Professional Documents
Culture Documents
ARIETTA-70 ServiceManual Vol.1 NoRestriction
ARIETTA-70 ServiceManual Vol.1 NoRestriction
Service Manual
Volume 1
English Version
MN2-2075 Rev.1
MN2-2075 Rev.1
ARIETTA 70 Service Manual VOLUME 1
1/6
MN2-2075 Rev.1
ARIETTA 70 Service Manual VOLUME 1
2/6
MN2-2075 Rev.1
ARIETTA 70 Service Manual VOLUME 1
3/6
MN2-2075 Rev.1
ARIETTA 70 Service Manual VOLUME 1
4/6
MN2-2075 Rev.1
ARIETTA 70 Service Manual VOLUME 1
Chapter 2 SCHEMATICS
Chapter 4 ADJUSTMENT
5/6
MN2-2075 Rev.1
ARIETTA 70 Service Manual VOLUME 1
(Blank page)
6/6
Chapter 1
INTRODUCTION
APPENDIX
MN2-2075 Rev.1
Chapter 1 Introduction
This service manual is applicable to ARIETTA 70, ARIETTA 70a, ARIETTA S70, ARIETTA S70a,
ARIETTA V70 and ARIETTA V70a.
The word, ARIETTA70 in the manual is deemed to be ARIETTA 70, ARIETTA 70a, ARIETTA S70,
ARIETTA S70a, ARIETTA V70 and ARIETTA V70a, if the case may be.
This service manual has been prepared for persons in charge of repair at the field.
This service manual is compiled according to the following basic principle. ”For service, pick out a
faulty PCB and replace it with a new PCB.”
Make the best use of this service manual, making also reference to available technical support
information such as “Technical Bulletin”.
* Technical Bulletin, the Technical Notes, and HISTORY of the equipment which are described on this
manual are released for Service Engineer who has taken appropriate training.
The equipment is repaired by PCB replacement. Therefore this service manual does not include the
circuit diagrams of the PCB unit. For the explanation of functions, Block Diagrams and signal list of
each PCB, whose circuit diagram is not included, refer to ”VOLUME 2: Chapter 1 SYSTEM
OPERATION”. The Specification of System and System Block Diagrams are described in “VOLUME
1: Chapter 4 SYSTEM OVERVIEW”.
However, “Cable Connection Diagram”, “Circuit Diagram of PCB equipped with the panel switches
which are easily exchangeable at the field” and ”Circuit Diagrams composed of general circuit such as
Power Supply unit” are described in ”VOLUME 2: Chapter 2 SCHEMATICS”.
For changes and modifications of as well as additions to specifications, if any, prompt information will
be given to you by means of “APPENDIX Manual Change Information”.
IMPORTANT Always observe the manner specified for replacement, addition, or deletion of
“Manual Change” to prevent missing of necessary information and keeping of
erroneous information.
VOLUME 1
General instructions on carrying out maintenance service, correspondence at the time of trouble,
procedure that is necessary for operation check are described.
VOLUME 2
Detailed information for actual repair work, available service information are described.
1-1
MN2-2075 Rev.0
Chapter 1 Introduction
VOLUME 1
Chapter 1 INTRODUCTION
Describing the purpose of the Service Manual
Chapter 3 INSTALL/DISASSEMBLE
Disassembling Procedure illustrates the disassembly and assembly of main components. Be sure to
follow working procedures if specified
Chapter 5 TROUBLESHOOTING
Describing precautions on actual repair work and shows the necessary tools and measuring
instruments. Also, it includes many hints on primary diagnosis and measures to be taken in the
field.
VOLUME 2
Chapter 1 SYSTEM OPERATION
Describing PCB Block diagram and the Signal List, additional detailed explanation to the
“VOLUME 1: Chapter 4 System Overview”.
Chapter 2 SCHEMATICS
Giving the cable connection diagram including all cables used, the circuit diagram of PCB
equipped with switches and the circuit diagram of and Power Supply unit.
Chapter 4 ADJUSTMENT
Giving guides for adjustment of PCB and units that are required when they are replaced.
1-2
MN2-2075 Rev.0
Chapter 1 Introduction
When disassembling the equipment after checking it for a trouble symptom, give care to the following:
1) Be sure to unplug the equipment before disassembly.
2) Be sure to turn off the main switch on the equipment when removing electrical parts such as PCBs,
probe, and cable.
3) Safety alert symbols
4) The indication used on this equipment and in this service manual has the following meaning
“ Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate
injury. ”
“ A caution message is inserted here. ”
5) Perfectness in grounding, screw tightening, and cover installation is essential. Negligence of it could
cause a possibility of leakage current from outer fitting which may lead to serious damage to a
patient being diagnosed.
When disassembling the equipment, give care to the following to protect Service Engineer or User from
hazards:
1) Keep the working environment neat.
2) Wear working gloves to protect your hands from getting injured by burrs on the unit and casing.
3) Use only proper tools suited to work being made.
4) Be sure to observe the disassembly procedure shown in VOLUME 2: Chapter 3.
5) Take sufficient care not to damage component with undue load.
6) Be sure to observe equipment is re-assembled properly after disassembly.
7) Use only the specified screws and nuts. Using any other screws and/or nuts would affect not only
mechanical performance, but also electrical performance of the equipment.
8) In case of the equipment has movable unit internally, take sufficient care not to pinch your hands,
ties, wristbands in movable unit. Be sure to zip-up/fasten your fastener and button. Do not put tools
and screws around movable unit.
9) Take care not to touch Fan when covers of equipment are off.
10) Fix the moving part appropriately when you transport or move the machine. If the machine has
transportation position, set this position appropriately.
1-3
MN2-2075 Rev.0
Chapter 1 Introduction
1) When it is necessary to touch the equipment, options and/or other peripheral devices at a customer
who uses intracorporeal (transesophageal, transurethral, transvaginal, transrectal) probes that need
sterilization, take special care to protect your hands against germs, irrespective of the usage of the
equipment: whether it is used in the operation room or not.
2) Service tools are subject to germ pollution in hospitals and, therefore, need periodical sterilization.
3) Be careful not to directly touch anything assumable to have germ pollution. If necessary, ask the
customer for effective protection against germs.
4) Be sure to confirm the equipment, options and peripheral devices are washed, disinfected or
sterilized appropriately when you take them back from customer site.
5) In case the equipment radiates X-ray, pay attention to the circumference and take care not exposed
X-ray indiscreetly when the equipment is radiating. You must put a film badge for monitoring the
personal exposure at proper position when you do the repair work.
Whenever grease, oil or other chemicals is used for maintenance service, options and/or peripheral
devices, be sure to clean the equipment and/or devices after service work.
OS (Operating System, such as Windows) operation is allowed Service Engineer who has taken
appropriate training. Illegal change on OS or our program files, Illegal copy of file/folder/partition
which NOT instructed, are prohibited.
Computer controlled medical equipment that involves starting up an operating system from an internal
storage drive could become infected with computer viruses.
Such equipment is usually infected via peripheral storage, media or connections to a network.
Examples detailing route of infection
A) Infections caused during upgrades or maintenance by service engineer.
An infected USB memory device was connected.
An infected floppy disk was inserted.
A virus-infected USB hard disk drive was connected to remove data.
B) Infections caused by copying/backing up the preset data or stored images
An infected USB memory device was connected.
An infected floppy disk was inserted.
A virus-infected USB hard disk drive was connected to remove data.
1) Scan all media for viruses before connecting them to or inserting them in the equipment.
2) When an infection is detected, investigate the route of infection and its scope before removing the
virus.
3) Any connections to a network should as far as possible be routed via a firewall.
4) Software, files or services other than those designated by us must not be installed on or uninstalled
from the equipment. Nor must files other than those specified by us be modified or edited.
5) DO NOT connect the internal storage device (hard disk, CF card and so on) to your PC.
It may cause the computer virus infection to the machine.
1-4
MN2-2075 Rev.0
Chapter 1 Introduction
These maintenance tools refer to the laptop computers and auxiliary storage devices that service
engineers carry around. (Any computer used for creating CD-Rs for installing upgrades is also regarded
as a maintenance tool.) Thus maintenance tools will include laptop computers, floppy diskettes, USB
memory devices, external hard disks, CD-Rs, etc.
1) Perform a virus scan of any tool that will be used to ensure that they are not infected by computer
viruses.
Regularly perform virus scans of maintenance tools.
2) In unavoidable cases when for some reason safety cannot be confirmed, or when an unknown
computer or memory device is connected, be sure to perform a virus check after use.
3) Update with the most recent virus pattern data prior to performing a virus scan.
4) If a virus scan does not remove the cause for anxiety, use another virus scanner to check.
Extreme care must be taken to ensure that data (image data, patient database, DICOM communication
log files) that has been saved to another media during equipment repair and may include patient data is
guarded against leakage, loss and theft. Delete any data that is no longer needed.
1) Work involving use of patient data should be performed in a room that can be locked to prevent
leakage, loss or theft of such data.
2) Such data must be stored in a locked shelf or similar container.
3) If no lockable shelves are available, encrypt the files to reduce possible harm in the event of theft,
leakage or loss.
2) Go over the “Technical Bulletin” and “Technical Notes” to see whether the complained trouble can
be mended by means of regular repairing method.
1-5
MN2-2075 Rev.0
Chapter 1 Introduction
It is our policy that neither repair nor modification of PCBs used for S.M.D. is made in the field as a
rule because of the following reasons:
[REMARKS]
PCB does not need repairing or modifying in the field as a rule.
Reuse of chip devices (including resistors, capacitors, diodes, etc.) is strictly inhibited.
CAUTION When handling a PCB, avoid touching the IC and connector pins on the devices
to prevent ESD (Electro Static Discharge) damage.
A service person should preferably wear an ESD wrist strap correctly grounded
when handling a PCB.
Applied to WEEE Directive. Equipment which has this label must be rejected or
recycled by manufacturer.
WEEE(Waste Electrical and Electronic Equipment) was adopted Feb 2003 by
European Union. The purpose of this directive is to prevent incidence of waste of
electrical and electronic devices and in order to reduce rejection, “re-cycle”,
“re-use” and/or “re-cycle in other way” are demanded.
1-6
MN2-2075 Rev.0
Chapter 1 Introduction
The PCBs (printed circuit boards) inside and other parts use lead-free (Pb free) solder and lead-free
RoHS compliant components.
Use the label shown above for products Use the label shown above for products that
containing only RoHS compliant mounted contain some RoHS compliant mounted
components that are soldered using lead-free components that are soldered using lead-free
solder solder.
EMI (Electro Magnetic interferences): An interferences generated from electronic and electrical
equipment.
Immunity: Tolerance against external Electromagnetic interferences.
1-7
MN2-2075 Rev.0
Chapter 1 Introduction
Symbols used by Aloka are described below, together with reference to IEC publication(s).
6
Be careful of acoustic power
IEC60417-5007
IEC60417-5008
IEC60417-5009
11 Alternating current
IEC60417-5032
1-8
MN2-2075 Rev.0
Chapter 1 Introduction
IEC60417-5021
13 Protected against the effects of continuance immersion in water
Labeled on Foot Switch MP-2345B, MP-2614B
IEC60529
14 Electrostatic discharge (ESD) symbol:
Follow the ESD guide line
1-9
MN2-2075 Rev.0
Chapter 1 Introduction
(Blank Page)
1 - 10
Chapter 2
SERVICE PROCESS
APPENDIX
MN2-2075 Rev.0
Chapter 2 Service Process
The typical processes for the repair work are shown as the Flow Chart on the next page. Do the repair
work according to this procedure. In the case of modification of the Technical Bulletin, Technical Notes
or Upgrade Kit, see the next item 2-2.
Each procedures of flow chart are numbered to refer its detail shown from page 2-3. Furthermore, the
Flow Chart and its explanation show the time when each section of service manual are required on
repair work. This is a guide for the usage of service manual.
The service manual is very important for the repair work, especially readjustment and performance
check after completion of repair work. This is to keep the safety and quality of equipment. If you make
them, you have to describe that the treatment has been done according to the applied section of service
manual, on the repair report or the like.
The circled numbers shown in the Flow Chart on next page are corresponded to the procedure number
shown from page 2-3.
2-1
MN2-2075 Rev.0
Chapter 2 Service Process
Demand of repair
(START)
5 VOLUME 1 Chapter 1, 5
Repair and adjustment VOLUME 2 Chapter 4
7
No
Work as normal?
Yes
8 10
Check by customer Yes
Fill repair report Repair again
Approve by customer
No
9 Present repair report 11 Show comments of the
Demand to repair the prohibition to use on
defective part(s) the equipment
2-2
MN2-2075 Rev.0
Chapter 2 Service Process
Accept the repair request from the customer or distributor. At this time, the following points have to be
confirmed and checked,
• Model name/number, and serial number
• Name of customer (Hospital), address, phone number, and name of person in charge
• Configuration of the connection of peripheral devices
• Software version or the like shown on the Maintenance display (if possible)
• Detail of phenomenon appeared on the function of equipment
Make an examination what circuit may be defective as the function of equipment based on the above
information. If you need to know about the basic operation and special information for the maintenance,
refer to the following sections, or ask to the Technical Support,
The reported phenomenon may be the original problem on the equipment. Because, refer to the
Technical Bulletin or the Technical Notes separately issued to check it whether defectiveness or not. If it
has been reported as the original problem, make a work according to the Technical Bulletin or the
Technical Notes.
* The Technical Bulletin and the Technical Notes are released for Service Engineer who has taken
appropriate training.
If you find the doubtful circuit, order the necessary parts. Then check the delivery date and decide the
date to visit on the consultation with the customer.
For the selection and order of parts, refer to the following sections
♦ VOLUME 1 Chapter 5 TROUBLE SHOOTING
♦ VOLUME 2 Chapter 5 PARTS LIST
For the electrical parts such as UNIT, check the history information on the HISTORY of this equipment
separately issued.
* The HISTORY of this equipment is released for Service Engineer who has taken appropriate
training.
2-3
MN2-2075 Rev.0
Chapter 2 Service Process
Check the required tools, measuring devices and parts to be replaced before the visiting the customer.
Then check the special information for the equipment reference with the following section,
♦ VOLUME 1 Chapter 2 SERVICE PROCESS
Confirm the appeared phenomenon and condition to happen it with the customer. If you don’t know
about the operation of equipment, refer to the Operation Manual attached to the equipment.
