal Region


The system automatically sets this region based upon the
measured forearm length.
Use this adjustment for research applications only.

KP0886-1016

To adjust the 1/3 distal region:


Step Action
1 Click the 1/3 Distal button.
The Medial boundary of the global ROI displays as active
(dashed yellow line).

Performing and Analyzing the Forearm Scan 10-21


QDR Reference Manual

Step Action

KP1068_002-1041

2 Click and drag the dashed yellow line up or down to


increase or decrease the 1/3 region.
As the line moves, the parallel boundary of the 1/3 distal
region also moves.

Viewing/Editing the Bone Map


The system creates a yellow bone map of the forearm over the
ROI defined in the previous steps. It automatically fills holes in
the yellow bone map.
KP0886-0821

To view the bone map:


Step Action
1 Click the Bone Map button.
The Bone map displays in yellow.

10-22 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Step Action

KP1068_002-1017

2 Verify whether the bone map is acceptable or needs to be


edited.
Evaluate the bone map:
There must be air (dark area) beyond the soft tissue on the ulna
side of the ROI. If there is not enough air, the system displays a
poor bone map image. If this occurs, click the ROI button and
increase the amount of air beyond the soft tissue on the ulna side.
To edit the bone map:
To add or delete bone from the image, use the Bone Map Toolbox
portion of the Analysis window. The operator can also adjust the
cursor size, fill holes, undo changes, or reset the image to the
original bone map if unintentionally changed.
If there are no bone areas to be filled in (the vast majority of
forearm scans), proceed to Adjusting the Ultra Distal Region.

Performing and Analyzing the Forearm Scan 10-23


QDR Reference Manual

KP0886-1018

To add bone to the bone map:


Step Action
1 Click the Add Bone radio button.
The cursor displays as a small green rectangular paint
brush when placed inside the global ROI on the image.
2 Select the cursor size using the Cursor Size tool.
Hologic recommends the default paint brush (size 2) be
used.
3 Move the cursor to the starting point.
Note: Be careful to follow the 4 Click and drag the cursor to connect outside edges to
bone edge to bridge all gaps. If complete the bone map.
there are any gaps, the Fill
Holes function will not work.
5 To automatically fill the remaining holes, click the Fill
Holes button.

10-24 Performing and Analyzing the Forearm Scan


QDR Reference Manual

To delete bone from the bone map:


Step Action
1 Click the Delete Bone radio button.
The cursor displays as a small green rectangular paint
brush when placed inside the global ROI on the image.
2 Select the cursor size using the Cursor Size tool.
Hologic recommends the default paint brush (size 2) be
used.
3 Move the cursor to the starting point.
Note: Be careful to delete a 4 Click and drag the cursor to erase a line through the
complete line through the bone yellow bone map.
map until the image separates.
This creates a small island—an isolated section of bone.
Note: If you leave any 5 To automatically remove the small island, click on the
connection, only the line you Sink Islands button.
deleted is removed.
Small island is removed.

Adjusting the MID/UD Regions


The next step in the analysis is adjusting the MID/UD regions.
The size of the regions and their position can be adjusted. The
KP0886-1019 UD (ultra-distal) default position is 10mm from the end of the
ROI. This position may have to be adjusted to exclude the end
plate of the radius (dense cortical bone). Cortical bone appears
much denser (lighter) on the monitor.

Performing and Analyzing the Forearm Scan 10-25


QDR Reference Manual

KP1068_002-1022

Although not usually necessary, the size of the UD (ultra-distal)


region can be increased or decreased to exclude as much cortical
bone as possible.
The MID/UD Toolbox
To check the MID/UD region
The MID/UD Toolbox includes tools that help define the MID/
UD regions:
• Whole Mode
Use this mode to move the UD (ultra-distal region) as a
whole.
• Line Mode
Use this mode to increase or decrease the size of the
MID/UD region.
To check the MID/UD region
Step Action
1 Click the MID/UD button.
The ROI box displays in the image area with the UD
region displayed as active (dashed yellow lines).
2 Verify that the UD region excludes the end plate of the
radius (dense cortical bone).
KP0886-1021

10-26 Performing and Analyzing the Forearm Scan


QDR Reference Manual

To adjust the MID/UD region:


Step Action
Note: The UD region is the area 1 The Ultra Distal button is the default and should be
between the two horizontal active (click if not already selected).
dashed yellow lines.
The Ultra Distal region displays as active (yellow dashed
lines).
2 The Whole Mode button is the default and should be
active (click if not already selected).
Note: Cortical bone appears 3 Click and drag the UD region to exclude the end plate of
much lighter (denser). the radius (dense cortical bone).

About the Ulna/Radius Divider


The system automatically computes the division and no
adjustment is required under normal conditions. If no changes
are necessary, proceed to Viewing Results on page 10-28.

KP0886-1021

KP0886-1022

To automatically reposition the ulna/radius divider:


Step Action
1 Click the Divider button in the MID/UD Toolbox.
The Auto Position button displays.

Performing and Analyzing the Forearm Scan 10-27


QDR Reference Manual

Step Action
2 Click the Auto Position button.
The system recalculates the divider line position and
displays it on the image.
To manually reposition the ulna/radius divider:
Step Action
1 Click the Divider button in the MID/UD Toolbox.
2 The Line Mode button is the default and should be active
(click if not already selected).
The divider line displays as active (dashed yellow line).
Note: This line cannot be moved 3 Click and drag the line left or right.
outside of the ROI boundary.
4 Click the Point Mode button.
A plus sign appears at each end of the divider line.
5 Click a point on the divider line and drag as necessary to
separate the ulna and radius.
The selected point displays as active (yellow).

Viewing Results
Results of the analysis display in the lower right portion of the
Analysis window (see Figure 10-8).
KP0886-0825

The forearm results window shows:


• Pertinent Statistical Information
The headings of the results section indicate the statistical
figures that are used in maintaining accuracy and
diagnostic precision for this machine, and are set at the
factory.
• Radius/Ulna
Indicates regions of the radius and/or ulna included in the
analysis and points to the results.
• Area
Area expressed in cm² (centimeters squared).
• BMC
Bone mineral content expressed in grams.
• BMD
Bone mineral density expressed in g/cm² (grams per
centimeter squared).

10-28 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Figure 10-8 Results button


Analysis Window

Analysis results KP1068_002-1023

To generate results:
Step Action
1 Click the Results button.
The Results Toolbox displays with three buttons: Radius
+ Ulna Results, Radius Results, and Ulna Results.

KP0886-1024

Performing and Analyzing the Forearm Scan 10-29


QDR Reference Manual

Step Action
2 The Radius + Ulna Results button is the default and
should be active (click if not already selected).
The system displays the combined radius and ulna
results.

KP1068_002-1025

3 To display only radius results, click the Radius Results


button.
The system displays the radius results only.

KP1068_002-1026

10-30 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Step Action
4 To display only ulna results, click the Ulna Results
button.
The system displays the ulna results only.

KP1068_002-1027

To reanalyze the current scan with different settings:


Step Action
Note: Once an Analysis Step 1 Click the appropriate Analysis Step Button (see Figure
Button is modified, each 10-5) and modify.
subsequent button (analysis
step) must be performed or 2 Repeat for each subsequent Analysis Step Button in the
results will not be accurate. Global ROI Toolbox.

To reanalyze the current scan from the main window:


Step Action
1 Click the Analyze Scan button.
2 Click on the Analyzed Scans tab.
3 Select the desired scan.
4 Click the Next >> button.
5 Continue with analysis as described earlier in this
chapter.

Performing and Analyzing the Forearm Scan 10-31


QDR Reference Manual

Exiting the Analysis


After performing the analysis, the Exit Analysis window displays.
To select a report type to print, click on the Report button in the
Exit Analysis window. Refer to Exiting the Analysis on page 7-27
of Chapter 7 for detailed instructions.
To exit the Analysis window after obtaining results:
Step Action
1 Click the Close button in the lower left corner of the
Analysis window.

To exit the Analysis window before obtaining results:


Step Action
Note: The scan may be analyzed 1 Click the Cancel button in the lower left corner of the
at a later time from the main Analysis window.
window (see above).

Generating and Printing Reports


A variety of reports can be generated from the results of the
analysis, customized to fit the needs of the examination. Generate
and print the desired report. Refer to Generating and Printing
Reports on page 7-32 of Chapter 7 for more information.

Evaluating the Image


This section provides examples that illustrate unacceptable scan
images.
Figure 10-9
Unacceptable Forearm
Scan: Crooked and Positioner
Seen on Image

KP0886-1028

10-32 Performing and Analyzing the Forearm Scan


QDR Reference Manual

Figure 10-10
Unacceptable Forearm Scan:
Clipped at the Top and Off
Center

KP0886-1029

Figure 10-11
Unacceptable Forearm
Scan: Started Too High

KP0886-1030

Performing and Analyzing the Forearm Scan 10-33


QDR Reference Manual

10-34 Performing and Analyzing the Forearm Scan


Chapter 11
Performing and Analyzing the
Whole Body Scan
This chapter provides procedures to perform a Whole Body scan and
analysis.
The following models can perform scans of the whole body:
• Discovery-A
• Discovery-W
• Discovery-Wi
• Explorer (with Whole Body Option)

Content
Subject Page
The Whole Body Examination 11-2
Auto Whole Body 11-2
Preparing the Patient 11-2
Creating/Retrieving a Patient Biography 11-3
Selecting the Scan Type and Mode 11-3
Positioning the Patient and C-Arm 11-4
Performing the Examination 11-6
Exiting the Examination 11-7
Performing the Analysis 11-7
Exiting the Analysis 11-17
Generating and Printing Reports 11-18
Evaluating the Image 11-18

11-1
QDR Reference Manual

The Whole Body Examination


The Whole Body examination includes all activities that begin
when the patient enters the exam room, and ends when an
acceptable image is obtained. The activities involved in
performing the exam include:
• Preparing the Patient
• Creating/Retrieving a Patient Biography
• Selecting the Scan Type and Mode
• Positioning the Patient and the C-arm
• Performing the Examination
• Exiting the Examination
• Performing the Analysis
• Generating and Printing Reports

Auto Whole Body


Note: Refer to “Analyze Tab” on The QDR system Auto Whole Body feature automatically
page 5 of Chapter 18 on how to adjusts the analysis for patients weighing from 17.6 lbs (8 Kg) to
disable the Auto Whole Body
88 lbs (40 Kg). Above 88 lbs, the Auto Whole Body analysis
feature.
provides results identical to previous whole body analysis
versions. The software has not been evaluated for patients
weighing less than 17.6 lbs and use on patients below this weight
is not recommended.
Because Auto Whole Body automatically provides an improved
analysis for patients less than 88 lbs (40 Kg), the Pediatric Whole
Body analysis method is no longer recommended. Patients
previously analyzed with Pediatric Whole Body should be re-
analyzed using Auto Whole Body. Note that while it is still
possible to use Pediatric Whole Body, reference database
comparisons and T and Z scores will not be valid.
Auto Whole Body is only available on systems that have the
body composition analysis option. If the QDR system acquiring
the scan does not have body composition option, then Whole
Body measurements on subjects less than 40 kg are not
recommended.

Preparing the Patient


Note: Remember to ask the Prepare the patient by asking the pre-examinations questions and
patient if they are, or might be, ensuring that they are not wearing anything that will interfere
pregnant. Postpone the exam
with the scan. Refer to Preparing the Patient on page 7-2 of
until pregnancy is ruled out.
Chapter 7 for detailed instructions.

11-2 Performing and Analyzing the Whole Body Scan


QDR Reference Manual

Creating/Retrieving a Patient Biography


If this is a new patient, create a new patient biography (refer to
Creating/Editing a Patient Record on page 6-9 of Chapter 6 for
detailed instructions). If this patient already has a biography,
retrieve it now (refer to Retrieving a Patient Record on page 6-7
of Chapter 6).

Selecting the Scan Type and Mode


After a patient biography has been created, or retrieved, the Select
Scan Type window displays.

KP0886-1101

The scan type for the Whole Body scan is Whole Body or HP
Whole Body.

Choosing the Scan Mode


Alternate scan modes:
Note: The Use Default Scan There are no alternate scan modes for the Whole Body
Mode check box is grayed and examination.
cannot be unchecked.

To choose the scan mode:


Step Action
1 Click on Whole Body or HP Whole Body in the scan
type list.
The scan type is highlighted.
2 Click the Next >> button.
The Scan Parameters window displays.

Performing and Analyzing the Whole Body Scan 11-3


QDR Reference Manual

Confirming Scan Parameters


Note: Typically, it should not be The only scan parameter that can be modified is the scan length
necessary to adjust the scan (see Figure 11-1). The scan length for a Whole Body exam is
length unless performing a
approximately 77 inches (195 cm). The scan width is
Whole Body scan on a child.
approximately 27 inches (69 cm). The system may change the
number entered to match the pre-programmed step size of the
scanning mechanism.
Figure 11-1
Whole Body Scan
Parameters

KP1035A-1102

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
Note: Typically, it should not be 2 To change scan length, place the cursor in the Scan
necessary to adjust the scan Length field and type the new length.
length unless performing a
Whole Body scan on a child.
3 Stop here and position the patient and the C-arm.

KP0886-stop

Positioning the Patient and C-Arm


This section provides information for positioning the patient, the
table, and the C-arm for a Whole Body exam. It is important to
follow each set of directions carefully since they are designed to
help acquire an ideal scan by describing:
• How to maximize patient comfort and safety.

11-4 Performing and Analyzing the Whole Body Scan


QDR Reference Manual

• How to assist the patient onto the table for proper


alignment.
• Suggestions for correct positioning techniques.

Placing the Patient on the Table


Note: Refer to Chapter 7, page Press the Patient On/Off switch on the Control Panel and seat
7-21, for a detailed description the patient on the table between the scan field indicators on the
of placing the patient on the
table pad. Their right shoulder should be adjacent to the C-arm.
table.
Continue with positioning the patient below.

Positioning the Patient


Positioning the patient for a Whole Body scan involves aligning
the patient’s body with the center lines at the right (head) and left
(foot) sides of the table pad (see Figure 11-1).

To position the patient:


Step Action
1 Lie the patient on their back with the head at the right end
of the table. Instruct the patient to look at the ceiling to
maintain head position.
Note: On Explorer the Center 2 Press the Center Table button on the Control Panel.
Table button must be held and The table and C-arm move to the center.
only the table moves.
Note: If the patient is taller than 3 Check the patient’s position on the table pad (see
6’6” (198 cm), extend the Figure 11-1), ensuring that the following conditions are
patient's feet beyond the end of
met:
the table. Instruct large patients
to place their hands vertically at • Body straight on the table pad using the center
their sides with the fifth finger on lines at the head and foot ends of the table pad as
the table pad. Their hands
the gauge.
should be next to the thighs to
ensure the hands and arms are • Body, including the feet, are positioned entirely
within the table limits. within the scan limit border line.
• Arms are at their sides, with the palms down,
separated from the thighs, and within the scan
limit border.
• Feet are pointing up.
4 Rotate the patient’s legs inward 25° until the toes touch,
then place tape around the feet to maintain position.
5 Instruct the patient to remain still and breathe normally.

Performing and Analyzing the Whole Body Scan 11-5


QDR Reference Manual

Performing the Examination


An acceptable whole body scan includes the following:
• Entire body
For patients larger than 6’6”, the feet may be excluded
from the scan.
• Arms
For patients who may be larger than the width of the exam
table, parts of the arms may be excluded from the scan.

Starting the Whole Body Scan


Once the patient is positioned correctly, return to the computer to
start the Whole Body scan. At this time, the Scan Parameters
window should be displayed.
To start the Whole Body scan:
Step Action
Note: If the Control Panel X-ray 1 Click the Start Scan button.
lamp fails to extinguish within The Scan window displays with the image appearing on
10 seconds after the end of the
the left side. Flashing X-rays On indicator at the top of
scan, press the red Emergency
Stop button on the Control Panel the window continues until the scan stops.
immediately. Then call Hologic
Service before resuming
operation.

2 Make sure the patient’s arms are included in the scan on


the first and last pass of the C-arm. The patient should
remain still until the scan is complete.

11-6 Performing and Analyzing the Whole Body Scan


QDR Reference Manual

Step Action
Note: Discovery-A makes 3 3 Allow the scan to complete.
passes while Discovery-W, When the scan completes, the Exit Exam window
Discovery-Wi and Explorer (with
displays.
Whole Body option) make 7
passes to complete the Whole
Body scan.

While the scan is being acquired the operator has two options:

Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the Exit Exam
window displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP0886-0809 proceeds to the main window.

Exiting the Examination


After acquiring an acceptable image, the Exit Exam window
displays. To perform the analysis of the scan just completed, click
the Analyze Scan button in the Exit Exam window. Refer to
Exiting the Examination on page 7-23 in Chapter 7 for detailed
instructions.

Performing the Analysis


Note: If the analysis of the scan The Whole Body analysis includes all activities that begin with a
is to be postponed to a later satisfactory image and ends with the calculation and summary of
time, the scan can be searched
the results of the scan. After completing a scan, the system
for after clicking Analyze Scan.
automatically saves the scan in the Scan Drawer. The three parts
of analysis include:
• Choosing the analysis method
Note: If the patient has a Whole Body baseline scan, the
software proceeds directly to the compare analysis
function instead of choosing the analysis method.
Compare analysis is discussed in detail in Chapter 14.
Whole Body scan analysis is discussed beginning on page
11-9.
• Performing the analysis.
• Generating a report.

Performing and Analyzing the Whole Body Scan 11-7


QDR Reference Manual

Choosing the Method of Analysis


The first analysis activity involves selecting the Method of
Analysis. Under method of analysis for whole body you have 2
selections:
• Whole Body Fan Beam. This method of analysis is used
to analyze scans acquired on an adult patient.
• Pediatric Whole Body. Although no longer
recommended, Pediatric Whole Body is available for
legacy purposes (see Auto Whole Body on page 11-2).

To select analyze setup for a first visit:


Step Action
1 From the Exit Exam window, click the Analyze Scan
button.
The Analyze Setup window displays.

KP0886-1105

Note: The Analyze Setup window 2 Select the method of analysis appropriate for the age of
displays with the Choose the patient (either Whole Body Fan Beam or Pediatric
Analysis Method radio button
Whole Body).
selected as the default.
Comparison is covered in detail
on page 14-8 of Chapter 14.

KP0886-1106

3 Click the Next >> button.


The Analysis window displays.

11-8 Performing and Analyzing the Whole Body Scan


QDR Reference Manual

Choosing the Compare to Previous Analysis


If this scan is a follow-up visit, select the Method of Analysis
appropriate for the age of the patient then select the Compare to
Previously Analyzed Scan radio button. The system then
presents two images. The left image is the unanalyzed scan, and
the right image is a previous scan with the region of interest
(ROI) defined. This allows the system to perform the new
analysis on the same ROI as the previous scan and maximize
accurate reporting of changes. Follow-up visits are discussed in
detail on page 14-7 of Chapter 14.

Performing the Whole Body Analysis


Once the setup method has been selected, continue to the
Analysis window to adjust the quality of the image and properly
define the area of analysis called the region of interest (ROI).
A specific set of tools called the Analysis Step Buttons, located
on the left side of the window (see Figure 11-2), allow the
operator to proceed step-by-step through each task. These tools
can maximize image quality and accuracy, preventing the need
for re-scanning.
Figure 11-2 Analysis Step buttons
Whole Body
Analysis Window

KP0886-1107
Toolbox
Brightness/Contrast button

The steps to analyze the Whole Body scan include:


1. Adjusting the image display brightness and contrast (Bright-
ness/Contrast button).

Performing and Analyzing the Whole Body Scan 11-9


QDR Reference Manual

2. Defining the Regions (Regions button).


3. Viewing analysis results (Results button).
4. Exiting the Analysis window (Close button).

Adjusting the Image Display


The contrast and brightness of the image can be adjusted to
obtain the best definition of important anatomical features.
Adjusting the contrast and brightness has no effect on
KP0886-0817 calculations; it is used to optimize the display and for visual
adjustments only. Refer to Contrast and Brightness on page 7-
25, in Chapter 7, for detailed instructions

Defining the Regions


The Regions are predefined as a template overlay and can be
adjusted to match the patient’s anatomy.
Figure 11-3
Regions Image in
the Analysis Window

KP0886-1108

11-10 Performing and Analyzing the Whole Body Scan


QDR Reference Manual

Regions Toolbox
The Regions Toolbox includes tools to help define the regions of
interest.
• Line Mode
When selected, one line on the regions template initially
displays as a dashed yellow line. Any of the template
lines can be selected and then moved. Top or bottom
(horizontal) lines can be moved up or down. Side
(vertical) lines can be moved left or right.
• Whole Mode
When selected, the entire regions template initially
displays as a dashed yellow line. This allows the entire
template to be moved. This tool is usually used during a
compare scan.
• Point Mode
When selected, a plus sign displays at the end of each line
in the regions template. These end points can be moved in
any direction.
KP0886-1109

Performing and Analyzing the Whole Body Scan 11-11


QDR Reference Manual

Figure 11-4
Positioning Horizontal Lines

KP0886-1110

To position horizontal template lines in the Line Mode:


Step Action
1 Click the Regions button in the top left corner of the
window, if not already selected.
The regions template displays in the image area.
2 The Line Mode button is the default and should be active
(click if not already selected).
The middle top line of the regions template displays as an
active line (dashed yellow).
3 Click and drag the middle top line of the region (labeled
A on Figure 11-4) to just under the patient’s jaw.
Note: Dashed yellow lines are 4 Click and drag the small horizontal line in the spine
active and can be edited. Solid (labeled B on Figure 11-4) to the approximate level of
red lines are inactive.
T12-L1.
5 Click and drag the horizontal line above the pelvis
(labeled C on Figure 11-4) to just above the iliac crest.

11-12 Performing and Analyzing the Whole Body Scan


QDR Reference Manual

Figure 11-5
Positioning Vertical Lines

KP0886-1111

To position vertical template lines in the Line Mode:


Step Action
1 Click and drag the vertical line on either side of the chest
(labeled A on Figure 11-5) to bring each one close to the
chest.
2 Click and drag the vertical line on either side of the spine
(labeled B on Figure 11-5) to bring each one close to the
spine.
3 Click and drag the vertical line on either side of the legs
(labeled C on Figure 11-5) to bring each one close to the
leg.
4 Click and drag the bottom center vertical line (labeled D
on Figure 11-5) to separate the legs and feet.

Performing and Analyzing the Whole Body Scan 11-13


QDR Reference Manual

Figure 11-6
Fine-tune the Vertical Lines

C
D

E
Note: Must scroll down to
see the 2 points for E
(not shown in figure).

KP0886-1112

To position and fine-tune template lines in the Point Mode:


Step Action
1 Click the Point Mode button.
Plus signs display at the end of each line in the regions
template.
2 Click the point on the left shoulder (labeled A on
Figure 11-6) and drag so that it is positioned between the
head of the humerus and scapula at the glenoid fossa.
Repeat this step for the right shoulder (also labeled A on
Figure 11-6).
A total of two points will have been repositioned when
finished with this step (one on the left and one on the
right).
3 Click the left points along the spine (labeled B on
Figure 11-6) and drag them close to the spine, matching
the curvature if possible. Repeat this step for the spine’s
right side (also labeled B on Figure 11-6).
A total of six points will have been repositioned when
finished with this step (three on the left and three on the
right).

11-14 Performing and Analyzing the Whole Body Scan


QDR Reference Manual

Step Action
4 If necessary, click the left point above the iliac crest
(labeled C on Figure 11-6) and drag it out at the sides to
include soft tissue of the chest and thighs. Repeat for the
right point (also labeled C on Figure 11-6).
A total of two points will have been repositioned when
finished with this step (one on the left and one on the
right).
5 Click and drag the lower point of the triangle below the
pelvis (labeled D on Figure 11-6) to bisect both femoral
necks.
A total of one point will have been repositioned when
finished with this step.
6 Use the scroll bar to scroll the image to the bottom of the
scan. Click and drag the points near the feet (labeled E on
Figure 11-6) to include as much of the soft tissue in the
thighs as possible without including the patient’s hand
and fingers.
A total of two points will have been repositioned when
finished with this step (one on the left leg and one on the
right leg).

Viewing Results
Results of the analysis display in the lower right portion of the
Analysis window (see Figure 11-7).
KP0886-0825

The whole body results area of the Analysis window shows:


• Pertinent Statistical Information
The headings of the results section indicate the statistical
figures that are used in maintaining accuracy and
diagnostic precision for this machine, and are set at the
factory.
• Regions
Indicates individual regions included in the analysis and
points to the results.
• Area
Area expressed in cm² (centimeters squared).
• BMC
Bone mineral content expressed in grams.

Performing and Analyzing the Whole Body Scan 11-15


QDR Reference Manual

• BMD
Bone mineral density expressed in g/cm² (grams per
centimeter squared).
Figure 11-7 Results button
Analysis Window

Analysis Results KP1068_002-1113

While the analysis is being performed, an Automatic Low


Density analysis will be used (unless disabled) if system software
determines it is needed. Upon completion of Auto Analysis,
analysis results are displayed. If the software determined
Automatic Low Density analysis was required, it will be noted in
the analysis results label.
To view results:
Step Action
1 Click the Results button.
The results display in lower right corner of the Analysis
window.

11-16 Performing and Analyzing the Whole Body Scan


QDR Reference Manual

Step Action
Note: If the Body Composition
Analysis (BCA) option is
installed on the QDR system
then a BCA report can be
generated.

KP1068_002-1114

To reanalyze the current scan with different settings:


Step Action
Note: Once an Analysis Step 1 Click the appropriate Analysis Step Button (see
Button is modified, each Figure 11-2) and modify.
subsequent button (analysis
step) must be performed or 2 Repeat for each subsequent Analysis Step Button in the
results will not be accurate. Regions Toolbox.

To reanalyze the current scan from the main window:


Step Action
1 Click the Analyze Scan button.
2 Click on the Analyzed Scans tab.
3 Select the desired scan.
4 Click the Next >> button.
5 Continue with analysis as described earlier in this
chapter.

Exiting the Analysis


After performing the analysis, the Exit Analysis window displays.
To select a report type to print, click on the Report button in the
Exit Analysis window. Refer to Exiting the Analysis in Chapter 7
for detailed instructions.

Performing and Analyzing the Whole Body Scan 11-17


QDR Reference Manual

Generating and Printing Reports


A variety of reports can be generated from the results of the
analysis, customized to fit the needs of the examination. Generate
and print the desired report. Refer to Generating and Printing
Reports on page 7-32 of Chapter 7 for more information.

Evaluating the Image


An acceptable image should show the patient’s entire body
including the hands and feet. If the patient exceeds the table
limits, the hands and feet can be excluded from the image:
• For patients taller than 6 ft. 6 in. (198 cm), an acceptable
image may not include the feet.
• For patients wider than the exam table scan field
indicators, an acceptable image may not show the arms in
their entirety.

11-18 Performing and Analyzing the Whole Body Scan


Chapter 12
Performing and Analyzing the
AP/Lateral Spine Scan
This chapter provides procedures to perform an AP/Lateral Spine
scan and analysis.
The following models can perform an AP/Lateral Spine scan and
analysis:
• Discovery-A
• Discovery-SL
The AP/Lateral Spine scan is used to perform a measurement of
the spine in the lateral view with a parent AP scan.

Content
Subject Page
The AP/Lateral Spine Examination 12-2
Preparing the Patient 12-2
Creating/Retrieving a Patient Biography 12-2
Placing the Patient on the Table 12-2
Selecting the Scan Type and Mode 12-3
Positioning the Patient 12-4
Positioning the C-arm 12-5
Performing the AP Scan 12-7
Analyzing the AP Scan 12-9
Performing the AP Analysis 12-11
Performing the Lateral Scan 12-25
Exiting the Examination 12-28
Analyzing the Lateral Scan 12-28
Exiting the Analysis 12-42
Generating and Printing Reports 12-42
Alternate Scan Modes 12-42

12-1
QDR Reference Manual

The AP/Lateral Spine Examination


The AP/Lateral Spine examination includes all activities that
begin when the patient enters the exam room, and ends when
acceptable images of the AP and Lateral spines are obtained. The
activities involved in performing the scans include:
• Preparing the Patient
• Creating/Retrieving a Patient Biography
• Placing the Patient on the Table
• Selecting the Scan Type and Mode
• Positioning the Patient
• Positioning the C-arm
• Performing the AP Scan
• Analyzing the AP Scan
• Performing the Lateral Scan
• Analyzing the Lateral Scan
• Exiting the Analysis
• Generating and Printing Reports

Preparing the Patient


Note: Remember to ask the Prepare the patient by asking the pre-examinations questions and
patient if they are, or might be, ensuring that they are not wearing anything that will interfere
pregnant. Postpone the exam
with the scan. Refer to Preparing the Patient on page 7-2 of
until pregnancy is ruled out.
Chapter 7 for detailed instructions.

Creating/Retrieving a Patient Biography


If this is a new patient, create a new patient biography (refer to
Creating/Editing a Patient Record on page 6-9 of Chapter 6 for
detailed instructions). If this patient already has a biography,
retrieve it now (refer to Retrieving a Patient Record on page 6-7
of Chapter 6).
After creating or retrieving a patient biography, the Select Scan
Type window displays, but before selecting the scan type,
continue with Placing the Patient on the Table below.

Placing the Patient on the Table


This section provides instructions for placing the patient on the
table before beginning the AP/Lateral exam. Follow the
directions carefully to acquire the best possible scan.

12-2 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

To place the patient on the table:


Step Action
Note: Refer to Chapter 7 for a 1 Press the Patient On/Off switch on the Control Panel
detailed description of placing and seat the patient on the table between the scan field
the patient on the table.
indicators on the table pad. Their right shoulder should be
adjacent to the C-arm.
Note: Refer to Introducing the 2 Lie the patient on their back with their head at the right
Positioning Aids on page 1-11 of end of the table and place the Head Positioner under the
Chapter 1 for more information
patient’s head (see Figure 12-1).
about the positioning aids.
3 Press the Center Table button on the Control Panel.
The table and C-arm move to the center.
4 Continue with Selecting the Scan Type and Mode below.

Selecting the Scan Type and Mode


After creating, or retrieving a patient biography, the Select Scan
Type window displays.

KP0886-1201

The scan type for the AP Lumbar Spine scan is AP/Lateral that is
selected in the Select Scan Type window. A check mark () in the
Use Default Scan Mode box designates the Express mode that
takes 10 seconds.

Choosing the Scan Mode


To choose an alternate scan mode:
To perform the scan in a different mode, remove the check mark
and a window will display that allows selection of an alternate
scan mode. See “Alternate Scan Modes” on page 12-42.

Performing and Analyzing the AP/Lateral Spine Scan 12-3


QDR Reference Manual

To choose the default scan mode (check mark in box):


Step Action
1 Click on the AP/Lateral in the scan type list.
The scan type is highlighted.
2 Click the Next >> button.
The Continuously Press ENABLE... message displays.
3 Press and hold the Enable button on the Control Panel
until motion is complete.
4 Continue with Positioning the Patient below.

Positioning the Patient


When performing the AP scan, it is important to center the patient
on the table so that they do not have to move between the AP and
lateral scans. During the system setup for the AP scan, the table
will lock in the centered position. The operator will still have the
ability, however, to move the C-arm to the correct starting point
for the scan.
Figure 12-1
Patient Positioning Goal

Contour Pillow

Knee Positioner
Area to be scanned

KP0886-1202

12-4 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

To position the patient:


Step Action
1 Position the patient’s body so that the spine is straight on
the table pad.
The patient’s shoulders should be at the upper scan limit
line and the patient must be centered on the table.
2 Verify that the patient’s pelvis and shoulders are aligned
straight on the table pad and centered to the marks on the
table pad.
Note: Refer to Introducing the 3 Place the positioner under the patient’s lower legs.
Positioning Aids in Chapter 1
for more information about the
positioning aids.
4 Adjust the positioner by rotating it until the femurs are as
vertical as possible. This will help reduce the lordic curve
of the lumbar spine.

o
90

KP0886-1203

5 Place patient’s arms over their head and into the contour
pillow (see Figure 12-1 and Figure 12-2).
6 Instruct the patient to remain still and to breathe
normally.

Positioning the C-arm


Now that the patient and table are in position, and the scan type
has been selected, the next step is to position the C-arm and align
the laser.

Performing and Analyzing the AP/Lateral Spine Scan 12-5


QDR Reference Manual

To position the C-arm using the laser:


Step Action
Note: The table is locked in the 1 Press the Laser button on the Control Panel.
center position. You cannot use
the Table buttons on the control 2 Verify alignment of the patient’s midline, extending from
panel. the patient’s xiphoid process to symphysis, to the
Note: The horizontal line is the horizontal laser line. If necessary, center the patient’s
long line that extends head to midline to the horizontal laser line. There should only be
foot. minimal positioning required if the patient is centered
properly on the table.
3 Determine the position of the patient’s iliac crest.
4 Press the Laser button on the Control Panel.
5 Use the Arm Controls on the Control Panel to place the
cross hair of the laser 1 to 2” below the iliac crest and
centered in the mid-line of the patient.
6 Press the Continue button.
The Scan Parameters window displays.

Confirming Scan Parameters


The only scan parameter that can be modified is the scan length
(see Figure 12-2). The scan length for an AP/Lateral exam is
approximately 8 inches (20 cm). If the patient is taller than six
feet, you may want to increase the scan length to 9.4” (23.9 cm).
The system may change the number you enter to match the pre-
programmed step size of the scanning mechanism.
Figure 12-2
AP/Lateral Scan Parameters

KP1035A-1205

12-6 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan
Length field and type the new length.

Performing the AP Scan


An acceptable AP scan (see Figure 12-3) includes the following:
• The scan starts in the middle of L5.
• The iliac crest is evenly displayed in both lower corners of
the image area.
• The AP Lumbar Spine is centered in the middle of the
scan window.
• There are even amounts of soft tissue on each side of the
spine.
• The scan stops where ribs are attached to T12 (usually the
middle of T12).
Figure 12-3
AP Scan

KP0886-1206

Performing and Analyzing the AP/Lateral Spine Scan 12-7


QDR Reference Manual

Starting the AP Scan


To start the AP scan:
Step Action
Note: If the Control Panel X-ray 1 Click the Start Scan button
lamp fails to extinguish within The Scan window displays with the image appearing on
10 seconds after the end of the
the left side. Flashing X-rays On indicator at the top of
scan, press the red Emergency
Stop button on the Control Panel the window continues until the scan stops.
immediately. Then call Hologic
Service before resuming
operation.

KP0886-1207

2 Inspect the image and verify that the center of L2-L4 is


between the blue guide lines.
If not, the patient is not centered on the table and needs to
be repositioned and re-scanned.
3 When you see where ribs are attached to T12 on the
image, click the Stop Scan button.
While the scan is being acquired the operator has four options:

Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the Exit Exam
window displays.
Stop Scan button Terminates the scan at the point at
which the button is pressed and the
KP0886-0808 Exit Exam window displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP0886-0809 proceeds to the main window. To
start another scan the C-arm must
be repositioned.

12-8 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

Option Description
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP0886-0810 either the window image or the
patient. Refer to the following
instructions.

Repositioning the Scan


If your scan is positioned correctly it will not be necessary to
reposition. If you need to reposition, refer to Repositioning the
Scan on page 8-9 of Chapter 8.

Analyzing the AP Scan


After acquiring an acceptable image, the system automatically
proceeds to the AP analysis.
The patient must remain in position on the exam table while
performing the AP analysis and setting up the computer for the
Lateral part of the exam.

Performing the Analysis


The AP Lumbar Spine scan analysis includes all activities that
begin with a satisfactory image and ends with the calculation and
summary of the results of the scan. After completing a scan, the
system automatically saves the scan, in the Scan Drawer, for
analysis. The three parts of analysis include:
• Choosing the analysis method
Note: If the patient has an AP Lumbar Spine baseline
scan, the software proceeds directly to the manual
compare analysis function instead of choosing the
analysis method. Compare analysis is discussed in detail
in Chapter 14. Manual AP Lumbar Spine scan analysis is
discussed beginning on page 12-13.
• Performing the analysis.
• Generating a report.

Choosing the Method of Analysis


The first analysis activity involves selecting the analyze setup.
The analyze setup determines the method that the system uses to
perform the analysis. It can perform the analysis either as a single
new scan, or as a comparison to one that has been previously
performed.

Performing and Analyzing the AP/Lateral Spine Scan 12-9


QDR Reference Manual

If this scan is the patient’s first visit, select the default method
(Lumbar Spine) and the system presents a single scanned image
for analysis. If this scan is a follow-up visit, select the Compare
to Previously Analyzed Scan radio button. The system then
presents two images. The left image is the unanalyzed scan, and
the right image is a previous scan with the region of interest
(ROI) defined.
This allows the system to perform the new analysis on the same
ROI as the previous one and maximize accurate reporting of
changes. Follow-up visits are discussed in detail on page 14-7 of
Chapter 14.
To select analyze setup for a first visit:
Step Action
1 After an acceptable image has been acquired, the Analyze
Setup window displays.

KP0886-1208

Note: The Analyze Setup window 2 Click the Choose Analysis Method radio button, if
displays with the Choose necessary.
Analysis Method radio button
selected as the default. 3 Verify the analysis method (see Methods of Analysis
Comparison is covered in detail below).
on page 14-8 of Chapter 14.
4 Click the Next >> button.
The Analysis window displays.

Methods of Analysis
The Lumbar Spine analysis is the standard method for analyzing
AP scans. This method should be used first before trying Low
Density Spine analysis. The Low Density Spine analysis is used
when the Lumbar Spine analysis produces a poor yellow bone
outline.

12-10 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

To select a method of analysis (other than default):


Step Action
1 Click the drop down list arrow on the Analyze Setup
window.
List of methods displays.
Note: Subregion Array Spine
analysis is available as an
option.

KP1068_002-0813

2 Click on the desired method.


List disappears, and selected method displays in field.
3 Click the Next >> button to continue.
The Analysis window displays.

Performing the AP Analysis


Note: Refer to “Analyze Tab” on Unless disabled, once the method of analysis is chosen the
page 18-5 of Chapter 18 on how software will automatically perform the analysis. This is called
to disable One-Time™ Auto
One-Time™ Auto Analysis and is a feature of all models in the
Analysis.
QDR Series.
After the One-Time™ Auto Analysis is completed the operator
can choose to perform a manual analysis of the acquired scan.
Both analysis methods are described below.

One-Time™ Auto Analysis


The AP Lumbar Spine scan auto analysis includes all activities
that begin with a satisfactory image and ends with the calculation
and summary of the results of the scan. After completing a scan,
the system automatically saves the scan, in the Scan Drawer, for
analysis.

While One TimeAuto Analysis is being performed, no operator


interaction is allowed. Upon completion of Auto Analysis,
analysis results are displayed.

Performing and Analyzing the AP/Lateral Spine Scan 12-11


QDR Reference Manual

Figure 12-4
AP Lumbar Spine
OneTime Auto Analysis
Results

Note: Evaluate the analysis for


correct placement of regions and
bone outline. It may be
necessary to adjust the image
display.

KP1068_002-1210
Analysis results

Manual Analysis
The operator may perform a manual analysis of the lumbar spine
scan after the auto analysis is completed. In most cases it will not
be necessary to perform manual analysis. However, upon
examination of the auto analysis results, if one of the following
conditions is seen, manual analysis may be required:
• regions are positioned incorrectly
• poor bone mapping
To perform manual analysis after auto analysis:
Step Action
1 Click on the Global ROI button (see Figure 12-4).
2 Perform the Manual Analysis starting with Performing
the Manual Analysis (page 12-13).

A specific set of tools, the Analysis Step Buttons, located on the


left side of the window (see Figure 12-5), allow the operator to
proceed step-by-step through each task. These tools can
maximize image quality and accuracy, preventing the need for re-
scanning.

12-12 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

Figure 12-5 Step buttons


AP Lumbar Spine Analysis
Window

KP0886-1211
Toolbox
Brightness/Contrast button

Performing the Manual Analysis


The steps to manual analyze the AP scan include:
1. Adjusting the image display brightness and contrast (Bright-
ness/Contrast button).
2. Defining the Global Region of Interest (Global ROI button).
3. Viewing/editing the bone map (Bone Map button).
4. Marking the Intervertebral spaces (Vertebral Line button).
5. Labeling the Vertebral bodies and view analysis results
(Results button).
6. Exiting the Analysis window (Close button).

Adjusting the Image Display


The contrast and brightness of the image can be adjusted to
obtain the best definition of important anatomical features.
Adjusting the contrast and brightness has no effect on
KP0886-0817 calculations; it is used to optimize the display and for visual
adjustments only. Refer to Contrast and Brightness on page 7-
25, in Chapter 7, for detailed instructions.

Performing and Analyzing the AP/Lateral Spine Scan 12-13


QDR Reference Manual

Defining the Region of Interest


The global ROI refers to the defined boundaries of the image that
is being analyzed. The ROI appears on the image as a box: the
dashed line is active and can be adjusted; the solid lines are
KP0886-0818
inactive and cannot be adjusted.
Figure 12-6 Dashed yellow line
AP ROI Image in the Area to be analyzed
Analysis Window

KP0886-1212

Solid red lines


A properly positioned Global ROI includes:
• The spine centered within the ROI.
• A preset width of 116 lines (automatically preset).
• The top line of the ROI positioned between T12 and L1.
• The bottom line of the ROI positioned between L4 and L5

12-14 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

The Global ROI Toolbox is used to define the region of interest:


• Line Mode
When selected, one line on the ROI box initially displays
as a dashed yellow line. Any of the ROI box lines can be
selected and then moved. Top or bottom (horizontal) lines
can be moved up or down. Side (vertical) lines can be
moved left or right.
• Whole Mode
When selected, the entire ROI box initially displays as a
dashed yellow line. This allows the entire box to be
moved to center the spine while maintaining the 116 line
width.
• Point Mode
When selected, a plus sign displays in each corner of the
ROI box, as well as at the center of the top and bottom
lines. The point mode should not be used when analyzing
KP0886-1213 an AP/Lateral pair.
To define the ROI in Line Mode:
Step Action
1 Click the Global ROI button in the top left corner of the
window, if not already selected.
The ROI box displays in the image area.
2 The Line Mode button is the default and should be active
(click if not already selected).
The top line of the ROI box displays as an active line
(dashed yellow).
Note: Dashed yellow lines are 3 Click and drag the top line to its position between T12
active and can be edited. Solid and L1 intervertebral space. (Evenly spaced, if possible.
red lines are inactive.
Note: The top line remains 4 Click and drag the bottom line to its position between L4
dashed until the bottom line is and L5 Intervertebral space. (Evenly spaced, if possible.)
clicked.

To define the ROI in Whole Mode:


Step Action
1 Click the Global ROI button in the top left corner of the
window, if not already selected.
The ROI box displays in the image area.
2 Click the Whole Mode button in the Global ROI
Toolbox.
The ROI box displays as active lines (dashed yellow).

Performing and Analyzing the AP/Lateral Spine Scan 12-15


QDR Reference Manual

Step Action
3 Position the cursor arrow anywhere within the ROI box.
The crossed arrows cursor displays.
4 Center the ROI box over the spine by clicking and
dragging it to the correct position.

Viewing/Editing the Bone Map


The system creates a yellow bone map of the spine over the ROI
defined in the previous steps. It automatically fills holes in the
yellow bone map and removes bone artifacts.
KP0886-0821

To view the bone map:


Step Action
1 Click the Bone Map button.
The bone map displays in yellow.

KP0886-1214

2 Verify whether the bone map is acceptable or needs to be


edited.
Evaluate the bone map:
It is strongly recommended that the operator not alter the bone
edges of the spine that are displayed in yellow. This degrades the
reproducibility of the spine image. If large areas of the bone map
are automatically excluded, the operator must edit the bone map.
To edit the bone map:

12-16 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

To add or delete bone from the image, use the Bone Map Toolbox
portion of the Analysis window. The operator can also adjust the
cursor size, fill holes, undo changes, or reset the image to the
original bone map if unintentionally changed.

KP0886-1215

To add bone to the bone map:


Step Action
Important: Do not alter the bone 1 Click the Add Bone radio button.
edges of the spine, as it degrades The cursor displays as a small green rectangular paint
the reproducibility of the scan. If
brush when placed inside the global ROI on the image.
necessary, use the Low Density
method to improve bone 2 Select the cursor size using the Cursor Size tool.
mapping. See the Scanning Hologic recommends the default paint brush (size 2) be
Special Conditions section of
used.
this chapter.
3 Move the cursor to the starting point.
Note: Be careful to follow the 4 Click and drag the cursor to connect outside edges to
bone edge to bridge all gaps. If complete the bone map.
there are any gaps, the Fill
Holes function will not work. 5 To automatically fill the remaining holes, click the Fill
Holes button.

Performing and Analyzing the AP/Lateral Spine Scan 12-17


QDR Reference Manual

To delete bone from the bone map:


Step Action
Note: For spurs or prominent 1 Click the Delete Bone radio button.
Transfers Processes, retain the The cursor displays as a small green rectangular paint
image as painted by the system
brush when placed inside the global ROI on the image.
to maintain the reproducibility of
the scan. 2 Select the cursor size using the Cursor Size tool.
Hologic recommends the default paint brush (size 2) be
used.
3 Move the cursor to the starting point.
Note: Be careful to delete a 4 Click and drag the cursor to erase a line through the
complete line through the bone yellow bone map.
map until the image separates.
This creates a small island—an isolated section of bone.
Note: If you leave any 5 To automatically remove the small island, click on the
connection, only the line you Sink Islands button.
deleted is removed.
Small island is removed.

Marking Intervertebral Spaces


Marking each intervertebral space with a line allows each
individual vertebra to be analyzed separately. The Vertebral
Lines button is used to mark the intervertebral spaces within the
KP0886-0824
region of interest. Up to three lines appear on the image that can
be used to mark the spaces. You may insert up to three lines or
delete lines as necessary.

To mark intervertebral lines using Line Mode:


Step Action
1 Click the Vertebral Lines button, if not already selected.
Three horizontal lines display in the image area.
Note: Dashed yellow lines are 2 The Line Mode button is the default and should be active
active and can be edited. Solid (click if not already selected).
red lines are inactive.
The top line displays as an active line (dashed yellow).
Note: The yellow dashed line is 3 Click and drag the top dashed yellow line and place it in
active until the next line is the L1/L2 space, evenly spaced between vertebrae.
selected.
The top line displays as active until you select the next
line.
Note: Active yellow dashed lines 4 Click and drag the middle line and place it in the L2/L3
can not cross a solid line except space, evenly spaced between vertebrae.
in point mode.
The middle line displays as active until you select a
different line.

12-18 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

Step Action
5 Click and drag the lowest intervertebral line and place it
in the L3/L4 space, evenly spaced between vertebrae.
The lowest line displays as active until you select a
different line.

KP0886-1216

To remove a line:
Step Action
Note: Use if analysis is three 1 Click on the line.
vertebrae or less. Line displays as active (dashed yellow).
2 Click the Delete Line button.
Line disappears.

To insert a line:
Step Action
Note: A maximum of three 1 Click the Insert Line button.
vertebral lines can appear on the A new line displays as active (dashed yellow), just above
image.
the previous active line, if any.

Point Mode
Point mode is used to mark intervertebral spaces of the Scoliotic
space and is explained in Scoliotic Spine Scans on page 8-30 of
Chapter 8.

Performing and Analyzing the AP/Lateral Spine Scan 12-19


QDR Reference Manual

Labeling The Vertebral Bodies


The Results button is used to automatically label the marked
vertebral bodies. Vertebra labels are automatically placed in
areas outside the bone region when possible.
KP0886-0825

Labeling Guidelines
Use the following guidelines to verify labels:
• Labels are automatically assigned numbers starting at the
top with L1.
• The top-most vertebra within the ROI is always labeled
L1.

Including or Excluding Vertebrae from Analysis


Once the operator verifies that the labels correspond to the correct
vertebral bodies, specific vertebrae can be included or excluded
from the analysis. Some reasons for excluding vertebral bodies
include: foreign objects (artifacts) within the scan area, the scan
area is a surgical location, or a physician requested the exclusion
(physician’s discretion).
To automatically label vertebrae:
Step Action
1 Click the Results button, if not already selected.
The vertebral bodies are automatically labeled starting
with L1.
2 Verify that the labels correspond correctly to the anatomy.
Usually no changes are necessary.

KP0886-1217

12-20 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

To change vertebrae labels:


Step Action
1 To change the label numbering, click the ▲ or ▼ arrow in
the First Label area of the Label Toolbox.
The label numbering on the image increments or
decrements accordingly. The first (top-most) label
displays in the First Label box.
2 To include or exclude a vertebra, click anywhere within
the boundary of the labeled vertebra on the image.
The ROI boundary for the selected vertebra displays as
active (dashed yellow).
3 To exclude, click the Exclude button.
The label on the image turns red indicating that the
vertebra has been removed.
4 To include previously excluded vertebra, click the
Include button.
The label on the image turns yellow indicating that the
vertebra has been included.

Viewing Results
Results of the analysis display in the lower right portion of the
Analysis window (see Figure 12-7).
KP0886-0827

The spine results window shows:


• Pertinent Statistical Information
The headings of the results section indicate the statistical
figures that are used in maintaining accuracy and
diagnostic precision for this machine, and are set at the
factory.
• Regions
Indicates individual vertebra included in the analysis and
point to the results.
• Area
Results display in cm² (centimeters squared).
• BMC
Bone mineral content expressed in grams.
• BMD
Bone mineral content expressed in g/cm² (grams per
centimeter squared).

Performing and Analyzing the AP/Lateral Spine Scan 12-21


QDR Reference Manual

To view results:
Step Action
1 The Results button should have already been selected
when labeling vertebrae. (Click if not already selected).
The results display in lower right corner of the Analysis
window.
Figure 12-7 Results button
AP Analysis Window
Results Area

KP1068_002-1218
Results area

Verifying the Start Position


The Start Position button allows the operator to verify the
starting point for the Lateral scan. The starting position is
calculated from the placement of the bottom of the ROI. If the
bottom line of the ROI is incorrectly placed, then the starting
position for the Lateral scan will be incorrect.
When the Start Position button is clicked, the system calculates
the centerline. If the patient positioning was incorrect, it may be
necessary to stop the analysis, reposition the patient, and repeat
the AP scan.
Centerline Warnings
The system displays the following warnings, depending on the
calculation of the centerline:

12-22 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

Centerline Angle
Note: If the centerline angle is
greater than 5 degrees the
system will not allow the
acquisition of a supine lateral
BMD scan.

KP0886-1219

The system displays this message if the centerline angle is greater


than the recommended maximum of 5 degrees from vertical. If
the system displays this warning message, decide whether to
reposition the patient and repeat the AP scan, or continue the
analysis.
• Click the Yes button to repeat the AP scan. If Yes is
selected, the system does not save the analysis and returns
you to the scan process.
• Click the No button to continue. If No is selected, the
system saves the analysis and returns to the Exit Exam
window.
Centerline Outside of Range

KP0886-1220

The system displays this message if the centerline extends outside


of the acceptable range. If this warning occurs, stop the analysis,
reposition the patient, and repeat the AP scan. Click the OK
button to return to the Exam setup window.

Performing and Analyzing the AP/Lateral Spine Scan 12-23


QDR Reference Manual

Centerline not Calculated

KP0886-1221

The system displays this message if it is unable to calculate the


centerline. If this warning message occurs, stop the analysis,
reposition the patient, and repeat the AP scan. Click the OK
button to return to the Exam setup window.
To check the start position:
Step Action
1 Click the Start Position button.
The system displays the Start Position Toolbox with a
Reset button. The scan image displays with a centerline
within L2 and L4 and an arrow indicating the start of the
Lateral scan.
Note: The start position of the 2 If the start position of the AP portion of the AP/Lateral
AP portion of the AP/Lateral scan is incorrect, reanalyze the AP to include the correct
scan should not be changed.
vertebrae.

