05_Clean_Install_Procedure

You might also like

Download as pdf or txt
Download as pdf or txt
You are on page 1of 46

HEALTHCARE Chapter 5

Imaging Services Software Options


Document No: DD+DIS050.08E

NX 2008
Type 4406/303

Clean Install Procedure for NX 2008

► Purpose of this document


This document describes the installation of an NX system starting from scratch.

REQUIRED TIME:
Before starting the Clean Install Procedure of NX 2008, please take into account that
the Clean Install Procedure takes approx. 3 h including performing and filling out the
Acceptance Test Protocol during installation/configuration.

► Document History

Edition. Release Changes


Revision Date compared to previous version 1.1
1.2 09-2008 • Implemented RAID (Redundant Array of Independent
Disks). Details see mainly section 8.
• Implemented HP PC. Details see section 3.1 and 3.2.

► Referenced Documents
Depending on the system, following documentation is needed on site:

Document Title
DD+DIS043.08E NX 2008 Service Manual
DD+DIS227.04E Mammography Solution Manual
QS 3.0: DD+DIS273.04E QS 3.X Service Manuals
QS 3.5: DD+DIS302.05E

DD+DIS012.06E Licensing Service Documentation


DD+DIS032.06E DX-Si System Manual

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2
09-2008 printed in Germany Document Node ID: 21273806-42
sw_05_software-options_e_template_v04
Agfa Company Confidential Copyright © 2008 Agfa HealthCare N.V.
DD+DIS050.08E Software Options

► Manufacturer
Agfa HealthCare N.V.

Publisher
Agfa-Gevaert HealthCare GmbH
Tegernseer Landstraße 161
D - 81539 München
Germany
Copyright © 2008 Agfa HealthCare N.V.
All rights reserved.
Technical modifications reserved.
AGFA and the Agfa-Rhombus are trademarks of Agfa HealthCare N.V.

WARNING:
Improper operation or service activities may cause damage or injuries.

INSTRUCTION:
1. Read the "Generic Safety Directions" document
(see MEDNET GSO => General Info => Agfa HealthCare => Publications =>
Service Manual) prior to attempting any operation, repair or maintenance task on
the equipment.
2. Strictly observe all safety directions within the "Generic Safety Directions" and on
the product.

NOTE:
To verify the latest version of single documents and of Service Manuals refer to the
Document Type ‘Order List’ in the GSO library.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 2


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

LIST OF CONTENTS

1 FLOWCHART OF CLEAN INSTALL PROCEDURE ................................................................5


2 BEFORE SITE VISIT ................................................................................................................6
2.1 How to obtain the Clean Install Kit ............................................................................................6

2.2 Obtaining additionally needed Equipment ................................................................................7

2.2.1 Tools .........................................................................................................................................7

2.2.2 Documentation ..........................................................................................................................8

2.2.3 DX-S or CR 30-X Digitizer Device Software .............................................................................8

2.2.4 X-ray Device Integration (XRDI) Software ................................................................................8

2.3 Obtaining Application License File (ALF) (if needed) ...............................................................9

3 CLEAN INSTALL PROCEDURE ..............................................................................................9


3.1 Prerequisites .............................................................................................................................9

3.2 Installing the Operating System (OS) .....................................................................................11

3.3 Performing additional Partitioning ...........................................................................................15

3.4 Configuring Basic XP Settings ................................................................................................19

3.5 Installing additional Languages...............................................................................................22

3.6 Installing the latest available Hotfixes.....................................................................................24

3.7 Configuring Network ...............................................................................................................24

3.7.1 Changing TCP/IP Settings ......................................................................................................25

3.7.2 Changing the Hostname and/or workgroup/domain of the Operating System .......................26

3.8 Installing Application and Documentation...............................................................................28

3.8.1 Installing Application ...............................................................................................................29

3.8.2 Installing NX Software Update (SU) (if available) ...................................................................32

3.8.3 Installing NX Online Help ........................................................................................................33

3.9 Running Hotfix installation again.............................................................................................34

3.10 Creating Fingerprint File (only in case of CR QS upgrade) ....................................................34

3.11 Loading Licenses (ALF) ..........................................................................................................34

3.12 Installing XRDI Software (if applicable) ..................................................................................36

3.13 Installing Digitizer Software.....................................................................................................36

3.14 Installing an Uninterruptible Power Supply (UPS) ..................................................................37

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 3


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

4 ACTIVATION ..........................................................................................................................38
5 UPGRADING ID TABLET FIRMWARE ..................................................................................40
6 CONFIGURATION..................................................................................................................40
6.1 Manual Configuration (if needed)............................................................................................41

6.1.1 GenRad...................................................................................................................................41

6.1.2 Mammography ........................................................................................................................41

7 BACKUP .................................................................................................................................42
8 ACTIVATING RAID (REDUNDANT ARRAY OF INDEPENDENT DISKS).............................43
9 FINALIZING NX ACCEPTANCE TEST ..................................................................................46

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 4


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

1 Flowchart of Clean Install Procedure

NOTE:
The outlined timing can significantly vary depending on the complexity of the given
situation on site. On a RAID system, two hours extra are required to perform a
clean install.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 5


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

2 Before Site Visit

2.1 How to obtain the Clean Install Kit

NOTE:
From NX 2.0 staged on DELL Optiplex 755 on, the Clean Install Material is delivered
with the system.