Repair the defective circuit with the brought parts. For the repair work, read the following section
carefully,
♦ VOLUME 1 Chapter 1 INTRODUCTION
♦ VOLUME 1 Chapter 2 SERVICE PROCESS
And, examine the trouble reason depending on the situation with following section,
♦ VOLUME 1 Chapter 5 TROUBLESHOOTING
The electrical or mechanical readjustment may be requested depending on the replaced parts. Because,
refer to the following section after completion of repair,
♦ VOLUME 2 Chapter 4 ADJUSTMENT
Check the system behavior to keep its condition as same as before in trouble, reference with the
following section. Be sure to do according to the description because check items are depending on the
portion to be treated.
♦ VOLUME 1 Chapter 6 PERFORMANCE CHECK
2-4
MN2-2075 Rev.0
Chapter 2 Service Process
If the result of “Procedure 6” is passed to the all standards, do the next “Procedure 8”. On the other side,
if not, make a judgment of “Procedure 10”.
Reconfirm the solution of trouble phenomenon with the customer. Then make a repair report and obtain
approval of customer.
The repair report shows not only the treatment but also the method of readjustment and operation check.
If they have been done according to the service manual, the followings have to be shown,
“Readjusted according to the VOLUME 2 Chapter 4 of service manual.”
“Checked according to the VOLUME 1 Chapter 6 of service manual, and passed.”
Fill the repair report with necessary item, and present it according to the certain procedure.
If the defective parts that trouble cause included is available to use again by repair, make an order to do.
If you cannot judge whether the part can be used again or not, ask to the Technical Support.
As the result of judgment on “Procedure 7”, if the trouble is not solved, judge the possibility to make
the repair work again.
If available, return to “Procedure 5” and continue to work.
If unavailable, go to “Procedure 11”.
As the result of judgment on “Procedure 10”, if you judge that it is impossible to continue the repair
work at this time, indicate that the equipment is still out of order, and also show the prohibition to use,
on the equipment.
Report the reason why the trouble cannot be solved to the customer. Then consult about the plan of next
repair work.
And do the same way from “Procedure 2”.
2-5
MN2-2075 Rev.0
Chapter 2 Service Process
The typical processes for the upgrade work are shown as the Flow Chart on the next page. Do the
upgrade work according to this procedure. In the case of repair work, see the previous item 2-1.
Each procedures of flow chart are numbered to refer its detail shown from page 2-8. Furthermore, the
Flow Chart and its explanation show the time when each section of service manual are required on
upgrade work. This is a guide for the usage of service manual.
The service manual is very important for the upgrade work, especially readjustment and performance
check after completion of upgrade work. This is to keep the safety and quality of equipment.
The circled numbers shown in the Flow Chart on next page, are corresponded to the procedure number
shown from page 2-8.
2-6
MN2-2075 Rev.0
Chapter 2 Service Process
Demand of Upgrade
(START)
14
4
Work as normal? No
11
Yes Can recover? Yes
8 Check by customer
No
2-7
MN2-2075 Rev.0
Chapter 2 Service Process
Accept the upgrade request from the customer, distributor or person in charge of sales. At this time, the
following points have to be confirmed and checked to decide the parts and kits,
• Document name that announced the upgrade or kit requested
• Model name/number, and serial number
• Name of customer (Hospital), address, phone number, and name of person in charge
• Configuration of the connection of peripheral devices
• Software version or the like shown on the Maintenance display
Make an examination what parts or kits are required based on the above information. For the selection,
refer to the following document separately issued, or ask to the Technical Support,
♦ Technical Bulletin
♦ Technical Notes
To confirm the detail of upgrade, see the Installation Procedure attached with applied Technical Bulletin
or Technical Notes.
Depending on the upgrade, hardware, or software, the other upgrade may be required. Check it with
the Technical Bulletin or Technical Notes.
Then, confirm the delivery date of required parts or kits, and decide the date to visit on the consultation
with the customer.
* The Technical Bulletin and the Technical Notes are released for Service Engineer who has taken
appropriate training.
Check the required tools, measuring devices and parts or kits to be used before the visiting the customer.
Then check the special information for the equipment reference with the following section and
document,
♦ VOLUME 1 Chapter 1 INTRODUCTION
♦ VOLUME 1 Chapter 2 SERVICE PROCESS
♦ Technical Bulletin, Technical Notes and/or Installation Procedure
2-8
MN2-2075 Rev.0
Chapter 2 Service Process
On the basis of work, the upgrade to the defective equipment is prohibited. Because, before upgrade
work, check the behavior of equipment whether normal or not according to following section and
document,
♦ VOLUME 1 Chapter 6 PERFORMANCE CHECK
♦ Operation Manual
If the result of “Procedure 3” is passed to the all standards, do the next “Procedure 5”. On the other side,
if not, go to “Procedure 10”.
Check the system behavior to keep its condition as same as before the upgrade, reference with the
following section. Be sure to do according to the description because check items are depending on the
portion to be treated.
♦ VOLUME 1 Chapter 6 PERFORMANCE CHECK
♦ VOLUME 1 Chapter 3 INSTALL/DISASSEMBLE
♦ Installation Procedure
If the result of “Procedure 6” is passed to the all standards, do the next “Procedure 8”. On the other side,
if not, make a judgment of “Procedure 11”.
2-9
MN2-2075 Rev.0
Chapter 2 Service Process
Reconfirm any functions of equipment with the customer. Then, if need, introduce and explain about the
new functions and specification added by this upgrade.
Furthermore, if need, make a report to be approved by the customer. The report shows not only the
treatment but also the method of operation check. If it has been done according to the service manual,
the following has to be shown,
“Checked according to the VOLUME 1 Chapter 6 of service manual, and passed.”
According to the Technical Bulletin or Technical Notes, return the unnecessary replaced or unused parts
as soon as possible if suggested.
And, if the report of upgrade is suggested on the same document, report it with the information required.
On the result of judgment in “Procedure 4”, if the equipment does not work normal, solve the problem
according to item 2-1 “Repair work on the description of service manual” shown in this section.
When the problem is solved, return to “Procedure 5” of this item and continue to do the upgrade work.
As the result of judgment on “Procedure 7”, if the problem has been made by this upgrade, judge the
possibility to recover it.
If available, return to “Procedure 5” and continue to work.
If unavailable, go to “Procedure 12”.
2 - 10
MN2-2075 Rev.0
Chapter 2 Service Process
As the result of judgment on “Procedure 11”, if you judge that it is impossible to recover at this time,
indicate that the equipment is the out of order, and also show the prohibition to use, on the equipment.
Report to the customer that the upgrade has not been completed because of the problem on the upgrade
work. Then make a schedule to fix and complete it
Report to the Technical Support about the happening of problem on the upgrade work, make an
examination to solve and order the additional parts. Before the asking, check the following points,
• Name of kit, or the issue number of Technical Bulletin or Technical Notes showing the upgrade
• Model name/number, and serial number
• Configuration of the connection of peripheral devices
• Software version or the like shown on the Maintenance display
• Indication of equipment such as Upgrade
• Detail of phenomenon appeared on the function of equipment
2 - 11
MN2-2075 Rev.0
Chapter 2 Service Process
(Blank page)
2 - 12
Chapter 3
INSTALL/DISASSEMBLE
APPENDIX
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
In order to proceed the work smoothly, refer this document with keeping the following points in
mind.
1) Index
Index number
2) Flowchart
Number of
Flowchart
· Flowchart represents the flow of disassembling parts/units. It flows from left side to right
side.
· It is necessary that the part/unit at the left side, connected with a line, is already
disassembled when you need to do it on a particular part/unit.
• The number indicated on each part/unit corresponds to the index No. of
Disassembly Diagram.
3-1
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3) Disassembly Diagram
Index No.
· It represents the disassembly of parts/units and the circled numbers put on the parts
indicate the order of dismantlement.
· The comment: (Details x-x) under the part name indicates there is Detailed Procedure of
Disassembly x-x in latter part of this document for your reference.
· The description of “BNK3x8 Fe,Ni,3” or like, indicates the type of screws and the number at
the end represents the amount of it.
· The index No. of Disassembly Diagram corresponds to the number of Flowchart.
Index No.
Sub-Index No.
· It provides more detailed procedure that could not see in the Disassembly Diagram.
2) The Index No. of Detailed Procedure of Disassembly corresponds to the Index No.
of Disassembly Diagram and the Sub-Index No. corresponds to the number of
(Details x-x).
3-2
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3-3
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
2. Dismounting Flowchart
3. Removing of Covers, Dust Filters, Rear Handle, Speaker and Pedal Unit
12. Removing of Probe Holder, Holder Base, Cable Hook, Panel locking System and Monitor
Arm Base
Caution
3-4
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Monitor
IPF-2101*
Operation Panel
PNL-132*
Physiological
Signal display unit
PEU-ARIETTA70*
3-5
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
2. Dismounting Flowchart
The Disassembling Instruction is composed based on the Flowchart so follow the order of this chart
to disassemble the equipment.
The numbers of Flowchart correspond to the Index No. of Disassembly Diagram and Detailed
Procedure of Disassembly.
3-6
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3. Removing of Covers, Dust Filters, Rear Handle, Speaker and Pedal Unit
[ Disassembly Diagram ]
Removing of Covers
Blind Cap
1
Cover: Arm
2
BNK3×8Fe-Ni
(Detail 3-1)
Cover: Right
6
BNK3×8Fe-Ni, 3
Cover: Foot
6 (Detail3-7)
(Detail 3-8)
Dust Filter Cover: Right Base
4 5
(Detail 3-5) BNK3×16Fe-Ni,2
(Detail 3-6)
3-7
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Removing of Covers
Cover: Top
9
BNK3×8Fe-Ni, 3
Dust Filter
2
(Detail 3-2)
Cover: Left
4
BNK3×8Fe-Ni, 3
Cover: PEU
6 (Detail 3-4)
BNK3×8Fe-Ni, 4
Cover: Left Base
Cable Clip Cover: Rear Assy 3
5 BNK3×16Fe-Ni,2
8 BNK4×10Fe-Ni,4 (Detail 3-3)
3-8
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Cover: Front
3
BNK3×8Fe-Ni, 6
(Detail 3-11)
Cover: Front(Plate)
2
(Detail 3-10)
Lock Knob
1
(Detail 3-9)
3-9
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Handle Grip
1
BNK3×8Fe-Ni, 4
Rear Handle
2
HB6×12SUS P1, 4
(Detail 3-12)
Speaker Cable
2
BNK3×8Fe-Ni, 8
Speaker Fixture
1
BNK3×8Fe-Ni, 8
(Detail 3-13)
3 - 10
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Pedal Unit
3
BNK4×8Fe-Ni, 2
(Detail 3-16)
Pedal
1
BNK4×10Fe-Ni, 3
(Detail 3-14)
Pedal Cover
2
BNK4×10Fe-Ni, 3
(Detail 3-15)
3 - 11
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
①
①With the up-down pedal of operation panel,
raise the operation panel.
Operation Panel
② Blind Cap, 2
BNK3×8Fe-Ni, 2
3 - 12
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
④
③
3 - 13
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 14
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
①Unfasten 3 screws.
2
Slide the Left Side Cover forward and
remove the hooks followed the procedures
described in figure and remove the Left Side
3
1 Cover.
BNK3×8Fe-Ni, 3
1
4
3 - 15
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Dust Filter
3 - 16
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
①Unfasten 3 screws.
2 Slide the Right Side Cover forward and
remove the hooks followed the procedures
described in figure and remove the Right
3 Side Cover.
1
BNK3×8Fe-Ni, 3
1
4
3 - 17
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
① ②
②
Foot Cover
3 - 18
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Lock Knobs, 5
①
D-cut
3 - 19
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
②
②Remove hooks at the upper of the Front
Cover (Plate).
3 - 20
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
<Positions of Boss>
Front Cover
3 - 21
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
① HB6×12SUS P1, 4
NOTE:
3 - 22
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Speaker Fixture
BNK3×8Fe-Ni, 8
< Attention on assembly >
Speaker Yellow (long) of the cable is left.
3 - 23
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Pedal
①Unfasten 3 screws and remove the Pedal.
BNK4×8 Fe-Ni,3
Pedal Cover
BNK4×10 Fe-Ni,2
3 - 24
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK4×10 Fe-Ni,2
BNK4×8 Fe-Ni,4
3 - 25
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
(Blank Page)
3 - 26
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
TGC Knobs, 8
2
JOYSTICK KNOB, 5
LCD Panel
1 2
BNK3×10 Fe-Ni, 4
(Detail 4-4)
DIAL (SMALL), 4
(Detail 4-1) 2
KNOB (M), 1
KNOB (S), 1 2
2
パネルベースキャップ,2
3
(Detail 4-4)
SW Panel DIAL(LARGE), 1
3 2
BNK3×10 Fe-Ni, 3 (Detail 4-1)
BNK3×8 Fe-Ni, 4
(Detail 4-5)
3 - 27
MN2-2075 Rev.0
Chaoter 3 Install/Disassemble
Key Caps
1
(Detail 4-2,3)
Track Ball
1
BNK3×10 Fe-Ni, 2
(Detail 4-7)
3 - 28
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 29
MN2-2075 Rev.0
Chaoter 3 Install/Disassemble
KEY TOP
1
Translucent Label
Panel Case Lower 2
7
(Detail 4-11)
DIAL HOLDER(SMALL)
4 Panel Frame
3T3×3 Fe-Black-Cut Point
BNK3×6Fe-Ni, 11
6
(Detail 4-12) (Detail 4-11)
PUDDLE
5
(Detail 4-11)
DIAL HOLDER(LARGE)
4
3T3×3 Fe-Black-Cut Point
(Detail 4-12)
3 - 30
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 31
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Key Cap
Ditch
Ditch
3 - 32
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Cap
3 - 33
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Key Cap
Ditch Ditch
3 - 34
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 35
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×10 Fe-Ni, 2
3 - 36
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 37
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
⑦Unplug 4 connectors.
BNK3×10 Fe-Ni
3 - 38
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×10 Fe-Ni, 3
②Unfasten 4 screws.
BNK3×8 Fe-Ni, 4
3 - 39
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 40
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×10 Fe-Ni, 2
3 - 41
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
② Unplug 2 connectors.
3 - 42
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×8 Fe-Ni, 6
3 - 43
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×8 Fe-Ni, 6
BNK3×8 Fe-Ni, 3
3 - 44
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×8 Fe-Ni, 3
⑦Loosen 3 screws.
3 - 45
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×8 Fe-Ni, 14
3 - 46
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
③Unplug 4 Connectors.
BNK3×8 Fe-Ni, 6
3 - 47
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
S3×6 Fe-Ni-EC, 4
BNK2×6 Fe-Ni, 4
3 - 48
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×6 Fe-Ni, 2
BNK3×6 Fe-Ni, 15
②Unfasten 15 screws, and remove the PANEL
SWITCH PCB.
