Discipline Instrument
Unit: Analyser
Element Gas and water analyzer

1. IMPLEMENTATION OF MAINTENANCE PROCEDURES AND ADJUSTMENT


Câu hỏi What kind of online analyzer are used in NCSP?

Lựa chọn

1 Hydrocarbon dewpoint, Water dewpoint, Oil in water, Oxygen, Instrument air dewpoint, Chlorine
(Câu đúng) and Gas chromatograph analyzer.

2 Hydrocarbon dewpoint, Water dewpoint, Oil in water, Oxygen, Instrument air dewpoint, Chlorine.

3 Hydrocarbon dewpoint, Water dewpoint, Oil in water, Oxygen, Instrument air dewpoint, Chlorine,
H2S and Gas chromatograph analyzer.
4 Water dewpoint, Oil in water, Oxygen, Instrument air dewpoint, Chlorine and Gas chromatograph
analyzer.
Câu hỏi What is the PMR are applied to online Hydrocarbon dewpoint and Water dewpoint analyzer in NCSP
gas process?
Lựa chọn

1 I 085 - I 086 - I100 – I110 and I111


(Câu đúng)

2 I 085 - I 086 - I096S

3 I 085 - I 086 - I100 – I097S

4 I 085 - I 086 – I096S – I097S

Câu hỏi What is the method to measure hydrocarbon dewpoint by Ametek analyzer ?

Lựa chọn

1 Chilled mirror technology


(Câu đúng)

2 Laser technology

3 Infrared technology

4 None Above

2. INTERROGATION AND FAULT-FINDING SKILLS


Câu hỏi How do we check the zero point of online Oxygen analyzer in NCSP?

Lựa chọn

1 Using Nitrogen 99,99%


(Câu đúng)

2 Using instrument air

3 Using nature gas

4 All are not correct

Câu hỏi What we should to do when the online water dewpoint analyzer fails to verification?

Lựa chọn

1 All above
(Câu đúng)

2 Check enclosure temperature

3 Check flow rate

4 Adjust moisture span

Câu hỏi What we should to do when the online hydrocarbon dewpoint analyzer fails to verification?

Lựa chọn

1 All Above
(Câu đúng)

2 Check RTD

3 Check optical cell and clean the mirror

4 Check LED performance

3. INSPECT, REPAIR SYSTEMS FOR RESTORATION TO REQUIRED PERFORMANCE


Câu hỏi How often does the oxygen sensor replaced?

Lựa chọn

1 Yearly
(Câu đúng)

2 Six monthly
3 Three monthly

4 Monthly

Câu hỏi How long does filter block of online hydrocarbon dewpoint analyser required to replace repair kits?

Lựa chọn

1 Every six months


(Câu đúng)

2 Yearly

3 Every Two years

4 Every Three years

Câu hỏi Generally, the preventive and general maintenance for Ametek hydrocarbon dewpoint consists of:

Lựa chọn

1 All above
(Câu đúng)

2 Cleaning and changing out the Filter block Assembly components

3 Cleaning the Electronics Enclosure’s flame path and checking it for


damage

4 Determining if the Sample and Vent Lines are plugged, damaged, or


sagging
4. RETURN EQUIPMENT TO SERVICE BY COMPONENT REMOVAL AND REPLACEMENT
Câu hỏi Can we return Oxygen analyzer when it is reading 16%?

Lựa chọn

1 NO
(Câu đúng)

2 Yes

3 No required

4 Sometime

Câu hỏi What should we do when replace moisture generator for water dewpoint analyzer (Ametek type)?

Lựa chọn
1 Key in moisture generator code
(Câu đúng)

2 Key in Dryer code

3 Adjust moisture span

4 None above

Câu hỏi What is the maximum error for hydrocarbon dewpoint analyzer verification?

Lựa chọn

1 ± 1 degree C
(Câu đúng)

2 ± 1,5 degree C

3 ± 0.5 degree C

4 ± 1.2 degree C

5. MONITOR AND ASSESS THE PERFORMANCE AND CONDITION OF COMPONENT


Câu hỏi Where can you find the performance data of analyzers in NCSP?

Lựa chọn

1 Maximo and Metering record


(Câu đúng)

2 Logbook

3 Handbook

4 Prolog

Câu hỏi Who can be perform adjustment or calibration for analyzers in NCSP?

Lựa chọn

1 Instrument technician
(Câu đúng)

2 Mechanical tech

3 Electrical tech

4 Production tech
Câu hỏi What is the standard applied to hydrocarbon dew point ?

Lựa chọn

1 ISO/TR 11150 -2007


(Câu đúng)

2 ISO 11150 -2007

3 ISO/TR 11150

4 ISO/TR 11150 -2015