Professional Documents
Culture Documents
W2-W3 X-ray Radiography
W2-W3 X-ray Radiography
W2-W3 X-ray Radiography
Radiography
X-rays
• X-rays is electromagnetic radiation.
2
Reducing patient dose is
important, but the image quality
also is essential.
3
Production of X-ray
• Involves the bombardment of a target with energetic electrons.
• Electrons undergo collisions and scattering processes during the
slowing down process which results in the production of
– Bremsstrahlung and
– Characteristic Radiation
4
Production of X-ray
• Energetic electrons are mostly slowed down in matter
by:
▪ Collisions and
▪ Excitation interactions
5
Production of X-ray
❑ The energy of the emitted photon is
subtracted from the kinetic energy of the
electron.
❑ Bremsstrahlung depend on
• Attractive Coulomb forces
• Distance of the electron from
the nucleus
6
Production of X-ray
• Electron bombardment on target yields a constant energy fluence (Ψ)
from zero up to the initial electron kinetic energy (T).
• The superposition from rectangular distributions forms a triangular
energy fluence distribution for a thick target, known as Ideal Spectrum.
7
Production of X-ray
According to this model
• An increase in electron energy increases the number
of thin layers each radiating X rays
• The triangular area grows proportional to the square
of the electron energy
8
Production of X-ray
• Characteristic radiation
• Fast electron knock-out electron
(atomic shell)
• The KE > BE of atomic electron
• Vacancy shell filled out higher
energy state.
• During the transition, an X-ray
photon is emitted.
The binding energy is Highest in the most
inner K-shell and decreases for the outer
shells (L, M, ..)
9
Production of X-ray
• For each element binding energies and the Monoenergetic radiation
resulting from such interactions, are unique and Characteristic for that
element
10
Production of X-ray
a) Ideal Bremsstrahlung
spectrum for a tungsten
anode
b) An actual spectrum at the
beam exit port with
characteristic X rays
(anode angle: 20°, inherent
filtration: 1 mm Be)
c) The spectrum filtered with
an equivalent of 2.5 mm Al
11
Production of X-ray
❑ The Mass Stopping Power of electrons is proportional to
Z².
❑ Integration along the electron path gives the total X ray
energy fluence as
I .Z .U 2
12
X-ray Tube
13
X-ray Tube
❑ There are two main circuits operate within the XRT:
• Filament circuit
• Tube voltage circuit
❑ The Filament is heated by a current that controls the
thermionic emission of electrons (Tube or Anode
Current)
e.g. <10 mA in fluoroscopy and 100 to >1000 mA in single exposures
14
X-ray Tube (Cathode)
• The arrangement of tube (filament, focusing cup, anode) generates an
electric field to accelerate the electron.
• Biasing and unbiased focusing cup leads to a compression of the
trajectories giving a smaller focus and broad focus.
15
X-ray Tube (Cathode)
• The spiral-Wound filament is
typically made from tungsten wire
of 0.2 to 0.3 mm diameter and
operates at around 2700o K.
• Thermionic emission of electrons
increases with temperature
(Richardson’s law) and produces a
cloud of electrons (Space Charge)
enclosing the filament.
−W
J = AGT e 2 kT
16
X-ray Tube
Space Charge shields the filament from the anode voltage.
17
X-ray Tube
18
X-ray Tube (Anode)
• Stationary anodes
– Usually for low anode current or infrequent low power exposures (e.g
dental units, portable X-ray)
– Here small tungsten is brazed to copper block to dissipate the heat
efficiently
– The maximum loading for stationary focal spot is determined by anode
temperature and temp gradients.
19
X-ray Tube
• The continuous bombardment to the anode will led to melting and
destruction of anode target (which reduce the fluence).
• Require dissipation of the energy into much larger volume.
• The rotating anode is introduces to increase the target area (the
circumference of focal tract)
20
X-ray Tube
❑ The Limiting Factor in the use of X ray tubes is given mainly by the
thermal loading capacity of the anode
Maximum Permissible
Tube Load vs. time for
a single exposure,
constant current.
(100 kV tube voltage
and a 50-kW tube)
Q = m C p T
21
X-ray Tube (Housing)
❑ The XRT (often referred to as the Insert) is installed in a
Housing providing the structural support required
• Typical housing assembly for
a general purpose XRT.