• The Clean Install material can be ordered via CM+number. Refer to to the NX
2008 Service Plan on MedNet GSO library or the DRS assortment list that is
distributed to all marketing and service managers for the correct order number.
• The NX 2008 Clean Install material can also be downloaded from the
MedNet FTP-Server:
Path: Computed Radiography Æ CR NX 2008 Æ Software Æ
NX 2008 SW for Clean Install
Link: (http://ftp.agfa.be/HE/software)
Only applicable for Agfa employees:
Users who need to access the Global FTP Server via the http interface will
have to sign on using the Username and Password listed in:
MedNet Bulletin PACS 2004-11-23 Global FTP Server Password.
For non Agfa employees:
Order via Spare Part Supply Chain or contact your local Agfa Partner.

Content of Clean Install Material:


• NX Starterkit DVD 1 – containing following folders:
o Images Intro
o Navigation files
o Online Help (contains the installer for the NX Online Help)
o PDF (contains User Documentation in different languages)
o Service Documentation (contains the documentation for
following scenarios):
ƒ Installation Documentation Clean Install
ƒ Installation Documentation Kit GENRAD, MAMMOGRAPHY or CMS
ƒ Installation Documentation NX Upgrade Kit
ƒ Installation Documentation QS Upgrade Kit
ƒ Installation Documentation Diagnostic monitors

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 6


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

o Service Software, contains:


ƒ Dose export tool (for different versions)
ƒ ID tablet firmware
ƒ NX 1.0 SP1 (for exporting images on NX 1.0 workstations)
ƒ NX 2008 Generic Installer
ƒ Automatic Updates (FUSE) Agent
ƒ NX Service Updates (if applicable)

• NX Starterkit DVD 2 – containing following folders:


o DRIVERS, contains:
ƒ Niowatch
ƒ Barco drivers
ƒ Lansafe driver
ƒ RAID Tool
o HOTFIX (contains the latest Hotfix)
o LANGUAGES (contains files for installing additional languages)
o Prepart
o The installation files for the XP Operating System (autostart)

NOTE
Chapter 01 NX 2008 SW + HW Compatibility Chart (DD+DIS043.08E) specifies the
software versions that are compatible with a specific NX profile
(Intranet Link / Extranet Link).

2.2 Obtaining additionally needed Equipment

Additionally, take following equipment with you on site.

2.2.1 Tools

• A standard serial cable to upgrade the firmware of the ID Tablet.


If the ID Tablet is connected via USB, make sure you take a serial cable on site.
• A portable medium (USB Stick or CD-R/RW).
• Default configuration file of the NX workstation of your region or country
(to be setup by the country during the initial NX workstation installations).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 7


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

2.2.2 Documentation

• Technical Service Manual for the respective CR QS workstation


(QS 3.0: DD+DIS273.04E; QS 3.5: DD+DIS302.05E)
• NX 2008 Service Manual (DD+DIS043.08E)
• For Mammography Systems: Mammography Solution Manual (DD+DIS227.04E)
• The latest Hotfix version on release date of NX 2008 is contained in the NX
Starter Kit DVD 2. The installation procedure can be found in the NX 2.0 Service
Bulletin No. 16 (DD+DIS395.07E) on MedNet GSO library
(Intranet Link / Extranet Link). Check the NX 2008 Service Bulletins for a possibly
more recent Hotfix release.

2.2.3 DX-S or CR 30-X Digitizer Device Software

If the NX Workstation is connected to a DX-S or CR 30-X digitizer, the compatible


device software has to be downloaded from Mednet GSO library and installed on the
workstation.
(1) Refer to chapter 1 “HW/SW compatibility chart” (DD+DIS043.08E) to find the
compatible version number(s).
(2) Go to the Mednet GSO Library (Intranet Link / Extranet Link) to download one of
the following device software packages:
• GSO Library Æ Computed Radiography Æ CR Digitizers Æ DX-S Æ Freeware
• GSO Library Æ Computed Radiography Æ CR Digitizers Æ CR 30-X Æ
Freeware

2.2.4 X-ray Device Integration (XRDI) Software

Information on which version of XRDI software is released and how it can be obtained:
• In case of a Mammography system, refer to the Mammography Solution Manual
(DD+DIS227.04E).
• In case of a DX-Si system to the DX-Si System Documentation (DD+DIS032.06E).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 8


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

2.3 Obtaining Application License File (ALF) (if needed)

IMPORTANT:
When performing a clean installation, the ALF should normally still be available and
can be used for the same software version after the clean install.
This is not the case if:
• The ALF got lost after a system crash:
o Fetch it from the Electronic License Management System (ELMS).

• You are upgrading a CR QS system:


You have two possibilities:
o Create it with the Simplified ELMS. In this case the Fingerprint File (FPF) has to
be created during the Clean Install (see section 3.10).
o Request it from the NSO.
Refer to the “Installation Instruction for Upgrade of QS to NX 2008”
(DD+DIS055.08E) for the information to be gathered to obtain the ALF.

INSTRUCTION:
For a detailed description of the procedures to obtain a License File (ALF) refer to the
Licensing Service Manual (DD+DIS012.06E) on MedNet GSO Library
(Intranet Link / Extranet Link).

3 Clean Install Procedure

3.1 Prerequisites

Check following items before you start the installation:

• Check hardware to see if it is supported for NX OS XP.


Install the NX only on machines that are on the supported HW list.
The list of supported hardware is specified in NX 2008 Service Manual, chapter 1
HW/SW Compatibility Chart (DD+DIS043.08E).