3 - 49
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 50
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 51
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
(Blank Page)
3 - 52
MN2-2075 Rev.1
Chapter 3 Install/Disassemble
DBF
EP572500**
8
CELL
EU-9159*/EU-9158*
CONT 12
(Detail 5-6)
EP572600**
9
(Detail 5-4)
TVIF
EP573000**
11
BNK3×8 Fe-Ni, 4
HDD
10
BNK3×6 Fe-Ni, 4
(Detail 5-5)
RX
EP572300**
5 FootSWStack
EP573300**
JUMPER 4
EP572400** BNK3×8 Fe-Ni, 2
2 (Detail 5-3)
BNK3×8 Fe-Ni, 2
(Detail 5-2) Probe Selector
USM Cover CWSRV EP572100**
1 EP572900** 3
(Detail 5-3)
BNK3×8 Fe-Ni, 5(Unfasten) 6
BNK3×8 Fe-Ni, 25(Loosen)
(Detail 5-1)
TX
EP572200**
7
3 - 53
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×8 Fe-Ni,5
(Unfasten)
BNK3×8 Fe-Ni,25
(Loosen)
JUMPER
①Unfasten 2 screws, and remove the
JUMPER.
BNK3×8 Fe-Ni, 2
3 - 54
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
① BNK3×8 Fe-Ni,2
Fixture
HB4×16 SUS,1
②
BNK3×8 Fe-Ni, 8
3 - 55
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Probe Selector
④
④Unfasten 2 screws, and remove the
FootSWStack.
FootSWStack
BNK3×8 Fe-Ni, 2
3 - 56
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
LAN Cable
Cable Clamp
② LAN Cable
3 - 57
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
③
CONT
① BNK3×6 Fe-Ni, 4
HDD
3 - 58
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Fixture
CELL
3 - 59
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
(Blank Page)
3 - 60
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
UNIT FAN
3
BNK3×30 Fe-Ni, 8
(Detail 6-4)
PeripheralIO
EP572800**
1
BNK3×6 Fe-Ni, 4
(Detail 6-1)
BACKPLANE ASSY
EP572700**
3
BNK3×8 Fe-Ni, 23
(Detail 6-3)
USM UNIT
(USM-40*)
2
BNK4×8 Fe-Ni, 3
(Detail 6-2)
3 - 61
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Cable Clamp
①
①Unplug connector of the FAN cable, and
remove the Cable Clamp.
3 - 62
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Connector
②
LAN Cable Connector
BNK3×8 Fe-Ni,2
3 - 63
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
③ BNK3×6 Fe-Ni,4
①Unfasten 4 screws.
BNK3×6 Fe-Ni,4
3 - 64
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK4×8 Fe-Ni,3
②
FAN Cable
3 - 65
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
CNK3×8 Fe-Ni,23
①
BACKPLANE ASSY
3 - 66
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Cable Clamp
3 - 67
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Breaker Fixture
1
BNK3×8 Fe-Ni, 2
Lock Fittings
2
BNK4×30 Fe-Ni, 2
3 - 68
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
PEU UNIT
1
BNK3×8Fe-Ni, 3
※Refer Installation Procedure(MS5-5148)
3 - 69
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
CO-PEU-AR70-B
3
S2.6×6, 4
CO-PEU-AR70-A
CO-PEU-AR70-C 2
4 S3×8 Fe-Ni, 2
S2.6×6 Fe-Ni, 4
3 - 70
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Cover: Rear
1
Monitor Unit BNK3×8 Fe-Ni, 4
3 (Detail 9-1)
BNK4×10 Fe-Ni-EC, 6
(Detail 9-2)
Monitor Module
6
(Detail 9-4)
Cover: Escutchen
5
BNK3×8 Fe-Ni, 10
(Detail 9-4)
3 - 71
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK3×8 Fe-Ni,4
3 - 72
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
①
①Fasten 2 screws, remove the Lower Cover.
Lower Cover
BNK3×8 Fe-Ni,4
DVI Cable ②
③ Cable Clamp
3 - 73
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
④ Power Cable
Grounding Clamp
⑤
BNK3×8 Fe-Ni,1
BNK4×10 Fe-Ni-EC,2
(Loosen) ⑥
BNK4×10 Fe-Ni-EC,4
(Unfasten)
3 - 74
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Handle
BNK4×10 Fe-Ni,2
BNK3×8 Fe-Ni,10
Escutcheon
3 - 75
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
①Abnormal display, showing the monitor display on the touch panel, is caused in cases the
equipment is started without detection of the monitor as listed below. In case of the
abnormal display, follow the “Recovery of the monitor display” below.
3 - 76
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Cover: SECONDARM(U)
8
Cover: TILT(Upper) No Screws, fitness
Cover: TILT(Lower) 12 (Detail 10-5)
11 No Screws, fitness
CNK3×8, 4 (Detail 10-9) Cover: BASETILT(Left)
(Detail 10-9) 10
BK3×6, 2
Cover: (Detail 10-8)
SECONDSHAFT (Right) Cover:
6 SECONDSHAFT(Left)
BK3×6, 1 7
(Detail 10-4) BK3×6, 1
(Detail 10-6)
Cover:
BASETILT(Right)
9
BK3×6, 2
(Detail 10-7)
Cover:
SECONDARM(Left)
Cover: 5
FIRSTARM(Right) CNK3×8, 2
2 (Detail 10-3)
CNK3×8, 2
(Detail 10-1)
Cover:
FIRSTARM(Left)
Cover: SECONDARM(Right) 3
4 CNK3×8, 2
CNK3×8, 2
(Detail 10-2) (Detail 10-1)
3 - 77
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
CNK3×8Fe-Ni, 4
②
②Remove the First Arm Cover.
3 - 78
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
CNK3×8Fe-Ni, 2
3 - 79
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
CNK3×8Fe-Ni, 2
3 - 80
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BK3×6Fe-Ni, 1
3 - 81
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BK3×6Fe-Ni, 1
BK3×6Fe-Ni, 2
3 - 82
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BK3×6Fe-Ni, 1
3 - 83
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
CNK3×8Fe-Ni, 4
3 - 84
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
S3×8Fe-Ni, 2
Power Cable
DVI Cable
3 - 85
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
③
③Unfasten 4 screws, and remove the Monitor
Arm.
HB4×12SUS-EC, 4
3 - 86
MN2-2075 Rev.0
Chapter 3 Install/Dosassemble
Cover: ARM(T)
3
HB3×6SUS, 2
(Detail 11-1)
Cap×2
2
No Screws, fitness
(Detail 11-1)
Cover: BASEPANEL
6
BNK3×6, 2
(Detail 11-4)
Cover: ARM(L)
5
Cover: ARM(R) BNK4×10, 2
4 (Detail 11-2)
BNK4×10, 2
(Detail 11-3)
3 - 87
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Cap, 2
②Unfasten 2 screws.
HB3×6SUS, 2
3 - 88
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
③
③Down the panel Arm to the lowest position.
3 - 89
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK4×8Fe-Ni, 2
3 - 90
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK4×8Fe-Ni, 2
BNK3×Fe-Ni, 2
3 - 91
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
BNK4×10Fe-Ni, 1
BNK4×10Fe-Ni, 1
3 - 92
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
③
③Unfasten 2 screws from frontward.
BNK4×10Fe-Ni, 2
BNK4×8Fe-Ni, 2
3 - 93
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
HB6×20SUS_P=1, 6
3 - 94
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 95
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Holder Base
5
HB6×12 SUS P=1, 4
Tray
1
BNK4×10 Fe-Ni, 4
Panel Locking Lever
4
S3×8 Fe-Ni, 2
3 - 96
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Removing of Casters
Caster: Rear, 2
①
Caster: Front, 2
①
N16, 2 N16, 2
W16, 2 W16, 2
(Detail 13-1)
3 - 97
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Nut
Put a jack in
this position
3 - 98
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Nut
Washer
3 - 99
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Nut
Put a jack in
this position
3 - 100
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Nut
Washer
③ Unfasten Nut and Washer, remove the Rear
Caster.
3 - 101
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
(Blank Page)
3 - 102
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
B/W Printer
Removing of One-device
B/W Printer
BNK4×10Fe-Ni, 2
①
B/W Printer
BNK4×10Fe-Ni, 2
②
Color Printer
BNK4×10Fe-Ni, 2
③
3 - 103
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
B/W Printer
BNK4×10Fe-Ni, 2
①
Color Printer
BNK4×10Fe-Ni, 2 ②
B/W Printer
BNK4×10Fe-Ni, 2
①
Color Printer
BNK4×10Fe-Ni, 2
②
B/W Printer
BNK4×10Fe-Ni, 2
①
HDD RECORDER
BNK4×10Fe-Ni, 2
②
3 - 104
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
Color Printer
BNK4×10Fe-Ni, 2
①
2-device / 3-device Mounting Box (M)
S4×10Fe-Ni, 4
②
B/W Printer
BNK4×10Fe-Ni, 4
③
Side Mount BW Printer Mounting Rack (70)
BNK4×10Fe-Ni, 2 ④
3-device Mounting Rack (70)
BNK4×10Fe-Ni, 12
⑤
HDD RECORDER
BNK4×10Fe-Ni, 2 ①
Color Printer
BNK4×10Fe-Ni, 2
②
2-device / 3-device Mounting Box (M)
S4×10Fe-Ni, 4
③
B/W Printer
BNK4×10Fe-Ni, 4
④
Side Mount BW Pronter Mounting Rack (70)
BNK4×10Fe-Ni, 2
⑤
3-device Mounting Rack (70)
BNK4×10Fe-Ni, 12 ⑥
※Refer Installation Procedure:
MP-FX-AR70-36/-37/-38(MS5-5170)
3 - 105
MN2-2075 Rev.0
Chapter 3 Install/Disassemble
3 - 106
Chapter 4
SYSTEM OVERVIEW
APPENDIX
MN2-2075 Rev.1
Chapter 4 System Overview
Operating mode B
M (includes FAM)
D (PW Doppler, CW Doppler, HPRF PW, Dual Gate
Doppler)
Color Flow
Power Doppler (Directional Power Doppler)
eFlow (Directional eFlow)
RT-3D (SOP-ARIETTA70-4, EU-9163 are required)
Panoramic View (SOP-ARIETTA70-1 is required)
Freehand 3D (SOP-ARIETTA70-37 is required)
Real-time Tissue Elastography
(SOP-ARIETTA70-43 is required)
Real-time Virtual Sonography
(SOP-ARIETTA70-62 is required)
Image Display mode B, 2B, 4B
M, B/M
D, B/D
B(Flow), B(Power Doppler), B(eFLOW), 2B(Flow),
4B(Flow)
2B(Power Doppler), 4B(Power Doppler)
2B(eFLOW), 4B(eFLOW), M (Flow), M (Power Doppler),
M (eFLOW), B(Flow)/M (Flow),
B(Power Doppler)/M(Power Doppler),
B(eFLOW)/M (eFLOW), B(Flow)/D, B(Flow)/Dual,
B(Power Doppler)/D, B(Power Doppler)/Dual,
B(eFLOW)/D, B(eFLOW)/Dual, B(Flow)/D simultaneous
real-time display (Triplex mode), B(Power Doppler)/D
simultaneous real-time display (Triplex mode),
B(eFLOW)/D simultaneous real-time display (Triplex
mode), B/B(Flow) simultaneous real-time display (Dual
Flow), B/B(Power Doppler) simultaneous real-time
display (Dual CF), B/B(eFLOW) simultaneous real-time
display (Dual Flow)
4-1
MN2-2075 Rev.1
Chapter 4 System Overview
4-2
MN2-2075 Rev.1
Chapter 4 System Overview
4-1-4 B-mode
Display Gray Scale 256 levels
Scanning area 100% to 25%, 1 degree step
Zoom HI zoom (real-time image):
Max. Up to 0.5 cm display image
The maximum magnification is 8.
PAN zoom (real time and frozen image):
Max. Up to 0.5 cm display image
The maximum magnification is 8.
Depth range selections 0.75/1.0/1.5/2.0/2.5/3.0/3.5/4.0/4.5/5.0, 1 cm intervals
from 5.0 to 24 cm , and 2cm intervals from 24cm to 40cm
(probe dependent)
Longitudinal and lateral inversion, Rotation by 90 degrees (probe dependent)
Gain 80 dB range (Gain and Dynamic range can be changed
after freezing)
Echo enhancement 8steps
Texture 2 steps(Smooth/Sharp), not available in M-mode
Persistence 8steps, including Off
TGC (time gain control) Gain versus depth curve control: 8 slide controls
TGC curve memory function
LGC (lateral gain control) Gain versus angle curve control: 8 sectors (Sector probes
only)
Dynamic range 40-90 dB
Gray map 8 types, changeable after frozen
AGC 4steps
Receiving filter 3 steps
View gamma 4types
Spatial Compound Imaging (by selected linear and convex sector probes)
Trapezoidal scan (possible by selected linear probes)
B steer function This function enables you to use oblique function not only
Flow/ Power Doppler mode but B mode.
Coded Function (Probe dependent) Transmitting and receiving of coded ultrasound pulse
increase energy and provides high S/N Image without
degradation of spatial resolution.
Adaptive Imaging HI REZ function 8 steps
Adaptive imaging is the technique to generate optimum
image with changing filter characteristic depending on
input signal. This technique which installed this ultrasound
diagnostic scanner reduces speckle noise on B-mode
images and provides uniformly and high contrast images.
Auto-optimizer (Optimization of gain, sound velocity adjustment)
4-3
MN2-2075 Rev.1
Chapter 4 System Overview
4-1-5 M-mode
Sweep method Moving bar
Sweep speed 300.0, 200.0, 133.3, 100.0, 66.7, 50, 40.0 mm/sec
Gain B-gain ±30 dB
Dynamic Range 40 dB-90 dB
AGC 4 steps
FAM (Free Angular M-mode) (SOP-ARIETTA70-5 is required)
4-4
MN2-2075 Rev.1
Chapter 4 System Overview
4-1-8 Manual
It is applicable to browse instruction manual on this system.
On-board operation manual(English / Japanese)
4-5
MN2-2075 Rev.1
Chapter 4 System Overview
4-6
MN2-2075 Rev.1
Chapter 4 System Overview
Patient data
Displayed information Conforms to DICOM 3.0 standard
Patient information ID (up to 64 characters), Name (up to 64 characters,
including middle name), Birthday, Sex, Age, Height,
Weight, Occupation
Study information Procedure ID, Accession, Study ID, Study Description
Referring physician, Reporting Phys, Sonographer
Data storage media
Main unit hard disk Usable space: Up to about 100 GB (About 72,000 still
images are storable.)