• The housing also provides
i. Radiation Shielding -
prevent any radiation
except the primary
beam from leaving the
housing
ii. Lead Sheets - to
minimize leakage
radiation
22
X-ray Tube (Filtration)
• Filtration, is the removal of x-rays as the beam passes
through a layer of material.
• Basically
– Inherent filtration (SiO2, Z=14 which attenuate 15 keV X-rays)
– Added filtration (Al, filtered the low-energy X-rays)
• At a given kVp, added filtration increases the HVL.
= Wi ( )i
0.693 ⎯⎯ →
HVL = ⎯⎯
23
X-ray (Generator)
Generally, the essential components are:
▪ a Filament Heating circuit to determine anode current
▪ a High Voltage supply
▪ a Motor Drive circuit for the stator windings required for a rotating
anode tube
▪ an Exposure Control providing the image receptor dose required
▪ an Operational Control
25
X-ray (Generator)
26
X-ray (Generator)
27
X-ray (Generator)
• The Operational Control is often accomplished by a
microprocessor system, but electromechanical devices
are still in use.
• Modern generators provide control of the Anode
Temperature by
▪ Monitoring the power applied to the tube and
▪ Calculating the cooling times required according to
the tube rating charts
28
X-ray (generator)
❑ Irrespective of the waveform the tube voltage is defined
as the Peak Voltage, kVp, of the voltage train
– The Voltage Ripple, R, is given as the relative difference of the
minimum voltage, kVmin, from the peak voltage:
29
Tube Voltage
• Single-phase generators • Three phase
• Step-Up Transformer with a fixed • mains supply three AC-voltages
winding ratio. each with a Phase-Shift
• Half-Wave Rectification gives a 1- • Full-Wave Rectification
Pulse waveform • 6 half-waves per period (6-Pulse
• pulse is a half-wave per period of waveform) with a nominal ripple of
frequency (50 or 60 Hz) 13.4%
Tube Voltage
High-frequency generators, stabilized the power supply in the front end
of the device
i. rectified and filtered AC to produce a DC-supply voltage for Inverter
Circuit.
ii. Inverter transformed, rectified the pulse
iii. Capacitor collect the pulse to give higher voltage.
iv. Inverter Pulse Rate is used to control the tube voltage.
31
Tube Voltage
32
Exposure Timing (AEC)
• Most of the system employed digital timers
– High reproducibility
– Accuracy of microsecond
– The timer activates and terminates a switch
• The accuracy of X-ray exposure time depend on the
switches.
• The countdown timer is used to terminate exposure
when the exposure switch failure.
33
Exposure Timing (AEC)
• Switches for three phase and constant potential
generators, known as triode, tetrode and pentode.
– To initiate and terminate the exposure
– Control by the electronic timer or photo timer
• Most of high-frequency generator utilize electronic
switching.
• Alternatively, a grid can be used to any generator to
switch the exposure.
34
Exposure Timing (AEC)
• Phototimer
• Use to measure the actual amount of radiation incident on the
image receptor (e.g., screen-film detector)
• terminate x-ray production when the proper amount is obtained.
• Phototimer consist of
• Radiation detectors
• Amplifier
• Density selector
• Comparator circuit
• Termination switch
• Backup timer
35
Exposure Timing (AEC)
36
Exposure Timing (AEC)
• Falling load generator works
simultaneously with AEC
system.
• The falling-load allows
maximum possible mA for the
selected kVp, by considering
the heat load characteristics
of X-ray tube.
• 50 mAs of 80 kVp (400 mA
with 125 msec) ~initially 600
mA, with 50 mAs in 100
msec)
37
Factors Affecting X-ray
• The output of X-ray is based on terms quality,
quantity and exposure.
– Quality describes the penetrability of an x-ray beam,
with higher energy x-ray photons having a larger HVL
and higher "quality.“
– Quantity refers to the number of photons comprising
the beam.
– Exposure, is nearly proportional to the energy fluence
of the x-ray beam and therefore has quality and
quantity associated characteristics.
38
Factors Affecting X-ray
• The quality, quantity and exposure can be
determined by six major factors
i. Anode target “affects the quantity of
bremsstrahlung photons and the quality of the
characteristic radiation”.
ii. Tube voltage (kVp) “determines the quality of
spectrum, the efficiency and exposure of X-ray
Exposure kVp 2
39
• The quality, quantity and exposure can be
determined by six major factors
iii. Tube current (mA) “the exposure and filtration is
proportional to the tube current”.
iv. Exposure time “duration of X-ray production and
proportional to the product of tube current and
exposure time (mAs)
v. Beam filtration “modifies the quantity and quality of
X-ray beam, by removing the low energy photons”
vi. Generator waveform “ a single phase has lower
potential difference compared to three phase or high
generator frequency. Both quality and quantity are
affected.
