Inspection instructions:
1. If finishing one maintenance step successfully, check it on the option of FINISH OR NOT with√, otherwise
check it with×, and write down the reason as well as the solution.
2. Tool to clean components is dirt-free cotton, cotton stick or soft hair brush; tool to clean PCB boards is the
mini hair dryer; tool to clean the tube is syringe; tool to clean the clot is the special needle and syringe.
syringe used to clean the wash tube and waste tube cannot be mixed.
3. Using 75% of alcohol to clean all components and tube; using WD-40(Water Displacement, 40th formula)
to clean or remove rust of the axle and loader.
A. System Safety
1/8
Shenzhen New Industries Biomedical Engineering Co., Ltd.
2/8
Shenzhen New Industries Biomedical Engineering Co., Ltd.
and cuvette entrance detection sensor with swab and 75% ethanol.
4.8 Check the tubes, cables and fixing clip of washer. Confirm if
they are fixed well on the washer component. If they are loosen or
damaged, please adjust their position or replace them.
4.9 Check the peristaltic pump’s tube. Run pipeline test for 20 times.
The requirement of the result is CV<3% and the average result of
each needle is between 65-85. Adjust the tightness if out of range.
5. Chamber
5.1 Wipe the outer surface of waste liquid needle and signal
needle with dust-free cloth and 75% ethanol.
5.2 Wipe the nozzle with swab and 75% ethanol after opening the
cover of chamber.
5.3 Wipe the upper surface disk and working position with swab
and 75% ethanol.
5.4 Wipe the side and bottom of disk with dust-free cloth and 75%
ethanol.
5.5 Wipe the groove and cuvette detection sensor with dust-free
cloth, swab and 75% ethanol.
5.6 Confirm the torsion spring of light blocker. If the blocker can’t
be bounced, please try to solve it by rotating the needle. If it can’t
be solved, replace it.
5.7 Check the tubes of starters. Replace them if bend.
5.8 Check the cable of aspiration detection. Replace it if damaged.
5.9 Check the rotation of chamber disk. If it can’t rotate smoothly,
or its synchronize belt faces rapture or damaged by wearing,
please replace them.
6. Sample Disk
6.1 Wipe the sampling disk with dust-free cloth and 75% ethanol.
6.2 Wipe the outer surface and inner surface of vortexer with swab
and 75% ethanol by passing the holes on disk.
6.3 Disassembly the sampling disk, check if the bearing has some
fragment on it. Clean if necessary with dust-free cloth.
6.4 Disassembly the sampling disk and clean the initialization
sensor with swab.
6.5 Clean the cuvette detection sensor with swab.
7. Claw
7.1 Wipe the gripper with dust-free cloth and 75% ethanol.
Change the gripper if the shape is changed seriously. Adjust the
position in service software if the distance between the two clips is
too far. If adjustment can’t let the claw hold the cuvette, please
change the gripper.
7.2 Check the synchronous belt of claw. If the synchronize belt
faces rapture or damaged by wearing, replace them if damaged.
8. Cuvette Storage
8.1 Clean the dust on lifter, chip on the top and push rod with dust-
3/8
Shenzhen New Industries Biomedical Engineering Co., Ltd.
free cloth and 75% ethanol. Adjust or replace the chip if its shape
is changed.
8.2 Clean the six sensors with swab and 75% ethanol as shown
below:
a) Initializing sensor for push rod (left side baffle is required to be
disassembled).
b) Cuvette box detection sensor at the entrance and exit.
c) Initialization sensor, bottom sensor and top sensor of lifter.
8.3 Clean the synchronous belt of new box transfer belt. Check
the condition of belt and replace or adjust the tightness if
necessary.
9. Cabinet
9.1 Wipe the baffle of analyzer, the door of sample reagent area,
the dust bin, the starter box holder and the inside of consumable
door with dust-free cloth and 75% ethanol.
9.2 Check the hinge of consumable door to see if its tightness is
suitable.
9.3 Check if the movement of dustbin is smoothly.
9.4 Check if the force required to open and close the cover of
instrument is suitable. Adjust the support force of the cover.
9.5 Take out the dust proof net and clean the dust with hair dryer.
Clean the dust on heatsink with brush.
9.6 Put the dustbin inside the instrument and wipe the channel for
cuvette discarding with dust - free cloth and 75% ethanol. Let the
dustbin collect the waste generated in this process.
10. Washing System Liquid Tank
10.1 Inject 80ml sodium hypochlorite (NaClO) solution with effective
chlorine content 5.5-6.5% and add pure water to 2L. Let the solution
soak all the inner walls of system liquid’s tank for 2-3min.
10.2 Wash the tank with pure water for at least 3-5 times after
soaking. Add system liquid back to the tank.
10.3 Wash the joints of system liquid tank.
D. Component Lubrication
4/8
Shenzhen New Industries Biomedical Engineering Co., Ltd.
3. Washer
3.1 Clean the dust on the guide rail and slider of washer lifter with
dust-free cloth. Provide lubrication grease Multemp PS NO.2 or
AS2.
3.2 Clean the dust on washer mixers‘Z-axis with dust-free cloth.
Provide lubrication grease Multemp PS NO.2 or AS2.
3.3 Clean the dust on the guide rail and slider of washer mixers with
dust-free cloth. Provide lubrication grease Multemp PS NO.2 or
AS2.
3.4 Rotate the peristaltic pump manually and check each roller. If
resistance is vast, please clean the rollers with WD40 and add
grease for lubrication.
4. Chamber
Clean the dust on the guide rail and rack of chamber lifter with dust-
free cloth. Check if the rack can be driven smoothly. Provide
lubrication grease Multemp PS NO.2 or AS2.
5. Claw
5.1 Clean the dust on the screw of claw’s Z-axis with dust-free
cloth. Provide lubrication grease HSS17 and drive the claw
manually to smear the grease. Please wipe the grease
accumulate at the two ends after smearing.
5.2 Clean the dust on the guiderails of claw’s X and Y-axis with dust-
free cloth. Provide lubrication grease Multemp PS NO.2 or AS2 and
drive the claw manually to smear the grease. Please wipe the
grease accumulate at the two ends after smearing.
6. Cuvette Storage
6.1 Clean the dust on the guide rail of cuvette storage’s lifter with
dust-free cloth. Provide lubrication grease Multemp PS NO.2 or
AS2 and drive the lifter manually to smear the grease.
6.2 Clean the dust on the guide rail of push rod with dust-free cloth.
Provide lubrication grease Multemp PS NO.2 or AS2 and drive the
push rod manually to smear the grease.
5/8
Shenzhen New Industries Biomedical Engineering Co., Ltd.
F. Clearance of tubing
G. Alignment
6/8
Shenzhen New Industries Biomedical Engineering Co., Ltd.
7/8
Shenzhen New Industries Biomedical Engineering Co., Ltd.
K. Performance Evaluation
Engineer User
Date Date
8/8