USB memory
USB HDD
CD-R
DVD-RAM
Network interface (DICOM format) 10 BASE/T or 100 BASE/TX, (automatically switched)
DICOM network communication (SOP-ARIETTA70-10 is required)
Conformity to DICOM service class Ultrasound image storage SCU
Ultrasound multi-image storage SCU
Storage media FSC/FSR
Print management SCU
Modality worklist management SCU
(For details, please refer to the DICOM Conformance
Statement issued by Hitachi Aloka Medical, LTD.)
Modality performed procedure step (MPPS) SCU
Storage Possible to store patient information directly to DICOM
file server
Print Possible to printout images with DICOM compatible
printer directly
Work list management Retrieval of patient and reservation information from
hospital information system (HIS)
NOTE: The HIS needs to be compatible with DICOM
standard supplement 10. The HIS network and the
DICOM network need to be linked.
Router setting possible
Compatible with SR (Structured Report) for OB, cardiology, vascular and abdominal
measurements (SOP-ARIETTA70-21 is required)
Query/Retrieve (SOP-ARIETTA70-59 is required)
Compatible with ED(Evidence Documents)
Profile
IHE (Integrated Healthcare Enterprise)
SWF (Scheduled Work Flow)
PIR (Patient Information Reconciliation)
Security measures
User authentication function is available.
3 Types of user authority can be set.It is possible to set
whether password is required or not at the start of
operation.
Teaching file can be made
4-7
MN2-2075 Rev.1
Chapter 4 System Overview
4-8
MN2-2075 Rev.1
Chapter 4 System Overview
4-9
MN2-2075 Rev.1
Chapter 4 System Overview
4 - 10
MN2-2075 Rev.1
Chapter 4 System Overview
4 - 11
MN2-2075 Rev.1
Chapter 4 System Overview
This ultrasound diagnostic machine (ARIETTA 70) consists of those units below
Main Body PSC-150
Ultrasound main unit USM-40
Frontend part (Tx/Rx, Digital Beam Former)
CONT part
Backend part
Power supply unit EU-6053
Main panel PNL-132
Viewing LCD monitor IPF-2101
Standard software
Real Time Doppler Auto Trace
eFlow
Spatial Compound
Trapezoidal scan
AIP
Optional devices
Physiological signal display unit PEU-ARIETTA70
CW & 3D unit EU-9163
Analog video signal output unit EU-9167
Independent probe connection unit
EU-9166
Alphanumeric keyboard unit EU-9165
Real-time Virtual Sonography EZU-RV3S
L-CABLE-940-12
EZU-RVF1
Foot switch MP-2345B
3-channel Foot switch MP-2819
Junction Box for HI VISION Probe
JB-293*
Junction Box for ProSound LN/CV Probe
JB-294*
Junction Box for ProSound Sector Probe
JB-295*
Jelly warmer JW-3000 (Mounting kit is required)
MP-FX-AR70-2-R or MP-FX-AR70-2-L
4 - 12
MN2-2075 Rev.1
Chapter 4 System Overview
DVD recorder
Sony HVO-500MD/HVO-550MD (Ver. 1.2.0 or later)
DVI-D connection
(PM-AR70-H003 and MP-FX-AR70-28 are required)
Software Options
Contrast Harmonic Imaging software CHM-ARIETTA70
Panoramic View software SOP-ARIETTA70-1
Real Time 3D software SOP-ARIETTA70-4 (EU-9163 is required)
Free Angular M software SOP-ARIETTA70-5
Flow Profile measurement software SOP-ARIETTA70-7
DICOM network communication software
SOP-ARIETTA70-10
eTracking software (Ver. 1.2.0 or later) SOP-ARIETTA70-11
(PEU-ARIETTA70 is required)
TDI analysis software SOP-ARIETTA70-13
Stress Echo software SOP-ARIETTA70-15
(PEU-ARIETTA70 is required)
FMD software (Ver. 1.2.0 or later) SOP-ARIETTA70-16
(PEU-ARIETTA70 is required)
3D volume measurement software SOP-ARIETTA70-20
DICOM Structured Report software SOP-ARIETTA70-21
(SOP-ARIETTA70-10 is required)
Wave Intensity software (Ver. 1.2.0 or later)
SOP-ARIETTA70-34
(PEU-ARIETTA70 is required)
Freehand 3D software SOP-ARIETTA70-37
IMT auto measurement software SOP-ARIETTA70-38
STIC software SOP-ARIETTA70-41
(EU-9163 and SOP-ARIETTA70-4 are required)
NT auto measurement software SOP-ARIETTA70-42
Real-time Tissue Elastography software SOP-ARIETTA70-43
TVF software SOP-ARIETTA70-47
McAfee Embedded Control software SOP-ARIETTA70-48
(PM-AR70-S001 is required)
2D TT Analysis software SOP-ARIETTA70-49
4D shading software SOP-ARIETTA70-51
(EU-9163 and SOP-ARIETTA70-4 are required)
CW software SOP-ARIETTA70-55 (EU-9163 is required)
Tissue Doppler Image software SOP-ARIETTA70-56
Dynamic Slow-motion Display software SOP-ARIETTA70-57
EyeballEF software(Cannot be activated on ARIETTA 70a)
SOP-ARIETTA70-58
DICOM QR software SOP-ARIETTA70-59
4 - 13
MN2-2075 Rev.1
Chapter 4 System Overview
(SOP-ARIETTA70-10 is required)
Real-time Tissue Elastography Strain Histogram software
SOP-ARIETTA70-60
(SOP-ARIETTA70-43 is required)
Real-time Virtual Sonography software SOP-ARIETTA70-62
(EZU-RV3S, L-CABLE-940-12, and
HMC-EZU-RVF1 are required)
Picture in Picture software SOP-ARIETTA70-63 (EU-9167 is required)
4 - 14
MN2-2075 Rev.1
Chapter 4 System Overview
Optional Probes
4 - 15
MN2-2075 Rev.1
Chapter 4 System Overview
4 - 16
MN2-2075 Rev.1
Chapter 4 System Overview
Bi-plane Probe
4 - 17
MN2-2075 Rev.1
Chapter 4 System Overview
There are the lists of all PCB/unit and the system block diagram mentioned from the next page.
4 - 18
ARIETTA70 PCB LIST (1/2)
Ultrasound main unit EP5721 PROBE SELECTOR Probe switching, special probe function control, and probe code interface
EP5722 TX Transmission focus control, optional waveform transmitter
Low-noise preamplifier, variable gain amplifier, and high-speed analog/digital
EP5723 RX
converter
EP5724 JUMPER Probe selector board - Signal connection between TX and RX boards
4 - 19
The reception signal has phasing added and is output. The transmission signal
is generated from the transmission focus data. Also equipped with a standard
EP5725 DBF clock generation circuit for transmission timing, amplifier control circuit for
the Rx board, and an analog/digital / digital/analog voltage reference
generation circuit.
Processing of the RF signal output from the USP and the Doppler baseband
signal output from the CWS, control of the ultrasound transmission and
reception sequence, processing of biological data output from the body unit,
EP5726 CONT
ultrasound image data and graphic signal overlay, startup and stopping
sequence control, all hardware control interfaces, Video Clip/RT-3D/RF data
DMA transmission
EU-9159(2CELL) Software processing of the IQ signal received by the CONT board (black and
white, color flow, and Doppler beam processing, cine memory, scan
or CELL
conversion, brightness adjustment) and output to the CONT board as RGB
EU-9158(1CELL) data for display
EP5728 Peripheral I/O External input/output connector (network, USB)
EP5727 Backplane Motherboard
MN2-2075 Rev.1
Chapter 4 System Overview
ARIETTA70 PCB LIST (2/2)
Power Supply unit EU-6053 Power Supply unit transmission amplifier, and insulated AC 100 V or AC 200 V systems
provided for peripheral devices
[Op] Physiological
PEU-ARIETTA70 EP5509 Physio Amp. Biosignal (ECG, PCG, Pulse, Resp) amplifier, biosignal A/D
signal display unit
Chapter 4 System Overview
STCW reception signal A/D, Tx & Rx (quadrature detection) and A/D for
[Op] CW & 3D Unit EU-9163 EP5729 CWSRV
IND probe, 3D probe motor control
[Op] Independent Probe
EU-9166 EP558701 Indepe Staqck Connector board for independent probe connection
connection unit
[Op] Analog video
EU-9167 EP5730 TVIF Y/C signal output, Composite signal output
signal I/O unit
4 - 20
MN2-2075 Rev.1
Chapter 4 System Overview
4 - 21
MN2-2075 Rev.1
Chapter 4 System Overview
(Blank page)
4 - 22
MN2-2075 Rev.0
Chapter 4 System Overview
The microprocessor and CPLD on the CONT board primarily carry out system power control.
Operation using standby power, and control of the main power ON/OFF sequence for the
diagnostic system by the operation panel or the signal from the application.
4 - 23
MN2-2075 Rev.0
Chapter 4 System Overview
1/1
System Control
TITLE
4 - 24
MN2-2075 Rev.0
Chapter 4 System Overview
A board having 4 probe connectors installed that can connect to electronic probes (electronic
sector, electronic linear, and electronic convex) and one probe connector for connecting an
independent probe [option]. In addition, a FootSW connector is also installed on this board.
The electronic connectors have been changed from the previous ProSound and Hivision
types and a change has been made to the small type connector used by Noblus. In addition,
in order to make use of the existing ProSound and Hivision probes, a JB-BOX has been
made available and able to be connected.
4 - 25
MN2-2075 Rev.0
Chapter 4 System Overview
4 - 26
MN2-2075 Rev.0
Chapter 4 System Overview
There are two systems of the same circuit to carry out parallel
processing (maximum of 8 directions) on black and white receiving
signals.
Flow signal: RF gain processing (short-distance gain correction), pulse compression,
quadrature detection, LPF
There are two systems of the same circuit to carry out parallel
processing (maximum of 4 directions) on Flow receiving signals.
PW signal: RF gain processing (short-distance gain correction), quadrature
detection, LPF, sample gate addition
EDA (Expanded Doppler Analysis):
To increase the display quality of PW Doppler, this device divides
sample gate into multiple segments (maximum of 8 divisions), and
then adds it after carrying out simultaneous parallel processing of I/Q
signal data.
Illustration signal: RF gain processing (short-distance gain correction), quadrature
detection, LPF
CW signal: LPF, thinning processing
eTracking signal:BPF, zero cross detection and tracking processing after over-sample
processing
Eventually, the I/Q signal (in case of B/W signal, envelope signal) and eTracking data are
converted to the designated format, and then sent to the Backend board.
4 - 27
MN2-2075 Rev.0
Chapter 4 System Overview
4 - 28
MN2-2075 Rev.0
Chapter 4 System Overview
Transmission and reception signal timing and control signal generation section.
The ultrasound transmission/reception timing is generated by the timing control section on
the CONT board, and the transmission sequence is controlled based on the beam table and
beam sequence specified in the software.
4 - 29
MN2-2075 Rev.0
Chapter 4 System Overview
(Blank page)
4 - 30
MN2-2075 Rev.0
Chapter 4 System Overview
HUM R-WAVE
ECG/RESP R-SIGNAL
Filter Detect
ISOLATIOR
ECG AMP
A/D
MUX PHYSIO DATA
RESP AMP
AD CONT
PCG
MUX_SEL
PCG AMP
PULSE
DC-DC
PULSE AMP Converter
R-WAVE
DCIN R-SIGNAL
Detect
ECG DCIN
A/D DCIN DATA
ECG DCIN AMP MUX
4 - 31
MN2-2075 Rev.0
Chapter 4 System Overview
(Blank page)
4 - 32
MN2-2075 Rev.0
Chapter 4 System Overview
Incidentally, the overall system power ON/OFF is controlled as below via the CONT.
Power ON sequence
1) When the breaker SW is turned ON, +5 Vsb is output and the power-SW on the panel
turns orange.
2) When the power-SW is pressed, the power-SW changes to white together with a beep
sound and the power ON sequence begins.
3) The CONT requests a power supply start from the power supply unit EU-6053*.
4) The power supply unit begins the supply of electricity to each section and after the power
output has been fully launched, a PWR_OK signal is transmitted to the CONT.
5) The system (OS and application programs) starts.
4 - 33
MN2-2075 Rev.0
Chapter 4 System Overview
AC IN AC/DC J120
J800 ±200 V floating
200 V_ONx
+5 Vsb, Stand-by
J850 +5 Vsbbk,
Battery buck-up
Stand-by
Batt (option)
BTT_ENx
J851
PSU_ERRx
Voltage check
Latching Fault Protection J810
Over Current PANEL/
Over Voltage Monitor
Short Circuit
Temperature
Error Detect J820 FAN for
J821 Rack
12 VC1+12 VL
AC/AC AC Poly switch (F502)
12 Vf FAN_SIG 1, 2
on/off FAN for
Poly switch (F503) J823
relay
J824 Power
Supply
J801 Isolated
VOLTAGE Transformer VOLTAGE J802 AC-out
Selector (with Thermal cutoffs) Selector J803 0 to 320 VA
Switch (S1) Switch (S2) J804
4 - 34
MN2-2075 Rev.0
Chapter 4 System Overview
4 - 35
MN2-2075 Rev.0
Chapter 4 System Overview
(Blank page)
4 - 36
MN2-2075 Rev.0
Chapter 4 System Overview
DVI
Post
Monitor 12 V 5V PWR PNL PNL HDD USB comm.
12 V 5 V LVD STB SW# MNT# RST# ACC# (480 Mbps)
J530 J500 J501
4 - 37
MN2-2075 Rev.0
Chapter 4 System Overview
Post
・ 12 V power supply
Gel-warmer
・ Connection detection
signal
・ Short circuit detection
signal
・ Each power switch board information ・ TGC signal
and signal ・ Illumination intensity signal
・ PS/2 communications signal
・ Board type, revision, S/N
・ Remote control signal
・ Connection information ・ Board type, revision, S/N
・ LED rank information ・ LED rank information
・ LED brightness correction value ・ LED brightness correction
information value information
TGC board
Control ・ TGC lamp flashing signal
Board ・ Storage information
・ Power switch signal
・ Reset signal ・ Key signal
・ HDD lighting signal (In the board) ・ Rotary encoder signal
・ Key signal ・ Touch panel signal
・ Rotary encoder ・ Connection detection signal
signal ・ Board type, revision, S/N
・ Key lamp lighting ・ LED rank information
・ LED lamp lighting control data signal
Switch board ・ Key input information data ・ LED brightness correction
・ Board information value information
・ Encoder count information data
・ Type information
Power
UI bus ・ Revision information
(In the board) ・ Power switch lamp lighting switch
・ S/N information
・ Key signal signal
・ Sound source control
・ Rotary encoder signal ・ Key lamp lighting signal Board
signal
・ Key lamp lighting ・ Storage information
signal
・ Board information ・ LVDS video signal
・ Type information ・ BL brightness control Touch panel
・ Revision information signal signal
・ S/N information ・ BL lighting control signal
LCD
・ LED rank information ・ LCD type information +
・ LED brightness ・ LCD connection Touch panel
correction value information
information ・ Key signal
・ Joy stick signal
・ Rotary encoder signal Joy stick
Rotary encoder
(×5)
・ 5 V power supply
・ Control signal, information
Trackball
・ Operation signal, information
・ Status, trackball information
・ 5 V power supply
・ Status lighting information
・ Key lamp lighting control signal
PS/2
・ PS/2 communications Full keyboard
4 - 38
Chapter 5
TROUBLE SHOOTING
APPENDIX
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5-1. Introduction
This troubleshooting is a part of the Service Manual that is produced for providing a direction of repair
work. The description in this chapter is based on replacing PCB and unit.