To reanalyze the current scan with different settings:


Step Action
Note: Once an Analysis Step 1 Click the appropriate Analysis Step Button (see Figure
Button is modified, each 12-5) and modify.
subsequent button (analysis
step) must be performed or 2 Repeat for each subsequent Analysis Step Button in the
results will not be accurate. Global ROI Toolbox.

To reanalyze the current scan from the main window:


Step Action
1 Click the Analyze Scan button.
2 Click on the Analyzed Scans tab.
3 Select the desired scan.
4 Click the Next >> button.

12-24 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

Step Action
5 Continue with analysis as described earlier in this
chapter.
To continue:
Step Action
1 Click the Close button.
The Continuously Press ENABLE... message displays.

Performing the Lateral Scan


Important: If the patient moves, The Lateral setup prepares the system for lateral scanning. It can
the AP setup and AP scan must only be performed after completion of the AP analysis for the
be repeated before performing
same patient, without the patient moving.
the lateral scan.
Performing the Lateral part of the AP/Lateral exam involves the
following activities:
• Rotating the C-arm to the lateral scan position
• Scanning
• Analyzing the Lateral scan

Rotating the C-Arm for the Lateral Scan


After completing the analysis of the AP scan by clicking the
Close button in the AP Analysis window, the system
automatically proceeds to the Lateral part of the exam and
displays the following message.

Continuously press ENABLE switch to move system into position

Abort

KP0886C-1222

Continuously press the Enable button on the Control Panel until


the C-arm has fully rotated to the lateral scan position.
Once the C-arm is in the correct position, the Scan Parameters
window displays.

Performing and Analyzing the AP/Lateral Spine Scan 12-25


QDR Reference Manual

Remind the patient not to touch the sides of the table during C-
arm rotation.

Table Safety Feature


Note: If the Table button does The Discovery-A and -SL provides a built-in safety feature. If
not turn back on, call the you or the patient touch the safety strips along either edge of the
Hologic Help desk.
table, the C-arm motion stops and emits a musical tone.
Once the tone stops, the Table button on the Control Panel turns
back on and you can continue. Press the Enable button to
complete the C-arm rotation.

Confirming Scan Parameters


The only scan parameter that can be modified is the scan length
(see Figure 12-8). The scan length for a Lateral spine exam is
approximately 6 inches (15 cm).
Figure 12-8
Lateral Scan Parameters

KP1035A-1223

To verify scan parameters:


Step Action
1 Verify patient name and scan type in the upper left corner
of the Scan Parameters window.
2 Verify the scan length.

Starting the Lateral Scan


Once the patient and C-arm are positioned correctly, return to the
computer to start the Lateral scan.

12-26 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

To start the Lateral scan:


Step Action
Note: If the Control Panel X-ray 1 Click the Start Scan button
lamp fails to extinguish within The Scan window displays with the image appearing on
10 seconds after the end of the
the left side. Flashing X-rays On indicator at the top of
scan, press the red Emergency
Stop button on the Control Panel the window continues until the scan stops.
immediately. Then call Hologic
Service before resuming
2 When the lateral scan begins, the system evaluates the
operation.
patient’s size and recommends the appropriate scan array
mode.

KP0886C-1224

3 Inspect the image.


4 As soon as L2–L4 are fully displayed, click the Stop
Scan button
The Continuously Press ENABLE... message displays.
Caution: The patient should be 5 Continuously press the Enable Lateral button on the
instructed to not touch the sides Control Panel until motion stops.
of the table during C-arm
The C-arm moves to the correct position.
rotation.

While the scan is being acquired the operator has three options:
Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the Exit Exam
window displays.

Performing and Analyzing the AP/Lateral Spine Scan 12-27


QDR Reference Manual

Option Description
Stop Scan button Terminates the scan at the point at
which the button is pressed and the
KP0886-0808 Exit Exam window displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP0886-0809 proceeds to the application main
window. To start another scan the
C-arm must be repositioned.

Removing the Patient From the Table


When an acceptable Lateral image has been obtained, perform the
following steps to allow the patient to get off the table.

Step Action
1 Continuously press the Enable Lateral button on the
Control Panel until motion stops.
The C-arm rotates and the table moves to the home
position.
2 Press the Patient On/Off button to move the C-arm and
table to permit the patient to get off the table.
3 Instruct the patient to get off the table.

Exiting the Examination


After acquiring an acceptable image, the Exit Exam window
displays. To perform the analysis of the scan just completed, click
the Analyze Scan button in the Exit Exam window. Refer to
Exiting the Examination on page 7-23 of Chapter 7 for detailed
instructions.

Analyzing the Lateral Scan


The Lateral scan analysis includes all activities that begin with a
satisfactory image and ends with the calculation and summary of
the results of the scan. After completing a scan, the system
automatically saves the scan, in the Scan Drawer, for analysis.
The three parts of analysis include:

12-28 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

• Choosing the analysis method.


Note: If the patient has a Lateral baseline scan, the
software proceeds directly to the compare analysis
function instead of choosing the analysis method.
Compare analysis is discussed in detail in Chapter 14.
Lateral scan analysis is discussed beginning on page 12-
31.
• Performing the analysis.
• Generating a report.

Choosing the Method of Analysis


The first analysis activity involves selecting the analyze setup.
The analyze setup determines the method that the system uses to
perform the analysis. It can perform the analysis either as a single
new scan, or as a comparison to one that has been previously
performed.
If this scan is the patient’s first visit, select the default method
(Lumbar Spine) and the system presents a single scanned image
for analysis. If this scan is a follow-up visit, select the Compare
to Previously Analyzed Scan radio button. The system then
presents two images. The left image is the unanalyzed scan, and
the right image is a previous scan with the region of interest
(ROI) defined.
This allows the system to perform the new analysis on the same
ROI as the previous one and maximize accurate reporting of
changes. Follow-up visits are discussed in detail in Chapter 13.

Performing and Analyzing the AP/Lateral Spine Scan 12-29


QDR Reference Manual

Step Action
1 From the Exit Exam window, click the Analyze Scan
button.
The Analyze Setup window displays.

KP0886C-1226

Note: The Analyze Setup window 2 Click the Choose Analysis Method radio button, if
displays with the Choose necessary.
Analysis Method radio button
selected as the default. 3 Verify the analysis method (there is only one choice).
Comparison is covered in detail
on page 14-8 of Chapter 14. 4 Click the Next >> button.
The Analysis window displays.

12-30 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

Performing the Lateral Analysis


Once the setup method has been selected, continue to the
Analysis window to adjust the quality of the image and properly
define the area of analysis called the region of interest (ROI).
A specific set of tools, the Analysis Step Buttons, located on the
left side of the window (see Figure 12-9), allow the operator to
proceed step-by-step through each task. These tools can
maximize image quality and accuracy, preventing the need for re-
scanning.
Figure 12-9 Step buttons
Lateral ROI Image in the
Analysis Window

Toolbox KP0886C-1226

Brightness/Contrast button

The steps to analyze the Lateral scan include:


1. Adjusting the image display brightness and contrast (Bright-
ness/Contrast button).
2. Defining the Global Region of Interest (Global ROI button).
3. Adjusting the Vertebral boundaries (Vertebral Boundaries
button).
4. Adjusting the Vertebral bodies (Vertebral Bodies button).
5. Viewing analysis results (Results button).
6. Exiting the Analysis window (Close button).

Performing and Analyzing the AP/Lateral Spine Scan 12-31


QDR Reference Manual

Adjusting the Image Display


The contrast and brightness of the image can be adjusted to
obtain the best definition of important anatomical features.
Adjusting the contrast and brightness has no effect on
KP0886-0817 calculations; it is used to optimize the display and for visual
adjustments only. Refer to Contrast and Brightness, in Chapter 7,
for detailed instructions.

Defining the Region of Interest (ROI)


The global ROI refers to the defined boundaries of the image that
is being analyzed. The ROI appears on the image as a
parallelogram with a bisecting line running down the middle and
KP0886-0818
the top and bottom edges initially positioned according to the AP
scan: the dashed line is active and can be adjusted; the solid lines
are inactive and cannot be adjusted.
Figure 12-10
Lateral ROI Image in the
Analysis Window

Spinous
Process

5 Lines
KP0886C-1227

A properly positioned Global ROI includes:


• The bisecting line positioned close to the vertebral bodies,
without touching their posterior edges.
• The right edge of the ROI positioned ½ of a vertebral
width anterior to the anterior edges of the vertebral
bodies.
• The left edge of the ROI positioned 5 lines to the right of
the posterior edge of the shortest spinous processes, inside
of the processes (see Figure 12-10).

12-32 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

The Global ROI Toolbox


The Global ROI Toolbox is used to define the region of interest:
• Line Mode
When selected, one line on the ROI box initially displays
as a dashed yellow line. Any of the ROI box lines can be
selected and then moved. To preserve the synchronization
of the AP and Lateral scans, all lines can only be moved
left or right.
• Whole Mode
When selected, the entire ROI box initially displays as a
dashed yellow line. This allows the entire box to be
moved as a unit.
• Point Mode
When selected, a plus sign displays in each corner of the
ROI box, as well at each end of the Posterior Boundary
line.

KP0886C-1228

Reanalysis Warning
If the AP scan has been reanalyzed, the system displays a warning
when the Global ROI button is selected. You have the option to
reset the ROI to match the new AP baseline scan analysis.

KP0886C-1229

If the lateral scan about to be analyzed was previously analyzed,


the system bypasses the Analysis Step Buttons and displays only
the Results button.
To define the ROI in Line Mode:
Step Action
1 Click the Global ROI button in the top left corner of the
window, if not already selected.
The ROI box displays in the image area.

Performing and Analyzing the AP/Lateral Spine Scan 12-33


QDR Reference Manual

Step Action
2 The Line Mode button is the default and should be active
(click if not already selected).
The top line of the ROI box displays as an active line
(dashed yellow).
Note: Dashed yellow lines are 3 Click and drag the top line until the bisecting line is
active and can be edited. Solid against the posterior edge of L2. Note, that the sides of
red lines are inactive.
the ROI and bisecting line remain parallel.
Note: The top line remains 4 Click and drag the bottom line until it is against the
dashed until the bottom line is posterior edge of L4.
clicked.
5 Click and drag the left (posterior) line so that it is five
lines to the right of the shortest posterior edge of the
spinous processes, as shown in Figure 12-10.
To define the ROI in Whole Mode:
Step Action
1 Click the Global ROI button in the top left corner of the
window, if not already selected.
The ROI box displays in the image area.
2 Click the Whole Mode button in the Global ROI
Toolbox.
The ROI box displays as active lines (dashed yellow).
Note: Be careful not to set the 3 Position the cursor arrow anywhere within the ROI box.
posterior edge of the ROI The crossed arrows cursor displays.
outside of the spinous process as
the inclusion of air in the ROI 4 Position the ROI so that the posterior boundary line
degrades the precision and includes as much of the posterior spinous process as
accuracy of the measurement. possible without including air in the ROI.

Adjusting the Vertebral Boundaries


Clicking the Vertebral Boundaries button allows the operator to
place the anterior boundary of the vertebral bodies and confirm
the vertebral lines that divide the vertebral bodies.
To adjust the Anterior Boundary:
Step Action
1 Click the Vertebral Boundaries button, then choose the
Ant. Boundary radio button in the Boundaries Toolbox.
The Anterior Boundary displays on the image as a
dashed yellow line, along with the Global ROI, Posterior
Boundary, vertebral bodies regions, Vertebral Spacer
lines and labels.

12-34 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

Step Action
2 Click the Line Mode button and move the dashed yellow
Anterior Boundary line to the most anterior point of L2-
L4.
3 Click the Point Mode button and adjust the anterior
points to match the curve of the spine.
4 Click the Vertebral Lines radio button to adjust the
spacer lines between the vertebrae. This step is not
usually necessary as the line placement is based on the
location of the AP scan.

Adjusting Vertebral Bodies


The Vertebral Bodies button allows the operator to mark the
boundaries of the vertebral bodies. The most important edge is
the anterior side. Be careful not to make the box larger than the
actual vertebral body.
To adjust the Vertebral Bodies in Whole Mode:
Step Action
Note: Line Mode is the default 1 Click the Vertebral Bodies button.
when the Vertebral Bodies The system displays three Vertebral Region boxes—one in
button is selected.
each vertebral body.
Note: It may be necessary to 2 Click and drag the active box (dashed yellow lines) just
adjust the size and shape of each inside the vertebral body.
vertebral body using Line and
Point Mode.
3 Repeat for each remaining vertebral body.

To adjust the Vertebral Bodies in Line Mode:


Step Action
1 Click the Line Mode button.
The top line of the selected vertebral body displays as
active (dashed yellow line).

Performing and Analyzing the AP/Lateral Spine Scan 12-35


QDR Reference Manual

Step Action
2 Adjust each side of each box so that it fits just inside the
vertebral body.

KP0886C-1231

To adjust the Vertebral Bodies in Point Mode:


Step Action
1 Click the Point Mode button.
A plus sign displays in each corner of the selected
vertebral body, one of which is active (yellow).
2 Adjust each corner of each box so that it conforms to the
shape of the vertebral body.

Confirming Lumbar Vertebrae Labels


The system automatically labels the lumbar vertebrae according
to the AP scan. Do not change the labels.

KP0886C-1230

12-36 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

If necessary, you can remove L4 from the analysis if the iliac


crest is overlapping all, or a portion of L4.
To exclude/Include from analysis:
Step Action
1 In Whole Mode, select the vertebral body to be excluded
or included in the analysis.
The vertebral body is highlighted with an interior dashed
box.
2 Perform one of the following:
KP0886C-1230a To remove the vertebral body from the analysis
click the Exclude button.
To add the vertebral body to the analysis click the
KP0886C-1230b
Include button.

Viewing/Editing the Bone Map


The system creates a yellow bone map of the spine over the ROI
defined in the previous steps. It automatically fills holes in the
yellow bone map and removes bone artifacts.
KP0886-0821

To view the bone map:


Step Action
1 Click the Bone Map button.
The bone map displays in yellow.

KP0886C-1232

Performing and Analyzing the AP/Lateral Spine Scan 12-37


QDR Reference Manual

Step Action
2 Verify whether the bone map is acceptable or needs to be
edited. If the image is unacceptable, click the Vertebral
Bodies button and re-adjust R2-R4 so that they are just
inside the bone edges.
Evaluate the bone map:
It is strongly recommended that the operator not alter the bone
edges of the spine that are displayed in yellow. This degrades the
reproducibility of the spine image. If large areas of the bone map
are automatically excluded, the operator must edit the bone map.
To edit the bone map:
If you properly adjusted the Vertebral Bodies in the previous step,
then you should not need to edit the bone map. For instructions on
editing the bone map, refer to the bone map editing topic in the
AP analysis section of this chapter. The bone map on a lateral
scan is not usually edited.

Adjusting the Mid Regions


If present, use the Mid Regions button to adjust the regions that
report the bone density in the interior (middle) of the vertebral
bodies.
To adjust the Mid Regions in Whole Mode:
Step Action
1 Click the Mid Regions button.
The Global ROI, Posterior, and Anterior Boundaries
display along with one mid region in each of the vertebral
spaces.
2 The Whole Mode button is the default and should be
active (click if not already selected).
One of the mid regions displays as active (dashed yellow
lines).
3 Click on the desired mid region and adjust as needed.
Place the mid regions in the center and slightly anterior in
each vertebral body.

12-38 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

To adjust the Mid Regions in Line Mode:


Step Action
1 Click the Mid Regions button, if necessary.
The Global ROI, Posterior, and Anterior Boundaries
display along with one mid region in each of the vertebral
spaces.
2 Click the Line Mode button.
The top line of one of the mid regions displays as active
(dashed yellow line).
Note: Do not move any other 3 Click and drag each line of the mid region as necessary.
edge.
4 Repeat for each remaining mid region.

To adjust Mid Regions in Point Mode:


Step Action
1 Click the Mid Regions button, if necessary.
The Global ROI, Posterior, and Anterior Boundaries
display along with one mid region in each of the vertebral
spaces.
2 Click the Point Mode button.
Plus signs appear at the corners of one of the mid
regions.
3 Click and drag corner points to adjust the mid region as
necessary.
4 Repeat for each remaining mid region.

Viewing Results
Results of the analysis display in the lower right portion of the
Analysis window (see Figure 12-11).
KP0886-0825

The Lateral results window shows:


• Pertinent Statistical Information
The headings of the results section indicate the statistical
figures that are used in maintaining accuracy and
diagnostic precision for this machine, and are set at the
factory.

Performing and Analyzing the AP/Lateral Spine Scan 12-39


QDR Reference Manual

• Regions—Vertebral Bodies
Indicates individual vertebra included in the analysis and
points to the results.
• Area
Area expressed in cm² (centimeters squared).
• BMC
Bone mineral content expressed in grams.
• BMD
Bone mineral density expressed in g/cm² (grams per
centimeter squared).
Figure 12-11 Results button
Lateral Lumbar Spine
Analysis Window

Results area KP1068_002-1233

Results toolbox
To view results:
Step Action
Note: Information about Mid 1 Click the Results button.
Regions only displays if the Mid The system displays the Results Toolbox with two buttons:
Regions option is turned on.
BMD and WA-BMD.

12-40 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

Step Action
2 • Click the BMD button to display Bone Mineral
Density results.

KP1068_002-1234

• Click WA-BMD button to display Width-


Adjusted BMD results.

KP1035A-1235 -

Performing and Analyzing the AP/Lateral Spine Scan 12-41


QDR Reference Manual

To reanalyze the current scan with different settings:


Step Action
Note: Once an Analysis Step 1 Click the appropriate Analysis Step Button (see
Button is modified, each Figure 12-5) and modify.
subsequent button (analysis
step) must be performed or 2 Repeat for each subsequent Analysis Step Button in the
results will not be accurate. Global ROI Toolbox.

Exiting the Analysis


After performing the analysis, the Exit Analysis window displays.
To select a report type to print, click on the Report button in the
Exit Analysis window. Refer to Exiting the Analysis on page 7-27
of Chapter 7 for detailed instructions.
To exit the Analysis window after obtaining results:
Step Action
1 Click the Close button in the lower left corner of the
Analysis window.

To exit the Analysis window before obtaining results:


Step Action
1 Click the Cancel button in the lower left corner of the
Analysis window.

Generating and Printing Reports


A variety of reports can be generated from the results of the
analysis, customized to fit the needs of the examination. Generate
and print the desired report. Refer to Generating and Printing
Reports on page 7-32 of Chapter 7 for more information.

Alternate Scan Modes


You may need to use an alternate scan mode if:
• The patient was previously scanned using a scan mode
other than the default.
• The patient is very large (a longer scan time may improve
the image quality).

12-42 Performing and Analyzing the AP/Lateral Spine Scan


QDR Reference Manual

The scan modes include:

Scan Mode Time


High Definition (h) 120 seconds
Array (a) 60 seconds
Fast Array (f) 30 seconds

Express (x) 1 10 seconds


1 Default

To choose a specific scan mode from the Select Scan Type


window:
Step Action
1 Click on the AP/Lateral exam type in the list.
The selected scan type is highlighted.
2 Remove the default scan mode check mark by clicking in
the check box.
Check mark removed.
3 Click the Next >> button to continue.
The Select AP Lumbar Spine Scan Mode for AP/Lateral
Exam window displays.

KP1068A-1236

4 Click on the desired scan mode in the list.


The selected mode is highlighted.

Performing and Analyzing the AP/Lateral Spine Scan 12-43


QDR Reference Manual

Step Action
5 Click the Next >> button to continue.
The Select Lateral Scan Mode for AP/Lateral Exam
window displays.

KP0886C-1237

6 Click on the desired scan mode in the list.


The selected mode is highlighted.
7 Click the Next >> button to continue.
The Continuously Press ENABLE... message displays.
8 Continuously press and hold the Enable button on the
Control Panel.
The C-arm moves to the center position and the table
locks into place.
9 Press the Continue button.
The system displays the following message, and turns the
laser on.

KP0886C-1238

12-44 Performing and Analyzing the AP/Lateral Spine Scan


Chapter 13
Instant Vertebral Assessment
and Quantitative Morphometry
This Chapter provides procedures to perform Instant Vertebral
Assessment (IVA HD and IVA) and Quantitative Morphometry
(QM) examinations on the QDR systems. IVA provides
functionality for the acquisition of scans that allow the
determining of the presence or absence of vertebral fractures. QM
provides the ability to analyze the shape of vertebra using
vertebral height measurements and calculations of those heights
to compute wedge ratios. The anatomical sites that are available
to view are the AP and lateral vertebral bodies from T4 to L4. The
process includes performing the scan and viewing the image.
IVA scans are intended for the visualization or quantitative
assessment of vertebral bone deformities. IVA also allows the
visualization of abdominal aortic calcification, and, if present,
clinical correlation may be advised since abdominal aortic
calcification may be associated with cardiovascular disease.

Contents
Subject Page
Introduction 13-2
The IVA Examinations 13-4
IVA Imaging on an A or SL System 13-4
IVA Imaging on a C, Ci, W or Wi System 13-21
IVA with BMD 13-38
IVA Image Viewer 13-55
Using Markers 13-61
Generating and Printing Images and Reports 13-69
Interpreting the IVA Image 13-74

13-1
QDR Reference Manual

Introduction
Quantitative Morphometry (QM) analyzes the shape of the
vertebra using vertebral height measurements and calculations of
those heights to compute wedge ratios. This information can be
used directly or compared over time.

Specification
IVA HD and IVA specifications are found in the Discovery Series
and Explorer Technical Specifications Manual.

Patient Position and Scanning


Positioning the patient is different on an A and SL system than on
a C, Ci, W and Wi system. See the description below for your
system.

Patient Positioning on A or SL System


To complete an AP and Lateral IVA examination on an A or SL
system, the patient is placed in the AP Lumbar Spine position
(Figure 13-1) and the AP IVA scan is performed. Between the AP
IVA scan and the Lateral IVA scan the patient must remain
motionless while the C-arm is rotated into the lateral position and
the Lateral IVA scan is performed.
Figure 13-1
Patient Positioning Goal
Knee Positioner
Contour Pillow

Area to be
scanned

KP1068A-QM-01

13-2 IVA and QM


QDR Reference Manual

Patient Positioning on C, Ci, W or Wi System


Note: The patient’s head must be To complete an AP and Lateral IVA examination on a C, Ci, W or
towards the right side of the table. Wi system, the patient is placed in the AP Lumbar Spine position
and the AP IVA scan is performed. Between the AP IVA scan and
the Lateral IVA scan the patient must be repositioned from the
lumbar Spine scan position (on their back) to a lateral scan
position (on their side). The spine should remain in relatively the
same position on the examination table (Figure 13-2).
Figure 13-2 Patient in Lumbar Spine scan position
Patient Positioning

Lower limits for Spine should remain


scanning spine in the same relative
on the mattress. position for both scans.

Patient in Lateral scan position

IVA and QM 13-3


QDR Reference Manual

The IVA Examinations


The IVA option allows two types of IVA examinations to be
performed on the QDR systems:
• IVA Imaging
• IVA with BMD

IVA Imaging Exam


The IVA Imaging examination can be performed on any
Discovery A/SL/C or W system. The examination consists of
performing a Lateral IVA scan, an AP IVA scan, or both. A report
or an image of the scan and the analysis can be printed.

IVA with BMD Exam


The IVA with BMD examination can be performed on an A or SL
system only. The examination is the same as the IVA Imaging
examination except that it includes AP/Lateral pair BMD.

IVA Imaging on an A or SL System


The IVA Imaging examination includes all activities that begin
when the patient enters the exam room, and ends when an
acceptable image is obtained. The activities involved in
performing the exam include:
• Preparing the Patient
• Creating/Retrieving a Patient Biography
• Starting the Examination
• Selecting the Scan Type and Mode
• The AP IVA Scan
• The Lateral IVA Scan

13-4 IVA and QM


QDR Reference Manual

Preparing the Patient


Note: Remember to ask the Prepare the patient by asking the pre-examination questions and
patient if they are, or might be, ensuring that they are not wearing anything that will interfere
pregnant. Postpone the exam
with the scan. Refer to “Preparing the Patient” on page 7-2 of
until pregnancy is ruled out.
Chapter 7.

Creating/Retrieving a Patient Biography


If this is a new patient create a new patient biography. If this
patient already has a biography, retrieve it now. Refer to
“Creating/Retrieving a Patient Biography” on page 7-4 of
Chapter 7.

Starting the Examination


To start the examination:
Step Action
1 On the Main Screen click the Perform Exam button.
The Select a Patient window will appear.

KP1068A-QM-07

KP1068A-QM-08

IVA and QM 13-5


QDR Reference Manual

Step Action
2 Highlight the name of the patient and click the OK
button.
The Patient Confirmation window will appear.

Note: The Accession Number 3 Fill in the information required (such as the Accession
must be present in order to use Number and other information) and click on the OK
DICOM.
button.
The Select Scan Type window will appear.

Selecting the Scan Type and Mode


After a patient biography has been created, or retrieved, the Select
Scan Type window displays. The operator can select the IVA
Imaging examination and determine which IVA scans to acquire.

KP1068A-QM-10

13-6 IVA and QM


QDR Reference Manual

If the check mark () is not removed from the Use Default Scan
Mode check box the examination will continue using the scan
defaults. The default is Lateral IVA scan only, and takes 10
seconds.
To run the IVA Lateral only:
Step Action
1 Highlight IVA Imaging in the Scan Type window.
2 Click Next >>.
The system prompts the operator to continuously press
Enable.
3 Continuously press Enable until table motion stops. The
system prompts the operator to align the patient.
4 Align the patient the same way as if performing an AP
Lumbar Spine scan (“Positioning the Patient” on
page 12-4 in Chapter 12).
5 When the patient is properly positioned, press Continue.
The system prompts the operator to continuously press
Enable.
6 Continuously press Enable until C-arm motion stops.
Continue with Confirming Scan Parameters for AP IVA
Scan on page 13-10.
To select IVA Imaging AP/Lateral:
Step Action
1 Click to remove the  mark in the Use Default Scan
Mode box.
The  mark disappears.

IVA and QM 13-7


QDR Reference Manual

Step Action
2 Click the Next >> button.
The following window appears for the AP IVA scan (SE
AP Image):

The operator has two choices for the AP IVA scan:


None - do not perform an AP IVA scan (automatically
defaults to performing a Lateral IVA scan)
SE AP Image - perform an AP IVA Scan

KP1068A-QM-11

3 Make a selection from the above window and click


Next>>.
Selecting None means you do not want to perform an AP
scan (only a lateral scan will be performed).
If None is selected, continue with Confirming Scan
Parameters for AP IVA Scan on page 13-10.
Selecting SE AP Image means you want to perform an
AP scan.
If SE AP Image is selected, the following window will
appear for IVA Lateral scan (SE Lateral Image).

13-8 IVA and QM


QDR Reference Manual

Step Action

KP1068A-QM-12

The operator has two choices for the IVA Lateral scan:
None - do not perform an SE Lateral scan
SE Lateral Image - perform an SE Lateral Scan
4 Make a selection from the above window and click
Next>>.
Selecting None means you do not want to perform a
lateral scan (only an AP scan will be performed). For this
choice the scan parameters window for the AP IVA scan
will appear. Continue with Confirming Scan Parameters
for AP IVA Scan on page 13-10.
Selecting SE Lateral Image here means you want to
perform an AP scan and a lateral scan (both scans will
be performed). For this choice, the system prompts the
operator to continuously press Enable.
5 Continuously press Enable until table motion stops. The
system prompts the operator to align the patient as
closely as possible with the laser.
6 Continue with Positioning the Patient for the AP IVA
Scan on page 13-10.

The AP IVA Scan


The activities involved in performing the AP IVA Scan include:
• Confirming Scan parameters for AP IVA Scan
• Positioning the Patient for the AP IVA Scan
• Performing the AP IVA Scan

IVA and QM 13-9


QDR Reference Manual

• Using the IVA Image Viewer for AP IVA Scan


• Exiting the IVA Image Viewer

Confirming Scan Parameters for AP IVA Scan


The only scan parameter for the AP IVA (f SE AP Image) scan
that can be modified is the scan length. The scan length is 16.1
inches (40.8 cm). Figure 13-3 is an example of the Scan
Parameter/Positioning screen.
Figure 13-3
Scan Parameters/Positioning
for Discovery-A and -SL

KP1068A-QM-14

To verify scan parameters


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan Length
field and type the new length.
3 Stop here and position the patient and the C-arm.

KP0886-stop

Positioning the Patient for the AP IVA Scan


The procedure used to place the patient on the table for an AP
IVA scan is the same as for an AP Lumbar Spine examination and
is described in “Positioning the Patient” on page 12-4 of Chapter
12.

13-10 IVA and QM


QDR Reference Manual

Performing the AP IVA Scan


Once the patient and C-arm are positioned correctly, return to the
computer to start the AP IVA scan. At this time, the Scan
Parameters window is displayed.
To start the AP IVA scan:
Step Action
Note: If the Control Panel X-ray 1 Ask the patient to hold their breath during the ten second
lamp fails to extinguish within 10 scan time (single energy).
seconds after the end of the scan,
press the red Emergency Stop 2 Click the Start Scan button
button on the Control Panel The Scan window displays with the image appearing on the
immediately. Then call Hologic left side. Flashing X-rays On indicator at the top of the
Service before resuming operation.
window continues until the scan stops.

3 Inspect the image.


4 When you see L4 through T4 on the image, click the Stop
Scan button and then advise the patient that they may
breathe normally.
IVA Image Viewer window displays.
While the scan is being acquired the operator has four options:

Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the IVA Image
Viewer window displays.

IVA and QM 13-11


QDR Reference Manual

Stop Scan button Terminates the scan at the point at


which the button is pressed and the
KP1068A-QM-16
IVA Image Viewer window
displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP1068A-QM-17
proceeds to the main window. To
start another scan the C-arm must
be repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP1068A-QM-18 either the window image or the
patient. Refer to the following
instructions.

Repositioning the Scan


If the scan is positioned correctly it will not be necessary to
reposition. Continue with “Using the IVA Image Viewer for AP
IVA Scan” on page 13-46.
If repositioning is required, follow the procedure below.
Two basic activities of repositioning include:
• Repositioning the image
If the image starts too far above or below L5 or is not
centered, move the image up, down, left, or right using the
scroll bars or cursor hand to reposition the scan for the
correct starting point.
• Repositioning the patient
If the spine is not straight, move the patient’s upper torso
either left or right to straighten the spine.
To reposition the image:
Step Action
Note: In order to reposition the 1 Click the Reposition Scan button
scan, the Reposition Scan button The image acquired so far displays with scroll bars on the
must be clicked before the scan is
right and bottom.
completed.
Note: You can also use the scroll 2 Position the cursor over the spine image.
bars to reposition the image. The arrow cursor changes to a hand.

13-12 IVA and QM


QDR Reference Manual

Step Action

Cursor
Scroll
Hand
Bars

KP1068A-1327

3 To reposition the lumbar spine:


• Click and drag the image so that the iliac crest is
within the lower portion of the scan field.
• The center of the lumbar spine should be straight
and centered within the scan field.
4 When the spine is positioned correctly, click the Restart
Scan button.
The Scan Parameters window displays.
Note: If the Control Panel X-ray 5 Ask the patient to hold their breath, and click the Start
lamp fails to extinguish within 10 Scan button to start a new scan at the new position.
seconds after the end of the scan,
The Scan window displays with a flashing X-rays On
press the red Emergency Stop
button on the Control Panel message. The image displays.
immediately. Then call Hologic 6 When you see L4 through T4 on the image, click the Stop
Service before resuming operation.
Scan button, and then tell the patient to breathe normally.
When the scan completes, the IVA Image Viewer window
displays.

While repositioning the scan the following buttons are displayed


in the window:

Button Description
Restart Scan button Proceeds to the Scan Parameters
KP1068A-QM-20
window to commence X-ray scan
from the new position.

IVA and QM 13-13


QDR Reference Manual

Accept Scan button Saves the scan in progress without


KP1068A-QM-21
repositioning the scanner.

Cancel Scan button Cancels the exam and returns to


KP1068A-QM-22
the main window.

Using the IVA Image Viewer for AP IVA Scan


When the scan is complete the program will automatically go to
the IVA Image Viewer (Figure 13-4).
Figure 13-4 Control Bar Image Patient Data
IVA Image Viewer Window for
AP IVA Scan

For a description of the IVA Image Viewer and its controls refer
to “IVA Image Viewer” on page 13-55.
Continue with Exiting the IVA Image Viewer below.

Exiting the IVA Image Viewer


To exit the IVA Image Viewer click Cancel.

If Only AP IVA Scan Selected


If you are performing ONLY an AP IVA scan, the Exit Analysis
window displays. Click Exit to return to the main window.

If Both AP and Lateral IVA Scans Selected


If you are performing BOTH an AP IVA scan and a Lateral IVA
scan, the system prompts you to continuously press Enable.

13-14 IVA and QM


QDR Reference Manual

To perform the Lateral after AP scan:


Step Action
Important: The patient must 1 Continuously press Enable until C-arm motion stops.
remain motionless between scans. The C-arm rotates to the lateral position.
2 Continue with Confirming Scan Parameters for Lateral
IVA Scan below.

The Lateral IVA Scan


The activities involved in performing the Lateral IVA Scan
include:
• Confirming Scan parameters for Lateral IVA Scan
• Performing the Lateral IVA Scan
• Using the IVA Image Viewer for Lateral IVA Scan
• Returning the C-arm to the AP Position

Confirming Scan Parameters for Lateral IVA Scan


The only scan parameter for the Lateral IVA Scan (f SE Lateral
Image) that can be modified is the scan length. The scan length is
16.1 inches (40.8 cm). Figure 13-5 is an example of Scan
Parameters/Positioning for Lateral IVA scan.
Figure 13-5
Scan Parameters/Positioning
for Lateral IVA scan

KP1068A-1325

IVA and QM 13-15


QDR Reference Manual

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan Length
field and type the new length.

Performing the Lateral IVA Scan


With the patient and C-arm positioned correctly, return to the
computer to start the Lateral IVA scan. At this time, the Scan
Parameters window should be displayed.
To start the Lateral IVA scan:
Step Action
1 Ask the patient to hold their breath during the ten second
scan time (single energy).
Note: If the Control Panel X-ray 2 Click the Start Scan button.
lamp fails to extinguish within 10 The Scan window displays with the image appearing on the
seconds after the end of the scan,
left side. Flashing X-rays On indicator at the top of the
press the red Emergency Stop
button on the Control Panel window continues until the scan stops.
immediately. Then call Hologic
Service before resuming operation.

KP1068A-1326

3 Inspect the image.


4 When you see L4 through T4 on the image, click the Stop
Scan button and then tell the patient to breathe normally.
IVA Image Viewer window displays.

13-16 IVA and QM


QDR Reference Manual

While the scan is being acquired the operator has four options:
Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the IVA Image
Viewer window displays.
Stop Scan button Terminates the scan at the point at
which the button is pressed and the
KP1068A-QM-16
IVA Image Viewer window
displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP1068A-QM-17
proceeds to the main window. To
start another scan the C-arm must
be repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP1068A-QM-18 either the window image or the
patient. Refer to the following
instructions.

Repositioning the Scan


If the scan is positioned correctly it will not be necessary to
reposition. Continue with Using the IVA Image Viewer for AP IVA
Scan on page 13-29.
If repositioning is required, follow the procedure below.
To reposition the image:
Step Action
Note: In order to reposition the 1 Click the Reposition Scan button
scan, the Reposition Scan button The image acquired so far displays with a scroll bar on
must be clicked before the scan
the right.
is completed.

IVA and QM 13-17


QDR Reference Manual

Step Action
Note: You can also use the scroll 2 Position the cursor over the spine image.
bars to reposition the image. The arrow cursor changes to a hand.

Cursor
Scroll
Hand
Bars

KP1068A-1327

3 To reposition the lumbar spine, click and drag the image


so that the iliac crest is within the lower portion of the
scan field.
The center of the lumbar spine should be straight and
centered in the scan area.
4 When the spine is positioned correctly, click the Restart
Scan button.
The Scan Parameters window displays.
Note: If the Control Panel X-ray 5 Ask the patient to hold their breath and click the Start
lamp fails to extinguish within Scan button to start a new scan at the new position.
10 seconds after the end of the
The Scan window displays with a flashing X-rays On
scan, press the red Emergency
Stop button on the Control Panel message. The image displays.
immediately. Then call Hologic 6 When you see L4 through T4 on the image, click the Stop
Service before resuming
Scan button, and then tell the patient to breathe normally.
operation.
When the scan completes, the IVA Image Viewer window
displays.

13-18 IVA and QM


QDR Reference Manual

While repositioning the scan the following buttons are displayed


in the window:

Button Description
Restart Scan button Proceeds to the Scan Parameters
KP1068A-QM-20
window to commence X-ray scan
from the new position.
Accept Scan button Saves the scan in progress without
KP1068A-QM-21
repositioning the scanner.

Cancel Scan button Cancels the exam and returns to


KP1068A-QM-22
the main window.

Using the IVA Image Viewer for Lateral IVA Scan


When the scan is complete the program will automatically go to
the IVA Image Viewer (Figure 13-6).
Figure 13-6
IVA Image Viewer Window for
Lateral IVA Scan

For a description of the IVA Image Viewer and its controls refer
to “IVA Image Viewer” on page 13-55. To perform the DE Scan
function refer to “DE Scan” on page 13-59.
Continue with Returning the C-arm to the AP Position below.

IVA and QM 13-19


QDR Reference Manual

Returning the C-arm to the AP Position


To return the C-arm to the AP position:
Step Action
1 From the IVA Image Viewer, click Cancel.
The system prompts you to continuously press Enable.
2 Continuously press Enable until C-arm and table motion
stops.
The C-arm rotates to the AP position and the Exit Analysis
window displays.
3 At the Exit Analysis window, click Exit to return to the
main window.

13-20 IVA and QM


QDR Reference Manual

IVA Imaging on a C, Ci, W or Wi System


The IVA Imaging examination includes all activities that begin
when the patient enters the exam room, and ends when an
acceptable image is obtained. The activities involved in
performing the exam include:
• Preparing the Patient
• Creating/Retrieving a Patient Biography
• Starting the Examination
• Selecting the Scan Type and Mode
• The AP IVA Scan
• The Lateral IVA Scan

Preparing the Patient


Note: Remember to ask the Prepare the patient by asking the pre-examination questions and
patient if they are, or might be, ensuring that they are not wearing anything that will interfere
pregnant. Postpone the exam
with the scan. Refer to the Preparing the Patient on page 7-2 of
until pregnancy is ruled out.
Chapter 7 for detailed instructions.

Creating/Retrieving a Patient Biography


If this is a new patient create a new patient biography. If this
patient already has a biography, retrieve it now. Refer to the
“Creating/Retrieving a Patient Biography” on page 7-4 of
Chapter 7 for detailed instructions.

Starting the Examination


To start the examination:
Step Action
1 On the Main Screen click the Perform Exam button.
The Select a Patient window will appear.

KP1068A-QM-07

KP1068A-QM-08

IVA and QM 13-21


QDR Reference Manual

Step Action
2 Highlight the name of the patient and click the OK
button.
The Patient Confirmation window will appear.

Note: The Accession Number 3 Fill in the information required (such as the Accession
must be present in order to use Number and other information) and click on the OK
DICOM.
button.
The Select Scan Type window appears.

Selecting the Scan Type and Mode


After a patient biography has been created, or retrieved, the Select
Scan Type window displays. The operator can select the IVA
Imaging examination and determine which IVA scans to acquire.

KP1068A-QM-10

13-22 IVA and QM


QDR Reference Manual

If the check mark () is not removed from the Use Default Scan
Mode check box the examination will continue using the scan
defaults. The default is the Lateral IVA scan and takes 10
seconds.
To run the IVA Imaging examination using defaults:
Step Action
1 Highlight IVA Imaging in the Scan Type window
2 Click the Next >> button.
Continue with Confirming Scan Parameters for AP IVA
Scan on page 13-25.

To select IVA Imaging scan modes:


Step Action
1 Click to remove the  mark in the Use Default Scan
Mode box.
The  mark disappears.
2 Click the Next >> button.
The following window appears for the AP IVA scan (SE AP
Image):

KP1068A-QM-11

The operator has two choices for the AP IVA scan:


None - do not perform an AP IVA scan (automatically
defaults to performing a Lateral IVA scan)
SE AP Image - perform an AP IVA Scan

IVA and QM 13-23


QDR Reference Manual

Step Action
3 Make a selection from the above window and click
Next>>.
Selecting None means you do not want to perform an AP
scan (only a lateral scan will be performed).
If None is selected, continue with Confirming Scan
Parameters for AP IVA Scan Scan on page 13-25.
Selecting SE AP Image means you want to perform an AP
scan.
If SE AP Image is selected, the following window will
appear for IVA Lateral scan (SE R/L Lateral Image).

KP1068A-QM-12
The operator has two choices for the IVA Lateral scan:
None - do not perform an SE R/L Lateral scan
SE R/L Lateral Image - perform an SE R/L Lateral
Scan.
4 Make a selection from the above window and click
Next>>.
Selecting None means you do not want to perform a lateral
scan (only an AP scan will be performed).
Selecting SE R/L Lateral Image here means you want to
perform an AP scan and a lateral scan (both scans will be
performed).
For either choice the scan parameters window for the AP
IVA scan will appear. Continue with Confirming Scan
Parameters for AP IVA Scan Scan on page 13-25.

The AP IVA Scan


The activities involved in performing the AP IVA Scan include:
• Confirming Scan parameters for AP IVA Scan

13-24 IVA and QM


QDR Reference Manual

• Positioning the Patient for the AP IVA Scan


• Performing the AP IVA Scan
• Using the IVA Image Viewer for AP IVA Scan
• Exiting the IVA Image Viewer

Confirming Scan Parameters for AP IVA Scan


The only scan parameter for the AP IVA (f SE AP Image) scan
that can be modified is the scan length. The scan length is 16.1
inches (40.8 cm).
Figure 13-7
Scan Parameters/Positioning
for C, Ci, W and Wi

KP1068A-QM-13

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan Length
field and type the new length.
3 Stop here and position the patient and the C-arm.

KP0886-stop

Positioning the Patient for the AP IVA Scan


The procedure used to place the patient on the table for an AP
Lumbar Spine examination and position the C-arm is described in
the Positioning the Patient on page 8-5 of Chapter 8.

IVA and QM 13-25


QDR Reference Manual

Performing the AP IVA Scan


Once the patient and C-arm are positioned correctly, return to the
computer to start the AP IVA scan. At this time, the Scan
Parameters window is displayed.
To start the AP IVA scan:
Step Action
Note: If the Control Panel X-ray 1 Ask the patient to hold their breath during the ten second
lamp fails to extinguish within 10 scan time (single energy).
seconds after the end of the scan,
press the red Emergency Stop 2 Click the Start Scan button.
button on the Control Panel The Scan window displays with the image appearing on the
immediately. Then call Hologic left side. Flashing X-rays On indicator at the top of the
Service before resuming operation.
window continues until the scan stops.

3 Inspect the image.


4 When you see L4 through T4 on the image, click the Stop
Scan button and then advise the patient that they may
breathe normally.
IVA Image Viewer window displays.

While the scan is being acquired the operator has four options:

Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the IVA Image
Viewer window displays.

13-26 IVA and QM


QDR Reference Manual

Stop Scan button Terminates the scan at the point at


which the button is pressed and the
KP1068A-QM-16
IVA Image Viewer window
displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP1068A-QM-17
proceeds to the main window. To
start another scan the C-arm must
be repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP1068A-QM-18 either the window image or the
patient. Refer to the following
instructions.

Repositioning the Scan


If the scan is positioned correctly it will not be necessary to
reposition. If it needs to be repositioned, follow the procedure
below.
Two basic activities of repositioning include:
• Repositioning the Image
If the image starts too far above or below L5, or is not
centered, move the image up, down, left, or right using the
scroll bars or cursor hand to reposition the scan for the
correct starting point.
• Repositioning the Patient
If the spine is not straight, move the patient’s upper torso
either left or right to straighten the spine.
To reposition the image:
Step Action
Note: In order to reposition the 1 Click the Reposition Scan button
scan, the Reposition Scan button The image acquired so far displays with scroll bars on the
must be clicked before the scan is
right and bottom.
completed.

IVA and QM 13-27


QDR Reference Manual

Step Action
Note: You can also use the scroll 2 Position the cursor over the spine image.
bars to reposition the image. The arrow cursor changes to a hand.

Cursor
Hand

Scroll
Bars

KP1068A-1319

3 To reposition the lumbar spine:


• Click and drag the image so that the iliac crest is
within the lower portion of the scan field.
• The center of the lumbar spine should be straight
and centered within the scan field.
4 When the spine is positioned correctly, click the Restart
Scan button.
The Scan Parameters window displays.
Note: If the Control Panel X-ray 5 Ask the patient to hold their breath. Click Start Scan to
lamp fails to extinguish within 10 start a new scan at the new position.
seconds after the end of the scan,
The Scan window displays with a flashing X-rays On
press the red Emergency Stop
button on the Control Panel message. The image displays.
immediately. Then call Hologic 6 When you see L4 through T4 on the image, click the Stop
Service before resuming operation. Scan button and then tell the patient to breathe normally.
When the scan completes, the IVA Image Viewer window
displays.

13-28 IVA and QM


QDR Reference Manual

While repositioning the scan the following buttons are displayed


in the window:
Button Description
Restart Scan button Proceeds to the Scan Parameters
KP1068A-QM-20
window to commence X-ray scan
from the new position.
Accept Scan button Saves the scan in progress without
KP1068A-QM-21
repositioning the scanner.

Cancel Scan button Cancels the exam and returns to


KP1068A-QM-22
the main window.

Using the IVA Image Viewer for AP IVA Scan


When the scan is complete the program will automatically go to
the IVA Image Viewer (Figure 13-8).
Figure 13-8 Control Bar Image Patient Data
IVA Image Viewer Window for
AP IVA Scan

For a description of the IVA Image Viewer and its controls refer
to “IVA Image Viewer” on page 13-55.

Exiting the IVA Image Viewer


To exit the IVA Image Viewer click Cancel.
If both an AP and Lateral IVA scan are to be performed (i.e. SE
R/L Lateral Image was selected as a scan mode), the system

IVA and QM 13-29


QDR Reference Manual

displays the Scan Parameters screen. Continue with The Lateral


IVA Scan below.
If only an AP IVA scan was to be performed (i.e. SE R/L Lateral
Image was NOT selected as a scan mode), the scan is complete
and the system displays the Exit Analysis window. Click Exit to
return to the main window.

The Lateral IVA Scan


The activities involved in performing the Lateral IVA Scan
include:
• Confirming Scan parameters for Lateral IVA Scan
• Positioning the Patient for the Lateral IVA Scan
• Performing the Lateral IVA Scan
• Using the IVA Image Viewer for Lateral IVA Scan
Confirming Scan Parameters for Lateral IVA Scan
The only scan parameter for the Lateral IVA Scan (f SE R/L
Lateral Image) that can be modified is the scan length. The scan
length is 16.1 inches (40.8 cm). Figure 13-9 is an example of
Scan Parameters and patient positioning.
Figure 13-9
Lateral IVA Scan Parameters
and Patient Positioning for C, Ci,
W and Wi

13-30 IVA and QM


QDR Reference Manual

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan Length
field and type the new length.
3 Stop here and position the patient and the C-arm.

KP0886-stop

Positioning the Patient for the Lateral IVA Scan


For an IVA Lateral scan on a Discovery-C or -W system the
patient is placed on their side (laterally) on the exam table. If the
patient is on their right side, facing the rear of the unit, their head
is to the right and their back is toward the front (Figure 13-9). The
spine should be straight and parallel to the long edge of the table.
The remainder of this section provides detailed information for
positioning the patient, the table, and the C-arm for an IVA
Lateral exam. It is important to follow each set of directions
carefully since they are designed to help acquire an ideal scan by
describing:
• How to assist the patient onto the table for proper
alignment.
• How to maximize patient comfort and safety.
• Suggestions for correct positioning techniques.

Placing the Patient on the Table


If the patient is not already on the table, press the Patient On/Off
switch on the Control Panel. When all scanner motion has
stopped, seat the patient on the table between the scan field
indicators on the table pad. Their right shoulder should be
adjacent to the C-arm.

Positioning the Patient


Position the patient as described below:

IVA and QM 13-31


QDR Reference Manual

Figure 13-10
Patient Positioning for an IVA Pillow
Spine Exam

To position the patient:


Step Action
1 Lie the patient on their right side with their head at the right
end of the table.
Place pillow(s) under the head to square the shoulders and
to keep the spine parallel to the table.
2 Check to see that top of the shoulders are within the scan
area as marked on the table top.
Note: It is convenient to have the 3 Check to see that the patient’s back is parallel to the front
patient dressed in an open-backed edge of the scanner and that the midcoronal plane is near
gown so that the vertebral column
the center of the table.
is exposed and can be adjusted
properly. 4 Flex the patient’s hips and knees to a comfortable position.
Flex the knees about 90 degrees. Flex the hips enough so
that the patient does not have a tendency to roll over.

The C-arm and table drive motors provide the table and arm
motion required to scan the X-ray beam along the patient’s spine.
Move the C-arm and turn on the laser using the buttons on the
control panel.
To position the C-arm:
Step Action
1 Move the C-arm to the proximity of the patient’s spine.
2 Turn on the laser.

13-32 IVA and QM


QDR Reference Manual

Note: If interested in 3 Position the cross hairs of the laser 2” below the iliac
determining the presence of a crest and centered over the spine.
calcified abdominal aorta,
center over the spine to ensure 4 Move the laser cross hairs 1” toward the back of the
that the aorta is included. patient.
This is to include both the thoracic and lumbar part of the
spine in the scan field.
5 With the patient and C-arm positioned, the scan type and
mode can be selected on the computer.
6 Continue with Performing the Lateral IVA Scan below.

Performing the Lateral IVA Scan


Once the patient and C-arm are positioned correctly, return to the
computer to start the Lateral IVA scan. At this time, the Scan
Parameters window should be displayed.
To start the Lateral IVA scan:
Step Action
1 Ask the patient to hold their breath during the ten second
scan time (single energy).
Note: If the Control Panel X-ray 2 Click the Start Scan button
lamp fails to extinguish within 10 The Scan window displays with the image appearing on the
seconds after the end of the scan,
left side. Flashing X-rays On indicator at the top of the
press the red Emergency Stop
button on the Control Panel window continues until the scan stops.
immediately. Then call Hologic
Service before resuming operation.

3 Inspect the image.

IVA and QM 13-33


QDR Reference Manual

4 When you see L4 through T4 on the image, click the Stop


Scan button and then tell the patient to breathe normally.
IVA Image Viewer window displays.

While the scan is being acquired the operator has four options:

Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the IVA Image
Viewer window displays.
Stop Scan button Terminates the scan at the point at
which the button is pressed and the
KP1068A-QM-16
IVA Image Viewer window
displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP1068A-QM-17
proceeds to the main window. To
start another scan the C-arm must
be repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP1068A-QM-18 either the window image or the
patient. Refer to the following
instructions.

Repositioning the Scan


If the scan is positioned correctly it will not be necessary to
reposition. If it needs to be repositioned, follow the procedure
below.
Two basic activities of repositioning include:
• Repositioning the image
If the image starts too far above or below L5 or is not
centered, move the image up, down, left, or right using the
scroll bars or cursor hand to reposition the scan for the
correct starting point.
• Repositioning the patient
If the spine is not straight, move the patient’s upper torso
either left or right to straighten the spine.