• Use the correct set of the Installation DVDs:

NX Starterkit DVD 1
NX Starterkit DVD 2

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 9


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

• Windows License Key:

For all installations you need to provide the Windows license key for the
operating system. This is often a ‘hologrammed’ sticker attached to the system
itself by the manufacturer. The key itself is referred to as the ‘Product ID’ and
has the form 5 characters in 5 blocks, for example:

HVYF3-T28PW-HWGKB-33J4W-B4VPM

• Network connected:

When installing, the NX PC must have a network connection.


The minimum is a connection to a hub so the system can ‘see’ a network
connection and another PC on the network.

The system can be connected to a larger network when installing, like the
hospital or company network, but be careful. If other NX systems are installing
at the same time, an IP conflict can occur.

• Change the keyboard to a QWERTY (US international) keyboard:

When you log on as administrator you will always have the English version of
Windows XP. Therefore it is easier working with a QWERTY keyboard
(US international) attached when you have to type something.

To configure the keyboard on the NX:


(1) Go to Start Æ Settings Æ Control Panel Æ Regional Settings.
(2) Go to the tab Languages and click <Details>.
(3) Select English (US) as Default Input Language.

• BIOS Settings for DELL PCs:

(1) Enter the BIOS setup by pressing <F2> during booting.


(2) Make sure that the boot sequence in the BIOS features the hard drive (HD)
before the CD/DVD drive.
(3) Set ATA/SATA mode:
o For machines that use serial ATA drives you can find a BIOS option to
set the ATA mode. Here you can choose normal or combined. Make
sure to set this setting to ‘combined’.
o For Dell 745: SATA operation to be set to ‘Legacy’.
o For Dell 755: SATA operation to be set to 'RAID/Autodetect/ATA'.
(4) Activate SMART Reporting: Set <Drives\SMART Reporting> to <ON>.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 10


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(5) Deactivate Hyper-threading (not applicable for multiple core CPUs): Select
<Performance\Hyper-Threading> and set to <Off>.
(6) Activate Multiple CPU core (only applicable for multiple core CPUs (e.g.
Dell D830 and Dell Optiplex 755)): Set <Multiple CPU core> to <ON>.
(7) Exit with <ESC> and save the changes.

• BIOS Settings for HP PCs:

(1) Enter the BIOS setup by pressing <F10> during booting.


(2) Make sure that the boot sequence in the BIOS features the hard drive (HD)
before the CD/DVD drive.
(3) Set Storage Options: SATA emulation to be set to 'IDE'.
(4) Go to File Æ Save Changes & Exit, press <Enter> and accept with
<F10>.

IMPORTANT:
For both, Installation and Configuration, have your Acceptance Test Protocol document
at hand and fill it in on-the-fly at every installation and configuration step instead of
filling it in afterwards. This saves a lot of time.
The document is available on MedNet GSO library (Intranet Link / Extranet Link).

3.2 Installing the Operating System (OS)

IMPORTANT:
Installing the OS leads to a removal of all available data on the complete hard disk,
including all existing partitions!

(1) Insert the NX Starterkit DVD 2.


(2) Reboot the PC and
• on a DELL PC: press <F12> during the boot sequence (custom boot).
• on a HP PC: press <F9> during the boot sequence (custom boot).
(3) Select to boot from CD-ROM and press <Enter>.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 11


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(4) Confirm the installation by pressing <Y>.


+-------------------------------------------------------------+
! AGFA NX System !
! Multi Language !
+-------------------------------------------------------------+
! !
! New Installation. !
! !
! (Y) Installation !
! (N) Exit !
+-------------------------------------------------------------+

figure 1

(5) Confirm the installation once more by pressing <Y>. Be aware that once you
enter <Y> the (re-)installation is starting, and all data will be erased.
+-------------------------------------------------------------+
! AGFA NX on XP !
! Multi Language !
+-------------------------------------------------------------+
! !
! Installation will erase all current systems !
! !
! (Y) This machine will be totally reinstalled !
! (N) Exit to Operating system !
+-------------------------------------------------------------+

figure 2

(6) After approx. 5 minutes, the RAID BIOS setting instruction is displayed.
Remove the DVD from the drive and press <CTRL+ALT+DEL> to reboot.

figure 3

IMPORTANT:
Perform the next two steps only on a DELL 755 system. Also perform them
if you are installing a DELL 755 system without RAID!
(7) Enter the BIOS setup by pressing <F2> during booting.
(8) Set ATA/SATA mode to 'RAID ON'.

IMPORTANT:
Perform the next two steps only on a HP system. Also perform them if you
are installing an HP system without RAID!
(9) Enter the BIOS setup by pressing <F10> during booting.
(10) Set Storage Options/SATA emulation to 'RAID'.

(11) From now on the installation is unattended. Your presence is no longer required
until you are prompted to log in.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 12


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

REQUIRED TIME OF UNATTENDED INSTALLATION:


Approx. 60 min.

IMPORTANT:
Don’t click anything until you get the log on screen described below in item 7. It could
interrupt the installation.
The unattended installation consists of the Norton Ghost Restore Procedure, the
Windows XP Setup and post-installation customization.
• Norton Ghost Restore Procedure:
A progress bar indicates the progress of the restore procedure. The PC will
automatically reboot afterwards.
• Windows XP Setup:
During the installation you’ll see a lot of screens of the Windows XP Setup.
Don’t worry, the installation runs in the background.
• Post-installation customization:
Automatic scripts run once the Windows XP Setup process has completed. Wait
until you see the log on screen. Do not interrupt these scripts.