40
Detector Image System
(Image Formation)
• X ray imaging is the collection of attenuation shadows that are
projected from an ideal X ray point source onto an image receptor
• This simple form is true for all X ray imaging modalities, including
complex ones that involve source and receptor movement such as
CT.
• This simplified view however is made vastly more complex by:
– the Non-Ideal point source
– the consequences of Projecting a 3D object onto a 2D detector
– the presence of Scattered Radiation, generated within the patient, which will
degrade any image that is captured
41
Detector Image System
(Image Formation)
• The principal components of a system for X ray
projection radiography:
42
Detector Image System
(Image Formation)
• The X ray intensity is related to thickness by the
attenuation law, therefore the Primary Contrast:
where:
xd: thickness of the detail
μd: linear attenuation coefficient of the
detail
μb: linear attenuation coefficient of the
background material
43
Detector Image System
(Image Formation)
Energy absorbed in a small region of the image receptor:
Ep: energy absorbed due to primary rays
Es: energy absorbed due to secondary rays
Es
Scatter to Primary Ratio Sp =
Ep
44
Components of an Imaging System
• In the presence of scattered radiation, the Primary
Contrast equation becomes
46
Effect of Tube Voltage on Contrast,
Noise & Dose
• In digital systems the choice of suitable kV and mAs
combinations requires that for each projection:
• the kV and mAs produce the correct value of the EI and that
• the maximum value of the kV is chosen that will allow
diagnostically acceptable CNR
• If a suitable phantom is radiographed at a low kV with
suitable mAs, and a range of further images is obtained
at increased kV settings at the same mAs, the images
will appear very similar.
47
Scattering In Projection Radiography
• It is often stated that photon Scattering is of no benefit
for projection radiography, leading only to fogging of the
image.
• However, this is incorrect, as the appropriate contrast for
each projection is chosen by setting a suitable kV to
provide the Correct Proportion of photoelectric and
scatter interactions.
• At low kV settings the contrast is high, due to the
predominance of the Photoelectric Effect, while at high
kV it is low, due to the predominance of Scattering
interactions.
48
Scattering In Projection Radiography
These effects are illustrated below - showing the
contribution of photoelectric and scattering interactions for
the primary contrast of a 1 mm sphere of calcium
embedded in water, calculated over a range of energies
from 20-100 keV
Calculated
49
Origins of Scattered Radiation
50
Magnitude of the scatter
51
Magnitude of the scatter
Dependence upon Radiographic Procedure
• The radiographic procedure itself has a strong influence
on the proportion of scatter, depending on whether the
subject is a region consisting largely of bone, or soft
tissue, or some intermediate combination.
• This is because the scatter interactions are predominant
for soft tissue, but the photoelectric interaction is
predominant for bone over much of the diagnostic
energy range (except mammography).
52
Magnitude of the scatter
Dependence on Field Size
As the field size increases from a narrow beam with almost no
scatter to very large, the Scatter Fraction increases until a
saturation level is reached, beyond which little or no further
increase in scatter fraction occurs
Dependence of scatter
fraction on beam area for four
phantom thicknesses at 80 kV
53
Magnitude of the scatter
Dependence on Thickness
The Scatter Fraction increases rapidly with patient
thickness, but tends to saturate for very large patients
Dependence of scatter
fraction on phantom thickness
for four X ray field sizes at 80
kV
54
Magnitude of the scatter
Dependence on Position (in 3D)
• the Scatter Fraction (at the image receptor) has been
considered so far, it is also important to quantify the
scattered radiation in all directions from the patient, as
this affects the dose to Personnel
• Knowledge of scatter levels is required in order to
determine appropriate Radiation Shielding levels
• Scatter may be categorized as forward scatter and side
and back scatter
55
Magnitude of the scatter
Dependence on Energy
• Conversely, the energy of the scattered photons
increases as the energy increases
• So that they are less likely to be attenuated in the body
and are more likely to escape
• The overall effect is that side-scatter, back-scatter and
forward- scatter all increase as energy is increased
• However, the primary beam attenuation also decreases
as energy is increased
56
Effect of Scatter
The Effect of Scatter: Noise
• The Quantum Noise in the image arises from both
primary and scattered photons
• Both of these image contributions therefore affect the
SNR
• Digital chest imaging with a grid shows that SP varies
between
• ~2 in the region of the mediastinum to ~0.6 in the lung fields
• leading to a value of SNR per pixel behind the mediastinum
of ~20, and in the lung fields ~60
57
Methods for Scatter Reduction
Anti Scatter Grids
• The scatter component of the image may be considered
to consist of the primary image convolved with a Scatter
Spread Function which gives a highly blurred version of
the image
• The resulting image may be considered to be the Sum of
these two images
• Efforts are being made to employ this idea for
Computerized Scatter Removal, rather than using grids
or other methods
58
Methods for Scatter Reduction
Grid Construction
59
Methods for Scatter Reduction
Grid Construction
Construction and
principle of operation
of a focused anti-
scatter grid (not to
scale)
60
WHY QUALITY CONTROL (QC)?