13-34 IVA and QM


QDR Reference Manual

To reposition the image:


Step Action
Note: In order to reposition the 1 Click the Reposition Scan button
scan, the Reposition Scan button The image acquired so far displays with scroll bars on the
must be clicked before the scan
right and bottom.
is completed.
Note: You can also use the scroll 2 Position the cursor over the spine image.
bars to reposition the image. The arrow cursor changes to a hand.

Cursor
Hand

Scroll
Bars

KP1068A-1334

3 To reposition the lumbar spine, click and drag the image


so that the iliac crest is within the lower portion of the
scan field.
The center of the lumbar spine should be straight and
centered in the scan area.
4 When the spine is positioned correctly, click the Restart
Scan button.
The Scan Parameters window displays.

IVA and QM 13-35


QDR Reference Manual

Note: If the Control Panel X-ray 5 Ask the patient to hold their breath and click the Start
lamp fails to extinguish within Scan button to start a new scan at the new position.
10 seconds after the end of the
The Scan window displays with a flashing X-rays On
scan, press the red Emergency
Stop button on the Control Panel message. The image displays.
immediately. Then call Hologic 6 When you see L4 through T4 on the image, click the Stop
Service before resuming
Scan button, and then tell the patient to breathe normally.
operation.
When the scan completes, the Exit Exam window
displays.

While repositioning the scan the following buttons are displayed


in the window:

Button Description
Restart Scan button Proceeds to the Scan Parameters
KP1068A-QM-20
window to commence X-ray scan
from the new position.
Accept Scan button Saves the scan in progress without
KP1068A-QM-21
repositioning the scanner.

Cancel Scan button Cancels the exam and returns to


KP1068A-QM-22
the main window.

Using the IVA Image Viewer for Lateral IVA Scan


When the scan is complete the program will automatically go to
the IVA Image Viewer (Figure 13-11).
Figure 13-11 Control Bar Image Patient Data
IVA Image Viewer Window for
Lateral IVA Scan

13-36 IVA and QM


QDR Reference Manual

For a description of the IVA Image Viewer and its controls refer
to “IVA Image Viewer” on page 55. To perform the DE Scan
function refer to “DE Scan” on page 59.
Click Cancel to display the Exit Analysis window.
At the Exit Analysis window, click Exit to return to the main
window.

IVA and QM 13-37


QDR Reference Manual

IVA with BMD


Note: IVA with BMD is The IVA Imaging examination includes all activities that begin
available on A or SL systems for when the patient enters the exam room, and ends when an
AP/Lateral pair.
acceptable image is obtained. The activities involved in
performing the exam include:

The IVA Imaging examination includes all activities that begin


when the patient enters the exam room, and ends when an
acceptable image is obtained. The activities involved in
performing the exam include:
• Preparing the Patient
• Creating/Retrieving a Patient Biography
• Starting the Examination
• Selecting the Scan Type and Mode
• The AP BMD Scan
• The Lateral BMD Scan
• The Lateral IVA Scan

Preparing the Patient


Note: Remember to ask the Prepare the patient by asking the pre-examination questions and
patient if they are, or might be, ensuring that they are not wearing anything that will interfere
pregnant. Postpone the exam
with the scan. Refer to Preparing the Patient on page 7-2 of
until pregnancy is ruled out.
Chapter 7.

Creating/Retrieving a Patient Biography


If this is a new patient create a new patient biography. If this
patient already has a biography, retrieve it now. Refer to Creating/
Retrieving a Patient Biography on page 7-4 of Chapter 7.

13-38 IVA and QM


QDR Reference Manual

Starting the Examination


To start the examination:
Step Action
1 On the Main Screen click the Perform Exam button.
The Select a Patient window will appear.

KP1068A-QM-07

KP1068A-QM-08

2 Highlight the name of the patient and click the OK button.


The Patient Confirmation window will appear.

Note: The Accession Number must 3 Fill in the information required (such as the Accession
be present in order to use DICOM. Number and other information) and click on the OK
button.
The Select Scan Type window will appear.

IVA and QM 13-39


QDR Reference Manual

Selecting the Scan Type and Mode


After a patient biography has been created, or retrieved, the Select
Scan Type window displays. The operator can select the IVA with
BMD Imaging examination and determine which IVA scans to
acquire.

KP1068A-1336

If the check mark () is not removed from the Use Default Scan
Mode then the examination will continue using the scan defaults.
The default is the AP BMD, followed by Lateral BMD, followed
by Lateral IVA.
To run the IVA with BMD examination using defaults:
Step Action
1 Highlight IVA Imaging in the Scan Type window.
2 Click Next >>.
The system prompts the operator to continuously press
Enable.
3 Continuously press Enable until table motion stops. The
system prompts the operator to align the patient.
4 Align the patient the same way as if performing an AP
Lumbar Spine scan.
5 When the patient is properly positioned, press Continue.
The system prompts the operator to continuously press
Enable.

13-40 IVA and QM


QDR Reference Manual

To select IVA Imaging scan modes:


Step Action
1 Click to remove the  mark in the Use Default Scan
Mode box.
The  mark disappears.
2 Click Next >>.
The following window appears for the AP IVA scan (SE
AP Spine):

KP1068A-1337

There are two choices:


None - do not perform an AP IVA scan (automatically
defaults to performing a Lateral IVA scan)
SE AP Image - perform an AP IVA Scan
3 Make a selection for the AP IVA scan and click Next >>.
For either choice the following screen will be displayed.

KP1068A-1338

Make a selection for the AP part of the AP Lateral pair


(screen above) and click Next >>.
The following screen will be displayed.

IVA and QM 13-41


QDR Reference Manual

Step Action
4 Make a selection for the AP part of the AP Lateral pair
(screen above) and click Next >>.
The following screen will be displayed.

KP1068A-1338

This screen is the scan modes for the lateral part of the
AP Lateral pair, the description of the scan modes is
found in Choosing the Scan Mode on page 12-3 of
Chapter 12.
5 Make a selection for the lateral part of the AP Lateral pair
(above window) and click Next >>. One of the following
will occur:
If None was selected for the AP IVA scan:
• The system prompts the operator to continuously
press Enable.
• Continuously press Enable until table motion
stops. The system prompts the operator to align
the patient.
• Align the patient the same way as if performing
an AP Lumbar Spine scan (described in
Positioning the Patient on page 12-4 of Chapter
12.)
The scan parameters for AP BMD appears,
continue with “Confirming Scan Parameters for
AP BMD Scan” on page 47.
If SE AP Image was selected for the AP IVA scan, the
following window appears for the Lateral IVA scan (SE
Lateral Scan).

13-42 IVA and QM


QDR Reference Manual

Step Action

KP1068A-1339

The operator has two choices for the Lateral IVA scan:
None - do not perform a Lateral IVA scan.
Selecting None means you do not want to perform a
lateral scan (only the BMD scan pair and an AP IVA
scan will be performed).
SE Lateral Image - perform a Lateral IVA Scan.
Selecting SE Lateral Image means you want to per-
form the BMD scan pair, and both the AP IVA scan and
lateral IVA scan.
6 Make a selection from the above window and click
Next>>.
For either choice, the system prompts the operator to
continuously press Enable.
7 Continuously press Enable until table motion stops. The
system prompts the operator to align the patient as
closely as possible with the laser.
8 Align the patient the same way as if performing an AP
Lumbar Spine scan (described in Positioning the Patient
on page 13-38 of Chapter 12.)
9 When the patient is properly positioned, press Continue.
Continue with Confirming Scan Parameters for AP IVA
Scan below.

The AP IVA Scan


The activities involved in performing the AP IVA Scan include:
• Confirming Scan parameters for AP IVA Scan
• Performing the AP IVA Scan
• Using the IVA Image Viewer for AP IVA Scan

IVA and QM 13-43


QDR Reference Manual

Confirming Scan Parameters for AP IVA Scan


The only scan parameter for the AP IVA (f SE AP Image) scan
that can be modified is the scan length. The scan length is 16.1
inches (40.8 cm). Figure 13-12 is an example of Scan Parameters
and Patient Positioning for an AP IVA Scan.
Figure 13-12
Scan Parameters and Patient
Positioning for AP IVA Scan

KP1068A-1341

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan Length
field and type the new length.

Performing the AP IVA Scan


Once the patient and C-arm are positioned correctly, return to the
computer to start the AP IVA (f SE AP Image) scan. At this time,
the Scan Parameters window is displayed.
To start the AP IVA scan:
Step Action
Note: If the Control Panel X-ray 1 Ask the patient to hold their breath during the ten second
lamp fails to extinguish within 10 scan time (single energy).
seconds after the end of the scan,
press the red Emergency Stop 2 Click the Start Scan button.
button on the Control Panel The Scan window displays with the image appearing on the
immediately. Then call Hologic left side. Flashing X-rays On indicator at the top of the
Service before resuming operation.
window continues until the scan stops.

13-44 IVA and QM


QDR Reference Manual

3 Inspect the image.


4 When you see L4 through T4 on the image, click the Stop
Scan button and then tell the patient that they may breathe
normally.
IVA Image Viewer window displays.

While the scan is being acquired the operator has four options:
Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the IVA Image
Viewer window displays.
Stop Scan button Terminates the scan at the point at
which the button is pressed and the
KP1068A-QM-16
IVA Image Viewer window
displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP1068A-QM-17
proceeds to the main window. To
start another scan the C-arm must
be repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP1068A-QM-18 either the window image or the
patient. Refer to the following
instructions.

IVA and QM 13-45


QDR Reference Manual

Repositioning
If the scan is positioned correctly it will not be necessary to
reposition.
If the scan is not positioned correctly the only option is to
reposition the patient. If the spine is not straight, move the
patient’s upper torso either left or right to straighten the spine.

Using the IVA Image Viewer for AP IVA Scan


When the scan is complete the program will automatically go to
the IVA Image Viewer (Figure 13-13)
Figure 13-13 Control Bar Image Patient Data
IVA Image Viewer Window for
AP IVA Scan

For a description of the IVA Image Viewer and its controls refer
to IVA Image Viewer on page 13-55.
Press Cancel from the IVA viewer and the scan parameters screen
for AP BMD (f Lumbar Spine) scan appears. Continue with
Confirming Scan Parameters for AP BMD Scan below.

The AP BMD Scan


The activities involved in performing the AP BMD Scan include:
• Confirming Scan parameters for AP BMD Scan
• Performing the AP BMD Scan
• Analyzing the AP BMD Scan
• Exiting the AP BMD Analysis

13-46 IVA and QM


QDR Reference Manual

Confirming Scan Parameters for AP BMD Scan


The only scan parameter that can be modified is the scan length.
The scan length is 8 inches (20.4 cm). Figure 13-14 is an example
of Scan Parameters and Patient Positioning for an AP BMD Scan.
Figure 13-14
Scan Parameters and Patient
Positioning for an AP BMD Scan

KP1068A-1344

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan
Length field and type the new length.
3 Click Start Scan to start the AP Lumbar Spine Scan

Performing the AP BMD Scan


The procedure used to perform the AP Lumbar Spine
examination is described in Performing the AP Scan on page 12-7
of Chapter 12.

Analyzing the AP BMD Scan


The procedure to analyze the AP Lumbar Spine is described in
Analyzing the AP Scan on page 12-9 of Chapter 12.

Exiting the AP BMD Analysis


To exit the AP BMD Analyze window:
Step Action
1 From the AP BMD Analyze window click on the Close
button.
2 The system prompts the operator to continuously press
Enable to reposition the C-arm and table.

IVA and QM 13-47


QDR Reference Manual

Important: Remind the patient 3 Continuously press Enable until C-arm and table motion
not to move, or touch the sides of stops. The scan parameter window for the Lateral BMD
the table, during C-arm rotation.
scan appears.
4 Continue with Confirming Scan Parameters for Lateral
BMD Scan below.

The Lateral BMD Scan


The activities involved in performing the Lateral BMD Scan
include:
• Confirming Scan parameters for Lateral BMD Scan
• Performing the Lateral BMD Scan

Confirming Scan Parameters for Lateral BMD Scan


The only scan parameter for the Lateral BMD (f Lateral) that can
be modified is the scan length. The scan length is 6 inches (15.3
cm). Figure 13-15 is an example of Scan Parameters and Patient
Positioning for a Lateral BMD Scan
Figure 13-15
Scan Parameters and Patient
Positioning for Lateral BMD
Scan

KP1068A-1345

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan
Length field and type the new length.

13-48 IVA and QM


QDR Reference Manual

Performing the Lateral BMD Scan


Once the C-arm is positioned correctly, return to the computer to
start the Lateral BMD (f Lateral) scan. At this time, the Scan
Parameters window is displayed.
To start the Lateral BMD scan:
Step Action
Note: If the Control Panel X-ray 1 Click the Start Scan button
lamp fails to extinguish within 10 The Scan window displays with the image appearing on the
seconds after the end of the scan,
left side. Flashing X-rays On indicator at the top of the
press the red Emergency Stop
button on the Control Panel window continues until the scan stops.
immediately. Then call Hologic
Service before resuming operation.

2 Inspect the image.


3 When you see where ribs are attached to T12 on the image,
click the Stop Scan button.
4 Depending upon whether or not a Lateral IVA scan will be
performed, one of the following screens displays:
• Continuously press Enable...
If no Lateral IVA scan is to be performed, (i.e.,
None was selected from the Lateral IVA scan mode
window), the system prompts the operator to
continuously press Enable to reposition the C-arm
and table.
Continue with Exiting the Lateral BMD Scan on
page 13-50.

IVA and QM 13-49


QDR Reference Manual

• Scan Parameters
If a Lateral IVA scan is to be performed, the system
displays the Scan Parameters window for that scan.
Continue with Confirming Scan Parameters for
Lateral IVA Scan on page 13-51.

While the scan is being acquired the operator has four options:
Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the IVA Image
Viewer window displays.
Stop Scan button Terminates the scan at the point at
which the button is pressed and the
KP1068A-QM-16
IVA Image Viewer window
displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP1068A-QM-17
proceeds to the main window. To
start another scan the C-arm must
be repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP1068A-QM-18 either the window image or the
patient. Refer to the following
instructions.

Exiting the Lateral BMD Scan


To exit the Lateral BMD scan:
Step Action
1 At the Continuously press Enable... screen, continuously
press Enable until C-arm and table motion stops.
The C-arm returns to the AP position and the Exit Exam
window appears.
2 At the Exit Exam window, click Analyze Scan to analyze
the lateral scan, or Exit Analysis to exit to the main
window.
The procedure to analyze the Lateral Spine scan is
described in Analyzing the Lateral Scan on page 12-28 of
Chapter 12.

13-50 IVA and QM


QDR Reference Manual

The Lateral IVA Scan


The activities involved in performing the Lateral IVA Scan
include:
• Confirming Scan parameters for Lateral IVA Scan
• Performing the Lateral IVA Scan
• Using the IVA Image Viewer for the Lateral IVA Scan

Confirming Scan Parameters for Lateral IVA Scan


After the Lateral BMD scan is complete, the scan parameter
screen for the Lateral IVA (f SE Lateral Image) will appear.
The only scan parameter for the Lateral IVA that can be modified
is the scan length. The scan length is 16.1 inches (40.8 cm).
Figure 13-16
Scan Parameters and Patient
Positioning for a Lateral IVA
Scan

KP1068A-1347

To verify scan parameters:


Step Action
1 Verify the patient name and scan type in the upper left
corner.
2 To change scan length, place the cursor in the Scan
Length field and type the new length.

IVA and QM 13-51


QDR Reference Manual

Performing the Lateral IVA Scan


To start the Lateral IVA scan:
Step Action
1 Ask the patient to hold their breath during the ten second
scan time (single energy).
Note: If the Control Panel X-ray 2 Click the Start Scan button
lamp fails to extinguish within The Scan window displays with the image appearing on
10 seconds after the end of the
the left side. Flashing X-rays On indicator at the top of
scan, press the red Emergency
Stop button on the Control Panel the window continues until the scan stops.
immediately. Then call Hologic
Service before resuming
operation.

3 Inspect the image.


4 When you see L4 through T4 on the image, click the Stop
Scan button and then tell the patient to breathe normally.
The IVA Image Viewer window displays.
While the scan is being acquired the operator has four options:

Option Description
(let the scan complete) The scan completes the length
designated on the Scan Parameters
window and then the IVA Image
Viewer window displays.

13-52 IVA and QM


QDR Reference Manual

Stop Scan button Terminates the scan at the point at


which the button is pressed and the
KP1068A-QM-16
IVA Image Viewer window
displays.
Abort Scan button Exits the scan immediately. An
exit window displays then
KP1068A-QM-17
proceeds to the main window. To
start another scan the C-arm must
be repositioned.
Reposition Scan button Stops the scan in progress and
enables the operator to reposition
KP1068A-QM-18 either the window image or the
patient. Refer to the following
instructions.

Using the IVA Image Viewer for the Lateral IVA Scan
When the scan is complete the program will automatically go to
the IVA Image Viewer (Figure 13-17).
Figure 13-17
IVA Image Viewer Window for
Lateral IVA Scan

For a description of the IVA Image Viewer and its controls refer
to IVA Image Viewer on page 13-55. To perform the DE Scan
function refer to DE Scan on page 13-59.
Continue with Returning the C-arm to the AP Position below.

IVA and QM 13-53


QDR Reference Manual

Returning the C-arm to the AP Position


To return the C-arm to the AP position:
Step Action
1 From the IVA Image Viewer, click Cancel.
The system prompts you to continuously press Enable.
2 Continuously press Enable until C-arm and table motion
stops.
The C-arm rotates to the AP position and the Exit
Analysis window displays.
3 At the Exit Analysis window, click Analyze Another
Scan to analyze the lateral BMD scan or Exit to return to
the main window.
The procedure to analyze the Lateral Spine scan is
described in the Analyzing the Lateral Scan on page 12-
28 of Chapter 12.

13-54 IVA and QM


QDR Reference Manual

IVA Image Viewer


The IVA Image Viewer allows the operator to view the IVA
image. There are controls on the Viewer to change viewing
modes, zoom, adjust the brightness and contrast of the IVA
image, place markers, print a report or image and perform a DE
Lateral Scan after running a SE Lateral Scan in an exam.

The IVA Image Viewer

Note: The IVA Image Viewer is The IVA Image Viewer window is divided into three areas:
used after completing an IVA
scan, or by selecting the Analyze • the Image
all Scans tab and then selecting
• the Patient Data
an IVA scan.
• the Control Bar

The Image
The scanned image is shown in the middle of the IVA window.
Figure 13-18 shows the scanned image for an AP IVA scan and
Figure 13-19 shows the scanned image for a Lateral IVA scan.
Figure 13-18 Control Bar Image Patient Data
IVA Image Viewer Window for
AP IVA Scan

IVA and QM 13-55


QDR Reference Manual

Figure 13-19 Control Bar Image Patient Data


IVA Image Window for
Lateral IVA Scan

The Patient Data


The patient data, for the scan, includes patient name, ID, and
other biographical information and is shown on the right side of
the window.

The Control Bar


The control bar occupies the left side of the window. It remains
the same size, even if the operator changes the window size.
The following describes the controls and their uses:
Control Description
DE Scan button This button is used differently for
the AP and Lateral IVA scan:
KP1068A-1350
• AP IVA Scan. This button
is grayed out and not used.
• Lateral IVA Scan. This
button is active. See page
13-59 for a description of
DE Scan.
Markers button Click Markers to show and edit the
QM data markers. How to use the
KP1068A-1351 markers is described in Using
Markers on pages 13-61 to 13-68.

13-56 IVA and QM


QDR Reference Manual

W-L Control The Window-Level control allows


you to change contrast and
brightness-type parameters of the
image. You can select each value
and type in a new value. You can
also dynamically change the W-L
values by dragging the black dot
with the mouse until you reach the
desired image.
For finer adjustment control, use
W-L Adjust (see below).

KP1068A-1352

Revert button Disregards all changes and returns


all viewing parameters to their
KP1068A-1353 initial values. This includes:
• W-L values
• Flip
• Invert
• Zoom: goes to Fit to
Window
Invert button Inverts the gray scale value of each
pixel creating a negative of the
KP1068A-1354 image.
Flip button Flips the image about the center
vertical axis. When Markers are
KP1068A-1355 initialized you cannot flip the
image.
Zoom Control buttons Enlarges or reduces the image
about its center point.
KP1068A-1356

Print Image button Print the image on the selected


printer (see page 13-69).
KP1068A-1357

Print Report button Print the report on the selected


printer (see page 13-72).
KP1068A-1358

IVA and QM 13-57


QDR Reference Manual

Cancel button The Cancel button exits the


analysis window and returns to the
KP1068A-1359 main window.
Close button If you perform a rescan, the
Cancel button changes to the
KP1068A-1360
Close button. Close exits the
Important: Once you have made analysis window and returns to the
any edits or re-scanned, you main window, saving any changes
cannot cancel to discard any
changes. Any changes you made
made to the scan.
are saved. If DE is selected, it returns to scan.

Mouse Pointer Controls


Placing the mouse pointer in the image area and right-clicking
displays the image control selection menu. An explanation of
each selection follows:
Zoom and W-L Selection Click one of the choices to
Menu (upper section) activate.
See descriptions below.

KP1068A-1361

Drag Zoom Drag the magnifying glass over the


area of the image that you want to
enlarge. When you release the left
mouse button, the captured area
expands to fit the window.

KP1068A-1362

Drag Pan Click and drag the image to place


it anywhere in the window.

KP1068A-1363

W-L Adjust This W-L Adjust allows finer


control over the W-L control
described earlier. Move it in any
direction to change the W and L
values to the left, but at a finer rate.

KP1068A-1364

13-58 IVA and QM


QDR Reference Manual

Zoom Control (lower Select any of the zoom control


section) settings to enlarge or reduce the
image size in the window.
If the image becomes too large for
the width or height of the window,
scroll bars appear at the bottom or
right of the window.

KP1068A-1365

If you are analyzing existing IVA scans, make sure that the spine
is on the left before continuing.

DE Scan
If a more detailed scan is required of part of the spine or a
vertebrae, the operator can define a region on the IVA scan to be
re-scanned using Dual Energy. This region of the patient is re-
scanned at a higher resolution
To perform this rescan the patient must remain on the table and in
the same position as the Lateral IVA scan. If the patient has
moved, or been removed from the table, another Lateral IVA scan
must be acquired and the rescan taken immediately after the
Lateral IVA scan with the patient in the same position.
To rescan:
Step Action
1 Click DE Scan in the upper left-hand corner of the
window.
KP1068A-1366
Two horizontal green lines appear in the upper and lower
Note: If the image is magnified areas of the image.
larger than the window, you may
not see the green lines. Select
“Fit To Window” if you do not
see the lines.
KP1068A-1368

2 Place the mouse pointer over the upper line and drag it to
the top of the region that you want to rescan.
3 Place the mouse pointer over the lower line and drag it to
the bottom of the region that you want to rescan.

IVA and QM 13-59


QDR Reference Manual

Upper Rescan Line


Area to be rescanned
Lower Rescan Line
KP1068A-1367

4 Click Close.
Because the DE Scan button was pushed at the time the
Close button was selected, the system returns to the Start
Scan dialog box to start the Dual Energy scan.
Note: It is assumed that the 54 Start the new scan. The system scans the patient and
patient and the C-arm have not returns to the IVA analysis window when done.
been moved since the last scan.
The patient may breathe normally during the Dual Energy
scan.

13-60 IVA and QM


QDR Reference Manual

Using Markers
Quantitative Morphometry (QM) allows you to define the shape
of one or more vertebrae by placing markers on the posterior,
anterior and midpoints of the vertebrae.
QM Markers can be placed on the AP and Lateral IVA scans.
Hologic does not recommend placing markers on the AP IVA
scan, however if this is done see “QM Results Block for AP IVA
Scans” on page 13-68 concerning information on determining
tags for the markers.

The Object of QM Markers


The object is to place a marker in the superior and inferior of the
anterior, posterior and midpoints (Figure 13-20) in order to mark
the size and shape of the vertebrae.
Figure 13-20 Superior Posterior Superior Midpoint Superior Anterior
Marker Placement

Inferior Posterior Inferior Midpoint Inferior Anterior

The proper placement of these six markers can be found in “The


Appendix to Chapter 20: Point Placement in Vertebral
Morphometric X-ray Absoptiometry” by Jacqueline A. Rea in
‘The Evaluation of Osteoporosis: Dual Energy Absorptiometry
and Ultrasound in Clinical Practice, Second Edition’, pages 456-
457.

Using QM Markers
QM Markers are placed on the image in the IVA Image Viewer
(Figure 13-20). To start placing markers the operator clicks on the
Markers bottom, then places the cursor on the scan image and
right clicks the mouse. A menu will appear that will allow the
operator to place and label the markers (procedures to place, label
and change the markers are included in this section). As markers
are placed on the scan image, a results block reports information
on the markers.

IVA and QM 13-61


QDR Reference Manual

Figure 13-21 Markers button


The IVA Image Viewer
Window

Results Block

Flipping the Image


Note: The image cannot be The IVA scan image should initially display the vertebrae on the
flipped after Markers are added. left. Before adding QM Markers make sure that the spine is on
the left of the image, clicking the Flip button if necessary.

13-62 IVA and QM


QDR Reference Manual

Adding QM Markers
To perform Quantitative Morphometry, you must insert QM
markers.
To insert the markers:
Step Action
1 On Control Bar in the IVA Image Viewer Analysis
Window, click on the Markers button (Figure 13-21).
KP1068A-1351
2 Right click on the image and select Add Marker.

KP1068A-1371

The markers appear on the image in the area of the mouse


pointer.

KP1068A-1372

The QM data for L4 also appears at the bottom of the QM


area of the window.

KP1068A-1373

Note: The first set of markers The software limits the marker labels to L4-L1, and T12-
that you add is always labeled T4. The software also inhibits you from changing the
L4. If you rename this set of
label of a marker to one that already exists.
markers to T12, for example, the
software assumes that you want
to label from T12 upward.
Therefore, the next set of
markers that you create is
labeled T11, and so forth.

IVA and QM 13-63


QDR Reference Manual

Selecting Markers
To select markers for a vertebrae:
Step Action
1 Place the mouse pointer in the middle of the six markers
and click,
-or-

KP1068A-1374

select the marker data in the QM area of the window.

KP1068A-1375

Note: You can select multiple The selected markers appear yellow (red if image is
markers in both the QM area or inverted). The unselected markers appear red (cyan if the
on the image using combinations image is inverted).
of the left mouse button and the
control and shift keys.

Moving Markers
To move one or more sets of markers:
Step Action
1 Select one or more sets of markers.
2 Click in the area described by the six markers of any of
the selected set(s) of markers.
3 Drag the set(s) of markers to their desired position.

Editing Markers
You can edit only one set of markers at a time.
To edit a set of markers:
Step Action
1 Click anywhere within the set of markers to be edited.
The set of markers become yellow (active).

13-64 IVA and QM


QDR Reference Manual

2 Click on one of the six markers and drag it to its desired


position.

KP1068A-1376

3 Repeat Step 2 for the remaining markers.

Labeling Markers Sets


The marker set that you create may not have the correct label
associated with it.
To label a set of markers:
Step Action
1 Select the set of markers.
2 Right-click within the markers of the selected set.
3 Select Label Marker from the menu.

KP1068A-1377

4 From the Label Vertebra dialog box, select the desired


label from the list and click OK.
The label associated with that set of markers changes and
the vertebra moves to the proper position in the QM data
area.

KP1068A-1378

IVA and QM 13-65


QDR Reference Manual

Deleting Marker Sets


To delete a set, or sets, of markers:
Step Action
1 Select one or more sets of markers.
2 Right click in any of the selected sets of markers.
Note: You can select multiple 3 Select Delete Marker from the menu.
markers by using combinations
of the left mouse button and the
control and shift keys.

KP1068A-1379

4 When the “Delete selected Markers?” dialog box appears,


click OK.
Deleting markers on the image also deletes them in the
QM area of the window.

QM Result Block
The result box for the Lateral and the AP IVA scans must be
interpreted differently because of the positions of the spine in the
scan.

QM Results Block for Lateral IVA Scans


As QM markers are placed on the vertebrae a report showing the
heights (Hght) of the vertebrae appears in the lower right hand
section of the IVA Image Viewer (Figure 13-22).

13-66 IVA and QM


QDR Reference Manual

Figure 13-22
The QM Result Block

KP1068A-1380

The following are descriptions of the labels in the above results


block.

Label Description
Post The height of the vertebrae between the superior and
Hght inferior posterior markers in millimeters.
(mm)
Mid The height of the vertebrae between the superior and
Hght inferior midpoint markers in millimeters.
(mm)
Ant The height of the vertebrae between the superior and
Hght inferior anterior markers in millimeters.
(mm)
Wedge The distance in millimeters when the Ant Hght is
divided by the Post Hght.
Mid The distance in millimeters when the Mid Hght is
Wedge divided by the Post Hght.

The numbers following Std Dev are the standard deviation of the
number above. For example, the Post Hght (mm) has a standard
deviation of +/- 1 mm; the Wedge has a standard deviation of +/-
0.05 mm.

IVA and QM 13-67


QDR Reference Manual

QM Results Block for AP IVA Scans


Note: Hologic does not The Result Block for the AP IVA scan will appear the same as
recommend placing QM markers the above Lateral IVA scan but the labels are used differently
on the AP IVA scan.
because of the position of the spine. The Mid Hght, Wedge and
Mid Wedge are the same but the Post Hght must be interpreted
as the Right Side of the vertebrae and the Ant Hght must be
interpreted as the Left Side of the vertebrae according to the
following:

Lateral IVA Scan AP IVA Scan


Post Right
Hght Side
(mm)
(mm)
Ant Left
Hght Side
(mm) (mm)
The Left Side and Right Side labels will not appear in the AP IVA
scan Result Block report or be printed on any reports.

13-68 IVA and QM


QDR Reference Manual

Generating and Printing Images and Reports


On the IVA Image Viewer for both the IVA Imaging and IVA with
BMD examinations there are buttons that will print a report or an
image.
IVA reports can also be printed from the Reports button on the
Main screen (see below).

Printing IVA Images from the Reports Button


To print IVA Reports from the Reports Button:
Step Action
1 From the Main screen click on the Reports button.
The Select Patient window appears.

KP1068A-1392

KP1068A-1381

2 Highlight the name of a patient whose report is to be


printed then click on the Next >> button.
The Select Scan window appears.

KP1068A-1382

IVA and QM 13-69


QDR Reference Manual

3 Highlight the scan to be printed then click on the Next>>


button.
The Print window appears.

KP1068A-1383

4 Click on the DICOM/IVA Report button


The DICOM Report window appears.

5 To send the report (via DICOM) to the Physician’s


Viewer, click Send.
If an accession number was assigned, the report is sent. If
no accession number was assigned, the Accession No.
Entry window appears (see below).
To preview the report, click Preview,
The Accession No. Entry window appears.

13-70 IVA and QM


QDR Reference Manual

KP1068A-1385

6 Click on the OK button.


The DICOM Report Preview window appears.

7 Click on the Print button.


The IVA Report is printed.

See below for sample printouts.

IVA and QM 13-71


QDR Reference Manual

Sample Printouts
Sample Image

13-72 IVA and QM


QDR Reference Manual

Sample Report

IVA and QM 13-73


QDR Reference Manual

Interpreting the IVA Image


IVA images are to be interpreted by a physician or properly
licensed practitioner. IVA images are not intended for general
radiological diagnosis, but are intended to be used for evaluation
of vertebral deformities.
The anatomy of the spine is shown in Figure 13-23, including
vertebral level labels. IVA images typically include levels T4 to
L4. Following the classification scheme of Genant (see reference
below), Figure 13-24 shows examples of a typical vertebral body
shape and examples of deformed vertebral shapes.

Figure 13-23 C1
The Human Spine C1
7 cervical C2
C4
vertebrae C5
C6
C7
T1
T2
T3
T4
T5
12 thoracic T6
vertebrae T7
T8
T9
T10
The human spine usually consists T11
of 7 cervical, 12 thoracic, and 5 T12
lumbar vertebrae.
L1

L2
5 lumbar L3
vertebrae
L4

L5

Ossacrum

KP1068A-1389

13-74 IVA and QM


QDR Reference Manual

Figure 13-24
Vertebrae Deformities
Typical Vertebral Body

Typical vertebral body shown


with severely deformed Severe Wedge Deformity
vertebrae.
Adapted from Genant, H.K., C.Y.
Wu, et al. (1993). “Vertebral
fracture assessment using a
semiquantitative technique.” J.
Bone Miner Res 8(9): 1137-48. Severe Biconcavity Deformity
KP1068A-1390

For detailed information regarding vertebral deformity evaluation


and the utility of vertebral deformity assessment in clinical
osteoporosis evaluation, see the references and organizations
listed below.
For Information on Spine Anatomy, Vertebral Deformity
Evaluation, and Vertebral Deformity Classification:
• The Evaluation of Osteoporosis: Dual Energy
Absorptiometry and Ultrasound in Clinical Practice,
Second Edition; Blake, G. M., Walgner, H. W., Fogelman,
I., © Martin Duritz Ltd 1999, “The Appendix to Chapter
20: Point Placement in Vertebral Morphometric X-ray
Absoptiometry” Jacqueline A. Rea: 456-457.
• Genant, H. K., C. Y. Wu, et al. (1993). “Vertebral fracture
assessment using a semiquantitative technique.” J Bone
Miner Res 8(9): 1137-48.
• Genant HK, Jergas M, van Kuijk C (Eds.): Vertebral
Fracture in Osteoporosis. San Francisco, CA, University
of California Osteoporosis Research Group, 1995
• Merrill’s Atlas of Radiographic Positions and Radiologic
Procedures; P. W. Ballinger and Ed Frank, Eds.
(Mosby, New York) 1999
• Diagnosis of Bone and Joint Disorders; R. Resnick, Ed.
(W.B. Saunders Co., Philadelphia) 1995

IVA and QM 13-75


QDR Reference Manual

• Radiological Society of North America:


Phone: 30-571-2670
Website: http://www.rsna.org
For Information on Clinical Patient Evaluation and the
Relationship between Vertebral Deformities and Subsequent
Fractures:
• National Osteoporosis Foundation:
Phone: 202-223-2226
Website: http://www.nof.org
• International Osteoporosis Foundation:
Phone: +33 472 91 41 77 (France)
Website: http://www.osteofound.org
• International Society for Clinical Densitometry:
Phone: 202-828-6056
Website: http://www.iscd.org
• American Association of Clinical Endocrinologists:
Phone:
Website: http://www.aace.com
• Black DM, Arden NK, Palermo L, Pearson J, Cummings
SR (1999). “Prevalent vertebral deformities predict hip
fractures and new vertebral deformities but not wrist
fractures.” J Bone Miner Res 1999 14:890-902.
• Davis, J. W., J. S. Grove, et al. (1999). “Spatial
relationships between prevalent and incident spine
fractures.” Bone 24(3): 261-4.
• Melton, L. J., 3rd, E. J. Atkinson, et al. (1999). “Vertebral
fractures predict subsequent fractures.” Osteoporosis Int
10(3): 214-21.
• Nevitt, M. C., P. D. Ross, et al. (1999). “Association” of
prevalent vertebral fractures, bone density, and
alendronate treatment with incident vertebral fractures:
effect of number and spinal location of fractures. The
Fracture Intervention Trial Research Group.” Bone 25(5):
613-9.

13-76 IVA and QM


Chapter 14
Compare and Follow-up
This chapter provides information on using the Compare function of
the QDR system. Compare is an analysis of the current scan using a
previous scan of the same patient. The previous scan is used as a
reference to duplicate the region of interest.

Content
Subject Page
Introducing Compare 14-2
Restoring the Patient’s Baseline Scan 14-3
Evaluating the Patient’s Baseline Scan 14-4
Performing the Follow-up Scan 14-7
Analyzing the Follow-up Scan 14-8
Generating a Rate of Change Report 14-12
Rate of Change Report for Unlike Scan 14-13
Modes
Rate of Change Report for Spine Scans with 14-14
Different Regions

14-1
QDR Reference Manual

Introducing Compare
Compare allows comparison of the current scan to a baseline scan
to determine the BMD rate of change over a period of time. The
baseline scan is generally the first scan acquired for the patient.
All follow-up scans are compared to the baseline scan.
When following bone mineral density (BMD) in a patient over a
period of time, it is important to compare the same region of
interest each time. The system helps by displaying the current
scan next to the baseline scan of the same patient. It uses the bone
map of the baseline scan as an overlay on the current scan so you
can visually match the region of interest in the current scan
image.
It is important to duplicate the size, shape, and location of the
region of interest as closely as possible.

Steps
Comparing a scan to the baseline scan involves the following
steps:
1. Locating and Restoring the patient’s baseline scan.
2. Evaluating the patient’s baseline scan.
3. Performing the follow-up scan.
4. Analyzing the follow-up scan using the Compare function.
5. Generating a rate of change report.
The steps are the same for all scan types. This chapter uses the
spine scan as an example, but the Compare function procedures
described here can be applied to all scans.
If necessary, changes can be made to correct the regions.
However, if changes are made, the ROI template will disappear
because it is no longer a direct compare. Examples of when this
may be necessary include a child that has grown or a patient that
has lost vertebral body height. It is still recommended to use
Compare on the follow-up.

14-2 Compare and Follow-up


QDR Reference Manual

Restoring the Patient’s Baseline Scan


Before acquiring a new scan, it is necessary to restore the
patient’s baseline scan. The baseline scan is the first scan
conducted on the patient. Each subsequent or follow-up scan is
compared to this baseline scan to determine changes in bone
mineral density.

Restoring a Baseline Scan


Suggestion: At the start of the The patient’s baseline scan is either on the computer’s hard disk
workday, it may be desirable to drive or has been archived onto other media.
locate and restore the baseline
scans for all follow-up patients Use the method described below if the baseline scan is still
scheduled for exams during the located on the computer’s hard drive, that is, it has not yet been
day. As patients arrive, time can archived and deleted from the computer’s hard drive.
be spent updating their records
and taking their history instead If the baseline scan is not on the computer’s hard disk, locate and
of searching for scans. restore the archived scan (see Chapter 16, Archiving, Locating
and Restoring Scans, for detailed instructions).

KP1035A-0313

To restore a baseline scan from the computer’s hard drive:


Step Action
1 Click the Analyze Scan button in the main window.
The system displays the Select a scan to analyze window
with three tabs: Unanalyzed Scans, Analyzed Scans, and
All Scans.
Note: If the baseline scan is not 2 Click the Analyzed Scans tab.
on the computer’s hard disk, The system displays a list of scans sorted alphabetically
locate and restore the archived
by patient last name.
scan (see “Restoring Scans” on
page 16-16 of Chapter 16, for 3 Locate the patient’s name on the list using either or both
detailed instructions). of the following methods:
Note: Case (upper case, lower • Start typing the patient’s name (last name,
case) is ignored when typing a
patient’s name to search for it.
comma, space, first name, space, middle initial)
until the scans for that patient appear.
• Use the scroll bars to bring the patient’s scans into
view.

Compare and Follow-up 14-3


QDR Reference Manual

Step Action
4 Click the scan with the oldest date (this is the baseline
scan).
The scan becomes highlighted.
Note: Scans can be compared 5 Click the Next >> button on the Select a scan to analyze
even if they are not the same window.
scan mode.
The Analyze Setup window displays.
Note: Do not change the 6 Click the Next >> button on the Analyze Setup window.
analysis method displayed in the to display the analysis window with the baseline scan
Analyze Setup window.
image.

Evaluating the Patient’s Baseline Scan


Before acquiring a new scan, an evaluation must be made of the
patient’s baseline scan. There are two objectives when evaluating
the patient’s baseline scan:

1. Ensure that patient positioning in the baseline scan is dupli-


cated as closely as possible in the follow-up scan. Changes,
even minor, can affect the precision of the BMD measure-
ments.
Note: If the baseline scan 2. Ensure that the baseline scan is analyzed correctly. If the
analysis is incorrect, all analysis is not correct, the baseline scan should be re-ana-
subsequent scans should be
lyzed and re-archived.
reanalyzed and archived.

Evaluating Baseline Scan Patient Positioning


Evaluate the baseline scan image to detect any abnormalities in
patient positioning during the baseline scan. For example:
• Before performing a spine scan, check to see if the spine
is straight and centered on the baseline image.
• Before performing a hip scan, check the baseline image. Is
the Lesser Trochanter clearly visible? How much of it is
visible? Is the leg straight? Evaluating the baseline hip
scan will help in gauging the amount of leg rotation
needed for the new scan.
When evaluating the baseline scan, review the operator or
physician comments, if any. When all of the information about
the baseline scan is known, determine if the baseline positioning
is acceptable.

Reviewing Operator or Physician Comments


The operator who performed the baseline scan may have entered
comments pertaining to positioning the patient. Review the scan

14-4 Compare and Follow-up


QDR Reference Manual

comments on the baseline scan to determine if there is any


information that may be helpful in positioning the patient, or that
may affect the comparison.
For example, the comments may state that the previous operator
was unable to elevate the patient’s legs due to patient discomfort.
You may need to duplicate this positioning in the new scan.

Baseline Positioning Acceptable


If the baseline positioning is acceptable, proceed to Evaluating
Baseline Scan Analysis on page 14-6.

Baseline Positioning Unacceptable


If the baseline positioning is unacceptable, one of the following
three options must be chosen. Consult with the physician to
determine which option to choose for this patient.
1. Take one new scan duplicating the incorrect baseline posi-
tioning.
2. Scan the patient twice. In the first scan, duplicate the incor-
rect baseline positioning. In the second scan, position the
patient correctly, and archive this scan as the new baseline
for future follow-up exams.
3. Take one new scan with correct positioning, and archive this
scan as the new baseline for future follow-up exams.

Scanning Once to Duplicate Incorrect Baseline Positioning


Duplicating the incorrect positioning of the baseline scan obtains
the best possible BMD rate of change. The disadvantage of this
option is that the incorrect baseline scan is not corrected for
subsequent follow-up exams. Consult with the physician before
choosing this option.

Scanning Twice
In the first scan, repeat the incorrect positioning of the baseline
scan. This allows a good BMD rate of change to be obtained for
this exam. In the second scan, position the patient correctly and
archive this scan as the new baseline for future follow-up exams.
The disadvantage of this option is that the patient must be
scanned twice. Consult with the physician before choosing this
option.

Scanning Once with Correct Positioning


Position the patient correctly and archive this scan as the new
baseline scan. This corrects the baseline scan for future follow-up
exams. The disadvantage of this option is that BMD rate of

Compare and Follow-up 14-5


QDR Reference Manual

change is compromised for this exam. Consult with the physician


before choosing this option.

Evaluating Baseline Scan Analysis


Note: If the baseline scan is Evaluate the baseline scan to determine if it must be re-analyzed.
reanalyzed, it must be Re-analysis of the baseline scan may be necessary for a number
rearchived as well. Also
of reasons:
reanalyze and rearchive all
subsequent scans. • The patient’s anatomy has changed since the baseline
scan. For example, a vertebra has collapsed, or there is
more space between vertebra in the new scan.
• The baseline analysis was performed incorrectly, or could
be improved upon.

To reanalyze the current scan:


Step Action
Note: If you need help in 1 Click the appropriate Analysis Step Button and modify.
reanalyzing the scan, refer to the
chapter on performing and 2 Repeat for each subsequent Analysis Step Button,
analyzing the scan for the ending with the Results button.
specific anatomy.
3 Click the Close button.

14-6 Compare and Follow-up


QDR Reference Manual

Performing the Follow-up Scan


In addition to the normal process of performing and analyzing a
scan, there are several important considerations, described below,
that could impact the follow-up scan report.

Updating the Patient’s Biography Record


In addition to the standard questions asked when taking the
patient’s history, ask the patient if they have had any injuries or
surgery in the exam area since the last exam.

Note: If the baseline scan is Some examples are:


reanalyzed and rearchived, all
subsequent scans should be • For a spine exam, verify that the patient has not had any
reanalyzed and rearchived. compression fractures since the baseline scan. If they
have, re-analyze the baseline scan and exclude that area.
• For a hip exam, if the patient has had surgery on their
right hip since the baseline scan, that hip can no longer
used as the baseline for comparison. Scan the left hip and
use that as the baseline from now on.

Any new or pertinent information may have to be included in the


patient record. Refer to Chapter 6 for instructions on retrieving
and updating the patient’s Biography file.

Performing the Follow-Up Scan


After retrieving the patient’s baseline scan and evaluating it,
perform the follow-up exam. To perform the appropriate scan
refer to the following:

Scan to be performed Refer to Chapter


Spine 8
Hip 9
Forearm 10
Whole Body 11
AP/Lateral 12

Compare and Follow-up 14-7


QDR Reference Manual

Analyzing the Follow-up Scan


After performing the follow-up scan on the patient, the Exit
Exam window displays.

Note: The steps for Compare


Analysis are the same for all
scan types. This section uses a
spine scan as an example, but
the procedures described here
can be applied to all scans.

KP0886C-1302

To Perform the Compare Analysis


Step Action
1 Click the Analyze Scan button.
The Analyze Setup window displays.

KP0886C-1303
Compare to Previously Analyzed
Scan radio button

14-8 Compare and Follow-up


QDR Reference Manual

Step Action
Note: Scans can be compared 2 Click the Compare to Previously Analyzed Scan radio
even if they are not the same button.
scan mode.
A list of scans for that patient will appear in the window
with the oldest scan highlighted. The oldest scan should
be the baseline scan, if it is not then click and highlight
the scan which is the baseline scan.

Baseline scan KP0886C-1304

3 Click the Next >> button.


The analysis window will appear with the baseline scan
on the right and the follow-up scan on the left.

KP0886C-1305
Follow-Up Scan Baseline Scan

4 Click the Global ROI button in the top left corner of the
window, if not already selected.
A template consisting of the ROI border and bone map
outline displays on the follow-up scan as active (dashed
and solid yellow lines).

Compare and Follow-up 14-9


QDR Reference Manual

Step Action
Note: The ROI may not appear if Global ROI with template
the ROI on the compare scan is
too large, or if the point size and
line size do not match in the
scans.

KP0886C-1306

5 The Whole Mode button is the default and should be


active (click if not already selected).
6 Position the cursor arrow anywhere within the ROI box
of the follow-up scan (to the left).
The crossed arrows cursor displays.
Note: For finer control in 7 Click and drag the template until it matches the anatomy
positioning the template, the on the baseline scan (to the right).
keyboard arrow keys may be
used to move the template one 8 Click the Bone Map button.
line at a time. The bone map is displayed on the follow-up scan (to the
left).
9 Click on the Vertebral Lines button.
Three horizontal lines display in the image area of the
follow-up scan (to the left).
10 Compare the Vertebral lines on the follow-up scan (to the
left) to those on the baseline scan (to the right). Adjust the
Vertebral lines in the follow-up scan to match those in the
baseline scan.
11 Click the Results button.

14-10 Compare and Follow-up


QDR Reference Manual

Step Action
Note: During the compare Results button
analysis it may be necessary to
make minor adjustments to
match the baseline analysis. If
so, return to the appropriate step
and adjust accordingly.

KP1068_002-1307
Current scan results

12 Click the Close button in the lower left corner of the


Analysis window.

Compare and Follow-up 14-11


QDR Reference Manual

Generating a Rate of Change Report


Note: Refer to “Rate of Change After a Compare analysis of the follow-up scan, a Rate of
Report” on page 15-15 of Change report can be generated that provides the change
Chapter 15 for more information
between the baseline analysis and the follow-up analysis.
on the Rate of Change report.
After performing the Compare analysis, the Exit Analysis
window displays. To generate and print a Rate of Change Report,
click the Report button in the Exit Analysis window.

To generate a Rate of Change report:


Step Action
1 Click the Report button in the Exit Analysis window.
The Print window displays.
Filing check box

KP0886C-1308
Rate of Change check box

2 Click the Rate of Change check box.


A check appears in the check box.
3 Select the number of copies to be printed in the number
box to the right of the Rate of Change check box.
4 Edit the physician’s comments, if desired.

14-12 Compare and Follow-up


QDR Reference Manual

Step Action
5 Click the Print button to print the Rate of Change Report
along with the default Filing Report (check box marked
“Filing”).
Note: Refer to Chapter 15 for
detailed information about
reports.

KP0886C-1309
T-Score
Rate of Change

Rate of Change Report for Unlike Scan Modes


Rate of Change reports for Unlike Scan Modes are generated the
same way as for any Compare Analysis (see Generating a Rate of
Change Report on page 14-12). Areas that are dissimilar on the
reports will be marked by an “*” (see Figure 14-1).

Compare and Follow-up 14-13


QDR Reference Manual

Figure 14-1
Unlike Scan Mode
Rate of Change Report

KP1035A-1314

Rate of Change Report for Spine Scans with Different Regions


When a Rate of Change report for Spine Scans with Different
Regions is generated, a message appears stating that only regions
that have been analyzed will be displayed (Figure 14-2).
Figure 14-2
Different Region Message

KP1068_002-1415

14-14 Compare and Follow-up


QDR Reference Manual

Click OK to generate a rate of change report for spine scans with


different regions.
In the following example (Figure 14-3) the most recent scan has
regions L1, L3-L4 while the prior scan had L1-L4. The Rate of
Change report shows the results for only L1, L3-L4 as indicated
by the graph title. The scan image will reflect the analyzed
regions of the most recent scan.
Figure 14-3
Rate of Change Report
for a Spine Scan
with Different Regions

Note: The DXA Results


Summary table displays BMD
values associated with the
regions that have been analyzed
(circled).

Note: To avoid confusion, a


disclaimer is printed on the KP1068_002-1416
report (circled).

Compare and Follow-up 14-15


QDR Reference Manual

14-16 Compare and Follow-up


Chapter 15
Reports
This chapter provides information about the types of reports available
in the QDR system. It also includes instructions for printing and
configuring these reports.

Content
Subject Page
Introducing Reports 15-2
Printing Reports 15-6
Previewing Reports 15-11
Report Examples 15-13
DICOM Reports 15-18
Alternate Navigation 15-39

15-1
QDR Reference Manual

Introducing Reports
This section provides an overview of the types of reports
available and a description of the information contained on them.

Report Types
Note: DICOM is standard on The QDR system includes four basic types of reports, which can
Discovery-A, -SL, -W and -C be printed, previewed (viewed on the monitor screen), and
models, and available as an
configured to meet specific needs. An additional type of report is
option on other models.
available if the DICOM Option has been installed (refer to
DICOM Reports on page 15-18).

Filing Report
The Filing Report is used primarily for filing examination
information and includes:
• Detailed patient and scan information
• Area, BMC, BMD, and T & Z-score results for each
region
• Scan image with ROI and bone map outline
• BMD vs. Age reference curve graph
• k and D0 calculations
• Physician’s comments
• QDR system technical information

Interpreting Report
The Interpreting Report is primarily for the interpreting physician
and includes:
• Detailed patient and scan information
• Area, BMC, BMD and T & Z-score results for each region
• Scan image with ROI and bone map outline
• BMD vs. Age reference curve graph
• Physician’s comments

Referring Report
The Referring Report is primarily for the referring physician and
includes:
• Summarized patient and scan information

15-2 Reports
QDR Reference Manual

• BMD and T-score results for the default region only (as
configured during system setup) for each scan image
• Scan images (up to three/page) with ROI and bone map
outlines
• BMD vs. Age reference curve graph for each scan image

Rate of Change Report


The Rate of Change Report is primarily for tracking changes in
results over time and includes:
• Detailed patient and scan information
• Visit date, patient age, BMD, and T-score results for each
visit
• Changes in results vs. Baseline and previous scans
• Scan image with ROI and bone map outline for current
scan
• BMD vs. Age reference curve graph for each baseline and
subsequent scans
• Physician’s comments

Report Information Blocks


Reports contain blocks of information which vary slightly
depending upon the type of report you choose. Refer to
Figure 15-1 and the following text for an explanation of report
blocks.