(5) After the unattended installation and the final reboot the log on screen is
displayed. Press <Ctrl-Alt-delete> and log in using Administrator account:

figure 4

(6) Enter following information:


• The Licensed Organization AGFA and click <OK>.

figure 5

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 13


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

• The Licensed Owner AGFA and click <OK>.

figure 6

• The Product ID. This can be found on the Windows XP Prof.


License sticker. This is delivered in a separate envelope. This sticker is to
be attached to the upgraded workstation next to the Windows 2003 Server
License sticker (if not yet done).

figure 7

(7) Click <OK> on following message:

figure 8

(8) Remove the NX Starterkit DVD 2 from the disk drive.

Result Here ends the Operating System installation. It is now necessary to carry out
some system configuration.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 14


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

3.3 Performing additional Partitioning

REQUIRED TIME:
Approx. 5 min.

The operating system has now been installed on the System & Configuration partition
(drive letter C:).
For NX we need to define 1 more partition (Data & Backup partition) with
following parameters:
• If possible create the partition on a second hard drive.
• Drive letter: D:
• Volume Label: DATA
• Size: 160 GB

NOTE:
The screenshots below are made on a two hard disk system. If the platform contains
only one harddisk, the previous OS installation step has allocated 25 Gbyte of the first
hard disk found for the C: partition (applicable from XOS 1.0.18 on). The D: partition
has be made manually!

IMPORTANT:
RAID: If the system contains a second harddisk for RAID mirroring, do NOT create
partitions on the second disk.

To create a new partition, perform the following steps:

(1) Right-click ”My Computer” and select “Manage” from the menu:

figure 9

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 15


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(2) In the Computer Management window, select “Disk Management”:

figure 10 – Example – The disc sizes may differ on your system.

NOTE:
In case of reuse of CR QS hardware, there is still the old CR QS D: and E:
partition. Delete both partitions.

(3) Right-click an unallocated disk partition.


(4) Select ”New Partition” from the menu.
(5) Click <Next> in the welcome window:

figure 11

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 16


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(6) Select “Primary Partition” and click <Next>:

figure 12

(7) If the size of the harddisk allows it, set the partition size to 160000. Otherwise
select less. It is not advised to choose a value higher than 160000.

NOTE:
• There might be more disk space available. Anyway, the partition D: should be
limited to 160 GB and partition C: to 25 GB. Those are the capacities that the
software has been tested and verified with.
• Excessive disk space can be used to create an extra partition for
backup purposes.
• When using a laptop platform (e.g. Dell Latitude D830), it may have insufficient
disk space to allocate 160 GB for the second partition. Allocate the full remaining
size to the second partition.

(8) After entering the parition size click <Next>:

figure 13

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 17


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(9) Choose the drive letter D: and click <Next>.

figure 14

(10) Enter following settings:


• Choose the partition formatting.
• Fill in the appropriate Volume Label DATA.
• check “Perform a quick format”, uncheck “Enable file and folder
compression”:

figure 15

(11) Click <Next>.


(12) Click <Finish> in the summary window to create the partition:

figure 16

Result You will now see your new partition in the computer management window
and the remaining unpartitioned space.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 18


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(13) Close window Computer Management.

NOTE:
For any additional partition you wish to create, perform these steps again.

3.4 Configuring Basic XP Settings

REQUIRED TIME:
Approx. 5 min.

IMPORTANT:
Whenever you change language or unicode settings it is advised to use the
Administrator account.
Reason: The service language should always be English. In case service is performed
remotely, the service engineer might not be able to understand the local language.

(1) Log in using the Administrator account (don't use “CRService” account here).
(2) Go to “Start Æ Settings Æ Control Panel”. If the Control Panel is in “Category
View”, click “Switch to classic view”.
(3) Double-click Date & Time and adjust the date and time properties.

figure 17

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 19


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(4) Open the tab Time Zone.

figure 18

(5) Select the correct time zone and check the checkbox “Automatically adjust for
daylight saving changes”.
(6) Click <OK>.
(7) Go back to the Control Panel and choose Regional & Language Settings.
(8) Set the appropriate setting in the tabs Regional Options and Languages.
(9) Go to the tab Advanced and select your language to match the language version
of non-unicode programs:

figure 19

(10) Check the box Apply all settings to the current user account and to the default user
profile.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 20


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

NOTE:
This has only effect when users are not in a domain.

For instructions to
• Change the language of single users,
• Change to non-standard string formats (e.g. Date&Time, figure formats),
• Configure the default language for new NX users,
refer to chapter 04 Appendices “Changing the regional settings of the NX”.

(11) Go to the tab Languages and click <Details...>.


(12) If the correct keyboard is not listed under "Installed services", click <Add...>,
select the keyboard layout and click <OK>.
(13) Select in the "Default Input Language" dropdown list, the correct input language.
(14) Click <OK>.
(15) A dialog may be displayed to ask if files should be copied from Windows CR-ROM,
or if the files already installed on the hard disk can be used.
Confirm that the files already installed on the hard disk can be used.
(16) Reboot the system of you are asked to.

NOTE:
The following steps are only applicable to PCs with SCSI disk, not for SATA disks:
(17) On the desktop, click right on My Computer and select Properties.
(18) Go to tab Hardware and click Device Manager.
(19) Select the disk drive, click right and select Properties.
(20) In the tab 'Policies', enable write caching on the disk.

figure 20

(21) Click <OK>.