5-2. Precaution
Please pay attention to the following instructions when you repair an equipment with referring this
troubleshooting in order to avoid occurring unnecessary trouble.
1) When you remove any of electrical part, like PCB, probe, cable, you should turn of the power of
the equipment.
When you remove parts that are biased by “standby power”, unplug the power cable from the
mains or turn off the circuit breaker and confirm that the power switch on the operation panel is off
the light.
2) When you disassemble an equipment, be sure to follow the disassembling procedure in order to
avoid damaging the equipment by disassembling it incorrectly.
3) To check/measure voltages and/or waveforms, you should be familiar with the specification and
handling of the measuring instrument.
4) When grounding a test probe of a measuring instrument or any other instrument, you need to know
the locations where measured signal grounds are. Be sure to identify the type of ground (analog,
digital, AC, DC, high voltage, or low voltage) before using it.
CAUTION If a ground of test probe is not connected to the correct ground, it may not
be possible to observer the correct voltage value or waveform, or the
relevant measuring instrument and/or the ultrasound diagnostic equipment
may burn out.
5) Be sure to turn off the power to the ultrasound diagnostic equipment before connecting a test probe
of any measuring instrument to a measuring test point.
6) Before replacing or repairing a PCB, determine the compatibility or other information of PCBs
following the “Software/Hardware History”.
CAUTION Inserting an incompatible PCB into the equipment may cause it to burn
out.
7) Avoid changing the setting of a variable resistor or switch on a PCB unless you are sure the
relevant component is faulty.
See "VOLUME 1: Chapter 6, Performance Check" to determine whether the relevant component is
faulty.
See "VOLUME 2: Chapter 4, Adjustment" if readjustments are required.
CAUTION Changing a variable resistor or switch without any reason may produce a
new problem, making it impossible to repair.
5-1
MN2-2075 Rev.0
Chapter 5 Troubleshooting
8) As you proceed with the troubleshooting as instructed here, you may need to consult with the
technical support. In such a case, provide at least the following information.
i) Model name of the equipment
ii) Serial number of the equipment
iii) History (repairs and upgrades) of the equipment and software version
iv) Description of the problem (send photos if necessary)
IMPORTANT When you contact the technical support, be sure to send the System
Information file to allow the technical support to make a correct decision
regarding the equipment information. To determine failure symptoms,
obtain an image or video of the equipment and send it with the description.
Standard tools and measuring instruments that are required for repair are:
1) Oscilloscope
Sensitivity : 5mV /div.
Frequency range : DC~50MHz
Maximum input voltage : 400V or more
2) Multi-meter
Class : 0.5 class
Mode : ACV, DCV, DCA, Ω
3) Test Phantom
Made by Gammex RMI
Gammex 403GS or equivalent
4) Probes
Convex : C251
Electronic sector : S12
5-2
MN2-2075 Rev.0
Chapter 5 Troubleshooting
It is summarizing the locations of each unit and module that belongs to the main body in the figure
below. In the case of actual maintenance (repairing and/or upgrading), refer to the Disassembling
Instruction and the Parts List.
Monitor
IPF-2101*
Operation Panel
PNL-132*
Physiological
signal display unit
PEU-ARIETTA70*
5-3
MN2-2075 Rev.1
Chapter 5 Troubleshooting
Rear Front
/EU-9158*
EU-9159*
Cell
5-4
MN2-2075 Rev.0
Chapter 5 Troubleshooting
CONT board
Do not remove/disassemble the CPU unit on the CONT board (EP572600**) that is mounted on it.
EP572600**
CPU
Disassembling
and/or replacement
is not possible
5-5
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5-6
MN2-2075 Rev.0
Chapter 5 Troubleshooting
There are some DIP switches and JP (Jumper connector) on a PCB that are for setting the initial
configuration and changing specification among various regional requirements. If the setting of them
were not in the state that is instructed, this will cause not only incorrect behave of the equipment but
some damage on it. These settings should not be altered unless you are instructed except for
maintenance purpose.
Several PCB have LED mounted on it and you could recognize the status of operation on the board
and surmise a defect of it as well. There are two type of LED indication that one is indicating the
status of the circuit independently, the other is observing the status between the board and other unit so
that it isn’t necessarily telling a problem in the board that is mounted.
This section describes the locations of LED and the statuses.
B-side
LED1
5-5-3 TX
EP572200**
There is no LED, DIP switch or jumper connector.
5-7
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5-5-5 RX
EP572300**
There is no LED, DIP switch or jumper connector.
5-5-6 JUMPER
EP572400**
There is no LED, DIP switch or jumper connector.
5-5-7 DBF
EP572500**
LD1(Red): It lights immediately after starting the equipment (Starting configuration)
It will be off when the configuration of EP5725 is done.
There is no DIP switch or jumper connector
LED1
EP572500**
B-side
A-side
5-8
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5-5-8 CONT
EP572600**
LED1(Green): Normal:”A” (It is combined with LED2 indicating 2 digit code; 7 segmental
display: Upper digit)
It indicates the status of booting sequence on BIOS/CPU.
LED2(Green): Normal:”0” (It is combined with LED1 indicating 2 digit code; 7 segmental
display: Lower digit)
It indicates the status of booting sequence on BIOS/CPU.
LED3(Green): Normally off; indicating “non-reset” status
It indicates the resetting status of PCIe from COMe module.
LED4(Green): Normally on; Indication “configuration is done”
It indicates the configuration status of FPGA(U8).
LED5(Green): Normally on; Indication “configuration is done”
It indicates the configuration status of FPGA(U32).
LED6(Green): Normally off; indicating “non-reset” status
It indicates the resetting status of PCIe from CELL module.
LED7(Green): Normally on; indicating “running”.
It indicates the status of firmware that is implemented on CELL module.
LED8(Green): Normally on; indicating non-reset status
It indicates the resetting status of CELL module.
LED9(Green): It is on when “Power Down” is done, and vice versa when it isn’t done.
It indicates the status of power supply on CELL module.
LED10(Green): Normally blink; indicating “running”.
It indicates the status of processor (U20) mounted on the module.
LED11(Green): Normally blink; indicating “normal”.
It indicates the status of timing processor on FPGA (U32).
LED12(Green): Normally off; indicating “lock”
It indicates the status of “PLL LOCK” (PCIe reference clock) of COMe
LED13(Green): Normally off; indicating “L0” status (Link established).
It indicates the status of LTSSM on COMe PCIe.
LED14(Green): Normally off.
It indicates the status of “Link Width” on COMe PCIe.
LED15(Green): Normally off; indicating “PLL lock” and “LTSSM L0”.
It indicates the status of “PLL LOCK” (PCIe reference clock) of CELL and
LTSSM status of PCIe
LED16(Green): Normally off
It indicates the status of “Link Width” on CELL PCIe.
LED17(Green): It blinks 8 times at the booting and PCIe reset (COMe).
LED on; indicating the command is being processed.
LED off; indicating that is waiting for new command.
It indicates the status of microcomputer for DMA controller
LED18(Green): It blinks 16 times at the booting and PCIe reset (CELL)
LED on; indicating the command is being processed.
LED off; indicating that is waiting for new command.
It indicates the status of microcomputer for FE controller
5-9
MN2-2075 Rev.0
Chapter 5 Troubleshooting
LED19(Green): It indicates the status of PN_ON that is passed to power supply from CONT
LED on: indicating the system is turned OFF (PS_ON is inactive)
LED off: indicating the system is turned ON (PS_ON is active)
LED20(Green): Normally off. It is for debugging
* When the base clock sig. for CONT (EP5726) isn’t provided from DBF (EP5725), all
LED11~LED18 blink. Otherwise the LEDs behave as described above.
U95 (DIP SW): #1 to #4 shows the board revision. #5 to #8 are "OFF" (fixed).
In the following picture, only #2 is ON and the board revision is "B."
U95
5 - 10
MN2-2075 Rev.1
Chapter 5 Troubleshooting
5-5-9 CELL
EU-9159*/EU-9158*
LED1 (Yellow): Cell Application
It is on at booting and freezing the image.
LED2 (Red): HW Error
It is on when it detects hardware error.
LED3 (Green): 3.3V AUX
It is on when 3.3V_AUX is running.
LED4 (Orange), CR5~10 (Green):These are for debugging.
BACKPLANE side
LED1
LED2
LED1
LED3
LED3 LED3
LED4
5 - 11
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Output threshold
+5Vsb 2V
+12VC1 8V
+5.1VS1 3V
+12V 8V
-12V -8V
+12V 8V
+5.1VS1P 3V
+75V 50V
-75V -50V
+100V 50V
-100V -50V
5 - 12
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5-5-15 Panel
PNL-132*
Operation Panel checks itself at booting and if it detected a problem, certain keys on it will blink
rapidly for indicating the problem.
(1) Flash memory abnormality
The orange LEDs of the Freeze key and B mode key flash rapidly.
(2) SRAM abnormality
The orange LEDs of the Freeze key and Dual key flash rapidly.
(3) I2C bus abnormality
In case there is a problem with the I2C (channel 0) itself
The white LEDs of the Freeze key and TGC0 flash rapidly.
In case there is a problem with the I2C (channel 1) itself
The white LEDs of the Freeze key and TGC1 flash rapidly.
In case there is a problem with the I2C (channel 3) itself
The white LEDs of the Freeze key and TGC3 flash rapidly.
In case there is a problem in communication with the illumination intensity sensor
The white LEDs of the Freeze key and TGC5 flash rapidly.
In case there is a problem in communication with the LCD electronic potentiometer
The white LEDs of the Freeze key, M, and TGC6 flash rapidly.
In case there is a problem in communication with the sound source electronic
potentiometer
The white LEDs of the Freeze key, M, and TGC7 flash rapidly.
In case there is a problem in communication with the trackball
The white LEDs of the Freeze key and TBF flash rapidly.
In case there is a problem in communication with the maintenance memory
(EEPROM)
The white LEDs of the Freeze key, M, and update flash rapidly.
In case there is a problem in communication with the control board information
memory (EEPROM)
The white LEDs of the Freeze key, M, and AUTO flash rapidly.
In case there is a problem in communication with the switch board information
memory (EEPROM)
The white LEDs of the Freeze key, M, and REC flash rapidly.
In case there is a problem in communication with the power switch board information
memory (EEPROM)
The white LEDs of the Freeze key, M, and PRINT flash rapidly.
In case there is a problem in communication with the TGC board information memory
(EEPROM)
The white LEDs of the Freeze key, M, and STORE flash rapidly.
(4) Panel controller abnormality
In case there is a problem with the panel controller (CPLD) register access
In case there is a problem with the PNLCNT1
The Freeze key LED and the white LED of B flash rapidly.
In case there is a problem with the PNLCNT2
The Freeze key LED and the white LED of PROBE flash rapidly.
(5) Trackball ball detection state abnormality
In case there is a problem with the trackball ball detection state (lack of sensitivity, no
ball)
The orange LEDs of the FREEZE key and TBF flash rapidly.
(6) Key scan bus abnormality
In case there is a problem with the bus for the key scan
The orange LEDs of the FREEZE key and U1 flash rapidly.
5 - 13
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5 - 14
MN2-2075 Rev.0
Chapter 5 Troubleshooting
SW1 SW2
Dip SW setting
SW1: Panel Test
Toggling enable/disable of “CTRL+ALT+DEL” and/or “F10”
on the full keyboard (PS2)
OFF: Disable (Default setting)
ON: Enable
SW2: Monitor
Select the video signal that is displayed on the monitor
NORM: Video output from VP block (Default)
MAINTE: Video output from CPU module
(For BIOS and Windows booting display)
Important When you set the SW1; ON then, ”Enter” and “UNDO” switch on the
Operation Panel will blink continually.
UNDO Enter
5 - 15
MN2-2075 Rev.0
Chapter 5 Troubleshooting
J810
A B
1 1
2 2
3 3
4 4
5 5
J851
3 6
LED 2 5
1 4
Backside of
equipment
5 - 16
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Instruction
1. Turn off the power to the equipment and turn off the circuit breaker and then, confirm
the power LED on the operation panel turns off.
2. Remove the covers to pull the power
supply unit EU-6053* out from USM
unit. Circuit Breaker
3. Take the EU-6053* out of the
system's main unit and reconnect the
circuit breaker and the power cable.
Power cable
J120
4. Short pin 1 (GND) and pin 4 (ALL_ONx) of J851 connector (Voltage Check).
J851
3 6
2 5
1 4
GND ALL_ONx
5. Turn on the circuit breaker for the equipment. As the power supply unit starts by itself,
power voltage is supplied. Check the voltage of each check terminal J120 referring to
the below section.
5 - 17
MN2-2075 Rev.0
Chapter 5 Troubleshooting
J120
Check Point
Output Voltage Standard
Connector No. Pin GND
P1 GND +12VC1 +11.4V ~ +12.6V
P4 GND +5.1VS1 +4.8V ~ +5.4V
P7 GND +5Vsb +4.5V ~ +5.5V
P8 GND +5Vsbbk +4.5V ~ +5.5V
P10 GND -12V -12.6V ~ -11.4V
P12 GND +12V +11.4V ~ +12.6V
P14 GND -5.1VL -5.4 V ~ -4.8V
P16 GND +5.1VL +4.8V ~ +5.4V
J120
P18 GND -12VL -12.6V ~ -11.4V
P20 GND +12VL +11.4V ~ +12.6V
D6 A6 -100V -105V ~ -95V
D8 A6 -75V -78V ~ -72V
D10 A6 +75V +72V ~ +78V
A10 A6 +100V +95V ~ +105V
D18 A16 -200V -210V ~ -190V
A18 A16 +200V +190V ~ +210V
6. After measuring voltage, turn off the circuit breaker and assemble the equipment by
reversing way from step 2) through 5).