Reports 15-3
QDR Reference Manual

Figure 15-1
Report Blocks Header

Patient
Information
Scan
Information

Image

DXA
Results
Reference
Graph

Physician
Comment
KP0886C-1401

Header
The Header block displays at the top of each report. It includes
the name, address, telephone, and fax number of the institution or
facility generating the report.
Note: Refer to To configure the The header can be turned off if your facility prefers to print
Header Block on page 18-21 for reports on letterhead.
information on turning the
header off.

Patient Information
The Patient Information block displays information from the
patient biographical record, and any updated information added
on the day the scan was performed. This includes the patient’s
age, height, and weight.

Scan Information
The Scan Information block contains scan record information
about the analyzed scan(s) including scan date, scan type, date of
analysis, operator, model, and serial number of the scanner.

15-4 Reports
QDR Reference Manual

Image
Note: Reports printed within the The Image block displays the analyzed image of the scan with
United States include the the region of interest, bone map outline, and global dimension of
disclaimer statement “Image not
the ROI. If the scan has not been analyzed, the image displays
for diagnostic use.”
without the ROI and bone map outline.

DXA Results
The DXA Results summary displayed in this block include the
BMC and BMD of a particular scanned region, and the Z-score
and T-score. The Z-score is a measure of the difference between
the patient’s BMD and that of age, sex, and ethnicity-matched
peers. The T-score is a measure of the difference between the
patient’s BMD and young normals of the same sex and ethnicity.

Reference Graph
The Reference Graph allows comparison of patient data to young
normal adults, and to an age-matched reference population.
The standard deviations appear shaded in different colors. If the
patient’s data point is within the colored portion of the graph, it is
shown as a black cross inside a circle. If the patient’s data point is
outside the colored graph, it is shown as a red cross inside a
circle. The source of the reference curve and scores is shown
below the graph.

Rate of Change Graph


The Rate of Change Graph (not shown on Figure 15-1) identifies
the rate of change of the BMD for the patient at the chosen
anatomical site of the selected scans. Results are reported for the
individual regions, and for the total of the anatomical site. The
Rate of Change graph is used for Follow-Up exams and Compare
procedures and is shown on the Rate of Change Report.

Physician’s Comment
The Physician’s comment block displays any comments
authorized by the physician. Physician’s comments can be
selected when generating the report. They can be selected from a
predefined list, or a new comment can be entered.

Reports 15-5
QDR Reference Manual

Printing Reports
APEX allows the operator to access and print reports a number of
different ways. This section includes instructions for printing a
report after completing an analysis and printing a report using the
Report button on the main window.

Printing a Report After Analysis


Note: To preview a report prior After completing the analysis of a scan, one of the options on the
to printing, refer to page 15-11. Exit Analysis window is the Report button. Clicking this button
will bring up the Print window from which reports may be
printed singly or as multiple copies.
Refer to Generating and Printing Reports on page 7-32 for
information on generating a report immediately following the
analysis.

Printing a Report Using the Report Button


Use the Report button in the main window to retrieve a specific
scan and print a report. Clicking the Report button and selecting
the patient’s name displays a list of analyzed scans from which to
choose.

KP1035A-0315

15-6 Reports
QDR Reference Manual

Note: Refer to pages 16-12 and The analyzed scan must reside on the computer’s hard drive. If
16-16 for more information on the scan has been deleted, use the Locate and Restore functions
locating and restoring an
to load it back onto the computer’s hard drive.
archived scan.

To print a report using the Report button:


Step Action
Note: Alternate ways to access 1 Click the Report button in the main window.
and print reports are described in The system displays the Select Patient window with a list
the Alternate Navigation section
of patient records, sorted alphabetically by patient last
of this chapter.
name.
2 Click the patient’s name for whom report(s) are to be
printed, then click the Next >> button.
The system displays the Select Scans window with the
patient’s name and ID at the top.
Note: Multiple scans may be 3 Click the scan(s) for which report(s) are to be printed, then
selected by holding down the Ctrl click the Next >> button.
key while clicking each scan with
The system displays the Print window with the patient’s
the mouse.
name and ID at the top.
Note: One or more types of 4 Choose the type(s) of reports to be printed by clicking in
reports may be selected. the box next to the report type.
A check mark indicates that the specified report will print.
Note: The number of copies may 5 For each report type selected, choose the number of copies
also be entered directly into the
to print by clicking the adjacent number box button to
number box.
increase the number of copies, or the button to decrease
the number of copies.
6 To have comments appear on the printed report, click the
Edit physician comments... button (see directions
below).

Reports 15-7
QDR Reference Manual

Step Action
7 Click the Preview... button to display a preview of the
report before printing it (see the Previewing Reports
section of this chapter).
8 Click the Configure... button to change the settings of the
report (see “Report Tab” on page 18-10).
9 Click either:
• the Print... button to print the report;
• the << Back button to return to the previous
window; or
• the Cancel button to cancel the operation without
printing.

Printing the Patient Questionnaire


The Questionnaire button appears if the Mobility option is
installed.
Printing the Patient Questionnaire
Step Action
1 From the Print screen, click on the Print Questionnaire
button.
The Print screen for the Questionnaire appears.

2 The following operations can be performed from the


Questionnaire:
• Print button. Clicking this button prints the
questionnaire to the desired printer and returns the
user to the previous screen.
• Cancel button. Clicking this button returns the
user to the previous screen without saving
answers.

15-8 Reports
QDR Reference Manual

Editing the Physician Comment


The Edit physician comment window allows the operator to
choose from a list of predefined comments or enter a new one.
The predefined comments are created by the physician. Any one
of these comments can be added to the report being printed. If
necessary, a new comment can be entered.
The comment selected or entered is saved with the scan record
and displayed in the printed report(s).

KP0886C-1404

To add or change a physician comment:


Step Action
1 In the Print window, click the Edit physician comment...
button.
The system displays the Edit physician’s comment window
with information and results pertaining to the selected
scan(s).
Note: New comments are not 2 • To select from the list of predefined comments,
added to the list of predefined click the button in the Physician’s Comment text
comments.
box and click on the comment you want to include
• To enter a new comment, click in the Physician’s
Comment text box and type the new comment.
3 Click either:
• the Update button to update the scan with the new
comment; or
• the Close button to exit the Edit physician’s
comment window without changing the comment.

Reports 15-9
QDR Reference Manual

Print Dialog Box


When the Print... button is clicked, either from the Print window
or the Preview Report window, the system displays a standard
print dialog box for your printer, similar to the one shown below.

KP0886C-1405

Reports can only be printed in portrait mode and on either letter


size (8.5¨ x 11¨) or A4 size paper.
To select the Print dialog settings:
Step Action
Note: It is usually not necessary to 1 If it is necessary to change printer properties, click the
change printer properties or Properties button to display and select the properties for
printer type. This information is
the printer. The properties include settings such as Page
pre-configured.
and Graphics that apply to the selected printer. If printer
properties are changed by mistake, click the Restore
Defaults button to return to the default settings.
Note: If a radio button is grayed 2 To specify which pages to print:
(dimmer than the others), it • Click the All radio button to print all the pages
indicates that option is not
available. • Click the Pages radio button and enter the first and
last page of the range to print
• Click the Selection radio button to print only the
highlighted (selected) information from the report
3 Click the OK button to begin printing the report(s) on the
selected printer.

15-10 Reports
QDR Reference Manual

Previewing Reports
The system provides the opportunity to preview each report
before it is actually printed. Clicking the Preview... button in the
Print window displays the report preview window.

Function buttons for Preview

KP0886C-1406

To display a preview of the report:


Step Action
1 In the Print window, click the Preview... button.
The system displays the report preview window showing
the first page of the report.
2 Refer to the table below for a description of each button in
the Report window.

Button Function
Print... Displays the standard print dialog window to print
the selected report.
Next Page Displays the next page of the report, if more than
one page exists. If there is only one page to the
report, this button is inactive (gray).
Prev Page Displays the previous page of the report, if more
than one page exists. If there is only one page to
the report, this button is inactive (gray).
Two Pages Displays two pages in the window. If there is only
one page to print, this button is inactive (gray).

Reports 15-11
QDR Reference Manual

Button Function
Note: Clicking within the area of Zoom In Zooms in on the report. If already at maximum
the report causes it to zoom, this button is inactive (gray).
successively zoom in, centered on
the cursor location, until at Zoom Out Zooms out on the report. If already at minimum
maximum zoom. Clicking once zoom, this button is inactive (gray).
more resets the display to
minimum zoom.
Close Closes the print preview window without printing
the report, returning to the Print window.
In addition to using the Next Page and Prev Page buttons, you
can use the scroll bar to the right of the report to scroll through the
pages.

15-12 Reports
QDR Reference Manual

Report Examples
This section provides examples of reports for various scan and
report types.

Filing Report
The Filing Report, shown for a spine exam in the example below,
contains detailed information about the patient record and scan
record. The report displays one scan on each page.
Figure 15-2
Filing Report Example:
Spine Exam

KP0886C-1407

Reports 15-13
QDR Reference Manual

Interpreting Report
The Interpreting Report, shown for a forearm exam in the
example below, contains detailed information about the patient
record and scan record, and is used by the interpreting physician.
The report displays one scan on each page.
Figure 15-3
Interpreting Report Example:
Forearm Exam

KP0886C-1408

15-14 Reports
QDR Reference Manual

Referring Report
The Referring Report, shown in the example below, contains
summarized information about the patient and scan record. This
report is generally used by the referring physician. The report
displays at least three scans per page.
Figure 15-4
Referring Report Example

KP0886C-1409

Rate of Change Report


The Rate of Change Report, shown in the example below,
contains the results of selected scans over a period of time. The
baseline scan is the oldest scan among the selected scans. The
report includes a column for change versus baseline scan, system
technical information and change versus the previous scan. The
most recent scan image is displayed.

Reports 15-15
QDR Reference Manual

Figure 15-5
Rate of Change Report
Example: Left Hip

KP0886C-1410

15-16 Reports
QDR Reference Manual

Patient Questionnaire
The Patient Questionnaire is shown in the example below.
Figure 15-6
Example of the
Patient Questionnaire

Reports 15-17
QDR Reference Manual

DICOM Reports

Your APEX software treats a DICOM report as an additional


report type that complements its standard Filing, Interpreting
Physician, and Referring Physician report formats.
From the DICOM Report screen you can print the History
Questionnaire (Hx) report.
The DICOM Report screen is shown in Figure 15-7:
Figure 15-7
DICOM Report Screen

To Create and Send a DICOM Report:


Step Action
Note: Alternate ways to access 1 Click the Report button on the main window.
and print reports are described The Select Patient window displays.
in the Alternate Navigation
section of this chapter.
2 Click on the patient’s name for whom you want to create
and send a DICOM report, then click on the Next>>
button.
The Select Scan(s) window displays.
Note: Multiple scans may be 3 Click on the scans that you want to report using DICOM,
selected by holding down the then click on the Next>> button.
Ctrl key while clicking each scan
The Print window displays.
with the mouse.

15-18 Reports
QDR Reference Manual

Step Action
Note: If the DICOM / IVA 4 Click on the DICOM / IVA report button that appears
report button is not visible, the below the Edit physician’s comment… button on the
DICOM option may not have
right side of the Print window.
been properly installed in your
system. The DICOM Report window displays.

5 Select the scan(s) that you want to create DICOM reports


for by clicking on each scan. Use the Select All button to
select all of the scans that appear in the list of scans.
Note: Standardized BMD and 6 Select the DICOM Report Type.
Extended Spine single analysis
reports are available for AP • Single Analysis
Lumbar Spine scans only.
• Rate of Change Report
If you want remove private patient information from the
report, check the De-identify protected health
information check box.
If you want to include the Ancillary Report and/or
patient’s Questionnaire in the study, check the respective
check box.
Note: To preview a report prior 7 Click on the Send button to transmit the report(s).
to transmission, you must select
a single scan then click on the The DICOM Report(s) is (are) placed in the Send Queue
Preview button. for transmission over your network.
Note: You may be prompted to
enter Accession number and
other user defined entries. (see
“Sending a DICOM Report” on
page 15-29).

Choosing the Report Type


DICOM supports the Single Analysis and Rate of Change reports.
These reports are similar to printed reports, but generally contain
more information. DICOM also supports Standardized BMD
(sBMD) and Extended Spine.
Single Analysis reports provide all the quantitative results from a
single scan and analysis—including the scan image and reference
population curves. The Single Analysis Report includes all the
information you need to interpret a scan.
Rate of Change Reports simplifies comparing visit-to-visit rates
of change. Significant changes are noted on the report to facilitate
clinical management. The Rate of Change report also includes the
Single Analysis report for the most recent scan and analysis.

Reports 15-19
QDR Reference Manual

To Choose a Single Analysis Report Type:


Step Action
1 On the DICOM Report screen, click on the radio button
Single analysis for the single analysis report.

Note: Standardized BMD and 2 If you want a standardized BMD and/or extended spine
Extended Spine single analysis report, check the Standardized BMD and/or Extended
reports are available for AP
Spine check boxes.
Lumbar Spine scans only.
4 Click on Preview to view the report or Send to transmit
the report.
Note: The accession number and 5 When Send is clicked and the selected scan does not have
any additional user defined an accession number associated with it already, the
entries for a scan are listed in
Accession Number Entry dialog box will appear. Enter an
the scan list on the DICOM
Reports screen. accession number and click OK.

KP1068A-2017

Note: You may be prompted to enter additional user


defined entries. If so, additional entry dialog boxes will
appear. Enter the requested entry and click OK for each
dialog box. (User defined entries can be left blank.) See
“HL7 Tab” on page 18-34 for details on enabling/
disabling the display of these fields and defining their
label names.

15-20 Reports
QDR Reference Manual

Choose a Rate of Change Report Type:


Step Action
1 On the DICOM Report screen, click on the radio button
Rate of change for the most recent value report.

2 Click on Preview to view the report or Send to transmit


the report.
Note: The accession number and 3 When Send is clicked and the selected scan does not have
any additional user defined an accession number associated with it already, the
entries for a scan are listed in
Accession Number Entry dialog box will appear. Enter an
the scan list on the DICOM
Reports screen. accession number and click OK.

KP1068A-2019

Note: You may be prompted to enter additional user


defined entries. If so, additional entry dialog boxes will
appear. Enter the requested entry and click OK for each
dialog box. (User defined entries can be left blank.) Refer
to “HL7 Tab” on page 18-34 for details on enabling/
disabling the display of these fields and defining their
label names.

Reports 15-21
QDR Reference Manual

Step Action
Note: The Skip button will allow 4 The Select Rate of Change Scans dialog box will appear
you to skip the selection process with a list of matching scans.
for a particular scan. The
resulting DICOM report will be
similar to a Single Analysis
report.

KP1068A-2020

5 Select the scans to be included in the DICOM report.


Hold down the <Ctrl> key + click to select each scan.
6 Click on OK.

Viewing a Scan’s Details


The Scan Details button allows you to view a scan’s detail such
as the patient biography information, scan date, scan ID, etc. It
also allows you to enter/edit certain patient biography fields.
To View a Scan’s Details:
Step Action
1 Click on the scan whose details you want to view from
the scan list on the DICOM Reports screen.
2 Click on Scan Details.
The Scan Property dialog box will be displayed. There
KP1068A-2021
are 2 tabs for viewing the details: Details and
Identification.
3 Click on the Details tab to view the patient biography
information.

15-22 Reports
QDR Reference Manual

Step Action
Note: Accession Number label
and field appear on the dialog
box only if the DICOM option is
installed on your system.

Note: HL7 Field 1, HL7 Field 2,


and HL7 Field 3 are examples of
user definable label names.
These labels and their entry
fields appear on the dialog box
when enabled. See “HL7 Tab”
on page 18-34 for details on
enabling/disabling the
displaying of these fields and
defining their label names.

KP1068A-2022

4 On the Details tab, edits are allowed on the following


fields:
• Accession Number – 16 characters maximum
• HL7 Field 1 – 64 characters maximum
• HL7 Field 2 – 64 characters maximum
• HL7 Field 3 – 64 characters maximum
• Operator – 5 characters maximum
• Height – 5 characters maximum
• Weight – 5 characters maximum
• Scan Comment – 100 characters maximum
Note: The scan-related fields are 5 Click on the Identification tab to view the scan
not editable. information.

Reports 15-23
QDR Reference Manual

Step Action

KP1068A-2023

6 Click on OK to save any edits made on the Details tab


and close the Scan Property dialog box.
7 To close the Scan Property dialog box without saving any
changes, click on Cancel.

Entering the Accession Number and User Definable Entries


The Accession number is uniquely associated with a patient visit,
and is typically generated by a Radiological Information System
(RIS) or a Hospital Information System (HIS).
Three user definable entries may be available for adding
information to the DICOM file’s Comments field. These entries
are not required and there is no restriction on their content. See
“HL7 Tab” on page 18-34 for details on enabling/disabling the
displaying of these fields and defining their label names.
To Enter the Accession Number:
Step Action
1 Select a scan from the scan list on the DICOM Reports
screen.
2 Click on Save As or Send.

15-24 Reports
QDR Reference Manual

Step Action
Note: The accession number for 3 If the selected scan does not have an accession number
a scan is listed in the scan list on associated with it already, the Accession Number Entry
the DICOM Reports screen.
dialog box will appear.

Note: When sending multiple


scans to be included in a single
report, all scans must have the
same Accession Number.

KP1068A-2024

Note: The maximum length for 4 Enter the accession number, and then press Enter or click
the Accession Number is 16 OK.
characters.
For example, if the Accession Number is 7888123, type:
7888123 <Enter>
The scan now has the accession number and the selected
function will continue.
5 Click Cancel if the accession number is unknown or if it
is to be entered at a later time. However, the selected
function (Save As, Send) will not work unless the scan
has an accession number.
Notes: Any additional user 6 You may be prompted to enter additional user defined
defined entries for a scan are entries. If so, additional entry dialog boxes will appear.
listed in the scan list on the
Enter the requested entry and click OK for each dialog
DICOM Reports screen.
box. (User defined entries can be left blank.) See
The maximum length for the user “HL7 Tab” on page 18-34 for details on enabling/
defined entries is 64 characters
disabling the display of these fields and defining their
label names.

Previewing a DICOM Report


The Preview button allows you to view the DICOM report before
you save or send it.

Reports 15-25
QDR Reference Manual

To Preview a DICOM Report:


Step Action
1 Click on the scan you want to preview from the scan list
on the DICOM Reports screen.
2 Click on Preview.

KP1068A-2025

3 The DICOM Report Preview screen will be displayed.

4 Click and drag with the mouse or use the scroll bars to
view the report.
5 Press and hold the Shift key, then click with the mouse to
zoom/de-zoom the report.
6 Click on <<Back to close the DICOM Report Preview
screen.

Printing a DICOM Report


The Print button on the DICOM Preview screen allows you to
print the DICOM report to the local default printer.
To Print a DICOM Report:
Step Action
1 Click on the scan you want to print from the scan list on
the DICOM Reports screen.
2 Click on Preview.

15-26 Reports
QDR Reference Manual

Step Action
3 The DICOM Preview Report screen is displayed.

4 Click on Print.
The DICOM Report is sent to your default printer.
KP1068A-2028

5 Click on <<Back to close the DICOM Report Preview


screen.

Saving a DICOM Report


The Save As button saves a DICOM report as a file to your
desired location.
To Save a DICOM Report:
Step Action
1 Click on the scan you want to save from the scan list on
the DICOM Reports screen.
2 Click on Save As.
KP1068A-2029

Reports 15-27
QDR Reference Manual

Step Action
Notes: The maximum length for 3 If the scan does not have an accession number, the
the Accession Number is 16 Accession Number Entry dialog box will appear. Enter in
characters.
an accession number for the scan and click OK to
proceed.

KP1068A-2030

The maximum length for the user Note: You may be prompted to enter additional user
defined entries is 64 characters defined entries. If so, additional entry dialog boxes will
appear. Enter the requested entry and click OK for each
dialog box. (User defined entries can be left blank.) See
“HL7 Tab” on page 18-34 of Chapter 18 for details on
enabling/disabling the display of these fields and defining
their label names.
Note: DICOM files are 4 The Save As dialog box will be displayed with a default
automatically saved with a .dcm file name.
extension. The default file name
is composed of: Patient name,
date, and time of the scan. You
can save the report to a different
file name, if you wish.

KP1068A-2031

5 Select the drive and folder you wish to save the report.
The example above saves the report in folder Scan.
6 Click on Save.
A dialog box will be displayed to confirm that the
DICOM report has been successfully saved.

15-28 Reports
QDR Reference Manual

Step Action

KP1068A-2032

7 Click on OK.

Sending a DICOM Report


The Send Report button allows you to send the DICOM report to
your configured destination remote node(s).
To Send a DICOM Report:
Step Action
Note: Multiple scan selections 1 Select the scan(s) you want to send from the scan list on
are allowed for the Send the DICOM Reports screen.
function.
2 If two or more destinations are shown in the Destination
list, highlight the desired destination by clicking on its
name.
If you want to include the Ancillary Report and/or
patient’s Questionnaire in the study, check the respective
check box.

3 Click on Send.

KP1068A-2033

Reports 15-29
QDR Reference Manual

Step Action
Note: The Accession Number 4 If a scan does not have an accession number, the
Entry dialog box will be Accession Number Entry dialog box will appear. Enter in
presented for each selected scan
an accession number for the scan and click OK to
that does not have an accession
number associated with it. proceed.
The maximum length for the
Accession Number is 16
characters.

KP1068A-2034

Note: You may be prompted to enter additional user


The maximum length for the user defined entries. If so, additional entry dialog boxes will
defined entries is 64 characters appear. Enter the requested entry and click OK for each
dialog box. (User defined entries can be left blank.) See
“HL7 Tab” on page 18-34 of Chapter 18 for details on
enabling/disabling the display of these fields and defining
their label names.
5 For each selected scan, a DICOM report will be generated
and placed in the queue. DICOM then sends the report in
the order it was placed in the queue.
6 To view the send status of the report, see the sections
Viewing the Queue on page 15-32 and Viewing a History
of Sent Reports on page 15-32.

Sorting the Scan List


The scan list consists of: Scan ID, Scan Date, Scan Time,
Analysis Date, and Accession Number. If the scan list contains
more than one scan, it may be sorted by any heading in either
ascending or descending order.

15-30 Reports
QDR Reference Manual

To Sort the Scan List:


Step Action
1 Click on the title bar of the heading you wish to sort the
scan list by.

KP1068A-2035

Note: The default sort order for 2 Click on the title bar of the heading again to toggle the
the scan list is by Scan Date in sort order. For example, if the scan list is sorted by the
ascending order.
Scan ID in ascending order, clicking on the Scan ID title
bar again will sort the scan list by Scan ID in descending
order.

Selecting/Deselecting All Scans


The Select All button allows you select, with one click, all of the
scans that are currently in the scan list. The Deselect All button
will deselect all currently selected scans.
To Select/Deselect All Scans:
Step Action
Note: Multiple scan selections 1 Click on Select All.
will only work with the Send
function. All scans in the scan list will be highlighted.

KP1068A-2036

Note: When sending multiple scans to be included in a


single report, all scans must have the same Accession
Number. Therefore the selection shown above will cause
multiple reports to be sent when using the Send function.

Reports 15-31
QDR Reference Manual

Step Action
2 Click on Deselect All.
All highlighted scans in the scan list will be un-
highlighted.

KP1068A-2037

Viewing the Queue


The View Queue button allows you to view the scans in the
queue waiting to be sent.
To View the Queue:
Step Action
1 If two or more destinations are shown in the Destination
list, highlight the desired destination by clicking on its
name.
2 Click on View Queue.
The View Queue dialog box is displayed.
KP1068A-2038

Note: The View Queue displays


the DICOM report name, the IP
address of the destination node,
and the submission timestamp.

3 Click on Close.

Viewing a History of Sent Reports


The View Log button on the View Queue dialog box allows you to
view a history of reports that had been previously sent.
To View a History of Sent Reports:
Step Action
1 If two or more destinations are shown in the Destination
list, highlight the desired destination by clicking on its
name.

15-32 Reports
QDR Reference Manual

Step Action
2 Click on View Queue.
The View Queue dialog box is displayed.

Note: The View Log displays the 3 Click on View Log.


DICOM report name, the IP
address of the destination node, The View Log dialog box is displayed.
the transmission attempt
timestamp, and the transmission
status.

4 Click on Close to close the View Log dialog box.


5 Click on Close.

Refreshing DICOM Reports in the Queue


The Refresh button on the View Queue dialog box updates the
DICOM reports in the queue list box to reflect the current queue
status.
To Refresh DICOM Reports in the Queue:
Step Action
1 If two or more destinations are shown in the Destination
list, highlight the desired destination by clicking on its
name.

Reports 15-33
QDR Reference Manual

Step Action
Note: In this example, there are 2 Click on View Queue.
two reports in the queue waiting
to be sent. The View Queue dialog box is displayed.

Note: Multiple report selections 3 Click on the DICOM report(s) in the queue to select it.
are allowed. Hold down the
<CTRL> key and click each
report you wish to select.

3 Click on Refresh.
The queue list box is updated and the transmitted report is
removed from the queue since it has been sent.

4 Click on Close.

Deleting a DICOM Report from the Queue


The Delete button on the View Queue dialog box allows you to
delete a DICOM report from the queue.

15-34 Reports
QDR Reference Manual

To Delete a DICOM Report from the Queue:


Step Action
1 If two or more destinations are shown in the Destination
list, highlight the desired destination by clicking on its
name.
2 Click on View Queue.
The View Queue dialog box is displayed.

Note: Multiple report selections 3 Click on the DICOM report(s) in the queue to select it.
are allowed. Hold down the
<CTRL> key and click each
report you wish to select.

5 Click on Delete.
The DICOM report(s) will be deleted from the queue.

6 Click on Close.

Reports 15-35
QDR Reference Manual

Viewing a History of Sent Storage Commitment Reports


The View SC Log button on the View Queue dialog box allows
you to view a history of storage commitment reports that had
been previously sent.
To View a History of Sent Storage Commitment Reports:
Step Action
1 If two or more destinations are shown in the Destination
list, highlight the desired destination by clicking on its
name.
2 Click on View Queue.
The View Queue dialog box is displayed.

Note: The View SC Log displays 3 Click on View SC Log.


the DICOM report name, the IP
address of the destination node, The View Log dialog box is displayed.
the transmission attempt
timestamp, and the transmission
status.

4 Click on Close to close the View SC Log dialog box.


5 Click on Close.

15-36 Reports
QDR Reference Manual

Printing the HxReport


HxReport is the patient Questionnaire filled out during the
patient’s examination. The report is automatically sent with the
DICOM Report.
To Print the HxReport
Step Action
1 On the DICOM Report screen click Preview Hx.
The DICOM Report Preview window appears.

2 To zoom in or out, shift click either left or right button on


the mouse.
Click and hold then drag to scroll up or down.
3 The following operations can be performed from the
Questionnaire:
• Print button. Clicking this button prints the
questionnaire to the desired printer.
• << Back button. Clicking this button returns the
user to the previous screen.

Reports 15-37
QDR Reference Manual

Closing DICOM
The Cancel button and the <<Back button on the DICOM
Reports screen allows you to close the DICOM option.
To Close DICOM:
Step Action
1 Click on Cancel.

KP1068A-2048
The DICOM Reports screen is closed and you are
returned to the Main Menu screen.
2 – OR – Click on <<Back.
KP1068A-2049
The DICOM Reports screen is closed and you are
returned to the previous screen (i.e. the Reports Print
screen).

15-38 Reports
QDR Reference Manual

Alternate Navigation
This section describes alternate methods of accessing and printing
reports.

Drag and Drop Records


You can generate a report by selecting a patient record or scan
record, dragging it, and dropping it onto the Report button in the
main window.
This method is useful if you already have the patient or scan
record displayed and decide to print a report.

Drag and Drop Patient Records


This section provides instructions for dragging and dropping a
patient record onto the Report button.
To drag and drop a patient record:
Step Action
1 Click the Patients button in the main window.
The system displays an alphabetical list of patients.

KP0886C-1411

2 Click a patient name. While holding down the left mouse


button, drag the record and drop it (release the mouse
button) on the Report button.
The system displays the Select Scan window with the
patient’s name at the top, and a list of scans. Scans are
listed by date with the most recent scan at the top.
3 Continue with Step 3 of To print a report using the Report
button: on page 15-7. Or, continue with Step 3 of To
Create and Send a DICOM Report: on page 15-18.
From this point on the procedure is the same.

Reports 15-39
QDR Reference Manual

Drag and Drop Scan Records


This section provides instructions for dragging and dropping a
scan record onto the Report button.
To drag and drop a scan record:
Step Action
1 Click the Scans button in the main window.
The system displays the Scan Drawer window with three
tabs: Analyzed Scans, Unanalyzed Scans, and Both.

KP0886C-1412

2 Click the Analyzed tab and select the scan record to be


printed. While holding down the left mouse button, drag
the record and drop it (release the mouse button) onto the
Report button.
The system displays the Print window with the patient
name and ID at the top.
3 Continue with Step 4 of To print a report using the Report
button: on page 15-7. Or, continue with Step 4 of To
Create and Send a DICOM Report: on page 15-18.
From this point on the procedure is the same.

15-40 Reports
Chapter 16
Archiving, Locating and
Restoring Scans
This chapter provides information and instructions for archiving,
locating, and restoring scans, as well as managing the archive
locations. This chapter also provides a troubleshooting section if
you encounter a problem transferring scans from one location to
another.

Content
Subject Page
Overview 16-2
Archiving Scans 16-8
Locating Scans 16-12
Restoring Scans 16-16
Copying Scans 16-19
Query/Retrieve Scans 16-22
Managing Archive Locations 16-27
Troubleshooting Transfer Errors 16-28

16-1
QDR Reference Manual

Overview
This section provides an explanation of terms and concepts used
when archiving, locating, restoring, and copying scans. The
following definitions may be helpful when reviewing this
chapter:
• Archiving
The process of duplicating scans contained on the
computer hard drive and transferring them to uniquely
identified removable media, to a uniquely identified
directory on a computer network, or to a Picture Archival
and Communications System (PACS) server for
permanent storage.
• Locating
The process of determining the storage location(s) of
archived scans.
• Restoring
The process of duplicating archived scans and transferring
them to the computer hard drive.
• Copying
The process of duplicating scans contained on the
computer hard drive and transferring them to another
location without electronically recording the location to
which the scans data have been copied.
• Query/Retrieve
The process of locating and restoring scans archived to a
PACS server.
Note: See “Running the System It is extremely important to understand the difference between
Backup” on page 17-5 of archiving scans and performing a system back up. The archive
Chapter 16 for information on
process transfers a copy of the selected scans to removable
how to perform a System
Backup. media—such as a CD R/W—or to a directory on a computer
network. The system back up process transfers a copy of data
contained in the QDR system database, which is different from
scan data files —it does not save scan image files.

Handling Read/Write CDs


Formatting a Read/Write CD
Warning: To ensure the integrity The first time you use a read/write CD in the CD R/W it must be
of the data on a read/write CD, formatted. For information on formatting a CD refer to the
the CD must be formatted the
manual that came with the CD R/W.
first time it is used.

16-2 Archiving, Locating and Restoring Scans


QDR Reference Manual

Removing a Read/Write CD
If your CD does not eject by pressing the button on the front of
the CD R/W drive, perform the following.
To Eject a CD from a CD R/W
Step Action
1 Minimize the APEX software by pressing the minimize
button in the upper right corner.

Minimize button
KP1035A-1501

2 Start Windows Explorer.


3 Right click on the icon for the CD R/W.
A menu appears.
CD Drive icon
Menu

KP1035A-1502
Eject command

4 Click on the Eject command in the menu and wait for the
CD tray to open.

Archiving Scans
Archiving scans is the process of duplicating scans contained on
the computer hard drive and transferring them to uniquely
identified removable media, to a uniquely identified directory on
a computer network, or to a Picture Archival and
Communications System (PACS) server for permanent storage.
Scan image files consume a large amount of space on the
computer hard drive. Scans should be archived for permanent
storage on a daily basis. It is strongly recommended that scans be
archived to two different locations. For example, perform the

Archiving, Locating and Restoring Scans 16-3


QDR Reference Manual

initial archive to a CD R/W and the second archive to a network


server or read/write CD. Two separate archive locations for each
scan will help to ensure that these scans will be accessible for use
during future patient visits.

Note: See Chapter 6 for Once scans have been archived, they can be deleted from the
information on how to delete computer hard drive to free up space.
scan files from your system.
Scans may be automatically deleted after they have been
archived (“Archive Tab” on page 18-31).

Locations
The archive process duplicates scans that are in the system
location and transfers the duplicates to an archive location, as
described below:
• System Location
Scans that are in the system location reside on the
computer hard drive and are available for analysis and
reporting. Use the Scans button in the main window to
view the list of scans that are currently on the system (that
is, in the system location). The system location is the point
of origin for the archive process.
• Archive Location
This is the destination location for the archive process.
There can be two archive locations for each scan—
primary and secondary. Archive locations can be on
removable media such as CDs or 3-1/2 inch diskettes.
They can also be in different directories on a computer
network or on a Picture Archival and Communiction
System (PACS) server. The APEX software will keep
track of two separate archive locations for each scan. An
archive location for removable media or a network drive
consists of two pieces of information:
• A fully qualified directory path (different for
PACS server)
The directory path is the complete specification of
a single file storage location. Fully qualified
means that the path starts with a drive
specification (such as a:) and ends with the name
of the directory that contains the file. For example,
a:\ is a directory path that instructs the system to
place archive files on a 1.44MB, 3.5 inch diskette.
• An archive label (not used for PACS)
The archive label is used to uniquely identify the

16-4 Archiving, Locating and Restoring Scans


QDR Reference Manual

archive location. Removable media should be


marked with the archive label immediately.

Archive Labels
Warning: A read/write CD can The first time a scan is archived to a removable media location,
be damaged if you do not use a the system creates an archive label from the date (of the archive),
fine-point permanent ink CD
a sequence number (to identify the number of removable storage
marker or a CD label when
labelling the CD. media needed for the archive) and the Serial Number of the
machine (see Figure 16-1). This label uniquely identifies the
removable storage media for future reference in locating and
restoring scans. The system will prompt you to write the archive
label using a fine-point permanent ink CD marker on a read/write
CD or on the label affixed to the storage media (such as a
1.44MD diskette). Figure 16-1 illustrates the archive label
information for a QDR serial number 80231.
Figure 16-1
Archive Label Sequence Number

092603-01-SN80231

Date (mmddyy) QDR


Serial Number
00158-002

Locating Scans
Locating scans is the process of determining the storage
location(s) of archived scans. Use the Locate Scans button to
locate scans. The Locate function provides the following
information about each scan:
• Analysis time and date
• Archive time and date
• Archive location
• Archive label
Note: Scans archived to a Picture Archival and Communications
(PACS) server can be located using the Query/Retrieve Scans
function. (Refer to “Query/Retrieve Scans” on page 16-22).

Archiving, Locating and Restoring Scans 16-5


QDR Reference Manual

Restoring Scans
Note: New patients may bring Restoring a scan is the process of duplicating an archived scan
electronic copies of Hologic and transferring it to the computer hard drive. The normal
scans performed at another
method for restoring a scan is to first locate the archived scan
facility. These scans should be
restored, viewed and archived. using the Locate function, which enables the Restore function.
The alternate method for restoring a scan is to select Restore
Scans from the Archive pull-down menu. Use this method to
restore a scan whose archive location is not contained in the
system’s database (for example, a new patient who has an
electronic copy of a scan performed at another facility). Once a
scan has been restored to the system, you can reanalyze it, create
a new report for the scan, and/or use the scan as the basis for
comparison with another scan.

Note: Scans archived to a Picture Archival and Communications


(PACS) server can be restored using the Query/Retrieve Scans
function. (Refer to “Query/Retrieve Scans” on page 16-22).

Copying Scans
Copying a scan is the process of duplicating a scan without
electronically recording the location to which the scan has been
copied. The Copy function should be used whenever custody of
the copied scan is not being retained. For example, use the Copy
function to provide a patient with an electronic copy of their
scan(s) or to submit scans to a third party for review.

About Archiving and Restoring Scans


This section provides an explanation of terms and concepts used
when archiving or restoring scans from one location to another.
To simplify the explanation of these processes, they will be
referred to as data transfer processes.

How Scans are Transferred


There are several data files associated with each scan, and they
must always be transferred as a set. Scans are transferred one at a
time, even when transferring multiple scans. The transfer of each
scan is not affected by the transfer of other scans.
When the system is in the process of transferring scans, it
prevents other operators from accessing the location to which the
scans are being transferred. If an operator tries to access a locked
location, their system will display a message prompting them to
chose another location or to try again later. Once the scan transfer

16-6 Archiving, Locating and Restoring Scans


QDR Reference Manual

has been completed, normal access will be restored to that


location. This feature is extremely important for scans that have
been archived to a directory on a computer network.
Transfer Progress Message
While scans are being transferred, the system displays a progress
message and provides the ability to cancel the transfer process.
The progress message displays the progress of the transfer.
Canceling a Transfer
The transfer process may be stopped at any time. The archiving of
the scan being transferred at the time the process was stopped is
canceled. All other pending transfers are canceled as well. Scan
transfers that were completed prior to the time the process was
stopped are not affected.
Errors During Transfer
If an error occurs during the transfer, the system cancels the
current transfer and all pending transfers, and displays an error
message. Refer to Troubleshooting Transfer Errors on page 16-28
for more information.
Notification of Successful Transfers
When the entire transfer has been successfully completed, the
system displays a message stating that the transfer is complete.

How the System Handles Data Inconsistencies


When the Locate or Restore function is used, the patient’s
biographical information and other data contained in the scan
being restored is compared to the information in the system
database. If the system detects inconsistencies, it automatically
corrects the information. For patient information, the data in the
scan file will be updated to agree with the system data. For
example, if the last name of a patient has changed from the date
of their last exam, the system will automatically update the
archive location data to reflect the patient’s new last name. When
data pertaining to the scan or analysis is inconsistent, the data in
the system database will be updated to agree with the data in the
scan file.
How the System Handles Archive Data
The system database keeps track of the two most recent archive
locations for a given scan. The database contains archive data
including the location, label, archive date, and analysis date. If an
attempt is made to archive a scan that has already been archived
to two locations, the system automatically alerts the operator that

Archiving, Locating and Restoring Scans 16-7


QDR Reference Manual

they are about to overwrite the archive information for the oldest
archive. The operator is then prompted to either manually record
the archive location that is about to be overwritten or cancel the
archive process.

Archiving Scans
Archiving a scan is the process of duplicating the scan contained
on the computer hard drive and transferring it to uniquely
identified removable media or to a uniquely identified directory
on a computer network for permanent storage. Once a scan has
been successfully archived, it may deleted from the system
(computer hard drive) to make space for new scans.

Archive Location
The archive location can be any of the following, depending on
which type(s) of device have been installed on the system:
• Read/write CD
• 3.5 inch diskette
• Optional Magneto-Optical or JAZ® cartridge
• A directory on a network to which the system is connected
• A Picture Archival and Communiction System (PACS)
server using the Report - DICOM Send function before
exiting and exam or accessing Reports from the Main
screen (refer to “DICOM Reports” on page 15-18).

Archive Process
The process of archiving scans is as follows:
1. Select the archive location
2. Select the scan(s) to archive
3. Start the archive function

16-8 Archiving, Locating and Restoring Scans


QDR Reference Manual

To archive scans:
Step Action
Warning: If the Archive 1 Click the Archive Scans button in the main window.
procedure was interrupted by a The Archive Selected Scan(s) window displays.
power failure the Archive may
not be complete on the storage
media. Check the media and
repeat the Archive if necessary.

KP1035A-0321

KP0886C-1503

Note: If you use a removable 2 Determine whether the path shown in the Archive
archive media that has never Location field at the top of the window is the location
been used for archiving scans,
(path and archive label) to which the scans should be
the Label field will be blank.
archived. To change the archive location, do the
Note: Refer to Managing following:
Archive Locations on page 16-
27 for information on adding • Click the down arrow on the Path pull-down box
archive locations. to display the available archive locations.
• Drag down and click the desired archive location.
The selected location appears in the pull-down box and
the associated label appears in the Label field.
Note: The Unarchived tab is the 3 Click one of the following tabs:
default selection.
• Unarchived to display a list of unarchived scans
Note: Once a scan has been
archived, it appears on the • Archived Once to display scans that have been
Archived Once tab, even if archived at least one time
archived again, to a secondary
location. • All Scans to display all scans (both Unarchived
If using new media for the first and Archived Once)
time, format the media from A list of scans of the selected type displays.
Windows Explorer.
4 Select the scan(s) to be archived using one or more of the
following techniques:

Archiving, Locating and Restoring Scans 16-9


QDR Reference Manual

Step Action
Note: Daily QC Phantom scans • Click the Select All button to select all the scans
provide information about your appearing on the list.
system that can be used if
problems occur. Archive all • Locate the scans either by
daily QC Phantom scans with
patient scans to provide a Using the scroll bars to bring scans into view; or
history of your system’s Typing the patient’s name (last name, comma,
operation.
space, first name, space, middle initial) until the
scans for that patient appear.
Note: To deselect individual • To select multiple scans that are not adjacent to
scans that have been selected, each other, depress and hold down the Ctrl key
depress and hold down the Ctrl
while clicking on each scan to be archived.
key, then click on the scans to be
deselected. Use the Deselect All • To select a range of adjacent scans, click on the
button to deselect all selected first scan in the range, depress and hold down the
scans.
Shift key, then click on the last scan in the range.
The selected scans display as highlighted (white text on a
blue background).
Note: The system will notify the 5 Click the Archive Scans button.
operator if there is no archive
media installed in the archive If using removable archive media that has never been
location, or if a network location used before for archiving, the system will prompt the
is not accessible. operator to manually record on the media label the
archive label name assigned by the system.
The Scan Transfer in Progress window is displayed while
the selected (highlighted) scans are duplicated and
transferred to the archive location specified in the
Archive Location field. The progress of the transfer
operation is displayed in this window.

KP0886C-1504

6 To stop the transfer process before it completes, click the


Cancel Transfer button in the Scan Transfer in Progress
window.
The archiving of the scan being transferred at the time
the Cancel Transfer button was clicked is canceled. All
other pending transfers are canceled as well. Scan
transfers that were completed prior to the time the button
was clicked are not affected.

16-10 Archiving, Locating and Restoring Scans


QDR Reference Manual

Step Action
7 When the Transfer results window displays, click OK.
8 It is highly recommended that you do an immediate
second archive of these scans to another cartridge or disk.
This second archive protects from loss of scans in the
event of damage to the first cartridge or disk.

Archiving, Locating and Restoring Scans 16-11


QDR Reference Manual

Locating Scans
The Locate function allows the operator to search the system
database for specific scans that have been archived. The Locate
function typically precedes the Restore function. Once a scan has
been located, it may be restored from its archive location to the
system location on the computer hard drive.
Use the Locate Scans button to search for and restore scans.
Note: Scans archived to a Picture Archival and Communications
(PACS) server may also be located using the Query/Retrieve
Scans function. (Refer to “Query/Retrieve Scans” on page 16-
22).
To locate and restore scans:
Step Action
1 Click the Locate Scans button in the main window.
The Select Patient for Scan Location window displays.

KP1035A-0311

KP0886C-1506

16-12 Archiving, Locating and Restoring Scans


QDR Reference Manual

Step Action
Note: To view the patient’s 2 Find the name of the patient for which scans are to be
biographical information, click restored either by
on the Details button in the
upper right corner of the • Using the scroll bars to bring patient names into
window. view; or
• Typing the patient’s name (last name, comma,
space, first name, space, middle initial) until the
scans for that patient appear.
Click on the patient’s name to select the patient, then
click the Locate Scans button.
The system displays the Locate Selected Scan(s) window
with the patient’s name and ID at the top.

KP0886C-1507

Note: You should always select 3 The Primary Archive tab window will appear with a list
scans to restore from the of scans for the selected patient.
Primary Archive tab, the
Secondary Archive tab can be The system displays a detailed list of scans of the selected
used if you cannot restore scans patient that includes the Scan Date, Scan Type, Scan ID,
from the Primary Archive tab. Analysis Date and Label (of the media) for each scan.
Note: To switch between the
Primary Archive Tab screen and
the Secondary Archive Tab
screen use Ctrl-Tab or the
mouse.

Archiving, Locating and Restoring Scans 16-13


QDR Reference Manual

Step Action
Note: To deselect individual 4 Select the scan(s) to be restored using one or more of the
scans that have been selected, following techniques:
depress and hold down the Ctrl
key, then click on the scans to be • Click the Select All button to select all the scans
deselected. Use the Deselect All appearing on the list.
button to deselect all selected
• Click on a single scan to select it.
scans.
• To select multiple scans that are not adjacent to
each other, depress and hold down the Ctrl key
while clicking on each scan to be restored.
• To select a range of adjacent scans, click on the
first scan in the range, depress and hold down the
Shift key, then click on the last scan in the range.
The selected scans display as highlighted (white text on a
blue background).
5 Note: Skip this step when the archived scan(s) are located on a
network drive or a Picture Archival and Communications (PACS)
server.
Locate and place the cartridge or disk with the correct
Label into the disk drive.
Label

KP0886C-1508

16-14 Archiving, Locating and Restoring Scans


QDR Reference Manual

Note. If you can not restore 6 Click the Restore Scans button.
scans from the Primary Archive
media screen, contact the If the list of selected scans includes removable media
Hologic Help Desk before using archive locations with different archive labels, the system
the Secondary Archive media. will automatically prompt the operator to insert
specifically labeled archive media as necessary.
The Scan Transfer in Progress window is displayed while
the selected (highlighted) scans are duplicated and
transferred to the system location. The progress of the
transfer operation is displayed in this window.

KP0886C-1509

7 To stop the transfer process before it completes, click the


Cancel Transfer button in the Scan Transfer in Progress
window.
The restoration of the scan being transferred at the time
the Cancel Transfer button was clicked is canceled. All
other pending transfers are canceled as well. Scan
transfers that were completed prior to the time the button
was clicked are not affected.
8 When the Transfer results window displays, click OK.

Archiving, Locating and Restoring Scans 16-15


QDR Reference Manual

Restoring Scans
The Restore function differs from the Locate function in that it
allows the operator to search for and restore scans by specifying
the location. This includes importing archive scans that may or
may not be resident in the system (that is, on the hard drive). For
example, the Restore function would be used for a new patient
who has an electronic copy of a scan performed at another
facility. When the Restore function is started, the archive location
from which the scans are to be restored must be specified. The
system searches that location and displays a list of all scans stored
there. One or more scans can be selected and restored from that
archive location to the system location on the computer hard
drive.
Note: Scans archived to a Picture Archival and Communications
(PACS) server can be restored using the Query/Retrieve Scans
function. (Refer to “Query/Retrieve Scans” on page 16-22).
To restore scans from a specified archive location:
Step Action
Note: A new patient’s scans 1 Click the Archive pull-down menu on the main window,
restored onto the system should then drag down to select Restore Scans.
be reviewed and then archived
The Restore Selected Scan(s) window displays.
onto the system’s archive media.

KP0886C-1510

16-16 Archiving, Locating and Restoring Scans


QDR Reference Manual

Step Action
Note: If the current path name is 2 Determine whether the path shown in the Archive
for a removable media drive, the Location field at the top of the window is the location
system will check the drive and
(path and archive label) from which the scans should be
prompt for a cartridge or
diskette to be inserted, if restored. To change the archive location, either:
necessary. • Click the down arrow on the Path pull-down box
to display the available archive locations, then
drag down and click the desired archive location;
or
• Click the Browse... button and navigate to the
directory that contains the archived scans to be
restored.
The selected location appears in the pull-down box and
the associated label appears in the Label field.
3 Select the scan(s) to be restored using one or more of the
following techniques:
• Click the Select All button to select all the scans
appearing on the list.
• Locate the scans either by
Using the scroll bars to bring scans into view; or
Typing the patient’s name (last name, comma,
space, first name, space, middle initial) until the
scans for that patient appear.
• To select multiple scans that are not adjacent to
Note: To deselect individual each other, depress and hold down the Ctrl key
scans that have been selected, while clicking on each scan to be archived.
depress and hold down the Ctrl
• To select a range of adjacent scans, click on the
key, then click on the scans to be
deselected. Use the Deselect All first scan in the range, depress and hold down the
button to deselect all selected Shift key, then click on the last scan in the range.
scans. The selected scans display as highlighted (white text on a
blue background).

Archiving, Locating and Restoring Scans 16-17


QDR Reference Manual

Step Action
4 Click the Restore Scans button.
If the list of selected scans includes removable media
archive locations with different archive labels, the system
will automatically prompt the operator to insert
specifically labeled archive media as necessary.
The Scan Transfer in Progress window is displayed while
the selected (highlighted) scans are duplicated and
transferred to the system location. The progress of the
transfer operation is displayed in this window.

KP0886C-1511

5 To stop the transfer process before it completes, click the


Cancel Transfer button in the Scan Transfer in Progress
window.
The restoration of the scan being transferred at the time
the Cancel Transfer button was clicked is canceled. All
other pending transfers are canceled as well. Scan
transfers that were completed prior to the time the button
was clicked are not affected.
6 When the Transfer results window displays, click OK.

16-18 Archiving, Locating and Restoring Scans


QDR Reference Manual

Copying Scans
The Copy Scans function allows the operator to copy scans that
currently reside on the computer hard disk to a specified location.
This function should be used whenever custody of the copied
scans is not being retained. For example, use the Copy Scans
function to provide a patient with an electronic copy of their
scan(s) or to submit scans to a third party for review. The system
will copy the selected scans to the specified location without
recording the location in the system database.
Any scans copied to an offline location can be later restored using
the Restore Scans function.
To copy scans:
Step Action
1 Click the Archive pull-down menu on the main window,
then drag down to select Copy Scans.
The Copy Selected Scan(s) window displays.

KP0886C-1512

Note: The Copy Location is 2 Determine whether the path shown in the Copy Location
where the selected scans will be field at the top of the window is the location to which the
copied.
scans should be copied. To change the Copy location,
either:
• Click the down arrow on the Path pull-down box
to display the available copy locations, then drag
down and click the desired copy location; or
• Click the Browse... button and navigate to the
directory that contains to which the scans are to
be copied.
The selected location appears in the pull-down box.

Archiving, Locating and Restoring Scans 16-19


QDR Reference Manual

Step Action
Note: The Unarchived tab is the 3 Click one of the following tabs:
default selection.
• Unarchived to display a list of unarchived scans
Note: Once a scan has been
archived, it appears on the • Archived Once to display scans that have been
Archived Once tab, even if archived at least one time
archived again, to a secondary • All Scans to display all scans (both Unarchived
location. and Archived Once)
A list of scans of the selected type displays.
Note: To deselect individual 4 Select the scan(s) to be copied to the specified location
scans that have been selected, using one or more of the following techniques:
depress and hold down the Ctrl
key, then click on the scans to be • Click the Select All button to select all the scans
deselected. Use the Deselect All appearing on the list.
button to deselect all selected
• Locate the scans either by
scans.
Using the scroll bars to bring scans into view; or
Typing the patient’s name (last name, comma,
space, first name, space, middle initial) until the
scans for that patient appear.
• To select multiple scans that are not adjacent to
each other, depress and hold down the Ctrl key
while clicking on each scan to be archived.
• To select a range of adjacent scans, click on the
first scan in the range, depress and hold down the
Shift key, then click on the last scan in the range.
The selected scans display as highlighted (white text on a
blue background).
Note: The system will notify the 5 Click the Copy Scans button.
operator if there is no archive
media installed in the archive The Scan Transfer in Progress window is displayed while
location, or if a network location the selected (highlighted) scans are copied to the location
is not accessible. specified in the Archive Location field. The progress of
the transfer operation is displayed in this window.