2. System-level
assessment
Including components of the system
that affect the eventual quality:
• Grid
• AEC
• Post-processing
ANTI-SCATTER GRID
01 Acceptance testing
• Grid ratio verification
• Grid focus verification
• Primary and scatter transmission verification
76
Methods for Scatter Reduction
Grid Construction
• Scattered photons whose angle of incidence is less than
tan-1 r will hit a lead strip, so that scatter rejection will
increase with increasing grid ratio
78
Methods for Scatter Reduction
Grid Construction
TYPICAL CONSTRUCTION DATA FOR THREE ANTI-SCATTER GRIDS
79
Methods for Scatter Reduction
Measures of Grid Performance
80
Methods for Scatter Reduction
Measures of Grid Performance
81
Methods for Scatter Reduction
Measures of Grid Performance
The Contrast Improvement Factor is given by:
83
Methods for Scatter Reduction
Grid Artefacts & Alignment
• There are several possible Misalignments that will lead
to Artefacts in projection images
• Additionally, a Damaged grid will generate artefacts and
must be replaced
84
Methods for Scatter Reduction
Grid Artefacts & Alignment
Possible misalignments of
the grid:
85
Other Methods for Scatter Reduction
86
Other Methods for Scatter Reduction
Collimation
Since smaller Field Sizes reduce the scatter fraction, it is
good practice to collimate the X-Ray Beam to as Small an
Area as Possible for Each Projection. It also helps to:
• improving image quality
• reducing patient dose
87
Other Methods for Scatter Reduction
Compression
• Scatter Fraction increases with Patient Thickness
• Therefore, if the patient thickness can be Reduced
during exposure by applying a Compression Band for
example, then the amount of scatter will be reduced
• This has further benefits, because a Shorter exposure
time can be used, reducing movement blur and patient
dose
• Alternatively, the kV setting may be reduced, improving
the contrast, or SNR in the image
88
Other Methods for Scatter Reduction
Air Gap
89
Other Methods for Scatter Reduction
Air Gap
92
Image Quality Assurance
• The preemptive measures taken to improve the quality of
treatment and services while being cost-effective are
characterized as quality assurance or quality
improvement.
93
Image Quality Assurance
Contrast:
• A radiographic contrast is a subtle difference in signal strength or
brightness between an object of interest and its environment.
Dynamic range:
• The range of different X-ray intensities that the detector can capture
is known as the dynamic range.
Spatial resolution
• The scanning system's spatial resolution is its capacity to tell apart
similar structures from one another.
94
Image Quality Assurance
Noise
• The variations in an image that are random or structured that do not
resemble the variations in an object's X-ray attenuation are known
as radiographic noise.
Artifacts:
• Poor image quality is a result of artifacts as well as low resolution,
noise, and SNR.
95
Image Quality Assurance
Noise
• The variations in an image that are random or structured that do not
resemble the variations in an object's X-ray attenuation are known
as radiographic noise.
Artifacts:
• Poor image quality is a result of artifacts as well as low resolution,
noise, and SNR.
96
Image Quality Assurance
Factors affecting image quality
97
Image Quality Assurance
Factors affecting image quality
98
Image Quality Assurance
Factors affecting image quality
99
Image Quality Assurance
Clinical Significance
• The radiation dose required to produce high-quality
digital radiographs can be reduced by changing kVp,
mAs, and focal spot size. Even though a larger radiation
dose results in less noise and improved image quality,
the patient should still be exposed to minimal radiation.
100
Image Quality Assurance
Clinical Significance
101
TERIMA KASIH / THANK YOU
www.upm.edu.my