5 - 18
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Check Point
Output Voltage Standard
Connector No. Note
J801
J802
Switched AC input voltage +/-10%
J803
J804
* Switched = No output with “Standby” condition
J801 J803
J802 J804
5 - 19
MN2-2075 Rev.0
Chapter 5 Troubleshooting
1 picture element
(1 pixel)
Each R, G, B sell is
called “dot”.
Bright dot: A defect that the dot is always displayed in brighter color (White, red,
green, blue, purple, light blue, yellow) on the LCD screen when it displays
“black” or dark color.
You can check it during the start-up of the machine (When the LCD is
black) or in the PRESET screen (Dark background).
5 - 20
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Dark dot: A defect that the dot always displayed in darker color (Red, green, blue,
purple, light blue and yellow) or it doesn’t light at all on the LCD screen
when it display “White” or brighter color. You can check it with “EXT”
screen or a test pattern display (If available).
MENU
- +
5 - 21
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Is the backlight lit? No Is the power supply cable No Connect the power
properly connected? supply cable properly.
Yes Yes
No
Connect the DVI cable properly.
Yes
Supposed to DVI cable disconnection
or connector malfunction.
Is the operation panel
properly worked?
5 - 22
MN2-2075 Rev.0
Chapter 5 Troubleshooting
If the system does not start or locks up during startup, refer to this section, and then perform repair work.
5 - 23
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5 loading and
starting OS
7 Starting Logon
process
11 Starting
transmission
Starting
monitor of Tx
voltage
12 loading Displaying *1
completion of ultrasound
measurement image
module
5 - 24
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Displaying
Lights in
1 Machine is running ultrasound
White
image
Sequentially Blinking in
Turn Power SW OFF light in white → White
2 Starting shutdown Lights off for
process some LEDs on
full-keyboard
Shutdown Lights in
4 All OFF
completed Orange
5 - 25
MN2-2075 Rev.0
Chapter 5 Troubleshooting
There are two types of messages are displayed on the equipment, the one is “dialogue message” and
the other is “assist message”. Some messages are used for indicating non-error information.
5 - 26
MN2-2075 Rev.0
Chapter 5 Troubleshooting
If there is a network error (DICOM communication error), see the following, and then perform repair
work.
1. Check for hardware and software failures
Follow the procedure in the section 5-10-1, and check that the network function on the Backend
unit and network function on the operating system are working normally. If an error is found,
replace the Backend unit or hard disk drive.
Reference For the ping test, use the network protocol named ICMP. Depending on the
computer settings, it is possible to set no response to echo requests in the ICMP
protocol.
If there are no problems with the DICOM function, and the ping test is
unsuccessful, there are no particular problems for usage.
If there are no problems with any of the settings, check with the network administrator of the
facility for unplugged network cables (LAN cables) and problems with hubs, routers, and
connection destination network devices.
If infrequent network problems are occurring, see 5-10-5 and then acquire and check the
communication data (packets) on the network. Consult with our technical support regarding
analysis of acquired data, and so on.
Reference For the C-Echo test, a service called Verification for DICOM communication is
used. Depending on the connection destination DICOM device, this service
may not be supported, and may not respond to C-Echo.
If there are no problems with the DICOM function, and the C-Echo test is
unsuccessful, there are no particular problems for usage.
5 - 27
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Check that the BE unit network function and operating system network function are operating
normally.
1. CHECKING THE BE unit
See 5-11, and then use the CPU self-diagnostic tool to diagnose the hardware. If this function
does not find any errors in the “PCI Network”, the network function installed in the Backend
unit is operating normally.
(1) Use a network cable (cross cable) to directly connect the ultrasound machine and
computer.
Cross cable
(2) Perform a ping test from the ultrasound machine to the computer (or vice versa). If a
response is received without issue, the operating system network function is operating
normally.
Reference Depending on the computer network settings, it is possible to obtain no
response to echo requests for ping command.
If the ping test fails from the ultrasound machine but succeeds from the
computer, there are no problems.
5 - 28
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Items
AE Title: Specify the AE title of the ultrasound machine. (maximum 16 characters)
Station Name: Specify the computer name of the ultrasound machine. (maximum 16
characters)
Port#: Specify the TCP port number used by the ultrasound machine for DICOM
communication. This port is used to receive data from external sources (an
inbound port). (value from 0 to 2349)
Obtain an IP address a DHCP server / Use the following IP address:
To automatically obtain the device IP address, select “Obtain an IP address a DHCP
server”. To use a fixed IP address, select “Use the following IP address”, and then
configure the section below manually.
- IP Address: Specify the IP address of the ultrasound machine.
- Router1/Router2: Configure this setting when the DICOM device to which you want to
connect is situated on a network address other than that of the
ultrasound machine. Otherwise, leave the fields blank.
- Subnet Mask: Specify the subnet mask of the ultrasound machine.
Test function
[Local Ping] Send a ping command to the loopback address.
Message Examination results
Ping check to **** successful. Shows the TCP/IP is operated.
Ping check to **** failed. Shows that TCP/IP failed to operate.
Contact your system administrator (IT
personnel) for the hospital network.
5 - 29
MN2-2075 Rev.0
Chapter 5 Troubleshooting
At this window, up to five image servers can be configured. However, it is possible to send images
to the specified server only.
Items
Connect: Select one button from [Remote1] to [Remote5] to configure the server to be
used.
Appl. Entry: Specify the AE title of the image server.
Station Name: Specify the computer name of the image server.
IP Address: Specify the IP address of the image server.
Port#: Specify the TCP port number to be used by the image server for DICOM
communication. This port is used to receive data from external sources (an
inbound port).
Test function
[Remote Ping] Send a ping command to the server (Remote1 to 5) specified under Connect,
and check that TCP/IP is performed correctly.
Message Examination results
Ping check to **** successful Show that there is a response for echo reply on ICMP
from image server. Generally, this is a guide that
network connection with a server is currently
established correctly.
Ping check to **** failed. Show that there is NO response for echo reply on
ICMP from image server. If the server has forbidden
sending an echo reply, there is no problem in this case.
5 - 30
MN2-2075 Rev.0
Chapter 5 Troubleshooting
[Remote C-ECHO] Send a C-echo command to the server (Remote1 to 5) specified under
Connect, and check that the DICOM communication function is
running.
メッセージ 結果
Echo check to **** Show that there is a response for echo reply on DICOM
successful. communication from image server. Generally, this is a
guide that DICOM communication with a server is
currently established correctly.
Echo check to **** failed. Show that there is NO response for echo reply on
DICOM communication from image server. If the
server does not support verification service, there is no
problem in this case.
[Send to Storage]
Photometric: Configure the storage format when sending color images over the network.
RGB This is color data. Specify the color of each pixel using its RGB
value.
Monochrome2
Black and white image data
JPEG Q Factor (50 - 99):
Configure the compression rate of images when JPEG Baseline is selected as
the transfer syntax. The smaller the number, the higher the compression and
the lower the image quality. (value from 50 to 99)
DICOM+Raw: When you store image in DICOM format, select presence of Raw image
data with it.
On Present
Off Not present
5 - 31
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5 - 32
MN2-2075 Rev.0
Chapter 5 Troubleshooting
The Worklist server can be configured separately from the image server. Confirm that the Box is
checked when you use it.
Items
Appl. Entry: Specify the AE title of the Worklist server.
Station Name: Specify the computer name of the Worklist server.
IP Address: Specify the IP address of the Worklist server.
Port#: Specify the TCP port number to be used by the Worklist server for DICOM
communication. This port is used to receive data from external sources (an
inbound port).
Test function
[Worklist Ping] Send a ping command to the Worklist server, and check that TCP/IP is
performed correctly.
Message Examination results
Ping check to **** successful Show that there is a response for echo reply on ICMP
from worklist server. Generally, this is a guide that
network connection with a server is currently
established correctly.
Ping check to **** failed. Show that there is NO response for echo reply on
ICMP from worklist server. If the server has forbidden
sending an echo reply, there is no problem in this case.
5 - 33
MN2-2075 Rev.0
Chapter 5 Troubleshooting
[Worklist C-ECHO] Send a C-echo command to the Worklist server, and check that the
DICOM communication function is running.
Message Examination results
Echo check to **** Show that there is a response for echo reply on
successful DICOM communication from worklist server.
Generally, this is a guide that DICOM communication
with a server is currently established correctly.
Echo check to **** failed. Show that there is NO response for echo reply on
DICOM communication from worklist server. If the
server does not support verification service, there is no
problem in this case.
Check that the check boxes of search keys to be sent to the Worklist server are selected.
Patient ID Modality
Patient Name Scheduled Station AE Title
Accession Number Scheduled Procedure Step Start Date
Requested Procedure ID
If “Scheduled Procedure Step Start Date” is selected, check that the examination date to be
searched is configured correctly.
Today:
Today +/- 1 day:
This week: Sunday to Saturday
Specified date from ** To **: Term from the "from" field date to the "to" field date
5 - 34
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Comply to VistA
If this check box is selected, Worklist operations comply with VistA, the electronic
health record system of the United States Department of Veterans Affairs.
With this setting, examination orders obtained from the Worklist server are all
displayed on the Worklist window, regardless of the “Patient ID”, “Patient Name”, or
“Accession Number” entered as search keys.
5 - 35
MN2-2075 Rev.0
Chapter 5 Troubleshooting
At this window, up to five DICOM printers can be configured. However, it is possible to send
images to the specified printer only.
Items
Model Name: Specify the name of the printer to be used in the settings. If the name does not
appear in the list, use General.
AE Title: Specify the AE title of the DICOM printer. (Max. 16 characters)
Station Name: Specify the computer name of the DICOM printer. (Max. 16 characters)
IP Address: Specify the IP address of the DICOM printer.
Port#: Specify the TCP port number to be used by the DICOM printer for DICOM
communication. This port is used to receive data from external sources (an
inbound port).
Test function
[Ping] Send a ping command to the specified DICOM printer, and then check that
TCP/IP is performed normally.
Message Examination results
Ping check to **** successful Show that there is a response for echo reply on ICMP
from DICOM printer. Generally, this is a guide that
network connection with a printer is currently
established correctly.
Ping check to **** failed. Show that there is NO response for echo reply on
ICMP from DICOM printer. If the printer has forbidden
sending an echo reply, there is no problem in this case.
5 - 36
MN2-2075 Rev.0
Chapter 5 Troubleshooting
[C-ECHO] Send a C-echo command to the specified DICOM printer, and check that the
DICOM communication function is running.
Message Examination results
Echo check to **** Show that there is a response for echo reply on
successful DICOM communication from DICOM printer.
Generally, this is a guide that DICOM
communication with a printer is currently
established correctly.
Echo check to **** failed. Show that there is NO response for echo reply on
DICOM communication from DICOM printer. If the
printer does not support verification service, there is
no problem in this case.
5 - 37
MN2-2075 Rev.0
Chapter 5 Troubleshooting
[DICOM Printer]
Items
Model Name: Select the DICOM printer to be configured.
Num of Copies: Specify the number of copies.
Medium type: Specify the print medium.
Film size: Specified the print page size.
Orientation: Specify the orientation of the print page.
Landscape
Portrait
Display format: Specify the displayed format when printing multiple images on a single
page.
Magnification: Specify the type of magnification processing.
Smoothing: Specify the type of smoothing processing.
Trim: Specify the white space.
No: No white space between images.
Yes: Specify the white space between images when printing multiple
images on a single page.
Configuration: Enter the configuration information of the printer.
Appl. Entity: Display the AE title of the DICOM printer.
Printer priority: Specify the priority of the print process.
Destination: Specify the film address.
Min density: Specify the minimum brightness for printing.
Max density: Specify the maximum brightness for printing.
Border density: Specify the brightness of the image border.
Empty density: Specify the brightness of areas not containing an image.
Image Polarity: Specify the polarity of the print image.
5 - 38
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5-10-2-5 Checking the DICOM settings for the SR (Structured Report) server
(1) Basic settings
IN THE "Preset Setup" - "SystemPreset" - "DICOM" - "SR", check that the SR server settings are
configured correctly.
Up to two SR servers can be configured. However, it is possible to send SR files to the specified
server only.
Items
Connect: Select one button from [Remote1] to [Remote2] to configure the SR server to
be used.
AE Title: Specify the AE title of the SR server. (Max. 16 characters)
Station Name: Specify the computer name of the SR server. (Max. 16 characters)
IP Address: Specify the IP address of the SR server.
Port#: Specify the TCP port number to be used by the SR server for DICOM
communication. This port is used to receive data from external sources (an
inbound port).
SR Auto Creation:
When set to ON and the New Patient switch is pressed after an examination,
an SR file is created based on the results of the automatic measurement, and
then the file is sent to the server.
When set to OFF, an SR file is not created automatically. With this setting,
you must manually create an SR file based on a report page if a SR file is
necessary.
5 - 39
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Test function
[Ping] Send a ping command to the server (Remote1 or 2) specified under Connect,
and check that TCP/IP is performed correctly.
Message Examination results
Ping check to **** successful Show that there is a response for echo reply on ICMP
from SR server. Generally, this is a guide that network
connection with a server is currently established
correctly.
Ping check to **** failed. Show that there is NO response for echo reply on
ICMP from SR server. If the server has forbidden
sending an echo reply, there is no problem in this case.
5 - 40
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Up to five QR server can be configured. It is possible to exchange the configured server from
the “Import” screen. The radio buttons are for selecting the server when you do Ping and
C-ECHO checking.
Items
AE Title: Specify the AE title of the QR server. (Max. 16 characters)
Station Name: Specify the computer name of the QR server. (Max. 16 characters)
IP Address: Specify the IP address of the QR server.
Port#: Specify the TCP port number to be used by the QR server for DICOM
communication. This port is used to receive data from QR server to the
equipment.
* Precaution
The reception TCP port# for “C-MOVE” DICOM communication is fixed to 2350. This port#
can be changed by the procedure below.
(1) Refer to 5-10-43: The handling of institution names during Worklist operations, run
Windows explorer.
(2) Create a file F:\UDE\pc_data\usdir\QR_LocalPort.ini and edit the contents of it as described
below.
Write the number you want to use at “port#”.
[QR]
LocalPort=port#
(3) Change the Port# in SystemPreset - DICOM – Common, “Ultrasound Equipment” to
temporal number and save it.
5 - 41
MN2-2075 Rev.0
Chapter 5 Troubleshooting
(5) Change the Port# again in SystemPreset - DICOM – Common, “Ultrasound Equipment” to
the value that is originally set at (3) then, save it.