KP0886C-1513

16-20 Archiving, Locating and Restoring Scans


QDR Reference Manual

Step Action
6 To stop the transfer process before it completes, click the
Cancel Transfer button in the Scan Transfer in Progress
window.
The copying of the scan being transferred at the time the
Cancel Transfer button was clicked is canceled. All other
pending transfers are canceled as well. Scan transfers
that were completed prior to the time the button was
clicked are not affected.
7 When the Transfer results window displays, click OK.

Archiving, Locating and Restoring Scans 16-21


QDR Reference Manual

Query/Retrieve Scans
The Query/Retrieve function allows the operator to locate and
retrieve scans meeting given parameters and filters from a
configured PACS server into the QDR system. Scans are stored
on PCAS servers using the DICOM Send functionality. (See
“DICOM Reports” on page 15-18.)
Note: Scans archived to a Picture Archival and Communications
(PACS) server may also be restored using the Locate Scans
function. (Refer to “Locating Scans” on page 16-12).

When the Query/Retrieve function is started, the PACS server


must be specified. The system queries that server and displays a
list of all scans meeting the given parameters and filters stored on
that server. One or more scans can be selected and retrieved from
that server to the system location on the computer hard drive.
To locate and retrieve scans from a PACS server

Step Action
1 Select Query/Retrieve from the Archive drop-down
menu on the main screen.
DICOM Query/Retrieve Studies displays.

2 Complete Query parameters as desired. (See “Query


Parameters for details.)
3 Click Optional Filters to add study level filters to the
query or go to Step 5.
4 Complete Study level filters as desired.

16-22 Archiving, Locating and Restoring Scans


QDR Reference Manual

Step Action
5 If more than one active location is configured, select
archive location (Destination). Click Destination field
down arrow and click desired destination.
6 Click Query. All studies meeting the query parameters
and filters archived on the destination PACS server
display in the Retrieve section of the DICOM Query/
Retrieve Studies window.
7 In the Retrieve section, select the study or studies to
retrieve.
• Select a single study by clicking on it.
• Select multiple studies that are not adjacent to each
other by holding the CTRL key down while clicking
on each study to be retrieved.
• Select a range of studies adjacent to each other by
clicking on the first study in the range, depressing and
holding down the Shift key, then clicking on the last
study in the range.
8 Click Retrieve.
9 Click View Queue to check status of current retrieve
requests. (See “Retrieve Queue” on page 16-25.)

Query Parameters

Patient ID Identification data such as medical record


number, social security number, etc.
Last Name, Last Name can include suffix where
First Name, applicable.
Middle Initial
Birth Date To include birth date as a parameter, first
check the checkbox. Change month, day and
year by clicking the respective section and
over typing.
Monthly calendar can be displayed for the
given month/year by clicking down arrow.

Archiving, Locating and Restoring Scans 16-23


QDR Reference Manual

Date Range Date range to query. Change month, day and


year by clicking the respective section and
over typing. Monthly calendar can be
displayed for the given month/year by
clicking down arrow.
Destination Archive location to query.

Study Level Filters

Accession Identifies a patient visit.


Number
Study ID Identifies a study.
Study Time To include study time as a filter, first check
the checkbox. Change hour, minute, second
and AM/PM by clicking the respective section
and over typing or clicking up/down arrow.
Study Instance Unique identifier for this study instance.
UID
Modalities in Type of system recognized by the HIS/RIS.
Study Since “bone densitometer” is not a standard
defined type, the QDR system is defined as
“OT” for “Other”.
Referring Name.
Physician

16-24 Archiving, Locating and Restoring Scans


QDR Reference Manual

Retrieve Queue
Allows the user to check status and delete current retrieve
requests, and view a log file of all retrieve requests.

Queue list Selection list of retrieve request


information that can be filtered to display
all requests or requests with a selected
status depending on the radio button
selected in the Display section.
Display Filters the selection list depending on the
button selected.
Select All/Deselect Toggles between Select All and Deselect
All All - Selects all Retrieve requests listed in
the Queue List.Deselect All - Deselects all
Retrieve requests listed in the Queue
Delete Deletes the selected Retrieve request(s).
View Log Displays the Retrieve Queue Log. See
“Retrieve Queue Log” on page 16-26
OK. Closes window. Returns to DICOM
Query/Retrieve Studies window.

Archiving, Locating and Restoring Scans 16-25


QDR Reference Manual

Retrieve Queue Log


Displays detailed information regarding the processing of
DICOM retrieve requests in reverse chronological order (most
recent entries first).

Click OK to return to the Retrieve Queue window.

16-26 Archiving, Locating and Restoring Scans


QDR Reference Manual

Managing Archive Locations


This section provides information and instructions on configuring
the system to manage archive locations.
An archive location for removable media or a network drive
consists of two pieces of information:
• A fully qualified directory path (different for PACS
server)
The directory path is the complete specification of a single
file storage location. Fully qualified means that the path
starts with a drive specification (such as a:) and ends with
the name of the directory that contains the file. For
example, a:\ is a directory path that instructs the system to
place archive files on a 1.44MB, 3.5 inch diskette.
• An archive label (not used for PACS)
The archive label is used to uniquely identify the archive
location. Removable media should be marked with the
archive label immediately.
These two pieces of information are entered in the system
database each time a scan is archived. For each scan, the system
keeps a record of the location of the two most recent archive
operations.
Archive locations are managed through the System Configuration
function located on the Utilities menu at the top of the main
window. The Archive tab of this window provides the tools
necessary to modify the following archive location information:
• Add a New Archive Location
Adds a new fully qualified path specification to the Path
pull-down box.
• Set the Default Archive Location
The default location is the path that initially appears in the
Path pull-down box when a window opens.
• Delete an Archive Location
Deletes an archive path from the Path pull-down box list.
• Include Raw Data Files
When checked, causes raw data files to be included when
transferring scans. Raw data files should not be
transferred with the scan files unless you have been
instructed to do so by Hologic, Inc.
The “Archive Tab” on page 18-31 section of Chapter 18 contains
detailed instructions on adding a new location, setting and
changing the default location, and deleting a location.

Archiving, Locating and Restoring Scans 16-27


QDR Reference Manual

Troubleshooting Transfer Errors


This section lists and describes the types of messages that the
system may display if it encounters problems when transferring
scans.

Problems Accessing an Archive Location


When scans are being restored and the system is unable to access
an archive location, it may display one the following messages,
depending on the circumstances.

Message Meaning Action


The archive Another operator is Click the OK
location at <Path> using that archive button. Select
is in use; Please location and it is another location, or
chose another locked to prevent try accessing the
location or try later. duplicate transfers. archive location
later.
The archive loca- The archive loca- Click the OK
tion at <Path> can- tion is not accessi- button. Verify the
not be accessed. ble. validity of the
Please choose archive location.
another location. Select a valid
location.

Transferring Scans
There are a number of reasons for encountering problems with
scan transfers. Reasons may be due to operating system errors, or
insufficient space on the removable media. Refer to the following
table for information on specific messages.
Message Meaning Action
Unable to An error occurred Click the OK
<Archive/Restore> while attempting to button and follow
scan due to an archive or restore instructions as
operating system the scan. The sys- prompted by the
error. tem will provide system.
specific informa-
tion about the error.

16-28 Archiving, Locating and Restoring Scans


QDR Reference Manual

Message Meaning Action


A file for scan The file is missing Click the OK
<Scan ID> is miss- from the directory button.
ing. Pending trans- but the system will
fers of other scans continue to transfer
will continue. other scans.
Cannot archive data The data at the Click the OK
at this location; specified archive button and contact
please contact your location is corrupt your service
service representa- and cannot be read. representative or
tive. the Hologic Help
desk.

Archive Locations
If the system encounters problems trying to transfer scans to an
archive location, it will display a message depending on the
circumstances.
Message Meaning Action
Unable to archive The location the Try archiving the
scan to <Path>; system is attempt- scan(s) to a
The selected ing to transfer to different location.
location is out of does not have
space; try another enough space to
location. accommodate the
data.
The cartridge or The removable Click the Cancel
diskette in drive media diskette or button to stop the
<drive letter> is cartridge is full. transfer and insert a
out of space; Please new diskette or
insert a different cartridge. Note: If
cartridge or diskette transferring a large
and click ‘OK’ to volume of data, try
continue. Click using a read/write
‘Cancel’ to stop the CD instead of a 3.5
archive process. inch diskette.
The archive loca- The specified Click the Yes
tion at <Path> archive location button to create a
does not exist. Cre- does not exist. The new location and
ate? system prompts the resume the transfer,
operator to create a or click the No
new location. button to cancel the
operation.

Archiving, Locating and Restoring Scans 16-29


QDR Reference Manual

Message Meaning Action


Cannot create an The system is Click the OK
archive location at unable to create an button. Refer to the
<Path>; please archive at the speci- specific error
choose another fied location. message for more
location and try information.
again.
Unable to access The system cannot Click the OK
<Path>. Please access the speci- button. Select
choose another fied archive loca- another location, or
location. tion. It may be in wait a few minutes
use by another and try again.
operator.
Scans produced by The specified Click the OK
an older version are archive location button.
archived at location contains a legacy
<Path>. Cannot archive.
create a new
archive. Please
choose another
location.

The location speci- The specified Click OK. Choose


fied by <Path> is a archive location is a valid location and
file and cannot be actually a file and continue.
used as an archive not a directory.
location; please
choose another
location.

Advanced User Information


To simplify the locate and restore process, the APEX software has
been set up to generate a single archive label (the name of an
archive location) for each removable storage media diskette or
cartridge containing scan archives. It is possible to use separate
folders (directories) within a single removable storage media
diskette or cartridge to organize the scans. Each folder (and sub-
folder if used) will be assigned a unique archive label which must
be recorded on the physical label of the storage media. It will be
extremely difficult to locate and restore scans that have been
archived within a folder in removable storage media without
having recorded on the storage media label each archive label and
the folder with which it is associated.

16-30 Archiving, Locating and Restoring Scans


QDR Reference Manual

To create a folder containing scan archives within a removable


storage media diskette or cartridge:
Step Action
Note: The folder must first be 1 Click the Utilities pull-down menu in the main window,
created on the archive then drag down to select System Configuration.
removable storage media
The System Configuration window displays.
diskette or cartridge.
2 Click on the Archive tab.
The archive configuration settings display.
3 Click in the Add New Location text box and type the
drive letter and folder name using the following format:
drive letter, colon, backslash, folder name, backslash.
Example: e:\foldername\
4 Click the Add button.
The removable media drive letter and folder name
appears in the Location Pool list.
5 Repeat steps 3 and 4 to add additional folders within the
same removable storage media diskette or cartridge.
6 Click the OK button to save the changes.
The main window displays.
Note: Refer to the Archiving 7 The new archive locations may now be used by following
Scans section of this chapter. the standard procedures for archiving scans.

Archiving, Locating and Restoring Scans 16-31


QDR Reference Manual

16-32 Archiving, Locating and Restoring Scans


Chapter 17
System Backup and Recovery
Content
Subject Page
Overview 17-2
Running the System Backup 17-5
Errors and Exceptions 17-6
System Recovery 17-8

17-1
QDR Reference Manual

Overview
System Backup is the function used to periodically save your
system files in case there is a system failure such as a hard drive
failure.
System Recover is the function used to restore your system
configuration files and database files (needed to rebuild your
system) if a system failure should occur.

Files Saved

System Disk
CD R/W

Or Or Or

SuperDisk TM

Floppy Disk
Calibration files
Patient scan database
Reference database
Reports database
Step phantom info
Network Disk
QDR registry info
Computer
KP1068A-1601

System Backup saves the following files, compressing them into


one file.
• Calibration files
• Patient scan database
• Reference database
Note: The files saved during • Report database
System Backup cannot be
selected. The system always • Step phantom information
saves the same files from a list
• APEX registry information
internal to the software.

System Backup does not overwrite any files, such as older backup
files. If a duplicate file name is found at the backup location, the
backup will not continue until one of the files is renamed.

17-2 System Backup and Recovery


QDR Reference Manual

Media Formatting
Note: System Backup permits If the media is not formatted, the program aborts. System Backup
backup to the hard drive (C:), does not format any media. This must be done by the operator
but this would defeat the purpose
prior to System Backup using the Windows operating system.
of saving files on other media, so
is not recommended.

Management of Media
Backups should always be done to multiple sets of diskettes or
cartridges. If one should become unreadable or get misplaced, the
other can be used to recover files.

Floppy Diskette
Note: Remember that all files If backing up to 3.5 inch diskettes, it is recommended that a set of
must be removed from the four diskettes be used. Label them #1 through #4. The first week,
diskette or cartridge before it
backup to the first diskette. The second week, backup to the
can be used for backup.
second diskette, and so forth. On the fifth week, erase all files on
diskette #1 and backup to it again. On the sixth week, erase all
files on diskette #2 and backup to it again, and so forth.
Because 3.5 inch diskettes wear out, they should be replaced with
new ones annually.

SuperDisk™
SuperDisk diskettes hold at least 120 Megabyte (MB) of data and
System Backup can save multiple backups on them. It is good
practice, however, to use at least two SuperDisk diskettes,
alternating them at each backup.

Read/write CD
Warning: To ensure the integrity When backing up to a CD R/W drive, the CD may contain other
of the data on a read/write CD, files (backup files or system archives).
the CD must be formatted the
first time it is used.

Network Drive
Note: Using multiple media and If network disks are available, a backup can be made to any of
replacing them periodically does them. These disks are usually backed up by your IS department
not apply here.
periodically.

When to Run System Backup


System Backup should be run at least once a week. If it has been
longer than one week since the last system backup, a message

System Backup and Recovery 17-3


QDR Reference Manual

similar to the following displays when you the system is turned


on.

KP0886C-1602

Click the Yes button to begin backing up your system. If you click
the No button, the system begins normal operation. During
normal operation, a message is displayed in the ticker tape area
that the system needs to be backed-up. Also, the System Backup
button (shown below) on the right side of the main window
blinks. These events continue during normal operation until you
run System Backup.

KP1035A-0323

Naming the Backup File


System Backup saves all files in a compressed format in one file.
The file name is in the format YYYYMMDDCCBB.CAB, where:
YYYYMM DD CC BB. CAB
Year Day This file extension indicates that
this file is stored in CAB
Month compression format.
This number indicates the number of
pieces of media needed to backup your
system files.
This number starts at 01 and increments
every time you perform a backup in a given
day.
KP0886C-1603a

17-4 System Backup and Recovery


QDR Reference Manual

For example:
If... The file name becomes...
Note: If it is necessary to use two Today is June 16, 2002 200206160101.CAB
3.5 inch diskettes for the backup,
do not use the next diskette in the You ran System Backup a sec- 200206160201.CAB
four diskette set. Instead, create ond time today
four more diskettes and label You ran System Backup on 200207290101.CAB
them Disk#1A, 2A, etc. The
July 29, 2002
system backup will now consist
of a set of eight 3.5 inch Your backup requires 2 floppy 200207290101.CAB (1st disk)
diskettes. diskettes
200207290102.CAB (2nd disk)

Running the System Backup


To run the System Backup:
Step Action
Warning: If the System Backup 1 Click the System Backup button on the main window.
procedure was interrupted by a The System Backup window displays.
power failure the System Backup
may not be complete on the
storage media. Check the media
and repeat the System Backup if
necessary.
KP1035A-0323

MAN-00214-1705

System Backup and Recovery 17-5


QDR Reference Manual

Step Action
2 Select the location (path) to which the system files are to
be backed-up. The location text box displays the default
location. It can be changed by deleting the current
location and typing the new one, or by clicking the
Browse button and navigating to the desired backup
location (illustrated below).

KP0886C-1605

Note: Changing the backup file 3 Accept the default backup file name (preferred), or enter
name makes restoring the a different backup file name (not recommended).
correct file difficult.
4 Click the OK button.
The backup process starts.

Errors and Exceptions


This section describes some of the error messages that may be
encountered while running System Backup, and explains what to
do to continue.

Backup file with the same name


If the backup file name already exists at the backup location, the
following warning displays:

KP0886C-1606

17-6 System Backup and Recovery


QDR Reference Manual

Either select a new location, delete the existing file, or change the
backup file name.

Drive is not accessible


If an attempt is made to backup to a drive containing no media
(for example, there is no 3.5 inch diskette in the drive or another
process is using the drive) the following warning appears:

KP0886C-1607

Put a diskette in the drive or wait for the other process to finish
using the drive.

Drive does not have enough space


If the backup diskette does not have enough free space, the
following warning displays:

KP0886C-1608

3.5 inch diskette


When backing up to a 3.5 inch disk drive, the diskette must be
formatted, blank (contain no files or directories), and have no
bad sectors (1,457,664 bytes available).
SuperDisk™ diskette
When backing up to a SuperDisk™ drive, the diskette may
contain files (usually other backup files).
Read/write CD
When backing up to a CD R/W drive, the CD may contain files
(other backup files or system archives).
Network drive
When backing up to a network disk, there is no issue with files.

System Backup and Recovery 17-7


QDR Reference Manual

Drive is not accessible


If an attempt is made to backup to a drive that is not accessible
(for example, a drive that is read-only), a warning similar to the
following displays:

KP0886C-1609

The problem must be corrected before continuing. In the example


shown above, click the Cancel button and choose another backup
location.

System Recovery
It is not necessary to do a System Recover unless there is a
serious problem with the hardware or software. If this happens,
contact Hologic Customer Support for help. Do not attempt to do
a System Recover yourself.
No additional information about System Recover is presented in
this guide.

17-8 System Backup and Recovery


Chapter 18
Configuring the System
APEX lets you change system configuration settings for a variety
of system parameters. You do this by setting default parameters
which stay in the system even after exiting and restarting the
software. You can change the default information at anytime the
software is running. You can also change parameters while
running different functions. If you do, they become the new
default parameters.

Content
Subject Page
Introduction to System Configuration 18-2
System Tab 18-3
Analyze Tab 18-5
CAD BMD Tab 18-8
Report Tab 18-10
Archive Tab 18-31
HL7 Tab 18-34
Scan Conversion 18-39
History 18-41
DICOM Tab 18-44
Modality Worklist 18-44
DICOM Send Destinations 18-53
DICOM Storage Commitment Destinations 18-59
DICOM Query/Retrieve Destinations 18-63
Host 18-66
Privacy Tools 18-67
Study Tab 18-68

18-1
QDR Reference Manual

Introduction to System Configuration


Your QDR system and APEX software come from the factory
with certain settings (for analysis, reports, archive, etc.) preset so
you can begin scanning and developing reports immediately.
However, you may change these settings at any time from the
System Configuration dialog box.
To enter System Configuration:
Step Action
1 Select System Configuration from the Utilities drop-down
menu on the main window.

2 Select the category (tab) that you want to change and


following the instructions below for that category.
Clicking on any of the tabs under System Configuration (see
Figure 18-1) displays the options that you can change.
Figure 18-1
System Configuration
Tabs

18-2 Configuring the System


QDR Reference Manual

System Tab
Select the System tab to configure system information. Not all
information on the System tab can be changed by the operator.
Some informational fields are automatically updated when
software is installed, system backup is performed, or service is
performed.
The three buttons on the lower right of the System tab perform the
following:
• OK button. Select to close this dialog box, save any
changes you may have made, and return to the main
window.
• Cancel button. Select to close this dialog box without
saving any changes you may have made, and return to the
main window.
• Help button. Select to display help information pertinent
to this dialog box.

Date Format
This field allows you to change the format of the date from US to
international (or vice versa).
The present format is identified by a dot in the radio button. To
change to a different format click either on the words or the radio
button, the dot in the radio button will appear in the selected field.

Patient Biography
This field allows you to change how the values are recorded on
the Patient Biography, as either US (inches, pounds) or
international (cms, kilograms). To change format, click the
appropriate radio button.

Autopositioning for Hip Scans


If checked, hip scans are automatically positioned (applies only to
Discovery systems).

Report Dose Area Product


If checked, reports total x-ray dosage applied to a patient when
the entire scan flow is completed (applies only to systems with
DAP Meter Option installed).

Configuring the System 18-3


QDR Reference Manual

Language
This field allows you to select the language the system will run
(applies only to systems with Language option installed).

Customer
This field allows you to enter the name of your firm, clinic or
hospital. Click in the text field, erase any information there and
type in the name of your firm, clinic or hospital. This text field is
limited to 32 characters.

Default Ethnicity
This field allows you to select the default ethnicity that displays
when entering biography for new patients.

Model, S/N, System ID


These fields provide information about your system that was
recorded during software installation. The fields cannot be
changed by the operator.

Data Directory
This field displays the location of the system database. This field
cannot be changed by the operator.

Software
This field is informational only and cannot be changed by the
operator. This field displays the version of the software installed
on your system. Also, to display a drop down list showing all the
software options installed on your system, click the down arrow.

Last System Backup Date


The date and time when you last ran the System Backup utility.
This field is updated each time you run the System Backup utility.

Last Service Date


The date and time when Hologic Customer Service last serviced
your system. The software sets it when service personnel run the
Service utilities.

18-4 Configuring the System


QDR Reference Manual

Analyze Tab
Select the Analyze tab to configure the methods of analysis on
your system. The Analyze tab is broken into various areas
according to the type of analysis. These are explained below.

Default
When you first select the Analyze tab the system will have
choices already selected. These are indicated by a check mark, or
radio button selection, and are the system defaults. You can
change the system defaults at any time, and your new entries will
then determine how your system will handle analysis.
The system defaults are:
• Prevent vertical motion of ROI under the Lateral
Options.
• Image Pro and Display Hip Axis Length under the
Image Display Options.
• Enable Body Composition Analysis under the General
Options (this default displays only if the Body
Composition option is loaded on your machine).
• Never under the Analysis Options - Disable Auto Low
Density
• Discovery under Image Style.
To change a configuration selection or a default setting:
Step Action
1 Click on the title of the selection you want
A check mark appears in the selected box (or dot in the
radio button).
2 If you want to remove a presently selected item
(including defaults), click on the selection name
The check mark (or dot) is removed and the selection is
removed.
3 Click OK.
The system returns to the Main window with your choices
programmed to handle the tasks selected.
To revert back to the system default at any time click on the
Defaults button.

Configuring the System 18-5


QDR Reference Manual

General Options
There are two General Options.

Display Region of Interest Coordinates


Select this option to display the region of interest (ROI)
coordinates and show the ROI values under the image. The
default is not to display region of interest coordinates.

Enable Body Composition Analysis


Select this option to calculate body composition whenever you
analyze a whole body scan.

Image Display Options


There are five Image Display Options.

Initially display unanalyzed Lateral Image as Single Energy


Select this option to initially display unanalyzed lateral images as
single energy images.

Initially display unanalyzed Hip image as Single Energy


The default for the hip is dual energy (the box is checked).
Display pseudo color image
To view pseudo color on the scan image when the analysis is in
the results step, check this box. The default is not to view pseudo
color.

Image Pro
The default is to use an enhanced image processing algorithm to
analyze IVA images. This feature can be overridden when an IVA
image is selected to be analyzed.

Display Hip Axis Length


When checked, hip axis length is displayed in the Analysis
Results screen and in hip reports.

18-6 Configuring the System


QDR Reference Manual

Lateral Options
There are two Lateral Options.

Display MID regions


Select this option to display the MIDs for lateral analysis. With
this option selected, you can also edit the MIDs. The default is not
to display MIDs.

Prevent vertical motion of ROI


Select this option to prevent vertical motion while adjusting
lateral ROI. If not selected, you can adjust the vertical size of the
lateral ROI. The default is to prevent vertical motion.

Analysis Option
There are four Analysis Options.

Disable Auto Low Density


Select Next Scan to disable the automatic low density calculation
for the next spine or hip analysis. Select Always to disable the
automatic low density calculation for all spine or hip analysis.
Select Never to enable auto low density analysis if needed by the
software during analysis. Default is to enable.

Disable Auto Analysis


Select this option to disable the Auto Analysis of the AP Spine,
the Hip or the AP Spine portion of the AP/Lateral. Default is to
enable Auto Analysis on the AP Spine and Hip.

Disable Auto Whole Body


Select this option to disable the Auto Whole Body. Default is to
enable the Auto Whole Body.

Enable NHANES BCA


Select this option to apply a correction factor to the lean tissue
results at the end of Whole Body - Body Composition and correct
the fat and percent fat results accordingly. When enabled,
NHANES BCA will be noted in the BCA results section.

Image Style
Select the desired style for image display.

Configuring the System 18-7


QDR Reference Manual

CAD BMD Tab


The CADBMD tab allows the user to select CADBMD messages
that may be displayed during analysis of scans. These messages
are designed to alert the user to any possible errors that can be
made during the analysis.
There are five CADBMD messages that may be selected for
display and one used to automatically compare the baseline scan,
if available, during analysis of scans:
Check vertebral Z-scores
When checked, displays a CADBMD message during analysis of
spine scans if any of the vertebrae have different Z-scores.
Check vertebral labels
When checked, displays a CADBMD message during analysis of
spine scans if any of the vertebrae have inconsistent labels as
compared to the compare scan.
Check vertebral regions
When checked, displays a CADBMD message during analysis of
spine scans if the regions differ than that of the compare scan.
Check analysis methods
When checked, displays a CADBMD message during analysis of
any type scan if the selected analysis method is different than that
of the compare scan.
Check midline angles
When checked, displays a CADBMD message during analysis of
hip scans if the midline angle is different than that of the compare
scan.
Auto compare with baseline scan
When checked, during analysis of any type scan the baseline
scan, if available, shall automatically be used to do a compare
analysis if it is configured.

Any CADBMD setting with a check mark is selected for display if


the error exists during the analysis of scans. The default for the
CADBMD tab is that all settings are selected for display (have a
check mark).

Press OK when done with the CADBMD tab.

18-8 Configuring the System


QDR Reference Manual

Figure 18-2
Typical CADBMD Dialog

CADBMD dialogs can be manually closed by the user clicking the


close button.

Configuring the System 18-9


QDR Reference Manual

Report Tab
Refer to Chapter 15 for a description of the various reports that
can be generated using the APEX software.
Each report is made up of a series of blocks such as the Header,
the Patient Information, Results, etc. (see Figure 18-3) which
contain various pieces of information. This section describes how
to configure that information to customize the various reports.
Figure 18-3
Report Blocks Header

Patient
Information

Scan
Information

Results

Physician
Comment
KP0886C-1705

System Configuration - Report provides three tab selected


dialogs: General, WHO Criteria, and Least Significant Change.

Report - General Tab


The Report - General dialog is used to set default selections that
are used when running the Report function. Any values changed
here become the new default settings and will display on the Print
window accessed form the Reports button.
To view the appearance of a report:
Click on the report name. A preview of the report's appearance
displays on the right.
To configure a report:
1. Click on the report name.

18-10 Configuring the System


QDR Reference Manual

2. Click Configure. The selected configure window appears.


Refer to “Configuring Reports” on page 18-14 for details.
To specify display of color fracture risks graphs:
Check the “Display colored Fracture risk” check box.
Checking this box will display the fracture risk graph in three
colors across the graph indicating normal risk (top color),
increased risk (middle color), and high risk (bottom color). The
colors in DICOM reports will be displayed in shades of gray.
To specify reports are to be ISCD:
Check the “Display ISCD Reports” check box.
Checking this box sets:
• Display T-scores for the patient above age” to 50.
• T-score based on to Caucasian.
• Default values for the “WHO criteria” page as follows:
Hip checked
Report Based on Lowest T-score of neck or total checked
AP checked
Lateral unchecked
Forearm checked
• Default values for the Reference Block Hip configuration.
(Refer to “Reference Block Hip Configuration” on page
18-27)
• Default values for the Reference Block Forearm
configuration. (Refer to “Reference Block Forearm
Configuration” on page 18-28)
To specify ancillary reports are to be included when printing
and previewing reports:
Check the “Display Ancillary Report” check box.
Ancillary reports display selective graphs and corresponding
results which plot any data verses any data that QDR can
measure. Data from the reference curve database is used for
graphing. Ancillary reports are previewed and/or printed in
addition to the selected reports (interpreting, filing, referring, and
rate of change) on one or more pages.
To specify the minimum age for displaying T-scores in both
Report and analysis results:
In the “Display T-Scores for patients above age” field, highlight
the number and enter a new one.

Configuring the System 18-11


QDR Reference Manual

If “Display ISCD Reports” check box is selected the default age


shall be set to 50.
Changing the default setting in this section will cause the
“Display ISCD Reports” checkbox to be unchecked.
To specify reference curves T-score will be based on:
In the “T-score based on” section, click Ethnicity matched or
Caucasian:
Caucasian marks the ISCD reference curves as current for use in
the report.
Ethnicity matched marks the classic Hologic curves as current for
use in the report.
If “Display ISCD Reports” check box is selected this section
defaults to Caucasian.
Changing the default setting in this section will cause the
“Display ISCD Reports” checkbox to be unchecked.

Report - WHO Criteria Tab


The Report - WHO Criteria dialog is used to specify which
reports will display the WHO criteria for each scan.
Automatically set to the following default values when the
“Display ISCD Reports” checkbox on the General tab is checked.
Hip checked
Report Based on Lowest T-score of neck or total checked
AP checked
Lateral unchecked
Forearm checked
Changing any setting on this dialog will cause the “Display ISCD
Reports” checkbox on the General tab to be unchecked.
To change setting(s):
1. Position insertion pointer in respective checkbox and click to
check or uncheck.
2. Click OK

Report - Least Significant Change Tab


The Report - Least Significant Change dialog is used to change
the values used when calculating the significant changes for
reporting rate of change results.

18-12 Configuring the System


QDR Reference Manual

If any values are changed from their default setting, and it is used
in the Rate of change report, the report will indicate “Site
Specific” LSC.
To change value(s):
1. Position insertion pointer in respective field, highlight the
existing value, and type new value.
2. To return all values to their default setting, Click Restore
Defaults.
3. Click OK.

Configuring the System 18-13


QDR Reference Manual

Configuring Reports
The Report - General tab allows the operator to configure all four
types of reports (Filing, Interpreting, Referring and Rate of
Change) as well as common data on all reports (references,
header and physician’s comments).
These are explained below.

Configuring the Reports


Depending upon the changes you want to make, configuring a
report type will consist of one, or more, of the following
procedures:
• Configure the Patient, Scan and Results Blocks
• Configure the Optional Reports
• Configure the Common Block

Configure the Patient, Scan and Results Blocks


The four report types (Filing, Interpreting, Referring and Rate of
Change) consist of various blocks which contain information that
the operator can include or exclude on reports. The reports consist
of four blocks of information: Patient, Scan, Results and Optional
Reports (the scan block does not apply to the Referring and Rate
of Change reports) as shown in the following table:

Blocks You Can Configure


Optional
Report Type Patient Scan Results Reports
Filing X X X X
Interpreting X X X X
Referring X X X
Rate of Change X X X

Instructions for configuring the Patient Block, Scan Block,


Results Block, and Optional Report Block are identical for all
types of reports listed below.

18-14 Configuring the System


QDR Reference Manual

To configure the blocks on a report type:


Step Action
1 Click on the AP Lumbar Spine in the scan type list.
The scan type is highlighted.

KP0886C-1707

2 Click on the Configure button


A window displays which contains four tabs: Patient
KP0886C-1708
Block; Scan block; Results Block and Optional Reports.
Note: This report is using the
Filing Report type as an example
but the procedure is the same for
all report types.
Note: For a detailed description of
the tabs refer to:
Details of the Patient Block on
page 18-23
Details of the Scan Block on
page 18-24
Details of the Results Block on
page 18-25

KP0886C-1709

3 Click on any of the items under Add to: to highlight the


item.
The << Remove button will highlight.

Configuring the System 18-15


QDR Reference Manual

Step Action
Note: Any Item in the Select 4 Click on the << Remove button.
From: box is excluded from the The item is removed from the Add To: box and placed in
report block.
the Select From: box. The Add >> button will highlight.
Note: If more than one item is in 5 Continue to remove items you want to exclude.
the Select From: box then the Add The items are removed from the Add To: box and placed in
All: button will highlight.
the Select from: box.
6 If you want to return any item to the Add To: box,
highlight the item and click on the Add >> button
or
If you want to return all items in the Select From: box
click the Add All >> button.
Note: The Scan Block does not 7 When finished with the Patient Block, click on the Scan
apply to the Referring and Rate of Block tab and repeat steps 3-6 above.
Change report types.
8 When finished with the Scan Block, click on the Results
Block tab and repeat steps 3-6 above.
9 If you have not completed the task of configuring the
report types go to Configure the Optional Reports on page
18-16 or Configure the Common Block on page 18-19.
or
If you have completed configuring the report types, click
on the OK button to save all your changes.
The System Configuration window displays.

Configure the Optional Reports


The Optional reports for all report types consist of the following,
their names describe their function:
• Combined AP/Lateral Report
• Extended AP Lumbar Spine Report
• Standardized BMD Report
• Body Composition Report
• Body Composition with BMD Report
• Hip Structure Analysis® Report

18-16 Configuring the System


QDR Reference Manual

To configure the Optional Reports of a report type:


Step Action
1 On the Report Tab, click on the Report type you want to
configure to place the dot in the radio button.

KP0886C-1707

2 Click on the Configure button


A window displays that contains four tabs: Patient Block;
KP0886C-1708
Scan block; Results Block and Optional Reports.
Note: This procedure uses the
Filing Report type as an example
but the procedure is the same for
all report types.

KP0886C-1709

Configuring the System 18-17


QDR Reference Manual

Step Action
3 Click on the Optional Reports tab.
The Optional Reports tab displays.

Note: The Configure Rate of Change - Optional Reports


tab provides only one optional report, the Extended Hip
Rate of Change Report.
4 Click on the reports you want.
A check mark appears in the box next to the report name.
5 If you have not completed the task of configuring the
report types go to Configure the Patient, Scan and Results
Blocks on page 18-14 or Configure the Common Block on
page 18-19.
or
If you have completed configuring the report types, click
on the OK button to save all your changes.
The System Configuration window displays.

18-18 Configuring the System


QDR Reference Manual

Configure the Common Block


To configure the Optional Reports of a report type:
Step Action
1 On the Report Tab, click on the Common block
configuration to place the dot in the radio button.

KP0886C-1711

2 Click on the Configure button.


The Configure Common block options window displays.
KP0886C-1708

KP0886C-1712

The Configure Common block option window consists of three


tabs or options:
• Reference Block
• Header Block
• Physician’s comments

Configuring the System 18-19


QDR Reference Manual

Each of these options can be configured separately or in any


combination.
To configure the Reference Block:
Step Action
Note: A detailed description of the 1 On the Configure Common block option window, click
Reference blocks can be found on the Reference Block tab.
starting on page 18-26.
The Reference Block window displays.

KP0886C-1713

2 Click on and highlight the part of the anatomy you want to


configure (hip, forearm or lateral).
3 Click on the Edit button.
An Edit window displays with various options for the part
KP0886C-1714 of the anatomy you want to configure.
4 Select the options you want to include in the report by
clicking on them and placing a dot in the radio button.
5 If you have not completed the task of configuring the
common areas go to Header Block on page 18-21 or
Physician’s comments on page 18-22.
or
If you have completed configuring the report types, click
on the OK button to save all your changes.
The System Configuration window displays.

18-20 Configuring the System


QDR Reference Manual

To configure the Header Block


Step Action
1 On the Configure Common block option window, click
on the Header Block tab.
The Header Block window displays.

KP0886C-1715

2 Complete each text field on this window as you want it to


appear on the header of the reports. All text fields are self
explanatory by their titles.
Note: If a check mark is not in this 3 After finishing all text fields, ensure that there is a check
box your system will not use the mark in the box next to Use this header information. If
information on this window in the
there is no check mark then click on the box and the
header of your reports.
symbol will appear.

KP0886C-1716

4 If you have not completed the task of configuring the


common areas go to Reference Block on page 18-20 or
Physician’s comments on page 18-22.
or
If you have completed configuring the report types, click
on the OK button to save all your changes.
The System Configuration window displays.

Configuring the System 18-21


QDR Reference Manual

To configure the Physician’s comments


Step Action
Note: The Physician’s comment 1 On the Configure Common block option window, click
block is used to create on the Physician’s comments tab.
predetermined comments that can
The Physician’s comment window displays.
be selected when generating the
report.

KP0886C-1717

2 The following are the different operations that can be used


on the list of items in the display window (represented by
the three button under the window):
Add. Add a new comment.
Delete. Delete the selected comment (an item must be
selected in the window).
Edit. Edit selected comment (an item must be selected in
the window).
You can add a new item that will become a predetermined
comment on the report; you can delete an item which will
eliminate it from the report; or you can edit a comment and
change it.
3 If you have not completed the task of configuring the
common areas go to Reference Block or Header Blocks.
or
If you have completed configuring the report types, click
on the OK button to save all your changes.
The System Configuration window displays.

18-22 Configuring the System


QDR Reference Manual

Details of the Patient Block


The Patient Block (Figure 18-4) deals with information on the
patient biography. By configuring information on the Patient
Block you can remove items of information from the report and
later add them back in.

Figure 18-4
The Patient Block

KP0886C-1718

The items listed in the window named Add To: are the items
printed on your report. Three buttons are provided to control
these items:
Adds a highlighted item in the Select From:
window to the Add To: window. This item
KP0886C-1719
is included in the printing of your report.
Removes an item from the Add To: window
and places it in the Select From: window.
KP0886C-1720
This item is excluded from printing on your
report.
Takes all items from the Select From:
window and moves them to the Add To:
KP0886C-1721
window.

The Patient Block consists of the following items:


Name
Age
Ethnicity
Menopause Age
Height

Configuring the System 18-23


QDR Reference Manual

Weight
Patient ID
Referring Physician
DOB
Sex

Details of the Scan Block


Note: The Scan Block does not The Scan Block (Figure 18-5) deals with information used when
apply to the Referring or Rate of taking the scan. By configuring information on the Scan Block
Change Reports.
you can remove items of information from the report and later
add them back in.
Figure 18-5
Scan Block

KP0886C-1722

The items listed in the window named Add To: are the items
printed on your report. Three buttons are provided to control
these items:

Adds a highlighted item in the Select From:


window to the Add To: window. This item
KP0886C-1719
is included in the printing of your report.
Removes an item from the Add To: window
and places it in the Select From: window.
KP0886C-1720
This item is excluded from printing on your
report.
Takes all items from the Select From:
window and moves them to the Add To:
KP0886C-1721
window.

18-24 Configuring the System


QDR Reference Manual

The Scan Block consists of the following items:


Analysis
Comment
Model
Operator
Protocol
Scan Date
Scan Type

Details of the Results Block


The Results Block (Figure 18-6) deals with information on the
result block of the report. Configuring information on the Results
Block you can remove items of information from the report and
later add them back in.

Figure 18-6
The Results Block

KP0886C-1723

The items listed in the window named Add To: are the items
printed on your report. Three buttons are provided to control
these items:

Configuring the System 18-25


QDR Reference Manual

Adds a highlighted item in the Select From:


window to the Add To: window. This item
KP0886C-1719
is included in the printing of your report.
Removes an item from the Add To: window
and places it in the Select From: window.
KP0886C-1720
This item is excluded from printing on your
report.
Takes all items from the Select From:
window and moves them to the Add To:
KP0886C-1721
window.
The Results Block consists of the following items:
Region
BMC
BMD
Fat
%Fat
Lean+BMC
Lean
T-Score
Z-Score
Fracture Risk
Area

Details of the Reference Block for Configure Common Block Option


The Reference Block (Figure 18-7) is used for setting up the
default region and range for the different scan types. This block is
common to all report types.
Figure 18-7
The Reference Block

KP0886C-1724

18-26 Configuring the System


QDR Reference Manual

The Reference Block allows you to chose the default region and
range on the Hip, Forearm and AP/Lateral scans.

Reference Block Hip Configuration


Highlighting Hip in the Reference Block window and clicking on
the Edit button displays the following Edit window:

The Edit window for Hip contains two sections. The Reference
Region section designates a reference region for all reports. The
Reported Regions selects one or more regions for reporting.
For ISCD compatible reports the default settings for the Hip
Reference Region is Neck (radio button contains a dot) and the
Reported Regions are Total and Neck (both check boxes contain
check marks). These default values are set by checking “Display
ISCD Reports” on the Reports - General tab dialog.
You can change these default by clicking on any of the other
items in the respective section. Selecting a Reference Region
automatically disables it in the Reported Regions section.
Changing any setting from its default value unchecks “Display
ISCD Reports” on the Reports - General tab dialog.
Click the OK button when finished to exit and save the changes,
or the Cancel button to exit without saving the changes.

Configuring the System 18-27


QDR Reference Manual

Reference Block Forearm Configuration


Highlighting Forearm in the Reference Block window and
clicking on the Edit button will display the following Edit
window:

For ISCD compatible reports the default setting for the Forearm
Region is 1/3 and the Range is Radius (the radio buttons are
marked with a dot). These default values are set by checking
“Display ISCD Reports” on the Reports - General tab dialog.
You can change this default by clicking on any of the other items
on the edit window (a dot will appear in that radio button).
Changing any setting from its default value unchecks “Display
ISCD Reports” on the Reports - General tab dialog.
Click the OK button when finished to exit and save the changes,
or the Cancel button to exit without saving the changes.

18-28 Configuring the System


QDR Reference Manual

Reference Block Lateral Configuration


Highlighting Lateral in the Reference Block window and clicking
on the Edit button will display the following Edit window:

KP0886C-1727

The default setting for the Lateral is VB (the radio button is


marked with a dot). You can change this default by clicking on
any of the other items on the edit window (a dot will appear in
that radio button).
Click the OK button when finished to exit and save the changes
or the Cancel button to exit without saving the changes.

Configuring the System 18-29


QDR Reference Manual

Configuring the Ancillary Report


Configuring the Ancillary Report consists of selecting the
graph(s) to include in the report:

Step Action
1 On the System Configuration - Report tab, check the
“Display Ancillary Report” check box and click
Configure.

The Configure Ancillary Report window displays.

Note: Refer to “Reference Curve” 2 Check the Include check box for the graphs to include in
on page 19-42 for details on the report, and uncheck the check box for the graphs that
adding additional custom values
are not to be included.
for reporting.

18-30 Configuring the System


QDR Reference Manual

Archive Tab
The Archive Tab under System configuration allows you to setup
locations to store information when you use the archive menu
item. You can set up various locations with one being the primary,
or default, location that is used automatically when data is
archived.
This default location is also used by the Locate Menu on the Main
Screen. When Locate is selected, the location displayed is the
default location chosen when you configured Archive.
The default location is shown on the Archive Tab. You can
change the default location by adding a new location or by
selecting a location on the list on the Archive Tab and setting it as
the default.

Adding a New Location


To add a new location to the archive list:
Step Action
1 On the Archive Tab, type a new location in the Add new
Location text box.
or
Press the Browse button to the right of the text window and
a window displays showing locations on your computer
and network (if one is present). Select a location and press
the OK button to display the location in the text window.

KP0886C-1729

Configuring the System 18-31


QDR Reference Manual

Step Action
2 Click on the Add button next to the text window.
A message displays asking if you want to make the new
location the default location.

KP0886C-1730

If you click the OK button then the new location is


designated as the new default location and shown as such
on the Archive Tab, and the location is added to the archive
list. If you click the No button the location is added to the
archive list.

Changing the Default Location


Use the following procedure to change the default location to the
archive list of locations:
To change the default location
Step Action
1 In the list of locations in the window on the Archive tab,
click on the location you want as the new default.
The location is highlighted.
2 Click on Set Default Location button.
The highlighted location is designated as the new default
and the location is placed in the Default Location area on
the Archive Tab.

Deleting a Location from the Archive List


Use the following procedure to delete a location from the achieve
list:
To delete a location from the archive list:
Step Action
1 In list of locations in the window on the Archive tab, click
on the location you want to delete.
The location is highlighted.

18-32 Configuring the System


QDR Reference Manual

Step Action
Note: If the location to be deleted 2 Click on Delete button.
is the default location a message The highlighted location is deleted from the archive list.
appears informing the operator.

Including Raw Data Files


When this option is checked, raw data files will be included when
transferring scans. Raw data files should not be transferred with
the scan files unless you have been instructed to do so by
Hologic, Inc.

Automatically Deleting Scans After Archiving


In the Delete Scans After Archiving section:
Click the Do Not Delete Scans button to disable automatically
deletion of scans after archiving.
Or
Click the Delete All Scans or the Delete Only Non-Baseline
Scans button to automatically delete the respective scans after
archiving, and then click the Delete After Archiving Once or the
Delete After Archiving Twice button to specify the number of
times the scans are to be archive before they are deleted.

Configuring the System 18-33


QDR Reference Manual

HL7 Tab
The HL7 Tab allows the user to enter additional fields of
information for tracking patient accounts for the Hospital
Information System (HIS). The HL7 Tab allows you to place up
to three additional fields that appear on the Patient Confirmation
window (see Refer to “Confirming the Patient Information” on
page 7-6 of Chapter 7).
The information entered in these fields will be passed back to the
HIS through the Physician’s Viewer (see the Physician’s Viewer
manual for additional information).

Entering HL7 Fields


Use the following procedure to enable/disable displaying of HL7
fields:
To enter a HL7 field:
Step Action
1 From the menu bar select Utilities → System
Configuration and click on the HL7 Tab.
2 On the HL7 Tab, click on the box to the left of Display
Label 1.
Box

KP1035A-1732

A check mark appears in the box and the area to the right
of Display Label 1 turns from gray to white.

18-34 Configuring the System


QDR Reference Manual

Step Action
3 In the white area to the right of Display Label 1 type in
the desired name of the label.
The label “Billing Code” is used as an example below.

KP1035A-1733

Note: If you click OK without 4 Click OK.


entering a label the software The system returns to the Main window.
removes the 3 and no HL7 field
will appear on the Patient
Confirmation window.
5 To confirm that the HL7 label was created, click the
Perform Exam button.
Select any patient and click OK.
KP1035A-0309
The Patient Confirmation window appears with the HL7
field enabled (in this example “Billing Code”).

Configuring the System 18-35


QDR Reference Manual

Changing HL7 Fields


To Change a HL7 field:
Step Action
1 From the menu bar select Utilities → System
Configuration and click on the HL7 Tab.
2 In the white area to the right of Display Label 1 type in
the name of the new label.
In the example below “Insurance Code” replaces
“Billing Code.”

KP1035A-1735

3 Click OK and the following window appears.

KP1035A-1736

4 Click Yes to return to the main window.

18-36 Configuring the System


QDR Reference Manual

Step Action
5 To confirm that the new label is enabled, click the
Perform Exam button, select a patient and the Patient
Confirmation window appears with the new label.

Removing HL7 Fields


To Remove a HL7 field:
Step Action
1 From the menu bar select Utilities → System
Configuration and click on the HL7 Tab.
2 In the white area to the right of Display Label 1 remove
the label from the field.

KP1035A-1738

Configuring the System 18-37


QDR Reference Manual

Step Action
3 On the HL7 Tab, click on the box to the left of Display
Label 1.

KP1035A-1739

The check mark is removed and the box to the right of


Display Label 1 turns from white to gray.
4 Click OK and the following window appears.

KP1035A-1740

5 To confirm that the new label has been removed, click the
Perform Exam button, select a patient and the Patient
Confirmation window appears with the label removed.

18-38 Configuring the System


QDR Reference Manual

Scan Conversion
The Scan Conversion tab allows the QDR system to except a scan
acquired on another manufacturer’s system. Once the data
(coefficients) for the other manufacturer’s scan are entered into
the database, the QDR will automatically accept scans from that
manufacturer.

The QDR system contains the Scan Conversion for Hologic,


Norland and Lunar scans (as shown in the Manufacturer drop
down list).

KP1068_002-1808

When any of these manufacturers (including Hologic) are


selected, their coefficients appear in the Input Coefficients boxes
for the spine and hip. The Input Coefficients are grayed out and
cannot be changed. The values for the Input Coefficients were
derived from the following studies:
• Hui SL, Gao S, et al 1997 Universal Standardization of
Bone Density Measurements: A Method with Optimal
Properties for Calibration Among Several Instruments. J
Bone Miner Res 9:1463-1470.
• Lu Y, Fuerst T September 6, 1996 Summary of Femur
Standardization: Generation of Formulas for Standardized
BMD (sBMD) and Inter-Manufacturer Cross-Calibration.
(A report from the UCSF Osteoporosis Research Group).
These results were confirmed by the University of
Indiana. Presented to the Subcommittee for standardizing
hip DXA, ASBMR Annual Meeting, Seattle, WA, Sept. 7,
1995.
• Genant HK, Grampp S, et al 1994 Universal
Standardization for Dual X-ray Absorptiometry: Patient
and Phantom Cross-Calibration Results. J Bone Miner
Res:9:1503-1514.
The system manager can add additional manufacturers to the drop
down list. To do so they must have access to a study or studies

Configuring the System 18-39


QDR Reference Manual

that provide the Input Coefficients for that manufacturer versus


Hologic.
To Add Additional Manufacturers
Step Action
1 Enter the name of the manufacturer in the Manufacturer
window.

KP1068_002-1806

Note: All “m” values must be 2 Within the Input Coefficients section add the value of m
within the range of 0.5 to 2.0. for the Spine Total.

KP1068_002-1807

Note: All “b” values must be 3 Within the Input Coefficients section and the value of b
within the range of -0.9 to +0.9. for the Spine Total.
4 Within the Input Coefficients section add the value of m
for the Hip Total.
5 Within the Input Coefficients section add the value of b
for the Hip Total.
6 Within the Input Coefficients section add the value of m
for the Neck.
7 Within the Input Coefficients section add the value of b
for the Neck.
8 Within the Input Coefficients section add the value of m
for the Trochanter.
9 Within the Input Coefficients section add the value of b
for the Trochanter.
10 When all coefficients have been entered, click OK.
The new manufacturer will appear in the drop down menu
with data entered into the Input Coefficients.

18-40 Configuring the System


QDR Reference Manual

History
This tab appears if the Mobility option is installed.
The History tab of the System Configuration window allows the
user to configure the patient questionnaire.
The following describes the fields for configuring the History
Questionnaire.

Default Questionnaire Template

MAN-00175_003-1834

This is an edit box for the location of the Default Questionnaire


Template. This box is filled in at install and points to the template.
The default is C\QDR\HxQuestionnaire\template\HologicHx.xml
(located under the install directory).
Clicking Browse button and a window opens that allows the
selection of a path to another questionnaire template on your
system or network.

MAN-00175_003-1834

Configuring the System 18-41


QDR Reference Manual

Location of Completed Questionnaire

MAN-00175_003-1836

This box provides an edit box for the location of completed


Questionnaires. This is filled in at install time and points to the
location to store Questionnaires that are filled out. The default is
C\install directory\HxQuestionnaire\Results (located under the
QDR directory).
Clicking the Browse button and a window appears that allows
selection of a path to another location to store the completed
questionnaires.

MAN-00175_003-1837

Port

MAN-00175_003-1838

This is an edit box for the communication Port. This port


communicates with the Questionnaire Application and accepts
numeric data only. The default is 106.