(22) Repeat the steps from step (17) if the system contains more than one SCSI disk.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 21


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

3.5 Installing additional Languages

REQUIRED TIME:
Approx. 10 min. per language

Purpose This step installs additional languages for menus and dialogs of NX.

(1) Insert the appropriate NX Starterkit DVD 2 and from


the folder.
(2) Go to the folder Languages.
(3) Double-click “muisetup.exe”.
(4) Select “Accept the License Agreement”.
(5) Click <Continue>.
(6) Enter following settings:
• Select the language(s) you want to install.
• Check the checkbox “Match the language for non-Unicode programs with
the default user language”.

figure 21

(7) Click <OK> to start the installation.


(The installation of the languages takes up to 5 minutes per selected language).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 22


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

NOTE:
After you clicked <OK>, the system might ask for a “cplexe.exe”.
• Do not click <Cancel> (the installation would be cancelled).
• Workaround:
Point manually to the “C:\i386\lang” folder, so that the installation can be continued.

(8) Click <OK> after you got a confirmation that the installation is completed.
(9) If following message appears click <Cancel>:

figure 22

(10) Another dialog box appears, asking for a confirmation, concerning the new file
versions. Click <Yes>:

figure 23

(11) Click <OK> to finish the installation of the languages that are on this CD.

figure 24

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 23


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

NOTE:
• Keep the language for the “CRService” account to English
Install scripts depend on English as default language.
• The Administrator can define another language during the setup of the account.
Changing the default language for new users will set the logon screen in this
language.

3.6 Installing the latest available Hotfixes

REQUIRED TIME:
Approx. 50 min.

The latest Hotfix version on release date of NX 2008 is contained in the NX Starter Kit
DVD 2. The installation procedure can be found in the NX 2.0 Service Bulletin No. 16
(DD+DIS395.07E) on MedNet GSO library (Intranet Link / Extranet Link). Check the NX
2008 Service Bulletins for a possibly more recent Hotfix release.
NOTE:
Some of the Hotfix patches are skipped during installation since the NX application is
not installed yet. This is why the Hotfix installation will be repeated after installing the
application (section 3.9).

3.7 Configuring Network

REQUIRED TIME:
Approx. 10 min.

NOTE:
Follwing actions need to be performed to configure the XP network settings.
If you read this document on the computer, click the links to read step by step
instructions. If you read a print out, continue with next page:

Checkpoints: See section


1. Changing TCP/IP Settings 3.7.1

Changing the Hostname and/or workgroup/domain of the 3.7.2


2.
Operating System

Back to Flowchart

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 24


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

3.7.1 Changing TCP/IP Settings

(1) Log in using “CRService” account.


(2) Go to “Start Æ Settings Æ Control Panel. If Control Panel is in Category View,
switch to Classic View and select “Network Connections”.

figure 25

(3) Right-click “Local Area Connection” and go to Properties. The window Local
Area Connection Properties opens.
(4) Select “Internet Protocol (TCP/IP)”:

figure 26

(5) Click <Properties>.


The window “Internet Protocol (TCP/IP) Properties” opens.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 25


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(6) Change the values according to the customers needs:

figure 27

(7) Click <OK>. (“Internet Protocol (TCP/IP) Properties”).


(8) Click <OK>. (“Local Area Connection Properties”).

3.7.2 Changing the Hostname and/or workgroup/domain of the Operating System

NOTE:
Gather information about network environment in the hospital (Domain or Workgroup)
or refer to the information you have gathered in chapter 3 “Installation Planning”
(DD+DIS045.08E).

To change the hostname and or the workgroup of the NX PC, perform following steps:

(1) Right-click the “My Computer” icon on the desktop and select “Properties”.
(2) Select the tab “Computer Name”:

figure 28

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 26


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(3) Click <Change…>. The window “Computer Name Changes” opens.


(4) Enter the new Computer/Host Name and/or the Workgroup/Domain.

figure 29

(5) Click <OK>.


(6) If you changed the Workgroup/Domain you will get the following Welcome
message:

figure 30

(7) Click <OK>. A message appears asking you to reboot.


(8) Click <OK> (“Computer Name Changes” dialog box).
(9) Click <OK> (“System Properties”).
(10) Reboot the PC.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 27


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

3.8 Installing Application and Documentation

REQUIRED TIME:
Approx. 20 min. (without additional Software Update)

Prerequisites:
• Successful OS installation & partitioning (C and D)
• The “CRService” account language setting is 'English'

NOTE:
Follwing actions need to be performed to install Application and Documentation.
If you read this document on the computer, click the links to read step by step
instructions. If you read a print out, continue with next page:

Checkpoints: See section


1. Installing Application 3.8.1

2. Installing NX Software Update (if available) 3.8.2

3. Installing NX Online Help 3.8.3

Back to Overview

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 28


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

3.8.1 Installing Application

IMPORTANT:
Before you continue, make sure the partition D: (DATA) was created as described in
section 3.3 Performing additional Partitioning.

(1) Log in using “CRService” account.