Test function
[Ping] Send a ping command to the server (Remote1 or 2) specified under Connect,
and check that TCP/IP is performed correctly.
Message Examination results
Ping check to **** successful Show that there is a response for echo reply on ICMP
from MPPS server. Generally, this is a guide that
network connection with a server is currently
established correctly.
Ping check to **** failed. Show that there is NO response for echo reply on ICMP
from MPPS server. If the server has forbidden sending
an echo reply, there is no problem in this case.
5 - 42
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Up to two MPPS servers can be configured. However, it is possible to send MPPS data to the
specified server only.
Items
AE Title: Specify the AE title of the MPPS server. (Max. 16 characters)
Station Name: Specify the computer name of the MPPS server. (Max. 16 characters)
IP Address: Specify the IP address of the MPPS server.
Port#: Specify the TCP port number to be used by the MPPS server for DICOM
communication. This port is used to receive data from external sources (an
inbound port).
Test function
[Ping] Send a ping command to the server (Remote1 or 2) specified under Connect,
and check that TCP/IP is performed correctly.
Message Examination results
Ping check to **** successful Show that there is a response for echo reply on ICMP
from MPPS server. Generally, this is a guide that
network connection with a server is currently
established correctly.
Ping check to **** failed. Show that there is NO response for echo reply on ICMP
from MPPS server. If the server has forbidden sending
an echo reply, there is no problem in this case.
5 - 43
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5 - 44
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5-10-2-8 Checking the DICOM settings for the Storage Commitment server
(1) Basic settings
IN THE "Preset Setup" - "SystemPreset" - "DICOM" - "MPPS/Commitment", check that the
Storage Commitment server settings are configured correctly.
Up to two Storage Commitment servers can be configured. However, it is possible to send Storage
Commitment data to the specified server only.
Items
AE Title: Specify the AE title of the Storage Commitment server.
Station Name: Specify the computer name of the Storage Commitment server.
IP Address: Specify the IP address of the Storage Commitment server.
Port#: Specify the TCP port number to be used by the Storage Commitment server
for DICOM communication. This port is for receiving data from Storage
Commitment server.
Transaction Limit:
Specify the expiry period for Transaction UID used with Storage
Commitment. When a value is set, this value becomes the expiry period.
When “Unlimited” is set, there is no expiry period.
Note that when the expiry period has been exceeded, a “Transaction UID
expired” message is displayed.
5 - 45
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Retries: For Storage Commitment, configure the number of background retries before
displaying an error message. The operations of the settings are listed below.
0 When a sending retry fails, an error message is displayed
immediately. If “Retry” in the message box is selected, a retry is
sent immediately. When Retries set to “0”, the interval specified
for Retry Interval is not applied.
1~99 No error messages are displayed for retries up to the specified
number, and retries are sent automatically. For example, if
Retries is “2”, no error message is displayed for the first two
failed retries. A message is displayed after the third failed retry.
Note that if “Retry” is selected in the message box, no message
is displayed for the next two retries.
Unlimited No error message is displayed for failed sending retries, and
retries are sent until a retry is successful.
Test function
[Ping] Send a ping command to the server (Remote1 or 2) specified under Connect,
and check that TCP/IP is performed correctly.
Message Examination results
Ping check to **** successful Show that there is a response for echo reply on ICMP
from Storage Commitment server. Generally, this is a
guide that network connection with a server is
currently established correctly.
Ping check to **** failed. Show that there is NO response for echo reply on
ICMP from Storage Commitment server. If the server
has forbidden sending an echo reply, there is no
problem in this case.
5 - 46
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5 - 47
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Private Tag: Specify the attachment method for private tags (Hitachi Aloka tag
information)
None: Do not attach private tags to DICOM files to be output.
Put into File. Attach private tags to DICOM files to be output media
only.
Put into Network.
Attach private tags to DICOM files to be sent to the
image server only.
Put into Both. Attach private tags to DICOM files to be sent to media
and DICOM files to be sent to the image server.
Put into File + DICOMDIR
Attach private tags to DICOM files and DICOMDIR
file to be output to the media
Put into All. Attach private tags to DICOM files and DICOMDIR
file to be output to the media and DICOM files to be
sent to the image server.
Reference It is necessary to configure to add the private tags on the DICOMDIR file,
if you copy the Volume data (RT-3D) as DICOM format and will load it
on the machine.
Attach Thumbnail:
Setting of how to create the thumbnail data.
Off: Create thumbnail data only in the data folder.
On: Create thumbnail data both in the data folder and
DICOMDIR.
5 - 48
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Region Calibration:
Specify whether to include Ultrasound Region Calibration Module
(0018,6011) of each image in the image information.
When “ON”, the information is included. When “OFF”, the
information is not included.
5 - 49
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5 - 50
MN2-2075 Rev.0
Chapter 5 Troubleshooting
During association for typical DICOM communication, the server (SCP: Service Class Provider)
selects the transfer syntax it can receive from the transfer syntax list provided by the machine
(SCU: Service Class User), and then data sending starts using this transfer syntax. However, some
servers cannot receive data even when the machine sends data using the transfer syntax specified
by the server. In addition, data transfer using a particular transfer syntax may be required.
In such cases, it is possible to apply limitations to the transfer syntax proposed by the machine. On
the Ver.1, See the following section, and then change the settings. From Ver.2, it is possible to
apply this limitation from Maintenance screen, see the “Volume 2: Chapter 3, Service
information” for more details.
5 - 51
MN2-2075 Rev.0
Chapter 5 Troubleshooting
As mentioned in the section 5-10, there are two types of RGB data formats: Pixel and Plane.
Some DICOM servers and viewer software use only one of these data formats. As a result, data
structures may be disturbed and images may not be displayed correctly.
The following example show abnormal display due to a Planer Configuration recognition error
when the following image was sent to the DICOM server. When such an image is displayed by a
server or viewer software, see the section 5-10, and then change the Planer Configuration
settings.
[Image sent to the DICOM server]
5 - 52
MN2-2075 Rev.0
Chapter 5 Troubleshooting
As mentioned above, when a Worklist is acquired with default settings, the machine AE title and
"US" as the modality and the current date is sent as the search key as the search conditions from
the machine. However, in some cases no worklist is acquired, and a "No Worklist" message is
displayed because there is no data matching the search conditions with these search keys,
depending on the Worklist server.
In such cases, check the search key necessary to acquire the Worklist by asking the person
responsible for the Worklist server. Next, see the section 0, and then change the search key to be
sent from the machine.
Note that the search key required by the Worklist server cannot be checked using the DICOM
communication logs. You must always ask the person responsible for the Worklist server.
5 - 53
MN2-2075 Rev.0
Chapter 5 Troubleshooting
You can specify whether to request the “Institution Name, (0008,0080)” in the Worklist by
manually overwriting the settings file in the machine.
When the name is not requested (default setting), the Hospital Name of the machine is added to
the image tag. When the name is requested, the institution name acquired from the Worklist is
added to the image tag, and the Hospital Name set on the machine is overwritten by the acquired
value. However, if the acquired value is “Null”, Hospital Name is not overwritten by the value
acquired from the Worklist.
To request the Institution Name in the Worklist, use the following procedure to change the settings
file.
5 - 54
MN2-2075 Rev.0
Chapter 5 Troubleshooting
5. Press Ctrl and [T] key at the same time on the full keyboard, the setting item “Disable Task
Manager” will appear. Then select “Enable”.
5 - 55
MN2-2075 Rev.0
Chapter 5 Troubleshooting
;----------------------------------------------------
[IDWorklist]
AutoAccession = 0 ; *****
InstitutionInfo = 1 ; *****
AppendCaseData = 1 ; *****
PQNullCheck = 0 ; *****
;(0008,0005)SpecificCharacterSet *******
;CharSet = ISO_IR 100
; ISO_IR 100*******
; ISO_IR 6
12. When you don’t need to request “Institution Name” to Worklist server, return the
description that you edited to the following.
InstitutionInfo = 0
5 - 56
MN2-2075 Rev.0
Chapter 5 Troubleshooting
Procedure
1. To access the maintenance switch, remove the rear cover.
2. Set maintenance Dip SW1 to “ ON”
and Dip SW2 to “ MAINTE”.
If this message does not appear, and the operating system starts, after the ultrasound
image is displayed, turn the machine off and on in the usual manner, and then perform
Step 5. again.
7. On the PC USB full-keyboard, press the “Ctrl” and “T” keys simultaneously.
5 - 57
MN2-2075 Rev.0
Chapter 5 Troubleshooting
8. If the key input is recognized correctly, the message “OK” appears. Ultra-POST starts,
and the following window appears.
9. After the software starts, the check starts automatically. The test finishes in about 5
minutes. If there are no particular problems, the message “Test Passed” appears. If there
is a problem with the hardware, the message “Test Failed” appears, and a device with
the problem is highlighted in red.
10. The “Test Failed” test result indicates that there is some kind of hardware error on the
Backend board.
11. After the test has finished, switch off the power switch on the operation panel, and
return the machine to its original state.
5 - 58
Chapter 6
PERFORMANCE CHECK
APPENDIX
MN2-2075 Rev.0
Chapter 6 Performance Check
6-1 Introduction
This performance check indicates the items to check to maintain the quality and safety of the device for
the following conditions.
When repair work has been performed
When modifications have been performed, such as defect countermeasures
When update work has been performed to, for example, improve performance and specifications
Periodic inspections and similar work
Before carrying out this check, refer to page 6-3 and prepare the tools and other items necessary for the
objectives.
6-2 Precautions
Caution When the disk light ( ) of the operation panel of the ultrasound diagnostic
equipment is on, do not switch off the power of the ultrasound diagnostic equipment.
If the power is switched off when the hard disk is being accessed, some or all of the
data written to the disk may be corrupted.
Unless otherwise instructed, carry out this performance check with all covers attached and under the
environment indicated below.
After setting the power switch to ON, wait 30 minutes or more.
If you will use the DICOM network, connect appropriately.
After completing repairs, modifications, and upgrades, check the following items before you carry out
this performance check.
Removed PCBs have all been reconnected normally.
Removed connectors have all been reconnected normally.
No tools, probes of measurement instruments, or other items have been left inside the device.
The earth cables for option devices, network cables, and other items have been reconnected
appropriately.
No screws have been left inside the device.
After completing the check, write the results on the service report and other forms, and then present
them to the user. Furthermore, retain a copy.
6-1
MN2-2075 Rev.0
Chapter 6 Performance Check
Fundamentally, the items that should be carried out as a performance check depend on the content of
work that requires the items. In accordance with the following chart, identify the necessary items (items
in the chart marked by ), and then carry out the check for each item. Some words in the chart are
defined below.
[Power supply unit] .............. Power supply unit, isolation transformer, power supply switch
[Other units] ......................... Power supply, biosignal display unit, and units other than PCBs
Check item
Work that must be checked Remarks
A B C D E F
Part replacement, modification, or No target work for this
upgrade of the power supply unit machine
interior
Replacement of the entire power
supply unit
Part replacement, modification, or Includes PCB
upgrade of the biosignal display unit exchange within unit
interior
Replacement of the entire biosignal
display unit
Replacement of PCBs inside units
other than those mentioned above
Part replacement, modification, or Includes ROM
upgrade on the PCB inside units replacement
other than those mentioned above
Replacement of an entire unit not Includes probes and
mentioned above scanners
Operation check before modification
or upgrade work
Periodic inspections
The content of each check raised in the check items are described starting from the pages listed below.
See these pages for more details.
A. Exterior/mechanism check...................... From Section 6-4-1
B. Cleaning .................................................. From Section 6-4-2
C. Device information/power supply voltage check From Section 6-4-3
D. Function check ........................................ From Section 6-4-4
E. Image quality check ................................ From Section 6-4-5
F. Safety check ............................................. From Section 6-4-6
*Carry out this check by following the Electrical safety test implementation manual (MN2-0204)
If the result of the check indicates that designated operations or standards are not met, the device is
judged to be malfunctioning. See "Volume 1: Chapter 5 Troubleshooting" to deal with the malfunction.
6-2
MN2-2075 Rev.0
Chapter 6 Performance Check
Monitor
The monitor operates within the movable range without any unusual noise.
6-3
MN2-2075 Rev.0
Chapter 6 Performance Check
Monitor arm
・ The 1st axis (bottom), 2nd axis (middle), 3rd axis (top), tilt axis, pivot mechanism, and
arm lock of the monitor arm operate smoothly without any unusual noise.
3rd axis
2nd axis
Arm lock
1st axis
Figure: Monitor arm unit Figure: Tilt axis Figure: Pivot mechanism
Casters
・ All casters can be locked and unlocked smoothly without any unusual noise.
・ When the casters are unlocked, the device moves smoothly without any unusual noise.
・ When the front wheels are locked, the caster locks are applied securely.
・ When the rear-wheel rotation is locked, the casters do not rotate.
・ When the rear-wheel turning is locked, the device does not turn.
6-4
MN2-2075 Rev.0
Chapter 6 Performance Check
6-4-2 Cleaning
Caution Do not sterilize the device with chemical solution or gas. The device may be
damaged.
Caution Do not splash liquid on the device surface or interior. This may cause electric
shock or a short circuit.
1) Disassembly
Switch off the power supply of the device, check the condition of the interior and exterior of
the device, and then clean the device as necessary. When you disassemble the device, refer to
"Volume 1: Chapter 3 Installation/disassembly".
3) Monitor
Soak a soft cloth or gauze in water, wring it well, and then wipe the device gently.
4) Gel-warmer
a) Remove the yellow silicon rubber from the gel-warmer.
6-5
MN2-2075 Rev.0
Chapter 6 Performance Check
5) Dust filters
Pull the left and right storage pockets upwards to remove them.
Use a vacuum cleaner to remove dust and rinse the filters with running water. Remove excess
water and then dry the filters in a well-ventilated location. After confirming the front and back
surfaces of the filters, attach the filters at their original position, and then attach the storage
packets. (Push the filters all the way inside.)
6) Trackball
Rotate the ring counterclockwise, and then remove the trackball from the operation panel.
6-6
MN2-2075 Rev.0
Chapter 6 Performance Check
Remove the trackball from the ring, and then wipe the trackball with a soft cloth.
Clean the support balls (3) and sensors (2) with a cotton swab. If it is difficult to remove dirt,
apply a very small amount of neutral solvent to the cotton swab before cleaning.
8) Probes
The cleaning, disinfection, and sterilization methods differ by probe. Refer to the
documentation included with the probe.
6-7
MN2-2075 Rev.0
Chapter 6 Performance Check
b) Wipe with a soft dry cloth. When there is a lot of dirt, soak a soft cloth in a neutral
solvent diluted with water. Wring the cloth well, and then wipe gently with this cloth
to remove the dirt. Finally, wipe off the cleanser.