18-42 Configuring the System


QDR Reference Manual

Print Patient Name and ID

MAN-00175_003-1837

This is a checkbox that prints the patient name and ID on multiple


page reports. A checkmark in the box prints the Patient Name and
ID on pages 2 until the last page of the questionnaire (this is the
default). Clicking in this area will remove the checkmark and the
Patient Name and ID will not print.

Configuring the System 18-43


QDR Reference Manual

DICOM Tab
Digital Imaging and Communications in Medicine (DICOM) is a
powerful tool that provides:
• Interpreting physicians with the ability to view electronic
QDR bone density scan and analysis results on a Picture
Archiving and Communications System (PACS) viewer. The
DICOM option allows results to be transmitted automatically
over a facility’s network directly to a physician’s DICOM
viewing station for interpretation and report dictation. The
results can also be archived on the PACS, making them avail-
able for future reference and for distribution to others on the
PACS network.
• The QDR system with the ability to retrieve schedule and
patient demographic information when the Modality Worklist
option is installed on the system.
• Locating and retrieving of scans that have been previously
archived to a remote storage system (PACS) when the Query/
Retrieve option is installed on the system.

The following sections describe Settings for DICOM functions are controlled using the System
how to configure the Modality Configuration – DICOM tab found under the Utilities pull
Worklist; add, edit, and delete
down menu in the main window.
DICOM Report Send remote
destinations; add, edit, and delete The DICOM tab contains four tabs used to configure:
DICOM Report Storage • Modality Worklist option (when installed)
Commitment remote destinations;
• DICOM report send destinations (when installed)
add, edit, and delete Query/
Retrieve remote destinations; and • DICOM report storage commitment destinations (when
configure the Host Machine (local installed)
system). • Query/Retrieve option (when installed)
• Host machine

Modality Worklist
The modality worklist option adds two tabs to the APEX
software:
• A Worklist tab is added to the System Configuration -
DICOM window to allow the configuration of the
modality worklist.
• A Worklist tab is added to the Select Patient for Exam
window to allow the operator to receive schedules from
the HIS/RIS to perform tasks on the QDR system.

18-44 Configuring the System


QDR Reference Manual

Configuring Modality Worklist

Caution: Changing information Modality Worklist is configured by selecting System


that configures the modality Configuration – DICOM tab – Worklist tab found under the
worklist can cause serious
Utilities pull down menu in the main window.
communication disruptions with
the HIS/RIS. Only authorized
personnel should change settings.

The Worklist tab is divided into 8 areas used to control


communication to and from the HIS/RIS and one area providing
an option for obtaining worklist data from an input file rather
than from a remote worklist provider.
• Query Parameters
• Auto Query Interval
• Query Retry Parameters
• Purge Interval
• Input from File
• Enable Detailed Query
• Configure Worklist Provider
• Map Worklist Fields
• Confirm network and PACS connection
Each of these areas is explained below.

Configuring the System 18-45


QDR Reference Manual

Query Parameters Use controls in this area to perform the following:


• Filter the query by modality and AE Title.
• Determine a limit of entries to the Worklist database over a specified
period of time.
• Determine if “Detailed Query” and “Extended Details” buttons will
appear in the Select a Patient for this Exam - Worklist dialog.
Days Back and This provides the valid time range of the query. There are a pair of drop
Forward down menus labeled Back and Forward. If both of these are set to 0
(zero) then the valid range is for today’s date only. If the Back drop-
down menu is set to 7 and the Forward drop-down menu is set to 0
(zero) then the valid range of the query is one week, starting with today’s
date going back seven (7) days. The range for the drop-down menu for
Back is 0 to 9 and the range for the drop-down menu for Forward is 0 to
8. Both Back and Forward drop-down menus can manually be set from
0 to 99.
Modality Modality is the type of system recognized by the HIS/RIS. Since “bone
densitometer” is not a standard defined type, the QDR System is usually
defined as “OT” for “Other.”
AE Title AE stands for Application Entity. This is a text entry box that provides a
unique name for the QDR system. Every QDR system will (or should)
have an AE name that uniquely identifies that specific system.
Maximum Hits This is an entry box that accepts a numeric value only. The number is the
Per Query maximum number of query hits that will be passed to the QDR system
specified in the Days Back and Forward. If there are more hits than the
maximum, only the number specified in this box are passed to the QDR
system.
Enable Detailed A check box determining if an “Detailed Query” button and an
Query “Extended Details” button will appear in the Select a Patient for this
Exam - Worklist dialog when performing an exam.
When checked both buttons will appear.

18-46 Configuring the System


QDR Reference Manual

Auto Query Use controls in this area to set up a specific time interval during which
Interval the QDR system queries the provider to update the worklist.
Auto Query Interval area contains three radio buttons, only one of which
may be selected.
Note: These controls remain enabled when Input From file is enabled.
Every Day At This entry provides two drop-down menus labeled HR and Min that
specify a specific time each day that the QDR system will query the
provider to update the worklist. The HR drop-down menu has a range of
0 to 23. The Min drop-down menu can be set to 0, 10, 20, 30, 40 or 50.
Every This entry provides a drop-down menu labeled Hours that specifies an
hourly interval at which the QDR system will query the provider to
update the worklist. The drop-down menu has a range of 1 to 12 (manual
entry is also possible).
Never If Never is selected, the QDR system will not automatically query the
provider to update the worklist. With this selected, queries must be done
manually by the operator.
Query Retry If the provider did not respond to a query to update the worklist for some
Parameters reason (e.g., busy, off-line), controls in this area will determine how long
the QDR system will wait for a response and give a specific time before
trying the query again.
This entry contains a checkbox and three drop-down menus that control
how long the QDR system will wait for the provider to respond to a
query.
Note: These controls remain enabled when Input From file is enabled.
Query Retry In order for the QDR system to perform a retry after a time-out period,
Checkbox they must be a check mark in this box. If there is no check mark in this
box then the QDR system will continue to wait until the provider
responds to the query to update the worklist. To place a check mark in
the box, click in the box.
Query Timeout This is a drop-down menu labeled Min. The value in this drop-down
menu tells the QDR system how long it can wait before retrying the
query. Min has a range of 0 to 60 minutes in 5 minute intervals. The
operator can enter a number from 0-99 manually.
Retry Number This is a drop-down menu labeled Times that tells the QDR system how
many times it may issue a retry. This drop-down menu allows a selection
of 0, 1, 2, 3, 4, 5, 6, 7, 8, 9 or 10 for the number of retries.The operator
can enter a number from 0-99 manually.

Configuring the System 18-47


QDR Reference Manual

Retry Interval This is a drop-down menu labeled Min that tells the QDR system how
long to wait between retries.This drop-down menu has a range of 10 to
90 minutes in 10 minute intervals.The operator can enter a number from
0-99 manually.
Purge Interval Each time the provider responds to a query from the QDR system the
worklist entries are stored in a database on the QDR system. Use
controls in this area to allow the database to be automatically purged
(data removed) at a specific time.
Note: These controls remain enabled when Input From file is enabled.
Used Entries This is a drop-down menu labeled Days. The value in this drop-down
menu provides a limit for storing those studies that have already been
performed. The studies are purged (deleted) after the specified number
of days. The range for Used Entries drop-down menu is 0 to 9 days. The
Used Entries drop-down menu can be set manually to any number of
days from 0 to 999.
Unused Entries This is a drop-down menu labeled Days. The value in this drop-down
menu provides a limit for storing those studies that have not been
performed. These are purged (deleted) after the specified number of
days. The range for the drop-down menu for Unused Entries is 0 to 9
days. The Unused Entries drop-down menu can be set manually to any
number of days from 0 to 999.
Input From File Provides an option for obtaining worklist data from an input file
generated by an electronic Medical Reporting System rather than from a
remote Worklist provider.
Enable A check box determining whether or not the Input from File feature is
enabled. When checked, worklist data is obtained from an input file.
All controls on the Worklist tab not used for the Input from File feature
are disabled when this box is checked.
Input File Name Displays a full path to the current worklist file. This field is populated or
changed using Browse to select a file path.
... (Browse) Displays a “File Open” dialog allowing the user to locate a worklist
input file on the local system or network.
Configre Worklist Click Configure Worklist Provider to define the worklist provider. The
Provider worklist provider supplies worklist entries for the QDR system.
Refer to “Configure Worklist Provider” on page 18-49 for details.

18-48 Configuring the System


QDR Reference Manual

Map Worklist Different hospitals and clinics may use the same DICOM attributes in
Fields various ways to identify their patients. Map Worklist Fields is used to
ensure that data in the QDR system and in the HIS/RIS are addressing
the same patient.
Click Map Worklist Fields, on the Worklist Tab, to display a window
called Map Worklist Keys that allows fields to be selected for patient
verification by the QDR system and the HIS/RIS database. Once these
entries are made, they will be checked on each study to verify that the
QDR system and the HIS/RIS have identified the same patient.
Refer to “Map Worklist Fields” on page 18-51 for details.
Local Ping Confirms if the local system is connected to a network.

Configure Worklist Provider


Clicking Configure Worklist Provider displays a window used
to define the worklist provider. The worklist provider supplies
worklist entries for the QDR system.

Worklist Provider Use controls in this area to define the worklist provider.
AE Title This is the Application Entity title of the worklist provider.
Remote Host This is the host name or IP address of the worklist provider.
The host name or IP address must be on the same network as the QDR
workstation.
Accepts up to 120 alphanumeric characters.
Remote Port This is the port number of the worklist provider.
Number
Performed This area provides an option to automatically respond to a performed
Procedure Step procedure provider once a specific task is performed in the worklist.
Use Performed When checked, each time a study is completed a response is sent to the
Procedure Step provider indicating that the task has been completed. To place a check
mark in the box, click in the box.
Provider Use controls in this area to define the performed procedure step provider.
The performed procedure provider may be the same as the worklist
provider or different.
Use Worklist When checked, the performed procedure step provider is the same as the
Provider Data worklist provider.

Configuring the System 18-49


QDR Reference Manual

AE Title This is the Application Entity title of the performed procedure step
provider when Use Worklist Provider Data is unchecked.
Remote Host This is the host name or IP address of the performed procedure step
provider when Use Worklist Provider Data is unchecked.
The host name or IP address must be on the same network as the QDR
workstation.
Accepts up to 120 alphanumeric characters.
Remote Port This is the port number of the performed procedure step provider when
Number Use Worklist Provider Data is unchecked.
Remote Ping Confirms if the configured Worklist provider or Performed Procedure
Step provider system is connected to the same network as the local
system.
C-ECHO Confirms if the configured Worklist provider or Performed Procedure
Step provider system is a Picture Archival and Communications system
(PACS).

18-50 Configuring the System


QDR Reference Manual

Map Worklist Fields


Clicking Map Worklist Fields on the displays the following
window:
Caution: Do not change settings
in the Map Worklist Fields
without specific directions from
your HIS/RIS department.

KP1043A-11

This window consist of 15 different drop-down menus with


specific labels on the left. The labels indicate data that is in the
Patient Biography on the QDR systems (some information, such
as HL7 fields, may not apply to each QDR system). The
information in the drop-down menus is information that may
appear in the HIS/RIS database for the patient. The goal is to find
information in the Patient Biography that matches information in
the HIS/RIS database to be used as a key to verify that the patient
in the Patient Biography is the same as the patient in the HIS/RIS
database.
This task of matching information must be performed by
someone with knowledge of both the QDR system and the
operations of the HIS/RIS.
There are two sets of drop-down menus:

Configuring the System 18-51


QDR Reference Manual

• Patient Matching Keys - six drop-down menus for


Patient Matching
• Other Keys - nine drop-down menus for Key Mapping
The Patient Matching drop-down menus allow specific data in the
Patient Biography of the QDR system to be mapped to data in the
HIS/RIS database. These drop-down menus have check boxes to
the left of the label. A check mark in any of these boxes means
that the information in the Patient Biography must match the
information in the HIS/RIS database to identify the same patient.
The Key Mapping drop-down menus provide data from the HIS/
RIS database that can be mapped to specific information in the
Patient Biography.
“Not Mapped” in any drop-down menu indicates that the field in
the Patient Biography will not be mapped with any field in the
HIS/RIS database.

18-52 Configuring the System


QDR Reference Manual

DICOM Send Destinations


Send destinations define destinations to which the DICOM
reports are transmitted when the Send function is used.

This section describes how to Send destinations are configured by selecting System
configure, add, edit, and delete a Configuration – DICOM tab – Send tab found under the
destination remote nodes and how
Utilities pull down menu in the main window.
to configure all Send destinations.

Configure DICOM Selection list of existing DICOM Send destinations configured on the
Send Destinations local system.
Add Destination Displays a dialog that allows configuring a new Send destination
Edit Destination Enabled when one destination is selected.
Displays a dialog that allows changing the selected Send destination’s
configuration.
Delete Destination Enabled when one or more destinations are selected.
Deletes the selected Send destination(s).
Local Ping Confirms whether or not the local system is connected to a network.
Remote Ping Enabled when one destination is selected.
Confirms whether or not the configured Send destination system is
connected to the same network as the local system.
C-ECHO Enabled when one destination is selected.
Confirms whether or not the configured Send destination system is a
Picture Archival and Communications system (PACS).
Configure Click Configure Parameters to define the Send destination auto retry
Parameters parameters and when to purge log entries.
Refer to “Configure DICOM Send” on page 18-57 for details.

Configuring the System 18-53


QDR Reference Manual

Using the System Configuration - DICOM - Send tab:


Step Action
1 To add a new destination: click Add Destination,
complete the Add / Edit DICOM Send Destination dialog
fields (refer to “Add / Edit DICOM Send Destination
Dialog” on page 18-55 for details), and click OK.
To reconfigure an existing destination: select the
destination in the selection list, click Edit Destination,
edit the Add / Edit DICOM Send Destination dialog box
as required (refer to “Add / Edit DICOM Send
Destination Dialog” on page 18-55 for details), and click
OK.
To enable/disable a destination for use: click the check
box in the Active column of the selection list for the
destination to check/uncheck (check to enable).
To verify the local system is connected to a local
network: click Local Ping. Click OK to close the
resulting message.
To verify the destination is connected to the same
network as the local system: select the destination in the
selection list and click Remote Ping. Click OK to close
the resulting message.
To verify the destination is a PACS: select the destination
in the selection list and click C-ECHO. Click OK to
close the resulting message.
To delete a destination: select the destination in the
selection list and click Delete Destination.
To define auto query parameters for all configured Send
destinations: click Configure Parameters, edit the
Configure DICOM Send dialog box as required (refer to
“Configure DICOM Send” on page 18-57 for details) and
click OK.
2 Click OK to return to the main screen.

18-54 Configuring the System


QDR Reference Manual

Add / Edit DICOM Send Destination Dialog

AE Title Application Entity. Provides a unique name for the destination system.
Accepts up to 16 alphanumeric characters.
Note: To add the QDR workstation as a destination node, type local in
the AE Title box.
Host Name or IP Name or IP address of destination.
Address The host name or IP address must be on the same network as the QDR
workstation.
Accepts up to 120 alphanumeric characters.
Note: To add the QDR workstation as a destination node, type localhost
in the Host Name or IP Address box.
SCP Port Port number on destination.
The default for the Service Class Provider Port number is 104.
Accepts up to 5 numeric characters in the range of 1 - 65535.
Destination Name Provides an alias name used for selecting a destination.
Accepts up to 120 alphanumeric characters.
Interpreting An optional entry that provides the name of the physician interpreting
Physician the scan.
Accepts up to 120 characters.

Configuring the System 18-55


QDR Reference Manual

Grayscale Only When checked, DICOM reports are converted into grayscale when
sending to the selected destination.
When unchecked, images for all report types that may contain graphs
will be created in color. Images for all other report types (IVA,
questionnaire, etc) will be created in grayscale.
Presentation File When checked, a DICOM GSPS file will be sent along with an image
DICOM file if needed for IVA scan.
If checked and the node is also marked for storage commitment, then the
commitment request will be issued for both image and GSPS DICOM
files.
IVA Results File When checked, a DICOM IVA Results file will be sent along with an
image DICOM file if needed for IVA scan.
If checked and the node is also marked for storage commitment, then the
commitment request will be issued for both image and IVA Results
DICOM files.
Unicode When checked, DICOM files are sent with Unicode coding.
When using Unicode, an Extended Character Set attribute will be
present in DICOM files. Text attributes will be populated with single
byte strings converted from double byte strings using UTF-8 encoding.
When not using Unicode, an Extended Character Set attribute will not be
present in DICOM files. Text attributes will be populated with single
byte strings converted from double byte strings using UTF-8 encoding.
Storage When checked, designates the destination is committed to storing
Commitment information sent to it.
Provider When checked a storage commitment provider must be designated. To
designate a storage commitment provider, select an existing provider
from the drop-down list, or click Add New Provider to designate add a
new storage commitment provider (refer to “Add / Edit Storage
Commitment Destination Dialog” on page 18-61 for details on adding a
new provider). Storage commitment providers can also be added using
the System Configuration - DICOM tab - Commit tab (refer to “DICOM
Storage Commitment Destinations” on page 18-59 for details).

18-56 Configuring the System


QDR Reference Manual

Scan Archive Displays if the Enterprise Data Management option is installed.


Location When checked, designates the destination is a scan archiving location
and will implement the inclusion of P&R files to the DICOM file.
P files are QDR scan files that contain processing data for the scan.
R files are QDR scan files that contain the raw data for the image.
When checked, the Select Existing Provider field can be left blank to
indicate the Send destination is also the scan archive location. Or, an
existing scan archiving location can be selected from the drop-down list,
or click Add New Provider to designate add a new scan archive location.
Scan archive locations can also be added using the System
Configuration - DICOM tab - Query/Retrieve tab (refer to “DICOM
Query/Retrieve Destinations” on page 18-63 for details).
OK Validates data. If validation passes, the dialog is closed and the configure
DICOM Send Destinations selection list is updated with new or changed
data. If validation fails, a warning message displays.
Cancel Ignores all edits and closes the dialog.

Configure DICOM Send

Query Retry If the Send destination did not respond to a request for some reason (e.g.,
Parameters busy, off-line), controls in this area will determine how long the QDR
system will wait for a response and give a specific time before trying the
query again.
Retry Number This is a entry box labeled Times that tells the QDR system how many
times it may issue a retry. The Retry Number entry box can be set
manually to any number of days from 0 to 99.

Configuring the System 18-57


QDR Reference Manual

Retry Interval This is a entry box labeled Min that tells the QDR system how long to
wait between retries.The Retry Interval entry box can be set manually
to any number of minutes from 1 to 1440.
Purge log entries This is a entry box labeled days. The value in this box provides a limit
after for storing DICOM Send log entries. The entries are purged (deleted)
after the specified number of days. The Purge log entries after entry
box can be set manually to any number of days from 0 to 99.

18-58 Configuring the System


QDR Reference Manual

DICOM Storage Commitment Destinations


Storage commitment destinations define destinations to which the
DICOM reports can be transmitted and stored when the Send
function is used.
Storage Commitment destinations must be defined before
defining DICOM Send destinations as storage commitment
destinations.

This section describes how to Send destinations are configured by selecting System
configure, add, edit, and delete a Configuration – DICOM tab – Commit tab found under the
storage commitment destination
Utilities pull down menu in the main window.
remote nodes and how to
configure all Storage Commitment
destinations.

Configure DICOM Selection list of existing DICOM Storage Commitment destinations


Send Destinations configured on the local system.
Add Destination Displays a dialog that allows configuring a new Storage Commitment
destination
Edit Destination Enabled when one destination is selected.
Displays a dialog that allows changing the selected Storage Commitment
destination’s configuration.
Delete Destination Enabled when one or more destinations are selected.
Deletes the selected Storage Commitment destination(s).
Local Ping Confirms whether or not the local system is connected to a network.
Remote Ping Enabled when one destination is selected.
Confirms whether or not the configured Storage Commitment
destination is connected to the same network as the local system.
C-ECHO Enabled when one destination is selected.
Confirms whether or not the configured Storage Commitment
destination is a Picture Archival and Communications system (PACS).
Configure Click Configure Parameters to define the Storage Commitment
Parameters destination auto retry parameters, when to purge log entries, and how
DICOM storage commitment files are sent to a destination.
Refer to “Configure Storage Commitment” on page 18-62 for details.

Configuring the System 18-59


QDR Reference Manual

Using the System Configuration - DICOM - Commit tab:


Step Action
1 To add a new destination: click Add Destination,
complete the Add / Edit Storage Commitment dialog
fields (refer to “Add / Edit Storage Commitment
Destination Dialog” on page 18-61 for details), and click
OK.
To reconfigure an existing destination: select the
destination in the selection list, click Edit Destination,
edit the Add / Edit Storage Commitment dialog box as
required (refer to “Add / Edit Storage Commitment
Destination Dialog” on page 18-61 for details), and click
OK.
To enable/disable a destination for use: click the check
box in the Active column of the selection list for the
destination to check/uncheck (check to enable).
To verify the local system is connected to a local
network: click Local Ping. Click OK to close the
resulting message.
To verify the destination is connected to the same
network as the local system: select the destination in the
selection list and click Remote Ping. Click OK to close
the resulting message.
To verify the destination is a PACS: select the destination
in the selection list and click C-ECHO. Click OK to
close the resulting message.
To delete a destination: select the destination in the
selection list and click Delete Destination.
To define auto query parameters for all configured
Storage Commitment destinations: click Configure
Parameters, edit the Configure Storage Commitment
dialog box as required (refer to “Configure Storage
Commitment” on page 18-62 for details) and click OK.
2 Click OK to return to the main screen.

18-60 Configuring the System


QDR Reference Manual

Add / Edit Storage Commitment Destination Dialog

AE Title Application Entity. Provides a unique name for the destination system.
Accepts up to 16 alphanumeric characters.
Note: To add the QDR workstation as a destination node, type local in
the AE Title box.
Host Name or IP Name or IP address of destination.
Address The host name or IP address must be on the same network as the QDR
workstation.
Accepts up to 120 alphanumeric characters.
Note: To add the QDR workstation as a destination node, type localhost
in the Host Name or IP Address box.
SCP Port Port number on destination.
The default for the Service Class Provider Port number is 104.
Accepts up to 5 numeric characters in the range of 1 - 65535.
Destination Name Provides an alias name used for selecting destinations.
Accepts up to 120 alphanumeric characters.
OK Validates data. If validation passes, the dialog is closed and the configure
DICOM Storage Commitment Destinations selection list is updated with
new or changed data. If validation fails, a warning message displays.
Cancel Ignores all edits and closes the dialog.

Configuring the System 18-61


QDR Reference Manual

Configure Storage Commitment

Query Retry If the Storage Commitment destination did not respond to a request for
Parameters some reason (e.g., busy, off-line), controls in this area will determine
how long the QDR system will wait for a response and give a specific
time before trying the query again.
Retry Number This entry box labeled Times tells the QDR system how many times it
may issue a retry. The Retry Number entry box can be set manually to
any number of days from 0 to 99.
Retry Interval This entry box labeled Min tells the QDR system how long to wait
between retries.The Retry Interval entry box can be set manually to any
number of minutes from 1 to 1440.
Mode DICOM files can be sent to a storage commitment destination as one file
or all files per request.
Image-by-image This radio button, when selected, tells the QDR system to issue a single
storage commitment request for all DICOM files to be sent.
Batch This radio button, when selected, tells the QDR system to issue a storage
commitment request for each DICOM file to be sent.
Purge log entries This is a entry box labeled days. The value in this box provides a limit
after for storing DICOM Send log entries. The entries are purged (deleted)
after the specified number of days. The Purge log entries after entry
box can be set manually to any number of days from 0 to 99.

18-62 Configuring the System


QDR Reference Manual

DICOM Query/Retrieve Destinations


Query/Retrieve allows the operator to query a remote location
(PACS) for scans meeting given parameters and filters and to
retrieve selective scans into the current computer. The scans must
have been stored to that remote location prior to using the Query/
Retrieve function.

This section describes how to Query/Retrieve destinations are configured by selecting System
configure, add, edit, and delete a Configuration – DICOM tab – Query/Retrieve tab found
destination remote node.
under the Utilities pull down menu in the main window.

Configure DICOM Selection list of existing DICOM Query/Retrieve destinations


Query/Retrieve configured on the QDR system.
Destinations
Add Destination Displays a dialog that allows configuring a new Query/Retrieve
destination
Edit Destination Enabled when one destination is selected.
Displays a dialog that allows changing the selected Query/Retrieve
destination’s configuration.
Delete Destination Enabled when one or more destinations are selected.
Deletes the selected Query/Retrieve destination(s).
Local Ping Confirms whether or not the local system is connected to a network.
Remote Ping Enabled when one destination is selected.
Confirms whether or not the configured Query/Retrieve system is
connected to the same network as the local system.
C-ECHO Enabled when one destination is selected.
Confirms whether or not the configured Query/Retrieve system is a
Picture Archival and Communications system (PACS).

Configuring the System 18-63


QDR Reference Manual

Using the System Configuration - DICOM - Query/Retrieve tab:


Step Action
1 To add a new destination: click Add Destination,
complete the Add DICOM Query/Retrieve Destination
dialog fields (refer to “Add / Edit DICOM Query/
Retrieve Destination Dialog” on page 18-65 for details),
and click OK.
To reconfigure an existing destination: select the
destination in the selection list, click Edit Destination,
edit the Edit DICOM Query/Retrieve Destination dialog
box as required (refer to “Add / Edit DICOM Query/
Retrieve Destination Dialog” on page 18-65 for details),
and click OK.
To enable/disable a destination for use: click the check
box in the Active column of the selection list for the
destination to check/uncheck (check to enable).
To verify the local system is connected to a local
network: click Local Ping. Click OK to close the
resulting message.
To verify the destination is connected to the same
network as the local system: select the destination in the
selection list and click Remote Ping. Click OK to close
the resulting message.
To verify the destination is a PACS: select the destination
in the selection list and click C-ECHO. Click OK to
close the resulting message.
To delete a destination: select the destination in the
selection list and click Delete Destination.
2 Click OK to return to the main screen.

18-64 Configuring the System


QDR Reference Manual

Add / Edit DICOM Query/Retrieve Destination Dialog

AE Title Application Entity. Provides a unique name for the destination system.
Accepts up to 16 alphanumeric characters.
Host Name or IP Name or IP address of destination.
Address The host name or IP address must be on the same network as the QDR
workstation.
Accepts up to 120 alphanumeric characters.
SCP Port Port number on destination.
The default for the Service Class Provider Port number is 104.
Accepts up to 5 numeric characters in the range of 1 - 65536.
OK Validates data. If validation passes, the dialog is closed and the
Configure DICOM Query/Retrieve Destinations selection list is updated
with new or changed data. If validation fails, a warning message
displays.
Cancel Ignores all edits and closes the dialog.

Configuring the System 18-65


QDR Reference Manual

Host
Host Machine defines your system when DICOM functions are
used.
Important: If changes are made to either the AE Title or
Listening Port data, the APEX application must be restarted for
the changes to take effect.

This section describes how to The Host Machine is configured by selecting System
configure a Host Machine. Configuration – DICOM tab – Host tab found under the
Utilities pull down menu in the main window.

AE Title Application Entity. Provides a unique name for the QDR system.
Accepts up to 16 alphanumeric characters.
Station Name Name of the QDR system.
Accepts up to 120 alphanumeric characters.
Listening Port Port number on which the QDR system listens.
The default port number is 104.
Accepts up to 5 numeric characters in the range of 1 - 65536.
OK Validates data. If validation passes, the System Configuration window
closes returning to the main screen. If validation fails, a warning
message displays.
Cancel Ignores all edits, closes the System Configuration window, and returns
to the main screen.

Using the System Configuration - DICOM - Host tab:


Step Action
1 To add or edit your QDR system information, complete
the Host fields as required.
2 Click OK to return to the main screen.

18-66 Configuring the System


QDR Reference Manual

Privacy Tools
The Privacy Tools tab allows the user to select QDR data fields
for privacy information. There are eight QDR data fields that may
be selected for privacy:
• Patient’s name
• Patient ID
• Patient ID2
• Patient Date of Birth
• Patient Address
• Patient Insurance Information
• Scan Data
• All Comments (including scan comments, physician
comments and patient comments)
Any comment marked with a check mark () is selected for
patient privacy and will be removed from DICOM Reports,
copied scans and exported database. The default for the Privacy
Tools tab is that all fields have a check mark, all are selected for
privacy.
Click on the check mark in the box to remove it and allow this
information to be added to DICOM Reports, copied scans and
exported database.
The button marked Restore Defaults will place a check mark in
all eight QDR Data Fields.
Press OK when done with the Privacy Tools tab.

Configuring the System 18-67


QDR Reference Manual

Study Tab
Note: If the worklist option is The Study Tab is used to develop a study that has multiple
installed on your QDR system, examinations attached to it. The studies are saved on the QDR
refer to “Using Modality
system and can be used from the Select Scan Type window.
Worklist with a Study” on
page 7-13.

Adding a Study
To Add a Study
Step Action
1 From the menu bar select Utilities → System
Configuration and click on the Study Tab.
2 Click New and the following Scan Add window appears.
In the following example a study will be developed using
KP1035A-1743
the AP/Lumbar Spine, Forearm and Left Hip
examinations and be called “Test01.”

3 On the Scan Add window, in the field next to Study


Name, enter the name of the study. This name may have
a maximum of 16 characters (spaces count as a
character).
In our example, the study is named “Test01”.

18-68 Configuring the System


QDR Reference Manual

Step Action
4 Click Add and the Select Scan Type window lists the
examinations on your QDR system.
KP1035A-1745

KP1035A-1746

Cancel returns to the Scan Add window.


5 If you are using a Scan Mode other than the default, click
on the check mark next to the Use Default Scan Mode.
The check mark is removed.
6 Highlight the scan type to be included in the study,

KP1035A-1747

The OK button becomes active.


Cancel returns to the Scan Add window.
7 Click OK.

Configuring the System 18-69


QDR Reference Manual

Step Action
Note: Some scan types have only 8 If you did not remove the check mark in step 5 above, go
one scan mode (the default), if to step 9.
one of these scan types is part of
a Study then this step will not If you removed the check mark in step 5 above, the
appear, go to step 9 (for more following window appears to allow selection of the scan
information refer to the mode (for the scan type selected in step 7).
appropriate chapter of this
manual for the scan type and see
“Selecting the Scan Type and
Mode”).
Note: Some scan types have
more than one scan mode, if one
of these scans is selected as part
of a study this step may appear a
number of times to satisfy the
conditions of the examination
(for more information refer to KP1035A-1748

the appropriate chapter of this Select the scan mode and click OK.
manual:
9 The Scan Add screen appears with the selected scan type
Chapter 8 for AP Lumbar
listed.
Spine
Chapter 9 for Hip
Chapter 10 for Forearm
Chapter 11 for Whole Body
Chapter 12 for AP/Lateral
Spine
See “Selecting the Scan Type
and Mode”).

KP1035A-1749

Note: You may add up to 10 scan 10 Repeat steps 4 through step 9 for each additional scan
types per study. type to be added to the study.
Note: Clicking the Cancel 11 To change the order of examinations, highlight the
button will return to the Study examination to be moved.
Tab window and abort the study.
In our example the Left Hip examination will be done
last.

Arrows

KP1035A-1750a

18-70 Configuring the System


QDR Reference Manual

Step Action
Use the arrow keys to move the examination up or down.
In the example the down arrow is used to move the Left
Hip Examination after the Left Forearm examination.

KP1035A-1750b

Click OK to accept the study.


In the example, the study Test01 now consists of an AP/
Lumbar Spine, Left Forearm and Left Hip examination,
performed in that order.
Click Cancel to reject changes to the order of
examinations
12 When the study is complete, click OK and the Study Tab
appears with the study name listed along with all scan
KP1035A-1751
types for that study.

13 Click OK to accept the study and return to the main


screen.

Configuring the System 18-71


QDR Reference Manual

Editing an Existing Study


Adding Items to an Existing Study
Adding Items to an Existing Study
Step Action
1 From the menu bar select Utilities → System
Configuration and click on the Study Tab.
2 Highlight the study to be changed and click the Edit
button.
Edit button

3 The Scan Add window lists the scan(s) for the study.

KP1035A-1770

18-72 Configuring the System


QDR Reference Manual

Step Action
4 Click the Add button and the Select Scan Type window
lists the examinations on your QDR system.

KP1035A-1747

5 Highlight the scan type to be added to the study and click


on the OK button.
In the example a Dual Hip Examination is added as the
third part of the study named “Test02”.
6 The Scan Add window appears with the new scan type
added to the list.

KP1035A-1771

7 Repeat steps 5 and 6 of this step for each new scan type to
be added to the study.

Configuring the System 18-73


QDR Reference Manual

Step Action
8 When you have added all the scan types to the study,
click OK on the Scan Add window. The Study Tab
window appears listing the Studies and all scans
associated with the studies.

9 Click OK to return to the main screen.

Deleting Items from an Existing Study


Deleting Items from an Existing Study
Step Action
1 From the menu bar select Utilities → System
Configuration and click on the Study Tab.

18-74 Configuring the System


QDR Reference Manual

Step Action
2 Highlight the study to be changed and click the Edit
button.
In our example the Dual Hip scan will be deleted from
the study named “Test02”.
Edit button

The Edit and Delete buttons become active.


3 The Scan Add window lists the scans for the study.

KP1035A-1774

Configuring the System 18-75


QDR Reference Manual

Step Action
4 Highlight the scan type to be deleted from the study.
In the example

KP1035A-1775

The Delete button becomes active.


5 Click Delete and the highlighted scan type is deleted
from the Study.
KP1035A-1776

KP1035A-1777

6 On the Add Scan window, click OK and the Study Tab


window appears listing all Studies and scans associated
with the studies.

18-76 Configuring the System


QDR Reference Manual

Step Action
7 Click OK to return to the main screen.

Deleting a Study
To Delete a Study
Step Action
1 From the menu bar select Utilities → System
Configuration and click on the Study Tab.
2 On the Study Tab window, highlight the study to be
deleted.
In the example study named “Test05” will be deleted.

The Edit and Delete buttons become active.

Configuring the System 18-77


QDR Reference Manual

Step Action
3 Click Delete and a confirmation screen will appear
asking you if you want to delete the study.
KP1035A-1780

Click Yes and the study is deleted from the Study Tab
window.

4 Click OK to return to the main screen.

Renaming a Study
To Rename a Study
Step Action
1 From the menu bar select Utilities → System
Configuration and click on the Study Tab.

18-78 Configuring the System


QDR Reference Manual

Step Action
2 On the Study Tab window, highlight the study to be
changed.
In the example study “Test03” is renamed to “Test05”.

3 Click on the Edit button and the Scan Add window will
appear.
KP1035A-1783

KP1035A-1784

Configuring the System 18-79


QDR Reference Manual

Step Action
4 In the window with the Study name, highlight the scan
name and type in the new name.
In the example “Test03” was highlighted then “Test05”
was typed.
Change name in this window

KP1035A-1785

5 Click OK to return to the Study Tab.


The name of the Study has been changed.
KP1035A-1751

6 Click OK to return to the main screen.

18-80 Configuring the System


QDR Reference Manual

Locating a Study from a Long List


If there are more that 15 studies on the Study Tab window then
you will have to use the side scroll bar to view any studies not on
the window.

Side
Scroll
Bar

In a long list of studies on the Study Tab window, the operator


can use the Scan Name search engine to locate a specific study.

KP1035A-1788

Typing the name of the study into the text box will find that study
and highlight it in the window.

Configuring the System 18-81


QDR Reference Manual

18-82 Configuring the System


Chapter 19
Utilities
Utilities are programs that find, move, store, and edit patient
biographies, patient scans and data, as well as system
information. Utilities are a subset of the main functions of the
system that do the “cleanup” work after the scanning and analysis
are done.
Content
Subject Page
Starting a Utility 19-2
System Configuration 19-2
Usage 19-3
Database Tools 19-9
Patient Management 19-9
Export/Import 19-25
Export 19-27
Import 19-29
Reconcile 19-31
Patient Callback 19-33
Scan File Look 19-36
Scan File Plot 19-36
Emergency Motion 19-40
AP Reposition 19-40
Reference Curve 19-42
To start the Reference Curve Editor: 19-43
Creating New Reference Curve Records 19-44
Copying a Reference Curve Record 19-49
Editing Reference Curve Records 19-51
Deleting Reference Curve Records 19-52
Adding New Ethnic Groups 19-53
Rebuild Archive Index 19-55
Install Options 19-57
Utilities NOT covered in this chapter
Some utilities may appear on the pull-down menu but are not
available to the user. These utilities are intended for
manufacturing and service personnel.

19-1
QDR Reference Manual

Starting a Utility
This section describes how to access the utilities available in the
System software.
To access the Utilities menu:
Step Action
1 Move the pointer to the Main Window Menu Bar and
click on Utilities.
A pull-down menu appears listing the available utilities.

2 Using the mouse, move the pointer down to highlight the


desired utility.
3 Click on the utility to start the operation.

System Configuration
The system software lets you change system configuration
settings for a variety of system parameters. You do this by setting
default parameters that are applied to your system and stay in
effect even after exiting and restarting the software.
The procedures for setting the system configuration are described
in Chapter 18, Configuring the System, in this User’s Guide.

19-2 Utilities
QDR Reference Manual

Usage
Note: Usage is for Strategic Usage is an application that displays and prints billing
Alliance Program (SAP) information captured on your QDR system. It allows you to view
customers only.
and print a summary graph or detailed list of the usage history of
your QDR system.

Hologic’s Strategic Alliance Program requires users of QDR X-


ray bone densitometers who are in the program to report
equipment usage each month. Usage reports for the month must
be faxed to Hologic at (781) 280-0158, no later than the eighth
calendar day of the following month.
The Usage utility is accessed through the Utilities pull-down
menu.
To start Usage:
Step Action
1 Select Utilities from the Menu Bar of the main window.
A list of utilities displays.

Utilities 19-3
QDR Reference Manual

Step Action
2 Select Usage from the pull-down menu.
The Usage program menu displays.

KP1068A-1803

Usage Program Menu


The Usage Program menu, shown above, allows you to select
one of two Usage functions as well as set the Start and End dates
that apply to those functions.
• Show Summary
Show Summary displays a graph showing the number of
studies for a specified period. The summary is the primary
usage report; the Strategic Alliance Program requires that
the report be faxed to Hologic each month.
• Show Detail
Show Detail displays a detailed report for a specified
period. This report lists the patient’s name, date of scan,
type of scan, and study number for each scan performed
during the specified period. The list is grouped by scan
date. Use this report to claim credit for studies that are
technically unacceptable.
Both functions allow you to print the information displayed.

Usage Summary Report


To display the Summary Report, follow the procedure below.
To display the Summary Report:
Step Action
1 Select the Show Summary radio button.
2 Enter the start and end date for the period the report will
cover, and click OK.
The following message displays.

19-4 Utilities
QDR Reference Manual

Step Action

KP0886C-1804

3 Click Yes to show the Summary Report. Clicking No


returns to the Usage program menu.

KP0886C-1805

Print the Usage Summary Report, shown above, to report QDR


system usage for the Strategic Alliance Program. To print the
report, click the Print button.
The Usage Summary Report displays a graph showing the count
of studies for the period specified in the Start and End Date
fields. It also displays the minimum number of studies required
by the Strategic Alliance Program, and additional information,
described below.
• Your institution’s name.
• A line graph showing the number of studies performed
during the specified period. The number of studies is
presented on the vertical axis while the time is shown on
the horizontal axis.
A study may comprise several scans on a single patient
conducted on the same day (for example, spine and hip).
• The date the report was run.
• The model and serial numbers of the scanner.
• The version number of the APEX software.
• A control code for use by Hologic, Inc.

Utilities 19-5
QDR Reference Manual

• The version number of the Usage software.


• The start and end dates of the summary.
• A running monthly average of the number of studies
performed over the most recent six months. If the period
specified in the start and end date is less than six months,
the average number of studies is computed over the
selected time span.
• A total count of the last month of studies.
If the end date is not the default date (i.e. last day of the
previous month) the last month count is left blank.
• Space is provided on the report to enable the user to
request credit for technically unacceptable studies (see
below). The number of technically unacceptable studies
must be hand written or typed on the printed copy.
• A Net Scans space is provided to enable the user to
calculate net scans for billing by subtracting the quantity
of technically unacceptable studies.

Technically Unacceptable Studies


A technically unacceptable study is one that has no diagnostically
usable scans. To request credit for technically unacceptable
studies, write the number of such studies in the Technically
Unacceptable field on the printed report. Subtract this number
from the monthly total of studies to determine the net study count
and write the result in the Net Scans field on the printed report.
In order for Hologic to issue credit for technically unacceptable
studies you must FAX a copy of the detailed report, with all the
technically unacceptable studies marked (circled or underlined)
together with the summary report. Hologic may request a copy of
these studies on diskette.
See the section Claims for Technically Unacceptable Studies on
page 19-8 of this chapter for the fax number and address.

Usage Detail Report


To display the Detail Report, follow the procedure below.
To display the Detail Report:
Step Action
1 Select the Show Detail radio button.

19-6 Utilities
QDR Reference Manual

Step Action
2 Enter the start and end date for the period the report will
cover, and click OK.
The following message displays.

KP0886C-1804

3 Click Yes to show the Detail Report. Clicking No returns


to the Usage program menu.

KP0886C-1806

The Usage Detail Report, shown above, is a detailed report of the


studies performed during the period you specified in the Start and
End Date fields. To print the report, click the Print button.
The Usage Detail Report lists the patient’s name, date of scan,
type of scan, and study count number. Note that multiple scans of
a patient performed on one day are all assigned the same study
count number. The list is sorted in chronological order.
The Usage Detail Report also includes the following:
• The date the report was run.
• Your institution’s name.
• The date the detailed list was prepared.
• The start and end dates of the summary.

Utilities 19-7
QDR Reference Manual

• The total number of studies performed during the given


period (excluding phantom scans).

Claims for Technically Unacceptable Studies


To request credit for unacceptable studies, you must:
• Produce a hard copy of the Usage Detail Report.
• Mark (circle or underline) any studies in which there are
no diagnostically usable scans.
• Fax a marked copy of the Usage Detail Report along with
the Monthly Summary Report to Hologic at (781) 280-
0158.
Hologic may request a copy of all scans in the unacceptable
studies on a diskette. Use the procedure describe in Archiving
Scans on page 16-8 of this user’s guide to produce the copy. Send
the diskettes containing the unacceptable studies in a diskette
mailer to:
Hologic, Inc.
35 Crosby Drive
Bedford, MA 01730
Attention: Strategic Alliance Program

19-8 Utilities
QDR Reference Manual

Database Tools
Database tools are used to move patient, reference, and QC data
to and from other databases. The database tools also allow you to
manage data created by earlier versions of the Hologic QDR
software.

Patient Management
You can manage patient and scan data using the Patient
Management database tool. With Patient Management you can:
• select a baseline scan for each type of scan in the database
for a selected patient
• move scan records from one patient to another.
• delete scan records.
• delete patient records.
• edit patient biographies.
• edit patient insurance information.
• list scan details and edit some entries.
• list scan identification information.

Starting Patient Management


To start Patient Management:
Step Action
1 Select Utilities>Database Tools>Patient Management
from the Menu Bar of the main window.

Utilities 19-9
QDR Reference Manual

Managing Patients
Database Manager dialog box (shown below) displays all the
patient and scan data on the system. The display has several
navigation and control tools to move and edit data. The following
table lists these and their function:

Name Function
Patient Name Any input from the keyboard is displayed in
Text Box this text box. You can select a patient by
typing part or all of the name.
Patient Name This column lists the patients in ascending
alphabetical order.
Patient ID This column lists the patient’s ID number.
Click on Patient ID to sort the list in
ascending numerical order.
Birth Date This column lists the patient’s birth date.
Click on Birth Date to sort the list in
descending chronological order.
Sex This column lists the patient’s sex. Click on
Sex to sort the list by female and then by male
Ethnicity This column lists the patient’s ethnicity. Click
on Ethnicity to sort the list in ascending
alphabetical order.
Referring This column lists patient’s referring physician.
Physician Click on Referring Physician to sort the list in
ascending alphabetical order.

19-10 Utilities
QDR Reference Manual

Name Function
Edit Click on this button to edit the patient’s
biography and insurance information. You
must first select the desired patient (see Item
9). You can also edit by double clicking on the
patient’s name.
Patient Select a patient by:
Selection • Clicking on the patient’s name or any
information to the right of the patient’s
name.
• Entering the name in the Patient Name
text box (see Item 1).
• Select the patient using the
<UpArrow> and <DownArrow> keys.
The patient is selected when a blue
background appears behind it.
Manage Scans This button enables if the selected patient has
button scan data in the system. Click on this button to
see a list of all the patient’s scans. The Select
Scans window appears and you can begin
scan management.
Manage Click this button to select a baseline scan for
Baseline button each type of scan in the database for a selected
patient. Refer to “Managing Patient Baseline
Scans” on page 19-17.
The selected baseline scan displays on the
Scan Parameters window and the scan and
reposition screens when another exam of that
type scan is performed for that patient.
Erase Patient Click on this button to delete the selected
button patient. If the selected patient has scan
information in the system, you can not delete
the patient–you must delete the scan(s) first
(see Item 10 above).
Refer to “Removing a Patient From the
System” on page 19-18.
Close button Click on this button to close the Database
Manager dialog box.

Utilities 19-11
QDR Reference Manual

Managing Patient Information


When you select a patient, the Edit button enables (Item 8 above).
Click this button (or double-click the selection) to enable the
Patient dialog box. The dialog box has two tabs, you can select
Biography and Insurance.

Biography
In the Biography dialog box, you can change any entry by
selecting it either by using the Tab key or the mouse, and
changing the information.
The fields and buttons in this box are:

1
2
3 11
4 12
5 13
6 14
7 15
8 16
9 17

10
18
19
20

Item Name Function


1 Biography The patient dialog box has two tabs–the
Tab Biography tab and the Insurance tab. The
Biography tab displays the items listed
below.

19-12 Utilities
QDR Reference Manual

Item Name Function


Note: If you change the First 2 First Name These fields display the patient’s first and
Name, Last Name, Middle last name and middle initial. You can
Initial, and DOB information so 3 Last Name
change this information by selecting it
that they are identical to a
4 Middle using the mouse pointer or the Tab key
patient that already exists on the
system, a warning appears (see Initial and typing in the new information.
below). Select the appropriate
5 Ethnicity You must select the ethnicity from the
action.
drop list which is generated by the system.
If the appropriate ethnicity is not
available, you can select the blank in the
KP0886C-1810
list.

Note: If you change the Patient 6 Patient ID You can change this filed using a string of
ID information so that it is up to 64 characters.
identical to a patient that
already exists on the system, a 7 Identifier 2 Enter a string of up to 15 characters.
warning appears (see below).
Select the appropriate action. 8 Referring This field is a combination drop list and
Physician text box. You can select any system
generated name from the list or select the
text and enter a string of up to 25
characters.
KP0886C-1811
9 Menopause If the sex of the patient is male, this field
Age is disabled. Otherwise, enter a number
from 00 to 99, or leave blank.
10 Patient Enter a string of up to 100 characters.
Comment
11 Sex Select the sex (F or M) from the drop list.
Note: If you enter inappropriate 12 DOB: Select the month from the drop list.
information for the day and/or Month Select the day and year from the spin
year, an error message appears
controls to the right of the field; or select
(see below). Select OK and 13 DOB: Day
correct the error or select the information in the field and enter the
Cancel to discard any changes. 14 DOB: Year new information.
15 Weight Enter the patient’s weight in lb. (for USA)
or kg (for Europe).
16 Height Enter the patient’s height in inches (for
KP0886C-1812
USA) or centimeters (for Europe).
17 Body Mass Automatically calculated from Height/
Index Weight entries.
18 OK Select OK to save any changes you have
button made and return to the Database Manager.

Utilities 19-13
QDR Reference Manual

Item Name Function


19 Cancel Select Cancel to ignore any changes you
button may have made and return to the Patient
Manager.
20 Help Select Help to display information about
button using this dialog box.

Insurance
Select the Insurance tab to display any insurance information for
the selected patient.
The fields and buttons in this box are:

1
2
3

5
6
7
8
9 11
10 12
13

KP0886C-1813

Item Name Function


1 Name The box displays the patient’s name and
ID. You cannot change these fields here.
2 Patient ID

19-14 Utilities
QDR Reference Manual

Item Name Function


3 Plan You can change all these fields by
selecting them and entering the new
4 Group
information. The system limits each field
5 Insurance to a string of characters of an appropriate
6 Address1 length.

7 Address2
8 City
9 State
10 Postal Code
11 OK button See the descriptions of these buttons under
Biography above.
12 Cancel
button
13 Help button

QC Phantom Details
In the Patient Management dialog box, if you select a Phantom
scan and click on the Edit button, the “QC Phantom Details”
dialog box appears.

KP0886C-1814

The fields and buttons in this box are:


Name Function
Phantom Name This is a concatenation of the information
in the Phantom Type field and the
Phantom Number field. You cannot
change the information in this field.

Utilities 19-15
QDR Reference Manual

Name Function
Phantom Type The system defines the phantom types.
You can change the Phantom Type using
the drop list. The types are limited to
Spine, Block, Hip, and Other. You cannot
enter a type that is not already defined.
Phantom Number You can enter or change this field using a
string of 6 characters or less.
Comment You can enter a comment up to 100
characters.
OK button See the descriptions of these buttons under
Biography above.
Cancel button

19-16 Utilities
QDR Reference Manual

Managing Patient Baseline Scans


Manage Baseline Scans allows the user to select a baseline scan
for each type of scan in the database for a selected patient.
The selected baseline scan displays on the Scan Parameters
window and the scan and reposition screens when an exam of that
type scan is performed for that patient.

This window displays a list of all scan types and a list of all scans
of a selected scan type in the database for the patient that you
selected on the Database Manager window. Use it to:
Select patient baseline scans:
1. In the Select Scan Type list box, click on the desired scan
type. The scan list populates with a list of all scans of a
selected scan type in the database for the selected patient.
The selected baseline scan is designated by a superscript 1 in
the Scan Type column.
2. In the scan list, click on the scan to designate as the baseline
scan and click Set As Baseline.
View a selected scan's details:
1. In the Select Scan Type list box, click on the desired scan
type.
2. In the scan list, click on the scan to designate as the baseline
scan and click Scan Details. The Scan Property window
appears.

Utilities 19-17
QDR Reference Manual

Removing a Patient From the System


You can remove a patient (or a phantom) from the system by:

Step Action
1 Select (highlight) the patient using the mouse or the
arrow keys.
Note: If the selected patient has
no associated scans in the
system, the Erase Patient button
is enabled.

Note: If the Erase Patient button


is not enabled then there are
scans associated with the
patient. All scans listed under
the patient name must be deleted
before the patient can be
removed from the system.

KP0886C-1815

2 Click on the Erase Patient button.


A warning message displays.

KP0886C-1816

3 Select Yes to remove the patient from the database.


The screen updates the patient records minus the patient
just erased.

Managing Scans
If you select a patient in the Patient Management dialog box with
associated scans, the Manage Scans button enables. Select this
KP0886C-1817
button to enable the following dialog box:

19-18 Utilities
QDR Reference Manual

1 2 3 5 6 7 8

KP0886C-1818
9 10 11 12 13 14
The Select Scans dialog box has several editing and navigation
tools to manage the scans. The following table lists these and
their functions:

Item Name Function


1 Patient Name This field indicates the patient’s name.
It can not be changed here.
2 Scan Date This indicates the creation date of the
(button) corresponding scan. You can click on
this button to arrange all the scans in
reverse chronological order. An
asterisk (*) appears after “Date” to
indicate that all the scans listed are
sorted by this field.
3 Scan Type This indicates the type of the
(button) corresponding scan. You can click on
this button to arrange all the scans in
alphabetical order. An asterisk (*)
appears after “Type” to indicate that all
the scans listed are sorted by this field.