(2) Insert the NX Starterkit DVD 1.
(3) Browse in the CD-drive to the folder Service Software Æ NX 2008.
(4) Double-click NX 2008.exe. After some seconds, the Welcome window opens.
(5) Click <Next>:

figure 31

(6) Click <I Agree>:

figure 32

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 29


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(7) The installer checks for already installed packages and displays a summary of the
results. Click <Next>:

figure 33

(8) The current status of the NX is displayed. Click <Next>:

figure 34

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 30


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(9) Following window lets you decide on which partition the examination data
(Pixel Files) will be stored. The examination data must always be stored on
partition D. If you haven’t created partition D: yet, cancel the installation and
perform the steps described in section 3.3 Performing additional Partitioning
Æ Do not check the box but make sure you have created the DATA partition D:
and click <Install>.

figure 35

(10) The installation starts. This takes some time. When it is finished, the button
<Next> becomes active. Click <Next>:

figure 36

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 31


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(11) Click <Finish>.

figure 37

NOTE:
The logging information of the process application installation can be found in the
log file: C:\Agfa\Healthcare\Log\NXMain.log

3.8.2 Installing NX Software Update (SU) (if available)

NOTE:
This is the moment when available NX software updates are installed. Check the NX
2008 Service Bulletins to see if there is a SU released. The most recent SU should be
installed. The documentation and software is available on NX Starterkit DVD 1.

(1) Insert the NX Starterkit DVD 1.


(2) Browse to the CD-drive to the folder "Service Documentation" to find the
respective installation instructions.
(3) Browse to the CD-drive to the folder "Service Software" to find the respective
installation package.
IMPORTANT:
Make sure that the ALF file you are going to activate, matches the version of the
software which is installed! How to pre-create an ALF file is described in section
2.3 "Obtaining Application License File (ALF)".

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 32


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

3.8.3 Installing NX Online Help

REQUIRED TIME:
Approx. 5 min.

(1) Insert the NX Starterkit DVD 1.


Internet Explorer is automatically started, showing the Documentation CD
Navigation screen. If the Internet Explorer is not automatically started, browse to
the CD drive and double click "index.htm".
(2) Click "Install Online Help".
Three messages can pop up:
• Internet Explorer: “Active content in html …” Æ Click <Yes>.
• File download security warning Æ Click <Run>.
• Internet Explorer security warning: “Publisher could not be verified …”
Æ Click <Run>.
(3) Click <Next>.
(4) The online Help installer will guide you through the Online Help installation
on your system.
• A “Typical Setup” will install Online Help in English on your system.
• A “Custom Setup” will open a next dialog where you can make a choice
of the languages you want to install.
• A “Complete Setup” will install all Online Help for all languages supported.
(5) Click <Install>.
(6) Click <Finish> to end the installation procedure.
(7) Remove the NX Starterkit DVD 1.
(8) After finishing the upgrade inform the customer about the usage of the
NX Online Help.
NOTE:
• If additional languages are not installed, the online help and CD navigation
screens may show strange characters, but this has no influence on the usability
of the installed languages.
• Installing the Online Help (any language) will also install a special Thai font on
the NX Workstation that is required to read the Thai Online Help. This implies that
browsing Thai Online Help directly from the CD will not work before the Online
Help is actually installed on the Workstation.
• If required, the font can be manually installed by copying this file from the CD into
the Windows font folder: ”C:\…\Nx Online Help\Windows\Fonts\Angsd___.ttf”

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 33


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

3.9 Running Hotfix installation again

REQUIRED TIME:
Approx. 5 min.

Run the hotfix installer a second time to install patches that were skipped during the
first installation (section 3.6). Patches that are already installed will be skipped.

3.10 Creating Fingerprint File (only in case of CR QS upgrade)

NOTE:
This step is only required if you are performing an upgrade from a CR QS and want to
create the ALF via ELMS yourself.

(1) Go to Start Æ AGFA Æ Service Æ License Manager.


(2) Click on <Generate Fingerprint>.
(3) Adapt file name and location e. g. external Memory stick and click <OK>.
(4) Check the status bar of the License manger until the remark creating …
disappears.
(5) Wait until the License manger is on status ready.
(6) Close the License Manager.
(7) Generating the Fingerprint File is complete.

3.11 Loading Licenses (ALF)

CAUTION:
Due to wrong date & time and time zone settings license may become invalid.
Before activating a license key file, make sure the correct time zone, date and time
have been configured! (Described in section 3.4).

REQUIRED TIME:
Approx. 10 min.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 34


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

NOTE:
If the installation wizard for the license dongle has not appeared yet, you have to
unplug the dongle and reconnect it. The wizard will pop up. Click through it using the
default settings.

figure 38

To load the licenses the Application License File (*.alf) is needed (see section 2.3)
This file has to be imported and the license has to be loaded into the system:

(1) Copy the ALF to “C:\AGFA\Healthcare\NX\Configuration\Licensing”.


(2) Start up the license manager tool from
Start Æ AGFA Æ Service Æ License manager.
(3) Click the <Add license file> button.
(4) Browse to the “C:\AGFA\Healthcare\NX\Configuration\Licensing” folder.
(5) Select the ALF and click <Open>.

Result You now have successfully loaded the license file.

NOTE:
The logging information of the Licensing process can be found in the log file:
C:\Agfa\Healthcare\Log\LicensingService.log

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 35


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

3.12 Installing XRDI Software (if applicable)

NOTE:
• This step is only required for DX-Si Workstations or Mammography workstations
requiring the XRDI software. A dedicated license needs to be ordered.
• If the XRDI software is de-installed and then reinstalled (e.g. to adapt the
SW version), the NX displays an application error: License is not ok
Æ when checking in the Configuration Tool, the Connectivity License is disabled.
Don’t forget to enable it again before starting the NX application!