When rinsing with water, rinse away the dirt with running water. For example, when
gel has adhered to the probe holder, use a sponge or gauze, and then rinse with water.
Next, wipe away the water with a clean cloth.
c) At position of (1), align and insert the probe holder. Align the position of the hook
(2), and then push until it makes a sound. Next, align the position of the hook (3).
Push until you hear a sound and attach the transvaginal/transrectal probe holder.
6-8
MN2-2075 Rev.0
Chapter 6 Performance Check
b) Wipe with a soft dry cloth. When there is a lot of dirt, soak a soft cloth in a neutral
solvent diluted with water. Wring the cloth well, and then wipe gently with this cloth
to remove the dirt. Finally, wipe off the cleanser.
When rinsing with water, rinse away the dirt with running water. For example, when
gel has adhered to the adapter, use a sponge or gauze, and then rinse with water.
Finally, wipe off the water.
c) Align the projection of the adapter with the hole of the probe holder, and then insert
the projection. Next, attach the back side of the adapter to the probe holder to attach
the adapter.
6-9
MN2-2075 Rev.0
Chapter 6 Performance Check
Expanded view of
dirty location
Cleaning
Confirm the dirty locations, and then clean location
using a cleaning sheet or ethyl alcohol.
Thermal head
b) Cleaning using a cleaning sheet
Use the cleaning sheet included with the
printer accessories for cleaning.
Set the printed side of the cleaning sheet
down, and then align the bottom edge of
the sheet with the depression.
6 - 10
MN2-2075 Rev.0
Chapter 6 Performance Check
Thermal head
6 - 11
MN2-2075 Rev.0
Chapter 6 Performance Check
b) SSZ-D710
Follow the steps below to clean the SSZ-D710.
a) Thermal head position/cleaning location
The thermal head is located near the
middle section inside the printer. You can
access it through the opening on the
bottom surface of the printer.
Thermal head
b) Cleaning by using a cleaning cartridge/sheet (mainly used for dealing with poor
quality printing)
Prepare the cleaning cartridge and
protective sheet.
* The cleaning cartridge is included with
the accessories of the printer unit.
Cleaning cartridge
* One protective sheet is included with the print paper of the
color print pack.
This sheet protects from soiling the
print surface of paper when inserting
paper in the paper supply tray. The
protective sheet is removed after print
paper has been inserted in the paper
supply tray. However, do not discard
this sheet because it is used for
cleaning. Store it along with the
cleaning cartridge. Protective sheet
Install the protective sheet (cleaning sheet).
With the back side (side with no printing)
6 - 12
MN2-2075 Rev.0
Chapter 6 Performance Check
c) Cleaning with ethyl alcohol (for dealing with poor printing and poor paper
supply) Securing
On the bottom surface of the printer,
remove the bottom panel and service Bottom
cover. l
* Remove the paper supply tray and
ribbon cartridge, and then turn the
printer over.
* Loosen the 2 screws securing the
bottom panel, and then remove the
bottom panel.
6 - 13
MN2-2075 Rev.0
Chapter 6 Performance Check
Platen roller
* There is a thermal head and platen
roller inside.
Thermal head
6 - 14
MN2-2075 Rev.0
Chapter 6 Performance Check
* As shown in the figure below, the paper ejection rollers and paper supply rollers
can be seen. Apply alcohol to a cotton swab and then clean the rollers.
Paper ejection rollers
Cleaning cartridge
* The cleaning sheet is packaged with the
printer accessories shipped at the end of
2012 or later.
This plate is used to clean paper supply
roller dirt and adhered matter caused by
usage conditions and prevent paper
supply errors and poor quality printing.
Cleaning plate
* Printers shipped at the end of 2013 and
later have the sticker shown in figure to
the right and are equipped with an
automatic cleaning function. Therefore,
a cleaning plate is not included in the
packaging. Printers equipped with this
function do not need cleaning using the
cleaning plate.
Identifying the presence of the
equipped automatic cleaning function
Remove the paper supply tray.
6 - 15
MN2-2075 Rev.0
Chapter 6 Performance Check
If print paper remains in the paper supply tray insertion section, remove the print
paper.
Insert the cleaning plate into the paper supply tray in the direction shown in the
figure.
If the separation plate of the paper supply
tray is standing, lower it.
With the label side of the cleaning plate
facing up, set the black pad on the inner
side of the tray and insert the cleaning
plate.
Insert the paper supply tray into the printer until you hear a click.
[CLEAN EXEC]
Press the "EXEC" button.
Cleaning of the interior rollers starts. (The
cleaning operation takes about 30
seconds.)
During cleaning, the following message
appears on the LCD.
[CLEANING]
[WAIT]
6 - 16
MN2-2075 Rev.0
Chapter 6 Performance Check
6 - 17
MN2-2075 Rev.0
Chapter 6 Performance Check
4) B mode
a) Basic image quality, convex
When probe C251 is connected, and the probe is touched to the body under the
following conditions, B mode of abdominal tissue is displayed. Organ outlines, vascular
walls, and other items are displayed with good connections, and there is no brightness
unevenness, noise, or unnecessary writing in the image.
<Conditions>
- Connector: PROBE 1
- Application Preset: Adult Abdomen
- B Gain: Arbitrary
- TGC: All center (no setting outside center for TGC memory)
- Dynamic Range (B): Near 70 dB (acceptable at preset default)
d) Acoustic Power
If the Acoustic Power rotary encoder is rotated to the left, the Acoustic Power automatic
display value on the main display drops, and the initial ultrasound weakens.
If the Acoustic Power rotary encoder is rotated to the right, the Acoustic Power
automatic display value on the main display rises, and the initial ultrasound strengthens.
5) Color Flow mode
a) Basic image quality: color flow
When probe C251 is connected, and the probe is touched to the body under the
following conditions, Color Flow mode of the abdominal tissue is displayed. Blood flow
is displayed with good connections, and there is no brightness unevenness, noise, or
unnecessary writing in the image.
The B mode image does not change by switching Color Flow mode ON and OFF.
<Conditions>
- Connector: PROBE 1
- Application Preset: Adult Abdomen
- B Gain: Arbitrary
6 - 18
MN2-2075 Rev.1
Chapter 6 Performance Check
- TGC: All center (no setting outside center for TGC memory)
- Dynamic Range (B): Near 70 dB (acceptable at preset default)
- CF Gain: Arbitrary
- Velocity Range (Color): Arbitrary
- Flow ROI size and location: Arbitrary
- Color Map: Preset Default
6) PW Doppler mode
a) Basic image quality: PW
When probe S12 is connected, and the probe is touched to the body under the following
conditions, PW Doppler mode of the heart is displayed. PW Doppler outlines are clear
and displayed with good connections, and there is no brightness unevenness, noise, or
unnecessary writing in the image.
<Conditions>
- Connector: PROBE 2
- Application Preset: Adult Heart
- B Gain: Arbitrary
- TGC: All center (no setting outside center for TGC memory)
- Dynamic Range (B): Near 70 dB (acceptable at preset default)
- Sample volume size and location: Arbitrary
- Baseline Shift (D) location: Arbitrary
- Velocity Range (D): Arbitrary
- D Gain: Arbitrary
b) Acoustic Power: PW
・ If the Acoustic Power rotary encoder is rotated to the left, the Acoustic Power
automatic display value on the main display drops, and the sensitivity of the PW
Doppler waveform weakens.
・ If the Acoustic Power rotary encoder is rotated to the right, the Acoustic Power
automatic display value on the main display rises, and the sensitivity of the PW Doppler
waveform strengthens.
6 - 19
MN2-2075 Rev.1
Chapter 6 Performance Check
・ The ECG Sense setting is controllable from 0 through 30, the baseline of the ECG wave
display is fixed, and the amplitude only can be changed.
6 - 20
MN2-2075 Rev.0
Chapter 6 Performance Check
b) Acoustic Power: CW
・ If the Acoustic Power rotary encoder is rotated to the left, the Acoustic Power
automatic display value of the main display changes from H to M to L, and the
sensitivity of the CW Doppler waveform weakens.
・ If the Acoustic Power rotary encoder is rotated to the right, the Acoustic Power
automatic display value of the main display changes from L to M to H, and the
sensitivity of the CW Doppler waveform strengthens.
11) IND probe connection unit (when EU-9163 and EU-9166 are installed)
a) When probe UST-2265-2 is connected, and the probe is touched to the body under the
following conditions, CW Doppler mode of the heart is displayed. CW Doppler outlines
are clear and displayed with good connections, and there is no brightness unevenness,
noise, or unnecessary writing in the image.
<Conditions>
- Connector: PROBE 5
- Application Preset: Adult Heart
- B Gain: Arbitrary
- TGC: All center (no setting outside center for TGC memory)
- Dynamic Range (B): Near 70 dB (acceptable at preset default)
- Sample volume size and location: Arbitrary
- Baseline Shift (D) location: Arbitrary
- Velocity Range (D): Arbitrary
- D Gain: Arbitrary
b) Acoustic Power: CW
・ If the Acoustic Power rotary encoder is rotated to the left, the Acoustic Power
automatic display value of the main display changes from H to M to L, and the
sensitivity of the CW Doppler waveform weakens.
・ If the Acoustic Power rotary encoder is rotated to the right, the Acoustic Power
automatic display value of the main display changes from L to M to H, and the
sensitivity of the CW Doppler waveform strengthens.
12) DVI-D in/out
When there is a peripheral device connected via DVI-D, the connected device shows the
same image as the main display.
13) Y/C in/out and Audio in/out (when EU-9167* is installed)
To check the Y/C input and output, connect the SSZ-709 or other analog color printer using a
6 - 21
MN2-2075 Rev.0
Chapter 6 Performance Check
Y/C cable. When setting Freeze at the maximum CF Gain in B Color Flow mode, the printout
is the same as the monitor display.
When there is a peripheral device connected using the Audio input/output, audio is recorded
to the connected device. Furthermore, when audio is played back on the connected equipment,
the audio is played back on the device.
14) Composite out (when EU-9167* is installed)
To check the composite signal output path, connect the SSZ-709 or other analog color printer
using a BNC cable. When setting Freeze at the maximum CF Gain in B Color Flow mode, the
printout is the same as the monitor display.
6 - 22
MN2-2075 Rev.0
Chapter 6 Performance Check
Coin check
Conditions: B mode, PRESET: arbitrary
Carry out an inspection by switching the focus between the shallowest section and
deepest section. Apply a thin coating of gel on the oscillator surface of the probe. Touch a
fine metal rod at a right angle to the longitudinal direction of the probe. If the rod is
moved slowly on the surface top, the ultrasound scan appears and moves as shown below.
Perform once at the shallowest focus section and once at the deepest focus section.
Make sure there are no ultrasound scan omissions or displays in 2 or more locations.
Check using all connected probes and all probe connectors.
6 - 23
MN2-2075 Rev.0
Chapter 6 Performance Check
Probe
Ultrasound
phantom
RMI model 403GS
Tomographical image of
measurement-point display section
(using probe C251)
6 - 24
MN2-2075 Rev.0
Chapter 6 Performance Check
Resolution
At the following settings, confirm that the distance resolution is 2.0 mm or less and the
direction resolution is 3.0 mm or less.
Distance resolution
Setting conditions:
PRESET: General
Mode: B-mode
THI: OFF
Frequency: FND-5.0R
Line Density: 2
TGC: All-stage center position (TGC memory not
used)
Acoustic Power: 10%
B Gain: 60
Monitor the resolution target of the phantom with a depth range of 4.0 cm.
Set the probe so that the B mode section is at a right angle to the target.
Direction resolution
Setting conditions:
PRESET: General
Mode: B-mode
THI: OFF
Frequency: FND-5.0R
Line Density: 2
TGC: All-stage center location
Acoustic Power: 10%
B Gain: 60
Monitor the resolution target of the phantom with a depth range of 4.0 cm.
Set the probe so that the B mode section is at a right angle to the target.
Probe
Ultrasound phantom
RMI model 403GS
6 - 25
MN2-2075 Rev.0
Chapter 6 Performance Check
Important The safety test identifies the quality of the electrical safety performance of the
device. The test methods and standard values of the safety test differ depending on
the category of degree of protection against electric shock.
In the case of the ARIETTA70 ultrasound image diagnostic equipment, the
category of degree of protection against electric shock corresponds to "BF-type
applied part".
Note that when the safety check is carried out in Japan, it is based on JIS T 0601-1.
1) Checking the conduction of the protective earth wire
Measure between the protective earth terminal edge of the power supply plug of the
ARIETTA70 ultrasound image diagnostic equipment and the following locations.
Power supply
cable
6 - 26
MN2-2075 Rev.0
Chapter 6 Performance Check
3) Leakage current
Patient measurement current (carried out only when the PEU-ARIETTA70 is connected)
Measure between the black lead and red lead/green lead bundle of the ECG cable.
Black
To leakage current meter (red Test probe)
Red
Green
6 - 27
MN2-2075 Rev.0
Chapter 6 Performance Check
6 - 28
MN2-2075 Rev.0
Chapter 6 Performance Check
6 - 29
MN2-2075 Rev.0
Chapter 6 Performance Check
6 - 30
Revision History of the Service Manual
APPENDIX
MN2-2075 Rev.1
ARIETTA 70 Service Manual VOLUME 1 Revision History
1 Introduction
Revisions
Rev. Date Note
Contents Pages
0 08-Dec.- 2014 1st issue
1 29-Oct.- 2015 Revision updated on Cover replaced Volume 1 Cover
Contents in Volume 1 replaced 1/6~6/6
Chapter 1 Introduction
Added the information to section “1-1 Service
Manual” replaced 1-1
Chapter 3 Install/Disassemble
Added the information for the EU-9158 replaced 3-53
Chapter 4 System Overview
4-1 System Specifications, 4-2 System
Configuration, 4-3 System Block Diagram
Updated due to additional specification on Ver.2
replaced 4-1~4-22
Chapter 5 Troubleshooting
Added the information to section “5-4 Location of
the units and modules” replaced 5-4
Added the information to section “5-11 CELL”
replaced 5-11
Chapter 6 Performance Check
Added the information to section “7) DMA
transfer” and corrected an error replaced 6-19~6-20
Revision history of Volume 1 replaced 1/2~2/2
Revision and addrees updated on rear cover
replaced Volume 1 Rear cover
1/2
MN2-2075 Rev.1
ARIETTA 70 Service Manual VOLUME 1 Revision History
(Blank page)
2/2
Service Manual Volume 1
www.hitachi-aloka.com
© Hitachi Aloka Medical, Ltd. 2014-2015. All rights reserved.
MN2-2075 Rev.1