Utilities 19-19
QDR Reference Manual

Item Name Function


4 Scan selection Select the desired scan(s) by using the
Up Arrow and the Down Arrow keys or
the mouse pointer. A blue background
highlights the scan(s) when you select
them. You can double-click on the
selected scan to get scan details. You
can also highlight the desired scan(s)
and select the Scan Details button to
get the details of that scan.
5 Scan ID This indicates the ID of the
(button) corresponding scan. You can click on
this button to arrange all the scans in
descending alpha-numerical order. An
asterisk (*) appears after “ID” to
indicate that all the scans listed are
sorted by this field.
6 Analysis Date This indicates the date that the
(button) corresponding scan was analyzed. You
can click on this button to arrange all
the scans in chronological order. An
asterisk (*) appears after “Date” to
indicate that all the scans listed are
sorted by this field.
7 Patient ID This field indicates the patient’s ID. It
can not be changed here.
8 Scan Details This button remains inactive until you
button select a scan. When you select this
button, the Scan Property dialog box
for that scan appears.
9 << Back button Select this button to return to the
Patient Management dialog box.
10 Move Scans>> This button enables only if you select
button one or more scans. Select this button
when you want to move one or more
scans to another patient.
11 Select All This button enables only when you
button select all the scans. Selecting this
button selects all the scans.
12 Deselect All This button enables only when you
button select one or more scans. Selecting this
button deselects all the scans.

19-20 Utilities
QDR Reference Manual

Item Name Function


13 Erase Scan This button enables only when you
button select one or more scans. Selecting this
button deletes the selected scans after
warning you that it will do so.
Selecting Yes deletes the scans from
the system permanently. Selecting No
returns you to the Patient Management
dialog box.
14 Cancel button Select this button to end this session
and return to the Hologic Main
Window.

Moving Scans
Scans can be moved from the one patient’s drawer to another
using the following procedure. Use this procedure if the operator
has forgotten to enter a new patient into the system and stored the
scans under the wrong patient by mistake.
To move scans:
Step Action
Shortcut: <Alt>udm 1 Select Utilities > Database Tools > Patient
Management from the Menu Bar of the main window.

2 From the Database Management dialog box, highlight


(select) the patient name that contains the scans to be
moved.

Utilities 19-21
QDR Reference Manual

Step Action

KP0886C-1819
Manage Scans button
Note: If this button is not active, 3 Click on the Manage Scans button.
there are no scans under this A Scan tab will be displayed showing all scans under the
person’s name to move.
patient’s name.

KP0886C-1820

4 Highlight (select) the scans to be moved.


The Move Scans>> button will become active.

19-22 Utilities
QDR Reference Manual

Step Action
5 Click on the Move Scans>> button.
The Select Target Patient screen is displayed.
KP0886C-1821

KP0886C-1822

6 Highlight (select) the destination (patient name) where


the selected scans are to be moved.
The Finish button will become active.
7 Click the Finish button.
A message is displayed.

KP0886C-1823

8 Click Yes to move the selected scans.


The Database Management dialog box is displayed.

Erasing Scans
Note: Once a scan is erased it is Scans can be erased (deleted) from the patient’s drawer using the
permanently deleted from the following procedure.
system.

To erase (delete) scans:


Step Action
Shortcut: <Alt>udm 1 Select Utilities > Database Tools > Patient
Management from the Menu Bar of the main window.

Utilities 19-23
QDR Reference Manual

Step Action

2 From the Database Management dialog box, highlight


(select) the patient name that contains the scans to be
deleted.

KP0886C-1819
Manage Scans button
Note: If this button is not active, 3 Click on Manage Scans button.
there are no scans under this A Scan tab will be displayed showing all scans under the
person’s name to delete.
patient’s name.

19-24 Utilities
QDR Reference Manual

Step Action

KP0886C-1820

4 Highlight (select) the scans to be deleted.


The Erase Scans button will become active.
5 Click on the Erase Scans button.
A message is displayed.
KP0886C-1824

KP0886C-1825

6 Click Yes to erase the selected scans.


The Database Management dialog box is displayed.

Export/Import
The Export/Import utilities move data to or from any storage
device available to your system. You select the type of data that
you want to move.
Export/Import allows incremental data transfers (new data only).
This saves time and disk space. The data can also be moved to
and from another system. It can be moved to a new or an existing
database.

Utilities 19-25
QDR Reference Manual

Types of Databases
Export/Import allows you to transfer the current data (MS
Access,.mdb extension) or data from an earlier version of Hologic
software (dBase,.dbf extension). Export/Import allows you to
move data:
• to/from any storage device.
• to/from another system.
• from an earlier version of Hologic software (Import only).

Types of Data
Export/Import allows you to move the following types of data:

Patient Data Data in the system that pertains to real


patients (not phantoms). This includes the
patient’s biography and any scan data.
Scan Data Data in the system that pertains to the
scanning of a patient or a phantom.
QC Scan Data All the phantom scan data and the
corresponding QC parameters.
Reference Data Collection of data grouped by age,
ethnicity, and sex. A patient’s scan data is
compared with the reference data from the
same group for evaluation.

Types of Transfers
Note: You can only move Export/Import lets you:
incremental data to/from • choose to add new data to existing data or
previously Exported /Imported
data on the current system. • overwrite the existing data.
Caution: If you elect to move all
the data, you may overwrite
newer data with older data. The
system prompts you if this is
about to happen.

19-26 Utilities
QDR Reference Manual

Export
To export data:
Step Action
Shortcut: <Alt>ude 1 Select Utilities>Database Tools>Export from the Menu
Bar of the main window.

The Export dialog box appears.


Note: You cannot select scans 2 In the Export Data field, click any or all of Patient Data,
without first selecting patients. Scan Data, QC Data and Reference Data for the data to
be exported.
Within the Patient Data and Scan Data fields, click the
check box, or boxes, to filter data as desired.

KP1068_002-1904

3 In the Export Options field:


• click De-identify protected health information
if de-identification is desired
• click All the Data or New Data Only for the files
to be exported.

Utilities 19-27
QDR Reference Manual

Step Action
Note: The software stores files 4 Click Export when done.
as MS Access Files with a.mdb A “Save As” window (similar to the following) appears:
extension. If you want to change
the format, you must do it
directly through the MS Access
application.

KP0886C-1829

Warning: Do not select QDR 5 Double click on the location (folder) that is the
data as a storage folder for your destination for the exported database.
data. The system reserves this
The Save in: window changes to the destination (“Temp”
folder for storing the system
data. in the following example):
Destination location

KP0886C-1830

Note: If an error occurs stating 6 In the “File name” text box, enter a name for the
that the name already exists, database.
type in another name for the
data file.
7 Click on Save button.
The window displays the progress of the database being
saved, when completed the Close button will be active.
8 Click on the Close button to exit Export.
The main window is displayed.

If you choose Reference Data along with QC and/or Patient/Scan


Data, Export prompts you twice for the desired database. This is

19-28 Utilities
QDR Reference Manual

because the software exports and stores the reference data


separately from the other data.

Import
To import data:
Step Action
Shortcut: <Alt>udi 1 Select Utilities>Database Tools>Import from the Menu
Bar of the main window.

The Import dialog box appears:

KP0886C-1832

2 Select the data types from the Data Sets that match the
data files to be loaded (click on the radio button).
3 Select either All the Data or New Data Only
(incremental data) under the Options heading.

Utilities 19-29
QDR Reference Manual

Step Action
Note: Import also restores data 4 Click the Import button when done.
from earlier QDR operating The “Open” window is displayed (similar to the
systems. This data has the.dbf
following:
extension and can be selected by
dropping the Save as type: list
and selecting DOS SCANS...
from the list.

KP0886C-1833

5 Double click on the folder that contains the database files


to be imported
The Look in: part of the window displays the databases
that exist in that folder (as seen in the following
example):

KP0886C-1834

6 Highlight (select) the database to be loaded and click on


the Open button.
The window displays the progress of the database being
loaded. When completed the Close button will be active.
7 Click on the Close button of the Import Dialog box.
The main window is displayed.

19-30 Utilities
QDR Reference Manual

If you choose Reference Data along with QC and/or Patient/Scan


Data, Import prompts you twice for the desired database. This is
because the software imports and stores the reference data
separately from the other data.

Reconcile
Reconcile is a helpful database tool that automatically compares
the system database with the scan files in your system directory.
Any discrepancies are automatically corrected by your system.
To perform the Reconcile function:
Step Action
1 Select Utilities>Database Tools>Reconcile from the
Menu Bar of the main window.

The following warning appears:

KP0886C-1836

Utilities 19-31
QDR Reference Manual

Step Action
2 Click on the Continue button to start the database
reconciliation.
The Database Reconcile window appears showing the
progress of Part 1 of the program.

KP0886C-1837
Cancel button - click to stop Reconcile procedure

Upon completing Part 1, a window appears showing the


progress of Part 2 of the program.

KP0886C-1838

Cancel button - click to stop Reconcile procedure

When Part 2 is complete the Close button is active.

KP0886C-1839

3 Click on the Close button


The main window is displayed.

19-32 Utilities
QDR Reference Manual

Patient Callback
Patient Callback provides a list of patients matching specified
criteria to be used for scheduling follow-up exams. The criteria
for this list can be a scanned date range, a T-score range or both
for the following scan types:
• Total Hip
• Femoral Neck (Hip
• Total Spine
To perform the Patient Callback function:
Step Action
1 Select Utilities>Database Tools>Patient Callback from
the Menu Bar of the main window.

The Patient Callback screen appears:

KP1068_002-1911

Utilities 19-33
QDR Reference Manual

Step Action
2 Enter a start and end date ranges into Date Range field,

KP1068_002-1912

or,
enter a minimum and maximum T-Score range into the T-
Score Range field.

KP1068_002-1913

or,
input ranges into both of the fields.
3 Click Generate List and a message appears stating that
KP1068_002-1914
the Callback Query is in progress.

KP1068_002-1915

A list of patients appears that fall within the ranges of


date, T-Score or both.

KP1068_002-1916

19-34 Utilities
QDR Reference Manual

Step Action
4 Click on any patient to add that patient to the Callback
list,
or,
use the Ctrl key to select multiple patients,
or,
click Select All to select all of the patients in the list.
KP1068_002-1917

5 After selecting the patient, you can provide the


information as follows:
• Print the list by clicking Print.

KP1068_002-1918

• Export the list as a text file by clicking Text File


Export (you will be prompted to name the file).

KP1068_002-1918

• Export the list as an Access database file by


clicking DB Export (you will be prompted to
name the file).

KP1068_002-1919

6 The Cancel button has two functions:


1. Clicking this button before performing step 5 above
KP1068_002-1920 returns to the main screen and any patients selected
for a list will be lost.
2. Clicking this button after performing step 5 above
returns to the main screen.

Utilities 19-35
QDR Reference Manual

Auto Baseline Utility


Sets the baseline scan of all restored scans (patients and scan
types) to the oldest scan.
To perform the Auto Baseline Utility function:
Step Action
1 Select Utilities>Database Tools>Auto Baseline Utility
from the Menu Bar of the main window.
A message displays when the function is completed.
2 Click OK.

Scan File Look


Data in scan files is stored in records.
Scan File Look displays records in any of the three types (P, R,
and A) of scan files and data contained in a record can be
displayed. It is a tool intended for use by people performing
research and Hologic personnel performing system diagnostics.
To perform the Scan File Look function:
Step Action
1 Select Utilities>Scan File Look from the Menu Bar of
the main window.

A scan file selection window displays.

19-36 Utilities
QDR Reference Manual

Step Action
2 Select a scan file in the selection list, select the file type
from the File Type drop-down field, and click OK.

The scan file viewer window displays a summary of all


records in the selected scan file.

Utilities 19-37
QDR Reference Manual

Step Action
3 To view the data contents of a record, click on the desired
record in the record selection list.

4 To print the data contained in the selected record, click


the print button or select Print from the File menu.

5 To copy text contained in the data view area for pasting


into another application, select (highlight) the desired
text, and click the copy button or select Copy from the
Edit menu.
6 To view the records in a different file type for the same
scan file, click the down arrow on the Open Scan File
button and select a corresponding file type.

7 To view the records in a different scan file, click Open


Scan File, located and select the desired scan file, and
click Open.

19-38 Utilities
QDR Reference Manual

Scan File Plot


Scan File Plot displays the bone image contained in a scan R file
and a graph representing the data in the lines of the scan. It is a
tool intended for use by people performing research and Hologic
personnel performing system diagnostics.
To perform the Scan File Plot function:
Step Action
1 Select Utilities>Scan File Plot from the Menu Bar of the
main window.

The Plot dialog appears.

2 Select a scan file in the selection list and click OK.

Utilities 19-39
QDR Reference Manual

Emergency Motion
The Emergency Motion utility should only be used under the
supervision of Hologic personnel or its authorized representative.
This utility allows the operator to move the table and the C-arm
independently. It is available in the unlikely event that your table
and/or C-arm are positioned outside of their normal operating
range.
To perform the Emergency Motion function:
Step Action
1 Select Utilities>Emergency Motion from the Menu Bar
of the main window.

Follow Hologic Help Desk instructions.

AP Reposition
The AP Reposition utility is available on Discovery-A and -SL,
systems only. It should only be used under the supervision of
Hologic personnel or its authorized representative. This utility
allows the operator to return the C-arm from the lateral position to
the AP position in the unlikely event that a scan is aborted during
the lateral portion of an AP/Lateral scan.

19-40 Utilities
QDR Reference Manual

To perform the AP Reposition function:


Step Action
1 Select Utilities from the Menu Bar of the Main Window.
Select AP Reposition from the drop-down menu.

Utilities 19-41
QDR Reference Manual

Reference Curve
Standard reference curves are provided by Hologic based on
studies performed on Hologic QDR bone densitometers.
References curves are sets of data points for a given sex,
ethnicity, and scan type/region and specify standard deviation and
skew value for the point.
Reference Curve allows users to setup and manipulate custom
reference curve data.
Using Reference Curve you can:
• view reference curve record data
• create new reference curve records
• modify reference curve records (Hologic supplied
reference curve records cannot be modified)
• delete reference curve records (Hologic supplied
reference curve records cannot be deleted)
• create new ethnicity groups
• restore database to Hologic-supplied reference curves
Hologic-supplied reference curves cannot be edited or deleted.
However, Hologic-supplied curves can be marked as current or
non-current and they can be copied and edited to create a new
reference curve.

19-42 Utilities
QDR Reference Manual

Starting Reference Curve Editor


To start the Reference Curve Editor:
Step Action
1 Select Utilities>Reference Curve>Editor from the Menu
Bar of the main window.

The following message displays.

KP0886C-1863

Note: An “H” in the Hologic field 2 Click OK to continue. The Reference Curve Editor dialog
indicates a Hologic provided box displays.
reference curve record that cannot
be modified or deleted.

Utilities 19-43
QDR Reference Manual

Viewing Reference Curve Data


To view reference curve data:
Step Action
1 Start the Reference Curve Editor (refer to “Starting
Reference Curve Editor,” on page 19-43).
2 Locate and click on the reference curve record line to view.
The line is highlighted.
3 Click View. The View Reference Curve dialog box appears.
Note: No changes can be made
from this screen. Refer to Editing
Reference Curve Records on Page
19-51 for details on editing curves.

Note: Refer to the Reference


Curve Description Fields table on
Page 19-47 for descriptions of the
fields on this screen.

The upper section contains the reference curve description


information. The lower section contains the reference
curve point data. No changes can be made on this dialog
box.
6 Click Close to return to the Reference Curve Editor dialog
box.
7 Click Close to return to the main screen.

Creating New Reference Curve Records


Note: New reference curve records can be easily created by
copying an existing reference curve record where most of the data
is to be the same as that of the existing record. Refer to Copying a
Reference Curve Record on Page 19-49.
To create new reference curve records:
Step Action
1 Start the Reference Curve Editor (refer to “Starting
Reference Curve Editor,” on page 19-43).

19-44 Utilities
QDR Reference Manual

Step Action
2 Click New.
A message similar to the following appears.

KP0886C-1849

3 Click OK to continue.The New Reference Curve dialog box


appears.
Note: Refer to the Reference
Curve Description Fields table on
Page 19-47 for descriptions of the
fields on this screen.

The upper section contains fields to enter the reference


curve description information. The lower section lists the
reference curve point data as it is added.
4 Complete the reference curve description information in
the upper section. Use the drop-down lists where available.
Use the tab key to move between fields.

Utilities 19-45
QDR Reference Manual

Step Action
5 Click Select X, Y Labels. The following dialog appears.

6 Expand the label selection trees as required.


7 Click on one label in both the X Label and Y Label section.
8 Click OK.
9 Click Input to add a new set of points to the reference
curve.
Input Record dialog box similar to the following appears.

10 Complete the information fields, S.D. (standard deviation),


and L (skew value for the point) fields and click OK.
The point set appears in the lower section sorted by the X-
axis selection.
11 Repeat Step 10 as required to add additional point sets. Or
click Cancel to close the Input Record dialog box and
continue.

19-46 Utilities
QDR Reference Manual

Step Action
12 If necessary, edit a point set by clicking on the desired
point set and clicking the Edit button.
The Edit Record dialog box appears.

13 Change the information fields as required and click OK.


The point set appears in the lower section sorted by the X-
axis selection.
14 If necessary, delete a point set by clicking on the desired
point set and clicking the Delete button.
A warning message appears.
15 Click Yes to continue.
The point set is removed from the lower section.
16 Repeat Steps 14 and 15 as required to delete additional
point sets.
17 When you have completed adding the reference curve,
click Close to record the curve data and return to the
Reference Curve Editor dialog
The message similar to the following appears.

KP0886C-1855

18 Click Close to return to the main screen.

Reference Curve Description Fields


Field Description
Sex Select from drop-down list.
Ethnicity Select from drop-down list.

Utilities 19-47
QDR Reference Manual

Field Description
Date Set by the system when a curve is created
or modified. Cannot be edited.
Author Identifier for person creating or
modifying curve. Enter up to five
characters.
Source Identifier for supplier providing the
reference curve data. Enter up to 61
characters.
Comment Comments pertaining to the reference
curve.
Select X, Y Labels This button brings up the Select X, Y
Labels window.
X-Axis
Label X-axis label to display on reports.
Display from - to Range of X-axis data to display on
reports between which the reference
curve is considered to be valid. This does
not necessarily correspond to the low and
high points that define the curve.
Y-Axis
Label Y-axis label to display on reports.
Display from - to Range used for the Y-axis in the
graphical display. Does not effect the
operation of Normals.
Is this curve current? Lets more than one reference curve (for
the same sex, ethnic group, scan type,
and bone region) on the system at the
same time. Only one of these curves can
be marked as current. Only current
curves are used by Normals.
Method Analysis method for the curve. Select
from drop-down list.
Scan Type Set by the system depending on the X, Y
label selection when a curve is created or
modified. Cannot be edited.

19-48 Utilities
QDR Reference Manual

Field Description
Age Peak BMD Age of maximum bone density used to
compute T-score. Visible when X-axis
label is selected as “Age” and Y-axis
label is selected as “BMD”.

Copying a Reference Curve Record


Copying an existing reference curve record makes it easy to
create a new curve record where most of the data is to be the same
as that of the existing record.
To create new reference curve records from existing records:
Step Action
1 Start the Reference Curve Editor (refer to “Starting
Reference Curve Editor,” on page 19-43).
2 Locate and click on the reference curve record line to copy.
The line is highlighted.
3 Click Copy.
A message similar to the following appears:

KP0886C-1859

Utilities 19-49
QDR Reference Manual

Step Action
4 Click OK.
The Copy Reference Curve dialog box appears:

The upper section contains fields to enter or change the


reference curve description information. The lower section
lists the current reference curve point data.
5 Change the reference curve description information in the
upper section as required. Use the drop-down lists where
available. Use the tab key to move between fields.
6 Please refer to Creating New Reference Curve Records -
Steps 5 through 8 beginning on Page 19-46 for selecting
X, Y labels.
7 Please refer to Creating New Reference Curve Records -
Steps 9 through 16 beginning on Page 19-46 for adding,
editing, and/or deleting reference curve point data.
Then continue with the following steps.
8 When you have completed changing the copied reference
curve, click Close to record the curve data and return to the
Reference Curve Editor dialog.
A message similar to the following appears.

KP0886C-1855

9 Click Close to return to the main screen.

19-50 Utilities
QDR Reference Manual

Editing Reference Curve Records


Note: With the exception of the Is the curve current? field in the
reference curve description section, Hologic provided reference
curve records cannot be modified.
To edit existing reference curve records:
Step Action
1 Start the Reference Curve Editor (refer to “Starting
Reference Curve Editor,” on page 19-43).
2 Locate and click on the reference curve record line to edit.
The line is highlighted.
3 Click Edit.
The Edit Reference Curve dialog box appears.
Note: Refer to the Reference
Curve Description Fields table on
Page 19-47 for descriptions of the
fields on this screen.

The upper section contains fields to enter or change the


reference curve description information. The lower section
lists the current reference curve point data.
4 Enter or modify the reference curve description
information in the upper section. Use the drop-down lists
where available. Use the tab key to move between fields.
5 Please refer to Creating New Reference Curve Records -
Steps 5 through 8 beginning on Page 19-46 for selecting
X, Y labels.
6 Please refer to Creating New Reference Curve Records -
Steps 9 through 16 beginning on Page 19-44 for adding,
editing, and/or deleting reference curve point data.
Then continue with the following steps.

Utilities 19-51
QDR Reference Manual

Step Action
7 When you have completed changing the copied reference
curve, click Close to record the curve data and return to the
Reference Curve Editor dialog.
A message similar to the following appears.

KP0886C-1855

8 Click Close to return to the main screen.

Deleting Reference Curve Records


Note: Hologic provided reference curve records cannot be
deleted.
To delete existing reference curve records:
Step Action
1 Start the Reference Curve Editor (refer to “Starting
Reference Curve Editor,” on page 19-43).
2 Locate and click on the reference curve record line to edit.
The line is highlighted.
3 Click Delete.
The following message appears:

KP0886C-1870

4 Click Yes to delete the selected record and return to the


Reference Curve Editor dialog box.
5 Repeat Steps 2 through 4 to delete additional records. Or
click Close to return to the main screen.

19-52 Utilities
QDR Reference Manual

Adding New Ethnic Groups


To add new ethnic groups:
Step Action
1 Select Utilities>Reference Curve>Add Ethnicity from
the Menu Bar of the main window.
3

The following message appears.

4 Click OK to continue.
The Add New Ethnicity dialog box appears:

KP0886C-1873

Utilities 19-53
QDR Reference Manual

Step Action
5 Type name and code (two alphanumeric characters) for the
new ethnicity group into their respective fields and click
OK to add the group and return to the Main screen.

Warning: If you plan to exchange scan data with other


users, make sure your new ethnicity code does not match
any of their ethnicity codes unless you are actually using
the same reference curves for that ethnicity.

Restoring Reference Curve Database


Note: This option restores the reference curve database to the
original state as supplied by Hologic. Any changes that may have
been made are lost.
To restore the reference curve database:
Step Action
1 Select Utilities>Reference Curve>Restore from the
Menu Bar of the main window.

The following message appears.

KP0886C-1875

2 Click Yes to restore the database. Or click No to stop the


restore. The system returns to the Main screen

19-54 Utilities
QDR Reference Manual

Rebuild Archive Index


Rebuild Archive Index is used to rebuild the index on an archive
media if it is corrupted.
To rebuild archive index:
Step Action
1 Select Utilities>Rebuild Archive Index.
Rebuild Archive Index window appears.

KP1068_001-1977

2 In the text box to the left of the Browse button, type in


the location of the archive
or
click Browse to select the location from the Browse for
Folder window.

KP1068_001-1978

Utilities 19-55
QDR Reference Manual

Step Action
3 Click Rebuild Index.
A window appears showing the index being rebuilt.

KP1068_001-1979

4 When the index has been rebuilt the following window


appears:

KP1068_001-1980

Click OK to return to the main screen.

19-56 Utilities
QDR Reference Manual

Install Options
Installs a selected APEX option on the QDR system. Before
installing an option, obtain a license key from Hologic for that
option.
To perform the Install Options function:
Step Action
1 Select Utilities>Install Options from the Menu Bar of
the main window.

2 Type the license key obtained from Hologic in the


License Key field.

3 Click Install Option.


4 Follow instructions on the screen.
5 Select another option to install or click Close

Utilities 19-57
QDR Reference Manual

19-58 Utilities
Chapter 20
Digital Imaging and
Communications in Medicine
(DICOM)
Digital Imaging and Communications in Medicine (DICOM) is a
powerful tool that provides interpreting physicians with the
ability to view electronic QDR bone density scan and analysis
results on a Picture Archiving and Communications System
(PACS). The DICOM option allows results to be transmitted
automatically over a facility’s network directly to a physician’s
DICOM viewing station for interpretation and report dictation.
The results can also be archived on the PACS, making them
available for future reference and for distribution to others on the
PACS network.

Contents
Subject Page
DICOM Selections 20-2
Configuring DICOM 20-4
Using DICOM 20-4
Troubleshooting 20-5

20-1
QDR Reference Manual

DICOM Selections
The following table shows the different scan types and options for
generating a DICOM report.
Scan Type
Group Lumbar Hip Forearm Whole AP/
Spine Body1 Lateral2
Report Type
Most Recent Value     
Rate of Change     
Range
Radius + Ulna 
Radius 
Ulna 
Region for Normal Plot
Neck 
Trochanter 
Inter 
Ward’s 
Total    
1/3 
MID 
UD 
VB 
MVB 
Note Availability
1 A and W models only
2 A and SL models only
The DICOM report displays all of the information necessary for
interpreting BMD results on a single screen. The report consists
of the radiographic image, detailed analysis results and
summaries, reference population graphs, patient demographic
data, referring physician information, and details about each scan
and analysis. Refer to Figure 20-1 for an example of a DICOM
report.

20-2 DICOM
QDR Reference Manual

Figure 20-1
DICOM Report Example

KP1068A-2001

DICOM 20-3
QDR Reference Manual

Configuring DICOM
DICOM is configured using the System Configuration –
DICOM tab found under the Utilities pull down menu in the
main window.
For details on configuring DICOM, refer to “DICOM Tab” on
page 18-44.

Using DICOM
DICOM can be used to:
• Send reports to remote destinations (refer to “DICOM
Reports” on page 15-18).
• Locate and retrieve patient studies previously archived to a
PACS server using the DICOM Send functionality (refer to
“Query/Retrieve Scans” on page 16-22).
• Select a patient for examination from a worklist schedule
received from the HIS/RIS (refer to “Using the Worklist Tab”
on page 7-8).

20-4 DICOM
QDR Reference Manual

Troubleshooting
This section covers frequently asked questions and answers.

Q When I click on Send, I get the error message: “Please


select a destination node.” What does this mean?
A You were trying to send a DICOM report to an undefined
destination node. Please refer to the Configuring DICOM
section for more details on setting up a destination node.
Q I sent a report a while ago, but the report is still waiting in
the queue. Why wasn’t the report transmitted?
A Make sure that your destination is configured properly.
The AE Title, IP Address, and SCP Port you entered in
the Add/Edit Remote Node dialog box must match those
of the DICOM receiver you are sending to. Check with
your System Administrator for the correct values.
Verify that the DICOM receiver you are sending to is
active. The destination machine must be running in order
for you to send a report successfully.
Check with your Network Administrator to make sure
that there are no network problems. If the network is
down, the DICOM software attempts continuously to
send the report until it succeeds.
Q I sent a report to a destination with an incorrect IP
address, and the report is still in the queue. How do I
delete it?
A Delete the report from the queue using the Delete button
on the View Queue dialog box.
Tip: Clicking on the View Queue button on the DICOM
Reports dialog box displays the View Queue dialog box.
Q When I send or save a report, an error message is
displayed: “Accession number already in use.” What
does this mean?
A The Accession number is uniquely associated with a
patient visit. You were trying to enter an accession
number for a scan. The accession number that you were
using is already being used by another scan. Enter a
different accession number or double-check the accuracy
of the accession number.

DICOM 20-5
QDR Reference Manual

Q My Send, Save As, Preview, and Scan Details buttons


are inactive (dimmed). How do I activate them?
A Highlight only one scan from the scan list on the DICOM
Reports screen. The Save As, Preview, and Scan Details
buttons only work with single scans.
The Send button can be used with multiple selected
scans.
Q The Standardized BMD check box is active (dimmed).
How do I activate it?
A The Standardized BMD (sBMD) report is only an option
for Single Analysis reports. Click on the Single Analysis
radio button to activate the Standardized BMD check
box.
Q How come the BMD reference curve graph is missing
from my Rate of Change DICOM report?
A The Skip button was pressed when the Select Rate of
Change Scans dialog box appeared. This skips over the
scan selection process to include in the report. When the
Select Rate of Change Scans dialog box appears, make
sure you select at least one scan to include and click OK.
Another reason could be that the scan that you selected
for the Rate of Change report does not have any other
matching scans for that patient. Select a scan that has
matching scans for the Rate of Change report.

20-6 DICOM
Appendix A
Options
This appendix provides information on the standard features and options available on QDR
systems.

Content
Subject Page
Options List A-2
Forearm A-2
IVA A-3
IVA HD A-3
DICOM Store A-3
DICOM Modality Worklist A-3
General Region of Interest A-3
Body Composition Analysis A-3
Decubitus Lateral A-3
Small Animal A-3
Prosthetic Hip A-4
Enterprise Data Management A-4
Hip Axis Length A-4
Infant Whole Body A-4
Low Density Spine A-4
International A-4
DAP Meter A-4
Hip Structure Analysis® (HSA) A-4
Mobility A-4

A-1
QDR Reference Manual

Options List
A listing of standard and optional features for QDR systems is presented in the table below. A
brief description of each software option has also been provided for your information. Note that,
in this Appendix, models shown only as A, SL W and C refer collectively to Discovery, QDR
4500 and Delphi models A, SL, W or C. Models Wi and Ci refer to Discovery only.

Features Models Work-


station 1
A SL W Wi C Ci Ex2
Forearm S S S S S S O S
IVA3 S S S O S O NA S
IVA HD3 S S S NA S NA NA S
DICOM Store S S S O S O O O
DICOM Modality Worklist O O O O O O O O
General Region of Interest O O O O O O O O
Body Composition Analysis O NA O O NA NA O O
Decubitus Lateral BMD NA NA O O O O O O
Small Animal O O O O O O NA O
Prosthetic Hip O O O O O O O O
Enterprise Data Management O O O O O O O O
Hip Axis Length O O O O O O O O
Infant Whole Body O NA O O NA NA NA O
Low Density Spine O O O O O O O O
International O O O O O O O O
DAP Meter O O O O O O O O
Hip Structure Analysis O O O O O O O O
Mobility O O O O O O O O
S: Standard O: Option NA: Not Available
1
QDR Workstations are unable to scan patients (reference to standard features)
2
Explorer
3
Requires DICOM option

Forearm
Hologic’s Forearm option provides the ability to perform a Forearm scan and analysis. The
anatomical sites that are measured for the forearm scan are: the distal portion of the radius and
ulna, the wrist bones, and the first row of carpal bones.

A-2 Options
QDR Reference Manual

IVA
The IVA (Instant Vertebral Assessment) option provides functionality for the acquisition of scans
that allow the determining of the presence or absence of vertebral fractures.

IVA HD
The IVA HD (Instant Vertebral Assessment - High Definition) option provides functionality for
high definition imaging for the determination of the presence or absence of vertebral fractures.

DICOM Store
Hologic’s DICOM Store allows the QDR system to transfer DICOM files to a PACS or Physician
Viewer.

DICOM Modality Worklist


Hologic’s DICOM Worklist allows the QDR system to retrieve schedule and patient demographic
information using the DICOM Modality Worklist service.

General Region of Interest


Hologic’s General Region of Interest option, which is sometimes referred to as Subregions,
provides the ability to use a nonstandard region of interest to analyze a scan. This feature is
available for spine, hip and forearm scan types but not for supine lateral scans.

Body Composition Analysis


Hologic’s Body Composition Analysis option provides the ability to analyze the soft tissue
composition of the entire body and to follow the changes in soft tissue composition over time in
response to intervention such as diet and exercise. Analysis of fat mass, lean mass, and percent fat
mass can be reported for the entire body and head, arms, trunk, pelvis and legs.

Decubitus Lateral
Hologic’s Decubitus Lateral option allows the acquisition, display and analysis of scans of the
lateral view of the spine on the W, Wi, C, Ci, and Explorer models.

Small Animal
Hologic’s Small Animal option provides the ability to perform two types of small animal
measurements:
• You can perform whole body bone mineral and body composition measurements in rats.
This software is optimized for adult rats weighing between 200 and 750 grams.
• You can also perform BMC and BMD measurements of small animal bones using regional
high resolution. The regional high resolution software is optimized for excised rat femurs
and spines but may also be used to measure these bones in vivo.

Options A-3
QDR Reference Manual

Prosthetic Hip
Hologic’s Prosthetic Hip option provides the ability to scan and analyze prosthetic hips and
anatomy that includes metal, for example a femur with a prosthesis. Changes in bone mass
surrounding metal implants can be assessed by using the specially designed algorithm. The metal
components of the implant is removed from the bone mass measurement.

Enterprise Data Management


Hologic's Enterprise Data Management option allows the QDR system to archive, locate and
query/retrieve raw scan data from a configured PACS server.

Hip Axis Length


Hologic's Hip Axis Length option measures the distance along the femoral neck axis, extending
from the bone edge at the base of the trochanter to the bone edge at the inner pelvic rim.

Infant Whole Body


Hologic’s Infant Whole Body option allows the acquisition, display and analysis of scans of
infants on the A, W and Wi models.

Low Density Spine


Hologic’s Low Density Spine option is available for legacy customers who have used this
function in the past and wish to continue using it. For all other users, this function is replaced by
the Auto Low Density analysis feature.

International
Hologic’s International option allows the QDR system to utilize and display metric units as well
as international date and time conventions.

DAP Meter
The Dose Area Product (DAP) Meter displays the radiation dose applied to a patient when a scan
is performed.

Hip Structure Analysis® (HSA)


The Hip Structure Analysis (HSA) option is a proprietary program for computing the structural
properties of various cross sections of the proximal femur.

Mobility
Includes generation and filling in of a History Questionnaire on a remote Tablet PC, electronic
sending from Tablet PC to QDR, inclusion with scans sent via DICOM to the Physician's Viewer
and inclusion of data from the questionnaire into the DxReport.

A-4 Options
Appendix B
Operator Maintenance
This appendix provides information and instructions for maintaining the QDR systems.

Content
Subject Page
Periodic Cleaning B-2
Periodic Visual Inspection B-3
Cleaning Accidental Spills B-4

B-1
QDR Reference Manual

Periodic Cleaning
This section provides instructions for periodically cleaning the QDR system and its components.

Cleaning the QDR and Computer Components


It is important to keep the exterior of the QDR system and the computer components clean.
To clean the system exterior and computer components:
Step Action
1 Switch off the power at the main breaker.
2 Use a soft, damp cloth to wipe the surfaces. If necessary,
use a mild detergent to remove caked-on dirt or debris.
3 Restore power at the main breaker.

The table pad is the area of the QDR system where the patient lies during an exam. It is important
to keep the pad clean.

Cleaning the Table Pad


Step Action
Note: Do not remove the cover 1 Use a simple solution of neutral soap and lukewarm
from the table pad to clean or water or a diluted (10 to 1) bleach solution.
disinfect.
Note: If cleaning does not 2 Let dry thoroughly before scanning.
produce satisfactory results,
contact your Hologic
representative to order a
replacement table pad.

Disinfecting the Table Pad


Step Action
Note: Disinfectants applied at 1 Use a phenolic, or quaternary type disinfectant in the
full concentration, or in highly concentration recommended by the manufacturer.
concentrated solutions may
damage the fabric. Do not use Idophor or any iodine bearing disinfectants.
2 Let dry thoroughly before scanning.

B-2 Operator Maintenance


QDR Reference Manual

Periodic Visual Inspection


It is important to periodically inspect the QDR system for proper clearance of moving parts, wear
and tear, and parts that may need to be replaced.
Visual inspections include:
• Checking for Adequate Clearance Around the Unit
• Checking Cables and Outlets
• Inspecting the Table Pad

Checking for Adequate Clearance


Warning: Obstructions can The QDR system incorporates moving parts. Check to make sure
cause a hazardous condition to there is enough clearance around the unit so that all parts move
the patient and unit.
freely.

Checking Cables and Outlets


Warning: Overloading the outlet Make sure that only computer equipment and accessories
strip can cause a hazardous supplied by Hologic are connected to the outlet strip.
condition and/or damage to the
equipment. Check all cables for worn and/or damaged insulation. Each cable
should be positioned so as not to constitute a hazard to the patient
or operator. Contact Hologic to replace worn or damaged cables.

Inspecting the Table Pad


Note: A damaged or stained Periodically inspect the table pad for damage, such as rips or
table pad can distort X-ray tears, and stains. Use a mild detergent to clean the pad, and a
transmission and produce
10% (or less) bleach solution to disinfect the pad.
erroneous results in the analysis.
Refer to page B-3 for Contact Hologic if you need to replace the table pad.
instructions on cleaning the
table pad.

Operator Maintenance B-3


QDR Reference Manual

Cleaning Accidental Spills


It is important to discourage liquids (such as coffee, water, or soft drinks) in the vicinity of the
QDR system to avoid accidental spills.
To clean accidental spills:
Step Action
Note: Contact the Hologic Help 1 In the event of a liquid spill, wipe the spill immediately
Desk if you need assistance. with a slightly dampened sponge. If the spill penetrates the
QDR system’s interior, shut off power at the main breaker
immediately.
Note: Moisture on the table pad 2 Allow the table pad to dry thoroughly before performing
can distort X-ray transmission and patient scans.
produce erroneous results in the
analysis.
3 If you turned off power at the main breaker in Step 1,
restore power when the unit is thoroughly dry.

B-4 Operator Maintenance


Glossary
This section provides an alphabetical listing of terms and acronyms, used in this User’s Guide,
with explanations.

Content

Glossary of Terms

G-1
QDR Reference Manual

Acetabulum A hollow area or socket in the pelvis in which the femoral head
fits.
Accuracy Ability of an instrument to measure the same bone mineral
content of an object as measured by another independent method
or “gold standard.”
Archive Process of transferring scan files, including the scan’s raw data,
onto removable media as permanent storage. The scans can then
be retrieved at a later date, and can also be re-analyzed.
ASIS Anterior Superior Iliac Spine. A palpable bone projection found
about one inch below the iliac crest of the pelvis.
Biography Contains a patient’s name, date of birth, and other demographic
information. Each patient must have only one entry within the
database, or else rate of change calculations are not possible.
Biochemical Blood and urine tests that can measure the rate of bone resorption
markers and formation. It provides physicians with information on how
quickly bone is lost by resorption, and is used to evaluate the
patient’s response to treatment.
BMD Bone Mineral Density
BMC Bone Mineral Content
Bone Map A yellow computer generated overlay, representing a region or
regions of bone, that is identified by the system during analysis.
Calcitonin A naturally occurring hormone secreted by the thyroid gland,
responsible for inhibiting bone resorption. Synthetic salmon
Calcitonin, an injectable drug treatment, is used to slow the bone
resorption rate in patients.
Cortical bone Compact dense bone that makes up eighty percent of the skeletal
mass.
CV Coefficient of variation is the standard deviation expressed as a
percentage of the mean.
(CV = 100 x (SD/Mean))
Database A set of unique HOLOGIC system files that include Biography,
QC, Locate, and reference database information.
Dbarchive Copies all the internal database files onto a floppy disk to protect
against data loss after a computer failure. This procedure should
be performed weekly.

G-2 Glossary
QDR Reference Manual

Detector Converts light into electrical impulses that are sent to the
computer for processing.
DEXA See DXA
Diskette A flexible, magnetized plastic disk, used to store system data,
and patient scan data.
DPA Dual-Photon Absorptiometry. The first dual energy clinical
densitometer that was capable of measuring the spine and hip
using a radioactive source, such as Gd 153. This radioactive
source naturally emits photons of two significantly different
energies (40 and 100 KeV).
DXA Dual energy X-ray Absorptiometry. A bone density system that
uses two X-ray energies to measure bone density, with a low
patient exposure and a fast scan time. Replaces DPA technique.
Epiphysis Endplate or growth plate. Located at the ends of the long bones
of the skeleton.
Foramen Naturally occurring hole or opening found in a bone.
e.g. Ischial foramen found in the pelvis.

Fosamax® Nonhormonal treatment for post menopausal women with low


bone density, using an aminobisphosphonate or alendronate
sodium.
Greater Trochanter A protuberance located on the lateral aspect of the femur. It is
palpable on the outer region of the thigh by rotating the leg.
Hard Drive The hard drive of the computer contains the system’s operating
software, the HOLOGIC patient database, and patient scans. In
order to view patient scans, the actual data must be present on the
hard drive.
HRT Hormone Replacement Therapy
Internal Reference Patented continuous pixel by pixel calibration throughout the
System scan, as it interposes bone and tissue equivalent material into the
beam. This ensures data compatibility and system stability.
IVA Instant Vertebral Assessment
IVA HD Instant Vertebral Assessment High Definition
Magneto Optical A high capacity device to store scan images, with a storage
Disk capacity of approximately 7,500 scans per disk.
Main Menu Appears on the system’s computer screen after the system is
turned on or boots up. From this menu, the operator has access to
all system functions.

Glossary G-3
QDR Reference Manual

Miacalcin® Nasal spray (calcitonin-salmon) manufactured by Sandoz


Pharmaceutical Corporation. see calcitonin
Normals Male and Female reference population curves used to compare
patient measurements with young and age matched normal
values.
Osteomalacia General term that defines a group of disorders in which there is
defective mineralization of bone.
Osteopenia Any condition of reduced bone mass.
Osteoporosis “A disease that is characterized by an absolute decrease in the
amount of bone to a level below that required for mechanical
support of normal activity and by the occurrence of non
traumatic skeletal deformities. Aging, menopause, lifestyle, local
factors regulating bone turnover, genetic factors, and additional
sporadic factors are considered to contribute to this bone loss.”1
Parathyroid Group of six small glands situated around the thyroid gland
Glands concerned with calcium and phosphorus regulation in the body.
Diminished function or removal of the parathyroid glands may
result in low calcium levels in the blood.
Parathyroid Potent stimulator of bone resorption. Increases bone formation
Hormone and activity.
Patient database Part of the internal HOLOGIC database containing patient
biographical information, scan results, and archive location for
each scan.
Precision Ability of an instrument to reproduce the same results in repeat
measurements.
QC Quality Control procedure used to monitor the stability of the
system.
QCT Quantitative Computed Tomography. Requires software
modification to an existing CT scanner to enable measurement
and quantitative determination of trabecular bone mineral
content in the lumbar spine. It requires the patient to lie on top of
a phantom, containing imbedded known reference material
during the scan.

QDR® Quantitative Digital Radiography. Uses an X-ray tube as a


source of dual energy photons, pulsing the X-rays on and off, to
measure the bone mineral content within a region. QDR is a
registered trademark of HOLOGIC.
Radius Large bone of the forearm, located on the thumb side.

G-4 Glossary
QDR Reference Manual

Region of interest User defined region surrounding the area of interest on a hip or
spine scan.
Scan Method of acquiring an areal measurement of a specific region
for bone mineral evaluation.
SPA Single Photon Absorptiometry. The first clinical type of bone
densitometer, using a sealed radionuclide source (I125) for a
single energy measurement. Scan time was almost fifteen
minutes and measured only the forearm.
SXA Single energy X-ray Absorptiometry. Modification of the SPA
forearm scanner, using X-rays instead of a radioactive source.
Supine Lateral Provides assessment of vertebral bone mass, without the
Spine influence of cortical bone posterior elements, or other artifacts
present in an AP spine scan.
T Score Indicates the amount of bone loss, by quantifying the difference
between the patient’s BMD at her/his current age, and the peak
bone mass achieved by young normal individuals.
Trabecular bone Spongy bone, comprising twenty percent of the skeletal mass and
very metabolically active.
Ulna Small bone of the forearm, located on the fifth finger side.
Vitamin D Active metabolite produced in the kidney. Primary effects
exerted on the gut where it increases calcium and phosphorus
absorption. Required for normal bone mineralization.
Wrist Two rows of small bones found between the hand and the
forearm.
Xiphoid Bottom of the sternum, that is palpable in the patient’s midline, at
the junction of the left and right side of the rib cage.
Z score Normalizes the patient’s BMD by assessing the amount of bone
loss compared to the expected loss for age matched peers.
1.
The evaluation of osteoporosis: Dual energy X-ray absorptimetry in clinical practice, H. Wahner and I.
Fogelman, 1994, Martin Dunitz, Ltd.