REQUIRED TIME:
Approx. 5 min.

• For Mammography workstations refer to the CR Mammography Solution Manual


(DD+DIS227.04E), Chapter 3.4 for instructions on how to install the XRDI software.
On MedNet GSO library (Intranet Link / Extranet Link).
• For DX-Si workstations refer to the DX-Si System Manual (DD+DIS032.06E),
chapter 02 – Installation and Setup for instructions on how to install the XRDI
software.On MedNet GSO library (Intranet Link / Extranet Link).

3.13 Installing Digitizer Software

REQUIRED TIME:
Approx. 30 min.

(1) Switch off the digitizer.


(2) Install the latest digitizer software (depending on connected digitizer) on the NX
processing station. Follow the instructions in the digitizer device software package
downloaded from Mednet. See also IMPORTANT note below.
(3) Restore the latest digitizer backup. See IMPORTANT note below if backup is
not available.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 36


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

IMPORTANT:
When the configuration of the NX processing station is finished, and the digitizer is
switched on, it is possible, that the digitizer shows following error messages:

Symptom 1 Error indicating, that the registration of the serial number failed.
Reason: Restore not performed: NX processing station does not know the serial
number of the connected device.
Cure: Fill in site specific data (Digitizer service client menu “Configuration –
Site specific data). When done, select buttons “update site specific data”
and “configure site with new EMS serial number”.

Symptom 2 Error indicating, that software versions loaded onto machine and
installed on PC are not identical.
Reason Software mismatch between digitizer software loaded on digitizer and
installed on NX processing station.
Cure: Perform software upload to the digitizer as described in the instructions
in the digitizer device software package. If the software on the digitizer is
higher than on the processing station, install the corresponding digitizer
software on the processing station.

3.14 Installing an Uninterruptible Power Supply (UPS)

REQUIRED TIME:
Approx. 45 min.
Refer to Chapter 4 Appendices of the NX 2008 Service Manual DD+DIS043.08E for
the installation and configuration procedure.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 37


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

4 Activation

REQUIRED TIME:
Approx. 15 min.

The NX workstation is setup to be configured as one of these profiles, based on the


information retrieved from the loaded ALF file:
• In-room mixed use NX workstation: depending on the available licenses, the NX
workstation will be able to perform:
o only GenRad examinations.
o only Mammography examinations.
o both GenRad and mammography examinations.
• In-room veterinary NX workstation: allows only configuring veterinary examinations.
• Central monitoring system: allows to monitor a set of in-room NX workstations.

(1) Log out and in again using the “CRService” account.


(2) Start the NX Activation Utility using the shortcut on the desktop, or from
Start Menu Æ Agfa Æ NX Æ Service Æ Install Tools.

figure 39

(3) The NX Activation Utility shows a warning message that a fresh NX will be
activated. Click <Next>:

figure 40

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 38


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

NOTE:
The NX Activitation Utility retreives the licensing information and checks if a license for
a TWAIN digitizer and the proper digitizer software is installed.
If digitizer software has been installed manually, the NX Activation Utility will display a
reminder to check if the software matches the digitizer type. You can still update the
installed digitizer software after Activation, if required.

(4) Decide if the NX will be configured to work with LGM values or with EI values for
dose calculation.
Decision should be based on these rules:
1. if a CMS is installed, all connected NX in-room workstations must have
the same setting as the CMS.
2. if a speed class free digitizer is installed, the setting must be to use EI
values. For more information about speed class free digitizers, refer to
chapter 1 “Product Description” and “HW/SW Compatibility Chart” of the
NX 2008 Service Manual (DD+DIS043.08E).
(5) If the right digitizer software has not yet been installed, the NX Activation Utility
shows an error message about unavailable Digitizer Software Installer. E.g:

figure 41

Click <OK> on this message , select 'none' for TWAIN digitizer and click
<Next>. Install the required software after Activation according to section 2.2.3
"DX-S or CR 30-X Digitizer Device Software".

(6) Select the default language for the NX workstation and click <Next>.
(7) Check the overview of specified selections. To change options, click <Back>, to
cancel the activation, click <Cancel>. If all selections are correct, click
<Activate>. The activation starts:

figure 42

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 39


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(8) After the activation, click <Next> and <Finish>.


(9) The NX workstation is rebooted.

NOTE:
The logging information of the process NX activation can be found in the log file:
C:\Agfa\Healthcare\Log\Nx\NXactivate.log

5 Upgrading ID Tablet Firmware

REQUIRED TIME:
Approx. 10 min.

NOTE:
If the ID tablet firmware is 1.12.07 or later this step is not required.
Please refer to Chapter 4 Appendices of the NX 2008 Service Manual
DD+DIS043.08E: ‘Upgrading ID tablet firmware’ which also describes how to check the
ID tablet firmware version.

6 Configuration

REQUIRED TIME:
Approx. 5 – 15 min.

IMPORTANT:
• If you do not have an existing configuration file available, the NX has to be
configured manually. In this case refer to section 6.1.
• Have your Acceptance Test Protocol document at hand and fill it in on-the-fly at
every configuration step instead of filling it in afterwards. This saves a lot of time.

(1) Log in using the “CRService” account.