Glossary G-5
QDR Reference Manual

G-6 Glossary
Index
This section provides an alphabetical index of subjects with page number(s) indicating where
information on the subject can be found,

Content

Alphabetical index of Subjects

For Accessing and Printing Reports , 15-39 to


15-40
For Finding Records , 6-3
Numerics Alternate Scan Modes
AP Lumbar Spine Examination , 8-27 to 8-29
4500A, Control Panel , 1-7 Hip Examination , 9-29 to 9-30
4500C, Control Panel , 1-7 Lateral Scan , 12-42 to 12-44
4500SL, Control Panel , 1-7 Analysis
4500W, Control Panel , 1-7 Exiting , 7-27
Performing , 6-23
Performing the , 7-24
A Window
Abdominal aortic calcification , 13-1 AP Lumbar Spine Analysis , 8-11
About 1/3 Distal Region, forearm examination , Forearm Analysis , 10-12, 10-29
10-21 to 10-22 Hip Analysis , 9-14
Accessing window
Records , 6-2 AP scan , 12-22
Lateral scan , 12-31
Reports , 15-39 to 15-40
Whole Body examination , 11-9, 11-16
Add Quality Control Phantom, window , 5-15
Analysis Reports, Descriptions , 6-2
Field Descriptions , 5-16
Analysis Setup, Window
Adding
AP Lumbar Spine Examination , 8-12
New Ethnic Group , 19-53
Forearm Examination , 10-13
Scan Comments , 7-23 to 7-24
Hip Examination , 9-15
Adjusting the MID/UD Regions, Forearm
Analyze Scan Button , 3-8
Examination , 10-25 to 10-27
Analyzing
Adjusting Trochanter Region, His Examination , 9-
AP Lumbar Spine Analysis , 8-11 to 8-27
37
Manual Analysis , 8-15
Adjusting, Image Contrast and Brightness , 7-25 to
One-Time Auto Analysis , 8-14
7-26
Forearm Analysis , 10-12 to 10-31
Alternate Methods
Hip Analysis , 9-14 to 9-29
QDR Reference Manual

Manual Analysis , 9-17 Labeling Vertebrae , 8-23


One-Time Auto Analysis , 9-16 Marking Intervertebral Spaces , 8-21 to 8-23
Analyzing the Follow-up Scan, See Compare, Inserting a Line , 8-22
Analyzing the Follow-up Scan Line Mode , 8-21
Aortic calcification, abdominal , 13-1 point Mode , 8-23
AP Lumbar Spine Removing a Line , 8-22
Examination , 8-7 to 8-11 Selecting the Setup , 8-12
Steps to Performing Manual Analysis , 8-16
Performing the Analysis
Vertebral Bodies
Defining the ROI for Scoliotic Spine, Line
Excluding , 8-23
Mode , 8-31
Including , 8-23
Defining the ROI for Scoliotic Spine, Point
Labeling , 8-23
Mode , 8-32
Viewing Results , 8-25 to 8-27
Editing a Bone Map , 8-20 to 8-21
Viewing the Bone Map , 8-19 to 8-21
Evaluating the Bone Map , 8-20
Reanalyzing Current Scan from the Main Performing the Examination , 8-7 to 8-11
Window , 8-27 Placing Patient on Table , 8-5 to 8-6
Reanalyzing Current Scan using Auto Positioning Patient and C-Arm , 8-5 to 8-7
Analysis , 8-26 Positioning the C-Arm , 8-6
Reanalyzing Current Scan using Different Preparing the Patient , 8-2
Settings , 8-26 Reports
Viewing the Bone Map , 8-19 Generating , 8-27
AP Lumbar Spine Examination Printing , 8-27
Alternate Scan Modes , 8-27 to 8-28 Reposition the Image , 8-9 to 8-10
Analysis , 8-11 to 8-27 Reposition the Scan , 8-9 to 8-11
Manual Analysis , 8-15 Repositioning the Scan
One-Time Auto Analysis , 8-14 Reposition the Patient , 8-9
Analysis Setup Window , 8-12 Scan, Window , 8-8
Analysis, Window , 8-26 Select AP Lumbar Spine Scan Mode, Window ,
Auto Analysis Window , 8-14 8-28
Choosing a Scan mode , 8-3 Select AP Lumbar Spine Scan Type, Window ,
Choosing an Alternative Scan Mode , 8-3 8-3
Configuring Scan Parameters , 8-3 to 8-4 Select Scan Type and Mode , 8-2 to 8-3
Creating/Retrieving a Patient Biography, See Special Scanning Conditions , 8-28 to 8-34
Patient Biography Starting the AP Lumbar Spine Scan , 8-8 to 8-9
Evaluating the Image , 8-28 to 8-29 Windows
Exiting the Analysis , 8-27 Analysis , 8-14, 8-26
Exiting the Examination , 8-11 Analysis Setup , 8-12
Generating a Report , 8-27 Select AP Lumbar Spine Scan Mode , 8-28
Global ROI Toolbox , 8-18 to 8-19 Select Scan Type , 8-3
Manual Analysis , 8-15 AP Lumbar Spine Scan, Performing and Analyzing
the , 8-2 to 8-34
Method of Analysis , 8-12
One-Time Auto Analysis , 8-14 AP Reposition, function , 8-9 to 8-11
AP Scan
Performing Manual Analysis , 8-16
Analysis Setup window , 12-10
Performing the Analysis
Adjusting the Image Display , 8-16 ANalyzing the AP Scan , 12-9
Changing Vertebral Labels , 8-24 Method of Analysis , 12-9 to 12-11
Choosing the Method of , 8-11 to 8-13 Performing the AP Scan , 12-7 to 12-9
Defining the ROI , 8-11 Performing the AP Scan Analysis , 12-9 to 12-25
Editing a Bone Map , 8-19 to 8-21 Adjusting the Image Display , 12-13
Adding to the Bone Map , 8-20 Analysis Window , 12-22
Deleting a Bone Map , 8-21 Bone Map , 12-16 to 12-18
Editing a Bone Map , 8-20 Add Bone to the Bone Map , 12-17

I-2 Index
QDR Reference Manual

Deleting Bone from the Bone Map , 12-18 Managing Archive Locations , 16-27
Editing the Bone Map , 12-16 Overview , 16-2 to 16-8
Evaluating the Bone Map , 12-16 About Archiving and Restoring Scans , 16-6
Viewing the Bone Map , 12-16 Canceling a Transfer , 16-7
Defining the ROI , 12-14 to 12-16 Errors During Transfer , 16-7
Global ROI Toolbox , 12-15 How Scans are Transferred , 16-6
Line Mode, 12-15 How the System Handles Archive Data , 16-
Whole Mode, 12-15 7
Labeling the Vertebral Bodies , 12-20 to 12-21 How the System Handles Data
Including and Excluding Vertebrae from Inconsistencies , 16-7
Analysis , 12-20 Notification of Successful Transfers , 16-7
Changing Vertebrae labels, 12-21 Transfer Progress Message , 16-7
To Automatically Label Vertebrae, 12-20 Archiving Labels , 16-5
Making Intervertebral Spaces , 12-18 to 12-19 Archiving Scans
Line Mode , 12-18 to 12-19 Locations , 16-4
Inserting a Line, 12-19 Archive Location, 16-4
Removing a Line, 12-19 System Location, 16-4
Point Mode , 12-19 Copying Scans , 16-6
Verifying the Start Position , 12-22 to 12-25 Handling Read/Write CDs
Adjusting the Start Position , 12-24 Formatiing a Read/Write CD , 16-2
Centerline Warnings , 12-22 to 12-24 Hnading Read/Write CDs
Centerline Angle, 12-23 Removing a Read/Write CD , 16-3
Centerline not Calculated, 12-24 Locating Scans , 16-5
Centerline Outside of Range, 12-23 Restoring Scans , 16-6
Reanalyzing the Current Scan from the Restoring Scans , 16-16 to 16-18
QDR Main Window , 12-24 Restoring Scans from a Specified Archive
Reanalyzing the Current Scan with Differ- Location , 16-16 to 16-18
ent Settings , 12-24 Troubleshooting Advanced User Information ,
Viewing Results , 12-21 to 12-22 16-30 to 16-31
Viewing Results , 12-22 Troubleshooting Transfer Errors , 16-28 to 16-31
AP scan Problems Accessing an Archive Location , 16-
Analysis 28
window , 12-22 Transferring Scans , 16-28 to 16-29
windows Archiving Locations , 16-29 to 16-30
Analysis , 12-22 Arm, Switch/Lamp
AP/Lateral Spine examination DELPHI A and SL , 4-25
confirming scan parameters , 12-6 DELPHI C and W, Left , 4-24
AP/Lateral Spine, Examination , 12-2 to 12-44 DELPHI C and W, Right , 4-24
Archive Menu QDR 4500A and SL , 4-17
Query/Retrieve Scans , 16-22 to 16-24 QDR 4500C and W , 4-16
Archive Scans Button , 3-8 Auto Analysis
Archive Scans, deleting , 6-22 AP Lumbar Spine , 8-14
Archive Scans, Tab , 6-22 Hip , 9-16
Archiving QC Pantom Scans , 5-20, 16-10 Auto Low Density, Enable/Disable , 18-7
Archiving Scans, See Archiving, Locating and Auto QC
Restoring Scans Auto QC, Passed
Archiving, Locating and Restoring Scans Auto QC Plot
Archiving Scans , 16-8 to 16-11 About the QC Plot
Copying Scans , 16-19 to 16-21 Data Points, 5-8
To Copy Scans , 16-19 to 16-21 Mean Value at Calibration, 5-8
Locating Scans , 16-12 to 16-15 Plot Statisics, 5-8
To Locate and Restore Scans , 16-12 to 16-15 CV, 5-8

Index I-3
QDR Reference Manual

Mean, 5-8 Changing


Number of Points, 5-8 QC Phantom Details , 19-15
S.D., 5-8 QC Plot Regression , 5-20
QC Limit Value, 5-8 Sort Order , 3-11
Reference Values, 5-8 Changing the Sort Order , 3-12
Scan Dates, 5-7
Check Box , 3-7
Setup, 5-8
Checking the MID/UD Region, Forearm
Auto Whole Body Analysis, Enable/Disable , 18-7 Examination , 10-26
Choosing Alternate Scan Mode
AP Lumbar Spine Examination , 8-27
B AP/Lateral Spine Examination , 12-42
Bars, Scroll , 3-11 Hip Examination , 9-29
Biography Tab, Patient Window Choosing Menu Items , 3-8 to 3-9
Field Descriptions , 6-4 to 6-5 Choosing Scan Mode
Box, Check , 3-7 AP Lumbar Spine Examination , 8-3
Button Descriptions, Select Quality Control Scans, Forearm Examination , 10-4
See Quality Control Scans Hip Examination , 9-3
Buttons Whole Body Examination , 11-3
Command , 3-10 Choosing the Scan Mode
Dragging and Dropping onto a Button , 3-3 AP Lumbar Spine Examination , 8-3
Emergency Stop Forearm Examination , 10-4
DELPHI-A and -SL , 4-26 Hip Examination , 9-3
DELPHI-C and -W , 4-25 Whole Body Examination , 11-3
QDR 4500A and SL , 4-18 Circuit Breaker
QDR 4500C and W , 4-17 DELPHI , 4-23
Flashing QDR 4500 , 4-15
Daily QS , 3-13
Cleaning
System Backup , 3-13
Spills , B-4
Function , 3-5
the QDR System , B-2
Main Window , 3-7
Clicking the Mouse , 3-3
Radio , 3-10
Closing Menus , 3-9
Buttons, Main System
Command Buttons , 3-10
Analyze Scan , 3-7
Compare , 14-2
Archive Scans , 3-8
Analyzing the Follow-up Scan , 14-8 to 14-11
Daily QC , 3-7, 5-2 To Perform the Compare Analysis , 14-8 to
Exit , 3-8 14-11
Locate Scans , 3-7 Evaluating Baseline Scan Analysis , 14-6
Patient Exam , 3-7, 7-5, 7-9 To Reanalyze the Current Scan , 14-6
Patients , 3-8, 6-7 Evaluating the Baseline Scan , 14-4 to 14-6
Report , 3-7 Evaluating Baseline Scan Patient
Scans , 3-8, 6-16 Positioning , 14-4
System Backup , 3-8 Baseline Positioning
Acceptable, 14-5
Scanning Once with Correct Positioning,
C 14-5
Unacceptable, 14-5
Center Table, Switch/Lamp Scanning Once to Duplicate Incorrect
DELPHI-A and -SL , 4-26 Baseline Positioning, 14-5
DELPHI-C and -W , 4-24 Scanning Twice, 14-5
QDR 4500A and SL , 4-18 Reviewing Operator or Physician’s
QDR 4500C and W , 4-16 Comments , 14-4

I-4 Index
QDR Reference Manual

Generating a Rate of Change Report , 14-12 to Automatic Analysis of Spine Phantom , 5-5 to 5-
14-13 9
To Generate a Rate of Change Report , 14-12 Button , 5-2
to 14-13 Procedure , 5-2
Introduction , 14-2 Automatic System Test , 5-4
Performing the Follow-up Scan , 14-7 Position the Spine Phantom , 5-3
Performing the Follow-up Scan , 14-7 Start Daily QC , 5-2
Updating the Patient’s Biography Record , 14- System Test, Window , 5-4
7 Daily QC button , 3-7, 5-2
Restoring a Baseline Scan , 14-3 to 14-4 Daily QC, Flashing Button , 3-13
Restoring a Baseline Scan from the Daily System Operation , 4-1 to 4-8
Computer’s Hard Drive , 14-3 Circuit Breaker and Indicator
steps , 14-2 Figure , 4-2, 4-9
Components, Main Window , 3-4 Data Management, QC , 5-11
Computer Power Switch , 4-15 Data Unconsistencies, How the System Handles ,
Configuring 16-7
DICOM , 20-4 Decubitus Lateral Option, Overview , A-4
Confirming Scan Parameters Deleting
AP Lumbar Spine Examination , 8-3 to 8-4 Archive Scans , 6-22
Forearm Examination , 10-4 to 10-5 Reference Curve Records , 19-52
Hip Examination , 9-5 Scan Files , 6-22
Whole Body Examination , 11-4 Scans, Window , 6-21
Confirming scan parameters Unarchived Scans , 6-22
AP/Lateral Spine examination , 12-6 Description
Confirming the Patient Information , 7-6 Analysis Report , 6-2
Context Sensitive Help , 3-14 Patient Record , 6-2
Context Sensitive Help, System Software Help Detail Report, Usage Program , 19-6 to 19-8
Online , 3-14 DICOM , 20-2 to 20-6
Copying Reference Curve Records , 19-49 Closing DICOM , 15-38
Copying Scans , 16-6 Configuring , 20-4
Copying Scans, See Archiving, Locating and Printing the DX Report , 15-37
Restoring Scans Reports , 15-18 to 15-38
Correcting a Failed System Test , 5-5 Choosing the Report Type , 15-19 to 15-22
Creating Creating and Sending , 15-18 to 15-19
A Patient Record , 6-9 to 6-11 Entering the Accession Number and User
New Phantom Record , 5-15 to 5-17 Definable Entries , 15-24 to 15-25
Patient Record , 7-4 Viewing a Scans’s Details , 15-22 to 15-24
Reference Curve Record , 19-44 to 19-47 Troubleshooting , 20-5 to 20-6
Creating/Retrieving A Patient Biography , 7-4 Using , 20-4
AP Lumbar Spine Examination , 8-2 Using DICOM
Forearm Examination , 10-2 Deleting a DICOM Report from the Queue ,
Hip Examination , 9-2 15-34 to 15-35
Entering the Accession Number and User
Whole Body Examination , 11-3
Definable Entries , 15-24 to 15-25
Creating/Retrieving a Patient Record, See QDR
Previewing a DICOM Report , 15-25 to 15-26
Patient Examination, Creating/Retrieving a
Printing a DICOM Report , 15-26 to 15-27
Patient Record
Refreshing DICOM Reports in the Queue , 15-
33 to 15-34
Saving a DICOM Report , 15-27 to 15-29
D Sending a DICOM Report , 15-29 to 15-30
Daily QC Sorting the Scan List , 15-30

Index I-5
QDR Reference Manual

Sselecting/Deselecting All Scans , 15-31 to Dual Hip


15-32 Choosing the Default Scan Mode , 9-4
Viewing a History of Sent Reports , 15-32 Exiting the Analysis , 9-29
Viewing a History of Sent Storage Positioning The C-Arm , 9-8
Commitment Reports , 15-36 Positioning The Patient , 9-7 to 9-8
Viewing the Queue , 15-32
Repositioning for , 9-12 to 9-13
Discovery System Introduction , 1-1 to 1-12 DXA Results Report , 15-5

QDR for Windows XP Software , 1-9 to 1-10


Main Window Features , 1-9 to 1-10
Daily Function buttons, 1-10 E
File Cabinet buttons, 1-10 Editing a Patient Record , 6-13
Help, 1-10 Emergency motion, Function , 19-40
Main Work Area, 1-9 Emergency Stop Procedure
Menu Bar, 1-9
DELPHI Systems , 4-29
System Function buttons, 1-10
QDR Systems , 4-21
System Message ticker tape, 1-10
Main Window Overview , 1-9 Entering
System Hardware Components , 1-3 to 1-7 Insurance Information , 6-12
C-Arm , 1-5 Patient Information , 6-11
Laser, 1-5 Text , 3-9
X-Ray Source and Detector, 1-5 Entering Forearm Length, Forearm Examination ,
Control Panel , 1-6 10-15
Discovery Control Panels, 1-6 Entering Text , 3-9
Discovery A/SL Control Panel Error Messages
Figure, 1-6 When Running System backup , 17-6 to 17-8
Discovery C/Ci/W/Wi Control Pane Ethnic Groups, Adding New , 19-53 to 19-54
Figure, 1-6 Evaluating a Baseline Scan, See Compare,
Explorer Control Panel Evaluating a Baseline Scan
Figure, 1-6
Evaluating Image
Switches and Indicators, 1-6
Forearm Examination , 10-32 to 10-33
Examination Table , 1-4 to 1-5
Table Pad, 1-4 Hip Examination , 9-30 to 9-32
Operator Console , 1-4 Whole Body Examination , 11-18
Computer Hardware and Peripherals, 1-4 Examinations
Models , 1-2 to 1-3 AP Lumbar Spine , 8-2 to 8-27
Discovery-A , 1-2 Choosing an Alternate Scan Modes , 8-3
Discovery-C , 1-2 Confirming Scan Mode , 8-3 to 8-4
Discovery-Ci , 1-2 Creating/Retrieving Patient Biography , 8-2
Discovery-SL , 1-2 Exiting the Examination , 8-11
Discovery-W , 1-2 Methods of Analysis , 8-12
Discovery-Wi , 1-2 Performing Manual Analysis , 8-16
Explorer , 1-2 Performing the Analysis , 8-11 to 8-27
Disinfecting the Table Pad , B-3 Adjusting the Image Display , 8-16
Display, Usage Choosing the Method of Analysis , 8-11 to
8-13
Detail Report , 19-6 to 19-8
Defining the ROI , 8-16 to 8-19
Summary Report , 19-4 to 19-6
Selecting the Setup , 8-12
Double-clicking the Mouse , 3-3 Steps of the Analysis , 8-16
Dragging and Dropping , 3-3 Performing the Examination , 8-7 to 8-11
Onto the Record Button , 15-39 Positioning the Patient and C-Arm , 8-4 to 8-7
Patient Records , 15-39 Placing the Patient on the Table , 8-5
Scan Records , 15-40 Positioning the C-Arm , 8-6
Drop-down List , 3-10 Positioning the Patient , 8-5

I-6 Index
QDR Reference Manual

Preparing the Patient , 8-2 Choosing Dual Hip , 9-4


Repositioning the Scan , 8-9 to 8-11 Choosing Left or Right Hip , 9-3
Repositioning the Image , 8-9 to 8-11 Choosing the Scan Mode , 9-3
Repositioning the Patient , 8-9 Confirming Scan Parameters , 9-5
Selecting the Scan Type and Mode , 8-2 to 8-4 Creating/Retrieving Patient Biography , 9-2
Starting the AP Lumbar Spine Scan , 8-8 to 8- Dual Hip, Repositioning the Scan , 9-11 to 9-
9 12
AP/Lateral Spine , 12-2 to 12-44 Evaluating the Image , 9-30 to 9-32
confirming scan parameters , 12-6 Exiting the Analysis , 9-29
Forearm Examination , 10-2 to 10-32 All Scans Except Dual Hip , 9-29
About 1/2 Distal Region , 10-21 Dual-Hip Scans , 9-29
About Ulma/Radius Divider , 10-27 Exiting the Examination , 9-13
Choosing Scan Mode , 10-4 Generating Reports , 9-29
Confirming Scan Parameters , 10-4 to 10-5 Methods of Analysis , 9-14 to 9-15
Creating/Retrieving Patient Biography , 10-2 Perform the Analysis , 9-14 to 9-29
Entering Forearm Length , 10-15 Adjusting the Image Display , 9-18
Evaluating the Image , 10-32 to 10-33 Bone Map , 9-21 to 9-24
Exiting the Analysis , 10-32 Adding Bone to the Bone Map, 9-23
Exiting the Examination , 10-11 Deleting Bone from the Bone Map, 9-24
Generating Reports , 10-32 Determining a Starting Point when Deleting
Measuring the Patient’s Forearm , 10-4 Bone, 9-24
Performing the Analysis , 10-12 to 10-31 Editing the Bone Map, 9-22
Adjusting the Display image , 10-15 Evaluating Bone Map, 9-22
Adjusting the MID/UD Regions , 10-25 to Acceptable Bone Map, 9-22
10-28 Unacceptable Bone Map, 9-22
The MID/UD Toolbox, 10-26 to 10-28 Viewing the Bone Map, 9-21
About the Ulma/Radius Divider, 10-27 Choosing the Method of Analysis , 9-14 to
Adjusting the MID/UD Region, 10-27 9-15
Automatically Repositioning the Defining the ROI , 9-18 to 9-21
Ulma.Radius Divider, 10-27 Line Mode, 9-20 to 9-21
Checking the MID/UD Regions, 10-26 Whole Mode, 9-21
Manually Repositioning the Ulma.Radius Hip Structural Analysis , 9-38 to 9-40
Divider, 10-28 HSA Toolbox, 9-39
Bone Map , 10-22 to 10-25 Manually adjusting HSA Regions, 9-39 to
Adding Bone to the Bone Map, 10-24 9-40
Deleting Bone from the Bone Map, 10-25 Viewing Results, 9-41
Editing Bone Map, 10-23 Position the Neck Box , 9-24 to 9-27
Evaluating Bone Map, 10-23 Adjusting the Neck Box, 9-26 to 9-27
Viewing the Bone Map, 10-22 Line Mode
Choosing the Method of Analysis , 10-12 to Move Neck Box away from Femoral
10-13 Head, 9-27
Defining the ROI , 10-16 to 10-22 Move Neck Box away from Ischum, 9-
Generating Results , 10-29 26
Reanalyzing the Current Scan from the Recalculate the Position of All Hip Regions
QDR Main Window , 10-31 via Auto Position, 9-25
Reanalyzing the Current Scan with Differ- Whole Mode, 9-26
ent Settings , 10-31 Viewing Results , 9-27 to 9-29
Steps to Analyze the Forearm Scan , 10-14 Reanalyze the Current Scan from the QDR
Performing the Examination , 10-7 to 10-11 Main Window, 9-29
Hip Examination , 9-2 to 9-29 Reanalyze the Current Scan with Different
Adjusting the Trochanter Region , 9-37 Settings, 9-29
Alternative Scan Modes , 9-29 to 9-30 Perform the Examination , 9-8 to 9-13
Choosing an Altenative Scan Modes , 9-3 Performing Manual Analysis , 9-17
Placing the Patient on the Table , 9-7

Index I-7
QDR Reference Manual

Positioning for Dual-Hip Scan , 9-8 Field, Text , 3-7, 3-9


Positioning the C-Arm , 9-8 Filing Report , 15-2
For Dual-Hip , 9-8 Example , 15-13
For Left or Right Hip , 9-8 example , 15-13
Positioning the Patient , 9-7 Finding Records, Alternate Procedure , 6-3
Positioning the Patient and C-Arm , 9-6 to 9-8
Follow-up, See Compare
Preparing the Patient , 9-2
Printing Reports , 9-29
Reposition the Image , 9-10 to 9-12
Reposition the Patient , 9-12 G
Reposition the Scan , 9-10 to 9-12 Generating
Selecting Scan Type and Mode , 9-3 to 9-4 AP Lumbar Spine examination
Starting the Hip Scan , 9-9 to 9-10 reports , 8-27
Verifying Scan Parameters , 9-5
reports , 8-27
Exit Button , 3-8
Generating a Rate of Change Report, See Compare,
Exit the Analysis. Window , 7-27 Generating a Rate of Change Report
Exit the Examination, Window , 7-23 Generating and Printing a Report, See QDR Patient
Exiting , 7-27 Examination, Generating and Printing a
The Analysis , 7-27 Report,
The AP Lumbar Spine Examination , 8-11 Generating reports
The Examination , 7-23 AP Lumbar Spine examination , 8-27
Operator Options , 7-23
The Forearm Examination , 10-11
The Hip Examination , 9-13 H
Exiting Examination, Operator Options , 7-23
Exiting the Analysis , 7-27 Help, See System Software Help
AP Lumbar Spine Analysis , 8-27 Hip examination
Forearm Analysis , 10-32 Analysis
Hip Analysis window , 9-28
For All Scans Except Dual-Hip , 9-29 choosing scan mode , 9-3
For Dual-Hip Scans , 9-29 creating/retrieving patient biography , 9-2
Hip Examination performing the analysis
For All Scans Except Dual-Hip , 9-29 moving neck box away from the femoral
Exiting the Analysis, See QDR Patient head , 9-27
Examination, Exiting the Analysis repositioning the image , 9-10
Exiting the Examination, See QDR Patient repositioning the patient , 9-11
Examination, Exiting the Examination windows
Analysis , 9-28
Hip Structure Analysis , 9-38
F
Failed System Test, Correcting , 5-5
Field Descriptions
I
Add Quality Control Phantom, Window , 5-16 Identifying the Sort Order , 3-11
Patient Window Instant Vertebral Assessment See QM
Biography Tab , 6-4 to 6-5
Insurance Tab , 6-6
Quality Control Plot Parameter, Window , 5-12 K
Quality Control Plot Setup, Window , 5-18
Key Elements of the QDR System , 2-2
Scan property Window, Identification Tab , 6-19 Accuracy and Precision , 2-3
to 6-20
DXA Technology , 2-2
Scan Records , 6-17

I-8 Index
QDR Reference Manual

Principles of Operation , 2-3


O
On line help , 3-14
L Operator Maintenance , B-1 to B-4
Cleaning Accidental Spills , B-4
Lateral scan Periodic Cleaning , B-2
Analysis Cleaning the QDR and Computer
window , 12-31 Components , B-2
windows Cleaning the Table Pads , B-2
Analysis , 12-31 Disinfecting the Table Pad , B-3
List items Periodic Visual Inspection , B-3
selecting , 3-12 Checking Cables and Outlets , B-3
multiple , 3-12 Checking for Adequate Clearance , B-3
one , 3-12 Inspecting the Table Pad , B-3
range of , 3-12 Options , A-1 to A-4
Locating Scans, See Archiving, Locating and Body Composition Analysis , A-2
Restoring Scans DAP Meter , A-4
General Region of Interest , A-3
High Power Whole Body , A-3
M Infant Whole Body , A-4
Main window List , A-2
components , 3-7 Low Density Spine , A-4
Managing Network , A-4
patient records , 6-1 Prosthetic Hip , A-3
Measuring the Patient’s Forearm , 10-4 Small Animal , A-3
Menu items Worklist , A-3
choosing , 3-8 Other Manufacture’s System, Adding , 6-24 to 6-
Menus , 3-8 to 3-9 27
Choosing Menu Items , 3-8 Entering data for Hip , 6-27
Closing , 3-9 Entering data for Spine , 6-26 to 6-27
closing , 3-9 Entering data from , 6-24 to 6-26
Using , 3-8 Other Window Components , 3-9 to 3-12
using , 3-8 Changing the Sort Order , 3-12
Using Sub-Menus , 3-9 Check Boxes , 3-11
Models , 1-2 to 1-3 Command Buttons , 3-10
4500A , 1-2 Drop-down Lists , 3-10
4500C , 1-3 Entering Text , 3-9
4500SL , 1-2 Identifying the Sort Order , 3-11
4500W , 1-3 Radio Buttons , 3-10
DELPHI-A , 1-3 Scroll Bars , 3-11
DELPHI-SL , 1-3 Sort List items , 3-11
Discovery-A , 1-2 Tabs , 3-10
Discovery-C , 1-2 Other window components
Discovery-Ci , 1-2 understanding , 3-9
Discovery-SL , 1-2
Discovery-W , 1-2
Discovery-Wi , 1-2 P
Explorer , 1-2 Patient
Mouse, See Understanding the Mouse examination , 7-1
Patient Record, Introduction , 6-2 to 6-3

Index I-9
QDR Reference Manual

Accessing the Records , 6-2 to 6-3 AP Lumbar Spine


Patient Record, Updating a Patient Record , 6-12 to editing a bone map , 8-20
6-14 evaluating a bone map , 8-20
Patient Record, Working with , 6-4 to 6-15 viewing a bone map , 8-19
Creating a Patient Record , 6-9 to 6-10, 7-4 to 7- Hip examination
7 moving neck box away from the femoral
Creating a New Patient Record , 6-10 head , 9-27
Entering Insurance Information , 6-12 Performing the Analysis, See QDR Patient
Entering Patient Biographical Information , Examination, Performing the Analysis
6-11 Performing the Examination, See QDR Patient
Overview , 6-4 to 6-6 Examination, Performing the Examination
Biography Tab , 6-4 to 6-5 Performing the Follow-up Scan, See Compare,
Window , 6-4 Performing the Follow-up Scan
Body Mass Index, 6-5 Phantom
DOB, 6-5 Spine , 5-3
Ethnicity, 6-5 positioning the , 5-3
First Name, 6-4 Positioning
Height/Weight, 6-5 the Phantom Spine , 5-3
Identifier 2, 6-5
Positioning the Patient and C-Arm, See QDR
Last Name, 6-4
Patient Examination, Positioning the Patient
Menopause Age, 6-5
and C-Arm,
MI (Middle Initial), 6-4
Power failure
Patient Comment, 6-5
operator action during
Patient ID, 6-5
QDR 4500 , 4-20
Referring Physician, 6-5
Sex, 6-5 Printing
Insurance Tab , 6-6 AP Lumbar Spine examination
Window , 6-6 reports , 8-27
Address, 6-6 reports , 8-27
Address 2, 6-6 Printing reports
City, 6-6 AP Lumbar Spine examination , 8-27
Group, 6-6
Insurance, 6-6
Name, 6-6
Patient ID, 6-6
Q
Plan, 6-6 QC Data Management , 5-11 to 5-19
Postal Code, 6-6 QC Plot , 5-7 to 5-9
State, 6-6 selecting scans to include , 5-13
Retrieving a Patient Record , 6-7 to 6-9 QC Plot Progression , 5-20
Go to the Patient Drawer , 6-7 QC Spine Phantom , 5-10
Searching the Patient Drawer , 6-7 to 6-9 QDR 4500
Using the Scroll Bar , 6-8 main window , 6-22
using the Search Text Field , 6-9 QDR Patient Examination , 7-2 to 7-33
Updating a Patient Record Creating/Retrieving a Patient Biography , 7-4 to
Editing a Patient Record , 6-13 7-7
Performing , 14-7 Confirming the Patient Information , 7-6
Performing analysis Creating a Patient Biography , 7-4
Forearm examination Retrieving a Patient Biography , 7-4
steps to , 10-14 Retrieving a Patient File , 7-4
the AP Lumbar Spine examination Searching for a Patient’s Name , 7-5
selecting the setup , 8-12 Exiting the Analysis , 7-27
steps to , 8-16 Exit Analysis Window , 7-27
Performing the analysis Analyze Another Scan button , 7-27

I-10 Index
QDR Reference Manual

Exit button , 7-27 Specifications , 13-2


New Scan button , 7-27 IVA Image Viewer , 13-55 to 13-60
Report Button , 7-27 DE Scan , 13-59 to 13-60
Exiting the Examination , 7-23 to 7-24 The IVA Viewer , 13-55 to 13-59
Adding Scan Comments , 7-24 The Image , 13-55 to 13-56
Exit Exam Window , 7-23 Mouse Pointer Control, 13-58
Add Scan Comments , 7-23 The Control Bar, 13-56 to ??, 13-56 to 13-
Analyze Scans , 7-23 58
New Scan , 7-23 Cancel button, 13-58
Generating and Printing Reports , 7-32 to 7-33 Close button, 13-58
Configuration the Report , 7-32 DE Scan button, 13-56
Editing Physician’s Comments , 7-33 Invert button, 13-57
Generating a Report , 7-32 Markers button, 13-56
Performing the Analysis , 7-24 to 7-26 Print Report button, 13-57
Contrast and Brightness , 7-25 to 7-26 Revert button, 13-57
Adjusting Contrast and Brightness , 7-26 W-L Control, 13-57
Performing the Examination , 7-22 Zoom Control button, 13-57
Positioning the Patient and C-Arm , 7-21 to 7-22 The Control Print Image button, 13-57
Placing the Patient on the Table , 7-21 The Patient Data, 13-56
Positioning for the Specific Examination , 7- IVA Imaging on an A or SL System , 13-4 to 13-
22 20
Preparing the Patient , 7-2 to 7-4 Creating/Retrieving a Patient Biography , 13-
Interview , 7-2 5
Patient Questions , 7-2 Preparing the Patient , 13-5
Preparation , 7-3 Selecting the Scan Type and Mode , 13-6 to
13-9
Selecting the Scan Type and Mode , 7-21
To select IVA Imaging scan modes , 13-7 to
QDR User Interface , 3-4 to 3-7
13-9
Dialog Windows , 3-6 Starting the Examination , 13-5 to 13-6
Main Window , 3-4 The AP IVA Scan
Tab Windows , 3-5 Conforming Scan Parameters for AP IVA
QM , 13-2 to 13-76 Scan , 13-10
Generating and Printing Images and Reports , To Verify Scan Parameters, 13-10
13-69 to 13-73 Performing the AP IVA Scan , 13-11 to 13-
Printing IVA Images from the Reports 12
Button , 13-69 to 13-71 Positioning the Patient for the AP IVA
Sample Printouts , 13-72 to 13-73 Scan , 13-10
Sample Image , 13-72 Repositioning the Scan , 13-12 to 13-14
Sample Report , 13-73 The Lateral IVA Scan , 13-15 to 13-20
Imaging on an A or SL System Confirming Scan Patameters for Lateral
Selecting the Scan Type and Mode IVA Scan , 13-15 to 13-16
To run the IVA Imaging examination using To Verify Scan Parameters, 13-16
defaults , 13-7 Performing the Lateral IVA Scan , 13-16 to
The AP IVA Scan , 13-9 to 13-15 13-17
Interpreting the IVA Image , 13-74 to 13-76 Repositioning the Scan , 13-17 to 13-19
Introduction , 13-2 to ?? Returning the C-Arm to the AP Position ,
Disclaimer on IVA Scan , 13-4 13-20
Patient Position and Scanning , 13-2 to 13-3 Using the IVA Viewer for Lateral IVA
Patient Position on an A or SL System , 13- Scan , 13-19
2 IVA Imaging on an C or W System , 13-21 to 13-
Patient Position on an C or W System , 13-3 37
Positioning Aids for the C and W System Creating/Retrieving a Patient Biography , 13-
Head Rest , 13-4 21

Index I-11
QDR Reference Manual

Preparing the Patient , 13-21 Exiting the AP BMD Analysis , 13-47 to


Selecting the Scan Type and Mode , 13-22 to 13-48
13-24 Performing the AP BMD Scan , 13-47
To run the IVA Imaging examination using The AP IVA Scan , 13-43 to 13-46
Defaults , 13-23 Confirming the Scan Parameters for AP
To select IVA Imaging scan modes , 13-23 IVA Scan , 13-44
to 13-24 Performing the AP IVA Scan , 13-44 to 13-
Starting the Examination , 13-21 to 13-22 46
The AP IVA Scan , 13-24 to 13-30 Repositioning, 13-46
Confirming Scan Parameters for AP IVA Using the IVA Image Viewer for the AP
Scan , 13-25 IVA Scan , 13-46
Position the Patient for the AP IVA Scan, 13- The Lateral BMD Scan , 13-48 to 13-50
25 Confirming Scan Parameters for Lateral
Exiting the IVA Image Viewer , 13-29 BMD Scan , 13-48
Performing the AP IVA Scan , 13-26 to 13- Exiting the Lateral BMD Scan , 13-50
29 Performing the Lateral BMD Scan , 13-49
Repositioning the Scan, 13-27 to 13-29 to 13-50
To Start the AP IVA Scan, 13-26 The Lateral IVA Scan , 13-51 to 13-54
Using the IVA Image Viewer for AP IVA Confirming Scan Parameters for Lateral
Scan , 13-29 IVA Scan , 13-51
The Lateral IVA Scan , 13-30 to 13-37 Performing the Lateral IVA Scan , 13-52 to
13-53
Performing the Lateral IVA Scan, 13-33 to Returning the C-Arm to the AP Position, 13-
13-36 54
To start the Lateral IVA Scan, 13-33 to Using the IVA Image Viewer for the Lateral
13-34 IVA Scan, 13-53
Repositioning the Scan, 13-34 to 13- The IVA Examination , 13-4 to 13-54
36 IVA Image Exam , 13-4
Placing the Patient on the Table, 13-31 to IVA with BMD Exam , 13-4
13-33 Using Markers , 13-61 to 13-68
Positioning the Patient, 13-31 to 13-32 QM Result Block , 13-66
To position the C-Arm, 13-32 QM QM Results Block for AP IVA Scans ,
To position the patient, 13-32 13-68
To use the positioning aids properly, The Object of QM MaRKERS , 13-61
13-32 Using QM Markers , 13-61 to 13-68
Positioning the Patient for the Lateral IVA Adding QM Markers , 13-63
Scan, 13-31 Deleting Marker Sets , 13-66
Confirming Scan Parameters for Latera Editing Markers , 13-64
IVA Scan , 13-30 to 13-31 Flipping the Image , 13-62
Using the IVA Image Viewer for Lateral IVA Labeling Marker Sets , 13-65
Scan , 13-36 Moving Markers , 13-64
IVA With BMD , 13-38 to 13-54 Selecting Markers , 13-64
Creating/Retrieving a Patient Biography , 13- QM
38 to 13-39 Using Markers
Preparing the Patient , 13-38 QM Result Block
Selecting the Scan Type and Mode , 13-40 to QM Results Block for Lateral IVA Scans ,
13-43 13-67
To Run the IVA with BMD examination us- Quality Control Plot Parameters
ing defaults , 13-40
window , 5-12
To select IVA Imaging scan modes , 13-41
Quantitative Morphometry See QM
to 13-43
Query/Retrieve Scans , 16-22 to 16-24
The AP BMD Scan , 13-46 to 13-48
Confirming Scan Parameters for AP BMD Query Parameters , 16-23 to 16-24
Scan , 13-47

I-12 Index
QDR Reference Manual

Printing a Report Using the Report Button ,


R 15-6 to 15-9
Radio Buttons , 3-10 Editing the Physician’s Comments , 15-8 to
Rate of Change Report 15-9
example , 15-15 To Print a Report Using the Report Button ,
Reference Database , 2-8 15-7
Reminders Repositioning the image
system software , 3-13 Forearm examination , 10-9
flashing buttons , 3-13 Hip examination , 9-10
Reports , 15-2 to 15-40 Repositioning the patient
Alternate Navigation , 15-39 to 15-40 Forearm examination , 10-10
Drag and Drop Records , 15-39 to 15-40 Hip examination , 9-11
Patient Records , 15-39 Restoring a Baseline Scan, See Compare,
Scan Records , 15-40 Restoring a Baseline Scan
DICOM , 15-18 to 15-38 Restoring Scans, See Archiving, Locating and
Creating and Sending , 15-18 to 15-19 Restoring Scans
Examples , 15-13 to 15-16 Retrieving
Filing Report , 15-13 scan records , 6-22
Interpreting Report , 15-14
Rate of Change Report , 15-15
Referring Report , 15-15
Introduction , 15-2 to 15-5
S
Report Information Blocks , 15-3 to 15-5 Scan Record, Introduction , 6-16
Report Blocks, Window , 15-4 Scan Drawer , 6-16
DXA Results, 15-5 Scan Record, Working with , 6-17 to 6-23
Rate of Change Graph, 15-5 Retrieving Scan Records , 6-22 to 6-23
Reference Graph, 15-5 Performing an Analysis , 6-23
Header, 15-4 To go to the Scan Drawer , 6-22
Image, 15-5 To View a Scan , 6-23
Patient Information, 15-4 Scan Deleting Scan Files , 6-21 to 6-22
Physician’s Comments, 15-5 Archive Scans Tab , 6-22
Scan information, 15-4 Unarchive Scans Tab , 6-22
Report Types , 15-2 to 15-3 Scan Details , 6-17 to 6-20
Filing Report , 15-2 Details Tab , 6-18 to ??
Interpreting Reports , 15-2 Accession Number , 6-18
Rate of Change Report , 15-3 Height , 6-18
Reference Report , 15-2 HL7 Field 1 , 6-18
Preview Reports , 15-11 to 15-12 HL7 Field 2 , 6-18
Report Preview, Window , 15-11 HL7 Field 3 , 6-18
Close button , 15-12 Operator , 6-18
Next Page button , 15-11 Patient Name , 6-18
Prev Page button , 15-11 Scan Comments , 6-18
Print button , 15-11 Weight , 6-18
Two Pages button , 15-11 Identification Tab , 6-19 to 6-20
Zoom In button , 15-12 Analysis Date , 6-20
Zoom Out button , 15-12 Archive 1 Date , 6-20
Printing Reports , 15-6 to 15-10 Archive 2 Date , 6-20
Editing the Physician’s Comments , 15-8 Machine Type , 6-19
Print Dialog Box , 15-10 Patient ID , 6-19
To Select the Print Dialog Setting , 15-10 Patient Name , 6-19
Printing a Report After Analysis , 15-6 Scan Date , 6-19
Printing a Report Print Dialog Box , 15-10 Scan ID , 6-19
Scan Mode , 6-19

Index I-13
QDR Reference Manual

Scan Type , 6-19 Running the System Backup , 17-5 to 17-6


Serial Number , 6-19 System Recovery , 17-8
Window , 6-20 System Backup, Flashing Button , 3-13
Scan Sites , 2-4 to 2-7 System Backup, See System Backup and Recovery
Forearm , 2-6 System Configuration , 18-2 to 18-81
Lumbar Spine , 2-4 Analyze Tab , 18-5 to 18-7
Proximal femur , 2-5 Analysis Options
Whole Body , 2-7 , 18-7
Scoliotic Spine Scan, AP Lumbar Spine Default , 18-5
Examination , 8-30 to 8-34 Changing a Configuration Selection on a
Adjusting the ROI Box , 8-30 to 8-32 Default Setting , 18-5
Defining ROI using Line Mode , 8-31 General Options , 18-6
Defining ROI using Point Mode , 8-32 Display Region of Interest Coordinates ,
Marking Interventebral Lines , 8-33 to 8-34 18-6
Marking Interventebral Spaces , 8-33 Enable Body Composition Analysis , 18-6
Image Display Options , 18-6
Pivoting Interventebral Lines , 8-33 to 8-34
Display Hip Axis Length , 18-6
Pivoting Intervertebral Line , 8-33 to 8-34
Display Pseudo Color Image , 18-6
Scroll Bars , 3-11 Image Pro , 18-6
Selecting Items Initially Display Unanalyzed Hip Image as
Selecting a Range of Items , 3-12 Single Energy , 18-6
Selecting Multiple Items , 3-12 Initially Display Unanalyzed Lateral Image
Selecting One Item from a List , 3-12 as Single Energy , 18-6
Selecting scan type and mode Image Style
Whole Body examination , 11-3 , 18-7
Selecting the Scan Type and Mode, See QDR Lateral Options , 18-7
Patient Examination, Selecting the Scan Type Display MID Regions , 18-7
and Mode Prevent Vertical Motion of ROI , 18-7
Sort List Items , 3-11 Archive Tab , 18-31 to 18-33
Study
Configuring. See System Configuration, Study Adding a New Location , 18-31 to 18-32
Tab Adding a New Location to the Archive List,
Using See Using a Study 18-31
Changing the Default location, 18-32
System Backup and Recovery , 17-2 to 17-8
Deleting a Location from the Archive List,
Errors and Exceptions , 17-6 to 17-8
18-32
Backup File with the Same Name , 17-6
Including Raw Data Files, 18-33
Drive Not Accessible , 17-7
CAD BMD Tab , 18-8 to 18-9
Drive Does Not have Enough Space , 17-7
Configuration Settings , 18-8
3.5 inch diskette, 17-7
Typical CAD BMD Dialog , 18-9
Network Drive, 17-7
DICOM Tab , 18-44 to 18-66
Read/Write CD, 17-7
Host Machine , 18-66
SuperDisk, 17-7
Modality Worklist , 18-44 to 18-52
Drive not Accessible , 17-8
Auto Query Interval , 18-47
Overview , 17-2 to 17-5
Configure Worklist Provider , 18-49 to 18-
Files Saved , 17-2
52
Media Formatting , 17-3
Input From File , 18-48
Management of Media , 17-3
Map Worklist Fields , 18-49, 18-51 to 18-52
Floppy Diskette , 17-3
Performed Procedure Step , 18-49 to 18-50
Network Drive , 17-3
Provider, 18-49 to 18-50
Read/Write CD , 17-3
Purge Interval , 18-48
SuperDisk , 17-3
Query Parameters , 18-46
Naming the Backup File , 17-4
Query Retry Parameters , 18-47 to 18-48
When to Run System Backup , 17-3

I-14 Index
QDR Reference Manual

Use Performed Procedure Step , 18-49 Reference Block Lateral Configuration, 18-
Worklist Provider , 18-49 29
Query/Retrieve Destinations , 18-63 to 18-65 Configuring the Ancillary Report , 18-30
Add / Edit DICOM Query/Retrieve Desti- Scan Conversion , 18-39 to 18-40
nation Dialog , 18-65 Adding Additional Manufacturers , 18-40
Send Destinations , 18-53 to 18-58 Study Tab , 18-68 to 18-81
Add / Edit DICOM Send Destination Adding a Study , 18-68 to 18-71
Dialog , 18-55 to 18-57 Deleting a Study , 18-77 to 18-78
Add / Edit Storage Commitment Destina- Editing an Existing Study , 18-72 to 18-74
tion Dialog , 18-61 Adding Items , 18-72 to 18-74
Configure DICOM Send Dialog , 18-57 to Deleting Items , 18-74 to 18-77
18-58 Locating a Study from a Long List , 18-81
Storage Commitment Destinations , 18-59 to Renaming a Study , 18-78 to 18-80
18-62 System Tab , 18-3 to 18-4
Configure Storage Commitment Dialog , System Tab Window
18-62 Autopositioning for Hip Scans , 18-3
HL7 Tab , 18-34 to 18-38 Customer , 18-3, 18-4
Changing HL7 Fields , 18-36 to 18-37 Data Directory , 18-4
Entering HL7 Fields , 18-34 to 18-35 Date Format , 18-3
Removing HL7 Fields , 18-37 to 18-38 Default Ethnicity , 18-4
Overview Language , 18-4
Entering System Configuration , 18-2 Last Service Date , 18-4
Privacy Tools , 18-67 Last System Backup Date , 18-4
QDR Modality Worklist , 7-8 to 7-20 Model , 18-4
Modality Worklist Operations , 7-16 to 7-20 Patient Biography , 18-3
Adding a Patient , 7-17 to 7-18 Report Dose Area Product , 18-3
Adding a Study , 7-18 S/N , 18-4
Matching an Existing Patient , 7-18 to 7-20 Software , 18-4
Using Modality Worklist , 7-8 to 7-13 System ID , 18-4
Using Modality Worklist with a Study , 7-13 System Hardware Components
to 7-16 AC Power isolation Module , 1-4
Report Tab , 18-10 to 18-30 DELPHI Systems , 4-23 to 4-30
Configuring Reports , 18-14 to 18-30 AC Power Isolation Module , 4-23
Report Blocks Window , 18-10 Circuit Breaker , 4-23
Report Configuring Reports Indicator , 4-23
Configure the Optional Reports Control Panels , 4-23 to 4-26
Configuring the Common Block, 18-19 to DELPHI-A and -SL , 4-25 to 4-26
18-20 Arm, 4-25
Configuring the Header Block, 18-21 Center Table, 4-26
Configuring the Optional Reports of a Emergency Stop button, 4-26
Report Type, 18-19 Emergency Table Lift, 4-26
Configuring the Physician’s Comments, Enable Lateral, 4-26
18-22 Laser, 4-26
Configuring the Reference Block, 18-20 Patient On/Off, 4-26
Details of the Patient Block, 18-23 to 18-24 Power Switch, 4-26
Details of the Reference Block for Configure Table, 4-26
Common Block Option, 18-26 to 18- X-ray on Lamp, 4-26
27 DELPHI-C and -W , 4-23 to 4-25
Details of the Results Block, 18-25 to 18-26 Arm Left, 4-24
Details of the Scan Block, 18-24 to 18-25 Arm Right, 4-24
Reference Block Forearm Configuration, 18- Center Table, 4-24
28 Emergency Stop button, 4-25
Reference Block Hip Configuration, 18-27 Instrument On/Off, 4-25

Index I-15
QDR Reference Manual

Laser, 4-24 Operator Action During a Power Failure ,


Laser On/Motion Enabled (green indicator), 4-20 to 4-21
4-25 After Power is Restored, 4-21
Patient On/Off, 4-24 Shutting down, 4-20
Table In, 4-24 Power Module , 4-14 to 4-15
Table Out, 4-24 Rear Panel , 4-15
X-ray On (amber indicator), 4-24 Side Panel , 4-14
Emergency Procedures , 4-28 to 4-30 Circuit Breakers, 4-15
If Equipment Fails while Operating , 4-29 Computer Power Switch, 4-15
Emergency Stop Procedure., 4-29 Instrument Power Switch, 4-15
If Power has been Turned Off , 4-30 X-ray Emissions Ind., 4-15
Applying Power to System, 4-30 X-ray Enable Key, 4-15
Operator Action During a Power Failure , X-ray Enable Lamp, 4-15
4-28 to 4-29 QDR 4500C and W
After Power is Restored, 4-29 Control Panel
Shutting Down, 4-29 Arm, 4-16
Turning Power On and Off , 4-26 to 4-28 Turning Power On and Off , 4-18 to 4-20
Beginning of the Workday , 4-27 Beginning of the Workday , 4-18
End of the Workday , 4-27 End of the Workday , 4-19
Examination Table System Recovery
Control panel function , 17-1
DELPHI , 1-7 System Recovery, See System Backup and
Indicators , 1-8 Recovery
QDR 4500 , 1-7 System Software Help , 3-14 to 3-15
Switches , 1-8 Online
Operator Console Ticker Tape , 3-15
Power Module , 1-4 Online Context Sensitive Help , 3-14
QDR 4500 Systems , 4-14 to 4-22 Tool Tips , 3-14
Control Panel , 4-15 to 4-18 System Software Reminders , 3-13
Laser , 4-16
Flashing Buttons , 3-13
QDR 4500A and SL , 4-17
Daily QC , 3-13
Arm, 4-17
System Backup , 3-13
Center Table, 4-18
Emergency Stop button, 4-18
Emergency Table, 4-18
Enable Lateral, 4-18 T
Laser, 4-17 Tabs , 3-10
Patient On/Off, 4-18 Tool Tips, System Software Help , 3-14
Table, 4-17
T-Score , 2-8
X-ray on Lamp, 4-18
QDR 4500C and W , 4-16
Center Table, 4-16
Emergency Stop button, 4-17 U
Patient On/Off, 4-16 Understanding
Table, 4-16 main window components , 3-7
X-ray on Lamp, 4-16
other window components , 3-9
Emergency Procedures , 4-20 to 4-22
Understanding Scan Results , 2-8
Equipment Fails While Equipment is
Operating , 4-21 Reference Database , 2-8
Emergency Stop Procedure, 4-21 T-Score , 2-8
Equipment has been Turned Off , 4-21 Z-Score , 2-8
Applying Power to the System, 4-22 Understanding the Mouse
Holding the Mouse , 3-2

I-16 Index
QDR Reference Manual

Mouse Tasks , 3-2 Height, 19-13


Clicking the Mouse , 3-3 Help button, 19-14
Double-Clicking the Mouse , 3-3 Identifier 2, 19-13
Dragging and Dropping , 3-3 Last Name, 19-13
Enter Text in a Field , 3-2 Menopause Age, 19-13
Move an Object on the Desktop , 3-2 Middle Initial, 19-13
Perform an Action , 3-2 OK button, 19-13
Pointing with the Mouse , 3-3 Patient Comments, 19-13
Using Patient ID, 19-13
menus , 3-8 Referring Physician, 19-13
sub-menus , 3-9 Weight, 19-13
Using a Study , 7-28 to 7-31 Erasing Scans, 19-23 to 19-25
Insurance, 19-14 to 19-15
Utilities , 19-2 to 19-57
Window, 19-14
AP Reposition , 19-40 to 19-41
Address1, 19-15
Database Tools , 19-9 to 19-32
Address2, 19-15
Auto Baseline Utility , 19-36
Cancel button, 19-15
Export , 19-27 to 19-29
City, 19-15
Export/Import , 19-25 to 19-26
Group, 19-15
Help button, 19-15
Types of Data
Insurance, 19-15
Patient Data, 19-26
Name, 19-14
QC Scan Data, 19-26
OK button, 19-15
Referring Data, 19-26
Patient ID, 19-14
Scan Data, 19-26
Plan, 19-15
Types of Transfers, 19-26
Postal Code, 19-15
Types of Data , 19-26
Managing Patient Baseline Scans, 19-17
Export/Import Function , 19-25 to 19-26
Managing Scans, 19-18 to 19-25
Import , 19-29 to 19-31
Window, 19-19
Patient Callback , 19-33 to 19-35
Analysis Date (button), 19-20
Patient Management , 19-9 to 19-25
Back button, 19-20
Managing Patients , 19-10 to 19-11
Cancel button, 19-21
Window, 19-10
Deselect All button, 19-20
Birth Date, 19-10
Erase Scan button, 19-21
Close button, 19-11
Move >> button, 19-20
Edit, 19-11
Patient ID, 19-20
Erase Patient button, 19-11
Patient Name, 19-19
Ethnicity, 19-10
Scan Date (button), 19-19
Patient ID, 19-10
Scan Details button, 19-20
Patient Name Text Box, 19-10
Scan ID (button), 19-20
Patient Selection, 19-11
Scan Location, 19-20
Referring Physician, 19-10
Scan Type (button), 19-19
Sex, 19-10
Select All button, 19-20
Managing Patients Information , 19-12 to Moving Scans, 19-21 to 19-23
19-15
QC Phantom Details, 19-15
Biography, 19-12 to 19-14
Window, 19-15
Window, 19-12
Cancel button, 19-16
Biography Tab, 19-12
OK button, 19-16
Cancel button, 19-14
Phantom Name, 19-15
DOB Day, 19-13
Phantom Number, 19-16
DOB Month, 19-13
Phantom Type, 19-16
DOB Year, 19-13
Removing a Patient from the System, 19-18
Ethnicity, 19-13
Starting , 19-9
First Name, 19-13
Reconcile , 19-31 to 19-32

Index I-17
QDR Reference Manual

Emergency Motion , 19-40 Displaying the Summary Report, 19-4 to


Import , 19-29 to 19-31 19-5
Install Options , 19-57 Technically Unacceptable Studies, 19-6
Rebuild Archive Indexr , 19-55 to 19-56 Usage Claims for Technically Unacceptable
Reference Curve , 19-42 to 19-54 Studies, 19-8
Adding New Ethic Groups , 19-53 to 19-54 Usage Detail Report, 19-6 to 19-8
Starting Reference Curve Editor , 19-43 Displaying the Detail Report, 19-6 to 19-
7
Reference Curve Editor
Copying a Reference Curve Record , 19-49 to Utilities not covered , 19-1
19-50
Creating New Reference Curve Records , 19-
44 to 19-49 V
Creating a New Reference Curve Record ,
Visualization of abdominal aortic calcification ,
19-44 to 19-47
13-1
Table
Reference Curve Description Fields, 19-47
to 19-49
Age Peak BMD, 19-49 W
Author of Curve, 19-48 Whole Body auto low density, Enable/Disable , 18-
Comment, 19-48 7
Date Last Modification, 19-48
Whole Body examination
Ethnicity, 19-47
Analysis
Is this curve current?, 19-48
window , 11-9, 11-16
Method, 19-48
Scan Site, 19-48 selecting scan type and mode , 11-3
Select X, Y Labels Button, 19-48 windows
Sex, 19-47 Analysis , 11-9, 11-16
Source, 19-48 Windows
X-Axis Display from - to, 19-48 Add Quality Control Phantom , 5-16
X-Axis Label, 19-48 Analysis
Y-Axis Display from - to, 19-48 AP scan , 12-22
Y-Axis Label, 19-48 Forearm examination , 10-14, 10-29
Deleting Reference Curve Records , 19-52 Hip examination , 9-28
Editing Reference Curve Records , 19-51 to Lateral scan , 12-31
19-52 Whole Body examination , 11-9, 11-16
Restoring Reference Curve Databases , 19-54 Analysis Setup
Viewing Reference Curve Data , 19-44 AP Lumbar Spine examination , 8-12
Scan File Look , 19-36 to 19-38 main
Scan File Plot , 19-39 components , 3-7
Starting a Utility , 19-2 QDR 4500
Accessing the Utilities Menu , 19-2 main , 6-22
System Configuration , 19-2 Quality Control Plot Parameters , 5-12
Usage , 19-3 to 19-8 Select a Patient for this Exam , 6-13
Starting Usage , 19-3 to 19-4
Usage Program Menu , 19-4
Show Detail, 19-4
Show Summary, 19-4
Z
Usage Summary Report , 19-4 to 19-8 Z-Score , 2-8

I-18 Index

You might also like