(2) Open the NX Configuration Tool.
(3) Select Load configuration Æ Load a configuration from an external file.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 40


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(4) Browse to the location of the config file you want to import and click <OK>.
The config file is being imported.
(5) Configure all site specific settings (e.g devices) in the configuration tool. For help
refer to chapter 04 “Installation and Configuration” of the NX 2008 Service Manual
(DD+DIS043.08E) and the Configuration Tool Online Help.
(6) Go to File and click <Activate Configuration> to activate the configuration.

NOTE:
• On a laptop PC, special power settings are required. They have to be configured
manually on each individual user account for users. Check the procedure in Chapter
4 Appendices, 'Laptop Power Settings'. These settings also have to be applied on
the automatically created accounts: “CRService” and “Administrator”.
• Additional configuration tasks may be needed depending on the settings for that
specific site/system (e.g. additional users, PCAnyWhere configuration, ...). In this
case, refer to chapter 4 of the NX 2008 Service Documentation (DD+DIS043.08E).

6.1 Manual Configuration (if needed)

REQUIRED TIME:
Refer to the Service/Solution Manuals. The required time depends on the complexity
of the exam tree.

6.1.1 GenRad
Configure the NX application as described in section "Configuration" in chapter 4
of the NX 2008 Service Documentation (DD+DIS043.08E).

6.1.2 Mammography
Goal of this section is to configure the upgraded workstation correctly so it will be
part of the CR Mammography solution.

(1) Refer to section "4. Configuration" in chapter 4 of the NX 2008 Service


Documentation (DD+DIS043.08E).
(2) Follow the procedures and stop before you reach section
4.12 Configuring Devices (Part 1).
(3) Refer to the CR Mammography Solution Manual (DD+DIS227.04E), Perform
the procedures of the chapter 3.4 Configuration of the Mammo Solution from
sub-section 5.2.6 “Device Configuration” until including
5.2.9 “Film Text Box Configuration”.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 41


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(4) Refer back to the NX 2008 Service Documentation chapter 4 “Installation and
Configuration” (DD+DIS043.08E) and perform the the procedure of section 4.18
Activating the Configuration.
(5) Continue with this document.

7 Backup

REQUIRED TIME:
Approx. 10 min.

After installation and after any change to the configuration of NX, it is important to
make a backup. Follow this procedure:
(1) Open the Configuration Tool and load the active configuration.
(2) In the Configuration Tool menu select “File Æ Save as…”.
(3) Choose a portable medium to save the configuration file on and store this file on a
save place. Name the file that it can be identified later on: e.g.
<Site>_initial_configuration_<date>.xml
The file should be kept outside the NX hard disk to have it available in case the
NX hard disk crashes.
(4) Run the script ‘Backup all data’:
Agfa Æ NX Æ Service Æ Backup Restore Æ Backup all data.
For more information about backup scripts, see NX 2008 Service Manual
DD+DIS043.08E, Chapter 4 ‘Installation and Configuration Appendices’.

NOTE:
A restore of the system (e.g. after hardware replacement) is performed using the
Clean Install Procedure (this document).

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 42


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

8 Activating RAID (Redundant Array of Independent Disks)

IMPORTANT:
RAID is supported ONLY on hardware that has been ORIGINALLY delivered as RAID
system and equipped with a second harddisk.
NX 2008 Service Manual - Chapter 01 SW/HW Compatibility Chart (DD+DIS043.08E) -
specifies the hardware types that are compatible with RAID.
It is not allowed to install RAID on systems not originally delivered as RAID system,
even if the hardware is compatible with RAID.

REQUIRED TIME:
Approx 120 min (waiting time included).

(1) Insert the NX Starterkit DVD 2.


(2) Go to the folder DRIVERS\RaidTool.
(3) Double click the file R173416.exe.
(4) A security warning may be displayed. Click <Run>:

figure 43

(5) Click <Continue> on the next dialog:

figure 44

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 43


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(6) A location is requested where to unzip the files to - enter "C:\DRV\R173416"


and click <Ok>.

figure 45

(7) Create the location by clicking <Yes>.

figure 46

(8) Click <Ok> on the next dialog.

figure 47

(9) The Intel Matrix Storage Manager setup wizard is started. All screens can be
confirmed and the PC will be automatically rebooted.

[...]
figure 48

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 44


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

(10) Log in using “CRService” account.


(11) Click the Start button. Go to All Programs Æ Intel Matrix Storage Manager Æ Intel
Matrix Storage Console to launch the Intel Storage Console.
(12) Select "Protect data from a hard drive failure with RAID 1":

figure 49

(13) Click <Create a RAID 1 volume>.


(14) Click <Yes> to start migrating:

figure 50

(15) The progress of migration is displayed.

figure 51

IMPORTANT:
Do not perform a reboot of the system until the migration process has
finished completely!

NOTE:
The computer can be used normally during the migration process.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 45


09-2008 Type 4406/303 Agfa Company Confidential
DD+DIS050.08E Software Options

9 Finalizing NX Acceptance Test

REQUIRED TIME:
Approx. 30 min.

IMPORTANT:
Check, if all required fields of the Acceptance Test have been filled out and complete it
and hand it over to the customer. The Acceptance Test is available on MedNet GSO
library (Intranet Link / Extranet Link).

IMPORTANT:
Refer to the respective Solution Manual (e.g. CR Mammography Solution Manual) for
additional Acceptance Test on Solution level.

DOCUMENT CONTROL NOTE:


The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.

Edition 1, Revision 2 NX 2008 Chapter 5 / 46